You are on page 1of 9

850_Sailer.

qxp 9/17/09 12:31 PM Page 850

In Vitro Study of the Influence of the Type of


Connection on the Fracture Load of
Zirconia Abutments with Internal and External
Implant-Abutment Connections
Irena Sailer, Dr Med Dent1/Thomas Sailer, Dr Med Dent2/Bogna Stawarczyk, Dipl Ing3/
Ronald Ernst Jung, Dr Med Dent1/Christoph Hans Franz Hämmerle, Prof Dr Med Dent4

Purpose: To determine whether zirconia abutments with an internal connection exhibit similar fracture
load as zirconia abutments with an external connection. Materials and Methods: The following zirconia
abutments were divided into four groups of 20 each: StraumannCARES abutments on Straumann
implants (group A), Procera abutments on Brånemark implants (group B), Procera abutments on Nobel-
Replace implants (group C), and Zirabut SynOcta prototype abutments on Straumann implants (group
D). The abutments were fixed on their respective implants either internally via a secondary abutment
(A) or a metallic coupling (C) (two-piece) or directly externally (B) and internally (D) (one-piece). In each
group, 10 abutments were left unrestored (A1 to D1). Ten received glass-ceramic crowns (A2 to D2).
Static loading was performed according to the ISO norm 14801 until failure. The bending moment was
calculated for comparison of the groups and subjected to statistical analysis (Student t test). Results:
The mean bending moments of the unrestored abutments were 371.5 ± 142.3 Ncm (A1), 276.5 ± 47.6
Ncm (B1), 434.9 ± 124.8 Ncm (C1), and 182.5 ± 136.5 Ncm (D1). Two-piece internally connected abut-
ments exhibited higher bending moments than one-piece internally (C1 versus D1 P = .003, A1 versus
D1 P = .03) or externally (C1 versus B1 P = .004) connected abutments. The groups with restorations
did not show different bending moments than those without restorations. The mean bending moments
of the restored abutments were 283.3 ± 44.8 Ncm (A2), 291.5 ± 31.7 Ncm (B2), 351.5 ± 58 Ncm (C2),
and 184.3 ± 77.7 Ncm (D2). Group C2 exhibited the highest bending moment (P < .05). Internally con-
nected one-piece abutments (D2) were weaker than all other groups (D2 versus A2 P = .002; D2 versus
B2 P = .001; D2 versus C2 P = .0003). Conclusions: The type of connection significantly influenced the
strength of zirconia abutments. Superior strength was achieved by means of internal connection via a
secondary metallic component. INT J ORAL MAXILLOFAC IMPLANTS 2009;24:850–858

Keywords: ceramic abutments, connection, fixation, implant reconstructions, stability, zirconia

he good long-term clinical results of dental


T implants allow their application in various regions
of the arches and for diverse indications.1 Following
1Assistant Professor, Clinic for Fixed and Removable Prosthodon-
the functional and biologic success of implant-sup-
tics and Dental Material Science, Center for Dental and Oral
Medicine, University of Zurich, Switzerland. ported reconstructions, additional criteria, primarily
2Private Practice, Ludwigsburg, Germany. esthetic aspects, gain importance.
3Dental Technologist, Clinic for Fixed and Removable Prosthodon-
Standardized titanium abutments exhibit high sur-
tics and Dental Material Science, Center for Dental and Oral vival rates because of their excellent physical proper-
Medicine, University of Zurich, Switzerland.
4Professor and Head, Clinic for Fixed and Removable Prosthodon- ties. 2 However, their application can impair the
tics and Dental Material Science, Center for Dental and Oral esthetic result. In case of soft tissue recession, expo-
Medicine, University of Zurich, Switzerland. sure of the gray titanium abutment can lead to a fail-
ure of the reconstruction in highly visible anterior
Correspondence to: Dr Irena Sailer, Clinic for Fixed and Remov- regions.3 Furthermore, when titanium abutments are
able Prosthodontics and Dental Material Science, Center for Den-
tal and Oral Medicine, University of Zurich, Plattenstrasse 11,
used in patients with a thin labial mucosa, a grayish
8032 Zurich, Switzerland. Fax: +41-44-634-43-05. discoloration of the mucosa can occur owing to the
Email: irena.sailer@zzmk.uzh.ch gray metal color showing through it.3,4

850 Volume 24, Number 5, 2009

© 2009 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY. NO PART OF THIS ARTICLE
MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
850_Sailer.qxp 9/17/09 12:31 PM Page 851

