Te ch ni cal N ote 6 4 2 8 7
Orbitrap Mass Spectrometry –
Definitions, Opportunities, and Advantages
Kerstin Strupat, Olaf Scheibner, and Maciej Bromirski
Thermo Fisher Scientific (Bremen) GmbH, 28199 Bremen, Germany
Relative Abundance
60
Isopyrine
Introduction C14H19N3O
Resolution and resolving power are terms used in mass 40
50
Q Exactive HF MS Q Exactive MS high-end Q-ToF
350,000
300,000
0
m
m/z 250,000
m
Resolution (FWHM)
200,000
b Accuracy 150,000
Probability Density
100,000
50,000
0
100 200 300 400 500 600 700 800 900 1,000
m/z
Precision Measured Value Figure 3. Resolution vs. m/z for small molecule mass range.
- dimethoate (C5H12NO3PS2), m/z 230.00690 as (M+H)+ Table 1. Elements in use to retrieve the number of possible
and elemental compositions of the ion at m/z 202.04335.
60
19
F 0 3
79
Br 0 3
40
R=25k R=35k R=65k
With the information considered above in mind, a First, the mass spectrometer itself needs to deliver precise,
database search (such as on ChemSpider), easily reveals reproducible data. Precise mass determination is easiest to
the identity of the compound, which is the commonly measure in a (neat) standard sample. This is the most
used fungicide thiabendazole, also named tiabendazole or common type of sample utilized for mass calibration to
2-(1,3-thiazol-4-yl)-1H-benzimidazole. ensure not only the best mass precision but also mass
accuracy.
Quantification
Usually, extracted ion chromatograms (XIC) of ions of Secondly, one must consider the mass analyzer’s ability to
interest are created by extracting a theoretical m/z value resolve the ions of interest from all possible interferences
with an MS-analyzer-suited mass tolerance window from originating from both matrix ions as well as other
the data set. Considering the high mass accuracy of chemical background in complex samples. If the mass
Orbitrap-based instrumentation, a suited tolerance resolution of the mass spectrometer is sufficient to
window for XIC extraction is ±2 to ±5 ppm. For less separate ions of interest from the chemical background,
accurate MS analyzer types, the tolerance window should the molecular ions can be measured with high precision
be set wider; as a consequence, more chemical noise is and accuracy in a wide range of complex samples in
extracted as well and is visible in the XIC. A wider mass different matrices.
tolerance window and more chemical noise can lead to
Sufficient mass precision can be achieved only if the
overlapping peaks and incorrect data interpretation, such
system provides enough mass resolution and thereby the
as false positives. The more narrow the mass tolerance
ability to distinguish and separate the peak of interest
window can be set, the less chemical noise is expected to
from adjacent, irrelevant signals. For acquisitions done
appear in the XICs of selected m/z values.
with insufficient mass resolution, the compound of
Mass precision is one main prerequisite to achieve proper interest can be hidden in the background. This is
XIC traces to be utilized for quantification. Precise illustrated in Figure 7 for the protonated ion of the
quantification is only possible if the selected m/z value can pesticide iprovalicarb at m/z 321.21713 with the
be extracted from all scans in the raw data with the same elemental composition of C18H28N2O3 of the (M+H)+
narrow mass tolerance window. Figure 6 shows an XIC of species.
the compound carbaryl. The individual measurements of
the eluting LC peak are displayed. Carbaryl shows a
signal at m/z 202.0863 and is extracted here with a
window of ±3 ppm. The mass precision across the LC
peak ensures an interference-free extraction of the signal,
and only because the scan-to-scan precision is high (all
data points are within 1 ppm), an undistorted peak shape
is extracted, enabling interpretation of the correct
quantitative value.
0.62 ppm
100 0.62 ppm
95
90
0.54 ppm
RMS = 0.62 ppm
85
0.69 ppm across the peak
80
75 0.69 ppm
0.84 ppm
70
65
0.62 ppm
60
0.69 ppm
Relative Abundance
55
50 0.47 ppm
45
0.69 ppm
40
35 0.77 ppm
30
25 0.54 ppm
0.24 ppm
20
15 0.32 ppm
0.84 ppm
10 0.16 ppm
5
0.84 ppm 0.24 ppm
0
2.00 2.02 2.04 2.06 2.08 2.10 2.12 2.14
Time (min)
Figure 6. Precision of mass determination displayed for the XIC of m/z 202.0863 ±3 ppm,
the (M+H)+ ion of carbaryl.
a 7.72
321.21713
100 NL: 1.14E4
C18 H29 O3 N2
Te ch ni cal N ote 6 4 2 8 7
321.20609 -0.43 ppm
100
50 321.16946
50
Relative Abundance
0 0
b 321.20383
100
NL: 0 C18 H29 O3 N2
False negative - 41.85 ppm
100
50
50
0 0
Figure 7. XIC of m/z 321.21713 (protonated ion of iprovalicarb) with (a) narrow (±5 ppm) and
(b) wider (±150 ppm) extraction window due to resolution chosen; (a) acquired with
70,000 mass resolution at m/z 200, (b) acquired with 10,000 mass resolution at m/z 200.
www.thermofisher.com
©2016 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This
information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these
products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change.
Not all products are available in all countries. Please consult your local sales representative for details.
Africa +43 1 333 50 34 0 Denmark +45 70 23 62 60 Japan +81 45 453 9100 Russia/CIS +43 1 333 50 34 0
Australia +61 3 9757 4300 Europe-Other +43 1 333 50 34 0 Korea +82 2 3420 8600 Singapore +65 6289 1190
Austria +43 810 282 206 Finland +358 9 3291 0200 Latin America +1 561 688 8700 Spain +34 914 845 965
Belgium +32 53 73 42 41 France +33 1 60 92 48 00 Middle East +43 1 333 50 34 0 Sweden +46 8 556 468 00
Canada +1 800 530 8447 Germany +49 6103 408 1014 Netherlands +31 76 579 55 55 Switzerland +41 61 716 77 00
China 800 810 5118 (free call domestic) India +91 22 6742 9494 New Zealand +64 9 980 6700 UK +44 1442 233555
400 650 5118 Italy +39 02 950 591 Norway +46 8 556 468 00 USA +1 800 532 4752
TN64287-EN 12/16S