Professional Documents
Culture Documents
2J313-002E"P
TOSHIBA
PERFORMANCE Ct-tECKMANUAL
FOR
(23313-OOZE*P>
TQSHIRA CORPORATION
%? TOSHIBA
CORPORATION
1986
ALLRIGHTS
RESERVED
No. 2J313-002E*P
REVISION RECORD
REV. DATE! REASON PAGE DOC
CHANGED No. PRODUCI
/AUTHOR
INI. --__-_-
'86
Apr. 18 Mr. Wakai NC
:A
:B '87 Correction
Sep. 11 Mr. Wakai
:C '88 Correction
Feb. 20 Mr. Ishiyama
:D '88 Correction
May 11 Mr. Kimura
FE '88 Correction
Dec. 23 Mr. Ishiyama
:F '89 Correction
May 26 Mr. Kimura
*H '90 Correction
Mar. 22 Mr. Kimura
.I '90 Correction
May 21 Mr. Mochizuki
:K '90 Correction
Sep. 20 Mr. Ishiyama
FL '90 Correction
Nov. 30 Mr. Kimura
FM '91 Correction
Oct. 15 Mr. Kimura
:N '92 Correction
Jan. Mr. Kimura
:O 03/'92 Correction NC
Mr. Ishiyama
*
R-l
No, 25313-002E*J
CONTENTS
Page
1. INTRODUCTION ______________-_-__---~-------~~~-~--~--~~-------~ 1
2. PREPARATION ______________________________I_________-~~-~---~-- 1
-a-
No. 2J313-002E”B
CONTENTS
Page
--_--_-_---___ 31
5.4.3 Fluoroscop ic tube voltage adjl .is tment
_________-__---_-__----- 33
7.1 PWB Setting and Adjustment Tables
-b-
No. 2J313-002E*A
1. INTRODUCTION
Before adjusting the generator as described in this manual, check that the
equipment has been installed according to the installation manual
(2C313-020E) and understand operating method given in the operation
manual.
Refer to the wiring diagram supplied with the equipment when performing
the adjustment or test.
2. PREPARATION
Measuring instrument
No. - Remarks
Item Hodel
i<Others>
I
/ Acrylic phantom 180mm thickness
-l-
No. 2J313-002E*M
2.2 Precautions
2.2.1 Introduction
- 2 -
No. 2J313-002E*M
- 3 -
No. 2J313-002E;kB
(3) Recommendations
-4-
No. 2J313-002E*B
(1) All setscrews and distribution terminal assembly screws should not be
loose.
. Since the cables were connected for the three-phase 415 V power source
prior to shipment, reconnect the cables according to the installed
power supply (in accordance with section 6.4 of manual No. 2C313-020E).
4. SYSTEM SETTING
-5-
No. 2J313-002EfcD
5. PERFORMANCE CHECK
Note: Check to see that the shielding terminal of the cable (for serial
transmission) is connected to the cabinet for noise prevention. If
it is not connected, error 700 (a communication error> may occur.
Chassis Y
-6-
No. 2J313-002EkB
i Start 1
I
I
Disconnect the connector from the trigger PWB within the INV unit.
t
System setting (for all items indicated as “necessary” in the
“when installed” columns of tables in section 7.1. >
i Energizing )
I
t
I
) Unit test (sections 5.1.1 to 5.1.6)
I sequence check I
Filament heating check and heating data (cable length and tube)
setting
t
r- Power off I
Connect the connector to the trigger PWB within the INV unit.
I
Connect the connector after the capa-
citor voltage become less than 5V.
ABC/AEC adjustment
ATR adjustment
5.1.1 Power ON
Turn ON the main switch of the distribution panel and the power unit
breaker within the control box. Check for abnormality by turning the
panel or stand power switch ON/OFF.
(1) KXO-30R
Make sure that the voltage between CH3 (+24 V> and CHl (GND) is
+24 v+3
- v.
p2 (+5 v), ~3 (+I5 V> and P4 (-15 V> within Power Unit 1
(2) KXO-30F or 50 F
-8-
No. 2J313-002E*B
Tl input tap
VR to be Measurement j
1 adjusted point
200 v 220 V, 380 v, 415 V, 440 V /
1
-9-
No. 2J313-OOZEW
Other nominal
Nominal 200V values = Vs
INV-3
Radiography flu LV LEVEL
LV>zzoVQc LV>V$$-10%
scopy disabled area
s
m u
% R64
LED6
,Q u
LED5
zu%
ii: u
19O>LV vs-lO%>LV
scopydisabledarea
(2) ID function
Repeating turning power ON, OFF, and ON, check that the following
indications equal those before turning power OFF.
. R kV value
. R mAs value
. R mA value
. R TIME value
. TEC. SW
. SF/LF
Make sure that the contents written in the ID are not destroyed by
turning power ON and OFF.
- 10 -
No. 2J313-002E"cA
Check that the panel of stand focus indication varies with the set
tube current.
NSS and F-ROT are controlled by the normal speed starter within the
X-ray control. The starter frequency equals the power source frequency
(50 or 60 Hz).
HSS is controlled by the high speed starter unit (RS-118A: option), and
the starter frequency is 180 Hz.
Check the wiring, phase advance capacitor value, and set time of the
installed X-ray tubes as described below.
( 1) Check the type of the X-ray tube combined. There are two types of
tubes depending on the X-ray tube stator.
The relation between the phase advance capacitor value and the
X-ray tube type is given below.
- 11 -
No. 2J313-002E”cA
The relation between the boost time and the X-ray tube type is
given below.
I I
/ 1
X-ray tube type / Boost time
I +o. 1
TOSHIBA U coil 1 0.8 set
I -0
Block TB2 of the power unit within the control box. (100 VI
div, 0.2 sec/div, storage mode)
(e) Put into the Ready state. Observe the waveform on the
oscilloscope and check t’ne boost time.
[HSS mode 1
(a) Set the stator coil, start voltage, R start time, F start time,
control voltage, and control time as specified in the high
speed starter installation manual.
(d) Put into the Ready state, and check the high speed anode
rotation according to the said installation manual.
- 12 -
No. 2J313-OOZE*1
The adjustment method for tube voltage and current is given below.
The R.mA and R.kV adjustment procedures are given separately below. R.mA
and R.kV should be adjusted simultaneously at each mA step.
(1) KXO-30R
<l> Set up the required units as illustrated below and adjust both
the tube voltage and current.
Control box
Secondary tube current
\
meter (to be mounted
to the X-ray tube)
\
G--J
Power source
9
\
-- --
generator
Anode
Bleeder
resist-
ance
f-A--
Cathode1
X-ray
tube
DG-303
Power ground
terminal
/
Storage
oscillo-
scope
- 13 -
No. 25313-002E*l
<2> Set up the required units as illustrated below, check the X-ray
tube current and readjust it if necessary.
