You are on page 1of 25

40 J. Giebułtowicz et al. / J. Chromatogr.

A 1423 (2015) 39–46

[16] and the only use of CPE and LC–MS combination was reported purification system (Milli-Q, Billerica, US). Blank plasma was
[3] for APCI as the ionization source. LC–ESI–MS/MS is the state- obtained from the Regional Center of Blood Donation and Treat-
of-the-art bioanalytical technique and is more frequently used in ment (Warsaw, Poland).
the bioanalysis of drugs than LC–APCI–MS/MS. However, the sur-
factants introduced to the sample may cause matrix effects and ESI 2.2. Chromatographic and mass spectrometric conditions
is more sensitive to this phenomena than APCI [17]. Therefore, as
regards further applications of CPE in the bioanalysis of drugs, it is The instrumental analysis was performed using Agilent 1260
of critical importance that the possibility of its compatibility with Infinity (Agilent Technologies, Santa Clara, CA, US), equipped with
LC–ESI–MS/MS is evaluated. a degasser, autosampler, binary pump coupled to a hybrid triple
Bisoprolol is a selective ␤-1-adrenoceptor antagonist bereft of quadrupole/linear ion trap mass spectrometer QTRAP 4000 (AB
any partial agonist effect [18]. Taking into account the social impor- Sciex, Framingham, MA, US). The Turbo Ion Spray source was oper-
tance of bisoprolol, novel methods for its rapid, convenient and ated in the positive mode. The curtain gas, ion source gas 1, ion
environmentally safe determination in human plasma are still source gas 2 and collision gas (all high purity nitrogen) were set
needed. Previously reported methods of LC–MS/MS detection of at 345 kPa, 207 kPa, 276 kPa and “high” instrument units, respec-
bisoprolol [19–28] did not use CPE as serum or plasma prepara- tively. The ion spray voltage and source temperature were 5000 V
tion technique. The most common one was protein precipitation and 600 ◦ C, respectively. The target compounds were analyzed in
by the addition of organic solvents (e.g. acetonitrile) [20,21,24–26]. the MRM mode (Table 1).
Although this technique is rapid and simple, its disadvantages Chromatographic separation was achieved with a Symmetry
result from insufficient sample clean-up and include a shorter HPLC C18 column (150 mm × 4.6 mm, 3.5 ␮m, Waters, Milford, US). The
column life, increased risk of significant matrix effects and con- column was maintained at 40 ◦ C at the flow rate of 0.5 mL min−1 .
tamination of a mass spectrometer [29]. The SPE procedures using The mobile phases consisted of HPLC grade water with 0.1% formic
Oasis polymeric mix sorbent cartridges were also applied for the acid as the eluent A and methanol with 0.1% formic acid as the elu-
determination of bisoprolol alone or among other beta-antagonists ent B. The gradient (%B) was as follows: 0 min. 10%; 1 min. 10%;
and beta-agonists [28]. SPE based on the hydrophilic-lipophilic bal- 8 min. 90%, 16 min. 90%. The re-equilibration of the column to the
anced sorbent beds is successfully used in the multi-analysis of initial conditions lasted 4 min. The injection volume was 10 ␮L.
compounds of different polarity [30]. However, this technique may
be considered time consuming and rather expensive, especially if 2.3. Standard solutions, calibration standards and quality control
there is a need for analyzing large numbers of samples in a pharma- samples
cokinetic study. LLE was also successfully applied in the bioanalysis
of bisoprolol [19,22,23]. This technique is often used in bioanaly- The standard stock solutions of bisoprolol and IS (1 mg mL−1 )
sis as it is fast and inexpensive. Our previous experience confirms were prepared by dissolving an appropriate amount of the sub-
that its application allows to avoid the contamination of an MS stance in methanol. The working solutions (10 ng mL−1 ) were
ion source and extends the duration of an HPLC column [29,31,32]. prepared by diluting standard stock solutions with water. The solu-
However, LLE requires the use of organic solvents which are toxic tions were stored at 4 ◦ C. The calibration standards for bisoprolol
to laboratory staff and the environment, as well as expensive to dis- were made at the following concentrations: 0.3, 0.9, 3.0, 10, 20, 30,
pose of. CPE is an alternative technique with all the advantages of 45, 60 and 70 ng mL−1 . The quality control samples were prepared
LLE. Additionally, CPE is regarded safer and more environmentally at the following levels: 0.9, 30 and 60 ng mL−1 . The calibration stan-
friendly due to the extraction being based on surfactants which are dards and quality control samples were stored at −80 ◦ C until use.
less toxic and less expensive to dispose of [33]. An extensive search Sodium citrate was used as an anticoagulant.
did not return any reports on the application of CPE to the biso-
prolol determination nor the statistical comparison of CPE and the 2.4. Plasma samples from the pharmacokinetic study
well-established sample preparation techniques.
In this paper we introduce a novel combination of sample prepa- Plasma samples from the pharmacokinetic study in humans
ration with the separation and detection techniques. The aim of the were obtained from the Pharmaceutical Research Institute [34].
study was to assess the compatibility of CPE and LC–ESI–MS/MS and All samples were previously analyzed by LLE and the LC–MS/MS
the applicability of CPE to the bisoprolol determination in spiked method adapted from Ding et al. [35]. Briefly, plasma sample (1 mL)
human plasma and clinical samples. was mixed with the IS (metoprolol, 50 ␮L), 1 M NaOH (0.1 mL) and
shaken for 5 min with ethyl acetate (5 mL). After centrifugation the
2. Materials and methods
organic layer was evaporated to dryness. The residue was recon-
2.1. Chemicals stituted in 60% aqueous methanol (200 ␮L) and 30 ␮L aliquot was
injected onto the LC–MS/MS consisting of Alliance 2695 liquid chro-
The reference standard of bisoprolol fumarate was a kind gift matograph (Waters, Milford, MA, USA) coupled to a Quattro micro
from ICN Polfa Rzeszów S.A. (Poland). Metoprolol tartrate (internal API triple quadrupole mass spectrometer equipped with an electro-
standard, IS) and Trition X-114 was purchased from Sigma-Aldrich spray ion source (Waters, Manchester, UK). The method was fully
(St. Louis, US). The solvents, HPLC gradient grade methanol, acetoni- validated according to the international guidelines [36] and all the
trile and formic acid 98% were purchased from Merck (Darmstadt, analysis were performed in accordance with the Good Laboratory
Germany). Ultrapure water was obtained from the Millipore water Practice (GLP).

Table 1
Optimized MRM conditions and ion transitions.

Substance [M + H] + ion DP [V] EP [V] Quantitative CE [V] CXP [V] Qualitative CE [V] CXP [V]
product ion product ion

Bisoprolol 326 91 10 116 27 8 222 21 12


Metoprolol (IS) 268 86 10 116 27 8 133 37 10

DP—declustering potential, EP—entrance potential, CE—collision energy, CXP—cell exit potential.


J. Giebułtowicz et al. / J. Chromatogr. A 1423 (2015) 39–46 41

2.5. Sample preparation post-extraction spiked samples. Various lots of matrix used for the
test included haemolyzed and hyperlipidaemic plasma samples.
2.5.1. CPE
50 ␮L of the IS working solution, 300 ␮L of 7% Triton X-114 2.7. Statistical analysis
solution and 400 ␮L of 0.2 M NaOH were added to 250 ␮L plasma
sample and vortexed for 5 min. Next, the sample was incubated in The statistical evaluation of the results was performed with the
a water bath at 55 ◦ C for 20 min and centrifuged at 10 000 rpm at SAS software version 9.4. for Windows. The Shapiro–Wilk test was
25 ◦ C for 5 min. Then the aqueous phase was separated by decant- used to evaluate normal distribution of QC results. The hypothesis
ing. After the CPE 700 ␮L of acetonitrile was added to the micellar on the equality of variances was verified by the F-test (p ≥ 0.05).
phase. Alternatively to create more environmental friendly method The linear regression for bisoprolol concentration measured using
ethanol can be used in that step as well. The sample was incubated the CPE and LLE methods was calculated using the least squares
at −20 ◦ C for 20 min and centrifuged at 10 000 rpm at 25 ◦ C for method, the 95% confidence limit band of the mean response at
5 min. The supernatant (500 ␮L) was diluted with an equal volume particular LLE points and the 95% prediction interval limits.
of water (500 ␮L) and transferred to the vial. Ten ␮L aliquot was
injected onto the LC–MS/MS. All CPE experiments were performed
3. Results and discussion
at the Bioanalysis and Drugs Analysis Department of the Medical
University of Warsaw.
3.1. Optimization of the CPE procedure

Triton X-114 is more frequently chosen for the analysis of bio-


2.5.2. LLE (LLE2)
logical fluids [1,8–10,12,16,20,38–40] than other surfactants such
The LLE2 experiments were performed at the Bioanalysis and
as Triton X-100 and Genapol X-080. All experiments were per-
Drugs Analysis Department of the Medical University of War-
formed using Triton X-114 due to its main advantages namely low
saw. Plasma sample (250 ␮L) was mixed with the IS (metoprolol,
cloud-point temperature (23 ◦ C) and high density which facilitate
50 ␮L), 1 M NaOH (25 ␮L) and shaken for 5 min with ethyl acetate
phase separation by centrifugation [16]. During the optimization of
(1.25 mL). After centrifugation the organic layer was evaporated to
the CPE conditions the following changes in the sample preparation
dryness. The residue was reconstituted in 60% aqueous methanol
protocol were made: plasma samples were fortified with bisoprolol
(250 ␮L) and 10 ␮L aliquot was injected onto the LC–MS/MS.
and water instead of the IS working solution. The IS working solu-
tion was added in the last step of the extraction process instead of
2.6. Method validation water.

The method previously developed at the Pharmacology Depart- 3.1.1. Effect of the equilibrium temperature
ment of the Pharmaceutical Research Institute was fully validated The CPE is based on the separation of two phases, when the
and the stability of bisoprolol was confirmed in various conditions aqueous solutions of the surfactant are heated above the cloud-
[34]. Therefore, only partial validation was performed for this study point temperature. Theoretically, the optimal temperature for the
according to the European Medicines Agency (EMA) [37] and US extraction is 15–20 ◦ C greater than the cloud-point of the surfac-
Food and Drug Administration (FDA) [36] guidelines. Briefly, the lin- tant. With the increase of the equilibrium temperature, the volume
earity range was selected as 0.3–70 ng mL−1 as described previously of the surfactant-rich phase decreases due to the disruption of the
[34]. The accuracy and precision of the method were determined hydrogen bonds and the dehydration of the phase [16]. However,
within-run and between-run using LLOQ (0.3 ng mL−1 ) and qual- too high temperature may reduce the recovery—because of both the
ity control (QC) samples (0.9, 30 and 60 ng mL−1 ). Carry over and decomposition of the heat labile compounds as well as the thermal
recovery was also studied. instability of the surfactant aggregates [8]. The range of 40–55 ◦ C
As the aim of the study was to evaluate the compatibility of was selected for the optimization experiments. Due to a possi-
CPE and LC–ESI–MS/MS, particular attention was paid to the matrix ble analyte instability, thermal stability problems of the surfactant
effect assessment. Ion suppression or enhancement was initially aggregates and denaturation of plasma proteins it was decided
studied by the steady post-column infusion of the analyte, i.e. biso- not to perform experiments in higher temperatures. The analyti-
prolol or the IS at the concentration 60 ng mL−1 and the injection cal signal of bisoprolol increased with increasing the equilibrium
of the extracted blank plasma sample. The second approach was temperature, thus 55 ◦ C was selected as the working equilibrium
based on the method by Matuszewski et al. and allowed the quan- temperature (Fig. 1A). One can observe that standard deviation
titative assessment of the absolute and relative matrix effect by the for selected temperature was the highest one, still during method
comparison of the analyte peak areas in the standard solutions and validation acceptance criteria for precision were met.

Fig. 1. Optimization of the CPE method (n = 5). Experimental conditions: (A) 7% Triton X-114, 0.5 M NaOH; (B) extraction temperature at 55 ◦ C, 0.5 M NaOH; (C) extraction
temperature at 55 ◦ C, 7% Triton X-114.
42 J. Giebułtowicz et al. / J. Chromatogr. A 1423 (2015) 39–46

