You are on page 1of 194

************************************************************

* Errorlisting SIREMOBIL 2000 software version > = VB01D *

************************************************************

*******************

* Contents view *

*******************

1. Definitions

1.1 Overview Memoskop boards

2. Error list

2.1 Initialization

2.2 Host

2.3 S3C-Bus

2.4 Memory

2.5 Television

2.6 Console

2.7 Download TV

2.8 Download Console

2.9 Download Memory

2.10 Download Harddisk

2.11 Harddisk

2.12 Generator
2.12.1 Reference list Generator

2.13 Generator Interface

2.14 Intensity acutal value Interface

2.15 Terminal handler

2.16 Motor controller

2.17 Unknown Component

3. Power up message

4. Multispot 2000 status and error messages

*******************

* 1. Definition *

*******************

Fatal errors in the Siremobil 2000

These errors are shown on the display and prevent from


radiation

until users initiation for error reset.

The error list is structured as follows and also used in the


service

pc:
XYYY reference list

with X: integration to a component

X = 0 Host

X = 1 S3C

X = 2 Image storage device (Memoskop)

X = 3 television

X = 4 console / download TV/console/

Memoskop D1/Memoskop D2/ D21 /D22

with: 40YY console

41YY download host

42YY download console

43YY download Memoskop D1

44YY download Memoskop D2 /D21 /D22

45YY harddisk error

X = 5 generator

X = 6 generator interface in Host

X = 7 intensity actual value interface

X = 8 host terminal handler

X = 9 Host motor controller

and YYY error numbers within a component


**********************************

* 1.1 Overview Memoskop boards *

**********************************

MEMOSKOP 2: D18 interface board

D17 memory

D17-D18 connector panel

MEMOSKOP 3: D1 interface board

D21 storage

D4 power supply +5V

D5 bus pc board

D6 text module

MEMOSKOP 3E: D1 interface board

D21 storage

D4 power supply +5V

D5 bus pc board

MEMOSKOP 50: D1 interface board

D2 storage with scsi controller

D3 power supply +12V


D4 power supply +5V

D5 bus pc board

D6 text module

MEMOSKOP SUB: D1 interface board

D22 storage with scsi controller

D23 arithmetic board (subtaction)

D3 power supply +12V

D4 power supply +5V

D5 bus pc board

D6 text module

**********************

* 2. ERROR LIST *

**********************

************************************

* 2.1 ERROR LIST INITIALIZATION *

************************************

After startup of the Siremobil 2000, runup messages appear

on the display of the control console and the D6 board. The


current digit
may correspond to a D2 bios initialization message, a positive

acknowledgement, an error message or wait message.

- D2 BIOS INITIALISAZION MESSAGES

are the init messages of the board D2 (SLOT CPU) and are shown
by the

LED¥s D6, J62 and J63, but not by the display.

The bios initialization messages appeare if an initialisazion


section is

started.

If any error was detected during this initialisazion section,

the processor system stops.

This suggests an error on board D2 (SLOT CPU).

Check CMOS battery (see D2 Bios initialisazion message 2F),

check also the power supply voltages.

- POSITIVE ACKNOWLEDGEMENTS

appear only if an initialization section was terminated


successfully

by the processor system. If any error was detected during this

unit initialization section, the preceding positive


acknowledgement is

either

retained or an error message is signalled.


- ERROR MESSAGES

are signalled only if an error in the corresponding

unit initialization section was recognized by the processor


system.

On successful termination of this unit initialization section


the

preceding

positive acknowledgement remains on the display until the next


positive

acknowledgement or error message.

- WAIT FOR MESSAGES

are issued if the host computer is waiting for the

acknowledge message from another assembly.

- WARNING MESSAGE

is signalled if "learning" is requested by the generator.

The Operation of the system is ensured.

Display:

Note: The bios initialization messages of the board D2 are shown


by the

LEDs D6 J62/J63 only. After it, the unit iniatialisation


messages of
the components is starting with 10 (hex) again.

As visible difference to the bios initialisazion messages,

the LED V21 on board D3 starts flashing.

D2 Bios initialisazion messages SLOT CPU (80286 CPU), part no.:


11 89 682:

01: D2 Bios initialisazion message: All 80286 registers tested


ok.

02: D2 Bios initialisazion message: BIOS checksum ok.

03: D2 Bios initialisazion message: Start of CMOS RAM test ok.

04: D2 Bios initialisazion message: 8254 Timer test started.

05: D2 Bios initialisazion message: 8254 Timer test faulty.

06: D2 Bios initialisazion message: Start of RAM test ok.

07: D2 Bios initialisazion message: RAM refresh test ok.

08: D2 Bios initialisazion message: Setting the video vectors


started

09: D2 Bios initialisazion message: Test of interrupts started

0A: D2 Bios initialisazion message: Test timer 0 interrupt


started

0B: D2 Bios initialisazion message: Seeting the interrupt table


1 ok.

0C: D2 Bios initialisazion message: Setting the interrupt table


2 ok.

0D: D2 Bios initialisazion message: CGA controller faulty

0E: D2 Bios initialisazion message: Test interrupt controller


ok.
0F: D2 Bios initialisazion message: Console controller selftest
started

D2 Bios initialisazion messages SLOT CPU (80386SX CPU),

part no.: 11 90 805:

01: D2 Bios initialisazion message: Processor test 1: Status


verification

02: D2 Bios initialisazion message: Keybard interface buffer


empty

03: D2 Bios initialisazion message: 8042 interface clear

04: D2 Bios initialisazion message: 8042 reset

05: D2 Bios initialisazion message: Get 8042 MANUFACTURE STATUS


read

input port 8042

06: D2 Bios initialisazion message: INIT CMOS and INTERRUPT


chips

07: D2 Bios initialisazion message: Processor test 2: Register


test

08: D2 Bios initialisazion message: INIT CMOS chip

09: D2 Bios initialisazion message: EPROM checksum for 64K


bytes

0A: D2 Bios initialisazion message: Initialize video 6845


controller

0B: D2 Bios initialisazion message: Test 8254 channel 0

0C: D2 Bios initialisazion message: Test 8254 channel 1

0D: D2 Bios initialisazion message: Test 8254 channel 2


0E: D2 Bios initialisazion message: Test CMOS SHUTDOWN BYTE,

sense CMOS battery power

0F: D2 Bios initialisazion message: Test extended CMOS, INIT


chipset,

date and time

10: D2 Bios initialisazion message: Test DMA-Channel 0

11: D2 Bios initialisazion message: Test DMA-Channel 1

12: D2 Bios initialisazion message: Test DMA Page Register

13: D2 Bios initialisazion message: Test 8042 UPI Keyboard


controller

interf.

14: D2 Bios initialisazion message: Test memory refresh toggle


circuit

15: D2 Bios initialisazion message: Autosize memory, test first


64kBytes

of system memory

16: D2 Bios initialisazion message: Set interrupt vector table


in first

64kB

17: D2 Bios initialisazion message: Set video I/O operation,

init video adapter cards

18: D2 Bios initialisazion message: Test video memory

- after a beep, the following initialisazion sections are


continued:
19: D2 Bios initialisazion message: Test 8259 channel 1 mask
bits

1A: D2 Bios initialisazion message: Test 8259 channel 2 mask


bits

1B: D2 Bios initialisazion message: Test CMOS battery level

1C: D2 Bios initialisazion message: Setup configuration from


CMOS

1E: D2 Bios initialisazion message: Sizing system memory and


compare

W/CMOS

1F: D2 Bios initialisazion message: Test system memory found

20: D2 Bios initialisazion message: Test stuck 8259¥s interrupt


bits

21: D2 Bios initialisazion message: Test stuck NMI bits (Parity


I/O

check)

22: D2 Bios initialisazion message: Test 8259 interrupt


functionality

23: D2 Bios initialisazion message: Test protected mode


(V86/Page)

and A20 gate

24: D2 Bios initialisazion message: Sizing extended memory


above 1MByte

25: D2 Bios initialisazion message: Test base and extended


memory found

26: D2 Bios initialisazion message: Test exception in protected


mode

27: D2 Bios initialisazion message: Test cache memory


28: D2 Bios initialisazion message: reserved

29: D2 Bios initialisazion message: reserved

2A: D2 Bios initialisazion message: Reset keyboard

2B: D2 Bios initialisazion message: Diskette init

2C: D2 Bios initialisazion message: Serial port init

2D: D2 Bios initialisazion message: Parallel port init

2E: D2 Bios initialisazion message: Harddisk init

2F: D2 Bios initialisazion message: Math coprocessor init

Possible help: Check


Battery,

if necessary change Battery


and

follow the next steps:

Power off. Insert D2.X2.


Remove

D2.X3.

Power on. Wait till LED


D3.V21 starts

flashing.

Power off.

Insert D2.X3 back again.

30: D2 Bios initialisazion message: reserved

31: D2 Bios initialisazion message: Adaptor ROM init

FE: D2 Bios initialisazion message: Password security init


FF: D2 Bios initialisazion message: Int 19H

Now the unit initialisazion messages are continued.

10: Error message: S3 bus faulty or I/O bus


fault

Initialization is continued; since, however,

communication (with S3 bus or I/0 bus) was

impossible, further Init error messages are

shown on the display. Unit stops at Init 39.

The following unit initialization messages appear on the display


of the

SM2000.

11 Positive acknowledgement: Variable access was successful.

Pointers were successfully


initialized.

11 does not appear: Display on board D6 (S3C bus) <<1


0:

test supply voltages, plug for D1,


boards

correctly inserted?
Possibly D2 (Slot CPU), D4 (CGA
card).

Display on D6 (S3C bus) <<1 0

Plug for D1, boards correctly


inserted?

Possibly D3 (boot card)

The following is displayed:

Either

12 positive acknowledgement: Memory banks for NOVRAM and

EE-Prom initialized.

or

50 error message: NOVRAM or EE-Prom cannot be


initialized

plug for D1, boards correctly


inserted?

Possibly D3 (boot card).

13 positive acknowledgement: NOVRAM and EE-Prom initialized

13 not displayed: NOVRAM or EE-Prom cannot be


initialized.
Plug for D1, boards correctly
inserted?

Possibly D3 (bootstrap card).

14 Positive acknowledgement: Fault storage initialized.

14 not displayed: Fault storage cannot be


initialized.

Plug for D1, boards correctly


inserted?

Possibly D3 (bootstrap card).

15 Positive acknowledgement: Assert buffer initialized.

Not used.

16 positive acknowledgement: Bootstrap card status o.k.

16 not displayed: Bootstrap card status not o.k..

Plug for D1, boards correctly


inserted?

Possibly D3 (bootstrap card)

17 positive acknowledgement: Interface driver to service PC


initialized.
17 not displayed: Interface driver was not
initialized.

Plug for D1, boards correctly


inserted?

Possibly D3 (bootstrap card).

18 positive acknowledgement: Variable initialization.

Transfer of unit parameters


completed,

variable initialized.

18 not displayed: Unit parameters were not


transferred.

Access o EE-Prom of bootstrap board

impossible, EE-Prom data lose.

Plug for D1, boards correctly


inserted?

Possibly D3 (bootstrap card).

19 positive acknowledgement: Internal status initialized.

19 not displayed: Access to EE-Prom of bootstrap


board

impossible EE-Prom data lose.

Plug for D1, boards correctly


inserted?
Possibly D3 (bootstrap card).

1A positive acknowledgement: Internal RMos clock is right.

1A not display Hardware clock on D2 (slot CPU) is


out

of order.

Turn off the unit, unplug battery


connector

on D2, turn on the unit again,


reinsert

battery connector with the unit


switched on.

Possibly D2 (slot CPU).

1B positive acknowledgement: R-Mos driver o.k.,

interface to generator initialized.

1B not displayed: R-Mos driver out of order,

interface to generator cannot be


initialized.

Plug for D1, boards correctly


inserted?

Possibly D3 (bootstrap card).

1C positive acknowledgement: Service task, watchdog task, boot


tasks

started.

1C not displayed: It was impossible to start


individual tasks.

1D positive acknowledgement: Reset on S3C is cancelled.

(approx. 3 sec. visible)

LEDs on S3C are flashing again.

Reset applied to generator.

1D not displayed: Host software error.

1E positive acknowledgement: Interrupt and dualported RAM are

not released.

1E not displayed: Access to D6 (S3C bus) impossible.

Plug for D1, boards correctly


inserted?

Possibly D6 (S3C bus).

1F positive acknowledgement: Brightactualinterrupt release.

1F not displayed: Possibly D2 (slot CPU) interrupt


module

defective.

20 positive acknowledgement: Download tasks and 1sec task


started.

20 not displayed: No hardware error,

possibly caused by malfunction.

After the message 20 has been signalled, the turn on code "FFFF"
is

filed in the error log storage of the Siremobil.

Either

21 error message: Operation with S3Cdualported RAM is

impossible.

Error message remains being


signalled for

5 sec, messages 25 to 2B are


skipped,

the program branches to message 2C.

Plug for D1, boards correctly


inserted?

Possibly D6 (S3C bus).


or:

22 positive acknowledgement: D6, S3C bus o.k. is displayed.

Either

23 error message: Work with I/O bus dualported RAM on


D6

(S3C bus) is impossible.

Error message remains being


signalled for sec

Messages 25 to 2B are skipped,

the program branches to message 2C.

Plug for D1, boards correctly


inserted?

Possibly D6 (S3C bus).

or:

24 positive acknowledgement: D6, I/O bus o.k. is displayed

Either:
25 error message: Polling start on S3C impossible.

Error message remains being


signalled for

2 sec

or:

26 positive acknowledgement: Polling start on S3C o.k.

