SSD-AR70-21
Date: 23-March-2017

TECHNICAL BULLETIN
Page: 1 of 29
CATEGORY PARTS REQUIRED
Top urgent Attached
As soon as possible Separately sent to you
Next time you visit the customer Prepare at your side
When the symptom appears Apply to HITACHI when necessary [Until ]
Information Sales item. Order to the sales person in charge of your country

TITLE:
ARIETTA70, Software Ver.4.1.0 Update

MODEL AFFECTED:
ARIETTA 70
* This Technical Bulletin is applicable to ARIETTA 70/70a, ARIETTA S70/S70a and ARIETTA
V70/V70a.
SERIAL NUMBERS AFFECTED:
All systems running on software version ver.4.0.1 and lower (including updated machines).

PURPOSE:
This is to provide the information of software update of Ver.4.1.0. Please refer to the Newsletter for
the changes and improvements on the software ver.4.1.0.
For the solved problems, refer to the “History of Problems on ARIETTA 70” that is attached to this
technical bulletin.

 Support of EU-9190B (Battery unit) for sleep mode

The optional Battery unit: EU-9190B is added from the system ver.4.1.0 for keeping “sleep
mode” of the system after unplugging the equipment from the mains. Please refer the
Newsletter for more details about the changes and improvements on the software ver.4.1.0 and
the installation of EU-9190B on ARIETTA 70 lower than Ver.4.0.0.

When you transport ARIETTA 70 with this unit installed, you are asked to disconnect the plug
at J850. Please secure the battery cable (With tape etc.) for avoiding to touch the metallic part.
It doesn’t require removing EU-9190B itself.

J850 (Battery unit connector)

L-320・C-8-09-A4
Technical Bulletin No. SSD-AR70-21 Page: 2 of 29

ITEMS OF REQUIRED PARTS:


The connectivity of additional probes are NOT included in this service pack.
However, the instruction manuals for this software Ver.4.1.0 are downloadable from
InfoShare as PDF format. Please follow the instruction below and distribute them to the
customer.

 Software upgrade kit, SP-AR70-S410, 1 set


Please download this software package from Infoshare web site. For more details, please refer
to the section “SOFTWARE UPDATE”.

 ARIETTA70 Instruction manuals CD-ROM version, MN-CD-ARIETTA70-E rev.14, 1 set


Please download it from InfoShare web site and then, burn it into the CD-R media and distribute to the
customers.

 ARIETTA70 CD-ROM Accompanying Document, MN1-5961 Rev.10, 1 set


Please download it from InfoShare web site and then, print out and distribute this document to the
customers with above CD.
If you need complete set of paper based Instruction Manuals for ver.4.1.0, please contact to Hitachi
Healthcare office nearest you (This is NOT free of charge).

ATTACHED DOCUMENTS:
 Product Update Report, ................................................................................................ 1 sheet
 History of problems on ARIETTA 70, ............................................................... A4, 52 pages
 Procedure for adjusting of puncture guideline for EUP-B514 MS4-2925-E, Rev0 .... A4, 5 pages
 Procedure for adjusting of the grid position for Brachytherapy MS4-2927-E, Rev0 ........ A4, 6 pages
 Instructions for setup of Wi-Fi Network Adapter MS4-2926-E, Rev1 ............... A4, 21 pages
 Procedure for list editing of Body Part Examined, Rev0 ..................................... A4, 5 pages
 Procedure for setup of network folder output destination, Rev1 ........................ A4, 10 pages
* Please download them from the InfoShare web site.
ESTIMATED LABOR TIME:
Preparation: 1.0 hours
Upgrade Work: 0.5 hours
Since the software update is “self-extracting” archive file with an update tool, you need to
uncompress the archive and extract the files, then create Update DVD and USB memory stick with
using this tool. For more details, please refer to the section “SOFTWARE UPDATE” in this
document.