3.1.2. Effect of the surfactant concentration Table 2


Accuracy and precision of bisoprolol determination by CPE and LLE.
Increasing surfactant concentration improves the recovery, but
above a certain point it dilutes the compound of interest and Nominal bisoprolol concentration [ng/ml] 0.9 30.0 60.0
decreases the preconcentration rate [12]. The concentrations of Within-run accuracy [%]
Triton X-114 were optimized in the range of 5–20% which corre- CPE (n = 5) 89.9a 91.4 98.2
sponded to the concentration of 1.5–6% in the extraction mixture. LLE (n = 6) 107.9 97.6 95.2
No significant correlation between the bisoprolol peak area and the LLE2 (n = 5) 109.3 96.7 102.5
Within-run precision [%]
surfactant concentration was observed (Fig. 1B). It was observed
CPE (n = 5) 12.36a 3.55 3.98
that the surfactant concentration over 7% resulted in a lower signal LLE (n = 6) 3.62 4.08 2.70
of bisoprolol, probably due to a higher phase volume ratio (vol- LLE2 (n = 5) 8.67 6.42 2.13
ume of the surfactant-rich phase/volume of the aqueous solution) Between-run accuracy [%]
CPE (n = 15) 95.1b 93.1 98.4
resulting in the dilution of the extracted bisoprolol. Thus, based on
LLE (n = 9) 107.1 98.6 93.6
the results of the experiment, 7% of Triton X-114 was selected to LLE2 (n = 5) 104.4 93.3 100.5
achieve the best analytical signals. Between-run precision [%]
CPE (n = 15) 12.60b 6.30 3.80
3.1.3. Effect of the NaOH concentration LLE (n = 9) 3.67 5.03 6.50
LLE2 (n = 5) 10.44 8.24 4.36
The pH optimization is an important step in the CPE method
development. For organic molecules, especially ionizable ones, the CPE and LLE2 was performed in Medical University of Warsaw; LLE was performed
in Pharmaceutical Institute.
maximum extraction efficiency is achieved at those pH values a
n = 6.
where the uncharged form of the analyte prevails and therefore b
n = 16.
the analyte is favored to be partitioned into the micellar phase
[10]. Since bisoprolol is a weak base, the NaOH concentrations were concentration of bisoprolol. The values of regression parameters
tested in the range of 0.05–0.5 M that corresponded to the concen- for the curve, described by the equation: y = ax + b, were calculated
tration of 0.02–0.2 M in the extraction mixture. The best result was as: a = 0.00509 (0.00012), b = 0.00802 (0.00085) and r2 = 0.999. All
obtained when 0.2 M NaOH was added (Fig. 1C), however, the dif- regression parameters were statistically significant (p < 0.05). The
ferences between the analyte signal in 0.2 and 0.5 M NaOH were accuracy for LLOQ was 105% (RSD = 12%, n = 5) within one day and
within the measurement uncertainty. The phenomenon of a lower 107% (RSD = 9.8%, n = 15) within three runs. The accuracy for the QC
bisoprolol signal in higher NaOH concentrations might be caused samples (Table 2) within one day ranged 89–98% (RSD 3.6–13%)
by the co-extraction of the interfering compounds at a higher pH. while between-runs ranged 93–98% (RSD 3.8–13%). The carry-over
experiment in which blank human plasma samples were analyzed
3.2. Influence of the surfactants and selection of internal standard after the highest calibration standards did not show any signals
influencing quantification.
It is well known that the surfactants might seriously influence The recovery for bisoprolol in CPE was moderate and was
the accuracy and precision of bioanalytical methods. One of the rea- 46% (RSD = 12%) for high concentrations (60 ng mL−1 ) and 61%
sons might be difficulty of pipetting viscous liquids which results (RSD = 14%) for low concentrations (0.9 ng mL−1 ). The correspond-
in a relatively high uncertainty of the extraction yield of the com- ing recovery for LLE was 86% for high concentrations (60 ng mL−1 )
pound. Too low volume of the surfactant-rich phase might also and 74% for low concentrations (0.9 ng mL−1 ). In CPE the extraction
result in an unsatisfactory accuracy of the analytical method. This efficiency can be improved using salts like NaCl or Na2 SO4. These
can be overcome by the use of an internal standard. However, in salts are not compatible with LC–MS/MS technique and may cause
the case of methods coupling CPE and mass spectrometry there high matrix effect. Moreover, nonvolatile salts can deposit in source
is a problem of the surfactant’s negative effect on such ionization and plug capillaries thus requiring more cleaning and maintenance
types as MALDI or ESI [41]. Therefore, in CPE a very careful eval- operations. Alternatively violate salts like ammonium acetate or
uation of matrix effects should be an essential part of the method formate can be used, when low recoveries result in failing the vali-
validation. In order to reduce the matrix effect the samples were dation requirements for LLOQ. However, these salts has never been
diluted with acetonitrile and water after the extraction and the gra- used previously in CPE.
dient chromatographic separation was developed. No changes in During the evaluation of the matrix effect by the post-column
the quality of MS spectra nor in chromatograms after CPE exper- infusion of the analyte, the injection of the extracted matrix
iments were observed. There was no indication for additional ESI resulted in similar changes of the signal for both bisoprolol and
source cleaning. metoprolol (Fig. 2). Moreover, near the retention time of the com-
Generally, isotope-labeled compounds are recommended for pounds the signal was barely affected by the matrix and rather sta-
the LC–MS bioanalytical methods as internal standards to reduce ble. The observation was confirmed by the quantitative evaluation
matrix effects and increase the reliability of the results [37]. For this of the matrix effect studied in the bisoprolol concentrations of 1.0
study a compound of similar structure, i.e. metoprolol, was selected and 50 ng mL−1 using the calculation method by Matuszewski et al..
as the internal standard. This internal standard was proved to be an The absolute matrix effect for bisoprolol and the IS were 98% and
appropriate standard both in the SPE [27] and LLE extraction [22,34] 101% for the lower concentration and 105% and 108% for the higher
of bisoprolol from plasma. The structural analogue was selected for concentration, respectively. The relative matrix effect was assessed
this study to assess the compliance of CPE and LC–MS in more dif- as 11.5% for 1.0 ng mL−1 and 4.4% for 50 ng mL−1 , respectively.
ficult conditions than when the isotope labeled internal standard During the experiment we have also tested the influence of
is applied. NaOH and Triton concentrations on the absolute matrix effect. The
absolute matrix effect varied from 97% to 112% (RSD = 4.3%), but
3.3. Method validation no significant trends were observed. The phenomenon might be
a results of adequate chromatographic separation of analytes and
All validation experiments met the acceptance criteria. The inferences like, e.g., Triton. Thus, in CPE this aspect of analysis is
calibration curve obtained by the weighted linear regression probably more important than in other extraction techniques. In
analysis was linear in the range of 0.3–70 ng mL−1 regarding further study we aim to explore the effect of pH and chromato-
the peak area ratio of bisoprolol and the IS versus the nominal graphic separation on the absolute matrix effect in CPE.
J. Giebułtowicz et al. / J. Chromatogr. A 1423 (2015) 39–46 43

XIC of +MRM (7 pairs): 268.133/116.100 Da ID: metoprolol 8 from Sample 3 (matryca i 1' 60ng/ml bisopr i metoprolol) of 2014-04-30bisoprolol i m... Max. 3341.7 cps.

18.98
3342

3000 OH 12.19 13.87


H 17.93 18.59
O N
2500
+ 11.90
H
IS
Intensity, cps

2000 11.65 17.66


O 15.02
1500 11.26
16.98

1000 8.68 10.72


3.33 4.07 8.22 8.95
0.92 1.17 2.14 2.33 3.19 4.81 5.10
6.59 7.11 7.30
500

0
1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0 15.0 16.0 17.0 18.0 19.0
Time, min
XIC of +MRM (7 pairs): 326.259/116.000 Da ID: bisoprolol 1 from Sample 3 (matryca i 1' 60ng/ml bisopr i metoprolol) of 2014-04-30bisoprolol i met... Max. 4.8e4 cps.

OH 13.87 19.24
4.8e4 H 12.84 18.98
4.5e4 O N
18.19
4.0e4 11.77
+
3.5e4 H

3.0e4
O
Bisoprolol 11.32
17.25
Intensity, cps

15.05
2.5e4 O 11.13 16.99

2.0e4 9.13 9.49


2.33 3.33 4.26 4.42 6.97 7.26 8.24 8.91
1.5e4 0.26 1.28 2.13 3.19
1.0e4

5000.0

0.0
1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0 15.0 16.0 17.0 18.0 19.0
Time, min

Fig. 2. Chromatograms recorded during the evaluation of the matrix effect by the steady post-column infusion and the injection of the extracted blank plasma sample for
metoprolol (internal standard, top) and bisoprolol (bottom).

These results indicate that CPE is a suitable technique for sample results. The statistical evaluation by the F-test indicated that the
preparation prior to the LC–ESI–MS/MS analysis as no significant hypothesis of the equality of variances was rejected for the low QC
matrix effect was observed. However, the chromatographic sep- (p = 0.003), as opposed to the medium and high QCs (p = 0.647 and
aration is probably crucial to reduce the matrix effect caused by p = 0.106, respectively). Therefore, one can conclude that for lower
the surfactant. Similar results of a negligible matrix effect in the concentrations, around 0.9 ng mL−1 , the precision of a bioanalytical
CPE–LC–MS method were observed by Liu et al. [3] using the atmo- method is significantly lower when CPE is applied instead of LLE.
spheric pressure chemical ionization (APCI) source. The results of To confirm that hypothesis we performed additional analysis
the validation also revealed that the structural analogue may be an using LLE method in the same laboratory, where CPE experiments
appropriate internal standard when CPE is used as the extraction were conducted. Due to instrumental differences we modified LLE
technique. method (LLE2) as described in Section 2.5.2. The results confirmed
our previous observation that in lower concentration (low QC) CPE
has significantly lower precision than LLE2 method (p = 0.042).
3.4. Comparison of the CPE and LLE methods

Surprisingly, there were only two reports in literature regarding


3.4.2. Application of the CPE method to clinical samples
the comparison of the CPE and LLE methods for the determination of
The final assessment of the compatibility of CPE and
drugs in biological matrices. However, the comparisons were based
LC–ESI–MS/MS was performed using 28 clinical plasma sam-
on the presentation of a chromatogram only in Zhou et al. [40] or
ples selected from different subjects and representing both high
in the brief description (including linearity, accuracy, precision and
(around maximum plasma concentration Cmax ) and low (in the
recovery) by Zhang et al. [11]. An extensive search did not find any
elimination phase) bisoprolol concentrations. Representative chro-
statistical comparisons of both methods; the lack of such analyses
matogram is presented in Fig. 3. A strong linear correlation of the
is contrary to the current trends in chemical metrology. To fill this
LLE and CPE data is evidenced by the Pearson’s correlation coef-
niche, a cross-validation of the bioanalytical methods using CPE and
ficient of 0.9802 (n = 28, SD = 0.0076) as presented in Fig. 4. The
LLE, supported by a detailed statistical evaluation, was performed
compatibility of CPE and LLE was further supported by the fact
[37,42,43]. The difference in the sample preparation techniques
that only 2 of 28 (7%) observations were located outside the 95%
was assumed to be the main source of variability. Other factors pos-
prediction limits.
sibly influencing the method’s performance, such as using different
A more detailed evaluation was based on the approach similar to
laboratories and analysts, different sample preparation techniques
the incurred sample reanalysis. The LLE results were considered as
as well as different LC–MS/MS methods and instruments, were con-
initial values, while the CPE results were considered as repeat val-
sidered less significant. The original data from the bioanalytical
ues. The acceptance criteria of 67% of the CPE results within ±20%
method full validation and pharmacokinetic study obtained pre-
of the mean were defined according to the cross validation part
viously in a GLP certified laboratory for the regulatory purposes
of the EMA guideline [37]. It was observed that 75% of 28 results
were used as reference [34].
met the acceptance criteria, however, the majority of the results
with % difference exceeding ±20% were found among the lower
3.4.1. Statistical evaluation of the validation parameters concentration (Table 3, Fig. 5). When data was limited to the LLE
Both methods met the acceptance criteria for accuracy and pre- results with the measured bisoprolol concentration not lower than
cision at all QC concentration levels (Table 2), confirming that 1.0 ng mL−1 , the statistics were far better: 91% of 22 results met the
both sample preparation techniques are capable to deliver reliable acceptance criteria, the mean %difference was lower and the SD of
44 J. Giebułtowicz et al. / J. Chromatogr. A 1423 (2015) 39–46

XIC of +MRM (7 pairs): 326.259/116.000 Da ID: bisoprolol 1 from Sample 16 (18/15) of 2014-04-17bisoprolol pacjencji.wiff (Turbo Spray) Max. 2016.7 cps.

8.36
2000

1900 Bisoprolol
1800

1700

1600

1500

1400

1300

1200
In te ns ity , c p s

1100

1000

900

800

700

600

500

400

300

200

100

0
1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0 15.0 16.0 17.0 18.0 19.0
Time, min

XIC of +MRM (7 pairs): 268.133/116.100 Da ID: metoprolol 8 from Sample 16 (18/15) of 2014-04-17bisoprolol pacjencji.wiff (Turbo Spray) Max. 7.6e4 cps.

7.67
7.5e4

7.0e4 IS
6.5e4

6.0e4

5.5e4

5.0e4

4.5e4
In te ns ity , c p s

4.0e4

3.5e4

3.0e4

2.5e4

2.0e4

1.5e4

1.0e4

5000.0

0.0
1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0 15.0 16.0 17.0 18.0 19.0
Time, min

Fig. 3. Chromatogram of sample obtained from healthy volunteer after 5 h of oral administration of bisoprolol fumarate. The determined concentration was 2.4 ng mL−1 . The
blue line represents the main ion transitions m/z 326 → 116 and m/z 268 → 116 for bisoprolol (top) and metoprolol (internal standard, bottom). The red line represents the
second ion transition specific m/z 326 → 222 and m/z 268 → 133 for bisoprolol and metoprolol, respectively.
J. Giebułtowicz et al. / J. Chromatogr. A 1423 (2015) 39–46 45

60 60%
y = 0.9646x + 1.2289
R² = 0.9608 40%
50

% difference
CPE result [ng/ml]

20%
40
0%
30
-20%
0 10 20 30 40 50 60
20 LLE result [ng/ml]

Fig. 5. The %difference, i.e. (CPE result—average result)/average result × 100%, plot-
10 ted against bisoprolol concentrations measured using LLE.

0 The observations from the clinical samples analysis were in


0 10 20 30 40 50 60 line with the result of the statistical evaluation of the meth-
LLE result [ng/ml] ods’ precision. Therefore, it was concluded that the determination
of bisoprolol concentration in human plasma in the range
Fig. 4. Correlation of bisoprolol concentrations in clinical samples measured using 1.0–70 ng mL−1 by the CPE method leads to the results which are
CPE and LLE (n = 28). equivalent to those obtained by the widely used LLE method.

4. Conclusion
%difference decreased below 15%. The positive value of the mean
%difference may indicate that the application of CPE may lead to
It has been experimentally proved that cloud-point extraction
higher concentrations measured than in the case of LLE; however,
is compatible with LC–ESI–MS/MS. This novel sample prepara-
this difference seems insignificant, as in both data sets, i.e. n = 28
tion technique was successfully applied for the first time for the
and n = 22, the mean %difference was below 15%.
bisoprolol determination in human plasma. The bisoprolol plasma
concentrations in the clinical samples obtained by the CPE and LLE
methods were comparable in the range 1.0–70 ng mL−1 . The results
Table 3
Comparison of LLE and CPE results for clinical plasma samples.
indicate that further research on combining CPE and LC–MS is rec-
ommended and may lead to significant advances in the bioanalysis
Sample no. LLE result CPE result %Difference of drugs. The study provides the first report on the statistical com-
[ng/ml] [ng/ml]
parison of CPE and the conventional extraction method. CPE offers
1 1.51 2.60 27% significant practical advantages and can have a positive impact on
2 0.30 0.54 28%
the environment, therefore its wider application in future pharma-
3 0.30 0.56 30%
4 38.82 45.10 7% cokinetic studies is justifiable.
5 33.71 42.90 12%
6 33.71 36.10 3%
Conflict of interest statement
7 30.04 30.60 1%
8 3.26 4.14 12%
9 1.96 2.14 4% The authors declare no conflict of interest regarding the content
10 0.80 2.00 43% of this article.
11 24.31 23.10 −3%
12 36.44 36.30 0%
13 53.07 47.40 −6% Acknowledgments
14 43.43 47.50 4%
15 2.49 4.60 30% The authors are grateful to all investigators of the former biso-
16 6.90 7.02 1%
17 2.60 2.37 −5%
prolol bioequivalence study [34] for kindly providing their clinical
18 0.70 2.13 51% samples for the additional analysis.
19 46.44 42.60 −4%
20 42.73 47.50 5%
21 44.71 39.20 −7%
References
22 2.97 3.18 3%
23 0.89 1.30 19% [1] K. Madej, K. Persona, M. Wandas, E. Gomółka, Sequential cloud-point extraction
24 0.35 0.59 26% for toxicological screening analysis of medicaments in human plasma by high
pressure liquid chromatography with diode array detector, J. Chromatogr. A
25 13.34 15.10 6%
1312 (2013) 42–48.
26 28.11 33.00 8%
[2] K. Madej, K. Persona, Drug screening in human plasma by cloud-point extrac-
27 44.20 33.80 −13%
tion and HPLC, Cent. Eur. J. Chem. 11 (2013) 94–100.
28 37.95 36.70 −2% [3] W. Liu, K. Bi, X. Liu, J. Zhao, X. Chen, Cloud-point extraction combined with
All data n= 28 LC–MS for analysis of memantine in rat plasma, Chromatographia 69 (2009)
837–842.
% of samples meeting acceptance criteria 75%
[4] N. Gao, H. Wu, Y. Chang, X. Guo, L. Zhang, L. Du, Y. Fu, Mixed micelle cloud
mean difference 10%
point-magnetic dispersive ␮-solid phase extraction of doxazosin and alfuzosin,
SD of difference 16%
Spectrochim. Acta A: Mol. Biomol. Spectrosc. 134 (2015) 10–16.
CPE result over 1 ng/ml n= 22 [5] R. Heydari, N.S. Elyasi, Ion-pair cloud-point extraction: a new method for the
% of samples meeting acceptance criteria 91% determination of water-soluble vitamins in plasma and urine, J. Sep. Sci. 37
(2014) 2724–2731.
Mean difference 4%
[6] F. Nazari Serenjeh, P. Hashemi, M. Safdarian, Z. Kheirollahi, Semi-automated
SD of difference 10%
cloud point extraction with cold column trapping of surfactant-rich phase for
46 J. Giebułtowicz et al. / J. Chromatogr. A 1423 (2015) 39–46

