B Scan
Probe: 10MHz Magnetic driven noiseless
Scanning Mode: Sector
Resolution: Lateral ≤0.5mm;Vertical≤0.3mm
Geometric Location Precision: Lateral≤10%;Vertical≤5%
Depth: ≥50mm
Angle: 53°
Gray Scale: 256
Display Mode: B/B+A/B+B
A Scan
Probe: 10MHz With LED
Precision: ±0.06mm
Depth: 15-40mm
Eye mode: NORMAL/ DENSE/ APHAKIC/ PMMA/ SILICON/ ACRYLIC
IOL Formula:SRK-II, SRK-T, BINKHORST, HOLLADAY, HOFFER-Q, HAIGIS
Measurement: AL, ACD, L, V and the average
Curve Freeze: auto freeze and manual freeze by foot switch
Others:
Input: Touch screen or keyboard
Output: Video Printer
K1,K2 Input: Manual input or infrared transfer from accompanied keratometer.
Measuring Mode: Double cursor distance measure
Power Supply: AC220V±10%, 50Hz±2%
Power:25W+10%
Size:220mm*200mm*280mm
SW-2100 A/B Scan (Portable,easy to take)
Specification
B scan
Frequency: 10MHz ,Magnetic driven, noiseless
Scanning Mode: Sector Scanning
Magnify:Multi continuous magnification,Real-Time magnification
Resolution: Lateral ≤0.3mm;Vertical≤0.2mm
Depth:60mm
Scanning Angle:53°
Gray Scale: 256
False Color: Multi colors. OTC
Measure Mode: 4 distances, 2 angles, perimeter and area
Movies: 100 images movie review,AVI ZIP JPG format image output
A scan:
Frequency:10MHz, with LED
Depth: 15-40mm
Precision:±0.05 mm
Measurement: AL, ACD, L, V and the average
IOL Formula: SRK-II, SRK-T, HOFFER-Q, HOLLADAY,BINKHORST-II, HAIGIS
Stat. Calculation: Average and standard deviation
Store: 10 Scanning results for each eye
Others:
Display Mode :B、B+B、B+A、A
Case Report: PDF format case report archiving
Case Search:Multi-keywords
SW-100 Keratometer
Specification
Device type:Type B
Measuring range:3.000mm~12.908mm
Accuracy:≥±0.05mm
Single Measurement Time:0.03s
Output:Infrared output
Weight:<0.5Kg (include battery)
Dimension:240mm×90mm×60mm
Power:500mW+15%
Power Supply:3 AA battery
SW-3200 UBM
Specification
Frequency: 50MHz
Scaning Mode: Sector Scanning
Scaning Range: 16mm*9mm; 10mm*6.5mm
Image Resolution(full screen):50um-100um
Geometry Distortion: Both Vertical & Lateral rate<3%,no compensation, undistorted
image
Display: Full Screen and Micro screen, can be switched
Dsiplay Mode :UBM\ UBM+A.
Magnify: Special independent ultrasound magnify system,More clear in Anterior Segment.
Working Platform: Windows XP
Image Process: Measurements and edition.
Output: Examining report can be created
SW-6000 Corneal Topography
Specification
2 Coverage: 10.91mm(Diameter)
5 Precision: ±0.02mm
Display axial curvature map, tangential curvature map, elevation map, imitated
keratoscope map and 3D cornea map
Adjust range: left/right 86mm; forward/backward 40mm; up / down 30mm ; chinrest 50mm
Specification
2 Measure Error:
±1.5mmHg(3mmHg≤intraocular pressure≤25mmHg);
±2.5mmHg(25mmHg<intraocular pressure≤70mmHg);