Control box
Secondary tube current
meter (to be mounted
to the X-ray tube)
Anode
High X-ray
Power
voltage tube
source -_:l 1 Cathode
--
5
--Jr I
1
Power ground -I
terminal
- 13A -
No. 25313-002E*J
Tl Tl High lnode
X-ray
Control _ voltage
T2 T2 Bleeder tube
Power source n box gener-
resist-
\ ator Cathode Cathod
&-- --- ance
7
--A
Tube
-f- voltmeter
Power ground
terminal
Storage
oscillo-
scope
- 14 -
No. 25313-002E*J
TEST
x2> Mode selec tor SW
I MODE
- 15 -
No. 25313-002E*J
3 moo
0 ON (Lighted)
- 16 -
NO. 25313-002E*J
Mode 6
- 17 -
No. 25313-002E*J
--+<2> Set the tube No. to be changed via the operation panel of the
X-ray control unit.
<4> Enter the new cable length using the adjustment switches and
the S/R switch.
If the two-digit (00 through 99) input is completed, the mode
indicator stops blinking, resulting in the condition in <2>.
[Example: 6 ml
0 0 0 0
Press Press
Press
L 4 ---f
Second
time [z.][-zj
m 1-q
0 1 1 0
aI
S/R
0
<5> Perform the operations (2) through <4> when changing the high-
voltage cable length for the 2nd or 3rd tube.
0 0000
1 0001
2 0010
3 0011
4 0100
5 0101
6 0110
7 0111
8 1000
9 1001
- 18 -
No. 25313-002E*J
--+<2) Set the tube number to be set through the operation panel of
the X-ray control unit.
r \
<3> When the I/), switch is pressed, the mode indicator blinks, and
<4> Enter the binary code corresponding to the tube to be used from
Table 5.2.2.
[ Examp le I DRX-1203*
Do not press Press
--_)
First time
Binary code
aI 0
S/R
Press Press
L *
al 0 Second time
S/R
-<5> Perform this setting for the 2nd and 3rd tubes.
The above procedure is used to set the data with both large and
- 19 -
No. 23313-002E*J
Adjustment for each point in Mode 1 Adjustment for each MA line in Mode 2
The adjustment points for mammography The line adjustment for mammography
are 20, 25, 32 and 45 kV. should be a; 25 kV.
25kV
20kV 1 I]
800mA 1 I
<Mode l>
The standard data (for three tubes) is set before shipment.
Correct the data depending on combinations of tubes.
Since the preheating and preflash are not checked, adjust the
current after the filament heating is stabilized. (Generate X-rays
by pressing the READY switch and the X-ray switch at intervals of
more than 2 seconds. (Do not press them simultaneously. Take note
of the heating data as shown in table 10. (1) in APPENDLX 01 after
adjustment.)
- 20 -
No. 25333-OOZE*5
<3> Set R.kV and R.mA to be adjusted in the same way. Some
combinations of R.kV-R.mA are limited to Maximum Power (30
kW/SO kW).
R.kV: 40, 60, 90, 120 kV or 20, 25, 32, 45 kV
R.mA: 10, 13, 16, 20, 25, 32, 40, 50, 64, 80, 100, 125, 160,
200, 250, 320, 400, 500, 640 mA
<5> Press the Ready SW and the X-ray SW simultaneously and confirm
the set conditions again to see if it is necessary to perform
preheating or calibration of preflash level.
<Mode 2>
The adjustment in this mode causes 40, 60, 90 and 120 kV to
simultaneously increase or decrease.
Mode 2 adjustment is the same as Mode 1, except that this
adjustment should be made with the R.kV tube voltage set to 60 kV.
When the X-ray tube for mammography is used, the tube current
adjustment in Mode 2 should be done at 25kV.
The preflash level should be adjusted only when the tube current
does not rise within the specified time in (5) <6>.
Preflash is intended to increase the X-ray tube filament rise rate.
After the READY switch is pressed, the preflash is done prior to
main heating.
~Preflash
Main heating
I
Heating level
Preheating
I Time
READY w ON ))
- 21 -
No. 25313-002E*J
For this unit, the preflash time is fixed and the amount of
preflash can be obtained by multiplying the amount of main heating
by the preflash factor.
Therefore, the preflash level is adjusted by changing the preflash
factor.
PRFl, PRF2, and PRF3 for the three adjustment points can be
selected as follows:
--+<2> Set conditions under which the preflash level is short in (5)
<5>, via the operation panel of the X-ray control unit.
- 22 -
No. 25313-OOZE*5
L<5> When the S/R switch is pressed, the indicator stops blinking
and the state in <2> is reset. In this state, another preflash
factor, another focus or another tube preflash factor can be
adjusted.
r
<2> Set the conditions under which the preflash level adjustment is
not satisfactory, via the operation panel of the X-ray control
unit.
<3> When the switch is pressed, the mode indicator blinks and
Increase Decrease
(Note 11, and check th e tube current rise by generating X-rays.
<5> When the S/R switch is pressed, the indicator stops blinking
and the state returns to <2>. In this state, another focus or
another tube preflash factor can be adjusted.
- 23 -
No. 25313-002E*J
<2> Use the operation panel of the X-ray control unit to specify
the number of the tube whose fluorographic focal size needs to
be changed.
<4> To adjust the focal size to the large focus (LF), tilt the S/R
switch only.
For the small focus (SF), tilt the S/R switch, while pressing
the switch.
<5> Reset to mode “0” by means of the S/R switch when the focal
size adjustment has been completed.
<l> When the S/R switch is pressed while the alarm sounds, the
alarm stops and Mode 7 is set.
(2) When the A, switch is pressed, the indicator blinks and the
sl
recovery routine starts.
<4> Reset Mode 0 by pressing the S/R switch and check the data. If
the power supply is turned off in the adjustment mode during
tube current adjustment , data is judged to have be destroyed
- 24 -
No. 2J313-O02E”CJ
when the power supply is turned on again, and the alarm sounds.
(This is because it is not possible to distinguish data change
from data breakdown. > In such cases, recover by using the
above procedure. After recovery, data adjusted prior to
turning off the power supply are all effective.
> Alarm
- 24A -
No. 25313-002E"J
TEST
<2> Mode selector switch
MODE
- 24B -
No. 25313-002E*J
0
I l ** Standby state
3 moo
0 ON (Lighted)
- 24C -
No. 25313-OOZE”5
Mode 7
Mode 2
Mode 1 Mode 5
Mode 4
Note 1: When the X-ray generator is used with an X-ray tube with a
very small focus, such as the X-ray tube for magnification
radiography, do not connect tube H.T. cables until the tube
code is set. Failure to do this may damage the tube focus.
- 24D -
No. 25313-002E*J
<2> Set the tube number to be set through the operation panel of
the X-ray control unit.
<3> When the /) switch is pressed, the mode indicator blinks, and
CJ
the equipment goes into the code input state.
<4> Enter the binary code corresponding to the tube to be used from
Table 5.2.2.
[Example] DRX-1203B
Do not press Tilt
S/R
0 0 1 0
A
<5> Perform this setting for the 2nd and 3rd tubes.
The above procedure is used to set the data with both large and
- 24E -
No. 25313-002E*J
Adjustment for each point in Mode 1 Adjustment for each MA line in Mode 2
The adjustment points for mammography The line adjustment for mammography
are 20, 25, 32 and 45 kV. should be at 25 kV.