27 positive acknowledgement: Preset values were successfully


transferred

to the generator.

27 not displayed: No hardware error, possibly caused


by

malfunction.

28 error message: Sensual TV control unit


initialization not

properly terminated.

Preset values to central TV control


unit

were not correctly transferred, no

hardware error, possibly caused by

malfunction.
Error message remains being
signalled

for 2 s.

29 positive acknowledgement: Version inquiry at console


successfully

terminated.

29 not displayed: Software or dualported_RAM of S3C


cannot be

addressed.

Possibly D6 (S3C).

2A positive acknowledgement: Version inquiry of TV control


unit/console

successfully terminated.

2A not displayed: Software or dualported_RAM of S3C


cannot be

addressed.

Possibly D6 (S3C).

2B positive acknowledgement: Version inquiry at memory


successfully

terminated. Message remains being


signalled
only for a short time.

2B not displayed: Software or dualported_RAM of S3C


cannot be

addressed.

Possibly D6 (S3C).

2C positive acknowledgement: Host has transmitted error release


and KV

stop cancellation to generator.


Message

remains being signalled only for a


short

time.

2C not displayed: No hardware error, possibly caused


by

malfunction.

2D positive acknowledgement: Host resets AGC mode of central TV


control

unit.

2D not displayed: No hardware error, possibly caused


by

malfunction.
2E positive acknowledgement: Host transmits the default
operating mode and

characteristic to the generator.


Message

remains being signalled for 0.5


sec.

2E not displayed: No hardware error, possibly caused


by

malfunction.

2F positive acknowledgement: Terminal interface on bootstrap


card is

released. Message remains being


signalled

only for a short time.

2F not displayed: No hardware error, possibly due to

malfunction.

The inquiries to messages 30 to 34 are active only in switch


positions

0 and 5 of switch D6 (S3C bus) S1. In all other switch positions,

these component acknowledgements are not interrogated.


30 wait for message: Host inquires of generator about

configuration and waits for


acknowledge

message.

Display duration max. 4 sec.

31 wait for message: Host waits for brightactual


telegrams from

central TV unit. Display duration


max. 20

sec.

32 wait for message: Host inquires of console about


configuration

and waits for acknowledge message.

Display duration max. 5 sec.

33 wait for message: Host interrogates of central TV


control

unit about configuration and waits


for

acknowledge message.

Display duration max. 5 sec.

34 wait for message: Host inquires of memory about


configuration
and waits for acknowledge message.

Display duration max. 10 sec.

After the message 34 has been signalled, the PTW area dose meter
is

checked; however, only in switch position 0 of switch D6.S1.

If the calibraton factor stored in the EEProm differs from the

determined value by more than 10%, the area dose meter is taken
from

the unit configuration and the error message 00007 is entered in


the

fault storage. After run up, no <F128M>m<F255D>Gy/s display


appears any longer on the display. Readjustment or repair of the
area dose meter

is required.

35 error message Generator has not answered up to


now.

Error message remains for 2 sec,

error number 0028 is entered in the

fault storage. Runup is continued


until

Init message 3A.

Check the generator.

Error message remains for 2 sec.


36 error message: Console has not yet answered.

Error number 002A is entered in the

fault storage. Run up is continued


until

initialization message 3A.

Check console.

Message remains for 2 sec.

37 error message: Memoskop has not yet answered.

Error number 0029 is entered in the

fault storage. Run up is continued


until

initialization message 3A.

Check Memoskop

Message remains for 2 sec.

38 error message: Central TV control unit has not yet


answered.

Error number 0027 is entered in the

fault storage. Run up is continued


until

initialization message 3A.

Check central TV control unit.

Message remains for 2 sec.


39 error message: S3C bus and I/0 bus of D6 are
faulty.

Since the switch position D6.S1


could not be

read out, the switch position 0 is


simulated

independent of the actual switch


position.

D6 defective or incorrected
inserted.

With the error message, a 0.5 sec


beep tone

is sounded.

3A error message If an error message 35 through 38


was

visible, the unit stops in switch


position 0

of switch D6.S1. In the switch


position > 0,

run up is continued and radiation


enabled.

3B positive acknowledgement: run up is terminated.

Note: When the unit is run up in switch position

> 0 or = fatal error message 002C appears


on the display and must be acknowledged.

3C warning: Generator requests learning


program.

If requested by the generator, e.g.


after

single-tank replacement or D201.

Error number 002B is entered in the

fault storage.

In switch position 1 of D6.S1, the

service display is shown


automatically.

***************************

* 2.2 ERRROR LIST HOST *

***************************

0003 Console does not answer

Cause : HOST sends note to console, however, no acknow-

. ledgement (KBD Ready) is received from console.

Active: Constantly

Possible trouble / help : Console

. Console power supply

. Connection to console
.

0004 Memoskop does not answer

Cause : HOST sends note to Memoskop, no acknow-

. ledgement (IMG Ready) is received from Memoskop.

Active: Constantly

Possible trouble / help : Memoskop

. Memoskop power supply

. Connection to Memoskop

0005 TV system does not answer

Cause : HOST sends note to TV, no acknowledgement

. (IMG Ready) is received from TV.

Active: Constantly

Possible trouble / help : TV

. TV power supply

. Connection to TV

0007 Area dose measurement device test out of tolerance

Cause : HOST detects excessive deviation between the value stored

. in the D3-EEPROM and the value determined in


. the self test of the area dose measurement device.

Active: during power on

Possible trouble / help : Adjustment D16

. Stored parameter wrong

. see 'unit parameter components, Dose'

000C Radiation-on message without radiation request.

Cause : HOST receives message without radiation switch-on

. (XRAY_ON), i.e. without actuation of hand or

. foot switch.

Active: during radiation

Possible trouble / help : Operator error - the release

. switch was pressed either too briefly


or in

. short, rapid succession.

0017 Unknown message of an event

Cause : The HOST has received from Memoskop an unknown or

. wrong message of an event.

Active: Constantly

Possible trouble / help : Incompatibility of HOST and Memoskop


. software.

. Wrong decoding of an event in the

. Memoskop software

001A RTO - Send error

Cause :

Active:

Possible trouble / help :

001B I/O port error

Cause : Transmission of an incorrect command.

Active: during power up

Possible trouble / help : Error in HOST software

001C Mailbox error

Cause : The internal buffer storage receives faulty data.

Active: Constantly

Possible trouble / help : System overload

Error in HOST software


.

001D HOST-TV-READY with exitcode > 0

Cause : Exitcode of TV-Ready > 0

Active: Constantly

Possible trouble / help : TV-service-switch is active

. IW-Board is not ready

001E D11 +15V or -15V out of tolerance

Cause : Message from power supply board D11. The +15 V

. or -15V voltage is out-of-tolerance.

Active: Constantly

Possible trouble / help : Power supply board D11 defective

. Connection between D11 and D6

. (acknowledgement)

001F D11, +24V out of tolerance

Cause : Message from power supply board D11. The +24 V

. voltage is out-of-tolerance.

Active: Constantly

Possible trouble / help : Power supply board D11 defective


Connection between D11 and D6

(acknowledgement)

0020 D11, +28V out of tolerance

Cause : Message from power supply board D11.

. The +28 V voltage (operating voltage of I.I. power

. supply) is out-of-tolerance.

Active: Constantly

Possible trouble / help : Power supply board D11 defective

. Connection between D11 and D6

. (acknowledgement)

0021 I.I. power supply error

Cause : Message from I.I. power supply. The power supply for

. the image intensifier (high voltage) is out-of-

. tolerance.

Active: Constantly

Possible trouble / help : I.I. power supply

. Interruption in indication lead from

. I.I. to board D6; operating voltage


. missing (also triggers error 0020).

0022 Area dose measuring device error

Cause : Message from board D16. Operating voltage

. for dose measuring chamber out-of-tolerance.

Active: Constantly

Possible trouble / help : Wrong configuration (measuring

. chamber selected, but not available)

. Measuring-chamber operating voltage

. missing. Board D16

0023 Generator not ready

Cause : Generator transmits error message to HOST but

. receives no error acknowledgement from HOST.

Active: Constantly

Possible trouble / help : If it occurs occasionally, data

. transmission fault. If it occurs

. often, fault in generator.

0024 Memoskop does not answer to "Radiation On"


Cause : HOST sends 'IMG-START' signal to the Memoskop

. at the beginning of radiation. It must be answered by

. the IMG-ACQUISITION signal from the Memoskop.

. If the HOST doesn't receive it, this error is displayed.

Active: during radiation triggering

Possible trouble / help : Memoskop

. Operator error - release switch


actuated

. repeatedly and in quick succession.

0025 Real-time clock malfunction

Cause : Hardware clock on Slot CPU (board D2) not running.

. The same time always appears following switch-on.

. Then the software clock starts running.

Active: during power up

Possible trouble / help : Battery or clock chip defective.

. Clock chip incorrectly initialized (in


case

unplug. Battery connector on D2, switch


POWER

ON/OFF off and plug battery connector


again).

.
0026 Intensity signal missing

Cause : During radiation release, the HOST waits for the

. message 'BRIGHT-ACT' from the TV system. If this

. message is missing, an error is displayed.

Active: during radiation

Possible trouble / help : Line break; defective transmit/

. receive chip; defective interface chip


on D3;

cable to generator not plugged in;


BRIGHT-ACT

connector (X3 on D3) not plugged.

0027 TV system not ready

Cause : HOST sends note to the TV system, but receives

. no acknowledgement.

Active: during power up

Possible trouble / help : HOST-TV connection interrupted

. TV defective

. TV power supply

0028 Generator not ready


Cause : HOST sends note to generator, but receives no

. acknowledgement.

Active: during power up

Possible trouble / help : HOST-generator connection

. interrupted

. Generator defective

. Generator power supply

0029 Memoskop not ready

Cause : HOST sends note to Memoskop, but receives no

. acknowledgement.

Active: during power up

Possible trouble / help : HOST-Memoskop connection

. interrupted, Memoskop defective

. Memoskop power supply

002A Console not ready

Cause : HOST sends note to unit console, but receives no

. acknowledgement.

Active: during power up


Possible trouble / help : HOST-console connection interrupt

. Unit console defective

. Console power supply

. S3C-Bus Controller ; try download

002B Generator needs learning

Cause : Generator lost learning datas

Active: during power up

Possible trouble / help : Execute generator learning

002C Host D6 switch S1 not in position 0 or 5

Cause : S1 on D6 is in service-mode 1

Active: during power up

Possible trouble / help : switch S1 to 0 (normal)

. or switch S1 to 5 (no radiation)

002E 'Radiation on' during initialisation (foot/handswitch)

Cause : Footswitch or handswitch for radiation on was pressed

during initialisation of the unit.

Active: during power up


Possible trouble / help : control foot/handswitch

0032 GEN-CURVE not allowed

Cause : generator curve is no longer allowed

Active: during radiation

Possible trouble / help : change to another curve

0033 Host D3 FEPROM error during download

Cause : Prior to downloading, the FEPROM type and memory

. space on D3 are checked. If the test is negative,

. this error is displayed.

Active: during downloading

Possible trouble / help : Defective Flash-EPROM

. Board D3 defective

0034 Host D3 FEPROM delete error during download

Cause : All FEPROMs are completely cleared prior to

. downloading. If not this error is indicated.


Active: during downloading

Possible trouble / help : +12 V programming voltage

. Board D3 defective

0035 Host D3 FEPROM programming error during download

Cause : During each write operation to the FEPROM the input

. value will be read back. If the read value

. does not agree with the written value, an error is


displayed.

Active: during downloading

Possible trouble / help : FEPROM defective

. Board D3 defective

0038 Memory bank changeover error

Cause : The memory areas can not be correctly changed over.

. It is possible that the defective memory can no

. longer be read.

Active: Constantly

Possible trouble / help : +5V power supply voltage < 4,75V

. Board D3 defective

.
0039 Block check error during service PC data transmission

Cause : Error during data transmission from Service PC

. to HOST.

Active: during service

Possible trouble / help : Serial cable from PC to unit

. Board D2 defective

. Software error

003A Echo timeout error during service PC data transmission

Cause : Error during data transmission from service PC

. to HOST.

Active: during service

Possible trouble / help : Serial cable from PC to unit

. board D2 defective

. Software error

003B Wrong echo during service PC data transmission

Cause : Error during data transmission from service PC

. to HOST.

Active: during service


Possible trouble / help : Serial cable from PC to unit

. board D2 defective

. Software error

003C Interface error during service PC data transmission

Cause : Error during data transmission from service PC

. to HOST.

Active: during service

Possible trouble / help : Serial cable from PC to unit

. board D2 defective

. Software error

003D Host D3 EEPROM programming error

Cause : During each write operation to the EEPROM, the input

. value will be read back. If the read value

. does not agree with the written value,

. an error is displayed.

Active: always

Possible trouble / help : EEPROM on board D3 defective

. Board D3 defective
. +5V power supply voltage < 4.75 V

003E Host D3 EEPROM checksum error

Cause : The stored checksum does not agree with the current

. value, i.e. the parameter file has an error. In this

. case, the entire file is automatically deleted.

Active: during power up

Possible trouble / help : Tube surges; board D3 defective;

. copy last parameter- and info file

. from diskette.

003F Host D3 NOVRAM checksum error

Cause : The stored checksum does not agree with the current

. value, i.e. the error memory is defective. In this

. case, the entire file is automatically deleted.

Active: during power up

Possible trouble / help : Tube surges; board D3 defective

. Organ programs and error memory

. were deleted.

