You are on page 1of 102

© Siemens AG 2007

SITRANS T
measuring instruments
for temperature

3/74 Accessories
3/2 Product overview 3/74 - Measuring inserts, not explosion
3/6 Transmitter for mounting in sensor protected
head 3/76 - Measuring inserts, explosion
3/6 SITRANS TH100, two-wire system protected
(Pt100) 3/77 - Welding-type protective tubes,
3/9 SITRANS TH200, two-wire system neck tubes and connection heads
universal 3/79 Thermocouples
3/15 SITRANS TH300, two-wire system 3/79 Technical description
universal, HART
3/83 Temperature transmitters for moun-
3/21 SITRANS TH400, fieldbus transmitter ting in the connection head
3/26 Transmitter for rail mounting 3/84 Questionnaire for temperature sen-
3/26 SITRANS TR200, two-wire system sors (resistance thermometers and
universal thermocouples)
3/32 SITRANS TR300, two-wire system 3/85 Straight thermocouples
universal, HART 3/85 - to DIN 43733, with connection head
3/38 SITRANS TW, four-wire system, 3/86 - Individual parts and accessories
universal, HART 3/88 Jacket thermocouples
3/88 - with extension lead
3/50 Transmitter for field mounting with 3/89 - with connection head, form B
temperature sensor
3/90 - with socket
3/50 SITRANS TF2, Digital display ther-
3/87 Individual parts
mometer
3/92 Accessories
3/53 SITRANS TF, two-wire system
3/92 - Cold junction with built-in power
3/53 Field indicator supply unit
3/53 SITRANS TF, Field indicator for 4 to 3/93 - Cold junction thermostat for
20 mA 6 or 12 measuring points
3/94 - Cold junction thermostat for
3/60 Resistance thermometers 12 or 24 measuring points
3/60 Technical description
3/62 Protective tubes, installation 3/95 Resistance thermometers for food,
3/64 Temperature transmitters for moun- pharmaceuticals and
ting in the connection head biotechnology
3/65 Questionnaire for temperature sen- 3/95 Resistance thermometers for instal-
sors (resistance thermometers and lation in pipelines and tanks
thermocouples) 3/99 Resistance thermometers with
3/66 Flue gas resistance thermometers, clamp-on system
with connection head
3/67 Low-pressure screw-in resistance
thermometers, with connection head
3/67 - without neck tube
3/68 - with neck tube
3/69 High-pressure screw-in resistance
thermometers
3/70 High-pressure welding-type resis-
tance thermometer
3/71 Flange-type resistance thermometer
3/71 - with connection head You can download all instructions, cata-
logs and certificates for SITRANS T free
3/72 - with reduced response time, with
of charge at the following Internet
connection head
address:
3/73 Resistance thermometers for damp
www.siemens.com/sitranst
rooms

Siemens FI 01 · 2008
© Siemens AG 2007

SITRANS T Measuring instruments for temperature


Product overview

■ Overview
Application Mounting of transmitter with Page Software for
Ex protection parameterization
Transmitters for temperature for mounting in sensor heads
Transmitter Sensor

3
SITRANS TH100 Zone 2, Zone 2, 3/6 SIPROM T
Transmitters for Pt100 zone 1 zone 1,
zone 0
• Mounting in sensor head

SITRANS TH200 Zone 2, Zone 2, 3/9 SIPROM T


Transmitters for connection to zone 1 zone 1,
resistance thermometers, resis- zone 0
tance-based sensors, thermocou-
ples and DC voltages up to 1.1 V
• Two-wire system
• Universal

SITRANS TH300 Zone 2, Zone 2, 3/15 SIMATIC PDM


Transmitters for connection to zone 1 zone 1,
resistance thermometers, resis- zone 0
tance-based sensors, thermocou-
ples and DC voltages up to 1.1 V
• Two-wire system
• Universal
• HART

SITRANS TH400 Zone 2, Zone 2, 3/21 SIMATIC PDM for TH 400 with
Transmitters for connection to zone 1, zone 1, PROFIBUS PA
resistance thermometers, resis- zone 21 zone 0,
tance-based sensors, thermocou- zone 21,
ples and DC voltages up to 0.9 V zone 20
• Fieldbus transmitter
• PROFIBUS PA
• FOUNDATION Fieldbus

3/2 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T Measuring instruments for temperature


Product overview

Application Mounting of transmitter with Page Software for


Ex protection parameterization
Transmitters for temperature for rail mounting
Transmitter Sensor
SITRANS TR200 Zone 2, Zone 2, 3/26 SIPROM T
Transmitters for connection to zone 1, zone 1,
zone 21 zone 0,
3
resistance thermometers, resis-
tance-based sensors, thermocou- zone 21,
ples and DC voltages up to 1.1 V zone 20
• Two-wire system
• Universal

SITRANS TR300 Zone 2, Zone 2, 3/32 SIMATIC PDM


Transmitters for connection to zone 1, zone 1,
resistance thermometers, resis- zone 21 zone 0,
tance-based sensors, thermocou- zone 21,
ples and DC voltages up to 1.1 V zone 20
• Two-wire system
• Universal,
• HART

SITRANS TW Safe area Zone 1, 3/38 SIMATIC PDM


Transmitters for connection to zone 0,
resistance thermometers, resis- zone 21,
tance-based sensors, thermocou- zone 20
ples, DC voltages and DC currents
for:
• Four-wire system

Transmitter for temperature for field mounting


SITRANS TF2 - - 3/50 Local programming using keys
Digital display thermometer
• Transmitter with LCD display and
mounted Pt100

SITRANS TF Zone 2, Zone 2, 3/53 Depends on mounted transmit-


Transmitters for connection to zone 1 zone 1, ter TH200/TH300
resistance thermometers, resis- zone 0
tance-based sensors, thermocou-
ples and DC voltages up to 1.1 V
• In field housing for heavy industri-
al use

Field indicator for 4 to 20 mA signals


SITRANS TF Zone 2, - 3/53 -
Field indicator for 4 to 20 mA sig- zone 1
nals
Display of units can be user-
defined

Siemens FI 01 · 2008 3/3


© Siemens AG 2007

SITRANS T Measuring instruments for temperature


Product overview

Device Largest measuring range Page


Resistance thermometers
Temperature transmitters factory-installed in 3/64
connection head of a resistance thermome-
ter (selection list)
• SITRANS TH400 PA or FF

3
• SITRANS TH200/TH300
• SITRANS TH100

Flue gas resistance thermometers -50 ... +600 °C (-58 ... +1112 °F) 3/66

Low-pressure screw-in resistance thermo- 3/67


meters
• Without neck tube -50 ... +400 °C (-58 ... +752 °F)
• With neck tube -50 ... +600 °C (-58 ... +1112 °F)

High-pressure screw-in resistance thermo- -50 ... +600 °C (-58 ... +1112 °F) 3/69
meters

High-pressure welding-type resistance ther- -50 ... 550 °C (-58 ... +1022 °F) 3/70
mometers

Flange-type resistance thermometers -50 ... +600 °C (-58 ... +1112 °F) 3/71

Resistance thermometers for damp rooms -30 ... +60 °C (-22 ... +140 °F) 3/73

3/4 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T Measuring instruments for temperature


Product overview

Device Largest measuring range Page


Thermocouples
Temperature transmitters factory-installed in 3/83
connection head of a thermocouple (selec-
tion list)
• SITRANS TH400 PA or FF
• SITRANS TH200/TH300

3
Straight thermocouples 0 ... 1250 °C (32 ... 2282 °F) 3/85

Jacket thermocouples with extension lead 0 ... 1100 °C (32 ... 2012 °F) 3/88

Jacket thermocouples with connection 0 ... 1100 °C (32 ... 2012 °F) 3/89
head, form B

Jacket thermocouples with socket 0 ... 1100 °C (32 ... 2012 °F) 3/90

Cold junction/cold junction thermostat


Cold junction with built-in power supply unit Reference temperature 3/92
0 or 20 °C (32 ... 68 °F)

Cold junction thermostat Reference temperature


• For max. 6 or 12 measuring points 50, 60 ... 70 °C 3/93
(122, 144 ... 158 °F)
• For max. 12 or 24 measuring points 50, 60 or 70 °C 3/94
(122, 144 or 158 °F)

Resistance thermometers for food, pharmaceuticals and biotechnology


Resistance thermometers - 50 ... + 400 °C 3/95
• For installation in pipelines and tanks (-58 ... +752 °F)
• For temperature measurements with hy-
gienic requirements

Resistance thermometers with clamp-on - 20 ... + 200 °C 3/99


system (- 4 ... + 392 °F)
The temperature sensors with pipe sleeve
are primarily used for temperature monitor-
ing in sterile applications in the food and
pharmaceutical industries.

Siemens FI 01 · 2008 3/5


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitters for mounting in sensor head
SITRANS TH100
two-wire system (Pt100)

■ Overview ■ Application
Used in conjunction with Pt100 resistance thermometers, the
SITRANS TH100 transmitters are ideal for measuring tempera-
tures in all industries. Due to its compact size it can be installed
in the connection head type B (DIN 43729) or larger.
The output signal is a direct current from 4 to 20 mA that is pro-

3
portional to the temperature.
Parameterization is implemented over the PC using the parame-
terization software SIPROM T and the modem for SITRANS
TH100/TH200. If you already have a "modem for SITRANS TK"
(Order No. 7NG3190-6KB), you can continue using this to pa-
rameterize the SITRANS TH100.
Transmitters of the "intrinsically-safe" type of protection can be
installed within potentially explosive atmospheres. The devices
comply with the Directive 94/9/EC (ATEX), as well as FM and
CSA regulations.

■ Function
The SITRANS TH100 dispenses with electrical isolation and uni- Mode of operation
versal sensor connection to provide a low-cost alternative for
Pt100 measurements. The measured signal supplied by a Pt100 resistance thermome-
ter (2, 3 or 4-wire system) is amplified in the input stage. The volt-
For the parameterization, the SIPROM T software is used in com- age, which is proportional to the input variable, is then converted
bination with the modem for SITRANS TH100/TH200. into digital signals by a multiplexer in an analog/digital converter.
Its extremely compact design makes the SITRANS TH100 ideal They are converted in the microcontroller in accordance with the
for the retrofitting of measuring points or for the use of analog sensor characteristics and further parameters (measuring
transmitters. range, damping, ambient temperature etc.).
The transmitter is available as a non-Ex version as well as for use The signal prepared in this way is converted in a digital/analog
in potentially explosive atmospheres. converter into a load-independent direct current of 4 to 20 mA.
An EMC filter protects the input and output circuits against elec-
■ Benefits tromagnetic interferences.
• Two-wire transmitter
• Assembly in connection head type B (DIN 43729) or larger, or
on a standard DIN rail
• Can be programmed, which means that the sensor connec-
tion, measuring range, etc. can also be programmed
• Intrinsically-safe version for use in potentially explosive areas

8UHI Input
,F 3W 5HVLVWDQFHWKHUPRPHWHU
(0&B ,QSXWOHYHOZLWKSURWHFWLYHFRPSRQHQWV
,& &RQVWDQWFXUUHQWVRXUFH
 08; 0XOWLSOH[HU
$' $QDORJGLJLWDOFRQYHUWHU
$' Output

3W (0&B '$ 'LJLWDODQDORJFRQYHUWHU
 8, 9ROWDJHWUDQVIRUPHUVFXUUHQW
08;
 WUDQVIRUPHUVFRQVWDQWYROWDJH
¦&  DQGUHIHUHQFHYROWDJHVRXUFH
 (0&B 2XWSXWOHYHOZLWKSURWHFWLYHFRPSRQHQWV
8DX[ $X[LOLDU\SRZHUVXSSO\
,RXW 2XWSXWFXUUHQW
'$
Microcontroller
˩& /LQHDUL]DWLRQIXQFWLRQVDQGVWRUDJH
8UHI  RIDOOGDWD


8 (0&B P$ 8DX[,RXW
,


SITRANS TH100, function diagram

3/6 Siemens FI 01 · 2008


© Siemens AG 2008

SITRANS T measuring instruments for temperature


Transmitters for mounting in sensor head
SITRANS TH100
two-wire system (Pt100)

■ Technical specifications Certificate and approvals


Input Explosion protection ATEX
Resistance thermometer EC type test certificate PTB 05 ATEX 2049X
Measured variable Temperature • Intrinsically-safe" type of II 1G EEx ia IIC T6/T4
protection II 2(1)G EEx ia/ib IIC T6/T4
Sensor type PT100 to IEC 60751
• „Operating equipment that is non- II 3G EEx nAL IIC T6/T4
Characteristic Temperature-linear
3
ignitable and has limited energy“
Type of connection 2, 3 or 4-wire circuit type of protection

Resolution 14 bit Explosion protection to FM for USA


and Canada (cFMUS)
Measuring accuracy • FM approval PID 3024169
• Span <250 °C (450 °F) < 0.25 °C (0.45 °F) • Degree of protection IS Cl I, II, III, Div 1, GP ABCDEFG
• Span >250 °C (450 °F) < 0.1% of span T4/T5/T6
Repeatability < 0.1 °C (0.18 °F) IS Cl I, ZN 0, 1 AEx ia IIC T4/T5/T6
Nl Cl I, II, III, Div 2, GP ABCDFG
Measuring current approx. 0.4 mA T4/T5/T6
Measuring cycle < 0.7 s Cl I, ZN 2, GP IIC T4/T5/T6

Range -200 ... +850 °C Software requirements for


(-328 ... +1562 °F) SIPROM T

Measured span 25 ... 1050 °C (77 ... 1922 °F) PC operating system: Windows ME, 2000 and XP;
also Windows 95, 98 and 98SE,
Unit °C or °F but only in connection with
Offset programmable: RS-232 modem.
-100 ... +100 °C (-180 ... +180 °F) Factory setting:
Line resistance Max. 20 Ω (total from feeder and • Pt100 (IEC 751) with three-wire circuit
return conductor)
• Measuring range: 0 ... 100 °C (32 ... 212 °F)
Noise rejection 50 and 60 Hz • Error signal in the event of sensor breakage: 22.8 mA
Output • Sensor offset: 0 °C (0 °F)
• Damping 0.0 s
Output signal 4 ... 20 mA, two-wire
Power supply 8.5 ... 36 V DC (30 V for Ex) ■ Selection and Ordering data Order-No.
Max. load (Uaux - 8.5 V)/0.023 A SITRANS TH100 temperature
Overrange 3.6 ... 23 mA, continuously adjust- transmitters for Pt100
able (default value: For installation in the connection
3.84 ... 20.5 mA) head, Type B (DIN 43729)
Two-wire system 4 ... 20 mA, pro-
Error signal (in the event of sensor 3.6 ... 23 mA, continuously adjust-
grammable, without electrical iso-
breakage) able (default value: 3.6 mA or
lation
22.8 mA)
• Without explosion protection } 7NG3211-0NN00 C)
Damping time 0 ... 30 s (default value: 0 s)
• With explosion protection,
Protection Against reversed polarity
"Intrinsic safety" and for zone 2
Resolution 12 bit
- to ATEX } 7NG3211-0AN00 C)
Accuracy at 23 °C (73.4 °F) < 0.1% of span
- to FM (cFMUS) } 7NG3211-0BN00 C)
Temperature effect < 0.1%/10 °C (0.1%/18 °F)
Further designs Order code
Effect of auxiliary power < 0.01% of span/V Please add "-Z" to Order No. and
Effect of load impedance < 0.025% of max. span/100 specify Order code(s)

Long-term drift Customer-defined operating data Y01


• in the first month < 0.025% of max. span Test protocol (5 measuring points) C11
• after one year < 0.035% of max. span Accessories Order-No.
• after 5 years < 0.05% of max. span
Modem for SITRANS TH100 and
Ambient temperature TH200 incl. SIPROM T parame-
terization software
Ambient temperature range -40 ... +85 °C (-40 ... +185 °F)
• with USB connection } 7NG3092-8KU C)
Storage temperature range -40 ... +85 °C (-40 ... +185 °F)
• with RS 232 connection } 7NG3092-8KM C)
Relative humidity 98%, with condensation
CD for measuring instruments } A5E00364512 G)
Electromagnetic compatibility According to EN 61326 and
for temperature
NAMUR NE21
With documentation in German,
Construction English, French, Spanish, Italian,
Approx. weight 50 g Portuguese and SIPROM T
parameterization software
Dimensions See dimension drawing
DIN rail adapters for head } 7NG3092-8KA
Material Molded plastic transmitters
(Quantity delivered: 5 units)
Cross-section of cables Max. 2.5 mm2 (AWG 13)
Degree of protection to EN 60529 } Available ex stock.
C) Subject to export regulations AL: N, ECCN: EAR99.
• Enclosure IP40
G) Subject to export regulations AL: N, ECCN: 5D992B1.
• Terminals IP00
Power supply units see "SITRANS I supply units and input isolators".

Siemens FI 01 · 2008 3/7


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitters for mounting in sensor head
SITRANS TH100
two-wire system (Pt100)

■ Dimensional drawings Mounting on DIN rail

 
3
44 (1.73) ,QWHUQDOGLDPHWHU
&HQWHUKROH 

SITRANS TH100, mounting of transmitter on DIN rail


0RXQWLQJVFUHZ
0[
14 (0.55) 50,5 (1.99)
33 (1.30)

33 (1.3)

59,6 (2.35)
  DQG   $X[LOLDU\SRZHUVXSSO\8DX[RXWSXWFXUUHQW,2XW
DQG 3WVHQVRU IRUFRQQHFWLRQVHH
 6HQVRUFRQQHFWLRQDVVLJQPHQW

SITRANS TH100, dimensions in mm (inch)

■ Schematics
DIN rail adaptor, dimensions in mm (inch)

7ZRZLUHV\VWHP 7KUHHZLUHV\VWHP
SDUDPHWHUL]DEOHOLQH
UHVLVWDQFH

8DX[
 

)RXUZLUHV\VWHP &RQQHFWLRQRIDX[LOLDU\
SRZHUVXSSO\ 8DX[

SITRANS TH100, sensor connection assignment

3/8 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitters for mounting in sensor head
SITRANS TH200
two-wire system, universal

■ Overview • Configuration status stored in EEPROM


• Expanded diagnostic functions, such as slave pointer, opera-
ting hours counter, etc.
• Special characteristic
• Electromagnetic compatibility to EN 61326 and NE21

■ Application
SITRANS TH200 transmitters can be used in all industrial sec-
tors. Due to their compact size they can be installed in the con-
3
nection head type B (DIN 43729) or larger. The following sen-
sors/signal sources can be connected over their universal input
module:
• Resistance thermometers (two, three or four-wire system)
• Thermocouple elements
• Resistance-based sensors and DC voltage sources
The output signal is a direct current from 4 to 20 mA in accor-
dance with the sensor characteristic.
Transmitters of the "Intrinsically-safe" type of protection can be
Ultra flexible - with the universal SITRANS TH200 transmitter installed within potentially explosive atmospheres (zone 0). Type
• Two-wire devices for 4 to 20 mA of protection Ex n (zone 2) can also be achieved. Explosion-
• Mounting in the connection head of the temperature sensor proof versions for Europe, USA and Canada are also available.
• Universal input for virtually any type of temperature sensor
• Configurable over PC
■ Function
The SITRANS TH200 is configured over a PC. A USB or RS 232
■ Benefits modem is linked to the output terminals for this purpose. The
configuration data can now be edited using the SIPROM T soft-
• Compact design ware tool. The configuration data are then permanently stored in
• Flexible mounting and center hole allow you to select your pre- the non-volatile memory (EEPROM).
ferred type of installation Once the sensors and power supply have been correctly con-
• Electrically isolated nected, the transmitter outputs a temperature-linear output sig-
• Test sockets for multimeters nal and the diagnostics LED displays a green light. In the case
of a sensor short-circuit, the LED flashes red, an internal device
• Diagnostics LED (green/red) fault is indicated by a steady red light.
• Sensor monitoring
open circuits and short-circuits The test socket can be used to connect an ammeter at any time
for monitoring purposes and plausibility checks. The output cur-
• Self-monitoring rent can be read without any interruption, or even without ope-
ning the current loop.

6,75$167+7+

7HVW
7&57'
VHQVRU

$ '
µ& µ& P$ 8DX[,RXW
' $



Input Output   (OHFWULFDOO\LVRODWHG


$' $QDORJGLJLWDOFRQYHUWHU ˩& 0LFURFRQWUROOHUSULPDU\FLUFXLW   /('
6HQVRUV 5HVLVWDQFHWKHUPRPHWHUWKHUPRFRXSOH '$ 'LJLWDODQDORJFRQYHUWHU
 UHVLVWDQFHEDVHGVHQVRUP9VHQVRU 8DX[ $X[LOLDU\SRZHUVXSSO\
˩& 0LFURFRQWUROOHUVHFRQGDU\FLUFXLW ,RXW 2XWSXWFXUUHQW

SITRANS TH200 function diagram

Siemens FI 01 · 2008 3/9


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitters for mounting in sensor head
SITRANS TH200
two-wire system, universal

■ Technical specifications Open-circuit monitoring can be switched off


Short-circuit monitoring can be switched off (value is
Input adjustable)
Resistance thermometer Measuring range Parameterizable, max.
Measured variable Temperature 0 ... 2200 Ω (see Table "Digital
measuring errors")
Sensor type
Min. measured span 5 Ω ... 25 Ω (see Table "Digital

3
• to IEC 60751 Pt25 to Pt1000 measuring errors")
• to JIS C 1604; a = 0.00392 K-1 Pt25 to Pt1000 Characteristic curve Resistance-linear or special cha-
• to IEC 60751 Ni25 to Ni1000 racteristic

• Special type over special characteristic Thermocouple elements


(max. 30 points) Measured variable Temperature
Sensor factor 0.25 ... 10 (adaptation of the Sensor type (thermocouples)
basic type, e.g. Pt100 to version
Pt25 ... Pt1000) • Type B Pt30Rh-Pt6Rh to DIN IEC 584
Units °C or °F • Type C W5%-Re to ASTM 988
Connection • Type D W3%-Re to ASTM 988
• Standard connection 1 resistance thermometer (RTD) • Type E NiCr-CuNi to DIN IEC 584
in 2-wire, 3-wire or 4-wire system • Type J Fe-CuNi to DIN IEC 584
• Generation of average value 2 identical resistance thermome- • Type K NiCr-Ni to DIN IEC 584
ters in 2-wire system for genera-
tion of average temperature • Type L Fe-CuNi to DIN 43710
• Generation of difference 2 identical resistance thermome- • Type N NiCrSi-NiSi to DIN IEC 584
ters (RTD) in 2-wire system
(RTD 1 – RTD 2 or RTD 2 – RTD 1) • Type R Pt13Rh-Pt to DIN IEC 584

Interface • Type S Pt10Rh-Pt to DIN IEC 584

• Two-wire system Parameterizable line resistance • Type T Cu-CuNi to DIN IEC 584
≤ 100 Ω (loop resistance) • Type U Cu-CuNi to DIN 43710
• Three-wire system No balancing required Units °C or °F
• Four-wire system No balancing required Connection
Sensor current ≤ 0.45 mA • Standard connection 1 thermocouple element (TC)
Response time ≤ 250 ms for 1 sensor with open- • Generation of average value 2 thermocouple elements (TC)
circuit monitoring
• Generation of difference 2 thermocouple elements (TC)
Open-circuit monitoring can be switched off TC1 – TC2 or TC2 – TC1)
Short-circuit monitoring can be switched off (value is Response time ≤ 250 ms for 1 sensor with open-
adjustable) circuit monitoring
Measuring range Parameterizable (see table "Digi- Open-circuit monitoring can be switched off
tal measuring errors")
Cold junction compensation
Min. measured span 10 °C (18 °F)
• Internal With integrated Pt100 resistance
Characteristic curve Temperature-linear or special thermometer
characteristic
• External With external Pt100 IEC 60571
Resistance-based sensors (2-wire or 3-wire connection)
Measured variable Actual resistance • External fixed Cold junction temperature can be
Sensor type Resistance-based, potentiome- set as fixed value
ters Measuring range Parameterizable (see table "Digi-
Units Ω tal measuring errors")

Connection Min. measured span Min. 50 ... 100 °C (90 ... 180 °F)
(see table "Digital measuring
• Normal connection 1 resistance-based sensor (R) in errors")
2-wire, 3-wire or 4-wire system
Characteristic curve Temperature-linear or special
• Generation of average value 2 resistance-based sensors in characteristic
2-wire system for generation of
average value mV sensor

• Generation of difference 2 resistance thermometers in Measured variable DC voltage


2-wire system Sensor type DC voltage source (DC voltage
(R1 – R2 or R2 – R1) source possible over an exter-
Interface nally connected resistor)

• Two-wire system Parameterizable line resistance Units mV


≤ 100 Ω (loop resistance) Response time ≤ 250 ms for 1 sensor with open-
• Three-wire system No balancing required circuit monitoring

• Four-wire system No balancing required Open-circuit monitoring can be switched off

Sensor current ≤ 0.45 mA Short-circuit monitoring can be switched off (value is


adjustable)
Response time ≤ 250 ms for 1 sensor with open-
circuit monitoring

3/10 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitters for mounting in sensor head
SITRANS TH200
two-wire system, universal
Measuring range -10 ... +70 mV Certificates and approvals
-100 ... +1100 mV
Explosion protection ATEX
Min. measured span 2 mV or 20 mV
EC type test certificate PTB 05 ATEX 2040X
Overload capability of the input -1.5 ... +3.5 V DC
• "Intrinsic safety" type of protection II 1G EEx ia IIC T6/T4
Input resistance ≥ 1 MΩ II 2(1)G EEx ia/ib IIC T6/T4
Characteristic curve Voltage-linear or special charac- • "Operating equipment that is non- II 3G EEx nAL IIC T6/T4

3
teristic ignitable and has limited energy"
type of protection
Output
Explosion protection to FM for USA
Output signal 4 ... 20 mA, 2-wire and Canada (cFMUS)
Auxiliary power supply 11 ... 35 V DC (to 30 V with EEx) • FM approval PID 3024169
Max. load (Uaux – 11 V)/0.023 A • Degree of protection IS Cl I, II, III, Div 1, GP ABCDEFG
Overrange 3.6 ... 23 mA, infinitely adjustable T4/T5/T6
(default range: IS Cl I, ZN 0, 1 AEx ia IIC T4/T5/T6
3.84 mA ... 20.50 mA) NI Cl I, II, III, Div 2, GP ABCDFG
T4/T5/T6
Error signal (e.g. in the event of sen- 3.6 ... 23 mA, infinitely adjustable Cl I, ZN2, GP IIC T4/T5/T6
sor breakage) (default value: 22.8 mA)
Software requirements for
Sample cycle 0.25 s nominal SIPROM T
Damping Software filter 1st order 0 ... 30 s PC operating system Windows ME, 2000 and XP; also
(parameterizable) Windows 95, 98 and 98 SE, but
Protection Against reversed polarity only in connection with RS 232
modem.
Electrically isolated Input against output (1 kVeff)
Factory setting:
Measuring accuracy
• Pt100 (IEC 751) with three-wire circuit
Digital measuring errors See Table "Digital measuring
errors" • Measuring range: 0 ... 100 °C (32 ... 212 °F)
Reference conditions • Fault current: 22.8 mA
• Auxiliary power supply 24 V ± 1 % • Sensor offset: 0 °C (0 °F)
• Load 500 Ω • Damping 0.0 s
• Ambient temperature 23 °C
• Warming-up time > 5 min Digital measuring errors
Error in the analog output (digi- < 0.1 % of span Resistance thermometer
tal/analog converter)
Input Measuring range Min. measu- Digital
Error due to internal cold junction < 0.5 °C (0.9 °F) red span accuracy
Temperature effect < 0.1 % of max. span/10 °C °C (°F) °C (°F) °C (°F)
(18 °F)
to IEC 60751
Power supply effect < 0.005 % of span/V
Pt25 -200 ... +850 10 (18) 0,2 (0.36)
Effect of load impedance < 0.012 % of span/100 Ω (-328 ... +1562)
Long-term drift Pt50 -200 ... +850 10 (18) 0,15 (0.27)
• in the first month < 0.02 % of max. span (-328 ... +1562)
• after one year < 0.03 % of max. span Pt100 ... Pt200 -200 ... +850 10 (18) 0,1 (0.18)
(-328 ... +1562)
• after 5 years < 0.04 % of max. span
Pt500 -200 ... +850 10 (18) 0,15 (0.27)
Rated conditions (-328 ... +1562)
Ambient conditions Pt1000 -200 ... +350 10 (18) 0,15 (0.27)
Ambient temperature range -40 ... +85 °C (-40 ... 185 °F) (-328 ... +662)

Storage temperature range -40 ... +85 °C (-40 ... 185 °F) to JIS C1604-81

Relative humidity < 98 %, with condensation Pt25 -200 ... +649 10 (18) 0,2 (0.36)
(-328 ... +1200)
Electromagnetic compatibility acc. to EN 61326 and NE21
Pt50 -200 ... +649 10 (18) 0,15 (0.27)
Construction (-328 ... +1200)
Material Molded plastic Pt100 ... Pt200 -200 ... +649 10 (18) 0,1 (0.18)
(-328 ... +1200)
Weight 50 g (0.11 lb)
Pt500 -200 ... +649 10 (18) 0,15 (0.27)
Dimensions See "Dimensional drawings" (-328 ... +1200)
Cross-section of cables Max. 2.5 mm² (AWG 13) Pt1000 -200 ... +350 10 (18) 0,15 (0.27)
Degree of protection to IEC 60529 (-328 ... +662)
• Enclosure IP40 Ni 25 to Ni1000 -60 ... +250 10 (18) 0,1 (0.18)
(-76 ... +482)
• Terminals IP00

Siemens FI 01 · 2008 3/11


© Siemens AG 2008

SITRANS T measuring instruments for temperature


Transmitters for mounting in sensor head
SITRANS TH200
two-wire system, universal
Resistance-based sensors
■ Selection and Ordering Data Order No.
Input Measuring Min. measured Digital
range span accuracy Temperature transmitter
SITRANS TH200
Ω Ω Ω
for installation in the connection
Resistance 0 ... 390 5 0,05 head type B (DIN 43729)
Resistance 0 ... 2200 25 0,25 2-wire connection 4 ... 20 mA,
programmable, with electrical

3 Thermocouple elements
isolation
• Without explosion protection } 7NG3211-1NN00
C)
Input Measuring Min. measured Digital
• With explosion protection,
range span accuracy
"Intrinsic safety" and zone 2
°C (°F) °C (°F) °C (°F)
- acc. to ATEX } 7NG3211-1AN00
Type B 0 ... 1820 100 (180) 2 (5.40) C)
(32 ... 3308) - acc. to FM (cFMUS) } 7NG3211-1BN00
Type C (W5) 0 ... 2300 100 (180) 2 (3.60) C)
(32 ... 4172) Further designs Order code
Type D (W3) 0 ... 2300 100 (180) 2 (3.60) Please add "-Z" to Order No. and
(32 ... 4172) specify Order code(s)
Type E -200 ... +1000 50 (90) 1 (1.80) Customer-specific setting of oper- Y01
(-328 ... +1832) ating data (specify operating data
Type J -210 ... +1200 50 (90) 1 (1.80) in plain text)
(-346 ... +2192) with test protocol (5 measuring C11
Type K -230 ... +1370 50 (90) 1 (1.80) points)
(-382 ... +2498) Accessories Order No.
Type L -200 ... +900 50 (90) 1 (1.80) Modem for SITRANS TH100,
(-328 ... +1652) TH200 and TR200 incl.
Type N -200 ... +1300 50 (90) 1 (1.80) SIPROM T parameterization
(-328 ... +2372) software
Type R -50 ... +1760 100 (180) 2 (3.60) • With USB connection } 7NG3092-8KU
(-58 ... +3200) C)

Type S -50 ... +1760 100 (180) 2 (3.60) • With RS 232 connection } 7NG3092-8KM
(-58 ... +3200) C)

Type T -200 ... +400 40 (72) 1 (1.80) CD for measuring instruments } A5E00364512
(-328 ... +752) for temperature G)

Type U -200 ... +600 50 (90) 2 (3.60) with documentation in German,


(-328 ... +1112) English, French, Spanish, Italian,
Portuguese and SIPROM T
parameterization software

mV sensor DIN rail adapters for head } 7NG3092-8KA


transmitters
Input Measuring Min. measured Digital (Quantity delivered: 5 units)
range span accuracy
} Available ex stock.
mV mV µV C) Subject to export regulations AL: N, ECCN: EAR99.
mV sensor -10 ... +70 2 40 G) Subject to export regulations AL: N, ECCN: 5D992B1.
Power supply units see "SITRANS I supply units and isolation amplifiers".
mV sensor -100 ... +1100 20 400
Factory setting:
The digital accuracy is the accuracy after the analog/digital con-
version including linearization and calculation of the measured • Pt100 (IEC 751) with three-wire circuit
value. • Measuring range: 0 ... 100 °C (32 ... 212 °F)
An additional error is generated in the output current 4 to 20 mA • Fault current: 22.8 mA
as a result of the digital/analog conversion of 0.1% of the set • Sensor offset: 0 °C (0 °F)
span (digital-analog error). • Damping 0.0 s
The total error under reference conditions at the analog output is
the sum from the digital error and the digital-analog error (poss.
with the addition of cold junction errors in the case of thermo-
couple measurements).

3/12 Siemens FI 01 · 2008 Modification 01/2008


© Siemens AG 2008

SITRANS T measuring instruments for temperature


Transmitters for mounting in sensor head
SITRANS TH200
two-wire system, universal

■ Dimensional drawings Mounting on DIN rail

š 

 

(1)
(1)
3
(2) (2)

SITRANS TH200, mounting of transmitter on DIN rail


(3)
14 (0.55) 50,5 (1.99)
33 (1.30)
(4)
 

59,6 (2.35)
0[

  DQG   $X[LOLDU\SRZHUVXSSO\8DX[RXWSXWFXUUHQW,2XW
DQG 3WVHQVRU IRUFRQQHFWLRQVVHH6HQVRU
 FRQQHFWLRQDVVLJQPHQW
7HVW  7HVW   0HDVXUHPHQWRIWKHRXWSXWFXUUHQWZLWKDPXOWLPHWHU DIN rail adaptor, dimensions in mm (inch)
  7HVWWHUPLQDO
  0RXQWLQJVFUHZ0[
  /('IRURSHUDWLRQLQGLFDWLRQ
  ,QWHUQDOGLDPHWHURIFHQWHUKROH 

SITRANS TH200, dimensions and pin assignment, dimensions in mm


(inch)

Siemens FI 01 · 2008 3/13


© Siemens AG 2008

SITRANS T measuring instruments for temperature


Transmitters for mounting in sensor head
SITRANS TH200
two-wire system, universal

■ Schematics

Resistance thermometer Resistance Thermocouple

7ZRZLUHV\VWHP 7ZRZLUHV\VWHP &ROGMXQFWLRQFRPSHQVDWLRQ


,QWHUQDOIL[HGYDOXH


7KUHHZLUHV\VWHP 7KUHHZLUHV\VWHP &ROGMXQFWLRQFRPSHQVDWLRQZLWK


H[WHUQDO3WLQWZRZLUHV\VWHP

)RXUZLUHV\VWHP )RXUZLUHV\VWHP &ROGMXQFWLRQFRPSHQVDWLRQZLWK


H[WHUQDO3WLQWKUHHZLUHV\VWHP

*HQHUDWLRQRIDYHUDJH *HQHUDWLRQRIDYHUDJH *HQHUDWLRQRIDYHUDJHYDOXHGLIIHUHQFH


YDOXHGLIIHUHQFH YDOXHGLIIHUHQFH ZLWKLQWHUQDOFROGMXQFWLRQFRPSHQVDWLRQ

3URJUDPPDEOHOLQHUHVLVWDQFHIRUWKHSXUSRVHRIFRUUHFWLRQ

Connection of auxiliary
Voltage measurement Current measurement power supply (Uaux)
8DX[
 

SITRANS TH200, sensor connection assignment

3/14 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitters for mounting in sensor head
SITRANS TH300
two-wire system, universal, HART

■ Overview • Expanded diagnostic functions, such as slave pointer, opera-


ting hours counter, etc.
• Special characteristic
• Electromagnetic compatibility to EN 61326 and NE21

■ Application
3
SITRANS TH300 transmitters can be used in all industrial sec-
tors. Due to their compact size they can be installed in the con-
nection head type B (DIN 43729) or larger. The following sen-
sors/signal sources can be connected over their universal input
module:
• Resistance thermometers (two, three or four-wire system)
• Thermocouple elements
• Resistance-based sensors and DC voltage sources
The output signal is a direct current from 4 to 20 mA in accor-
dance with the sensor characteristic, superimposed by the digi-
tal HART signal.
Transmitters of the "Intrinsically-safe" type of protection can be
"HART" to beat - the universal SITRANS TH300 transmitter installed within potentially explosive atmospheres (zone 0). Type
• Two-wire devices for 4 to 20 mA, HART of protection Ex n (Zone 2) can also be achieved. Explosion-
proof versions for Europe, USA and Canada are also available.
• Mounting in the connection head of the temperature sensor
• Universal input for virtually any type of temperature sensor ■ Function
• Configurable over HART
The SITRANS TH300 is configured over HART. This can be done
■ Benefits using a handheld communicator or even more conveniently with
a HART modem and the SIMATIC PDM parameterization soft-
• Compact design ware. The configuration data are then permanently stored in the
non-volatile memory (EEPROM).
• Flexible mounting and center hole allow you to select your pre-
ferred type of installation Once the sensors and power supply have been correctly con-
• Electrically isolated nected, the transmitter outputs a temperature-linear output sig-
nal and the diagnostics LED displays a green light. In the case
• Test sockets for multimeters of a sensor short-circuit, the LED flashes red, an internal device
• Diagnostics LED (green/red) fault is indicated by a steady red light.
• Sensor monitoring The test socket can be used to connect an ammeter at any time
open circuits and short-circuits for monitoring purposes and plausibility checks. The output cur-
• Self-monitoring rent can be read without any interruption, or even without ope-
• Configuration status stored in EEPROM ning the current loop.

6,75$167+7+

7HVW
7&57'
VHQVRU

$ '
µ& µ& P$ 8DX[,RXW
' $



Input Output   (OHFWULFDOO\LVRODWHG


$' $QDORJGLJLWDOFRQYHUWHU ˩& 0LFURFRQWUROOHUSULPDU\FLUFXLW   /('
6HQVRUV 5HVLVWDQFHWKHUPRPHWHUWKHUPRFRXSOH '$ 'LJLWDODQDORJFRQYHUWHU
 UHVLVWDQFHEDVHGVHQVRUP9VHQVRU 8DX[ $X[LOLDU\SRZHUVXSSO\
˩& 0LFURFRQWUROOHUVHFRQGDU\FLUFXLW ,RXW 2XWSXWFXUUHQW

SITRANS TH 300 function diagram

Siemens FI 01 · 2008 3/15


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitters for mounting in sensor head
SITRANS TH300
two-wire system, universal, HART

■ Technical specifications Open-circuit monitoring can be switched off


Short-circuit monitoring can be switched off (value is
Input adjustable)
Resistance thermometer Measuring range Parameterizable, max.
Measured variable Temperature 0 ... 2200 Ω (see Table "Digital
measuring errors")
Sensor type
Min. measured span 5 ... 25 Ω (see Table "Digital

3
• to IEC 60751 Pt25 to Pt1000 measuring errors")
• to JIS C 1604; a = 0.00392 K-1 Pt25 to Pt1000 Characteristic curve Resistance-linear or special cha-
• to IEC 60751 Ni25 to Ni1000 racteristic

• Special type over special characteristic (max. Thermocouples


30 points) Measured variable Temperature
Sensor factor 0.25 ... 10 (adaptation of the Sensor type (thermocouples)
basic type, e.g. Pt100 to version
Pt25 ... Pt1000) • Type B Pt30Rh-Pt6Rh to DIN IEC 584
Units °C or °F • Type C W5%-Re to ASTM 988
Connection • Type D W3%-Re to ASTM 988
• Standard connection 1 resistance thermometer (RTD) • Type E NiCr-CuNi to DIN IEC 584
in 2-wire, 3-wire or 4-wire system • Type J Fe-CuNi to DIN IEC 584
• Generation of average value 2 identical resistance thermome- • Type K NiCr-Ni to DIN IEC 584
ters in 2-wire system for genera-
tion of average temperature • Type L Fe-CuNi to DIN 43710
• Generation of difference 2 identical resistance thermome- • Type N NiCrSi-NiSi to DIN IEC 584
ters (RTD) in 2-wire system
(RTD 1 – RTD 2 or RTD 2 – RTD 1) • Type R Pt13Rh-Pt to DIN IEC 584

Interface • Type S Pt10Rh-Pt to DIN IEC 584

• Two-wire system Parameterizable line resistance • Type T Cu-CuNi to DIN IEC 584
≤ 100 Ω (loop resistance) • Type U Cu-CuNi to DIN 43710
• Three-wire system No balancing required Units °C or °F
• Four-wire system No balancing required Connection
Sensor current ≤ 0.45 mA • Standard connection 1 thermocouple (TC)
Response time ≤ 250 ms for 1 sensor with open- • Generation of average value 2 thermocouples (TC)
circuit monitoring
• Generation of difference 2 thermocouples (TC) TC1 – TC2
Open-circuit monitoring can be switched off or TC2 – TC1)
Short-circuit monitoring can be switched off (value is Response time ≤ 250 ms for 1 sensor with open-
adjustable) circuit monitoring
Measuring range Parameterizable (see table "Digi- Open-circuit monitoring can be switched off
tal measuring errors")
Cold junction compensation
Min. measured span 10 °C (18 °F)
• Internal With integrated Pt100 resistance
Characteristic curve Temperature-linear or special thermometer
characteristic
• External With external Pt100 IEC 60571
Resistance-based sensors (2-wire or 3-wire connection)
Measured variable Actual resistance • External fixed Cold junction temperature can be
Sensor type Resistance-based, potentiome- set as fixed value
ters Measuring range Parameterizable (see table "Digi-
Units Ω tal measuring errors")

Connection Min. measured span Min. 50 ... 199 °C (90 ... 180 °F)
(see table "Digital measuring
• Normal connection 1 resistance-based sensor (R) in errors")
2-wire, 3-wire or 4-wire system
Characteristic curve Temperature-linear or special
• Generation of average value 2 resistance-based sensors in characteristic
2-wire system for generation of
average value mV sensor

• Generation of difference 2 resistance thermometers in 2- Measured variable DC voltage


wire system Sensor type DC voltage source (DC voltage
(R1 – R2 or R2 – R1) source possible over an exter-
Interface nally connected resistor)

• Two-wire system Parameterizable line resistance Units mV


≤ 100 Ω (loop resistance) Response time ≤ 250 ms for 1 sensor with open-
• Three-wire system No balancing required circuit monitoring

• Four-wire system No balancing required Open-circuit monitoring can be switched off

Sensor current ≤ 0.45 mA Short-circuit monitoring can be switched off (value is


adjustable)
Response time ≤ 250 ms for 1 sensor with open-
circuit monitoring

3/16 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitters for mounting in sensor head
SITRANS TH300
two-wire system, universal, HART
Measuring range -10 ... +70 mV Certificate and approvals
-100 … +1100 mV
Explosion protection ATEX
Min. measured span 2 mV or 20 mV
EC type test certificate PTB 05 ATEX 2040X
Overload capacity of the input -1.5 ... +3.5 V DC
• "Intrinsic safety" type of protection II 1G EEx ia IIC T6/T4
Input resistance ≥ 1 MΩ II 2(1)G EEx ia/ib IIC T6/T4
Characteristic curve Voltage-linear or special charac- • "Operating equipment that is non- II 3G EEx nAL IIC T6/T4

3
teristic ignitable and has limited energy"
type of protection
Output
Explosion protection to FM for USA
Output signal 4 ... 20 mA, 2-wire with communi- and Canada (cFMUS)
cation to HART Rev. 5.9
• FM approval PID 3024169
Auxiliary power supply 11 ... 35 V DC (to 30 V with EEx)
• Degree of protection IS Cl I, II, III, Div 1, GP ABCDEFG
Max. load (Uaux – 11 V)/0.023 A T4/T5/T6
Overrange 3.6 ... 23 mA, continuously adjus- IS Cl I, ZN 0, 1 AEx ia IIC T4/T5/T6
table (default range: NI Cl I, II, III, Div 2, GP ABCDFG
3.84 mA ... 20.50 mA) T4/T5/T6
Cl I, ZN2, GP IIC T4/T5/T6
Error signal (e.g. in the event of sen- 3.6 ... 23 mA, continuously adjus-
sor breakage) table (default value: 22.8 mA) Factory setting:
Sample cycle 0.25 s nominal • Pt100 (IEC 751) with three-wire circuit
Damping Software filter 1st order 0 ... 30 s • Measuring range: 0 ... 100 °C (32 ... 212 °F)
(parameterizable)
• Fault current: 22.8 mA
Protection Against reversed polarity • Sensor offset: 0 °C (0 °F)
Electrically isolated Input against output (1 kVeff) • Damping 0.0 s
Measuring accuracy
Digital measuring errors See Table "Digital measuring
errors" Digital measuring errors
Reference conditions Resistance thermometer
• Auxiliary power supply 24 V ± 1% Input Measuring range Min. measu- Digital accu-
red span racy
• Load 500 Ω
°C (°F) °C (°F) °C (°F)
• Storage temperature 23 °C
to IEC 60751
• Warming-up time > 5 min
Pt25 -200 ... +850 10 (18) 0.2 (0.36)
Error in the analog output (digi- < 0.1% of span
(-328 ... +1562)
tal/analog converter)
Pt50 -200 ... +850 10 (18) 0.15 (0.27)
Error due to internal cold junction < 0.5 °C (0.9 °F)
(-328 ... +1562)
Temperature effect < 0.1% of max. span/10 °C
Pt100 ... Pt200 -200 ... +850 10 (18) 0.1 (0.18)
(18 °F)
(-328 ... +1562)
Power supply effect < 0.005% of span/V
Pt500 -200 ... +850 10 (18) 0.15 (0.27)
Effect of load impedance < 0.012% of span/100 Ω (-328 ... +1562)
Long-term drift Pt1000 -200 ... +350 10 (18) 0.15 (0.27)
(-328 ... +662)
• in the first month < 0.02% of max. span
to JIS C1604-81
• after one year < 0.03% of max. span
Pt25 -200 ... +649 10 (18) 0.2 (0.36)
• after 5 years < 0.04% of max. span (-328 ... +1200)
Rated conditions Pt50 -200 ... +649 10 (18) 0.15 (0.27)
Ambient temperature (-328 ... +1200)
Ambient temperature range -40 ... +85 °C (-40 ... 185 °F) Pt100 ... Pt200 -200 ... +649 10 (18) 0.1 (0.18)
(-328 ... +1200)
Storage temperature range -40 ... +85 °C (-40 ... 185 °F)
Pt500 -200 ... +649 10 (18) 0.15 (0.27)
Relative humidity < 98%, with condensation (-328 ... +1200)
Electromagnetic compatibility acc. to EN 61326 and NE21 Pt1000 -200 ... +350 10 (18) 0.15 (0.27)
Construction (-328 ... +662)

Material Molded plastic Ni 25 to Ni1000 -60 ... +250 10 (18) 0.1 (0.18)
(-76 ... +482)
Weight 50 g (0.11 lb)
Dimensions See "Dimensional drawings"
Cross-section of cables Max. 2.5 mm² (AWG 13)
Degree of protection to IEC 60529
• Enclosure IP40
• Terminals IP00

Siemens FI 01 · 2008 3/17


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitters for mounting in sensor head
SITRANS TH300
two-wire system, universal, HART
Resistance-based sensors
■ Selection and Ordering Data Order No.
Input Measuring Min. measured Digital accu-
range span racy Temperature transmitter
SITRANS TH300
Ω Ω Ω
for installation in the connection
Resistance 0 ... 390 5 0.05 head type B (DIN 43729)
Resistance 0 ... 2200 25 0.25 2-wire connection 4 ... 20 mA,
communication-capable acc. to

3 Thermocouples
HART, with electrical isolation
• Without explosion protection } 7NG3212-0NN00
C)
Input Measuring Min. measured Digital
• With explosion protection;
range span accuracy
type of protection:
°C (°F) °C (°F) °C (°F) "Intrinsic safety" and EEx n
Type B 0 ... 1820 100 (180) 2 (5.40) - acc. to ATEX } 7NG3212-0AN00
(32 ... 3308) C)

Type C (W5) 0 ... 2300 100 (180) 2 (3.60) - acc. to FM (cFMUS) } 7NG3212-0BN00
(32 ... 4172) C)

Type D (W3) 0 ... 2300 100 (180) 2 (3.60) Further designs Order code
(32 ... 4172) Please add "-Z" to Order No. and
Type E -200 ... +1000 50 (90) 1 (1.80) specify Order code(s)
(-328 ... +1832) Customer-specific setting of oper- Y01
Type J -210 ... +1200 50 (90) 1 (1.80) ating data (specify operating data
(-346 ... +2192) in plain text)
Type K -230 ... +1370 50 (90) 1 (1.80) with test protocol (5 measuring C11
points)
(-382 ... +2498)
Accessories Order No.
Type L -200 ... +900 50 (90) 1 (1.80)
(-328 ... +1652) CD for measuring instruments } A5E00364512
for temperature G)
Type N -200 ... +1300 50 (90) 1 (1.80)
(-328 ... +2372) with documentation in German,
English, French, Spanish, Italian,
Type R -50 ... +1760 100 (180) 2 (3.60) Portuguese and SIPROM T
(-58 ... +3200) parameterization software
Type S -50 ... +1760 100 (180) 2 (3.60) HART modem
(-58 ... +3200)
• With RS 232 connection } 7MF4997-1DA
Type T -200 ... +400 40 (72) 1 (1.80) D)
(-328 ... +752)
• With USB connection } 7MF4997-1DB
Type U -200 ... +600 50 (90) 2 (3.60) D)
(-328 ... +1112)
SIMATIC PDM operating See Section 9
software
DIN rail adapters for head } 7NG3092-8KA
mV sensor
transmitters
Input Measuring Min. measured Digital (Quantity delivered: 5 units)
range span accuracy
} Available ex stock.
mV mV µV C) Subject to export regulations AL: N, ECCN: EAR99.
mV sensor -10 ... +70 2 40 D) Subject to export regulations AL: N, ECCN: EAR99H.
G) Subject to export regulations AL: N, ECCN: 5D992B1.
mV sensor -100 ... +1100 20 400 Power supply units see "SITRANS I supply units and isolation amplifiers".
The digital accuracy is the accuracy after the analog/digital con-
version including linearization and calculation of the measured
value.
An additional error is generated in the output current 4 to 20 mA
as a result of the digital/analog conversion of 0.1% of the set
span (digital-analog error).
The total error under reference conditions at the analog output is
the sum from the digital error and the digital-analog error (poss.
with the addition of cold junction errors in the case of thermo-
couple measurements).

3/18 Siemens FI 01 · 2008


© Siemens AG 2008

SITRANS T measuring instruments for temperature


Transmitters for mounting in sensor head
SITRANS TH300
two-wire system, universal, HART

■ Dimensional drawings Mounting on DIN rail

š 

 

(1)
(1)
3
(2) (2)
SITRANS TH300, mounting of transmitter on DIN rail

(3)
14 (0.55) 50,5 (1.99)
33 (1.30)
(4)
 

59,6 (2.35)
0[

  DQG   $X[LOLDU\SRZHUVXSSO\8DX[RXWSXWFXUUHQW,2XW
DQG 3WVHQVRU IRUFRQQHFWLRQVVHH6HQVRU
 FRQQHFWLRQDVVLJQPHQW
7HVW  7HVW   0HDVXUHPHQWRIWKHRXWSXWFXUUHQWZLWKDPXOWLPHWHU DIN rail adapter, dimensions in mm (inch)

  7HVWWHUPLQDO
  0RXQWLQJVFUHZ0[
  /('IRURSHUDWLRQLQGLFDWLRQ
  ,QWHUQDOGLDPHWHURIFHQWHUKROH 

SITRANS TH300, dimensions and pin assignment, dimensions in mm


(inch)

Siemens FI 01 · 2008 3/19


© Siemens AG 2008

SITRANS T measuring instruments for temperature


Transmitters for mounting in sensor head
SITRANS TH300
two-wire system, universal, HART

■ Schematics

Resistance thermometer Resistance Thermocouple

7ZRZLUHV\VWHP 7ZRZLUHV\VWHP &ROGMXQFWLRQFRPSHQVDWLRQ


,QWHUQDOIL[HGYDOXH


7KUHHZLUHV\VWHP 7KUHHZLUHV\VWHP &ROGMXQFWLRQFRPSHQVDWLRQZLWK


H[WHUQDO3WLQWZRZLUHV\VWHP

)RXUZLUHV\VWHP )RXUZLUHV\VWHP &ROGMXQFWLRQFRPSHQVDWLRQZLWK


H[WHUQDO3WLQWKUHHZLUHV\VWHP

*HQHUDWLRQRIDYHUDJH *HQHUDWLRQRIDYHUDJH *HQHUDWLRQRIDYHUDJHYDOXHGLIIHUHQFH


YDOXHGLIIHUHQFH YDOXHGLIIHUHQFH ZLWKLQWHUQDOFROGMXQFWLRQFRPSHQVDWLRQ

3URJUDPPDEOHOLQHUHVLVWDQFHIRUWKHSXUSRVHRIFRUUHFWLRQ

Connection of auxiliary
Voltage measurement Current measurement power supply (Uaux)
8DX[
 

SITRANS TH200/TH300, sensor connection assignment

3/20 Siemens FI 01 · 2008


© Siemens AG 2008

SITRANS T measuring instruments for temperature


Transmitters for mounting in sensor head
SITRANS TH400
fieldbus transmitter

■ Overview ■ Application
• Linearized temperature measurement with resistance ther-
mometers or thermocouple elements
• Differential, mean-value or redundant temperature measure-
ment with resistance thermometers or thermocouple elements
• Linear resistance and bipolar millivolt measurements
• Differential, mean-value or redundant resistance and bipolar
millivolt measurements 3
■ Function
Features
common
• Mounting in connection head, type B, to DIN 43729, or larger
• Polarity-neutral bus connection
• 24-bit analog-digital converter for high resolution
• Electrically isolated
SITRANS TH400 fieldbus transmitters • Intrinsically-safe version for use in potentially explosive areas
• Special characteristic
Versions:
• Sensor redundance
• for FOUNDATION Fieldbus and
Transmitter with PROFIBUS PA communication
• for PROFIBUS PA
• Function blocks: 2 x analog
The SITRANS TH400 temperature transmitter is a small field bus
transmitter for mounting in the connection head of form B. Exten- Transmitter with FOUNDATION Fieldbus communication
sive functionality enables the temperature transmitter to be pre- • Function blocks: 2 x analog and 1 x PID
cisely adapted to the plant’s requirements. Operation is very • Functionality: Basic or LAS
simple in spite of the numerous setting options. Thanks to its uni-
versal concept it can be used in all industries and is easy to in- Mode of operation
tegrate in Totally Integrated Automation applications.
The following function plan explains the mode of operation of the
Temperature transmitters with type of protection "Intrinsic safety" transmitter.
may be installed within potentially explosive atmospheres
(zone 1) or in zone 0. The devices have an EC type test certifi- The only difference between the two versions of the SITRANS
cate and meet the appropriate harmonized European standards TH400 (7NG3214-... and 7NG3215-...) is the type of fieldbus
(ATEX). In addition they offer approvals for USA (FM) and Can- protocol used (PROFIBUS PA or FOUNDATION fieldbus).
ada (CSA).
Installing SITRANS TH400 in temperature sensors turns them
into complete, bus-capable measuring points; compact - and in
a single device.

7UDQVIRUPHU
,QSXW
,QSXW
2SWLRQDOLQSXWV 'LIIHUHQWLDO
0HDQYDOXH
5HVLVWDQFHWKHUPRPHWHU 5HGXQGDQF\
7HUPLQDOWHPSHUDWXUH
7KHUPRFRXSOH (QJLQHHULQJXQLWV
P9VHQVRU 'LDJQRVWLFIXQFWLRQV
7DEOHOLQHDUL]DWLRQ
5HVLVWDQFHEDVHG 3RO\QRPLDOOLQHDUL]DWLRQ
3URFHVVFDOLEUDWLRQ

(OHFWULFDOO\
LVRODWHG

([SRZHUFLUFXLW

,QSXW 
 $'
FRQYHUWHU &38 &RPPXQLFDWLRQ %XVFRQQHFWLRQ


,QSXW


,QWHUQDO ((3520 352),%86SURWRFRO 1*


3W &RPSOH[FRQILJXUDWLRQ RU
&RUUHFWLRQFRHIILFLHQW )RXQGDWLRQ)LHOGEXVSURWRFRO 1*
)DFWRU\VHWWLQJV

SITRANS TH400, function diagram

Siemens FI 01 · 2008 3/21


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitters for mounting in sensor head
SITRANS TH400
fieldbus transmitter
System communication Thermocouple
to IEC 584 Measuring range

2XWSXW • Type B 400 .. +1820 °C (752 ... 3308 °F)


%XV • Type E -100 ... +1000 °C (-148 ... 1832 °F)
WHUPLQDWRU 352),%863$
• Type J -100 ... +1000 °C (-148 ... 1832 °F)
6HJPHQW • Type K -100 ... +1200 °C (-148 ... +2192 °F)

3  
FRXSOHU
• Type N
• Type R
-180 ... +1300 °C (-292 ... 2372 °F)
-50 ... +1760 °C (-58 ... +3200 °F)
s
6,75$16
7+3$ • Type S -50 ... +1760 °C (-58 ... +3200 °F)
• Type T -200 ... +400 °C (-328 ... +752 °F)
%XVFRQQHFWLRQ to DIN 43710
• Type L -200 ... +900 °C (-328 ... +1652 °F)
%XV • Type U -200 ... +600 °C (-328 ... +1112 °F)
WHUPLQDWRU )281'$7,21)LHOGEXV
to ASTM E988-90
6HJPHQW • Type W3 0 ... 2300 °C (32 ... 4172 °F)
FRXSOHU • Type W5 0 ... 2300 °C (32 ... 4172 °F)
 
to IEC 60751
s
6,75$16
7+)) • External cold junction compen- -40 ... +135 °C (-40 ... +275 °F)
sation
Sensor fault detection
%XVFRQQHFWLRQ
• Sensor break detection Yes
SITRANS TH400, communication interface • Sensor short-circuit detection Yes, < 3 mV

■ Technische Daten • Sensor current in the event of


open-circuit monitoring
4 µA

Input mV sensor - voltage input


Analog-to-digital conversion Measuring range -800 ... +800 mV
• Measurement rate < 50 ms Input resistance 10 MΩ
• Resolution 24 Bit Output
Resistance thermometer Filter time (programmable) 0 ... 60 s
Pt25 ... Pt1000 to IEC 60751/JIS C 1604 Update time < 400 ms
• Measuring range -200 ... +850 °C Measuring accuracy
(-328 ... +1562 °F)
Accuracy is defined as the higher value of general values and basic
Ni25 ... Ni1000 to DIN 43760 values.
• Measuring range -60 ... +250 °C General values
(-76 ... +482 °F)
Type of input Absolute accu- Temperature
Cu10 ... Cu1000, α = 0,00427 racy coefficient
• Measuring range -50 ... +200 °C All ≤ ±0,05% of ≤ ±0,002% of
(-58 ... +392 °F) measured value measured
Line resistance per sensor cable Max. 50 Ω value/°C

Sensor current Nominal 0.2 mA Basic values

Sensor fault detection Type of input Basic accuracy Temperature


coefficient
• Sensor break detection Yes
Pt100 and Pt1000 ≤ ±0.1 °C ≤ ±0.002 °C/°C
• Sensor short-circuit detection Yes, < 15 Ω
Ni100 ≤ ±0.15 °C ≤ ±0.002 °C/°C
Resistance-based sensors
Cu10 ≤ ±1.3 °C ≤ ±0.02 °C/°C
Measuring range 0 Ω... 10 kΩ
Resistance-based sensors ≤ ±0.05 Ω ≤ ±0.002 Ω/°C
Line resistance per sensor cable Max. 50 Ω
Voltage source ≤ ±10 µV ≤ ±0.2 µV/°C
Sensor current Nominal 0.2 mA
Thermal element, type: ≤ ±0.5 °C ≤ ±0.01 °C/°C
Sensor fault detection E, J, K, L, N, T, U
• Sensor break detection Yes Thermal element, type: ≤ ±1 °C ≤ ±0.025 °C/°C
B, R, S, W3, W5
• Sensor short-circuit detection Yes, < 15 Ω
Cold junction compensation ≤ ±0.5 °C
Reference conditions
Warming-up time 30 s
Signal-to-noise ratio Min. 60 dB
Calibration condition 20 ... 28 °C (68 ... 82 °F)

3/22 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitters for mounting in sensor head
SITRANS TH400
fieldbus transmitter
Rated conditions Explosion protection: FM for USA
Ambient temperature • for 7NG3214-0AN00 or 7NG3215-0AN00
Permissible ambient temperature -40 ... +85 °C (-40 ... +185 °F) - FM approval FM 3015609
Permissible storage temperature -40 ... +85 °C (-40 ... +185 °F) - Degree of protection • IS Class I, Div 1 Groups A, B, C,
D T4/T5/T6, FISCO
Relative humidity ≤ 98%, with condensation
• IS Class I, Zone 0, AEx ia, IIC
Insulation resistance T4/T5/T6, FISCO
• Test voltage
• Continuous operation
500 V AC for 60 s
50 V AC/75 V DC
• NI Class I, Div. 2, Groups A, B,
C, D T4/T5/T6, FNICO
3
• for 7NG3214-0NN00 or 7NG3215-0NN00
Mechanical testing
- FM approval FM 3015609
• Vibrations (DIN class B) to IEC 60068-2-6 and
IEC 60068-2-64 - Degree of protection NI Class I, Div 2, Groups A, B, C,
4 g/2 ... 100 Hz D T4/T5/T6, FNICO
Electromagnetic compatibility Explosion protection for Canada
EMC noise voltage influence < ±0,1% of span • for 7NG3214-0AN00 or 7NG3215-0AN00
Extended EMC noise immunity: < ±1% of span - CSA approval CSA 1418937
NAMUR NE 21, criterion A, Burst
- Degree of protection • IS Class I, Groups A, B, C, D
EMC 2004/108/EC Emission and EN 61326 T4/T5/T6
Noise Immunity to • Ex ia IIC T4/T5/T6 und
Construction Ex ib [ia] IIC T4/T5/T6
Dimensions Ø 44 x 26.3 mm • for 7NG3214-0NN00 or 7NG3215-0NN00
(Ø 1.73 x 1.04 inch) - CSA approval CSA 1418937
Degree of protection - Degree of protection • Class I, Div 2, Groups A, B, C, D
• Transmitter enclosure IP40 T4/T5/T6
• Terminal IP00 • Ex nA IIC T4/T5/T6 und
Ex ib [ia] IIC T4/T5/T6
Approx. weight 55 g (0.12 lb)
Communication
Auxiliary power supply
Parameterization interface
Power supply
• PROFIBUS PA connection
• Standard DC 9.0 ... 32 V
- Protocol Profile 3.0
• ATEX, FM, UL and CSA DC 9.0 ... 30 V
- Address (for delivery) 126
• In FISCO installation DC 9.0 ... 17.5 V
• FOUNDATION Fieldbus connec-
Power consumption < 11 mA tion
Max. increase in power consump- < 7 mA - Protocol FF Protocoll
tion in the event of a fault
- Functionality Basic or LAS
Certificate and approvals
- Version ITK 4.6
ATEX 94/9/EG to EN 50014, EN 50020,
EN 60079-15, EN 50284, - Function blocks 2 x Analog and 1 x PID
IEC 60079-27 (FISCO) Factory setting for SITRANS TH400 PA and SITRANS TH400 FF
FM to 3600, 3610, 3611 Sensor Pt100 (IEC)
CSA, CAN/CSA to C22.2 No. 142, No. 157, No. 213 Type of connection Three-wire system
CAN/CAS to E79-0, -11, -15 Unit °C
Explosion protection ATEX Failure mode Last valid value
• for 7NG3214-0AN00 or 7NG3215-0AN00 Filter time 0s
- Type of protection: "Approved for • II 1 GD or II 2 (1) GD, only for SITRANS TH400 PA
operation in potentially explosive T65 °C ... T105 °C
atmospheres, Zone 0 and Zone PA address 126
• EEx ia IIC or EEx ib [ia] IIC
1" T4 ... T6 PROFIBUS Ident No. Manufacturer-specific
- EC type test certificate KEMA 06 ATEX 0264 X only for SITRANS TH400 FF
• for 7NG3214-0NN00 or 7NG3215-0NN00 Node address 22
- Type of protection: "Approved for EEx nA [nL] IIC T4 ... T6
operation in potentially explosive
atmospheres, Zone 2"
- EC type test certificate KEMA 06 ATEX 0263 X

Siemens FI 01 · 2008 3/23


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitters for mounting in sensor head
SITRANS TH400
fieldbus transmitter

Selection and Ordering data Order No.

Temperature transmitter
SITRANS TH400
for installation in the sensor head, with
electrical isolation, order instruction
manual separately.

3
• Bus-capable to PROFIBUS PA
- without explosion protection, } 7NG3214-0NN00
EEx n for zone 2
- with explosion protection „intrinsic } 7NG3214-0AN00
safety to ATEX/FM/CSA“
• Bus-capable to FOUNDATION Fieldbus
- without explosion protection, } 7NG3215-0NN00
EEx n for zone 2
- with explosion protection „intrinsic } 7NG3215-0AN00
safety to ATEX/FM/CSA“
Further designs Order code
Please add „-Z“ to Order No. and specifiy
Order code (s) and plain text.
• Customer-specific setting of operating Y011)
data (specify in plain text)
• With test protocol (5 measuring points) C111)
Accessories Order No.
CD for measuring instruments for } A5E00364512
temperature G)
with documentation in German, English,
French, Spanish, Italian, Portuguese and
SIPROM T parameterization software
SIMATIC PDM operating software see chapter 9
For additional PA components see catalog IK PI
} Available ex stock.
1)
For TH400 FF available soon.
G) Subject to export regulations AL: N, ECCN: 5D992B1.

■ Dimensional drawings

s
1 2
0[
3 6
3-W

4 + - 5

š 
 
 
š 

7HUPLQDOV)LHOGEXVFRQQHFWLRQ
7HUPLQDOV6HQVRUFRQQHFWLRQ

SITRANS TH400 dimensions in mm (inches) and connections

3/24 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitters for mounting in sensor head
SITRANS TH400
fieldbus transmitter

■ Schematics

Resistance thermometer Thermocouple Resistance

3
 
7ZRZLUHV\VWHP  ,QWHUQDO 7ZRZLUHV\VWHP
FROGMXQFWLRQFRPSHQVDWLRQ

 

7KUHHZLUHV\VWHP &ROGMXQFWLRQFRPSHQVDWLRQ 7KUHHZLUHV\VWHP


ZLWKH[WHUQDO3WLQWZRZLUHV\VWHP

 

)RXUZLUHV\VWHP &ROGMXQFWLRQFRPSHQVDWLRQ )RXUZLUHV\VWHP


ZLWKH[WHUQDO3WLQWKUHHZLUHV\VWHP

  
  
 

0HDQYDOXHGLIIHUHQWLDORU 0HDQYDOXHGLIIHUHQWLDORU 0HDQYDOXHGLIIHUHQWLDORUUHGXQGDQF\JHQHUDWLRQ


UHGXQGDQF\JHQHUDWLRQ UHGXQGDQF\JHQHUDWLRQZLWKLQWHUQDO UHVLVWRULQWZRZLUHV\VWHP
[WZRZLUHV\VWHP FROGMXQFWLRQFRPSHQVDWLRQ UHVLVWRULQWKUHHZLUHV\VWHP

Voltage measurement


  
  

0HDQYDOXHGLIIHUHQWLDORU 0HDQYDOXHGLIIHUHQWLDORU
UHGXQGDQF\JHQHUDWLRQ UHGXQGDQF\JHQHUDWLRQDQG
VHQVRULQWZRZLUHV\VWHP FROGMXQFWLRQFRPSHQVDWLRQ  
VHQVRULQWKUHHZLUHV\VWHP ZLWKLQWHUQDO3W
LQWZRZLUHV\VWHP 2QHYROWDJHVRXUFH


  
 

3URJUDPPDEOHOLQHUHVLVWDQFHIRUWKHSXUSRVHRIFRUUHFWLRQ
0HDVXUHPHQWRIPHDQYDOXHGLIIHUHQWLDODQG
UHGXQGDQF\ZLWKYROWDJHVRXUFHV

SITRANS TH400, sensor connection assignment

Siemens FI 01 · 2008 3/25


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitters for rail mounting
SITRANS TR200
two-wire system, universal

■ Overview ■ Application
SITRANS TR200 transmitters can be used in all industrial sec-
tors. Their compact design enables simple mounting on stan-
dard DIN rails on-site in protective boxes or in control cabinets.
The following sensors/signal sources can be connected over
their universal input module:

3
• Resistance thermometers (two, three or four-wire system)
• Thermocouple elements
• Resistance-based sensors and DC voltage sources
The output signal is a direct current from 4 to 20 mA in accor-
dance with the sensor characteristic.
Transmitters of the "Intrinsically-safe" or Ex nAL (zone 2) type of
protection can be installed within potentially explosive atmos-
pheres.