50mA
lOOmA lOOmA l I I I
.
.
.
800mA 1 800mA
<Mode l>
The standard data (for three tubes) is set before shipment.
Correct the data depending on combinations of tubes.
Since the preheating and preflash are not checked, adjust the
current after the filament heating is stabilized. (Generate X-rays
by pressing the READY switch and the X-ray switch at intervals of
more than 2 seconds. (Do not press them simultaneously. Take note
of the heating data as shown in table 10. (1) in APPENDIX 01 after
adjustment.)
- 24F -
No. 23313-OOZE*5
:I
L- <4>
<3> Set R.kV
combinations
kW/50 kW).
R.kV:
Correct
switches
200, 250,
set conditions
current.
the
(Note
and
Adjust
R.mA to
of R.kV-R.mA
320, 400,
data
in order
all
are limited
500, 640 mA
with
1) by generating
to adjust
conditions
be
the
adjusted
to Maximum Power (30
X-rays
the
to be
100, 125,
y/
cl ;
in
160,
(increase)
the same way.
or
n
/(/I
(decrease)
<5> Press the Ready SW and the X-ray SW simultaneously and confirm
the set conditions again to see if it is necessary to perform
preheating or calibration of preflash level.
<Mode 2>
The adjustment in this mode causes 50, 60, 90 and 120 kV to
simultaneously increase or decrease.
Mode 2 adjustment is the same as Mode 1, except that this
adjustment should be made with the R.kV tube voltage set to 60 kV.
When the X-ray tube for mammography is used, the tube current
adjustment in Mode 2 should be done at 25kV.
The preflash level should be adjusted only when the tube current
does not rise within the specified time in (4) <6>.
Preflash is intended to increase the X-ray tube filament rise rate.
After the READY switch is pressed, the preflash is done prior to
main heating.
/ Preflash
.
Main heating
I
Heating level
Preheating
1 Time
READY mON ))
- 24G -
No. 25313-002E*J
For this unit, the preflash time is fixed and the amount of
preflash can be obtained by multiplying the amount of main heating
by the preflash factor.
Therefore, the preflash level is adjusted by changing the preflash
factor.
PRFl, PRF2, and PRF3 for the three adjustment points can be
selected as follows:
<2> Set conditions under which the preflash level is short in (5)
<5>, via the operation panel of the X-ray control unit.
- 24H -
No. 25313-002E*J
Increase Decrease
will sound and the preflash factor will increase or
decrease by 1%. If either switch continues to be
pressed, the buzzer will automatically start sounding
so that the result can be achieved as if the switch
were pressed by the number of times the buzzer sounded.
---+<2> Set the conditions under which the preflash level adjustment is
not satisfactory, via the operation panel of the X-ray control
unit.
<3> When the switch is pressed, the mode indicator blinks and
Increase Decrease
(Note l), and check th e tube current rise by generating X-rays.
<5> When the S/R switch is pressed, the indicator stops blinking
and the state returns to <2>. In this state, another focus or
another tube preflash factor can be adjusted.
- 241 -
No. 25313-002E*J
<2> Use the operation panel of the X-ray control unit to specify
the number of the tube whose fluorographic focal size needs to
be changed.
<4> To adjust the focal size to the large focus (LF), tilt the S/R
For the small focus (SF), tilt the S/R switch, while pressing
the
A buzzer
Ed. switch.
<5> Reset to mode “0” by means of the S/R switch when the focal
size adjustment has been completed.
T Tilt
Hold down this switch
cl> On the operating panel of the generator, specify the tube No.
for which the preset compensation current level needs to be
adjusted.
<3> Enter the tube current which is to be corrected. (See Note 1. >
- 245 -
No. 25313-002E*J
<6> When the current level is correct, set the S/R switch down;
the indicator stops blinking.
<7> Generate X-rays to confirm that the adjusted current level for
the dropout tube current is correct. (See Note 5.)
Repeat Procedure <4> if the compensation current level requires
further adjustment, or repeat from Procedure <3> if it is
necessary to correct the compensation current level for other
tube currents.
Note 1: Tube current values of 20, 50, 100, 200, 500 and 64OmA
can be entered in Procedure <3>. The tube voltage
should be set to 90kV.
Note 2: Increase the compensation current level to correct for
a fall in tube current during X-ray exposure and
decrease it to correct for a rise in tube current.
<l> When the S/R switch is tilted while the alarm sounds, the
alarm stops and Mode 7 is set.
- 24K -
No. 25313-002E*J
<4> Reset Mode 0 by tilting the S/R switch and check the data. If
the power supply is turned off in the adjustment mode during
tube current adjustment, data is judged to have be destroyed
when the power supply is turned on again, and the alarm sounds.
(This is because it is not possible to distinguish data change
from data breakdown.) In such cases, recover by using the
above procedure. After recovery, data adjusted prior to
turning off the power supply are all effective.
(10) Alarm
- 24L -
No. 2J313-002E*P
If the alarm sounds, turn off the READY switch decrease the
heating level.
.
No. Binary code Tube name No. Binary code Tube name
0 00 0000 32 10 0000 DRX-3535HC
1 00 0001 33 10 0001 DRX-3545HC
2 00 0010 DRX-1203B 34 10 0010
3 00 0011 DRX-1403B 35 10 0011 DRX-3603HD
4 00 0100 DRX-1415B 36 10 0100 DRX-3624HC
5 00 0101 37 10 0101 DRX-3634HC
6 00 0110 DRX-1603B 38 10 0110 1
- 25 -
No. 2J313-OOZE*1
Table 5.2.3
Setting position
Power supply
of SW1
Single-phase 200 V 0
(190 to 215 V>
Single-phase 220 V 1
(216 to 240 V)
Three-phase 380 V 4
(380 to 400 V)
Three-phase 415 V 5
(401 to 425 V>
Three-phase 440 V 6
(426 to 1
Set SW5 on the INV2 PWB according to the table below. (No setting
is required for PWBs without SW5.j
to 15m 3, 2, 1 or 0
16 to 20m 2, 1 or 0
r- 21 to 25m 1 or 0
26 to 0
- 26 -
No. 25313-002E*I
[Adjustment procedure]
<2> Look at the display of 7--segment LED on the INV-2 PWB. This is
tube voltage shift data and stored into the RAM. If an
increase or decrease is perfomed around this numeric value, the
tube voltage will also be increased or decreased. One step
corresponds to about 2 kV. Thus, determine a value to be
written anew.
<3> Ad just SW&3 (rotary switch) on the INV-2 PWB to the value
determined in <2>, then press SW3. Consequently, the RAM is
rewritten. Check that the display of 7-segment LED has changed
to this value.
<4> Generate X-rays again and measure the tube voltage. If the
tube voltage is not found within the tolerance, repeat the
above from Item (1).
0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
- 26A -
No. 2J313-002E*I
Press SW3.
