LOGIQt 400
Basic Users Manual Addendum

CopyrightE 1999 By General Electric Co.

Operating Documentation
Regulatory Requirement

This product complies with regulatory requirements of the following European


Directive 93/42/EEC concerning medical devices

This Basic Users Manual Addendum is a reference for the LOGIQ 400 MD
MR3 and the LOGIQ 400CL. It applies to all versions of 4.31 software for the
LOGIQ 400. and 4.32 software for the LOGIQ 400CL.

The “Basic Users Manual Addendum” provide the supplemental information for
t
the LOGIQ 400 Basic Users Manual (2233660–1XX)

GE Medical Systems

GE Medical Systems: Telex 3797371


P.O. Box 414, Milwaukee, Wisconsin 53201 U.S.A.
(Asia, Pacific, Latin America, North America)

GE Medical Systems Ċ Europe: Tel: +49 (0) 212 28 02 208


Beethovenstraße 239, Postfach 11 05 60, D-42655 Solingen
GERMANY
Revision History

Revision History
REV DATE REASON FOR CHANGE

0 July 29, 1999 Initial Release


1 Aug 18, 1999 Correction
2 Oct 18, 1999 Correction
3 Dec 20, 1999 Addition of LD probe information

LIST OF EFFECTIVE PAGES


PAGE REVISION
NUMBER NUMBER

Title Page 3
Revision History A and B 3
Table of Contents 1 and 2 3
1 thru 18 3

Please verify that you are using the latest revision of this document. Information
pertaining to this document is maintained on GPC (GE Medical Systems Global
Product Configuration). If you need to know the latest revision, contact your
distributor, local GE Sales Representative or in the USA call the GE Ultrasound
Clinical Answer Center at 1-800-682-5327 or 262-524-5698.

t
LOGIQ 400 Basic Users Manual Addendum
2248601–100 Rev. 3 Revision History A
Revision History

This page left blank intentionally.

Revision History B t
LOGIQ 400 Basic Users Manual Addendum
2248601–100 Rev. 3
Table of Contents

Table of Contents

LOGIQ 400/LOGIQ 400CL Functionality . . . . . . . . . . . . . . . . . . . . . . . . . 1


Text References . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
ATO Map Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
ATO Create . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
ATO On/Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
THI (Tissue Harmonic Imaging) Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Enhance DICOM Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Enhance Image Archive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Search criteria (MO, HD, Queue) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
File discriminators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Send an image to DICOM device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Copy an image . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Type of an image . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Status of DICOM Queued Images . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Storage on Hard Disk . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
LD Probe Availability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
CFM/PDI Map Transparency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Spatial Filter Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Higher Frame Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
EFW Using User Programmable Table Measurements . . . . . . . . . . . . . . . . 12
Ovary Follicles Measurements and Report in Gynecology . . . . . . . . . . . . . 13
Ovary Follicles Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Ovary Follicles Report Page . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Language Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Set Up Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Custom Display Page 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Custom Display Page 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Custom Display Page 6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Custom Display Page 7 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Custom Display Page 8 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
System Parameters Page 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
System Parameters Page 5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Acoustic Output Tables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

t
LOGIQ 400 Basic Users Manual Addendum
Table of Contents 1
2248601–100 Rev. 3
Table of Contents

This page left blank intentionally.

Table of Contents 2 t
LOGIQ 400 Basic Users Manual Addendum
2248601–100 Rev. 3
Basic Users Manual Addendum

Basic Users Manual Addendum

LOGIQ 400/LOGIQ 400CL Functionality


The added features to this version of the LOGIQ 400 are
listed below.

Item LOGIQ 400 LOGIQ 400CL


Color System Black/White System
ATO Map Option Available Available Available
THI (Tissue Harmonic Available Available Available with C358
Imaging) Option Probe Only
Enhance DICOM Option Available Available Available
Enhance Image Archive Available Available Available
Follicles Report Page Available Available Available

.
Available: Available as Option or Standard
Not Available: Not available as Option or Standard

NOTE: ATO Map Option is only available in Europe.

Text References

LOGIQ 400CL The LOGIQ 400CL system, available only in Europe and Asia,
does not support every feature of the LOGIQ 400. When a
feature does not apply to the LOGIQ 400CL, the symbol on
CL the left appears.

