You are on page 1of 26

BRITISH STANDARD PD 6519-2:

1988
Licensed Copy: Institute Of Technology Tallaght, Institute of Technology, Tue Apr 03 04:04:41 GMT+00:00 2007, Uncontrolled Copy, (c) BSI

IEC 479-2:
1987
Incorporating
Amendment No. 1

Guide to

Effects of current on
human beings and
livestock —
Part 2: Special aspects relating to
human beings

UDC [612.795 + 612.014.42]:621.3.011.21.014.12:614.826:[621.3.025 + 621.3.029.45]:001.4


PD 6519-2:1988
Licensed Copy: Institute Of Technology Tallaght, Institute of Technology, Tue Apr 03 04:04:41 GMT+00:00 2007, Uncontrolled Copy, (c) BSI

Committees responsible for this


Published Document

This Published Document was prepared by the General Electrotechnical


Engineering Standards Committee upon which the following bodies were
represented:

Associated Offices Technical Committee


Association of Control Manufacturers (TACMA) (BEAMA Ltd.)
BEAMA Ltd.
British Approvals Service for Electric Cables
British Cable Makers’ Confederation
British Radio and Electronic Equipment Manufacturers’ Association
Department of Energy (Electricity Division)
Department of Trade and Industry (National Physical Laboratory)
Department of Trade and Industry (Radiocommunications Division)
Department of Trade and Industry [Shipbuilding and Electrical Engineering
Division (SEE)]
Electrical Contractors’ Association
Electrical Contractors’ Association of Scotland
Electrical Installation Equipment Manufacturers’ Association (BEAMA Ltd.)
Electricity Supply Industry in England and Wales
Electronic Components Industry Federation
Electronic Engineering Association
Energy Industries Council
Engineering Equipment and Materials Users’ Association
ERA Technology Ltd.
GAMBICA Ltd.
Health and Safety Executive
Institution of Electrical Engineers
Ministry of Defence
Trades Union Congress

This Published Document,


having been prepared under
the direction of the General
Electrotechnical Engineering
Standards Committee, was
published under the authority
of the Board of BSI and comes Amendments issued since publication
into effect on
29 April 1988
Amd. No. Date of issue Comments
© BSI 03-1999
8649 June 1995 Indicated by a sideline in the margin
The following BSI reference
relates to the work on this
Published Document:
Committee reference GEL/-

ISBN 0 580 16454 3


PD 6519-2:1988
Licensed Copy: Institute Of Technology Tallaght, Institute of Technology, Tue Apr 03 04:04:41 GMT+00:00 2007, Uncontrolled Copy, (c) BSI

Contents

Page
Committees responsible Inside front cover
National foreword ii
Chapter 4: Effects of alternating current with frequencies above 100 Hz
1 General 1
2 Scope 1
3 Definitions 1
4 Effects of alternating current in the frequency
range above 100 Hz up to and including 1 000 Hz 1
5 Effects of alternating current in the frequency
range above 1 000 Hz up to and including 10 000 Hz 1
6 Effects of alternating current in the frequency
range above 10 000 Hz 1
Chapter 5: Effects of special waveforms of current
1 General 4
2 Scope 4
3 Definitions 4
4 Effects of alternating current with d.c. components 4
5 Effects of alternating current with phase control 5
6 Effects of alternating current with multicycle control 6
Chapter 6: Effects of unidirectional single impulse currents of
short duration
1 General 10
2 Scope 10
3 Definitions 10
4 Effects of unidirectional impulse currents of short duration 11
Figure 9 — Variation of the threshold of perception within
the frequency range 50/60 Hz to 1 000 Hz 2
Figure 10 — Variation of the threshold of let-go within the
frequency range 50/60 Hz to 1 000 Hz 2
Figure 11 — Variation of the threshold of ventricular
fibrillation within the frequency range 50/60 Hz to 1 000 Hz,
shock-durations longer than one heart period and longitudinal
current paths through the trunk of the body 3
Figure 12 — Variation of the threshold of perception within
the frequency range 1 000 Hz to 10 000 Hz 3
Figure 13 — Variation of the threshold of let-go within
the frequency range 1 000 Hz to 10 000 Hz 3
Figure 14 — Waveforms of currents 7
Figure 15 — Waveforms of rectified alternating currents 7
Figure 16 — Waveforms of alternating currents with phase control 8
Figure 17 — Waveforms of alternating currents with multicycle control 8
Figure 18 — Threshold of ventricular fibrillation (average values)
for alternating current with multicycle control for various degrees
of power control (results of experiments with young pigs) 9
Figure 19 — Forms of current for rectangular impulses,
sinusoidal impulses and for capacitor discharges 14

© BSI 03-1999 i
PD 6519-2:1988
Licensed Copy: Institute Of Technology Tallaght, Institute of Technology, Tue Apr 03 04:04:41 GMT+00:00 2007, Uncontrolled Copy, (c) BSI

Page
Figure 20 — Rectangular impulse, sinusoidal impulse and
capacitor discharge having the same specific fibrillating energy
and the same shock-duration 14
Figure 21 — Threshold of perception and threshold of pain for
capacitor discharges (dry hands, large contact areas) 15
Figure 22 — Threshold of ventricular fibrillation 16
Bibliography 17
Publications referred to Inside back cover

ii © BSI 03-1999
PD 6519-2:1988
Licensed Copy: Institute Of Technology Tallaght, Institute of Technology, Tue Apr 03 04:04:41 GMT+00:00 2007, Uncontrolled Copy, (c) BSI