phenazopyridine determination in human serum, J. Iran. Chem. Soc. 11 (2014) [25] M. Liu, D. Zhang, Y. Sun, Y. Wang, Z. Liu, J. Gu, Liquid
733–739. chromatographic–electrospray tandem mass spectrometric determination of
[7] Z. Pourghobadi, R. Heydari, R. Pourghobadi, M. Rashidipour, Determination of bisoprolol in human plasma, Biomed. Chromatogr. 21 (2007) 508–513.
gabapentin in human plasma using simultaneous cloud point extraction and [26] S. Li, G. Liu, J. Jia, Y. Liu, C. Pan, C. Yu, Y. Cai, J. Ren, Simultaneous determina-
precolumn derivatization by HPLC, Monatsh. Chem. Chem. Mon. 144 (2013) tion of ten antiarrhythic drugs and a metabolite in human plasma by liquid
773–779. chromatography—tandem mass spectrometry, J. Chromatogr. B 847 (2007)
[8] H. Ma, J. You, Y. Liu, Cloud-point extraction combined with HPLC for deter- 174–181.
mination of larotaxel in rat plasma: a pharmacokinetic study of liposome [27] J. Bhatt, G. Subbaiah, S. Kambli, B. Shah, M. Patel, A. Saxena, A. Baliga,
formulation, J. Sep. Sci. 35 (2012) 1539–1546. S. Nigam, H. Parekh, G. Yadav, A high throughput and sensitive liquid
[9] W.-J. Zhao, W. Liu, J.-B. Chen, Z.-M. Zhou, M.-M. Yang, Use of cloud point extrac- chromatography–tandem mass spectrometry (LC–MS/MS) method for the
tion with derivatizing reagent for the extraction and determination of isoniazid, estimation of bisoprolol in human plasma using multiplexing technique, J.
J. Chromatogr. Sci. 49 (2011) 154–158. Chromatogr. B 852 (2007) 374–381.
[10] F. Han, R. Yin, X.-L. Shi, Q. Jia, H.-Z. Liu, H.-M. Yao, L. Xu, S.-M. Li, Cloud point [28] M. Josefsson, A. Sabanovic, Sample preparation on polymeric solid phase
extraction-HPLC method for determination and pharmacokinetic study of flur- extraction sorbents for liquid chromatographic-tandem mass spectrometric
biprofen in rat plasma after oral and transdermal administration, J. Chromatogr. analysis of human whole blood—a study on a number of beta-agonists and
B 868 (2008) 64–69. beta-antagonists, J. Chromatogr. A 1120 (2006) 1–12.
[11] H. Zhang, H.-K. Choi, Analysis of meloxicam by high-performance liquid chro- [29] J. Musijowski, M. Filist, P.J. Rudzki, Sensitive single quadrupole LC/MS method
matography with cloud-point extraction, Anal. Bioanal. Chem. 392 (2008) for determination of lapatinib in human plasma, Acta Pol. Pharm. 71 (2014)
947–953. p7.
[12] X. Liu, X.-H. Chen, Y.-Y. Zhang, W.-T. Liu, K.-S. Bi, Determination of arbidol in rat ˛
[30] J. Giebułtowicz, G. Nałecz-Jawecki, Occurrence of antidepressant residues in
plasma by HPLC–UV using cloud-point extraction, J. Chromatogr. B 856 (2007) the sewage-impacted Vistula and Utrata rivers and in tap water in Warsaw
273–277. (Poland), Ecotoxicol. Environ. Safe. 104 (2014) 103–109.
[13] M.D. Rukhadze, S.K. Tsagareli, N.S. Sidamonidze, V.R. Meyer, Cloud-point [31] J. Musijowski, E. Piórkowska, P.J. Rudzki, Determination of sunitinib in human
extraction for the determination of the free fraction of antiepileptic drugs in plasma using liquid chromatography coupled with mass spectrometry, J. Sep.
blood plasma and saliva, Anal. Biochem. 287 (2000) 279–283. Sci. 37 (2014) 2652–2658.
[14] A.B. Tabrizi, S. Naini, K. Parnian, S. Mohammadi, F.E. zad, S.P. Anvarian, A. Abdol- [32] M. Filist, K. Buś-Kwaśnik, H. Ksycińska, P.J. Rudzki, Simplified LC–MS/MS
lahi, Determination of triamterene in human plasma and urine after its cloud method enabling the determination of azithromycin in human plasma after
point extraction, Quím. Nova 37 (2014) 1182–1187. a low 100 mg dose administration, J. Pharm. Biomed. Anal. 100 (2014)
[15] H. Wu, G.-Y. Zhao, L.-M. Du, Determination of ofloxacin and gatifloxacin 184–189.
by mixed micelle-mediated cloud point extraction-fluorimetry combined [33] H. Filik, S.D. Çekiç, Cloud point extraction of pesticide residues, in: M.
methodology, Spectrochim. Acta, A: Mol. Biomol. Spectrosc. 75 (2010) Stoytcheva (Ed.), Pesticides in the Modern World—Trends in Pesticides Analy-
1624–1628. sis, InTech, Rijeka, Croatia, 2011.
[16] X.Y. Qin, J. Meng, X.Y. Li, J. Zhou, X.L. Sun, A.D. Wen, Determination of ven- [34] K. Buś-Kwaśnik, H. Ksycińska, A. Leś, K. Serafin-Byczak, P.J. Rudzki, J. Raszek, T.
lafaxine in human plasma by high-performance liquid chromatography using Łazowski, A. Bielak, A. Wybraniec, Bioequivalence and pharmacokinetics of two
cloud-point extraction and spectrofluorimetric detection, J. Chromatogr. B 872 10-mg bisoprolol formulations as film-coated tablets in healthy white volun-
(2008) 38–42. teers: a randomized, crossover, open-label, 2-period, single-dose, fasting study,
[17] B.K. Matuszewski, M.L. Constanzer, C.M. Chavez-Eng, Strategies for the Int. J. Clin. Pharmacol. Ther. 50 (2012) 909–919.
assessment of matrix effect in quantitative bioanalytical methods based on [35] L. Ding, X. Zhou, X. Guo, Q. Song, J. He, G. Xu, LC–ESI–MS method for the deter-
HPLC−MS/MS, Anal. Chem. 75 (2003) 3019–3030. mination of bisoprolol in human plasma, J. Pharm. Biomed. Anal. 44 (2007)
[18] G.-F. Li, K. Wang, R. Chen, H.-R. Zhao, J. Yang, Q.-S. Zheng, Simulation of the phar- 520–525.
macokinetics of bisoprolol in healthy adults and patients with impaired renal [36] FDA Guidance for Industry, Center for Drug Evaluation and Research
function using whole-body physiologically based pharmacokinetic modeling, (CDER).Bioanalytical Method Validation, Food and Drug Administration,
Acta Pharmacol. Sin. 33 (2012) 1359–1371. Rockville, MD,U.S, 2001.
[19] H. Chang, J. Li, J. Li, X. Guan, F. Sun, Z. Qian, K. Bi, G. Fan, Simulta- [37] Guideline on bioanalytical method validation, EMEA/CHMP/EWP/192217/
neous determination of amlodipine and bisoprolol in rat plasma by a liquid 2009, European Medicines Agency. Committee for Medicinal Products for
chromatography/tandem mass spectrometry method and its application in Human Use (CHMP), 2011.
pharmacokinetic study, J. Pharm. Biomed. Anal. 71 (2012) 104–110. [38] Z.-M. Zhou, D.-Y. Zhao, J. Wang, W.-J. Zhao, M.-M. Yang, Study of cloud point
[20] G.-Y. Liu, W. Wang, J.-Y. Jia, C. Lu, Y.-M. Liu, M.-Q. Zhang, Y. Liu, S.-J. Li, C. Yu, extraction and high-performance liquid chromatographic determination of iso-
Liquid chromatography tandem mass spectrometry method for determination niazid based on the formation of isonicotinylhydrazone, J. Chromatogr. A 1216
of bisoprolol in human plasma using d5-bisoprolol as the internal standard, (2009) 30–35.
Biomed. Chromatogr. 24 (2010) 574–580. [39] G. Ren, Q. Huang, J. Wu, J. Yuan, G. Yang, Z. Yan, S. Yao, Cloud point
[21] O. Gonzalez, G. Iriarte, E. Rico, N. Ferreirós, M.I. Maguregui, R.M. Alonso, R.M. extraction—HPLC method for the determination and pharmacokinetic study
Jiménez, LC–MS/MS method for the determination of several drugs used in of aristolochic acids in rat plasma after oral administration of Aristolochiae
combined cardiovascular therapy in human plasma, J. Chromatogr. B 878 Fructus, J. Chromatogr. B 953–954 (2014) 73–79.
(2010) 2685–2692. [40] J. Zhou, P. Zeng, H.H. Tu, F.Q. Wang, Development and application of
[22] G. Peste, N. Bibire, M. Apostu, A. Vlase, C. Oniscu, A New Liquid high-performance liquid chromatography for the study of ampelopsin phar-
Chromatography-Tandem Mass Spectrometry Method for Determination of macokinetics in rat plasma using cloud-point extraction, J. Sep. Sci. 34 (2011)
Bisoprolol in Human Plasma Samples, J. Biomed. Biotechnol. 2009 (2009), 160–168.
http://dx.doi.org/10.1155/2009/736327, Article ID 736327, 8 pages. [41] A. 43 Nordström, J.V. Apon, W. Uritboonthai, E.P. Go, G. Siuzdak, Surfactant-
[23] G. Peste, C. Oniscu, A. Vlase, Experimental research for determination of enhanced desorption/ionization on silicon mass spectrometry, Anal. Chem. 78
bisoprolol fumarate in human plasma samples using liquid chromatography- (2005) 272–278.
tandem mass spectrometry (LC–MS/MS) technique, Rom. Biotechnol. Lett. 15 [42] D. Gansser, Experience with a cross-validation approach, Chromatographia 55
(2010) 5141–5145. (2002) S71–S74.
[24] M.F. Tutunji, H.M. Ibrahim, M.H. Khabbas, L.F. Tutunji, Simultaneous determina- [43] M.T. Gilbert, I. Barinov-Colligon, J.R. Miksic, Cross-validation of bioana-
tion of bisoprolol and hydrochlorothiazide in human plasma by HPLC coupled lytical methods between laboratories, J. Pharm. Biomed. Anal. 13 (1995)
with tandem mass spectrometry, J. Chromatogr. B 877 (2009) 1689–1697. 385–394.
Talanta 83 (2011) 980–987

Contents lists available at ScienceDirect

Talanta
journal homepage: www.elsevier.com/locate/talanta

Tyrosinase immobilized magnetic nanobeads for the amperometric assay of


enzyme inhibitors: Application to the skin whitening agents
Veronica Harceaga Sima a , Stephanie Patris b , Zeynep Aydogmus c , Ahmad Sarakbi b ,
Robert Sandulescu a , Jean-Michel Kauffmann b,∗
a
University of Medicine and Pharmacy, Faculty of Pharmacy, 400012 Cluj-Napoca, Romania
b
Université Libre de Bruxelles (ULB), Faculty of Pharmacy, 1050 Brussels, Belgium
c
Istanbul University1 , Faculty of Pharmacy, 34116 Istanbul, Turkey

a r t i c l e i n f o a b s t r a c t

Article history: The immobilization of tyrosinase onto glutaraldehyde activated streptavidine magnetic particles and
Received 22 September 2010 subsequent retention onto a magnetized carbon paste electrode for the amperometric assay of tyrosi-
Received in revised form 26 October 2010 nase inhibitors is described. Tyrosine was used as substrate as it is the first substrate in the melanogenesis
Accepted 1 November 2010
process. The sensing mode is based on monitoring the decrease of the amperometric signal correspond-
Available online 10 November 2010
ing to the electrochemical reduction of dopaquinone enzymatically generated. This current decrease is
due to the presence of inhibitors acting directly on the enzyme or inhibitors acting on the product of the
Keywords:
enzymatic reaction, i.e. dopaquinone. The methodology is designed for the evaluation of the inhibitory
Magnetic nanoparticles
Tyrosinase
potency of the most frequently used active substances in cosmetic marketed products against hyper-
Skin whitening agent pigmentation such as kojic acid, azelaic acid and benzoic acid. These compounds bind to the tyrosinase
Biosensor active center. Ascorbic acid is also investigated as it interrupts the synthesis pathway of melanin by reduc-
ing the melanin intermediate dopaquinone back to l-dopa. By comparing the obtained IC50 , under the
same experimental conditions, the order of their inhibitory potency was: kojic acid (IC50 = 3.7 × 10−6 M,
Ki = 8.6 × 10−7 M), ascorbic acid (IC50 = 1.2 × 10−5 M), benzoic acid (IC50 = 7.2 × 10−5 M, Ki = 2.0 × 10−5 M)
and azelaic acid (IC50 = 1.3 × 10−4 M, Ki = 4.2 × 10−5 M) in close agreement with literature spectrophoto-
metric inhibition data using the soluble tyrosinase.
© 2010 Elsevier B.V. All rights reserved.

