You are on page 1of 7

O R I G I N A LA R T I C L E

Transparent Film Intravenous Line Dressing


Incorporating a Chlorhexidine Gluconate
Gel Pad: A Clinical Staff Evaluation
Tarja J. Karpanen, PhD, RGN
Department of Clinical Microbiology, University Hospitals Birmingham NHS Foundation Trust, Birmingham, UK
Anna L. Casey, PhD
Department of Clinical Microbiology, University Hospitals Birmingham NHS Foundation Trust, Birmingham, UK
Ira Das, MD, MBBS, FRCPath
Department of Clinical Microbiology, University Hospitals Birmingham NHS Foundation Trust, Birmingham, UK
Tony Whitehouse, MD, MBBS, FRCA, FFICM
Department of Anaesthetics and Intensive Care, University Hospitals Birmingham NHS Foundation Trust,
Birmingham, UK
Peter Nightingale, PhD
Wolfson Computer Laboratory, Institute of Translational Medicine, University Hospitals Birmingham NHS
Foundation Trust, Birmingham, UK
Thomas S. J. Elliott, PhD, DSc, FRCPath
Corporate Division, University Hospitals Birmingham NHS Foundation Trust, Birmingham, UK

Abstract
Background: Chlorhexidine gluconate (CHG) has been shown to reduce the microbial load at intravascular (IV)
catheter insertion sites and the risk of catheter-related infections. The Centers for Disease Control and Prevention and
the UK National Institute for Health and Care Excellence subsequently recommended CHG-containing IV dressings for
specific clinical indications.
Aim: To evaluate clinical staff perceptions of a standard transparent IV dressing in comparison to a transparent IV
dressing incorporating a 2% (w/w) CHG gel pad when used at the insertion site of short-term central venous catheters
and vascular access catheters for dialysis in adult critical care patients.
Methods: Following a 9-month trial period during which a CHG dressing was introduced to critical care patients at a
university hospital, the staff perception of this dressing in comparison to a standard transparent IV dressing was
evaluated by a questionnaire. The number of dressing changes required and skin condition under the dressing was also
determined in a proportion of patients.
Results: The majority of the clinical staff (70 out of 81 respondents) considered the performance of the IV dressing
containing a CHG gel pad better or much better than the standard dressing, and 77 out of 78 of the respondents
recommended continuing its use. Both types of dressing performed well when applied to the insertion site of IV
catheters in the internal jugular, subclavian, or femoral vein.
Conclusions: Staff satisfaction with the IV dressing incorporating a CHG gel pad was rated good, and the dressing
performed well in a diverse group of critical care patients. Keywords: CHG, critical care, CVC, IV dressing, vascath

Correspondence concerning this article should be addressed to Tom.Elliott@uhb.nhs.uk


http://dx.doi.org/10.1016/j.java.2016.03.008
Copyright © 2016, ASSOCIATION FOR VASCULAR ACCESS. Published by Elsevier Inc. All rights reserved.

