You are on page 1of 5

Automated Method for Subtraction of Fluorescence from

Biological Raman Spectra

CHAD A. LIEBER and ANITA M AHADEVAN-JANSEN*


Department of Biomedical Engineering, Vanderbilt University, Station B, Box 351631, Nashville, Tennessee 37235

One of the challenges of using Raman spectroscopy for biological terference in Raman spectra, but these methods require
applications is the inherent  uorescence generated by many biolog- modiŽ cations to the spectroscopic system. 1–3 M athemati-
ical m olecules that underlies the measured spectra. This  uores- cal methods implemented in software require no such
cence can sometimes be several orders of magnitude more intense
system modiŽ cations and thus have become the norm for
than the weak Raman scatter, and its presence must be minimized
in order to reso lve and analyze the Raman spectru m. Several tech-
 uorescence reduction. These m ethods include Ž rst- and
niques involving hardware and software have been devised for this second-order differentiation,4,5 frequency-domain Ž lter-
purpose; these include the use of wavelength shifting, time gating, ing, 2 wavelet transformation,6,7 and polynomial Ž tting.8–10
frequency-dom ain Ž ltering, Ž rst- and second-order derivatives, and Though each of these methods has been shown to be
simple curve Ž tting of the broadband variation with a high-order useful in certain situations, they are not without limita-
polynomial. Of these, polynomial Ž tting has been found to be a tions. Differentiation is an unbiased and efŽ cient m ethod
simple but effective method. However, this technique typically re- for  uorescence subtraction, yet this method severely dis-
quires user intervention and thus is time consuming and prone to torts Raman line shapes and relies on complex mathe-
variability. An automated method for  uorescence subtraction,
matical Ž tting algorithms to reproduce a traditional spec-
based on a modiŽ cation to least-squares polynomial curve Ž tting, is
described. Results indicate that the presented automated method is
tral form. 2 Frequency-based techniques can under- or
proŽ cient in  uorescence subtraction, repeatability, and in reten tion over-Ž lter, or cause artifacts to be generated in the pro-
of Raman spectral lineshapes. cessed spectra if the frequency elements of the Raman
Index Headings: Raman spectroscopy; Fluorescence rejectio n; Back- and noise features are not well separated. 2 Wavelet trans-
ground removal; Tissue diagnosis. formation is highly dependent on the decomposition
method used and the shape of the  uorescence back-
ground.6 Polynomial Ž tting is often used for its preser-
INT RODUCTIO N vation of traditional Raman line shapes, yet m ost pub-
lished records rely on sample-dependent user intervention
Recent years have seen an explosion in the use of Ra- for assignment of ‘‘non-Raman’’ locations on which to
man spectroscopy for biological purposes such as tissue Ž t the curve.
diagnosis, blood analyte detection, and cellular exami-
nation. The greatest beneŽ t of this technique is its high
POLYNOM IAL CURVE-FITTING
sensitivity to subtle molecular (biochemical) changes, as
well as its capability for nonintrusive application. How- Polynomial curve-Ž tting has a distinct advantage over
ever, biological applications of Raman spectroscopy in- other  uorescence reduction techniques in its ability to
volve turbid, chemically complex, and widely varying retain the spectral contours and intensities of the input
target sites. Thus, the challenge in using Raman spec- Raman spectra. However, simply Ž tting a polynomial
troscopy for biological purposes is not only the acquisi- curve to the raw Raman spectrum in a least-squares man-
tion of viable Raman signatures but also the suppression ner does not efŽ ciently reproduce the  uorescence back-
of inherent noise sources present in the target media. Per- ground, as the Ž t will be based on m inimizing the dif-
haps the greatest contributor of noise to biological Raman ferences between the Ž t and the measured spectrum,
spectra is the intrinsic  uorescence of many organic mol- which includes both the  uorescence background and the
ecules in biological materials. This  uorescence is often
Raman peaks. Thus, subsequent subtraction of this Ž t
several orders of m agnitude more intense than the weak
polynom ial results in a spectrum that varies about the
chemical transitions probed by Raman spectroscopy and,
zero baseline. It is therefore paramount that the polyno-
if left untreated, can dominate the Raman spectra and
mial Ž ts the spectral regions containing only background
make analysis of sample biochemistry impractical. There-
 uorescence while ignoring spectral regions containing
fore, in order to extract Raman signal from the raw spec-
Raman bands. The polynomial Ž tting technique tradition-
trum acquired, it is necessary to process the spectrum to
rem ove this  uorescence. ally relies on user-selected spectral locations on which to
A number of methods, both instrumentational and base the Ž t. Unfortunately, this subjective intervention
mathematical, have been proposed for  uorescence sub- has several drawbacks. It is time consuming, as the user
traction from raw Raman signals. M ethods implemented must process each spectrum individually to identify non-
in hardware such as wavelength shifting and time gating Raman-active spectral regions to be used in the Ž t. In
have been shown to effectively m inimize  uorescence in- addition, identiŽ cation of non-Raman-active frequencies
is not always trivial, as biological Raman spectra some-
Received 29 January 2003; accepted 17 June 2003.
times contain several adjacent peaks or peaks that are not
* Author to whom correspondence should be sent. E-mail: immediately obvious. The end effect is a m ethod that is
anita.mahadevan-jansen@ vanderbilt.edu. prone to variability.