.
0040 Host D6 FEPROM error during download

Cause : The type and storage capacity of the FEPROM are

. tested prior to downloading. If the test is

. negative, an error is indicated.

Active: during downloading

Possible trouble / help : FEPROM defective

. Unit board D6 defective

0041 Host D6 FEPROM delete error during download

Cause : All Flash-EPROMs will be completely cleared prior to

. downloading. If not, an error is indicated.

Active: during downloading

Possible trouble / help : +12 V programming voltage

. Unit board D6 defective

0042 Host D6 FEPROM programming error during download

Cause : After each write operation to the FEPROM, the input

. value is read back. If the read value does not agree

. does not agree with the written value,


. an error is displayed.

Active: during downloading

Possible trouble / help : Flash-EPROM defective

. +12 V programming voltage defective

. Board D6 defective

0043 Console D20 does not answer during download

Cause : HOST transmits data to unit console, but receives

. no answer.

Active: during downloading

Possible trouble / help : Load lead for unit console does

. not answer

. Flash-EPROM defective

. Board D20 defective

. Download-connection defective

0044 Console D20 FEPROM delete error during download

Cause : Console FEPROM is completely cleared prior

. to downloading.

. If this proves impossible, an error is indicated.

Active: during downloading


Possible trouble / help : +15 V voltage

. Board D20 defective

. Appears in connection with error 420B

0045 Console D20 FEPROM programming error during download

Cause : After each write operation to the FEPROM, the input

. value is read back. If the value read does not agree

. with the value written, an error is displayed.

Active: during downloading

Possible trouble / help : Flash-EPROM defective

. +12 V programming voltage defective

. Board D20 defective

. Appears with error 420C

0046 TV does not answer during download

Cause : HOST transmits data to TV, but receives no answer.

Active: during downloading

Possible trouble / help : Load lead for TV does not answer

. TV system defective

. Download-connection defective
.

0047 TV NOVRAM delete error during download

Cause : All NOVRAMs are completely cleared prior to

. downloading. If this proves impossible, an error is

indicated.

Active: during downloading

Possible trouble / help : +15 V voltage

. TV system defective

0048 TV NOVRAM programming error during download

Cause : After each write operation to the NOVRAM, the input

. value is read back. If the value read does not agree

. with the value written, an error is displayed.

Active: during downloading

Possible trouble / help : NOVRAM defective

TV system defective

0049 Memoskop D1/D18 does not answer during download

Cause : HOST transmits data to Memoskop, but receives no answer.

Active: during downloading


Possible trouble / help : Download indication D6.V12 dark

. Download connection defective

. Memoskop defective

004A Memoskop D1/D18 FEPROM delete error during download

Cause : All FEPROMs are completely cleared prior to

. downloading. If not, an error is indicated.

Active: during downloading

Possible trouble / help : +12 V programming voltage

. Memoskop defective

004B Memoskop D1/D18 FEPROM programming error dur. download

Cause : After each write operation to the FEPROM, the input

. value is read back. If the read value does not agree

. with the written value, an error is displayed.

Active: during downloading

Possible trouble / help : FEPROM defective

. +12 V programming voltage

. Memoskop defective

.
004C Memoskop D2 does not answer during download

Cause : HOST transmits data to hard disk, but receives

. no answer.

Active: during downloading

Possible trouble / help : Load lead for hard disk

. does not answer

. Hard disk controler defective

. Download-connection defective

004D Memoskop D2 FEPROM delete error during download

Cause : All FEEPROMs are completely cleared prior to down-

. loading.If this proves impossible, an error is indicated.

Active: during downloading

Possible trouble / help : +12 V programming voltage

. Hard disk controler defective

004E Memoskop D2 FEPROM programming error during download

Cause : After each write operation to the FEPROM, the input

. value is read back. If the value read doesn't agree

. with the value written, an error is displayed.


Active:during downloading

Possible trouble / help : FEPROM defective

. +12 V programming voltage

. Hard disk controler defective

004F No console 'READY' during download

Cause : HOST sends note to console, but receives no

. acknowledgement.

Active: during downloading

Possible trouble / help : Unit console

. Power supply for console defective

. Connection to console defective

. Internal procedure error during

. downloading

0050 No TV 'READY' during download

Cause : HOST sends note to TV, but receives no

. acknowledgement.

Active: during downloading

Possible trouble / help : TV system defective


. TV system power supply defective

. Connection to TV system defective

0051 No Memoskop D1/D18 'READY' during download

Cause : HOST sends note to Memoskop, but receives no

. acknowledgement.

Active: during downloading

Possible trouble / help : Memoskop defective

. Connection to Memoskop defective

0052 No Memoskop D2 'READY' during download

Cause : HOST sends note to harddisk, but receives no

. acknowledgement.

Active: during Download

Possible trouble / help : Harddisk controler defective

. Connection to harddisk controler


defective

0053 Host D3 EEPROM faulty

Cause : EEPROM has been initialized. All adjustments of the


. testfield and from service have been reseted.

Active:

Possible trouble / help : load latest parameter-file

0054 Unit Software version different

Cause : The stored unit software version string in menu

. <options><software ver> is different to the

. actual unit software version

Active:

Possible trouble / help : change software version in menu

. <options> <software ver> <set new>

0055 Host D6 hardware not correct

Cause : processor checks D6

Active: power up

Possible trouble / help : D6 not pluged-in correctly

. AT-bus interface D6 defective

0056 Host D6 Watchdog Reset

Cause : Processor out of cycle


Active: Constantly

Possible trouble / help : Reset

0057 GEN D201 FEPROM delete error during download

Cause :

Active: During DOWNLOAD of generator

Possible trouble / help : Check +12V on D201

. FEPROM defective, change D201

0058 HOST-GEN communication error during download

Cause : Communication HOST - generator

Active: During DOWNLOAD of generator

Possible trouble / help :

0059 Generator checksum failure after download

Cause :

Active: During DOWNLOAD of generator

Possible trouble / help : Download once more

.
005A Generator communication error during download

Cause :

Active: During DOWNLOAD of generator

Possible trouble / help :

005B Generator communication error during download

Cause :

Active: During DOWNLOAD of generator

Possible trouble / help :

005C Generator FEPROM wrong internal ID

Cause : FEPROM type not released or FEPROM defektive

Active: During DOWNLOAD of generator

Possible trouble / help : Board D201 defektive

******************************

* 2.3 ERROR LIST S3C (D3) *

******************************
1070 Host D6 S3B-receive-buffer unknown status

Cause : S3C-software doesn't work properly

Active: during power up

Possible trouble / help : download S3C software

. D6 defective

1071 Host D6 S3B-receive-buffer has init status

Cause : S3C-software doesn't work properly

Active: during power up

Possible trouble / help : download S3C software

. D6 defective

1075 Host D6 S3B-undefined receive-message

Cause : S3C-software doesn't work properly

Active: during power up

Possible trouble / help : download S3C software

. D6 defective

1076 Host D6 S3B-class-error


Cause : S3C-software doesn't work properly

Active: during power up

Possible trouble / help : download S3C software

. D6 defective

1077 Host D6 S3B-code-error

Cause : S3C-software doesn't work properly

Active: during power up

Possible trouble / help : download S3C software

. D6 defective

1078 Host D6 S3B-code-extension-error

Cause : S3C-software doesn't work properly

Active: during power up

Possible trouble / help : download S3C software

. D6 defective

1079 Host D6 S3B-function-code-error

Cause : S3C-software doesn't work properly

Active: during power up


Possible trouble / help : download S3C software

. D6 defective

107A Host D6 S3B-undefined feedback

Cause : S3C-software doesn't work properly

Active: during power up

Possible trouble / help : download S3C software

. D6 defective

107B Host D6 S3B-send-buffer full-message

Cause : S3C-software doesn't work properly

Active: during power up

Possible trouble / help : download S3C software

. D6 defective

107C HOST-S3 bus not initialized

Cause : S3C-software doesn't work properly

Active: during power up

Possible trouble / help : download S3C software


. D6 defective

107D Host D6 S3B-send-buffer undefined feedback

.Cause : S3C-software doesn't work properly

Active: during power up

Possible trouble / help : download S3C software

. D6 defective

107E Host D6 S3B-send-buffer unknown status

Cause : S3C-software doesn't work properly

Active: during power up

Possible trouble / help : download S3C software

. D6 defective

107F Host D6 S3B-send-buffer unknown error

Cause : S3C-software doesn't work properly

Active: during power up

Possible trouble / help : download S3C software

. D6 defective

.
1081 Host D6 S3B-undefined input-feedback

Cause : S3C-software doesn't work properly

Active: during power up

Possible trouble / help : download S3C software

. D6 defective

1082 Host D6 S3B-undefined input-message

Cause : S3C-software doesn't work properly

Active: during power up

Possible trouble / help : download S3C software

. D6 defective

1083 Host D6 S3B-undefined command-address

Cause : S3C-software doesn't work properly

Active: during power up

Possible trouble / help : download S3C software

. D6 defective

.
1084 Host D6 S3B-undefined input-error

Cause : S3C-software doesn't work properly

Active: during power up

Possible trouble / help : download S3C software

. D6 defective

1085 Host D6 S3B-undefined input-receive-feedback

Cause : S3C-software doesn't work properly

Active: during power up

Possible trouble / help : download S3C software

. D6 defective

1086 HOST - S3 bus communication error

Cause : Communication between HOST and S3 bus disturbed

Active: Constantly

Possible trouble / help : Board D6 defective

. Normal following Download

. set the unit Online

1087 Host D6 S3B-input wrong-length-error


Cause : S3C-software doesn't work properly

Active: during power up

Possible trouble / help : download S3C software

. D6 defective

1088 Host D6 S3B-input unknown-type-error

Cause : S3C-software doesn't work properly

Active: during power up

Possible trouble / help : download S3C software

. D6 defective

108A Host D6 S3B-input wrong-checksum-error

Cause : S3C-software doesn't work properly

Active: during power up

Possible trouble / help : download S3C software

. D6 defective

108B Host D6 S3B-input undefined feedback

Cause : S3C-software doesn't work properly


Active: during power up

Possible trouble / help : download S3C software

. D6 defective

10D4 Host D6 S3C-IOB unknown receive status

Cause : S3C software doesn't work properly

Active: during power up

Possible trouble / help : download S3C software

. D6 defective

10D5 Host D6 S3C-IOB receive-init-status

Cause : S3C-software doesn't work properly

Active: during power up

Possible trouble / help : download S3C software

. D6 defective

10D9 Host D6 S3C-IOB receive undefined message

Cause : S3C-software doesn't work properly

Active: during power up

Possible trouble / help : download S3C software


. D6 defective

10DA Host D6 S3C-IOB class error

Cause : S3C-software doesn't work properly

Active: during power up

Possible trouble / help : download S3C software

. D6 defective

10DB Host D6 S3C-IOB code error

Cause : S3C-software doesn't work properly

Active: during power up

Possible trouble / help : download S3C software

. D6 defective

10DE Host D6 S3C-IOB receive undefined feedback

Cause : S3C-software doesn't work properly

Active: during power up

Possible trouble / help : download S3C software

. D6 defective
.

10DF Host D6 S3C-IOB send-buffer full

Cause : S3C-software doesn't work properly

Active: during power up

Possible trouble / help : download S3C software

. D6 defective

10E0 Host D6 I/O RAM not initialized

Cause : Dual ported RAM on D6 can't be initialized

Active: during power up

Possible trouble / help : Check FEPROM for S3C

. Download S3C software

. Check RAM on D6

. D6 defective

10E1 Host D6 S3C-IOB send undefined feedback

Cause : S3C-software doesn't work properly

Active: during power up

Possible trouble / help : download S3C software

. D6 defective
.

10E2 Host D6 S3C-IOB send unknown status

Cause : S3C-software doesn't work properly

Active: during power up

Possible trouble / help : download S3C software

. D6 defective

10E3 Host D6 S3C-IOB send unknown error

Cause : S3C-software doesn't work properly

Active: during power up

Possible trouble / help : download S3C software

. D6 defective

10E5 Host D6 S3C-IOB-input undefined feedback

Cause : S3C-software doesn't work properly

Active: during power up

Possible trouble / help : download S3C software

. D6 defective

.
10E6 Host D6 S3C-IOB-input undefined message

Cause : S3C-software doesn't work properly

Active: during power up

Possible trouble / help : download S3C software

. D6 defective

10E7 Host D6 S3C-IOB-input undefined command-address

Cause : S3C-software doesn't work properly

Active: during power up

Possible trouble / help : download S3C software

. D6 defective

10E8 Host D6 S3C-IOB-input undefined error

Cause : S3C-software doesn't work properly

Active: during power up

Possible trouble / help : download S3C software

. D6 defective

10E9 Host D6 S3C-IOB-input-receive undefined feedback


Cause : S3C-software doesn't work properly

Active: during power up

Possible trouble / help : download S3C software

. D6 defective

10EA HOST - I/O bus communication error

Cause : Communication between HOST and I/O bus disturbed

Active: Constantly

Possible trouble / help : Board D6 defective

. Normal following Download

. and set the unit Online

10EB Host D6 S3C-IOB-input-receive wrong-length

Cause : S3C-software doesn't work properly

Active: during power up

Possible trouble / help : download S3C software

. D6 defective

10EC Host D6 S3C-IOB-input-receive unknown-type


Cause : S3C-software doesn't work properly

Active: during power up

Possible trouble / help : download S3C software

. D6 defective

10EE Host D6 S3C-IOB-input-receive wrong-checksum

Cause : S3C-software doesn't work properly

Active: during power up

Possible trouble / help : download S3C software

. D6 defective

10EF Host D6 S3C-IOB-input-receive undefined feedback

Cause : S3C-software doesn't work properly

Active: during power up

Possible trouble / help : download S3C software

. D6 defective

******************************

* 2.4 ERROR LIST MEMOSKOP *

******************************
2000 Memoskop IMG communication error

2101 Memoskop D1/D18 processor hardware error

Cause : Internal processor test negative.

