You are on page 1of 7

See

discussions, stats, and author profiles for this publication at: https://www.researchgate.net/publication/231067943

Calibration of high-heat-flux sensors in a solar


furnace

Article in Metrologia · November 2006


DOI: 10.1088/0026-1394/43/6/003

CITATIONS READS

12 215

7 authors, including:

Jesús Ballestrín José Rodríguez


Plataforma Solar de Almería Plataforma Solar de Almería
57 PUBLICATIONS 302 CITATIONS 99 PUBLICATIONS 578 CITATIONS

SEE PROFILE SEE PROFILE

Inmaculada Cañadas J. Barbero


Centro Investigaciones Energéticas, Medioa… Universidad de Almería
95 PUBLICATIONS 337 CITATIONS 43 PUBLICATIONS 383 CITATIONS

SEE PROFILE SEE PROFILE

Some of the authors of this publication are also working on these related projects:

Solar furnace: preparation of Ti-TiB2 composites and thermal shock behaviour of TiB2 coating on
steel View project

SFERA: Solar Facilities for the European Research Area View project

All content following this page was uploaded by J. Barbero on 02 June 2014.

The user has requested enhancement of the downloaded file.


INSTITUTE OF PHYSICS PUBLISHING METROLOGIA
Metrologia 43 (2006) 495–500 doi:10.1088/0026-1394/43/6/003

Calibration of high-heat-flux sensors in a


solar furnace
J Ballestrı́n1 , M Rodrı́guez-Alonso1 , J Rodrı́guez1 , I Cañadas1 ,
F J Barbero2 , L W Langley3 and A Barnes3
1
CIEMAT—Plataforma Solar de Almerı́a, Aptdo. 22, E-04200 Tabernas, Almerı́a, Spain
2
Departamento de Fı́sica Aplicada, Universidad de Almerı́a, Spain
3
Vatell Corporation, PO Box 66, Christiansburg, VA 24073, USA

Received 26 October 2005


Published 6 November 2006
Online at stacks.iop.org/Met/43/495

Abstract
The most common sensors used for the measurement of high solar
irradiance are the Gardon gauges, which are usually calibrated using a black
body at a certain temperature as the radiant source. This calibration
procedure is assumed to produce a systematic error when solar irradiance
measurements are taken using these sensors. This paper demonstrates a
calorimetric method for calibrating these high-heat-flux gauges in a solar
furnace. This procedure has enabled these sensors to be calibrated under
concentrated solar radiation at higher irradiances under non-laboratory
conditions in the CIEMAT solar furnace at the Plataforma Solar de Almerı́a.
Working at higher irradiances has allowed the uncertainty in the calibration
constant of these sensors to be reduced. This work experimentally confirms
the predicted systematic errors committed when measuring high
solar irradiances using Gardon sensors calibrated with a black body.
(Some figures in this article are in colour only in the electronic version)