Sailer et al

The esthetic shortcomings of titanium led to the quently, the type and architecture of the implant-
development of ceramic materials as an alternative abutment connection might have a substantial influ-
for esthetically demanding anterior and premolar ence on the stability and fixation of brittle ceramic
regions. The first all-ceramic implant abutments were abutments.
developed in 1993 and were made of alumina, a high- Implants are designed with different types of
strength ceramic. Among the ceramics, alumina implant-abutment connections. The abutments can
exhibits favorable physical material properties.5,6 The either be fixed onto an external connecting part of
bending strength of alumina reaches 547 MPa and its the implant (eg, an external hexagon) or internally
fracture toughness is 3.55 MPa m 1/2. 7 These early into the implant (eg, an internal cone). The internal
abutments were customized manually to meet the connection of zirconia abutments can be accom-
anatomical requirements of each individual site. In plished either by the abutment itself (one-piece) or
vitro studies demonstrated high fracture resistance of by means of secondary components (two-piece).
alumina abutments.8 In clinical investigations, how- One-piece abutments are made entirely of ceramic,
ever, abutment fractures were observed in 7% of sin- whereas for two-piece abutments the internal con-
gle-implant crowns and 1.9% of implant-supported necting part can be either a secondary titanium abut-
fixed dental prostheses.9,10 In the search for a ceramic ment (eg, CARES, Straumann) or a separate metallic
abutment material with improved physical proper- insert (eg, Replace, Nobel Biocare) mounted on the
ties, yttria-stabilized zirconia was introduced in implant together with the abutment and fixed by
1996.11 This high-strength ceramic exhibits fracture means of one abutment screw.
toughness and bending strength that are almost The influence of the type of connection on the
twice as high as alumina ceramic7,12: The bending long-term stability of the abutment-implant complex
strength of zirconia is 900 MPa and its fracture tough- has been analyzed for titanium abutments in several
ness reaches 9 MPa m 1/2 . 13 Zirconia abutments studies. With this type of abutment, mechanical prob-
showed resistance to high loads of up to 738 N in one lems such as loosening or fracture of the abutment
in vitro study.14 In comparison, the naturally occur- screw can occur with external connections.21–23 In
ring mean inciso-occlusal loads in anterior regions one clinical study of an external-connection implant
amount to 110 N for teeth and 370 N for implants.15,16 system, loosening of the abutment screw was the
With these data from in vitro studies, zirconia was most frequent technical complication observed after
expected to reduce the risk of fracture. Indeed, no 3 years of service.22 In contrast, the internal conical
fractures of zirconia abutments were reported after 4 connection was demonstrated to exhibit significantly
years of clinical service in one study.17 higher strength in vitro than the external hexagonal
Because of its different material properties, the connection owing to its higher resistance to bending.
mechanism of ceramic abutment failure differs from The occurrence of abutment screw fracture was lower
that seen with titanium abutments. Ceramics are brit- with an internal connection.24–26 In one clinical study
tle and therefore do not withstand tensile forces very analyzing an internal-connection implant system, the
well. Fracture occurs when the tensile forces exceed cumulative survival rate for the abutment screws and
the limits given by the fracture toughness. In contrast, the restorations supported by titanium abutments
metals are ductile. Their ductility enhances the toler- was 100% after 18 months, and no screw loosening or
ance for both compressive and tensile forces. Prior to fracture occurred.27
fracture, first elastic deformation followed by plastic The stability of zirconia abutments with external
deformation occurs. This property is the reason for connections has been analyzed in detail.14,28 In con-
the excellent loading capacity of metals. trast, to date, the stability of internally connected
The nature and the direction of the load have a ceramic abutments has not been specifically investi-
major influence on the stability of ceramic implant gated. It might be expected that one- and two-piece
abutments. During occlusal loading of an implant- internally connected zirconia abutments exhibit dif-
supported reconstruction, the region around the ferent resistance to loading as a result of a different
abutment screw head is the area of the highest distribution of the loading forces. The aim of this
torque and stress concentrations, and it has been study was to determine whether or not zirconia abut-
demonstrated to be the most critical region for the ments with an internal connection, provided in either
stability of ceramic abutments. 14,18,19 High tensile a one-piece or a two-piece design, exhibit a fracture
forces occurring in this region during function were strength similar to that of one-piece zirconia abut-
the most frequent origin of fracture of ceramic abut- ments with an external connection.
ments in several in vitro studies.14,18 At metal abut-
ments, the same forces first led to deformation and
then to fracture of the abutment screws.20 Conse-

The International Journal of Oral & Maxillofacial Implants 851

© 2009 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY. NO PART OF THIS ARTICLE
MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
850_Sailer.qxp 9/17/09 12:31 PM Page 852

Sailer et al

Table 1 Types of Zirconia Abutments and Their Corresponding Implants, Connection Designs, and Torque
Used for Fixation of the Abutments
Groups Abutments Implants Type of connection Secondary components Fixation torque

A1 and A2 CARES RN Straumann Standard Plus RN, Internal connection 1.5-mm titanium SynOcta 35 Ncm
4.1 ⫻ 12 mm (cone and octagon)
B1 and B2 Procera RP Brånemark MK III TiUnite RP, External connection None 35 Ncm
3.75 ⫻ 13 mm (hexagon)
C1 and C2 Procera RP NobelReplace Select Straight Internal connection Metallic insert 35 Ncm
TiUnite RP, 4.3 ⫻ 13 mm (triangle)
D1 and D2 Zirabut Straumann Standard Plus RN, Internal connection None 35 Ncm
SynOcta RN 4.1 ⫻ 12 mm (cone and octagon)
RN = regular neck; RP = regular platform.