j j
2016 j
Vol 21 No 3
JAVA 133
Background used in our hospital, a transparent IV dressing (3M Tegaderm
T he incidence of bloodstream infections associated with the use of
intravascular (IV) catheters is decreasing, partly due to enhanced
IV dressing; 3M Health Care), which is an adhesive,
semiperme-able, polyurethane film dressing (8.5 cm 11.5 cm
adherence to catheter care bundles and in size), was studied in 2 phases (7 months before and 6
1 months after introduction of the CHG dressing) to reduce any
improved technology. These preventative measures include
education and training of health care personnel who insert and coincidental temporal effect unrelated to the study dressings.
maintain catheters; use of maximal sterile barrier precautions
during central venous catheter (CVC) insertion; application of Staff Training
> 0.5% (w/v) chlorhexidine with alcohol for skin antisepsis, All staff working on the critical care unit and involved in
and avoiding routine replacement of CVCs. These recommen- the care of patients with a CVC had training and were
dations are based on strong clinical evidence; however, there is a competent in the care of CVCs, according to our hospital
paucity of data to support the selection of type of dressing. 2,3 policy. The hos-pital guidance on CVC care reflects the UK
There is a substantial body of evidence available on the use of epic3 and Saving Lives High Impact Interventions, which are
dressings containing chlorhexidine gluconate (CHG) that similar to the Cen-ters for Disease Control and Prevention
demonstrate reduction in the incidence of both catheter-related guidelines on Preven-tion of Intravascular Catheter-Related
infections and catheter site microbial colonization.3-6 The Infections.2,3 These guidelines highlight the main principles
Centers for Disease Control and Prevention Guidelines for the in the prevention of IV catheter-related infections, including
Prevention of IV Catheter-Related Infections3 recommend the correct hand decon-tamination, strict aseptic technique and
use of a CHG-impregnated sponge dressing on short-term rigorous skin prepara-tion, meticulous catheter and site care,
catheters, and antiseptic/antibiotic-impregnated catheters, if in a correct replacement strategy and prompt catheter removal,
particular unit the central line-associated bloodstream infec-tion and regular catheter site observation (minimum of 8 hourly
rate remains high despite adherence to basic preventative observations for inpa-tients). These guidelines also include
measures. The CHG-impregnated sponge, which is placed around indications for dressing change, which recommend that
the insertion site of IV catheters, requires a secondary, dressings are changed every 7 days or earlier if they become
transparent IV dressing to be applied over it. More recently other soiled, loosened, or fluid appears under the dressing.2,3 To
antimicrobial dressings have been developed, including reiterate the correct CVC site care, posters describing the
a transparent IV dressing incorporating 2% (w/w) CHG in an correct application and removal of the IV catheter dressings,
aqueous gel pad.7-9 The use of the dressing containing CHG gel indications for dressing change, and regular observation of the
has been associated with a reduction in catheter-related sepsis.4 CVC site were also displayed on the unit throughout the study
Following a review of the available clinical efficacy data and (during both the standard and CHG dressing study periods).
economic evidence, the CHG gel dressing has been recently Before implementation of the CHG dressing, nursing staff
recommended by the UK National Institute for Health and Care competent in IV catheter care and anesthetists who insert IV
Excellence to be used on critically ill adults who catheters were given training in groups of 1-5. The training
require a CVC or arterial catheter in intensive care or high encompassed the correct method for applying and removing the
dependency units.10,11 dressing with emphasis on the differences between the 2 study
dressings. The training also included ensuring that the skin
Aim of the Study antiseptic had fully dried before applying the dressing, applying
To evaluate clinical staff experience following the imple- the dressing without stretching it, positioning the CHG gel pad
mentation of a transparent IV dressing containing a CHG gel over the CVC insertion site, molding the gel pad around the CVC
pad after the use of a standard transparent IV dressing. The to ensure maximal skin contact, smooth-ing down the dressing
performance of both dressings at 3 different anatomic CVC with a full understanding of the effects of the pressure-sensitive
insertion sites on 2 types of IV catheters inserted into a adhesive in the dressing, and overlap-ping the sections behind the
diverse group of critical care patients was also assessed. lumens to close the gap around the CVC lumens. In addition, the
importance of observing the level of the CHG gel pad fluid
Methods saturation was emphasized. This was performed by lightly
Study Dressings pressing the gel pad, and if the resulting pressure mark in the gel
An adhesive, semipermeable, transparent polyurethane film pad did not dissipate or if the gel itself became displaced, the
dressing incorporating a transparent gel pad containing 2% (w/w) dressing required replacement. To prevent skin trauma when the
CHG (3M Tegaderm CHG IV Securement Dressing; 3M Health dressing was being removed, the importance of gently folding the
Care, St Paul, MN), was implemented for 9 months on the dressing back on itself and slowly peeling the dressing toward
critical care unit at a university hospital. All patients who had a the inser-tion site or in the direction of hair growth was
short-term CVC or vascular access catheter for dial-ysis inserted emphasized. When the dressing incorporating a CHG gel pad
on the critical care unit or in operating theatres had a CHG was being removed, the gel pad was moistened with 1-2 drops of
dressing applied to the catheter insertion site. Two sizes of the sterile fluid, such as saline, which facilitated removal. The
CHG dressing were utilized: 10 cm 15.5 cm (incor-porating a 7.5 importance of regular monitoring of the skin condition around
cm 3.0 cm gel pad) and 8.5 cm 11.5 cm (4.0 cm 3.0 cm gel pad). the CVC insertion site and under the dressing, including
In comparison, the standard dressing inspection of

j j j
134 JAVA Vol 21 No 3 2016
the skin for evidence of irritation or erythema and skin macer- hospital), followed by 9 months’ use of the IV dressing incor-
ation, was also highlighted. porating a CHG gel pad, a questionnaire was distributed to the
trained nurses on critical care and anesthetists in operating
Study Approval the-atres who were competent in CVC site care and had
Research ethics committee (NRES Committee North West experience of using both types of dressings. The questionnaire
- Greater Manchester South, No. 11/H1003/4) and hospital was given directly to eligible staff. Staff members completed
research and development department approvals were the ques-tionnaire anonymously and sealed boxes were used
obtained before commencing the study. Informed written for collec-tion of completed forms to ensure anonymity of the
patient con-sent was obtained before inclusion in this study. respondents.