■ Function
The SITRANS TR200 is configured over a PC. A USB or RS 232
Ultra flexible - with the universal SITRANS TR200 transmitter modem is linked to the output terminals for this purpose. The
configuration data can now be edited using the SIPROM T soft-
• Two-wire devices for 4 to 20 mA ware tool. The configuration data are then permanently stored in
• Enclosure for rail mounting the non-volatile memory (EEPROM).
• Universal input for virtually any type of temperature sensor Once the sensors and power supply have been correctly con-
• Configurable over PC nected, the transmitter outputs a temperature-linear output sig-
nal and the diagnostics LED displays a green light. In the case
■ Benefits of a sensor short-circuit, the LED flashes red, an internal device
fault is indicated by a steady red light.
• Compact design
The test socket can be used to connect an ammeter at any time
• Electrically isolated for monitoring purposes and plausibility checks. The output cur-
• Test sockets for multimeters rent can be read without any interruption, or even without ope-
ning the current loop.
• Diagnostics LED (green/red)
• Sensor monitoring
open circuits and short-circuits
• Self-monitoring
• Configuration status stored in EEPROM
• Expanded diagnostic functions, such as slave pointer, opera-
ting hours counter, etc.
• Special characteristic
• Electromagnetic compatibility to EN 61326 and NE21

6,75$167575

7
1
7HVW
6HQVRU
7&57' 2 3

$ '
µ& µ& P$ 8DX[,RXW
' $
4
5 6

 6HQVRUVXFKDVUHVLVWDQFHWKHUPRPHWHU  (OHFWULFDOLVRODWLRQ 8DX[ $X[LOLDU\SRZHUVXSSO\


 WKHUPRFRXSOHUHVLVWDQFHEDVHG  0LFURFRQWUROOHUSULPDU\FLUFXLW ,RXW 2XWSXWFXUUHQW
 VHQVRUP9VHQVRU  'LJLWDODQDORJFRQYHUWHU 7HVW 7HVWWHUPLQDOVIRUWHPSRUDU\
 $QDORJGLJLWDOFRQYHUWHU  /('  FRQQHFWLRQRIDQDPSHUHPHWHU
 0LFURFRQWUROOHUVHFRQGDU\FLUFXLW 


SITRANS TR200 function diagram

3/26 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitters for rail mounting
SITRANS TR200
two-wire system, universal

■ Technical specifications Response time T63 ≤ 250 ms for 1 sensor with open-
circuit monitoring
Input Open-circuit monitoring Always active (cannot be disab-
Resistance thermometer led)
Measured variable Temperature Short-circuit monitoring Can be switched off (value is
adjustable)
Sensor type
Measuring range Parameterizable, max.
0 ... 2200 Ω (see Table "Digital
3
• to IEC 60751 Pt25 to Pt1000
• to JIS C 1604; a=0.00392 K-1 Pt25 to Pt1000 measuring errors")

• to IEC 60751 Ni25 to Ni1000 Min. measured span 5 ... 25 Ω (see Table "Digital
measuring errors")
• Special type Over special characteristic (max.
30 points) Characteristic curve Resistance-linear or special cha-
racteristic
Sensor factor 0.25 ... 10 (adaptation of the
basic type, e.g. Pt100 to version Thermocouple elements
Pt25 ... Pt1000) Measured variable Temperature
Units °C or °F Sensor type (thermocouples)
Connection • Type B Pt30Rh-Pt6Rh to DIN IEC 584
• Standard connection 1 resistance thermometer (RTD) • Type C W5%-Re to ASTM 988
in 2-wire, 3-wire or 4-wire system
• Type D W3%-Re to ASTM 988
• Generation of average value 2 identical resistance thermome-
ters in 2-wire system for genera- • Type E NiCr-CuNi to DIN IEC 584
tion of average temperature • Type J Fe-CuNi to DIN IEC 584
• Generation of difference 2 identical resistance thermome- • Type K NiCr-Ni to DIN IEC 584
ters (RTD) in 2-wire system
(RTD 1 – RTD 2 or RTD 2 – RTD 1) • Type L Fe-CuNi to DIN 43710
Interface • Type N NiCrSi-NiSi to DIN IEC 584
• Two-wire system Parameterizable line resistance • Type R Pt13Rh-Pt to DIN IEC 584
≤ 100 Ω (loop resistance)
• Type S Pt10Rh-Pt to DIN IEC 584
• Three-wire system No balancing required
• Type T Cu-CuNi to DIN IEC 584
• Four-wire system No balancing required
• Type U Cu-CuNi to DIN 43710
Sensor current ≤ 0.45 mA
Units °C or °F
Response time T63 ≤ 250 ms for 1 sensor with open-
circuit monitoring Connection

Open-circuit monitoring Always active (cannot be disab- • Standard connection 1 thermocouple element (TC)
led) • Generation of average value 2 identical thermocouple ele-
Short-circuit monitoring Can be switched off (value is ments (TC)
adjustable) • Generation of difference 2 identical thermocouple ele-
Measuring range Parameterizable (see table "Digi- ments (TC) TC1 – TC2 or
tal measuring errors") TC2 – TC1

Min. measured span 10 °C (18 °F) Response time T63 ≤ 250 ms for 1 sensor with open-
circuit monitoring
Characteristic curve Temperature-linear or special
characteristic Open-circuit monitoring Can be switched off

Resistance-based sensors Cold junction compensation

Measured variable Actual resistance • Internal With integrated Pt100 resistance


thermometer
Sensor type Resistance-based, potentiome-
ters • External With external Pt100 IEC 60571
(2-wire or 3-wire connection)
Units Ω
• External fixed Cold junction temperature can be
Connection set as fixed value
• Normal connection 1 resistance-based sensor (R) in Measuring range Parameterizable (see table "Digi-
2-wire, 3-wire or 4-wire system tal measuring errors")
• Generation of average value 2 resistance-based sensors in Min. measured span Min. 50 ... 100 °C (90 ... 180 °F)
2-wire system for generation of (see table "Digital measuring
average value errors")
• Generation of difference 2 resistance thermometers in Characteristic curve Temperature-linear or special
2-wire system characteristic
(R1 – R2 or R2 – R1)
mV sensor
Interface
Measured variable DC voltage
• Two-wire system Parameterizable line resistance
≤ 100 Ω (loop resistance) Sensor type DC voltage source (DC voltage
source possible over an exter-
• Three-wire system No balancing required nally connected resistor)
• Four-wire system No balancing required Units mV
Sensor current ≤ 0.45 mA Response time T63 ≤ 250 ms for 1 sensor with open-
circuit monitoring

Siemens FI 01 · 2008 3/27


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitters for rail mounting
SITRANS TR200
two-wire system, universal
Open-circuit monitoring Can be switched off Certificates and approvals
Measuring range Parameterizable max. Explosion protection ATEX
-100 … 1100 mV
EC type test certificate PTB 07 ATEX 2032X
Min. measured span 2 mV or 20 mV
• "Intrinsic safety" type of protection II 2(1)GD EEx ia/ib/ic IIC T6/T4
Overload capability of the input -1.5 ... +3.5 V DC
• Type of protection, "equipment has II 3G EEx nL IIC T6/T4
Input resistance ≥ 1 MΩ limited energy"

3 Characteristic curve Voltage-linear or special charac-


teristic
• Type of protection, "equipment is
non-arcing"
II 3G EEx nA T6/T4

Output • Type of protection, "associated II (1)GD [EEx ia] IIC T6/T4


equipment"
Output signal 4 ... 20 mA, 2-wire
Software requirements for
Auxiliary power supply 11 ... 35 V DC (to 30 V with Ex) SIPROM T
Max. load (Uaux – 11 V)/0.023 A PC operating system Windows ME, 2000 and XP; also
Overrange 3.6 ... 23 mA, infinitely adjustable Windows 95, 98 and 98 SE, but
(default range: only in connection with RS 232
3.84 mA ... 20.50 mA) modem.
Error signal (e.g. in the event of sen- 3.6 ... 23 mA, infinitely adjustable Factory setting:
sor breakage) (default value: 22.8 mA)
• Pt100 (IEC 751) with three-wire circuit
Sample cycle 0.25 s
• Measuring range: 0 ... 100 °C (32 ... 212 °F)
Damping Software filter 1st order 0 ... 30 s
(parameterizable)
• Fault current: 22.8 mA
• Sensor offset: 0 °C (0 °F)
Protection Against reversed polarity
• Damping 0.0 s
Electrically isolated Input against output 2.12 kV DC
(1.5 kV AC)
Measuring accuracy Digital measuring errors
Digital measuring errors See Table "Digital measuring
errors" Resistance thermometer
Reference conditions Input Measuring range Min. measu- Digital
red span accuracy
• Auxiliary power supply 24 V ± 1 %
°C (°F) °C (°F) °C (°F)
• Load 500 Ω
to IEC 60751
• Ambient temperature 23 °C
Pt25 -200 ... +850 10 (18) 0.2 (0.36)
• Warming-up time > 5 min (-328 ... +1562)
Error in the analog output (digi- < 0.1 % of span Pt50 -200 ... +850 10 (18) 0.15 (0.27)
tal/analog converter) (-328 ... +1562)
Error due to internal cold junction < 0.55 °C (0.9 °F) Pt100 ... Pt200 -200 ... +850 10 (18) 0.1 (0.18)
(-328 ... +1562)
Temperature effect < 0.1 % of max. span/10 °C
(18 °F) Pt500 -200 ... +850 10 (18) 0.15 (0.27)
(-328 ... +1562)
Power supply effect < 0.005 % of span/V
Pt1000 -200 ... +350 10 (18) 0.15 (0.27)
Effect of load impedance < 0.012 % of span/100 Ω
(-328 ... +662)
Long-term drift
to JIS C1604-81
• in the first month < 0.02 % of max. span
Pt25 -200 ... +649 10 (18) 0.2 (0.36)
• after one year < 0.03 % of max. span (-328 ... +1200)
• after 5 years < 0.04 % of max. span Pt50 -200 ... +649 10 (18) 0.15 (0.27)
(-328 ... +1200)
Rated conditions
Pt100 ... Pt200 -200 ... +649 10 (18) 0.1 (0.18)
Ambient conditions (-328 ... +1200)
Ambient temperature range -40 ... +85 °C (-40 ... +185 °F) Pt500 -200 ... +649 10 (18) 0.15 (0.27)
Storage temperature range -40 ... +85 °C (-40 ... +185 °F) (-328 ... +1200)
Relative humidity < 98 %, with condensation Pt1000 -200 ... +350 10 (18) 0.15 (0.27)
(-328 ... +662)
Electromagnetic compatibility According to EN 61326 and
NAMUR NE21 Ni 25 to Ni1000 -60 ... +250 10 (18) 0.1 (0.18)
(-76 ... +482)
Construction
Material Plastic, electronic module potted
Weight 122 g
Dimensions See "Dimensional drawings"
Cross-section of cables Max. 2.5 mm² (AWG 13)
Degree of protection to IEC 60529
• Enclosure IP20

3/28 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitters for rail mounting
SITRANS TR200
two-wire system, universal
Resistance-based sensors
■ Selection and Ordering Data Order No.
Input Measuring Min. measured Digital accu-
range span racy Temperature transmitter
SITRANS TR200
Ω Ω Ω
for mounting on a standard DIN
Resistance 0 ... 390 5 0.05 rail, two-wire system, 4 to 20 mA,
Resistance 0 ... 2200 25 0.25 programmable, with electrical iso-
lation, with documentation on CD

Thermocouple elements
• Without explosion protection }
D)
7NG3032-0JN00
3
• With explosion protection to } 7NG3032-1JN00
Input Measuring Min. measured Digital ATEX; type of protection: "Intrin- D)
range span accuracy sic safety" EEx ia/ib, non-arcing
°C °C (°F) °C (°F) (nA), limited energy (nL)
(°F) Further designs Order code
Type B 0 ... 1820 100 (180) 2 (3.60)1) Please add "-Z" to Order No. and
(32 ... 3308) specify Order code(s)
Type C (W5) 0 ... 2300 100 (180) 2 (3.60) Customer-specific setting of oper- Y01
(32 ... 4172) ating data (specify operating data
in plain text)
Type D (W3) 0 ... 2300 100 (180) 1 (1.80)2)
(32 ... 4172) with test protocol (5 measuring C11
points)
Type E -200 ... +1000 50 (90) 1 (1.80)
(-328 ... +1832) Accessories Order No.
Type J -210 ... +1200 50 (90) 1 (1.80) Modem for SITRANS TH100,
(-346 ... +2192) TH200 and TR200 incl. SIPROM
T parameterization software
Type K -230 ... +1370 50 (90) 1 (1.80)
(-382 ... +2498) • With USB connection } 7NG3092-8KU
C)
Type L -200 ... +900 50 (90) 1 (1.80)
(-328 ... +1652) • With RS 232 connection } 7NG3092-8KM
C)
Type N -200 ... +1300 50 (90) 1 (1.80)
(-328 ... +2372) CD for measuring instruments } A5E00364512
for temperature G)
Type R -50 ... +1760 100 (180) 2 (3.60)
(-58 ... +3200) with documentation in German,
English, French, Spanish, Italian,
Type S -50 ... +1760 100 (180) 2 (3.60) Portuguese and SIPROM T
(-58 ... +3200) parameterization software
Type T -200 ... +400 40 (72) 1 (1.80) } Available ex stock.
(-328 ... +752)
C) Subject to export regulations AL: N, ECCN: EAR99.
Type U -200 ... +600 50 (90) 2 (3.60) D) Subject to export regulations AL: N, ECCN: EAR99H.
(-328 ... +1112) G) Subject to export regulations AL: N, ECCN: 5D992B1.
1) Power supply units see "SITRANS I supply units and isolation amplifiers".
The digital accuracy in the range 0 to 300 °C (32 to 572 °F) is
3 °C (5.4 °F).
2) The digital accuracy in the range 1750 to 2300 °C (3182 to
4172 °F) is 2 °C (3.6 °F).

mV sensor
Input Measuring Min. measured Digital
range span accuracy
mV mV µV
mV sensor -10 ... +70 2 40
mV sensor -100 ... +1100 20 400

The digital accuracy is the accuracy after the analog/digital con-


version including linearization and calculation of the measured
value.
An additional error is generated in the output current 4 to 20 mA
as a result of the digital/analog conversion of 0.1% of the set
span (digital-analog error).
The total error under reference conditions at the analog output is
the sum from the digital error and the digital-analog error (poss.
with the addition of cold junction errors in the case of thermo-
couple measurements).

Siemens FI 01 · 2008 3/29


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitters for rail mounting
SITRANS TR200
two-wire system, universal

■ Dimensional drawings ■ Schematics



3
4..20mA

SIEMENS

8
4

SITRANS
TR200

7
3

1 2 3 4

6
2
TEST

TEST 4..20mA

5
1

SIEMENS
SITRANS
 TR200


Made in Croatia

3W RTD CJC
with external
2W RTD CJC
with external
U sensor
int./fixed CJC
5 6 7 8
TC sensor TC sensor TC sens. with
+

+ + +

8 7 6 5 8 7 6 5 8 7 6 5

4W sensor 3W sensor 2W sensor




RTD or R RTD or R RTD or R




8 7 6 5 8 7 6 5 8 7 6 5

CJC 2W sensor & test


with int./fixed 2x RTD or R 4..20mA loop
2x TC sensor
S1
4 3 2 1
S2 S1
+ + S2
A
+ -
8 7 6 5

$VVVLJQPHQWV
8 7 6 5 TEST +UB -

Connection Plan
s
  DQG   7HVWWHUPLQDOV 7HVW IRUPHDVXUHPHQWRIWKHRXWSXW
 FXUUHQWZLWKDPXOWLPHWHU
  DQG   3RZHUVXSSO\8DX[2XWSXWFXUUHQW,RXW
DQG 6HQVRUDVVLJQPHQWVHHVFKHPDWLFV

SITRANS TR200, dimensions in mm (inch) SITRANS pin assignment, TR200

3/30 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitters for rail mounting
SITRANS TR200
two-wire system, universal

Resistance thermometer Resistance Thermocouple


 
3
5 7&
 
57'

7ZRZLUHV\VWHP 7ZRZLUHV\VWHP &ROGMXQFWLRQFRPSHQVDWLRQ


,QWHUQDOIL[HGYDOXH


 
7&
 

57' 5 57'

7KUHHZLUHV\VWHP 7KUHHZLUHV\VWHP &ROGMXQFWLRQFRPSHQVDWLRQZLWK


H[WHUQDO3WLQWZRZLUHV\VWHP


 
7&
 

57' 5
57'
)RXUZLUHV\VWHP )RXUZLUHV\VWHP &ROGMXQFWLRQFRPSHQVDWLRQZLWK
H[WHUQDO3WLQWKUHHZLUHV\VWHP

 
5 
57'
7& 7&
   
57' 5

*HQHUDWLRQRIDYHUDJH *HQHUDWLRQRIDYHUDJH *HQHUDWLRQRIDYHUDJHYDOXHGLIIHUHQFH


YDOXHGLIIHUHQFH YDOXHGLIIHUHQFH ZLWKLQWHUQDOFROGMXQFWLRQFRPSHQVDWLRQ

3URJUDPPDEOHOLQHUHVLVWDQFHIRUWKHSXUSRVHRIFRUUHFWLRQ

Voltage measurement Current measurement Test terminals Power supply/


4 ... 20 mA (Uaux)

8DX[
 $ 
 
 

 
  5
8  
,

SITRANS TR200, sensor connection assignment

Siemens FI 01 · 2008 3/31


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitters for rail mounting
SITRANS TR300
two-wire system, universal, HART

■ Overview • Special characteristic


• Electromagnetic compatibility to EN 61326 and NE21

■ Application
SITRANS TR300 transmitters can be used in all industrial sec-
tors. Their compact design enables simple mounting on stan-
dard DIN rails on-site in protective boxes or in control cabinets.
3 The following sensors/signal sources can be connected over
their universal input module:
• Resistance thermometers (two, three or four-wire system)
• Thermocouple elements
• Resistance-based sensors and DC voltage sources
The output signal is a direct current from 4 to 20 mA in accor-
dance with the sensor characteristic, superimposed by the digi-
tal HART signal.
Transmitters of the "Intrinsically-safe" or Ex nAL (zone 2) type of
protection can be installed within potentially explosive atmos-
pheres.
"HART" to beat - the universal SITRANS TR300 transmitter
• Two-wire devices for 4 to 20 mA, HART ■ Function
• Device for rail mounting The SITRANS TR300 is configured over HART. This can be done
• Universal input for virtually any type of temperature sensor using a handheld communicator or even more conveniently with
a HART modem and the SIMATIC PDM parameterization soft-
• Configurable over HART ware. The configuration data are then permanently stored in the
non-volatile memory (EEPROM).
■ Benefits Once the sensors and power supply have been correctly con-
• Compact design nected, the transmitter outputs a temperature-linear output sig-
• Electrically isolated nal and the diagnostics LED displays a green light. In the case
of a sensor short-circuit, the LED flashes red, an internal device
• Test sockets for multimeters fault is indicated by a steady red light.
• Diagnostics LED (green/red)
The test socket can be used to connect an ammeter at any time
• Sensor monitoring for monitoring purposes and plausibility checks. The output cur-
open circuits and short-circuits rent can be read without any interruption, or even without ope-
• Self-monitoring ning the current loop.
• Configuration status stored in EEPROM
• Expanded diagnostic functions, such as slave pointer, opera-
ting hours counter etc.

6,75$167575

7
1
7HVW
6HQVRU
7&57' 2 3

$ '
µ& µ& P$ 8DX[,RXW
' $
4
5 6

 6HQVRUVXFKDVUHVLVWDQFHWKHUPRPHWHU
6HQVRUZLH:LGHUVWDQGVWKHUPRPHWHU  *DOYDQLVFKH7UHQQXQJ
(OHFWULFDOLVRODWLRQ 8DX[8 DX[ +LOIVHQHUJLH
 $X[LOLDU\SRZHUVXSSO\
 7KHUPRHOHPHQW:LGHUVWDQGVJHEHU
WKHUPRFRXSOHUHVLVWDQFHEDVHG  0LNURFRQWUROOHU3ULP¦UVHLWH
0LFURFRQWUROOHUSULPDU\FLUFXLW ,RXW ,RXW$XVJDQJVVWURP
2XWSXWFXUUHQW
 0LOOLYROWJHEHU
VHQVRUP9VHQVRU  'LJLWDO$QDORJ:DQGOHU
'LJLWDODQDORJFRQYHUWHU 7HVW 7HVW3U¾INOHPPHQ]XPNXU]]HLWLJHQ
7HVWWHUPLQDOVIRUWHPSRUDU\
 $QDORJ'LJLWDO:DQGOHU
$QDORJGLJLWDOFRQYHUWHU  /('
/('   $QVFKOXVVHLQHV6WURPPHVVJHU¦WV
FRQQHFWLRQRIDQDPSHUHPHWHU
 0LNURFRQWUROOHU6HNXQG¦UVHLWH
0LFURFRQWUROOHUVHFRQGDU\FLUFXLW 


SITRANS TR300 function diagram

3/32 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitters for rail mounting
SITRANS TR300
two-wire system, universal, HART

■ Technical specifications Open-circuit monitoring Can be switched off


Short-circuit monitoring Can be switched off (value is
Input adjustable)
Resistance thermometer Range Parameterizable, max.
Measured variable Temperature 0 ... 2200 Ω (see Table "Digital
measuring errors")
Sensor type
Min. measured span 5 Ω ... 25 Ω (see Table "Digital

3
• to IEC 60751 Pt25 to Pt1000 measuring errors")
• to JIS C 1604; a=0.00392 K-1 Pt25 to Pt1000 Characteristic Resistance-linear or special cha-
• to IEC 60751 Ni25 to Ni1000 racteristic

• Special type over special characteristic (max. Thermocouple elements


30 points) Measured variable Temperature
Sensor factor 0.25 ... 10 (adaptation of the Sensor type (thermocouples)
basic type, e.g. Pt100 to version
Pt25 ... Pt1000) • Type B Pt30Rh-Pt6Rh to DIN IEC 584
Units °C or °F • Type C W5%-Re to ASTM 988
Connection • Type D W3%-Re to ASTM 988
• Standard connection 1 resistance thermometer (RTD) • Type E NiCr-CuNi to DIN IEC 584
in 2-wire, 3-wire or 4-wire system • Type J Fe-CuNi to DIN IEC 584
• Generation of average value 2 identical resistance thermome- • Type K NiCr-Ni to DIN IEC 584
ters in 2-wire system for genera-
tion of average temperature • Type L Fe-CuNi to DIN 43710
• Generation of difference 2 identical resistance thermome- • Type N NiCrSi-NiSi to DIN IEC 584
ters (RTD) in 2-wire system
(RTD 1 – RTD 2 or RTD 2 – RTD 1) • Type R Pt13Rh-Pt to DIN IEC 584

Interface • Type S Pt10Rh-Pt to DIN IEC 584

• Two-wire system Parameterizable line resistance • Type T Cu-CuNi to DIN IEC 584
≤ 100 Ω (loop resistance) • Type U Cu-CuNi to DIN 43710
• Three-wire system No balancing required Units °C or °F
• Four-wire system No balancing required Connection
Sensor current ≤ 0.45 mA • Standard connection 1 thermocouple (TC)
Response time T63 ≤ 250 ms for 1 sensor with open- • Generation of average value 2 thermocouples (TC)
circuit monitoring
• Generation of difference 2 thermocouples (TC) TC1 – TC2
Open-circuit monitoring can be switched off or TC2 – TC1
Short-circuit monitoring can be switched off (value is Response time ≤ 250 ms for 1 sensor with open-
adjustable) circuit monitoring
Range Parameterizable (see table "Digi- Open-circuit monitoring Can be switched off
tal measuring errors")
Cold junction compensation
Min. measured span 10 °C (18 °F)
• Internal With integrated Pt100 resistance
Characteristic Temperature-linear or special thermometer
characteristic
• External With external Pt100 IEC 60571
Resistance-based sensors (2-wire or 3-wire connection)
Measured variable Actual resistance • External fixed Cold junction temperature can be
Sensor type Resistance-based, potentiome- set as fixed value
ters Range Parameterizable (see table "Digi-
Units Ω tal measuring errors")

Connection Min. measured span Min. 50 ... 199 °C (90 ... 180 °F)
(see table "Digital measuring
• Normal connection 1 resistance-based sensor (R) in errors")
2-wire, 3-wire or 4-wire system
Characteristic Temperature-linear or special
• Generation of average value 2 resistance-based sensors in characteristic
2-wire system for generation of
average value mV Sensor

• Generation of difference 2 resistance thermometers in Measured variable DC voltage


2-wire system Sensor type DC voltage source (DC voltage
(R1 – R2 or R2 – R1) source possible over an exter-
Interface nally connected resistor)

• Two-wire system Parameterizable line resistance Units mV


≤ 100 Ω (loop resistance) Response time T63 ≤ 250 ms for 1 sensor with open-
• Three-wire system No balancing required circuit monitoring

• Four-wire system No balancing required Open-circuit monitoring Can be switched off

Sensor current ≤ 0.45 mA Short-circuit monitoring Can be switched off (value is


adjustable)
Response time T63 ≤ 250 ms for 1 sensor with open-
circuit monitoring

Siemens FI 01 · 2008 3/33


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitters for rail mounting
SITRANS TR300
two-wire system, universal, HART
Range -10 ... 70 mV Certificate and approvals
-100 ... 1100 mV
Explosion protection ATEX
Min. measured span 2 mV or 20 mV
EC type test certificate PTB 07 ATEX 2032X
Overload capacity of the input -1.5 ... +3.5 V DC
• "Intrinsically-safe" type of protec- II 2(1)GD EEx ia/ib/ic IIC T6/T4
Input resistance ≥ 1 MΩ tion
Characteristic Voltage-linear or special charac- • "Operating equipment that is non- II 3G EEx nL IIC T6/T4

3
teristic ignitable and has limited energy"
type of protection
Output
• Type of protection, "equipment is II 3G EEx nA T6/T4
Output signal 4 ... 20 mA, 2-wire non-arcing"
Communication acc. to HART
Rev. 5.9 • Type of protection, "associated II (1)GD [EEx ia] IIC T6/T4
equipment"
Power supply 11 ... 35 V DC (to 30 V with EEx)
Explosion protection to FM for USA
Max. load (Uaux – 11 V)/0.023 A and Canada (cFMUS)
Overrange 3.6 ... 23 mA, continuously adjus- • FM approval PID 3024169
table (default range:
3.84 ... 20.50 mA) • Degree of protection IS Cl I, II, III, Div 1, GP ABCDEFG
T4/T5/T6
Error signal (e.g. in the event of sen- 3.6 ... 23 mA, continuously adjus- IS Cl I, ZN 0, 1 AEx ia IIC T4/T5/T6
sor breakage) table (default value: 22.8 mA) NI Cl I, II, III, Div 2, GP ABCDFG
T4/T5/T6
Sample cycle 0.25 s nominal Cl I, ZN2, GP IIC T4/T5/T6
Damping Software filter 1st order 0 ... 30 s Software requirements for
(parameterizable) SIPROM T
Protection Against reversed polarity PC operating system: Windows ME, 2000 and XP; also
Electrically isolated Input against output (1 kVeff) Windows 95, 98 and 98 SE, but
only in connection with RS 232
Measuring accuracy modem.
Digital measuring errors See Table "Digital measuring
errors" Factory setting:
Reference conditions • Pt100 (IEC 751) with three-wire circuit
• Power supply 24 V ± 1%
• Measuring range: 0 ... 100 °C (32 ... 212 °F)
• Fault current: 22.8 mA
• Load 500 Ω
• Sensor offset: 0 °C (0 °F)
• Storage temperature 23 °C
• Damping 0.0 s
• Warming-up time > 5 min
Error in the analog ouput (digi- < 0.1 % of span
tal/analog converter) Digital measuring errors
Error due to internal cold junction < 0.5 °C (0.9 °F) Resistance thermometer
Temperature effect < 0.1% der max. span/10°C
input Range Min. measu- Digital accu-
(18 °F)
red span racy
Power supply effect < 0.005% of span/V
°C (°F) °C (°F) °C (°F)
Effect of load impedance < 0.012% of span/100 Ω
to IEC 60751
Long-term drift
Pt25 -200 ... +850 10 (18) 0.2 (0.36)
• in the first month < 0.02% of max. span (-328 ... +1562)
• after one year < 0.03% of max. span Pt50 -200 ... +850 10 (18) 0.15 (0.27)
(-328 ... +1562)
• after 5 years < 0.04% of max. span
Pt100 ... Pt200 -200 ... +850 10 (18) 0.1 (0.18)
Rated conditions (-328 ... +1562)
Ambient temperature Pt500 -200 ... +850 10 (18) 0.15 (0.27)
Storage temperature -40 ... +85 °C (-40 ... 185 °F) (-328 ... +1562)
Functional temperature -40 ... +85 °C (-40 ... 185 °F) Pt1000 -200 ... +350 10 (18) 0.15 (0.27)
(-328 ... +662)
Relative humidity < 98%, with condensation
to JIS C1604-81
Electromagnetic compatibility acc. to DIN EN 61326 and NE21
Pt25 -200 ... +649 10 (18) 0.2 (0.36)
Construction (-328 ... +1200)
Material Molded plastic Pt50 -200 ... +649 10 (18) 0.15 (0.27)
Approx. weight 50 g (0.11 lb) (-328 ... +1200)

Dimensions See "Dimension drawings" Pt100 ... Pt200 -200 ... +649 10 (18) 0.1 (0.18)
(-328 ... +1200)
Cross-section of cables Max. 2.5 mm² (AWG 13)
Pt500 -200 ... +649 10 (18) 0.15 (0.27)
Degree of protection to EN 60529 (-328 ... +1200)
• Enclosure IP40 Pt1000 -200 ... +350 10 (18) 0.15 (0.27)
(-328 ... +662)
• Terminals IP00
Ni 25 to Ni1000 -60 ... +250 10 (18) 0.1 (0.18)
(-76 ... +482)

3/34 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitters for rail mounting
SITRANS TR300
two-wire system, universal, HART
Resistance-based sensors
■ Selection and Ordering Data Order No.
Input Range Min. measured Digital accu-
span racy Temperature transmitter
SITRANS TR300
Ω Ω Ω
for mounting on a standard DIN rail,
Resistance 0 ... 390 5 0.05 two-wire system, 4 ... 20 mA, HART,
Resistance 0 ... 2200 25 0.25 with electrical isolation, with docu-
mentation on CD

Thermocouple elements
• Without explosion protection }
D)
7NG3033-0JN00
3
• With explosion protection to } 7NG3033-1JN00
input Range Min. measured Digital accu- ATEX; type of protection: "Intrin- D)
span racy sic safety" EEx ia/ib, non-arcing
°C °C (°F) °C (°F) (nA), limited energy (nL)
(°F) Further designs Order code
Type B 0 ... 1820 100 (180) 2 (5.4) Please add "-Z" to Order No. and
(32 ... 3308) specify Order code(s)
Type C (W5) 0 ... 2300 100 (180) 2 (3.6) Customer-specific setting of oper- Y01
(32 ... 4172) ating data (specify operating data
in plain text)
Type D (W3) 0 ... 2300 100 (180) 2 (3.6)
(32 ... 4172) with test protocol (5 measuring C11
points)
Type E -200 ... +1000 50 (90) 1 (1.8)
(-328 ... +1832) Accessories Order No.
Type J -210 ... +1200 50 (90) 1 (1.8) CD for measuring instruments } A5E00364512
(-346 ... +2192) for temperature G)

Type K -230 ... +1370 50 (90) 1 (1.8) with documentation in German,


(-382 ... +2498) English, French, Spanish, Italian,
Portuguese and SIPROM T
Type L -200 ... +900 50 (90) 1 (1.8)
parameterization software
(-328 ... +1652)
HART modem
Type N -200 ... +1300 50 (90) 1 (1.8)
(-328 ... +2372) • With RS 232 connection } 7MF4997-1DA
D)
Type R -50 ... +1760 100 (180) 2 (3.6)
(-58 ... +3200) • With USB connection } 7MF4997-1DB
D)
Type S -50 ... +1760 100 (180) 2 (3.6)
(-58 ... +3200) SIMATIC PDM operating software See Section 9

Type T -200 ... +400 40 (72) 1 (1.8) } Available ex stock.


(-328 ... +752) D) Subject to export regulations AL: N, ECCN: EAR99H.
G) Subject to export regulations AL: N, ECCN: 5D992B1.
Type U -200 ... +600 50 (90) 2 (3.6)
(-328 ... +1112) Power supply units see "SITRANS I supply units and isolation amplifiers".

mV Sensor
input Range Min. measured Digital accu-
span racy
mV mV µV
mV Sensor -10 ... +70 2 40
mV Sensor -100 ... +1100 20 400

The digital accuracy is the accuracy after the analog/digital con-


version including linearization and calculation of the measured
value.
An additional error is generated in the output current 4 to 20 mA
as a result of the digital/analog conversion of 0.1% of the set
span (digital-analog error).
The total error under reference conditions at the analog output is
the sum from the digital error and the digital-analog error (poss.
with the addition of cold junction errors in the case of thermo-
couple measurements).

Siemens FI 01 · 2008 3/35


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitters for rail mounting
SITRANS TR300
two-wire system, universal, HART

■ Dimensional drawings ■ Schematics



3
4..20mA

SIEMENS

8
4

SITRANS
TR300

7
3

1 2 3 4

6
2
TEST

TEST 4..20mA

5
1

SIEMENS
SITRANS
 TR300


Made in Croatia

3W RTD CJC
with external
2W RTD CJC
with external
U sensor
int./fixed CJC
5 6 7 8
TC sensor TC sensor TC sens. with
+

+ + +

8 7 6 5 8 7 6 5 8 7 6 5

4W sensor 3W sensor 2W sensor




RTD or R RTD or R RTD or R




8 7 6 5 8 7 6 5 8 7 6 5

CJC 2W sensor & test


with int./fixed 2x RTD or R 4..20mA loop
2x TC sensor
S1
4 3 2 1
S2 S1
+ + S2
A
+ -
8 7 6 5 8 7 6 5 TEST +UB -

Connection Plan $VVLJQPHQWV


s
  DQG   7HVWWHUPLQDOV 7HVW IRUPHDVXUHPHQWRIWKHRXWSXW
 FXUUHQWZLWKDPXOWLPHWHU
  DQG   3RZHUVXSSO\8DX[2XWSXWFXUUHQW,RXW
DQG 6HQVRUDVVLJQPHQWVHHVFKHPDWLFV

SITRANS TR300, dimensions in mm (inch) SITRANS TR300, pin assignment

3/36 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitters for rail mounting
SITRANS TR300
two-wire system, universal, HART

Resistance thermometer Resistance Thermocouple


 
3
5 7&
 
57'

7ZRZLUHV\VWHP 7ZRZLUHV\VWHP &ROGMXQFWLRQFRPSHQVDWLRQ


,QWHUQDOIL[HGYDOXH


 
7&
 

57' 5 57'

7KUHHZLUHV\VWHP 7KUHHZLUHV\VWHP &ROGMXQFWLRQFRPSHQVDWLRQZLWK


H[WHUQDO3WLQWZRZLUHV\VWHP


 
7&
 

57' 5
57'
)RXUZLUHV\VWHP )RXUZLUHV\VWHP &ROGMXQFWLRQFRPSHQVDWLRQZLWK
H[WHUQDO3WLQWKUHHZLUHV\VWHP

 
5 
57'
7& 7&
   
57' 5

*HQHUDWLRQRIDYHUDJH *HQHUDWLRQRIDYHUDJH *HQHUDWLRQRIDYHUDJHYDOXHGLIIHUHQFH


YDOXHGLIIHUHQFH YDOXHGLIIHUHQFH ZLWKLQWHUQDOFROGMXQFWLRQFRPSHQVDWLRQ

3URJUDPPDEOHOLQHUHVLVWDQFHIRUWKHSXUSRVHRIFRUUHFWLRQ

Voltage measurement Current measurement Test terminals Power supply/


4 ... 20 mA (Uaux)

8DX[
 $ 
 
 

 
  5
8  
,

SITRANS TR300, sensor connection assignment

Siemens FI 01 · 2008 3/37


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitters for rail mounting
SITRANS TW
four-wire system, universal, HART

■ Overview • Temperature-linear characteristic can be selected for all tem-


perature sensors
• Automatic correction of zero and span
• Monitoring of sensor and cable for open-circuit and short-cir-
cuit
• Sensor fault and/or limit can be output via an optional sensor
fault/limit monitor

3 • Hardware write protection for HART communication


• Diagnostic functions
• Slave pointer functions
Mode of operation

TC

RTD 9

8
4 10
The user-friendly transmitters for the control room HART modem/
D communicator
The SITRANS TW universal transmitter is a further development A 5
µC Output
of the service-proven SITRANS T for the 4-wire system in a 11
U or I
mounting rail housing. With numerous new functions it sets new D A
1 2 6 12
standards for temperature transmitters. A Current test
0/4 ... 20 mA
Power
With its diagnostics and simulation functions the SITRANS TW EEPROM supply 13
provides the necessary insight during commissioning and oper- UH
3 unit
ation. And using its HART interface the SITRANS TW can be 7
conveniently adapted with SIMATIC PDM to every measurement
task.
All SITRANS TW control room devices are available in a non-in- The signal output by a resistance-based sensor (two-wire, three-
trinsically safe version as well as in an intrinsically safe version wire, four-wire system), voltage source, current source or ther-
for use with the most stringent requirements. mocouple is converted by the analog-to-digital converter
(1, function diagram) into a digital signal. This is evaluated in the
■ Application microcontroller (2), corrected according to the sensor character-
istic, and converted by the digital-to-analog converter (6) into an
The SITRANS TW transmitter is a four-wire rail-mounted device output current (0/4 to 20 mA) or output voltage (0/2 to 10 V). The
with a universal input circuit for connection to the following sen- sensor characteristics as well as the electronics data and the
sors and signal sources: data for the transmitter parameters are stored in the non-volatile
• Resistance thermometers memory (3).
• Thermocouple elements AC or DC voltages can be used as the power supply (13). Any
• Resistance-based sensors/potentiometers terminal connections are possible for the power supply as a re-
• mV sensors sult of the bridge rectifier in the power supply unit. The PE con-
ductor is required for safety reasons.
• As special version:
- V sources A HART modem or a HART communicator permit parameteriza-
- Current sources tion of the transmitter using a protocol according to the HART
specification. The transmitter can be directly parameterized at
The 4-wire rail-mounted SITRANS TW transmitter wire is de- the point of measurement via the HART output terminals (10).
signed for control room installation. It must not be mounted in po-
tentially explosive atmospheres. The operation indicator (4) identifies a fault-free or faulty operat-
ing state of the transmitter. The limit monitor (9) enables the sig-
All SITRANS TW control room devices are available in a non-in- naling of sensor faults and/or limit violations. In the case of a cur-
trinsically safe version as well as in an intrinsically safe version rent output, the current can be checked on a meter connected
for use with the most stringent requirements. to test socket (12).