Active: during power up

Possible trouble / help : processor defective

. Board D1 defective.

2102 Memoskop D1/D18 external processor RAM test failed

Cause : All RAMS are cleared during initialization.

. If this is not possible, an error is indicated.

Active: during power up

Possible trouble / help : RAM defective

. Board D1 defective.

2103 Memoskop D1/D18 checksum test configur. byte failed

Cause : Current checksum does not agree with stored

. checksum. In this case, the standard configuration

. is automatically selected.
Active: during power up

Possible trouble / help : NOVRAM defective

. Replace board D1

. Restore old configuration

2104 Memsokop D1/D18 FEPROM checksum error

Cause : Current checksum does not agree with

. stored checksum.

Active: during power up

Possible trouble / help : Flash-EPROM defective

. Replace board D1

. Try to download

2105 Memoskop D2 (D21/D17/D22) not addressable

Cause : When data are sent to D2 (D21/D17/D22) the data

. transmitted are automatically polled. If the feedback

. is not correct, an error message appears.

Active: during power up

Possible trouble / help : Board D2 (D21/D17/D22) defective

. (not plugged in)

. Backplane D5 defective
. Board D1 defective

2106 Memoskop D2 (D21/D22/D17) CPU2-BUS defective

Cause : Error detected in CPU 2 during initialization.

. Internal test images no longer selectable. Image can

. no longer be cleared for "Next Patient".

Active: during power up

Possible trouble / help : Board D2 (D21/D17/D22) defective

. Backplane D5 defective

. Board D1 defective

2107 Memoskop D1 NOVRAM (Hospital line IMG) checksum error

Cause : Current checksum does not agree with the stored one

. Only an empty line is displayed instead of the name.

Active: during power up

Possible trouble / help : NOVRAM defective

. Board D1 defective

2108 Memoskop D1/D18 Stackoverflow


Cause : Internal Software Error

Active: Constantly

Possible trouble / help :

2109 Memoskop IMG-text-loading checksum test failed

Cause: During loading an image from the disk there was

. a checksum failure of the text of this image.

. No text will be loaded.

Active: during loading of an image

Possible trouble / help: There wasn't ever stored any text

. on this image.

. Disk is full

. Change 50Hz to 60Hz or 60Hz to 50Hz

210A Memoskop D22 not addressable

Cause : When data are sent to D22 the transmitted data are

. automatically polled. If the feedback is not

. correct, an error massage apears.

Active: during power up

Possible trouble / help : Board D22 defective

. (not plugged in)


.

2301 No START following HS_MODE (exposure telegram start)

Cause : HOST has transmitted the HS_MODE signal, but

. the answer (hardware lead START) is not transmitted

. within 5 sec.

Active: during radiation

Possible trouble / help : Operator error - the release

switch was pressed either too briefly


or in

short, rapid succession.

2302 Excessive START pause during pulsed-mode operation

Cause : HS_STOP (end of exposure) message not received

. within 5 sec. in pulsed mode. This may result

. in a HOST reset.

Active: during pulsed mode operation

Possible trouble / help : Operator error - the release

. switch was pressed either too briefly or in

. short, rapid succession.


.

2303 VR-sync. missing in HS_MODE

Cause : The Memoskop-internal VR pulse is missing on board D1

. at the beginning of the exposure. The green LED on

. the Memoskop does not light up.

Active: during start of exposure

Possible trouble / help : Connector X13 on board D1

. Board D1 defective

. Power supply power supply

2304 Internal VR sync. in HS_MODE

Cause : The Memoskop-external VR pulse is missing on

. board D1 at the start of the exposure. The Memoskop

. is synchronised to the internal VR sync.

Active: during start of exposure

Possible trouble / help : BAS signal from TV system missing

. Plug X9 not connected

. Board D1 defective

.
2401 Memsokop CPU1/2 communication error

Cause : Communication error between CPU1 (D1) and CPU2 (D2)

. The checksum is not ok. Error detected by D1.

Active: during hard-disk activities

Possible trouble / help : Board D1, D2, or D5 defective

. Different software on D1 and D2.

. Try Download.

2402 Memsokop, no communication with CPU2 following RESET

Cause : Communication error between CPU1 (D1) and CPU2 (D2).

. Error detected by D1.

Active: during initialization

Possible trouble / help : Board D1, D2, or D5 defective.

. Different software on D1 and D2

. Try Download

. Possible subsequend error resulting

. from error 2103.

2403 Memoskop, timeout during communication with CPU2

Cause : Timeout during communication between D1 and CPU2

Active: during all hard-disk activities


Possible trouble / help : Board D2 (CPU2) defective

. Different software on D1 and D2

. Try Download Memoskop.

2404 Memoskop, error during polling of CPU2 version

Cause : Wrong software detected in CPU2 (D2).

Active: during power up

Possible trouble / help : Different software versions on

. boards D1 and D2

. Try Download Memoskop.

2420 Memoskop IMG_SUB Diskoverflow

Cause : All images on disk are 'protected';

. Nop images will be written to disk.

. (only warning)

Active: when store an image on disk

Possible trouble / help : 'unprotect' patients

2421 Memoskop IMG_SUB disk full


Cause : scene is longer than amount of 'unprotected

. images'. No more images will be stored on disk.

. (only warning)

Active: when store an image on disk

Possible trouble / help :'unprotect' patients

2422 Memoskop IMG_SUB Unexepected Image type

Cause : Subtrtaction image found without mask

. (only warning)

Active: during loading an image from disk

Possible trouble / help : Overwrite image

2423 Memoskop IMG_SUB no stored image available

Cause : No image found on disk; all images cleared

. (only warning)

Active: during loading an image from disk

Possible trouble / help :

2424 Memoskop disk directory init error

Cause : Failure during initialization of disk directory


Active: during power up

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

2425 Memoskop IMG_SUB disk directory checksum error

Cause : Checksum error during reading the disk directory

. Disk directory will be initialized, all 'protections'

. will be cancled, deleted images become visible

. (noname patients with date 0.JAN)

Active: during power up

Possible trouble / help : Unit was switched off during

. exposure or number of images on disk was changed

2480 Memoskop harddisk-bus error

Cause : bus not free

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective
.

2481 Memoskop harddisk-bus error

Cause : no arbitration phase

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

2482 Memoskop harddisk-bus error

Cause : no selection phase

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

2483 Memoskop harddisk-bus error

Cause : no reselection phase

Active:

Possible trouble / help : Board D2 defective

. Cable defective
. Harddisk defective

2484 Memoskop harddisk-bus error

Cause : no command phase 1

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

2485 Memoskop harddisk-bus error

Cause : no command phase 2

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

2486 Memoskop harddisk-bus error

Cause : no data in phase 1

Active:
Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

2487 Memoskop harddisk-bus error

Cause : no data in phase 2

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

2488 Memoskop harddisk-bus error

Cause : no data out phase 1

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

2489 Memoskop harddisk-bus error

Cause : no data out phase 2


Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

248A Memoskop harddisk-bus error

Cause : no status phase

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

248B Memoskop harddisk-bus error

Cause : no message in phase

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

.
248C Memoskop harddisk-bus error

Cause : no message out phase

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

248D Memoskop harddisk-bus error

Cause : no data out LRAM phase 1

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

248E Memoskop harddisk-bus error

Cause : no data out LRAM phase 2

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

.
248F Memoskop harddisk-bus error

Cause : no data in LRAM phase 1

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

2490 Memoskop harddisk-bus error

Cause : no data-in-LRAM phase 2

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

2491 Memoskop harddisk-bus timeout error

Cause : CPU1 I/O data timeout

Active:

Possible trouble / help : Board D2 defective

. Cable defective
. Harddisk defective

2492 Memoskop harddisk-bus error

Cause : CPU1 no out-data phase 1

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

2493 Memoskop harddisk-bus timeout error

Cause : CPU1 OUT-data phase timeout

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

2494 Memoskop harddisk-bus timeout error

Cause : CPU1 IN-data timeout

Active:

Possible trouble / help : Board D2 defective


. Cable defective

. Harddisk defective

2495 Memoskop harddisk-bus timeout error

Cause : CPU1 IN-data timeout

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

2496 Memoskop harddisk-bus timeout error

Cause : CPU1 IN-data phase timeout

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

24A0 Memoskop CPU2-SW-Error

Cause : No Arbitration phase


. Internal Software Error

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

24A1 Memoskop CPU2-SW-Error

Cause : No Reselection phase

. Internal Software Error

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

24A3 Memoskop CPU2-SW-Error

Cause : No Message Out Phase Impl

. Internal Software Error

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective
.

24B0 Memoskop harddisk bus timeout error

Cause : Arbitration timeout

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

24B1 Memoskop harddisk bus timeout error

Cause : Selection timeout

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

24B2 Memoskop harddisk bus timeout error

Cause : Reselection timeout

Active:

Possible trouble / help : Board D2 defective


. Cable defective

. Harddisk defective

24B3 Memoskop harddisk bus timeout error

Cause : Command entry timeout

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

24B4 Memoskop harddisk bus timeout error

Cause : Command phase timeout

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

24B5 Memoskop harddisk bus timeout error

Cause : Data-IN timeout

Active:
Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

24B6 Memoskop harddisk bus timeout error

Cause : Data-IN phase timeout

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

24B7 Memoskop harddisk bus timeout error

Cause : Data-OUT timeout

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

24B8 Memoskop harddisk bus timeout error


Cause : Data-OUT phase timeout

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

24B9 Memoskop harddisk bus timeout error

Cause : Status entry timeout

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

24BA Memoskop harddisk bus timeout error

Cause : Status phase timeout

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

.
24BB Memoskop harddisk bus timeout error

Cause : Message IN-entry timeout

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

24BC Memoskop harddisk bus timeout error

Cause : Message IN-phase timeout

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

24BD Memoskop harddisk bus timeout error

Cause : Message Out-entry timeout

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective
.

24BE Memoskop harddisk bus timeout error

Cause : Message OUT-phase timeout

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

24BF Memoskop harddisk bus timeout error

Cause : Data-OUT-LRAM-entry timeout

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

24C0 Memoskop harddisk bus timeout error

Cause : Data-OUT-LRAM-phase timeout

Active:

Possible trouble / help : Board D2 defective

. Cable defective
. Harddisk defective

24C1 Memoskop harddisk bus timeout error

Cause : Data-IN-LRAM-entry timeout

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

24C2 Memoskop harddisk bus timeout error

Cause : Data-IN-LRAM-phase timeout

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

24C3 Memoskop harddisk bus timeout error

Cause : CMD-exit-Message entry timeout

Active:
Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

24D0 Memoskop harddisk error

Cause : Check condition

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

24D1 Memoskop harddisk error

Cause : Unexpected Status

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

24D2 Memoskop harddisk error

Cause : Unexpected IN-message


Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

24D3 Memoskop harddisk error

Cause : Data parity error

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

24D4 Memoskop harddisk error

Cause : Harddisc not on SCSI bus

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

. 12V-Power Supply

.
24D5 Memoskop harddisk error

Cause : Hardisk not ready

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

. 12V Power Supply

24D8 Memoskop harddisk bus error

Cause : SCSI Stop error

Active:

Possible trouble / help : Board D2 defective.

. Cable defective

. Harddisk defective

24D9 Memoskop harddisk bus error

Cause : SCSI stop: parity error

Active:

Possible trouble / help : Board D2 defective

. Cable defective
. Harddisk defective

24DA Memoskop harddisk bus error

Cause : SCSI stop: Data phase error

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

24DB Memoskop harddisk bus error

Cause : SCSI stop: error

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

24DC Memoskop harddisk bus error

Cause : SCSI stop: LRAM stop error

Active:
Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

24DD Memoskop harddisk bus error

Cause : SCSI stop error

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

24DE Memoskop harddisk bus error

Cause : SCSI stop error

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

24DF Memoskop harddisk bus error

Cause : SCSI stop error


Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

24E0 Memoskop harddisk bus error

Cause : CMD exit with check condition

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

24E1 Memoskop harddisk bus error

Cause : CMD Exit no message in phase

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

.
24E2 Memoskop harddisk bus error

Cause : CMD exit unexpected IN-message

Active:

Possible trouble / help : Board D2 defective

. Cable defective

. Harddisk defective

24F0 Memoskop CPU2-SW-error

Cause : Unknown DBI unknown request from CPU1

Active:

Possible trouble / help : Internal SW error

24F1 Memoskop CPU2-SW-error

Cause : Error in 3 byte ADD

Active:

Possible trouble / help : Internal SW error

24F2 Memoskop CPU2-SW-error

Cause : Checksum Error in FEPROM

Active:
Possible trouble / help : Board D2 defective

. Try download

24F3 Memoskop CPU2-SW-error

Cause : Wrong request

Active:

Possible trouble / help : Internal SW Error

24F4 Memoskop CPU2-SW-error

Cause : Wrong request number

Active:

Possible trouble / help : Internal SW Error

24F8 Memoskop CPU1-CPU2 communication error

Cause : Receiver timeout

Active:

Possible trouble / help : Board D1 defective

. Board D2 defective

. Board D5 defective
.