- 27 -
No. 2J313-002EycB
Adjustment
30kV 40kV 60kV 90kV 120kV 140kV
value
Tube voltage
TP2
waveform
Gradual increase
(2)
T Fine-adjust VR3 by turning
I-
-we--_ it CW (towards "3").
Gradual decrease
-I _L
(3) If a similar waveform is
observed, do not turn VR3.
F=
I
In a single-phase 2OOV I
system, the waveform may i
show a gradual decrease I
under a large load.
Check that the radiographic timer operates correctly for several set
radiographic times as described below.
(1) Connect the monitor output terminal of the tube voltmeter and the
input terminal of the oscilloscope with a coaxial cable.
(4) Set the kV, mA, and time values as specified in Item 7.2.2 (2), and
generate X-ray.
(5) Read each radiation time from the waveform on the oscilloscope, and
check that it is within the allowable range.
The radiation time is the interval from Point A, where the tube voltage
increases to 75% of its peak value, to Point B, where the tube volt age
decreases to 75% of its peak value.
75% / 100%
75%
\ I I
High potential
waveform
Tolerance Tolerance
Fig. 5.2.4
- 29 -
No. 2J313-002E;kB
The followings are necessary only when automatic exposure control (AEC)
is installed.
(2) Check that the settings of the AEC unit are correct.
Refer to the AEC-lOA/lOB installation manual, 2C376-519E.
@ Set the time value longer than an appropriate AEC radiation time.
If this occurs, the panel or desk timer functions as an AEC backup
timer.
lmll
I I
2mA I I !
3rnA I I I
- 30 -
No. 253 13-002E+B
(1) Select SPOT (T u b e 1) via the operation panel of the X-ray control
unit.
(3) Set the F.kV and F.mA values to be adjusted with the spot operation
part of the conventional R/F table or the operation part of F-unit.
Combinations:
F.kV: 50, 90, 120kV
F.mA: 05, 1, 2, 3, 4mA
(5) When the adjustment is completed, reset Mode 0 by pressing the S/R
switch.
Note 1: When the [z] or [z] switch is used, the value can be
If the X-ray tube current is over, it may not be possible to obtain the
correct tube voltage. In this case the tube current must be adjusted.
Operation (1) below is the main adjustment procedure and operation (2)
is the supplementary adjustment procedure.
Fine-adjust VR6 on the INV-1 PWB so that the measured value of the
tube voltmeter becomes 90kV+lkV under the conditions of 90kV and
1mA. LED4 should indicate sT8” in the fluoroscopic mode.
By manually setting all mA and kV values, check that they are
within 7% of the kV accurate value.
(2) If the tube voltage accuracy is not yet within 7% after adjustment
(1) above, use SW2 and SW3 on the INV-2 PWB as in the case with the
radiographic tube voltage adjustment. This adjustment, however, is
possible only in the fluoroscopic mode. When this mode is turned
off, the radiographic tube voltage adjustment is possible.
In addition this adjustment is performed for each voltage section
shown below at 0.5, 1, 2, 3 and 4mA respectively.
No. 2J313-002E*B
Note : Take care not to obtain the very discontinuous tube voltage
at the boundaries at each section.
After all of the preliminary and final tests to be necessary for the
installation system have been successfully completed, the HHS test phase
may be performed.
The HHS test is necessary only when finally checking the ORH compliance.
The test method, equipment required and rejection limit are shown
in APPENDIX 03.
The test method, equipment required and rejection limit are shown
in APPENDIX 04.
Refer to APPENDIX 06
The test method, equipment required and rejection limit are shown in
APPENDIX 06.
Refer to APPENDIX 07
The test method, equipment required and rejection limit are shown in
APPENDIX 07.
- 32 -
No. 2J313-002E*B
6. SYSTEM ACCURACY
I I 1
/ Maximum allowable
No. Technique factor Applicable range
deviation
I
I I
I
I
I
1 (Radiographic mA / all
I -+lO% I
1 ~ 27%
c
TRadiographic
3 / Radiographic
kV
1
all
7. ATTACHED DATA
PWBs which are not listed in 7.1 require no setting and adjustment when
they are installed. For the automatic exposure control PWB, refer to the
AEC-lOA/lOB installation manual.
- 33 -
No. 2J313-002E":D
(Panel)
PWB name: [PANEL CONT l] or
A: KXO-30R B: KXO-3OF/5OF
TW CONT] l/8
Equipment When
Setting and adjustment installed
B
Switch settings
I Baud rate
1 2 3 4
Equipment
i
/
B
I
i-
LON
OFF
EP
1 2
A
'&El
3 4 5
PU
6 7
Q
8
i
OFF: 90% Rating
100% Rating can be used for only testing; this switch shouid be
- 34 -
No. 2J313-002E9cD
Equipment
When
Setting
installed
A B
- 36 -
No. 2J313-002E>kD
ONP1
OFF
U P Cl 51 Q Q P
-2 3 4 5 6 7 8
L-
L
X-ray tube type (12
types max.>
Refer to next page.
I
l-i-4
SW No.
1 OFF
4
1 OFF
Upper limit
1 Max. -0kV
SW No.
--r---l
1
1 ON
2
Mode
1 ON 1 Automatic change-mode 1
I I - 37 -
No. 2J313-002E"N
ON OFF ON ON B DXB-0434ES(A256>(USA>
ON ON OFF ON D DRX-3724HD/HC(USA)
ON ON ON OFF E DRX-3624HC(USA)
ON ON ON ON F DXB-0324CS(A192)(USA)
ON OFF 2 DRX-3224HD-H
ON OFF ON ON B DXB-0434EScA256)
ON ON ON ON F DRX-1724B
Equipment
When
Setting
installed
A B
- 38 -
No. 2J313-002E"D
Equipment
Setting
A B
I i
-~
- 39 -
No. 2J313-002E*K
Equipment
When
Setting
installed
A B
SW setting Contents
- 40 -
No. 2J313-002E'kB
(Stand)
Equipment
When
Setting
installed
A
j B
Switch settings
Baud rate
1 4
0 Jumper Unnecessary
- 41 -
No. 2J313-002E*B
(Stand)
Equipment
When
Setting
installed
A B
0 Japan
1 USA
3
Undefined
z
F
- 42 -
No. 2J313-002E
Equipment
When
Setting
installed'
A 1 B
I
I I
I ,
I
I SW7 thru SW12: ’ Necessary
I o Setting the minimum time of mAs for
each TECHNIQUE
I
/These switches set minimum time of mAs for each TECHNIQUE.
Switch settings
Minimum time
(msec>
112 3 4 5 6 7 81
I
i /
I
1 : 1
ON / ON 1 ON /ON ON ON OFF ON / 253
I I 1
ON / ON / ON / ON 1 ON / ON ON OFF j 254
I
ON 1 ON 1 ON ON ON ON ON ON 255
- 43 -
No. 2J313-002E*K
Equipment
When
Setting
installed
A B
0 sw13: Unnecessary
5 6 7 8 Limitation
ON ON ON ON No limit up to 40kW
02 KXO-50F
SW NO.