■ Function Diagnosis and simulation functions


The SITRANS TW comes with extensive diagnosis and simula-
Features tion functions.
• Transmitter in four-wire system with HART interface Physical values can be defined with the simulation function. It is
• Housing can be mounted on 35 mm rail or 32 mm G rail thus possible to check the complete signal path from the sensor
• Screw plug connector input to inside the control system without additional equipment.
The slave pointer functions are used to record the minimum and
• All circuits electrically isolated maximum of the plant’s process variable.
• Output signal: 0/4 to 20 mA or 0/2 to 10 V
• Power supplies: 115/230 V AC/DC or 24 V AC/DC
• Explosion protection [EEx ia] or [EEx ib] for measurements
with sensors in the hazardous area
• Temperature-linear characteristic for all temperature sensors

3/38 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitters for rail mounting
SITRANS TW
four-wire system, universal, HART

■ Integration ■ Technical specifications


System configuration Input
Selectable filters to suppress the 50 Hz, 60 Hz, also 10 Hz for spe-
line frequency cial applications (line frequency
filter is similar with measuring fre-
SIMATIC S7-400 quency)

3
Resistance thermometer

SIMATIC PDM
Measured variable Temperature
ET 200M Measured range Parameterizable
Control system
Measured span Min. 25 °C (45 °F) x 1/scaling fac-
UH or tor
A Sensor type
4...20 mA Load
SITRANS TW • DIN IEC 751 Pt100 (DIN IEC 751)
transmitter
• Acc. to JIS C 1604-81 Pt100 (JIS C1604-81)
• Acc. to DIN 43760 Ni100 (DIN 43760)
PC/laptop with
HART HART SIMATIC PDM
communi- modem • Special type (RRTD ≤ 500 Ω) Multiples or parts of the defined
cator or characteristic values can be
RS-232-C parameterized (e.g. Pt500, Ni120)
Voltage measurement Temperature-linear, resistance-lin-
Possible system configurations ear or customer-specific
The SITRANS TW transmitter as a four-wire rail-mounted device Type of connection • Normal connection
can be used in a number of system configurations: as a stand- • Sum or parallel connection
alone version or as part of a complex system environment, e.g. • Mean-value or differential con-
with SIMATIC S7. All device functions are available via HART nection
communication.
Connection 2, 3 or 4-wire circuit
Communication options through the HART interface: Measuring range limits Depending on type of connected
• HART communicator thermometer (defined range of
resistance thermometer)
• HART modem connected to PC/laptop on which the appropri-
ate software is available, e.g. SIMATIC PDM Sensor breakage monitoring Monitoring of all connections for
open-circuit (function can be
• HART-compatible control system (e.g. SIMATIC S7-400 with switched off)
ET 200M)
Sensor short-circuit monitoring Parameterizable response thresh-
old (function can be switched off)
Resistance-based sensors,
potentiometers
Measured variable Ohmic impedance
Measured range Parameterizable
Measured span Min. 10 Ω
Voltage measurement Resistance-linear or customer-
specific
Type of connection • Normal connection
• Differential connection
• Mean-value connection
Connection 2, 3 or 4-wire circuit
Input range 0 ... 6000 Ω;
with mean-value and difference
circuits: 0 ... 3000 Ω;
Sensor breakage monitoring Monitoring of all connections for
open-circuit (function can be
switched off)
Sensor short-circuit monitoring Parameterizable response thresh-
old (function can be switched off)

Siemens FI 01 · 2008 3/39


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitters for rail mounting
SITRANS TW
four-wire system, universal, HART
Thermocouple elements µA, mA sources
Measured variable Temperature Measured variable DC voltage
Measured range Parameterizable Measured range Parameterizable
Measured span Min. 50 °C (32.22 °F) x 1/scaling Voltage measurement Current-linear or customer-
factor specific
Measuring range limits Depend. on type of thermocouple Input range/min. span
element
3
• Devices with 7NG3242-xxxx4 -12 µA ... +100 µA/0.4 µA
Thermocouple Type B: Pt30%Rh/Pt6%Rh
• Devices with 7NG3242-xxxx5 -120 µA ... +1000 µA/4 µA
(DIN IEC 584)
• Devices with 7NG3242-xxxx6 -1.2 mA ... +10 mA/0.04 mA
Type C W5%-Re (ASTM 988)
• Devices with 7NG3242-xxxx7 or -12 mA ... +100 mA/0.4 mA
Type D W3%-Re (ASTM 988)
7NG3242-xxxx 0 with U/I plug
Type E: NiCr/CuNi (DIN IEC 584)
• Devices with 7NG3242-xxxx8 -120 mA ... +1000 mA/4 mA
Type J: Fe/CuNi (DIN IEC 584)
Sensor breakage monitoring Not possible
Type K: NiCr/Ni (DIN IEC 584)
Output
Type L Fe-CuNi (DIN 43710)
Output signal Load-independent direct current
Type N: NiCrSi-NiSi (DIN IEC 584) 0/4 ... 20 mA, can be switched to
load-independent DC voltage
Type R: Pt13%Rh/Pt
0/2 ... 10 V using plug-in jumpers.
(DIN IEC 584)
Current 0/4 ... 20 mA
Type S: Pt10%Rh/Pt
(DIN IEC 584) • Overrange -0.5 ... +23.0 mA, continuously
adjustable
Type T: Cu/CuNi (DIN IEC 584)
• Output range following sensor fault -0.5 ... +23.0 mA, continuously
Type U: Cu/CuNi (DIN 43710)
adjustable
Special type
• Load ≤ 650 Ω
(-10 mV ≤ UTC ≤ 100 mV)
• Open-circuit voltage ≤ 30 V
Voltage measurement Temperature-linear, voltage-linear
or customer-specific Voltage 0/2 ... 10 V
Type of connection • Normal connection • Overrange -0.25 ... +10.75 V, continuously
• Sum connection adjustable
• Mean-value connection • Output range following sensor fault -0.25 ... +10.75 V, continuously
adjustable
• Differential connection
• Load resistance ≥ 1 kΩ
Cold junction compensation None, internal measurement,
external measurement or pre- • Load capacitance ≤ 10 nF
defined fixed value • Short-circuit voltage ≤ 100 mA (not permanently short-
Sensor breakage monitoring Function can be switched off circuit-proof)
mV sensors • Electrical damping
Measured variable DC voltage - Adjustable time constant T63 0 ... 100 s, in steps of 0.1 s
Measured range Parameterizable • Current source/voltage source Continuously adjustable within the
total operating range
Measured span Min. 4 mV
Sensor fault/limit signalling By operation indicator, relay out-
Input range -120 ... +1000 mV put or HART interface
Voltage measurement Voltage-linear or customer-spe- Operation indicator Flashing signal
cific
• Limit violation Flashing frequency 5 Hz
Overload capacity of inputs Max. ± 3.5 V
• Overrange Flashing frequency 1 Hz
Input resistance ≥ 1 MΩ
Relay output Either as NO or NC contact with
Sensor current Approx. 180 µA 1 changeover contact
Sensor breakage monitoring Function can be switched off • Switching capacity ≤ 150 W, ≤ 625 VA
V sources • Switching voltage ≤ 125 V DC, ≤ 250 V AC
Measured variable DC voltage • Switching current ≤ DC 2.5 A
Measured range Parameterizable Overrange Signalling of sensor or line break-
Voltage measurement Voltage-linear or customer-spe- age and sensor short-circuit
cific Limit monitoring
Input range/min. span • Operating delay 0 ... 10 s
• Devices with 7NG3242-xxxx1 or -1.2 V ... +10 V/0.04 V • Monitoring functions of limit • Sensor fault (breakage and/or
7NG3242-xxxx 0 with U/I plug module short-circuit)
• Devices with 7NG3242-xxxx2 -12 V ... +100 V/0.4 V • Lower and upper limit
• Devices with 7NG3242-xxxx3 -120 V ... +140 V/4.0 V • Window (combination of lower
Sensor breakage monitoring Not possible and upper limits)
• Limit and sensor fault detection
can be combined
• Hysteresis Parameterizable between 0 and
100% of measuring range

3/40 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitters for rail mounting
SITRANS TW
four-wire system, universal, HART
Power supply Rated conditions
Universal power supply unit 115/230 V AC/DC or 24 V AC/DC Installation conditions
Tolerance range for power supply Location (for devices with explosion
protection)
• With 115/230 V AC/DC PSU 80 ... 300 V DC; 90 ... 250 V AC
• Transmitters Outside the potentially explosive
• With 24 V AC/DC PSU 18 ... 80 V DC; 20.4 ... 55.2 V AC
atmosphere
(in each case interruption-resis-
tant up to 20 ms in the complete • Sensor Within the potentially explosive

Tolerance range for mains fre-


tolerance range)
47 ... 63 Hz
atmosphere zone 1 (also in zone
0 in conjunction with the pre-
scribed protection requirements
3
quency
for the sensor)
Power consumption with
Ambient conditions
• 230 V AC ≤ 5 VA
Permissible ambient temperature -25 ... +70 °C (-13 ... +158 °F)
• 230 V DC ≤5W
Permissible storage temperature -40 ... +85 °C (-40 ... +185 °F)
• 24 V AC ≤ 5 VA
Climatic class
• 24 V DC ≤5W
• Relative humidity 5 ... 95%, no condensation
Electrical isolation
Design
Electrically isolated circuits Input, output, power supply and
Weight Approx. 0.24 kg (0.53 lb)
sensor fault/limit monitoring out-
put are electrically isolated from Enclosure material PBT, glass-fibre reinforced
one another. The HART interface
Degree of protection to IEC 529 IP20
is electrically connected to the
output. Degree of protection to VDE 0100 Protection class I
Working voltage between all electri- The voltage Ueff between any two Type of installation 35-mm top hat rail (1.38 inch)
cally isolated circuits terminals must not exceed 300 V (DIN EN 50022) or 32-mm G-type
rail (1.26 inch) (DIN EN 50035)
Measuring accuracy
Electrical connection/process con- Screw plug connectors,
Measurement error
nection max. 2.5 mm² (0.01 inch²)
• Error in the internal cold junction ≤ 3 °C ± 0,1 °C/10 °C
Parameterization interface
(≤ 5.4 °F ± 0,18 °F/18 °F)
Protocol HART, version 5.9
• Error of external cold junction ter- ≤ 0.5 °C ± 0,1 °C/10 °C
minal 7NG3092-8AV (≤ 0.9 °F ± 0.18 °F/18 °F) Load with connection of
• Digital output See "Digital error" • HART communicator 230 ... 650 Ω
• Analog output IAN or UAN ≤ 0.05% of span in addition to • HART modem 230 ... 500 Ω
digital error
Software for PC/laptop SIMATIC PDM version V5.1 and
Influencing effects (referred to the Compared to the max. span: later
digital output)
• Temperature drift ≤ 0.08%/10 °C (≤ 0.08%/18 °F)
≤ 0.2% in the range -10 ... +60 °C
(14 ... 140 °F)
• Long-term drift ≤ 0.1%/year
Influencing effects referred to the Compared to the span:
analog output IAN or UAN
• Temperature drift ≤ 0.08%/10°C (≤ 0.08%/18 °F)
≤ 0.2% in the range -10 ... +60 °C
(14 ... 140 °F)
• Power supply ≤ 0.05%/10 V
• Load with current output ≤ 0.05% with change from
50 ... 650 Ω
• Load with voltage output ≤ 0.1% with change in load cur-
rent from 0 ... 10 mA
• Long-term drift (start-of-scale val- ≤ 0.03%/month
ue, span)
Response time (T63 without electri- ≤ 0.2 s
cal damping)
Electromagnetic compatibility According to EN 61326 and
NAMUR NE21
Certificate and approvals
ATEX To DIN EN 50014: 1997,
EN 50020: 1994
Intrinsic safety to EN 50 020
• for 7NG3242-xAxxx II (1) G D [EEx ia/ib ] IIB
• for 7NG3242-xBxxx II (1) G D [EEx ia/ib ] IIC
EC-Type Examination Certificate TÜV (German Technical Inspec-
torate) 01 ATEX 1675

Siemens FI 01 · 2008 3/41


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitters for rail mounting
SITRANS TW
four-wire system, universal, HART
Digital error
Resistance thermometer Thermocouple elements
Input Measured range Max. permissi- Digital error Input Measured range Digital error1)
ble line resis-
tance °C (°F) °C (°F)

°C (°F) Ω °C (°F) Type B 0 ... +1820 3 (5.4)


(+32 ... +3308)

3
DIN IEC 751 Type C 0 ... +2300 2 (3.6)
• Pt10 -200 ... +850 20 3.0 (5.4) (+32 ... +4172)
(-328 ... +1562) Type D 0 ... +2300 1 (1.8)
• Pt50 -200 ... +850 50 0.6 (1.1) (+32 ... +4172)
(-328 ... +1562) Type E -200 ... +1000 1 (1.8)
• Pt100 -200 ... +850 100 0.3 (0.5) (-328 ... +1832)
(-328 ... +1562) Type J -210 ... +1200 1 (1.8)
• Pt200 -200 ... +850 100 0.6 (1.1) (-346 ... +2192)
(-328 ... +1562) Type K -200 ... +1372 1 (1.8)
• Pt500 -200 ... +850 100 1.0 (1.8) (-328 ... +2502)
(-328 ... +1562)
Type L -200 ... +900 2 (3.6)
• Pt1000 -200 ... +850 100 1.0 (1.8) (-328 ... +1652)
(-328 ... +1562)
Type N -200 ... +1300 1 (1.8)
JIS C 1604-81 (-328 ... +2372)
• Pt10 -200 ... +649 20 3.0 (5.4) Type R -50 ... +1760 2 (3.6)
(-328 ... +1200) (-58 ... +3200)
• Pt50 -200 ... +649 50 0.6 (1.1) Type S -50 ... +1760 2 (3.6)
(-328 ... +1200) (-58 ... +3200)
• Pt100 -200 ... +649 100 0.3 (0.5) Type T -200 ... +400 1 (1.8)
(-328 ... +1200) (-328 ... +752)
DIN 43760 Type U -200 ... +600 2 (3.6)
(-328 ... +1112)
• Ni50 -60 ... +250 50 0.3 (0.5)
(-76 ... +482) 1)
Accuracy data refer to the largest error in the complete measuring range
• Ni100 -60 ... +250 100 0.3 (0.5)
(-76 ... +482) Voltage/current sources
• Ni120 -60 ... +250 100 0.3 (0.5) Input Measuring range Digital error
(-76 ... +482)
mV sources (linear) mV µV
• Ni1000 -60 ... +250 100 0.3 (0.5)
(-76 ... +482) -1 ... +16 35
-3 ... +32 20
Resistance-based sensors -7 ... +65 20
-15 ... +131 50
Input Measured Max. permissi- Digital error
range ble line resis- -31 ... +262 100
tance -63 ... +525 200
Ω Ω Ω -120 ... +1000 300

Resistance 0 ... 24 5 0.08 V sources (linear) V mV


(linear) 0 ... 47 15 0.06 -1.2 ... +10 3
0 ... 94 30 0.06 -12 ... +100 30
0 ... 188 50 0.08 -120 ... +140 300
0 ... 375 100 0.1 µA/mA sources (linear) µA/mA µA
0 ... 750 100 0.2 -12 ... +100 µA 0.05
0 ... 1500 75 1.0 -120 ... +1000 µA 0.5
0 ... 3000 100 1.0 -1.2 ... +10 mA 5
0 ... 6000 100 2.0 -12 ... +100 mA 50
-120 ... +1000 mA 500

3/42 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitters for rail mounting
SITRANS TW
four-wire system, universal, HART

■ Ordering examples ■ Ordering information


Desired transmitter Parameter: Ordering The order number structure shown below is used to specify a
design fully functioning transmitter. The selection of the operating data
Standard Special
(type of source, measuring range, characteristic etc.) is made
Example 1: according to the following rules:
SITRANS TW, • Operating data already set in factory to default values:
transmitter in four-wire system

3
The default settings can be obtained from the list of parame-
• with explosion protection ATEX
terizable operating data (see "Special operating data"). The

7NG3242-1AA00
• 230 V AC/DC power supply presets can be modified by the customer to match the require-

(stock item)
• current output ments precisely.
• without sensor fault/limit monitor
- Sensor PT100, three-wire circuit X
• Operating data set on delivery according to customer require-
ments:
- Measuring range 0 ... 150 °C X
Supplement the Order No. by "-Z" and add the Order code
- Temperature-linear characteristic X
"Y01". The operating data to be set can be obtained from the
- Filter time 1 s X list of parameterize operating data. The Order codes A 7 7 to
- Output 4 ... 20 mA, line filter 50 Hz X K 7 7 for operating data to be set need only be specified in the
- Output driven to full-scale in event X order if they deviate from the default setting.
of like breakage The default setting is used if no Order code is specified for op-
Example 2: erating data.
SITRANS TW, The selected parameters are printed on the transmitter's rating
transmitter in four-wire system
plate.
Y01 + S76 + A05 + Y30 + H10

• without explosion protection


Y30: MA=0; ME= 950; D=C

• 24 V AC/DC power supply


Y01: see Order code
7NG3242-0BB10-Z

• Voltage output
• Sensor fault/limit monitor
- Rating plate in English S76
- Sensor NiCr/Ni, type K A05
- Cold junction internal X
- Measuring range 0 ... 950 °C Y30
- Temperature-linear characteristic X
- Filter time 1 s X
- Output 0 ... 10 V, line filter 50 Hz H10
- Output driven to full-scale in event X
of like breakage
- Limit monitoring switched off X
Example 3:
Y01 + A40 + Y32 + G07 + H11 + J03

SITRANS TW,
transmitter in four-wire system
Y32: MA=0; ME= 5; D=V

• without explosion protection


Y01: see Order code
7NG3242-0BA01-Z

• 24 V AC/DC power supply


• Current output
• without sensor fault/limit monitor
- Voltage input, measuring A40
range -1.2 V ... +10 V
- Measuring range 0 ... 5 V Y32
- Source-proportional characteristic X
- Filter time 10 s G07
- Output 0 ... 20 mA, line filter 60 Hz H11
- No monitoring for sensor fault (X) J03

Siemens FI 01 · 2008 3/43


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitters for rail mounting
SITRANS TW
four-wire system, universal, HART

Selection and Ordering data Order No. Selection and Ordering data Order No.
SITRANS TW universal transmitter } 7 NG 3 2 4 2 - 77777 Accessories
for rail mounting, in four-wire system Instruction Manual for SITRANS TW
(order instruction manual separately)
• German/English } A5E00054075
Explosion protection • French/Italian/Spanish } A5E00064515
• without } 0
• for inputs [EEx ia] or [EEx ib] } 1 Cold junction terminal } 7NG3092-8AV

3 Power supply
• 115/230 V AC/DC } A
U/I plug
(-1.2 ... +10 V DC pr -12 ... +100 mA)
} 7NG3092-8AW

• 24 V AC/DC } B SIMATIC PDM operating software see Chapter 9


Output signal
HART modem
• 0/4 ... 20 mA (can be switched to 0/2 ... 10 V)} A
• 0/2 ... 10 V (can be switched to 0/4 ... 20 mA) B • with RS232 interface } 7MF4997-1DA
Sensor fault/limit monitor D)
• without (retrofitting not possible) } 0 • with USB interface } 7MF4997-1DB
• relay with changeover contact 1 D)
Input for
• Temperature sensor, resistance-based sen- } 0 D) Subject to export regulations AL:N, ECCN: EAR99H.
sor and mV sensor with measuring range
-120 ... +1000 mV DC and with U/I plug
• Voltage input (V sources) 1)
Measuring range:
- DC -1.2 ... +10 V 1
- DC -12 ... +100 V (not Ex version) 2
- DC -120 ... +140 V (not Ex version) 3
• Current input (µA, mA sources) 1)
Measuring range:
- DC -12 ... +100 µA 4
- DC -120 ... +1000 µA 5
- DC -1.2 ... +10 mA 6
- DC -12 ... +100 mA 7
- DC -120 ... +1000 mA 8
Further designs Order code
Please add "-Z" to Order No. and specify
Order code(s) (see "List of parameterizable
operating data").
• Customer-specific setting of operating data Y01
(see "List of parameterizable operating
data")
Note:
specify in plain text: „see Order code“
• Meas. point description (max. 16 char.) Y23
• Text on front of device (max. 32 char.) Y24
• HART tag (max. 8 characters) Y25
• with test report P01
• with shorting plug to HART communication S01
for 0 mA or 0 V
• with plug for external cold junction compen- S02
sation
• with U/I plug S03
(-1.2 ... +10 V DC or -12 ... +100 mA)
Language of rating plate
(together with Y01 order code only)
• Italian S72
• English S76
• French S77
• Spanish S78
1) Observe max. values with Ex version.
} Available ex stock.

3/44 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitters for rail mounting
SITRANS TW
four-wire system, universal, HART

■ List of parameterizable operating data (Order codes A 7 7 + B 7 7 ... E 7 7)


Operating data acc. to default setting Order No. with Order code: 7NG3242 - 7 7 7 7 7 -Z Y01
Order codes: A 7 7 ... E 7 7 777 + 777 + 777 + 777 + 777
Sensor
Thermocouples Connection Cold junction Measuring
Type Temperature range compensation ranges
B: Pt30%Rh/Pt6%Rh
C:W5%Re
0 ... 1820 °C
0 ... 2300 °C
A 0 0 Standard
A 0 1 Sum n 1) n=2
B 0 1 None
B 0 2 Internal
C00
C10
-30 ... +60 °C
-20 ... +20 °C
E00
E01 3
D:W3%Re 0 ... 2300 °C A02 ... . . . Fixed val. 0 °C C20 0 ... 40 °C E02
E:NiCr/CuNi -200 ... +1000 °C A03 n = 10 B10 20 °C C22 0 ... 60 °C E03
J:Fe/CuNi (IEC) -210 ... +1200 °C A 0 4 Difference 2) Diff1 B31 50 °C C25 0 ... 80 °C E04
K:NiCr/Ni -200 ... +1372 °C A05 Diff2 B32 60 °C C26 0 ... 100 °C E05
L: Fe/CuNi (DIN) -200 ... +900 °C A 0 6 Mean-val. 2) MW B41 70 °C C27 0 ... 120 °C E06
N:NiCrSi/NiSi -200 ... +1300 °C A07 Special value 7) Y10 0 ... 150 °C E07
R:Pt13%Rh/Pt -50 ... +1760 °C A08 External meas. Y11 0 ... 200 °C E08
S:Pt10%Rh/Pt -50 ... +1760 °C A09 (through Pt100 0 ... 250 °C E09
T:Cu/CuNi (IEC) -200 ... +400 °C A10 DIN IEC 751) 7) 0 ... 300 °C E10
U:Cu/CuNi (DIN) -200 ... +600 °C A11 0 ... 350 °C E11
0 ... 400 °C E12
Resistance thermometer Connection Connection Line resis- 0 ... 450 °C E13
(or max. permissible line resistance see tance 3) 0 ... 500 °C E14
„Technical specifications“) 0 ... 600 °C E15
Pt100 (DIN IEC) -200 ... +850 °C A 2 0 Standard B 0 1 2-wire-system C32 0Ω D00 0 ... 700 °C E16
Pt100 (JIS) -200 ... +649 °C A 2 1 Sum n 4) n=2 B 0 2 3-wire-system C 3 3 10 Ω D10 0 ... 800 °C E17
Ni100 (DIN) -60 ... +250 °C A22 ... . . . 4-wire-system C 3 4 20 Ω D20 0 ... 900 °C E18
n = 10 B10 50 Ω D50 0 ... 1000 °C E19
Parallel n 5) n = 0.1 B21 Special val.7) Y20 0 ... 1200 °C E20
n= 0.2 B22 0 ... 1400 °C E21
n= 0.5 B25 0 ... 1600 °C E22
Special value 6) 7) Y00 0 ... 1800 °C E23
Difference 2) Diff1 B51
Diff2 B52 50 ... 100 °C E24
50 ... 150 °C E25
Mean-val. 2) MW B61
100 ... 200 °C E26
100 ... 300 °C E27
100 ... 400 °C E28
200 ... 300 °C E29
200 ... 400 °C E30
200 ... 500 °C E31
300 ... 600 °C E32
500 ... 1000 °C E33
600 ... 1200 °C E34
800 ... 1600 °C E35
Special range 7) Y30

Resistance-based sensors, potentiome- Connection Connection Line resis- Measuring


ters tance 3) ranges
(or max. permissible line resistance see A 3 0 Standard B 0 1 2-wire-system C32 0Ω D00 0 ... 100 Ω E40
„Technical specifications“) Difference 2) Diff1 B 5 1 3-wire-system C 3 3 10 Ω D10 0 ... 200 Ω E41
Diff2 B 5 2 4-wire-system C 3 4 20 Ω D20 0 ... 500 Ω E42
Mean val.2) MW B61 50 Ω D50 0 ... 1000 Ω E43
Special val. 7) Y 2 0 0 ... 2500 Ω E44
0 ... 5000 Ω 8) E45
0 ... 6000 Ω 8) E46
Special range 7) Y21

mV, V and µA, mA sensors 9) A 4 0 Meas. range with Order No. 7NG 3242 - 7 7 7 7 7 -Z Y01 E50
0 -120 ... +1000 mV
1 -1,2 ... +10 V 10)
1) n = number of thermocouple elements to be connected in series 2 -12 ... +100 V 10)
2)
See „Circuit diagrams“ for meaning of type circuit 3 -120 ... +140 V 10)
3)
Line resistance of channels 1 and 2, for max. permissible line resistance see 4 -12 ... +100 µA 10)
„Technical specifications“ (only with C32, not with C33 and C34)
4) 5 -120 ... +1000 µA 10)
n = number of resistance thermometers to be connected in series
5) 6 -1,2 ... +10 mA 10)
1/n = number of resistance thermometers to be connected in parallel
6) 7 -12 ... +100 mA 10)
Combination of series and parallel connection of resistance thermometers
7) 8 -120 ... +1000 mA 10)
Operating data: see „Special operating data“
8) Special range 7) Y 3 2
This range does not apply to mean-value and difference circuits.
9)
The max. permissible currents and voltages according to conformity cer-
tificate must be observed in devices with explosion protection.
10) Without detection of line breakage

Siemens FI 01 · 2008 3/45


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitters for rail mounting
SITRANS TW
four-wire system, universal, HART

■ List of parameterizable operating data (Order codes F 7 7 ... K 7 7)


Operating data according to default setting Order No. with Order code: 7NG3242 - 7 7 7 7 7 -Z Y01
Order codes: F 7 7 ... K 7 7 777 + 777 + 777 + 777 + 777
Sensor
Thermocouple elements Voltage Filter Output sig- Failure signal Limit
measure- time1) nal and line monitor 3)

3 Type Temperature range


ment filter 2)

B: Pt30%Rh/ 0 ... 1820 °C A00 Temperature- F 0 0 0s G0 0 4 ... 20 mA/ with line break- Limit monitor- K 0 0
C:W5%Re 0 ... 2300 °C A01 linear 0.1 s G0 1 2 ... 10 V age/fault: ing ineffective
D:W3%Re 0 ... 2300 °C A02 Voltage- F10 0.2 s G0 2 with line filter: (but sensor
E:NiCr/CuNi -200 ... +1000 °C A03 linear 0.5 s G0 3 50 Hz H00 to full scale J 0 0 fault signalling
J:Fe/CuNi (IEC) -210 ... +1200 °C A04 1s G0 4 60 Hz H01 to start of scale J 0 1 with closed-
K:NiCr/Ni -200 ... +1372 °C A05 2s G0 5 10 Hz 4) H02 hold last value J 0 2 circuit opera-
tion)
L: Fe/CuNi (DIN) -200 ... +900 °C A06 5s G0 6 0 ... 20 mA/
N:NiCrSi/NiSi -200 ... +1300 °C A07 10 s G0 7 0 ... 10 V no monitoring J03
R:Pt13%Rh/Pt -50 ... +1760 °C A08 20 s G0 8 with line filter: Effective 5) Y70
S:Pt10%Rh/Pt -50 ... +1760 °C A09 50 s G0 9 50 Hz H 1 0 Safety value 5) Y60
T:Cu/CuNi (IEC) -200 ... +400 °C A10 100 s G1 0 60 Hz H11
U:Cu/CuNi (DIN) -200 ... +600 °C A11 Special Y50 10 Hz H12
time 5)

Resistance thermometer Voltage Failure signal


(max. permissible line resistances see measurment
„Technical specifications“)

Pt100 (DIN IEC) -200 ... +850 °C A 2 0 Temperature- F 0 0 with line break-
Pt100 (JIS) -200 ... +649 °C A 2 1 linear age/fault:
Ni100 (DIN) -60 ... +250 °C A 2 2 Resistance- F 2 0 to full scale J00
linear to start of scale J01
hold last value J02

no monitoring J03

Safety value 5) Y60


with line break-
age or short-cir-
cuit/fault:
to full scale J10
to start of scale J 1 1
hold last value J 1 2

no monitoring J13

Safety value 5) Y61

Resistance-based sensors, potenti- Voltage Failure signal


ometers measurment
(max. permissible line resistances see A 3 0 Resistance- F 2 0 with line break-
„Technical specifications“) linear age/fault:
to full scale J00
to start of scale J01
hold last value J02

no monitoring J03

Safety value 5) Y60


mV, V and µA, mA sources A 4 0 Voltage
measurment
Source pro- F 3 0
portional
1) Software filter to smooth the result
2) Filter to suppress line disturbances on the measured signal.
3)
If signalling relay present
4)
for special appliciations
5)
Operating data: see „Special operating data“

3/46 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitters for rail mounting
SITRANS TW
four-wire system, universal, HART

■ Special operating data


Order Plain text Options
code required
Y00 N=. Factor N for multiplication with the charac-
teristic values of resistance thermometers
Range of values: 0.10 to 10.00
1. Example: 3 x Pt500 parallel:

Y10 TV=.
N = 5/3 = 1.667;
2. Example: Ni120: N = 1.2
Temperature TV of the fixed cold junction
3
D= Dimension; range of values: C, K, F, R
Y11 RL=. Line resistance RL in Ω for compensation of
cold junction line of external Pt100
DIN IEC 751
Range of values: 0.00 to 100.00
Y20 RL1=. Line resistances RL of channel 1 (RL1) and
RL2=. channel 2 (RL2) in Ω if the resistance ther-
mometer or the resistance-based sensor is
connected in a two-wire system
Range of values depending on type of sen-
sor: 0.00 to 100.00
Y30 MA=. Start-of-scale value MA and full-scale value
ME=. ME for thermocouples and resistance ther-
mometers
(Range of values depending on type of sen-
sor)
D= Dimension, range of values: C, K, F, R)
Y31 MA=. Start-of-scale value MA and full-scale value
ME=. ME for resistance-based sensors or potenti-
ometers in Ω
Range of values: 0.00 to 6,000.00
Y32 MA=. Start-of-scale value MA and full-scale value
ME=. ME for mV, V, µA and mA sources
Range of values depending on type of sen-
sor: -120.00 to 1,000.00
D= Dimension (mV entered as MV, V as V,
µA as UA, mA as MA)
Y50 T63=. Response time T63 of software filter in s
Range of values: 0.0 to 100.0
Safety value S of signal output in mA or in V
corresponding to the set type of output.
Range of values
- with current output: -0.50 to 23.00
- with voltage output: -0.25 to 10.75
Y60 S= . Safety value S with line breakage of sensor
Y61 S=. Safety value S with line breakage or short-
circuit of sensor
Y70 UG=. Lower limit value (dimension as defined by
measuring range)
OG=. Upper limit value (dimension as defined by
measuring range)
H=. Hysteresis (dimension as defined by mea-
suring range)
K= Switch on/off combination of limit function
and sensor fault detection; J=on; N=off
(standard: J)
A= Type of relay output: A=open-circuit opera-
tion; R=closed-circuit operation (standard: R)
T=. Switching delay T of relay output in s
Range of values: 0.0 to 10.0 (standard: 0.0)

Siemens FI 01 · 2008 3/47


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitters for rail mounting
SITRANS TW
four-wire system, universal, HART

■ Schematics
Sensor input connections

1 2 3 4
4 3 2 1 4 3 2 1
3
4 3 2 1 4 3 2 1

R, RTD R, RTD R, RTD R2, RTD2 R1, RTD1

5 6 7 8
4 3 2 1 4 3 2 1 4 3 2 1 4 3 2 1

- + - + Pt100 - + TC1
Pt100
TC TC
4 1 - + TC2
- +
TC 7NG3092-8AV

9 10 11

4 3 2 1 4 3 2 1 4 3 2 1

- + - + - +
-120 ... 1000 mV

12 13
4 3 2 1 4 3 2 1

U+ U- I+ I-
+ - - +
7NG3092-8AW
-1,2 ... 10 V -12 ... 100 mA

Resistance thermometers, resistance-based Thermocouples: Further sources:


sensors, potentiometers: 5 Determination of cold junction temperature using 9 mV sources with two-wire system (7NG3242-xxxx0)
1 Two-wire system; resistance can be built-in Pt100 or fixed reference temperature 10 V sources with two-wire system (7NG3242-xxxx[1-3])
parameterized for line compensation 6 Determination of cold junction temperature using 11 mA/mA sources with two-wire system
2 Three-wire system external Pt100; resistance can be parameterized (7NG3242-xxxx[4-8])
for line compensation
3 Four-wire system 12 Voltage measurement -1,2 to 10 V with
7 Determination of cold junction temperature using U/I plug 7NG3092-8AW (7NG3242-xxxx0)
4 Difference/mean-value circuit; cold junction terminal 7NG3092-8AV
2 resistors can be parameterized 13 Current measurement -12 to 100 mA with
for line compensation 8 Difference/mean-value circuit with internal U/I plug 7NG3092-8AW (7NG3242-xxxx0)
cold junction temperature

Connection diagram for the input signal


Channel 1 is the measured variable between the terminals 2 and 3 on the input plug. With a difference or mean-value circuit, the
calculation of the measured value is defined by the type of measurement. Otherwise the measured value is determined via channel
1. The following code is used for the type of measurement:
The short-circuit jumpers shown in the circuits must be inserted
Type of measurement Calculation of measured value
in the respective system on site.
Single channel Channel 1
Differential connection 1 Channel 1 - Channel 2
Differential connection 2 Channel 2 - Channel 1
Mean-value 1 ½ ⋅ (Channel 1 + Channel 2)

3/48 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitters for rail mounting
SITRANS TW
four-wire system, universal, HART

Power supply Limit


connection monitor HART/ Sensor

14 13 12 11 10 9 8 7 6 5 4 3 2 1

~ ~ + -
+U- Relay output HART

3
H

1 to 4 Signal input (see "Sensor input connections" for possible types of connection)
5, 6 Analog output (U or I output parameterizable using plug-in jumpers)
7, 8 Connection with HART communication for local parameterization
9 to 11 Output for sensor fault/limit monitor as relay contact
(see below for possible parameterization)
12 PE connection
13, 14 Power supply input (protected against reverse polarity)

Connection diagram for power supply, input and outputs


Relay output
Connected terminals
Closed-circuit operation
(relay opens when error)
• Device switched off 10 and 11
• Device switched on and no error 9 and 11
• Device switched on and error 10 and 11
Open-circuit operation
(relay closes when error)
• Device switched off 10 and 11
• Device switched on and no error 10 and 11
• Device switched on and error 9 and 11

■ Dimensional drawings

Dimensions for control room mounting, rail mounting in mm (inches)

Siemens FI 01 · 2008 3/49


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitter for field mounting with temperature sensor
SITRANS TF2
Digital display thermometer

■ Overview ■ Design
The SITRANS TF2 has a stainless steel housing (diam. 80 mm)
with protective glass. The stainless steel protective tube with
screw socket contains the temperature sensor Pt100. By using
stainless steel for the protective tube it displays high chemical
resistance, which means that the temperature sensor is well pro-
tected against external effects
3 The protective tube is supplied as standard in lengths of 170 mm
or 260 mm; a customer-specific version is also possible. Simi-
larly, the protective tube can be supplied in the material of the
customer’s choice.
At the rear of the housing is the electrical connection for the volt-
age supply using a current loop of 4 to 20 mA. The connection
is made with plug connectors to EN 175301-803A.
At the front of the housing is the 5-digit display behind a glass
cover. Underneath the display are the 3 keys for parameterizing
the SITRANS TF2. Above the display are a green and a red LED
for indicating the operating status.
The SITRANS TF2 is available in two versions (see "Dimensional
drawings“):
• In the radial version (type A) the display is fitted in parallel with
the protective tube. The display can be rotated by up to ±120°
relative to the protective tube.
• In the axial version (type B) the display is at right angles to the
The temperature transmitter SITRANS TF2 integrates three ele- protective tube. The display can be rotated by 360° relative to
ments in one device: the protective tube.
• a Pt100 resistance thermometer in a stainless steel protective
tube, ■ Function
• a stainless steel housing with a high degree of protection, and
Mode of operation
• a built-in transmitter with LCD and three keys for parameter-
ization.
Green LED Red LED
It is used to indicate and monitor the temperature measured at IK LCD
the point of installation.
The SITRANS TF2 is available in an axial and a radial version. 200.00
RTD µC °C
A/D
■ Benefits
• Robust stainless steel housing with two connection versions
• High measuring accuracy 3 keys
EEP RO M
• Precise display with a resolution of 1/100 °C in the highest mea- UK
EMV
suring range +
D/A UA U I 4 ... 20 mA UH, IA
• Measuring ranges from -50 to +200 °C (-58 ... +392 °F) pa- -
rameterizable
3 keys Configuring of parameters
• Customer-specific lengths and materials possible for the pro- A/D Analog-to-digital converter
tective tube D/A Digital-to-analog converter
EEPROM Memory for all parameters
• Stainless steel protective tube with high resistance to chemi- EMC Output stage with protective components
IA Current output
cals IK Constant-current source
LCD Display of measured values with dimension
• Signaling of limit violation in the LCD as well as with a red LED Green LED Display of normal mode
Red LED Display of error messages or limit violation

■ Application RTD
UH
µC
Pt100 resistance thermometer
Power supply
Microcontroller for computing functions ond monitoring
The SITRANS TF2 is used for indicating and monitoring a tem-
perature variable at the point of installation. Applications are all The outside lying temperature sensor Pt100 is supplied with cur-
process engineering branches, e.g.: rent from the constant current course Ik. A temperature-related
• Chemical industry voltage drop is thus created over the sensor.
• Energy industry The voltage drop is converted on the analog/digital converter
(A/D) into a digital signal.
• Long-distance heating
• Water supply In the microcontroller (µC) the digital signal is linearized and
evaluated in accordance with the data saved in the EEPROM.
• Sewage works The processed data are shown in the display.
• Food industry
In addition the values are converted on the digital/analog con-
• Steelworks and the cement industry verter (D/A) and the voltage/current transformer (U/I) into a tem-
• Pharmaceutical industry perature-linear current signal IA (4 to 20 mA).
• Biotechnology