Two-piece One-piece Two-piece One-piece


internal external internal internal
(CARES, (Procera, (Procera, (Zirabut,
Straumann) Brånemark) Replace) Straumann)

Fig 1 Master cast of the clinical situation used for the production Fig 2 Detailed view of the four types of zirconia abutments and
of the master abutment. The abutment shoulder was designed to their methods of implant-abutment connection.
follow the scalloping of the peri-implant mucosa.

MATERIALS AND METHODS that the prospective crown margin would follow the
soft tissue scallop with a slightly submucosal shoul-
Four types of zirconia abutment-implant combina- der (Fig 1). The shape of this master abutment was
tions were included in this study and divided into transferred to all other abutments as follows (Fig 2):
four groups of 20 each: Straumann CARES abutments
for Straumann implants (Straumann), Procera abut- • Group A: Two-piece, connected internally (Strau-
ments for Brånemark implants (Nobel Biocare), Pro- mann CARES abutments on Straumann RN
cera abutments for NobelReplace implants (Nobel implants). The data from the master abutment
Biocare), and Zirabut SynOcta prototype abutments were used for production of an additional 20 iden-
(Wohlwend) for Straumann implants (Table 1). tical zirconia abutments. For the internal connec-
A clinical case in which the maxillary left central tion, corresponding secondar y titanium
incisor had been replaced with a Straumann RN abutments were used (SynOcta 1.5-mm titanium
implant (Straumann) was selected for the design of a abutment, Straumann).
master abutment. In this patient the peri-implant • Group B: One-piece, connected externally (Procera
mucosa had been conditioned by means of an abutments on Brånemark implants). The external
implant-supported provisional prosthesis, resulting in form of the master abutment was mechanically
a scalloped soft tissue margin. The master zirconia scanned (Procera Forte 1.1 scanner, Nobel Biocare).
abutment was digitally designed for this patient’s sit- Subsequently, this copy of the master abutment
uation using three-dimensional abutment fabrication was transferred to the Brånemark MK III Regular-
software (Infinident, Sirona). To achieve this, a scan Platform (RP) hexagonal platform and by means of
abutment was mounted on the implant replica of the computer-aided design (CAD) software (Procera
master cast and scanned (Cerec InEos scanner, CAD Design C3D version 1.1.0, Nobel Biocare).
Sirona). The abutment was designed in such a way Twenty identical abutments were produced.

852 Volume 24, Number 5, 2009

© 2009 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY. NO PART OF THIS ARTICLE
MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
850_Sailer.qxp 9/17/09 12:31 PM Page 853