Performance of the Dressings Results


The performance of the 2 types of dressings was evaluated Performance of the Dressings
in a portion of the critical care patients. This evaluation Dressing performance was evaluated on 273 patients who
included determining the number of dressing changes and the gave their written consent to participate in the study. The
phlebitis scores around the catheter insertion site and su-tures num-ber of dressing changes per patient did not vary
securing the CVC. significantly between the 2 types of dressings (P ¼ .96, based
on Mann-Whitney U test) or between CVC when inserted into
Clinical Environment the inter-nal jugular, subclavian, or femoral vein sites (P ¼
The IV dressings were evaluated on a critical care unit with .75, based on Kruskal-Wallis test) (see the Table). Likewise,
up to 75 critical care beds, at a single university hospital. Pa- there was no significant difference in the number of dressing
tient groups included surgical (general, cardiac, neurology, changes per patient when single or multiple CVCs were
and maxillofacial), medical (general medicine, cardiology, inserted into the same anatomic location (P ¼ .24, based on
and hepatology), trauma, burns, and solid organ transplant Kruskal-Wallis test) or when a quad lumen CVC or vascular
(liver, heart, and lung) patients. access catheter for dialysis was inserted (P ¼ .62, based on
Kruskal-Wallis test) (see the Table).
User Evaluation Similarly, when the number of dressing changes was
After 12 months’ use of the standard transparent dressing adjusted per catheter days there was no significant difference
(which was the standard dressing used throughout the between the CHG-containing and standard transparent

Table. Performance of the Dressings: The Median Number of Dressing Changes Per
Patient With Interquartile Ranges (IQRs) (N ¼ 273 Patients)
Median number of dressing changes (IQR)
CHG Standard P value Test method

All catheters 1 (1-2) (n ¼ 136) 1 (1-2) (n ¼ 137) .96 Mann-Whitney U test

Location of the catheter insertion site

Internal jugular vein 1 (1-2) (n ¼ 112) 1 (1-2) (n ¼ 119) .75 Kruskal-Wallis test

Subclavian vein 2 (1-2)


(n ¼ 17) 1 (0.75-1.25) (n ¼ 12)
Femoral vein 1 (0.25-2.5) (n ¼ 7) 0.5 (0-1) (n ¼ 6)
Number of catheters at the same insertion site

1 1 (0-2)
(n ¼ 98) 1 (0-2) (n ¼ 98) .24 Kruskal-Wallis test

2 1 (1-2) (n ¼ 38) 1 (1-2) (n ¼ 39)


Type of catheter

Quad/5-lumen central venous 1 (1-2) (n ¼ 123) 1 (1-1) (n ¼ 125) .62 Kruskal-Wallis test
catheter

Vascath* 2 (1-2) (n ¼ 13) 1 (0-2) (n ¼ 12)