3/50 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitter for field mounting with temperature sensor
SITRANS TF2
Digital display thermometer
Display
Rated conditions
Display Ambient conditions
The SITRANS TF2 has a 5-digit display behind a glass cover. Ambient temperature -25 ... +85 °C (-13 ... +185 °F)
The following data are shown on the display:
Temperature range for best read- -10 ... +70 °C (14 ... 158 °F)
• measured temperature ability
• unit (°C, °F, °R or K and mA or %) Storage temperature -40 ... +85 °C (-40 ... +185 °F)
• limit violation, indicated by LED and arrow symbols in the dis-
play
Degree of protection
Electromagnetic compatibility
IP65 to EN 60529
EN 61326/A2 Appendix A (2001)
3
Settings
Displays and controls
The SITRANS TF2 is set using the 3 input keys behind the glass Display LCD, max. 5 digits, digit height
cover underneath the display. 9 mm (0.354 inch)
The key "M" is used to selected the operating mode. Following Resolution at max. measuring range 0.01 °C (0.01 °F)
modes of operation are available:
Decimal point Freely parameterizable
• Measured value
Limit values Freely parameterizable
• Password
Limit violation display Red LED and message on LCD
• Unit of measurement (↑ symbol /↓ symbol in case of
• Start of scale and end limit violation in upward/down-
ward direction)
• Upper and lower limit value
Parameterization With 3 keys
• Offset
Units mA or % or Ω or physical variable:
• Output current calibration °C, °F, °R, K
• Upper and lower current saturation limit Damping Between 0.1 and 100 s
• Electrical damping (increment: 0.1 s) freely parame-
The other two keys are used to set the values in the individual terizable
operating modes. Design
Monitoring Weight ≈ 0.7 kg (≈ 1.54 lb)
Two LED indicators are fitted above the display to monitor the set Non-wetted parts materials
range and the status: • Field housing Diam. 80 mm (diam. 3.15 inch),
• The green LED signals that the measured temperature lies stainless steel, mat. No. 1.4016
within the set limits. • Cover Stainless steel, mat. No. 14016
• The red LED lights up when the measured temperature lies with glass
outside the set limits and when there is an error. Wetted parts materials
• Protective tube To DIN 43772 form 8 (March
■ Technical specifications 2000), diam. 14 x 1.5 mm
(diam. 0.55 x 0.06 inch)
Measuring principle
- Material Stainless steel (mat. No.
Resistance thermometer Pt100 class B acc. to DIN IEC 751 1.4571/316Ti)
Input • Protective tube screw socket G½B to DIN 3852-2 form A or
Measured variable Temperature ½“-14 NPT

Max. measuring range -50 °C ... +200 °C - Material Stainless steel


(-58 ... +392 °F) (mat. No. 1.4571/316Ti)

Min. measured span 50 K (90 °F) Measuring insert Length to fit the ordered protec-
tive tube, stainless steel
Output
Connection of display to the protec- radial (type A), can be swiveled
Output signal 4 ... 20 mA, 2-wire tive tube by max. ±120° (α)
Lower current limit min. 3.6 mA axial (type B), can be swiveled by
max. ± 360°
Upper current limit max. 23 mA
Length of the protective tube(U1) see Ordering data
Output protected against reversed polarity, overvoltage and
short-circuiting Electrical connection Using 2-pole plug connector
made of plastic with
max. load (UH – 12 V) / 0.023 A M16x1.5-cable entry to
Voltage measurement Temperature-linear EN 175301-803A or ½“-14 NPT
Measuring accuracy Power supply
Error in measurement at 23 °C ± 5 K < ± (0.45 K + 0.2% of full-scale Terminal voltage on temperature 12 ... 30 V DC
(73.4 ± 9 °F) value in K + 1 digit in K) transmitter (UH)
(< ± (0.81 °F + 0.2% of full-scale Operating limits
value in °F + 1 digit in °F)
Pressure max. 40 bar (580 psi)
Measuring cycle time ≤ 100 ms
Temperature effect < ± 0.15%/10 K (< ± 0.15%/18 °F)
Power supply effect < ±0,01% of full-scale value/V
Vibration influence < ±0.05%/g to 500 Hz in all direc-
tions (to IEC 68-2-64)

Siemens FI 01 · 2008 3/51


© Siemens AG 2008

SITRANS T measuring instruments for temperature


Transmitter for field mounting with temperature sensor
SITRANS TF2
Digital display thermometer

Selection and ordering data Order No. Order code Selection and ordering data
Temperature transmitter SITRANS TF2, 7 NG 3 1 4 0 - Further design Order code
field device
Manufacturer’s test certificate M to DIN C11
Temperature transmitter with LCD in stain- 7777 0 777 55340, Part 18 and ISO 8402 (calibration
less steel housing, degree of protection certificate 3 points), add "-Z" to Order No.
IP65, stainless steel protective tube, and Order code.
resistance thermometer with Pt100 sensor,
Order No.
3
measuring range -50 ... +200 °C
(-58 ... +392 °F), local parameterization,
output signal 4 ... 20 mA Manufacturer’s test certificate M to DIN 7MF1564-8CC11
55340, Part 18 and ISO 8402 (calibration
Display/cable entry certificate 1 point) supplied later,
• Radial version (type A), } 1 specify factory no. of transmitter.
parallel to protective tube/M16x1.5 Additional data Order code
• Axial version (type B), 2
at right angles to protective tube/M16x1.5 Add „-Z“ to Order No. and specify Order
code and plain text
• Radial version (type A), 3
parallel to protective tube/½"-NPT Measuring range to be set Y01
• Axial version (type B), 4 Y01: .......................
at right angles to protective tube/½"-NPT Measuring point label made of stainless Y15
Process connection steel, please specify inscription in plain text.
• Connection shank G½B } A
• Connection shank ½"-14 NPT B

■ Schematics
• Other version (on request) Z J1Y
add Order code and plain text:
connection shank: ...
Length of the protective tube (U1)
• 170 mm (6.70 inch) } A 1 (+)
• 260 mm (10.24 inch) } B 1 I0 +
• 4.5" (114 mm) K UB
• 7.5" (190 mm) P 2 (-) RL
• 10.5" (266 mm) T 2
• Other version (on request) Z K1Y
add Order code and plain text:
length: ...
Material of the protective tube UB Power supply
• Stainless steel (mat. No. 1.4571/316Ti) } 0 RL Load
• Other version (on request) 9 L1Y I0 Output current
add Order code and plain text:
mat. No.: ...
SITRANS TF2, connection diagram
} Available ex stock

■ Dimensional drawings
Ø 82 (3.23) Ø 82 (3.23)

45 (1.77) 48,5 (1.91)


48,5 (1.99)
Ø 80 (3.15)

45 (1.77)

M16x1.5
67 (2.64)

or
½"-NPT
M16x1.5
SW 27 or
½"-NPT
SW 27

G½B
or
½"-14 NPT G½B
or
½"-14 NPT
Ø 14 (0.55)
U1
U1

Type A Type B

SITRANS TF2, dimensions in mm (inches)

3/52 Siemens FI 01 · 2008 Modification 01/2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitter for field mounting / field indicator
SITRANS TF - Transmitter, two-wire system
and SITRANS TF - Field indicator for 4 to 20 mA

■ Overview ■ Function
Configuration
The communication capability over the HART protocol V 5.9 of
the SITRANS TF with an integrated SITRANS TH300 permits pa-
rameterization using a PC or HART communicator (hand-held
communicator). The SIMATIC PDM makes it easy.
Parameterization is carried out using a PC for SITRANS TF - with
the integrated and programmable SITRANS TH200. Available
3
for this purpose are a special modem and the software tool
SIPROM T.
Mode of operation
Mode of operation of SITRANS TF as temperature transmitter
The sensor signal, whether resistance thermometer, thermocou-
ple or Ω and/or V signal, is amplified and linearized. Sensor and
output side are electrically isolated. An internal cold junction is
integrated for measurements with thermocouples.
The device outputs a temperature-linear direct current of 4 to
20 mA. As well as the analog transmission of measured values
from 4 to 20 mA, the HART model also supports digital commu-
nication for online diagnostics, measured value transmission
and configuration.
SITRANS TF automatically detects when a sensor should should
be interrupted or is indicating a short-circuit. The practical test
Our field devices for heavy industrial use terminals allow direct measurement of 4 to 20 mA signals over
• HART, Universal an ammeter without interrupting the output current loop.
• 4 to 20 mA, universal Mode of operation of SITRANS TF as field indicator
• Field indicator for 4 to 20 mA signals Any 4 to 20 mA signal can be applied to the generous terminal
The temperature transmitter SITRANS TF works where others block. As well as a range of predefined measurement units, the
feel uncomfortable. adjustable indicator also supports the input of customized units.
This means that any 4 to 20 mA signal can be represented as
■ Benefits any type of unit, e.g. pressure, flow rate, filling level or tempera-
ture.
• Universal use
- as transmitter for resistance thermometer, thermoelement, Power supply
SITRANS TF
Ω or mV signal
0000
- as field indicator for any 4 to 20 mA signals Sensor SITRANS TH300 8
00°C
TC RTD 1 3 4 9 5
• Local sensing of measured values over digital display 2
A D
µP Load
• Rugged two-chamber enclosure in die-cast aluminium or D A
stainless steel
• Type of protection IP68 7
• Test terminals for direct read-out of the output signal without 6
breaking the current loop 1 Analog-to-digital converter
2 Electrical isolation
• Can be mounted elsewhere if the measuring point 3 Microprocessor
- is not easily accessible 4 Digital-to-analog converter
- is subject to high temperatures 5 Power supply
- is subject to vibrations from the system 6 PC/Laptop
- or if you want to avoid long neck tubes and/or protective 7 HART modem
8 Digital display (option)
tubes 9 Test connector
• Can be mounted directly on American-design sensors
• Wide range of approvals for use in potentially explosive atmo-
spheres. "Intrinsically safe, non-sparking and flameproof" type Operating principle: SITRANS TF with an integrated transmitter and
of protection, for Europe and USA. digital display

■ Application
SITRANS TF can be used everywhere where temperatures need
to be measured under particularly adverse conditions, or where
a convenient local display is ideal. Which is why users from all
industries have opted for this field device. The rugged enclosure
protects the electronics. The stainless steel model is even resis-
tant to sea water and other aggressive elements The inner work-
ings offer high measuring accuracy, universal input and a wide
range of diagnostic options.

Siemens FI 01 · 2008 3/53


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitter for field mounting / field indicator
SITRANS TF - Transmitter, two-wire system
and SITRANS TF - Field indicator for 4 to 20 mA

■ Technical specifications Measuring range Parameterizable (see table "Digi-


tal measuring errors")
Input Min. measured span 5 ... 25 Ω (see table "Digital mea-
Resistance thermometer suring errors")
Measured variable Temperature Characteristic Resistance-linear or special char-
acteristic
Sensor type
Thermocouple

3
• to IEC 60751 Pt25 to Pt1000
• to JIS C 1604; a=0.00392 K-1 Pt25 to Pt1000 Measured variable Temperature
• to IEC 60751 Ni25 to Ni1000 Sensor type (thermocouples)
Units °C and °F • Type B Pt30Rh-Pt6Rh to DIN IEC 584
Connection • Type C W5%-Re to ASTM 988
• Normal connection 1 resistance thermometer (RTD) • Type D W3%-Re to ASTM 988
in 2-wire, 3-wire or 4-wire system • Type E NiCr-CuNi to DIN IEC 584
• Generation of average value Series or parallel connection of • Type J Fe-CuNi to DIN IEC 584
several resistance thermometers • Type K NiCr-Ni to DIN IEC 584
in a two-wire system for the gen-
eration of average temperatures • Type L Fe-CuNi to DIN 43710
or for adaptation to other device • Type N NiCrSi-NiSi to DIN IEC 584
types • Type R Pt13Rh-Pt to DIN IEC 584
• Generation of difference 2 resistance thermometers (RTD) • Type S Pt10Rh-Pt to DIN IEC 584
in 2-wire system (RTD 1 – RTD 2 • Type T Cu-CuNi to DIN IEC 584
or RTD 2 – RTD 1)
• Type U Cu-CuNi to DIN 43710
Interface
Units °C or °F
• Two-wire system Parameterizable line resistance
Connection
≤ 100 Ω (loop resistance)
• Standard connection 1 thermocouple (TC)
• Three-wire system No balancing required
• Generation of average value 2 thermocouples (TC)
• Four-wire system No balancing required
• Generation of difference 2 thermocouples (TC)
Sensor current ≤ 0.45 mA
(TC 1 – TC 2 or TC 2 – TC 1)
Response time ≤ 250 ms for 1 sensor with open-
Response time ≤ 250 ms for 1 sensor with open-
circuit monitoring
circuit monitoring
Open-circuit monitoring Can be switched off
Open-circuit monitoring Can be switched off
Short-circuit monitoring Can be switched off (value is
Cold junction compensation
adjustable)
• Internal With integrated Pt100 resistance
Measuring range Parameterizable (see table "Digi-
thermometer
tal measuring errors")
• External With external Pt100 IEC 60571 (2-
Min. measured span 10 °C (18 °F)
wire or 3-wire connection)
Characteristic Temperature-linear or special
• External fixed Cold junction temperature can be
characteristic
set as fixed value
Resistance-based sensors
Measuring range Parameterizable (see table "Digi-
Measured variable Actual resistance tal measuring errors")
Sensor type Resistance-based, potentiome- Min. measured span Min. 50 ... 100 °C (90 ... 180 °F)
ters (see table "Digital measuring
errors")
Units Ω
Characteristic Temperature-linear or special
Connection characteristic
• Normal connection 1 resistance-based sensor (R) in mV Sensor
2-wire, 3-wire or 4-wire system
Measured variable DC voltage
• Generation of average value 2 resistance-based sensors in
2-wire system for generation of Sensor type DC voltage source (DC voltage
average value source possible over an exter-
nally connected resistor)
• Generation of difference 2 resistance-based sensor in
2-wire system (R 1 – R 2 or Units mV
R 2 – R 1)
Response time ≤ 250 ms for 1 sensor with open-
Interface circuit monitoring
• Two-wire system Parameterizable line resistance Open-circuit monitoring Can be switched off
≤ 100 Ω (loop resistance)
Short-circuit monitoring Can be switched off (value is
• Three-wire system No balancing required adjustable)
• Four-wire system No balancing required Measuring range -10 ... +70 mV
-100 ... +1100 mV
Sensor current ≤ 0.45 mA
Min. measured span 2 mV or 20 mV
Response time ≤ 250 ms for 1 sensor with open-
circuit monitoring Overload capacity of the input -1.5 ... +3.5 V DC
Open-circuit monitoring Can be switched off Input resistance ≥ 1 MΩ
Short-circuit monitoring Can be switched off (value is Characteristic Voltage-linear or special charac-
adjustable) teristic

3/54 Siemens FI 01 · 2008


© Siemens AG 2008

SITRANS T measuring instruments for temperature


Transmitter for field mounting / field indicator
SITRANS TF - Transmitter, two-wire system
and SITRANS TF - Field indicator for 4 to 20 mA
Output Certificate and approvals
Output signal 4 ... 20 mA, 2-wire Explosion protection ATEX
Communication with to HART Rev. 5.9 • "Intrinsically-safe" type of With digital indicator:
SITRANS TH300 protection II 2 (1) G EEx ia IIC T4
Digital display Without digital indicator:
II 2 (1) G EEx ia IIC T6
Digital display (optional) in current loop
- EC type test certificate ZELM 99 ATEX 0007
Display
Digit height
max. 5 digits
9 mm (0.35")
• "Operating equipment that is
non-sparking and has limited energy
II 3G EEx nAL IIC T6/T4 3
Display range -99999 ... +99999 for zone 2" type of protection

Units Any (max. 5 char.) - EC type test certificate ZELM 99 ATEX 0007

Setting: With 3 keys • Flame-proof enclosure" type of II 2 G EEx d IIC T5/T6


Zero point, upper range value and protection
unit - EC type test certificate CESI 99 ATEX 079
Measuring accuracy Explosion protection to FM Certificate of Compliance
Digital measuring errors See table "Dig. measuring errors" 3017742
• Identification (XP, DIP, NI, S) • XP / I / 1/BCD / T5 Ta = 85 °C
Reference conditions (185 °F), T6 Ta = 50 °C (112 °F),
• Auxiliary power supply 24 V ± 1 % Type 4X
• DIP / II, III / 1 / EFG / T5 Ta =
• Load 500 Ω 85 °C (185 °F), T6 Ta = 50 °C
• Storage temperature 23 °C (73.4 °F) (112 °F), Type 4X
• NI / I / 2 / ABCD / T5 Ta = 85 °C
• Warming-up time > 5 min (185 °F), T6 Ta = 50 °C (112 °F),
Error in the analog ouput < 0.1% of span Type 4X
(digital/analog converter) • S / II, III / 2 / FG / T5 Ta = 85 °C
(185 °F), T6 Ta = 50 °C (112 °F),
Error due to internal cold junction < 0.5 °C (0.9 °F) Type 4X
Temperature effect < 0.1% of max. span/10 °C Hardware and software require-
(18 °F) ments
Power supply effect < 0.005% of span/V • For the parameterization software
Long-term drift SIPROM T for SITRANS TH200
• in the first month < 0.02% of max. span - Personal computer PC with CD-ROM drive and
USB/RS 232 interface
• after one year < 0.03% of max. span
- PC operating system Windows 98, NT, 2000, XP
• after 5 years < 0.04% of max. span
• For the parameterization software See chapter 9, "Software",
Rated conditions SIMATIC PDM for SITRANS TH300 "SIMATIC PDM"
Ambient temperature Communication
Storage temperature -40 ... +85 °C (-40 ... +185 °F) Load for HART connection 230 ... 1100 Ω
Condensation Permissible • Two-core shielded ≤ 3.0 km (1.86 mi)
Electromagnetic compatibility According to EN 61326 and • Multi-core shielded ≤ 1.5 km (0.93 mi)
NAMUR NE21
Protocol HART protocol, version 5.x
Degree of protection to EN 60529 IP68
Design Factory setting (transmitter):
Approx. weight Approx. 1.5 kg (3.3 lb), without
• Pt100 (IEC 751) with three-wire circuit
options • Measuring range: 0 ... 100 °C (32 ... 212 °F)
Dimensions See "Dimensional drawings" • Fault current 22.8 mA
Enclosure material Die-cast aluminum, low in copper, • Sensor offset: 0 °C (0 °F)
GD-AlSi 12 or stainless steel, • Damping 0.0 s
polyester-based lacquer, stain-
less steel rating plate
Electrical connection, Screw terminals, cable inlet via
sensor connection M20 x 1.5 or ½-14 NPT threaded
gland
Mounting bracket (optional) Steel, galvanized and chrome-
plated or stainless steel
Power supply
Without digital display 11 ... 35 V DC (30 V with Ex)

With digital display 13.1 ... 35 V DC (30 V with Ex)


Electrically isolated Between input and output
• Test voltage U eff = 1 kV, 50 Hz, 1 min

Modification 01/2008 Siemens FI 01 · 2008 3/55


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitter for field mounting / field indicator
SITRANS TF - Transmitter, two-wire system
and SITRANS TF - Field indicator for 4 to 20 mA
Digital measuring errors Thermocouple elements
Resistance thermometer Input Measuring range Min. mea- Digital
sured span accuracy
Input Measuring range Min. mea- Digital
sured span accuracy °C (°F) °C (°F) °C (°F)
°C (°F) °C (°F) °C (°F) Type B 0 ... 300 100 (180) 3 (5.4)
(32 ... 572)
according to
300 ... 1820 100 (180) 2 (3.6)
3
IEC 60751
(572 ... 3308)
Pt25 -200 ... + 850 10 (18) 0,2 (0.36)
(-328 ... +1562) Type C (W5) 0 ... 2300 100 (180) 2 (3.6)
(32 ... 4172)
Pt50 -200 ... + 850 10 (18) 0,15 (0.27)
(-328 ... +1562) Type D (W3) 0 ... 1750 100 (180) 1 (1.8)
(32 ... 3182)
Pt100 ... Pt200 -200 ... + 850 10 (18) 0,1 (0.18)
(-328 ... +1562) 1750 ... 2300 100 (180) 2 (3.6)
(3182 ... 4172)
Pt500 -200 ... + 850 10 (18) 0,15 (0.27)
(-328 ... +1562) Type E -200 ... +1000 50 (90) 1 (1.8)
(-328 ... +1832)
Pt1000 -200 ... + 350 10 (18) 0,15 (0.27)
(-328 ... +662) Type J -210 ... +1200 50 (90) 1 (1.8)
(-346 ... +2192)
according to
JIS C1604-81 Type K -200 ... +1370 50 (90) 1 (1.8)
(-328 ... +2498)
Pt25 -200 ... + 649 10 (18) 0,2 (0.36)
(-328 ... +1200) Type L -200 ... +900 50 (90) 1 (1.8)
(-328 ... +1652)
Pt50 -200 ... + 649 10 (18) 0,15 (0.27)
(-328 ... +1200) Type N -200 ... +1300 50 (90) 1 (1.8)
(-328 ... +2372)
Pt100 ... Pt200 -200 ... + 649 10 (18) 0,1 (0.18)
(-328 ... +1200) Type R -50 ... +1760 100 (180) 2 (3.6)
(-58 ... +3200)
Pt500 -200 ... + 649 10 (18) 0,15 (0.27)
(-328 ... +1200) Type S -50 ... +1760 100 (180) 2 (3.6)
(-58 ... +3200)
Pt1000 -200 ... + 350 10 (18) 0,15 (0.27)
(-328 ... +662) Type T -200 ... +400 40 (72) 1 (1.8)
(-328 ... +752)
Ni 25 ... Ni1000 -60 ... + 250 10 (18) 0,1 (0.18)
(-76 ... +482) Type U -200 ... +600 50 (90) 2 (3.6)
(-328 ... +1112)

Resistance-based sensors
mV sensors
Input Measuring range Min. mea- Digital
sured span accuracy Input Measuring range Min. mea- Digital
sured span accuracy
Ω Ω Ω
mV mV µV
Resistance 0 ... 390 5 0,05
mV sensors -10 ... +70 2 40
Resistance 0 ... 2200 25 0,25
mV sensors -100 ... +1100 20 400

The digital accuracy is the accuracy after the analog/digital


conversion including linearization and calculation of the mea-
sured value.
An additional error is generated in the output current 4 to 20 mA
as a result of the digital/analog conversion of 0.1% of the set
span (digital-analog error).
The total error under reference conditions at the analog output is
the sum from the digital error and the digital-analog error (poss.
with the addition of cold junction errors in the case of thermocou-
ple measurements).

3/56 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitter for field mounting / field indicator
SITRANS TF - Transmitter, two-wire system
and SITRANS TF - Field indicator for 4 to 20 mA

■ Dimensional drawings
13 (0.51) 144 (5.67) 55 80 (3.15)
15* (0.59) 138* (5.43) (2.17) b 100* (3.94)

(0.79)
52 (2.05) 28 6 2

20
50* (1.97) (1.10)

(1.97)
3

50
80 (3.15)

117 (4.61)

a
237 (9.33)
8 5 4

68 (2.68)

120 (4.72)
1

36,5 (1.44)
72 (2.83)

a: max. 164 (6.46) (M20x1.5)


max. 189 (7.44) (½-14 NPT)
b: max. 25 (0.98) (M20x1.5)
max. 50 (1.97) (½-14 NPT)
123 (4.84) 7
*) Dimensions for stainless Æ 50 ... 60
steel enclosure (1.97 ... 2.36)

105 (4.13)
1 Sensor connection (screwed gland M20x1,5 or ½-14 NPT)
2 Blanking plug 6 Protective cover (without function)
3 Electrical connection (screwed gland M20x1,5 or ½-14 NPT) 7 Mounting bracket (option) with clamp for
4 Terminal side, output signal securing to a vertical or horizontal pipe
5 Terminal side, sensor 8 Cover with window for digital display

SITRANS TF, dimensions in mm (inches)

Siemens FI 01 · 2008 3/57


© Siemens AG 2008

SITRANS T measuring instruments for temperature


Transmitter for field mounting / field indicator
SITRANS TF - Transmitter, two-wire system
and SITRANS TF - Field indicator for 4 to 20 mA

Selection and Ordering data Order No. Selection and Ordering data Order No.
Temperature transmitter in field housing D) 7 NG 3 1 3 7 - 77777 Accessories
Two-wire system 4 ... 20 mA, with electrical Modem for SITRANS TH200
isolation, with documentation on incl. parameterization software T
CD-ROM
• with USB interface } 7NG3092-8KU
Integrated transmitter C)
• SITRANS TH200, programmable
3 - without Ex protection
- with EEx ia
5
5
0
1
• with RS 232 interface } 7NG3092-8KM
C)

CD for measuring instruments } A5E00364512


- with EEx nAL for zone 2 5 2 for temperature G)
- total device SITRANS TF EEx d1) 5 4 with documentation in German, English,
- total device SITRANS TF according to FM 5 5 French, Spanish, Italian and Portuguese, and
(XP, DIP, NI, S)1) parameterization software SIPROM T
• SITRANS TH300, communication capability (included in delivery with SITRANS TF)
according to HART V 5.9
HART modem
- without Ex-protection 6 0
- with EEx ia 6 1 • with RS 232 interface } 7MF4997-1DA
D)
- with EEx nAL for zone 2 6 2
- total device SITRANS TF EEx d1) 6 4 • with USB interface } 7MF4997-1DB
D)
- total device SITRANS TF according to FM 6 5
(XP, DIP, NI, S)1) SIMATIC PDM parameterization software see chap. 9
SITRANS TF field indicator 7 NG 3 1 3 7 - 77777 also for SITRANS TH300
for 4 ... 20 mA signals, Mounting bracket and securing parts
with documentation on CD-ROM • made of steel for 7NG313.-..B.. 7MF4997-1AC
• without Ex-protection 0 0 1 • made of steel for 7NG313.-..C.. 7MF4997-1AB
• with EEx ia 0 1 1 • made of stainless steel for 7NG313.-..B.. } 7MF4997-1AJ
• with EEx nAL for zone 2 0 2 1 • made of stainless steel for 7NG313.-..C.. 7MF4997-1AH
• total device SITRANS TF EEx d1) 0 4 1 Digital indicator1) 7MF4997-1BS
• total device SITRANS TF according to FM 0 5 1 } Available ex stock.
(XP, DIP, NI, S)1) Power supply units see "SITRANS I supply units and input isolators".
Enclosure 1)
It is not possible to upgrade devices with Ex protection
• die-cast aluminium A
• stainless steel precision casting E C) Subject to export regulations AL: N, ECCN: EAR99.
D) Subject to export regulations AL: N, ECCN: EAR99H.
Connections/cable inlet
G) Subject to export regulations AL: N, ECCN: 5D992B1.
• screwed glands M20x1.5 B
• screwed glands ½-14 NPT C
Digital indicator Factory setting (transmitter):
• without 0 • Pt100 (IEC 751) with three-wire circuit
• with 1
• Measuring range: 0 ... 100 °C (32 ... 212 °F)
Mounting bracket and securing parts • Fault current 22.8 mA
• without 0
• made of steel 1
• Sensor offset: 0 °C (0 °F)
• made of stainless steel 2 • Damping 0.0 s
Further designs Order code
Please add "-Z" to Order No. and specify
Order code(s) and plain text.
Customer-specific setting of operating data Y 0 1 2)
Inscription on measuring-point label
• measuring range (max. 27 characters) Y2 2 3)
• meas. point description (max. 16 char.) Y2 3 3)
• measuring point text (max. 27 char.) Y2 4 3)
Test protocol (5 measuring points) C1 1
Power supply units see "SITRANS I supply units and input isolators".
1)
Without cable gland.
2)
Y01: Please specify all data that does not correspond to factory settings
(see above) (e.g. Y01 = thermocouple element type K; internal cold junc-
tion; 0 ... 800 °C; fault current 3.6 mA).
3)
Y22, Y23, Y24: If no order is placed for Y01, these data are only noted on
the measuring point label and are not programmed in the transmitter.

3/58 Siemens FI 01 · 2008 Modification 01/2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Transmitter for field mounting / field indicator
SITRANS TF - Transmitter, two-wire system
and SITRANS TF - Field indicator for 4 to 20 mA

■ Schematics

Resistance thermometer Resistance Thermocouple

7ZRZLUHV\VWHP  7ZRZLUHV\VWHP  &ROGMXQFWLRQFRPSHQVDWLRQ


,QWHUQDOIL[HGYDOXH


7KUHHZLUHV\VWHP 7KUHHZLUHV\VWHP &ROGMXQFWLRQFRPSHQVDWLRQZLWK


H[WHUQDO3WLQWZRZLUHV\VWHP 

)RXUZLUHV\VWHP )RXUZLUHV\VWHP &ROGMXQFWLRQFRPSHQVDWLRQZLWK


H[WHUQDO3WLQWKUHHZLUHV\VWHP

*HQHUDWLRQRIDYHUDJH *HQHUDWLRQRIDYHUDJH *HQHUDWLRQRIDYHUDJHYDOXHGLIIHUHQFH


YDOXHGLIIHUHQFH  YDOXHGLIIHUHQFH  ZLWKLQWHUQDOFROGMXQFWLRQFRPSHQVDWLRQ

3URJUDPPDEOHOLQHUHVLVWDQFHIRUWKHSXUSRVHRIFRUUHFWLRQ

Voltage measurement Current measurement

SITRANS TF, sensor connection assignment

Siemens FI 01 · 2008 3/59


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Resistance thermometers
Technical description

■ Design ■ Function
A resistance thermometer comprises Measuring resistor
• the measuring resistor (metal; platinum, Pt or nickel, Ni) and Measuring resistors Measuring resistors are suit-
• the mounting and connection parts required in each case. able for temperatures from

In the standard version, the measuring resistors are embedded Made of platinum -200 ... +850 °C
(-328 ... +1562 °F)
in ceramic. In the case of special vibration resistance require-
3 ments, the Pt measuring resistors are double-wound and fused
into glass.
Made of nickel -60 ... +150 °C (-76 ... 302 °F),
briefly up to 180 °C (356 °F)
• Measuring resistors of class B are supplied. A range of resis- The measuring resistor changes with the temperature in accor-
tors of class A or 1/3 to 1/10 class B is available on request. dance with a certain reproducible series of calibration data (see
• Single and double resistance thermometers are available. the table "Calibration data for platinum measuring resistors (to
DIN EN 60751)“ under "Technical data“).
The changes in resistance are transmitted as changes in voltage
to indicators, recorders or controllers directly via copper wires or
via transmitters. The type of measuring circuit depends on the
instrument to be connected and the required measuring range.
The measuring resistors are balanced at 0 °C (32 °F) to
100 Ω ± 0.12 Ω . The calibration values of the resistors (i.e. the
dependence of the resistor on the temperature) and the permis-
sible deviations are specified in DIN EN 60751 (IEC 751) (see
the table "Error limits according to DIN EN 60751“ under "Tech-
nical Specifications“).
Measuring resistors of class B are supplied. A range of resistors
of class A or 1/3 to 1/10 class B is available on request.
Measuring principle of the resistance temperature measure-
ment
The thermometer is heated up by the thermometer current com-
pared to the material to be measured. The heating-up error thus
produced increases with the square of the thermometer current
and linearly with the resistance of the measuring resistor. Apart
from the magnitude of the thermometer current, the error de-
pends on the design of the thermometer and on the heat transfer
between the protective tube and the medium. A high measuring
power is required for output instruments operating according to
the deflection method. To keep the heating-up error within per-
missible limits, the thermometer current should not be greater
than 10 mA in this case.
When measuring the temperature of gases with very small flow
velocities, a considerably greater heating-up error occurs than
in measurements with very fast gases or liquids. The heating-up
error is negligibly small with high flow velocities.
Components and circuitry of a resistance thermometer
Protection fitting/protective tubes
To protect the measuring resistor for industrial measurements
and to enable easy replacement, it is fixed in a measuring insert Suitable protective fittings are used for installation in pipelines,
(4) which in turn is fitted in a protective tube (5). The measuring tanks etc. depending on the mechanical or chemical require-
insert is spring-mounted in the connection head (1) of the pro- ments.
tective tube using two screws. The internal conductor (10) in the The materials for the protective tubes or combinations of differ-
measuring insert connects the measuring resistor (11) to the ter- ent protective tube materials must be carefully selected in order
minals on the terminal block. to meet the requirements due to static pressure, flow and tem-
perature Furthermore, the indication response should be as fast
Depending on the measuring range and the accuracy required, as possible.
the thermometers are connected in two-wire, three-wire or four-
wire systems to the output devices. Installation examples with the suitable protective tube materials
can be found under "Technical data" in the table "Installation ex-
The measuring inserts are therefore available with two, three or amples and material of protective tubes".
four internal conductors. If the resistance of the internal conduc-
tor is negligibly small, measuring inserts with only two internal The type of installation of the protective tubes depends on the
conductors can be used for the three-wire and four-wire sys- application. The protective tubes are screwed into the pipelines
tems. in the case of operating pressures up to approx. 90 bar. Tapered
protective tubes which can be welded in are available for higher
Exact balancing of the internal conductor under operating con- pressures. Thermometers for measuring furnace temperatures
ditions is only possible with three conductors. If the resistance of are fixed using flanges.
the internal conductor is greater than 0.2 Ω, its magnitude is in-
dicated on the mounting flange of the measuring insert. Owing to the different operating conditions, no guarantee can be
given for protective fittings. The manufacturer is responsible for
damages and measuring errors caused by wrong installation in
compliance with the General Terms of Delivery if the instruments
have been installed by the manufacturer and if the specifications
for the operating conditions furnished by the customer were cor-
rect and sufficiently detailed.

3/60 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Resistance thermometers
Technical description

■ Integration
Instrument combination for measuring and controlling tem-
perature

Sense
Resistance thermometer

3
Match
Electric transmitter
Catalog FI 01

Output Electric Catalog


Control indicator MP 12
limit monitor
recorder MP 20

controller MP 31

Instrument combination with resistance thermometer as sensor


Installation examples and materials of protective tubes
Measuring point Max. operat- Protective tube material Measuring point Max. operat- Protective tube material
ing temp. ing temp.
°C (°F) name No. °C (°F) name No.
A. Steam power plants D. Dye works
Water and steam lines 300 (572) Bronze Sn Bz 6 (only 2.1020 Jigger, automatic yarn 110 (230) X 6 CrNiMoTi 17 122 1.4571
(Screw-in and welding- for water) skein dying machine
type thermometers) 400 (752) St 35.8 1.0305 E. Food and drink industries
540 (1004) 13 CrMo 44 1.7335
570 (1058) 10 CrMo 9 10 1.7380 1. Breweries

Flue gas 550 (1022) St 35.8, enamelled 1.0305 Brewing water 80 (176) Bronze Sn Bz 6 or X 6 2.1020
CrNiTi 18 10
Pulverized coal/air mix- 100 (212) St 35.8 (with baffle rod) 1.0305
ture line Mash

Water treatment 30 (86) X 6 CrNiTi 18 10 or 1.4541 • Hot wort 100 (212) Bronze or X 6 CrNiTi 18 1.4541
X 6 CrNiMoTi 17 122 1.4571 10

B. Paper mills • Cold wort 4 (39,2) X 6 CrNiTi 18 10 1.4541

In paper pulp (cylindri- 60 (140) X 6 CrNiMoTi 17 122 1.4571 2. Sugar factories


cal paper mills, hand- Salt removal in sugar 100 (212) X 6 CrNiTi 18 10 1.4541
made paper, refiner) juice
C. Pulp production For all tanks with inter- 3. Cereals production
nal lining: only flange-
type thermometers Salt removal in whey 20 (68) X 6 CrNiTi 18 10 1.4541
1. Sulfite pulp 4. Malt houses
Boiling acid in vat, 150 (302) X 6 CrNiMoTi 17 122 1.4571 Steeping water 100 (212) Bronze Sn Bz 6 2.1020
peetz tank and acid 5. Yeast production
tower
Yeast cooling 4 (39,2) X 6 CrNiTi 18 10 1.4541
Hypochlorite tower, 40 (104) X 6 CrNiMoTi 17 122 1.4571
alkali tower Yeast fermentation 33 (91,4) X 6 CrNiMoTi 17 122 1.4571
Sulphite lye evapora- 140 (284) X 6 CrNiMoTi 17 122 1.4571 F. Chemical and petrochemical industries
tion, heat exchanger, Corrosion-resistant steel, material Nos. 1.4541 and 1.4571, can be
preheater and lye col- used for many applications. The numerous substances to be measured
lecting tank in these industrial branches render it sometimes difficult to recommend
2. Sulfate pulp suitable protective tube materials. If required, please contact our repre-
sentatives in these cases.
Vat, lye heater In vat 180 X 6 CrNiTi 18 10 or 1.4541
Black, green and white (356) other- X 6 CrNiMoTi Ti 17 122 1.4571
lye tanks wise
80 (176)
Cellulose multiple- 40 (104) Hastelloy C (59 Ni; 16 1.4571
stage bleaching (moist Mo; 15,5 Cr; 5,5 Fe; 3,8
chloric gases present) W) or X 6 CrNiMoTi 17
122 with Ti protective
sleeve
Sulfate lye evaporation, 140 (284) X 6 CrNiMoTi 17 122 1.4571
heat exchanger, pre-
heater and lye collect-
ing tank

Siemens FI 01 · 2008 3/61


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Resistance thermometers
Protective tubes, installation

■ Technical specifications Error limits acc. to DIN EN 60 751


The resistance thermometers are divided into two classes ac-
Calibration data for platinum measuring resistors cording to their error limits:
(to DIN EN 60751)
Class Error limits in °C
°C (°F) Ω
A 0,15 + 0,002 |t| 1)
-200 (-328) 18.52
-180 (-292) 27.10 B 0,3 + 0,005 |t|

3 -160
-140
(-256)
(-220)
35.34
43.88
1)
|t| is the numerical value of the temperature in °C without consideration of
the sign
-120 (-184) 52.11
-100 (-148) 60.26 ■ Integration
-80 (-112) 68.33
-60 (-76) 76.33 Protective tubes in a pipeline
-40 (-40) 84.27
-30 (-22) 88.22
-20 (-4) 92.16
-10 (14) 96.09
0 (32) 100.00
10 (50) 103.90
20 (68) 107.79
30 (86) 111.67
40 (104) 115.54
50 (122) 119.40
60 (140) 123.24
80 (176) 130.90
100 (212) 138.51
120 (248) 146.07
140 (284) 153.58
160 (320) 161.05
180 (356) 168.48
200 (392) 175.86 Flue gas resistance thermometer
220 (428) 183.19
240 (464) 190.47
260 (500) 197.71
280 (536) 204.90
300 (572) 212.05
320 (608) 219.15
340 (644) 226.21
360 (680) 233.21
380 (716) 240.18
400 (752) 247.09
420 (788) 253.96
440 (824) 260.78
460 (860) 267.56
480 (896) 274.29
500 (932) 280.98
520 (968) 287.62
540 (1004) 294.21
560 (1040) 300.75
580 (1076) 307.25
600 (1112) 313.71
620 (1148) 320.12
640 (1184) 326.48
660 (1220) 332.79
700 (1292) 345.28
750 (1382) 360.64
800 (1472) 375.70
850 (1562) 390.48

Fitted in a sheet-metal duct (top) and a flue gas duct (bottom)

3/62 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Resistance thermometers
Protective tubes, installation
Screw-in resistance thermometer Welding-type protective tubes, mounted

In a pulverized coal line, with baffle rod Top: Wall thickness ≤ 35 mm; mounting using welded coupling; bottom:
Wall thickness > 35 mm
Welding-type protective tubes, mounted

At top for low-pressure, with screw socket G½; at bottom for high-pres-
sure, with screw socket G1
It must be possible to weld the pipe coupling and pipeline. Ma-
terial according to data provided by pipeline or vessel manufac-
turer.

