System Overview Cat. No. 86100


Analyzer type Real-Time PCR analyzer
Reagent system Open
Samples per run 1-96
Length per run Depends on parameter
Languages available English, French*, Spanish*, Russian* (*coming soon)
User interface External PC
HumaCycler
Real-Time PCR Analyzer

Tests Measuring modes Fluorescence detection


Bottom detection technology for short distance
Analysis modes Absolute quantification
Relative quantification
Single-Nucleotide-Polymorphisms analysis
Melting Curve
Calculation modes Baseline Threshold, 2nd Derivative Maximum, Linear Regression
# of programmable tests Unlimited
# of programmable profiles Unlimited
# of standards per test Up to 4 (for absolute quantification)

Sample / Control / Sample type Serum, EDTA-plasma


Calibrator Sample vessels 0.2 ml PCR tubes, 12 x 8 PCR tube stripes, 96-well plates
Sample positions Total of 96 positions for samples


Reaction unit Reaction wells 0.2 ml PCR tubes / 12 x 8 PCR tube stripes / 96-well plates
Reaction volume 5-100 l

Temperature Range of block working 4...105C


Minimum division 0.1C
Heating /Cooling rate 4C / s
Accuracy 0.2C
Uniformity 0.4C
Fluctuation 0.1C


Reading Optical system 4 channel system
Optical adjustment Automatic by software
Light source LED
Wavelengths F1: 470 nm 15 nm (excitation), 525 nm 10 nm (emission)
F2: 523 nm 10 nm (excitation), 564 nm 10 nm (emission)
F3: 571 nm 10 nm (excitation), 612 nm 10 nm (emission)
F4: 628 nm 20 nm (excitation), 692 nm 20 nm (emission)
Max. # of wavelengths filters 4
Detector Photomultiplier

HumaCycler page 2 of 4
Data Processing Memory for Experiments, templates, standard curves, patient data
Memory capacity Unlimited
Reports for Patient results, standard and amplification curves, work sheet, methods
Warnings Temperature control, room temperature control, module position
Printer Printer connected to external PC

General PC min. requirements External PC, Pentium IV or Intel I3,


Hard disk capacity min. 10 GBytes,
Operating system: Microsoft Windows XPSP2/Vista/7
Application software: Excel 2000/2003/2007 or Access 2000/2003/2007
Physical dimensions (W x D x H) Instrument without any components: 43 x 39.5 x 35.2 cm
Packaging: 58 x 67 x 46 cm (metal box)
79 x 82 x 69 cm (wooden box)
Weight: Gross: 78.0 kg, net: 28.0 kg
Mains fuse F / 250 V, 8 A, 5 x 20 mm
Electrical requirements 100...240 VAC; 50/60 Hz
Environmental Operating: temperature 10...30C, humidity < 70 % rel. non condensing
Transport: temperature -20...50C, humidity < 80 % rel. non condensing

Scope of Supply Unit/Size Cat.No.


HumaCycler 1 86100
Packlist
User Manual 1 86100/1
Quick Guide 1 86100/4
Frequently Asked Questions 1 86100/7
Software CD 1 ...................
Power Cable 250V 10A 1 ...................
Power Cable 125V 12A 1 ...................
RS232C Converter Box 1 ...................
USB Converter Box 1 ...................
Bluetooth Converter Box 1 ...................
Rubber Typhoon 1 ...................
PVC Socket 1 ...................
Fuse 250V 8A 2 ...................
Anti-Condensate Water Seal Ring 1 ...................

page 3 of 4 HumaCycler
HumaCycler
Real-Time PCR Analyzer

Obligatory Items Accessories


Personal Computer with monitor is necessary for the operation of this instrument.
HUMAN offers them as optional items below.

Optional Items Accessories Cat.No


HumaCycler Dye Calibration Kit 86150
Personal Computer with Windows 7 (32Bit) incl. Keyboard and Mouse 18992P
Monitor LCD 19 inch 17901M
LCD Touch Screen Monitor 18991MT
HP Laser Printer (USB + Parallel) 18993L


86100/2015-04 2015 Human

Legal statement
While Human is making every effort to include accurate and up-to-date information, we make

no representations or warranties, express or implied, as to the accuracy or completeness of the

information provided in this document and disclaim any liability for the use of it.

You might also like