LOGIQ 400 B/W The LOGIQ 400 Black/White system does not come standard
with every feature. When a feature is available as an option,
B/W the symbol on the left appears.
Option

t
LOGIQ 400 Basic Users Manual Addendum
2248601–100 Rev. 3 1
Basic Users Manual Addendum

ATO Map Option

ATO Create

Description—ATO, Automatic Tissue Optimization, optimizes


the image based upon a specified Region Of Interest (ROI) or
anatomy within the display. ATO gray scale map is available
only in B-Mode.

Benefits—Improves productivity and patient throughput with


one touch image optimization. Improves the contrast of the
display image data.

Accessing/Changing—To create an ATO gray scale map:

1. Press the ATO Create Sub-Menu up/down arrow key


on B-Mode Sub-Menu page 1. The ROI box cursor is
displayed on the image.

B 1/4

Dynamic Gray Focus ATO ATO


Range Map Number Create On/Off

42 B–2 2

Figure 1. B-Mode Sub-Menu Page 1

2. Use the trackball for changing the size and location of


the ROI box cursor. Press Scan Area to toggle
between sizing and moving the cursor.
3. Press Set. The system automatically optimizes the
entire image without pressing ATO On/Off; the
highlighted sub-menu selection automatically changes
to ATO On/Off.

2 t
LOGIQ 400 Basic Users Manual Addendum
2248601–100 Rev. 3
Basic Users Manual Addendum

ATO On/Off
Description—The ATO On/Off selection is highlighted when
the ATO map is active on the image. The ATO Display can be
toggled on and off to display that same ATO map again.

Benefits—Ability to access the previously created ATO map


without any time delay.

When the ATO gray scale map is active on the image, an


asterisk (*) is displayed on the right side of the gray scale map
number.

ATO gray scale map is available until changing the following:

.
Probe, New Patient, Exam Category, User Preset, Gray Scale
Map, Power Off, or new ATO map creation. Only one ATO gray
map is available at a specific time.

NOTE: ATO Map Option is only available in Europe.

t
LOGIQ 400 Basic Users Manual Addendum
2248601–100 Rev. 3 3
Basic Users Manual Addendum

THI (Tissue Harmonic Imaging) Option


Description—Tissue Harmonic Imaging utilizes the natural
harmonic characteristics inherent in non-linear sound
propagation properties in the human body. This function does
not permit the use of any contrast agent. Currently, it is only
available on the S222, S317 and C358 probes.

S222 and S317 probes


CL

Benefits—Harmonic imaging diminishes low frequency high


amplitude noise and often improves imaging technically difficult
patients. Tissue Harmonic Imaging is beneficial to contrast
resolution and cystic clearing. Harmonic 1 is a higher harmonic
frequency, whereas Harmonic 2 is lower.

Accessing/Changing—To activate Tissue Harmonic Imaging:

1. Connect the S222, S317 or C358 probe.


2. Press the Imaging Freq Sub-Menu up/down arrow key
on B-Mode Sub-Menu page 2.

B 2/4

Frame Imaging Image Color 3D


Average Freq Softner Mode

OFF 3.5MHz

Figure 2. B-Mode Sub-Menu Page 2

3. Press the Imaging Freq Sub-Menu up/down arrow key


until the appropriate frequency, indicated by the letter
‘T’ for tissue, appears.
Refer to the Frequency of Operation Table on the next
page.

4 t
LOGIQ 400 Basic Users Manual Addendum
2248601–100 Rev. 3
Basic Users Manual Addendum

THI (Tissue Harmonic Imaging) Option (cont’d)