SPECIAL TOOLS AND TEST EQUIPMENT REQUIRED:


 Microsoft Windows PC with CD/DVD drive,
DVD-R writing software is installed, available hard disk space is 2Gbytes or more,
virus protection software with the latest virus definition file is installed and the latest OS
updates/patches are applied.
In addition, “.NET Framework 4.5” supplied by Microsoft must first be installed on your PC to
use the tools.
 Portable USB DVD drive (Pioneer, DVR-XD10 is recommended)
 DVD-R disk (Blank), 1 disk
 USB memory stick (capacity is more than 1GBbytes)
Virus scan is applied with Virus protection software with latest virus definition file.
ESSENTIAL CONDITION:
Refer to the section “SOFTWARE UPDATE” in this document.

L-321・C-8-09-A4
Technical Bulletin No. SSD-AR70-21 Page: 3 of 29

UPDATE:
Ver.4.1.0 update procedure with free update kit (service pack) is summarized as following. For
more details, please refer to the section “SOFTWARE UPDATE” in this document.
System software update
 Checking the machine information
 Getting a license key for this update
 Downloading the necessary files from InfoShare web site.
 Making DVD disk and USB memory stick for software update
 Update the software of the machine

There are sales upgrade kits other than this service-pack that updates for Ver.4.1.0. When sales
update kit is used, it is different from the conventional update procedure. The sales upgrade kits
should be applied after updating with service pack. The update step is summarized as following.

ARIETTA 70 Ver.1.1.0~Ver.1.2.3, Ver.2.0.0~Ver.2.3.0, Ver.3.0.0~Ver.3.0.2, Ver.4.0.0, Ver.4.0.1


Including updated machine
ARIETTA 70a Ver.2.1.0~Ver.2.3.0, Ver.3.0.0~Ver.3.0.2, Ver.4.0.0, Ver.4.0.1

Service pack (free update)


Ver.4.1.0 update
(SP-AR70-S410)

ARIETTA 70 Ver.4.1.0 Additional probe compatibility on ver.4.1.0 is


with free update kit NOT included in the service pack.

Ver.4.1.0 probe update Sales update kit


(PM-AR70-S410xxx)

ARIETTA 70 Ver.4.1.0
With sales update kit

The table below describes the differences between free update and sales update kit.
Service Pack (free) Sales update kit
Update kit
SP-AR70-S410 PM-AR70-S410xxx
Additional probe NOT included Included
Instruction manual NOT included Instruction manual CD is
However, CD version of Included
instruction manuals are available
from InfoShare web site
Note You can download the You can download the
installation procedure from installation procedure from
Infoshare web site. Infoshare web site, also it is
included in Sales update kit.

PRODUCT UPDATE REPORT:


After this software update, please send/fax “Product Update Report” to Hitachi Healthcare Business
Unit Quality Assurance Department 1 Office directly.

BEGINNING OF PRODUCTION CHANGE:


The following serial numbers are upgraded to software version Ver.4.1.0 or higher before shipment.
20599704 and higher

L-321・C-8-09-A4
Technical Bulletin No. SSD-AR70-21 Page: 4 of 29

SOFTWRAE UPDATE:
Update procedure is summarized as following.
 Downloading the necessary files from InfoShare web site
 Creating the DVD ISO image file with using the tool
 Creating the DVD for this update
 Creating the USB memory for this update
 Getting the machine information
 Getting a license key for this update
 Checking the machine operation
 Checking the machine configuration
 Updating the software of the machine
 Checking the machine operation and software version
1. Creation of update DVD disk
Create the update DVD disk referring to this and following pages.
1.1 Merge and decompress the downloaded files
SP-AR70-S410
(1) After download all necessary files for
ARIETTA 70 update SP-AR70-S410 from SP-AR70-S410
SP-AR70-S410
InfoShare web site, you can confirm totally 8 SP-AR70-S410.001
FMDToolForSP-AR70-S410
files like a picture shown right.
SP-AR70-S410.002 SP-AR70-S410.003
In this case, please launch the tool named
FMDToolForSP-AR70-S410.EXE by SP-AR70-S410.004 SP-AR70-S410.004

double-clicking the icon.


