Professional Documents
Culture Documents
JOURNAL OF PACING AND CLINICAL ELECTROPHYSIOLOGY , Volume 24, No. 1, January 2001
Copyright © 2001 by Futura Publishing Company, Inc., Armonk, NY 10504-0418.
SWEENEY, M.O., ET AL.: Prospective Randomized Comparison of 50%/50% Versus 65%/65% Tilt Bipha-
sic Waveform on Defibrillation in Humans. It is unknown if there is a single optimal biphasic waveform
for defibrillation. Biphasic waveform tilt may be an important determinant of defibrillation efficacy. The
purpose of this study was to compare acute defibrillation success with a three-electrode configuration in
humans using 50%/50% versus 65%/65% tilt truncated exponential, biphasic waveforms delivered
through a 110-mF capacitor. Acute DFTs for biphasic waveforms with 50%/50% versus 65%/65% tilt were
measured in random order in 60 patients using a binary search method. The electrode configuration con-
sisted of a RV coil as the cathode, and a SVC coil plus a pectoral active can emulator (CAN) as the anode.
The waveforms were derived from an external voltage source with 110-mF capacitance, and the leading
edge voltage of phase 2 was equal to the trailing edge voltage of phase 1. Stored energy DFT (9.2 6 5.7
[50%/50%] vs 10.8 6 6.4 [65%/65%] J, P 5 0.007), current DFT (10.9 6 4.0 [50%/50%] vs 12.0 6 4.4
[65%/65%] A, P 5 0.002) and voltage DFT (391 6 118 [50%/50%] vs 424 6 128 [65%/65%] V, P 5 0.004)
were significantly lower for the 50%/50% tilt waveform versus the 65%/65% tilt waveform using this
three-electrode configuration and a 110-mF capacitor. For an RV (2)/SVC plus CAN (1) electrode config-
uration and a 110-mF capacitor, a 50%/50% tilt biphasic waveform results in a 15% reduction in energy
DFT, 9% reduction in current DFT, and 8% reduction in voltage DFT versus a 65%/65% tilt biphasic
waveform. (PACE 2001; 24:60–65)
defibrillation, waveforms, implantable cardioverter defibrillators
Methods
Table II.
The study population consisted of 60 patients
undergoing ICD implantation at five centers. All ICD Electrodes Used in the Study
patients gave written, informed consent according ICD Electrodes N 5 60 Patients
to the rules and regulations of the Human Re-
search Committees at the investigative sites. Base- Right ventricular endocardial, %, (N)
line characteristics of the study population are Medtronic Model 6932 52 (31/60)
shown in Table I. A three-electrode pectoral Medtronic Model 6936 28 (17/60)
transvenous ICD system was used. In 59 patients, Ventritex Model RVO2 17 (10/60)
a right ventricular (RV) sensing/pacing/defibrilla- CPI Model 0125 2 (1/60)
tion electrode was positioned with the tip at the Intermedics Model 497-19 2 (1/60)
distal RV apex and a separate transvenous coil Superior vena cava, %, (N)
electrode was positioned with the tip at the junc- Medtronic Model 6933 13 (8/60)
tion of the superior vena cava (SVC) and the right Medtronic Model 6937 67 (40/60)
atrium (Table II). In one patient, a single-pass dual Ventritex Model SVO2 17 (10/60)
coil electrode with a fixed separation between the CPI Model 0125 2 (1/60)
distal and proximal coil was used (Table II). An Intermedics Model 497-22 2 (1/60)
active can emulator (CAN) was placed in the left Pectoral active can emulator, %, (N)
mid-pectoral position. The RV coil served as the Medtronic Model 7221 27 (16/60)
defibrillation cathode (2) and the SVC coil plus Medtronic Model 7223 73 (44/60)
CAN served as the defibrillation anode (1). Defib-
rillation waveforms were derived from a Ventritex, Sunnyuale, CA, USA; CPI, Cardiac Pacemakers, Inc.,
St. Paul, MN, USA. ICD 5 implantable cardioverter defibrillator.
Medtronic (Minneapolis, MN, USA) Model 2934
external cardioverter defibrillator with 110-mF ca-
pacitance. Ventricular fibrillation induction and
termination testing was performed under tran- RV endocardial electrode. Ventricular fibrillation
sient light general anesthesia with intravenous was defined as rapid, irregular, disorganized sur-
agents or general anesthesia with inhaled agents face electrocardiographic activity accompanied by
and endotracheal intubation. Ventricular fibrilla- a mean cycle length , 200 ms on simultaneous in-
tion was induced using a low energy monophasic tracardiac electrograms. Ventricular fibrillation
shock delivered between the RV coil and the SVC was required to persist for 10 seconds prior to de-
coil/CAN and synchronized to the T wave during livery of a defibrillating pulse. DFT metrics (en-
ventricular pacing or 50-Hz burst pacing via the ergy, voltage, current, and resistance) were deter-
mined for truncated exponential, biphasic
waveforms with 50%/50% versus 65%/65% tilts.
