You are on page 1of 6

Stefan Vandeweghe Accuracy of digital impressions of

Valentin Vervack
Melissa Dierens
multiple dental implants: an in vitro
Hugo De Bruyn study

Authors’ affiliations: Key words: dental implant, digital impression, edentulous, intra-oral scan
Stefan Vandeweghe, Valentin Vervack, Hugo De
Bruyn, Department of Periodontology, Oral
Implantology, Removable and Implant Prosthetics, Abstract
Dental School, Faculty of Medicine and Health Introduction: Studies demonstrated that the accuracy of intra-oral scanners can be compared with
Sciences, Ghent University, Ghent, Belgium
conventional impressions for most indications. However, little is known about their applicability to
Melissa Dierens, Oral and Maxillofacial Imaging
Unit, Dental School, University Hospital Ghent, take impressions of multiple implants.
Ghent, Belgium Aim: The aim of this study was to evaluate the accuracy of four intra-oral scanners when applied
Hugo De Bruyn, Department of Prosthodontics,
for implant impressions in the edentulous jaw.
Malm€ o University, Malm€o, Sweden
Material and methods: An acrylic mandibular cast containing six external connection implants
Corresponding author: (region 36, 34, 32, 42, 44 and 46) with PEEK scanbodies was scanned using four intra-oral scanners:
Stefan Vandeweghe, DDS, PhD
Department of Periodontology, Oral Implantology, the Lava C.O.S. and the 3M True Definition, Cerec Omnicam and 3Shape Trios. Each model was
Removable and Implant Prosthetics scanned 10 times with every intra-oral scanner. As a reference, a highly accurate laboratory scanner
University Hospital Ghent (104i, Imetric, Courgenay, Switzerland) was used. The scans were imported into metrology software
De Pintelaan 185, Poli 8
9000 Ghent (Geomagic Qualify 12) for analyses. Accuracy was measured in terms of trueness (comparing test
Belgium and reference) and precision (determining the deviation between different test scans). Mann–
Tel.: +32 9 332 59 22 Whitney U-test and Wilcoxon signed rank test were used to detect statistically significant
Fax: +32 9 332 15 26
e-mail: stefan.vandeweghe@ugent.be differences in trueness and precision respectively.
Results: The mean trueness was 0.112 mm for Lava COS, 0.035 mm for 3M TrueDef, 0.028 mm for
Trios and 0.061 mm for Cerec Omnicam. There was no statistically significant difference between
3M TrueDef and Trios (P = 0.262). Cerec Omnicam was less accurate than 3M TrueDef (P = 0.013)
and Trios (P = 0.005), but more accurate compared to Lava COS (P = 0.007). Lava COS was also less
accurate compared to 3M TrueDef (P = 0.005) and Trios (P = 0.005).The mean precision was
0.066 mm for Lava COS, 0.030 mm for 3M TrueDef, 0.033 mm for Trios and 0.059 mm for Cerec
Omnicam. There was no statistically significant difference between 3M TrueDef and Trios
(P = 0.119). Cerec Omnicam was less accurate compared to 3M TrueDef (P < 0.001) and Trios
(P < 0.001), but no difference was found with Lava COS (P = 0.169). Lava COS was also less accurate
compared to 3M TrueDef (P < 0.001) and Trios (P < 0.001).
Conclusions: Based on the findings of this in vitro study, the 3M True Definition and Trios scanner
demonstrated the highest accuracy. The Lava COS was found not suitable for taking implant
impressions for a cross-arch bridge in the edentulous jaw.

The introduction of CAD/CAM has simplified these errors. The virtual model used by the
and improved the workflow of fixed prosthetic cad software is almost immediately created
dentistry. Although many studies demonstrate using the data of the intra-oral scanner. As
a significant improvement in the accuracy of there is no need for a stone cast or a conven-
CAD/CAM compared to conventional cast tional impression, the dimensional errors
frameworks (Karl & Holst 2012), a significant that take place during these procedures can
error is still present. In the conventional work- be avoided (Johnson & Craig 1985; Millstein
flow, half of the misfit is introduced during the 1992). Theoretically, this could improve the
impression procedure and production of the accuracy and fit of the final prosthesis.
stone cast, while the other half occurs while Although passive fit is difficult to describe
manufacturing the prosthesis (Schneider et al. (Heckmann et al. 2004), Jemt & Lie (1995)
Date: 2001; Heckmann et al. 2004). Although cad- defined it as an accuracy level that does not
Accepted 20 March 2016
cam improved the accuracy of the latter, the cause any long-term clinical complications.
To cite this article: initial step in the workflow, taking an impres- According to the authors, discrepancies up to
Vandeweghe S, Vervack V, Dierens M, De Bruyn H. Accuracy
of digital impressions of multiple dental implants: an in vitro sion and pouring the cast, remained the same. 150 lm were acceptable. For implant-sup-
study.
Making a digital impression by means of ported reconstructions, the level of fit is even
Clin. Oral Impl. Res. 00, 2016, 1–6
doi: 10.1111/clr.12853 an intra-oral scanner may help to overcome more important compared to teeth-supported

© 2016 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd 1
Vandeweghe et al  Accuracy of digital implant impressions

prostheses. Dental implants have a reduced angulation did not affect the accuracy (Gime- The Trios scanner (3Shape, Copenhagen,
mobility which only exists because of the nez et al. 2014). Denmark) is based on confocal microscopy
flexibility of the bone (Kim et al. 2005). The The aim of this study was to evaluate the and continuously captures 2D images from
absence of a periodontal ligament prevents accuracy of various intra-oral scanners when different positions to create a 3D surface.
the implant from adapting to the ill-fitting used for implant impressions in the edentu- The recent version captures in colour and
framework, and as a consequence, stress will lous jaw. does not use contrast spraying.
be induced in the implant and framework
(Bacchi et al. 2013). Combined with an Material and methods Data analyses
increase in occlusal loading (Hammerle et al. All scans were imported into metrology soft-
1995), it is not surprising that the complica- Study protocol
ware (Geomagic Qualify 12, 3D Systems,
tion rate is higher for implant restorations Six regular dental implants with an external Rock Hill, SC, USA) for data analyses (Fig. 2).
compared to tooth-born restorations (Good- hex connection (IBT, Southern Implantsâ, The CAD file of the scan body was aligned
acre et al. 2003). Irene, South Africa) were placed in an acrylic to the six scanbodies, using a best-fit algo-
Several laboratory investigations have eval- model of the edentulous mandible at the posi- rithm with the tolerance set at 1 lm. Next,
uated the accuracy of intra-oral scanners. tion of the 1st molar, 1st premolar and lateral the six aligned scanbodies were saved as a
Ender & Mehl (2011) compared the accuracy incisor (Fig. 1). Inter-scan body distance and new file. This allowed to compare only the
of the Lava COS and Cerec Bluecam intra-oral angulation are depicted in Table 1. Six cus- scanbodies, thereby excluding all other irrele-
scanners with a conventional polyether tom-made intra-oral scanbodies in PEEK vant parts of the model.
impression and found that the accuracy of the (=polyether ether ketone) (Proscan, Zonhoven, As defined by Ender & Mehl (2011), accu-
digital impressions was better. Similarly, Belgium) were connected to the implants and racy consists of two parameters: trueness
Guth et al. (2013) compared the accuracy of tightened by hand to approximately 10Ncm. describes how close a test scan resembles the
the traditional CAD/CAM workflow with The model was scanned with a highly scan taken by a reference scanner, while pre-
intra-oral scanning and found the highest accurate optical digitizer (104i, Imetric, Cour- cision describes how much the various test
accuracy for the latter. However, most studies genay, Switzerland) and a 3D image was cre- scans differ from each other.
evaluating the accuracy of intra-oral scanners ated and exported as an open-format STL file. For this study, the trueness was evaluated
focus on tooth-supported restorations or eval- Similarly, the model was scanned 15 times by by comparing the various test scans obtained
uate only the data output of the digitizers. each intra-oral scanner according to the manu- from the intra-oral scanner with the reference
Making digital impressions of dental facturer’s recommendation. The five first scans scan from the Imetric 104i. Precision was
implants requires the use of scanbodies, were not used for analyses, to avoid inaccuracies evaluated by comparing the scans from each
which are easy to capture as most of the object caused by a learning curve. The remaining scans intra-oral scanner with one another. The pri-
is located supragingivally. On the other hand, were exported to an open file format (STL, Ply). mary outcome is thus to evaluate the accu-
the accuracy will also depend on the fit of For the Lava COS and 3M TrueDef, these files racy, in terms of trueness and precision, at
these scanbodies on the implants. Stimmel- could be downloaded directly from the cloud, the level of the implant scanbodies.
mayr et al. (2012) reported an average discrep- while for the 3Shape Trios and Cerec Omnicam, Test and reference scans were superim-
ancy in the fit of the scanbodies of 39 lm on the data needed to be imported into additional posed and aligned, using a best-fit algorithm
the original implants and only 11 lm on the software (Exocad, Darmstadt, Germany) to with the tolerance set at 0.001 lm. Then, a
implant analogues. Consequently, one could allow creation of an open STL fi 3D comparison was done, thereby calculating
assume that laboratory scanning of the stone the absolute mean deviation from the mean
cast would lead to fewer errors. Another Intra-oral scanners positive and negative deviation. The trueness
aspect of the implant workflow that may The LavaTM Chairside Oral Scanner (Lava was based on 10 comparisons per scanner,
cause small deviations in accuracy is the COS) was launched in 2008 by 3MTMESPE while the precision was calculated from 45
matching of the virtual scan body from the (Seefeld, Germany) and is based on the princi- comparisons.
cad software’s library on the virtual model. If ple of active wavefront sampling with struc-
the original scan lacks quality, this may result tured light projection. It allows data Statistics
in an incorrect matching of the scan body and capturing in a video sequence and models the Statistics were performed using SPSS v 20
an error in the position of the analogue in the data in real time. Light powder dusting of the (IBMâ, Armonk, NY, USA) with the level of
virtual model. dental arch is necessary to locate reference significance set at P = 0.05. Differences in
Scanning an edentulous jaw may be chal- points for the scanner. trueness and precision were first evaluated
lenging due to the lack of anatomical land- In 2012, the Lava COS was replaced by the using Friedman’s 2-way ANOVA test. Post
marks and the fact that all scanbodies are 3MTM True Definition (3M TrueDef) scanner hoc analyses to identify significant differ-
identical. Andriessen et al. (2014) experienced (3MTMESPE), which is an upgraded version of ences in trueness and precision in between
several problems when scanning two the Lava COS, with updated software and an the different intra-oral scanners were done
implants in the edentulous mandible because improved wand with a larger focus depth. It using Wilcoxon signed rank test.
the intra-oral scanner had difficulties to also uses light powder dusting.
make a distinction between both implants. The Cerec Omnicam (Sirona, Long Island
As a result, the majority of scans were use- City, NY, USA) is based on the concept of Results
less. active triangulation and uses a white light to
The inter-implant distance also seems to project a pattern on the object. It captures Mean values for trueness and precision are
affect the accuracy. A longer scanning dis- data continuously in colour, without the based on the pooled findings for all six scan-
tance resulted in an increased error. Implant need for contrast spraying. bodies analysed. Friedman’s test detected

2 | Clin. Oral Impl. Res. 0, 2016 / 1–6 © 2016 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd
Vandeweghe et al  Accuracy of digital implant impressions

(Schneider et al. 2001; Heckmann et al.


2004). To eliminate these errors, the concept
of digital impressions was introduced.
When evaluating the fit of a tooth-sup-
ported crown, digital impressions resulted in
a better fit compared to conventional impres-
sions and stone casts for single crowns (Syrek
et al. 2010; Ng et al. 2014) as well as for
fixed partial dentures (Almeida e Silva et al.
2014; Svanborg et al. 2014). Although digital
Fig. 1. The study model represented an edentulous jaw with six external connection implants. The intraoral scan- impressions clearly improved the framework
bodies were made in PEEK (polyether ether ketone).
fit of tooth-born restorations, its final accu-
racy will still vary, depending on the shape of
Table 1. Overview of the distance and angle between the adjacent implant scanbodies the preparations and the span of the frame-
Scan body 46–44 44–42 42–32 32–34 34–36 work.
Little research was done on the use of
Angle 0.57° 1.65° 4.62° 4.79° 4.22°
Distance, mm 9.51 6.61 10.28 7.28 5.70 intra-oral scanners for the full arch. Using a
full-arch model containing 14 tooth prepara-
tions, Patzelt et al. (2014) tested the accuracy
of four intra-oral scanners. The mean true-
ness was between 38 and 332.9 lm, while
the precision ranged from 37.9 to 99.1 lm.
The authors concluded that only one intra-
oral scanner (Lava C.O.S.) could be recom-
mended for use in the edentulous jaw. Ender
& Mehl (2015) evaluated several conventional
impression materials and intra-oral scanners.
The digital impressions of a full arch had a
trueness between 29 and 45 lm and a preci-
sion ranging from 19 to 63 lm, which was
Fig. 2. The metrology software produces a colour-map, demonstrating the size of the deviations between test and
not significantly better than the conventional
reference through different colours.
impressions. The authors stated that the
intra-oral scans demonstrated more local
significant differences for trueness (P < 0.001) Omnicam. There was no statistical signifi- deviations and their accuracy depended lar-
and precision (P < 0.001). cant difference between 3M TrueDef and gely on the scanning technique. Su & Sun
3Shape Trios (P = 0.119). Cerec Omnicam (2016) evaluated the precision of the Trios
Trueness was less accurate compared to 3M TrueDef scanner and compared it with a laboratory
The mean trueness was 0.112 mm (SD 0.025, (P < 0.001) and 3Shape Trios (P < 0.001), but scanner. Not only was the precision signifi-
range 0.084–0.174) for Lava COS, 0.035 mm no difference was found with Lava COS cantly lower for the intra-oral scanner, the
(SD 0.012, range 0.023–0.061) for 3M True- (P = 0.169). Lava COS was also less accurate deviation also increased with number of
Def, 0.028 mm (SD 0.007, range 0.021–0.044) compared to the 3M TrueDef (P < 0.001) and teeth scanned.
for 3Shape Trios and 0.061 mm (SD 0.023, 3Shape Trios (P < 0.001) (Fig. 3). Digital implant impressions may be even
range 0.029–0.099) for Cerec Omnicam. more challenging. Scanning a single implant
There was no statistically significant differ- can be done with high predictability, as was
ence between 3M TrueDef and 3Shape Trios Discussion shown in several studies and case reports
(P = 0.262). Cerec Omnicam was less accu- (Lee & Gallucci 2013; Lin et al. 2013; Abdel-
rate than 3M TrueDef (P = 0.013) and 3Shape A good fitting prosthesis is important to Azim et al. 2014; Joda et al. 2014). When
Trios (P = 0.005) but more accurate compared avoid complications and assure the longevity using a monolithic restoration, a complete
to Lava COS (P = 0.007). Lava COS was also of the construction. However, this is not easy digital workflow is possible as the abutment
less accurate compared to 3M TrueDef to achieve. The application of CAD/CAM in and crown are virtually designed and manu-
(P = 0.005) and 3Shape Trios (P = 0.005) dentistry has improved the accuracy of the factured in their final shape (Joda & Bragger
(Figs 3 and 4). frameworks, when compared to the conven- 2014). However, when scanning multiple
tional casted frames (de Franca et al. 2015). implants in an edentulous jaw, some difficul-
Precision Despite this improvement, implant-supported ties may arise. As multiple, identical scan-
The mean precision was 0.066 mm (SD frameworks still have demonstrated micro- bodies are used, it may be difficult for the
0.025, range 0.001–0.132) for Lava COS, gaps up to 38 lm, depending on the span of intra-oral scanner to distinguish one from
0.030 mm (SD 0.011, range 0.013–0.054) for the construction (Katsoulis et al. 2015). another and thus identify the correct location
3M TrueDef, 0.033 mm (SD 0.012, range The remaining misfit is largely caused by in the jaw. Intra-oral scanners that work with
0.005–0.057) for 3Shape Trios and 0.059 mm errors that occur during the impression tak- a photosystem may paste images of different
(SD 0.024, range 0.009–0.115) for Cerec ing and the production of the stone cast scanbodies on top of each other (Andriessen

© 2016 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd 3 | Clin. Oral Impl. Res. 0, 2016 / 1–6
Vandeweghe et al  Accuracy of digital implant impressions

Fig. 3. Boxplot representing the overall trueness and precision for the different scanners.

Fig. 4. Graph representing the mean trueness at every implant location for the different scanners.

4 | Clin. Oral Impl. Res. 0, 2016 / 1–6 © 2016 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd
Vandeweghe et al  Accuracy of digital implant impressions

et al. 2014). Gimenez et al. (2015) scanned of the arch due to an accumulation of regis- et al. 2014). Similarly, a larger inter-implant
six implants in the edentulous jaw, using the tration errors in the patched surfaces. distance combined with a flat mucosal sur-
Cerec Bluecam intra-oral scanner. Although Lava COS demonstrated the highest devia- face may result in a lack of reference points
the first quadrant could be captured quite tions for trueness as well as for precision and to enable correct stitching (Gimenez et al.
accurately, the trueness significantly wors- performed significantly worse compared to 2015). In our study, the implants were posi-
ened when also the second quadrant was the other scanners. However, its successor tioned rather close to each other. One can
recorded. Papaspyridakos et al. (2015) com- 3M TrueDef demonstrated the best results assume that, if the inter-implant distance
pared a digital impression system (Trios) with together with 3Shape Trios. This improve- would increase, the scanning process would
several conventional (polyether splinted and ment in accuracy is probably due to changes become more difficult, which could decrease
non-splinted) impression techniques for the made in the wand and software. The focus the accuracy.
registration of five implants in the edentu- depth was significantly enlarged, which sim- It is difficult to determine the acceptable
lous mandible. The authors found no signifi- plified the scanning procedure and, at the level of fit for implant-supported prosthesis.
cant differences between the digital and same time, reduced scanning errors. According to Jemt & Lie (1995), discrepancies
conventional impression method and con- One important difference between the vari- up to 150 lm will not induce clinical compli-
cluded that digital impressions could be used ous scanners was the use of contrast powder. cations. In contrast, others put this threshold
for implant impressions in the edentulous Lava COS and 3M TrueDef scanners used con- a lot lower, between 50 and 75 lm (Mor-
jaw. The mean deviation for the Trios intra- trast powder, while Cerec Omnicam and mann et al. 1985; Denissen et al. 2000; van
oral scanner was 19.18 lm, which is lower 3Shape Trios did not. When looking at the out- der Zel et al. 2001). As these numbers also
than the values observed in our study. This come, one cannot say that powdering resulted include the errors of the final processing and
can be explained by differences in the num- in a better or worse result. However, when a production of the framework, the scanning
ber of implants, a different implant connec- reflective or translucent material needs to be deviation must be below this threshold.
tion or a different design and fit of the scan scanned, it was shown that the contrast pow- Based on these numbers, one can conclude
body. Stimmelmayr et al. (2012) reported a der improved the accuracy by facilitating the that Lava COS cannot be used to take digital
significant difference in scan body fit reflection of the projected light pattern impressions for a large-span implant-sup-
between the actual original implants and lab- (Nedelcu & Persson 2014). The PEEK scanbod- ported construction. The other scanners
oratory analogues, in favour of the latter. In ies used in our study did not cause these issues demonstrated a level of accuracy which
the study by Papaspyridakos et al. (2015), and therefore explain why the powdering did seems clinically acceptable. Consequently,
conventional extra-oral scan bodies were not influence the outcome. clinicians should be aware that not every
used, which were significantly longer than Although the results of this in vitro study intra-oral scanner can be used for every indi-
the ones used in our study. As was reported are very promising, there are some limita- cation.
by Fluegge et al. (2015), the precision will tions when it comes to scanning in the oral
decrease when shorter and smaller (intra-oral) cavity. Scanning in the mouth may doubles
scanbodies are used. the error compared to scanning a model due Conclusions
The largest discrepancies were found at the to the different environment (Flugge et al.
extremities of the model, being the scanbod- 2013). Another difference between in vivo There was a significant difference in accuracy
ies located at position 36 and 46. As these and in vitro scanning is the stability of the between the different scanners. One intra-
are the endpoints of our scanned surface, all scanning surface. The shape of the mucosa oral scanner did not achieve the necessary
errors will occur somewhere in between and may change depending on the jaw move- level of accuracy to be used for large-span
thus lead to a maximal error at one of the ments, which complicates the scanning pro- implant-supported reconstructions. The other
endpoints. van der Meer et al. (2012) reported cedure because the latter depends on the scanners demonstrated an acceptable level of
an increase in distance error over the length presence of fixed reference points (Andriessen trueness and precision for this indication.

References
Abdel-Azim, T., Zandinejad, A., Elathamna, E., Lin, Bacchi, A., Consani, R.L., Mesquita, M.F. & Dos of obtaining full-arch dental impressions. Quin-
W. & Morton, D. (2014) The influence of digital Santos, M.B. (2013) Effect of framework tessence International 46: 9–17.
fabrication options on the accuracy of dental material and vertical misfit on stress distribu- Fluegge, T., Att, W., Metzger, M. & Nelson, K. (2015)
implant-based single units and complete-arch tion in implant-supported partial prosthesis A novel method to evaluate precision of optical
frameworks. International Journal of Oral and under load application: 3-d finite element analy- implant impressions with commercial scan bodies-
Maxillofacial Implants 29: 1281–1288. sis. Acta Odontologica Scandinavica 71: 1243– an experimental approach. Journal of Prosthodon-
Almeida e Silva, J.S., Erdelt, K., Edelhoff, D., Ara- 1249. tics. doi: 10.1111/jopr.12362. [Epub ahead of print]
ujo, E., Stimmelmayr, M., Vieira, L.C. & Guth, Denissen, H., Dozic, A., van der Zel, J. & van Waas, Flugge, T.V., Schlager, S., Nelson, K., Nahles, S. &
J.F. (2014) Marginal and internal fit of four-unit M. (2000) Marginal fit and short-term clinical per- Metzger, M.C. (2013) Precision of intraoral digital
zirconia fixed dental prostheses based on digital formance of porcelain-veneered cicero, cerec, and dental impressions with itero and extraoral digiti-
and conventional impression techniques. Clinical procera onlays. Journal of Prosthetic Dentistry zation with the itero and a model scanner. Amer-
Oral Investigations 18: 515–523. 84: 506–513. ican Journal of Orthodontics and Dentofacial
Andriessen, F.S., Rijkens, D.R., van der Meer, Ender, A. & Mehl, A. (2011) Full arch scans: con- Orthopedics 144: 471–478.
W.J. & Wismeijer, D.W. (2014) Applicability and ventional versus digital impressions–an in-vitro de Franca, D.G., Morais, M.H., das Neves, F.D. &
accuracy of an intraoral scanner for scanning study. International Journal of Computerized Barbosa, G.A. (2015) Influence of cad/cam on the
multiple implants in edentulous mandibles: a Dentistry 14: 11–21. fit accuracy of implant-supported zirconia and
pilot study. Journal of Prosthetic Dentistry 111: Ender, A. & Mehl, A. (2015) In-vitro evaluation of cobalt-chromium fixed dental prostheses. Journal
186–194. the accuracy of conventional and digital methods of Prosthetic Dentistry 113: 22–28.

© 2016 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd 5 | Clin. Oral Impl. Res. 0, 2016 / 1–6
Vandeweghe et al  Accuracy of digital implant impressions

Gimenez, B., Ozcan, M., Martinez-Rus, F. & Pra- scanbody technique” for emergence profile sup- ysis. Journal of Prosthetic Dentistry 112: 1461–
dies, G. (2014) Accuracy of a digital impression port. Clinical Oral Implants Research 25: 395– 1471.
system based on parallel confocal laser technol- 397. Ng, J., Ruse, D. & Wyatt, C. (2014) A comparison
ogy for implants with consideration of operator Johnson, G.H. & Craig, R.G. (1985) Accuracy of of the marginal fit of crowns fabricated with digi-
experience and implant angulation and depth. four types of rubber impression materials tal and conventional methods. Journal of Pros-
International Journal of Oral and Maxillofacial compared with time of pour and a repeat pour of thetic Dentistry 112: 555–560.
Implants 29: 853–862. models. Journal of Prosthetic Dentistry 53: Papaspyridakos, P., Gallucci, G.O., Chen, C.J.,
Gimenez, B., Ozcan, M., Martinez-Rus, F. & Pra- 484–490. Hanssen, S., Naert, I. & Vandenberghe, B.
dies, G. (2015) Accuracy of a digital impression Karl, M. & Holst, S. (2012) Strain development of (2015) Digital versus conventional implant
system based on active triangulation technology screw-retained implant-supported fixed restora- impressions for edentulous patients: accuracy
with blue light for implants: effect of clinically tions: procera implant bridge versus convention- outcomes. Clinical Oral Implants Research 27:
relevant parameters. Implant Dentistry 24: 498– ally cast restorations. The International Journal 465–472.
504. of Prosthodontics 25: 166–169. Patzelt, S.B., Emmanouilidi, A., Stampf, S., Strub,
Goodacre, C.J., Bernal, G., Rungcharassaeng, K. & Katsoulis, J., Muller, P., Mericske-Stern, R. & Blatz, J.R. & Att, W. (2014) Accuracy of full-arch scans
Kan, J.Y. (2003) Clinical complications with M.B. (2015) Cad/cam fabrication accuracy of long- using intraoral scanners. Clinical Oral Investiga-
implants and implant prostheses. Journal of Pros- vs. Short-span implant-supported fdps. Clinical tions 18: 1687–1694.
thetic Dentistry 90: 121–132. Oral Implants Research 26: 245–249. Schneider, A., Kurtzman, G.M. & Silverstein, L.H.
Guth, J.F., Keul, C., Stimmelmayr, M., Beuer, F. Kim, Y., Oh, T.J., Misch, C.E. & Wang, H.L. (2005) (2001) Improving implant framework passive fit
& Edelhoff, D. (2013) Accuracy of digital mod- Occlusal considerations in implant therapy: clini- and accuracy through the use of verification
els obtained by direct and indirect data captur- cal guidelines with biomechanical rationale. Clin- stents and casts. Journal of Dental Technology
ing. Clinical Oral Investigations 17: 1201– ical Oral Implants Research 16: 26–35. 18: 23–25.
1208. Lee, S.J. & Gallucci, G.O. (2013) Digital vs. Con- Stimmelmayr, M., Guth, J.F., Erdelt, K., Edelhoff,
Hammerle, C.H., Wagner, D., Bragger, U., Lussi, ventional implant impressions: efficiency out- D. & Beuer, F. (2012) Digital evaluation of the