5 6 7 8 Limitation
ON ON ON ON No limit up to 60kW
ON 00000000
OFF 12345678
I 1
Not us!d A ’ ” ‘l Not used
MAMMO
TOMOGRAPHY
- 44 -
NO. 2J313-002E*C
Equipment
When
Setting
installed
B
OOH 10
OlH 13
02H 16
03H 20
!
OFH 320
- 44A -
No. 23313-002EAG
Equipment When
Setting installed
A B
ON PO 0 0 P RRP
OFF 1 2 3 4 5 6 7 a
1 ’ 1 1 ’ A , 1 AL Unused ’
I
I__ All TECHNIQUES are
operated.
/ // I I
MAMMO TECHNIQUE
TOMOGRAPHY TECHNIQUE
SPOT TECHNIQUE
j I BKY2 TECHNIQUE
BKYl TECHNIQUE
GEN/CASSETTE TECHNIQUE
- 45 -
No. 2J313-002E*B
r
Equipment
Setting
When
installed
A B
6, 7 and 8 combi-
Contents
nations of SW13
1 . .. . . . . . .. Setting for SW1 of AU-APR unit
2 .......... Setting for SW2 of AU-APR unit
3 .......... Setting for SW3 of AU-APR unit
4 .......... Setting for SW4 of AU-APR unit
5 .......... Setting for SW5 of AU-APR unit
6 .......... Setting for SW6 of AU-APR unit
7 .......... Setting the maximum in 1.1. fluoroscopy mode
8 of SW28 Contents
0 .......... AU-APR spot radiography mode
1 .......... AU-APR fluoroscopy mode
SW32 ........... Data writing
SW17 combinations Contents
1 .......... Selecting the head
2 .......... Selecting the breast
3 .......... Selecting the abdomen
4 .......... Selecting the spinal cord
5 .......... Selecting the lumbar vertebrae
6 .......... Selecting the arms
7 .......... Selecting the legs
SW7 combinations Contents
1 to 10, 15 or 20 . Selecting an organ from those set with SW17
0 to 12 . . . . . . . . . . . Determining the maximum mA code for 1.1.
fluoroscopy
. Setting procedure
a First set the numeric on SW13 corresponding to the switch
number on the AU-APR unit to be set.
(@ Set SW28 (spot radiography/fluoroscopy), SW17 and SW7 to the
desired position/organ numbers.
3 Press the SW32 pushbutton switch (for write operation).
84 If the buzzer sounds, the setting is completed. If the buzzer
does not sound, a SW setting error has occurred, and go back
to @ .
@ Terminate the setting operation by setting 6, 7 and 8 of SW13
to "0".
@ After the setting operation, reset SW13, SW28, SW17 and SW7 tc
their original position.
- 46 -
No. 25313-OOZE"0
Equipment When
Setting installed
A B
0 SW32: Unnecessary
ON ON ON ON ON ON ON ON AUX.
- 47 -
No. 2J313-002EfcB
Equipment When
Setting
installed
A B
ONEl1
OFF
Q
2
P
3
P
4
u u
5 6
El P
7 8
ATR
MAMMO
1 Turn switches used
TOM0 for the individual
GENE l TECHNIQUES ON, and set
BKY2 the other switches to
BKYl OFF.
SPOT
NOTE
This switch can be used for 100% of OLP testing only
in the USA.
ON: 100% Rating
OFF: 90% Rating
100% Rating can be used for only testing; this switch
should be turned OFF after completing the 100% Rating
test.
- 48 -
No. 2J313-002EfcB
Equipment When
Setting
installed
- 49 -
No. 2J313-002E
Equipment
When
, Setting
installed
A B
- 50 -
No. 23313-002E
Equipment When
Setting
A B
L
installed
ON a E u u a u 0 E
OFF 1 2 3 4 5 6 7 8
J I
+
X-ray tube type (16
types max.>
Refer to next page.
I SW No. I
Mode
I
1
,2 I
- 51 -
No. 2J313-002E*P
Sw8: TUBE 1
sw9: TUBE 2
SWlO: TUBE 3 (for USA and CANADA)
ON ON OFF ON D DRX-3724HD/HC(USA)
ON ON ON OFF E DRX-3624HC(USA)
ON ON . ON ON F DXB-0324CS(A192>(USA>
SW8: TUBE 1
sw9: TUBE 2
SWlO: TUBE 3 (for other countries except USA and CANADA)
OFF ON ON ON 7 DRX-3745HD-H
ON OFF ON ON B DRX-2924HD
ON ON OFF ON D DRX-2703HD/HC
ON ON ON OFF E DRX-ll24B-MO
ON ON ON ON F DRX-1724B
1 i
- 52 -
No. 2J313-002E*F
Equipment
when
Setting
installed
A B
- 52-1 -
No. 25313-002E
When
installed
0 ! 0
1 ! 1 1
I 2 2 I
3 3
4 4 -Th is va lue is set
fore shipment.
F I 15sec. 1 I
- -I-
I
- 53 -
No. 2J313-002E:cA
(Control box)
Equipment
When
Setting
installed
A B,
0 1 Xl : Short 1
I
I Unnecessary
I
X26: X26- Al3 Selecting Ml ROM
1
27128
I
X27: X27- +5
Equipment' When
Setting
installed
q---J
- 54 -
No. 2J313-002EkD
Equipment When
Setting
installed
A
LRZ-D HRZ-F
LRl-C HRl-E
LTlw---+B HTZ-B
LT2-A HTl-A
Equipment
When
Setting
installed
Aj
When
Setting installed
- 55 -
No. 25313-OOZE
Equipment When
Setting and adjustment
installed
A B
0 0 ! Jumper Jl Unnecessar}
' ;Jl .
. Short (only for USA.)
Equipment When
Setting
installed
A B ’
I Necessary
0 I Jumpers Jl, 52: Setting TOMO, BUCKY selection
- 56 -
No. 2J313-002E"cL
Equipment When
Setting and adjustment installed
A B
SPOT Unused
BKYl
BKY2
GENE
TOM0
MAMMO
105mm spot
camera
When selecting the continuous radiography mode for
TECHNIQUES, turn the switches corresponding to the
individual TECHNIQUES ON.
, (The switches of three TECHNIQUES of SPOT, TOMO, and
105mm Spot Camera are set to ON before shipment.)
TUBE 2
- 57 -
No. 2J313-OOZE*M
Equipment When
Setting and adjustment installed
A B
F*kV
SW1 =A
SW2 =B
SW3 = c
SW4 = D
SW5 = E
F mA
When setting the switches. measure F.kV values
exceeding the limti of 2.58 x 10-3C/(kg=min)
(lOR/min} at individual mA values (binary), and set
them in the individual switches.
- 58 -
No. 2J313-002E*M
Equipment When
Setting and adjustment
installed
A B
125
110
0100
85
80
0 / ,946, sw7: Setting the dose limit in Fully Auto mode Necessary
/
~ These switches should be set for USA and other
1 countries where the dose limit is imposed.
/( SW7 requires no setting.)