0003-7028 / 03 / 5711-1363$2.00 / 0
Volume 57, Number 11, 2003 q 2003 Society for Applied Spectroscop y
APPLIED SPECTROSCOPY 1363
F IG . 2. Spectra of the blue sponge used as the  uorescence phantom
measu red using a long integration time (1 min) and moderate power
(80 m W), using a short integration time (8 ms) and low excitation power
(5 m W), and using a sliding window mean Ž lter on the short integration
spectrum. Measured spectra were normalized to their mean intensities
for visualization.

TEST AND VALIDATION


F IG . 1. Polynomial curve Ž tting as it is applied to rhodamine 6G. (A)
Raw measured Raman spectrum and a least-squared polynom ial curve M easurement System. All test spectra were recorded
Ž t of the baseline  uorescence. (B) Raw m easured Raman spectrum and using a Ž ber-probe-based Raman spectroscopic system.
two iterations of the modiŽ ed curve Ž t applied until a smoothed poly-
nomial curve along the base of the measured spectrum is obtained. In this system, a 300-mW, 785 nm GaAlAs diode laser
(Process Instruments, UT) is used to excite samples using
a 400-mm-core-diameter fused-silica optical Ž ber. The de-
livery Ž ber is contained in an optical probe (Visionex,
M ODIFIED POLYFIT M ETHOD GA) consisting of seven 300-mm fused-silica beam-
steered collection Ž bers around the central delivery Ž ber,
To address these limitations, the m odiŽ ed polyŽ t meth- with in-line Ž lters in the delivery and collection Ž bers for
od for  uorescence subtraction was developed. This rejection of signals generated in the Ž bers themselves.
method smoothes the spectrum in such a way that Raman The collection Ž bers are aligned linearly and imaged
peaks are automatically eliminated, leaving only the base- on to a 100 mm entrance slit of the detection system,
line  uorescence intact, to be subtracted from the raw consisting of an imaging spectrograph (Kaiser Optical
spectrum. Systems Inc, M I) and a deep-depletion, back-illuminated,
The basis for this method is a least-squares-based poly-
nomial curve-Ž tting function. However, to eliminate the
Raman bands from the Ž t, this function is modiŽ ed such
that all data points in the generated curve that have an
intensity value higher than their respective pixel value in
the input spectrum are automatically reassigned to the
original intensity (Fig. 1). This process (curve Ž tting and
subsequent reassignment) is repeated typically between
25 and 200 iterations (depending on factors such as the
relative amount of  uorescence to Raman), gradually
eliminating the higher-frequency Raman peaks from the
underlying baseline  uorescence. In the ideal case, the
Ž ltering process would cease when there are no longer
any data points in the Ž t curve that require reassignment
(all values equal to or less than respective smoothed spec-
trum intensities). However, due to noise factors and other
intrinsic artifacts, it is sometimes impossible to affect all
pixel values with the Ž lter. Thus, the Ž ltering process can
also be halted if there is convergence in the number of
data points affected by each iteration, as determined by
a standard root test for convergence. The processed base- F IG . 3. Mathematically generated pure Ram an spectrum created using
line spectrum is then subtracted from the raw spectrum a series of Lorentzian peaks on a null baseline in a distribution sim ilar
to yield the Raman bands on a near-null baseline. to that seen in tissue spectra; peak locations are labeled.