Siemens FI 01 · 2008 3/63


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Resistance thermometers
Temperature transmitters
for mounting in the connection head

■ Overview ■ Selection and Ordering Data


Detailed information on the transmitters can be found for the re-
spective products under "Transmitters for temperature“.

Transmitter to be fitted Order code

3
To order the sensor with a built-in
temperature transmitter, add "-Z" to
the Order No. of the sensor, and
supplement by the following Order
code:
• SITRANS TH100
- without Ex T10
- EEx ia IIC and EEx n for zone 2 T11
- FM T13
• SITRANS TH200
- without Ex T20
- EEx ia IIC and EEx n for zone 2 T21
The following temperature transmitters are available for mount-
- FM (IS, I, NI) T23
ing in the connection head:
• SITRANS TH300
SITRANS TH100
- without Ex T30
Programmable two-wire temperature transmitter (4 to 20 mA),
- EEx ia IIC und EEx n for zone 2 T31
without electrical isolation, only for Pt100 resistance thermome-
ters. - FM (IS, I, NI) T33

SITRANS TH200 • SITRANS TH400 PA


- without Ex T40
Programmable two-wire temperature transmitter (4 to 20 mA),
electrical isolation for resistance thermometers and thermocou- - EEx ia T41
ple elements. • SITRANS TH400 FF
SITRANS TH300 - without Ex T45
Two-wire temperature transmitter with HART communication - EEx ia T46
(4 to 20 mA), electrical isolation for resistance thermometers and Customer-specific setting of the Y111)
thermocouple elements. built-in transmitter (specify settings
in plain text)
SITRANS TH400
1)
For TH400 FF available soon
Temperature transmitter with PROFIBUS PA or FOUNDATION
Fieldbus connection, electrical isolation for resistance thermom-
eters and thermocouple elements.
Note:
• SITRANS TH100/TH200/TH300/TH400 can be fitted instead of
the terminal block or in the high hinged cover. Additional fitting
only possible in high hinged cover.
• If using intrinsically-safe temperature sensors any installed
temperature transmitters must also be intrinsically-safe.

3/64 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Resistance thermometers
Questionnaire for temperature sensors
(resistance thermometers and thermocouples)

■ General information
Customer: ....................................................................................
Address: ......................................................................................
Contact partner: .........................................................................
Purchasing dept.: ........................................................................ Tel.: ...........................................................................
Sales dept.: ................................................................................. Tel.: ........................................................................... 3
Process dept.: ............................................................................. Tel.: ...........................................................................
Inquiry:.........................................................................................
Quotation: ....................................................................................
Place and date: ...........................................................................

■ Operating conditions ■ Miscellaneous


1. Application: ...................................................................... Please additionally provide the following: rough sketch, installa-
(e.g. exhaust gas measurement) tion diagram, section of drawing, photo

2. Location:...........................................................................
■ Sensor design
(e.g. pipe bend, tank)
1. Measuring element...........................................................
3. Mounting position:............................................................ (type and standard) (e.g. Pt100 or TC type K)
(e.g. vertical, 45° against flow)
1.1. Tolerance:.........................................................................
4. Temperature (measuring point): ......................................
1.2. Design: .............................................................................
Operating temperature:....................................................
(e.g. Pt100 or 2, 3 or 4-wire system)
Temperature range: .........................................................
1.3. Degree of protection/type of protection: ..........................
5. Medium: ...........................................................................
2. Protective fitting:...............................................................
6. Pressure: ..........................................................................
Nominal pressure: ............................................................ 2.1. Protective tube: ................................................................
Operating pressure: ......................................................... (dimensions/material)
7. Flow: ................................................................................. 2.2. Mounting: .........................................................................
(dimensions/material)
8. Vibrations: ........................................................................
2.3. Neck tube:........................................................................
9. Miscellaneous: .................................................................
(dimensions/material)
(e.g. vessel or pipe materials, PTFE lining)
2.4. Mounting length/nominal length:......................................
■ Ambient conditions 3. Material certificates: .........................................................
(e.g. seawater atmosphere, chemical plant)
4. Connection: ......................................................................
Definition:.................................................................................
4.1. Connection head/box:......................................................
.................................................................................................
4.2. Cable:...............................................................................
.................................................................................................
(dimensions/insulation/standard)
■ Special information 4.3. Other: ...............................................................................
1. Mounting of temperature transmitter in connection head: .........................................................................................
......................................................................................... 5. Tests: ................................................................................
......................................................................................... .........................................................................................
2. Packaging regulations: .................................................... 6. Accessories:.....................................................................
......................................................................................... .........................................................................................
.........................................................................................
7. Supplementary requirements:..........................................
.........................................................................................

Siemens FI 01 · 2008 3/65


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Resistance thermometers
Flue gas resistance thermometers
with connection head

■ Overview Selection and Ordering data Order No.


Flue gas resistance thermometer
Measuring resistor
(winding) embedded in ceramic
1 Pt100 measuring resistor,
three-wire circuit
Mounting length/ Weight/

3 mm (inch):
• 500 (19.7)
• 710 (28.0)
kg (lb):
0.9 (1.98)
1.1 (2.43)
7 MC 1 0 0 0
7 MC 1 0 0 0
-
-
1 BA 2 7
2 BA 2 7
• 1000 (39.4) 1.5 (3.31) 7 MC 1 0 0 0 - 3 BA 2 7
• 1400 (55.1) 1.9 (4.19) 7 MC 1 0 0 0 - 4 BA 2 7
• 2000 (78.7) 2.7 (5.95) 7 MC 1 0 0 0 - 5 BA 2 7
Connection head, form B,
The flue gas resistance thermometer with connection head is made of cast light alloy,
suitable for the temperature range from -50 to +600 °C with 1 cable inlet and
(-58 to +1112 °F) and can also be supplied with a built-in tem- • screw cover 1
perature transmitter. • standard hinged cover 4
Please order mounting flange or threaded sleeve separately. • high hinged cover 6
Further designs Order code
■ Technical specifications Please add "-Z" to Order No. and specify
Order code(s) and plain text.
Design According to DIN 43764:
Thermometer without mount • Different design (mounting length, protec- Y01
tive tube material etc.), specify in plain text.
Protective tube • TAG plate made of stainless steel Y15
• Form 1, DIN 43772; cylindrical, 15 mm specify TAG No. in plain text
diameter (0.59 inch), wall thick- • Calibration carried out at one point, specify Y33
ness 3 mm (0.12 inch), seamless desired temperature in plain text (order
equivalent number of times for several cal-
• Material St 35.8, mat. No. 1.0305, enam- ibration points).
elled
Accessories Order No.
• Loading capacity 1 bar (14.5 psi) above atmo-
spheric, to DIN 43772 Mounting flange 7 MC 2 9 9 8 - 5 C A
Adjustable, to DIN 43734;
Measuring insert Replaceable, with measuring
Material: GTW 35, mat. No. 0.8035,
insert tube (8 mm diameter
for protective tube diameter
(0.31 inch)) made of stainless
15 mm (0.59 inch),
steel; terminal block with clamp-
0.3 kg (0.66 lb)
ing springs
Gas-tight threaded sleeve
■ Dimensional drawings Material: 9 SMnPb 28
Material No. 1.0718,
for protective tube diameter
15 mm (0.59 inch),
0.4 kg (0.88 lb)
• G¾ internal thread with gasket 7 MC 2 9 9 8 - 5 D A
• G½ internal thread with gasket 7 MC 2 9 9 8 - 5 D C
To order a temperature transmitter installed in the connection head,
see "Temperature transmitters for installation in the connection
head" (page 3/64).
Individual parts: Measuring inserts, see "Accessories".

Flue gas resistance thermometer with connection head, dimensions in


mm (inches)

3/66 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Resistance thermometers
Low-pressure screw-in resistance thermometers
with connection head, without neck tube

■ Overview Selection and Ordering data Order No.


Low-pressure screw-in resistance
thermometer with connection head,
without neck tube
Protective tube and screw socket
X 6 CrNiMoTi 17 122, mat. No. 1.4571
Protective tube to DIN 43,772,
form: similar to 2G
Measuring resistors embedded in ceramic,
one Pt100 measuring resistor: three-wire
circuit, Two Pt100 measuring resistors:
3
two-wire circuit
Measuring insert not explosion protected
One Pt100 measuring resistor,
Mounting length U1/ Weight/
The low-pressure screw-in resistance thermometer with connec- mm (inch): kg (lb):
tion head (no neck tube) is suitable for the temperature range • 100 (3.9) 0.6 (1.32) } 7 MC 1 0 0 6 - 1 DA 1 7
from -50 to +400 °C (-58 to +752 °F) and can also be supplied • 160 (6.3) 0.7 (1.54) 7 MC 1 0 0 6 - 2 DA 1 7
with a built-in temperature transmitter. • 230 (9.1) 0.8 (1.76) 7 MC 1 0 0 6 - 3 DA 1 7
• 360 (14.2) 0.9 (1.98) 7 MC 1 0 0 6 - 4 DA 1 7
■ Technical specifications • 510 (20.1) 1.0 (2.20) 7 MC 1 0 0 6 - 5 DA 1 7
Two Pt100 measuring resistors
Design According to DIN 43765: Mounting length U1/ Weight/
Screw-in thermometer mm (inch): kg (lb):
Protective tube • 100 (3.9) 0.6 (1.32) 7 MC 1 0 0 6 - 1 DB 1 7
• Form Similar to 2G, DIN 43772; cylindri- • 160 (6.3) 0.71 (1.57) 7 MC 1 0 0 6 - 2 DB 1 7
cal, 9 mm (0.35 inch) diameter, • 230 (9.1) 0.81 (1.79) 7 MC 1 0 0 6 - 3 DB 1 7
wall thickness 1 mm (0.04 inch) • 360 (14.2) 0.91 (2.01) 7 MC 1 0 0 6 - 4 DB 1 7
• Loading capacity to 20 bar (290.1 psi) (loading • 510 (20.1) 1.01 (2.23) 7 MC 1 0 0 6 - 5 DB 1 7
capacity dependent on material, Measuring insert with explosion protection
temperature, flow rate, mounting
One Pt100 measuring resistor
length etc., see DIN 43772 for
details) Mounting length U1/ Weight/
mm (inch): kg (lb):
Screw socket G½; suitable is gasket 21 x 26, • 100 (3.9) 0.6 (1.32) } 7 MC 1 0 0 6 - 1DE 17
similar to form C or D, DIN 7603
• 160 (6.3) 0.7 (1.54) 7 MC 1 0 0 6 - 2DE 17
Measuring insert Replaceable, with measuring • 230 (9.1) 0.8 (1.76) 7 MC 1 0 0 6 - 3DE 17
insert tube (6 mm diameter • 360 (14.2) 0.9 (1.98) 7 MC 1 0 0 6 - 4DE 17
(0.24 inch) made of stainless
steel; terminal block with clamp- • 510 (20.1) 1.0 (2.20) 7 MC 1 0 0 6 - 5DE 17
ing springs Two Pt100 measuring resistors
Mounting length U1/ Weight/
Response times (to VDI/VDE 3 522) mm (inch): kg (lb):
• In water with flow velocity t 0.5 = 25 s, t0.9 = 75 s • 100 (3.9) 0.6 (1.32) 7 MC 1 0 0 6 - 1DF 17
v = 0.4 m/s (1.31 ft/s) • 160 (6.3) 0.71 (1.57) 7 MC 1 0 0 6 - 2DF 17
• In air with flow velocity v = 1 m/s t 0.5 = 2 min, t0.9 = 6.3 min • 230 (9.1) 0.81 (1.79) 7 MC 1 0 0 6 - 3DF 17
(3.28 ft/s) • 360 (14.2) 0.91 (2.01) 7 MC 1 0 0 6 - 4DF 17
Explosion protection II 1/2G EEx ia IIC T4/T6 • 510 (20.1) 1.01 (2.23) 7 MC 1 0 0 6 - 5DF 17
Connection head, form B,
■ Dimensional drawings • made of cast light alloy,
with 1 cable inlet and
- screw cover 1
- standard hinged cover 4
- high hinged cover } 6
• made of stainless steel, 7
with 1 cable inlet and screw cover
Further designs Order code
Please add "-Z" to Order No. and specify
Order code(s) and plain text.
• Different design (mounting length, protec- Y01
tive tube material etc.), specify in plain text.
• TAG plate made of stainless steel Y15
Low-pressure screw-in resistance thermometer with connection head, specify TAG No. in plain text
without neck tube, dimensions in mm (inches) • Calibration carried out at one point, specify Y33
desired temperature in plain text (order
equivalent number of times for several
calibration points).
} Available ex stock
To order a temperature transmitter installed in the connection head,
see "Temperature transmitters for installation in the connection head"
(page 3/64).
Individual parts: Measuring inserts and connection heads,
see "Accessories".

Siemens FI 01 · 2008 3/67


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Resistance thermometers
Low-pressure screw-in resistance thermometers
with connection head, with neck tube

■ Overview Selection and Ordering data Order No.


Low-pressure screw-in resistance ther-
mometer with connection head and neck
tube
Protective tube and screw socket
X 6 CrNiMoTi 17 122, mat. No. 1.4571
Protective tube to DIN 43772,
form: 2G
3 Measuring resistors embedded in ceramic,
one Pt100 measuring resistor: three-wire
circuit, Two Pt100 measuring resistors:
two-wire circuit
Measuring insert not explosion protected
One Pt100 measuring resistor
Mounting length U1/ Weight/
The low-pressure screw-in resistance thermometer with connec- mm (inch): kg (lb):
tion head and neck tube is suitable for the temperature range • 160 (6.3) 0.8 (1.76) 7 MC 1 0 0 7 - 5 D A 1 7
from -50 to +600 °C (-58 to +1112 °F) and can also be supplied • 250 (9.84) 0.9 (1.98) 7 MC 1 0 0 7 - 6 D A 1 7
with a built-in temperature transmitter. • 400 (15.7) 1.0 (2.20) 7 MC 1 0 0 7 - 7 D A 1 7
Two Pt100 measuring resistors
■ Technical specifications Mounting length U1/ Weight/
mm (inch): kg (lb):
Design According to DIN 43765: • 160 (6.3) 0.9 (1.98) 7 MC 1 0 0 7 - 5 D B 1 7
Screw-in thermometer • 250 (9.84) 1.0 (2.20) 7 MC 1 0 0 7 - 6 D B 1 7
Protective tube • 400 (15.7) 1.1 (2.43) 7 MC 1 0 0 7 - 7 D B 1 7
• Form 2G, DIN 43772; cylindrical, 9 mm Measuring insert with explosion protection
(0.35 inch) diameter, wall thick- One Pt100 measuring resistor
ness 1 mm (0.04 inch) Mounting length U1/ Weight/
• Loading capacity to 20 bar (290.1 psi) (loading mm (inch): kg (lb):
capacity dependent on material, • 160 (6.3) 0.8 (1.76) 7 MC 1 0 0 7 - 5 D E 1 7
temperature, flow rate, mounting • 250 (9.84) 0.9 (1.98) 7 MC 1 0 0 7 - 6 D E 1 7
length etc., see DIN 43772 for
• 400 (15.7) 1.0 (2.20) 7 MC 1 0 0 7 - 7 D E 1 7
details)
Two Pt100 measuring resistors
Screw socket G½; suitable is gasket 21 x 26, Mounting length U1/ Weight/
similar to form C or D, DIN 7603 mm (inch): kg (lb):
Measuring insert Replaceable, with measuring • 160 (6.3) 0.9 (1.98) 7 MC 1 0 0 7 - 5 D F 1 7
insert tube (6 mm diameter • 250 (9.84) 1.0 (2.20) 7 MC 1 0 0 7 - 6 D F 1 7
(0.24 inch) made of stainless • 400 (15.7) 1.1 (2.43) 7 MC 1 0 0 7 - 7 D F 1 7
steel; terminal block with clamp-
ing springs Connection head, form B,
Response times (to VDI/VDE 3 522) • made of cast light alloy,
with 1 cable inlet and
• In water with flow velocity v = 0.4 t 0.5 = 25 s, t0.9 = 75 s - screw cover 1
m/s (1.31 ft/s) - standard hinged cover 4
• In air with flow velocity t 0.5 = 2 min, t0.9 = 6.3 min - high hinged cover 6
v = 1 m/s (3.28 ft/s) • made of stainless steel, 7
Explosion protection II 1/2G EEx ia IIC T4/T6 with 1 cable inlet and screw cover
Further designs Order code
■ Dimensional drawings Please add "-Z" to Order No. and specify
Order code(s) and plain text.
• Different design (mounting length, protec- Y01
tive tube material etc.), specify in plain text.
• TAG plate made of stainless steel Y15
specify TAG No. in plain text
• Calibration carried out at one point, specify Y33
desired temperature in plain text (order
equivalent number of times for several
calibration points).
To order a temperature transmitter installed in the connection head,
see "Temperature transmitters for installation in the connection
head" (page 3/64).

Low-pressure screw-in resistance thermometer with neck tube, dimen- Individual parts: Measuring inserts and connection heads, see "Acces-
sions in mm (inches) sories".

3/68 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Resistance thermometers
High-pressure screw-in resistance thermometers

■ Overview Selection and Ordering data Order No.


High-pressure screw-in resistance ther-
mometer with connection head and neck
tube
Protective tube and screw socket
X 6 CrNiMoTi 17 122, mat. No. 1.4571
Protective tube to DIN 43772,
form: 2G
One Pt100 measuring resistor
embedded in ceramic, three-wire circuit
3
Mounting length U1/ Weight/
mm (inch): kg (lb):
• 160 (6.3) 0.83 (1.83) 7 MC 1 0 0 8 - 6 D A 1 7
• 250 (9.84) 0.93 (2.05) 7 MC 1 0 0 8 - 7 D A 1 7
The high-pressure screw-in resistance thermometer with con- Two Pt100 measuring resistors
nection head and neck tube is suitable for the temperature embedded in ceramic, two-wire circuit
range from -50 to +600 °C (-58 to +1112 °F) and can also be Mounting length U1/ Weight/
supplied with a built-in temperature transmitter. mm (inch): kg (lb):
• 160 (6.3) 0.86 (1.20) 7 MC 1 0 0 8 - 6 D B 1 7
■ Technical specifications • 250 (9.84) 0.94 (2.07) 7 MC 1 0 0 8 - 7 D B 1 7
Connection head, form B,
Design According to DIN 43765:
Screw-in thermometer • made of cast light alloy,
with 1 cable inlet and
Protective tube - screw cover 1
• Form 2G, DIN 43772; cylindrical, - standard hinged cover 4
11 mm (0.43 inch) diameter, wall - high hinged cover 6
thickness 2 mm (0.08 inch) • made of stainless steel, 7
• Loading capacity to 50 bar (725.2 psi) (loading with 1 cable inlet and screw cover
capacity dependent on material, Further designs Order code
temperature, flow rate, mounting Please add "-Z" to Order No. and specify
length etc., see DIN 43772 for Order code(s) and plain text.
details)
• Different design (mounting length, protec- Y01
Screw socket G1; suitable is gasket 33 x 39, tive tube material etc.), specify in plain text.
similar to form C or D, DIN 7603
• TAG plate made of stainless steel Y15
Measuring insert Replaceable, with measuring specify TAG No. in plain text
insert tube (6 mm diameter • Calibration carried out at one point, specify Y33
(0.24 inch) made of stainless desired temperature in plain text (order
steel; terminal block with clamp- equivalent number of times for several
ing springs calibration points).
Response times (to VDI/VDE 3 522) To order a temperature transmitter installed in the connection head,
• In water with flow velocity t 0.5 = 32 s, t0.9 = 96 s see "Temperature transmitters for installation in the connection
v = 0.4 m/s (1.31 ft/s) head" (page 3/64).

• In air with flow velocity v = 1 m/s t 0.5 = 2.2 min, t0.9 = 6.8 min Individual parts: Measuring inserts and connection heads, see "Acces-
(3.28 ft/s) sories".

■ Dimensional drawings

High-pressure screw-in resistance thermometer with neck tube, dimen-


sions in mm (inches)

Siemens FI 01 · 2008 3/69


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Resistance thermometers
High-pressure welding-type
resistance thermometers

■ Overview
u

Selection and Ordering data Order No.


High-pressure welding-type resistance
thermometer
Measuring resistor: Winding embedded in
ceramic, protective tube to DIN 43772, form 4
1 Pt100 measuring resistor/three-wire circuit
• Measuring insert not explosion protected
3 - max. 540 °C (1004 °F), protective tube
13 CrMo 44, mat. No. 1.7335
7 MC 1 0 1 0 - 7GA 2 7

- max. 550 °C (1022 °F), protective tube X 6 7 MC 1 0 1 0 - 7 F A 2 7


CrNiMoTi 17 122, mat. No. 1.4571
• Measuring insert with explosion protection
- max. 550 °C (1022 °F), protective tube X 6 7 MC 1 0 1 0 - 7 F E 2 7
CrNiMoTi 17 122, mat. No. 1.4571
The high-pressure welding type resistance thermometer with
2 Pt100 measuring resistors/two-wire circuit
connection head and neck tube is suitable for the temperature
range from -50 to 540 or 550 °C (-58 to 1004 or 1022 °F) and can • Measuring insert not explosion protected
also be supplied with a built-in temperature transmitter. - max. 540 °C (1004 °F), protective tube 7 MC 1 0 1 0 - 7GB 2 7
13 CrMo 44, mat. No. 1.7335
■ Technical specifications - max. 550 °C (1022 °F), protective tube X 6
CrNiMoTi 17 122, mat. No. 1.4571
7 MC 1 0 1 0 - 7 F B 2 7

Design According to DIN 43767: • Measuring insert with explosion protection


Welding-type thermometer - max. 550 °C (1022 °F), protective tube X 6 7 MC 1 0 1 0 - 7 F F 2 7
Protective tube CrNiMoTi 17 122, mat. No. 1.4571
• Form 4, DIN 43772; tapered, for weld- Mounting Protective Total length Weight with
ing in; with female thread length U tube without one/two
M18 x 1.5 for neck tube length L connection measuring
head l5 resistors
• Loading capacity to 450 bar (6527 psi) (loading
capacity dependent on material, mm (inch) mm (inch) mm (inch) kg (lb)
temperature, flow rate, mounting 305 (12.0) 0.78 (1.7)/ 1
length etc., see DIN 43772 for 0.79 (1.7)
details)
365 (14.4) 0.82 (1.8)/ 2
140 (5.5)
Neck tube Stainless steel, unscrewable 0.83 (1.8)
Measuring insert Replaceable, with measuring 395 (15.6) 0.85 (1.9)/ 3
65 (2.6) 0.86 (1.9)
insert tube made of stainless
steel; terminal block with clamp- 365 (14.4) 0.95 (2.1)/ 4
ing springs 0.96 (2.1)
200 (7.9)
Response times (to VDI/VDE 3 522) 395 (15.6) 0.98 (2.2)/ 5
1.00 (2.2)
• In water with flow velocity t 0.5 = 25 s, t0.9 = 80 s
v = 0.4 m/s (1.31 ft/s) 365 (14.4) 0.95 (2.1)/ 6
0.96 (2.1)
Explosion protection II 1/2G EEx ia IIC T4/T6 200 (7.9)
395 (15.6) 0.98 (2.2)/ 7
125 (4.9) 1.00 (2.2
■ Dimensional drawings 260 (10.2) 395 (15.6) 1.15 (2.5)/ 8
1.20 (2.7)
Connection head, form B
• made of cast light alloy, with 1 cable bushing
and
- screw cover 1
- standard hinged cover 4
- high hinged cover 6
• made of stainless steel, with 1 cable bushing 7
and screw cover
Further designs Order code
Please add "-Z" to Order No. and specify
Order code(s) and plain text.
• Different design (mounting length, protective Y01
tube material etc.), specify in plain text.
• TAG plate made of stainless steel Y15
specify TAG No. in plain text
• Calibration carried out at one point, specify Y33
desired temperature in plain text (order
equivalent number of times for several cali-
bration points).
High-pressure welding-type resistance thermometer with connection
head and neck tube, dimensions in mm (inches) To order a temperature transmitter installed in the connection
head, see "Temperature transmitters for installation in the
connection head" (page 3/64).
Individual parts: Measuring inserts and connection heads, see "Acces-
sories".

3/70 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Resistance thermometers
Flange-type resistance thermometer
with connection head

■ Overview Selection and Ordering data Order No.


Flange-typeresistance thermometer
With connection head
Protective tube to DIN 43772, form: 2F
one Pt100 measuring resistor: three-wire
circuit, two Pt100 measuring resistors:
two-wire circuit
Measuring insert not explosion protected
One Pt100 measuring resistor
Mounting length/ Weight/
3
mm (inch): kg (lb):
• 160 (6.3) 1.5 (3.31) } 7 MC 1 0 1 7 - 1 F A 1 7
• 250 (9.84) 1.5 (3.31) 7 MC 1 0 1 7 - 2 F A 1 7
Two Pt100 measuring resistors
Mounting length/ Weight/
The flange-type resistance thermometer with connection head mm (inch): kg (lb):
can be installed in tanks and pipelines; it is suitable for a temper-
• 160 (6.3) 1.6 (3.53) 7 MC 1 0 1 7 - 1 F B 1 7
ature range from -50 to +600 °C (-58 to +1112 °F) and can also
be supplied with a built-in temperature transmitter. • 250 (9.84) 1.6 (3.53) 7 MC 1 0 1 7 - 2 F B 1 7
Measuring insert with explosion protection
■ Technical specifications One Pt100 measuring resistor
Mounting length/ Weight/
Protective tube mm (inch): kg (lb):
• Form 2F, DIN 43772; cylindrical, 11 mm • 160 (6.3) 1.5 (3.31) } 7 MC 1 0 1 7 - 1 F E 1 7
(0.43 inch) diameter, wall thick- • 250 (9.84) 1.5 (3.31) 7 MC 1 0 1 7 - 2 F E 1 7
ness 2 mm (0.08 inch) Two Pt100 measuring resistors
• Material X 6 CrNiMoTi 17 122, mat. No. Mounting length/ Weight/
1.4571 mm (inch): kg (lb):
• 160 (6.3) 1.6 (3.53) 7 MC 1 0 1 7 - 1 F F 1 7
• Loading capacity to 40 bar (580 psi) (loading
capacity dependent on material, • 250 (9.84) 1.6 (3.53) 7 MC 1 0 1 7 - 2 F F 1 7
temperature, flow rate, mounting Connection head, form B
length etc., see DIN 43772 for • made of cast light alloy,
details) with 1 cable inlet and
Flange Nominal diameter DN 25, nominal - screw cover 1
pressure PN 40 - standard hinged cover 4
(ASME on request) - high hinged cover } 6
Measuring insert Replaceable, with measuring • made of stainless steel, 7
insert tube (6 mm diameter with 1 cable inlet and screw cover
(0.24 inch) made of stainless
steel; terminal block with clamp- Further designs Order code
ing springs Please add "-Z“ to Order No. and specify
Order code(s) and plain text.
Response times (to VDI/VDE 3 522)
• Different design (mounting length, protec- Y01
• In water with flow velocity t 0.5 = 32 s, t0.9 = 96 s tive tube material etc.), specify in plain text.
v = 0.4 m/s (1.31 ft/s) • TAG plate made of stainless steel Y15
• In air with flow velocity v = 1 m/s t 0.5 = 2.2 min, t0.9 = 6.8 min specify TAG No. in plain text
(3.28 ft/s) • Calibration carried out at one point, specify Y33
desired temperature in plain text (order
Explosion protection II 1/2G EEx ia IIC T4/T6 equivalent number of times for several cal-
ibration points).
■ Dimensional drawings } Available ex stock
To order a temperature transmitter installed in the connection head,
see "Temperature transmitters for installation in the connection
head" (page 3/64).
Individual parts: Measuring inserts and connection heads,
see "Accessories".

Flange-type resistance thermometer with connection head, dimensions


in mm (inches)

Siemens FI 01 · 2008 3/71


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Resistance thermometers
Flange-type resistance thermometer with reduced
response time, with connection head

■ Overview Selection and Ordering data Order No.


Flange-type resistance thermometer
with reduced response time
with connection head
Protective tube to DIN 43772, form: 3
one Pt100 measuring resistor: three-wire
circuit, two Pt100 measuring resistors:
two-wire circuit
3 Measuring insert not explosion protected
One Pt100 measuring resistor
Mounting length/ Weight/
mm (inch): kg (lb):
• 225 (8.9) 1.5 (3.31) 7 MC 1 0 4 1 - 1 A A 0 7
• 285 (11.2) 1.5 (3.31) 7 MC 1 0 4 1 - 2 A A 0 7
• 345 (13.6) 1.5 (3.31) 7 MC 1 0 4 1 - 3 A A 0 7
The flange-type resistance thermometer with connection head Two Pt100 measuring resistors
with reduced response time can be installed in tanks and pipe-
Mounting length/ Weight/
lines; it is suitable for a temperature range from -50 to +600 °C mm (inch): kg (lb):
(-58 to +1112 °F). The sensor is also available with a built-in tem-
• 225 (8.9) 1.6 (3.53) 7 MC 1 0 4 1 - 1 A B 0 7
perature transmitter.
• 285 (11.2) 1.6 (3.53) 7 MC 1 0 4 1 - 2 A B 0 7

■ Technical specifications • 345 (13.6) 1.6 (3.53)


Measuring insert with explosion protection
7 MC 1 0 4 1 - 3 A B 0 7

Protective tube One Pt100 measuring resistor


• Form 3, DIN 43772; cylindrical, 12 mm Mounting length/ Weight/
(0.47 inch) diameter, wall thick- mm (inch): kg (lb):
ness 2.5 mm (0.1 inch), tapered • 225 (8.9) 1.5 (3.31) 7 MC 1 0 4 1 - 1 E A 0 7
towards tip; tip cylindrical over • 285 (11.2) 1.5 (3.31) 7 MC 1 0 4 1 - 2 E A 0 7
length of 47 mm (1.85 inch), • 345 (13.6) 1.5 (3.31) 7 MC 1 0 4 1 - 3 E A 0 7
9 mm (0.35 inch) diameter, wall
thickness 1.5 mm (0.06 inch) Two Pt100 measuring resistors
Mounting length/ Weight/
• Material X 6 CrNiMoTi 17 122, mat. No. mm (inch): kg (lb):
1.4571 • 225 (8.9) 1.6 (3.53) 7 MC 1 0 4 1 - 1 E B 0 7
• Loading capacity to 40 bar (580 psi) (loading • 285 (11.2) 1.6 (3.53) 7 MC 1 0 4 1 - 2 E B 0 7
capacity dependent on material, • 345 (13.6) 1.6 (3.53) 7 MC 1 0 4 1 - 3 E B 0 7
temperature, flow rate, mounting
length etc., see DIN 43772 for Connection head, form B
details) • made of cast light alloy,
with 1 cable inlet and
Flange Nominal diameter DN 25, nominal
pressure PN 40 - screw cover 1
(ASME on request) - standard hinged cover 4
- high hinged cover 6
Measuring insert Replaceable, with measuring
insert tube (6 mm diameter • made of stainless steel, 7
(0.24 inch) made of stainless with 1 cable inlet and screw cover
steel; terminal block with clamp- Further designs Order code
ing springs
Please add "-Z" to Order No. and specify
Response times (to VDI/VDE 3 522) Order code(s) and plain text.
• In water with flow velocity t 0.5 = 22 s, t0.9 = 66 s • Different design (mounting length, protec- Y01
v = 0.4 m/s (1.31 ft/s) tive tube material etc.), specify in plain text.
• TAG plate made of stainless steel Y15
• In air with flow velocity v = 1 m/s t 0.5 = 2.1 min, t0.9 = 6.5 min
specify TAG No. in plain text
(3.28 ft/s)
• Calibration carried out at one point, specify Y33
Explosion protection II 1/2G EEx ia IIC T4/T6 desired temperature in plain text (order
equivalent number of times for several
■ Dimensional drawings calibration points).
To order a temperature transmitter installed in the connection head,
see "Temperature transmitters for installation in the connection
head" (page 3/64).
Individual parts: Measuring inserts and connection heads,
see "Accessories".

Flange-type resistance thermometer with reduced response time; with


connection head, dimensions in mm (inches)

3/72 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Resistance thermometers
for damp rooms

■ Overview Selection and Ordering data Order No.

The resistance thermometer for damp rooms is suitable for a Resistance thermometer for
damp rooms
temperature range from -30 to +60 °C (-22 to +140 °F).
stainless steel protective tube

■ Technical specifications • with one Pt100 measuring resistor


0.1 kg (0.22 kg)
} 7MC1027-1AA

Protective tube Made of stainless steel • with two Pt100 measuring resistors 7MC1027-1AB
Connection head Made of cast light alloy, with
cable bushing; made of plastic on
0.1 kg (0.22 kg)
Further designs Order code
3
request Please add "-Z" to Order No. and specify
Measuring insert 1 or 2 Pt measuring resistors to Order code(s) and plain text.
DIN EN 60751, connection in • Different design (mounting length, protec- Y01
three-wire or two-wire system, tive tube material etc.), specify in plain text.
class B • TAG plate made of stainless steel Y15
Degree of protection IP65 acc. to DIN EN 60529 specify TAG No. in plain text
• Calibration carried out at one point, specify Y33

■ Dimensional drawings
desired temperature in plain text (order
equivalent number of times for several
calibration points).
} Available ex stock
To order a temperature transmitter installed in the connection head,
see "Temperature transmitters for installation in the connection
head" (page 3/64).
Note:
Additional fitting of head mounted transmitter of SITRANS TH series is
possible.

Resistance thermometer for moist rooms, dimensions in mm (inches)

Siemens FI 01 · 2008 3/73


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Resistance thermometers
Accessories
Measuring inserts, not explosion protected

■ Design ■ Dimensional drawings


Components (design to DIN 43762)
• Measuring insert with resistor,
• insert tube with retaining plate,
• terminal block with clamping springs and 2 or 3 internal con-
ductors of Cu-Ni sheathed wire for each Pt100 measuring re-
3 sistor.
The resistance of the internal conductors is stated on the mea-
suring insert if it is greater than 0.2 Ω.
The measuring inserts have a filling of Al2O3 powder which sur-
rounds the measuring resistor and the internal conductors and
results in a high vibration resistance (for temperatures up to Measuring insert for flue gas resistance thermometers, dimensions in mm
600 °C (1112 °F). The winding is embedded in a ceramic body. (inches)

If the vibrations at the measuring location are greater than nor-


mal, the special vibration-proof measuring inserts should be
used (for temperatures up to 450 °C (842 °F)). The measuring re-
sistor is embedded in an homogeneous, fused glass body.

Measuring insert for low-pressure, high-pressure and flange-type resis-


tance thermometers, dimensions in mm (inches)

Selection and Ordering data Order No.


Measuring insert for flue gas resistance thermometer 7MC1000
for temperatures to 600 °C (1112 °F), stainless steel measuring insert tube
Measuring resistor: Winding embedded in ceramic
Measuring insert not explosion protected
One Pt100 measuring resistor/three-wire circuit
Measuring insert length Mounting length of the resistance thermometer Weight
mm (inch): mm (inch): kg (lb):
• 525 (20.7) 500 (19.7) 0.22 (0.49) 7MC1900-1EA
• 735 (28.9) 710 (28.0) 0.27 (0.60) 7MC1900-2EA
• 1025 (40.4) 1000 (39.4) 0.32 (0.71) 7MC1900-3EA
• 1425 (56.1) 1400 (55.1) 0.42 (0.93) 7MC1900-4EA
• 2025 (79.7) 2000 (78.7) 0.62 (1.37) 7MC1900-5EA

Selection and Ordering data Order No.