1. Introduction Hyperpigmentation is a complex process. There are many fac-


tors that influence the activity of melanocites, but it is well known
Melanin, the dark pigment produced by skin cells in the most and unanimously accepted that tyrosinase activity plays the main
inner layer of the epidermis, plays an important role in protect- role in the pathway of melanin biosynthesis. It catalyzes the first
ing human skin from the harmful effects of the UV sun radiations. two reactions of the melanogenesis process namely: the hydroxy-
Melanin also determines our phenotypic appearance. However, lation of l-tyrosine to l-dopa as monophenolase, using molecular
abnormal accumulation of melanin in the basal layer of the dermis oxygen as co-substrate, and the oxidation of l-dopa to dopaquinone
is responsible for hyperpigmentation including melasma, freckles as diphenolase. These steps, and mostly the first oxidizing step, are
and senile lentigines [1]. These undesirable conditions of the human the rate-limiting steps in melanin synthesis since the subsequent
skin are serious esthetic problem for human beings nowadays since reactions can proceed spontaneously at physiological pH [3].
both appearance and quality of life are becoming more and more A number of tyrosinase and melanogenesis inhibitors, from both
important. This problem is particularly prevalent in middle aged natural and synthetic sources, have been identified [4]. However,
and elderly individuals [2] and this has encouraged researchers to the definition of “tyrosinase inhibitor” is sometimes misleading, the
seek for new potent inhibitors of the hyperpigmentation process. terminology being also used to refer to melanogenesis inhibitors,
Recently, a global market demand has developed on this subject. whose action mainly reside in some interference in melanin forma-
For the past few decades, tyrosinase inhibitors have been a great tion, regardless of any direct inhibitor–enzyme interaction [3]. Only
concern solely due to the key role of tyrosinase in mammalian specific tyrosinase inactivators and inhibitors should be regarded
melanogenesis [3]. as “true inhibitors”, which bind to the enzyme and inhibit its activ-
ity. Some skin whitening agents that act as tyrosinase inhibitors,
such as kojic acid and azelaic acid, or as melanogenesis inhibitors
∗ Corresponding author.
such as hydroquinone, arbutin and ascorbic acid are the most
E-mail address: jmkauf@ulb.ac.be (J.-M. Kauffmann).
well known to most dermatologists [5]. The interference with
1
On leave. melanin synthesis can occur in different ways. Kojic acid is a

0039-9140/$ – see front matter © 2010 Elsevier B.V. All rights reserved.
doi:10.1016/j.talanta.2010.11.005
V.H. Sima et al. / Talanta 83 (2011) 980–987 981

chelator of the copper of the enzyme active site [6], arbutin is a Different phenolic compounds can be enzymatically oxidized
pro-drug of hydroquinone that is an alternative substrate, azelaic by tyrosinase and several substrates have been proposed in the
acid blocks the tyrosine access to the active site and ascorbic acid literature for the evaluation of the enzyme inhibitors. Dopamine
chemically reduces the dopaquinone back to l-dopa, thus avoid- was used as a substrate in the construction of an amperometric
ing dopachrome and melanin formation [7]. Also benzoic acid, a biosensor for the detection of tyrosinase inhibitors such as kojic
preservative very frequently used in cosmetic and food industry [8], acid, benzoic acid and SCN− ion [14]. Catechol, p-cresol, m-cresol,
has been demonstrated to have inhibitory properties for tyrosinase phenol and p-chlorophenol were used as enzyme substrates for
[9]. the determination of benzoic acid based on its inhibition prop-
The need for the quantification and comparative analysis at high erties, using an amperometric biosensor device [19–21]. l-Dopa
throughput of the inhibitory activity of possible new products and was used as enzyme substrate in spectrophotometric evaluation of
of the existing ones is becoming more and more of interest for the skin-whitening agents [10]. Also l-tyrosine was used as a substrate
evaluation and development of efficient inhibitors of the hyperpig- in spectrophotometric evaluation of different inhibitors [22,23].
mentation process. Cellular tests with human melanocytes were used for the trans-
Until now, several methods for the screening of skin-whitening formation rate study of the tyrosinase substrate l-dopa using a
agents have been proposed. The method most frequently applied spectrophotometric method [24].
exploited the spectrophotometric quantification of dopaquinone, Recently, biosensors based on the inhibition of tyrosinase activ-
before and after the inhibition of soluble tyrosinase [6,10]. Another ity in the presence of a monophenol or o-diphenol substrate have
method with tyrosinase in solution relied on electron spin reso- been proposed for the determination of triazine pesticides [25], car-
nance spectrometry that measured the accumulation of eumelanin bamate and organophosphate pesticides [16], diuron and atrazine
radicals in melanin culture cells [11]. In vivo studies evaluated the [26], herbicides such as alachlor, diazinon and carbaryl [27], propyl
inhibitors potency by photometry [12]. Usually in each of the cur- gallate [28], 2,4-dichlorophenol [29], fluoride [30], benzoic acid
rently developed method, the inhibitory strength of the studied [19], cinnamic acid and sorbic acid [31].
compounds was compared to that of a standard inhibitor, namely Besides choosing the adequate transducer and enzyme in the
kojic acid [6]. Spectrophotometry is not a primary choice analyti- construction of a biosensor, an important role is played by the
cal method for the quantification of an on going high ratio reaction enzyme immobilization method. The use of magnetic particles
which is the case of the enzymatic formation of dopaquinone due (beads) in biomedical and pharmaceutical sciences has increased
to the low reproducibility of the obtained results. The use of spec- significantly in recent years [32]. The masterbeads are specially
trometry has the disadvantage of the need of culture cells and the designed for allowing simple and effective immobilization of lig-
photometric one is applied to in vivo studies so it can be used just for ands such as enzyme through activation with bi-functionalized
substances that have been approved for human use and the imple- reactants [33], like glutaraldehyde. Immobilization of biomolecules
mentation of such study involves ethic problems. Thus they are not onto the surface of magnetic particles can result in a number of
suitable for rapid assays in scientific research of novel potential skin additional functionalities: on one hand the nanostructure of the
whitening agents. The use of biosensors would be an interesting and magnetic particles permits a high enzyme loading without affect-
valuable alternative for the rapid screening and comparison of the ing its natural structure and consequently its activity and on the
inhibitory potency of the existing compounds and the novel ones other hand the beads can be attracted by a magnet and be strongly
due to its relative simplicity in use and the high reproducibility of retained in close proximity to the electrode surface and read-
the analytical data. ily washed away if needed [34]. The streptavidine masterbeads
To the best of our knowledge there have been very few tyrosi- (NMPS) are mainly designed for immunological analyses [35] but
nase based biosensors fabricated for the evaluation of the inhibitory they can very well be used for enzyme immobilization. Based on
potency of kojic acid. A Clark-type oxygen electrode that monitored the IUPAC definition [36] such a configuration is strictly speaking
oxygen consumption during catechol conversion in the presence not a biosensor, i.e. a self-contained integrated device, but since it
of kojic acid inhibitor was reported [13]. Another biosensor for has the enzyme, during the assays, in close spatial contact with the
kojic acid quantification used dopamine as substrate and it mon- electrode we may consider it as a biosensor.
itored the oxygen consumption with a Clark type electrode [14]. In the present work, a novel method for the evaluation of the
There have been a couple of tyrosinase biosensors developed for potency of the tyrosinase and melanogenesis inhibitors used as
the evaluation of the inhibitory potency of benzoic acid [8,15] skin whitening agents is proposed. The biosensor was fabricated
but none of them used l-tyrosine as enzyme substrate. Tyrosinase using tyrosinase immobilized via glutaraldehyde activated strep-
based biosensors usually used phenol or its derivatives as sub- tavidine magnetic nanoparticles and retained onto a “magnetized”
strate with the enzyme being combined with an electrochemical carbon paste electrode, mCPE. The interest in using a carbon paste
transducer to sense either the oxygen consumption of the overall electrode lies in its low background current, allowing for highly
enzymatic reaction [14] or the enzymatic production of the elec- sensitive assay to be performed in the applied potential range
troactive quinone species [16]. There was one article reported of close to 0.0 V versus an Ag/AgCl reference electrode. The employed
a tyrosinase based biosensor for quantitative analysis of phenols. tyrosinase was the mushroom tyrosinase as it is highly homology
Tyrosinase was immobilized to core–shell magnetic particles and with the human enzyme. The substrate used was the l-tyrosine,
subsequently attached to the surface of a carbon paste electrode the enzyme’s natural substrate in the skin and the investigated
with the help of a permanent magnet [17]. inhibitors were kojic acid, azelaic acid, ascorbic acid, the most
Tyrosinase is widely distributed in microorganisms, animals, frequently used active substances in marketed products for hyper-
humans and plants. The enzyme extracted from the mushroom pigmentation, and benzoic acid, a well known preservative in the
Agaricus bisporus is highly homologous with the mammalian one cosmetic industry. The obtained configuration was used to moni-
making it well suited as a model for studies on melanogenesis. tor the reduction current of the enzymatically generated oxidized
Practically all studies of inhibitors conducted so far have used species of l-tyrosine, i.e. the dopaquinone, in the presence of molec-
the mushroom tyrosinase likely because of its readily commercial ular oxygen (Fig. 1). The biosensor was applied to the screening
availability [3]. Yet alternative systems for testing can be obtained of inhibitors which induced a decrease of the reduction current
like mammalian tyrosinase, melanocytic cultures, co-cultures of proportional to the inhibitor concentration and strength.
keratinocytes and melanocytes, skin culture to finally in vivo appli- In practice, the steady state current for l-tyrosine was recorded
cation to animal skin [5,18]. as I0 . The biosensor response in the case of the addition of an
982 V.H. Sima et al. / Talanta 83 (2011) 980–987

A 10 mg/mL tyrosinase-bead suspension was used for l-


tyrosine response optimization and an approximately 1.25 mg/mL
tyrosinase-bead suspension was used in the inhibition assays.
The initial 1 mL amount of the tyrosinase-NMPSs was diluted and
it resulted finally in 8 mL suspension of 1.25 mg/mL tyrosinase-
NMPSs that allowed performing approximately 800 inhibition
assays.
Visible spectrophotometry was applied for the determination
of the efficiency of the immobilization method of tyrosinase onto
the NMPSs. The method consisted in determining the amount of
Fig. 1. Biosensor’s mechanism of action. immobilized tyrosinase by measuring the absorbance of tyrosinase
in a 60 times diluted tyrosinase solution used in the immobiliza-
tion process, before and after the enzyme was immobilized onto the
inhibitor was recorded as I1 . The concentration of inhibitors was activated NMPSs. The percentage of immobilization, calculated by
correlated with the percentage of inhibition (In %) which was calcu- referring to a calibration curve (absorbance versus tyrosinase con-
lated using the relationship: In (%) = [(I0 − I1 )/I0 ] × 100. With I0 and centration) realized in 0.01 M phosphate buffer pH 6.5 ( = 280 nm),
I1 being the biosensor current signals before and after the addi- was found to be around 40 ± 5%. Therefore it was inferred that
tion of the inhibitor, respectively, I1 was expected to be smaller the inhibition assays were realized with approximately 8 Units of
comparing to I0 as the process was inhibited. tyrosinase immobilized at the electrode surface.

2. Materials and methods 2.2.2. Biosensor fabrication


The working carbon paste electrode (mCPE) consisted of a home
2.1. Material made electrode: inside a plastic rod, a permanent magnet (Neody,
3 mm diameter) was inserted leaving a depression at the surface of
The mushroom tyrosinase (EC 1.14.18.1) 5370 Units/mg was approximately 2 mm for housing the solid carbon paste layer, and
purchased from Sigma. l-Tyrosine and l-dopa were provided by a copper wire was used as the electrical conductor.
Janssen Chimica. The Masterbeads Streptavidine (500 nm) were Before each experiment the solid carbon paste was manually
purchased from Ademtech France. Dopamine, glutaraldehyde 25% poured in the hole and the resulting mCPE was smoothed on a clean
wt and monosodium phosphate was provided by Sigma–Aldrich paper surface. The diameter of the active surface was 3 mm. A vol-
and disodium phosphate was provided by Merck. The carbon paste ume of 10 ␮L of the Tyrosinase-NMPSs suspension was spread over
was from Metrohm. Kojic acid was purchased from Kreglinger the surface of the electrode (in position up side down) allowing
Europe, azelaic acid and ascorbic acid from Acros Organics and the particles to settle, being attracted within a few seconds by the
benzoic acid from Fluka. magnet. Subsequently, the tyrosinase-NMPSs immobilized mCPE
All reagents were of analytical grade and were used as received. was oriented in the right position in a three-electrode cell. The
All solutions were prepared with distilled water and stored in the strong attraction of the Tyrosinase-NMPSs by the magnet housed
refrigerator (4 ◦ C) when not in use. The phosphate buffer was pre- inside the working electrode allowed appropriate stirring during
pared by mixing two stock solutions of monosodium phosphate the amperometric experiments.
0.1 M and disodium phosphate 0.1 M to the desired pH. The stock
solutions of l-tyrosine, l-dopa, kojic acid, benzoic acid, azelaic acid 2.3. Apparatus
and ascorbic acid were prepared in phosphate buffer.
The experiments were performed by using as potentiostat a
2.2. Construction of the tyrosinase-NMPS biosensor BASi EPSILON system with a C3 Cell Stand. Amperometry was per-
formed in a conventional three electrodes setup with the biosensor
2.2.1. Enzyme immobilization as working, a Pt wire as auxiliary and an Ag/AgCl 3 M KCl as refer-
The Streptavidin Masterbeads were monodispersed and were ence electrode, respectively. The cyclic voltammetry experiments
based on superparamagnetic particles composed of a magnetic core were performed in the same conventional three electrodes setup
encapsulated by a hydrophilic polymer shell coated with strepta- used in amperometry. All the cyclic voltammetry experiments were
vidin. The magnetic particles had a diameter of 500 nm (RSD max recorded at 100 mV s−1 or 10 mV s−1 in a potential window com-
20%), density of approx. 2.0 g/cm3 , magnetic susceptibility approx. prised between −0.6 V and +1.1 V versus Ag/AgCl 3 M KCl reference
40 emu/g, specific surface area of 5 m2 /g, iron oxide content approx. electrode. The spectrophotometric assays were performed using
70% m/m and the solid content of 10 mg/mL [33]. an Agilent Hewlett Packard Diode Array UV Vis Spectrophotometer
An amount of 1 mL 10 mg/mL NMPSs was washed with 0.01 M Model 8453. The pH of the solution was controlled with a Metrohm
phosphate buffer pH 6.5 (5 mL) then reacted with 1 mL of 2.5% 827 pH Lab system. All experiments were performed at room tem-
wt glutaraldehyde solution (in 0.01 M phosphate buffer, pH 6.5) perature (23 ◦ C).
at room temperature for 30 min. After washing with 0.01 M phos-
phate buffer pH 6.5 (5 mL), the glutaraldehyde-treated NMPSs 2.4. Amperometric assay
were reacted with 1 mL tyrosinase solution (3 mg/mL in phosphate
buffer 0.01 M, pH 6.5) during 1 h at room temperature to obtain During amperometric experiments the biosensor potential was
the tyrosinase-NMPSs. The resulting suspension was washed with kept at −100 mV under continuous stirring conditions with a mag-
0.01 M phosphate buffer pH 6.5 (5 mL) and stored in 1 mL 0.01 M netic spinbar at 300 rpm. The working potential was imposed and
phosphate buffer pH 6.5 at 4 ◦ C. During the washing steps, the the background current was allowed to reach a steady state value
beads were trapped by magnetic forces by placing the reacting within approx. 10 min. Different amounts of tyrosinase standard
Eppendorf tube close to a strong magnet and the supernatant was solution or inhibitor solution were added every 100 or 200 s and the
carefully pipetted off. The final concentration of tyrosinase-NMPSs current was recorded as a function of time in 10 mL 0.1 M phosphate
was 10 mg/mL (referring to the amount of NMPS) [35]. buffer solution.
V.H. Sima et al. / Talanta 83 (2011) 980–987 983