1364 Volume 57, Number 11, 2003


F IG . 4. Phantom spectra of varying  uorescence/Ram an ratios processed using the m odiŽ ed polyŽ t technique. (A) Smoothed sponge  uorescence
spectrum with pure Raman phantom in a ratio of 10:1 and 100:1, and (B) the corresponding extracted Raman spectra along with the pure Raman
phantom, which have R 2 5 0.994 and 0.726 compared to the pure phantom, respectively. (C ) Using a Ž fth-order polynomial Ž t of the sm oothed
sponge spectrum as the  uorescence with the pure Ram an phantom in a ratio of 1 000 000:1 and 10 000 000:1, and (D ) the corresponding extracted
Raman spectra along with the pure Raman phantom, which have R 2 5 0.997 and 0.961 compared to the pure phantom, respectively.

liquid-nitrogen-cooled charge-c oupled device (C CD ) a potential candidate was found in the blue-colored
camera (Roper ScientiŽ c Instruments, NJ). The detection sponge sometimes used in optics packaging. This sub-
system yields a spectral resolution of 7 cm 2 1 . stance yields moderately high intrinsic  uorescence in
Spectral calibration of each m easured spectrum was very short integration times, with no immediately resolv-
performed using a neon–argon lamp. Raman shift was able Raman peaks. Furthermore, the spectral shape of the
calibrated using 4-acetamidophenol and naphthalene. All  uorescence emission closely resembles that of human
spectra were binned to half the spectral resolution of the tissue. To further guarantee a Raman-free background
detection system for direct comparison. Since the modi- phantom, the sponge spectrum was measured at a low
Ž ed polyŽ t technique presented in this paper is designed excitation power (5 mW ) with an integration time of 8
only to reduce the low-frequency background  uores- ms (the minimum value for the detector and shutter used).
cence, a Ž rst-order Savitzky–Golay Ž lter was used to This creates a pseudo-time-gated spectrum in that any
minimize high-frequency noise (such as shot noise, read- Raman features present are obscured by thermal and
out noise, dark current, etc.) present in all Raman spectra. readout noise of the CCD array. Several iterations of a
After noise smoothing and subsequent binning, the spec-
simple sliding-window mean Ž lter (with window width
tra were treated with the m odiŽ ed polyŽ t method to re-
5 23 spectral resolution of the system) ser ved to both
move background  uorescence.
smooth the readout noise in the spectrum and to further
All mathematical processing was performed using in-
house code written in Matlab (The MathWorks Inc., MA). guarantee elimination of Raman contributions. Figure 2
Phantom Design. In order to test the efŽ cacy of the shows the smoothed sponge spectrum used as the  uo-
modiŽ ed polyŽ t  uorescence removal technique, a phan- rescence phantom, along with the raw pseudo-time-gated
tom system with known Raman and  uorescence spectral spectrum and a raw spectrum obtained at 80 mW and 1
contributions was needed. This posed a unique challenge min integration (both measured raw spectra norm alized
in that nearly all known  uorescent substances are also to their mean intensity), which clearly shows both the
Raman active. Literature searches and personal commu-  uorescence and Raman features.
nications yielded no likely candidates that were  uores- The Raman spectral components in the phantom were
cent but Raman inactive. In studying various materials mathematically generated as a series of Lorentzian peaks
that were available in a typical biomedical photonics lab, (Fig. 3), with a distribution similar to that seen in Raman

APPLIED SPECTROSCOPY 1365


F IG . 6. Tissue Ram an spectra of the human skin, cervix, and ovary
acquired in vitro (A) before and (B) after modiŽ ed polyŽ t processing.
Each spectrum is normalized to its mean intensity and all spectra are
offset for visualization.