SP-AR70-S410.006 SP-AR70-S410.007
007 File

8 items

(2) If total numbers of files are 10 and there is a file SP-AR70-S410

named FileMergeDecomp.exe in this folder,


SP-AR70-S410 SP-AR70-S410
please launch this file by double-clicking the
icon.
FMDToolForSP-AR70-S410 SP-AR70-S410.001

SP-AR70-S410.002 SP-AR70-S410.003

SP-AR70-S410.004 SP-AR70-S410.004

SP-AR70-S410.006 SP-AR70-S410.007
007 File

10 items

(3) After launching the tool by either above two


ways, the window like a picture shown right
appears.
SP-AR70-S410
Click [Start...] button to merge split files and
SP-AR70-S410
decompresses automatically.
7 files(s)

SP-AR70-S410.001
SP-AR70-S410.002
SP-AR70-S410.003
SP-AR70-S410.004
SP-AR70-S410.005
SP-AR70-S410.006
SP-AR70-S410.007

L-321・C-8-09-A4
Technical Bulletin No. SSD-AR70-21 Page: 5 of 29

(4) After clicking [Start…] button, the confirmation


message like a picture shown right appears.
Click [Yes] button when you go further. SP-AR70-S410

(5) Tool will merge downloaded split files, and


decompresses merged file automatically. After
this process is completed, click [OK] button on SP-AR70-S410,

the message like a picture shown right.

(6) To close the tool, click [Close] button.

SP-AR70-S410’

1.2 Confirm created files


(1) After closing the tool, please confirm SP-AR70-S410

that downloaded split files (file


extensions are .001 ~ .007) and other SP-AR70-S410 SP-AR70-S410

unnecessary files are deleted


automatically and created sub-folder
named “DVD”. FMDToolForSP-AR70-S410

(2) With this service pack, there is a file


named “SP-AR70-S410.iso” in
this sub-folder.
SP-AR70-S410

L-321・C-8-09-A4
Technical Bulletin No. SSD-AR70-21 Page: 6 of 29

(3) When you need to confirm that the file is not corrupted, please use following MD5 or SHA-
1 hash on a file.
File name Hash value
SP-AR70-S410.iso MD5: d7ce16d8c5b1b910bd0c732739034b02
SHA-1: 62fc7390211c4be84561d33afa40cba3faf35cf6

(4) The created file named “SP-AR70-S410.iso“ at step (2) is ISO 9660 image file. The ISO
9660 file is a disk image file and all information such as files and folders to be created on DVD
are contained in this image file.
You can burn this image file onto a DVD-R media by CD/DVD writing software (If your PC is
running on Windows 7, you can use “Windows Disc Image Burner” to write the image file).
Following example shows that burning a DVD with “Windows Disc Image Burner”. If you have
other CD/DVD writing software, use a similar function of “Burn ISO image file” in your writing
software.

Insert blank DVD-R media to the CD/DVD drive on your PC.


Right-click on the disk image file
SP-AR70-S410.iso, and then SP-AR70-S410.iso

select [Burn disk image] from


pop-up menu.
If there is no selection of this from
pop-up menu, you will find
[Windows Disk image Burner]
in the [Open with…] menu.

When the window like a picture shown right appears,


select DVD drive for burning, put the check mark on 
Verify disc after burning, and then click on the SP-AR70-S410.iso

[Burn] button. DVD-R Drive (E:)

To start burning the disc image, click Burn.

The window like a picture shown right appears and starts


DVD-R writing. SP-AR70-S410.iso

DVD-R Drive (E:)

L-321・C-8-09-A4
Technical Bulletin No. SSD-AR70-21 Page: 7 of 29

After the writing process is completed, Verifying for the


burned disk starts. SP-AR70-S410.iso

DVD-R Drive (E:)

Verifying disc…

After the whole process is completed, burned DVD-R media


is ejected automatically and then shows the message “The
SP-AR70-S410.iso
disc image has been successfully burned to disc.” in status
DVD-R Drive (E:)
window.
To close this tool, click on the [Close] button.

(5) With the steps described above, one update


DVD has been created.
Please confirm that a disk is in the good
SP-AR70-S410-USB
condition by using Windows Explorer on your File folder File folder

PC. The right picture shows the contents of


SP-AR70-S410.
After the confirmation of that the DVD has no
problem, put the indication on the DVD like a
picture shown right. Do not put it on the
DVD’s recording surface.
You can use this created DVD for multiple ver.4.1.0 ARIETTA 70
update work. Ver.4.1.0 update
DVD

SP-AR70-S410

L-321・C-8-09-A4
Technical Bulletin No. SSD-AR70-21 Page: 8 of 29

2. Creation of USB memory stick for this update


Create the update USB memory stick referring to this and following pages.