National foreword

This Published Document, which has been prepared under the direction of the
General Electrotechnical Engineering Standards Committee, is identical with the
Report of the International Electrotechnical Commission, IEC
Publication 479-2:1987 “Effects of current on human beings and livestock.
Part 2: Special aspects relating to human beings”. Although the content of the
IEC Publication is not judged suitable for publication as a British Standard, its
status in the IEC of “Basic safety standard” (see IEC Guide 104:1984 “Guide to
the drafting of safety standards, and the role of committees with safety pilot
functions and safety group functions”) is likely to result in many references to it
being made in other IEC standards concerned with safety and it is therefore
thought to be of interest to members of appropriate Technical Committees and to
those concerned with development of products. In particular, the number of
experiments on which the data in IEC 479 were based was stastically small and
there may have been a bias in respect of age, health and racial characteristics.
The most conservative figures in IEC 479 cover 95 % of the adult population so,
where safety of persons is a consideration, a substantial safety factor should be
applied. Attention is drawn to the introduction in PD 6519-1 which indicates
ways in which the PD should be used.
PD 6519 is published in two Parts. The other Part is:
— Part 1 General aspects
Terminology and conventions. The text of IEC 479-2 has been approved as
suitable for publication as a Published Document without deviation. Some
terminology and certain conventions are not identical with those used in British
Standard publications; attention is drawn especially to the following.

Cross-referencesa

International standard Corresponding British Standard

IEC 50 BS 4727: Glossary of electrotechnical, power,


telecommunication, electronics, lighting and colour
terms
IEC 50(551):1982 Part 2:Group 02:1983 Power electronics terminology
(Technically equivalent)
IEC 50(801):1984 Part 3:Group 08:1985 Acoustics and electroacoustics
terminology
(Technically equivalent)
a
Undated in the text.

A British Standard does not purport to include all the necessary provisions of a
contract. Users of British Standards are responsible for their correct application.
Compliance with a British Standard does not of itself confer immunity
from legal obligations.

Summary of pages
This document comprises a front cover, an inside front cover, pages i to iv,
pages 1 to 18, an inside back cover and a back cover.
This standard has been updated (see copyright date) and may have had
amendments incorporated. This will be indicated in the amendment table on
the inside front cover.

© BSI 03-1999 iii


Licensed Copy: Institute Of Technology Tallaght, Institute of Technology, Tue Apr 03 04:04:41 GMT+00:00 2007, Uncontrolled Copy, (c) BSI

iv
blank
PD 6519-2:1988
Licensed Copy: Institute Of Technology Tallaght, Institute of Technology, Tue Apr 03 04:04:41 GMT+00:00 2007, Uncontrolled Copy, (c) BSI

Chapter 4: Effects of alternating 4.2 Threshold of let-go


current with frequencies above 100 Hz For the threshold of let-go the frequency factor is
given in Figure 10.
1 General 4.3 Threshold of ventricular fibrillation
Electric energy in the form of alternating current of For shock-durations longer than the cardiac cycle,
higher frequencies than 50/60 Hz is increasingly the frequency factor for the threshold of fibrillation
used in modern electrical equipment, for example for longitudinal current paths through the trunk of
aircraft (400 Hz), power tools and electric welding the body is given in Figure 11.
(mostly up to 450 Hz), electrotherapy (using
mostly 4 000 Hz to 5 000 Hz), switching mode For shock-durations shorter than the cardiac cycle
power supplies (20 kHz to 1 MHz). no experimental data is available.
Little experimental data is available for this 5 Effects of alternating current in the
chapter, so that the information given herein should frequency range above 1 000 Hz up to
be considered as provisional only but may be used
for the evaluation of risks in the ranges of and including 10 000 Hz
frequencies concerned (see bibliography). Attention 5.1 Threshold of perception
is also drawn to the fact, that the impedance of For the threshold of perception the frequency factor
human skin decreases approximately inversely is given in Figure 12.
proportional to the frequency for touch voltages in
the order of some tens of volts, so that the skin 5.2 Threshold of let-go
impedance at 500 Hz is only about one tenth of the For the threshold of let-go the frequency factor is
skin impedance at 50 Hz and may be neglected in given in Figure 13.
many cases. This holds even more true for higher 5.3 Threshold of ventricular fibrillation
frequencies. The impedance of the human body at
such frequencies is therefore reduced to its internal Under consideration.
impedance Zi (see Chapter 1).
6 Effects of alternating current in the
2 Scope frequency range above 10 000 Hz
This chapter describes the effects of sinusoidal 6.1 Threshold of perception
alternating current within the frequency ranges: For frequencies between 10 kHz and 100 kHz the
— above 100 Hz up to and including 1 000 Hz threshold rises approximately from 10 mA
(see Clause 4); to 100 mA (r.m.s. values).
— above 1 000 Hz up to and including 10 000 Hz For frequencies above 100 kHz the tingling
(see Clause 5); sensation characteristic for the perception at lower
— above 10 000 Hz (see Clause 6). frequencies changes into a sensation of warmth for
current intensities in the order of some hundred
3 Definitions milliamperes.
In addition to the definitions given in Part 1, the 6.2 Threshold of let-go
following definition applies: For frequencies above 100 kHz there is neither
3.1 experimental data nor reported incidents
frequency factor Ff concerning the threshold of let-go.
ratio of the threshold current for the relevant 6.3 Threshold of ventricular fibrillation
physiological effects at the frequency f to the For frequencies above 100 kHz there is neither
threshold current at 50/60 Hz experimental data nor reported incidents
NOTE The frequency factor differs for perception, let-go and concerning the threshold of ventricular fibrillation.
ventricular fibrillation.
6.4 Other effects
4 Effects of alternating current in the Burns may occur at frequencies above 100 kHz and
frequency range above 100 Hz up to current magnitudes in the order of amperes
and including 1 000 Hz depending on the duration of the current flow.