1. Introduction 2. Heat-flux sensors at Plataforma Solar de Almerı́a


Measurement of high irradiances obtained by solar At the present time, the sensors used to measure high
concentrating systems is required in a growing variety of solar irradiance in the various Plataforma Solar de Almerı́a
research fields and technical applications, such as photovoltaic (PSA) facilities are Gardon-type gauges. Their transducer
and solar thermal energy. At the present time, the sen- is a differential thermocouple that measures the temperature
sors used at the Plataforma Solar de Almerı́a (PSA) [1] for difference between the centre and the circumference of a thin
this purpose are the circular foil or Gardon-type [2]. Their foil disc, the diameter and thickness of which vary with the
small size and short response time (<1 s) make them versatile range of irradiance [5]. The disc is bonded to a circular
devices, which may be used for many applications involving opening in a 2.54 cm long cylindrical heat sink with a 0.79 cm
the concentration of solar radiation without interfering with to 2.54 cm front-face diameter. The standard foil is made of
test procedures. At present, there is no international stan- constantan and the heat sink is copper. These materials produce
dard providing traceable high solar irradiance measurements an EMF output which is directly proportional to the heat flux
for calibrating these heat flux sensors. Several manufacturers absorbed. The usual signal is 0 mV to 10 mV regardless of the
and laboratories have made an effort to calibrate them [3, 4], heat flux range, which means a constantan–copper temperature
using black bodies and pyrometric temperature measurement difference of up to 214 ◦ C.
traceable to the National Institute of Standards and Tech- One of the few suppliers of Gardon-type sensors is the
nology (NIST) to determine heat flux, but these calibrations Vatell Corporation. The exposed face of the sensors is sprayed
are not valid in the solar context because the spectral radi- with a high-absorptance black coating to achieve absorption
ance of a black body is significantly different from the so- high enough to generate a significant signal. Zynolyte® [6]
lar spectral distribution. This paper describes a calorimet- is the preferred coating at Vatell because it creates a rough
ric method for calibrating high-heat-flux sensors in a solar surface that improves heat absorption; however, when heat flux
furnace. is over 3500 kW m−2 colloidal graphite (Electrodag® 154) is a

0026-1394/06/060495+06$30.00 © 2006 BIPM and IOP Publishing Ltd Printed in the UK 495
J Ballestrı́n et al

Figure 1. Spectral irradiance from a black body at 850 ◦ C and solar


spectral irradiance with AM2, both normalized to a solar constant of
1353 W m−2 . Reflectance of a typical mirror used to concentrate
solar radiation. Absorptance of the two coatings (Zynolyte® and
colloidal graphite) used to paint heat flux sensors.
Figure 2. Solar furnace diagram.
better option because the centre of the foil gets so hot that the
Zynolyte® would burn off. with these gauges has been estimated from all this information
The current procedure for calibrating these sensors at in a previous paper [6]. If the sensor coating is Zynolyte® , the
Vatell is based on a dual-cavity black body, which is basically sensor overestimates the solar irradiance by 3.6%. However,
an electrically heated double-ended cylindrical graphite tube if the coating is colloidal graphite, the overestimation of solar
with a centre partition that allows its internal temperature irradiance is 27.9% [6].
to be measured with an NIST-traceable pyrometer while