CAUTION

Change the “View” setting in “Folder Option” of Windows Explorer on your PC as


following.
 Select “Show hidden files, folders, and drives”
 Uncheck “Hide extensions for known file types”

(1) The update DVD is put in the DVD drive on


your PC.
Then, open the folder, “SP-AR70-S410-USB”.
SP-AR70-S410-USB
File folder File folder
(2) Connect the USB memory stick that you use
for update to the PC then, double click
ADAM.exe in this folder to execute the tool.

SP-AR70-S410-USB

SP-AR70-S410-USB SP-AR70-S410-USB.xml

959 KB

218 bytes

(3) After the completion of uncompressing, the tool


shown right appears.
Select the USB memory stick that you connect in the
drop-down list of “Destination” (In this example, it
is K: drive).
If the [Create] button does not become active and a
message “ERROR: USB memory not found!” is
appearing, No USB memory stick is connected on
your PC. Connect it and then click on the [Reload]
button.

Also, Make sure that “ARIETTA70 Ver.4.1.0” is


selected in the drop-down list of “Instrument”.

(4) After clicking on the [Create] button, the


confirmation message like a picture shown
right appears.

Click on the [Yes] button to go further.

L-321・C-8-09-A4
Technical Bulletin No. SSD-AR70-21 Page: 9 of 29

(5) The tool asks you whether the USB


memory stick can be formatted.
Make sure that all files in the USB
memory stick can be deleted then, click
on the [Yes] button.
Although you can cancel the “Media
Format” by clicking on the [No] button,
it is not recommended using an
unformatted USB stick for disk creation.

(6) The tool automatically starts the format program for the USB
Memory stick.
Please just click on the [Start] button without changing the
format conditions. The confirmation message will be appeared
again then, click on the [Yes] button to continue.
When you use the formatted USB memory stick, you may
check on [Quick Format] in “Format options” frame to make
the formatting process shorter.
After the format process is completed, the message “Format
Complete.” appears. Click on the [OK] button on this
window then, click on the [Close] button on format program
window.

(7) The tool will copy the necessary files into the USB
memory stick, and verify if the files are correctly copied
and then, shows the message “All Disks were successfully
created”.

The screen like a picture shown right appears after


clicking on the [OK] button. Then click on the
[Cancel] button to close the tool.

L-321・C-8-09-A4
Technical Bulletin No. SSD-AR70-21 Page: 10 of 29

If you use Windows7 and the screen


like a picture shown right appears,
click “This program installed
correctly”.

3. Checking the machine information


Make a note of the machine ID and the serial number of the machine that you will update the software.
(1) Press [PROBE] switch on the operation panel.
(2) Select [Preset Setup] button on touch-panel.
Note: Below (3) ~ (5) steps are possible to be performed also by the full-keyboard,
when Optional alphanumeric keyboard(EU-9165) is installed to the system.
(3) Press [CTRL] on touch-panel to make it selected, and then press [S] key on touch-panel.
(4) Move the cursor to Password entry field by trackball and press [ENTER] switch on the operation
panel, and then press [CTRL] on touch-panel to unselect it.
(5) Type “ moani ” by the Virtual keyboard on touch panel into Password entry field.
(6) At the [Setup] screen displayed on the LCD monitor, take a note of the information below.
 Eight (8) digits of characters in the "Serial No:" box (It is 203D7249 as shown in the photo
below in this example).
 Nine (9) digits of numbers in the "Machine ID:" box (It is 110001720 as shown in the
photo below in this example).
Setup screen:

70

203D7249

110001720

(7) Then click [Exit] at lower right of the screen to exit maintenance screen.