Frequency of Operation depends on Image Frequency Sub-Menu Selection

Further
Probe Resolution Normal Penetration Penetration Harmonic 1 Harmonic 2
B510 6.5MHz 5MHz 5MHz N/A N/A N/A
S317 4MHz 3.5MHz 2.5MHz N/A T5.0 T4.7
S222 3.5MHz 3MHz 2MHz N/A T3.5 T3.2
S611 6MHz 5MHz 4MHz N/A N/A N/A
S220 (W) 3MHz 2MHz 2MHz N/A N/A N/A
E721(MTZ) 7MHz 6MHz 5.5MHz N/A N/A N/A
C721 7MHz 6MHz 6MHz N/A N/A N/A
C551 (CAE) 6MHz 5MHz 4.4MHz N/A N/A N/A
C364 (CBF) 4MHz 3.5MHz 3MHz N/A N/A N/A
C386 5MHz 3.8MHz 3MHz N/A N/A N/A
C358 5MHz 4MHz 3MHz 2MHz T5.0 T4.5
739L 9MHz 7MHz 6MHz N/A N/A N/A
LA39 11MHz 9.6MHz 8.2MHz N/A N/A N/A
546L 6.6MHz 5.2MHz 4.6MHz N/A N/A N/A
LD 4MHz 3.5MHz 3MHz N/A N/A N/A
L764 (LH) 7MHz 7MHz 7MHz N/A N/A N/A
I739 8MHz 7MHz 6MHz N/A N/A N/A
T739 8MHz 7MHz 6MHz N/A N/A N/A
CWD2 N/A N/A N/A N/A N/A N/A
CWD5 N/A N/A N/A N/A N/A N/A
ERB7C 7MHz 6MHz 5.5MHz N/A N/A N/A
ERB7L 7MHz 6MHz 5.5MHz N/A N/A N/A

Table 1. Frequency of Operation

ERB7 Probe is not available in the USA.

Sector Probes
CL

t
LOGIQ 400 Basic Users Manual Addendum
2248601–100 Rev. 3 5
Basic Users Manual Addendum

Enhance DICOM Option


In the DICOM option, after setting the DICOM transfer key
sequence to the User Define key, the Patient Entry and Report
screens can be sent to the DICOM device by pressing the
defined key.

The maximum number of temporary images that can be stored


on the HDD for DICOM transfer is 600.

Enhance Image Archive


From the Image Archive Patient Search Menu, the following
functions are available:

S Show the storage place (MO, HD or Queue)


S Show image(s) in the DICOM queue
S

.
Copy image(s) to MO to/from HD from/to MO
S Show the type of an image

NOTE: DICOM related keys are not displayed if the DICOM


Storage Option is not installed.

Figure 3. Image Archive Patient Search Menu

6 t
LOGIQ 400 Basic Users Manual Addendum
2248601–100 Rev. 3
Basic Users Manual Addendum

Enhance Image Archive (cont’d)


The table below summarizes the functions on the Image
Archive Patient Search Menu as applied to different image
types.

Image Image Image


Search Tag/ Image Lock/ Image Image DICOM
Image Type (list up) Untag Recall Unlock Delete Copy send
on MO Ǹ Ǹ Ǹ Ǹ Ǹ Ǹ Ǹ
on HD Ǹ Ǹ Ǹ Ǹ Ǹ Ǹ Ǹ
in Queue Ǹ Ǹ Ǹ NA NA Ǹ Ǹ
LZW compressed Ǹ Ǹ Ǹ Ǹ Ǹ Ǹ NA

NA: Not Available

Table 2. Search Menu Functions on Different Image Types

Search criteria (MO, HD, Queue)


A search list can be displayed from selected file discriminators.
To select the search type, move the cursor over MO, HD and/or
QUE in the “From MO HD QUE” portion of the Image Archive
Patient Search Menu. Press the Set key. An asterisk mark (*)
indicates the selection. For example, when only MO is
selected, the screen display changes from “From MO HD QUE”
to “From *MO HD QUE”. The search list will show MO archived
images.

File discriminators

There are discriminators displayed in front of the file name that


indicate whether they are archived. The following file
discriminators are:

SYMBOL DESCRIPTION

None Indicates a file on Hard Disk.


+ Indicates a file on MO Disk.
/ Indicates a file on Hard Disk as DICOM queue.

t
LOGIQ 400 Basic Users Manual Addendum
2248601–100 Rev. 3 7
Basic Users Manual Addendum

Send an image to DICOM device


If the current selected device is a DICOM printer and the
DICOM printer is configured as Multi format print in the
Destination column, only images with the same patient name
and patient ID can be sent.