24F9 Memoskop CPU1-CPU2 communication error

Cause : Receiver checksum error

Active:

Possible trouble / help : Board D1 defective

. Board D2 defective

. Board D5 defective

24FA Memoskop CPU1-CPU2 communication error

Cause : Transmitter timeout

Active:

Possible trouble / help : Board D1 defective

. Board D2 defective

. Board D5 defective

24FB Memoskop CPU1-CPU2 communication error

Cause : Transmitter checksum error

Active:

Possible trouble / help : Board D1 defective

. Board D2 defective
. Board D5 defective

24FC Memoskop CPU1-CPU2 communication error

Cause : Transmitter HW Error

Active:

Possible trouble / help : Board D1 defective

. Board D2 defective

. Board D5 defective

2501 Memoskop unknown S3 bus telegram received

Cause : Unknown command on S3 bus

Active: Constantly

Possible trouble / help : Different software versions on

. boards D1 and host

. Try Download HOST and Memoskop.

2502 Memoskop S3 bus telegram with wrong length received

Cause : wrong command on S3 bus

Active: Constantly
Possible trouble / help : Different software versions on

. boards D1 and D2

. Try Download HOST and Memoskop.

2503 Memoskop IMG S3 bus transmit error

Cause : Error in data transmission between Memoskop and S3 bus

Active: Constantly

Possible trouble / help : Host - S3 bus - D6 input

. connection defective

. Output connection for board D1


defective

2504 Memoskop S3 bus receive error

Cause : Error in data transmission between S3 bus and Memoskop

Active: Constantly

Possible trouble / help : Output connection between host D6

. and S3 bus defective

. Input connection of board D1 defective

2601 Memoskop error during text controller test


Cause : Internal test of test controller negative.

Active: during power up

Possible trouble / help : Text module D6 in Memoskop defective

. Board D1 defective

2602 Memoskop error during text controller RAM test

Cause : Internal test of test controller RAMs negative.

Active: during power up

Possible trouble / help : Text module D6 in Memoskop defective

. Board D1 defective

2610 Memoskop keyboard interface test: baud rate too low

Cause : Error detected in SG carriage during keyboard test

Active: during power up

Possible trouble / help : Interface module for keyboard on

. board D1 defective.

2611 Memoskop keyboard interface test: baud rate too high

Cause : Error detected in SG carriage during keyboard test

Active: during power up


Possible trouble / help : Interface module for keyboard on

. board D1 defective.

2612 Memoskop keyboard interface test: Baud clock missing

Cause : Error detected in SG carriage during keyboard test

Active: during power up

Possible trouble / help : Interface module for keyboard on

. board D1 defective.

2620 Memoskop keyboard interface test: Feedback error

Cause : Error detected in SG carriage during keyboard test

Active: during power up

Possible trouble / help : Interface module for keyboard on

. board D1 defective.

2621 Memoskop keyboard interface test: Feedback timeout

Cause : Error detected in SG carriage during keyboard test

Active: during power up

Possible trouble / help : Interface module for keyboard on


. board D1 defective.

2710 Memoskop BAS sync. test: H sync. period too long

Cause : Error detected during synchronization test of

. storage TV system

Active: during power up

Possible trouble / help : Board D1 defective

. Cable X13 on Memoskop

2711 Memoskop BAS sync. test: H sync. period too short

Cause : Error detected during synchronization test of

. storage TV system

Active: during power up

Possible trouble / help : Board D1 defective

. Cable X13 on Memoskop

2712 Memoskop BAS sync. test: H sync. missing

Cause : Error detected during synchronization test of

. storage TV system

Active: during power up


Possible trouble / help : Board D1 defective

. Cable X13 on Memoskop

2720 Memoskop BAS sync. test: 50 Hz VRsync period too long

Cause : Error detected during synchronization test of

. storage TV system

Active: during power up

Possible trouble / help : Board D1 defective

. Cable X13 on Memoskop

2721 Memoskop BAS sync. test: 50 Hz VRsync period too short

Cause : Error detected during synchronization test of

. storage TV system

Active: during power up

Possible trouble / help : Board D1 defective

. Cable X13 on Memoskop

2722 Memoskop BAS sync. test: 60 Hz VRsync period too long

Cause : Error detected during synchronization test of


. storage TV system

Active: during power up

Possible trouble / help : Board D1 defective

. Cable X13 on Memoskop

2723 Memoskop BAS sync. test: 60 Hz VRsync period too short

Cause : Error detected during synchronization test of

. storage TV system

Active: during power up

Possible trouble / help : Board D1 defective

. Cable X13 on Memoskop

2724 Memoskop BAS sync. test: VR sync. missing

Cause : Error detected during synchronization test of

. storage TV system

Active: during power up

Possible trouble / help : Board D1 defective

. Cable X13 on Memoskop

2730 Memoskop BAS sync. test: H sync. level error


Cause : Error detected during synchronization test of

. storage TV system

Active: during power up

Possible trouble / help : Board D1 defective

. Cable X13 on Memoskop

2731 Memoskop BAS sync. test: PLL lock error

Cause : Error detected during synchronization test of

. storage TV system

Active: during power up

Possible trouble / help : Board D1 defective

. Cable X13 on Memoskop

2740 Memoskop BAS sync. test: VR interrupt missing

Cause : Error detected during synchronization test of

. storage TV system

Active: during power up

Possible trouble / help : Board D1 defective

. Cable X13 on Memoskop

.
2750 Memoskop BAS sync. test: Wrong callup parameter

Cause : Error detected during synchronization test of

. storage TV system

Active: during power up

Possible trouble / help : Software error

2800 Memoskop Voltage test: ADC ref. voltage too high

Cause : Reference voltage for analog/digital converter on

. board D1 too high.

Active: during power up

Possible trouble / help : CPU1 error

. Board D1 defective

2801 Memoskop Voltage test: ADC ref. voltage too low

Cause : Reference voltage for analog/digital converter on

. board D1 too high.

Active: during power up

Possible trouble / help : CPU1 error

. +5 V supply voltage too high

. Board D1 defective
.

2810 Memoskop Voltage test: VPP min. too high

Cause : Programming voltage for flash PROMs

. out of tolerance.

Active: during power up

Possible trouble / help : Voltage regulator for programming

. voltages on D1 defective

. Connector X13 on Memoskop

2811 Memoskop Voltage test: VPP min. too low

Cause : Programming voltage for flash PROMs

. out of tolerance.

Active: during power up

Possible trouble / help : Voltage regulator for programming

. voltages on D1 defective

. Connector X13 on Memoskop

2812 Memoskop Voltage test: VPP max. too high

Cause : Programming voltage for flash PROMs


. out of tolerance.

Active: during power up

Possible trouble / help : Voltage regulator for programming

. voltages on D1 defective

. Connector X13 on Memoskop

2813 Memoskop Voltage test: VPP max. too low

Cause : Programming voltage for flash PROMs

. out of tolerance.

Active: during power up

Possible trouble / help : Voltage regulator for programming

. voltages on D1 defective

. Connector X13 on Memoskop

2814 Memoskop Voltage test: VPP missing

Cause : Programming voltage for flash PROMs

. out of tolerance.

Active: during power up

Possible trouble / help : Voltage regulator for programming

. voltages on D1 defective

. Connector X13 on Memoskop


.

2820 Memoskop Voltage test: 5V analog too high

Cause : Supply voltage exceeds 6 V

Active: during power up

Possible trouble / help : Voltage regulator on board D1 def.

. Connector X13 on Memoskop

2821 Memoskop Voltage test: 5V analog too low

Cause : Supply voltage less than 4 V; LED +5V on Memoskop

. switches off.

Active: during power up

Possible trouble / help : Voltage regulator on board D1 def.

. Connector X13 on Memoskop

2830 Memoskop Voltage test: -5V2 too high

Cause : Supply voltage for clock too high

Active: during power up

Possible trouble / help : Voltage regulator on board D1 def.

. Connector X13 on Memoskop


.

2831 Memoskop Voltage test: -5V2 too low

Cause : Supply voltage for clock too low; LED on Memoskop

. -5V2 switches off.

Active: during power up

Possible trouble / help : Voltage regulator on board D1 def.

. Connector X13 on Memoskop

2840 Memoskop Voltage test: VEE analog (-5V) too high

Cause : Supply voltage for ADC/DAC too high.

Active: during power up

Possible trouble / help : Voltage regulator on board D1 def.

. Connector X13 on Memoskop

2841 Memoskop Voltage test: VEE analog (-5V) too low

Cause : Supply voltage for ADC/DAC too low. LED VEE on

. Memoskop switches off.

Active: during power up

Possible trouble / help : Voltage regulator on board D1 def.

. Connector X13 on Memoskop


.

********************************

* 2.5 ERROR LIST TELEVISION *

********************************

3000 TV communication failure

3001 TV reference voltage out of tolerance

Cause : Tolerances for reference voltage exceeded

Active: Constantly

Possible trouble / help : See troubleshooting instructions

. Power supply D10

. Board D110

3004 TV D112 NVRAM RAM-region error

Cause : Processor tries to overwrite a protected area

. in the NVRAM.

Active: Constantly

Possible trouble / help : Perform "source" test and switch

. unit off and on.


. Error eliminated: ok

. Error not eliminated: D112 board;

. +5V voltage (rise, drop).

3005 TV D112 NVRAM checksum error

Cause : Stored checksum of NVRAM does not agree with actual

. checksum.

Active: Constantly

Possible trouble / help : Try download and switch on/off

. several times

. Error eliminated: ok

. Error not eliminated: D112;

. +5V voltage (rise, drop).

3006 TV D112 NVRAM test error

Cause : Test of NVRAM is negative

Active: during Download

Possible trouble / help : NVRAM defective ->

. Replace D112

.
3009 TV interference during download

Cause : Illegal spikes on download line

Active: during Download

Possible trouble / help : Check download lines

3010 KK voltage out of tolerance

Cause : One of the KK voltages (+5 Vdigital, +12 Vanalog,

. +12 Vdigital, -10 Vdigital) is missing.

Active: Constantly . G5321-83

Possible trouble / help : Measure at D110 pin X4/b10:

.Voltage between +1.9 and +2.0 V: +5 Vdigital defective

.Voltage between - 0.6 and - 0.3 V: +12 Vdigital defective

.Voltage between +4.6 and +4.7 V: -10 Vdigital defective

.Voltage between +2.6 and +2.7 V: +12 Vanalog defective

.Voltage between +4.0 and +4.1 V: all voltages are o.k.

3011 TV CUP Peltier voltage out of tolerance

Cause : Faulty supply voltage of Peltier element

Active: Constantly

Possible trouble / help : Voltage at test pin X1/30 (CUP)


. not ok -> power supply defective

. If voltage at test pin X1/30 (CUP) is ok, then

. measure voltage at test pin X2/15 (GUP);

. If ok -> VV (preamp) board defective.

If voltage at test pin X2/15 (GUP) is not ok, KK def.

3012 KK B DC signal out of tolerance

Cause : DC value of B signal on board KK not ok

Active: Constantly

Possible trouble / help : Power supply of D110 (preamp)

. Test pin X1/VV ok: D110 defective

. Test pin X1/VV not ok: KK board defective.

3013 KK B-signal AC-value out of tolerance

Cause : AC value of B signal on board KK not ok

Active: Constantly

Possible trouble / help : KK board defective

3014 KK Peltier element does not cool the CCD chip

Cause : Measured temperature value with active


. peltier element not lower than with inactive

. peltier element.

Active: during initial loading of the TV camera

Possible trouble / help :Normal after worming up

. of the CCD chip KK and initial loading

. of the TV camera.

. Peltier element defective

3020 TV D110 X27TP D+12V out of tolerance

Cause : +12 Vdigital supply voltage for the KK board def.

Active: Constantly

Possible trouble / help : Test pin X27/D110 ok: D110

. defective.

. Test pin X27/D110 not ok but +15 V ok:

. D110 defective.

. Test pin X27/D110 not ok and +15 V not

. ok: power supply defective.

3021 TV D110 X26TP A+12V out of tolerance

Cause : +12 Vanalog supply voltage for KK board defective.


Active: Constantly

Possible trouble / help : D110 defective

3022 TV D110 LED V62, +5V out of tolerance

Cause : +5 V supply voltage for the VV board defective

Active: Constantly

Possible trouble / help : Test pin X1/25 on D110 ok:

. D110 defective

. Test pin X1/25 on D110 not ok:

. power supply defective

3023 TV D110 LED V63, -10V peletier (CUB) out of tolerance

Cause : Supply voltage for Peltier element (-10V) defective

Active: Constantly

Possible trouble / help : Test pin X1/29 on D110 ok:

. D110 defective

. Test pin X1/29 on D110 not ok:

. power supply defective

3024 TV D110 X28TP, A-12V out of tolerance


Cause : Fault in -12 V supply voltage for D110

Active: Constantly

Possible trouble / help : Test pin X28/D110 ok: D110 def

. Test pin X28/VV not ok but -15 V supply

. voltage ok: D110 defective

. Test pin X28/D110 not ok and -15 V supply

. voltage not ok: power supply def.

3025 TV D110 X29TP, D-12V out of tolerance

Cause: Fault in -12 V supply voltage for D110

Active: Constantly

Possible trouble / help : D110 defective

3027 TV D110 X6TP, H-sawtooth amplitude test error

Cause : Fault in H sawtooth on D110

Active: Constantly

Possible trouble / help : Test pin X6/D110 ok: D110 def.

.Test pin X6/D110 not ok and amplitude adjustable:

.Set amplitude.

.Amplitude not adjustable: Measure XH/C9 on H-pulse board


.If H-pulse ok: D110 defective; otherwise, D112.