Sailer et al

• Group C: Two-piece, connected internally (Procera


abutments, NobelReplace implants). The scan of
the master abutment, which had been done previ-
ously for implants of the Brånemark implant sys-
tem, was transferred to the NobelReplace RP
platform, again by means of CAD software (Procera
CAD Design C3D version 1.1.0, Nobel Biocare) and
20 abutments were fabricated. The zirconia abut-
ments were identical to those produced for the
Brånemark implants. For internal connection into
the Replace implants a corresponding removable
metallic insert was added.
• Group D: One-piece, connected internally (Zirabut
SynOcta Regular-Neck [RN] 1.5 abutments on Fig 3 Sample with glass-ceramic crown mounted in the steel
holder of the universal testing machine at an angle of 30
Straumann RN implants). The 20 Zirabut abut- degrees. To ensure an even distribution of the static force, tin foil
ments were produced by copy milling (Copy- was applied between the sample and the indenter.
Milling Machine, Wohlwend), which transferred the
shape of the master abutment to densely sintered
zirconia blanks (Metoxit). These one-piece proto-
type zirconia abutments were designed with a All samples (A1 through D1 and A2 through D2)
cone and an octagon that fit the internal part of were embedded according to ISO Norm 14801. 29
the Straumann RN implants. According to this norm, the implants were embedded
in an acrylic resin holder with 3 mm of vertical dis-
All abutments were fixed on their corresponding tance from the most coronal bone-to-implant border
implants using a torque controller with the torque to the top of the holder, thereby simulating vertical
recommended by the manufacturers. In each group, bone resorption of 3 mm (Fig 3). According to this
10 abutments were left unrestored to receive infor- requirement, cylindric acrylic glass holders (diameter
mation on the influence of the type of connection 16.4 mm) with a modulus of elasticity of more than 3
with emphasis on the abutments (A1 through D1), GPa were custom made for each implant system. The
whereas the other 10 received glass-ceramic crowns implants were placed in a corresponding drill hole in
to more closely resemble the clinical loading situa- the center of the specific holders and fixed with a
tion (A2 to D2) (Table 1). self-curing acrylic resin ( Technovit 4071, Haereus
For the fabrication of the crowns, the master abut- Kulzer). Likewise, the modulus of elasticity of this
ment was fixed on the implant replica of the master acrylic resin was within the requirements of the ISO
cast and scanned (Cerec InEos scanner, Sirona). The norm. 29,30 The flowable acrylic resin was applied
missing central incisor crown was designed virtually through a lateral drill hole using a syringe (Fig 4). Any
(Cerec 3D software, Sirona). Forty identical crowns overflow was meticulously removed. The cement was
were milled out of leucite-reinforced glass-ceramic left to set for 24 hours before the test. In groups A1 to
ingots (ProCAD, Ivoclar Vivadent) by means of a D1, the screw access hole was filled with a light-cur-
milling machine (Cerec InLab, Sirona). After the ing composite (Tetric Ceram, Ivoclar Vivadent).
milling, the crowns were fitted onto the abutments as The embedded specimens were mounted in a 30-
necessary and all crowns were glazed (ProCAD Glaze, degree angled steel holder of a universal testing
Ivoclar Vivadent) according to the manufacturer’s machine (Z 010, Zwick/Roell) (Fig 3). A piece of tin (Sn)
instructions. The crowns were adhesively cemented foil with a thickness of 0.5 mm was applied between
to the abutments. The internal side of the crowns was the indenter of the testing machine and the abut-
etched with a 9% hydrofluoric etching gel (Ultradent ments/crowns. With this procedure, an even distribu-
Porcelain Etch, Ultradent) and silanized according to tion of the load that avoided stress peaks was sought.
the manufacturer’s instructions (Clearfil Porcelain A static load was applied on the palatal side of the
Bond Activator and Clearfil SE Bond Primer, Kuraray). specimens, 2 mm below the incisal edge of the abut-
The zirconia abutments of groups A2 through D2 ments in groups A1 through D1 and of the crowns in
were cleaned with alcohol and silanized using the groups A2 through D2, until fracture or deformation
same silane. The crowns were cemented with a chem- occurred. The feed motion of the indenter was 1
ically curing resin cement (Panavia 21 TC, Kuraray) mm/min. The fracture load was captured in Newtons
according to the manufacturer’s instructions. by means of a specific software (testXpert V.11.02,
Zwick/Roell).

The International Journal of Oral & Maxillofacial Implants 853

© 2009 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY. NO PART OF THIS ARTICLE
MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
850_Sailer.qxp 9/17/09 12:31 PM Page 854

Sailer et al

F Fig 4 Schematic drawing of the setup for the static testing and
detailed information on the parameters needed for the calcula-
tion of the bending moment in Ncm. a = midline; b = direction of
load application; c = total length of sample incuding holder; l =
distance from acrylic resin holder to center of load; F = load; 1 =
b indenter; 2 = sample; 3 = acrylic resin holder.
30°
1