L-321・C-8-09-A4
Technical Bulletin No. SSD-AR70-21 Page: 11 of 29

4. Getting a license key for this update


Get the license key for this software update, based on the machine information that is taken at the previous
step and the unique number of this update (It is described in the later part of this procedure).
The license key is provided from the web site: "Licensing Center" that is on the Internet. You will get a
license key by accessing to it and type the necessary information in the application form. It requires having
a user account for accessing to it because of the security reason.
Get the license key with following procedure.
(1) Run web browser on your PC, and then access following web address.
https://infoshare.hitachi-aloka.co.jp/infoshare/alc/
(2) The screen below appears when you have an access to the "Licensing Cernter". Fill in your user
account (Your e-mail address: yourname@domain) and passward then, click on the [Login]
button.

(3) Select “ARIETTA 70” at the “Step 0” block in this page. Then enter the "Instrument serial
number" and the "Machine ID" at the "Step 1" block referring the note that is taken at section “1.
Checking the machine information” (They are 203D7249 and 110001720 as shown in the
photo below in this example). The information is validated and when it is correct then, the
product name and the number of licensed software appear at (A) in the photo below (It is
"ARIETTA 70, 1 software(s) licensed" in this example).

(A)

xxxx hospital
xxxx, xxxx, Tokyo, Japan

L-321・C-8-09-A4
Technical Bulletin No. SSD-AR70-21 Page: 12 of 29

(4) At the "Step 2" block in this page, enter 010923000010 for "Product ID" and 3597 for
"Check Code", and then click [Add] button.
Above product ID and check code is reusable and valid only for the ARIETTA70
Ver.4.1.0 update. When we release next software update, we will inform other
numbers for both. Based on the update software you want to use, Enter the proper
numbers at the “Step 2” block in this page.

010923000010 3597

The "End user name" and "End user address" are not essential but recommended to enter them
for your reference.
(5) The parts number of this update you are going to get the licese key appears as in the photo below.
If you wrongly selected the machine to apply this update, click [Return to Step 1] button and
then re-enter the product information.

ARIETTA 70 Ver.4.1.0 UPDATER SP-AR70-S410

(6) Make sure all information inputted at the Step-1 and -2 is correct then, click [Issue] button at
the "Step 3" to get the license key. If you checked on “ Send e-mail notification to
*******@******* as well”, the license key and related information will be sent to the login
user by e-mail.

abcd1234@hitachi-aloka.co.jp as well

(7) Click on the [OK] button when the window like a


picture shown right appears.

(8) The license key with thirty (30) characters will appear in the area of "License key issued: " at the
"Step 3" when it is issued successfully (See in the figure below). Then, take a note of the license
key.

L-321・C-8-09-A4
Technical Bulletin No. SSD-AR70-21 Page: 13 of 29

(9) The machine is able to read the issued license information from the file in USB memory stick.
Click on the [XML] button at the top of the screen.

(10) Click on the [OK] button when the


message like a picture shown right appears.

Save the file ALC_*********.dat into your desktop or somewhere. You need to copy this file
to the folder ”ALC” on the USB memory stick that you use for update after you create it.

ALC_**********.dat

ALC

ALC_**********.dat
ALC_********
DAT file
594 bytes

Caution
It is possible to place several license files for multiple machines in the same
“ALC” folders on the USB memory stick. However, DO NOT mix-up the license
files for both optional software and software update in this folder. Also, make
sure that all license files are in the folder “ALC”, and do not change the name of
license files.

(11) Click on the [Logout] button at the top of the screen, and then close web browser on your PC.

L-321・C-8-09-A4
Technical Bulletin No. SSD-AR70-21 Page: 14 of 29

5. Updating the machine


 Check the operation of the machine well before updating and take the backup of
necessary preset data (DICOM Store/Send etc.).
 Connect Probe(s) to the machine during the update, otherwise an error will happen.
 Do not turn the machine off / Do not remove the USB stick during the accessing the
hard disk and/or USB stick.
 When you connect some USB device into the machine, Disconnect these before update
work.
 When the error message appears during software update, Perform troubleshoot referring
to the section “6. Troubleshooting”.

Update the software version with following instruction.