the reference transducer is being calibrated. When the 3. Thermal balance calibration in a solar furnace
selected temperature of 850 ◦ C has stabilized, the transducer
is quickly inserted into the cavity opposite the pyrometer. The The validity of the thermal balance technique for calibrating
Stefan–Boltzman law supplies the resulting irradiance value heat flux sensors was presented in a previous paper [8], which
of about 90 kW m−2 on the front face of the sensor, assuming reported the calorimetric method applied in the laboratory.
hemispherical irradiation. Calibration of the reference heat- The main limitation was that the maximum irradiance level
flux sensor is performed with this single-point technique. was about 100 kW m−2 . This constraint has been overcome
This calibration is transferred to the commercial sensors by by working in the Plataforma Solar de Almerı́a (PSA) solar
comparison in a calibration furnace with a graphite plate that furnace.
radiates evenly and symmetrically when an electrical current
passes through it. The calibration constant obtained with this
3.1. PSA solar furnace
method translates voltage to irradiance on the front face of the
sensor. The manufacturer states that the accuracy of gauges Solar furnaces reach one of the highest concentration levels
calibrated in this way is within ±3% with a repeatability of obtainable with a solar concentrating system, having attained
±1% [6]. concentrations of over 10 000×. A solar furnace essentially
These heat flux sensors have been used since 1953 [5] consists of a flat solar-tracking heliostat, a parabolic collector
in different fields such as plasma diagnostics and combustion mirror, an attenuator or shutter and the test zone located in
research, which have to deal with dissimilar spectral radiances, the concentrator focus (figure 2). The flat collecting mirror,
and, therefore, compatibility with solar radiation is not given, or heliostat, reflects the parallel horizontal solar beams onto
as the spectral absorptance of these two coatings (Zynolyte® the parabolic dish, which in turn reflects them onto its focus
and colloidal graphite) is not constant over the whole where the test area is set up. The amount of incident light is
electromagnetic spectral range (figure 1). Unfortunately, at regulated by the shutter located between the concentrator and
850 ◦ C, the spectral radiance of a black body is significantly the heliostat. Below the focus there is a test table movable in
different from the solar spectral distribution and a systematic three directions (East–West, North–South, up and down) that
error arises due to the difference in power absorbed by the places the test samples in the focus with high precision.
coatings under these two dissimilar electromagnetic spectral The PSA solar furnace currently has one 140 m2 heliostat.
distributions [6, 7]. Figure 1 shows the spectral irradiance The reflective surface of the heliostat is made up of 32 non-
from a black body at 850 ◦ C and solar spectral irradiance with concentrating flat facets with 90% reflectivity that continually
air mass ratio 2 (AM2), both normalized to a solar constant track the solar disc and reflect its parallel horizontal beams onto
of 1353 W m−2 , the reflectance of a typical mirror used to the concentrator. The concentrator dish is the main component
concentrate solar radiation and the absorptance of the two of the solar furnace (figure 2). It concentrates the incident
coatings (Zynolyte® and colloidal graphite) used on heat flux light from the heliostat, multiplying the irradiance in the focal
sensors. The systematic error from measuring solar irradiance zone. Its optical properties especially affect the irradiance