2.0E-06 used l-tyrosine as a substrate, reported in the literature. Moreover

I (A)
1.5E-06
it has been demonstrated that the intensity of inhibition varied
b
considerably depending on the characteristics of the substrate [19].
1.0E-06
The biosensor response was tested comparatively both in the pres-
5.0E-07 a ence of l-tyrosine and l-dopa using 10 ␮L of the tyrosinase-NMPS
0.0E+00
suspension (10 mg/mL) deposited onto the surface of the mCPE. It
-1.0 -0.5 0.0 0.5 1.0 1.5 was observed that, in the concentration range 4.98 × 10−6 M and
-5.0E-07 2.91 × 10−5 M, the sensitivity to l-dopa (−5.56 × 10−4 A/M) was
E (V) vs Ag/AgCl
-1.0E-06 almost 6 times higher comparing to l-tyrosine (−9.37 × 10−5 A/M)
and the response time (the time necessary for reaching 95% of the
-1.5E-06
maximum response) to l-dopa (56 s) was faster comparing to l-
Fig. 2. Repetitive CV for l-tyrosine 4.76 × 10−5 M, 0.1 M phosphate buffer pH 6.5, tyrosine (86 s). l-Tyrosine, however, was chosen as model substrate
mCPE, scan rate, a: 10 mV s−1 , b: 100 mV s−1 . for subsequent experiments. This decision was based on the fact
that, in vivo, l-tyrosine is the natural substrate of tyrosinase and
After each amperometric inhibition assay, the tyrosinase- also the starting point of melanin synthesis and its oxidation, to
NMPSs were readily removed by flushing a water burst over the l-dopa, is the rate-limiting step of the whole process [3]. The sensi-
biosensor surface. The surface of the mCPE was renewed after each tivity to l-tyrosine was considered high enough for assuring reliable
inhibition assay. results and the lag period of its first enzymatic oxidizing step to l-
dopa was sufficiently short for further readily implementation of
the testing conditions.
3. Results and discussions
3.1.3. Selection of the pH
3.1. Working conditions
The biosensor response was studied in the pH range comprised
between 5.5 and 7. It is well known that enzyme activity is highly pH
The working conditions for an optimal response to the sub-
dependent and that the optimum pH for an enzymatic assay must
strate of the tyrosinase based biosensor have been tested. Since
be determined empirically [39]. By decreasing the pH from 7.0 to
the biosensor was designed for testing inhibitors of the skin pig-
5.5, the amperometric signal increased but no stable steady-state
mentation process it was taken into account to operate as close as
plateau was obtained at pH 6.0 and 5.5. For further experiments,
possible to the natural environment of the skin tyrosinase which is
the pH chosen was 6.5. At this pH, the current and the steady-state
a transmembrane protein of melanosomes [37].
of the biosensor response were considered optimal, in agreement
with different literature data where the working pH of the tyrosi-
3.1.1. Selection of the applied potential nase based assays vary between 6.0 and 7.0. [9,14,22,23]. The
It is well known that l-tyrosine enzymatic oxidation pro- selected pH corresponded to the Sigma declared tyrosinase stabil-
duces dopaquinone and it was expected that this product could ity pH and is also close to the normal pH range of the skin which is
be detected by electroreduction at the mCPE. Likewise the elec- considered to vary between 4 and 6.5.
trochemical oxidation of l-tyrosine was supposed to generate
dopaquinone. Cyclic voltammetry (CV) performed at 10 and 3.1.4. Amount of tyrosinase-NMPS spiked onto the mCPE
100 mV/s (from 0.0 mV to 1.1 V, and back to −0.6 V) in 0.1 M phos- The amperometric response was determined as a function of l-
phate buffer pH 6.5, showed the irreversible oxidation peak of tyrosine, in the concentration range 4.98 × 10−6 and 2.91 × 10−5 M,
l-tyrosine (1 × 10−3 M and 4.76 × 10−5 M) since no reduction peak using different amounts (5 ␮L, 10 ␮L and 20 ␮L) of the 10 mg/mL
was observed in the investigated potential domain (Fig. 2). Same tyrosinase-NMPS suspension spiked onto the surface of the mCPE.
experiments were performed also for l-dopa which gave an oxida- The response increased proportionally by raising the amount of
tion peak at 0.550 V but still no reduction peak was observed. The immobilized enzyme onto the surface of the electrode. The sen-
absence of a reduction current and the progressive decrease of the sitivities obtained were: −6.82 × 10−5 A/M, −9.37 × 10−4 A/M
l-tyrosine peak during multiple scanning suggested that the gen- and −1.68 × 10−4 A/M for 5, 10 and 20 ␮L spiked suspension,
erated dopaquinone escaped rapidly from the electrode solution respectively. The repeatability of 5 ␮L deposited beads was poor
interface likely due its high reactivity giving rise to polymer-like comparing to 20 ␮L and 10 ␮L deposits. The latter was selected
structures fouling the electrode surface [38]. Amperometry of l- in subsequent work since it offered good sensitivity and repeata-
tyrosine (2.44 × 10−5 M) at the tyrosinase-NMPS mCPE, thanks to bility (RSD = 2.8%, n = 3) and it consumed less beads then a 20 ␮L
the low background current compared to CV, permitted, however, deposit. For the subsequent inhibition assays it was decided to use
to detect a reduction current at potentials near 0 V versus Ag/AgCl a diluted suspension of tyrosinase-NMPS (1.25 mg/mL) since it gave
likely corresponding to the electroreduction of some enzymatically a good response to l-tyrosine (3.33 × 10−4 M) and allowed to reduce
generated dopaquinone remaining at the electrode solution inter- substantially the cost of the assays.
face. Amperometric experiments were realized at 0, −50, −100,
−150, −200 and −250 mV. By decreasing the applied potential the 3.1.5. Selection of the substrate concentration for the inhibition
reduction current increased but the response steady-state became studies
less stable. An applied potential of −100 mV was considered opti- The opinions about the substrate concentration to be consid-
mal taking into account the magnitude of the reduction current, ered in inhibition assays are quite contradictory. Kok et al. [40]
the ratio between signal and background current, the steady-state concluded, when measuring the inhibition potency of a competi-
of the response plateau and the possible interfering species at the tive inhibitor with an acetylcholinesterase and a choline oxidase
applied potential, in agreement with other tyrosinase based biosen- biosensor, that the inhibition percentage increased by raising the
sors described in the literature [21]. substrate concentration. Therefore they worked in enzyme satu-
rated substrate conditions. Shan et al. [19] demonstrated that for
3.1.2. Selection of the substrate a tyrosinase biosensor for the determination of benzoic acid, the
To the best of our knowledge there is no tyrosinase based amper- concentration of the catechol substrate did not affect the maxi-
ometric biosensor for the evaluation of skin-whitening agents that mum inhibition percentage but it affected the sensitivity of the
984 V.H. Sima et al. / Talanta 83 (2011) 980–987

method. It must be noted, however, that the use of a high sub- Table 1
IC50 for kojic acid, benzoic acid, azelaic acid and ascorbic acid.
strate concentration would not yield sensitive inhibition responses
when the quantification of a competitive inhibitor is performed by IC50 (␮M) Kojic acid Benzoic acid Azelaic acid Ascorbic acid
simultaneous addition of the inhibitor and the substrate. Because IC50 1 3.66 72.0 125 12.4
in competitive inhibition the substrate competes with the inhibitor IC50 2 3.57 72.0 127 11.6
for the enzyme active site, and the inhibition, especially at low IC50 3 3.71 72.6 128 11.0
inhibitor concentrations, would likely not be detected [40]. IC50 4 3.72 70.9 119 12.2
IC50 5 3.63 70.5 127 11.7
Two different l-tyrosine concentrations were studied;
IC50average 3.66 71.6 125 11.8
3.33 × 10−4 M and 4.76 × 10−5 M, and the obtained IC50 for
kojic acid were found to be equal within the experimental error
(RSD = 2.3%). The percentage of inhibition was evaluated from The proposed biosensor was intended to test inhibitors that
the response of the uninhibited form of the enzyme versus the have two different mechanisms of action, namely: true enzyme
response of the enzyme partially inhibited. Taking into account that inhibitors which bind to the active site of the enzyme and inhibitors
our device was not designed for the quantification of inhibitors but which consume the dopaquinone intermediate of the melanin syn-
for the comparative screening and quantification of their inhibitory thesis pathway.
potency we considered that all the immobilized enzyme molecules It was checked that the NMPS mCPE (i.e. without tyrosinase)
must take part in the reaction and this could only be possible in gave no response under the selected experimental conditions and
substrate concentration corresponding to the saturation portion in the presence of the studied compounds. It was found that l-
of the activity versus substrate curve [40]. Enzyme saturated tyrosine, kojic acid, benzoic acid, azelaic acid and ascorbic acid gave
substrate concentration, namely 3.33 × 10−4 M l-tyrosine, was no amperometric signal.
then selected for further inhibition studies. Also blank experiments at the tyrosinase-NMPS-mCPE without
Since the concentration of substrate was quite high, it had to be addition of the l-tyrosine substrate were performed. No ampero-
established if the quantity of the dissolved oxygen in the solution metric response was observed for the tested inhibitors.
was not a limiting factor for a 3.33 × 10−4 M l-tyrosine concentra-
tion. After 30 min of air bubbling into the working buffer solution 3.2.1. “True” enzyme inhibitors
the biosensor response did not change, but after 30 min of nitrogen Kojic acid, benzoic acid and azelaic acid inhibit the melanin for-
bubbling the biosensor response decreased dramatically. The latter mation by competing with tyrosinase natural substrate, l-tyrosine,
was also a further indirect evidence of the generation of an elec- for the binding to the enzyme active site. As illustrated in Fig. 3a
troactive species by the immobilized tyrosinase. It was concluded typical time-dependent response was obtained at the tyrosinase-
that in the selected experimental conditions the quantity of oxy- NMPSs biosensor by l-tyrosine injection followed by stepwise
gen was not a limiting factor and that the enzymatic process was additions of kojic acid. It can be seen that the biosensor baseline
saturated by the substrate. was reached within 600 s allowing the experiment to be per-
formed. Once the steady state amperometric response of l-tyrosine
3.1.6. Michaelis–Menten curve (3.33 × 10−4 M) was observed, successive additions of inhibitor
A typical Michaelis–Menten plot and its linearization by the produced a diminution of the reduction current. This clearly indi-
Lineweaver–Burk plot (y(A−1 ) = −7.91 × 103 × (M−1 )−8.53 × 107 , cated that kojic acid interfered with the production of dopaquinone
R2 = 1.000, RSD slope = 3.3%, RSD intercept = 7.7%, n = 3) was at the electrode surface. The inhibition percentage (In %) increased
obtained for l-tyrosine between 2.49 × 10−6 M and 4.41 × 10−4 M with the inhibitor concentration. Kojic acid was tested in the con-
in the above described optimal working conditions of the tyrosinase centration range 2.50 × 10−6 to 1.24 × 10−5 M where it inhibited
based biosensor, namely: applied potential (Eapp ) = −100 mV, 0.1 M between 33% and 88% of the response to l-tyrosine. Same experi-
phosphate buffer pH 6.5, 10 ␮L of 1.25 mg/mL tyrosinase-NMPS ments were performed for benzoic acid in the concentration range
suspension spiked onto the mCPE surface. The kinetic parameters 3.98 × 10−5 to 1.96 × 10−4 M where it inhibited between 34% and
Km app and Imax were calculated as the average of 3 consecu- 81% of the response and also for azelaic acid in the concentration
tive determinations: Km app = 9.7 × 10−5 M (RSD = 3.1%, n = 3) and range 9.90 × 10−5 to 4.76 × 10−4 M where it inhibited between 43%
Imax = −1.2 × 10−8 A (RSD = 7.5%, n = 3). and 88% of the response.
The obtained Km app values (Table 2) can be compared to The inhibition calibration curve of kojic acid is shown in
those reported in the literature for the enzyme in solution, i.e. the insert of Fig. 3. IC50 , the concentration of inhibitor which
Km app = 6.2 × 10−4 M [23] or 3.3 × 10−4 M [41] at pH 6.5, illustrating inhibited 50% of the l-tyrosine signal, was calculated from 5
the fact that the applied immobilization technique maintained the different curves by plotting 1/In(%) versus 1/inhibitor concen-
enzyme affinity for its substrate. tration (Table 1): kojic acid IC50 = 3.7 × 10−6 M, RSD = 1.6%, n = 5
(y = 4.54 × 10−8 x + 7.61 × 10−3 , R2 = 0.9999, RSD slope = 3.1%, RSD
3.2. Inhibitors assay intercept = 4.1%), benzoic acid IC50 = 7.2 × 10−5 M, RSD = 1.2%, n = 5
(y = 7.88 × 10−7 x + 8.99 × 10−3 , R2 = 0.9990, RSD slope = 5.1%, RSD
The developed biosensor was designed for testing inhibitors of intercept = 6.3%) and azelaic acid IC50 = 1.3 × 10−4 M, RSD = 3.0%,
the skin pigmentation process. When testing their potency, there n = 5 (y = 1.60 × 10−6 x + 7.76 × 10−3 , R2 = 0.9996, RSD slope = 9.3%,
are many factors which contribute to the final results: type of RSD intercept = 7.6%).
enzyme, quantity of immobilized enzyme, immobilization method, The study of the kinetics and of the mechanism of inhibition
type of substrate, substrate concentration, time of contact between of kojic acid, benzoic acid and azelaic acid was also carried out.
the enzyme, substrate and inhibitor, pH, temperature, applied The response of the tyrosinase-NMPSs biosensor to l-tyrosine was
potential, rate of solution stirring. In order to obtain a correct com- studied in the absence and in the presence of different concentra-
parison of the inhibition potency of the studied compounds, it was tions of inhibitor (Fig. 4). Table 2 data show that approximately
worked under the same experimental conditions [19]. the same maximum current was obtained but different values
As the developed device was a novel analytical tool, in order to for the apparent Michaelis–Menten constant (Km app ), determined
test its efficiency, it was applied to the study of compounds that following a Lineweaver–Burk plot, were calculated when various
have already demonstrated their activity in clinical practice or by amounts of kojic acid was added in the initial testing solution. The
using other analytical methods. same experiment was performed and the same conclusions were
Academic Press is an imprint of Elsevier
225 Wyman Street, Waltham, MA 02451, USA
525 B Street, Suite 1800, San Diego, CA 92101–4495, USA
125 London Wall, London, EC2Y 5AS, UK
The Boulevard, Langford Lane, Kidlington, Oxford OX5 1GB, UK

First edition 2015

© 2015 Elsevier Inc. All rights reserved.

No part of this publication may be reproduced or transmitted in any form or by any means,
electronic or mechanical, including photocopying, recording, or any information storage and
retrieval system, without permission in writing from the publisher. Details on how to seek
permission, further information about the Publisher’s permissions policies and our
arrangements with organizations such as the Copyright Clearance Center and the Copyright
Licensing Agency, can be found at our website: www.elsevier.com/permissions.

This book and the individual contributions contained in it are protected under copyright by
the Publisher (other than as may be noted herein).

Notices
Knowledge and best practice in this field are constantly changing. As new research and
experience broaden our understanding, changes in research methods, professional practices,
or medical treatment may become necessary.

Practitioners and researchers must always rely on their own experience and knowledge in
evaluating and using any information, methods, compounds, or experiments described
herein. In using such information or methods they should be mindful of their own safety and
the safety of others, including parties for whom they have a professional responsibility.