(1) Connect Probe(s) to the machine before booting-up the machine.
(2) Press [PROBE] switch on the operation panel.
(3) Select [Preset Setup] button on touch-panel.
Note: Below (4) ~ (7) steps are possible to be performed also by the full-keyboard,
when Optional alphanumeric keyboard(EU-9165) is installed to the system.
(4) Press [CTRL] on touch-panel to make it selected, and then press [S] key on touch-panel.
(5) Move the cursor to Password entry field by trackball and press [ENTER] switch on the operation
panel, and then press [CTRL] on touch-panel to unselect it.
(6) Type “ moani ” by the Virtual keyboard on touch panel into Password entry field.
(7) Connect DVD drive that update DVD is inserted in and USB memory prepared in previous step.
(8) Select [Upgrade] from the menu tree, then press [System Update] button by operating
trackball and [Enter] key.
(9) Press [Enter] key when the message on the right figure is
displayed.
(10) The system screen changes to right above, and then the
message shown at right bottom appears. Confirm that license
key for this update is correctly displayed on this window. Then
when you are ready, click on the [Update] button.

The updated license key can also be input by external full


keyboard or optional alphanumeric keyboard (EU-9165).

(11) The files in the USB memory stick are processed, copying
new version of programs, saving current software, and
converting the configuration file.

L-321・C-8-09-A4
Technical Bulletin No. SSD-AR70-21 Page: 15 of 29

(12) When process is finished, the figure right is displayed. Then


click on the [Version Up] button to update the PRESET
database.

Click on the [Yes] button when the window like a picture


shown right appears.

If the update of the PRESET database isn’t necessary,


[Version Up] button is not active, so click on the [Close]
button.

It starts the upgrading of the PRESET database.

Click on the [OK] button when the


window as in the picture right appears.

(13) Following above, the update process for Java is


automatically started. Click on the [Install >]
button to continue the update process.

Note: If updated Java is already installed to the


system, message right bottom appears. Click
[Yes] button in this case.

L-321・C-8-09-A4
Technical Bulletin No. SSD-AR70-21 Page: 16 of 29

When you get the message for completion of Java


update process, click on the [Close] button.

(14) Installation of Microsoft Visual C++2010 is


automatically started.
Check to the box of [I have read and accept
the license terms.] then, click [Install]
button.

Note:
When you update the system running on
Ver.3.0.* and higher, the message shown right
appears. Confirm [Repair Microsoft..] is
selected, then click [Next >] button.

L-321・C-8-09-A4
Technical Bulletin No. SSD-AR70-21 Page: 17 of 29

(15) [Installation Is Complete] is displayed when


the installation is finished. Click [Finish]
button to close the window.

Note:
If you update the system running on Ver.3.0.*
and higher, [Repair Is Complete] is displayed.
Click [Finish] button to close the window.

(16) Click on the [OK] button when the window as


in the picture right appears.

(17) Following above, the update process for


Graphics Driver is automatically started.
Click on the [NEXT >] button.

L-321・C-8-09-A4
Technical Bulletin No. SSD-AR70-21 Page: 18 of 29

Click on the [NEXT >] button when the


window as in the picture right appears.

Click on the [Yes] button when the


window as in the picture right appears.

Click on the [NEXT >] button when


the window as in the picture right
appears.

L-321・C-8-09-A4
Technical Bulletin No. SSD-AR70-21 Page: 19 of 29

The picture right is shown when the


installation is finished. Click [NEXT
>] button to close the window.

When the window as in the picture


right appears, select [No, I will
restart this computer later.],
then click [Finish] button.

(18) After completion of whole update


process, the message shown at right
appears. Click [Finish] button.

(19) Remove the USB memory stick and


then press [OK] button. The system
automatically shut down.

L-321・C-8-09-A4
Technical Bulletin No. SSD-AR70-21 Page: 20 of 29

(20) When restarting the system, the firmware


update will take place.
Depending on the software version before
update, the indication of the firmware is
varied. Shut down the system manually
when the message “Please shutdown
the system” is displayed.
Please note that it doesn’t display all the
progress message of the update on the
main LCD monitor but it can be displayed
on the touch screen when you turn “ON”
the Maintenance Dip SW#1.
Because the condition of the graphic
driver is temporarily undefined due to the
updating. This will be corrected on the
next boot.
When you turn “ON” the
Maintenance Dip SW#1

CAUTION
If shutdown didn’t start when you press Power key, wait for 4 minutes and press Power key
again. Or holding down the Power key with End Exam. key and Menu key 3 seconds at the
same time to activate force shutdown function.