yielded coefŽ cients of determination (R 2 ) of 0.994 and


F IG . 5. Comparison of conventional  uorescence subtraction tech- 0.726, respectively, with respect to the generated pure
niques with the modiŽ ed polyŽ t techniques using the sponge phantom Raman component added to the phantom spectra; this
with a 20:1  uorescence-to-Raman ratio. (A) Raw phantom spectrum; indicates that the Raman bands can be extracted from
(B) processed spectrum using Ž rst derivative (solid line) along with the
referen ce Raman component; (C ) processed spectrum using the manual signals in which  uorescence is at least one hundred
polynomial Ž t m ethod by three independent users (solid lines) along times more intense than the Raman features.
with the generated pure Raman component; and (D ) processed spectrum The modiŽ ed polyŽ t technique relies on the observa-
using the modiŽ ed polyŽ t method (solid line) along with the generated tions by many groups that biological  uorescence is best
pure Raman component.
approximated by a high-order polynomial. Thus, there are
two separate factors that affect the processing of a mea-
sured Raman spectrum using the modiŽ ed polyŽ t meth-
spectra of human tissue.† A linear Ž t of the baseline off-
od: (1) the ability to extract faint Raman spectral features
set produced by overlapping peaks was subtracted to en-
from a much higher background intensity, and (2) the
sure a null baseline. A series of phantom spectra were
dependence upon the background  uorescence’s strict ad-
then created by adding the artiŽ cial Raman spectrum to
herence to the contours of a high-order polynomial. In
the sm oothed sponge  uorescence spectrum in varying order to test only the former factor, phantom spectra were
ratios.
created by adding the generated pure Raman spectral
component to a Ž fth-order polynomial Ž t of the smoothed
RESULTS AND DISCUSSION sponge spectrum and were processed using the modiŽ ed
The phantom spectra with various  uorescence-to-Ra- polyŽ t technique. Figure 4C shows the phantom spectra
man ratios were processed using the developed modiŽ ed created for this experiment, each spectrum normalized to
polyŽ t technique. Since the phantom spectra contained its m ean intensity and offset, for illustration purposes
known quantities of both  uorescence background and only. The resultant modiŽ ed polyŽ t-processed spectra
Raman scatter, the perform ance of the processing tech- (Fig. 4D) show that Raman bands can be extracted even
nique could be evaluated by its ability to return only the at 10 000 000:1 ratios of  uorescence to Raman intensity
added Raman scatter. Figure 4A shows the raw spectra (R 2 5 0.961). This data also illustrates that approximation
of two different phantoms prior to  uorescence subtrac- of the  uorescence background by a high-order polyno-
tion where m aximum  uorescence intensity was 10 and mial is a limiting factor of this method. W hile biological
100 times more than the maximum Raman intensity (at  uorescence has been reported to be best approximated
1451 rel. cm 2 1), respectively. For illustration purposes, by a high-order polynomial on the order of 4 –5,8,10 this
each phantom spectrum was norm alized to its mean in- may not always be the case (in non-biological spectra this
tensity and the spectra were offset for clarity. After m od- is even more plausible). However, the results presented
iŽ ed polyŽ t processing (Fig. 4B), the Ž ltered spectra here are directed towards the application of this m ethod
for biological materials and thus show its validity using
† Centers (and respective full widths at half-maximum) of peaks used
a polynomial Ž t. For other types of materials, the basic
were 568.75 (70), 792.75 (59.5), 1002.75 (35), 1107.75 (105), premise of the modiŽ ed polyŽ t method can also be im-
1195.25 (94.5), 1450.75 (28), and 1646.75 (31.5) rel. cm 2 1. plemented using non-polynom ial background line shapes,