3028 TV D110 X8TP, digital measuring field test error

Cause : Fault in digital measuring field

Active: Constantly

Possible trouble / help : Test pin X8 on D110 ok:

. VV board defective

. Test pin X8 on D110 not ok:

. Measure test pin X4/a8.

. Test pin o.k.:D110 defective;

. otherwise D112 defective.

3029 TV IW analog measuring field test error

Cause : Error in analog measuring field

Active: Constantly

Possible trouble / help : D111 defective

. D110 possibly defective

302A TV D110 X9TP, 250kHz-test-image test error


Cause : Fault in digital test image

Active: during test image

Possible trouble / help : Activate 250 kHz test image and

. measure test pin X9 on D110.

. If ok: D110 defective; otherwise switch

. over S100/D110 and measure X9/D110 again.

. If ok: D112 defective; otherwise

. D110 defective.

302B TV D110 IW MIWB signal error

Cause : MIWD signal on D110 defective

Active: Constantly

Possible trouble / help : D110 defective

302C TV D110 BASA signal error

Cause : BASA contrast of D110 defective

Active: Constantly

Possible trouble / help : D110 defective

. Check connection to Memoskop

. (75-ohm terminal resistance)


.

302D TV D110 BASA linear signal error

Cause: BASA signal (linear) of D110 defective

Active: Constantly

Possible trouble / help : D110 defective

302E TV D110 BASA gamma signal error

Cause : BASA signal (gamma) of D110 defective

Active: Constantly

Possible trouble / help : D110 defective

3030 TV D111 actual value error

Cause : Fault on D111

Active: Constantly

Possible trouble / help : Check supply voltages on D110.

. If voltages ok: D111 defective.

3031 TV D111 actual value timeout


Cause : Fault on IW board

Active: Constantly

Possible trouble / help : Check supply voltages on D110.

. If voltages ok: D111 def.

3032 TV D111 actual value transfer error

Cause : transfer error on D111

Active: Constantly

Possible trouble / help : D111 defective

3033 TV D111 actual value timeout on transfer

Cause : transfer error on D111

Active: Constantly

Possible trouble / help : D111 defective

3035 TV D111 actual value digital error

Cause : Fault in IWB signal

Active: Constantly

Possible trouble / help : If error 302B also occurs,


. D110 is defective; otherwise

. D111 is defective.

3036 TV D111 actual value MW error

Cause : Error in actual mean value

Active: Constantly

Possible trouble / help : D111 defective

3037 TV D111 actual value SW error

Cause : Error in actual peak value, white

Active: Constantly

Possible trouble / help : D111 defective

3038 TV D111 actual value SS error

Cause : Error in actual peak value, black

Active: Constantly

Possible trouble / help : D111 defective

.
.

3039 TV D111 actual value SS-SWerror

Cause : Error in actual peak value, black and white

Active: Constantly

Possible trouble / help : D111 defective

303A TV D111 AGC error

Cause : fault in AGC control

Active: Constantly

Possible trouble / help : D111 defective

3101 TV D112 test switch active

3102 TV data failure

Active: Constantly

3103 TV DBI failure

Active: Constantly
3104 TV D111 actual value isnt't ready

Active: during power up

310A TV NOVRAM error during download

Active: during TV Download

310B TV HX_DLO_START signal not received during download

Active: during TV Download

310C TV NOVRAM checksum error during download

Active: during TV Download

*****************************

* 2.6 ERROR LIST CONOSOLE *

*****************************

4000 Console communication error

4001 Console FEPROM checksum error

Cause : Wrong checksum during data transfer from console

. to HOST.

Active: during power up

Possible trouble / help : FEPROM defective


. Software in FEPROM

. console defective

. Try Download.

4002 Console watchdog reset error

Cause : Console signals a RESET to the HOST

Active: Constantly

Possible trouble / help : FEPROM defective

. Software in FEPROM

. console defective

. Try Download.

4003 Console FEPROM +12V Vpp out of tolerance

Cause : Programming voltage for FEPROMs

. out-of-tolerance

Active: during power up

Possible trouble / help : Power supply for console not o.k.

. Console defective

. Break in cable
.

4004 Console +5V out of tolerance

Cause : Power supply for console out-of-tolerance

Active: during power up

Possible trouble / help : Power supply for console not o.k.

. Console defective

. Break in cable

4005 Console SW version not released

Cause : Software version of console could not be recognized.

Active: during power up

Possible trouble / help : Console software defective,

. Try download console

*********************************

* 2.7 ERROR LIST DOWNLOAD TV *

*********************************

4105 TV D112 NVRAM checksum error during download

Cause : Error detected during download

Active: during TV download


Possible trouble / help : Board D112 defective

. Repeat download.

4106 TV D112 NVRAM test error during download

Cause : Error detected during download

Active: during TV download

Possible trouble / help : Board D112 defective

. Repeat download.

4116 TV D112 data error during download

Cause : Error detected during download

Active: during TV download

Possible trouble / help :

4117 TV D112 DBI error during download

Cause : Unknown data word detected during download

Active: during TV download

Possible trouble / help : Operation of unit executed during

. download

.
411E TV D112 NVRAM error during download

Cause : Error detected during download

Active: during TV Download

Possible trouble / help : Repeat download.

. Board D112 may be defective.

411F TV D112 no download start

Cause : No data are sent for Download unless a download

. request is pending.

Active: during TV download

Possible trouble / help : Software error

4120 TV D112 checksum error in DBI during download

Cause : Error detected during download

Active: during TV download

Possible trouble / help : Malfunction of data line

. Check plug connections.

**************************************

* 2.8 ERROR LIST DOWNLOAD CONSOLE *

**************************************
4201 Console FEPROM checksum error during download

Cause : Wrong checksum for data transfer.

Active: during console download

Possible trouble / help : FEPROM defective

. Software in FEPROM

. console defective

4202 Console watchdog reset during download

Cause : Console signals a RESET to the HOST

Active: during console download

Possible trouble / help : FEPROM in console defective

. Try download.

4203 Console FEPROM +12V Vpp out of tolerance dur. download

Cause : Programming voltage for FEPROMs

. out-of-tolerance

Active: during console download

Possible trouble / help : Power supply for console not o.k.

. Console defective

. Break in cable
.

4204 Console +5V out of tolerance during download

Cause : Power supply for console out-of-tolerance

Active: during console download

Possible trouble / help : Power supply for console not o.k.

. Console defective

. Break in cable

4205 Console SW version not released during download

Cause : Software version of console could not be recognized.

Active: during console download

Possible trouble / help : Console software defective

. Try download

420B Console FEPROM erase error during download

Cause : The FEPROM is completely erased prior to

. programming. If this proves impossible, an error

. message appears.

Active: during console download

Possible trouble / help : +15 V supply voltage missing


. Board D20 defective

420C Console FEPROM programmming error during download

Cause : After each write operation to the FEPROM, the input

. value is read back. If the value read does not agree

. with the value written, an error is displayed.

Active during: Console download

. Possible trouble / help : FEPROM defective, D20

. +15 V supply voltage

. Console defective

420D Console checksum error in DBI during download

Cause : Data were transmitted with the wrong checksum.

Active: during console download

Possible trouble / help : Malfunction of the data line

. Check plug connections

420E Console FEPROM checksum error during download

Cause : Checksum of flash EPROM is not correct


Active: during console Download

Possible trouble / help : FEPROM defective

. Board D20

******************************************

* 2.9 ERROR LIST DOWNLOAD MEMOSKOP D1 *

******************************************

4314 Memoskop D1/D18 data checksum error during download

Cause : Checksum error detected in data transfer

Active: during Memoskop download

Possible trouble / help : Malfunction of data line

. Check plug connections

4321 Memoskop D1/D18 not ready for download

Cause : Request sent for download, however,

. Memoskop not ready.

Active: during download of Memoskop D1 (D18)

Possible trouble / help : Timing

. Try download again.

.
4322 Memoskop D1/D18 DBI too long during download

Cause : Excessively long data word detected

. during download.

Active: during download of Memoskop D1 (D18)

Possible trouble / help : Fault in Memoskop

. Board D1 (D18)

. S3-Bus error

4323 Memoskop D1/D18 address outside of area dur. download

Cause : Selected address lies outside of permissible area

Active: during download of Memoskop D1 (D18)

Possible trouble / help : Internal software error

4324 Memoskop D1/D18 incorrect data checksum dur. download

Cause : Data was transferred with an incorrect checksum.

Active: during download of Memoskop D1 (D18)

Possible trouble / help : Malfunction of data line (S3-Bus).

. Check plug connections.

.
4331 Memoskop D1/D18 FEPROM erase error during download

Cause : Memoskop can not be cleared

Active: during download of Memoskop D1 (D18)

Possible trouble / help : FEPROM defective.

. +15 V supply voltage, X13

. Board D1 defective

4332 Memoskop D1/D18 FEPROM programming error dur. download

Cause : After each write operation to the FEPROM, the input

. value is read back. If the value read does not agree

. with the value written, an error is displayed.

Active:during download of Memoskop D1 (D18)

Possible trouble / help : FEPROM defective

. +15 V supply voltage, X13

. Board D1 defective

4333 Memoskop D1/D18 FEPROM1 programming error dur download

Cause : No data can be written to FEPROM 1.

Active: during download of Memoskop D1 (D18)

Possible trouble / help : FEPROM defective


. Board D1 (D18) defective

. +15 V supply voltage, X13

4334 Memoskop D1/D18 FEPROM2 programming error dur download

Cause : No data can be written to FEPROM 2.

Active: during download of Memoskop D1 (D18)

Possible trouble / help : FEPROM defective

. Board D1 (D18) defective

. +15 V supply voltage, X13

4335 Memoskop D1/D18 FEPROM not complete erased dur. downld

Cause : The FEPROM is completely erased prior

. to programming.

. If this proves impossible, an error message appears.

Active: during download of Memoskop D1 (D18)

Possible trouble / help : +15 V supply voltage missing

. Board D1 (D18) defective

. FEPROM defective

.
4336 Memoskop D1/D18 FEPROM all bytes already 0

4341 Memoskop D1/D18 unknown test number during download

Cause :An unknown test number was called up.

Active: during download of Memoskop D1 (D18)

Possible trouble / help : Software error

4342 Memoskop D1/D18 unknown DBI in HX_REQUEST dur download

Cause : Unknown data bit detected during polling

. of configuration

Active: during download of Memoskop D1 (D18)

Possible trouble / help : Software error

4343 Memoskop D1/D18 wrong segment during download

Cause : Software error

Active: during download of Memoskop D1 (D18)

Possible trouble / help :

.
4345 Memoskop D1/D18 wrong segment during download

Cause : Software error

Active: during download of Memoskop D1 (D18)

Possible trouble / help :

4351 Memoskop D1/D18 unknown telegram from HOST dur downld

4352 Memoskop D1/D18 receive error during download

Cause : Error in data reception by HOST

Active: during download of Memoskop D1 (D18)

Possible trouble / help : Memoskop-HOST connection

. Software error

4353 Memoksop D1/D18 transmit error during download

Cause : Error in data transmission by HOST

Active: during download of Memoskop D1 (D18)

Possible trouble / help : Memoskop-HOST connection

4361 Memoskop D1/D18 +12V Vpp out of tolerance dur download


Cause : +12 V programming voltage not o.k.

Active: during download of Memoskop D1

Possible trouble / help : Connector X13 on board D1

. Board D1 (D18) defective

. Power unit voltage supply

4362 Memoskop D1/D18 +12.4V Vpp out of tolerance dur downld

Cause : +12 V programming voltage not o.k.

Active: during download of Memoskop D1 (D18)

Possible trouble / help : Connector X13 on board D1

. Board D1 (D18) defective

. Power unit voltage supply

4363 Memoskop D1/D18 internal test failed durning download

Cause : Error during internal test

Active: during download of Memoskop D1 (D18)

Possible trouble / help :

4371 Memoskop D1/D18 FEPROM checksum error during download


Cause : Checksum of FEPROMs incorrect

Active: during download of Memoskop D1

Possible trouble / help : FEPROM defective

****************************************************

* 2.10 ERROR LIST DOWNLOAD MEMOSKOP D2/ D21/ D22 *

****************************************************

4414 Memoskop D2 data checksum error during download

Cause : Error detected in checksum during data transfer.

Active: during download of Memoskop D2

Possible trouble / help : Malfunction of data line

. Check plug connections

. D1 defective

4421 Memoskop D2 not ready for download

Cause : Request sent for download, however,

. Memoskop not yet ready.

Active: during download of Memoskop D2

Possible trouble / help :

.
4422 Memoskop D2 DBI too long during download

Cause : Excessively long data word detected

. during download.

Active: during download of Memoskop D2

Possible trouble / help : Fault in Memoskop

. Board D2

. S3-Bus error

4423 Memoskop D2 address outside of area during download

Cause : Selected address lies outside of permissible area

Active: during download of Memoskop D2

Possible trouble / help :

4424 Memoskop D2 incorrect data checksum during download

Cause : Data transferred with incorrect checksum.

Active: during download of Memoskop D2

Possible trouble / help : Malfunction of data line.

. Check plug connections.

.
4431 Memoskop D2 FEPROM erase error during download

Cause : Memoskop can not be cleared

Active: during download of Memoskop D2

Possible trouble / help : Board D2, D1 defective.

. +15 V supply voltage, X13

. FEPROM defective

4432 Memoskop D2 FEPROM programming error during download

Cause : After each write operation to the FEPROM, the input

. value is read back. If the value read does not agree

. with the value written, an error is displayed.