Power key

End Exam. key

Menu key

L-321・C-8-09-A4
Technical Bulletin No. SSD-AR70-21 Page: 21 of 29

(21) Restart the machine then the system automatically starts converting the user data depending
on the software version before update. During this process, the message “Store
VersionUp : ***/***” is displayed at the upper left corner of screen. After the conversion is
completed, the message disappears and then, ultrasound image is displayed on the monitor,
so DO NOT turn the system off during this message is displayed.
Please be careful, the progress indication becomes invisible when you push [Enter] or
[Undo] switch on the operation panel. Even in this case, DO NOT turn the system off
until the ultrasound image is displayed. This process may take up to an hour.

Number of Processed file.

Total number of files to be processed.

(22) After restarting the system, check the system version in the “SYSTEM1” page of the
maintenance menu that is shown as in the pictures below.

EX

System Version 00-4.1.0


OS 00-4.1.0(09Nov) Database 00-4.1.0(r2430)
pcpf 00-4.1.0(17Nov) Manual 00-4.1.0(17Nov)EX

(23) Update process is finished, press [Exit] button to close the maintenance menu.

L-321・C-8-09-A4
Technical Bulletin No. SSD-AR70-21 Page: 22 of 29

6. Setting confirmation after the update


6-1 About setting changed after update
(1) Reverse Depth Control: (Direction of [PAN ZOOM/DEPTH] key rotation and direction
of the displayed image size)
When machine is upgrade from Ver.1 to Ver.2 and higher, Reverse Depth Control is added on
“System Preset”, “General”, “Common 1” tab, Range Zoom Encoder, then default setting is
“ON”.
ON : Turn the [PAN ZOOM/DEPTH] key clockwise to reduce the displayed image.
Off : Turn the [PAN ZOOM/DEPTH] key clockwise to magnify the displayed image.
Please exclude the check and becomes “Off” if you need to keep the same setting before update.

L-321・C-8-09-A4
Technical Bulletin No. SSD-AR70-21 Page: 23 of 29

(2) Pombo, Teichholz, Gibson on Cardio Preset : (When Cardiac Report is Normal)
When machine is upgrade from Ver.2 and lower to Ver.3 and higher, some setting of the
measurement of Cardio application is changed.
Each item of Pombo, Teichholz, and Gibson becomes "Auto" on “Measurement”, “CARDIO”,
“Create Measurement Tools”, “Application Measurement”, “Measured Method & Display
Items”, the 3rd page of B.Mode(Measured Method & Display Item (3/5)).
Please set to “Manual” if you need to keep the same setting before update.

(3) Mark Indicator: (Displayed the measured number to the lower right of the caliper
marks)
When machine is upgrade from Ver.2 and lower to Ver.3 and higher, setting of Mark Indicator is
changed to “Medium” on “Measurement”, each Application”, “Create Measurement Tools”,
“Basic Measurement”, “Measured Method & Display Items”, “Caliper Mark Control”.
Please set to “Small” on every Application if you need to keep the same setting before update.

L-321・C-8-09-A4
Technical Bulletin No. SSD-AR70-21 Page: 24 of 29

6-2 About the influence that added new function gives in images
The setting of some new functions affects image quality after updating from Ver.3.0.* and lower
to Ver.4.0.0 and higher. Please change setting as needed.
(1) Low Echo Reduction(B): (It compensates lower brightness of the B mode image)
Low Echo Reduction is added to “B1” tab on QSS, B1 MENU. In following Application, please
set the Low Echo Reduction to "0" if you need to keep the same setting before update.
 Adult Heart, Pediatric Heart, Cardio Analysis, Coronary
 Carotid, Artery LN, Artery CV, Vein LN, Vein CV
 Breast, Thyroid, Musculoskeletal, Rheumatoid Ar.
 Intestine
 Obstetrics
 Obstetrics3D

L-321・C-8-09-A4
Technical Bulletin No. SSD-AR70-21 Page: 25 of 29

(2) TGC Enhancement: (Optimization of the flow gain in the depth direction)
TGC Enhancement is added to “Common” tab on QSS, Color MENU. QSS Setting of “C”
(QSS Name: V/A) becomes “On” in Carotid Application after updating to Ver.4.0.0 and higher.
If you need to keep the same setting before update, please set the TGC Enhancement to "Off" or
adjust the Color Flow Gain.