The waveform voltage and duration were mea-
Table I.
sured directly by the test instrument. Resistance
Clinical Characteristics of the Study Population was calculated using leading and trailing edge
voltage and equations of exponential decay, and
Characteristic N 5 60 Patients current was calculated using Ohm’s Law. For both
Mean age, years 65 6 11
waveforms, phase1/phase 2 tilts were equal and
Male gender, %, (N) 90 (54/60)
phase 2 leading edge voltage equaled phase 1 trail-
Mean ejection fraction, % 32 6 15
ing edge voltage. In each patient the stored energy
Clinical arrhythmia, %, (N)
DFT, defined as the minimum stored energy that
Aborted sudden cardiac death 32 (19/60)
successfully terminated ventricular fibrillation on
Sustained VT 58 (35/60) a single occasion, was prospectively determined
Nonsustained VT 10 (6/60)
for each waveform in random order using a binary
Heart disease, %, (N)
search method (Fig. 1). For each episode of ven-
Coronary artery disease 70 (42/60)
tricular fibrillation, only the first pulse was used
Nonischemic dilated CMP 27 (16/60)
to determine the DFT. If the first pulse was unsuc-
Other 3 (2/60)
cessful, testing continued with the next prescribed
Antiarrhythmic drugs at implant, energy level after restoration of sinus rhythm and
%, (N)
a 5-minute rest period.
Amiodarone 25 (15/60) Statistical Methods
Class IA 3 (2/60)
Class IB 3 (2/60)
To detect a difference of 15% on a mean
stored energy DFT of approximately 10 J, (a 5
CMP 5 cardiomyopathy; VT 5 ventricular tachycardia 0.05, 1-b 5 0.80) and assuming a standard devia-
Table III.
Defibrillation Threshold Metrics for 50%/50% versus 65%/65% Tilt Biphasic Waveforms
10.1 6 6.5 J (P 5 0.62). This indicates that the or- ration and pathway resistance.10,17 Perhaps most
der of testing did not bias the stored energy DFT importantly, because these conflicting studies
for either waveform. used a much smaller sample size (9–19 patients)
than the current study, they would have had a low
Discussion probability of detecting 10%–15% differences in
This prospective, randomized comparison in defibrillation efficacy.
60 patients using a RV (2)/SVC plus CAN (1) The reason for the increased defibrillation ef-
electrode configuration and a 110-mF capacitance ficacy of the 50%/50% fixed tilt waveform versus
voltage source demonstrates that a 50%/50% tilt the 65%/65% fixed tilt waveform in our study is
biphasic waveform results in significantly lower unknown. The increased efficacy of the 50%/50%
stored energy, voltage, and current DFTs versus a tilt waveform may be related to better optimiza-
65%/65% tilt biphasic waveform. Capacitance, tion of pulse duration, higher negative phase peak
lead configuration, and testing protocol were held voltage, or both, as both waveform parameters are
constant within and between patients to minimize affected by altering tilt.