4.1 Threshold of perception


For the threshold of perception the frequency factor
is given in Figure 9.

© BSI 03-1999 1
PD 6519-2:1988
Licensed Copy: Institute Of Technology Tallaght, Institute of Technology, Tue Apr 03 04:04:41 GMT+00:00 2007, Uncontrolled Copy, (c) BSI

Figure 9 — Variation of the threshold of perception within the frequency


range 50/60 Hz to 1 000 Hz

Figure 10 — Variation of the threshold of let-go within the frequency


range 50/60 Hz to 1 000 Hz

2 © BSI 03-1999
PD 6519-2:1988
Licensed Copy: Institute Of Technology Tallaght, Institute of Technology, Tue Apr 03 04:04:41 GMT+00:00 2007, Uncontrolled Copy, (c) BSI

NOTE For shock-durations shorter than one heart period, other curves are under consideration.
Figure 11 — Variation of the threshold of ventricular fibrillation within the frequency
range 50/60 Hz to 1 000 Hz, shock-durations longer than one heart period and
longitudinal current paths through the trunk of the body

Figure 12 — Variation of the threshold of Figure 13 — Variation of the threshold of


perception within the frequency let-go within the frequency range 1 000 Hz
range 1 000 Hz to 10 000 Hz to 10 000 Hz

© BSI 03-1999 3
PD 6519-2:1988
Licensed Copy: Institute Of Technology Tallaght, Institute of Technology, Tue Apr 03 04:04:41 GMT+00:00 2007, Uncontrolled Copy, (c) BSI

Chapter 5: Effects of special 3.4


waveforms of current multicycle control factor p
the ratio between the number of conducting cycles
1 General and the sum of conducting and non- conducting
The increasing interest in special waveforms of cycles in the case of multicycle control
current derived from alternating current and direct (see Figure 17)
current is evidenced by the rising number of
applications of electronic controls causing such 4 Effects of alternating current with
types of current particularly in the case of an d.c. components
insulation fault. This holds true also for equipment 4.1 Waveforms and frequencies
using alternating currents with phase control and
Figure 14, shows typical waveforms which are dealt
multicycle control.
with in this clause. Pure d.c. and pure a.c. are
As is to be expected the effects of such currents on represented as well as combined waveforms of
the human body are between those caused by direct various ratios a.c. to d.c. The following current
and by alternating current; therefore equivalent magnitudes have to be distinguished:
current magnitudes with regard to ventricular
fibrillation can be established. Irms = r.m.s. value of the current of the
resultant waveform,
2 Scope Ip = peak value of the current of the
This chapter describes the effects of current passing resultant waveform,
through the human body for: Ipp = peak-to-peak value of the current of the
— alternating sinusoidal current with d.c. resultant waveform,
components, Iev = r.m.s. value of a sinusoidal current
— alternating sinusoidal current with phase presenting the same risk as regards
control, ventricular fibrillation as the waveform
— alternating sinusoidal current with multicycle concerned.
control. NOTE The current Iev is used instead of the current IB
NOTE Other waveforms are under consideration. in Figure 5 of Chapter 2 to estimate the risk of ventricular
fibrillation.
The information given is deemed applicable for
alternating current frequencies from 15 Hz up 4.2 Threshold of perception
to 100 Hz. The threshold of perception depends on several
parameters such as the area of the body in contact
3 Definitions with an electrode (contact area), the conditions of
In addition to the definitions given in Part 1, the contact (dry, wet, pressure, temperature) and also
following ones apply for the purpose of this chapter: on physiological characteristics of the individual.
3.1 Values for the threshold of perception are under
phase control consideration.
the process of varying the instant within the cycle at 4.3 Threshold of let-go
which current conduction begins The threshold of let-go depends on several
3.2 parameters, such as the contact area, the shape and
phase control angle (current delay angle) size of the electrodes and also on the physiological
characteristics of the individual.
the time expressed in angular measure by which the
Values for the threshold of let-go are under
starting instant of current conduction is delayed by
consideration.
phase control
3.3
multicycle control
the process of varying the ratio of the number of
cycles which include current conduction to the
number of cycles in which no current conduction
occurs

4 © BSI 03-1999
PD 6519-2:1988
Licensed Copy: Institute Of Technology Tallaght, Institute of Technology, Tue Apr 03 04:04:41 GMT+00:00 2007, Uncontrolled Copy, (c) BSI