Active: during download of Memoskop D2

Possible trouble / help : FEPROM defective

. +15 V supply voltage, X13

. Board D2, D1 defective

4433 Memoskop D2 FEPROM1 programming error during download

Cause : No data can be written to FEPROM 1.

Active: during download of Memoskop D2

Possible trouble / help : FEPROM defective


. +15 V supply voltage, X13

. Board D2, D1 defective

4434 Memoskop D2 FEPROM2 programming error during download

Cause : No data can be written to FEPROM 2.

Active: during download of Memoskop D2

Possible trouble / help : FEPROM defective

. +15 V supply voltage, X13

. Board D2, D1 defective

4435 Memoskop D2 FEPROM erase error during download

Cause : The FEPROM is completely erased prior

. to programming.

. If this proves impossible, an error message appears.

Active: during download of Memoskop D2

Possible trouble / help : FEPROM defective

. +15 V supply voltage, X13

. Board D2, D1 defective

.
4436 Memoskop D2 FEPROMS all bytes already 0

4441 Memoskop D2 unknown test number during download

Cause : An unknown test number was called up.

Active: during download of Memoskop D2

Possible trouble / help : Software error

4442 Memoskop D2 unknown DBI in HX_REQUEST during download

Cause : Unknown data bit detected during polling

. of configuration

Active: during download of Memoskop D2

Possible trouble / help : Software error

4443 Memoskop D2 wrong segment during download

Cause : Software error

Active: during download of Memoskop D2

Possible trouble / help : Software error

4444 Memoskop D2 wrong test DBI during download


Cause : Software error

Active: during download of Memoskop D2

Possible trouble / help :

4445 Memoskop D2 wrong segment during download

Cause : Software error

Active: during download of Memoskop D2

Possible trouble / help :

4451 Memoskop D2 unknown telegram from HOST during download

4452 Memoskop D2 receive error during download

Cause : Error in data reception by HOST

Active: during download of Memoskop D2

Possible trouble / help : Memoskop-HOST connection

. Software error

4453 Memoskop D2 transmit error during download

Cause : Error in data transmission by HOST


Active: during download of Memoskop D2

Possible trouble / help : Memoskop-HOST connection

4461 Memoskop D2 +12V Vpp out of tolerance during download

Cause : +12 V programming voltage not o.k.

Active: during download of Memoskop D1

Possible trouble / help : Connector X13 on board D1

. Board D1 defective

. Power unit voltage supply

4462 Memoskop D2 +12.4V Vpp out of tolerance dur. download

Cause : +12 V programming voltage not o.k.

Active: during download of Memoskop D1

Possible trouble / help : Connector X13 on board D1

. Board D1 defective

. Power unit voltage supply

4463 Memoskop D2 internal test failed during download

Cause : Error during internal test

Active: during download of Memoskop D2


Possible trouble / help :

4471 Memoskop D2 FEPROM checksum error during download

Cause : Checksum of FEPROMs incorrect

Active: End of download of Memoskop D2

Possible trouble / help : FEPROM defective

******************************

* 2.11 ERROR LIST HARDDISK *

******************************

4500 Memoskop D2 Harddisk communication error

*******************************

* 2.12 ERROR LIST GENERATOR *

*******************************

5000 GEN communication error

5002 X-RAY_ON timeout error


Cause : During radiation, the HOST processor sends

. an X-RAY_ON

. note every 100 ms. If the generator does not receive

. such a note within 300 ms, an error is displayed.

Active: during radiation

Possible trouble / help : Interference on the HOST-generator

. serial interface

5003 GEN intensity signal timeout error

Cause : During radiation, the TV processor sends

. an intensity

. value every 20 ms. If the generator does not receive

. a valid TV note within 100 ms, an error is displ..

Active: during radiation

Possible trouble / help : Interference on the

. TV-HOST-generator serial interface

5005 GEN D201 EEPROM can not be overwritten

Cause : After each write operation to the EEPROM, the input

. value will be read back. If the value read does not agree

. with the value written, an error is displayed.


Active: during standby

Possible trouble / help : EEPROM defective

5006 GEN time limit for cassette exposure

Cause : For cassette exposures, a max. time is calculated

. according to the mAs setting. If this setting is

. exceeded by 20%, the exposure is aborted. An error

. is then displayed.

Active: during cassette exposure

Possible trouble / help : Required tube current

. not attained.

. Line impedance too high.

5007 GEN D201 FEPROM checksum error

Cause : Checksum of FEPROM is wrong

Active: during power up

Possible trouble / help : download generator software

. Change FEPROM

.
5008 GEN-learning not possible

Cause :

Active: Service-mode

Possible trouble / help : Check filament current

. Short circuit tube

. Check x-ray tube and cable

500A Overload current in tube heating circuit

Cause : If the overcurrent monitor of the heating inverter

. constantly responds during a preset period of time,

. an error is displayed.

Active: during radiation, Standby

Possible trouble / help : Short circuit in inverter bridge

. Malfunction of controller

. (ground wiring, shielding)

500B Overload current in intermediate-circuit reduct. contr

Cause : If the overcurrent monitor of intermediate-circuit

. reduction controller constantly responds during a

. preset period of time, an error is displayed

. (V1.1 IZ D202 lights up for approx. 0.5 s).


Active: during radiation, standby

Possible trouble / help : Short circuit in reduction

. controller bridge

. Malfunction of controller (ground wiring, shielding)

500C Overload current in main inverter

Cause : If the overcurrent monitor of the main inverter

. constantly responds during a preset period of time, an

. error is displayed.(LED V1.1 IZ on D202 lights up for

. appr. 0.5 s).

Active: during radiation

Possible trouble / help : Short circuit in main inverter

. Malfunction of controller

. (ground wiring, shielding)

5014 Maximum tube current exceeded

Cause : If an illegal tube current is detected at the analog

. input of the computer during radiation, an error is

. displayed.

Active: during radiation


Possible trouble / help : Single tank not

. high-voltage-proof.

5015 Minimum tube current not attained

Cause : If an excessively low tube current or no current at

. all is detected at the analog input of the computer

. during radiation, an error is displayed.

Active: during radiation

Possible trouble / help : Single tank defective

. No actual mA value present

5016 Maximum kV exceeded

Cause : Controller D202 monitors the max. permissible value

. of the high voltage (120 kV). If this value is exceeded,

. an error is displayed.

Active: during radiation

Possible trouble / help : Single tank defective

. No actual kV value present

5017 Minimum kV not attained


Cause : Controller D202 monitors the min. permissible value

. of the high voltage. If this value is not attained,

. an error is displayed.

Active: during radiation

Possible trouble / help : No actual kV value present

. Inverter not functioning

. No drive pulses present.

5019 Maximum heating current exceeded

Cause : The HK inverter monitors the r.m.s. value of the

. heating current (D203 J11) with time-dependence.

. If the monitor constantly responds (LED V25 on D203),

. an error is displayed.

Active:

Possible trouble / help :

501A Minimum heating current not attained

Cause : If an excessively low heating current or no current

. at all is detected at the analog input of the computer

. during radiation, an error is displayed.


Active: during radiation, standby

Possible trouble / help : Oscillating branch of heating

. circuit not o.k. No actual tube values present.

501B Maximum intermediate-circuit voltage too high

Cause : If an illegal intermediate-circuit voltage

. is detecte at the analog input of the computer,

. an error is displayed.

Active: during radiation, standby

Possible trouble / help : Single tank defective

. No actual kV value present

. +/-15 V voltage supply

501C Minimum intermediate-circuit voltage too low

Cause : If an excessively low intermediate-circuit voltage

. or no itermediate-circuit voltage at all is detected

. at the analog input of the computer, an error is displ.

Active: during radiation, standby

Possible trouble / help : Drive pulses for intermediate-

. circuit reduction controller not present; oscillating

. circuit not o.k. (choke L1)


.

501D Maximum input voltage exceeded

Cause : If an excessive rectified input voltage is detected

. at the analog input of the computer (D200, X12 Un=324V)

. an error is displayed.

Active: during radiation, standby

Possible trouble / help : Input voltage D204, X1 too high.

501E Minimum input voltage not attained

Cause : If an excessively low rectified input voltage or no

. such voltage at all is detected at the analog input

. of the computer (D200, X12 Un=324V), an error is displ.

Active: during radiation, standby

Possible trouble / help : Input voltage D204, X1 too low.

. Line impedance too high.

5020 Maximum heating-inverter frequency

Cause : If the heating inverter operates at maximum

. frequency for a preset period of time,


. an error is indicated.

Active: during radiation, standby

Possible trouble / help : Actual IH value not present at

. analog input.

5028 GEN D202 +/- 15 V supply voltage missing

Cause : Controller D202 monitors the +/-15 V supply voltages

. (relays K15P, K15N). If the monitor responds,

. an error message is output.

Active: during radiation, standby

Possible trouble / help : +/- 15 V not present.

5029 GEN D202 power circuit supply voltages missing

Cause : Controller D202 monitors the input voltage for the

. intermediate circuit (Un=230 V). If the value measured

. lies beyond the prescribed limits, an error message comes

Active: during radiation, standby

Possible trouble / help : Input voltage D204, X1 missing

. Fuses D204, F1/F2 defective

. Fuse D200, F1 defective.

.
502B No radiation enable from controller

Cause : Controller D202 must send the XRAY_Q_R signal to

. enable radiation release. If this enable signal is

. not sent or is cancelled during radiation emission,

. an error message is displayed.

Active: during radiation

Possible trouble / help : Check the signals I_MAX_ZW,

. P_MAX_R,I_MAX_WR, VZW_OK, XRAY_ON on D202.

502C Inverter control not possible

Cause : The WR_OK signal indicates whether drive pulses are

. available for the inverter. If the computer detects that

. this signal is missing, an error is then displayed.

Active: during radiation

Possible trouble / help : D202, XRAY_ON_S1

. (HW release missing)

502D temperature sensor defective.

Cause : Current single tank sensor temperature < +17∞ Celsius


Aktive: during power up.

Possible trouble / help: Break or short circuit in sensor line.

single tank sensor < +17∞ Celsius

**************************************

* 2.12.1 REFERENCE LIST GENERATOR *

**************************************

5065 GEN temp max note

Cause : Max. temperature of single tank attained (70¯C)

Active: during radiation, standby

Possible trouble / help : Single tank overheated

. Break or short circuit in sensor line

5067 GEN pmax active note

Cause : Oil pressure switch on single tank responded;

. radiation is switched off.

Active: during radiation, standby

Possible trouble / help : Single tank overheated.

. Check D200, X5

. Break in line to oil pressure switch.


.

5068 GEN pmax o.k. note

Cause : Oil pressure switch deactivated after single tank

. cooled off.

Active: during radiation, standby

Possible trouble / help :

5069 GEN xray on S2 note

Cause : If the trigger switch is released during radiation

. (LIH, cassette exposure), a message is sent to the host.

Active: during radiation

Possible trouble / help : Operator prematurely aborted

. exposure.

. Line break

506A GEN missing learn data note

Cause : Learn data missing

Active: during standby

Possible trouble / help : Heating data learn procedure not


. performed.

. Heating data learn operation procedure


aborted.

. Perform generator learn procedure.

506B GEN rx acknowledge note

Cause : No acknowledge note received

Active: during radiation

Possible trouble / help : Interference with HOST-GEN

. communication

. Check connecting lines.

506C GEN telegram repeat note

Cause : Generator receives wrong acknowledge 3x following

. transmission of note.

Active: during radiation, standby

Possible trouble / help : Interference with HOST-GEN

. communication

506D GEN error data note


Cause : Error during storage of error data in EEPROM D201.

Active: during radiation, standby

Possible trouble / help : D201

506E GEN-pushfactor not exactly

Cause : pushfactor of filament current not o.k.

Active: service mode (learning of filament current)

Possible trouble / help : line resistance to high

. board D203 defective

506F GEN operating temperature to low

Cause : Current single tank sensor temperature < +17∞ Celsius

Aktive: during power up.

Possible trouble / help: Break or short circuit in sensor line.

single tank sensor < +17∞ Celsius

*****************************************

* 2.13 ERROR LIST INTERFACE GENERATOR *

*****************************************
6006 Generator Ready not received

Cause : Generator does not answer HOST note

Active: Constantly

Possible trouble / help : Connection to generator defective

. Generator defective

6007 Unknown ready timeout

6008 HOST received GEN-note with wrong DBI or parameter

Cause : generator sends wrong note to HOST

Active: Constantly

Possible trouble / help : Incompatible generator and

. HOST software.

6009 HOST received GEN-note with wrong checksum

Cause : Incorrect checksum during data transfer between

. generator and HOST.

Active: Constantly

Possible trouble / help : Transfer error

. Defective interface modules


.

600A HOST received GEN-note of incorrect length

Cause : Note from generator to HOST has incorrect length.

Active: Constantly

Possible trouble / help : Transmission error

. Fault in generator

. Check plug connection

600B Gen_in ACK received but no acceptance

Cause : The generator sends an acknowledge in response to a

. note even though no note was sent by the HOST.

Active: Constantly

Possible trouble / help : Generator-HOST timing defective,

. e.g. in short, rapid succession

. Release switch actuated.

600C XRAY_ON, but hand/foot switch not pressed

Cause : Radiation is triggered even though the

. release switch was not pressed longer than 4 sec.

Active: during radiation


Possible trouble / help : Operator error - the release

. switch was pressed either too briefly


or in

. short, rapid succession.

600D HOST received GEN-note with wrong DBI

Cause : If data is transmitted from the HOST to the

. generator, the generator subsequently sends an answer

. telegram.If the answer contains the wrong function

. code, an error message appears.