Bending Moments in the Unrestored


Abutments
The internal fixation via a metallic secondary compo-
2 nent (A1, C1) showed higher bending moments com-
3 pared to one-piece externally (B1) and internally fixed
l zirconia abutments (D1) (Table 4, Fig 5). The mean
bending moments were 371.5 ± 142.3 Ncm (A1),
276.5 ± 47.6 Ncm (B1), 434.9 ± 124.8 Ncm (C1), and
182.5 ± 136.5 Ncm (D1). Two-piece internally con-
a c nected zirconia abutments exhibited significantly
higher bending moments than one-piece internally
(C1 versus D1 P = .003, A1 versus D1 P = .03) or exter-
nally (C1 versus B1 P = .004) connected abutments.
Furthermore, zirconia abutments manufactured in
the same way showed significantly higher bending
moments with the internal connection by means of a
metallic insert (C1 versus B1 P = .004). No difference
in the bending moments was found between one-
piece internally and externally connected abutments.
Because of the different location of the crestal Analysis of the samples after the test revealed that
bone–implant border of the examined implant sys- the abutments represented the “locus minoris
tems, the distance from the acrylic resin to the loca- resistentiae” (area of least resistance) in both groups
tion of the indenter was not uniform for all groups, with one-piece external (B1) and internal connections
resulting in different lever arms. As a consequence, (D1). Fracture of the abutments occurred prior to
the bending moment was calculated using the for- deformation of the abutment screws or the implants
mula M = 0.5 ⫻ F ⫻ l (ISO Norm 14801) (Fig 4) and in this group. Interestingly, the internal fixation was
used for comparison.24 The distance l was measured associated with varying types of failures. In 20% of
for each implant-abutment combination in all groups the samples of groups A1 and C1 (two-piece internal
(Tables 2 and 3). connections), the metallic secondary component
failed because of plastic deformation prior to the
Statistical Analysis fracture of the ceramic abutment. In 10% of the sam-
The data were analyzed descriptively using SPSS 12.0 ples of group C1, the deformation occurred in combi-
for Windows (SPSS Inc), visualized using box plots with nation with a fracture of the abutment. In the
whiskers and outliers, and checked for normal distrib- remaining 70%, the abutments fractured without
ution. The fracture load within the groups as well as metal deformation.
the bending moments of the groups A1 through D1
and A2 through D2 were compared. Statistical signifi- Bending Moments in the Restored Abutments
cance was tested by means of the Student t test. The The mean bending moments for the restored abut-
level for significance was set at 5% (P < .05). ments were 283.3 ± 44.8 Ncm (A2), 291.5 ± 31.7 Ncm
(B2), 351.5 ± 58 Ncm (C2), and 184.3 ± 77.7 Ncm (D2)
(Table 5, Fig 6). The restoration of the abutments with
RESULTS adhesively fixed ceramic crowns (A2 through D2) did
not lead to significant changes of the bending
The mean fracture load of groups A1 through D1 moments compared to the unrestored abutments
ranged from 292 ± 218.4 N to 724.9 ± 207.9 N. In (A1 through D1). In the groups with a restoration,
groups A2 through D2, the fracture load ranged from group C2 exhibited a significantly higher bending
245.7 ± 103.6 N to 484.9 ± 80.0 N (Tables 2 and 3). moment (P < .05) than the other groups (A2, B2, D2).

854 Volume 24, Number 5, 2009

© 2009 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY. NO PART OF THIS ARTICLE
MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
850_Sailer.qxp 9/17/09 12:31 PM Page 855

Sailer et al

Table 2 Fracture Loads (Individual Values, Means, Table 3 Fracture Loads (Individual Values, Means,
and Standard Deviations, in N) for Unrestored and Standard Deviations, in N) for Restored
Abutments Abutments
Sample Group A1 Group B1 Group C1 Group D1 Sample Group A2 Group B2 Group C2 Group D2

1 677.5 557.5 600.1 163.7 1 242.4 349.7 433.6 134.4


2 1158.0 373.9 596.7 253.6 2 338.7 423.1 444.3 117.2
3 561.9 389.4 551.9 151.9 3 358.5 386.9 560.4 255.5
4 538.0 390.5 1038.4 220.4 4 383.1 399.0 505.2 210.4
5 507.9 465.5 582.9 234.2 5 457.6 427.5 650.8 175.4
6 475.3 493.4 594.5 279.0 6 411.3 491.6 542.2 408.2
7 515.5 632.3 756.4 215.9 7 380.7 466.6 385.1 178.7
8 738.5 514.9 1157.0 157.8 8 434.6 356.7 444.4 289.5
9 351.6 534.2 664.3 887.9 9 407.5 415.3 464.0 275.2
10 420.6 457.3 706.5 355.7 10 362.8 447.6 418.8 412.6
Mean 594.5 480.9 724.9 292.0 Mean 377.7 416.4 484.9 245.7
SD 227.7 82.8 207.9 218.4 SD 59.7 45.3 80.0 103.6
Mean l (mm) 12.5 11.5 12.0 12.5 Mean / (mm) 15.0 14.0 14.5 15.0
The distance from the top of the acrylic holder to the center of the sta- The distance from the top of the acrylic holder to the center of the sta-
tic load (l) was measured for each group (in mm). By means of l and tic load (l) was measured for each group in mm. By means of l and the
the fracture strength, the bending moment for each group was calcu- fracture strength, the bending moment for each group was calculated
lated and used for comparison. (in Ncm) and used for comparison.

Table 4 Bending Moments (Individual Values, 800


Means, and Standard Deviations, in Ncm) for
Unrestored Abutments Bending moment (Ncm)
600
Sample Group A1 Group B1 Group C1 Group D1

1 423.5 320.6 360.0 102.3


2 723.8 215.0 358.0 158.5 400
3 351.2 223.9 331.1 94.9
4 336.3 224.6 623.0 137.7
5 317.5 267.7 349.8 146.4 200
6 297.1 283.7 356.7 174.4
7 322.2 363.6 453.8 135.0
8 461.6 296.1 694.2 98.6 0
9 219.8 307.2 398.6 555.0 A1 B1 C1 D1
10 262.9 262.9 423.9 222.3 Group
Mean 371.5 276.5 434.9 182.5
SD 142.3 47.6 124.8 136.5 Fig 5 Box plots with whiskers and outliers of the bending
moments for the groups A1 through D1 (unrestored groups). A1 =
two-piece internal connection (Straumann CARES abutments and
Straumann implants); B1 = one-piece external connection (Pro-
cera abutments and Brånemark implants); C1 = two-piece inter-
nal connection (Procera abutments and NobelReplace implants);
D1 = one-piece internal connection (Zirabut SynOcta prototype
abutments and Straumann implants).