4.4 Threshold of ventricular fibrillation b) For durations shorter than 0.75 times the
4.4.1 Waveforms consisting of specific ratios of period of the cardiac cycle:
alternating to direct current I pp Ip
I ev = -------
- = ------
-
The fibrillation hazard may be taken as being 2 2
approximately the same as with an equivalent
Hence for half wave rectification Iev is related to
alternating current Iev having the following
the r.m.s. value of the rectified current Irms by:
characteristics:
a) For shock durations longer than I ev = 2 I rms
approximately 1.5 times the period of the cardiac and for full wave rectification by:
cycle, Iev is the r.m.s. value of a current having
the same peak-to-peak value Ipp as the current of Iev = Irms
the waveform concerned:
I pp
5 Effects of alternating current with
I ev = ----------
- phase control
2 2
5.1 Waveforms and frequencies
b) For shock durations shorter than
Figure 16, shows the waveforms for symmetrical
approximately 0.75 times the period of the
and asymmetrical control.
cardiac cycle, Iev is the r.m.s. value of a current
having the same peak value Ip as the current of 5.2 Threshold of perception and threshold of
the waveform concerned: let-go
Ip As described in the preceding Sub-clauses 4.2
I ev = ------
- and 4.3, these thresholds depend on different
2 parameters.
NOTE This correlation is the less applicable the smaller the
ratio a.c. to d.c. becomes. For pure d.c. shocks of a duration
The effect of the current in producing sensation or
less than 0.1 s the threshold is equal to the corresponding inhibiting let-go is about equal to a pure a.c. with
r.m.s. value of the alternating current (see Figure 5 and the same peak value Ip. For phase control angles
Figure 8 in Chapter 2 and Chapter 3 respectively). above 120° the peak values increase as a
c) In the duration range from 0.75 to 1.5 times the consequence of the decreasing duration of the
period of the cardiac cycle the amplitude current flow.
parameter changes from peak value to
5.3 Threshold of ventricular fibrillation
peak-to-peak value.
NOTE The details of the nature of the transition that takes
The thresholds differ for symmetrical and
place are subject to further studies. asymmetrical waveforms.
4.4.2 Examples of rectified alternating current 5.3.1 Symmetrical control
Figure 15, shows the waveforms for half wave and The fibrillation hazard may be taken as being
full wave rectification. For these waveforms the approximately the same as with equivalent
peak value of the current is identical with its alternating current Iev having the following
peak-to-peak value. characteristics:
The equivalent alternating current Iev is a) for shock-durations longer than
determined: approximately 1.5 times the period of the cardiac
a) For durations longer than 1.5 times the period cycle, Iev has the same r.m.s. value as the current
of the cardiac cycle by: of the relevant waveform concerned;
I pp Ip b) for shock-durations shorter than
I ev = ----------
- = ----------
- approximately 0.75 times the period of the
2 2 2 2 cardiac cycle, Iev is the r.m.s. value of a current
Hence for half wave rectification Iev is related to having the same peak value as the current of the
the r.m.s. value of the rectified current Irms by: relevant waveform concerned;
I rms
I ev = ----------
-
2
and for full wave rectification by:
I rms
I ev = ----------
-
2

© BSI 03-1999 5
PD 6519-2:1988
Licensed Copy: Institute Of Technology Tallaght, Institute of Technology, Tue Apr 03 04:04:41 GMT+00:00 2007, Uncontrolled Copy, (c) BSI

NOTE For phase control angles above 120° a rise of the 6.3 Threshold of ventricular fibrillation
threshold of fibrillation is to be expected.
Depending on the duration of shock and on the
c) in the duration range from 0.75 to 1.5 times the
degree of power control alternating currents with
period of the cardiac cycle, the amplitude
multicycle control are equally or less dangerous
parameter changes from peak to r.m.s. value.
than a.c. of the same shock duration and current
NOTE The details of the nature of the transition that takes
place are subject to further studies.
magnitude.
5.3.2 Asymmetrical control Figure 18, shows the threshold of ventricular
fibrillation measured on pigs for various degrees of
The fibrillation hazard may be taken as being power control.
approximately the same as with an equivalent
6.3.1 For shock-durations longer than
alternating current Iev having the following
approximately 1.5 times the period of the cardiac
characteristics:
cycle, the threshold depends on the degree of power
a) for shock-durations longer than control p. For p near unity it has the same r.m.s.
approximately 1.5 times the period of the cardiac value as a sinusoidal alternating current of the
cycle: Under consideration. same duration. For p near 0.1 the r.m.s. value of the
b) for shock-durations shorter than current during current conduction I1 rms is the same
approximately 0.75 times the period of the as the threshold for alternating current of a
cardiac cycle, Iev is the r.m.s. value of a current duration below 0.75 times the period of the cardiac
having the same peak value as the current of the cycle.
relevant waveform concerned. NOTE For intermediate values of p, the fibrillation threshold
NOTE 1 For phase control angles above 120° a rise of the rises from the low level shown in Figure 5 of Part 1 to the high
threshold of fibrillation is to be expected. level indicated for shock-durations below 0.1 s.
NOTE 2 Currents caused by asymmetrical control 6.3.2 For shock-durations shorter than
(see IEV 551-05-19)1) may also have d.c. components. approximately 0.75 times the period of the cardiac
cycle the r.m.s. value of the current during current
6 Effects of alternating current with conduction I1 rms is the same as that for a sinusoidal
multicycle control alternating current of the same duration.
6.1 Waveforms and frequencies
Figure 17, shows the waveforms for a degree of
power control of p = 0.67.
6.2 Threshold of perception and threshold of
let-go
As described in the preceding
Sub-clauses 4.2, 4.3, 5.2 and 5.3, these thresholds
depend on different parameters.
The threshold of perception and threshold of let-go
are under consideration.