the potentially confounding influence of these The Weiss-Lapicque relationship for cardiac
variables on defibrillation efficacy. Overall, fewer tissue stimulation using rectangular pulses pre-
patients had a high stored energy DFT ($ 24 J) dicts a minimum mean current or rheobase for
with the 50%/50% tilt versus the 65%/65% tilt stimulation and a chronaxie, which is the stimu-
waveform (5% vs 10%, respectively) although this lation pulse duration at twice rheobase cur-
was not statistically significant. rent.18,19 This relationship produces a U-shaped
The delivered energy for the 50%/50% tilt curve for energy as a function of pulse duration as
waveform was less than the 65%/65% tilt wave- has been demonstrated with cardiac pacing. The
form partly because of shorter average pulse dura- lowest energy requirement for electrical stimula-
tions. A 50%/50% tilt waveform delivers a lower tion of the heart using rectangular waveforms,
delivered percentage of stored energy than a therefore, occurs when the pulse duration approx-
65%/65% tilt waveform (93.7% vs 98.3%, respec- imates the chronaxie time.20 Animal and human
tively). Results also demonstrate that stored en- studies have shown a similar strength-duration re-
ergy, current, and voltage DFTs were also reduced lationship for defibrillation using a monophasic
by the 50%/50% tilt waveform. truncated exponential waveform17,21,22 and phase
Previous reports on the effect of biphasic 1 of biphasic truncated exponential defibrilla-
waveform tilt on defibrillation efficacy have been tion. 11,12,17,23 A pulse duration between 2 and 4
conflicting. Swartz et al.4 reported a 31% reduc- ms has been observed to yield a defibrillation en-
tion in the stored energy DFT using a 42%/42% ergy minima for monophasic shocks and phase 1
tilt versus a 65%/65% tilt biphasic waveform de- of biphasic shocks.11,12,17 Quantitative models of
livered from a 120-mF capacitor. That study used defibrillation based on this strength-duration rela-
two transvenous electrodes (RV and SVC or coro- tionship recite a necessary balance between the
nary sinus) and a subscapular patch electrode. In exponential-decay time constant of the capacitive
contrast, Poole et al.14 found no difference in de- discharge, pulse duration, and chronaxie
fibrillation efficacy between a 50%/50% tilt ver- time.5–8,17
sus a 65%/65% tilt biphasic waveform delivered To explain the increased efficacy of the
from a 120-mF capacitor using a unipolar high 50%/50% tilt waveform versus the 65%/65% tilt
voltage configuration (RV/CAN). Similarly, Li et waveform on the basis of pulse durations, several
al.15 found no difference in defibrillation efficacy reasonable but unproven assumptions must be
between 50%/50% versus 65%/65% tilt biphasic made. First, that phase 1 of a biphasic shock acts
waveforms using two transvenous electrodes as does a monophasic shock,6,24 and that the prin-
(RV/SVC). Natale et al.9 observed a reduction in cipal effect of phase 1 is to capture the my-
biphasic DFT with a three-electrode configuration ocardium, whereas phase 2 “burps” the residual
when the tilt was decreased from 65%/65% to charge left on the cell membrane by phase 1.6 Sec-
50%/50%, but no further reduction when tilt was ond, that monophasic shocks follow a Weiss-
decreased to 40%/40%. The disparate results of Lapicque relationship. If those assumptions hold
these studies may be related to differences in elec- true, then the increased efficacy of the 50%/50%
trode configurations, capacitances, DFT testing tilt waveform might be due to phase 1 pulse dura-
techniques, and sample size. Prior research has tions being nearer the chronaxie time. Animal
demonstrated, for example, that the increased de- models have shown a monophasic defibrillation
fibrillation efficacy of biphasic versus monopha- chronaxie time between 2 and 4 ms, and studies in
sic waveforms is influenced by electrode configu- humans have suggested a chronaxie time in a sim-
ration. 16 Further, the optimal phase 1 duration ilar range.17,25,26
and capacitance may vary with electrode configu- Numerical models of defibrillation with trun-
cated exponential waveforms indicate a linkage ing lower DFT with 50% tilt, no difference, or
between the exponential-decay time constant of lower DFT with 65% tilt was 38 6 9, 38 6 6, and
the capacitive discharge, the time-response pa- 35 6 5 V, respectively (P . 0.05). Thus, even with
rameter of cardiac myocytes to the shock, and the current study, the interpatient variability of re-
pulse duration.5–8,17,27–30 The time constant of the sistance does not explain the difference in tilt,
capacitive discharge, or ts, is the product of path- though this is perhaps confounded by the inaccu-
way resistance and capacitance. In the hyperbolic racy of a crude DFT estimation in one patient.
average current model5,6 the time-response pa- If there is a single optimal pulse duration for
rameter of cardiac myocytes is the chronaxie time defibrillation using a capacitive discharge wave-
tc (as defined by the Weiss-Lapicque relationship). form is unproven. This reflects the fact that the
In the leading edge-current exponential strength-duration relationship for transvenous de-
model,7,8,27–30 the time-response parameter is re- fibrillation in humans is incompletely character-
ferred to as the membrane time constant tm . tc and ized. At least one study in humans appeared to
tm are assumed to be related although this rela- show that the transvenous monophasic DFT volt-
tionship is inexact and may vary by waveform. age and stored energy declined exponentially with
DFT is predicted to be minimized when ts ap- increasing pulse durations but were insensitive to
proximates the cardiac myocyte time-response pa- pulse durations . 6 ms.25 This observation is
rameter (either tc or tm ) in both models.19 Because more compatible with the quantitative model de-
pathway resistance was nearly equal and output veloped by Blair27,28 that describes the heart mus-
capacitance was identical between waveforms, cle membrane as a capacitor and resistor in paral-
the difference in efficacy in the current study can- lel and predicts that DFTs are insensitive to
not have been due to a difference in ts . Using equa- altering pulse durations over a wide range.