Active: Constantly

Possible trouble / help : HOST and generator software

. incompatible

600E HOST received GEN-note with wrong parameter

Cause : If data is transmitted from the HOST to the

. generator, the generator subsequently sends an answer

. telegram.If the answer contains the wrong function

. code, an error message appears.

Active: Constantly

Possible trouble / help : HOST and generator software


. incompatible

600F HOST denied GEN-note

Cause : If data are sent from the HOST to the generator, the

. generator then sends an answer note. If the answer

. contains the wrong function code, an error appears.

Active: Constantly

Possible trouble / help : Incompatible HOST and

. generator software

6010 HOST received GEN-note with wrong checksum

Cause : If data are sent from the HOST to the generator, the

. generator then sends an answer note. If the answer

. contains the wrong function code, an error appears.

Active: Constantly

Possible trouble / help : Incompatible HOST and

. generator software

6011 HOST received GEN-note with wrong length


Cause : If data are sent from the HOST to the generator, the

. generator then sends an answer note. If the answer

. contains the wrong function code, an error appears.

Active: Constantly

Possible trouble / help : Incompatible HOST and

. generator software

6012 HOST received GEN-note with wrong function code

Cause : If data are sent from the HOST to the generator, the

. generator then sends an answer note. If the answer

. contains the wrong function code, an error appears.

Active: Constantly

Possible trouble / help : Incompatible HOST and

. generator software

6013: HOST tries to send undefined function code

Cause: The HOST tries to send an undefined function code to

. the generator.

Active: Constantly

Possible trouble / help : HOST software error

.
6014 HOST tries to send undefined parameter

Cause : HOST tries to send a defined function code to the

. generator, however, the parameters of the command

. are incorrect.

Active: Constantly

Possible trouble / help : HOST software error

6015 No ready response to HOST-GEN-note after n retries

Cause : The generator did not respond to a HOST note even

. though the HOST sent it several times.

Active: Constantly

Possible trouble / help : Fault in generator

. HOST-generator connection

. Interface in generator

6016 Host tries to send wrong message type

Cause : HOST tries to transmit wrong type of message.

Active: Constantly

Possible trouble / help : HOST software error


.

6018 HOST D3 GEN-interface parity/overrun/frame error

Cause : The interface chip for the generator on D3 detected

. an error. The note is discarded by the HOST.

Active: Constantly

Possible trouble / help : With infrequent entry - normal

. Board D3 defective

. Wrong baud rate from generator

6019 RMOS RIO receive error

Cause : Software driver in HOST detected receive error.

Active: Constantly

Possible trouble / help : System overload

601A RMOS RIO send error

Cause : Error detected by software driver in HOST.

Active: Constantly

Possible trouble / help : System overload

*****************************************
* 2.14 ERROR LIST INTENSITY INTERFACE *

*****************************************

7000 INT communication error intensity interface

7018 HOST D3 intensity interface parity/overrun/frame error

Cause : Interface chip for intensity interface on D3

. detected a fault. The note is discarded by the HOST.

Active: Constantly

Possible trouble / help : With infrequent entry - normal

. Board D3 defective

. Wrong TV baud rate

7019 HOST D3 intensity interface receive error

Cause : Error detected by software driver in HOST.

Active: Constantly

Possible trouble / help : System overload

702D intensity telegramm difference from menu-value


**************************************

* 2.15 ERROR LIST TERMINAL HANDLER *

**************************************

8000 HOST D3.X5-interface communication error

8018 HOST D3.X5-interface parity/overrun/frame error

Cause : Interface chip for service interface on D3 detected

. a fault. The note is discarded by the HOST.

Active: Constantly

Possible trouble / help : With infrequent entry - normal

. Board D3 defective

. Wrong baud rate from generator

8019 HOST D3.X5-interface RMOS RIO receive error

Cause : Receive error detected by software driver in HOST.

Active: Constantly

Possible trouble / help : System overload

.
801A HOST D3.X5-interface RMOS RIO send error

Cause : Error detected by software driver in HOST.

Active: Constantly

Possible trouble / help : System overload

**************************************

* 2.16 ERROR LIST MOTOR CONTROLLER *

**************************************

9000 Motorcontroller not ready

Cause :

Active: Constantly

Possible trouble / help : Board D5 defective

. (not plugged in)

. Hardware defective

9001 Collimator not on cassette format

Cause :

Active: Switch over to casette position

Possible trouble / help : Check mechanical movement


. Check parameters of colimator drive

. and calibrate colimator drive

. Check end position signal

. Check motor control D5

9010 At collimator-autoscanning no index from plane 1

Cause : No index will be found

Active: During power up

Possible trouble / help : Encoder disk 1 doesn't move

. Index mark not o.k.

. Light barrier defective

. Check parameters in FEPROM

9011 Plane 1 not exactly positioned

Cause : End position signal not o.k.

Active: Switch over to norm position or cassette position

Possible trouble / help : Check mechanical movement

. Check end position signal

. Check motor control D5

. Check parameters in FEPROM

.
9020 At collimator-autoscanning no index from plane 2

Cause : No index will be found

Active: During power up

Possible trouble / help : Encoder disk 2 doesn't move

. Index mark not o.k.

. Light barrier defective

. Check parameters in FEPROM

9021 Collimator Plane 2 not exactly positioned

Cause : End position signal not o.k.

Active: Switch over to norm position or cassette position

Possible trouble / help : Check mechanical movement

. Check end position signal

. Check motor control D5

. Check parameters in FEPROM

9030 At collimator-autoscanning no index from plane 3

Cause : No index will be found

Active: During power up


Possible trouble / help : Encoder disk 3 doesn't move

. Index mark not o.k.

. Light barrier defective

. Check parameters in FEPROM

9031 Collimator Plane 3 not exactly positioned

Cause : End position signal not o.k.

Active: Switch over to norm position or cassette position

Possible trouble / help : Check mechanical movement

. Check end position signal

. Check motor control D5

. Check parameters in FEPROM

9040 X-iris doesn't find index

Cause : No index will be found

Active: During power up

Possible trouble / help : Encoder disk doesn't move

. Index mark not o.k.

. Light barrier defective

. Check parameters in FEPROM

.
9041 X-iris > than zoom position

Cause : End position signal not o.k.

Active: Switch over to zoom position

Possible trouble / help : Check mechanical movement

. Check end position signal

. Check motor control D5

. Check parameters in FEPROM

9042 X-iris > than norm position

Cause : End position signal not o.k.

Active: Switch over to norm position

Possible trouble / help : Check mechanical movement

. Check end position signal

. Check motor control D5

. Check parameters in FEPROM

9043 X-iris position wrong

Cause : End position signal not o.k.

Active: Switch over to any position


Possible trouble / help : Check mechanical movement

. Check end position signal

. Check motor control D5

. Check parameters in FEPROM

9050 No index at camera-autoscanning

Cause : No index will be found

Active: During power up

Possible trouble / help : camera head doesn't move

. Index mark not o.k.

. Light barrier defective

9051 Camera-rotation driving to norm position not possible

Cause : End position signal not o.k.

Active: Switch over to norm position

Possible trouble / help : Check mechanical movement

. Check end position signal

. Check motor control D5

. Check parameters in FEPROM

.
9052 Camera-rotation position wrong

Cause : End position signal not o.k.

Active: Switch over to norm position

Possible trouble / help : Check mechanical movement

. Check end position signal

. Check motor control D5

. Check parameters in FEPROM

9060 Camera iris doesn't find index

Cause : No index signal (switch S32)

Active: During power up

Possible trouble / help : Check mechanical movement

. Check end position signal

. Check motor control D5

. Check parameters in FEPROM

9061 Camera iris can't be moved exactly to home position

home position

Cause : Movement from index to home (80 steps)

Active: During power up


Possible trouble / help : Check mechanical movement

. Check end position signal

. Check motor control D5

. Check D32, D30

9062 Camera iris not in zoom position

Cause :.

Active: During radiation

Possible trouble / help : Check mechanical movement

. Check end position signal

. Check motor control D5

9063 Camera iris not in norm position

Cause :

Active: During radiation

Possible trouble / help : Check mechanical movement

. Check end position signal

. Check motor control D5

9064 Camera-iris position wrong


Cause :.

Active: During power up or radiation

Possible trouble / help : comes with error 9061 to 9063

. Check mechanical movement

. Check end position signal

. Check motor control

9210 HOST D5 motor controler plane1 reset error

Cause :.

Active: During power up

Possible trouble / help : Only fatal error in service mode

. (Board D6 - S1 on 1)

. If there is also a problem with

. the motor - change board D5

9220 HOST D5 motor controler plane 2 reset error

Cause :.

Active: During power up

Possible trouble / help : Only fatal error in service mode

. (Board D6 - S1 on 1)
. If there is also a problem with

. the motor - change board D5

9230 HOST D5 motor controler plane 3 reset error

Cause :.

Active: During power up

Possible trouble / help : Only fatal error in service mode

. (Board D6 - S1 on 1)

. If there is also a problem with

. the motor - change board D5

9240 HOST D5 motor controler xiris reset error

Cause :.

Active: During power up

Possible trouble / help : Only fatal error in service mode

. (Board D6 - S1 on 1)

. If there is also a problem with

. the motor - change board D5

9250 HOST D5 motor controler camera rotation reset error


Cause :.

Active: During power up

Possible trouble / help : Only fatal error in service mode

. (Board D6 - S1 on 1)

. If there is also a problem with

. the motor - change board D5

9260 HOST D5 motor controler camera iris reset error

Cause :.

Active: During power up

Possible trouble / help : Only fatal error in service mode

. (Board D6 - S1 on 1)

. If there is also a problem with

. the motor - change board D5

9310 HOST D5 motor controler plane 1 INIT error

Cause :.

Active: During power up

Possible trouble / help : Only fatal error in service mode

. (Board D6 - S1 on 1)
. If there is also a problem with

. the motor - change board D5

9320 HOST D5 motor controler plane 2 INIT error

Cause :.

Active: During power up

Possible trouble / help : Only fatal error in service mode

. (Board D6 - S1 on 1)

. If there is also a problem with

. the motor - change board D5

9330 HOST D5 motor controler plane 3 INIT error

Cause :.

Active: During power up

Possible trouble / help : Only fatal error in service mode

. (Board D6 - S1 on 1)

. If there is also a problem with

. the motor - change board D5

9340 HOST D5 motor controler xiris INIT error


Cause :.

Active: During power up

Possible trouble / help : Only fatal error in service mode

. (Board D6 - S1 on 1)

. If there is also a problem with

. the motor - change board D5

9350 HOST D5 motor controler camera rotation INIT error

Cause :.

Active: During power up

Possible trouble / help : Only fatal error in service mode

. (Board D6 - S1 on 1)

. If there is also a problem with

. the motor - change board D5

9360 HOST D5 motor controler camera iris INIT error

Cause :.

Active: During power up

Possible trouble / help : Only fatal error in service mode

. (Board D6 - S1 on 1)
. If there is also a problem with

. the motor - change board D5

***********************************

* 2.17 ERROR LIST UNKNOWN ERROR *

***********************************

F000 Error of unknown component

Cause : HOST receives error message, however,

. can not allocate it to any component.

Active: Constantly

Possible trouble / help : Software error

***************************

* 3. POWER UP MESSAGE *

***************************

FFFF POWER up message

Cause :Power up identification for input of time and date.

. Useful in determining exactly when an error or fault

. occurred following switch-on.


Active during: Power up

Possible trouble / help : Normal, input to error memory

. following each switch-on.

***************************************************

* 4. Multispot 2000 status and error messages *

***************************************************

These messagges may depend on the software level used in the

Multispot 2000 camera.

WARM UP The camera runs through a warm up period.

As shown on the display the time is counted


down.

SELF TEST The camera is in the internal self test mode.

SYSTEM OK The program confirms that the camera


successfully

completed the self-test.

INSERT CASSETTE Program query prompting the user to insert the


film
cassette into the camera or to check whether
the

cassette was inserted correctly.

REMOVE DRK SLIDE The right dark slide is still in place. It has
to

be removed for exposure release.

CALIBRATING After the dark slide has been removed or the

brightness level has been changed, the camera

calibrates the internal brightness settings.

EXPOSING POS: 1/1 The video data is exposed on film

END OF FILM Last exposure depending on the film subdivision

selected. Reinsert the dark slide and remove


the

cassette.

NO VIDEO SIGNAL The diagnostic imaging device is not


transmitting

a Video signal to the camera. Verify that the

video signal is connected to the camera.


DARK SLIDE ERROR The exposure release button was pressed.
However,

the dark slide had not been removed. Pull out


the

dark slide.

NEG CAL ERROR The camera did not finish brightness


calibration

before exposure release. Try to increase the

brightness value and reduce the contrast.


Another

source for error may be the 75 Ohm termination


at

the video output of device. Ensure that the


cable is

connected.

POS CAL ERROR See NEG CAL ERROR

PHOTOCELL ERROR If this occurs immediately after the warm up


cycle

and the self test, press the exposure release


button.

If the display reads NO VIDEO SIGNAL, check the

video signal and the video connection.

BRT DRIVE ERROR The microprocessor of the camera does not


supply the

brightness control signal for the internal


monitor.

As a result, image generation is blocked.


Change camera.

CNT DRIVE ERROR The mictoprocessor of the camera does not


supply the

contrast control signal for the internal


monitor.

As a result, image generation is blocked.


Change camera.

MON 26V Fault Power supply defective.

MON P15V FAULT Power supply defective.

MON N15V FAULT Power supply defective.

MON 5V FAULT Power supply defective.

MEMORY FAULT Problems with the memory. All programmed


parameters

may have been deleted.

You might also like