*Vascular access catheter for dialysis.
j
6 Vol 21 No 3

j j
JAVA 135
dressings. The median (interquartile range [IQR]) number of to the structured questions and perceived improved CVC site
dressing changes per catheter-days were 0.18 (IQR, 0.11- securement and protection. Six staff members commented on
0.25) and 0.20 (IQR, 0.08-0.29), respectively (P ¼ .267, the size of the dressing, which 3 of the respondents
based on Mann-Whitney U test). Furthermore, there was no considered too large. One health care worker commented that
signifi-cant difference between the 2 standard dressing study she would not continue to use the CHG dressing was because
periods in the number of dressing changes per single CVC they were too large to apply. These respondents did not state
when adjusted per CVC days (period 1: median, 0.20 [IQR, 0- which size dressing was used. One staff member commented
0.25] and period 2: median, 0.17 [IQR, 0-0.25]) (P ¼ .226, by on dif-ficulties of visualizing the CVC insertion site through
Mann-Whitney U test). the dres-sing and 3 respondents commented on the stronger
There was also no significant difference in the CVC and su- CHG dressing adhesion (a single respondent considered the
ture site phlebitis scores between the 2 dressing groups. The adhesive to be too strong). Conversely, another respondent
median phlebitis scores for both dressing groups at the CVC believed that the CHG dressing poorly secured multiple
insertion site was 0 (IQR, 0-1; P ¼ .875, based on Mann- lumen catheters. There was no correlation between the health
Whitney U test) and at the suture site was 0 (IQR, 0-2; P ¼ care workers’ duration of caring for patients with CVCs and
.978, based on Mann-Whitney U test). their response to questions (P ¼ .942, based on Spearman’s
correlation r ¼ 0.009).
User Evaluation
In total, 71 nurses who had worked on critical care and 10 Discussion
anesthetists who had worked in operating theatres for at least Previous reports suggest that CHG-containing IV dressings
a year responded to the survey. This represented a response reduce microbial colonization at the IV catheter site and the risk
rate of 16.4%. It was also determined that during the study of catheter-related infection.2,3,10 In addition, cost-efficacy
each nurse cared for more than 200 patients who had a study analysis of the dressings support the use of these anti-microbial
dressing applied. dressings where infection rates remain high following
Overall, the staff members were satisfied with the perfor- implementation of the basic preventative measures, or with high-
mance of the CHG gel dressing, with 86.4% of respondents
risk patients.10 Cost-efficacy analysis has indeed suggested that
rating the overall performance of it as better (35.8%) or much
better (50.6%) than the standard transparent IV dressing (see the use of the CHG gel dressing would save £73 ($105) per
the Figure). patient compared with the use of standard non-antimicrobial
Furthermore, 77 out of 78 (98.7%) staff members who dressings.10,11 However, lack of clinical staff training and
responded recommended continuation of use of the CHG dres- associated knowledge has been shown to limit the adoption of
sing. The supplementary feedback from the users was mostly new technologies.12 Furthermore, although management of
supportive of the dressing incorporating a CHG gel pad (9 out of CVCs and other IV catheters is part of routine patient care in
15 responses), emphasizing their positive feedback acute-care settings, postinsertion CVC site care and compliance
with CVC dressing recommendations are

Figure. Critical care nurses’ and anesthetists’ perception of a transparent intravascular line
dressing incorporating a chlorhexidine gluconate gel pad in comparison with a transparent
intravascular line dressing (N ¼ 81). CVC ¼ Central venous catheter.