The following message is displayed when attempting to send


images with different patient name and/or patient ID to the print
queue:

Different PT Name/ID Print is rejected.

Only B&W or color images can be sent to a DICOM printer. A


combination of the two is not allowed.

The following message is displayed when both B&W and color


images are sent to a B&W or color DICOM printer.

Rejected. Select B&W or color images only.

To send an image to a DICOM device from an Archive Patient


Search Menu.

1. Move the cursor to the image which will be sent.


2. Press the Set key to select the image. If needed,
select multiple images.
3. Move the cursor to the right of the Destination column.
4. Press the Ellipse up/down arrow key to select the
desired DICOM device.

.
5. Move the cursor to the Send column.
6. Press the Set key.

NOTE: LZW compressed images cannot be sent to a DICOM


device. If there is an attempt to send a compressed image, the
message “Compressed image cannot be sent.” is displayed.

8 t
LOGIQ 400 Basic Users Manual Addendum
2248601–100 Rev. 3
Basic Users Manual Addendum

Copy an image
To copy an image from the HD (MO) to MO (HD):

1. Select the image to copy by pressing the Set key.


Multiple images can be selected.
2. Move the cursor to the Image Copy column.
3. Press the Set key. The image is copied to the MO (HD)
from HD (MO).
4. Move the cursor to List All and press the Set key to
reflect the copied image in the Image Archive Patient
Search Menu.

The following message is displayed if the capacity of the HD or


MO is not enough for an image to be copied.

Too much data. Copy Failed.

Type of an image

A symbol appears in front of the date to indicate whether the


image is color or gray scale. The symbols are as follows:

SYMBOL DESCRIPTION

G Gray scale image.


C Color image.
g LZW compressed gray scale image.
c LZW compressed color image.

Status of DICOM Queued Images


A list of DICOM Queue images can be displayed on the Image
Archive Patient Search Menu, when its status has been
changed to FAIL or COMP in the Queue Status Menu in the
DICOM option. DICOM queued images cannot be locked,
unlocked or deleted at the Image Archive Patient Search Menu.

t
LOGIQ 400 Basic Users Manual Addendum
2248601–100 Rev. 3 9
Basic Users Manual Addendum

Storage on Hard Disk


An image can be stored to a Hard Disk Drive. The capacity of
the HDD storage is up to 600 images.

To store the image to the Hard Disk Drive, press the HD


Storage up/down arrow key.

Archive 1/1

Patient Media MO HD DEFF MO


Search Search Archive Storage Format Eject

Figure 4. Archive Sub-Menu Page 1

The following message is displayed during the storage process:

In progress. Please wait. HD Free : xxx


xxx is the number of images available for future archival.

.
LD Probe Availability
LD probe is available for CFM and Doppler Modes.

NOTE: LD probe is available only in Europe.

CAUTION A portion of the low sensitivity scan image for the LD probe lies
along the hollow portion of its structure used for the biopsy
guide sleeve. Keep this fact in mind when performing all
biopsies with the LD probe.

10 t
LOGIQ 400 Basic Users Manual Addendum
2248601–100 Rev. 3
Basic Users Manual Addendum

CFM/PDI Map Transparency


The CFM Map Transparency and PDI Map Transparency
B/W presets, found on Set Up/Custom Display page 7, allow for the
Option default maps to be set. The higher the number, the higher the
transparency level.

Spatial Filter Setting


The Spatial Filter has the six settings (Off, 1 to 5). These filters
B/W allow for a smoother color flow image. This selection is
Option available in CFM Mode Sub-Menu page 3.

The CFM/PDI Spatial Filter preset is found on Set/Up Custom


Display page 6.

Higher Frame Rate


Frame rates have been increased by the implementation of the
B/W very-small packet size. This selection is available in the CFM
Option Mode Sub-Menu page 3.