- 59 -
No. ZJ313-002E$cB
Equipment
When
Setting
installed
A B
0 ’/ SW8, SW9: Setting the F.ABC parameter mode and the Necessary
full-auto mode
1
I
I
SWlO: F-ON test switch 1 Unnecessary
/ O
I ) The fluoroscopy can be done by this Test switch
I instead of the F-ON signal from the conventional
( R/F table.
- 60 -
No. 2J313-002E*L
Equipment
T Setting and adjustment
When
installed
B
Adjustment procedure
0 VR8: ABC gain adjustment VR for video signal type ABC Necessary
Adjust the ABC gain according to the system for video
requirements. signal
type ABC
I
- 61 -
No. 25313-002E
Equipment When
Setting
installed
A B 1
1 I
12 3 4 5 6 7 8
A h 4 ,I ih A
, ‘: ”
, i ; i
I I I ~ !
I 27 26 25 24 23 22 21 20
( ( I
[Example]
- 62 -
No. 25313-002E
E;i.,,;t 1
When
Setting
installed
,
I 1
I I SW2,6,10 (Jp$
““EIPPUUiiP :
12 34567 a,
/ h h m A A Ii 4
‘1
27 26 25 24 23 22 21 20
[Example]
Ic
Binary Hexa-
kV decimal
I .I 2’ 26 125 zL ! 2” 12’ I-9 12” ~ (H)
- 63 -
No. 2J313-002EfcB
-r---j
Equipment
A B
0
’
I
Setting
Necessary
- 64 -
No. 25313-OOZE
Equipment When
Setting and adjustment installed
A B
I
Adjust VR5 to set the P8 voltage to OV when F.mA=lmA. !
VR8 through VR13 adjustment (for spot film radiography)i
These VRs correct R.kV values for film division and
image intensifier (1.1.) magnification.
- 65 -
No. 25313-002EfcB
Equipment
When
Setting and adjustment
installed
A
When
Setting
installed
/
0 ! 0 j VR2: Frequency adjustment iunnecessary
r
I
I
o 1-0 / VR4: Offset adjustment IUnnecessary
When
Setting
installed
r 0
’
I
0 ’l/
I
VR12: D/A converter span j Unnecessary
I 0
I o j 128/256: 128 crossover span 1 Unnecessary
- 66 -
No. 2J313-002EkI
0 0 CNNA Unnecessary
- 67 -
No. 25313-002E*I
Equipment Adjustment
Setting and adjustment when
A B installed
0 0 I SW1 Necessary
POWER
Setting Position of SW1
source
19 2oov 0
I /
l$ 220v 1
3cP 380V 4
39 415v 5
39 440v 6
@ During charging
:
- 68 -
No. 25313-OOZE*1
Equipment Adjustment
Setting and adjustment when
A B installed
to 15m 3, 2, 1 or 0
16 to 20m 2, 1 or 0
21 to 25m 1 or 0
26 to 0
- 68A -
No. 2J313-002E*D
Equipment Adjustment
Setting and adjustment when
A B installed
- 69 -
No. 2J313-002Ey:B
Code , Meaning
1
.k _k
’ Transmission system error (Data communication between the panel/desk
and the box is not carried out correctly)
201 TECHNIQUE setting switch (on tghe panel or desk) is not set
202 Tube No. setting error (Tube setting switch on the panel or desk is
not set correctly)
301 TECHNIQUE setting switch (on the panel or desk) is not set
913 1 Normal starter error (Current was applied when the normal starter
/ was not ready)
I
96 Over kV (Output kV value has become greater than the specified value
98 Line voltage error (The power source voltage has become greater or
smaller than the specified value and X-rays cannot be generated.
- 70 -
No. 2J313-002E*B
8. MAINTENANCE SCHEDULE
APPENDICES
LIST
- 71 -
No. 2J313-OOZE*B
This data sheets are to be filled out and copied at the time of the
installation and maintenance. Record test results in copied sheets and return
to:
2. Installed by:
Branch Office:
Address:
3. Hospital:
Address:
0 DC-30R:
q DC-30F:
0 DC-50F:
17 DG-30F:
q DG-5OF:
Tube 1
Tube 2
Tube 3
Manufacturer:
Model name:
Serial No.:
- 72 -
No. 2J313-002E~kB
Manufacturer:
Model name:
Serial:
8 Power source
G 2OOV 0 22ov
0 38OV i? 415V 0 440v 3 48OV
Item Result
- 73 -
No. 2J313-002E*K
TUBE NO.
Tube model
lx” 71A
L kV mA kV mA kV mA kV mA
I
13
16
L I J
- 74 -
No. 2J313-OOZE*K
TUBE NO.
Tube model
120 140
kV mA kV mA
- 75 -
No. 2J313-002E*K
TUBE NO.
Tube model
kV mA kV mA kV mA kV mA kV mA
10
13
16
20
25
32
40
50
64
80
100
125
160
200
250
320
400
500
640
- 76 -
No. 2J313-002E"B
(2) R.time
111
+ 90kV lOOmA 1
0.2sec
Tube model
Table model
F. KV and F. mA
m&4 / mL4 mA
/ 1 I
I mA
mA mA
I 2
I mA ’ mA mA
I
j 3 /
mA mA InA I
4
- 77 -
No. 2J313-002E*B
X-ray tube:
Grid:
Intensifying screen:
Film:
mA correction 0.5mA kV
1mA 0 kV
2mA kV
3mA kV
4mA kV
Division l/2 kV
correction l/3 kV
l/4 kV
l/6 kV
Ml kV
M2 kV
- 78 -
No. 2J313-002E*B
._. -_ .-... . .
_ . ,
- . - ._ .
. _._-..__ _.
_ _. ..
._
_. _-.
._______.&_______ -- -----
-_--
. _.._-.
____L_~-.-
.
s-___._.-----_
i L :; ;_ I! ;_ __ .: .I - f F_ ATR.d.ata gheet:~‘;~___i_I~” :I : ::f _~~~:- -1. -f -_ --! --‘i t- I- :
-. ._. -.
. . _. .. __ __. ..- . . . __
: - _ _ __._. . . _.
:- :- .--. .- . _._ . . . _ _ .
T
Acrylic Radiographic conditions
Fluoroscopic conditions
phantom
12 cm kV mA kV mA
14
16
18
20
22
24
- 79 -
No. 23313-002E*M
b) ABC mode
limit F.mA . mA
limit F.kV kV
- 80 -
No. 2J313-002E*M
TUBE NO.
TUBE TYPE
kV 60 kV 60 kV 60 kV 60 kV 60 kV
mA InA mA mA mA mA
sec. S S S S S
Focus
I
Line regulation I -+% I -+% I -+% I -+% I -+% I
Exposure 1 mR mR mR mR mR
No.
2 mR mR mR mR mR
3 mR mR mR mR mR
4 mR mR mR mR mR
5 mR mR mR mR mR
6 mR mR mR mR mR
7 mR mR mR mR mR
8 mR mR mR mR mR
9 mR mR mR mR mR
10 mR mR mR mR mR
x mR mR mR mR mR
S I I I I I I
C
Xl +X2
(1 mR = 2.58 x 10-7C/kg)
- 81 -
No. 2J313-002E*M
TUBE NO.