j j j
136 JAVA Vol 21 No 3 2016
often inadequate.13 This was illustrated in a recent study of the dressing into clinical practice, to reiterate the correct
where nearly one-third of CVC had suboptimal dressings in technique for dressing removal, thereby reducing the risk of
place, with blood or moisture under the dressing, or the skin trauma and CVC dislodgement.
dressing did not adequately cover the CVC insertion site.13 The majority of respondents regarded the ability to visu-
In this study, staff training, including the use of posters alize the insertion site through the dressing incorporating a
describing the correct method for using the CVC dressings, CHG gel pad the same as or better than through the standard
was fully integrated into the implementation of the new CVC transparent dressing. However, a limited number of staff
dressing in the clinical environment. The training included members reported the transparency through the dressing
reiteration of the importance of postinsertion CVC site care incorporating a CHG gel pad as worse when compared with
and differences between the hospital standard dressing and the transparent film dressing. The aqueous CHG gel pad
the CHG-containing CVC dressing. Appropriate CVC absorbs fluid, which may have affected visualization. This
insertion site care included training to prevent premature further highlights the importance of staff education. The
dres-sing detachment and adverse skin reactions. The transparent yellow-colored gel pad allows direct obser-vation
importance of allowing a patient’s skin to fully dry following of the CVC insertion site without removing the dres-sing;
skin antisepsis was also emphasized in the training because however, the importance of adequate illumination needs to be
this not only re-duces the risk of adverse reactions associated highlighted. This is required to facilitate assessment of the
with the applica-tion of dressings onto a wet skin surface, skin condition at the CVC insertion site adequately. Simi-
such as skin maceration and erythema, but also prevents larly, a high level of fluid saturation in the gel pad is an indi-
dressings from detaching prematurely. cation for dressing replacement and if this is followed,
Education of staff regarding CVC site care is also of para- visualization of the insertion site should not be impaired. The
mount importance for the prompt detection of, and response to, gel pad’s ability to absorb fluid compared with a stan-dard
potential adverse events. CHG is a widely used antiseptic and has transparent film dressing may also prolong the wear time of
been included in body washes, preoperative skin prep-arations, the dressing as fluid accumulation under the film dressing is
and incorporated into some medical devices. However, adverse reduced. The vapor permeability (ie, breath-ability) of the
reactions to CHG have been reported.14-16 In this sur-vey, the dressing also assists in preventing excess fluid accumulation
majority of staff rated the patient’s skin condition under the CHG due to normal perspiration.
dressing as the same as, better than, or much better than while Although profiles of both of the study dressings are similar,
using the standard transparent IV dressing. This was confirmed there were additional handgrips for application of the dressing
by the regular assessment of the skin condition under the incorporating a CHG gel pad compared with the standard
dressing, with no severe adverse reactions being re-ported with transparent dressing. This may have been reflected in the staff
either dressing studied and the phlebitis scores be-ing similar in responses on speed and simplicity of the dressing application.
both study groups. This suggests that when used appropriately, It may also indicate the simplicity of the dressing application
the dressing containing a CHG gel pad is safe to use with diverse where the CHG gel pad is integrated in the dressing, enabling
critical care patient groups. This has been confirmed in other 1-step application.
studies by our group with the current version of the transparent
dressing incorporating a CHG gel pad.5
Conclusions
Although both study dressings were transparent, semiperme-
An IV dressing incorporating a CHG-containing gel pad
able, with a waterproof film, and with the same pressure-
was considered by clinical users to have performed well when
sensitive adhesive, the adherence to skin was rated better with
implemented on a critical care unit with diverse patient
the dressing incorporating a CHG gel pad than the film-dressing
groups. The value of appropriate levels of training in
without CHG. The majority of respondents also considered that
achieving this response was emphasized.
the CHG-containing dressing lasted for up to 7 days, or was
adequate for most patients care plan. These comments may
reflect the additional adherence of the CHG gel pad on the CVC Limitations
and skin compared with the transparent film dressing alone. This Product training is an important part of implementing any
also offers an explanation for why 10% of respondents new device or practice in a clinical environment. The effect
considered the dressing removal from the skin and catheter was alone of implementation of the new type of IV dressing in
more difficult with the dressing incorporating a CHG gel pad critical care without training and education of staff was not
compared with the standard transparent film dressing. Although studied and therefore no firm conclusions can be made on the
standard practice for CVC dressing removal was followed, the influence of the training alone on the study results. Also, the
removal of the gel pad from the skin and CVC should be training provided was not evaluated by the staff.
facilitated by the prior application of sterile fluid, such as saline, The comparator in this study was a transparent IV dressing;
between the CHG gel pad and the skin and CVC. This reduces further studies could compare the performance of different types
the adhesiveness of the CHG gel pad and it is seems probable of CHG dressings, as was recently carried out by another clinical
that this procedure may not have always been followed correctly. team.17 Furthermore, this study did not evaluate the
This highlights further the importance of staff training before cost implications of adopting the CHG dressing, although this
implementation has been carried out elsewhere.10,11