1)
IEC Publication 50 (551): International Electrotechnical Vocabulary (IEV), Chapter 551: Power Electronics.

6 © BSI 03-1999
Licensed Copy: Institute Of Technology Tallaght, Institute of Technology, Tue Apr 03 04:04:41 GMT+00:00 2007, Uncontrolled Copy, (c) BSI

© BSI 03-1999
Figure 14 — Waveforms of currents

Figure 15 — Waveforms of rectified alternating currents

7
PD 6519-2:1988
PD 6519-2:1988
Licensed Copy: Institute Of Technology Tallaght, Institute of Technology, Tue Apr 03 04:04:41 GMT+00:00 2007, Uncontrolled Copy, (c) BSI

Figure 16 — Waveforms of alternating currents with phase control

Ip
I1 rms = ------
- = r.m.s. value of current during current conduction
2
NOTE I1 rms is not to be confused with the r.m.s. value of current during working period I 2 rms = I 1 rms p .
Figure 17 — Waveforms of alternating currents with multicycle control

8 © BSI 03-1999
PD 6519-2:1988
Licensed Copy: Institute Of Technology Tallaght, Institute of Technology, Tue Apr 03 04:04:41 GMT+00:00 2007, Uncontrolled Copy, (c) BSI

NOTE Body current IB rms is the r.m.s. value of the current during current conduction I1 rms.
Figure 18 — Threshold of ventricular fibrillation (average values) for alternating current
with multicycle control for various degrees of power control (results of experiments
with young pigs)

© BSI 03-1999 9
PD 6519-2:1988
Licensed Copy: Institute Of Technology Tallaght, Institute of Technology, Tue Apr 03 04:04:41 GMT+00:00 2007, Uncontrolled Copy, (c) BSI

Chapter 6: Effects of unidirectional NOTE Fe is determined by the form of the impulse as the
integral
single impulse currents of short
duration
1 General Fe multiplied by the body resistance gives the energy dissipated
Unidirectional single impulse currents of short in the human body during the impulse.
duration in the form of rectangular and sinusoidal 3.2
impulses or capacitor discharges may be a source of specific fibrillating charge Fq (C or As)
danger in the case of an insulation fault of an the minimum It value of a unidirectional impulse of
electric appliance containing electronic components short duration which under given conditions
or when touching live parts of such equipment. It is (current-path, heart-phase) causes ventricular
therefore important to establish the danger limits fibrillation with a certain probability
for these types of currents.
NOTE Fq is determined by the form of the impulse as the
For a shock-duration of 10 ms the effects described integral
in this chapter correspond to those given in
Chapters 2 to 5 so that IEC Publication 479 covers
the whole range of shock-durations from 0.1 ms
to 10 s for nearly all current waveforms which are of 3.3
technical interest. The content of this chapter is time constant
based on the assumption derived from scientific
research that the principal factor for the initiation of the time required for the amplitude of an
ventricular fibrillation for the various forms of exponentially decaying field quantity to decrease
1
unidirectional impulse currents is the It or the I2t to --e- = 0.3679 times an initial amplitude
value as for shocks of up to 10 ms duration (IEV 801-01-44)2)
(see Bibliography). 3.4
shock-duration of a capacitor discharge (ti)
2 Scope the time interval from the beginning of the
This chapter describes the effects of current passing discharge to the time when the discharge current
through the human body in the form of single has fallen to 5 % of its peak value
unidirectional rectangular impulses, sinusoidal NOTE When the time constant of the capacitor discharge is
impulses and impulses resulting from capacitor given by T the shock-duration of the capacitor discharge is equal
discharges. to 3 T. During the shock-duration of the capacitor discharge
practically all the energy of the impulse is dissipated.
NOTE The effects of sequences of impulses are under
consideration. 3.5
The values specified are deemed to be applicable for threshold of perception
impulse durations from 0.1 ms up to and the minimum value for the charge of electricity
including 10 ms. For impulse durations longer which under given conditions causes any sensation
than 10 ms the values given in Figure 5 of to the person through whom it is flowing
Chapter 2 apply. 3.6
threshold of pain
3 Definitions
The maximum value of charge (It) or specific energy
In addition to the definitions given in Chapters 2 (I2t) that can be applied as an impulse to a person
to 5, the following ones apply for the purpose of this holding a large electrode in the hand without
chapter: causing pain
3.1 3.7
specific fibrillating energy Fe (Ws/7 or A2s) pain
the minimum I2t value of a unidirectional impulse of an unpleasant experience such that it is not readily
short duration which under given conditions accepted a second time by the subject submitted to it
(current-path, heart-phase) Causes ventricular NOTE Examples are an electric shock above the threshold of
fibrillation with a certain probability pain described in Sub-clause 4.3, the sting of a bee or burn of a
cigarette.

2) IEC Publication 50 (801): International Electrotechnical Vocabulary (IEV), Chapter 801: Acoustics and Electroacoustics.