To the fullest extent of the law, neither the Publisher nor the authors, contributors, or editors,
assume any liability for any injury and/or damage to persons or property as a matter of
products liability, negligence or otherwise, or from any use or operation of any methods,
products, instructions, or ideas contained in the material herein.

ISBN: 978-0-12-803300-5
ISSN: 1871-5125 (Series)

For information on all Academic Press publications


visit our website at http://store.elsevier.com/
Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy 151 (2015) 945–955

Contents lists available at ScienceDirect

Spectrochimica Acta Part A: Molecular and


Biomolecular Spectroscopy
journal homepage: www.elsevier.com/locate/saa

Continuous Wavelet Transform, a powerful alternative to Derivative


Spectrophotometry in analysis of binary and ternary mixtures: A
comparative study
Eman S. Elzanfaly, Said A. Hassan ⇑, Maissa Y. Salem, Badr A. El-Zeany
Department of Analytical Chemistry, Faculty of Pharmacy, Cairo University, Kasr El-Aini Street, 11562 Cairo, Egypt

h i g h l i g h t s g r a p h i c a l a b s t r a c t

 Numerical Differentiation and


Wavelet Transform as approaches for
derivative calculation.
 Wavelet Transform is a powerful
alternative to traditional derivative
algorithms.
 Simple, selective and precise
spectrophotometric methods for
binary and ternary mixtures.
st
 The 1 spectrophotometric method
for analysis of Amlodipine, Aliskiren
and Hydrochlorothiazide.
 Methods validated as per ICH
guidelines, parameters found to be
within the limits.

a r t i c l e i n f o a b s t r a c t

Article history: A comparative study was established between two signal processing techniques showing the theoretical
Received 21 April 2015 algorithm for each method and making a comparison between them to indicate the advantages and
Received in revised form 20 June 2015 limitations. The methods under study are Numerical Differentiation (ND) and Continuous Wavelet
Accepted 25 June 2015
Transform (CWT). These methods were studied as spectrophotometric resolution tools for simultaneous
Available online 30 June 2015
analysis of binary and ternary mixtures. To present the comparison, the two methods were applied for
the resolution of Bisoprolol (BIS) and Hydrochlorothiazide (HCT) in their binary mixture and for the
Keywords:
analysis of Amlodipine (AML), Aliskiren (ALI) and Hydrochlorothiazide (HCT) as an example for ternary
Derivative Spectrophotometry
Continuous Wavelet Transform
mixtures. By comparing the results in laboratory prepared mixtures, it was proven that CWT technique
Amlodipine is more efficient and advantageous in analysis of mixtures with severe overlapped spectra than ND.
Aliskiren The CWT was applied for quantitative determination of the drugs in their pharmaceutical formulations
Bisoprolol and validated according to the ICH guidelines where accuracy, precision, repeatability and robustness
Hydrochlorothiazide were found to be within the acceptable limit.
Numerical Differentiation Ó 2015 Elsevier B.V. All rights reserved.

1. Introduction

UV–VIS absorption spectroscopy is a well-established technique


for rapid and accurate determination of analytes in a mixture form
without prior separation, if the interferences between the spectra
⇑ Corresponding author.
can be eliminated. One of the techniques used for elimination of
E-mail addresses: said.hassan@pharma.cu.edu.eg, saidmonem_84@yahoo.com
(S.A. Hassan).
interference in spectroscopy is signal processing. The most

http://dx.doi.org/10.1016/j.saa.2015.06.100
1386-1425/Ó 2015 Elsevier B.V. All rights reserved.
946 E.S. Elzanfaly et al. / Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy 151 (2015) 945–955

commonly used signal processing technique is Derivative AmturnideÒ is a combination of Amlodipine besylate (AML),
Spectrophotometry (DS). Aliskiren hemifumarate (ALI) and Hydrochlorothiazide (HCT) indi-
Rutherford [1] was the first to apply derivative methods suc- cated for the treatment of hypertension to reduce the risk of
cessfully to experimental data, he suggested that the sensitivity strokes and myocardial infarctions [56]. Many reported methods
of mass spectroscopy would be increased on applying the first were reported for the determination of AML, ALI or HCT in different
derivative technique. Derivative methods were first applied to dosage forms [57–62]. HPLC and capillary electrophoresis methods
UV–VIS and IR spectroscopy in 1953 by Hammond and Price [2], were reported for simultaneous determination of this ternary mix-
followed by Morrison [3] and in 1955 Giese and French [4] ture [63,64]. No spectrophotometric methods were developed for
described the theory of the derivative technique. In 1959, Martin the determination of this mixture.
[5] introduced the theoretical basis for second and higher orders In this work, ND and CWT were applied for determination of
of derivative, and in 1968, Morrey [6] generated first to fourth Bisoprolol (BIS) and Hydrochlorothiazide (HCT) in their binary
order derivatives with a digital computer. In 1964, Savitzky and mixture and for determination of Amlodipine (AML), Aliskiren
Golay [7] introduced least-squares procedures for the smoothing (ALI) and Hydrochlorothiazide (HCT) in their ternary mixture as
and differentiation of fluctuating data. examples for complex spectra. A comparative study between the
Derivative calculation is a powerful technique used in analytical two methods was performed to compare their resolution power
chemistry to resolve spectra, sharpen peaks, remove background in analysis of overlapped spectra. Then the CWT was used for the
interference, and carry out quantitative analysis. The most impor- analysis of the cited mixtures in pharmaceutical formulations,
tant field is its use in resolution enhancement technique to facili- namely ConcorÒ Plus tablets and AmturnideÒ tablets.
tate the detection and location of poorly resolved components in
complicated experimental signals [8]. It was applied successfully
to chromatography [9], electrochemistry [10–12], flow-injection 2. Theoretical background
analysis [13] and in all fields of spectroscopy [14–16]. In the region
of UV–VIS spectroscopy, DS played an important role in resolution 2.1. Numerical Differentiation (ND) [65]
of overlapped spectra [17,18]. The zero-crossing derivative spectra
method was the most common procedure for the simultaneous For a discrete spectrum xi (i = 1, . . ., n), suppose that wi
determination of single components as well as binary and ternary (i = 1, . . ., n) are its sampling wavelengths, the direct-difference
mixtures from their overlapping spectra [19–21]. DS was also method can be expressed as follows:
applied to ratio spectra for analysis of binary and ternary mixtures
X iþ1  X i
[22–24]. yi ¼
W iþ1  W i
In the spectral studies, the application of the usual DS to the
original absorption spectra possesses several disadvantages such
Here yi (i = 1, . . ., n  1) represent the series of derivatives of spec-
as: peak intensity diminishes with higher order derivation, it
trum xi. If the sampling wavelength is equal, this equation can be
required the additional smooth function as well as the scaling
rewritten as yi = xi+1  xi (i = 1, . . ., n  1).
factor processes. These parameters may deform the obtained
derivative spectra from the original ones. As a result, the usual
derivative method produces several errors in the quantitative anal- 2.2. Wavelet Transform (WT): [27]
ysis [25]. These drawbacks suggested using other signal processing
techniques as powerful alternatives for derivative calculation such Wavelet Transform (WT) involves decomposition of a signal
as Wavelet Transform [26,27]. function or vector (spectrum) into simpler, fixed building blocks
A Wavelet Transform (WT) involves the decomposition of a sig- at different scales and positions. In WT treatment, all basis func-
nal function or vector e.g., a spectrum of a chemical species into tions Wa;b ðxÞ can be derived from a mother wavelet WðxÞ through
simpler, fixed building blocks at different scales and positions the following dilation and translation processes:
[28]. Different algorithms were proposed to achieve this purpose.  
In 1977, Coifman and Weiss [29] developed the automatic decom- 1 xb
Wa;b ðxÞ ¼ a2 W a; b 2 R and a–0
position method, based on work of Calderón and Zygmund [30], a
which had a very strong influence on the development of wavelet
theory. In 1984, Grossmann and Morlet [31] proposed the where a and b are the scale and position parameters, respectively,
Continuous Wavelet Transform (CWT). In 1989, Mallat [32] intro- expressed in real number R.
duced the multi-resolution signal decomposition MRSD algorithm. The basic idea of WT is to represent any arbitrary function f ðxÞ
Daubechies [26,33,34] adopted this approach to construct families as a superposition of wavelets. Decomposition of x with respect to
of compactly supported wavelets and coupled it to quadrature mir- the wavelet function series {Wj;k ðxÞ} is described by the following
ror filtering. formula:
WT has been applied successfully for signal processing in differ-
X
þ1 X
þ1
ent fields of analytical chemistry. It was applied to de-noising of f ðxÞ ¼
ðjÞ
C k Wj;k
data [35], compression of signals [36], analysis of electrochemical j¼1 k¼1
data [37], flow-injection analysis [38], multivariate calibration
[39] and quantitative analysis of near infrared spectra [40]. In the Therefore, the signal is represented by a set of coefficients {C j;k } in
field of UV–VIS spectrophotometry, Continuous Wavelet the wavelet domain.
Transform (CWT) combined either with a zero-crossing technique The approximate first derivative of X is assigned to be equal to
[41] or ratio spectra [13] was used for simultaneous determination the difference between two scale coefficients at the first resolution
of chemical species in binary and ternary mixtures [18,42–45]. level as
ConcorÒ Plus tablets contain both Bisoprolol hemifumarate
(BIS) and Hydrochlorothiazide (HCT) for the treatment of high xð1Þ  C 1;D2m  C 1;D2m~ ~
m–m
blood pressure [46]. There are reported methods for the determi-
nation of BIS or HCT in different dosage forms [47–50] and in their where, D2m and D2m~ represent any two Daubechies wavelet
binary mixtures [45,51–55]. ~ being any positive integer.
functions, with m and m
E.S. Elzanfaly et al. / Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy 151 (2015) 945–955 947

3. Experimental 3.3.3.1. Numerical Differentiation method (ND).

3.1. Apparatus Binary mixture

SHIMADZU dual beam UV–visible spectrophotometer For determination of BIS, the 3rd and 4th derivative spectra were
(Kyoto/Japan), model 1650 UV-PC, with matched 1-cm quartz cells obtained by ND method with Dk = 8 nm and scaling factor = 1000.
was used. The bundled software, UV-Probe personal spectroscopy The peak amplitude was measured at 283.6 and 287.5 nm,
software version 2.21 (SHIMADZU) is used. The spectral bandwidth respectively.
is 2.0 nm and wavelength scanning speed is 2800 nm/min with For determination of HCT, the first derivative spectra were
0.1 nm interval. obtained by ND method with Dk = 4 nm and scaling factor = 100.
The peak amplitude was measured at 272.3 nm.
3.2. Chemicals and reagents The calibration curves were constructed relating the peak
amplitudes at the specified wavelengths to the corresponding
3.2.1. Pure samples concentrations of BIS and HCT.

Ternary mixture
- BIS, HCT and AML; kindly supplied by Al-Hekma pharmaceuti-
cal Company, Cairo, Egypt, their purity was certified to be
For determination of AML, the first derivative spectra were
99.7 ± 0.8%, 100.6 ± 0.6% and 99.9 ± 0.5%, respectively.
obtained by ND method with Dk = 4 nm and scaling factor = 10.
- ALI; kindly supplied by Novartis Pharmaceuticals Corporation,
The peak amplitude was measured at 386.8 nm.
USA, its purity was certified to be 100.6 ± 0.3%.
For determination of ALI, the fourth derivative spectra were
obtained by ND method with Dk = 8 nm and scaling factor = 1000.
3.2.2. Market samples
The peak amplitude was measured at 287.5 nm.
For determination of HCT, the second derivative spectra were
- ConcorÒ 5 Plus labeled to contain 5(BIS)/12.5(HCT) mg, batch
obtained by ND method with Dk = 8 nm and scaling factor = 100.
number 1030039 and ConcorÒ 10 Plus labeled to contain
The peak amplitude was measured at 332.1 nm.
10(BIS)/25(HCT) mg, batch number 0795049, manufactured
The calibration curves were constructed relating the peak
by Pfizer Ltd., Cairo, Egypt.
amplitudes at the specified wavelengths to the corresponding
- AmturnideÒ tablet dosage forms; labeled to contain 10
concentrations of AML, ALI and HCT.
(AML)/300 (ALI)/25 (HCT) mg batch number F0005, manufac-
tured by Novartis Pharmaceuticals Corporation, USA.
3.3.3.2. Continuous Wavelet Transform method (CWT).

3.2.3. Methanol
Binary mixture
Spectroscopy grade (El-NASR Pharmaceutical Chemicals Co.,
Abu-Zaabal, Cairo, Egypt).
The first derivative spectra were obtained by CWT method with
Morlet family (morl) and scaling = 200. The calibration curves were
3.3. Procedures constructed relating the peak amplitudes at 278.3 and 283.6 nm to
the corresponding concentrations of BIS and HCT, respectively.
3.3.1. Standard solutions
Ternary mixture
- BIS, HCT, AML and ALI standard stock solutions; 1000 lg /mL in
methanol. For determination of AML, the first derivative spectra were
- BIS, HCT and AML standard working solutions; 50 lg/mL in obtained by CWT method with Gaussian-1 family (Gaus-1) and
methanol. scaling = 200. The peak amplitude was measured at 387.2 nm.
- ALI standard working solution; 500 lg/mL in methanol. For determination of ALI, the first derivative spectra were
- Preparation of laboratory prepared mixtures; aliquots of stan- obtained by CWT method with Gaussian-8 family (Gaus-8) and
dard working solutions were transferred into two series of scaling = 200. The peak amplitude was measured at 315.6 nm.
10-mL measuring flasks, completed to volume with methanol For determination of HCT, the first derivative spectra were
to prepare binary mixture of BIS and HCT and ternary mixture obtained by CWT method with Coiflet-3 family (Coif-3) and
of AML, ALI and HCT containing different ratios of the drugs. scaling = 200. The peak amplitude was measured at 289.1 nm.
The calibration curves were constructed relating the peak
3.3.2. Spectral characteristics of the drugs amplitudes at the specified wavelengths to the corresponding
The zero-order (D0) absorption spectra of 4 lg/mL BIS and concentrations of AML, ALI and HCT.
10 lg/mL HCT were recorded against methanol as a blank over
the range of 200–400 nm. 3.3.4. Application of the proposed methods for the determination of
The zero-order (D0) absorption spectra of 4, 120 and 10 lg/mL laboratory-prepared mixtures
solutions of AML, ALI and HCT, respectively, were recorded against The spectra of laboratory prepared mixtures were scanned from
methanol as a blank over the range of 200–400 nm. 200 to 400 nm and stored in the computer. The same procedure
under construction of calibration curves was applied and the con-
3.3.3. Construction of calibration curves centrations of the drugs were calculated from the corresponding
Aliquots equivalent to 20–400 lg BIS, 5–250 lg HCT, 10–350 lg regression equations.
AML and 150–2500 lg ALI were accurately transferred from their
respective standard working solutions into separate series of 3.3.5. Application of the proposed methods for the determination of the
10-mL volumetric flasks then completed to volume with methanol. drugs in dosage forms
The spectra of the prepared standard solutions were scanned from Ten tablets of both ConcorÒ 5 Plus, ConcorÒ 10 Plus were accu-
200 to 400 nm and stored in the computer. rately weighed, finely powdered and an amount of the powder
948 E.S. Elzanfaly et al. / Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy 151 (2015) 945–955