496 Metrologia, 43 (2006) 495–500


Calibration of high-heat-flux sensors

following equation can be employed with a negligible error:

Q̇ = V̇ ρ(Tav )cp (Tav )T , (2)

where T is the temperature increment in the coolant and Tav


the average temperature between Tin and Tout .
The heat flux absorbed per unit area on the front face of
the sensor is
Q̇ − Q̇Ins
q= , (3)
A
where A is the front area of the calorimeter and Q̇Ins the
power from the insulation to the calorimeter. Assuming
homogeneous hemispherical irradiation, it is possible to obtain
the solar irradiance on the front face of the sensor from
q
E= , (4)
α
Figure 3. Gardon sensor and calibration set-up. where α is the hemispherical solar absorptance of the coating.
Some samples of mirror and metal plates painted with
Zynolyte® and colloidal graphite were analysed in detail to
distribution at the focus. It is composed of 89 spherical
obtain the absorptance of this coating under concentrated solar
facets with a total surface of 98.5 m2 , 92% reflectivity and
radiation from [6]
focal distance of 7.45 m. The parabolic surface is achieved
with spherically curved facets, distributed along five radii with  2500
Gbλ ρλ αλ dλ
different curvatures depending on their distance from the focus. α = 300 2500 , (5)
The attenuator consists of a set of horizontal louvres that rotate 300 Gbλ ρλ dλ
on their axes, regulating the amount of sunlight incident on the
where λ is the wavelength, 300 nm to 2500 nm is the
concentrator. The total energy on the focus is proportional to
approximate solar spectral band, Gbλ is the direct solar
the radiation that passes through the attenuator. It is composed spectral irradiance, ρλ is the spectral specular reflectance of
of 30 louvres arranged in two columns of 15. In closed position the mirror and αλ is the spectral hemispherical absorptance
the louvres form a 55◦ angle with the horizontal and 0◦ when of the coating. The solar hemispherical absorptances of
open. Zynolyte® and colloidal graphite (Electrodag® 154) estimated
The concentrator and distribution of the irradiance in the using equation (5) are 95.4% and 84.7%, respectively.
focus characterize a solar furnace. The characteristics of the
focus when 100% open and with direct solar irradiance of
3.3. Experimental set-up and measurement equipment
1000 W m−2 are an irradiance peak of 3000 kW m−2 , total
power of 58 kW and focal diameter of 23 cm [10]. The sensors are placed successively in a stainless steel housing
in the test area of the solar furnace (figures 2 and 3). An
3.2. Gardon sensor details and basis for thermal balance automated low-conductivity shutter (figure 3) is placed in
front of the sensor to allow irradiance steps over the gauge.
Two 15.9 mm front-face diameter Vatell Gardon sensors The different apertures of shutter 1 (figure 2) make different
(#7915, #7918) were used for this experiment. The irradiance levels possible. The stainless steel enclosure is
ranges of the sensors are between 1200 kW m−2 and filled with low conductivity insulation and all gauge surfaces
5000 kW m−2 , respectively, the coatings are Zynolyte® are insulated except the entire front face which is exposed
and colloidal graphite, and the calibration constants to the radiation (figures 3 and 4). A Type-T thermocouple
supplied by the manufacturer are 113.59 kW m−2 mV−1 and measures the temperature of the insulation and a differential
541.74 kW m−2 mV−1 , respectively. Type-T thermocouple measures the temperature increment
Approximately 95% of the front face of the Gardon sensor between inlet and outlet water. Two PT-100 sensors also
is copper (figure 3). The excellent thermal conductivity of this measure both temperatures independently. The differential
metal allows this sensor to be used as a calorimeter because its thermocouple and the two PT-100 sensors are calibrated using
thermal time constant is about one second. The principle of methods specific to each. An uncertainty of less than 2% in the
calorimeters is based on the measurement of the power input measurement of the temperature difference is obtained after
into a cooling medium according to inclusion of the data acquisition system (DAS) error. The
 higher the temperature increment, the lower its uncertainty. An
Tout
offset adjustment is performed before starting measurement to
Q̇ = V̇ ρ(T )cp (T ) dT , (1)
Tin eliminate possible thermal input from water friction [9].
An electromagnetic flow meter, expressly selected for this
where V̇ , ρ and cp are the volume flow, the density and the experiment, measures the flow volume with an uncertainty
specific heat of the cooling medium, respectively. Tin and Tout of 0.25% in a range 1 litre min−1 to 20 litre min−1 , rising to
are the inlet and outlet temperatures of the coolant. When the 0.75% after the DAS error has been included. The higher
temperature difference is only a few degrees centigrade, the the flow volume, the higher the uncertainty. The uncertainty

Metrologia, 43 (2006) 495–500 497


J Ballestrı́n et al

3.5. Heat flux conduction from the insulation to the


calorimeter
The heat flux transfer between the insulation and the
calorimeter is calculated assuming heat conduction between
two concentric cylinders in a stationary state. The inner and
outer radii are the radius of the calorimeter and the distance
from the insulation thermocouple to the centre of the gauge,
respectively. The average cooling water temperature, obtained
from inlet temperature Tin and outlet temperature Tout , is
gauge considered to be the sensible measurement of the insulation
contact temperature. The thermal emission loss of the outer
enclosure is neglected. With the 12 mm insulation thickness,
the heat transfer coefficient of the insulation, 0.1 W m−1 K−1 ,
and the insulation temperature, TIns , the estimation of this
Figure 4. Gardon sensor under concentrated solar radiation. reduced power is continuously included in the thermal balance
obtained by equations (2)–(4) according to
 
Tin + Tout
litre
Q̇Ins = 1.72 × 10−2 TIns − . (6)
litre
2
litre
litre During all the tests, the average cooling water temperature
(<20 ◦ C) is lower than the temperature of the insulation
litre

(<30 ◦ C).