The internally connected one-piece zirconia abut- loosening of the abutment screw. Fracture of the
ments exhibited the lowest bending moment (D2 abutment was the reason for failure in 100% of the
versus A2 P = .002; D2 versus B2 P = .001; D2 versus one-piece externally or internally connected abut-
C2 P = .0003). ments (B2, D2).
In 70% of the samples of the two groups with
internal connection via a metallic secondary compo-
nent (A2 and C2), a plastic deformation of the metallic DISCUSSION
components (metallic insert, abutment screw,
implant shoulder) was found in addition to or prior to Two-piece zirconia abutments with a secondary cou-
the abutment fracture. In 50% of the group A2 sam- pling abutment or a metallic insert exhibited signifi-
ples and in 20% of group C2 samples, only deforma- cantly higher bending moments than one-piece
tion of the metallic parts (implant and/or secondary internally or externally connected abutments. Fur-
abutment/metallic insert) was found. This resulted in thermore, the internal connection via a metallic

The International Journal of Oral & Maxillofacial Implants 855

© 2009 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY. NO PART OF THIS ARTICLE
MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
850_Sailer.qxp 9/17/09 12:31 PM Page 856

Sailer et al

Table 5 Bending Moments (Individual Values, 500


Means, and Standard Deviations, in Ncm) for
Restored Abutments

Bending moment (Ncm)


400
Sample Group A2 Group B2 Group C2 Group D2

1 181.8 244.8 314.4 100.8 300


2 254.0 296.1 322.1 87.9
3 268.9 270.8 406.3 191.7 200
4 287.3 279.3 366.3 157.8
5 343.2 299.3 471.8 131.6
100
6 308.4 344.1 393.1 306.1
7 285.5 326.6 279.2 134.0
8 326.0 249.7 322.2 217.1 0
9 305.7 290.7 336.4 206.4 A2 B2 C2 D2
10 272.1 313.3 303.7 309.4 Group
Mean 283.3 291.5 351.5 184.3
SD 44.8 31.7 58.0 77.7 Fig 6 Box plots with whiskers and outliers of the bending
moments for the groups A2 through D2 (restored groups). A2 =
two-piece internal connection (Straumann CARES abutments and
Straumann implants); B2 = one-piece external connection (Pro-
cera abutments and Brånemark implants); C2 = two-piece inter-
nal connection (Procera abutments and NobelReplace implants);
D2 = one-piece internal connection (Zirabut SynOcta prototype
abutments and Straumann implants).

insert showed a significantly increased bending of zirconia abutment was in a prototype stage,
moment compared to the external fixation of the whereas all other abutments were already developed.
same abutment type. The restoration of the abut- In many studies the analysis of the stability of zir-
ments with adhesively cemented all-ceramic crowns, conia abutments has been limited to externally con-
however, did not influence the bending moment in nected abutments. In a recent investigation,
any of the groups. unrestored externally connected zirconia abutments
Therefore, it can be concluded that the type of exhibited a mean fracture load of 294 ± 53 N after
connection to the implant influences the stability of chewing simulation.28 In the present study, externally
ceramic abutments regardless of the presence or connected zirconia abutments exhibited a mean load
absence of restorations. This finding has been of 553 ± 95.3 N under static loading. When recon-
observed for titanium abutments before. The stability structed with an all-ceramic crown, the fracture load
of the implant-abutment complex of titanium abut- was not different. In a comparable investigation, the
ments was significantly higher with internal com- fracture load of the same type of abutments with
pared to external connection.24,26 adhesively luted crowns amounted to 738 ± 245 N
Internal connection has also been associated with under static force application.14 The results of these
a more favorable load distribution in the connection investigations are not consistent. Generally, data on
area.31 A finite element analysis demonstrated high the fracture stability of zirconia abutments are diffi-
tensile stresses in the abutment screw threads upon cult to compare between studies because of different
lateral loading of an externally connected abutment study designs. In the present study the implants and
(butt joint design), whereas with a tapered internal abutments were embedded in the acrylic resin hold-
connection, lateral loading was taken up by the taper, ers, simulating horizontal bone loss, whereas in the
thus protecting the thread portion of the abutment other investigations the implant-supported recon-
from load transfer.32 structions were embedded up to the implant shoul-
The results of the present study demonstrate a der. 14 As a result, the loads were applied with
superior distribution of the load with an internal different lever arms. Furthermore, variations in the
implant-abutment connection for zirconia abut- angle of the applied load, static or dynamic testing
ments. A metallic secondary component, however, methods, and the size and shape of the abutments
seems to be advantageous to transfer the forces. In and reconstructions can have an important influence
contrast to the two-piece internally connected abut- on the results.
ments, the one-piece internally connected abutments In the present study, zirconia abutments with
exhibited the lowest bending moment in the present internal fixation by means of a metallic insert exhib-
study. It has to be considered, though, that this type ited the highest bending moment. The fracture load