1366 Volume 57, Number 11, 2003


such as Gaussian, Weibull, logarithmic, exponential, or the Raman spectra of these in vitro human tissues before
other distributions. (Fig. 6A) and after processing with the modiŽ ed polyŽ t
Figure 5 compares the modiŽ ed polyŽ t method with method (Fig. 6B) and demonstrates the effectiveness of
two other conventional  uorescence subtraction tech- the presented method in subtracting tissue  uorescence
niques typically used to process biological spectra, Ž rst- and retaining Raman signatures. Although the intensity
order differentiation and m anual polynomial curve Ž tting. of the broad peak in the 800 –900 rel. cm 2 1 region varies
The smoothed sponge spectrum was used as the phantom considerably due to deviations in silica contamination
 uorescence background, at a 20:1 intensity ratio with from within the collection optics, the dominant tissue
the generated pure Raman spectrum. This ratio was se- peaks at ;1200 –1400, ;1450, and ;1660 rel. cm 2 1 are
lected for its approximation to that encountered in bio- clearly retained.
logical measurements. Figure 5 clearly illustrates the ad-
vantages of the modiŽ ed polyŽ t method as compared to CONCLUSION
the other methods in terms of consistency, peak retention, The goal of any  uorescence subtraction technique is
and baseline drift. The output of the modiŽ ed polyŽ t the unbiased removal of all  uorescence contribution,
method is seen to accurately retain the Raman spectral with no effect on the Raman bands. The derivative meth-
components of the phantom. In comparison, the deriva- od is quite adept at the former requirement; however, it
tive method m akes the lower intensity Raman bands fails miserably at the latter. Polynomial curve-Ž tting, in
(1050 –1300 rel. cm 2 1 ) difŽ cult to identify in the pro- principle, provides a solution to both, but its implemen-
cessed spectrum, and the conversion of Raman peaks to tation thus far has relied on user intervention, which
sinusoids makes traditional peak identiŽ cation difŽ cult, makes this technique unreliable and time consuming. The
especially between peaks of close proximity. The manual modiŽ ed polyŽ t m ethod described here automates the
polynom ial Ž t was performed by three separate individ- manual curve-Ž tting process, thus fulŽ lling the require-
uals so that variability due to user intervention could be ments of an effective  uorescence subtraction method.
realized. The individuals were given the raw phantom This m ethod can be completely automated and has been
spectrum with no a priori knowledge of the Raman peak shown to be proŽ cient in producing a null baseline and
locations. Even in this small number of cases, the vari- retaining Raman spectral contours, even at exceedingly
ability in the Ž t is evident. small signal-to-noise ratios.
W hile demonstration of the modiŽ ed polyŽ t technique
1. R. Van-Duyne, D. Jean maire, and D. Shriver, Anal. Chem. 46, 213
on a phantom demonstrates its capability, the technique (1974).
was designed for application to biological samples, spe- 2. P. Mosier-Boss, S. Lieberman, and R. Newberr y, Appl. Spectrosc.
ciŽ cally hum an tissues. Several human tissue specimens 49, 630 (1995).
(ovary, skin, and cervix) were obtained fresh-frozen for 3. J. Zhao, M. M. Carrabba, and F. S. Allen, Appl. Spectrosc. 56, 834
(2002).
this study, thawed in room temperature phosphate buff- 4. D. Zhang and D. Ben-Amotz, Appl. Spectrosc. 54, 1379 (2000).
ered saline (PB S) solution, and measured using the spec- 5. A. O’Grady, A. Dennis, D. Denvir, J. M cGar vey, and S. Bell, Anal.
troscopic system described (P 5 80 m W, 30 s integra- Chem. 73, 2058 (2001).
tion). Although the modiŽ ed polyŽ t method can be used 6. V. Barclay and R. Bonner, Anal. Chem. 69, 78 (1997).
7. T. Cai, D. Zhang, and D. Ben-Amotz, Appl. Spectrosc. 55, 1124
to automate  uorescence subtraction from measured Ra- (2001).
man signals of different types of materials, it was de- 8. A. Mahadevan-Jansen, M. F. M itchell, N. Ramanujam, A. Malpica,
signed speciŽ cally for use with tissue spectra. Tissue S. Thomsen, U. Utzinger, and R. Richards-Kortum, Photochem .
spectra typically display low signal-to-noise ratios (as Photobiol. 68, 123 (1998).
9. T. Vickers, J. R. E. Wambles, and C. Mann, Appl. Spectrosc. 55,
compared to chemical standards) and a  uorescence 389 (2001).
background that has been described as best approximated 10. J. F. Brennan, Y. Wang, R. R. Dasari, and M. S. Feld, Appl. Spec-
by a fourth- or Ž fth-order polynomial.8,10 Figure 6 shows trosc. 51, 201 (1997).

APPLIED SPECTROSCOPY 1367

You might also like