The presets for CFM Packet Size and PDI Packet Size are
found on Custom Display pages 6 and 8, respectively.

t
LOGIQ 400 Basic Users Manual Addendum
2248601–100 Rev. 3 11
Basic Users Manual Addendum

EFW Using User Programmable Table Measurements


If it is desired to use User Programmable Table measurements
in the calculation and reporting of EFW, perform the following
steps:

System Setup

1. Create the desired User Programmable Tables as


described on 14–75 of the Basic User Manual.
2. Assign the User Table measurement to the proper
Measurement Mode Menu as described on 14–84 of
the Basic User Manual.
3. Assign the User Programmable Table measurement to
the desired Report Page as described on 14–72 of the
Basic User Manual.

Calculating EFW

4. Perform the measurements used in the calculation of


EFW.
For example, if Tokyo method is selected in the OB
Report Format, found in System Parameters page 2,
EFW is calculated by measuring BPD, FL, APTDxTTD.
If user programmable tables are used in the
measurement of BPD, FL and/or APTDxTTD, these
measurements should be performed first prior to the
factory default measurements [i.e. FL (user1), BPD, FL,
APTDxTTD].

After all measurements have been completed, EFW is not


displayed in the measurements results area, but is displayed on
the Report Page.

If it is desired to use User Programmable Table measurements


Hints in the calculation of EFW, it would be helpful to program a User
Measurement Sequence as described on 14–79 of the Basic
User Manual.

If measurement averaging is active, EFW will be calculated


from the average of similar factory default and user
programmable table measurements.

12 t
LOGIQ 400 Basic Users Manual Addendum
2248601–100 Rev. 3
Basic Users Manual Addendum

Ovary Follicles Measurements and Report in Gynecology


New measurements and a Report Page for Ovary Follicles are
available.

Ovary Follicles Measurement

Ovary Follicle (LtOvFo or RtOvFo) Measurements are available.

3/4

LtOvFo RtOvFo Ov Fol


Report

Figure 5. GYN Calculation Sub-Menu (Follicles)

LtOvFo or RtOvFo measurements are calculated from one, two,


or three distances using the following formula:

Calc Mnemonic Calc Name Input Measurements Formula


LtOvFo[ml] Left Ovary Follicles One distance* D1[cm]3 x p/6
Two distances* D1[cm]2 x D2x[cm] p/6
: (D1 <D2)
D1[cm] x D2[cm]2 x p/6
: (D2 <D1)
Three distances D1[cm] x D2[cm]x
D3[cm] x p/6
RtOvFo[ml] Rith Ovary Follicles One distance* D1[cm]3 x p/6
Two distances* D1[cm]2 x D2[cm] x p/6
: (D1 <D2)
D1[cm] x D2[cm]2 x p/6
: (D2 <D1)
Three distances D1[cm] x D2[cm] x
D3[cm] x p/6

* To calculate LtOvFo or RtOvFo with one (or two distances), press the Clear key
after the first (or second distance) measurement(s).

t
LOGIQ 400 Basic Users Manual Addendum
2248601–100 Rev. 3 13
Basic Users Manual Addendum

Ovary Follicles Report Page


The Ovary Follicles Report Page is provided to improve
productivity and facilitate consistency in Gynecology Ovary
Follicles procedures. It is available by selecting Ov Fol Report
from the measurement Sub-Menu.

Figure 6. Follicles Report

D1 ~ 3—3 fields where measured distances are reported.

MEAN—Average of one, two, or three distance measurements.

VOLUME—Calculated from one, two or three distances.

FOLLICULAR OBSERVATIONS:
S Volume Changes—Input the description of volume
changes.
S Border Appearance—Input the description of border
appearance.
S Contents—Input the description of the contents.

ENDOMETRIUM THICKNESS—Displayed when the Endo


measurement is performed.

COMMENTS—Input general comments and findings.

14 t
LOGIQ 400 Basic Users Manual Addendum
2248601–100 Rev. 3
Basic Users Manual Addendum

Language Support
The system supports more languages for the displayed
graphics. The “Language” preset is found on System
Parameters page 1. The selections available are:

English, French, German, Spanish, Italian, Portuguese,


Swedish, Danish, Russian, Greek and Turkish.

To change the language selection in System Parameters


page 1:

1. Move the cursor over the language preset and press


the Set key.
2. Use the Ellipse up/down arrow key to display the
desired language.
3. Move the cursor to SAVE and press the Set key.

The language change will take effect after the system is turned
off and powered up again.