3.6. Calibration constant, linearity, uncertainty and


repeatability
The cycle described above (section 3.4) is repeated at different
shutter 1 apertures, obtaining 13–19 irradiance values on
the front faces of the Gardon sensors and the corresponding
voltages. Operating in this manner, the solar irradiance
achieved in this experimental set-up is limited to a maximum
of about 1300 kW m−2 . With this information, a linear fit
of the data is taken (figure 7) as the sensor calibration.
Figure 5. Calculated measurement uncertainty of the absorbed heat This hypothesis is confirmed by the correlation coefficient
flux density by the Gardon sensor at different cooling water flows.
R (tables 1 and 2). This procedure has been systematically
Laboratory and solar furnace conditions.
repeated for both sensors at different water flow rates (tables 1
and 2) with less than 1% uncertainty for the calibration constant
in measurement of the heat flux density absorbed can be B, (figure 7) and a repeatability of 0.6%. This repeatability at
calculated with equations (2) and (3) and is given in figure 5. different cooling conditions confirms the reduced contribution
This figure shows how, by working at higher irradiances, it is of convective and thermal effects on the front face of the
possible to reduce the uncertainty in this measurement. A 16- sensor to the thermal balance. A significant loss of heat to the
bit resolution DAS card connected to a PC receives the signal. environment (convection, thermal) would significantly modify
A DASYLab® program handles the data transfer, visualization, constant calibration under different cooling conditions [11].
on-line calculations and storage. Once the thermal balance calibration constant (B) of the
sensor has been obtained (tables 1 and 2), the deviation of
the corresponding black body calibration constant (BCC) is
estimated from
3.4. PSA solar furnace operating procedure BCC − B
SE = 100 (7)
B
The sensor is placed in the solar furnace test area as described
and the uncertainty of this deviation
above (figures 2–4) and the water flow is selected. Shutter 1 is
then opened to a predefined position and shutter 2 is removed ∂SE ∂SE
SE = BCC + B
for about five seconds (figure 6), during which time a stable ∂BCC ∂B 
level of irradiance is achieved. This level is simultaneously 1 BCC
= 100 BCC − B , (8)
obtained by thermal balance (equations (2)–(4)) and by the B B2
black body calibration associated with the signal voltage of the where B and BCC are the uncertainties of the two
transducer. After this short time, shutter 2 goes back in front calibration constants, respectively (tables 1 and 2). The
of the sensor and shutter 1 is closed. The voltage generated by predicted systematic errors in measuring high solar irradiances
the transducer is recorded throughout the process. using Gardon sensors calibrated with a black body of 3.6% and

498 Metrologia, 43 (2006) 495–500


Calibration of high-heat-flux sensors

Figure 6. Estimated solar irradiance by the thermal balance and by black body calibration when the Gardon sensors are under concentrated
solar radiation.

litre
litre

Figure 7. Calibration of both Gardon sensors by the thermal balance method.

27.9%, respectively, are in agreement with these experimental heat flux can modify the contribution of these factors to the
results (tables 1 and 2). thermal balance. As this contribution to the balance may
It should also be kept in mind that these heat flux sensors be negative (losses) or positive (gains) during calibration and
are calibrated according to their hemispherical absorptance. subsequent use of the gauge, the contributions to the calibration
Any alteration in this would mean a smaller solid angle of process are difficult to estimate. However, such contributions
vision, as in the solar furnace case, and possible consequent should be included as additional random uncertainty in the
change in absorptance, which would have to be considered for calibration. Detailed thermal simulation of the gauge in
every specific application of the gauges [7, 11]. Changes in a central receiver context has been performed in previous
absorptance at different temperatures would be another factor work, showing that the order of magnitude of the convective
to consider [11]. However, the good agreement between theory and thermal contributions to the uncertainty of the constant
and experimental results leads us to believe that the influence calibration is about 1% [13].
of these factors is minor and negligible.
4. Conclusions

3.7. Thermal and convective losses and gains This paper reports on a calorimetric method for calibrating
high-heat-flux sensors in a solar furnace. This technique
Thermal and convective effects on the front face of the sensor calibrates for the absorbed radiation. In order to get the
are difficult to analyse experimentally and during mid-range incident radiation it is necessary to know the absorptivity of
calibrations are usually assumed to be negligible [12]. Any the surface, which usually is a property not very well known.
change in cooling conditions, ambient temperature or incident This procedure has enabled these sensors to be calibrated