Measuring insert for low-pressure, high-pressure and flange-type resistance thermometers
Measuring resistor: Winding embedded in ceramic (max. 600 °C (1112 °F))
Measuring insert not explosion protected
One Pt100 measuring resistor/three-wire circuit
Measuring Screw-in ther- Screw-in Screw-in Flange-type Flange-type Weight
insert length mometer with- thermometer thermometer thermometer thermometer
out neck tube, with neck tube,
7MC1006 7MC1007 and 7MC1010 7MC1017 7MC1041
7MC1008
Mounting Mounting Total length Mounting Mounting
length length without con- length length
nection head
mm (inch) mm (inch) mm (inch) mm (inch) mm (inch) mm (inch) kg (lb)
• 145 (5.7) 100 (3.9) – – – – 0.15 (0.33) 7MC1910-1JA
• 205 (8.1) 160 (6.3) – – – – 0.16 (0.35) 7MC1910-2JA
• 275 (10.8) 230 (9.1) – – – – 0.17 (0.37) 7MC1910-3JA
• 315 (12.4) – 160 (6.3) 305 (12.0) 160 (6.3) 225 (8.9) 0.18 (0.40) 7MC1910-4JA
• 375 (14.8) – – 365 (14.4) – 285 (11.2) 0.19 (0.42) 7MC1910-5JA
• 405 (15.9) 360 (14.2) 250 (9.8) 395 (15.6) 250 (9.8) – 0.20 (0.44) 7MC1910-6JA
• 435 (17.1) – – – – 345 (13.6) 0.20 (0.44) 7MC1910-8JA
• 555 (21.1) 510 (20.1) 400 (15.8) – – – 0.21 (0.46) 7MC1910-7JA

3/74 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Resistance thermometers
Accessories
Measuring inserts, not explosion protected
Selection and Ordering data Order No.
Measuring insert for low-pressure, high-pressure and flange-type resistance thermometers
Measuring resistor: Winding embedded in ceramic (max. 600 °C (1112 °F))
Two Pt100 measuring resistors/three-wire circuit
Measuring insert Screw-in ther- Screw-in Screw-in Flange-type Flange-type Weight
length mometer with- thermometer thermometer thermometer thermometer
out neck tube, with neck tube,
7MC1006 7MC1007 and 7MC1010
7MC1008
7MC1017 7MC1041
3
Mounting Mounting Total length Mounting Mounting
length length without connec- length length
tion head
mm (inch) mm (inch) mm (inch) mm (inch) mm (inch) mm (inch) kg (lb)
• 145 (5.7) 100 (3.9) – – – – 0.15 (0.33) 7MC1910-1JB
• 205 (8.1) 160 (6.3) – – – – 0.16 (0.35) 7MC1910-2JB
• 275 (10.8) 230 (9.1) – – – – 0.17 (0.37) 7MC1910-3JB
• 315 (12.4) – 160 (6.3) 305 (12.0) 160 (6.3) 225 (8.9) 0.18 (0.40) 7MC1910-4JB
• 375 (14.8) – – 365 (14.4) – 285 (11.2) 0.19 (0.42) 7MC1910-5JB
• 405 (15.9) 360 (14.2) 250 (9.8) 395 (15.6) 250 (9.8) – 0.20 (0.44) 7MC1910-6JB
• 435 (17.1) – – – – 345 (13.6) 0.20 (0.44) 7MC1910-8JB
• 555 (21.1) 510 (20.1) 400 (15.8) – – – 0.21 (0.46) 7MC1910-7JB
Further measuring inserts on request.

Siemens FI 01 · 2008 3/75


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Resistance thermometers
Accessories
Measuring inserts, explosion protected

■ Overview ■ Technical specifications


An explosion protected measuring insert is suitable for installa- Measuring temperature -200 ... +450 °C (-328 ... +842 °F)
tion in a protective fitting and for connection to a certified intrin- The temperature depends on the
sically-safe circuit of category "ia" or "ib". maximum measuring current
which can flow in the connected
The measuring insert may only be used if the protective fitting measuring circuit if an error
has the degree of protection IP20 according to DIN 40050. occurs.

3 ■ Design
Resistance of internal conductor Per measuring circuit, i.e. for both
conductors of the two-wire sys-
tem: 0.17 Ω/m measuring insert
The measuring resistor is fitted in a 60 mm (2.36 inch) long stain- Resistance of measuring insert ≥ 1000 MΩ at room temperature
less steel sleeve with an outer diameter of 6 mm (0.24 inch). The
sleeve and the sheath of the supply cable are welded. The Outer diameter of sleeve 6 mm (0.24 inch)
sheath of the supply cable is also made of stainless steel, and Explosion protection II 2G EEx ia IIC T4/T6
its outer diameter is 5 mm (0.2 inch).
The conductors are made of special copper and are embedded
in highly compacted magnesium oxide powder. The connection ■ Dimensional drawings
points between the measuring resistor and the conductors are
additionally isolated. The conductors are potted in the terminal
base.
The rating plate is located underneath the retaining plate.

Explosion protected measuring insert for resistance thermometers, di-


mensions in mm (inches)
Selection and Ordering data Order No.
Measuring insert for low-pressure, high-pressure and flange-type resistance thermometers
Stainless steel measuring insert
Measuring resistor: Winding embedded in ceramic (max. 450 °C (842 °F))
Outer diameter of sleeve: 6 mm (0.24 inch)
Measuring insert with explosion protection
One Pt100 measuring resistor/three-wire circuit
Measuring Screw-in ther- Screw-in ther- Screw-in Flange-type Flange-type Weight
insert length mometer without mometer with thermometer thermometer thermometer
neck tube, neck tube,
7MC1006 7MC1007 7MC1010 7MC1017 7MC1041
Mounting length Mounting length Total length Mounting Mounting length
without length
connection head
mm (inch) mm (inch) mm (inch) mm (inch) mm (inch) mm (inch) kg (lb)
• 145 (5.7) 100 (3.9) – – – – 0.15 (0.33) 7MC1913-1AA22
• 205 (8.1) 160 (6.3) – – – – 0.16 (0.35) 7MC1913-2AA22
• 275 (10.8) 230 (9.1) – – – – 0.17 (0.37) 7MC1913-3AA22
• 315 (12.4) – 160 (6.3) 305 (12.0) 160 (6.3) 225 (8.9) 0.18 (0.40) 7MC1913-4AA22
• 375 (14.8) – – 365 (14.4) – 285 (11.2) 0.19 (0.42) 7MC1913-5AA22
• 405 (15.9) 360 (14.2) 250 (9.8) 395 (15.6) 250 (9.8) – 0.20 (0.44) 7MC1913-6AA22
• 435 (17.1) – – – – 345 (13.6) 0.20 (0.44) 7MC1913-7AA22
• 555 (21.1) 510 (20.1) 400 (15.8) – – – 0.21 (0.46) 7MC1913-8AA22
Two Pt100 measuring resistors/two-wire circuit
Measuring insert Screw-in Screw-in Screw-in Flange-type Flange-type Weight
length thermometer thermometer thermometer thermometer thermometer
without neck with neck tube,
tube, 7MC1006 7MC1007 7MC1010 7MC1017 7MC1041
Mounting Mounting Total length Mounting Mounting
length length without length length
connection head
mm (inch) mm (inch) mm (inch) mm (inch) mm (inch) mm (inch) kg (lb)
• 145 (5.7) 100 (3.9) – – – – 0.15 (0.33) 7MC1913-1AB12
• 205 (8.1) 160 (6.3) – – – – 0.16 (0.35) 7MC1913-2AB12
• 275 (10.8) 230 (9.1) – – – – 0.17 (0.37) 7MC1913-3AB12
• 315 (12.4) – 160 (6.3) 305 (12.0) 160 (6.3) 225 (8.9) 0.18 (0.40) 7MC1913-4AB12
• 375 (14.8) – – 365 (14.4) – 285 (11.2) 0.19 (0.42) 7MC1913-5AB12
• 405 (15.9) 360 (14.2) 250 (9.8) 395 (15.6) 250 (9.8) – 0.20 (0.44) 7MC1913-6AB12
• 435 (17.1) – – – – 345 (13.6) 0.20 (0.44) 7MC1913-7AB12
• 555 (21.1) 510 (20.1) 400 (15.8) – – – 0.21 (0.46) 7MC1913-8AB12
Further measuring inserts on request.

3/76 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Resistance thermometers
Accessories – Welding-type protective tubes,
neck tubes and connection heads

■ Welding-type protective tube ■ Neck tube


Welding-type protective tube for high-pressure resistance Neck tube for high-pressure screw-in resistance thermome-
thermometers to DIN 43 767, without neck tube, without con- ter
nection head • Made of stainless steel, mat. No. 1.4571
• Tapered shank with cylindrical welding stubs • With threads at both ends
• For measuring insert tube with 6 mm (0.24 inch) • For measuring insert tube with 6 mm (0.24 inch) OD
• OD female thread M18 x 1.5 (including steel screw plug)
3
■ Dimensional drawings
Connection heads for low and high-pressure resistance thermometers, flue gas and flange-type resistance thermometers

Connection head, form B, degree of protection IP54, made of cast light


alloy, with screw cover, dimensions in mm (inches)

Connection head, form B, degree of protection IP65, made of cast light


alloy, with high hinged cover, dimensions in mm (inches)

Connection head, form B, degree of protection IP54, made of plastic, with


screw cover, dimensions in mm (inches)

Connection head, form B-VA, degree of protection IP65, made of stain-


less steel, with screw cover, dimensions in mm (inches)

Connection head, form B, degree of protection IP65, made of cast light


alloy, with standard hinged cover, dimensions in mm (inches)

Siemens FI 01 · 2008 3/77


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Resistance thermometers
Accessories – Welding-type protective tubes,
neck tubes and connection heads

Selection and Ordering data Order No.


Welding protective tube for high-pressure resistance thermometers according to DIN 43767,
without neck tube, without connection head
tapered shank with cylindrical welding stub, for measuring insert tube with 6 mm (0.24 inch) OD;
female thread M18 x 1.5 (including steel screw plug)
Up to 540 °C (1004 °F)
Protective tube to DIN 43772, form 4 made of 13 CrMo 44, mat. No. 1.7335

3 Mounting
length U
Protective tube
length L
Weight
mm (inch)
mm (inch) mm (inch)
• 65 (2.56) 140 (5.51) 0.3 (0.66) 7MC1905-1GA
• 65 (2.56) 200 (7.87) 0.5 (1.1) 7MC1905-2GA
• 125 (4.92) 200 (7.87) 0.5 (1.1) 7MC1905-3GA
• 125 (4.92) 260 (10.24) 0.6 (1.32) 7MC1905-4GA
Up to 550 °C (1022 °F)
Protective tube to DIN 43772, form 4 made of 6 CrNiMoTi 17122, mat. No. 1.4571
Mounting Protective tube Weight
length U length L kg (lb)
mm (inch) mm (inch)
• 65 (2.56) 140 (5.51) 0.3 (0.66) 7MC1905-1DA
• 65 (2.56) 200 (7.87) 0.5 (1.1) 7MC1905-2DA
• 125 (4.92) 200 (7.87) 0.5 (1.1) 7MC1905-3DA
• 125 (4.92) 260 (10.24) 0.6 (1.32) 7MC1905-4DA

Selection and Ordering data Order No.


Neck tube for high-pressure screw-in resistance thermometer
made of stainless steel, mat. No. 1.4571, with thread at both ends, for measuring insert tube with 6 mm (0.24 inch) OD
Neck tube Total length of the resistance thermometer, Protective Weight
length without connection head tube length
mm (inch) mm (inch) mm (inch) kg (lb)
• 135 (5.31) 395 (15.55) 260 (10.24) 0.14 (0.31) 7MC1906-1AA
• 165 (6.50) 305/365 (12.01/14.37) 140/200 (5.51/7.87) 0.15 (0.33) 7MC1906-2AA
• 195 (7.68) 395 (15.55) 200 (7.87) 0.18 (0.40) 7MC1906-3AA
• 225 (8.86) 365 (14.37) 140 (5.51) 0.20 (0.44) 7MC1906-4AA
• 255 (10.04) 395 (15.55) 140 (5.51) 0.22 (0.49) 7MC1906-5AA

Selection and Ordering data Order No.


Connection heads for low-pressure,
high-pressure, flue gas and flange-type resis-
tance thermometers
Connection head, form B, degree of protection IP54
• made of cast light alloy, 7MC1907-1BA
with screw cover and with 1 cable bushing,
weight: 0.14 kg (0.31 lb)
• made of plastic, 7MC1907-1BK
with screw cover and with 1 cable bushing,
weight: 0.08 kg (0.18 lb)
Connection head, form B, degree of protection IP65
Weight: 0.3 kg (0.66 lb)
• made of cast light alloy, 7MC1907-1BF
with standard hinged cover and with 1 cable
bushing
• made of cast light alloy, 7MC1907-1BL
with high hinged cover and with 1 cable bushing
Connection head, form B-VA, degree of protec-
tion IP65
• made of stainless steel, 7MC1907-1BV
with screw cover and with 1 cable bushing,
weight: 0.65 kg (1.43 lb)
Accessories
for connection head, form B, degree of protection IP65
• Quick-release clamp (degree of protection of 7MC1907-1BS
connection head reduced to IP54)
Weight: 0.02 kg (0.04 lb)
Connection heads with a drilled hole of 15.5 mm diameter (0.61 inch)
instead of the female thread M24 x 1.5 on request.

3/78 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Thermocouples
Technical description

■ Design Mode of operation


The thermocouples are extended from the connection point to a
A thermocouple comprises point whose temperature is as constant as possible (the cold
• The thermocouple element (sensor) and junction) by means of extension leads.
• The mounting and connection parts required in each case. The extension leads have the same color code as the associated
The thermocouple element is formed by two conductors of dis- thermocouple elements; the positive pole is marked in red. Cor-
similar metals or metal alloys which are soldered or welded to- rect polarity must be ensured since otherwise large errors will
gether at one end, the measuring junction: occur. Up to 200 °C, the same calibration data and tolerances
apply to the extension leads as to the corresponding thermocou-
ples.
3
The influence of temperature changes at the cold junction can
be balanced by means of a compensating circuit, e.g. a com-
pensating box. The reference temperature is 0 (32 °F ) or 20 °C
(68 °F).
It is also possible to keep the cold junctions at a constant tem-
perature of 50, 60 or 70 °C (122, 140 or 158 °F) using a thermo-
stat (for several measuring junctions).
The connections from the cold junction to the measuring or pro-
cess instrument are made using copper leads. With energy-con-
suming instruments such as indicators or multipoint recorders,
the complete measuring circuit (thermocouple, extension lead
and copper lead) must be balanced in the operating condition
using a resistor. SITRANS T transmitters and
KOMPENSOGRAPH servo recorders for connection to thermo-
couple elements have a built-in compensating circuit for balanc-
ing the effect of the ambient temperature on the cold junction.
Lead balancing is not necessary in this case because of the high
input impedance.
Protection fitting/protective tubes
The thermocouple can be protected against mechanical stress
and chemical attack by a ceramic or metal protective tube which
may be mounted using flanges, screwed glands or by welding
into the pipeline or tank. The thermocouple element terminates
Thermocouple element
in the connection head.
■ Function Installation examples with specification of the recommended
thermocouples and protective tube materials are listed on pages
Measuring principle of the thermocouple element "Technical Data"and "Installation Examples".
If the measuring junction is exposed to a temperature different Owing to the different operating conditions, no guarantee can be
from that at the free ends of the thermocouple, a voltage (the given for protective fittings. The manufacturer is responsible for
thermoelectric voltage, Seebeck effect) is produced at these damages and measuring errors caused by wrong installation in
free ends. The magnitude of the thermoelectric voltage depends compliance with the General Terms of Delivery if the instruments
on the difference in temperature between the measuring junction have been installed by the manufacturer and if the specifications
and the free ends, and on the combination of materials in the for the operating conditions furnished by the customer were cor-
thermocouple. Since a thermocouple always measures a tem- rect and sufficiently detailed.
perature difference, the free ends of the thermocouple must be
Thermocouple elements are very compatible since it is almost
connected to a reference junction (cold conjunction) and held
always possible to adapt them in shape and size to the particular
constant at a known temperature.
problem. The temperature-responsive part is almost point-
Calibration data for thermoelectric voltages and permissible shaped. Thermocouple elements are therefore particularly suit-
deviations able for measuring rapidly changing temperatures.
The calibration data and the permissible deviations for com-
monly used thermocouples are defined isee Technical Data,
Table "Calibration data for thermoelectric voltages and error
limits").
The thermocouples Cu-CuNi and Fe-CuNi to DIN 43710 are
used for replacement purposes. Thermocouples of class 2 are
supplied as standard. For more accurate measurements, ther-
mocouples are available with half the DIN tolerance or with a test
certificate. The tolerances only apply to the condition upon de-
livery.
During operation at high temperatures, the tolerances of the
thermocouples may change due to absorption of foreign matter,
oxidation or evaporation of alloy components.

Siemens FI 01 · 2008 3/79


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Thermocouples
Technical description

■ Integration
Instrument combination for measuring and controlling temperature, with thermocouple element as sensor

Sense Thermocouple

3
Match Cold junction thermostat Cold junction with power supply unit
Catalog FI 01 Catalog FI 01

Electric transmitter
Catalog FI 01

Output Electric Catalog Electric Catalog


digital indicator analog indicator MP 12
Control digital limit monitor analog limit monitor MP 12
recorder MP 20

controller MP 31

Installation examples with specification of appropriate thermocouples and protective tubes


Measuring Permissible Thermocou- Protective tube Measuring Permissible Thermocou- Protective tube
point operating ple point operating ple
temperature temperature
°C (°F) °C (°F)
A. Iron and steel works Exhaust gas 1000 (1832) Ni Cr/Ni X 10 CrAl 24,
before or behind mat. No. 1.4762,
1. Blast furnace recuperator dished; installation
Hot blast 1000 (1832) Ni Cr/Ni Open protective tube with gas-tight screw
X 10 CrAl 24, socket
mat. No. 1.4762, bare Combustion air 700 (1292) Ni Cr/Ni X 10 CrAl 24,
soldering point, behind recuper- mat. No. 1.4762,
cemented ator dished; installation
Crude gas 300 (572) Ni Cr/Ni X 10 CrAl 24, with gas-tight screw
mat. No. 1.4762, socket
dished 5. Pusher, rotating hearth furnace, and other types of rolling mill
2. Air preheater furnace
Dome 1200 (2192) Pt 10% Rh/Pt Outside: X 15 CrNiSi Preheating zone 800 (1472) Pt 10% Rh/Pt Outside: Silicon car-
24 19, mat. No. or Ni Cr/Ni bide
1.4841, inside: KER 710 or
inside: KER 710 outside: Silicon car-
bide
Exhaust gas 300 (572) Ni Cr/Ni St 35.8, mat. No. inside: KER 610
1.0305, seamless,
enamelled Heating zone 1250 (2282) Pt 10% Rh/Pt As above
3. Open-hearth furnace Soaking zone 1350 (2462) Pt 10% Rh/P Outside: Silicon car-
bide
Exhaust gas 600 (1112) Ni Cr/Ni X 10 CrAl 24, inside: KER 710
ducts mat. No. 1.4762,
dished Exhaust gas 900 (1652) Ni Cr/Ni X 10 CrAl 24,
before air recu- mat. No. 1.4762,
Checkers 1350 (2462) Pt 10% Rh/Pt Outside: Silicon car- perator dished; installation
bide with gas-tight screw
inside: KER 710 socket
Melt (for short 1600 (2912) Pt 10% Rh/Pt Immersion-type ther- Exhaust gas 700 (1292) Ni Cr/Ni As above or KER 610
periods) mocouple element of before and
special design behind gas
4. Soaking pit recuperator

Furnace cham- 1350 (2462) Pt 10% Rh/Pt Outside: Silicon car- Preheated air 700 (1292) Ni Cr/Ni As above
ber bide behind recuper-
inside: KER 710 ator

3/80 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Thermocouples
Technical description
Installation examples with specification of appropriate thermocouples and protective tubes (continued)

Measuring Permissible Thermocou- Protective tube Measuring Permissible Thermocou- Protective tube
point operating ple point operating ple
temperature temperature
°C (°F) °C (°F)
6. Annealing furnace, roller-hearth furnace 11.2 In hardening shop baths
(protective tubes bored from the solid are recommended)
Furnace cham-
ber
800 (1472) Ni Cr/Ni X 10 CrAl 24,
mat. No. 1.4762,
dished
Salt and nitre 550 (1022) Ni Cr/Ni Soft iron,
mat. No. 1.1003
3
Exhaust gas 600 (1112) Ni Cr/Ni As above Cyanogen 950 (1742) Ni Cr/Ni Soft iron or
duct X 15 CrNiSi 24 19,
mat. No. 1.4841
7. Bell-type anneal. furn.
Baths contain- 1050 (1922) Pt 10% Rh/Pt Outside: X 10 CrAl
Directly heated 950 (1742) Ni Cr/Ni X 10 CrAl 24, ing chloride 24,
furnace mat. No. 1.4762, mat. No. 1.4762,
dished inside: KER 710
Stack for tin 600 ... 900 Ni Cr/Ni Special design, bare Lead baths 1200 (2192) Pt 10% Rh/P Outside: Chromium
plates (1112 ... 1652) thermocouple ele- nickel alloy
ment with thermal NiCr 60 15,
contact plate mat. No. 2.4867,
Stack for motor- 700 ... 920 Ni Cr/Ni As above inside: KER 710;
car body sheet (1292 ... 1688) bored from the solid
Protective gas 650 (1202) Ni Cr/Ni Bare thermocouple Barium chloride 1300 (2372) Pt 10% Rh/Pt Outside: Chromium
element baths nickel alloy
NiCr 60 15,
8. Sinter belts mat. No. 2.4867,
(Sucking boxes) 400 (752) Fe/Cu Ni (Jacket thermocou- inside: KER 710;
ple element) B. Metallurgical plants (only limited life time in molten metals)
X 10 CrAl 24,
mat. No. 1.4762 Copper melt 1250 (2282) Pt 10% Rh/Pt Outside: X 10 CrSi
29, mat. No. 1.4772,
For temperature 400, 500 Fe/Cu Ni As above, with dou- bored from the solid
maximum com- (752, 932) ble thermocouple inside: KER 710
puting circuit
Exhaust of cop- 1300 (2372) Pt 10% Rh/Pt Outside: Silicon car-
9. Tinning plants per melting fur- bide
Palm nut oil, 650 (1202) Ni Cr/Ni St 35.8, naces inside: KER 710
pickling vat, tin mat. No. 1.0305, Brass melts 900 (1652) Ni Cr/Ni X 10 CrAl 29,
bath seamless, enamelled mat. No. 1.4772,
10. Zinc coating plants bored from the solid

Zinc bath 480 (896) Fe/Cu Ni St 35.8, Aluminium melts 700 (1292) Ni Cr/Ni Pearlite iron GG 22,
mat. No. 1.0305, bored from the solid
seamless Die-casting, 700 (1292) Ni Cr/Ni Soft iron,
11. Heat treatment plants magnesium mat. No. 1.1003,
bored from the solid
11.1 In gases
Bearing metal, 600 (1112) Fe/Cu Ni St 35.8,
Tempering fur- 550 (1022) Ni Cr/Ni St 35.8, lead and tin mat. No. 1.0305,
naces mat. No. 1.0305, melting houses seamless, enamelled
seamless
Lead foundries 700 (1292) Fe/Cu Ni Chromium nickel
Annealing in oxi- 1050 (1922) Pt 10% Rh/Pt Outside: X 10 CrAl alloy
dizing gases (Ni Cr/Ni) 24, NiCr 60 15,
containing sul- mat. No. 1.4762, mat. No. 2.4867,
phur and car- dished bored from the solid
bon inside: KER 710
Zinc foundries 480 (878) Fe/Cu Ni St 35.8,
1200 (2192) Pt 10% Rh/Pt Outside: X 10 CrAl mat. No. 1.0305,
24, seamless
mat. No. 1.4762,
dished 600 (1112) Fe/Cu Ni Silicon carbide
inside: KER 710 Exhaust gases 1300 (2372) Pt 10% Rh/Pt Outside: Silicon car-
Annealing in 1200 (2192) Pt 10% Rh/Pt As above of zinc melting bide
reducing gases furnaces inside: KER 710
containing sul- C. Ceramic industry
phur
Ring kiln for 800 ... 1100 Ni Cr/Ni X 10 CrAl 24,
Annealing in 1200 (2192) Pt 10% Rh/Pt Outside: standard bricks (1472 ... 2012) mat. No. 1.4762,
nitrogen-con- X 15 CrNiSi 24 19, dished
taining gases mat. No. 1.4841,
poor in oxygen, dished; Ring kiln for clin- 1200 ... 1300 Pt 10% Rh/Pt Outside and inside:
also nitriding inside: KER 710 kers, retort fur- (2192 ... 2372) KER 710
furnaces with nace, tunnel
ammonia furnace, glaz-
ing kiln

Siemens FI 01 · 2008 3/81


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Thermocouples
Technical description
Installation examples with specification of appropriate thermocouples and protective tubes (continued)

Measuring Permissible Thermocou- Protective tube Measuring Permissible Thermocou- Protective tube
point operating ple point operating ple
temperature temperature
°C (°F) °C (°F)
D. Glass industry G. Steam power parts

3 1. Glass trough furnace


Checkers of the 1300 (2372) Pt 10% Rh/Pt Outside: KER 530
Water and steam lines (screw-in and welding-type thermocouple ele-
ments)
regenerative Inside: KER 710 Screw-in ther- 300 (572) Fe/Cu Ni Sn Bz 6 bronze, to
chambers, mocouple ele- DIN 1726 (only for air
above ment or water)
Checkers of the 600 (1112) Ni Cr/Ni Thermocouple porce- Screw-in ther- 400 (752) Fe/Cu Ni St 35.8,
regenerative lain mocouple ele- mat. No. 1.0305
chambers, ment
below
Screw-in ther- 500 (932) Fe/Cu Ni 113 CrMo 44,
Lateral walls, 1550 (2822) Pt 30% Rh/ Outside and inside: mocouple ele- mat. No. 1.7335
roof, bottom (in 1400 (2552) Pt 6% Rh KER 710 ment
drilled channels
down to 50 mm Welding-type 540 (1004) Fe/Cu Ni 113 CrMo 44,
below the bath) thermocouple mat. No. 1.7335
element
Leer 800 (1472) Ni Cr/Ni X 10 CrAl 24,
mat. No. 1.4762, Welding-type 570 (1058) Ni Cr/Ni 10 CrMo 9 10,
dished thermocouple mat. No. 1.7380
element
Annealing fur- 1200 (2192) Pt 10% Rh/Pt KER 710 750 (1382) Fe/Cu Ni X 8 CrMoNb 16 16,
nace (Ni Cr/Ni) mat. No. 1.4981

2. Pot furnace Flue gas 1000 (1832) Pt 10% Rh/Pt Megapyr,


Cr Al 20 5,
1500 (2732) Pt 30% Rh/ Pt Outside and inside: mat. No. 1.4767
6% Rh KER 710 or
outside: X 10 CrAl 24,
3. Gas generators mat. No. 1.4762,
Crude gas 750 (1382) Ni Cr/Ni X 10 CrAl 24, inside: KER 710 St
mat. No. 1.4762, 600 (1112) (Ni Cr/Ni) St 35.8,
dished mat. No. 1.0305,
Rotary grate 180 ... 200 Fe/Cu Ni Special design on enamelled
(Grate head) (356 ... 392) request (to be manu- Pulverized 100 (212) Fe/Cu Ni St 35.8,
For short peri- factured on site coal/air mixture mat. No. 1.0305,
ods: according to specifi- line (with baffle rod)
500 to 1000 cations)
(932 to 1832)
E. Cement industry Rotary kilns
Secondary air at 900 (1652) Ni Cr/Ni X 10 CrAl 24, ■ Technical specifications
cooler mat. No. 1.4762
Thermocouple designations to DIN 43710 and DIN IEC 584
Hot chamber 900 (1652) Ni Cr/Ni X 10 CrAl 24,
mat. No. 1.4762 DIN 43710, release 12.85
Drying chamber 400 (752) Fe/Cu Ni St 35.8, Cu-CuNi Type U
mat. No. 1.0305, or
similar, enamelled Fe-CuNi Type L

F. Cellulose factories DIN IEC 60584

Sulphur com- to 1500 (2732) Pt 30% Rh/ Pt Outside: Silicon car- Cu/Cu Ni Type T
bustion fur- 6% Rh bide Fe/Cu Ni Type J
naces, pyrite inside: KER 710
burnerr Ni Cr/Ni Type K
Behind waste 600 (1112) Ni Cr/Ni X 10 CrAl 24, Ni Cr Si-NiSi Type N
heat boiler mat. No. 1.4762 Pt 10% Rh/Pt Type S
Pt 13% Rh/Pt Type R
Pt 30% Rh/ Pt 6% Rh Type B

3/82 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Thermocouples
Temperature transmitters
for mounting in the connection head

■ Overview ■ Selection and Ordering Data


Detailed information on the transmitters can be found for the re-
spective products under "Transmitters for temperature“.

Transmitter to be fitted Order code

3
To order the sensor with a built-in
temperature transmitter, add "-Z" to
the Order No. of the sensor, and
supplement by the following Order
code:
• SITRANS TH200
- without Ex T20
- EEx ia IIC and EEx n for zone 2 T21
- FM (IS, I, NI) T23
• SITRANS TH300
- without Ex T30
- EEx ia IIC und EEx n for zone 2 T31
The following temperature transmitters are available for mount-
- FM (IS, I, NI) T33
ing in the connection head:
• SITRANS TH400 PA
SITRANS TH200
- without Ex T40
Programmable two-wire temperature transmitter (4 to 20 mA),
- EEx ia T41
electrical isolation for resistance thermometers and thermocou-
ple elements. • SITRANS TH400 FF

SITRANS TH300 - without Ex T45


- EEx ia T46
Two-wire temperature transmitter with HART communication
(4 to 20 mA), electrical isolation for resistance thermometers and Customer-specific setting of the Y111)
thermocouple elements. built-in transmitter (specify settings
in plain text)
SITRANS TH400 1) For TH400 FF available soon.
Temperature transmitter with PROFIBUS-PA or FOUNDATION
fieldbus connection, electrical isolation for resistance thermom-
eters and thermocouple elements.
Note:
• SITRANS TH200/TH300/TH400 can be fitted instead of the ter-
minal block (not with 7MC2000-.....) or in the high hinged
cover. Additional fitting only possible in high hinged cover.
• If using intrinsically-safe temperature sensors any installed
temperature transmitters must also be intrinsically-safe.

Siemens FI 01 · 2008 3/83


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Thermocouples
Questionnaire for temperature sensors
(resistance thermometers and thermocouples)

■ General information
Customer: ....................................................................................
Address: ......................................................................................
Contact partner: .........................................................................
Purchasing dept.: ........................................................................ Tel.: ...........................................................................
3 Sales dept.: ................................................................................. Tel.: ...........................................................................
Process dept.: ............................................................................. Tel.: ...........................................................................
Inquiry:.........................................................................................
Quotation: ....................................................................................
Place and date: ...........................................................................

■ Operating conditions ■ Miscellaneous


1. Application: ...................................................................... Please additionally provide the following: rough sketch, installa-
(e.g. exhaust gas measurement) tion diagram, section of drawing, photo

2. Location:...........................................................................
■ Sensor design
(e.g. pipe bend, tank)
1. Measuring element...........................................................
3. Mounting position:............................................................ (type and standard) (e.g. Pt100 or TC type K)
(e.g. vertical, 45° against flow)
1.1. Tolerance:.........................................................................
4. Temperature (measuring point): ......................................
1.2. Design: .............................................................................
Operating temperature:....................................................
(e.g. Pt100 or 2, 3 or 4-wire system)
Temperature range: .........................................................
1.3. Degree of protection/type of protection: ..........................
5. Medium: ...........................................................................
2. Protective fitting:...............................................................
6. Pressure: ..........................................................................
Nominal pressure: ............................................................ 2.1. Protective tube: ................................................................
Operating pressure: ......................................................... (dimensions/material)
7. Flow: ................................................................................. 2.2. Mounting: .........................................................................
(dimensions/material)
8. Vibrations: ........................................................................
2.3. Neck tube:........................................................................
9. Miscellaneous: .................................................................
(dimensions/material)
(e.g. vessel or pipe materials, PTFE lining)
2.4. Mounting length/nominal length:......................................
■ Ambient conditions 3. Material certificates: .........................................................
(e.g. seawater atmosphere, chemical plant)
4. Connection: ......................................................................
Definition:.................................................................................
4.1. Connection head/box:......................................................
.................................................................................................
4.2. Cable:...............................................................................
.................................................................................................
(dimensions/insulation/standard)
■ Special information 4.3. Other: ...............................................................................
1. Mounting of temperature transmitter in connection head: .........................................................................................
......................................................................................... 5. Tests: ................................................................................
......................................................................................... .........................................................................................
2. Packaging regulations: .................................................... 6. Accessories:.....................................................................
......................................................................................... .........................................................................................
.........................................................................................
7. Supplementary requirements:..........................................
.........................................................................................

3/84 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Thermocouples
Straight thermocouples to DIN 43733,
with connection head

■ Overview Selection and Ordering data Order No.


Straight thermocouple with
Ni Cr/Ni thermocouple (type K)
with metallic protective tube
to 1000 °C (1832 °F)
X 10 CrAl 24, mat. No. 1.4762
22 mm Ø x 2 mm (0.87 inch x 0.079 inch)

3
1 thermocouple
Leg diameter 2 mm (0.08 inch)
Weight: 1.1 ... 2.9 kg (2.4 ... 6.4 lb)
Nominal length in mm (inch):
• 500 (19.7) 7 MC 2 0 0 0 - 1 D C 0 7
• 710 (28.0) 7 MC 2 0 0 0 - 2 D C 0 7
• 1000 (39.4) 7 MC 2 0 0 0 - 3 D C 0 7
2 thermocouples
The straight thermocouple together with a metal protective tube Leg diameter 2 mm (0.08 inch)
is suitable for temperatures from 0 to 1250 °C (32 to 2282 °F) and Weight: 1.1 ... 3.2 kg (2.4 ... 7.0 lb)
can be supplied with a built-in temperature transmitter. Nominal length in mm (inch)
• 500 (19.7) 7 MC 2 0 0 0 - 1 D D 0 7
• 710 (28.0) 7 MC 2 0 0 0 - 2 D D 0 7
■ Technical specifications • 1000 (39.4) 7 MC 2 0 0 0 - 3 D D 0 7
Thermocouples Ni Cr/Ni type K to 1100 °C (2012 °F)
X 18 CrN28, material No. 1.4749
• Number 1 or 2 26 mm Ø x 4 mm (1.02 inch x 0.16 inch)
• Leg diameter 2 ... 3 mm (0.08 ... 0.12 inch) 1 thermocouple
Leg diameter 3 mm (0.12 inch)
• Insulation of legs Insulating beads Weight: 1.3 ... 2.2 kg (2.7 ... 4.8 lb)
Protective tube Metal Nominal length in mm (inch):
• 500 (19.7) 7 MC 2 0 0 0 - 1 E C 0 7
Connection head Form A, DIN 43729; made of cast • 710 (28.0) 7 MC 2 0 0 0 - 2 E C 0 7
light alloy, with one cable bushing • 1000 (39.4) 7 MC 2 0 0 0 - 3 E C 0 7
2 thermocouples
■ Dimensional drawings Leg diameter 3 mm (0.12 inch)
Weight: 1.4 ... 2.4 kg (3.1 ... 5.3 lb)
Nominal length in mm (inch):
• 500 (19.7) 7 MC 2 0 0 0 - 1 E D 0 7
• 710 (28.0) 7 MC 2 0 0 0 - 2 E D 0 7
• 1000 (39.4) 7 MC 2 0 0 0 - 3 E D 0 7
to 1200 °C (2192 °F)
X 15 CrNi Si 24 19, material No. 1.4841
22 mm Ø x 2 mm (0.87 inch x 0.079 inch)
1 thermocouple
Leg diameter 2 mm (0.08 inch)
Weight: 1.7 ... 2.9 kg (3.7 ... 6.4 lb)
Nominal length in mm (inch):
• 500 (19.7) 7 MC 2 0 0 0 - 1 F C 0 7
• 710 (28.0) 7 MC 2 0 0 0 - 2 F C 0 7
Straight thermocouple, dimensions in mm (inches) • 1000 (39.4) 7 MC 2 0 0 0 - 3 F C 0 7
2 thermocouples
■ Design Leg diameter 2 mm (0.08 inch)
Weight: 1.9 ... 3.1 kg (4.2 ... 6.8 lb)
Nominal length in mm (inch):
• 500 (19.7) 7 MC 2 0 0 0 - 1 F D 0 7
• 710 (28.0) 7 MC 2 0 0 0 - 2 F D 0 7
• 1000 (39.4) 7 MC 2 0 0 0 - 3 F D 0 7
To 1250 °C (2282 °F)
CrAl 205 (Megapyr), material No. 1.4767
22 mm Ø x 2 mm (0.87 inch x 0.079 inch)
1 thermocouple
Leg diameter 3 mm (0.12 inch)
Weight: 1 ... 2.9 kg (2.2 ... 6.4 lb)
Nominal length in mm (inch):
• 500 (19.7) 7 MC 2 0 0 0 - 1 H C 0 7
• 710 (28.0) 7 MC 2 0 0 0 - 2 H C 0 7
• 1000 (39.4) 7 MC 2 0 0 0 - 3 H C 0 7
Straight thermocouple with base-metal element Ni Cr/Ni with metal pro- 2 thermocouples
tective tube Leg diameter 3 mm (0.12 inch)
Weight: 1.1 ... 3.2 kg (2.4 ... 7.0 lb)
Nominal length in mm (inch):
• 500 (19.7) 7 MC 2 0 0 0 - 1 H D 0 7
• 710 (28.0) 7 MC 2 0 0 0 - 2 H D 0 7
• 1000 (39.4) 7 MC 2 0 0 0 - 3 H D 0 7
Connection head, form A,
• made of cast light alloy,
with 1 cable inlet and
- screw cover 1
- high hinged cover 6

Siemens FI 01 · 2008 3/85


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Thermocouples
Straight thermocouples
Individual parts and accessories

Selection and Ordering data Order No. Selection and Ordering data Order No.
Straight thermocouple with Thermocouples elements for straight
Ni Cr/Ni thermocouple (type K) thermocouple according to DIN 43733
for temperatures to 1250 °C (2282 °F);
with metallic protective tube Base-metal thermocouple with insulating
beads
Further designs Order code Wire diameter 3 mm (0.12 inch)
Please add "-Z" to Order No. and specify Ni Cr/Ni, to 1000 °C (maximal 1300 °C),

3
Order code(s) and plain text. (to 1832 °F (max. 2372 °F))
• Different design (mounting length, protec- Y01 0.55 ... 2.10 kg (1.21 ... 4.63 lb)
tive tube material etc.), specify in plain text. Nominal Thermocouple
• TAG plate made of stainless steel Y15 length L1 in length L2 in
specify TAG No. in plain text mm (inch): mm (inch):
• Calibration carried out at one point, specify Y33 • 500 (19.7) 540 (21.3) 7MC2903-1CA
desired temperature in plain text (order • 710 (28.0) 750 (29.5) 7MC2903-2CA
equivalent number of times for several cal- • 1000 (39.4) 1040 (40.9) 7MC2903-3CA
ibration points).
To order a temperature transmitter installed in the connection head,
see "Temperature transmitters for installation in the connection
head" (page 3/83).
Installation of a transmitter is only possible here in the versions with a
high hinged cover (7MC2000-....6).