Set Up Parameters

Custom Display Page 2

B Image Frequency EC, UP, P, R


Choose the B-Mode image frequency optimization setting.
Resolution Process image for best resolution.
Normal Compromise between resolution and penetration.
Penetration Process image for penetration.
Further Penetration C358 Probe Only—Process image for more penetration
Harmonic1 S222, S317 and C358 Probes Only—Process image with
tissue harmonic frequency.
Harmonic2 S222, S317 and C358 Probes Only—Process image with
tissue harmonic frequency.
The soft menu display shows this parameter as an imaging frequency depending on the
active probe.

t
LOGIQ 400 Basic Users Manual Addendum
2248601–100 Rev. 3 15
Basic Users Manual Addendum

Custom Display Page 3

M Image Frequency EC, UP, P, R


Choose the M-Mode image frequency optimization setting.
Resolution Process image for best resolution.
Normal Compromise between resolution and penetration.
Penetration Process image for penetration.
Further Penetration C358 Probe Only—Process image for more penetration.
Harmonic1 S222, S317 and C358 Probes Only—Process image with tissue
harmonic frequency.
Harmonic2 S222, S317 and C358 Probes Only—Process image with tissue
harmonic frequency.

This parameter is not selectable from the soft menu.

Custom Display Page 6

CFM/PDI Spatial Filter EC, UP, P


Choose the desired amount of averaging between color pixels.
Off, 1, 2, 3, 4, or 5

CFM Packet Size Survey, Map, SrvyDtl and MpDtl EC, UP, P
Choose the default packet size for each CFM diagnostic mode display. This will be for
Survey Mode, Map Mode, Survey-Detail Mode and Map-Detail Mode.
Very-Small Very Small packet size.
Small Small packet size.
Mid Medium packet size.
Large Large packet size.
NOTE: The larger the packet size, the better the color presentation but with a slower
frame rate.

Custom Display Page 7

CFM Map Transparency EC, UP, P


Choose the default CFM Transparency map. The higher the number, the more
transparent the map.
Off, 1, 2, 3, or 4

PDI Map Transparency EC, UP, P


Choose the default PDI Transparency map. The higher the number, the more
transparent the map.
Off, 1, 2, 3, or 4

16 t
LOGIQ 400 Basic Users Manual Addendum
2248601–100 Rev. 3
Basic Users Manual Addendum

Custom Display Page 8

PDI Packet Size Survey, Map, SrvyDtl and MpDtl EC, UP, P
Choose the default packet size for each PDI diagnostic mode display. This will be for
Survey Mode, Map Mode, Survey-Detail Mode and Map-Detail Mode.
Very-Small Very Small packet size.
Small Small packet size.
Mid Medium packet size.
Large Large packet size.
NOTE: The larger the packet size, the better the color presentation but with a slower
frame rate.

System Parameters Page 1

Language
Choose the desired language for all displayed graphics by using the Ellipse up/down
arrow keys.
English, French, German, Spanish, Italian, Portuguese, Swedish, Danish, Russian,
Greek or Turkish.
Power on Status
The “Keep–Latest” function is not available. The system always resets the
Exam/System Parameters at power up, even if “Keep–Latest” is selected.

System Parameters Page 5

Record 1 B&W or Color


Choose the type of device to be controlled by the Record 1 key for black and white or
color images. Use the Ellipse up/down arrow key to cycle through the selections.
Black and White Polaroid Camera, Color Polaroid Camera, Black and White Printer,
Color Printer, Multi Image Camera, MO Archive, HD Storage or Computer.
Record 2 B&W or Color
Choose the type of device to be controlled by the Record 2 key for black and white or
color images. Use the Ellipse up/down arrow key to cycle through the selections.
Black and White Polaroid Camera, Color Polaroid Camera, Black and White Printer,
Color Printer, Multi Image Camera, MO Archive, HD Storage or Computer.

t
LOGIQ 400 Basic Users Manual Addendum
2248601–100 Rev. 3 17
Basic Users Manual Addendum

Acoustic Output Tables


S317, C358 and S222 probe’s Acoustic Output Data has been
updated.

Refer to the Advanced Reference Manual (2248599–100) for


this information.

18 t