10 © BSI 03-1999
PD 6519-2:1988
Licensed Copy: Institute Of Technology Tallaght, Institute of Technology, Tue Apr 03 04:04:41 GMT+00:00 2007, Uncontrolled Copy, (c) BSI

4 Effects of unidirectional impulse I C (p)


NOTE The relationship I DC = ------------
- is derived as follows:
currents of short duration 6
4.1 Waveforms
Figure 19, shows the forms of currents for
rectangular impulses, sinusoidal impulses and for
capacitor discharges. The following current
magnitudes have to be distinguished:
IDC = magnitude of the current of the
rectangular impulse,
IAC rms = r.m.s. value of the current of the
4.3 Threshold of perception and threshold of
sinusoidal impulse,
pain for capacitor discharge
lAC (p) = peak value of the current of the
The thresholds depend on the form of the electrodes,
sinusoidal impulse,
on the charge of the impulse and on its peak current
IC rms = r.m.s. value of the current of the value. Figure 21, shows the threshold of perception
capacitor discharge for a duration of 3 T, and the threshold of pain as a function of the charge
IC (p) = peak value of the capacitor discharge. and the charging voltage of the capacitor for a
person holding large electrodes with dry hands.
NOTE If Uc is the voltage of the capacitor at the beginning of The threshold of pain in terms of specific energy is
the discharge through the human body and Ri the initial body
resistance, IC (p) is determined by:
in the order of 50 to 100 · 10–6 A2s for current paths
through the extremities and large contact areas.
U
I C (p) = -------c 4.4 Threshold of ventricular fibrillation
Ri
The threshold of ventricular fibrillation depends on
4.2 Determination of specific fibrillating
the form, duration and magnitude of the current of
energy Fe
the impulse, the heart phase in which the impulse
The specific fibrillating energy Fe for the different starts, the current path in the human body and on
forms of impulses dealt with in this chapter, is the physiological characteristics of the person.
determined: Experiments on animals show:
a) For rectangular impulses by Fe = IDC2ti — that for impulses of short duration ventricular
b) For sinusoidal impulses by fibrillation in general results only if the impulse
falls within the vulnerable period of the cardiac
cycle;
— that the specific fibrillating charge Fq or the
c) For a capacitor discharge with a time-constant specific fibrillating energy Fe determines the
initiation of ventricular fibrillation for
T by unidirectional impulses for shock-durations
shorter than 10 ms.
Figure 20, compares the current magnitudes for Thresholds for ventricular fibrillation are shown
rectangular impulses, sinusoidal impulses and a in Figure 22. For 50% probability of fibrillation, Fq
capacitor discharge with the time constant T having is of the order of 0.005 As and Fe rises from
the same specific fibrillating energy Fe and the same about 0.01 A2s at an impulse duration ti = 4 ms
shock-duration ti. In this case the following to 0.02 A2s for ti = 1 ms.
relationships exist:
4.5 Examples
In order to explain the practical application of the
relationships described in this chapter, two
examples are given. The first example deals with a
capacitor discharge with a time constant of T = 1 ms
and a shock-duration ti = 3 T = 3 ms and is within
the scope of this part. In the second example, the
time constant is T = 10 ms, i.e. ti = 30 ms which
means that the limits for ventricular fibrillation are
those given in Figure 5 of Chapter 2.

© BSI 03-1999 11
PD 6519-2:1988
Licensed Copy: Institute Of Technology Tallaght, Institute of Technology, Tue Apr 03 04:04:41 GMT+00:00 2007, Uncontrolled Copy, (c) BSI

Example 1
Effects of capacitor discharge on the human body:
Capacitor C = 1 4F, charging
voltages 10 V, 100 V, 1 000 V and 10 000 V.
Current-path: hand-foot, initial body resistance
assumed to be Ri = 1 000 7.3)
Time constant T = 1 ms, i.e. shock-duration
ti = 3 T = 3 ms.
WC
Specific fibrillating energy Fe = I2C rmsti ≈ --------
Ri
-
Effects of shocks:

Charging voltage Uc (V) 10 100 1 000 10 000

Discharge current
Peak value IC(p) (A) 0.01 0.1 1 10
Discharge current
r.m.s. value (A)
I C (p)
IC rms = -------------
6 0.004 0.04 0.4 4
Specific charge Fq (As) 0.01 · 10–3 0.1 · 10–3 10–3 10 · 10–3
Discharge energy WC (Ws) 0.05 · 10–3 5 · 10–3 0.5 50
Specific fibrillating energy Fe
(Ri = 1 000 7) (A2s) 0.048 · 10–6 4.8 · 10–6 0.48 · 10–3 48 · 10–3
Physiological effects slight disagreeable painful ventricular
fibrillation
likely

3)
The value of Ri of 1 000 7 has been arbitrarily chosen for the purpose of this example. Not to be confused with the value of Ri
for 5 % percentile rank of Clause 6 of Chapter 1.