tions from these models6,8,17 and time-response Higher peak negative phase voltage may also
parameters derived from human strength-duration have contributed to the increased defibrillation ef-
curves using transvenous leads,17 we calculated ficacy of the 50%/50% tilt waveform. Tomassoni
the optimal first pulse duration and tilt using the et al.31 demonstrated that for three commercially
values of capacitance and average resistance mea- available biphasic waveforms the waveform with
sured in this study. The leading edge current ex- the lowest negative phase peak voltage and
ponential model predicted that 50% tilt and 3.19 longest pulse duration resulted in the highest de-
ms would be optimal, while the average current fibrillation energy requirement. Further, when
hyperbolic and exponential models predicted op- phase 1 trailing edge/phase 2 leading edge volt-
timal values of 64% and 65% for tilt and 4.73 ms ages were held equal and phase 1 tilt was fixed at
and 4.80 ms for pulse duration, respectively. 65%, a shorter pulse duration in phase 2 did not
In our study, the average phase 1/phase 2 reduce defibrillation efficacy. However, reducing
pulse durations for the 50%/50% tilt waveform phase 2 leading edge voltage to 1/2 phase 1 trailing
were 3.23 6 0.58/3.35 6 0.54 ms versus 4.82 6 edge voltage, while holding phase1/phase 2 tilt
0.87/4.99 6 0.85 ms for the 65%/65% tilt. Thus, fixed at 65% and phase 1/phase 2 pulse duration
our data are in agreement with theoretical predic- equal, resulted in significantly higher DFTs. They
tions of the leading edge exponential model. concluded that negative phase peak voltage was
There are limitations to these models in pre- more important for defibrillation efficacy than
dicting the response to various pulse tilts, how- phase 2 pulse duration. In the present study,
ever. Despite one model predicting exactly which phase 1 trailing edge and phase 2 leading edge
tilt would have the lower DFT in our study of one voltages were equal for both waveforms. However,
discharge time constant, these models are all inac- starting with any phase 1 leading edge voltage, the
curate in predicting behavior across time con- phase 2 leading edge voltage will be relatively
stants. Each model predicts that when the system higher for the 50%/50% versus 65%/65% tilt.
time constant, ts, decreases the optimal tilt should
increase. Studies presented in the literature did Study Limitations
not show a difference between 50% or 65% tilt us- The principal methodological limitation of
ing two-electrode systems. However, when using this study is that we did not determine strength-
three-electrode systems, which have a decreased duration curves for defibrillation. This is imprac-
resistance (and therefore decreased ts ) compared tical in the context of ICD implantation in humans
with two-electrode systems, the lower DFT was as it would mandate an excessive, and potentially
associated with the lower tilt, not the higher tilt deleterious, number of ventricular fibrillation in-
value predicted by the models. We further in- ductions and terminations. The wide energy steps
spected our data to determine if individual patient in our DFT search protocol may have been inade-
resistances could explain differences observed in quate to resolve small but important differences
our study. The mean resistance for patients show- between waveforms. Although single shock DFTs
References
1. Schuder JC, McDaniel WC, Stoeckle H. Defibrillation of 100 kg 16. Bardy GH, Troutman C, Johnson G, et al. Electrode system influ-
calves with asymmetrical, bidirectional, rectangular pulses. Car- ence on biphasic waveform defibrillation efficacy in humans. Cir-
diovasc Res 1984; 18:419–426. culation 1991; 84:665–671.
2. Bardy GH, Ivey TD, Johnson G, et al. A prospective randomized 17. Swerdlow CD, Brewer JE, Kass RM, et al. Application of models of
evaluation of biphasic versus monophasic waveform pulses on defibrillation to human defibrillation data: Implications for opti-
defibrillation efficacy in humans. J Am Coll Cardiol 1989; 14: mizing implantable defibrillator capacitance. Circulation 1997;
728–733. 96:2813–2822.
3. Schuder JC, Stoeckle H, West JA, et al. Transthoracic ventricular 18. Weiss G. Sur la possibilite de render comparable entre eux les ap-
defibrillation in the dog with truncated and untruncated exponen- pareils survant a l’excitation electrique. Arch Ital Biol 1901;
tial stimuli. IEEE Trans Biomed Eng 1971; 18:410–415. 35:413–446.