equivalent to 10 mg BIS was weighed. Five tablets of AmturnideÒ ALI and HCT show highly overlapped spectral band in the region
were accurately weighed, finely powdered and amount of the pow- 200–350 nm (Fig. 2). The difficulty of BIS determination in the bin-
der equivalent to 300 mg ALI was weighed. The powder was dis- ary mixture and the severe overlap in the ternary mixture nomi-
solved in methanol by shaking in ultrasonic bath for about nated these combinations to be ideal mixtures presenting the
30 min, then they were filtered and transferred quantitatively into comparative study to show the advantages and disadvantages of
separate 100-ml volumetric flasks and the volume was completed both algorithms in analysis of overlapping spectra. HCT in the bin-
to the mark with methanol. Ten ml aliquots were transferred into ary mixture and AML in the ternary mixture can be directly deter-
separate 50-ml volumetric flasks and the volume was completed to mined in the zero order spectra at their kmax, 316 and 360 nm
the mark with methanol. Then necessary dilutions were made to respectively, without interference from other components, but
reach concentrations in the linearity range. The spectra of these they were determined by ND and CWT for investigational
solutions were scanned from 200 to 400 nm and stored in the com- purposes.
puter. Spectra obtained were analyzed by the proposed methods. For using ND method, the parameters of derivative calculation,
wavelength increment over which the derivative is obtained (Dk)
and the scaling factor, were optimized. Different Dk (2, 4 and 8)
4. Results and discussion
and scaling factors of 10, 100 and 1000 were tested, and the best
parameters, in terms of spectral shapes, linearity and recovery,
Signal processing techniques use different algorithms in trans-
are mentioned under procedure.
forming the spectra into new graphs where the spectra of different
For analysis of chosen mixtures using CWT algorithm, different
compounds are resolved. Derivative Spectrophotometry (DS) is the
wavelet families were tested such as Haar, Daubechies (db),
most commonly used processing technique for resolution of over-
Symlets (sym), Coiflets (coif), Meyer (meyr), Gaussian (gaus),
lapped spectra. Lately, Continuous Wavelet Transform algorithm
Mexican hat (mexh) and Morlet (morl). The only parameter that
(CWT) started to be used as an alternative to Numerical
should be optimized after choosing the proper wavelet is the scal-
Differentiation (ND) for calculation of DS.
ing parameter. No need for scaling factor as CWT has the advantage
The aim of this work was to establish a comparative study
of signal amplification compared to ND.
between the two signal processing techniques. For proper presen-
tation of the comparison, two different mixtures were chosen to
show the resolution power of the methods. The first mixture was 4.1. Binary mixture
a combination of BIS and HCT as an example of binary mixtures.
The spectra of BIS and HCT (Fig. 1) show severe overlap as the peak In this section, we will focus on determination of BIS being a
of HCT (kmax 226 nm) completely coincide with that of BIS (kmax challenge in presence of highly absorptive HCT. Thus, the choice
225 nm) showing the same features. This overlap, together with of derivative order and optimum parameters is mainly considered
the high absorptivity of HCT which is also found in higher level for determination of BIS.
in ConcorÒ Plus tablets, prevents direct determination of BIS. In For analysis of BIS in this mixture using ND algorithm, the 1st,
previous work, the transformed ratio spectra had to be combined 2 , 3rd and 4th order derivatives were obtained (Fig. 3). The 1st
nd

with zero-crossing technique for analysis of BIS in this mixture order did not show any possible zero-crossing points that can be
[45]. The second mixture composed of AML, ALI and HCT was used used for BIS determination (Fig. 3a), while the wavelengths 241.0
as an example of ternary mixtures. The absorption spectra of AML, and 270.0 nm did not show a real zero-crossing. Possible
Absorbance

Wavelength (nm)

Fig. 1. Zero order absorption spectra of 4 lg/mL BIS (—) and 10 lg/mL HCT (- - -) using methanol as blank.
E.S. Elzanfaly et al. / Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy 151 (2015) 945–955 949

Absorbance

Wavelength (nm)

Fig. 2. Zero order absorption spectra of 4 lg/mL AML (—), 120 lg/mL ALI (....) and 10 lg/mL HCT (- - -), using methanol as blank.

zero-crossing points were noticed in higher order derivatives, For analysis of the ternary mixture using CWT algorithm, differ-
which are 230.3 and 278.5 nm in 2nd order, 233.3 and 283.6 nm ent families were tried in different orders and scaling. The best
in 3rd order and 229.5, 235.3 and 287.5 nm in 4th order. In contrast results regarding presence of zero-crossing points, linearity and
to 1st order derivative, true zero-crossing points were detected in selectivity were obtained with 1st order derivatives using gaus-1,
2nd (278.5 nm), 3rd (283.6 nm) and 4th (287.5 nm) orders gaus-8 and coif-3 families and scaling = 200 for determination of
(Fig. 3b–d). The zero-crossing point in 2nd order showed poor lin- AML, ALI and HCT at 387.2, 315.6 and 289.1 nm, respectively
earity and recovery of laboratory prepared mixtures, while the (Fig. 6).
points in 3rd and 4th orders resulted in good recoveries in some The selectivity of the proposed ND and CWT methods was
laboratory-prepared mixtures, where BIS was found in higher ratio assessed by the analysis of laboratory prepared mixtures contain-
than HCT. For determination of HCT, the 1st order derivative was ing different ratios of the drugs (Tables 1 and 2). The proposed
enough to get zero-crossing point (272.3 nm) and good recoveries CWT method was applied for the determination of BIS and HCT
in all ratios of laboratory-prepared mixtures were obtained in ConcorÒ Plus tablets and AML, ALI and HCT in AmturnideÒ
Table 1. tablets and the validity was further assessed by applying the stan-
For analysis of BIS using CWT algorithm, the families were dard addition technique (Tables 3 and 4). Results obtained by CWT
tested in different orders and with different scaling parameters for analysis of the dosage forms were statistically compared to
and the best results, regarding linearity and recovery% in those obtained by the reported HPLC methods [51,63]. The results
laboratory-prepared mixtures, were obtained with 1st derivative showed no significant differences between the proposed method
calculated using morl family and scaling = 200 (Fig. 4) at and the reported ones as presented in Table 5. Validation was done
278.3 nm. Different wavelengths were detected as possible according to ICH recommendations and the validation parameters
zero-crossing points at 255.3, 265.7, 278.3, 292.2 and 306.3 nm. are shown in Table 6.
The points at 255.3 and 265.7 nm were not real zero-crossing
points and the last two wavelengths showed bad linearity and sen- 4.3. CWT versus ND
sitivity compared to 278.3 nm. HCT could be determined with the
same parameters at 283.6 nm. ND is the simplest method of derivative calculation. The major
problem is the choice of Dk. Wide Dk values can be used to sup-
4.2. Ternary mixture press noise and increase SNR, but they degrade resolution as well.
Finding the optimum Dk to compromise between the high resolu-
For analysis of this mixture using ND algorithm, the 1st order tion and good signal to noise ratio (S/N) is considered a challenge.
derivative spectra were obtained where a zero-crossing point Also the values of S/N obtained by ND are very small, so they are
(386.8 nm) was detected for AML determination (Fig. 5a). But in usually multiplied by scaling factor to be considerable.
the same order spectra, ALI and HCT were not resolved, so the The 1st order derivative fails to resolve severely overlapped
2nd order derivatives were calculated and a zero-crossing point spectra, so we use higher order derivatives. This approach can help
(332.1 nm) was noticed for determination of HCT (Fig. 5b). For in case the components to be determined are present in a reason-
determination of ALI, the 3rd and 4th order derivatives were tried. able ratio and have comparable absorptivities. In the case of BIS in
While no zero-crossing points were observed in the 3rd order spec- its binary mixture with HCT, the zero order spectrum of BIS is com-
tra (Fig. 5c), the 4th order showed a zero-crossing point (287.5 nm) pletely enclosed with that of HCT showing very similar features
that was used for ALI determination (Fig. 5d). and low absorptivity, so 1st order derivative failed to present a
950 E.S. Elzanfaly et al. / Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy 151 (2015) 945–955

a b

HCT BIS

c
d

BIS

BIS

Fig. 3. Different order absorption spectra of 40 lg/mL BIS (–) and 1–22 lg/mL HCT (- - -) using ND method showing zero-crossing points for determination of BIS and HCT. (a)
1st order (b) 2nd order (c) 3rd order (d) 4th order.

Table 1
Determination of BIS and HCT in laboratory prepared mixtures by ND and CWT methods.

Concentration (lg/mL) Recovery%a


BIS HCT
BIS HCT ND (3rd order) ND (4th order) CWT ND (1st order) CWT
b b
4 10 54.7 115.1 99.5 101.8 101.3
6 6 89.9b 103.9b 102.4 101.8 99.4
6 12 65.7b 107.8b 98.4 101.1 100.6
6 15 64.6b 109.7b 100.2 100.6 100.2
6 18 53.7b 92.2b 101.3 100.2 99.1
12 6 103.0 100.4 100.6 101.8 100.7
12 15 76.0b 92.6b 98.1 101.3 99.8
15 6 102.9 100.5 101.1 102.5 99.9
15 12 85.4b 99.7 99.0 101.9 100.2
15 15 77.3b 94.3b 98.5 100.9 99.3
18 6 101.6 97.3 101.6 102 99.0
Mean ± SD 102.5 ± 0.8 99.5 ± 1.5 100.1 ± 1.5 101.4 ± 0.7 100.0 ± 0.7
a
Average of three determinations.
b
Excluded according to rejection rule [67].

zero-crossing point for its determination. The 2nd, 3rd and 4th order In contrast to ND, CWT could resolve BIS from HCT in different
derivatives showed zero-crossing points for BIS determination, the ratios by 1st order derivative using morl family. Using CWT algo-
recoveries in laboratory prepared mixtures indicated that BIS can- rithm offers the flexibility of using several wavelet families, which
not be determined in some mixtures contained higher levels of results in transforming the spectra into 1st order derivative in dif-
HCT. This may be attributed to the poor absorptivity of BIS, at ferent shapes. This fact increases the possibility of finding
the wavelengths where zero-crossing points were observed, in zero-crossing points in the 1st order derivatives for compounds
addition to the fact that higher order derivatives diminish the sig- with very similar spectra. The CWT algorithm also has the advan-
nal to noise ratio (S/N). tage of signal amplification compared to ND [66]. In addition to
E.S. Elzanfaly et al. / Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy 151 (2015) 945–955 951

Amplitude

HCT

BIS

Wavelength (nm)

st
Fig. 4. 1 order absorption spectra of 40 lg/mL BIS (—) and 1–22 lg/mL HCT (- - -) using CWT method (morl) showing zero-crossing points for BIS and HCT determination.

a b
AML
HCT

c d

ALI

Fig. 5. Different order absorption spectra of 35 lg/mL AML (–), 250 lg/mL ALI (....) and 25 lg/mL HCT (- - -) using ND method showing zero-crossing points for determination
of AML, ALI and HCT. (a) 1st order (b) 2nd order (c) 3rd order (d) 4th order.

the fact that calculating higher order derivatives is not necessary as zero-crossing points for determination of BIS was not observed in
the 1st order derivative in CWT is enough to get zero-crossing ND 1st order derivative, in contrast to the 1st order morl wavelet.
points. Thus CWT maintains high S/N in contrast to ND calculation, The higher S/N of CWT spectra allowed the determination of BIS
which is the reason why CWT showed higher sensitivity as shown in mixtures where HCT is found in higher ratios, while 2nd, 3rd
from LOD and LOQ values in Table 6. These facts can explain why and 4th order derivatives obtained by ND failed.
952 E.S. Elzanfaly et al. / Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy 151 (2015) 945–955

AML

ALI

HCT

Fig. 6. First order absorption spectra of 35 lg/mL AML (–), 250 lg/mL ALI (....) and 25 lg/mL HCT (- - -), using CWT method showing zero-crossing point for determination of
AML, ALI and HCT. (a) gaus-1 (b) gaus-8 (c) coif-3.

The severe overlap between the spectra of AML, ALI and HCT in ALI and HCT show no interference in their original spectra. Upon
the region 200–350 nm prevented the resolution of the drugs in 1st using CWT families, gaus-1, gaus-8 and coif-3, the 1st order deriva-
order derivative in this region. So ALI and HCT were resolved only tives were enough to resolve the three drugs.
in the 4th and 2nd order derivatives, respectively. AML was excep- The advantages of CWT over ND in calculating derivatives can
tion being determined in 1st order derivative at 386.8 nm where be summarized in the following:
E.S. Elzanfaly et al. / Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy 151 (2015) 945–955 953

Table 2
Determination of AML, ALI and HCT in their ternary mixture by ND and CWT methods.

Concentration (lg/mL) Recovery%a


AML ALI HCT
AML ALI HCT ND CWT ND CWT ND CWT
1 30 2.5 79.6b 102.6 99.0 98.5 102.2 99.8
2 60 5 101.0 102.0 100.0 99.2 98.5 100.9
4 120 10 101.5 102.3 99.6 99.2 99.6 101.3
5 30 25 97.6 102.3 101.7 100.0 100.6 100.6
6 180 15 98.3 102.2 99.1 100.9 98.5 101.2
12 30 5 100.2 101.8 102.0 98.9 100.7 98.1
15 15 15 100.5 100.9 94.8b 105.0b 101.0 99.4
20 20 20 99.9 100.4 95.0b 101.1 98.0 99.7
35 30 5 98.5 100.0 102.0 98.1 99.3 100.5
35 70 5 99.1 100.4 99.3 99.6 97.0 99.3
Mean ± SD 99.6 ± 1.3 101.5 ± 0.9 100.3 ± 1.3 99.5 ± 1.0 99.5 ± 1.6 100.1 ± 1.0
a
Average of three determinations.
b
Excluded according to rejection rule [67].

Table 3
Determination of BIS and HCT in ConcorÒ Plus tablets by the proposed CWT and the reported HPLC method [51] with application of standard addition technique.

Product Drug CWT Reported methoda Standard addition CWT


Takenb (lg/mL) Added (lg/mL) Found (lg/mL) Recovery%c
ConcorÒ Plus 5/12.5 BIS 99.1 ± 0.4 99.4 ± 0.9 6 5 5.04 100.8
6 5.96 99.3
7 7.06 100.9
Mean ± SD 100.3 ± 0.9
HCT 99.7 ± 0.4 100.2 ± 0.8 10 8 7.97 99.6
10 9.94 99.4
12 12.11 100.9
Mean ± SD 100.0 ± 0.8
ConcorÒ Plus 10/25 BIS 100.5 ± 0.7 100.7 ± 0.4 6 5 4.96 99.2
6 6.03 100.5
7 6.97 99.6
Mean ± SD 99.8 ± 0.7
HCT 99.5 ± 0.4 99.9 ± 0.9 10 8 7.95 99.4
10 9.98 99.8
12 12.03 100.3
Mean ± SD 99.8 ± 0.5
a
RP-HPLC using C18 and 0.1 M potassium dihydrogen phosphate buffer and acetonitrile (70:30, v/v) as mobile phase. The UV detection was carried out at 228 nm at a flow
rate of 1.0 mL/min.
b
Standard addition were done on two different dilutions of the dosage form for the drugs to be within the linearity range.
c
Average of three determinations.

