You are on page 1of 8

USER’S MANUAL

TEMPERATURE CALIBRATOR
OTE-T700

For testing of:


* Temperature Indicators
* Temperature Switches
* Temperature Sensors
* Safety Equipment
* Etc.

Stalsbergvn. 9, 3128 NOTTEROY, NORWAY


 + 47 3338 2500  +47 3338 2525
 omicron.spare@autronicafire.no
Instructions

Connect the temperature calibrator to correct AC source.

Set the adjustment knob to desired temperature according to the conversion


table on top of the calibrator. Wait until the temperature has stabilized. If
necessary, tune to correct desired temperature.

When the temperature has stabilized on desired value, put the test object into
the test chamber, (se one of the 3 attached brass adapters), and let it remain
there for 2-3 minutes. If necessary, calibrate/adjust the test objects’ reading.

If more than one object is to be calibrated/ tested, start with the one with the
highest temperature value and test downwards.

If the object to be tested is a part of a safety system, make sure it can be tested
without interrupting the total process.

When finished, turn the adjusting knob to ”0”. Allow the unit to cool down for 1-
2 minutes before switching off the unit.
CALIBRATION PROCEDURE FOR
OMICRON’S TEMPERATURE CALIBRATOR
TYPE OTE-T700
Note: OMICRON’s Test Certificate shall be filled in when proceeding through
the calibration procedure.

ƒ First of all you will need a temperature reference calibration equipment with better
accuracy than the OTE-T700.

ƒ Remove all eight 3.5 mm screws which are located all around the side with the handle,
also remove all eight 3.5 mm screws around the side where the ON/OFF switch is
placed. Gently separate the unit into two parts.

After you have separated the unit, look at the side were the display is placed. Here
you will see a printed circuit board (pcb). There are placed two pot. meters near the
display. These two pot. meters are the ones that shall be adjusted.

ƒ On the top of the unit, there are placed three brass adapters, use the one that fits the
reference temperature sensor best. Insert the temperature sensor through the brass
adapter. Then place the brass adapter with the temperature sensor in the hole to the heat
chamber. Make sure that the heat sensor hits the bottom of the heat chamber, be gentle.
Then pull the heat sensor up approximately 2 cm. This way the heat sensor will be in the
middle of the heat chamber.

ƒ Turn on the OTE-T700 unit, adjust the knob on top of the unit to about
350-400°C. (Refer to the scaling information on top of the cabinet). Let the system
heat up for a few minutes.

Now turn down the temperature knob to zero. Let the system stabilise for ten minutes.
When the temperature is between 30-40°C, with ambient temperature 20°C, the system
has stabilised on the lowest possible value.

If the lower temperature deviates more than 1°C from the calibration reference,
you need to adjust the pot. marked OFF SET on the pcb until you are satisfied. Note
readings. (Test no 1 in the Test Certificate)

ƒ Adjust the temperature knob to about 350 - 420°C. (Refer again to the scaling
information on top of the unit).Let the system stabilise for ten minutes.

The pot. meter market CAL. adjust the temperature range read - out. Adjust
this pot. meter so that the display shows 1°C lower than your calibration reference
equipment. (Let it stabilise a couple of minutes). Note readings. (Test no 3 in Test
Certificate)

This adjustment may influence the off set setting. Therefore, check the off set setting
again and adjust if necessary. Repeat this procedure until both the upper and lower
temperature readings satisfies your requirements.

ƒ Then adjust temperature knob to test no 2 and note readings and finally adjust to
test no 4 and note readings.
Rev 1.
TEMPERATURE CALIBRATOR OTE-T700

Technical Specifications:

Measuring range : ~30 - 700oC


Medium : Heated Air
Display : 3,5 Digit LCD disp. 15mm
Dimensions : LxWxH, 410mm, 140mm, 160mm
Weight : 4 kg
Power : 230 V 50/60 Hz 2300 W
: 115 V 50/60 Hz 2700 W
Cable length : 2,5 m with plug
Hysteresis : 30 -100oC +/- 1oC F.s.
: 100-400oC +/- 1% F.s.
: 400-700oC +/- 2% F.s.

Stalsbergvn. 9, 3128 NOTTEROY, NORWAY


 + 47 3338 2500  +47 3338 2525
 omicron.spare@autronicafire.no
ANNUAL RE-CALIBRATION

According to different classification societies, re-calibration of the calibrators must take place once a year.
A calibration certificate/statement must be issued and kept onboard in office files in case of inspection.
Re-calibration can be done by qualified technicians at Autronica Fire and Security AS facilities in Norway.
In case of repair, only original parts will be used.

Before shipment, please contact Autronica Fire and Security AS at omicron.spare@autronicafire.no. Details
regarding invoice address and return address are important.

Please send the units to below office address and mark the package with RE-CALIBRATION.

Autronica Fire and Security AS


Stalsbergveien 9
NO-3128 Notteroy
Norway
Tel : +47 33382500
Fax : +47 33382525
Email : omicron.spare@autronicafire.no
Att: Support Dept.