TUBE TYPE
kV 90 kV 90 kV 90 kV 90 kV 90 kV
mA mA mA mA mA mA
sec. S S S S S
Focus
Line regulation +
- % +
- % -+ % +
- % -+ %
Exposure 1 mR mR mR mR mR
No.
2 mR mR mR mR mR
3 mR mR mR mR mR
4 mR mR mR mR mR
5 mR mR mR mR mR
6 mR mR mR mR mR
7 mR mR mR mR mR
8 mR mR mR mR mR
9 mR mR mR mR mR
10 mR mR mR mR mR
‘j;- mR mR mR mR mR
I)il - x21
Xl + x2
(1 mR = 2.58 x IO-7C/kg)
- 82 -
No. 2J313-002EkM
TUBE NO.
TUBE TYPE
mA mA mA mA mA mA
sec. S S S S S
Focus
Line regulation +
- % +
- % -+ % +
- % +
- %
Exposure 1 mR mR mR mR mR
No.
2 mR mR mR mR mR
3 mR mR mR mR mR
4 mR mR mR mR mR
5 mR mR mR mR mR
6 mR mR mR mR mR
7 mR mR mR mR mR
8 mR mR mR mR mR
9 mR mR mR mR mR
10 mR mR mR mR mR
x mR mR mR mR mR
Xl +X2
(1 mR = 2.58 x lo-7C/kg)
- 83 -
No. 2J313-002E*B
Measured time
1+2
1+3
2+3
/1+2+3/
- 84 -
No. 2J313-002E9cB
APPENDIX 02 RELATIONSHIP BETWEEN FOCAL SPOT SIZE AND TUBE CURRENT RANGE
Al92 (EIMK) 0.6 x 0.6mm 20 to 320mA 1.2 x 1.2mm 400 to 1OOOmA 40 to 150
kV
1
- 85 -
No, 2J313-002E*M
A. Objective
Fluoroscopic equipment which is provided with automatic exposure rate
control is not operable at any combination of tube potential and current
which results in an exposure rate in excess of 2.58 x lo-3C/ (kgemin)
{lOR/min} at the point where the center of the useful beam enters the
patient. (See the section 1020.32 (d).)
B. Test method
(1) The position of test equipment and diagnostic X-ray table assembly is
as follows.
I \ I i
3oomm Aluminum attenuation block
(200 x 200 x 19mm)
4Dosimeter 1
P
Beam-limiting device
X-ray tube
5. Place the lead sheet (300 x 300 x 2mm) at the center of the
X-ray field between the attenuation block and the imaging
system.
- 86 -
No. 2J313-002E*B
Beam-limiting device
7
_. If the source-to-image distance (SID) is variable, set the end
of the beam-limiting device as close to the tabletop as it can
be placed provided it shall not be closer than 30Omm.
5. Place the lead sheet (300 x 300 x 2mm) at the center of the
X-ray field on the tabletop.
- 57 -
No. 2J313-002E*M
4 - Imaging system
Beam-limiting device
t X-ray tube
2. Place the probe of the dosimeter 300mm (12in) from the input
surface of the imaging system.
3. Place the aluminum attenuation block (200 x 200 x 19mm) and the
lead sheet (200 x 200 x 2mm) at the center of the X-ray field as
shown in the above figure.
(3) Set the Rotary DIP SW, SW5 on the DESK CONTROL PWB in the control desk
to the position 0.
This setting means that the X-ray generator is excluded from the
entrance exposure rate limit.
(5) Select the spot-film radiography technique on the control panel, and
not select the “ABC” mode on the spot film device.
(6) Set the F.mA control for 4mA and F.kV control for 5OkV on the spot
film device.
(9) When the dosimeter reading reaches the limit level 23.22 x lo-*C/
(kgamin) {9R/min}, read the setting F.kV on the control panel and
record it.
(10) Repeat Steps (6) to (9) at the F.mA setting of 3, 2, 1 and 0.5mA.
- 88 -
No. 2J313-002E*M
(11) Set the limit kV which is recorded in Step (9) according to the Table
1 for each F.mA setting. The F.mA settings and the settings switch
numbers are shown below.
0.5 mA SW1
1mA SW2
2mA SW3
i
3mA SW4
4m.A SW5
(15) When the dosimeter reading reaches the limit level 23.22 x lo-4C/
(kgemin) {9R/min), read the indicated F.kV and F.mA on the control
panel and record them.
(16) Set the limit kV recorded above according to Table 1. Set the limit
mA recorded above according to Table 2. The setting switches are
shown below.
kV limit SW6
mA limit ! SW7
(17) Set the Rotary DIP SW, SW5 on the DESK CONTROL PWB in the control desk
to the position 1.
Note: When setting the DIP switch, the power should be off.
- 89 -
No. 2J313-002E*B
SW. No.
F.kV - 1 2 3 b 5 6 7 8 F.kV 1 2 3 ‘1. “;’ 6 7 8
I
50 0 0 1 1 0 0 1 0 88 o/101 1 ,
IO
t I I , I I 1 I I ,oo /
’
53 o/o l/1lOll:o / , j ( I/ 91 011 ! 0 1 1 IO L jl
60 0 0 1 1(1,1
’ i j 0 0 198 011 1/00/0/1~0
62 0 I 0 1 1 / 1 ( 1 / 1 0 1 100 1 0 / 1 / 1 1 0 j 0 1 1 1 0 10
[ 69 10 im l~o;O,o ~] ! : 0 I 1 1 1o7 / 0 / 1 1 1 1 0 / 1 /
0 / l/ 1 j 0 1 108 1 0 / 1 / 1 / 0 / 1 / ljOlO(
1 ( I
73 0 1 / 0 01 l ojo 111 0 1 1 0 1 I 1 1 1 1
75 0 ! 1 I 0 011 0 1
79 0 1 O/l 1 101110/l
1
80 0 1 / O/ 1 / 0 / 0 10 0 / 118 0 1 11 1 O/ 1 / 11 0
f
81 0 10 1;O 0 0 11 119 0 1 1 1 o/1/+
- 90 -
No. 2J313-002E*K
0.2 0 0 0 0 0 0 1 0 2.6 0 0 0 1 1 0 1 0
0.3 0 0 0 0 0 0 1 1 2.7 0 0 0 1 1 0 1 1
0.4 0 0 0 0 0 1 0 0 2.8 0 0 0 1 1 1 0 0
0.5 0 0 0 0 0 1 0 1 2.9 0 0 0 1 1 1 0 1
0.6 0 0 0 0 0 1 1 0
0.7 0 0 0 0 0 1 1 1 3.1 0 0 0 1
0.8 0 0 0 0 1 0 0 0 3.2 0 0 1 0
0.9 0 0 0 0 1 _O 0 1 3.3 0 0 1 0 0 0 0 1
1.0 0 0 0 0 I 0 1 0 3.4 0 0 I 0 0 0 I 0
1
1.1 0 0 0 0 1 0 1 1 1 3.5 / 0 / 0 1 1 j 0 1 0 ( 0 / 11 1
I I I I
1.2 0 0 0 0 1 1 0 0 3.6 0 0 1 0 0 1 0 0
LL
1.3
1.4
0
0
0
0
0
0
0
0
I
1
1
1
1
I
0
1
1
o( 3.8 1 0 1 0 1 1 1 0 1 0 1 1 1 1
1.5 0 0 0 0 1 1 1 1
1.6 0 0 0 1 0 0 0 0 4.0 0 0 1 0 1 0 0 0
1.7 0 0 0 1 0 0 0 1
1.8 0 0 0 1 0 0 1 0
1.9 0 0 0 1 0 0 1 1
2.0 0 0 0 1 0 1 0 0
2.1 0 0 0 1 0 1 0 1
2.2 0 0 0 1 0 1 1 0
2.3 0 0 0 1 0 1 1 i
2.4 0 0 0 1 1 0 0 0
2.5 0 0 0 1 1 0 0 1
1: ON, 0: OFF
Note: When mA is more than 4.OmA, turn on all switches.
- 91 -
No. 2J313-002E*M
(18) After adjustment of the entrance exposure rate limit, repeat the Steps
(6) to (8) for the F.mA setting of 4, 3, 2, 1 and 0.5mA and record the
tube voltage and the dosime ter readings at the limit points in manual
mode
(19) Perform fluoroscopy in the FULL-AUTO mode and record the tube voltage,
current and dosimeter readings at the limit point.
(1) Dosimeter
This measurement requires one of the following test instruments:
b. Victreen model 550 with 550-3, 550-4, 550-5 or 550-6 probe (Refer
to TOSHIBA APPENDIX 026A.)
C. Victreen model 555 with lmA, 1OmA or 1DAS chamber (Refer to TOSHIBA
APPENDIX 008A.j
D. Rejection limit
- 92 -
No. 2J313-002E*B
A. Objective
B. Test method
(1) Set the X-ray apparatus and test instrument according to the
following block diagram.
s Dosimeter
Focus detector
distance: 1OOcm
Collimator
Primary
voltmeter
(2) Turn on the equipment line switch and calibrate test instruments.
(3) Set the technique factors (radiographic kV, mA and time selectors)
and during the 10 consecutive measurements, at intervals of one or
two minutes the technique factors are adjusted to alternate settings
and reset the test setting after each measurement.
(4) While confirming that the respective percent line voltage regulations
are within +l% of the mean value for all the 10 measurements, record
the value of a dosimeter.
- 93 -
No. 2J313-002E*M
(6) Confirm that the above product “C” is not greater than 0.045.
(1 mR = 2.58 x lo-TC/kg)
(8) Confirm that the above product is not greater than 0.09 times.
Dosimeter
Beam
limiting device
__---- ___------
-----_____
-----_
Aluminum attenuation
block (200 x 200 x 38mm)
Primary
voltmeter
Power
H.P. - X-ray
Igenerator ] ( control
(2) Set the cassette with film (Sakura A type) and screen (Toshiba EM
type) posterior to the detector as shown in the above figure.
(3) Set the tube voltage for 90kV, the tube current for lOOmA and the
film density regulator for “0” posit ion and make an exposure.
(4) Adjust the AEC so that the film density of 0.85 to 1.15 is got.
(5) Adjust the beam limiting device that the exposure field conforms to
the inside size of the attenuation block.
- 94 -
No. ZJ313-002E$cB
Density
/ X-ray conditions (Tube voltage, Tube current, Exposure time)
I positions ,
- 95 -
No. 2J313402E*B
D. Rejection limit
2 0.09
Xl + x2
- 96 -
No. 2J313-002E+B
1. General
60kV ’
1OmA I-
3.0sec 1 13.0sec ] 13.0sec 1
1 1
Setting Resetting
- 70kV 60kV
I-
13mA /- 1OmA _I 1OmA -*
2.31sec I 3.0sec / / 3.0sec
/
(Alternate (Test (Alternate (Test
setting(2)) setting) setting(l)) setting)
Test Procedure
- 97 -
No. 2J313-OOZE*K
No. 1 2 3
4 6 7
No. 8 9
T 10 11 12 13 14
No. 15 16 17 18 19
- 98 -
No. 2J313-002E*B
A. Objective
To determine that the product of X-ray tube current and exposure time is
limited to not more than 6OOmAs per exposure when an automatic exposure
control is provided.
B . Test method
Power
SUPPl X-ray
tube
--& \)
High-voltage
cables
I 4
Exposure KV indi-
time cating
counter unit
(TD-3/-3A) (KV-201)
(2) Turn on the main circuit breaker and press the POWER ON pushbutton.
(3) Press the technique switch which can use the automatic exposure
control.
(6) Remove the detector for automatic exposure control outside the
X-ray field.
(7) Make an exposure and record the mA reading and exposure time when
back up operates.
- 99 -
No. 2J313-OOZE$cB
(8) Verify that the product of indicated value is not over 540mAs.
(2) Exposure time : ALCO Model TD-3/-3A (Refer to TOSHIBA APPENDIX 018A.)
counter
(3) High voltage : ALCO Model KV-201 (Refer to TOSHIBA APPENDIX 003A.)
meter
D. Rejection limit
The product of tube current and exposure time measured on the mA meter and
the exposure time counter must not exceeded the following limits:
- 100 -
No. 2J313-002E*B
A. Objective
When the X-ray tube potential is equal to or greater than 50kV, the minimum
exposure time for field emission equipment rated for pulsed operation shall
be equal to or less than a time interval equivalent to two pulses, and the
minimum exposure time for all other equipment shall be equal to or less
than l/60 second or a time interval required to deliver SmAs, whichever is
greater (See 102.31(a)(3)(ii)).
B. Test method
(1) Set the X- ray apparatus and test instruments according to Fig. 2.
X-ray detector
I
I Beam limiting device Acrylic phantom (5cm)
X-ray tube
Power
SUPPlY
I I
I
(3) Set the tube voltage for 90kV, the tube current for lOOmA and the film
density regulator for “0” position.
(4) Set the acrylic phantom (thickness : 5cm) in front of the detector.
In the case of the detector with multiple detecting fields, test should
be taken to each combination of detecting fields. For example, test
should be done under 7 conditions (23 - 1 = 7) for the detector with 3
fields.
(6) Confirm that the exposure time is within the rejection limits.
- 101 -
C. Test eqci?men: reqE:lrea
D. Rejection limit
The exposure time measured on the exposure time counter must no: exceed the
following limits.
- 102 -
AP?ENDIX OS TEST TOQLS AND INSTRUMENTS REQUIRED
Test equipment
Remarks
Item Type
I
AC Voltmeter Full range: 300/6OOV
The following test equipment are necessary cnly for HHS test in addition to
the above equipment.
'Test equipment
Remarks
Item Type
- 103 -
TOSHIBA
TOSHiBA CORPORATlON
1385,SHIMOISHIGAMI,OTAWARA-SHI,TOCHiGl-KEN 324-8550,JAPAN