12 © BSI 03-1999
PD 6519-2:1988
Licensed Copy: Institute Of Technology Tallaght, Institute of Technology, Tue Apr 03 04:04:41 GMT+00:00 2007, Uncontrolled Copy, (c) BSI

Example 2
Effects of capacitor discharge on the human body:
Capacitor C = 20 4F, charging
voltage 10 V, 100 V, 1 000 V and 10 000 V.
Current-path: hand-trunk of body, initial body
resistance assumed to be Ri = 500 7.4)
Time constant T = 10 ms, i.e. shock-duration
ti = 3 T = 30 ms.5)
Effects of shocks:

Charging voltage Uc (V) 10 100 1 000 10 000

Discharge current
Peak value IC (p) (A) 0.02 0.2 2 20
Discharge current
r.m.s. value (A)
I C (p)
I C rms = -------------
6 0.008 0.08 0.8 8
Specific charge Fq (As) 10–3 10–3 10–3 200 · 10–3
a
0.2 · 2· 20 ·
Discharge energy WC (Ws) 1 · 10–3 0.1 10 1 000
Specific fibrillating energy — — — —
Fe (A2s)a
Physiological effects slight painful dangerous, but dangerous, and
ventricular ventricular
fibrillation fibrillation
unlikely likely
a As the shock duration ti is longer than 10 ms, fibrillation thresholds are to be taken from Figure 5 in Chapter 2.

4)
The value of Ri of 500 7 has been arbitrarily chosen for the purpose of this example. Not to be confused with the value of Ri for
the 5 % percentile rank of Clause 6 of Chapter 1.
5)
As the shock duration ti is longer than 10 ms, fibrillation thresholds are to be taken from Figure 5 in Chapter 2.

© BSI 03-1999 13
PD 6519-2:1988
Licensed Copy: Institute Of Technology Tallaght, Institute of Technology, Tue Apr 03 04:04:41 GMT+00:00 2007, Uncontrolled Copy, (c) BSI

Figure 19 — Forms of current for rectangular impulses, sinusoidal impulses and for
capacitor discharges

Figure 20 — Rectangular impulse, sinusoidal impulse and capacitor discharge having


the same specific fibrillating energy and the same shock-duration

14 © BSI 03-1999
PD 6519-2:1988
Licensed Copy: Institute Of Technology Tallaght, Institute of Technology, Tue Apr 03 04:04:41 GMT+00:00 2007, Uncontrolled Copy, (c) BSI

Zone A: Threshold of perception. Curve B: Typical threshold of pain.


NOTE The diagonal axes are scaled for capacitance (C) and energy (W). From the intersection of the co ordinates for charging
voltage and capacitance the charge and the energy of the impulse can be read on the appropriate axes.
Figure 21 — Threshold of perception and threshold of pain for capacitor discharges
(dry hands, large contact areas)

© BSI 03-1999
PD 6519-2:1988
Licensed Copy: Institute Of Technology Tallaght, Institute of Technology, Tue Apr 03 04:04:41 GMT+00:00 2007, Uncontrolled Copy, (c) BSI

The curves indicate the probability of fibrillation risks for current flowing in the path left hand to feet.
For other current paths, see Clause 5 and Table III of Chapter 2.
below C1: no fibrillation,
above C1 up to C2:low risk of fibrillation (up to 5 % probability),
above C2 up to C3:average risk of fibrillation (up to 50 % probability),
above C3: high risk of fibrillation (more than 50 % probability).
Figure 22 — Threshold of ventricular fibrillation

16 © BSI 03-1999
PD 6519-2:1988
Licensed Copy: Institute Of Technology Tallaght, Institute of Technology, Tue Apr 03 04:04:41 GMT+00:00 2007, Uncontrolled Copy, (c) BSI

Bibliography 6)
Chapter 4
1 Dalziel, C.F. and T.H. Mansfield: Effect of frequency on perception currents. Electrical
Engineering, 69: 794–800 (Sept. 1950), AIEE Transactions, 69: pp. 1162-1168 (1950).
2 Dalziel, C.F., E. Odgen and C.E. Abott: Effect of frequency on let-go currents. AIEE Transactions
(Electrical Engineering), 62: pp. 745-750 (Dec. 1943).
3 Geddes, L.A., L.E. Baker, P. Cabler and Brittain: Response to passage of sinusoidal current through
the body. Journal of the Association for the Advancement of Medical Instrumentation,
Vol. 5 (1971), No. 1, pp. 13-18.
4 Weirich, J., St. Hohnloser and H. Antoni: Factors determining the susceptibility of the isolated guinea pig
heart to ventricular fibrillation induced by sinusoidal alternating current at frequencies from 1 to 1 000 Hz.
Basic Res. Cardiol. Vol. 78, No. 6 (1983), pp. 604–616.
Chapter 5
1 Knickerbocker, G.G.: Fibrillating Parameters of direct and alternating (20 Hz) currents separately and in
combination. Conference Paper IEEE, No. C 72-247-0 (1972).
2 Jacobsen, J., S. Buntenkötter und H.J. Reinhard: Experimentelle Untersuchungen an Schweinen zur
Frage der Mortalität durch sinusförmige, phasenangeschnittene sowie gleichgerichtete elektrische Ströme.
Biomedizinische Technik, Vol. 20 (1975), No. 3, p. 99.
3 Reinhold, K.: Die Gefährdung durch schwingungspaketartig gesteuerte elektrische Ströme. Institut zur
Erforschung elektrischer Unfälle, Berufsgenossenschaft der Feinmechanik und Elektrotechnik, Köln,
Medizinisch-Technischer Bericht 1976.
Chapter 6
1 Biegelmeier, G., E. Homberger: Über die Wirkungen von unipolaren Impulsströmen auf den menschlichen
Körper. Bull. ASE/UCS 73 (1982) 18, S. 958–967.
The effect of unipolar current pulses on the human body. Johns Hopkins University, Applied Physics
Laboratory, Laurel, Maryland 20707, 1983.
Effets des courants d’impulsions unipolaires sur le corps humain. Bull. ASE/UCS Vol. 74 (1983),
n° 22, p. 1298.
2 Stauss, O.: Die Wirkungen von Kondensatorentladungen auf den menschlichen Körper.
Elektrizitätswirtschaft (1934), H. 23, S. 508.
3 Kouwenhoven, W.B.: Effects of capacitor discharges on the heart. Trans. Amer. Inst. Electr. Eng.,
No. 56-6 (1956).
4 Peleska, B.: Cardiac arrhythmias following condenser discharges and dependence upon strength of
current and phase of cardiac cycle. Circulation research, Vol. XIII, July 1963, p. 21-31.
5 Peleska, B.: Cardiac arrhythmias following condenser discharges led through an inductance. Circulation
research, Vol. XVI, January 1965, p. 11-18.
6 Dalziel, Ch.F.: A study of the hazards of impulse currents. AIEE-Transactions, Part III, Power Apparatus
and Systems, Vol. 72, 1953,p. 1032-1043.
7 Green, H.L., J. Ross and P. Kurn: Danger levels of short electrical shocks from 50 Hz supply. International
conference Divetech. 1981, London.
8 Kouwenhoven, W.B., G.G. Knickerbocker, R.W. Chesnut, W.R. Milnor and D.J. Sass: A–C shocks on
varying parameters affecting the heart. Trans. Amer. Inst. Electr. Eng., Part I, Bd. 78 (1959), S. 163-169.