856 Volume 24, Number 5, 2009

© 2009 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY. NO PART OF THIS ARTICLE
MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
850_Sailer.qxp 9/17/09 12:31 PM Page 857

Sailer et al

of the abutments with the metallic insert and the Efforts should be made to conduct in vitro studies
ceramic crowns was within the range of the values based on ISO norms and standardizations to allow
reported in other studies. In these studies the stability better comparison of results. Clinical studies are nec-
of zirconia, alumina, and titanium abutments was essary to support the present findings.
analyzed after connection to the same type of
implants.19,20 The mean fracture load of the zirconia
abutments was 457 ± 252 N in one study20 and 470 ± ACKNOWLEDGMENTS
152 N in the other.19 However, in contrast to the pre-
sent investigation, the samples were artificially aged The authors gratefully acknowledge Mr Christian Lüscher for the
fabrication of the cylindric acrylic glass holders and Drs Andreas
before loading, making the results only partly compa-
Ender, Daniel Wolf, and Jörg Michel for their technical support dur-
rable. ing the fabrication of the Cerec crowns. Furthermore, they thank
In the groups with internal connection via a sec- the companies Straumann, Wohlwend, and Ivoclar Vivadent for
ondary metallic component, plastic deformations of the support of the study with implants, abutments, and ceramic
the metallic insert/abutment, the abutment screw, ingots.
and/or the implant shoulder of several samples were
observed. These might be one reason for the higher
standard deviations of the bending moments in REFERENCES
these groups. In these groups, the failure pattern
1. Buser D, Mericske-Stern R, Bernard JP, et al. Long-term evalua-
encompassed fracture of the ceramic abutment,
tion of non-submerged ITI implants: Part 1: 8-year life table
deformation of the metal parts, and/or loss of reten- analysis of a prospective multicenter study with 2,359
tion of the reconstruction. A variation in the fracture implants. Clin Oral Implants Res 1997;8:161–172.
pattern was also observed in other studies reporting 2. Andersson B, Ödman P, Lindvall AM, Brånemark P-I. Cemented
on alumina, zirconia, and titanium abutments with single crowns on osseointegrated implants after 5 years:
internal connection.19,20 In only one of those studies, Results from a prospective study on CeraOne. Int J Prostho-
dont 1998;11:212–218.
however, an implant neck distortion was found in a 3. Prestipino V, Ingber A. All-ceramic implant abutments: Esthetic
specimen bearing a titanium abutment.19 indications. J Esthet Dent 1996;8:255–262.
The abutment screw was identified as the weakest 4. Jung RE, Sailer I, Hämmerle CHF, Attin T, Schmidlin P. In-vitro
component in a previous study examining titanium color changes of soft tissues caused by restorative materials.
abutments.20 In another investigation testing alu- Int J Periodontics Restorative Dent 2007;27:251–257.
5. Prestipino V, Ingber A. Esthetic high strength implant abut-
mina and zirconia abutments, fracture of the abut- ments. Part I. J Esthet Dent 1993;5:29–36.
ment and/or fracture of both the abutment and 6. Prestipino V, Ingber A. Esthetic high strength implant abut-
crown were the main reasons for failure.19 In contrast ments. Part II. J Esthet Dent 1993;8:255–262.
to these results, both ceramic and metal components 7. Lüthy H. Strength and toughness of dental ceramics. In: Mör-
failed by fracture or deformation in the groups with mann WH (ed). CAD/CIM in Aesthetic Dentistry. Chicago: Quin-
tessence, 1996:229–239.
metallic internal connections in the present study. 8. McGlumphy EA, Wall JG, Elfers CL, Ingber A, Prestipino V. New
One reason for the differences found in these investi- ceramic core implant abutment: A comparison study [abstract
gations may be the difference in force application. 74]. J Dent Res 1992;71:115.
Whereas dynamic loads were used in the two previ- 9. Andersson B,Taylor A, Lang B, et al. Alumina ceramic implant
abutments used for single-tooth replacement: A prospective
ous studies,19 static loads were applied in the present
1- to 3- year multicenter study. Int J Prosthodont 2001;14:
study, allowing higher loads before failure. Neverthe- 432–438.
less, naturally occurring forces in patients remain far 10. Andersson B, Glauser R, Maglione M, Taylor A. Ceramic implant
below the forces recorded in these in vitro studies.15 abutments for short-span FPDs: A prospective 5-year multi-
In clinical situations, therefore, a plastic deformation center study. Int J Prosthodont 2003;16:640–646.
11. Wohlwend A, Studer S, Schaerer P. Das Zirkonoxidabutment:
of the metallic components is unlikely to occur.
Ein neues vollkeramisches Konzept zur ästhetischen
Verbesserung der Suprastruktur in der Implantologie. Quintes-
senz Zahntech 1996;22:364–381.
CONCLUSIONS 12. Filser F, Lüthy H, Schaerer P, Gauckler L. All-ceramic dental
bridges by direct ceramic machining (DCM). Mater Medicine
ETH Zürich 1997:165–189.
Internally connected abutments using a secondary
13. Rieger W. Medical Applications of Ceramics. London: Academic
metallic component exhibited the highest bending Press, 1989.
moments, whereas internally connected one-piece 14. Yildirim M, Fischer H, Marx R, Edelhoff D. In vivo fracture resis-
zirconia abutments were the weakest of those inves- tance of implant-supported all-ceramic restorations. J Prosthet
tigated. Restoration of the abutments with adhesively Dent 2003;90:325–331.
cemented reconstructions had no influence on the
bending moment in any of the groups.

