You are on page 1of 2

EX-200 Inductive Gauging Sensors

Accurate analog output of minute displacement

Inductive Gauging
The EX-200 Series measures target displacement with a resolution
of 0.04% F.S. (0.4 µm 0.00002" with the EX-305 sensor head). The
dual analog outputs (voltage/current) enable connection to external

Sensors equipment.

EX-200 Series Linearity of ±1% of F.S.


Using the built-in linearizer circuit, the EX-200 Series accurately
outputs absolute displacement values.
Refer to P.821 for a list of products
complying with EU Directives.

High response speed


The EX-200 Series offers a response frequency of up to 18 kHz (with
the EX-305), allowing measurement of rapidly vibrating targets.

FREE DOWNLOAD

www.keyence.com/GQR
Free downloads for product and technical support are readily
available in one convenient location

518
Inductive Gauging Sensors EX-200 GT

MEASUREMENT SENSORS
LK
IL
Lineup IA
Model LJ
Measuring Mounting
Type Appearance Resolution
Sensor head Controller range size SI
LT
ø5.4 x 18 mm 0 to 1 mm 0.4 µm
Cylindrical EX-305 EX-201 ø5.4 mm ø0.21" LS
ø0.21" x 0.71" 0.04" 0.00002"
TM
M10 x 18 mm 0 to 2 mm 0.8 µm IG
Threaded EX-110 EX-202 0.08" M10
0.71" 0.00003" IB
FW
ø14.5 x 20 mm 0 to 5 mm 2 µm EX
EX-416 EX-205 M16
ø0.57" x 0.79" 0.20" 0.00008"
Cylindrical/ GA
Threaded RD
ø22 x 35 mm 0 to 10 mm 4 µm
EX-422 EX-210 M12
ø0.87" x 1.38" 0.39" 0.00016"

Specifications
Cylindrical Threaded Cylindrical/Threaded
Type
ø5.4 ø0.21" M10 ø14.5 ø0.57"/M16 ø22 ø0.87"/M12
Sensor head EX-305 EX-110 EX-416 EX-422
Model
Controller EX-201 EX-202 EX-205 EX-210
Measuring range 0 to 1 mm 0.04" 0 to 2 mm 0.08" 0 to 5 mm 0.20" 0 to 10 mm 0.39"
Output voltage 0 to 5 V (Output impedance 100 Ω) New Products
Output current 4 to 20 mA (Applicable load: 0 to 350 Ω)
Resolution 0.04% of F.S. (Response frequency: LOW) Fiberoptic
Analog output
Linearity ±1% of F.S. Sensors

Response HIGH 18 kHz (-3 dB) 15 kHz (-3 dB) 13 kHz (-3 dB) 8 kHz (-3 dB)
frequency Photoelectric
LOW 1.3 kHz (-3 dB)
Sensors
Disconnection alarm output 2.
NPN: 100 mA (40 V) max., Residual voltage: 1 V max. (N.C.)
Functions Auto-zero function/Response frequency selecting function Proximity
Temperature Sensor head 1. 0.03% of F.S./°C Sensors
fluctuation Controller 1. 0.04% of F.S./°C
Power supply voltage 12 to 24 VDC ±10% Safety
Equipment
Current consumption 240 mA max.
Sensor head -10 to +60°C (14 to 140°F), No freezing Flow/
Ambient temperature Pressure/
Controller 0 to 50°C (32 to 122°F)
Temperature
Relative humidity 35 to 85%, No condensation
Enclosure rating Sensor head: IP67 Measurement
Sensor head Approx. 45 g Approx. 55 g Approx. 75 g Approx. 200 g
Sensors
Weight
Controller Approx. 285 g
PLCs
The above data was obtained using a steel target (S45C, SS41, t=1 mm 0.04" ).
When measuring aluminum, copper, or stainless steel targets, refer to the linear characteristics for these materials.
1. When the distance between the sensor head and the target is within 50% of the measuring range.
2. NPN output can easily be converted to PNP output by connecting the optional OP-5148 PNP output converter. Static
Eliminators

Dimensions Unit: mm inch


Vision
Systems
Controllers Sensor heads 0.71" 118.11"
18 3000
Marking
EX-201/202/ EX-305
ø5.4 Equipment
205/210 ø0.21" ø3.2 ø0.13"

EX-110 0.71" 118.11" Code Readers


112
4.41"
18 3000

Digital
M10, P=1 ø3.2 ø0.13"
Microscopes
0 to 180° Across-flats:14, t=3 0.55", t=0.12"

4-ø5 ø0.20" EX-416 0.55" Measurement


0.24" 118.11" Systems
mounting hole 6 14 3000

ø14.5
ø0.57" ø3.2 ø0.13"
112 M16, P=1 Across-flats: 21, t=4 0.83", t=0.16"
4.41"
99 35.4
3.90" 1.39" EX-422 0.79" 0.59" 118.11"
90 20 15 3000
3.54"
ø22
ø0.87"
ø5.4 ø0.21"
40 1.57" Across-flats: 19, t=3.5 M12, P=1
0.10"
60 2.36" 2.5 65 2.56" 0.75", t=0.14"

CAD DATA Download: