You are on page 1of 104

Instructions for Use

PACE 203H
Dual-Chamber Temporary Pacemaker

Manufacturer Distributor
Osypka Medical GmbH Dr.Osypka GmbH
Albert-Einstein-Strasse 3 Earl-H.-W ood-Strasse 1
D – 12489 Berlin, Germ any D - 79618 Rheinfelden, Germ any
Phone: +49 (30) 6392 8300 Phone: +49 (7623) 7405 - 0
Fax: +49 (30) 6392 8301 Fax: +49 (7623) 7405 - 160
e-m ail: m ail@osypkam ed.com e-m ail: m ail@osypka.de

5I-17-016X-B-16
Table of Contents

1 PREFACE 6

1.1 GENERAL 6
1.2 CHECKING THE DELIVERY 6
1.3 OPTIONAL ACCESSORIES 6
1.4 W RITING CONVENTIONS OF THIS MANUAL 6
2 PRODUCT DESCRIPTION 6

3 INDICATION 7

4 CONTRAINDICATION 8

5 POSSIBLE COMPLICATIONS 8

6 PRECAUTIONARY MEASURES AND WARNINGS 9

7 PATIENT SAFETY 11

8 ELECTROMAGNETIC COMPATIBILITY 11
TM
9 USE AND APPLICATION OF THE PACE 203H 13

9.1 DESIGN 13
9.2 OVERVIEW OF THE DISPLAY, DIALS AND KEYS 15
9.2.1 The Upper Parameter Display 15
9.2.2 The Lower Mode / Menu Display 15
9.2.3 The Dials 15
9.2.4 The Fixed Labeled Keys 16
9.2.5 The Softkeys and the Menus 16
9.2.6 Display Backlight 17
9.2.7 LEDs for Sensing and Stimulation 17
9.2.8 Acoustic Indications 17
TM
9.3 POWERING ON THE PACE 203H 17
TM
9.4 LOCKING/UNLOCKING THE PACE 203H 19
TM
9.5 POWERING OFF THE PACE 203H 19

2/104
9.6 MODES OF OPERATION 20
9.6.1 Ventricular Asynchronous (V00) Pacing 21
9.6.2 Atrial Asynchronous (A00) Pacing 21
9.6.3 A-V Sequential Asynchronous (D00) Pacing 21
9.6.4 Ventricular Inhibited (VVI) Pacing 21
9.6.5 Atrial Inhibited (AAI) Pacing 22
9.6.6 Atrial Synchronous (P-Tracking) (VDD) Pacing 22
9.6.7 Dual-Chamber Pacing and Sensing with Inhibition and Tracking DDD 22
9.6.8 The A-V Sequential, Atrial Inhibited (DAI) Pacing 22
9.6.9 The A-V Sequential, Ventricular Inhibited (DVI) Pacing 23
9.6.10 Atrial Sensing Atrial Synchronous (P-Tracking) (VAT) Pacing 23
9.6.11 Non-Pacing Mode 0D0 (Pause) 23
9.6.12 Atrial Trigger Modes DDD+AT and DAT 23
9.7 CHANGING THE MODE OF OPERATION 23
9.8 ADJUSTMENT OF RATE, STIMULATION AMPLITUDE, SENSITIVITY AND A-V DELAY USING THE
DIALS 25
9.8.1 Adjustment of Basic Rate 26
9.8.2 Adjustment of Stimulation Amplitude 27
9.8.3 Adjustment of Sensitivity 27
9.8.4 Adjustment of A-V Delay 27
9.9 USING THE AUTOMATIC FEATURES 28
9.9.1 Automatic Setting of A-V Delay with the Setting of Rate 29
9.9.2 Automatic Setting of PVARP with the Setting of Rate 29
9.9.3 Automatic Setting of MTR with the Setting of Rate 30
9.9.4 Automatic Adjustment of Sensitivity (Auto Sense) 30
9.9.4.1 How the Auto Sense Function Works 30
9.9.4.2 Activation of Auto Sense 31
9.9.4.3 Re-Initialization of Auto Sense 34
9.9.4.4 Deactivation of Auto Sense 34
9.10 THE HIGH RATE STIMULATION (ATRIAL OVERDRIVE) 35
9.11 STANDARD PROGRAMS 37
9.11.1 Parameters Stored in a Standard Program 38
9.11.2 Store Standard Programs 39
9.11.3 Recall Standard Programs 39
9.11.4 Recall Manufacturer’s Defaults 40
9.11.5 Change the Turn-On Program 40
9.12 OTHER PARAMETERS/OPTIONS SETTING 42
9.12.1 Adjustment of PVARP 42
9.12.2 Adjustment of MTR 43

3/104
9.12.3 Adjustment of Pulse Duration 44
9.12.4 Adjustment of Atrial Refractory Period ARP (AAI, AAT) 44
9.12.5 Setting Options 45
9.12.5.1 Enable and Disable Acoustic Indicator 45
9.12.5.2 Enable and Disable the Alarms 46
9.12.5.3 Enable and Disable Aatrial Trigger Function 46
9.12.5.4 Setting the User Language 47
9.13 STARTING EMERGENCY-STIMULATION 47
9.14 PAUSE FUNCTION 48
9.15 CONNECTING THE LEADS TO THE PACE 203H 49
9.15.1 Lead Types 49
9.15.2 Lead Connection Terminals 50
9.15.3 Connection Configurations 50
9.15.4 Connecting the Leads 51
9.16 DETERMINING THE SENSITIVITY THRESHOLD 52
9.17 DETERMINING THE CARDIAC CAPTURE THRESHOLD 52
9.18 MAINS OPERATION (OPTION) 53
9.19 CONTROLLING AN INTRA-AORTIC BALLOON PUMP (OPTION) 53
®
9.20 CONTROL BY MONITOR AESCULON (OPTION) 55
9.21 INTERNAL SURVEILLANCE AND SAFETY FEATURES 56
9.21.1 Battery Surveillance 56
9.21.2 Lead Surveillance 56
9.21.3 Interference and Mode Switching 58
9.21.4 Crosstalk and Ventricular Safety Pacing 59
9.21.4.1 Ventricular Blanking Period 59
9.21.4.2 Crosstalk Sensing Window and Ventricular Safety Pacing 59
9.21.5 Premature Ventricular Contraction (PVC) 60
9.21.5.1 Isolated PVC, or Initial PVC 61
9.21.5.2 Subsequent PVCs 61
9.21.5.3 Absolute und relative PVARP 61
9.21.6 Runaway Protection 62
9.21.7 Protection Against Abnormal Settings 62
9.21.7.1 Possible Conflicts 63
9.21.7.2 Mode Switching 65
9.21.8 Effects from Therapeutc and Diagnostic Energy Sources 66
9.21.8.1 Defibrillation 66
9.21.8.2 RF Surgery 66

4/104
9.21.9 Summary of Text Messages 66
9.21.9.1 Warning Messages Related to Conflicting Parameter Settings 66
9.21.9.2 Warning Messages Related to Application Errors 70
9.21.9.3 Warning Messages Related to Device Errors 72
9.21.9.4 Informative Messages 73
10 STORAGE 76

11 CARE AND MAINTENANCE 77

11.1 CLEAR AND CLEANING 77


11.2 CHANGING THE BATTERY 77
11.3 SAFETY CHECK-UPS OF THE PACEMAKER 78
11.4 PRODUCT RETURN POLICY 79
12 CUSTOMER SERVICE 79

13 TECHNICAL DATA 80

14 DELIVERY UNIT 84

15 CONDITIONS OF GUARANTEE AND LIABILITY RESTRICTIONS 84

16 APPENDIX A: GLOSSARY 86

17 APPENDIX B: MENU TREE 90

18 APPENDIX C: EMC GUIDANCE AND MANUFACTURER’S


DECLARATION 92

19 APPENDIX D: FIGURES 98

20 APPENDIX E: TABLES 100

21 APPENDIX F: DECLARATION OF CONFORMITY 104

5/104
TM
When the BPI 202 is used, commonly
1 Preface available intra-aortic balloon pulsation
pumps can be synchronized with the
intra-cardiac ECG, instead of surface
1.1 General ECG synchronization.
Read these instructions carefully before
using the product described within. 1.4 Writing Conventions of this
Should you have any questions about Manual
these instructions or the use of this
product, please contact the customer In this instruction manual, certain
service department before using the conventions are used.
product: Keys and displays are represented in
Phone +49 (7623) – 7405 - 0 the text as follows:

The product may only be placed in Fixed labeled keys and dials: ON, OFF,
service when its proper use can be Pause, Unlock/Lock, V-STIM
assured. Upper Display: AUTO, A-TRIG.
Lower display und soft-labeled keys:
According to European Standards, the
Main menu, START
PACE 203H is a Class IIb medical
product (Council Directive 93/42/EEC of Helpful hints and notes on the usage of
14 Juni 1993 (‘Medical Device the device and for understanding the
Directive’), Annex IX). modes of operation will be introduced by
’Notes’.
1.2 Checking the Delivery Important facts and warnings to be
observed are introduced by 'Warning'.
Unpack the product and carefully check
to see if any damage has occurred in
shipment. Check to see if everything 2 Product Description
was delivered as listed on the shipping
list. The PACE 203H™ is a dual chamber
external cardiac pacemaker with atrial
Please let us know immediately if controlled timing for routine temporary
something is missing or damaged. heart stimulation. All the usual modes of
Claims that are filed afterwards cannot stimulation are available for treatment of
be considered. acute brady-arrhythmia and for pre-,
intra-, and post-operative stimulation of
1.3 Optional Accessories the heart. The stimulation parameters
are easily adjustable by rotating dials
Optionally the PACE 203H™ can be through a wide range of values.
ordered with a wall power supply. This
allows the pacemaker to be powered by The PACE 203H™ offers the possibility
AC mains. for atrial overdrive stimulation, or rapid
TM
In addition, the BPI 202 intra-aortic atrial pacing, for terminating supra-
baloon pump (IABP) interface is ventricular tachycardia. The rate of the
available as an optional accessory. overdrive stimulation is adjustable within

6/104
a wide range and is independent of the An Unlock/Lock button protects against
selected stimulation rate. The overdrive accidental change of the set
rate can be determined before and parameters.
changed during overdrive therapy. If The set parameters and (error)
required, overdrive stimulation can be messages are shown on a liquid crystal
initiated with the touch of a button. The display.
overdrive stimulation is indicated The detection of the intrinsic heart
optically and acoustically. activity as well as the emission of
The battery operated device can be stimulation impulses are shown
affixed to the patient’s arm by the arm separately by blinking LEDs for both
strap included. The housing is protected atrium and ventricle. Additionally, a
against accidental liquid spills. beep-tone can be switched on whenever
desired.
Optionally, the pacemaker can be System malfunctions that occur are
connected to a wall power supply for indicated optically and acoustically.
AC mains operation. Then the inserted
battery ensures safe operation in case A lead surveillance system indicates
of mains power failure. In AC mains interruptions and short circuits.
operation the battery will not be When a battery change is required,
discharged. optical and acoustic alerts are provided.
Optional AC mains operation without
The enclosure is protected against discharging the inserted (safety-)
accidental wetting. battery.
The functional design of the During dual chamber pacing, an
PACE 203H™ allows safe and easy automatic mode for adapting A-V delay,
operation with regards to all maximum tracking rate (MTR), and
requirements of DDD stimulation. PVARP is available.
The PACE 203H™ offers further the An automatic mode for adjusting the
following features: sensitivity in both the atrium and
ventricle may be chosen.
During battery changes, stimulation will A Pause function is available for easy
be maintained for at least 30 s. determination and measurement of the
A non-volatile memory keeps any patient’s intrinsic heart activity.
desired stand-by program ready for use, For controlling a balloon pulsation pump,
even if the device is shut off. or other peripheral devices, there is an
A standard program, which can be optional interface available.
individualized, is available for each All further technical functions of the
primary pacing mode. PACE 203H™ will be described in detail
An emergency program can be “called in the following chapters
up” by pressing an emergency key.
A burst- and a ramp function are
available for atrial overdrive-stimulation. 3 Indication
The external pacemaker PACE 203H™
can be used together with a stimulation

7/104
lead system for temporary atrial, chaotic frequency of detected fibrillation
ventricular or A-V sequential stimulation waves.
in the clinical environment. The device
can be used for therapeutic or Overdrive-stimulation therapy must only
diagnostic purposes as well as be used in the atrium. Overdrive-
prophylactically. stimulation in the ventricle could cause
life threatening ventricular fibrillation.
Specific indications for temporary, anti-
bradycardia stimulation (among others)
are: 5 Possible Complications
Total or intermittent heart block; When using an external pacemaker
Symptomatic sinus-bradycardia; such as PACE 203H™, the following
complications can arise:
Sick sinus syndrome;
Atrial and/or ventricular ectopic Infection.
arrhythmia; ⇒ Sepsis
Acute myocardial infarction with heart Thrombosis and pulmonary embolism.
block; ⇒ Death.
Temporary stimulation during asystole; Perforation of the heart.
Temporary support for hemodynamic ⇒ Hemopericardium. Hemothorax.
optimization after heart surgery; Cardiac tamponade
Temporary use during the exchange of Muscle and nerve stimulation.
an implantable pacemaker; ⇒ Patient discomfort.
Temporary stimulation and control Dislocation of lead.
before implanting a pacemaker. ⇒ System malfunction. Failure to pace.
Indication for atrial overdrive-stimulation Disconnection or breakage of lead
(among others): Contact problems at connection sites.
Insufficient tightening of the collets.
Supra-ventricular tachycardia.
⇒ Intermittent or complete failure of
effective stimulation and/or sensing.
4 Contraindication Dislocation of LV lead near atrium.
There are no contraindications with ⇒ atrial oversensing and ventricular
regards to the use of the PACE 203H™ inhibition.
for temporary cardiac stimulation for Significant rise in the stimulation
therapy and prevention of arrhythmia. threshold
The state of health of the patient, ⇒ Loss of effectiveness of the
however, can restrict the choice of stimulation (exit block).
operational mode and stimulation Significant drop of the ECG-signal
parameters. amplitude after lead dislocation or
For example, a mode of operation with ingrown lead.
atrial sensing is not suitable or ⇒ Loss of sensing (entrance-block).
appropriate when atrial fibrillation
occurs. This is due to excessive and

8/104
Abnormal pacemaker settings. with defibrillators or electro-surgical
⇒ Erratic rhythm. Compromise in stroke instruments.
volume / cardiac output.
Inappropriately high sensitivity setting.
Sensing of R or T waves in the atrium or 6 Precautionary Measures and
P waves in the ventricle.
Detection of interference Warnings
(noise, electromagnetic interference). The following list presents important
⇒ Ventricular tachycardia, ventricular precautionary measures and warnings.
fibrillation, and death, if not recognized Further important precautionary
immediately. measures and warnings will be found in
Time mismatch between intracardiac the following chapters.
conduction and pacemaker settings. In order to prevent unnecessary
⇒ Pacemaker mediated tachycardia. (In complications, the PACE 203H™ should
order to prevent this, some decision only be applied and used by medical
overrides are implemented in the personnel with extensive experience in
PACE 203H™. However, it is not cardiac stimulation therapy. Additionally,
possible to completely prevent the the person using the device should be
possibility of a pacemaker mediated thoroughly familiar with the contents of
tachycardia.) this instruction manual.
Overdrive stimulation in the atrium = All lead systems are to be connected to
rapid atrial pacing. type CF devices only, because of the
⇒ Accidental conduction into the danger of current being diverted to the
ventricle can create ventricular heart. Devices that are connected to a
arrhythmia. main supply pose increased danger for
Battery failure or exhaustion. current diversions to the heart.
⇒ Failure of impulse emission. Failure Make sure that all devices in the vicinity
to pace. of the patient are properly grounded.
Technical defect in the PACE 203H™ The stimulation leads provide a direct,
(failure of components). low-resistance current path to the heart.
⇒ Failure or change in the impulse Therefore, it is absolutely necessary that
emission, changed (or no) sensing, the connector plug is not touched with
incorrect displays. Failure to pace. bare hands or come in contact with
Undetected programming errors. electrically conductive or wet surfaces.
⇒ Chaotic rhythm. All possible static electricity sources
must be kept away from the stimulation
Erroneous lead connection. system.
Device does not function properly.
Chaotic rhythm. Failure to pace as While the lead is being inserted and
intended. connected to the pacemaker, continuous
ECG-monitoring is mandatory. For
Influence of defibrillation and RF emergency situations, a defibrillator
surgery. should always be available in a ready-to-
⇒ See chapter 8 for the effects while use state.
simultaneous use of the PACE 203H™

9/104
ECG-monitoring should be continued If the PACE 203H is to be used for a
during all the time the PACE 203H™ is longer period of time on a patient, the
in use for the patient to signalise stimulation threshold should be checked
possible complications of different from time to time (the first time after a
reasons immediately to the medical few hours, then daily), since an increase
staff. of the threshold may occur.
In either case it is necessary to monitor An unnecessarily high sensitivity (small
the patient and to be prepared for a sensitivity value) increases the
failure of the pacemaker function. A probability that proper pacemaker
back-up device for the pacemaker must functioning will be affected by external
be available. interference and the device will switch to
Error messages and conflict warnings of asynchronous stimulation (see also
the PACE 203H™ does not replace an chapter 9.21.3). If there are strong
ECG monitoring. electromagnetic fields caused by
During atrial overdrive-stimulation an telecommunication devices (like mobile
accidental conduction into the ventricle phones) or other sources an
is possible and may cause ventricular asynchronous mode should be set with
tachycardia. Continuous ECG- a higher than the patient’s intrinsic rate.
monitoring is therefore mandatory. A During dual chamber pacing there is an
defibrillator should always be available inherent potential risk of cross-
and ready-to-use. stimulation which is a cross-talk of an
If the PACE 203H™ operates in an atrial stimulation or depolarization
asynchronous mode the pacing pulses impulse into the ventricle or vice versa.
may occur during the vulnerable phase The PACE 203H™ is designed in such a
of the patient’s intrinsic activity way that this cross-talk is minimized.
(corresponds approximately in the ECG Furthermore, the distance between atrial
with the T-wave) and may cause and ventricular lead system should be
ventricular fibrillation and ventricular not less than 4 cm. Due to differences in
flutter respectively. the anatomy, electro-physiological
conditions and the location of the
It is mandatory to continuously monitor electrodes, cross-stimulation can,
the patient and to be prepared for a however, not be completely prevented.
possible failure or malfunctioning of the Therefore, when the device is switched
pacemaker when the PACE 203H™ is on, or when the stimulation parameters
used in conjunction with electro-surgical are changed, the user must make sure
instruments or defibrillators. that no cross-stimulation occurs. If this
To protect the patient and the should happen, one can try to eliminate
pacemaker from current passing through the effect by adapting the stimulation
the pacemaker/lead-circuitry caused by amplitudes or by exchanging the polarity
defibrillation discharges, the stimulation of the leads. If this is not possible, the
circuit should always be opened during pacemaker must, in this special case, be
defibrillation, if possible. Current flows used in a single-chamber mode without
caused by defibrillator discharges can connection of the leads of the other
endanger the patient. High current can channel.
also damage the pacemaker.

10/104
The PACE 203H™ prevents abnormal can stop operating as a result of an
settings (see chapter 9.21.7). unpredictable component failure. The
In order to provide continuous operation patient should not left unattended by the
of the pacemaker during battery medical staff. In addition, the patient
changes, the battery must not be left in should be supervised by a monitoring
the device until it is completely drained system.
(see also 9.21.1).
If the device is switched off, or to stand- 7 Patient Safety
by, after the request to change the
battery appears, the battery must be The temporary cardiac pacemaker
changed before the pacemaker is turned PACE 203H™ meets all applicable
on again. international standards for patient
safety:
In case that the PACE 203H™ is not
used for long periods of time, the battery IEC 60601-1 Medical electrical
must be removed in order to prevent equipment.
damage from possible battery acid General requirements
leakage. (Such damage cannot be for safety
compensated under the guarantee). IEC 60601-1-2 Medical electrical
equipment.
The pacemaker must not be submerged Electromagnetic
in either water or any other cleaning compatibility
solution. Do not use any scrubbing IEC 60601-2-31 Medical electrical
powder/liquid on the device. equipment.
The device must not be sterilized in an External cardiac
autoclave. Sterilization with plasma, pacemakers
ultrasound or gamma radiation is also
not allowed. The PACE 203H™ can be The sum of leakage currents may
damaged by such procedures. exceed the allowable limits in the event
Connector cables, intended for single the patient is connected to multiple
use, cannot be re-sterilized and electrical devices.
reinserted into a patient. 8 Electromagnetic Compatibility
Only the manufacturer, or facilities The temporary cardiac pacemaker
authorized by the manufacturer, can PACE 203H™ meets all applicable
perform repairs or calibration of the international standards with respect to
PACE 203H™. electromagnetic compatibility (EMC).
All automatic settings provided by the
IEC 60601-1-2 Medical electrical
PACE 203H™ are only a guide to assist equipment.
the user in finding the appropriate Electromagnetic
settings. It is the responsibility of the compatibility
operator to determine if these settings CISPR 11 Industrial, scientific and
are correct. medical radio-frequency
All battery powered devices can loss equipment.
their function because of normal Electromagnetic
disturbances
discharge of the battery. Furthermore, it
must be considered that the pacemaker

11/104
IEC 61000 Electromagnetic
compatibility
Some sources of radiated emissions
may impact the proper function of
PACE 203H:
• Electrosurgical instruments
operating with radio frequency,
• Devices for diathermy therapy,
• Magnetic resonance imaging (MRI);
• Medical telemetry systems;
Special electromagnetic compatibility
(EMC) precautions should be taken
when using medical electrical equipment
such as PACE 203H™:
A Portable and mobile RF communications
equipment, such as mobile phones, can
affect the function of the AESCULON.
Mobile phones with a rated maximum
output power of 2 Watt and a transmitter
frequency up to 2.5 GHz should be used
no closer to any part of the AESCULON
(including patient cables and sensors)
than a recommended separation distance
of 10 m (30 ft).
B Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic materials, the relative humidity
should be at least 30 %.
C Power frequency magnetic fields should
be at levels characteristic of typical
commercial or hospital environments.

Please find further information and


guidance regarding EMC in chapter 18.

12/104
TM
9 Use and Application of the PACE 203H
9.1 Design
The PACE 203H™ with its displays, keys, dials and terminals is shown in Figure 1.

Figure 1 PACE 203H™ Display Face


On the rear side of the device (see Figure 2) there is a metal hanger. This hanger deploys
in three snap-in positions. Thus, it can be used as a hanger, as a stand on a table top, or it
can be folded away. The easiest way to adjust the hanger is to grasp it at the sides and
pull away from the pacemaker frame.
Furthermore, the rear side features a battery compartment release button. This button
opens the battery compartment. The battery compartment is located at the bottom of the

13/104
device. A button cover prevents the release button from being unintentionally pressed (see
chapter 11.2)

Figure 2 Bottom View

14/104
various messages These messages can
9.2 Overview of the Display, Dials be informative, as well as alerting the
and Keys user regarding errors and conflicts

9.2.1 The Upper Parameter Display If the lower display is off, then pressing
the key Lock/Unlock turns the display
The six parameters, basic rate, atrial on.
sensitivity, atrial pulse amplitude, A-V
delay, ventricular pulse amplitude, After 2½…3 min, without any user
ventricular sensitivity, as well as the operation (turn of dials or pressing of
pacing mode code (NBG code) and keys), the lower display disappears for
additional symbols, are continuously battery power saving.
displayed as shown in Figure 3.
9.2.3 The Dials
There are six dials corresponding to the
parameter segments in the upper
parameter display. The six parameters
basic rate (& atrial overdrive rate), atrial
sensitivity, atrial pulse amplitude, A-V
delay, ventricular pulse amplitude and
ventricular sensitivity are adjustable by
rotating these dials (see Table 1).
When the dials are rotated, there is both
a tactile (notched click) and acoustic
response (beep).
Dial Function

Adjusting the basic rate


Figure 3: Upper Display (HI-)RATE and the overdrive
stimulation rate.
On the left, a sample DDD mode display
with numerous extra symbols as Adjusting the atrial
A-SENSE
described later in this manual is shown. sensitivity.
Adjusting the atrial pulse
If parameters have no meaning in the A-STIM
amplitude.
mode chosen, for example, atrial
stimulation amplitude, atrial sensitivity A-V DLY Adjusting the A-V delay.
and A-V delay in VVI mode, the
corresponding section of the display will Adjusting the ventricular
V-STIM
pulse amplitude.
be empty. Such a sample display is
shown on the right of Figure 3.
Adjusting the ventricular
V-SENSE
sensitivity.
9.2.2 The Lower Mode / Menu Display
Table 1: Dials and their functions
The lower display is used to show
selection menus (see chapter 9.2.5) and

15/104
9.2.4 The Fixed Labeled Keys Mode Menu Key No.
Fixed labeled keys according to Table 2 DDD
are implemented. D00 DVI DAI  1

VVI
V00  2
Key Function
ON Turning on the device. AAI
OFF Turning off the device. A00 AAT  3
Setting emergency
Emergency VDD
stimulation parameters.
V00 VVI VAT  4
Unlocking/Locking the dials
Lock/Unlock
and the keys.
Disable stimulation (as long Main Menu 5
Pause
as pressed)
Table 2: Fixed labeled keys and their Table 3: Mode Menu
functions

9.2.5 The Softkeys and the Menus All other functions are placed in the
second, third or even forth menu level.
For setting of the pacing mode, and To access these features, the Main
other parameters, and functions, five Menu must first be entered by pressing
additional keys, so called softkeys, are the softkey labeled Main Menu. If the
implemented. These keys are used in lower display is off, then pressing the
conjunction with, and located next to, the
key Unlock/Lock turns the display on.
lower display. Each softkey selects the
The Main Menu consists of two parts. It
option that is displayed adjacent to it in
appears as shown in Table 4.
different menus shown in the lower
display.
As a general rule the Mode Menu
(Table 3). appears first when the lower
display is turned on. The mode selection
is located in the top menu level to allow
a quick access.

16/104
Main Menu 1 Key No. 9.2.7 LEDs for Sensing and
Stimulation
High-Rate 1 To the left of the upper display there are
four LEDs for indicating atrial and
ventricular sensing and stimulation
Auto 2
events. The LEDs for indicating sensing
flash green and those for indicating
Standard 3 stimulation flash yellow.
All LEDs appear brightly shortly after
Parameters
4 turning on the PACE 203H™ to verify
Options
that they are operational.

↑_ 5 9.2.8 Acoustic Indications


Table 4: Main Menu Key pressing results in a high-frequency
acoustic signal (beep).
From here, the High-Rate (overdrive)
stimulation, the general Warnings are acoustically indicated by
parameter/options settings, the three successive short high frequency
parameters of the cardiac acoustic signals (beeps), whereas errors
resynchronization therapy (CRT), the are indicated by one long high frequency
automatic settings feature, the recall of signal (long beep).
standard programs, the lead impedance If the Beep feature is enabled (see
measurement, and further parameter chapter 9.12.5.1), sensing and
settings and options can be accessed. stimulation events are acoustically
Note: Pressing the softkey ↵ in any indicated (beep). Sensing and
menu always returns to the Main Menu. stimulation are indicated by a different
high frequency signal (beep).
Specifically, sensing occurs with a lower
9.2.6 Display Backlight
frequency signal than stimulation.
Both the upper and lower displays are
backlit. The backlight is deployed as 9.3 Powering On The PACE 203H
TM
long as the device is unlocked.
The PACE 203H™ is powered on by
After 30 seconds without any user pressing the key labeled ON
operation (rotation of dials or pressing
keys), the backlight is automatically Subsequently, the PACE 203H™
turned off for battery power saving. This performs a self-test.
occurs in parallel upon locking the dials
and keys. If the device was set to Stand-by (see
chapter 9.5) it starts operation
If the battery voltage decreases below a immediately with the last saved
specified battery depletion level, the parameter setting.
backlight is automatically deactivated
and will not be turned on again until the If the device was OFF, the key
battery has been replaced. Lock/Unlock, which is located to the

17/104
right of the softkeys, must be pressed Release Lock/Unlock key
and released to indicate that it is request
functioning properly. The PACE 203H™
indicates this by the schematized key Release
shown in the lower display (Table 5 and
Table 6).

Press Lock/Unlock key


request
Version n.nn
Press
yyyy/mm/dd

Table 6: Release Lock/Unlock Key


Request

Version n.nn Note: Version n.nn indicates the


hardware and/or software version of the
yyyy/mm/dd device, and yyyy/mm/dd the
corresponding release date
(Year/Month/Day).
Table 5: Press Lock/Unlock Key
Request
When the key Unlock/Lock has been
pressed and released, the
PACE 203H™ begins to function with
the Turn-on program (see chapter
9.11.5). Upon shipping from the
manufacturer, a DDD mode with high
sensitivity and moderate stimulation
amplitudes is preset. This setting
provides safe therapy for most pacing
situations (see Table 34). The Turn-on
program, however, may be customized
according to the physician’s
preferences.
The language setting (see chapter
9.12.5.4) and acoustic indication of
sensing and stimulation (see chapter

18/104
9.2.8) will remain unchanged during any dials and menu operation of the
power off. PACE 203H™.
Whether the device is locked or
unlocked, these conditions will be
Note: If the indication to press and continuously indicated by the lock
release the key Unlock/Lock is not symbol (Figure 4) in the upper display.
followed within 30 sec, an error
message
Startup timeout
(Press Unlock)
Figure 4: Lock Symbol
will appear, and the device will switch
off. This prevents discharge of the
battery, if the key labelled ON has been If a key is pressed, or a dial is rotated
pressed accidentally. while the device is locked, a warning
Note: Please do not press any key beep will be emitted and the lock symbol
during the self-testing after turning on will blink for 2 seconds.
the PACE 203H™. Among other tests, a
key check is performed during this
period. If a key is pressed, a request will Note: For High-Rate Stimulation Stand-
be indicated as follows by (see chapter 9.10) the time of
operating freedom is 60 seconds after
Release all keys last operation.
If the operator is not in compliance with
this request, or if a key is ‚stuck’, the 9.5 Powering Off The PACE 203H
TM

message
The PACE 203H is powered off by
Keyboard error pressing the key labeled OFF.
will appear in the lower display after 10 In order to prevent powering off the
seconds. device by mistake, a menu (Table 7)
appears in the lower menu with the
9.4 Locking/Unlocking the PACE softkeys labeled Off and Stand-by.
TM
203H
In order to prevent an accidental change
of pacing parameters, the device will be
locked automatically, if no dial has been
turned or no key has been pressed for
30 sec.
If the device is locked, pressing the key
Lock/Unlock performs an unlock
function. If the device is unlocked,
pressing the key Lock/Unlock disables

19/104
Power-Off Menu Key No. of a complete pacing cycle, measured in
milliseconds (ms). It is determined as
Off
(no storage) 1 the inverse of the basic rate.
Wenckebach behavior is always
2 activated in atrioventricular modes
incorporating atrial sensing. In case of
Stand-by high spontaneous intrinsic atrial rates,
(store data) 3
the P-V delay will be prolonged in order
to prevent high ventricular stimulation
4 rates. This function limits the ventricular
rate to the so-called
Maximum Tracking Rate (MTR). As a
↑_ 5
consequence, Wenckebach behavior
Table 7: Turn-off Menu intermittently blocks the conduction of
intrinsic atrial beats to the ventricular
chamber.
Upon selecting Off (no storage) by The A-V delay (a pacemaker parameter)
pressing the corresponding softkey, the is the programmed atrioventricular
PACE 203H™ switches off without pacing interval, initiated by an atrial
storage of the actual settings, prompting stimulus.
the message “Off (no storage)“.
The P-V delay (a pacemaker parameter)
Upon selecting “Stand-by (store data)” is the programmed time interval between
by pressing the corresponding softkey, atrial sensing (P wave) and the
the PACE 203H™ stores the actual ventricular stimulus. The P-V delay can
settings, prompting the message, be interpreted as a programmable
“Stand-by (store data)”, and then conduction delay window. It is
switches off. The PACE 203H™ does determined by the A-V delay and by the
not consume any battery power in the Maximum Tracking Rate (MTR).
Stand-by mode.
The A-V interval is the measured time
When the PACE 203H™ is powered on interval between a sensed or paced
from the Stand-by mode, it will, after atrial event and the subsequent
switching on, function with these prior ventricular sensed or paced event. In
settings, and not with the Turn-on case of spontaneous ventricular
program. depolarization, the A-V interval is shorter
The PACE 203H™ switches off with a than the programmed A-V delay.
short acoustic signal (beep).
The V-A delay (a pacemaker parameter)
is the programmed ventriculoatrial
9.6 Modes of Operation interval, initiated by a sensed or paced
ventricular event. Known also as the
The following terms are defined:
Atrial Escape Interval, it is the interval
The beat-to-beat interval (a pacemaker from a sensed or paced ventricular
parameter) is the programmed interval event to an atrial paced event. Every

20/104
cycle, the V-A delay is calculated as the 9.6.1 Ventricular Asynchronous (V00)
difference between the set beat-to-beat Pacing
interval and the A-V interval. Ventricular asynchronous (V00) pacing
Atrial triggering means that a detected is the simplest of all pacing modes
P-wave causes an atrial stimulus. The because there is no sensing and no
programmable options for atrial mode of response. The ventricular
triggering is OFF or ON. pacing stimuli occur at the programmed
rate, regardless of any intrinsic cardiac
The blanking period is defined as the event.
time during and after a sensed or paced
event when the sensing channel and/or 9.6.2 Atrial Asynchronous (A00)
the opposite channel of an Pacing
atrioventricular pacemaker is insensitive.
The purpose is to avoid sensing of late Atrial asynchronous (A00) pacing
potentials and sensing the event of one behaves exactly like V00, but the pacing
channel in the opposite channel stimuli occur in the atrium. The atrial
(crosstalk). Thus, during the blanking pacing stimuli occur at the programmed
period, no events are recognized. rate, regardless of any intrinsic cardiac
event.
Refractory period is a set time in the
pacemaker, in which a signal in the 9.6.3 A-V Sequential Asynchronous
respective channel will be recognized (D00) Pacing
but not tracked or does not cause
inhibition. Dual-chamber, or A-V sequential
asynchronous (D00), pacing provides
Sensing phase is the period of time in atrial and subsequent ventricular stimuli
which a signal that occurred in the with a fixed, programmed delay. The
respective channel will be recognized, pacing stimuli occur at the programmed
interpreted as intrinsic, and tracked. rate, regardless of any intrinsic atrial or
Thus, this is the period outside blanking ventricular events.
or refractory periods.
The post-ventricular atrial refractory
period, or PVARP (a pacemaker 9.6.4 Ventricular Inhibited (VVI)
parameter), is the period after a sensed Pacing
or paced ventricular event during which
the atrial sensing circuit is refractory. Ventricular inhibited (VVI) pacing
Thus, any atrial event occurring during incorporates sensing on the ventricular
PVARP will not be sensed by the atrial channel, and pacemaker output is
sensing circuit. In biventricular inhibited by a sensed ventricular event.
stimulation the PVARP starts after the In VVI mode, the PACE 203H is
second ventricular stimulus. refractory for a period after a paced or
sensed ventricular event, the ventricular
refractory period. Any ventricular event
occurring within the ventricular refractory
period is not sensed. Any ventricular
event occurring outside the ventricular

21/104
refractory period is sensed and resets
the timing of the pacemaker to the
beginning of a new beat-to-beat interval. 9.6.7 Dual-Chamber Pacing and
Sensing with Inhibition and
Tracking DDD
9.6.5 Atrial Inhibited (AAI) Pacing This pacing mode provides dual-
chamber pacing and sensing with
In the absence of intrinsic activity,
inhibition and atrial tracking.
ventricular or atrial stimulation impulses
will be given with the set basic rate. The If intrinsic atrial and ventricular activity
recognition of a P-wave in AAI mode in occur before the beat-to-beat interval
the pacemaker’s sensing phase inhibits terminates, both channels are inhibited
the delivery of the next stimulation and no pacing occurs.
impulse, resets the timing of the
pacemaker to the beginning of the beat- If a P wave is sensed before the V-A
to-beat interval and starts the refractory interval is completed, output from the
period. atrial channel is inhibited. The A-V delay
is initiated. If no ventricular activity is
sensed before the A-V delay terminates,
a ventricular pacing stimulus is
9.6.6 Atrial Synchronous (P- delivered, that is, P-synchronous
Tracking) (VDD) Pacing ventricular pacing.
In atrial synchronous (P-tracking) (VDD) If no atrial activity is sensed before the
pacing mode, the PACE 203 H paces V-A delay is completed, an atrial pacing
only in the ventricle, senses in both stimulus is delivered, which initiates the
atrium and ventricle, and responds both A-V delay. If intrinsic ventricular activity
by inhibition of ventricular output by occurs before the termination of the A-V
intrinsic ventricular activity and by delay, the ventricular output from the
ventricular tracking of P waves. PACE 203 H is inhibited, that is, atrial
pacing. If no intrinsic ventricular activity
A sensed P wave initiates the P-V delay. occurs before the termination of the A-V
During the P-V delay, the atrial channel delay, a ventricular pacing stimulus is
is refractory. At the end of the P-V delay, delivered, that is, A-V sequential dual-
a ventricular stimulus is delivered if no chamber pacing.
intrinsic ventricular activity has been
sensed, that is, P wave tracking.
Ventricular activity, paced or sensed,
initiates the PVARP and the V-A delay. 9.6.8 The A-V Sequential, Atrial
If no P wave activity occurs during the V- Inhibited (DAI) Pacing
A delay, the A-V delay is initiated. A-V sequential, atrial inhibited (DAI)
During the A-V delay, the atrial channel pacing provides pacing in both the
is refractory. If no R-wave activity occurs atrium and the ventricle but sensing only
during the A-V delay, the PACE 203 H in the atrium. This mode is implicitly
delivers a ventricular pacing stimulus selected by setting first the DDD mode
upon termination of the A-V delay. and, second, the ventricular sensitivity to

22/104
infinite ("–.–"). Thus, sensing in the atrium only. If intrinsic atrial activity
ventricular channel is disabled. occurs, a ventricular stimulus is
delivered after the P-V delay.
The DAI mode may be useful in
situations where no ventricular sensing This mode is implicitly selected by first
is possible (for instance, because of setting the VDD mode and, second, the
interference), or when A-V synchrony ventricular sensitivity to infinite ("–.–").
must be maintained upon a complete A- Thus, sensing in the ventricular channel
V block. is disabled.
The VAT mode may be useful in
9.6.9 The A-V Sequential, Ventricular situations where no ventricular sensing
Inhibited (DVI) Pacing is possible (for instance, because of
A-V sequential, ventricular inhibited interference), or when A-V synchrony
(DVI) pacing provides pacing in both the must be maintained upon a complete A-
atrium and the ventricle but sensing only V block.
in the ventricle. This mode is implicitly
selected by setting first the DDD mode 9.6.11 Non-Pacing Mode 0D0 (Pause)
and, second, the atrial sensitivity to
infinite ("–.–"). Thus, sensing in the atrial In this mode, sensing is enabled in all
channel is disabled. channels but not pacing. The PACE
203H™ is ‘listening’ to intrinsic atrial and
Note: With the DVI mode (implicitly) set, ventricular events only.
the following exceptions are taking
The 0D0 mode is utilized to check the
place:
patient’s intrinsic activity (siehe chapter
1) The PACE 203 H - specific 9.14) or to get initial values for the Auto
algorithm for determining Sense feature (see chapter 9.9.4). It
extrasystoles is disabled. In DVI must be used with great caution on
mode, the PACE 203 H cannot patients depending on a pacemaker.
differentiate between high intrinsic
rates and occurrences of 9.6.12 Atrial Trigger Modes DDD+AT
extrasystoles, because the atrial and DAT
sensing is disabled.
If Atrial Trigger option (see chapter
2) If another R-wave is detected during 9.12.5.3) is activated, an atrial stimulus
the V-A delay, the V-A delay will be is delivered any time an atrial activity is
initiated with a value equal to the sensed, and with the atrial channel not
difference between the beat-to-beat being refractory.
interval and the A-V delay.
The Atrial Trigger is enabled when either
AAI (→ AAT) or DDD (→ DDD+AT) or
9.6.10 Atrial Sensing Atrial
DAI (→ DAT) modes are set.
Synchronous (P-Tracking) (VAT)
Pacing
9.7 Changing the Mode of Operation
In atrial synchronous (P-tracking) (VDD)
pacing mode, the PACE 203 H paces If the lower display is switched on, it
only in the ventricle but senses in the always displays the Mode Menu first.

23/104
Each of the primary modes (DDD, VVI, Mode Menu Key No.
AAI, VDD) can now be selected by the
touch of the corresponding softkey. DDD
D00 DVI DAI  1
(Table 8).
If the lower display is off, it can be VVI
V00  2
turned on pressing the key
Unlock/Lock.In the event that another AAI
menu is displayed, pressing the ↑_ A00 AAT  3
softkey will switch to the mode menu
immediately. VDD
V00 VVI VAT  4

Main Menu 5

Table 8: Mode Menu


The modes D00, V00, A00, DVI, DAI
and VAT are set by selecting the
"primary" mode (DDD, VVI, AAI, VDD)
and setting the sensitivity in the
respective channel(s) to infinite
(displayed as "–.–").

Primary Suppl.
Sensitivity
Mode Mode
DDD A-Sensitivity → "–.–" DVI
DDD V- Sensitivity → "–.–" DAI
DDD A- and V- Sensitivity → "–.–" D00
VVI V- Sensitivity → "–.–" V00
AAI A- Sensitivity → "–.–" A00
VDD A- Sensitivity → "–.–" VVI
VDD V- Sensitivity → "–.–" VAT
VDD A- and V- Sensitivity → "–.–" V00
Table 9: Supplementary Mode Setting

24/104
The selected mode is indicated with its solve the problem (see chapter 9.21.7
NBG code1 in the upper display. and 9.9)

In the new mode, the PACE 203H


continues to work with the parameters
set before the mode change. 9.8 Adjustment of Rate, Stimulation
Amplitude, Sensitivity and A-V
Note: In case of a change from a single
Delay Using the Dials
to a dual chamber mode or from one
chamber mode to another chamber One dial is assigned to each of the
mode (for instance VVI→AAI), the last parameters. From top to bottom:
used parameters are adopted. If the
Turn-on program (see chapter 9.11.5),  Rate (RATE)
has been re-defined to a single chamber • Atrial sensitivity (A-SENSE)
mode, in case of a first change of such  Atrial stimulation ampl. (A-STIM)
kind, the parameter set will be • A-V Delay (A-V DLY)
complemented from the standard  Ventricular stim. ampl. (V-STIM)
program of the new mode. • Ventricular sensitivity (V-SENSE).

The enabling of the atrial trigger function The actual values of the parameters,
in the AAI and DDD modes is described and the corresponding bar diagrams
in chapter 9.12.5.3. indicating signal intensity or magnitudes,
are located in the upper display, which is
A non-pacing mode (0D0) is accessible always turned on. These bar diagrams
by pressing the key labeled Pause (see indicate the set value within its range.
chapter 9.14).

General Notes:
Note: A standard program can be stored
for each primary mode. These standard In order to change a parameter, or
programs may be defined by the user parameters, the dials must be unlocked
(see chapter 9.11). by pressing the key Unlock/Lock (if the
PACE 203H™ is not yet unlocked).
Rotating the rate, amplitude, or A-V
Note: If conflicts (because of abnormal delay dials clockwise increases the
settings) occur as a result of a mode parameter value. In contrast,
change, the PACE 203H™ will make counterclockwise rotation decreases the
use of its Automatic setting feature to parameter value
Rotating a sensitivity dial clockwise
increases the sensitivity value, i.e.
1 Bernstein AD, Camm AJ, Fletcher RD, et decrease the sensitivity. In contrast,
al. The NASPE/BPEG generic pacemaker counterclockwise rotation decreases the
code for antibradyarrhythmia and adaptive- sensitivity value, i.e. increases the
rate pacing and antitachyarrhythmia sensitivity
devices. Pacing Clin Electrosphysiol 1987;
10: 794-799

25/104
If a maximum or minimum value is
reached, further rotation of the dial is
disregarded. Confirmation to increase Key No.
the Rate higher than 150
Each parameter change is visualized on ppm
the upper display, both as a number and
as a bar diagram
1
Any change of the parameters will
become effective after the next sensing Rate
> 150 ppm ?
or stimulation event. 2
Several parameters are allowed to
change simultaneously.
If a parameter has no pacing function in 3
YES →
the mode selected, for example, A-
Sense in the VVI mode, the 4
corresponding section in the upper NO →
display is blanked. If a dial, belonging to
parameter without pacing function in the 5
mode selected, is set, the respective,
previously blanked section will be
switched on. A numeric value will be Table 11: Message related to increase
briefly displayed, whereupon it will the Rate from 150 ppm
disappear in approximately 2 seconds.

9.8.1 Adjustment of Basic Rate


Each time the stimulation rate is
Rotating the RATE clockwise will increased (by rotating the dial RATE
increase the basic rate in the range clockwise) higher than 150 ppm, the
shown in Table 10. The actual rate value message “Rate > 150 ppm?” appears
and the intensity bar diagram can be and must be confirmed with YES in
seen in the upper display. The intensity order to increase the rate from 150 ppm.
bar diagram enlarges from left to right NO abandons the intention to increase
with rate increase. the stimulation rate, and the message
disappears. Once the threshold of 150
ppm is passed, increases of the
Parameter Adjustable Values Unit stimulation rate in the range beyond 150
Rate 30 (2) 220 ppm
ppm and less or equal than 180 ppm
require no confirmation.
Table 10: Adjustable values for rate
Note: The safety feature "protection
against abnormal settings" (see chapter
In order to avoid unintentional settings of 9.21.7) prevents increasing the basic
high stimulation rates, the desire to rate to a value larger then the MTR in
increase the stimulation rate from 150 DDD, VDD, DAI, VAT and DAT mode.
ppm (Table 11) and then from 180 ppm Limiting the basic rate value is
must be confirmed.

26/104
accompanied by a transient warning sensitivity value (mV value) decreases.
message in the lower display. This indicates that the sensitivity value
increases. The parameter intensity bar
9.8.2 Adjustment of Stimulation diagram expands from the right to left
Amplitude side.
By rotating the A-STIM dial or the V- The sensitivity is adjustable in the range
STIM dial clockwise the stimulation and with the steps shown in Table 13.
amplitude will be increased in the range
shown in Table 12.The actual
stimulation amplitude value and the Parameter Adjustable Values Unit
intensity bar diagram can be seen in the
upper display. The intensity bar diagram 0,2 (0,1) 2,0
enlarges from left to right with amplitude 2,0 (0,2) 5,0
increase. Atrial Sensitivity 5,0 (0,5) 10,0 mV
10,0 (1,0) 20,0
Parameter Adjustable Values Unit –.– (= no sensing)
0,1 (0,1) 2,0
Stimulation- 2,0 (0,2) 6,0
V 1,0 (0,1) 2,0
Amplitude 6,0 (0,5) 12,0
12,0 (1,0) 18,0 2,0 (0,2) 5,0
Ventricular
5,0 (0,5) 10,0 mV
Table 12: Adjustable values for Sensitivity
10,0 (1,0) 20,0
stimulation amplitude –.– (= no sensing)

Table 13: Adjustable values for


9.8.3 Adjustment of Sensitivity sensitivity

By rotating the A-SENSE dial or the V-


SENSE dial clockwise, the sensitivity Warning: Increasing the mV values
value (mV value) increases. This indicates decreasing the sensitivity. An
indicates that the sensitivity will atrial value of 0.2 mV and a ventricular
decrease. The actual sensitivity value value of 1.0 mV represent the highest
and the parameter intensity bar diagram sensitivity in the respective channels.
can be seen in the upper display. The The value "infinite" results in an
parameter intensity bar diagram asynchronous mode for the
diminishes from left to right, when the corresponding channel. An inhibition of
sensitivity value is increased (sensitivity the stimulation pulse in this channel is
is decreased). If the minimum sensitivity not possible.
is reached, the value "infinite" is
indicated by "–.–" in the display, and a
9.8.4 Adjustment of A-V Delay
mode change occurs. This is indicated
by the pacing mode code, visible in the By rotating the A-V DLY dial clockwise,
upper display (see chapter 9.7). the A-V delay will become extended
within the range shown in Table 14. The
By rotating the A-SENSE dial or the V- actual A-V delay and the parameter
SENSE dial counterclockwise, the intensity bar diagram can be observed in

27/104
the upper display. The parameter V-A interval in the D00 mode,
intensity bar diagram corresponds to the respectively. The limitation of the A-V
A-V delay. The leftmost bar represents delay value will be indicated by a
the timely occurrence of the atrial event transient warning message in the lower
and is always displayed. A second bar display.
represents the timely occurrence of the
ventricular event and is "shifted" right or
left, related to the increase or decrease
the A-V delay, respectively 9.9 Using the Automatic Features
The PACE 203H™ offers the capability
for automatic setting of A-V delay,
Parameter Adjustable Values Einheit PVARP, and maximum tracking rate
(MTR), with the setting of rate as well as
A-V Delay 5, 10 (10) 400 ms
for automatic adjustment of sensitivity.
Table 14: Adjustable values for A-V To enable these features, the Auto
delay Menu must first be entered by pressing
The P-V delay (PVD) is coupled with the the softkey labeled Main Menu and then
A-V delay (AVD). Adjusting the AVD the softkey labeled Auto.
also sets the PVD. As default the If the lower display is off, then pressing
difference AVD-PVD is equal to 30 ms, the key Unlock/Lock turns it on. The
that is, PVD is 30 ms shorter than AVD. Auto Menu appears as shown in Table
Note: When atrial Auto Sense is 15.
enabled, the A-V delay and P-V delay Then, with the corresponding keys, one
are limited to a minimum of 30 ms. This or more of the automatic features can be
is because Auto Sense requires a enabled or disabled individually.
specified time to measure the intrinsic
atrial activity. In case the atrial Auto
Sense is enabled, and the A-V delay
Auto Menu Key No.
was set to a value smaller than 30 ms,
then the A-V delay will be automatically Auto
AVD  1
extended to 30 ms. This action is
indicated by a transient informative Auto
PVARP  2
message in the lower display (see also
chapter 9.9.4). Auto
MTR  3
Auto
Sense 4
Note: Because of its safety feature
"protection against abnormal settings" ↑_ 5
(see chapter 9.21.7), the PACE 203H™
prevents increasing the A-V delay to a Table 15: Auto Menu
value, which would not guarantee a
minimum atrial sensing phase in dual
chamber demand modes, or a minimum

28/104
9.9.1 Automatic Setting of A-V Delay AUTO disappears from the upper
with the Setting of Rate display.
If this feature is enabled, the PACE To deactivate Auto A-V Delay in the
203H automatically adjusts the A-V Auto menu, the softkey labeled
delay with the setting of the basic rate. Auto AVD must be pressed again. This
To activate Auto A-V Delay, the softkey causes the check mark to disappear
labeled Auto AVD in the Auto Menu has from the corresponding check box and
to be pressed. The corresponding check AUTO from the upper display.
box then gets checked.
9.9.2 Automatic Setting of PVARP
The PACE 203H then adjusts the A-V with the Setting of Rate
delay depending on the basic rate
according to the formula in Table 16. If this feature is enabled, the PACE
and rounded to the nearest manually 203H™ automatically adjusts the post-
adjustable value (see chapter 0). In the ventricular atrial refractory period
corresponding A-V DLY segment in the (PVARP) with the setting of the basic
upper display the label AUTO indicates rate.
that the Auto A-V delay feature is To activate Auto PVARP, the softkey
activated. The calculated value for the labeled Auto PVARP in the Auto Menu
A-V delay is also indicated numerically. must be pressed. The corresponding
check box then gets checked.

Parameter Automatic A-V Range Unit


The PACE 203H™ now adjusts the
Delay Setting PVARP depending on the basic rate
Formula according to the formula in Table 17 and
rounded to the nearest manually
290 – (1,5 × adjustable value (see chapter 9.12.1). In
A-V Delay 50.. 250 ms
Rate/ppm) the upper display the label AUTO
Table 16: Automatic setting formula PVARP indicates that the Auto PVARP
for A-V delay feature is activated.

Auto A-V Delay can be deactivated Parameter Automatic Range Unit


either via the Auto menu or simply by PVARP Setting
rotating the (unlocked) A-V DLY dial a Formula
few clicks (approximately a quarter of
330 – (0.75 ×
one revolution). PVARP
Rate/ppm)
100.. 500 ms

If the A-V DLY dial has been rotated to Table 17: Automatic setting formula
deactivate Auto A-V delay, the transient
for PVARP
message
Auto AVD
turned off To deactivate Auto PVARP, the softkey
labeled Auto PVARP in the Auto menu
appears in the lower display
must be pressed again. This causes the
accompanied by a short beep, and

29/104
check mark to disappear from the Parameter Automatic Range Unit
corresponding check box and AUTO MTR Setting
PVARP from the upper display. The Formula
PVARP & MTR setting menu appears
showing the actual value which provides MTR Rate + 34 100.. 230 ppm
the opportunity to adjust a new value
manually (see chapter 9.12.1). Table 18: Automatic setting formula
for MTR

Note: The PVARP of the PACE 203H™


is divided in two parts: an absolute part To deactivate Auto MTR, the softkey
and a relative part. In the absolute part labeled Auto MTR in the Auto Menu
(which is 100 ms long), no atrial events must be pressed again. This causes the
are recognized. In the relative part (the check mark to disappear from the
remaining) atrial events are registered, corresponding check box and the label
but not tracked. This algorithm prevents AUTO MTR in the upper display. The
the misinterpretation of ventricular MTR setting menu appears showing the
events after premature atrial events as actual value which provides the
extrasystoles (PVCs), as described in opportunity to input a new value
chapter 9.21.5. manually (see chapter 9.12.2).

9.9.4 Automatic Adjustment of


Sensitivity (Auto Sense)
9.9.3 Automatic Setting of MTR with
the Setting of Rate
9.9.4.1 How the Auto Sense Function
If this feature is enabled, the PACE Works
203H™ automatically adjusts the
The Auto Sense function automatically
Maximum Tracking Rate (MTR) with the
adjusts the PACE 203H™ sensitivity to
setting of the basic rate.
maintain an approximate 3:1 sensing
To activate Auto MTR, the softkey safety margin, potentially reducing the
labeled Auto MTR in the Auto Menu number of under/over-sensing episodes.
must be pressed. The corresponding This feature may also reduce the need
check box then gets checked. to manually determine the sensitivity
threshold (see chapter 9.16).
The PACE 203H™ then adjusts the
MTR to a value 34 ppm higher than the The Auto Sense function is different
basic rate, but with a minimum of from the other automatic functions in
100 ppm and a maximum of 230 ppm, that the automatic sensing value is not
according to the formula and the range calculated from another parameter. The
in Table 18. In the upper display the PACE 203H™ adapts the sensing value
label AUTO MTR indicates that the Auto continuously to the patient’s intrinsic
MTR feature is activated. heart activity. For this intrinsic activity,
which is detected in the sensing phase
of the respective channel, the amplitude
is measured and the sensitivity value of

30/104
the pacemaker is adapted to maintain Important Notes for the use of the
the 3:1 sensing safety margin. Auto Sense function:
14 Because the intrinsic activity of the
12
10
safety
margin patient’s heart determines the sensitivity
8
6 value, the Auto Sense feature can, of
4
2
sensitivity
course, only be used in a patient who
0 has such intrinsic activity.
mV

-2
-4
-6
Auto Sense cannot be used together
-8 with Atrial Trigger, because the triggered
-10
-12 stimulation prevents proper
-14
Increase Sensitivity until lower margin is exceeded
measurement of the intrinsic activation
(see also 9.12.5.3)
Figure 5: The Auto Sense function Auto Sense requires a specific time to
increases sensitivity measure the intrinsic atrial activity. P-V
delay is therefore limited to a minimum
Figure 5 shows an example in which the of 30 ms when atrial Auto Sense is
sensitivity is increased (i.e. sensitivity activated. Because A-V delay cannot be
value is decreased) to maintain an shorter than P-V delay it is also limited
appropriate safety margin when the to 30 ms. In case atrial Auto Sense is
EGM amplitude drops. enabled, and the A-V delay was set to a
value smaller than 30 ms, then the A-V
14
12
safety
margin
delay will be automatically extended to
10
8 30 ms. This action will be indicated by a
6
4
sensitivity transient informative message in the
2
0
lower display (siehe auch 0).
mV

-2
-4
-6
-8
-10
-12 9.9.4.2 Activation of Auto Sense
-14

Decrease Sensitivity until upper margin is no more exceeded Because the Auto Sense algorithm
requires sensed intrinsic atrial or
Figure 6: The Auto Sense function ventricular events to become
decreases sensitivity operational, an initial appropriate
sensing value is required to enable the
Auto Sense feature. The following
Figure 6 shows an example in which the procedure has been established to
sensitivity is decreased (i.e. sensitivity obtain this initial value:
value is increased) when the EGM To enter the Auto Sense Menu the
amplitude goes up. An appropriate softkey labeled Auto Sense in the Auto
safety margin is kept without working Menu must be pressed (see Table 19).
with unnecessarily high sensitivity, which
may promote noise detection.

31/104
Auto Sense Menu Key No. selected, initialization request)

1
Atrial When the softkey labeled Initiate is
Auto Sense  2 pressed, the PACE 203H™ stops pacing
Ventricular as long as this key is held down.
Auto Sense  3 However, the PACE 203H™ continues
to sense and attempts to detect a
4 regular intrinsic activation of the
patient’s heart. While doing this, the
↑_ 5 lower display appears as shown in Table
21. A flashing heart symbol indicates
Table 19: Auto Sense Menu each detected event. Furthermore, the
acoustic indicator will automatically be
enabled temporarily, if it is not already
The softkey labeled Atrial Auto Sense switched on. Because the PACE 203H™
and/or Ventricular Auto Sense must be goes through an initialization algorithm,
in order to enable or disable the Auto individual intrinsic events may or may
Sense feature for the atrial and/or not be detected.
ventricular channel (Table 19).
If Auto Sense is be enabled in a channel
Auto Sense Menu Key No.
where it was not activated before, a
check mark appears in the respective
1
checkbox. The checkbox then starts
blinking, and the request to press and Searching

hold the softkey labeled Initiate is A-Sense 2
indicated in the lower display. Searching

V-Sense 3
Table 20 illustrates this scenario, if Auto
Sense is enabled in both channels. hold... 4

5
Auto Sense Menu Key No.
Table 21: Auto Sense Menu
1 (Searching)
Atrial
Auto Sense  2
Ventricular If the PACE 203H™ has found an
Auto Sense
 3 intrinsic heart rhythm, it will display the
sensed EGM amplitude as shown in
Initiate
(Press & Hold) 4 Table 22.

↑_ 5
Table 20: Auto Sense Menu (channels

32/104
Notes for the initialization of the Auto
Auto Sense Menu Key No. Sense function:

Sensed: 1 If the user continues to press the softkey


labeled Initiate, the PACE 203H™ will
A = 2.3 mV 2
continue to search for intrinsic activity
and find the optimal value.
V = 8.0 mV 3 If the PACE 203H™ detects frequencies
higher than the noise frequency
Finished (273 bpm) during the Auto Sense
(release key) 4
initialization, the corresponding
interference symbol ( ) is shown in the
5
upper display. Any P or R wave value
Table 22: Auto Sense Menu determined by the PACE 203H™ during
(intrinsic rhythm found) interference is cleared off the display. 2
seconds after the PACE 203H™
determines interference with the Auto
The softkey labeled Initiate can now be Sense function enabled, the sensitivity
released. An appropriate sensitivity setting will be doubled in order to find a
value will be automatically selected and sensitivity level where the heart activity
set which will be visible in the upper is sensed without interference.
display. Furthermore the sign AUTO will If no activity is sensed in a channel for
be shown in the respective section(s) of 3 seconds, a previously determined P or
the upper display. The lower display will R wave will be cleared off the display.
return to the Auto Sense Menu with the If no activity is sensed in a channel for
boxes checked (blinking ceases) as 3 seconds, a previously determined P or
shown in Table 23. R wave will be cleared off the display.
Table 24 shows an example were an
atrial rhythm could be found, but no
Auto Sense Menu Key No. ventricular rhythm.

Atrial
Auto Sense  2

Ventricular
Auto Sense  3

↑_ 5

Table 23: Auto Sense Menu (finished)

33/104
Auto Sense Menu Key No. The user can decide if another attempt
shall be made to find an intrinsic rhythm,
Sensed: 1 or if the procedure shall be terminated
using the ↑_ key. In the previous
example, the Auto Sense feature will
A = 2.3 mV 2 then be enabled only in the atrial
channel.
No V-Sense 3
9.9.4.3 Re-Initialization of Auto Sense
Finished
(release key) 4 Once activated, the Auto Sense feature
can be easily re-initialized by just
5 pressing and holding the key PAUSE
(see 9.14). However, this is only
Table 24: Auto Sense Menu necessary if the amplitude of the
intrinsic activity has decreased so
(no R-wave found)
dramatically and rapidly such that the
Auto Sense algorithm could not follow
this change.
If no intrinsic rhythm is detected in one
or both channels, the respective check
mark continues blinking after returning
Warning: If frequencies higher than the
to the previous menu. The initialization
request persists as shown in Table 25. noise frequency (273 bpm) are detected
during the re-initialization, the
corresponding interference symbol (…)
will be displayed in the upper display.
Auto Sense Menu Key No. The sensitivity setting will remain
unchanged and the measured frequency
1 will be displayed, because it may be a
high-frequency tachycardia. This
Atrial behavior is different from Auto Sense
Auto Sense  2 initialization (see above), where the
sensitivity value will be increased step
Ventricular
Auto Sense  by step in case of noise.
3
9.9.4.4 Deactivation of Auto Sense
Initiate.
(press & hold) 4 The Auto Sense function can be
disabled either via the Auto Menu or
simply by rotating the respective
↑_ 5 (unlocked) A-SENSE or V-SENSE dial a
Table 25: Auto Sense Menu (one few clicks (approx. a quarter of one
revolution).
channel initialization request)

34/104
If a SENSE dial has been turned to main menu via the softkey labeled Main
deactivate Auto Sense, the transient Menu, and, then, by selecting the high
message rate stand-by menu by pressing the
softkey labeled High-Rate. (Remember
to press the key Unlock/Lock in case
Atrial Ventricular the PACE 203H™ keypad is locked. In
case the lower display was turned off, it
Auto Sense or Auto Sense
is then turned on.) The High-Rate
turned off turned of Standby menu appears as shown in
Table 26 and the overdrive-stimulation
standby mode is entered. The acoustic
appears in the lower display, indicator will automatically be switched
accompanied by a short beep. The label on temporarily, if it is not already on.
AUTO disappears from the upper
display.
Note: Within the High-Rate Stand-By
Using the Auto menu, the Auto Sense menu, the pacemaker continues to
function is disabled by pressing the operate in the set mode.
corresponding softkey. As a
consequence, the corresponding check
box is unchecked. The label AUTO
Any previously selected overdrive
disappears from the upper display.
stimulation rate and the corresponding
beat-to-beat interval are displayed.
Note: Upon a mode change (for
instance VVI → AAI or DDD → VVI) that Note: After turning on, the PACE
disables a channel, the Auto Sense 203H™ overdrive-stimulation-rate is set
function, enabled in this particular to 240 ppm if the device has been
channel, is disabled. The Auto Sense powered off using the softkey labeled
function is not reactivated upon enabling Off (no storage). If the device has been
the particular channel, because the powered off with the softkey labeled
EGM amplitudes could have changed Stand-by (store data) then it will store
significantly during the time the Auto
the high rate frequency set before
Sense function was disabled.
powering off.

9.10 The High Rate Stimulation


(Atrial Overdrive)
The PACE 203H™ provides a function
for high rate stimulation, also known as
atrial overdrive stimulation, or rapid atrial
pacing. The atrial high rate stimulation
function is activated by entering the

35/104
High Rate Standby Key No. event, the pacemaker switches to the
Menu A00 mode and stimulates with the set
overdrive-stimulation rate, showing the
START 1 High-Rate Running menu (Table 28).
240 ppm During overdrive stimulation, the PACE
(250 ms) 2 203H™ switches temporarily (as long as
the START key is pressed) to A00
3 mode. To prevent confusion, the upper
Dial RATE display is not completely changed
to change accordingly. The display of atrial
High-Rate! 4 stimulation section, however, is enabled
on (if not already enabled) to allow a
↑_ 5 convenient pre-setting of the atrial
stimulation amplitude.
Table 26: High Rate Standby Menu
Warning: There is a risk of causing
ventricular tachycardia during atrial
The overdrive stimulation rate can be overdrive-stimulation. Continuous ECG-
adjusted by using the RATE dial. This monitoring of the patient is therefore
means that the RATE dial defaults into a mandatory. A defibrillator should always
HIGH-RATE dial for adjustment of the be available and ready to use.
overdrive rate as long as the High-Rate High Rate Stimulation with Ramp
Menu is shown. Warning is directed Function
here by a blinking symbol HI- in front of
the label RATE in the upper display. The PACE 203H™ provides a function
allowing the application of overdrive-
The overdrive stimulation rate can be stimulation with a ramp function. While
changed in 10 ppm increments between in the High Rate Running Menu, turning
70 ppm and 1000 ppm, as shown in the HIGH-RATE dial changes the
Table 27. overdrive stimulation rate. This feature
allows the user to manually set a ramp
function for atrial overdrive stimulation
Parameter Adjustable Values Unit Note: The PACE 203H™ ignores any
rotation of the HIGH-RATE dial beyond the
Atriale High Rate 70 (10) 1000 ppm defined minimum or maximum settings.

Table 27: Adjustable values for atrial


high rate

The PACE 203H™ delivers overdrive


stimulation impulses upon pressing and
holding the softkey labelled START After
the next sensed atrial or ventricular

36/104
High Rate Running- Key No. the atrial channel (see chapter
Menu 9.21.2) is enabled. If no atrial lead is
connected, the error message Atrial
running... 1 lead disconnected will consequently
appear. Nevertheless, the impulse
240 ppm
2 emission with overdrive-rate will be
(250 ms)
carried out. After releasing the
softkey labeled START, the
3 previous state of the lead
Dial RATE
to change surveillance will be restored.
High-Rate! 4 • The Pause function (see chapter
9.14) can be used to measure the
frequency of atrial tachycardia up to
↑_ 5
rates of about 700 bpm). Thus, the
Pause function assists in
Table 28: High Rate Running Menu determining the optimal overdrive
stimulation rate. (Please note that
the atrial frequency measurement
After releasing the softkey labeled can be affected by crossblanking
START, now labeled running…, the after ventricular senses.
PACE 203H™ abandons the delivery of • As long as the softkey labeled
overdrive stimulation impulses, and START is pressed, all other keys
returns to High-Rate standby. If the and dials are disabled.
overdrive rate has been changed during
High-Rate running, the basic rate prior
the start of the overdrive-stimulation is
9.11 Standard Programs
restored. Thus, a rate change during
running overdrive stimulation is only The PACE 203H™ provides the
temporary. Thus, a ramp can always be capability of storing and recalling a user
started from the same rate. defined standard set of parameters (a so
called "standard program") for each
The overdrive mode is abandoned by
primary mode of operation (DDD, VVI,
pressing the softkey ↑_, which forces the
AAI, VDD). That is, four different
PACE 203H™ to return to the Mode
standard programs can be stored.
Menu
Alternatively, the user can reset each
Note: primary mode to the manufacturer’s
default settings. In addition, the initial
• Within the High-Rate Stand-by mode and parameter settings can be
menu, the pacemaker does not lock configured in the Turn-on program.
automatically.
To perform one of these functions, the
• To return to the mode menu, please Standard Menu must first be entered via
press the key ↑_. the softkey labeled Main Menu, and then
via the softkey labeled Standard. If the
• After the start of an overdrive-
lower display is off, then pressing the
stimulation, the lead surveillance in

37/104
key Unlock/Lock turns the display on. 9.11.1 Parameters Stored in a
The Standard Menu appears as shown Standard Program
in Table 29. The parameters stored in each of the
four sets are marked with "+" in Table
30. The parameters marked with "–" are
Standard Menu Key No. not stored in the set and will remain
unchanged if a stored program is
Save as
1 recalled.
Turn-on

Save as XXX
standard 2

Recall XXX
standard 3

Manufacturer’s-
default 4

↑_ 5

Table 29: Standard Menu

Parameter DDD VVI AAI VDD Unit


Basic Rate + + + + ppm
Overdrive Stimulation Start Rate + – + – ppm
Atrial Amplitude + – + – V
Atriale Sensitivity (incl. Automatic state) + – + + mV
Atrial Pulse Duration + – + – ms
Ventricular Stimulation + + – + –
Ventricular Amplitude + + – + V
Ventricular Pulse Duration + + – + ms
Ventricular Sensing + + – +
Ventricular Sensitivity + + – + mV
(incl. Automatic state)
A-V Delay + – – + ms
(incl. Automatic state)
PVARP + – – + ms
(incl. Automatic state)
MTR + – – + ppm
(incl. Automatic state)
Atrial Trigger Option + – + – –
Table 30: Standard Programs Set

38/104
Note: Because the sensitivity and the Standard Saving Key No.
options are also stored, the pacing mode Confirmation
stored may also be one of the
supplementary modes (D00, DVI, VAT, 1
V00, A00, AAT and DDD+AT). Overwrite
standard XXX program? 2

YES 3
Warning: Because the automatic
condition of the sensitivity adjustment is
also stored, recalling a standard setting NO 4
with Auto Sense = ON is followed by the
request to search for initial sensitivity ↑_ 5
values as described in chapter 9.9.4.2.
Table 31: Standard Saving
Therefore, storing a standard program
Confirmation
with Auto Sense activated, is only
recommended for experienced users.
Note: Upon the manufacturer’s shipping After storing the user defined Standard
of the PACE 203H™, the user standard Program, the PACE 203H™ abandons
settings are identical with the the Standard Menu and returns to the
manufacturer’s defaults (see chapter Mode Menu.
9.11.4).
If the question Overwrite standard XXX
program? Is answered with “NO”, the
PACE 203H™ returns to the (previous)
9.11.2 Store Standard Programs Standard Menu.
By pressing the softkey labeled Save as Pressing the softkey ↑_ forces the PACE
XXX standard in the Standard Menu 203H to revert to the Mode Menu
(Table 29) the actual setting will become
the standard for the current primary 9.11.3 Recall Standard Programs
mode (XXX stands for one of the primary
modes DDD, VVI, AAI or VDD) upon By pressing the softkey labeled Recall
answering the question Overwrite XXX standard in the Standard Menu
standard XXX program? With “YES” (Table 29), the stored standard program
(Table 31). for the current primary mode (XXX
stands for one of the primary modes
DDD, VVI, AAI or VDD) becomes the
actual setting.
After pressing the softkey labeled Recall
XXX standard , the PACE 203H™
abandons the Standard Menu and
returns to the Mode Menu.

39/104
9.11.4 Recall Manufacturer’s Defaults After pressing the softkey
By pressing the softkey Manufacturer’s Manufacturer’s default, the PACE
default in the Standard Menu (Table 29), 203H™ abandons the Standard Menu
the manufacturer’s default for the current and returns to the Mode Menu.
primary mode (as shown in Table 32)
becomes the actual setting.

Parameter DDD VVI AAI VDD Unit


Basic Rate 72 72 72 72 ppm
Overdrive Stimulation 240 – 240 – ppm
Start Rate
Atrial Amplitude 5 – 5 – V

Atrial Sensitivity 1 – 1 0.5 mV

Atrial Pulse Duration 1 – 1 – ms


Ventricular Amplitude 8 8 – 8 V

Ventricular Pulse 0.75 0.75 – 0.75 ms


Duration
Ventricular Sensitivities 2 2 – 2 mV
A-V Delay AUTO – – AUTO ms
(i.e. 180) (i.e. 180)

PVARP AUTO – – AUTO ms


(i.e. 280) (i.e. 280)

MTR AUTO – – AUTO ms


(i.e. 106) (i.e. 106)

Automatic Sensing OFF OFF OFF OFF –


Atrial Trigger Option OFF – OFF – –
Table 32: Manufacturer’s Default Programs

9.11.5 Change the Turn-On Program


By pressing the softkey Save as Turn-on
in the Standard Menu (Table 29), the
actual setting will become the
permanent Turn-On program when the
question Overwrite Turn-on program? is
answered with "YES" (Table 33).

40/104
Turn-on Saving Key No. Parameter Value Unitt
Confirmation Mode DDD –
Basic Rate 60 ppm
1 Overdrive Stimulation
Overwrite 240 ppm
Turn-on Start Rate
program? 2 Atrial Amplitude 5 V
Atrial Sensitivity 0.5 mV

YES 3 Atrial Pulse Duration 1 ms

Ventricular Amplitude 8 V
NO 4
Ventricular Pulse
0.75 ms
Duration
↑_ 5 Ventricular Sensitivity 1 mV
Table 33: Turn-on Saving AUTO
A-V Delay ms
Confirmation (i.e. 200)
AUTO
PVARP ms
(i.e. 280)
After storing the user defined Turn-on AUTO
MTR ppm
Program, the PACE 203H™ abandons (i.e. 100)
the Turn-On Menu and returns to the Automatic Sensing OFF –
Mode Menu. Atrial Trigger Option OFF –

is answered with "NO", the PACE Table 34: Manufacturer’s Turn-on


203H™ returns to the (previous) Program
Standard Menu.
Pressing the ↑_ key forces the PACE Warning: The Manufacturer's Turn-on
203H™ to revert to the Mode Menu. program is intended to provide a safe
and conservative pacing therapy for the
patient. A change of this program is only
Note: Upon the manufacturer’s shipping recommended for experienced users.
of the PACE 203H™, the Turn-On
Program parameter are set according to Note: Because the sensitivity and the
Table 34. options are also stored, the pacing
mode stored may also be one of the
supplementary modes (D00, DVI, VAT,
DAT, V00, A00, AAT and DDD+AT).
Note: Because the automatic state of
the sensitivity adjustment is also stored,
turning on a pacemaker with a Turn-on
program with Auto Sense = ON is
followed by the request to search for
initial sensitivity values, as described in

41/104
chapter 9.9.4.2. Because the patient Parameters/Options Key No.
should be provided with a therapeutic Menü
pacing program immediately after
turning on the PACE 203H™, storing a PVARP & MTR 1
Turn-On Program with Auto Sense
enabled is NOT recommended. ARP (AAI, AAT) 2
Note: The Turn-on Program also stores
a changed Overdrive Stimulation Start Pulse duration 3
Rate.
Options 4
9.12 Other Parameters/Options
Setting ↑_ 5
The PACE 203H™ provides the Table 35: Parameters/Options-
capability of setting additional Menu 1
parameters and enabling or disabling
options. To accomplish this, the
Parameters/Options Menu must be
entered via the softkey labeled Main 9.12.1 Adjustment of PVARP
Menu and, then, via the softkey labeled
The post ventricular atrial refractory
Parameters/Options.
period (PVARP) can be adjusted by
If the lower display is off, then pressing pressing the softkey PVARP & MTR in
the key Unlock/Lock turns the display the Parameters/Options Menu. The
on. PVARP& MTR Menu appears as shown
in Table 36.
The Parameters/Options Menu appears
as shown in Table 35. PVARP & MTR Menu Key No.

At this point the corresponding softkeys


for the menu setting PVARP & MTR, PVARP ↑ 1
pulse duration, difference AVD-PVD,
ARP or further options can be chosen. 280 ms ↓ 2

MTR ↑ 3

106 ppm ↓ 4

↑_ 5
Table 36: PVARP & MTR Menu
The PVARP can be changed with the
softkey labeled ↑ or the softkey labeled
↓ in increments / decrements of 10 ms
between 100 ms and 500 ms,
respectively, as shown in Table 37.

42/104
The parameter change appears PVARP & MTR Menu Key No.
immediately in the second line of the
lower display and will become effective PVARP ↑ 1
with the next intervention interval.
Parameter Manually Adjustable Unit 280 ms ↓ 2
Values
MTR ↑ 3
PVARP 100 (10) 500 ms
Table 37: Manually Adjustable Values 106 ppm ↓ 4
for PVARP
↑_ 5
Note: Changing the PVARP value
manually will also switch off the Table 38: PVARP & MTR Menu
automatic setting, if it was activated (see The MTR can be changed with the
chapter 9.9.2).
softkey labeled ↑ or the softkey labeled
Note: The safety feature "protection ↓ in increments / decrements of 2 ppm
against abnormal settings" (see chapter between 80 ppm and 230 ppm,
9.21.7) prevents increasing the PVARP respectively, as shown in Table 39
to a value which would not guarantee a
The parameter change appears
minimum atrial sensing phase, or
immediately in the fourth line of the
Wenckebach behavior. Any limitation of
lower display and will become effective
the PVARP setting will be indicated by a
with the next intervention interval.
transient warning message in the lower
display.
Note: The PVARP of the PACE 203H™ Parameter Manually Adjustable Unit
is divided in two parts: an absolute part Values
and a relative part. In the absolute part MTR 80 (2) 230 ppm
(90 ms) no atrial events are recognized. Table 39: Manually Adjustable Values
In the relative part (the remaining) atrial for MTR
events are registered, but not tracked.
This algorithm prevents misinterpretation Note: Changing the MTR value manually
of ventricular events after premature will also switch off the automatic setting,
atrial events as extrasystoles (PVCs) as if it was activated (see chapter 9.9.3).
described in chapter 9.21.5.
Note: The safety feature "protection
against abnormal settings" (see chapter
9.12.2 Adjustment of MTR
9.21.7) prevents decreasing the MTR to
The maximum tracking rate (MTR) can a value smaller than the basic rate.
be adjusted by pressing the softkey Increasing the MTR will be limited to a
labeled PVARP & MTR in the value that still guarantees Wenckebach
Parameters/Options Menu. behavior. The limitation of the MTR
value will be indicated by a transient
The PVARP &MTR Menu appears as
warning message in the lower display.
shown in Table 38.

43/104
9.12.3 Adjustment of Pulse Duration 9.12.4 Adjustment of Atrial
The pulse durations can be adjusted by Refractory Period ARP (AAI,
pressing the softkey labeled Pulse AAT)
duration in the Parameters/Options By pressing the softkey labeled ARP in
Menu. the Parameters/Options Menu, the ARP
Menu appears as shown in Table 42.
The Pulse Duration Menu appears as
shown in Table 40. The ARP Menu can be used to change
the ARP default setting of 250 ms.
Pulse Duration-Menu Key No.

A ↑ 1 ARP (AAI, AAT) Menu Key No.

1.00 ms ↓ 2 ARP = 250 ms  1

V ↑ 3
ARP ↑ 2
0.75 ms ↓ 4
250 ms ↓ 3
↑_ 5
Table 40: Pulse Duration Menu 4
The pulse duration can be changed with
the softkey labeled ↑ or the softkey ↑_ 5
labeled ↓ in increments / decrements of
0.05 ms between 0.05 ms and 1.50 ms, Table 42: ARP (AAI, AAT) Menu
respectively, as shown in Table 41. The ARP can be changed with the
The parameter change appears softkey labeled ↑ or the softkey labeled
immediately in corresponding line of the ↓ in increments / decrements of 5 ms
lower display and will become effective between 240 ms and 400 ms,
with the next intervention interval. respectively, as shown in Table 43.

Parameter Manually Adjustable Unit Parameter Manually Adjustable Unit


Values Values
Atrial Pulse ARP 250 (5) 400 ms
0.05 (0.05) 1.50 ms
Duration Table 43: Manually Adjustable Values
Ventricular for ARP (AAI, AAT)
0.05 (0.05) 1.50 ms
Pulse Duration
Table 41: Manually Adjustable Values
for Pulse Duration The parameter change appears
immediately in corresponding line of the
lower display and will become effective
with the next intervention interval.

44/104
9.12.5 Setting Options Options Menu 2 Key No.
The Options Menu consists of two parts.
It appears as shown in Table 44 und Language 1
Table 45. By pressing the softkey more..
it can be switched between Options PMC  BPI  2
Menu 1 and Options Menu 2.
By pressing the softkey labeled Options 3
in the Parameters/Options Menu, the
Option Menu 1 for enabling and
disabling acoustic signals (beep), alarms more … 4
and atrial trigger appears, as shown in
Table 44. ↑_ 5

Table 45: Options Menu 2

Options Menu 1 Key No.


9.12.5.1 Enable and Disable Acoustic
Beep  1 Indicator
With the softkey labeled Beep in the
Alarms  2 Options Menu 1, the acoustic indicator
can be enabled or disabled. The
Atrial corresponding check box then gets
Trigger  3 checked or unchecked, respectively. If
the acoustic indicator is enabled, the
more … 4 symbol for a loudspeaker (see Figure 7)
appears in the upper display.

↑_ 5

Table 44: Options Menu 1


Figure 7: Acoustic Indicator (Symbol
for a Loudspeaker)
The user language can be set via the
Options Menu 2 by pressing the softkey With the acoustic indicator enabled, the
labeled Language as shown in Table 45. delivery of a stimulus is indicated by a
high-pitched tone (beep), whereas
The meaning of the second line in sensing of a P or R wave will be
Options Menu 2 is explained in sections signaled by a beep of lower frequency.
9.19 and 9.20.
Once enabled, the acoustic indicator is
disabled by pressing the softkey labeled
Beep.

45/104
The check mark disappears from the
corresponding check box. The
loudspeaker symbol disappears from the
upper display, as well.
Figure 8: Alarms Off (Crossed Bell)
Upon powering off the PACE 203H™, it Symbol
retains the enabled / disabled status of
the acoustic indicator. Thus, the enabled
/ disabled status of the last usage is
restored again upon powering on the To reactivate the lead surveillance
PACE 203H™. feature and the protection against
Upon the manufacturer’s shipping of the abnormal settings the Alarms key must
PACE 203H™, the acoustic indicator is be pressed again. This causes the
disabled according to Table 42. check mark to appear in the
corresponding check box, and the
Note: The warning and error signals are crossed bell symbol in the upper display
not influenced whether or not the disappears.
acoustic indicator is enabled or disabled.
Warning and error signals are always Note: If previously disabled Alarms are
accompanied by acoustic signals. enabled, the PACE 203H™ will solve
possible conflicts caused by abnormal
settings using its Automatic setting
feature. The fact that a parameter has
9.12.5.2 Enable and Disable the been adapted according (see chapter
Alarms 9.21.7 and 9.9).
With the softkey labeled Alarms in the
Options Menu, the lead surveillance
feature (see chapter 9.21.2) and the 9.12.5.3 Enable and Disable Atrial
protection against abnormal settings Trigger Function
(see chapter 9.21.7) ) can be enabled or
disabled. When the atrial trigger function is
enabled, the PACE 203H™ delivers an
Warning: Alarms should only be atrial stimulus immediately after an atrial
disabled in a controlled environment, sensing within the atrial sensing phase.
specifically, where medical professionals
continuously monitor the patient. With the softkey labeled Atrial Trigger in
the Options Menu, the atrial trigger
Upon turning on the PACE 203H™, the function can be enabled or disabled. In
alarms are always enabled. case the atrial trigger function is
enabled, the corresponding check box is
With alarms enabled, pressing the checked. Furthermore, the PACE
softkey labeled Alarms causes the
203H™ indicates the label A-TRIG in
check mark to disappear from the
the atrial stimulation section of the upper
corresponding check box. In the upper
display.
display, the crossed bell symbol (see
Figure 8) is shown.

46/104
Note: The atrial trigger function is Language Menu Key No.
available in the primary modes AAI and
DDD. English  1
To turn the atrial trigger function off, the
softkey labeled Atrial Trigger must be Deutsch  2
pressed again. The check mark
disappears from the corresponding
Français  3
check box, and the label A-TRIG
disappears from the upper display as
well. more... 4

Atrial Trigger is also disabled when the


primary mode is changed to VVI or VDD, ↑_ 5
because these modes do not provide
atrial stimulation. Atrial Trigger must be Table 46: Language Menu
activated again after changing back to
the primary modes AAI or DDD.
The manufacturer’s language setting
default is English.
Note: Atrial Trigger cannot be used
together with Atrial Auto Sense, because 9.13 Starting Emergency Stimulation
the triggered stimulation prevents proper
In case heart stimulation becomes
measurement of the intrinsic atrial
ineffective with the set parameters in
activation. (see also 9.9.4).
place, the PACE 203H™ provides an
Note: A sensed atrial event in the emergency stimulation program. This
relative PVARP (see chapter 9.12.1) is emergency program can be quickly and
never followed by a (triggered) atrial easily activated from all operational
stimulation – even if atrial trigger is states by pressing the key
enabled – because it may be a EMERGENCY.
ventricular echo.
Note: In case the PACE 203H™ is
locked, it must be unlocked first by
pressing the key Unlock/Lock.
9.12.5.4 Setting the User Language
The PACE 203H defaults immediately to
By pressing the softkey labeled the emergency stimulation program with
Language in the Options Menu 2, the the parameter shown in Table 47.
Language Menu appears as shown in
Table 46. The activated user language is
marked. By pressing the corresponding
key the language can be selected. The
softkey labeled more... toggles through
the different language menus.

47/104
Parameter V00 Unit The pause function is implemented as
Basic Rate 80 ppm the pacing mode 0D0 (see also chapter
Atrial Amplitude 12 or V 9.6.11).
set value, if
In case any intrinsic heart activity is
higher
sensed, atrial and ventricular rates, and
Atrial Pulse 1 or ms
Duration set value, if the corresponding intervals in ms are
longer shown on the lower display. The
Ventricular 12 or V parameter values are displayed after the
Amplitude set value, if second intrinsic sensed event. These
hgher events are detected up to frequencies of
Ventricular Pulse 0.75 or ms about 700 bpm. Furthermore, the
Duration set value, if measured P and R wave amplitudes are
longer displayed as shown in Table 48.
A-V Delay AUTO ms
(i.e. 170) If no heart activity can be sensed, or the
Table 47: Emergency Stimulation sensitivity is set to a value much smaller
or much larger than the actual P or R
Warning: If the device is operating in an wave amplitude, the PACE 203H™ will
atrial mode (AAI, AAT, A00), it defaults change the sensitivity temporarily. The
first to the V00 mode with the above PACE 203H™ will attempt to find
mentioned values upon pressing the intrinsic heart activity and to measure
emergency key. However, if the lead the amplitudes optimally. When the Auto
surveillance determines that no Sense function is deactivated, this
ventricular lead is connected, the device change of sensitivity thresholds is only
switches to A00 mode. temporary. Thus, after releasing the key
labeled Pause, the previous sensitivity
Note: By pressing the emergency key, threshold is restored. When the Auto
the PACE 203H™ irrevocably abandons Sense function is activated in one or in
the last parameter setting. both channels, the changed sensitivity
setting will be taken as new start value
for the Auto Sense algorithm (see 9.9.4).
Thus, the Auto Sense function is
9.14 Pause Function reinitialized with the determined
The PACE 203H™ provides, upon the sensitivity threshold.
user’s request, the capability of briefly The display is shown as long as the key
abandoning any stimulation therapy to
labeled Pause is pressed, plus for
allow undisturbed monitoring of the
approximately an additional second after
patient’s intrinsic heart activity. This
releasing the key.
PAUSE function is activated by pressing
and holding the key labeled Pause. The acoustic indicator will automatically
be enabled temporarily, if not already
The current stimulation therapy is enabled.
interrupted to check for intrinsic events.

48/104
Pause Menu Key No. may result in a rate determination being
of too low, or unstable. Thus, if a precise
Sensed: 1 measurement of the atrial frequency is
desired, the ventricular lead must be
2 disconnected.
230 bpm
A (261 ms) If frequencies higher than the noise
2.3 mV frequency (273 bpm) are detected while
3
the key labeled Pause is pressed, the
corresponding interference symbol
50 bpm 4 (Figure 12) is displayed in the upper
V (1200 ms) display. The measured frequency is
8.0 mV 5 displayed, nevertheless, because it may
be a high-frequency tachycardia.
Table 48: Pause Menu Accordingly, for the same reason, the
setting of the sensitivity threshold
remains unchanged (unlike during
The pause function is activated for up to initialization of the Auto Sense function).
10 seconds. Subsequently, the PACE When no intrinsic activity is sensed in
203H™ resumes pacing, independent if any channel for 3 seconds while the key
the user continues to hold the key labeled Pause is pressed, any
labeled Pause. measured values already displayed are
erased.
As long as the key labeled Pause is
Warning: Because pacing is interrupted, pressed, all other keys and dials are
the user must be aware that the patient disabled.
receives no stimulation therapy and
support by the pacemaker as long as
the key labeled Pause is pressed.
9.15 Connecting the Leads to the
PACE 203H
Notes:
9.15.1 Lead Types
The PAUSE function is always
For temporary stimulation of the heart
operational, except during application of
with the PACE 203H™, temporary
atrial overdrive stimulation. It does
transvenous leads or permanent leads,
deploy, however, during High-Rate
regardless of whether they are of bipolar
standby and can be used to measure
or unipolar configuration, can be used.
the frequency of the atrial tachycardia,
The use of myocardial leads (up to 6
which assists in finding the optimal
electrodes) and VDD single pass leads
overdrive stimulation rate.
is also permissible.
The determination of the intrinsic atrial
rate can be influenced by the cross Temporary, transvenous pacing
blanking after ventricular sensing. Thus, leads: These leads are advanced into
sensing of atrial activations can get the heart via a vein and are connected
"lost" due to the cross blanking, which

49/104
to the PACE 203H™ either directly or by connections are located to the left and
means of an extension cable. the atrial connections to the right.
Heart wires (myocardial pacing leads The indifferent terminals (+) are colored
Heart wires (myocardial leads) are red, while the different terminals (–) are
affixed to the heart during open heart colored black.
surgery, when it is expected that the
patient may need stimulation for a
limited period of time after surgery. The
heart wires are connected to the PACE
203H™ by means of an extension cable.
Permanent pacing leads: Prior to
implantation of a permanent pacemaker, Figure 9: Lead Connection Terminals
or during a pacemaker change,
stimulation may be properly maintained
with the assistance of the PACE 9.15.3 Connection Configurations
203H™. The permanent lead is
connected to the PACE 203H™ either Various connection configurations exist,
directly or by means of an extension depending on the leads used. One may
cable. choose between bipolar leads or
unipolar leads with indifferent electrode.
Bipolar: When using a bipolar,
Warning: All lead systems must be transvenous stimulation lead, the distal
connected to type CF devices only, pole of the lead is connected to the
because of the danger of current being different (–, black) terminal of the
diverted to the heart. Devices that are corresponding channel on the PACE
connected to a mains supply pose 203H™. The proximal pole is then
increased danger for current diversions connected to the indifferent (+, red)
to the heart. terminal.
For exact specifications of leads and When using heart wires, the connection
patient cables, please refer to our is arbitrary, since both electrodes are
product catalog. affixed in the myocardium. The
connection may be selected to obtain
9.15.2 Lead Connection Terminals the best sensing and stimulation
parameters.
At the top of the pacemaker are
protected terminals (collets) located for When a stimulus is delivered, the
plugs with a diameter of 0.9 mm to 2.0 current flows from the negative to the
mm. Viewing the display face of the positive electrode, and causes a
PACE 203H™, the atrial connections myocardial depolarization, which then
are located to the left and the ventricular leads to a heart muscle contraction.
connections to the right (see Figure 1). Unipolar: When using a unipolar lead, it
Viewing the top of the PACE 203H™, as is connected to the different (–, black)
shown in Figure 9, the ventricular terminal of the corresponding channel of

50/104
the PACE 203H™. In order to close the sources must be kept away from the
circuit and allow a stimulus, an stimulation system.
indifferent electrode must be connected
to the indifferent (+, red) terminal of the The user must follow the steps
corresponding channel of the PACE described below when connecting a
203H™. This electrode must have a stimulation lead to the PACE 203H™:
large surface and must be attached 1. Turn off the PACE 203H™.
subcutaneously. Open the collets (see Figure 9)
If unipolar stimulation is given in the 2. If a patient cable is used, this
atrium as well as in the ventricles, one cable must be connected first
indifferent electrode is sufficient. The to the PACE 203H™, before
indifferent terminals of all channels can the stimulation lead is
be connected to each other. connected to the patient cable.
Connect the ventricular pacing
Warning: The unipolar configuration is
leads, or the corresponding
more susceptible to external noise, as
extension cable, to the
compared to the bipolar configuration.
ventricular output terminals of
the PACE 203H™. Connect the
9.15.4 Connecting the Leads atrial pacing leads or the
The PACE 203H™ must be turned off corresponding extension cable,
while the stimulation leads are to the atrial output terminals of
connected to it. the PACE 203H™. Make sure
the polarity is correct.
Warning: While the leads are being
inserted into the patient, and the PACE 3. Secure the connections by
203H™ is connected, continuous ECG manually turning the collets
monitoring is mandatory. For emergency clockwise.
situations, a defibrillator must always be
4. If extension cables are used,
in a ready-to-use state. The user must
connect the leads to the
ensure that all devices in the vicinity of
extension cables.
the patient are properly grounded.
5. Turn on the PACE 203H™ and
Warning: Before handling the set the desired pacing mode.
pacemaker, the patient cable, or the
indwelling stimulation lead, the 6. Determine the sensitivity
electrostatic potential between user and threshold (see chapter 9.16).
patient must be equalized.
7. Determine the cardiac capture
The stimulation leads provide a direct, threshold (see chapter 9.17).
low-resistance current path to the heart.
8. Control the proper functioning
Therefore, it is absolutely mandatory
of the PACE 203H™ with the
that the connector plug is not touched
assistance of an ECG monitor
with bare hands or can come in contact
or recorder.
with electrically conductive or wet
surfaces. All possible static electricity

51/104
9.16 Determining the Sensitivity threshold. In order to create a
Threshold ‘safety cushion’, the sensitivity must
be raised further. The set value
Determining the sensitivity threshold can should be a 1/2 to 1/3 of the
be accomplished by using the Auto sensitivity threshold value.
Sense function (see chapter 9.9.4) or
manually. Both methods require that the
patient has intrinsic heart activity.
Warning: An unnecessarily high
For manual determination of the sensitivity (i.e. smaller sensitivity value)
sensitivity threshold, the patient must increases the probability that the proper
have heart activity with a rate that is pacemaker function will be affected by
hemodynamically tolerated over a longer external interference, and the
period of time (a few minutes). pacemaker will switch to asynchronous
stimulation (see also chapter 9.21.3).
The manual determination of the
sensitivity threshold is performed as
follows:
6. In case a dual chamber mode is
1. Set the stimulation amplitude (in chosen, the procedure must be
dual-chamber mode: both channels) repeated for the other channel.
to the smallest value so that any
asynchronous stimulation, which
occurs during the procedure, Note: When setting the sensitivity
remains ineffective. threshold, it is very helpful to turn on the
2. Set the basic rate 10 ppm below the acoustic signal of the PACE 203H™,
patient’s intrinsic rate. because it emits a different pitched tone
for stimulation and for sensing (see
3. If the PACE 203H™ is operating in chapter 9.12.5.1).
a dual chamber mode, the A-V
delay must be set longer than the
patient’s intrinsic A-V interval.
9.17 Determining the Cardiac
4. In case the pacemaker already
Capture Threshold
senses intrinsic heart activity in the
atrial or ventricular channel, set the Warning: If the patient has a sufficient
sensitivity at a low level (i.e. raise intrinsic heart rate, the determination of
the sensitivity value), such that the the sensitivity threshold must be carried
PACE 203H™ will not sense out before the cardiac capture threshold
intrinsic events in the atrium and can be determined. This is done in order
ventricle, respectively. Eventually, to make certain that no asynchronous
the PACE 203H™ operates in superimposition of intrinsic rhythm and
asynchronous pacing mode. stimulation occurs.
5. Raise the sensitivity (i.e. lower the To determine the cardiac capture
sensitivity value) again until threshold, the following steps should be
stimulation is inhibited. This taken:
sensitivity set equals the sensitivity

52/104
1. Set the basic rate at least 10 ppm inserted. Therefore, in case of mains
above the patient’s own rate. If the failure or accidentally interrupting mains
PACE 203H™ is already effectively connection the PACE 203H™ continues
stimulating, lower the stimulation operating and sustains its life supporting
amplitude, until the stimulus is no function. The battery will not be
longer effective. discharged during mains operation. The
PACE 203H™ indicates mains operation
2. In case the PACE 203H™ is
with a green LED labeled Mains
operating in a dual chamber mode,
between the keys Off and On.
the A-V delay must be set shorter
than the patient’s intrinsic A-V For connecting the PACE 203H™ to AC
interval. power, please adhere the following
instructions:
3. Raise the stimulation amplitude
slowly until the stimulus is effective 1. Turn off the PACE 203H™.
again. The amplitude set equals the
cardiac capture threshold. 2. Lift the cap that protects the
terminal CTRL. OUT (see Figure 2).
4. In order to create a ‘safety cushion’,
the stimulation amplitude must be 3. Insert the connector of the wall
raised further. The set value should power supply into the terminal
be from two to three times the CTRL. OUT . Secure the
cardiac capture threshold. connection by manually turning the
collets clockwise.
5. In case a dual-chamber mode is
chosen, the procedure must be 4. Insert the wall power supply into a
repeated in the other channel. mains socket.
5. Control the proper lighting of the
green Mains LED.
Warning: If the PACE 203H™ is to be
used for a longer period of time on a 6. Turn on the PACE 203H™ in the
patient, the stimulation threshold should usual way.
be checked from time to time (the first
Warning: If the green LED Mains does
time after a few hours, then daily),
not light, the PACE 203H™ operates
because an increase in the capture
battery powered with the specified
threshold can occur.
battery lifespan.

9.18 Mains Operation (Option)


9.19 Controlling an Intra-Aortic
Option not available in the U.S. Balloon Pump (Option)
The PACE 203H™ possesses the Option not available in the U.S.
capability for mains operation using a
The PACE 203H™ possesses the
particular, optionally available wall
capability of controlling the triggering of
power supply. For safety reasons mains
an intra-aortic balloon pump (IABP). The
operation is only possible with a battery

53/104
PACE 203H™ provides an interface to Warning: Because the ventricular
the IABP, which outputs a surface ECG- intracardiac electrogram is processed to
type signal but derived from the control the IABP, the output at the
intracardiac electrogram. terminal CTRL OUT is only enabled if a
pacing mode with sensing and / or
The PACE 203H™ utilizes the
stimulation in the ventricle is chosen.
intracardiac electrogram obtained from
The output at the terminal is disabled
the ventricular channel. The ventricular
any time the PACE 203H™ operates in
signal is processed in such a way that
a pacing modes using atrial sensing and
its characteristics (shape, impulse
/ or pacing (AAI, A00, AAT).
amplitude, base width) are similar to a
signal obtained from surface electrodes. Enable the balloon pump interface
communication by selecting the radio
The PACE 203H™ to connected to the
button assigned to BPI (manufacturer’s
IABP via an optional IABP interface
default) in the Options Menu 2 (Table
device (BPI 202), which is connected to
49). Pressing the PMC/BPI softkey in
the CTRL. OUT terminal of the PACE the will toggle between selection of the
203H™ (see Figure 1 and Figure 2). radio button assigned to PMC ( for for
The PACE 203H™ accomplishes the Pacemaker Clinic™, a program view of
®
synchronization of the IABP with the the AESCULON ) and the radio button
heart activity in both intrinsic and paced assigned to BPI (for Balloon Pump
ventricular events by a application of a Interface). Balloon pump interface
differential timing: communication is default.

Intrinsic event: Upon sensing of a Options Menu 2 Key No.


R wave, the PACE 203H™ provides a
control signal, whose processed Language 1
‘R wave’ is delayed by approximately 5
ms, assuming that the amplitude of the
PMC  BPI  2
R wave measured is twice the value of
the sensitivity set.
Paced event: Upon delivery of a 3
ventricular stimulus, the PACE 203H™
provides a control signal, whose more … 4
processed ‘R wave’ is delayed by
approximately 35 ms, accounting for the
latency period, that is, the time between ↑_ 5
actual delivery of the stimulus and the Table 49: Options Menu 2
occurrence of ventricular depolarization.
(BPI enabled)
Please see the instruction manual for
the IABP interface device for further Note: The Terminal CTRL. is protected
information. by a cap which will prevent pollution.
This cap should be in place, when the
CTRL OUT socket is not used.

54/104
®
9.20 Interfacing with AESCULON Table 50: Options Menu 2
Monitor (Option) (PMC enabled)
Option not available in the U.S. The communication between PACE
®
The PACE 203H™ can communicate 203H™ and AESCULON is established
®
with the AESCULON comprehensive via a custom interface cable connected
cardiovascular monitor, for example, to on one end to the CTRL. OUT terminal
record the pacemaker settings with the of the PACE 203H™ and on the other
®
AESCULON or synchronize PACE end to the AESCULON . Upon
® successful communication, the
203H™ and AESCULON during
automatic hemodynamic optimization of PACE 203H™ parameter settings are
®
patients. transmitted to AESCULON , and both
®
PACE 203H™ and AESCULON are
Enable the communication between synchronized regarding timely
®
PACE 203H™ and AESCULON by occurrences of atrial and ventricular
selecting the radio button assigned to stimulation.
PMC in the Options Menu 2 (Table 49). ®
The combination of and AESCULON
Pressing the PMC/BPI softkey in the provides a tool for determining the
Options Menu 2 (Table 49) will toggle hemodynamically optimal pacing
between selection of the radio button parameters. The user utilizes the
TM
assigned to PMC ( for for Pacemaker AESCULON view Pacemaker Clinic
Clinic™, a program view of the (PMC) to define a range for the atrio-
®
AESCULON ) and the radio button ventricular (AV) delay, a step width for
assigned to BPI (for Balloon Pump stepping through the aforementioned
Interface). range and a waiting period applied
between a change of the AV delay and a
period of hemodynamic measurements.
Options Menu 2 Key No. Upon a key stroke, the PACE 203H™
automatically applies the various pacing
parameter permutations for a limited
Language 1
time one after the other, while the
obtaining and recording the
PMC  BPI  2 corresponding hemodynamic
measurements. At the end of this ‘scan’,
®
the AESCULON presents the results in
3
a matrix and suggests the optimal
pacing parameter permutation, which
more … 4 can be accepted by the user, or ignored.
Please refer to the instructions for use of
®
the AESCULON monitor for more
↑_ 5
detailed information.

55/104
9.21 Internal Surveillance and Safety During a battery change, the PACE
Features 203H™ is capable of providing
continuous operation for at least
9.21.1 Battery Surveillance 30 seconds.
The PACE 203H™ is powered by a Warning: In order to provide continuous
standard 9 V battery. operation of the pacemaker during
battery changes, it is recommended that
The PACE 203H™ continuously the battery will be replaced before being
monitors the battery voltage. The battery completely drained (see also
symbol (see Figure 10), located in the chapter 11.2).
upper display, continuously indicates the
remaining battery voltage. The time-to- If the PACE 203H™ is switched off, or to
change-battery level is reached when the stand-by state after the first request
only one segment (the rightmost to change the battery has appeared, the
triangular one) remains and is blinking. battery must be replaced before the
Additionally, the warning message pacemaker is turned on again. The
Change battery! will appear battery can be changed without losing
approximately every 10 min in the lower the stored settings.
display, accompanied by an acoustic
warning signal. The battery surveillance cannot be
turned off with the Alarms on/off option
Note: The text message Change (see chapter 9.12.5.2)
battery! disappears only upon the user
pressing the key Unlock/Lock. 9.21.2 Lead Surveillance
In order to prevent system malfunctions
(Exit-Block), the PACE 203H™ checks,
during each delivery of a stimulus,
whether an interruption or a short circuit
Figure 10: Battery Symbol of the pacing system has occurred.
If the time-to-change-battery warning is
disregarded, and the voltage falls below
a critical value, then the empty battery Warning: The lead surveillance is
symbol blinks, requesting an immediate enabled only if the pacing amplitude is
change of the battery (see chapter 11.2). set greater than or equal to 2.0 V, and if
This action is accompanied by the the pulse duration is set to a value
warning message Hurry up! Change greater than or equal to 0.15 ms.
battery! and the acoustic warning
Short Circuit in Pacing System:
signals every 2 minutes.
A short circuit in the pacing system can
The lifespan of the battery depends on occur, for example, because of defective
the set stimulation parameters. In insulation. An interruption of the
chapter 13 "Technical Data" the lifespan stimulation system can be caused by
for 100 % stimulation in two usual lead or cable breakage, or an incorrectly
modes (DDD, VVI) are given for alkaline made connection
and lithium batteries.

56/104
A short circuit will be indicated by the three failed deliveries of pacing stimuli),
warning messages Atrial output short or the open lead symbol (see Figure 11)
Ventricular output short (see appears in the respective section of the
chapter 9.21.9) located in the lower upper display, accompanied with an
display. Each warning message will also acoustic warning signal. The warning
be accompanied by an acoustic warning signal is repeated as long as the
signal. Only pressing the key interruption persists. In addition, the
Unlock/Lock will discard the warning warning messages Atrial lead
messages for a few seconds, but will disconnected or Ventricular lead
reappear if the failure persists. disconnected. appear in the lower
display (only if the lower display is
Interruption in Pacing System: switched on). Once the problem for the
In case of an interruption of the malfunction is resolved, the warning
stimulation circuit – which is defined as message disappears automatically, and
3 successive stimulations not properly the acoustic warning signal ceases.
carried out – the open lead symbol (see
Figure 11) will be shown in the upper The user can confirm the error message
display, with respect to the by pressing the key Unlock/Lock.
corresponding atrial and/or ventricular Confirmation causes the warning
channel. message to disappear and the acoustic
signal to cease, despite the fact, for
example, that the lead remains
disconnected. In this case, the PACE
203H™ ceases the lead-breakage
Figure 11: Open Lead Symbol
warning message upon the user’s
confirmation, but continues to display
the open lead symbol if the malfunction
Additionally, the PACE 203H™ performs remains present.
lead surveillance with an intelligent lead-
breakage warning function, which If the disconnection is detected when
operates independently in the atrial and the lower display is switched off, only the
ventricular channels. The PACE 203H™ acoustic signal is enabled. A first press
enables this function in a channel of the key Unlock/Lock switches on the
automatically upon recognition that a lower display and the backlight, and
lead is connected. According to the unlocks the device. Then, the warning
definition of the function, a lead is message is displayed. A second press
connected upon three sensed or paced on the key Unlock/Lock confirms the
events. In case an atrial or ventricular error message.
lead, or any part of the circuit, fractures
(accidentally or on purpose, identified by If the disconnection is detected when
three failed deliveries of pacing stimuli), the lower display is switched on, but the
the open lead symbol. device is locked, a first press of the key
Unlock/Lock switches the backlight on,
In case an atrial or ventricular lead, or and unlocks the device. The user
any part of the circuit, fractures confirms the warning message with a
(accidentally or on purpose, identified by second press of the key Unlock/Lock.

57/104
Once the leads are connected and Warning: The lead surveillance function
functioning appropriately, the lead- should only be disabled if a medical
breakage warning function is enabled professional continuously monitors the
automatically after three sensed or patient
paced events.
9.21.3 Interference and Mode
Switching
Note: As long as the stimulation
The PACE 203H™ is designed for
amplitude is set to less than 2.0 V, or optimal sensing characteristics and
the pulse duration is set less than 0.15 filtering suppression of possible noise
ms, the lead-breakage surveillance and interference of frequencies beyond
remains in the previously held status the range of intrinsic and paced
(enabled or disabled). frequencies. However, there are types of
interference whose frequencies hardly
differ from those of intracardiac signals
When the PACE 203H™ resumes (for example, noise from mains supply
operation from the stand-by mode, the and strong muscle potentials), and are
lead-breakage surveillance is always of such magnitude that complete
activated. The PACE 203H™ assumes suppression cannot be achieved.
that the pacemaker is already connected
to the pacing leads and thus, to the As a consequence, the PACE 203H™
patient. In any other case of beginning implements an algorithm, which decides,
operation, the PACE 203H™ first by means of a rate analysis of the
disables the lead-breakage surveillance. observed signal, whether the intrinsic
Thus, the user is able to set up the events or interference is sensed:
patient without being bothered by lead- Signals with a frequency below 273 bpm
breakage warnings, while the (about 4.5 Hz) will be judged as intrinsic
pacemaker is not connected to the heart signals and lead to an inhibition
pacing leads. However, upon proper (or, according to the pacing mode, to a
connection of the PACE 203H™ to the triggering) of stimulation.
pacing leads and, thus, the patient, the
PACE 203H™ enables lead-breakage Signals with a frequency above 273 bpm
surveillance after three successfully (about 4.5 Hz) will be judged as
sensed or paced events. interference and force the PACE 203H™
to an asynchronous stimulation.
If the PACE 203H™ recognizes
Note: The lead surveillance can be interference in a channel, the basic rate
turned off using the Alarms on/off option will be increased by 10 ppm, but not
(see Chapter 9.12.5.2). However, it is above the MTR (only in the case the
not recommended to disable the lead MTR is applicable, and the MTR set to a
surveillance function. value less than 10 ppm above the basic
rate) and not above 220 ppm. The
PACE 203H™ displays the symbol for
interference (see Figure 12) ) in the
upper display (atrium and/or ventricle).

58/104
ventricular stimulation. This is known as
crosstalk, or A-V crosstalk. In a
pacemaker-dependent patient, inhibition
Figure 12: Symbol for Interference of ventricular output by crosstalk results
in asystole.
For a better understanding how the
Additionally, the pacemaker changes to
PACE 203H™ prevents, or operates in
the pacing mode due to interference.
the presence of crosstalk, the A-V delay
Table 51 shows the corresponding
should be considered as an interval with
reaction.
three subportions: The ventricular
The pacemaker will switch to blanking period, the ventricular
the following mode during triggering period, also known as the
Basic mode Interference in Interference in crosstalk sensing window, and the
ventricular atrial channel remaining period of the A-V delay, in
channel which a sensed event inhibits the
ventricular output. This remaining period
DDD D00 DVI may be not extant if the sum of
D00 No change No change ventricular blanking period and crosstalk
VVI V00 No change sensing equals the A-V delay.
V00 No change No change
AAI No change A00 9.21.4.1 Ventricular Blanking Period
A00 No change No change
To prevent crosstalk, the PACE 203H™
VDD V00 VVI
initiates, after an atrial stimulus, a brief
Table 51: Mode Switching Due to ventricular blanking period. As a
Interference consequence, any possible crosstalk
from the atrial pacing stimulus is
masked, or blanked, with respect to the
Note: Interference is not recognized ventricular sensing circuitry.
(and indicated) The blanking period is of short duration
If the sensitivity of the corresponding because it is important for the ventricular
channel is set to infinite ("–.–"), or sensing circuit to be returned to the
If the corresponding channel is “alert” state relatively early during the A-
refractory. V delay so that intrinsic ventricular
activity can inhibit the ventricular output.
This is because sensing is turned off in
all these cases.
9.21.4.2 Crosstalk Sensing Window
and Ventricular Safety Pacing
The crosstalk sensing window follows
9.21.4 Crosstalk and Ventricular the ventricular blanking period. From
Safety Pacing then on, the PACE 203H™ cannot
In dual-chamber pacing modes, an atrial differentiate crosstalk from intrinsic
stimulus could be sensed on the ventricular activity.
ventricular channel, and inhibit

59/104
To prevent ventricular systole, the PACE 9.21.5 Premature Ventricular
203H™ delivers a ventricular stimulus if Contraction (PVC)
either crosstalk or intrinsic ventricular The PACE 203H™ assumes an
activity is sensed within 40 ms after the extrasystole, or premature ventricular
termination of the blanking period. This contraction (PVC), if it detects two
interval of 40 ms is referred to as the ventricular events without an atrial event
crosstalk sensing window, and the in between. The detection of a PVC is
ventricular response to as ventricular indicated in the ventricular section of the
safety pacing. This ventricular pacing upper display by the symbol for an
stimulus is delivered early, that is, 100 extrasystole (siehe Figure 13).
ms after the atrial event. This
corresponds to an abbreviated A-V
delay of 100 ms.
If the signal sensed is indeed crosstalk,
the paced ventricular stimulus at the Figure 13: Symbol for Extrasystole
abbreviated A-V delay prevents (Premature Ventricular Contraction,
ventricular asystole. PVC)
If, on the other hand, intrinsic ventricular
activity occurs during this 40 ms window,
the safety mechanism results in delivery For patients with retrograde conduction,
of a ventricular pacing stimulus within, or the sensing of a ventricular event,
immediately after, the intrinsic beat. This conducted retrogradely to the atrium,
delivery is safe because the ventricle is can lead to pacemaker mediated
refractory. Consequently, no tachycardia.
depolarization results from the Note: Pacemaker mediated tachycardia
ventricular stimulus. Furthermore, the has been defined as a reentry
stimulus is too early to coincide with arrhythmia in which the dual-chamber
ventricular repolarization or a vulnerable pacemaker acts as the antegrade limb
period. of the tachycardia and the natural
Crosstalk window and safety pacing are pathway acts as the retrograde limb 2 3.
designed to supplement the ventricular
blanking function to ensure that The PACE 203H™ prevents pacemaker
crosstalk inhibition will not occur. mediated tachycardia by implementation

The PACE 203H™ indicates ventricular


safety pacing after A-V crosstalk through 2 Furman S, Fisher JD. Endless loop
a short time interval between the tachycardia in an AV universal (DDD)
blinking of the LEDs corresponding to pacemaker. Pacing Clin Electrophysiol
ventricular sensing and pacing, 1982; 5: 486-489.
respectively.
3 Den Dulk K, Lindemans FW, Bar FW,
Wellens HJ. Pacemaker related
tachycardias. Pacing Clin Electrophysiol
1982; 5: 476-485.

60/104
of a specific PVC algorithm. The 1
PVARPMAX = − AVD − 20 ms
response is different in case of isolated Basic Rate
PVC or the initial PVC of a series, and
subsequent PVCs. The maximum PVARP is in agreement
with the rules for "abnormal settings"
9.21.5.1 Isolated PVC, or Initial PVC (see 9.21.7).
After detection of a PVC, the PVARP is After subsequent PVCs, the V-A delay is
extended to 500 ms. Therefore, any calculated to
sensed intrinsic atrial events are not
1
tracked by the ventricle. After VA Delay = − AVD
termination of the extended PVARP, the Basic Rate
atrial channel is ‘alert’ and sensing
begins again. 9.21.5.3 Absolute und relative
PVARP
Despite the extended PVARP, the
duration of the sensing phase, or the If the patient’s heart is beating at a
width of the crosstalk sensing window, is relatively high intrinsic rate, an extended
set to at least 350 ms, independent of PVARP could have the consequence
set basic rate and A-V delay. The that, after the recognition of a single
duration of the sensing phase may be extrasystole, subsequent extrasystoles
even longer, if the difference between are determined only because within the
V-A delay and extended PVARP, i.e. atrial channel, the PVARP takes up most
of the V-A delay.
1
Sen sin g Phase = − AVD − 500 ms
Basic Rate The PACE 203H™ avoids this effect by
(where AVD means the A-V delay) dividing the PVARP in two subportions,
reveals a value exceeding 350 ms. in an absolute PVARP and a relative
PVARP. Within the absolute PVARP
A sensing phase of at least 350 ms (90 ms), no atrial events are sensed.
allows atrioventricular synchronization. If Within the relative PVARP (equals the
the sensing phase terminates without an remaining 400 ms), atrial events are
atrial event sensed, the PACE 203H™ sensed, but not tracked by the
delivers an atrial stimulus - unless a ventricular channel. Thus, the PACE
further ventricular extrasystole has been 203H™ interprets a ventricular sensed
detected. event, following an intrinsic atrial event
occurred within the relative PVARP, as a
9.21.5.2 Subsequent PVCs regular beat and not as an extrasystole
If a PVC is followed by another PVC As a consequence, ventricular
without intermediate atrial sensed or responses caused by premature atrial
paced event, the behavior is adapted for contractions are not counted as
subsequent PVCs, ensuring appropriate extrasystoles.
pacing at higher basic rates.
After subsequent PVCs, the PVARP is
extended up to 500 ms, but not
exceeding

61/104
9.21.6 Runaway Protection 9.21.7 Protection Against Abnormal
The PACE 203H™ can stimulate the Settings
heart with a maximum basic rate of The PACE 203H™ continuously
220 ppm or a maximum tracking rate monitors the setting of parameters so
(MTR) of 230 ppm. The output of higher that abnormal settings, as shown in
stimulation rates, for example, due to a Table 52 can be detected and avoided.
defect in the frequency generator, is Such settings can cause danger for the
limited by an independent safety patient.
function, to a maximum of 238 ppm. This
safety function is referred to as runaway
protection.

Abnormal Settings Reason for Mode Action


Preventing
a) Basic Rate > MTR Basic Rate must not DDD Transient warning message and
be larger than MTR VDD prevent abnormal setting
DAI
VAT
DAT
b) AVD + PVARP + 20ms > To guarantee DDD Transient warning message and
LRI minimum atrial VDD prevent abnormal setting
sensing phase DAI
VAT
DAT
c) AVD + 60ms > LRI To guarantee D00 Transient warning message and
minimum V-A-delay DVI prevent abnormal setting
d) PVD + PVARP ≥ URI To guarantee DDD Transient warning message and
Wenckebach- VDD prevent abnormal setting
(with URI = 1/MTR) behavior DAI
VAT
DAT
e) ARP + 20 ms > LRI To guarantee AAI Transient warning message and
minimum atrial AAT prevent abnormal settings
sensing phase

Table 52: Abnormal Settings


Note: The protection against abnormal
settings (warning and limitation) can be
Note: All conflicts are avoided, when turned off using the "Alarms on/off"
"Auto A-V Delay", "Auto PVARP" and option (see 9.12.5.2). ). If Alarms are off,
"Auto MTR" are activated (see chapter it is the responsibility of the user to set
9.9). appropriate parameter combinations.

62/104
9.21.7.1 Possible Conflicts "" - Value to be increased
Conflicts can occur in the process of "" - Value to be decreased
changing parameter settings. In the
tables below, all possible cases are "A" - Parameter is set to "Automatic
listed together with the transient conflict Setting"
warning message, which will be
displayed on the lower display. In case "M" - Parameter is set to "Manual
of a conflict, the parameter to be Setting"
changed is limited to a value still
permissible.
The acronyms in the tables have the
following definitions:

63/104
Basic Rate MTR Message

 M Increase MTR to increase Rate


M  Decrease Rate to decrease MTR
Table 53: Conflicts Type a)

Basic Rate AVD PVARP Message


 M M Decrease PVARP or AVD to increase Rate
 A M Decrease PVARP to increase Rate
 M A Decrease AVD to increase Rate
 A M Decrease PVARP to decrease Rate
 M A Decrease AVD to decrease Rate
M  M or A Decrease PVARP or Rate to increase AVD
M M or A  Decrease AVD or Rate to increase PVARP
Table 54: Conflicts Type b)

Basic Rate AVD Message


 M Decrease AVD to increase Rate
M  Decrease Rate to increase AVD
Table 55: Conflicts Type c)

64/104
Basic AVD PVARP MTR Message
Rate (PVD)
 M A A Decrease AVD to increase Rate
 A M A Decrease PVARP to increase Rate
 M M A Decrease PVARP or AVD to increase Rate
 M A A Decrease AVD to increase Rate
 A M A Decrease PVARP to decrease Rate
 A A M Decrease MTR to decrease Rate
 M A M Decrease AVD or MTR to decrease Rate
 A M M Decrease PVARP or MTR to decrease Rate
M  M or A M or A Decrease PVARP or MTR to increase AVD
M M or A  M or A Decrease AVD or MTR to increase PVARP
M M or A M or A  Decrease PVARP or AVD to increase MTR
Table 56: Conflicts Type d)

Basic Rate ARP Message


 M Decrease ARP to increase Rate
M  Decrease Rate to increase ARP
Table 57: Conflicts Type e)

”manual”, the PACE 203H™ will, first,


9.21.7.2 Mode Switching set the PVARP to the automatic value
(see chapter 9.9.2). If this does not
solve the conflict, the A-V delay will be
The PACE 203H™ avoids conflicts as a reset to the automatic value (see
result of mode switching by selecting chapter 9.9.1 The PACE 203H™
appropriate parameter values, according acknowledges that a parameter has
to the rules set for the automatic modes been changed appropriately, with
(A-V delay, PVARP and MTR) – regards to the basic rate given, by a
independent of whether these automatic short warning message in the lower
modes for the respective parameters are display. Example:
enabled or disabled. The basic rate is
the determining parameter and will
never be altered. If a conflict type b)
(Table 52) ) must be solved and both AVD
PVARP and A-V delay are set to adapted

65/104
In this example, the PACE 203H™ 9.21.8.2 RF Surgery
changed the A-V delay. When using RF-cautery and RF-surgery
Note: Because of this maneuver, the instruments, very strong electrical and
PACE 203H does not switch to an magnetic interference is generated
automatic mode. The PACE 203H™ which can influence, or even damage,
simply determines and sets one or more electronic instruments like the PACE
parameters, similar to what the 203H™. Fibrillation currents can also be
corresponding automatic mode would generated by crosstalk into the leads
have determined. and cables.
If simultaneous use of the PACE 203H™
and electro-surgical instruments is
9.21.8 Effects from Therapeutc and required, the PACE 203H™ should be
set to an asynchronous pacing mode
Diagnostic Energy Sources
(without any sensing).
9.21.8.1 Defibrillation
The PACE 203H™ is designed in a Warning: In any case, it is mandatory to
manner such that it withstands continuously monitor the patient, and to
defibrillation discharges according to be prepared for a possible failure or
ISO 14708-1. malfunctioning of the PACE 203H™.

Warning: Any protection against


defibrillation discharges is limited, due to 9.21.9 Summary of Text Messages
the low resistance required by the The PACE 203H™ differentiates
pacemaker output circuitry. In any case, between conflicts, application related
it is mandatory to monitor the patient for and device related errors, and warning
a period of time after defibrillation, and messages.
to be prepared for a possible failure or
malfunctioning of the pacemaker.

9.21.9.1 Warning Messages Related


To protect the patient and the to Conflicting Parameter
pacemaker from current passing through Settings
the pacemaker / lead-circuit, caused by Conflict warnings alert the user that he
defibrillation discharges, the stimulation or she must adjust unusual, conflicting
circuit should always be opened, if parameter settings. Such settings can
possible. result in improper pacing therapy for the
Excessively high currents can also patient. The PACE 203H™ blocks
damage the pacemaker. conflicting parameter settings as long as
the "Alarms" functions are not disabled
(see chapter 9.21.7 and 9.12.5.2).

66/104
When the PACE 203H™ determines a after a few seconds, or after
conflict, it displays a warning message confirmation by the user. The user
in the lower display, accompanied by an confirms the warning message by
acoustic warning signal. The warning pressing the key Unlock/Lock.
message alerts the user about the Confirmation causes the warning
problem, and how to solve it. Warning message to disappear. Table 58 shows
messages appear after the recognition a list of these warning messages.
of conflicts. Depending on the warning
message, it disappears automatically

PACE 203H Meaning / Cause Measures to Eliminate the Chapter


Warning Message Conflict / Error No.
Increase The user has attempted MTR must be increased or 9.21.7
MTR to to increase the rate. This set to "Auto MTR", before 1
increase led to a conflict (type a), the rate can be further
Rate because MTR is set at an increased.
excessively low value.

Decrease The user has attempted PVARP and/or A-V delay 9.21.7
PVARP or to increase the rate. This must be decreased or set to 2
A-V Delay led to a conflict (type b or "Auto PVARP" and "Auto
to increase d), because the sum of AVD" before the rate can be
Rate PVARP and A-V delay is further increased.
excessively high. Alternatively, in case of an
implicit conflict (type d)
between PVARP plus AVD
and automatically set MTR,
the manual setting of the
MTR to a lower value is
suggested.
Decrease The user has attempted A-V delay must be 9.21.7
A-V Delay to increase the rate. This decreased or set to "Auto 3
to increase led to conflict (type b or AVD", before the rate can be
Rate c), because A-V delay is further increased.
excessively high.

Decrease The user has attempted PVARP and/or rate must be 9.21.7
PVARP or to increase A-V delay. decreased, before A-V delay 4
Rate This led to conflict can be further increased.
to increase (type b), because there is
A-V Delay inadequate time for a long
A-V delay in the presence
of a high rate and a long
PVARP.

67/104
PACE 203H Meaning / Cause Measures to Eliminate the Chapter
Warning Message Conflict / Error No.
Decrease The user has attempted The rate must be decreased, 9.21.7
Rate to increase A-V delay. before A-V delay can be 5
to increase This led to conflict (type further increased.
A-V Delay c), because there is
inadequate time for a long
A-V delay in the presence
of high rate.

Decrease The user has attempted PVARP and/or MTR must be 9.21.7
PVARP or to increase A-V delay. decreased, before A-V delay 6
MTR This led to conflict (type can be further increased.
to increase d), because there is
A-V Delay inadequate time for a long
A-V delay in the presence
of a high MTR and a long
PVARP.

Decrease The user has attempted A-V delay and/or rate must 9.21.7
A-V Delay to increase PVARP. This be decreased, before 7
or Rate led to conflict (type b), PVARP can be further
to increase because there is increased.
PVARP inadequate time for a long
PVARP in the presence
of a high rate and a long
A-V delay.
Decrease The user has attempted A-V delay and/or MTR must 9.21.7
A-V Delay to increase PVARP. This be decreased, before 8
or MTR led to a conflict (type d), PVARP can be further
to increase because there is increased.
PVARP inadequate time for a long
PVARP in the presence
of a high MTR and long
A-V delay.

Devrease The user has attempted PVARP and/or A-V delay 9.21.7
PVARP or to increase MTR. This led must be decreased before 9
A-V Delay to a conflict (type d), MTR can be further
to increase because the sum of increased.
MTR PVARP and P-V delay is
too long.

Decrease The user has attempted Rate must be decreased, 9.21.7


Rate to decrease MTR. This before the MTR can be 10
to decrease led to a conflict (type a), further decreased.
MTR because rate is set to an
excessively high value.

68/104
PACE 203H Meaning / Cause Measures to Eliminate the Chapter
Warning Message Conflict / Error No.
Decrease The user has attempted PVARP must be decreased 9.21.7
PVARP to increase the rate. This or set to "Auto PVARP", 11
to increase led to a conflict (type b), before the rate can be
Rate because the sum of further increased.
PVARP and A-V idelay is
too long with respect to Alternatively, manual setting
the rate or MTR. of the A-V delay or (in case
of a conflict type d) MTR to a
lower value is suggested.

Decrease The user has attempted A-V delay must be 9.21.7


A-V Delay to increase rate. This led decreased or set to "Auto 12
to increase to a conflict (type b or d), AVD", before the rate can be
Rate because the sum of further increased.
PVARP and A-V delay is
too long with respect the Alternatively, manual setting
rate or MTR. of the PVARP or (in case of
a conflict type d) the MTR to
a lower value is suggested
Decrease The user has attempted PVARP must be decreased 9.21.7
PVARP to decrease the rate. This or set to "Auto PVARP", 13
to decrease led to a conflict (type b), before the rate can be
Rate because the sum of further decreased.
PVARP and A-V delay is
too long with respect to Alternatively, manual setting
rate or MTR. of the A-V delay or (in case
of a conflict type d) the MTR
to a lower value is
suggested.

Decrease The user has attempted The A-V delay must be 9.21.7
A-V Delay to decrease the rate. This decreased or set to "Auto 14
to decrease led to a conflict (type b or AVD", before the rate can be
Rate d), because the sum of further decreased.
PVARP and A-V delay is
too long with respect to Alternatively, manual setting
the rate. of the PVARP or (in case of
a conflict type d) the MTR to
a lower value is suggested.
Decrease The user has attempted The MTR must be 9.21.7
MTR to decrease the rate. This decreased or set to "Auto 15
to decrease led to a conflict (type d), MTR", before the rate can
Rate because the MTR is to be further decreased.
high with respect to the
sum of PVARP and A-V Alternatively, manual setting
delay. of the PVARP or of A-V
delay to a lower value is
suggested.

69/104
PACE 203H Meaning / Cause Measures to Eliminate the Chapter
Warning Message Conflict / Error No.
Decrease The user attempted to A-V delay and/or MTR must 9.21.7
A-V Delay decrease the rate. This be decreased or set to "Auto 16
or MTR led to a conflict (type d), AVD" and "Auto MTR",
to decrease because the sum of before the rate can be
Rate PVARP, P-V delay and further decreased.
V-V delay is too long with
respect to the MTR. Alternatively, manual setting
of the PVARP to a lower
value is suggested.

Decrease The user attempted to PVARP and/or MTR must be 9.21.7


PVARP decrease the rate. This decreased or set to "Auto 17
or MTR led to a conflict (type d), PVARP" and "Auto MTR",
to decrease because the sum of before the rate can be
Rate PVARP and A-V delay is further decreased.
too long with respect to
the MTR. Alternatively, manual setting
of the A-V delay to a lower
value is suggested.

Table 58: Warning Messages Related to Conflicting Parameter Settings

If an error is detected when the lower


9.21.9.2 Warning Messages Related display is switched off, the acoustic
to Application Errors signal is then the only available warning.
Generally, if the problem is identified, A first press of the key Unlock/Lock
any application related errors can be switches the lower display and the
solved by the user. Application related backlight on and unlocks the PACE
errors result in a warning message in the 203H™. The warning message is now
lower display accompanied by an visible. A second press of the key
acoustic warning signal, which will be Unlock/Lock then confirms the error
repeated periodically. The warning message.
message alerts the user about the
problem. It appears after recognition of If an error is detected when the lower
the error and disappears after the display is switched on, but the device is
problem has been solved, or after locked, a first press of the key
confirmation by the user. The user can Unlock/Lock switches the backlight on
confirm the error message by pressing and unlocks the PACE 203H™. A
the key Unlock/Lock. Confirmation second press of the key Unlock/ then
causes the warning message to confirms the error message.
disappear, and the acoustic signal to The message (except the "... change
cease. battery" and the "…lead disconnected"
messages) will reappear a few seconds

70/104
after confirmation, if the error persists. Table 59 shows these messages.

PACE 203H Meaning / Cause Measures to Eliminate the See


Warning Message Conflict / Error Chapter
Change battery! The battery change level is Insert a new battery. 9.21.1
reached. 11.2
Note: This message will be
repeated after confirmation
every 10 minutes.

Hurry up! The critical battery change Insert a new battery. 9.21.1
level is reached. 11.2
Change Note: This message will be
battery! repeated after confirmation
every 2 minutes.
Release A key is pressed while the Release all keys during the 9.3
all keys PACE 203H is turning on. power-on process of the device
to allow a proper self-test.

Key A key is pressed for more In case no cause is apparent, ---


timeout than 120 seconds. There is and the error message
a possibility that something persists, this indicates that a
heavy is pressing on the key is malfunctioning, and the
PACE 203H, or it is being device must be sent to the
abutted against an edge of manufacturer for inspection
some sort. and service..

Atrial The atrial lead or If the lead was not 9.21.2


lead ventricular lead has been disconnected intentionally,
disconnected disconnected. The last there is a possibility that lead
stimulation could not be breakage or a disconnection of
or properly carried out. the connector cable exists.
Ventricular Check all connections,
Note: This error message connection cables and leads.
lead will not be repeated after
disconnected confirmation, even if the
disconnection persists. The
open output is only
indicated by the open lead
symbol (see Figure 11).
Note: Surveillance can only
occur, if the stimulation
amplitude is set to a value
≥ 2.0 V, and if the pulse
duration is ≥ 0.15 ms.

71/104
PACE 203H Meaning / Cause Measures to Eliminate the See
Warning Message Conflict / Error Chapter
Atrial The atrial or ventricular Check all connections, 9.21.2
output output, respectively, is connection cables and leads,
short short-circuited or low- and attempt to find the short-
resistance bridged. The circuit.
or PACE 203H was unable to
Ventricular deliver the last stimulation
output pulse properly.
short Note: Surveillance can only
occur, if the stimulation
amplitude is set to a value
≥ 2.0 V, and if the pulse
duration is ≥ 0.15 ms.

Table 59: Warning Messages Related to Application Errors

If an error is detected when the lower


9.21.9.3 Warning Messages Related display is switched off, the acoustic
to Device Errors signal is then the only available warning.
Device related errors are those which A first press of the key Unlock/Lock
are probably caused by a malfunction of switches the lower display and the
the pulse generator itself. (Some may backlight on, and unlocks the PACE
also have external reasons.) Errors of 203H™. The error warning message is
this kind result in a warning message in now visible. This is despite the fact that
the lower display, accompanied by an the error no longer present. A second
acoustic error signal, which will be press of the key Unlock/Lock then
repeated periodically. The warning confirms the error message.
message alerts the user about the
problem. It appears after recognition of If an error is detected when the lower
the error and disappears only after display is switched on, but the device is
confirmation by the user, i.e. device locked, a first press of the key
related error messages must be Unlock/Lock switches the backlight on
confirmed. (The acoustic error signal and unlocks the PACE 203H™. A
will, however, cease when the error is no second press of the key Unlock/Lock
more present.) The user must confirm then confirms the error message.
the error message by pressing the key If the error persists, the warning
Unlock/Lock. Confirmation causes the message will reappear after
text message to disappear and the confirmation. Table 60 lists the possible
acoustic signal to cease. warning messages.

72/104
PACE 203H Meaning / Cause Measures to Eliminate the See
Warning Message Conflict / Error Chapter
Unexpected Most of the device related Attempt to switch the PACE ---
device error errors are reported by this 203H off and, on waiting after a
no. ### warning message. few seconds, on again.

Restart The error number "###" can If this doesn't solve the
device or give the manufacturer a problem, remove the battery
submit to helpful hint as to locating while the device is switched
inspection the cause of the error. on, wait for a minimum of
10 minutes and then reinsert.
Note: The PACE 203H
attempts to continue its Note:. The patient must not be
operation with the current connected to the PACE 203H
parameter set, or, if this is during this maneuver.
not possible, with the
emergency mode. If the error persists despite of
these maneuvers, the PACE
If the error repeats itself, 203H must be sent to the
this may lead to an erratic manufacturer for inspection
rhythm. and service.

Stored data The memory for the Stand- In case this error repeats itself, ---
lost ! by data and the standard the PACE 203H must be sent
Manufacturer's programs has lost its data. to the manufacturer for
defaults The manufacturer's inspection and service.
will be used. standard turn-on program
(see 9.11.5) has been set
and all standard programs
have been reset to their
manufacturer's defaults.
Keyboard While turning on the device, Releasing all keys during the 9.3
error a key was pressed power-on process of the device
constantly for more than is necessary to allow a proper
10 seconds, despite of the self-check.
message "Release all keys"
In case all keys were released
and the error message
persists, this implies that a key
is malfunctioning and the
PACE 203H must be sent to
the manufacturer for inspection
and service.

Table 60: Warning Messages Related to Device Errors

9.21.9.4 Informative Messages


The PACE 203H™ displays various
informative messages on the lower

73/104
display, which alert the user that certain message to disappear. Table 61 shows
actions must be taken. These messages al list of these messages.
are accompanied by an acoustic
warning signal and require no
confirmation. They disappear
automatically after a short period of
time, or after confirmation by the user.
The user can confirm an informative
message by pressing the key
Unlock/Lock. Confirmation causes the
PACE 203H Meaning / Cause Measures to Eliminate See
Message the Problem Chapter
Startup timeout The key Unlock/Lock was not In order to turn on the 9.3
(Press Unlock) pressed within 30 sec after pacemaker the key Unlock/
turning on the pacemaker. Lock must be pressed within
30 sec after pressing the key
Note: The PACE 203H will labeled ON.
turn itself off after this
message.
Auto AVD The automatic A-V delay If the automatic A-V delay 9.9.1
turned off setting has been turned off, setting is to remain enabled,
because the A-V DLY dial has activate it again via the Auto
been turned. Menu and avoid turning the
A-V DLY dial.

Atrial The automatic atrial sensing If the automatic atrial sensing 9.9.4
Auto Sense adjustment has been turned adjustment is to remain
turned off off, because the A-SENSE enabled, activate it again via
SENSE dial has been turned. the Auto Sense Menu and
avoid turning the A-SENSE
dial.
Ventricular The automatic ventricular If the automatic ventricular 9.9.4
Auto Sense sensing adjustment has been sensing adjustment is to
turned off turned off, because the remain enabled, activate it
V-SENSE dial has been again via the Auto Sense
turned. Menu and avoid turning the
V-SENSE dial.

74/104
PACE 203H Meaning / Cause Measures to Eliminate See
Message the Problem Chapter
Switch Atrial Atrial Auto Sense and Atrial Act according to the 9.9.4
Auto Sense Trigger cannot be enabled at message. 9.12.5.3
off to enable the same time, because the
Atrial Trigger triggered stimulation prevents
proper measurement of the
intrinsic atrial activation.

Switch Atrial Atrial Auto Sense and Atrial Act according to the 9.9.4
Trigger off to Trigger cannot be enabled at message. 9.12.5.3
enable the same time, because the
Atrial Auto Sense triggered stimulation prevents
proper measurement of the
intrinsic atrial activation

Switch Atrial If atrial Auto Sense is If a shorter A-V delay is 0


Auto Sense off activated, a minimum of 30 ms necessary, atrial Auto Sense 9.9.4
to decrease AVD P-V delay is required to allow cannot be employed and
proper measurement of the must be deactivated.
intrinsic atrial activation.
Because A-V delay cannot be
shorter than P-V delay, it is
also limited to 30 ms.

AVD To resolve a parameter conflict N/A (conflict resolved by the 9.21.7


adapted during mode switching, the device).
A-V delay has been changed,
i.e. adapted to the current Please check if the altered
rate*. A-V delay is appropriate for
the patient.
OR
A-V delay has been increased
to the minimum value of 30 ms
when atrial Auto Sense has
been activated.
PVARP To resolve a parameter conflict N/A (conflict resolved by the 9.21.7
adapted during mode switching, the device).
PVARP has been changed,
i.e. adapted to the current Please check if the altered
rate*. PVARP is appropriate for the
patient.

75/104
PACE 203H Meaning / Cause Measures to Eliminate See
Message the Problem Chapter
MTR To resolve a parameter conflict N/A (conflict resolved by the 9.21.7
adapted during mode switching, the device).
maximum tracking rate has
been changed, i.e. adapted to Please check if the altered
the current rate*. MTR is appropriate for the
patient

Table 61: Informative Messages


*) Sometimes only a combination of If damage to the package is detected
adaptations resolves the conflict before use, the cable must be repacked
completely. In this case the message in a gas-permeable container and re-
reports all adapted parameters. sterilized with ethylene oxide gas at a
maximum temperature of 50°C (122°F)
10 Storage and a maximum overpressure of 1.7 bar.
Sterilization must be accomplished in
The storage temperature range of the accordance with the instructions of the
PACE 203H is –20°C (-4°F)…+60°C gas sterilizer manufacturer. After
(+140°F). Note that the device must be sterilization, aerate the items before use
within the operational temperature range to permit escape of residual ethylene
before use (+10°C (+50°F)…+45°C oxide.
(+113°F)).
The enclosed connector cables are
sterilized prior to shipment. They are Warning: Connector cables intended for
packed in double transparent sterile single use must not be re-sterilized after
packages. use.

The sterilization method is ethylene


oxide gas. Warning In case the PACE 203H™ is
The sterility is guaranteed as indicated not to be used for longer periods of time,
by the expiration date, if the package is the battery must be removed in order to
not damaged and properly stored. prevent damage from possible battery
acid leakage. Damage because of a
The sterile connection cables must be leaking battery is not covered by the
stored in a dry, cool place at manufacturer’s guarantee.
temperatures between 10°C (50°F) and
25°C (77°F). Keep out of direct light.

76/104
11 Care and Maintenance 11.2 Changing the Battery
While the device is in use, the battery
11.1 Clear and Cleaning should be changed when the battery
depletion symbol shows only one
As a precision electronic device, the
blinking segment and the warning
PACE 203H™ must be handled with
message Change battery! appears. The
corresponding care. Although the device
amount of time remaining from this
is robustly constructed, it can be
moment, until the battery is completely
damaged by heavy mechanical stress,
drained depends implicitly on the type of
for example, by dropping on a hard
battery used, and therefore cannot be
surface.
predicted with absolute certainty.
The enclosure and the keypad of the Typically, one day of reserve energy can
PACE 203H™ are protected against be expected with the recommended
accidental liquid spills. To clean the battery, if the PACE 203H™ is set to a
device, use a towel or sponge mode with standard parameters.
moistened with water or alcohol.
During a battery change, the PACE
For disinfection, the enclosure of the 203H™ is capable of maintaining
PACE 203H™ can be cleaned with continuous operation for at least
alhydex, cydex, or with detergicide. 30 seconds. Typically, this duration is
even longer. The battery change should,
Warning:. The PACE 203H™ must not therefore, be quick but does not require
be submerged in either water or in any haste.
other cleaning solution. Do not use any
scrubbing powder/liquid on the device.
The device must not be sterilized in an Warning: In order to provide continuous
autoclave. Sterilization with ultrasound operation of the PACE 203H™ during a
or gamma rays is also not permitted. battery change, the battery must not
The PACE 203H™ can be damaged by remain in the PACE 203H™ until it is
such procedures. completely drained. During operation of
the PACE 203H™, the user must not
The PACE 203H™ can be sterilized with permit the critical depletion level to
gas. The battery has to be removed. A occur, indicated by battery symbol
the temperature of 55°C and a relative shown empty and blinking, and the
humidity of 90 % must not be exceeded. warning message Hurry up! Change
The pressure must not exceed battery! and then being forced to replace
atmospheric pressure. The procedure the depleted battery with a new one with
given in the instructions for use of the great haste.
sterilization equipment shall be followed.
For validation approved methods must If the PACE 203H™ is switched off or to
be used. stand-by mode after the request to
change the battery appears, the battery
Re-usable cables must be cleaned, must be changed before the PACE
disinfected and sterilized after each use. 203H™ is turned on again
Single use cables must not be re-used.

77/104
Note: After inserting a new battery, the 3. Remove the battery from the battery
PACE 203H™ requires 30 minutes to compartment.
recharge its internal power capacitor in
4. Place a new battery in the
order to be able to perform the bridging
compartment. The orientation of the
function again.
battery polarity is irrelevant.
5. Close the battery compartment lid
until it is audibly latched in place.
Rotate the protection cover over the
battery compartment release button.
6. It is mandatory to dispose of the
battery in an environmentally
friendly, legal manner.

Warning: Avoid handling with spillage, if


the battery compartment is open!
Warning: Only insert a 9 V-battery
(6LR61) produced by a respected and
reliable battery manufacturer!

11.3 Safety Check-Ups of the


Pacemaker
In order to assure safe operation of the
PACE 203H™, the following check-ups
Figure 14: Battery Compartment
must be carried out on a regular basis.
Check-ups to be made by the
For battery replacement, please adhere manufacturer or by designated
to the following instructions: authorized persons thereof:
1. Have a new 9 V-battery (6LR61) Yearly checks:
ready.
Measuring the auxiliary currents
2. On the bottom of the PACE 203H™, Measuring the stimulation parameters
turn the protection cover of the (amplitude, pulse width) in the atrial
battery compartment release button, and ventricular channels
and push the battery compartment Measuring the rate.
release button. Open the battery
Measuring the sensitivity in the atrial and
compartment lid (see Figure 14).
ventricular channels
Make certain that the inserted
battery does not drop on the floor. Inspecting the battery surveillance and
measuring the power maintenance time
Inspecting the lead surveillance.

78/104
Inspecting the runaway protection 11.4 Product Return Policy
We recommend having the In the event that a PACE 203H™ and /
manufacturer carry out the yearly safety or its accessories are defective and / or
check-ups and the function check. are beyond repair, the user (customer)
Check-ups to be done by the user: must obtain a Return-to-Manufacturer-
Authorization (RMA) number from the
Before each use: manufacturer and then return these
Visual inspection: items to the manufacturer for repair or
proper disposal.
Inspect the device and accessories for
visible damage.
Inspect the connections for visible 12 Customer Service
damage. If you have any questions, the customer
Funktionsprüfung: service can be reached at:
Inspect all connections to see if they
Dr. Osypka GmbH – Customer Service
affix and function properly.
Earl-H.-Wood-Str. 1
Inspect all operating elements and D – 79618 Rheinfelden
displays for perfect function.
After each use: Phone: +49 (7623) - 7405 – 0

Care and cleaning of the device and of Fax: +49 (7623) - 7405 – 160
the accessories is to be done according Email: mail@osypka.de
to chapter 11.
For inspection and repair service, please
Warning: The PACE 203H™ contains contact the manufacturer:
no parts that are repairable or can be Osypka Medical GmbH
calibrated by anyone other than those Albert-Einstein-Strasse 3
authorized in writing, by the D – 12489 Berlin
manufacturer. Repairs or replacements
made by anyone except by the Phone: +49 (30) 6392 8300
manufacturer, or his authorized Fax: +49 (30) 6392 8301
designees, void the guarantee of the
PACE 203H™. Email: mail@osypkamed.com

79/104
13 Technical Data
TM
PACE 203H Dual-Chamber External Pacemaker

Measurement conditions: Environment temperature: 20 ± 2°C


Voltage supply: 9 ± 0.5 V
Load resistance: 500 Ω ± 1%
Test impulse: as specified in
ISO 14702-8
(Triangle 2 ms / 13 ms)
Pacing modes: primary:
DDD, VVI, AAI, VDD
supplementary:
D00, V00, A00, DVI, DAI, DDD+AT,
DAT
Stimulation indicators: Yellow LEDs blink
synchronous to emission
of stimulation pulses
Sensing indicators: Green LEDs blink synchronous to
sensing of P-/R-waves
Basic rate: 30 … 220 ppm ± 2 %
Maximum tracking rate (MTR): 80 … 230 ppm ± 2 %
Atrial overdrive stimulation: 70 … 1000 ppm ± 5 %
Output pulse: Cathodic, biphasic, asymmetric,
capacitive coupled,
passive discharge
Nominal pulse amplitude (500 Ω): 0.1…18 V ± 10 % ± 0.05V V
Nominal pulse amplitude (200 Ω … 2000 Ω): 0.1…18 V ± 20 % ± 0.05V V
Pulse duration: 0.05 ms ... 1.50 ms ± 10 % ± 0.01 ms
Atrial sensitivity: 0.2 … 20 mV ± 20 % ± 0.1 mV, ∞
Ventricular sensitivity: 1.0 … 20 mV ± 20 % ± 0.1 mV, ∞
Input impedance: 22 kΩ ± 20 %
Output impedance: < 10 Ω
Load impedance range: 200 Ω…2000 Ω

80/104
Load impedance measurement: 200 Ω…2000 Ω ± 20 %
(for a set stimulation amplitude equal
or higher 2,0 V and a set pulse
duration equal or higher 0,15 ms)
A-V delay (AVD): 5 ms ... 400 ms ± 5 % ± 2 ms
(minimum. 30 ms when atrial
Auto Sense is activated)
P-V delay (PVD): Determined by A-V delay – 30 ms
(minimum 5 ms when atrial Auto
Sense is not activated, minimum
30 ms when atrial Auto Sense is
activatd); may be prolonged
depending on MTR
Escape interval: Determined by basic rate ± 5 %
Atrial escape interval (V-A delay): Determined by basic rate and
A-V interval ± 5 %
Extended atrial escape interval: Determined by basic rate
(V-A delay after first PVC) and A-V delay
(minimum 850 ms) ± 5 %
Atrial refractory period: 250 ms … 400 ms± 5 % (AAI,AAT)
A-V interval plus PVARP (DDD,
VDD, DAI, VAT, DAT)
PVARP: 100 ms ...500 ms ± 5 %
- absolute portion: 90 ms ± 5 %
- relative portion: PVARP - 90 ms ± 5 %
Extended PVARP (after PVC): 500 ms ± 5 % (or maximum possible
period)

Ventricular refractory period4: 250 ms ± 5 %


Blanking period in atrial channel: 85 ms ± 5 % + 2 ms / - 6 ms after
atrial as well as biventricular
stimulation and sensing

4 Retriggered by second stimulation pulse in case of biventricular stimulation.

81/104
Blanking period in ventricular channel: 85 ms ± 5 % + 2 ms / - 6 ms after
ventricular
stimulation and sensing,
55 ms ± 5 % + 2 ms / - 6 ms after
atrial stimulation
Discharge period: 45 ms ± 5 % ± 2 ms
Crosstalk detection window: 40 ms
Nonphysiologic A-V delay: 100 ms ± 5 % ± 2 ms
(for ventricular safety stimulation)
Emergency mode: V00 (A00),
rate 80 ppm,
all amplitudes 12 V or
set value, when higher
pulse duration atrium 1,0 / ventricles
0.75 ms or set value, when longer
AVD 170 ms (AUTO),
Standard setting: Standard parameter set for each
primary mode,
plus one turn-on parameter set
Acoustic signaling: Different for stimulation, sensing and
warnings; stimulation and sensing
indication can be turned on or off
Interference detection frequency: > 4.5 Hz ± 5 % (> 273 bpm)
Interference rate: Basic rate + 10 ppm ± 2 %
(max. MTR and max. 220 ppm)
Defibrillation protection: According to IEC 60601-2-31
Runaway protection: 238 ppm ± 3 ppm
Lead surveillance: Message in case of short circuit or
interruption in the stimulation circuit
(for a set stimulation amplitude equal
or higher 2.0 V and a set pulse
duration equal or higher 0.15 ms)
Batterie: 9 Volt,
(identification per IEC 86: 6LR61)
Recommended types:
Duracell Alkaline MN1604
Sonnenschein Lithium SLM

82/104
Life span of recommended battery: Alkaline:
minimal 10 days (VVI, standard
parameter)
minimal 8 days (DDD, standard
parameter)
Lithium:
minimal 19 days (VVI, standard
parameter)
minimal 15 days (DDD, standard
parameter)
Plus 1 day reserve after first
appearance of the battery change
message
Battery depletion indication: Continuous displaying of battery
depletion symbol,
blinking of last segment and acoustic
warning when the time-to-charge
level is reached (7.2 ± 0.2 V),
blinking the empty depletion symbol
and acoustic warning when an
immediate battery change is required
(5.5 ± 0.5 V),
Power maintenance during battery change: Minimum 30 s
Mains operation (option): with battery inserted, indication by
LED
Operating temperature: +10°C (+50°F)…+45°C (+113°F)
Storage temperature (w/o battery): –20°C (-4°F)…+60°C (+140°F)
Operation in explosion hazard areas: The device may not be used in areas
where flammable agents are present.
Housing dimensions (L×W×D): Approx. 200 mm × 96 mm × 38 mm
Overall dimensions (L×W×D): Approx. 212 mm × 96 mm × 51 mm
(with terminals and dials)
Wight without battery: Approx. 445 g
Wight with battery: Approx. 490 g
Lead connections: Protected terminals (collets) for plugs
with 0.9 mm – 2.0 mm diameter
Other interface: Intra-aortic balloon pump interface

83/104
Safety standards: IEC 60601-1, IEC 60601-2-31
Type CF protection against electric
shock
EMC standards: IEC 60601-1-2, IEC 6100-4-2,
IEC 61000-4-3
CISPR 11 class B
European Classification Class IIb
acc. EC-directive 93/42/ECC
US Classification Class III
acc. 513 FD&C
21 CFR 862-892 [807.87(c)]
Product code 74DTE
acc. 21 CFR 870.3600
We reserve the right to make changes without notice.

14 Delivery Unit
The PACE 203H is delivered in a carrying case with the following content:
• Dual-Chamber External Pacemaker PACE 203H
• 2x Extension cables XI.TME (D 2P-SP)
• Protection bag
• PACE 203H Instructions for Use
• PACE 203H User Notes

The following accessories are available optionally:


• Wall Power Supply PACE for mains operation of PACE 203H
• Interface BPI 202 for controlling an intra-aortic balloon pump

Nevertheless, should you recognize that


15 Conditions of Guarantee and a product under warranty performs
inefficiently, or improperly, you must
Liability Restrictions
return it to us within 30 days of
The medical technology products of the occurrence of the malfunction. Please
company Osypka Medical GmbH are enclose a description of the defect or
produced from high quality materials fault. The product in question will then
and under strict adherence to controlled be thoroughly examined in our factory.
and proven manufacturing processes. We will repair or replace, free of charge,
The quality is continuously verified all components that are found to be
during production and assured before defective.
delivery.

84/104
The guarantee expires 24 months after resulting from the failure or malfunction
delivery of the product to the user of the product or accessories, whether
(customer). such claim is based on warranty,
contract, tort, or otherwise. No person
This guarantee does not include any has the authority to bind the
batteries. manufacturer to any representation or
Product damage that occurs through warranty contrary to, or in addition to this
improper storage or use, arbitrary warranty. There are no other warranties,
alterations to the product, use other than which extend beyond the face hereof.
that which the product is intended for, or
by unauthorized re-use and re-
sterilization, are explicitly excluded from
the guarantee.
The right to this guarantee is voided, if
the stipulated safety check-ups are not
regularly carried out.
If inspections, interventions, alterations
or changes are made by parties other
than those authorized in writing by the
manufacturer, the guarantee becomes
void.
This guarantee only applies to the repair
or replacement of the device itself. All
further claims for replacement by the
purchaser and from third parties are
excluded. All risks that exist in
connection with the medical application
of our products are solely and explicitly
the responsibility of the purchaser, user
or patient, if applicable.
THIS WARRANTY IS GIVEN IN LIEU
OF ANY OTHER WARRANTY;
EXPRESSED OR IMPLIED;
INCLUDING, BUT NOT LIMITED TO,
ANY WARRANTY OF
MERCHANTABILITY OR FITNESS FOR
PARTICULAR PURPOSE.
The remarks set forth herein contain the
sole remedy available to any person.
The manufacturer will not be liable to
any person for any medical expenses, or
any direct or consequential damages,

85/104
16 Appendix A: Glossary
Atrial escape interval Time between ventricular sensed beat or a
ventricular pulse and the succeeding non-triggered
atrial pulse of a pulse generator, i.e. the time when
the pacemaker waits for a P-wave.
Atrial tracking In an atrioventricular mode, a ventricular stimulus
tracks an detected atrial P-wave (after P-V delay)
up to the programmed maximum tracking rate. This
results in an atrial controlled timing of the
pacemaker.
Atrial triggering A detected P-wave causes an atrial stimulus.
Atrioventricular interval Delay between an atrial pulse or the sensing of an
(AVI) atrial depolarization and the subsequent ventricular
pulse or the sensing of a ventricular depolarization.
The actual AVI can be shorter then the adjusted A-
V delay or P-V delay in case of a spontaneous
ventricular depolarization.
A-V crosstalk Detection of the atrial output pulse by the
ventricular sensing amplifier which may result in
ventricular output inhibition.
A-V delay (AVD) Programmed delay between an atrial pulse and the
subsequent non-triggered ventricular pulse under
the condition that no sensing of a ventricular
depolarization was detected by the pulse
generator.
Basic rate Pulse rate of an pulse generator, either atrial or
(short: rate) ventricular, unmodified by sensed cardiac or other
electrical influence.
Basic pulse interval Pulse interval unmodified by sensed cardiac or
other electrical influence.
Basic pulse interval = 1 / basic rate.
Blanking period Period during which a sensing function of a pulse
generator is disabled.
Cross blanking period Period during which the sensing function of the
pulse generator is disabled in one channel
because of an event in the other channel.

86/104
Escape interval Time between a sensed beat or a pulse and the
succeeding non-triggered pulse of a pulse
generator.
External pacemaker Non-Implantable pulse generator and patient
cables (if used).
Inhibition The effect of the pulse suppression in a demand-
pacemaker at sensing of a cardiac depolarization
or other electrical influence in the same chamber.
Interference pulse rate Pulse rate with which the pulse generator responds
when it senses electrical activity other than that
from the myocardium and recognizes that as
interference.
Lower rate interval (LRI) The longest period between consecutive (paced or
sensed) events occurring in the relevant chamber.
(In absence of sensed cardiac or other electrical
influence: equal to basic pulse interval.)
Maximum tracking rate Maximum ventricular pacing rate in response to
(upper rate, MTR) sensed atrial activity. Expressed differently it is the
maximum pulse rate at which the non-implantable
pulse generator will respond on a 1:1 basis to a
triggering signal.
Non-implantable pulse generator Medical electrical equipment with an internal
electrical power source which is intended for use
outside the body and which produces a periodic
electrical pulse intended to stimulate the heart
through a lead (or combination of lead and patient
cable).
Patient Cable Device attached to the terminals of a non-
implantable pulse generator so that the distance
between it and the pacing lead can be increased.

87/104
Post-ventricular atrial refractory Period after a ventricular event (whether sensed or
period (PVARP) paced), during which synchronous ventricular
pacing is disabled, regardless of any atrial event.
The PVARP of the PACE 203H is divided in two
parts: an absolute part and a relative part. In the
absolute part (90 ms) no atrial events are
recognized. In the relative part (the remaining)
atrial events are registered, but not tracked. This
algorithm prevents misinterpretation of ventricular
events after premature atrial events as
extrasystoles (PVCs).
Premature ventricular contraction A sensed ventricular event not preceded by a
(PVC) sensed atrial event. Also named extrasystole.
Pulse (Monophasic) electrical output of a pulse generator
intended to stimulate the myocardium.
Pulse amplitude Magnitude of the pulse expressed in volts or
milliamperes.
Pulse duration Duration of the pulse.
Pulse interval Time interval between identical points of two
consecutive pulses, expressed in milliseconds.
Pulse rate Number of pulses per minute
1
1 ppm = 1/60 s−
P-V delay (PVD) Delay between the sensing of an atrial
depolarization and the subsequent non-triggered
ventricular pulse under the condition that no
sensing of a ventricular depolarization was
detected by the pulse generator.
Refractory period Period during which a pulse generator will not
respond to a beat.
Sensitivity Minimum signal, expressed in millivolts, required to
consistently control the function of the pulse
generator.
Upper rate interval (URI) The shortest period allowable between paced or
sensed events, while still maintaining 1:1
atrioventricular synchrony (equal to 60 / MTRppm).
V-A delay (VAD) Other expression for atrial escape interval.

88/104
Ventriculoatrial interval (VAI) Delay between an ventricular pulse or the sensing
of a ventricular depolarization and the subsequent
atrial pulse or the sensing of a atrial depolarization.

89/104
17 Appendix B: Menu Tree

90/104
91/104
18 Appendix C: EMC Guidance and Manufacturer’s Declaration
The PACE 203H™ needs like every medical electrical device special precautions
regarding electromagnetic compatibility (EMC). This section is intended to give the user
guidance and information about all aspects to ensure that the PACE 203H™ operates as
intended in an electromagnetic environment. Guidance here given must be followed for
putting PACE 203H™ into service.
The user must be aware that the use of leads and patient cables other than those
specified in the instruction for use may result in increased electromagnetic emissions or
decreased electromagnetic immunity.
The PACE 203H™ should not be used adjacent to or stacked with other equipment. If
adjacent or stacked use is necessary, the PACE 203H™ should be observed to verify
normal operation in the configuration in which it will be used.
Electromagnetic Emission
Table 62 states the compliance classification for electromagnetic emissions and describes
the general electromagnetic environment that the PACE 203H™ is intended to operate in.
Guidance and manufacturer’s declaration – electromagnetic emissions
The PACE 203H™ is intended in the electromagnetic environment specified below.
The customer and the user of the PACE 203H™ should assure that it will be used in
such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions Group 1 The PACE 203H™ uses RF energy only for its
CISPR 11 internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference
in nearby electronic equipment.
RF emissions Class B The PACE 203H™ is suitable for use in all
CISPR 11 establishments, including domestic establishments
and those directly connected to the public low-
Harmonic Not voltage power supply network that supplies
emissions applicable buildings used for domestic purposes.
IEC 61000-3-2
Voltage Not
fluctuations/ applicable
Flicker
emissions
IEC 61000-3-3
Table 62: Guidance and manufacturer’s declaration – electromagnetic emission

92/104
Electromagnetic Immunity
Table 63 states the immunity test and compliance level for electrostatic discharge and
power frequency magnetic fields and describes the electromagnetic environment that the
PACE 203H is intended to operate in.
Guidance and manufacturer’s declaration – electromagnetic immunity
The PACE 203H™ is intended in the electromagnetic environment specified below. The customer
and the user of the PACE 203H™ should assure that it will be used in such an environment.
Immunity test IEC 60601 Compliance Electromagnetic environment -
test level level guidance
Electrostatic ± 6 kV contact ± 6 kV contact Floors should be wood, concrete or
discharge ceramic tile. If floors are covered with
(ESD) ± 8, ± 15 kV air ± 8, ± 15 kV air synthetic materials, the relative
IEC 61000-4-2 humidity should be at least 30 %.
Electrical fast ± 2 kV for power ± 2 kV for power Mains power quality should be that of
transient/burst supply lines supply lines typical commercial or hospital
IEC 61000-4-4 ± 1 kV for Not applicable for environment.
input/output lines input/output lines

Surge ± 1 kV ± 1 kV different. Mains power quality should be that of


IEC 61000-4-5 differential mode mode typical commercial or hospital
± 2 kV common ± 2 kV common environment.
mode mode
Voltage dips, <5 % UT Not applicable No applicable
short (>95 % dip in UT) (PACE 203H has an uninterruptible
interruptions for 0.5 cycle power system)
and voltage 40 % UT
variations on (60 % dip in UT)
power supply for 5 cycles
lines 70 % UT
IEC 61000-4-11 (30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
Power 3 A/m 3A/m Power frequency magnetic fields
frequency should be at levels characteristic of a
magnetic field typical commercial or hospital
environment.
NOTE UT is the a. c. mains voltage prior to application of the test level

Table 63: Guidance and manufacturer’s declaration – electromagnetic immunity I

93/104
The following Table 64 states the immunity test and compliance level for radiated RF
electromagnetic fields. Guidance is given for the electromagnetic environment regarding
those fields.
Guidance and manufacturer’s declaration – electromagnetic immunity
The PACE 203H™ is intended in the electromagnetic environment specified below. The
customer and the user of the PACE 203H™ should assure that it will be used in such an
environment.
Immunity test IEC 60601 Compliance Electromagnetic
test level level environment - guidance
Portable and mobile RF
communications equipment
should be used no closer to
any part of the PACE 203H,
including cables and leads,
than the recommended
separation distance
calculated from the equation
applicable to the frequency of
Conducted RF 10 Vrms 10 Vrms the transmitter.
IEC 61000-4-6 150 KHz to 80 MHz 150 KHz to 80 MHz
outside ISM bandsa outside ISM bandsa Recommended separation
distance
10 Vrms 10 Vrms
150 KHz to 80 MHz 150 KHz to 80 MHz
in ISM bandsa in ISM bandsa  3 .5 
d=  P
Radiated RF 10 V/m 10 V/m  V1 
IEC 61000-4-3 80 MHz to 2.5 GHz

 12 
d=  P
V 2 

d = [1.2] P 80 MHz to
800 MHz

d = [2.3] P 800 MHz


to 2.5 GHz

where P is the maximum


output power rating of the

94/104
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in metres (m).b

Field strength from fixed RF


transmitters, as determined
by an electromagnetic site
survey,c should be less than
the compliance level in each
frequency range.d

Interference may occur in the


vicinity of equipment marked
with the following symbol:

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a The ISM (industrial, scientific and medical) bands between 150 KHz and 80 MHz are
6.765 MHz to 6,798 MHz;
13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
b The compliance levels in the ISM frequency bands between 150 KHz and 80 MHz and in
the frequency range 80 MHz to 2.5 GHz are intended to decrease the likelihood that
mobile/portable communications equipment could cause interference if it is inadvertently
brought into patient areas. For this reason, an additional factor of 10/3 is used in
calculating the recommended separation distance for transmitters in these frequency
range
c Field strength from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the PACE 203H is used
exceeds the applicable RF compliance level above, the PACE 203H should be observed
to verify normal operation. If abnormal performance is observed, additional measures may
be necessary, such as reorienting or relocating the PACE 203H.
d Over the frequency range 150 KHz to 80 MHz, field strengths should be 10 V/m.

Table 64: Guidance and manufacturer’s declaration – electromagnetic immunity II

95/104
Table 65 lists minimum separation distances between PACE 203H™ and portable / mobile
RF communication equipment that should be maintained to avoid interference.
Recommended separation distances between
portable and mobile RF communication equipment
The equipment is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or user of the PACE 203H can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the PACE 203H as recommended below, according
to the maximum output power of the communications equipment.
Rated Separation distance according to frequency of transmitter
maximum m
output
power of 150 KHz to 80 MHz 150 KHz to 80 MHz 80 MHz to 800 800 MHz to 2.5
transmitter outside ISM bands in ISM bands MHz GHz
W  
3 . 5  
12
d=  P d=  P d = [1.2] P d = [2.3] P
 V1  V 2 
0.01 0.04 0.12 0.12 0.23
0.1 0.11 0.38 0.38 0.73
1 0.35 1.2 1.2 2.3
10 1.1 3.8 3.8 7.3
100 3.5 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation
distances d in metres (m) can be determined using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 The ISM (industrial, scientific and medical) bands between 150 KHz and 80 MHz are
6.765 MHz to 6,798 MHz;
13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
NOTE 3 An additional factor of 10/3 is used in calculating the recommended separation distance
for transmitters in the ISM frequency bands between 150 KHz and 80 MHz and in the frequency
range 80 MHZ to 2.5 GHz to decrease the likelihood that mobile/portable communications
equipment could cause interference if it is inadvertently brought into patient areas.
NOTE 4 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.

Table 65: Recommended separation distances between portable and mobile RF


communication equipment

96/104
97/104
19 Appendix D: Figures
Figure 1 PACE 203H™ Display Face ...............................................................................13
Figure 2 Bottom View ........................................................................................................14
Figure 3: Upper Display.....................................................................................................15
Figure 4: Lock Symbol.......................................................................................................19
Figure 5: The Auto Sense function increases sensitivity...................................................31
Figure 6: The Auto Sense function decreases sensitivity..................................................31
Figure 7: Acoustic Indicator (Symbol for a Loudspeaker) .................................................45
Figure 8: Alarms Off (Crossed Bell) Symbol .....................................................................46
Figure 9: Lead Connection Terminals ...............................................................................50
Figure 10: Battery Symbol.................................................................................................56
Figure 11: Open Lead Symbol ..........................................................................................57
Figure 12: Symbol for Interference....................................................................................59
Figure 13: Symbol for Extrasystole (Premature Ventricular Contraction, PVC) ................60
Figure 14: Battery Compartment .......................................................................................78

98/104
99/104
20 Appendix E: Tables
Table 1: Dials and their functions ......................................................................................15
Table 2: Fixed labeled keys and their functions ................................................................16
Table 3: Mode Menu..........................................................................................................16
Table 4: Main Menu...........................................................................................................17
Table 5: Press Lock/Unlock Key Request .........................................................................18
Table 6: Release Lock/Unlock Key Request .....................................................................18
Table 7: Turn-off Menu ......................................................................................................20
Table 8: Mode Menu..........................................................................................................24
Table 9: Supplementary Mode Setting ..............................................................................24
Table 10: Adjustable values for rate..................................................................................26
Table 11: Message related to increase the Rate from 150 ppm .......................................26
Table 12: Adjustable values for stimulation amplitude ......................................................27
Table 13: Adjustable values for sensitivity ........................................................................27
Table 14: Adjustable values for A-V delay ........................................................................28
Table 15: Auto Menu .........................................................................................................28
Table 16: Automatic setting formula for A-V delay ............................................................29
Table 17: Automatic setting formula for PVARP ...............................................................29
Table 18: Automatic setting formula for MTR ....................................................................30
Table 19: Auto Sense Menu..............................................................................................32
Table 20: Auto Sense Menu (channels selected, initialization request) ............................32
Table 21: Auto Sense Menu (Searching) ..........................................................................32
Table 22: Auto Sense Menu (intrinsic rhythm found) ........................................................33
Table 23: Auto Sense Menu (finished) ..............................................................................33
Table 24: Auto Sense Menu (no R-wave found) ..............................................................34
Table 25: Auto Sense Menu (one channel initialization request) ......................................34
Table 26: High Rate Standby Menu ..................................................................................36

100/104
Table 27: Adjustable values for atrial high rate .................................................................36
Table 28: High Rate Running Menu ..................................................................................37
Table 29: Standard Menu..................................................................................................38
Table 30: Standard Programs Set.....................................................................................38
Table 31: Standard Saving Confirmation ..........................................................................39
Table 32: Manufacturer’s Default Programs......................................................................40
Table 33: Turn-on Saving Confirmation ............................................................................41
Table 34: Manufacturer’s Turn-on Program ......................................................................41
Table 35: Parameters/Options- Menu 1 ............................................................................42
Table 36: PVARP & MTR Menu ........................................................................................42
Table 37: Manually Adjustable Values for PVARP ............................................................43
Table 38: PVARP & MTR Menu ........................................................................................43
Table 39: Manually Adjustable Values for MTR ................................................................43
Table 40: Pulse Duration Menu.........................................................................................44
Table 41: Manually Adjustable Values for Pulse Duration ................................................44
Table 42: ARP (AAI, AAT) Menu .......................................................................................44
Table 43: Manually Adjustable Values for ARP (AAI, AAT)...............................................44
Table 44: Options Menu 1 .................................................................................................45
Table 45: Options Menu 2 .................................................................................................45
Table 46: Language Menu ................................................................................................47
Table 47: Emergency Stimulation .....................................................................................48
Table 48: Pause Menu ......................................................................................................49
Table 49: Options Menu 2 (BPI marked) ...........................................................................54
Table 50: Options Menu 2 (PMC marked).........................................................................55
Table 51: Mode Switching Due to Interference .................................................................59
Table 52: Abnormal Settings .............................................................................................62
Table 53: Conflicts Type a) ...............................................................................................64
Table 54: Conflicts Type b) ...............................................................................................64
Table 55: Conflicts Type c)................................................................................................64

101/104
Table 56: Conflicts Type d) ...............................................................................................65
Table 57: Conflicts Type e) ...............................................................................................65
Table 58: Warning Messages Related to Conflicting Parameter Settings ........................70
Table 59: Warning Messages Related to Application Errors.............................................72
Table 60: Warning Messages Related to Device Errors ...................................................73
Table 61: Informative Messages .......................................................................................76
Table 62: Guidance and manufacturer’s declaration – electromagnetic emission............92
Table 63: Guidance and manufacturer’s declaration – electromagnetic immunity I..........93
Table 64: Guidance and manufacturer’s declaration – electromagnetic immunity II.........95
Table 65: Recommended separation distances between portable and mobile RF
communication equipment.......................................................................................96

102/104
103/104
21 Appendix F: Declaration of Conformity

We,

Osypka Medical GmbH


Albert-Einstein-Strasse 3
12489 Berlin

declare under our sole responsibility, that the medical device

TM
Dual-chamber External Pacemaker PACE 203H
including Power Supply PACE and accessories

is in conformity with the act on Medical Devices, the Medical Device Directive 93/42/EEC
and the applicable standards and regulation.
This Declaration of Conformity covers all in our company manufactured loads of the above
mentioned device which are labeled with the CE mark.
This Declaration of Conformity is based on the Full Quality Assurance System certification
with the registration number G1 08 12 39212 011 issued by the Notified Body No. 0123,
the TÜV Product Service GmbH in Munich, Germany.
Berlin, January 02, 2009

Dr.-Ing. Bernd Tröger Dipl.-Ing. (FH) Thilo Thümecke


Plant Manager Quality Manager

104/104

You might also like