Metrologia, 43 (2006) 495–500 499


J Ballestrı́n et al

Table 1. Calibration of a the Gardon sensor (#7915, Zynolyte) by the thermal balance method. Convective and thermal effects are not
considered.
Water flow/ Inlet temperature/ B/ B/
litre min−1 ◦
C kW m−2 mV−1 kW m−2 mV−1 100 × Uncertainty R 100 × SE 100 × SE
1.5 29.5 111.4 0.4 0.4 0.999 91 6 3
2 29.2 112.6 0.5 0.4 0.999 94 5 3
2.5 29.4 113.6 0.6 0.5 0.999 95 4 3

Table 2. Calibration of a Gardon sensor (# 7918, colloidal graphite) by the thermal balance method. Convective and thermal effects are not
considered.
Water flow/ Inlet temperature/ B/ B/
litre min−1 ◦
C kW m−2 mV−1 kW m−2 mV−1 100 × Uncertainty R 100 × SE 100 × SE
1.5 22.1 416 2 0.5 0.999 89 30 4
2 22.0 416 2 0.5 0.999 92 30 4
2.5 21.8 414 3 0.7 0.999 92 30 4

under concentrated solar radiation at higher irradiances under [3] Murthy A V, Tsai B K and Saunders R D 1998 High-heat-flux
non-laboratory conditions in the CIEMAT solar furnace at the sensor calibration using black-body radiation Metrologia
35 501–4
Plataforma Solar de Almerı́a. The thermal balance calibration
[4] Murthy A V, Tsai B K and Saunders R D 2000 Radiative
technique in the laboratory is limited to approximately calibration of heat flux sensors at NIST: facilities and
100 kW m−2 . Working at higher irradiances has allowed the techniques J. Res. Natl Inst. Stand. Technol. 105 293–305
uncertainty of the calibration constant of these sensors to [5] Diller T E 1999 Heat flux The Measurement, Instrumentation
be reduced. This work has experimentally confirmed the and Sensors Handbook ed J G Webster (Boca Raton, FL:
CRC Press) chapter 34, pp 1–15
predicted systematic errors in measuring high solar irradiances
[6] Ballestrı́n J et al 2003 Systematic error in the measurement of
using Gardon sensors calibrated with a black body. very high solar irradiance Sol. Energy Mater. Sol. Cells
80 375–81
[7] Ulmer S et al 2004 Calibration corrections of solar tower flux
density measurements Energy 29 925–33
Acknowledgments [8] Ballestrı́n J et al 2004 High-heat-flux sensor calibration using
calorimetry Metrologia 41 314–18
This work received financial support from the Spanish Ministry [9] Groer U and Neumann A 1999 Development and test of a high
of Science and Technology under the MEPSOCON project flux calorimeter at DLR Cologne Proc. 9th Int. Symp. Solar
(Ref DPI2003-03788). Thermal Concentrating Technologies, J. Phys. IV 9 643–8
(Les Ulis, France: EDP Sciences)
[10] Neumann A 1994 Flux densities in the focal region of the PSA
solar furnace SolarPACES Technical Report No III-2/94
References [11] Siegel R and Howell J R 1972 Thermal Radiation Heat
Transfer (New York: McGraw-Hill)
[1] Ballestrı́n J and Monterreal R 2004 Hybrid heat flux [12] Bryant R et al 2003 Radiative heat flux measurement
measurement system for solar central receiver evaluation uncertainty Fire Mater. 27 209–22
Energy 29 915–24 [13] Marcos M J et al 2005 Outdoor uncertainty on heat flux
[2] Gardon R 1953 An instrument for the direct measurement of measurement Solar World Congress (ISES, International
intense thermal radiation Rev. Sci. Instrum. 24 366–70 Energy Society) (Orlando, USA, 6–12 August, 2005)

500 Metrologia, 43 (2006) 495–500

View publication stats

You might also like