A., Karayiannis, A., Joss, A. & Lang, N.P. (1995) comes. Clinical Oral Implants Research 24: 111– reproducibility of implant scanbody fit–an
Threshold of tactile sensitivity perceived with 115. in vitro study. Clinical Oral Investigations 16:
dental endosseous implants and natural teeth. Lin, W.S., Harris, B.T. & Morton, D. (2013) The use 851–856.
Clinical Oral Implants Research 6: 83–90. of a scannable impression coping and digital Su, T.S. & Sun, J. (2016) Comparison of repeata-
Heckmann, S.M., Karl, M., Wichmann, M.G., Win- impression technique to fabricate a customized bility between intraoral digital scanner and
ter, W., Graef, F. & Taylor, T.D. (2004) Cement anatomic abutment and zirconia restoration in extraoral digital scanner: an in-vitro study.
fixation and screw retention: parameters of pas- the esthetic zone. Journal of Prosthetic Dentistry Journal of Prosthodontic Research 59: 236–242.
sive fit. An in vitro study of three-unit implant- 109: 187–191. Svanborg, P., Skjerven, H., Carlsson, P., Eliasson,
supported fixed partial dentures. Clinical Oral van der Meer, W.J., Andriessen, F.S., Wismeijer, D. A., Karlsson, S. & Ortorp, A. (2014) Marginal and
Implants Research 15: 466–473. & Ren, Y. (2012) Application of intra-oral dental internal fit of cobalt-chromium fixed dental pros-
Jemt, T. & Lie, A. (1995) Accuracy of implant- scanners in the digital workflow of implantology. theses generated from digital and conventional
supported prostheses in the edentulous jaw: PLoS One 7: e43312. impressions. International Journal of Dentistry
analysis of precision of fit between cast gold- Millstein, P.L. (1992) Determining the accuracy of 2014: 534382.
alloy frameworks and master casts by means gypsum casts made from type iv dental stone. Syrek, A., Reich, G., Ranftl, D., Klein, C., Cerny, B.
of a three-dimensional photogrammetric tech- Journal of Oral Rehabilitation 19: 239–243. & Brodesser, J. (2010) Clinical evaluation of all-
nique. Clinical Oral Implants Research 6: Mormann, W., Brandestini, M., Ferru, A., Lutz, F. ceramic crowns fabricated from intraoral digital
172–180. & Krejci, I. (1985) [marginal adaptation of adhe- impressions based on the principle of active
Joda, T. & Bragger, U. (2014) Complete digital sive porcelain inlays in vitro]. Schweizerische wavefront sampling. Journal of Dentistry 38: 553–
workflow for the production of implant-supported Monatsschrift f€ur Zahnmedizin (1984) 95: 1118– 559.
single-unit monolithic crowns. Clinical Oral 1129. van der Zel, J.M., Vlaar, S., de Ruiter, W.J. & David-
Implants Research 25: 1304–1306. Nedelcu, R.G. & Persson, A.S. (2014) Scanning son, C. (2001) The cicero system for cad/cam fab-
Joda, T., Wittneben, J.G. & Bragger, U. (2014) Digi- accuracy and precision in 4 intraoral scanners: an rication of full-ceramic crowns. Journal of
tal implant impressions with the “individualized in vitro comparison based on 3-dimensional anal- Prosthetic Dentistry 85: 261–267.

6 | Clin. Oral Impl. Res. 0, 2016 / 1–6 © 2016 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd

You might also like