Software Addendum
e360 Ventilator
Software 7.0
Addendum
OPR360U A0509 5-1
e360 Ventilator
Software 7.0
Addendum
Introduction............................................................ A-1
Respiratory Mechanics......................................... A-1
P0.1.................................................................... A-1
Negative Inspiratory Force (NIF)...................... A-4
Changes to Weaning Data Set......................... A-6
Capability to Disable O2 Sensor........................... A-7
Update of Inspiratory Tidal Volume and
Expiratory Tidal Volume ....................................... A-8
FlexCycle* (Automatic Expiratory Threshold
Management)......................................................... A-9
Introduction
Units with software version 7.0 offer the following additional features:
Respiratory Mechanics:
• P0.1
• Negative Inspiratory Force
• Changes to Weaning Data Set
Capability to Disable O2 Sensor
Update Inspiratory Tidal Volume (VTI) and Expiratory Tidal Volume (VTE)
displayed values simultaneously
FlexCycle (Automatic Expiratory Threshold Management)*
Respiratory Mechanics
P0.1 Measurement
This function is enabled via the P0.1 button on the Technical Screen. It
is available in all modes and breath types. It is not available when Non
Invasive Ventilation (NIV) is activated.
1. Press the Setup & Calibration menu button on the control panel.
2. Touch TECHNICAL.
3. Touch P0.1 to enable the function. (See Figure 1) The screen
will change to the Main Screen.
4. Touch START P0.1 to initiate the measurement. (Figure 2)
The measurement result will be displayed after the first patient
trigger is detected.
This function is enabled via the NIF button on the Technical Screen.
The NIF Maneuver is available in all modes, with and without Non
Invasive Ventilation (NIV) activated.
1. Press the Setup & Calibration menu button on the control panel.
2. Touch TECHNICAL.
3. Touch NIF to enable the function. (See Figure 4) The screen will
change to the main screen with a pressure waveform displayed.
(See Figure 5)
4. Touch and hold the NIF (menu button) to perform the maneuver.
The e360 will temporarily cease all flow delivery while the button
5-1 is being touched. OPR360U A0509
5. Upon release of the NIF button, choose the ideal NIF value by
using the adjustment knob to move the cursor to the
appropriate point on the pressure waveform. The NIF value
displayed is the measured value at the selected pressure point.
6. Touch the SAVE button to store the measurement result and
the screen image. The measurement will be logged
in the Event History and the screen image will be stored in the
Screen Files List. (See Figure 6)
Note: The Numeric Screen does not display the P0.1 and NIF
measurements.
The following table lists the parameters displayed in each data set. This
table replaces the version on page 4-17 in the operating manual.
5-1 1. Press the Setup & Calibration menu button on the control panel.
OPR360U A0509
2. Touch the Sensors button, then touch the O2 Sensor button.
(See Figure 7)
3. Touch Disable to disable the O2 sensor, then touch Yes to
confirm.
4. Touch Exit.
1. Press the Setup & Calibration menu button on the control panel.
2. Touch the Sensors button, then touch the O2 Sensor button.
(See Figure 7)
3. Touch Enable to enable the O2 sensor.
4. Touch Exit.
5-1The displays for both VTI and VTE are updated simultaneouslyOPR360U
at the A0509
end of the expiratory period of each breath.