Table 4
Determination of AML, ALI and HCT in AmturnideÒ tablets by the proposed CWT and the reported HPLC method [63] with application of standard addition technique.

Product Drug CWT Reported methoda Standard addition CWT


Takenb (lg/mL) Added (lg/mL) Found (lg/mL) Recovery%c
Ò
Amturnide 10/300/25 AML 100.6 ± 1.1 100.4 ± 0.9 6 4 3.98 99.5
6 6.09 101.5
8 8.03 100.4
Mean ± SD 100.5 ± 1.0
ALI 99.9 ± 0.6 100.3 ± 0.8 120 110 109.31 99.4
120 121.55 101.3
130 130.73 100.6
Mean ± SD 100.4 ± 1.0
HCT 100.4 ± 0.7 99.9 ± 0.7 10 5 5.01 100.2
10 10.11 101.1
15 14.88 99.2
Mean ± SD 100.2 ± 0.9
a
RP-HPLC using acetonitrile: methanol: 50 mM phosphate buffer (20:50:30% by volume) adjusted to pH 3 with orthophosphoric acid and flow rate of 1.0 mL/min at
239 nm.
b
Standard addition were done on two different dilutions of the dosage form for the drugs to be within the linearity range.
c
Average of three determinations.
954 E.S. Elzanfaly et al. / Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy 151 (2015) 945–955

Table 5
Statistical comparison for the results obtained by the proposed CWT and the reported HPLC [51,63] methods for the analysis of ConcorÒ Plus and AmturnideÒ tablets.

Value ConcorÒ Plus tablets AmturnideÒ tablets


a
CWT Reported method CWT Reported methodb
BIS HCT BIS HCT AML ALI HCT AML ALI HCT
Mean 99.8 99.6 100.0 100.0 100.6 99.9 100.4 100.4 100.3 99.9
RSD% 0.9 0.4 1.0 0.8 1.1 0.6 0.7 0.9 0.8 0.7
n 6 6 6 6 6 6 6 6 6 6
Variance 0.832 0.140 0.903 0.582 1.329 0.347 0.496 0.755 0.617 0.499
Student’s t testc 0.474 1.116 – – 0.405 0.845 1.228 – – –
(2.228)
F value c 1.085 4.157 – – 1.76 1.777 1.007 – – –
(5.05)
a
RP-HPLC using C18 and 0.1 M potassium dihydrogen phosphate buffer and acetonitrile (70:30, v/v) as mobile phase. The UV detection was carried out at 228 nm at a flow
rate of 1.0 mL/min.
b
RP-HPLC using acetonitrile: methanol: 50 mM phosphate buffer (20:50:30% by volume) adjusted to pH 3with orthophosphoric acid and flow rate of 1.0 mL/min at
239 nm.
c
The values in the parenthesis are the corresponding theoretical values of t and F at P = 0.05.

Table 6
Assay validation sheet of ND and CWT methods for the simultaneous determination of the mixtures under study.

Parameter Binary mixture Ternary mixture


BIS HCT AML ALI HCT
ND (3rd ND (4th CWT ND CWT ND CWT ND CWT ND CWT
order) order)
Accuracya 102.0 ± 0.3 99.7 ± 2.2 100.0 ± 0.6 99.9 ± 1.6 99.8 ± 0.5 100.6 ± 1.2 100.8 ± 1.2 99.9 ± 1.0 100.3 ± 0.9 98.5 ± 0.8 99.7 ± 1.0
(Mean ± SD)
Precision
– Repeatabilityb 1.3 1.2 0.6 0.7 0.6 0.9 0.9 1.0 0.9 0.7 0.8
– Intermediate 1.8 1.6 0.9 1.2 0.8 1.2 1.3 1.2 1.1 1.0 1.1
precisionc
Robustnessd 1.5 1.5 0.8 1.4 0.6 1.1 0.8 1.3 1.3 1.6 1.6
e
LOD (lg/mL) 1.42 1.41 0.57 0.44 0.22 0.57 0.26 4.50 4.45 1.48 0.59
LOQe (lg/mL) 4.75 4.69 1.90 1.48 0.73 1.89 0.88 15.01 14.84 4.93 1.95
Linearity
– Slope 0.0291 0.0086 0.0115 0.1743 0.1062 0.0049 0.1644 0.0113 0.0392 0.0027 0.1507
– Intercept 0.0021 0.0116 0.0014 0.0049 0.0258 0.0011 0.0030 0.3070 0.0070 0.0001 0.0190
– Correlation 0.9998 0.9997 0.9999 0.9998 0.9999 0.9999 1.000 0.9999 0.9998 0.9998 0.9999
coefficient (r)
– Range (lg/mL) 5.0–40.0 5.0–40.0 2.0–40.0 2.0–22.0 1.0–22.0 2.0–35.0 1.0–35.0 15.0– 15.0– 5.0–25.0 2.0–25.0
250.0 250.0
a
The accuracy (n = 3), average of three concentrations of each drug.
b
The intraday (n = 3), average of three concentrations of each drug repeated three times within day.
c
The interday (n = 3), average of three concentrations each drug repeated three times in three days.
d
Robustness (n = 3), average of three concentrations each drug analyzed using 75% and 70% methanol.
e
LOD and LOQ were calculated using the following equations: LOD = 3Sr and LOQ = 10Sr, r is the standard deviation of residuals and S is the slope.

1. Flexibility in calculating 1st order derivatives in different shapes similar spectra without the need of higher order derivative
using different wavelet families, that allows CWT to provide calculation.
zero-crossing points for complex mixtures.
2. Signal amplification provided by the algorithm together with its
flexibility in 1st order derivatives gives CWT high sensitivity References
compared to ND. [1] E. Dymond, Math. Proc. Cambridge Philos. Soc. 22 (1924) 405–408.
[2] V. Hammond, W. Price, J. Opt. Soc. Am. 43 (1953) 924–930.
[3] J. Morrison, J. Chem. Phys. 21 (1953) 1767–1772.
[4] A.T. Giese, C.S. French, Appl. Spectrosc. 9 (1955) 78–96.
5. Conclusion [5] A. Martin, Spectrochim. Acta A 14 (1959) 97–103.
[6] J.R. Morrey, Anal. Chem. 40 (1968) 905–914.
[7] A. Savitzky, M.J. Golay, Anal. Chem. 36 (1964) 1627–1639.
Using CWT offered many advantages over traditional ND. One [8] G. Talsky, Derivative Spectrophotometry: Low and High Order, VCH, California,
advantage is that CWT algorithm maintains high S/N in contrast 1994.
to ND calculation. In addition to the fact that calculating higher [9] N. Funasaki, S. Hada, S. Neya, J. Phys. Chem. B 103 (1999) 169–172.
[10] B. Ebeshi, E. Vaikosen, D. Tuma, Int. J. Biol. Chem. Sci. 7 (2014) 2591–2599.
order derivatives is not necessary in CWT as the 1st order deriva-
[11] Y. Ni, J. Bai, Talanta 44 (1997) 105–109.
tive is enough to get zero-crossing points for severely overlapped [12] A. Molina, M. López-Tenés, C. Serna, J. Electrochem. Soc. 147 (2000) 3429–
spectra. The flexibility of CWT algorithm allowing the use of sev- 3435.
eral wavelet families in different orders, results in transforming [13] M. Blanco, J. Gene, H. Iturriaga, S. Maspoch, J. Riba, Talanta 34 (1987) 987–993.
[14] F. Singleton, G. Collier, J. Appl. Chem. 6 (1956) 495–510.
the spectra into different shapes of 1st order derivatives, which [15] W. Snelleman, T. Rains, K. Yee, H.D. Cook, O. Menis, Anal. Chem. 42 (1970)
increases the possibility of finding zero-crossing points in very 394–398.
E.S. Elzanfaly et al. / Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy 151 (2015) 945–955 955

[16] T.C. O’Haver, Modulation and Derivative Techniques in Luminescence [47] A.D. Gudruman, N. Bibire, G. Tantaru, M. Apostu, M. Vieriu, V. Dorneanu, Rev.
Spectroscopy, Springer, New York, 1976. Chim. 64 (2013) 393–396.
[17] A.Y. El-Sayed, N.A. El-Salem, Anal. Sci. 21 (2005) 595. [48] H.W. Darwish, S.A. Hassan, M.Y. Salem, B.A. El-Zeany, Spectrochim. Acta A 122
[18] C.B. Ojeda, F.S. Rojas, Microchem. J. 106 (2013) 1–16. (2013) 744–750.
[19] M. Korany, A. Wahbi, M. Elsayed, S. Mandour, Farmaco Ed. Prat. 39 (1984) 243. [49] M. Emanual, J. Chromatogr. Sep. Tech. (2012).
[20] M. Korany, A. Wahbi, M. Elsayed, S. Mandour, Anal. Lett. 17 (1984) 1373–1389. [50] H.W. Darwish, S.A. Hassan, M.Y. Salem, B.A. El-Zeany, Int. J. Pharma Bio Sci. 4
[21] M.A. Korany, A.A.S. El-Din, N.A. Abael-salam, Anal. Lett. 17 (1984) 483–495. (2013) 345–356.
[22] F. Salinas, J. Nevado, A. Mansilla, Talanta 37 (1990) 347–351. [51] S.J. Joshi, P.A. Karbhari, S.I. Bhoir, K. Bindu, C. Das, J. Pharm. Biomed. Anal. 52
[23] J. Berzas, C. Guiberteau, F. Salinas, Talanta 39 (1992) 547. (2010) 362–371.
[24] E. Dinc, F. Onur, Anal. Chim. Acta 359 (1998) 93–106. [52] A. Singh, T. Sivakumar, D. Jain, U. Mishra, A. Yadav, T. Sharma, Int. J. Pharm.
[25] E. Dinç, D. Baleanu, Spectrochim. Acta A 63 (2006) 631–638. Res. Dev. 3 (2011) 11–22.
[26] I. Daubechies, Ten Lectures on Wavelets, SIAM, Pennsylvania, 1992. [53] J. Shi, Y. Chen, Y.P. Long, Cent. South Pharm. 11 (2012) 000.
[27] A.K.M. Leung, F.T. Chau, J.B. Gao, Anal. Chem. 70 (1998) 5222–5229. [54] S. Kurbanoglu, P.R. San Miguel, B. Uslu, S.A. Ozkan, Chromatographia (2013) 1–
[28] X.D. Dai, B. Joseph, R. Motard, Introduction to Wavelet Transform and Time- 7.
frequency Analysis, Springer, New York, 1994. [55] P.N. Bhoya, E.M. Patelia, J. Chromatogr. Sep. Tech. 4 (2013) 163–166.
[29] R.R. Coifman, G. Weiss, Bull. Am. Math. Soc. 83 (1977) 569–645. [56] Y. Lacourcière, S. Taddei, G. Konis, H. Fang, T. Severin, J. Zhang, J. Hypertens. 30
[30] A. Calderón, A. Zygmund, Stud. Math. 46 (1973) 297–299. (2012) 2047–2055.
[31] A. Grossmann, J. Morlet, SIAM J. Math. Anal. 15 (1984) 723–736. [57] F. Belal, M. Walash, N. El-Enany, S. Zayed, J. Chromatogr. B 933 (2013) 24–29.
[32] S.G. Mallat, IEEE Trans. Acoust. 37 (1989) 2091–2110. [58] H.W. Darwish, S.A. Hassan, M.Y. Salem, B.A. El-Zeany, Spectrochim. Acta A 113
[33] I. Daubechies, IEEE Trans. Inf. Theory 36 (1990) 961–1005. (2013) 215–223.
[34] I. Daubechies, SIAM J. Math. Anal. 24 (1993) 499–519. [59] H.W. Darwish, S.A. Hassan, M.Y. Salem, B.A. El-Zeany, Spectrochim. Acta A 104
[35] D.L. Donoho, I.M. Johnstone, C.R. Acad. Sci. I Math. 319 (1994) 1317–1322. (2013) 70–76.
[36] V. Barclay, R. Bonner, I. Hamilton, Anal. Chem. 69 (1997) 78–90. [60] M.I. Ezzeldien, E. Shokry, M.A. Fouad, R.I. Elbagary, Int. J. Adv. Chem. 1 (2013)
[37] X. Zou, J. Mo, Anal. Chim. Acta 340 (1997) 115–121. 13–20.
[38] M. Bos, E. Hoogendam, Anal. Chim. Acta 267 (1992) 73–80. [61] H.W. Darwish, S.A. Hassan, M.Y. Salem, B.A. El-Zeany, Int. J. Spectrosc. 2013
[39] H.W. Tan, S.D. Brown, J. Chemom. 16 (2002) 228–240. (2013) 8.
[40] U. Depczynski, K. Jetter, K. Molt, A. Niemöller, Chemom. Intell. Lab. Syst. 47 [62] H.W. Darwish, S.A. Hassan, M.Y. Salem, B.A. El-Zeany, Int. J. Pharma Bio Sci. 4
(1999) 179–187. (2013) 230–243.
[41] T. O’haver, G. Green, Anal. Chem. 48 (1976) 312–318. [63] G.K. Swamy, P. Sravanthi, M.L. Surekha, J.K.J. Rao, Int. J. ChemTech Res. 4
[42] E. Dinç, D. Baleanu, J. Pharm. Biomed. Anal. 31 (2003) 969–978. (2012) 1666–1673.
[43] E. Dinç, A. Özdemir, D. Baleanu, Talanta 65 (2005) 36–47. [64] M.M. Salim, W.M. Ebeid, N. El-Enany, F. Belal, M. Walash, G. Patonay, J. Sep. Sci.
[44] E. Dinç, M. Kanbur, D. Baleanu, Die Pharm. 60 (2005) 892–896. 37 (2014) 1206–1213.
[45] E.S. Elzanfaly, S.A. Hassan, M.Y. Salem, B.A. El-Zeany, Spectrochim. Acta A 140 [65] F.T. Chau, Y.Z. Liang, J. Gao, X.G. Shao, Chemometrics: From Basics to Wavelet
(2015) 334–343. Transform, John Wiley & Sons, New Jersey, 2004.
[46] W.H. Frishman, B.S. Bryzinski, L.R. Coulson, V.L. DeQuattro, N.D. Vlachakis, W.J. [66] E. Dinç, M. Kanbur, D. Baleanu, Spectrochim. Acta A 68 (2007) 225–230.
Mroczek, G. Dukart, J.D. Goldberg, D. Alemayehu, K. Koury, Arch. Internal Med. [67] F.J. Anscombe, Technometrics 2 (1960) 123–146.
154 (1994) 1461–1468.
Contributing Editor
ABDULLAH A. AL-BADR

Founding Editor
KLAUS FLOREY

You might also like