TGC Enh: Off TGC Enh: On

The sensitivity of deeper part


is improved, but there would
be unnecessary flow pattern
displayed in the Carotid
application.

L-321・C-8-09-A4
Technical Bulletin No. SSD-AR70-21 Page: 26 of 29

(3) Color Depth: (Color is used front to back to enhance the sense of depth in 3D images)
Color Depth is added to “4D 2” tab on QSS, 3D/4D MENU. In following Application, default
setting of Color Depth is “ON”. Please set the Color Depth to "Off" if you need to keep the
same setting before update.
- Obstetrics, Obstetrics TV, Obst TV 3D, Gynecology TV, Obstetrics 3D
(*In case of Obstetrics 3D, only A, B and D of QSS becomes ON)

L-321・C-8-09-A4
Technical Bulletin No. SSD-AR70-21 Page: 27 of 29

7. How to change the Language for message display


When you want to change [Language] after update, please change setting according to the following
procedures.
(1) Enter the maintenance menu.
(2) Select [Upgrade] from the menu tree, then press [Language Setting] button by trackball and
[Enter] key.

(3) When the Initial Configuration window is displayed, change the [Language] that you need.

(4) The system will be shut down automatically when the [Exit] is pressed.
(5) After restarting the system, check the [Language] in the “SYSTEM1” page on maintenance
menu that it is changed to the selected one.

8. Final operation check


After the update work, please perform the overall operation check including network
communication. In addition, make sure that the preset data, patient database and stored images are
the same as before update.
For the operation, please refer to the instruction manual.

L-321・C-8-09-A4
Technical Bulletin No. SSD-AR70-21 Page: 28 of 29

9. Troubleshooting
Perform troubleshoot referring to the table below when you encounter the difficulties during the
update work.
Process Symptom Solution
Pre-operation The system does not work correctly. Solve the problem first and then
check resume the update work.
Panel [Enter] and [Undo] switch alternating When SW1(Panel Test) for the
operation flashing on the operation panel. maintenance work is set to “↓ON”,
switches are flashing to indicate
that it is maintenance mode.

Update Following message is shown on screen. The update USB memory stick is
not connected to the USB port, or
the update USB is damaged.
Re-create update USB memory
stick and then try to update again.
Update Following message is shown on screen. The update USB memory stick is
damaged.
Re-create update USB memory
stick and then try to update again.

Update Following message is shown on screen. The update USB memory stick is
damaged.
Re-create update USB memory
stick and then try to update again.
Update Following message is shown on screen. The update DVD disk is not
inserted into the DVD drive.
Please insert update DVD disk into
the DVD drive, and then select
[OK] button.

When you select Cancel, message


shown at left bottom is displayed.
Press OK to exit to this update.

Update Following message is shown on screen. Current system software version is


(Version out of range of this update.
confirmation) When you select Cancel, message
shown at left bottom is displayed.
Press OK to exit to this update.

Make sure that current system


software version is correct.

L-321・C-8-09-A4
Technical Bulletin No. SSD-AR70-21 Page: 29 of 29

Process Symptom Solution


Update Following message is shown on screen. The current software version is
(Version Ver.4.1.0.
confirmation) Please select “Yes” to continue.

If you select “No”, the message


shown at left bottom appears and
please select ”Cancel” to exit this
update.

Update The license key is not automatically Make sure that there is a folder
(License key displayed. “ALC” in USB memory stick, and
confirmation) license key file is in this folder.
Also, make sure that file name of
license key is NOT changed from
the original.

Update Following message is shown on screen. The license key is incorrect.


(License key
Make sure that machine
confirmation)
information and license key are
correct.

Update Following message is shown on screen. The update USB is damaged.


Re-create update USB memory
stick and then try to update again.