6)
This list is not exhaustive. A more comprehensive bibliography has been published by UNIPEDE (Union Internationale des
Producteurs et Distributeurs d’Energie Electrique).

© BSI 03-1999 17
Licensed Copy: Institute Of Technology Tallaght, Institute of Technology, Tue Apr 03 04:04:41 GMT+00:00 2007, Uncontrolled Copy, (c) BSI

18
blank
Licensed Copy: Institute Of Technology Tallaght, Institute of Technology, Tue Apr 03 04:04:41 GMT+00:00 2007, Uncontrolled Copy, (c) BSI

© BSI 03-1999
See national foreword.
Publications referred to
PD 6519-2:1988
|
|
|
Licensed Copy: Institute Of Technology Tallaght, Institute of Technology, Tue Apr 03 04:04:41 GMT+00:00 2007, Uncontrolled Copy, (c) BSI

|
|
|
|
|
|
BSI Ð British Standards Institution
|
|
|
|
|
|
| BSI is the independent national body responsible for preparing British Standards. It
|
| presents the UK view on standards in Europe and at the international level. It is
| incorporated by Royal Charter.
|
|
| Revisions
|
|
| British Standards are updated by amendment or revision. Users of British Standards
|
| should make sure that they possess the latest amendments or editions.
|
|
| It is the constant aim of BSI to improve the quality of our products and services. We
|
| would be grateful if anyone finding an inaccuracy or ambiguity while using this
| British Standard would inform the Secretary of the technical committee responsible,
|
| the identity of which can be found on the inside front cover. Tel: 020 8996 9000.
|
| Fax: 020 8996 7400.
|
|
| BSI offers members an individual updating service called PLUS which ensures that
|
| subscribers automatically receive the latest editions of standards.
|
|
| Buying standards
|
| Orders for all BSI, international and foreign standards publications should be
|
| addressed to Customer Services. Tel: 020 8996 9001. Fax: 020 8996 7001.
|
|
| In response to orders for international standards, it is BSI policy to supply the BSI
|
| implementation of those that have been published as British Standards, unless
|
| otherwise requested.
|
|
| Information on standards
|
| BSI provides a wide range of information on national, European and international
|
| standards through its Library and its Technical Help to Exporters Service. Various
|
| BSI electronic information services are also available which give details on all its
|
| products and services. Contact the Information Centre. Tel: 020 8996 7111.
|
| Fax: 020 8996 7048.
|
|
| Subscribing members of BSI are kept up to date with standards developments and
| receive substantial discounts on the purchase price of standards. For details of
|
| these and other benefits contact Membership Administration. Tel: 020 8996 7002.
|
| Fax: 020 8996 7001.
|
|
| Copyright
|
|
| Copyright subsists in all BSI publications. BSI also holds the copyright, in the UK, of
|
| the publications of the international standardization bodies. Except as permitted
| under the Copyright, Designs and Patents Act 1988 no extract may be reproduced,
|
| stored in a retrieval system or transmitted in any form or by any means ± electronic,
|
| photocopying, recording or otherwise ± without prior written permission from BSI.
|
|
| This does not preclude the free use, in the course of implementing the standard, of
|
| necessary details such as symbols, and size, type or grade designations. If these
|
| details are to be used for any other purpose than implementation then the prior
| written permission of BSI must be obtained.
|
|
| If permission is granted, the terms may include royalty payments or a licensing
|
| agreement. Details and advice can be obtained from the Copyright Manager.
|
| Tel: 020 8996 7070.
|
|
|
|
|
|
|
|
|
BSI |
|
389 Chiswick High Road |
|
London |
|
W4 4AL |
|
|
|
|
|
|

You might also like