The International Journal of Oral & Maxillofacial Implants 857

© 2009 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY. NO PART OF THIS ARTICLE
MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
850_Sailer.qxp 9/17/09 12:31 PM Page 858

Sailer et al

15. Haraldson T, Carlsson GE, Ingervall B. Functional state, bite 24. Norton MR. An in-vitro evaluation of the strength of an inter-
force and postural muscle activity in patients with osseointe- nal conical interface compared to a butt joint interface in
grated oral implant bridges. Acta Odontol Scand implant design. Clin Oral Implants Res 1997;8:290–298.
1979;37:195–206. 25. Möllersten L, Lockowandt P, Lindén LA. Comparison of strength
16. Paphangkorakit J, Osborn JW.The effect of pressure on maxi- and failure mode of seven implant systems: An in-vitro test.
mum incisal bite force in man. Archs Oral Biol 1997;42:11–17. J Prosthet Dent 1997;78:582–591.
17. Glauser R, Sailer I, Wohlwend A, Studer S, Schibli M, Schaerer P. 26. Khraisat A, Stegaroiu R, Nomura S, Miyakawa O. Fatigue resis-
Experimental zirconia abutments for implant-supported tance of two implant/abutment joint designs. J Prosthet Dent
single-tooth restorations in esthetically demanding regions: 2002;88:604–610.
4-year results of a prospective clinical study. Int J Prosthodont 27. Drago CJ, O’Connor CG. A clinical report on the 18-month
2004;15:428–434. cumulative survival rates of implants and implant prostheses
18. Tripodakis AP, Strub JR, Kappert HF, Witkowski S. Strength and with an internal connection implant system. Compend Contin
mode of failure of single implant all-ceramic abutment Educ Dent 2006;27:266–271.
restorations under static load. Int J Prosthodont 1995;8: 28. Butz F, Heydecke G, Okutan M, Strub JR. Survival rate, fracture
265–272. strength and failure mode of ceramic implant abutments after
19. Att W, Kurun S, Gerds T, Strub JR. Fracture resistance of single- chewing simulation. J Oral Rehabil 2005;32:838–843.
tooth implant-supported all-ceramic restorations: An in vitro 29. ISO Norm 14801. Dentistry Fatigue Test for Endosseous Dental
study. J Prosthet Dent 2006;95:111–116. Implants. Geneva, Switzerland: International Organization for
20. Att W, Kurun S, Gerds T, Strub JR. Fracture resistance of single- Standardization, 2003.
tooth implant-supported all-ceramic restorations after expo- 30. Burstein AH, Wright TM. Fundamentals of Orthopedic Biome-
sure to the artificial mouth. J Oral Rehabil 2006;33:380–386. chanics. Baltimore: Lippincott Williams & Wilkins, 1994.
21. Jemt T, Lekholm U, Grondahl K. 3-year follow-up study of early 31. Maeda Y, Satoh T, Sogo M. In-vitro differences of stress concen-
single implant restorations ad modum Brånemark. Int J Peri- trations for internal and external hex implant-abutment con-
odontics Restorative Dent 1990;10:340–349. nections: A short communication. J Oral Rehabil 2006;33:75–78.
22. Jemt T, Pettersson P. A 3-year follow-up study on single implant 32. Merz BR, Hunenbart S, Belser UC. Mechanics of the implant-
treatment. J Dent 1993;21:203–208. abutment connection: An 8-degree taper compared to a butt
23. Scheller H, Urgell JP, Kultje C, et al. A 5-year multicenter study joint connection. Int J Oral Maxillofac Implants 2000;15:
on implant-supported single crown restorations. Int J Oral 519–526.
Maxillofac Implants 1998;13:212–218.

858 Volume 24, Number 5, 2009

© 2009 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY. NO PART OF THIS ARTICLE
MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.

You might also like