j j
2016 j
Vol 21 No 3
JAVA 137
Recommendation for Practice 5. Karpanen TJ, Casey AL, Whitehouse T, Nightingale P,
When accompanied by the appropriate training, a trans- Das I, Elliott TS. Clinical evaluation of a chlorhexidine
parent IV dressing incorporating a CHG gel pad can be safely intravascular catheter gel dressing on short-term central
implemented for the care of a diverse group of critical care venous catheters. Am J Infect Control. 2016;44(1):54-60.
patients. 6. Safdar N, O’Horo JC, Ghufran A, et al. Chlorhexidine-
impregnated dressing for prevention of catheter-related
Acknowledgments infection: meta-analysis. Crit Care Med. 2014;42(7):
The authors thank the anesthetists and critical care nurses at 1703-1713.
University Hospitals Birmingham NHS Foundation Trust for 7. Carty N, Wibaux A, Ward C, Paulson DS, Johnson P.
their continuous support with research. The authors also thank Antimicrobial activity of a novel adhesive containing
3M for assisting with the staff training. This work, in part, was chlorhexidine gluconate (CHG) against the resident
presented at the Third International Conference on Prevention micro-flora in human volunteers. J Antimicrob
& Infection Control, Geneva, Switzerland, June 16-19, 2015 Chemother. 2014;69(8):2224-2229.
(abstract No. IA). The antimicrobial performance of the dress- 8. Pfaff B, Heithaus T, Emanuelsen M. Use of a 1-piece
ings data were published elsewhere (Am J Infect Control. 2015 chlorhexidine gluconate transparent dressing on critically
Oct 8. pii: S0196-6553(15)00939-6. http://dx.doi.org/10.1016/j ill patients. Crit Care Nurse. 2012;32(4):35-40.
.ajic.2015.08.022).5 9. Jeanes A, Bitmead J. Reducing bloodstream infection
with a chlorhexidine gel IV dressing. Br J Nurs.
Disclosures 2015;24(Suppl 19):S14-S19.
This work and presentation of its results were supported by 10. Jenks M, Craig J, Green W, Hewitt N, Arber M, Sims A.
an educational grant from 3M Health Care, Neuss, Germany. Tegaderm CHG IV securement dressing for central
3M Health Care assisted with the design of the questionnaire, venous and arterial catheter insertion sites: a NICE
but did not have involvement in the study design; the data medical tech-nology guidance. Appl Health Econ Health
collection, analysis, and interpretation; in the writing of the Policy. 2016;14(2):135-149.
report; or in the decision to submit the article for publication. 11. Maunoury F, Motrunich A, Palka-Santini M, Bernatchez
Thomas Elliott and Tarja Karpanen have received honoraria to SF, Ruckly S, Timsit JF. Cost-effectiveness analysis of a
present at a symposium and attend an international confer-ence transparent antimicrobial dressing for manag-ing central
supported by 3M. Thomas Elliott has also received hon-oraria for venous and arterial catheters in intensive care units. PLoS
attendance at advisory board meetings. The other authors have no One. 2015;10(6):e0130439.
conflicts of interest to disclose. 12. Griswold S, Bonaroti A, Rieder CJ, et al. Investigation of a
safety engineered device to prevent needlestick injury: why
References has not StatLock stuck? BMJ Open. 2013 Apr 24;3(4).
1. Bion J, Richardson A, Hibbert P, et al. Matching Michigan 13. Rupp ME, Cassling K, Faber H, et al. Hospital-wide assess-
Collaboration & Writing Committee. ‘Matching Michigan’: a ment of compliance with central venous catheter dressing
2-year stepped interventional programme to minimise central recommendations. Am J Infect Control. 2013;41(1):89-91.
venous catheter-blood stream infections in intensive care 14. Patil PS, Branstiter AD, Tripathi RS, et al. Chlorhexidine
units in England. BMJ Qual Saf. 2013;22(2):110-123. mediated access site ulceration. Int J Crit Illn Inj Sci
2. Loveday HP, Wilson JA, Pratt RJ, et al. epic3: National [serial online]. 2014;4:184-185.
evidence-based guidelines for preventing healthcare- 15. Wall JB, Divito SJ, Talbot SG. Chlorhexidine gluconate-
associated infections in NHS hospitals in England. impregnated central-line dressings and necrosis in
compli-cated skin disorder patients. J Crit Care.
J Hosp Infect. 2014;86(Suppl 1):S1-S70.
2014;29(6):1130. e1-1130.e4.
3. O’Grady NP, Alexander M, Burns LA, et al. Healthcare
16. Weitz NA, Lauren CT, Weiser JA, et al. Chlorhexidine
Infection Control Practices Advisory Committee. Guide-lines
gluconate-impregnated central access catheter dressings
for the prevention of intravascular catheter-related in-
as a cause of erosive contact dermatitis: a report of 7
fections. Am J Infect Control. 2011;39(4 Suppl 1):S1-S34.
cases. JAMA Dermatol. 2013;149(2):195-199.
4. Timsit JF, Mimoz O, Mourvillier B, et al. Randomized
17. Eggimann P, Joseph C, Thévenin M-J. Fitness of use of
controlled trial of chlorhexidine dressing and highly
adhesive dressing for preventing catheter-related Biopatch and TegadermTM CHG for protecting central
infections in criti-cally ill adults. Am J Respir Crit Care venous catheters and arterial lines in critically ill patients.
Antimicrob Resist Infect Control. 2015;4(Suppl 1):O26.
Med. 2012;186(12): 1272-1278.

j j j
138 JAVA Vol 21 No 3 2016

You might also like