4. Swartz JF, Fletcher RD, Karasik PE. Optimization of biphasic 19. Lapicque L. Definition experimentalle de l’excitabilite. Proc Soc
waveforms for human nonthoracotomy defibrillation. Circulation Biol 1909; 77:280–285.
1993; 88:2646–2654. 20. Pearce JA, Bourland JD, Nielsen W, et al. Myocardial stimulation
5. Irnich W. Optimal truncation of defibrillation pulses. PACE 1995; with ultrashort duration current pulses. PACE 1982; 5:52–58.
18:673–688. 21. Bourland JD, Tacker WA, Geddes LA. Strength-duration curves for
6. Kroll MW. A minimal model of the monophasic defibrillation trapezoidal waveforms of various tilts for transchest defibrillation
pulse. PACE 1993; 16:769–777. in animals. Med Instrum 1978; 12:38–41.
7. Kroll MW. A minimal model of the single capacitor biphasic wave- 22. Chapman PD, Wetherbee JN, Vetter JW, et al. Strength-duration
form. PACE 1994; 17:1782–1792. curves of fixed pulse width variable tilt truncated exponential
8. Walcott GP, Walker RG, Cates AW, et al. Choosing the optimal waveforms for nonthoracotomy internal defibrillation in dogs.
monophasic and biphasic waveforms for ventricular defibrillation. PACE 1988; 11:1045–1050.
23. Feeser SA, Tang ASL, Kavanagh KM, et al. Strength-duration and
J Cardiovasc Electrophysiol 1995; 6:737–750.
probability of success curves for defibrillation with biphasic wave-
9. Natale A, Sra J, Krum D, et al. Relative efficacy of different tilts
forms. Circulation 1990; 82:2128–2141.
with biphasic defibrillation in humans. PACE 1996; 19:197–206.
24. Hillsley RE, Walker RG, Swanson DK, et al. Is the second phase of
10. Swerdlow CD, Kass RM, Chen PS, et al. Effect of capacitor size and
a biphasic defibrillation waveform the defibrillating phase? PACE
pathway resistance on defibrillation threshold for implantable de-
1993; 16:1401–1411.
fibrillators. Circulation 1994; 90:1840–1846. 25. Gold MR, Shorofsky SR. Strength-duration relationship for human
11. Schauerte PN, Schondube FA, Grossman M, et al. Influence of transvenous defibrillation. Circulation 1997; 96:3517–3520.
phase duration of biphasic waveforms on defibrillation energy 26. Shorofsky SR, Foster AH, Gold MR. Effect of waveform tilt on de-
requirements with a 70-uF capacitance. Circulation 1998; 97: fibrillation thresholds in humans. J Cardiovasc Electrophysiol
2073–2078. 1997; 8:496–501.
12. Schauerte PN, Ziegert K, Waldmann M, et al. Effect of biphasic 27. Blair H. On the intensity-time relations for stimulation by electric
shock duration on defibrillation threshold with different electrode currrents, I. J Gen Physiol 1932; 15:709–729.
configurations and phase 2 capacitances. Prediction by upper- 28. Blair H. On the intensity-time relations for stimulation by electric
limit-of-vulnerabilit y determination. Circulation 1999; 99: currrents, II. J Gen Physiol 1932; 15:731–755.
1516–1522. 29. Cleland B. A conceptual basis for pacing waveforms. PACE 1996;
13. Tang ASL, Yabe S, Wharton JM, et al. Ventricular defibrillation us- 19:1177–1185.
ing biphasic waveforms: The importance of phasic duration. J Am 30. Cleland B. A conceptual basis for defibrillation waveforms. PACE
Coll Cardiol 1989; 13:207–214. 1996; 19:1186–1195.
14. Poole JE, Bardy GH, Kudenchuk PJ, et al. Prospective randomized 31. Tomassoni G, Newby K, Deshpande S, et al. Defibrillation efficacy
comparison of biphasic waveform tilt using a unipolar defibrilla- of commercially available biphasic impulses in humans. Circula-
tion system. PACE 1995; 18:1369–1373. tion 1997; 95:1822–1826.
15. Li HG, Yee R, Mehra R, et al. The effect of biphasic shock wave- 32. McDaniel MC, Schuder JC. The cardiac ventricular defibrillation
form tilt upon defibrillation efficacy. (abstract) Circulation 1993; threshold: Inherent limitations in its application and interpreta-
88:I-113. tion. Med Instrum 1987; 21:170–176.