Selection and Ordering data Order No.


Metallic protective tubes for straight
thermocouple elements according
to DIN 43733
X 10 CrAl 24, material No. 1.4762
; 22 mm x 2 mm (; 0.87 inch x 0.08 inch),
0.55 ... 1.10 kg (1.21 ... 2.42 lb), dished
Nominal length Protective tube length
in mm (inch): in mm (inch):
• 500 (19.7) 520 (20.5) 7MC2900-1DA
• 710 (28.0) 730 (28.7) 7MC2900-2DA
• 1000 (39.4) 1020 (40.2) 7MC2900-3DA
X 10 CrAl 24, material No. 1.4749
; 26 mm x 4 mm (; 1.02 inch x 0.16 inch),
1.25 ... 2.20 kg (2.76 ... 4.85 lb), dished
Nominal length Protective tube length
in mm (inch): in mm (inch):
• 500 (19.7) 520 (20.5) 7MC2900-1EC
• 710 (28.0) 730 (28.7) 7MC2900-2EC
• 1000 (39.4) 1020 (40.2) 7MC2900-3EC
X 15 CrNiSi 25 20, material No. 1.4841
; 22 mm x 2 mm (; 0.87 inch x 0.08 inch),
1.05 kg (2.31 lb), dished
Nominal length Protective tube length
in mm (inch): in mm (inch):
• 1000 (39.4) 1020 (40.2) 7MC2900-3FA
CrAl 205 (Megapyr), material No. 1.4767
; 22 mm x 2 mm (; 0.87 inch x 0.05 inch),
0.55 ... 1.10 kg (1.21 ... 2.42 lb)
Nominal length Protective tube length
in mm (inch): in mm (inch):
• 500 (19.7) 520 (20.5) 7MC2900-1HA
• 710 (28.0) 730 (28.7) 7MC2900-2HA
• 1000 (39.4) 1020 (40.2) 7MC2900-3HA

3/86 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Thermocouples
Straight thermocouples
Individual parts and accessories

■ Connection heads ■ Mounting accessories for connection heads


Connection head, form A (without terminal block and terminals) • Terminal block
for protective tube diameter (bore = protective tube diameter • Terminal
+0.5 mm (0.02 inch)) • Set of gaskets
• Set of washers
Selection and Ordering data Order No. • Mounting flange
Connection head, form A,
(without terminal block and terminals)
• Threaded sleeve 3
1 Cable inlet, degree of protection IP53,
0.35 kg (0.77 lb)
Selection and Ordering data Order No.
Cast light alloy
Mounting accessories
fastener, unscrewable
for protective tube diameter in mm (inch) Terminal block without terminals 7MC2998-1AA
(bore = protective tube diam. +0.5 mm) for base-metal thermocouples;
(0.02 inch): 0.06 kg (0.13 lb)
• 22 (0.87) 7MC2905-1AA Terminal 7MC2998-1BA
• 26 (1.02) 7MC2905-1BA for base-metal thermocouples;
Cast light alloy 0.01 kg (0.02 lb)
high hinged cover Set of gaskets (100 off) 7MC2998-1CA
for protective tube diameter in mm (inch) for the connection head cover;
(bore = protective tube diam. +0.5 mm) 0.01 kg (0.02 lb)
(0.02 inch):
• 22 (0.87) 7MC2905-4AA Set of washers (100 off) 7MC2998-1CB
• 26 (1.02) 7MC2905-4BA for the terminal block; 0.01 kg (0.02 lb)
Mounting flange, adjustable; made of GTW
• for protective tube outer diameters 7MC2998-2CB
22 mm (0.87 inch); 0.35 kg (0.77 lb)
• for protective tube outer diameters 7MC2998-2CC
26 mm (1.02 inch); 0.32 kg (0.71 lb)
Threaded sleeve
Gas-tight up to bar (14.5 psi), adjustable,
materiall No. 1.0718, with gasket;
0.40 kg (0.88 lb)
• for protective tube outer diameters 7MC2998-2DB
22 mm (0.87 inch), G1
• for protective tube outer diameters 7MC2998-2DC
26 mm (1.02 inch), G1

■ Dimensional drawings

Mounting flange to DIN 43734 (left) and threaded sleeve (right) for install-
ing straight thermocouples, dimensions in mm (inches)

Siemens FI 01 · 2008 3/87


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Thermocouples
Jacket thermocouples with extension lead

■ Overview Selection and Ordering data Order No.

The jacket thermocouple with extension lead is suitable for a Jacket thermocouple element
with extension lead
temperature range from 0 to 700, 1000 or 1100 °C (32 to 1292,
1832 or 2012 °F); a temperature of 80 to 260 °C (176 to 500 °F) Maximum temperature of medium 700 °C
is permissible for the extension lead. (1292 °F), jacket outer diameter 0.5 mm
(0.02 inch)

■ Technical specifications Extension lead type L2KK

3 Thermocouple Ni Cr/Ni, type K, one off


Nominal length in mm (inch):
• 300 (11.8) 7MC2027-2BA10
Jacket Maximum temperature of medium 1000 °C
(1832 °F), jacket outer diameter 1 mm
• Form Outer diameter 0.5, 1.0, 1.5 or (0.04 inch)
3 mm (0.02, 0.04, 0.06 or
Extension lead type L2SS
0.12 inch)
Nominal length in mm (inch):
• Minimum bending radius fivefold outer diameter • 100 (3.94) 7MC2027-1AA20
• Material Inconel; NiCr 15 Fe, • 300 (11.8) 7MC2027-2AA20
mat. No. 2.4816 • 500 (19.7) 7MC2027-3AA20
Extension lead type L2KK
Extension lead Nominal length in mm (inch):
• Number of conductors: 2 • 100 (3.94) 7MC2027-1BA20
• Conductor cross-section: 0.22 mm2 (0.00034 inch2) • 300 (11.8) 7MC2027-2BA20
• Length: 3 m (9.84 ft) • 500 (19.7) 7MC2027-3BA20

Type Max. tem- Conductor Insulation Maximum temperature of medium 1100 °C


perature material (2012 °F), jacket outer diameter 1.5 mm
(0.06 inch)
°C (°F) single common Extension lead type L2SS
L2SS 180 (356) Ni Cr/Ni Silicon Silicon Nominal length in mm (inch):
• 100 (3.94) 7MC2027-1AA30
L2KK 80 (176) Ni Cr/Ni PVC PVC
• 300 (11.8) 7MC2027-2AA30
L2TGD 260 (500) Ni Cr/Ni PTFE Glass-filament • 500 (19.7) 7MC2027-3AA30
and external
Extension lead type L2KK
stainless steel
braiding Nominal length in mm (inch):
• 100 (3.94) 7MC2027-1BA30

■ Dimensional drawings • 300 (11.8)


• 500 (19.7)
7MC2027-2BA30
7MC2027-3BA30
Jacket outer diameter 3.0 mm (0.12 inch)
Extension lead type L2SS
Nominal length in mm (inch):
• 100 (3.94) 7MC2027-1AA40
• 300 (11.8) 7MC2027-2AA40
• 500 (19.7) 7MC2027-3AA40
Extension lead type L2KK
Nominal length in mm (inch):
• 100 (3.94) 7MC2027-1BA40
• 300 (11.8) 7MC2027-2BA40
Jacket thermocouple with extension lead, dimensions in mm (inches) • 500 (19.7) 7MC2027-3BA40
Extension lead type L2TGD
Nominal length in mm (inch):
• 100 (3.94) 7MC2027-1CA40
• 300 (11.8) 7MC2027-2CA40
• 500 (19.7) 7MC2027-3CA40
Further designs Order code
Please add "-Z" to Order No., and specify
Order code(s) and plain text.
• Different design (mounting length, protec- Y01
tive tube material etc.), specify in plain text.
• TAG plate made of stainless steel Y15
specify TAG No. in plain text
• Calibration carried out at one point, specify Y33
desired temperature in plain text (order
equivalent number of times for several cal-
ibration points).

3/88 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Thermocouples
Jacket thermocouples
with connection head, form B

■ Overview ■ Dimensional drawings

3
Jacket thermocouple with connection head, form B, made of cast light al-
loy, dimensions in mm (inches)
Selection and Ordering data Order No.
Jacket thermocouple element
The jacket thermocouple with connection head, form B is suit- with connection head, form B
able for the temperature range from 0 to 800 or 1,100 °C Fe/Cu Ni type J thermocouple
(32 to 1472 or 2012 °F) and can also be supplied with a built-in Jacket made of X 6 CrNiTi 18 10, material No.
temperature transmitter. 1.4541, temperature up to 800 °C (1472 °F)
1 thermocouple
■ Technical specifications Jacket outer diameter/mm (inch):
• 3.0 (0.12) 7MC2021-2CE-Z
Jacket thermocouple with connection head, form B • 6.0 (0.24) 7MC2021-4CE-Z
2 thermocouples
Thermocouple Fe/Cu Ni, Fe-CuNi or Ni Cr/Ni Jacket outer diameter/mm (inch):
Jacket Measuring junction insulated from base of • 3.0 (0.12) 7MC2021-2CF-Z
jacket • 6.0 (0.24) 7MC2021-4CF-Z
• Material X 6 CrNiTi 18 10, mat. Inconel Fe/Cu Ni type L thermocouple
No. 1.4541 Jacket made of X 6 CrNiTi 18 10, material No.
1.4541, temperature up to 800 °C (1472 °F)
• Temperature Up to 800 °C (1472 °F) Up to 1100 °C 1 thermocouple
(2012 °F) Jacket outer diameter/mm (inch):
Connection head Form B to DIN 43729 • 3.0 (0.12) 7MC2021-2CA-Z
• 6.0 (0.24) 7MC2021-4CA-Z
• Material Cast light alloy 2 thermocouples
• Temperature -30 ... +100 °C (-22 ... +212 °F) is permissible Jacket outer diameter/mm (inch):
• 3.0 (0.12) 7MC2021-2CB-Z
• Cable inlet For cables up to 15 mm (0.59 inch) diameter • 6.0 (0.24) 7MC2021-4CB-Z
• Terminal block Without clamping springs Ni Cr/Ni type K thermocouple
Jacket made of Inconel,
Thermocouple temperature up to 1,100 °C (2012 °F)
• Insulation resistance between conductors and jacket: 1 thermocouple
≥ 1000 MΩ/m at 20 °C (≥ 305 MΩ/ft at 68 °F) Jacket outer diameter/mm (inch):
• 3.0 (0.12) 7MC2021-2LC-Z
• Minimum bending radius: 5 x outer diameter of jacket • 6.0 (0.24) 7MC2021-4LC-Z
• Leak resistance of measuring junction: 2 thermocouples
Tested at 40 bar (580 psi) Jacket outer diameter/mm (inch):
• 3.0 (0.12) 7MC2021-2LD-Z
Thermo- Jacket No. of Thermo- Resis- • 6.0 (0.24) 7MC2021-4LD-Z
couple thermo- couple tance (for Further designs Order code
Outer Wall couples wire double
diameter thickness Please specify Order code(s) and plain text.
diameter conduc-
tor) • Length L in m (ft)
- 0.25 (0.82) A01
mm (inch) mm (inch) mm (inch) Ω/m (Ω/ft) - 1 (3.28) A02
Fe/Cu Ni 3.0 (0.12) 0.42 1 0.54 (0.02) 2.7 (8.9) - 5 (16.4) A03
(type J) and (0.017) - 10 (32.8) A04
Fe-CuNi • Different design (mounting length, protec- Y01
(type L) 0.42 2 0.48 (0.02) 3.5 (11.5) tive tube material etc.), specify in plain text.
(0.017) • TAG plate made of stainless steel Y15
6.0 (0.24) 0.55 1 1.07 (0.04) 0.66 (2.2) specify TAG No. in plain text
(0.022) • Calibration carried out at one point, specify Y33
desired temperature in plain text (order
0.89 2 0.81 (0.03) 1.14 (3.7) equivalent number of times for several cal-
(0.035) ibration points).
Ni Cr/Ni 3.0 (0.12) 0.42 1 0.54 (0.02) 4.3 (14.1) To order a temp. transmitter installed in the connection head, see
(type K) (0.017) "Temp. transm. for installation in the connection head" (page 3/83).
0.42 2 0.48 (0.02) 5.5 (18.0) Note: SITRANS T3K PA can only be used in a connection head with high
(0.017) hinged cover. Can be ordered als special version (...-Z Y01).
6.0 (0.24) 0.55 1 1.07 (0.04) 1.12 (3.7) Example of ordering:
(0.022)
The following is required:
0.89 2 0.81 (0.03) 1.94 (6.4) • Jacket thermocouple element
(0.035) jacket made of X 6 CrNiTi 18 10; outer diameter 3.0 mm (0.12 inch);
- with 1 Fe/Cu Ni thermocouple, type J; nominal length 5 m (16.4 ft)
Order as follows:
1 jacket thermocouple 7MC2021-2CE-Z A03

Siemens FI 01 · 2008 3/89


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Thermocouples
Jacket thermocouples with socket

■ Overview Selection and Ordering data Order No.

The sensor is suitable for a temperature range from 0 to 800, Jacket thermocouple element
with coupling socket
1000 or 1100 °C (32 to 1472, 1832 or 2012 °F); a maximum tem-
Ni Cr/Ni type K thermocouple
perature of 130 °C (266 °F) is permissible for the plug connec-
tion. Jacket material No. 1.4541
Temperature of medium 800 °C (1472 °F)
■ Technical specifications 1 thermocouple

3 Thermocouple Ni Cr/Ni
Jacket outer dia./
mm (inch):
Length L/m (ft):

• 3.0 (0.12) 0.3 (0.98) 7MC2028-1AC31


Jacket
• 3.0 (0.12) 0.5 (1.64) 7MC2028-2AC31
• Material X 6 CrNiTi 18 10, mat. Inconel, NiCr 15 Fe, 2 thermocouples
No. 1.4541 mat. No. 2.4816
Jacket outer dia./ Length L/m (ft):
• Temperature Up to 800 °C (1472 °F) Up to 1100 °C (2012 °F) mm (inch):
Coupling socket • 3.0 (0.12) 0.3 (0.98) 7MC2028-1AD31
• 3.0 (0.12) 0.5 (1.64) 7MC2028-2AD31
• Housing Made of polyamide • 6.0 (0.24) 0.3 (0.98) 7MC2028-1AD41
• Contact Made of temperature-resistant material • 6.0 (0.24) 0.5 (1.64) 7MC2028-2AD41
• Coupling plug See "Jacket thermocouples", "Individual parts"; Jacket material No. 2.4816
incorrect connection impossible Temperature of medium 1000 °C (1832 °F)
1 thermocouple
■ Dimensional drawings Jacket outer
diameter/mm (inch):
Length L/m (ft):

• 1.0 (0.04) 0.3 (0.98) 7MC2028-1AC12


• 1.0 (0.04) 0.5 (1.64) 7MC2028-2AC12
Jacket material No. 2.4816
Temperature of medium 1100 °C (2012 °F)
1 thermocouple
Jacket outer dia./ Length L/m (ft):
mm (inch):
• 1.5 (0.06) 0.3 (0.98) 7MC2028-1AC22
• 1.5 (0.06) 0.5 (1.64) 7MC2028-2AC22
• 3.0 (0.12) 0.3 (0.98) 7MC2028-1AC32
• 3.0 (0.12) 0.5 (1.64) 7MC2028-2AC32
• 6.0 (0.24) 0.3 (0.98) 7MC2028-1AC42
• 6.0 (0.24) 0.5 (1.64) 7MC2028-2AC42
Jacket thermocouple with socket; at top with one thermocouple, at bot- 2 thermocouples
tom with two thermocouples; dimensions in mm (inches)
Jacket outer dia./ Length L/m (ft):
mm (inch):
• 3.0 (0.12) 0.3 (0.98) 7MC2028-1AD32
• 3.0 (0.12) 0.5 (1.64) 7MC2028-2AD32
• 6.0 (0.24) 0.3 (0.98) 7MC2028-1AD42
• 6.0 (0.24) 0.5 (1.64) 7MC2028-2AD42
Further designs Order code
Please add "-Z" to Order No. and specify
Order code(s) and plain text.
• Different design (mounting length, protec- Y01
tive tube material etc.), specify in plain text.
• TAG plate made of stainless steel Y15
specify TAG No. in plain text
• Calibration carried out at one point, specify Y33
desired temperature in plain text (order
equivalent number of times for several
calibration points).
Different designs (mounting length, protective tube material etc.)
on request: Add Order code "Y01" and specify in plain text.
(Please order coupling plug separately (see "Individual parts"))

3/90 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Thermocouples
Individual parts
Selection and Ordering data Order No.
Extension lead for
jacket thermocouple with coupling socket
(7MC2028); twisted cores
for Ni Cr/Ni thermocouple, color code: green
Outer diameter 4 mm (0.16 inch) 7MC2921-1AC-Z
PVC isolation

3
Number of conductors and cross-section:
2 x 0.22 mm² (2 x 0.00034 inch²)
Temperature range -20 ... +80 °C
(-4 ... +176 °F)
Outer diameter 4 mm (0.16 inch) 7MC2921-2AC-Z
Silicone-silicone isolation d D l L SW
Number of conductors and cross-section:
2 x 0.22 mm² (2 x 0.00034 inch²) mm (inch) mm (inch) mm (inch)
Silicone-silicone isolation 1,5 (0.06) M8x1 8 (0.3) 32 (1.3) 12
Temperature range -40 ... +180 °C
(-40 ... +356 °F) 3,0 (0.12) M8x1 8 (0.3) 32 (1.3) 12
Outer diameter 6 mm (0.24 inch) 7MC2921-3AC-Z 6,0 (0.24) G¼ 10 (0.4) 48 (1.9) 19
PVC isolation
Compression joint with M8 x 1 or G¼ thread dimensions in mm (inches)
Number of conductors and cross-section:
4 x 0.22 mm² (2 x 0.00034 inch²)
Temperature range -20 ... +80 °C
(-4 ... +176 °F)
Outer diameter 6 mm (0.24 inch) 7MC2921-4AC-Z
Silicone-silicone isolation
Number of conductors and cross-section:
4 x 0.22 mm² (2 x 0.00034 inch²)
Temperature range -40 ... +180 °C
(-40 ... +356 °F)
Coupling socket, dimensions in mm (inches)
Further designs Order code
Please specify Order code(s) and plain text. Selection and Ordering data Order No.
• Length of extension lead in m (ft) Screw nipple
- 0.25 (0.82) A01 for soldering jacket thermocouple elements
- 1 (3.28) A02 The screw nipple has a centering hole and must be
- 5 (16.4) A03 drilled according to the outer diameter of the jacket
thermocouple element.
- 10 (32.8) A04 The thermocouple must be protected from exces-
sive heat when soldering (e.g. by blowing with air).
■ Other individual parts • with M8 x 1 thread, for jacket outer diameter up to 7MC2922-1EA
• Screw nipple 3.0 mm (0.12 inch); 0.01 kg (0.02 lb)
• Compression joint • with G¼ thread, for all jacket outer diameters list- 7MC2922-1FB
ed in the catalog, 0.02 kg (0.04 lb)
• Coupling socket
Compression joint, gas-tight
• Coupling plug Temperature up to 350 °C (662 °F)
Pressure-tight up to 80 bar (1160 psi)
■ Dimensional drawings Subsequent release and movement is not possible,
material X6 CrNi Mo Ti 17–122, material No 1.4571;
tapered ring like compression joint
• Thread M8 x 1; 0.03 kg (0.06 lb)
- jacket outer diameter 1.5 mm (0.06 inch) 7MC2922-3AA
- jacket outer diameter 3.0 mm (0.12 inch) 7MC2922-3BA
• Thread G¼; 0.04 kg (0.08 lb)
- jacket outer diameter 6 mm (0.24 inch) 7MC2922-3DB
Coupling socket
of jacket thermocouple 7MC2028 (one per element); 7MC2922-4BB
for Ni Cr/Ni thermocouple element, type K; 0.05 kg
(0.11 lb)
Screw nipple with M8 x 1 thread (left) or G¼ (right), dimensions in mm
(inches) Coupling plug
matching the jacket thermocouple 7MC2028 with 7MC2922-4BD
socket; one plug is required per thermocouple ele-
ment; for Ni Cr/Ni thermocouple, type K; 0.05 kg
(0.11 lb)

Coupling plug, dimensions in mm (inches)

Siemens FI 01 · 2008 3/91


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Thermocouples
Accessories
Cold junction with built-in power supply unit

■ Overview Power supply 47 ... 63 Hz, 24 V DC,


220 V, 110 V or 24 V AC, - 25%, + 33%
- 25%, + 33%
• Power consumption Approx. 0.1 VA 0.1 W
• Residual ripple – < 10%
Permissible ambient tem- -10 ... +65 °C (14 ... 149 °F)
perature

3 Permissible storage tem-


perature
-30 ... +80 °C (-22 ... +176 °F)

Degree of protection DIN 40050


• Housing IP40
• Terminals IP20
Weight Approx. 0.1 kg (0.22 lb)
The cold junction is suitable for a thermocouple element of type
J, K, L, R, S, T or U. A correction circuit is fitted which compen- Selection and Ordering data Order No.
sates the influence of changes in the cold junction temperature.
The device can be mounted on a rail. Cold junction M 7 2 1 6 6 - 77777
Note: positioning of the cold junction in the vicinity of the mea- with built-in power supply unit,
suring point saves expensive extension leads. for rail mounting
Power supply
■ Design 220 V AC
110 V AC
B1
B2
The cold junction has a terminal housing made of plastic which 24 V AC B3
is suitable for mounting on a standard rail DIN EN 50022-35 x 24 V DC B4
7.5.
The sensor for the cold junction temperature is positioned di- Connection to thermocouple element
rectly on the terminals for the extension lead. The device is Fe-CuNi Type L 1
matched to the type of thermocouple using built-in function Fe/Cu Ni Type J 2
plugs and adjusted according to the reference temperature. The Ni Cr/Ni Type K 3
built-in power supply unit provides the cold junction with power. Pt10% Rh/Pt Type S 4
Pt 13% Rh/Pt Type R 5
■ Function Cu-CuNi
Cu/Cu Ni
Type U
Type T
6
7
Mode of operation Ref. temperature
0 °C (32 °F) 00
20 °C (68 °F) 20
Caution! Compensation of changes in the cold junction temperature
should not be carried out for Pt 30% Rh/Pt 6% Rh thermocouple
elements. The extension leads should be routed to a position where a
temperature between –10 and +60 °C (14 to 140 °F) exists.

■ Dimensional drawings

Function diagram of cold junction


The power supply (1) stabilized in constant-voltage transformer
(4) the feeds a bridge containing a precision temperature sensor Cold conjunction, dimensions in mm (inches)
(5) in one of its branches. The resistance of the temperature sen-
sor changes with the temperature of the terminals, and thus also ■ Schematics
the diagonal voltage of the bridge. The bridge is designed ac-
cording to the characteristic of the respective thermocouple el-
ement.

■ Technical specifications
Signal input For thermocouple elements acc. to
DIN IEC 584 and DIN 43710
Ref. temperature 0 or 20 °C (32 or 68 °F)
Error limits 0.5 K (0.9 °F)
Temperature error <0.1 K / 10 K change (0.18°F/18 °F)
Connection diagram of cold junction

3/92 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Thermocouples
Accessories - Cold junction thermostat
for 6 or 12 measuring points

■ Overview Selection and Ordering data Order No.


Cold junction thermostat 7 MC 2 9 3 3 - 77777
Ref. temperature
50 °C (112 °F) 1
60 °C (140 °F) 2
70 °C (158 °F) 3
Thermocouple
Fe-CuNi Type L A 3
Ni Cr/Ni Type K B
Pt 10% Rh/Pt Type S C
Cu-CuNi Type U D
Number of cold junctions
6 A
The cold junction thermostat is used to keep the cold junction
12 B
temperature constant when measuring temperature using ther-
mocouples. Power supply
220 V AC 1
■ Design 110 V AC 2
24 V AC 3
The cold junction thermostat consists of an aluminium block in 24 V DC, 30 W 4
which 6 or 12 reference elements are fitted, a controller, a heater 24 V DC, 17 W 5
and a power supply unit. The aluminium block, controller, heater
and power supply unit are arranged in two blocks potted in syn- Temperature control
thetic resin. In this manner, uniform temperature distribution and without 0
protection against external effects are ensured. The cast blocks with resistance thermometer 1
are fitted to the base plate using metal brackets, thus ensuring with electronic monitoring unit 2
high resistance to vibration.
The device has a red lamp visible on the outside of the door to
permit checking of the temperature in the aluminium block (ther-
mostat). It is also possible to fit
■ Dimensional drawings
• a Pt100 resistance thermometer or
• an electronic monitoring unit with relay output whose NC con-
tact opens if there is a fault.
The cold junction thermostat has a sheet-steel housing.

■ Function
The temperature of the aluminium block is held constant at a ref-
erence temperature of 50, 60 or 70 °C (122, 140 or 158 °F) by an
electronic two-step controller with low hysteresis.
With a constant cold junction temperature, the difference be-
tween the voltage of the measuring element and that of the ref-
erence element is a measure of the temperature at the measur-
ing point. The thermocouples are electrically isolated from one
another and from the power supply.
Cold junction thermostat 7MC2933, dimensions in mm (inches)
■ Technical specifications
Number of measuring 6 or 12 ■ Schematics
points
Ref. temperature 50, 60 or 70 °C (122, 140 or 158 °F)
• Error limits 0.5 °C (0.9 °F)
Effect of ambient tempera- 0.2 °C / 10 K (0.36 °F/18 °F) temperature
ture change
Error limits of control 0.05 °C (0.09 °F)
Permissible ambient tem- -20 ... +45 °C (-4 ... +113 °F)
perature
Power supply (depending Warm-up time
on version)
48 ... 63 Hz,
24, 110 or 220 VAC ± 15%,
• Approx. 30 VA Approx. 10 min
• 24 V DC, approx. 30 W Approx. 10 min
• 24 V DC, approx. 17 W Approx. 20 min
Degree of protection IP54 to DIN 40050
Resistance to vibration Frequency Acceleration Test period
Cold junction thermostat 7MC2933, connection diagram
5 ... 55 Hz 1 g, constant 4 min
30 Hz 10 g 1 min
Weight Approx. 9.5 kg (20.9 lb)

Siemens FI 01 · 2008 3/93


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Thermocouples
Accessories - Cold junction thermostat
for 12 or 24 measuring points

■ Overview Selection and Ordering data Order No.


Cold junction thermostat 7 MC 2 9 3 0 - 77777
Ref. temperature
50 °C (112 °F) 1
60 °C (140 °F) 2
70 °C (158 °F) 3

3 Thermocouple
Fe-CuNi Type L A
Ni Cr/Ni Type K B
Pt10% Rh/Pt Type S C
Cu-CuNi Type U D
Number of cold junctions
12 A
The cold junction thermostat is used to keep the cold junction
24 B
temperature constant when measuring temperature using ther-
mocouples. Power supply
220 V AC 1
■ Design 110 V AC 2
24 V AC 3
The cold junction thermostat consists of an aluminium block in 24 V DC 4
which 12 or 24 reference elements are fitted, a controller, a
heater and a power supply unit. Temperature control
The aluminium block, controller, heater and power supply unit with resistance thermometer;
electronic monitoring of the temperature
are designed as crystal modules. In this manner, uniform tem-
perature distribution and protection against external effects are • in the thermostat
ensured. - with relay output 1
A Pt100 resistance thermometer or an electronic monitoring unit - without 24 V DC output 2
with relay output or 24 V DC output are fitted to permit checking
of the temperature in the aluminium block (thermostat). The mon-
itoring unit is triggered when the maximum or minimum temper-
■ Dimensional drawings
ature of the cold junction block is violated or if the power supply
fails.
A signalling lamp "Operation" is present in the door.
The cold junction thermostat has a sheet-steel housing.

■ Function
The temperature of the aluminium block is held constant at a ref-
erence temperature of 50, 60 or 70 °C (122, 140 or 158 °F) by an
electronic two-step controller with low hysteresis.
With a constant cold junction temperature, the difference be-
tween the voltage of the measuring element and that of the ref-
erence element is a measure of the temperature at the measur-
ing point. The thermocouples are electrically isolated from one
another and from the power supply.

■ Technical specifications Cold junction thermostat 7MC2930, dimensions in mm (inches)

Number of measuring points


Ref. temperature
12 or 24
50, 60 or 70 °C
■ Schematics
(122, 140 or 158 °F)
• Error limits 0.5 K (0.9 °F)
Permissible ambient temperature
with ref. temperature:
• 50 °C (122 °F) -20 ... +45 °C (-4 ... +113 °F)
• 60 or 70 °C (32 or 158 °F) -20 ... +65 °C (-4 ... +149 °F)
Effect of ambient temperature 0.2 K (0.36 °F) per 10 K (1.8 °F)
temperature change
Error limits of temperature control 0.05 K (0.09 °F)
Monitoring circuit
• Upper switching point ≤ Ref. temperature + 5 K (9 °F)
• Lower switching point ≤ Ref. temperature – 5 K (-9 °F)
Signalling relais NC contact opens in case of error
• Switching capacity Max. 10 VA (max. 240 V AC,
max. 0.5 A)
Power supply 47 ... 63 Hz, 24, 110 or 220 V AC
± 15%, approx. 30 VA;
24 V DC ± 10%, approx. 30 W Cold junction thermostat 7MC2930, connection diagram

Warm-up time Approx. 10 min


Degree of protection IP55 to DIN 40050
Weight Approx. 13 kg (28.7 lb)

3/94 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Resistance thermometers for food, pharmaceuticals and biotechnology
Resistance thermometers
for installation in pipelines and tanks

■ Overview ■ Technical specifications


Design Replaceable measuring insert
with connection head and protec-
tion fitting
Connection head Either:
• Form B standard, screw cover,

3
stainless steel 1.4301, IP67
• Form B, cover with 2 slotted
screws,
aluminium, IP54, standard
• Form B, screw cover,
plastic, IP54 (BK)
• Form B, hinged cover with slot-
The resistance thermometer is designed for installation in tanks ted screws,
and pipelines as well as for the measurement of temperature aluminium, IP65 (BUZ)
with hygiene requirements. The usual process connections are • Form B, hinged cover with
available. The rugged design means that it is suitable for a wide quick-release,
range of process applications in the food, pharmaceutical and aluminium, IP65 (BUS)
biotechnology industries. The resistance thermometer is also • Form B, high hinged cover with
available with a built-in transmitter. A versatile range of head slotted screw,
transmitters is available for this application. aluminium, IP65 (BUZH)

■ Design
Protective tube Stainless steel 1.4404/316L 6 or
9 mm (0.24 or 0.35 inch) diam.,
optionally with tapered tip, see
• Pt100 measuring resistor Selection and Ordering data for
• Stainless steel measuring insert mounting length U1
• Replaceable measuring insert Measuring insert Stainless steel, replaceable
• Process connections for food/pharmaceuticals/biotechnology Pt100 measuring resistor to
• Hygiene version, design corresponds to EHEDG recommen- DIN 43762
dations Rigid design or as jacket element
• Fast response available with reduced tip (mineral-insulated, flexible,
increased vibration resistance)
• Transmitter can be integrated (4 to 20 mA or PROFIBUS PA)
Accuracy of measuring resistor Class A according to
The resistance thermometer has a replaceable measuring in- DIN EN 60751
sert. The measuring insert contains either one or two Pt100 mea- Integration of transmitter Suitable Pt100 transmitters for
suring resistors which are connected to the base in the connec- head mounting can be fitted in
tion head with a two-wire, three-wire or four-wire system. The the connection head, see Selec-
change in resistance dependent on the measured temperature tion and Ordering data
can be recorded by a transmitter and converted into a standard- Process connections • DIN 11851 with slotted union nut
ized signal.
• Clamp connection to DIN 32676
• Clamp connection to ISO 2852
• Tri-clamp
• Varivent
• Neumo BioControl
• Spherical welding-type sleeve
cyl./sph. 30 x 40 mm
(1.18 x 1.57 inch)
• Aseptic connections
The gasket is not included in the
standard scope of delivery! Fur-
ther process connections on
request. Process connection
material: Stainless steel
1.4404/316L
Surface properties
• Standard Surface roughness Ra < 1.5 µm
(5.9 x 10-5 inch)
• Hygiene Surface roughness Ra < 0.8 µm
(3.1 x 10-5 inch)
• Welded seam < 1.5 µm (5.9 x 10-5 inch)

Siemens FI 01 · 2008 3/95


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Resistance thermometers for food, pharmaceuticals and biotechnology
Resistance thermometers
for installation in pipelines and tanks

Selection and Ordering data Order No. Order code Selection and Ordering data Order No. Order code
Pt100 resistance thermometer for food, 7 MC 8 0 0 5 - Pt100 resistance thermometer for food, 7 MC 8 0 0 5 -
pharmaceuticals and biotechnology pharmaceuticals and biotechnology
7777 0 - 777 0 777 7777 0 - 777 0 777
Connection head Neck tube length M
• Form B, cast light alloy, screw cover, 1 • 80 mm (3.15 inch) 1
IP54, cable gland • 145 mm (5.71 inch) 2

3
• Form B, plastic, screw cover, 2 • Special version: 9 N1Y
IP54, cable gland (add Order code and plain text)
• Form BUZ, cast light alloy, screw cover, 3
IP65, cable gland Mounting length U1
• Form BUZH, cast light alloy, high hinged 4 • 15 mm (0.59 inch) B
cover, IP 65, cable gland • 35 mm (1.38 inch) C
• Form B, stainless steel, standard, IP67, 5 • 50 mm (1.97 inch) D
cable gland • 100 mm (3.94 inch) E
• Special version: 9 H1Y • 160 mm (6.30 inch) F
(add Order code and plain text) • 250 mm (9.84 inch) G
Process connection, material 1.4404/316L • 400 mm (15.75 inch) H
• Milk pipe union to DIN 11851 with slotted • 4 inch J
union nut and nominal diameter/pressure • 6 inch K
- DN 25 / PN 40 AA • 9 inch L
- DN 32 / PN 40 AB • Special version: Z P1Y
- DN 40 / PN 40 AC (add Order code and plain text)
- DN 50 / PN 25 AD Sensor
• Clamp connection: Thin-film technology:
ISO 2852 DIN Tri-Clamp Outer dia- measuring range -50 ... +400 °C
32676 meter D (-58 ... +752 °F)
• 1 x Pt100, class A, three-wire F
– – ½“/ ¾“ 25.0 mm CA • 2 x Pt100, class A, three-wire G
DN 25/ DN 1“, 1½“ 50.5 mm CB • 1 x Pt100, class A, four-wire H
33.7/38 25/32/40
• Special version: Z Q1Y
DN 40/51 DN 50 2“ 64.0 mm CC (add Order code and plain text)
DN 63.5 – 2½“ 77.5 mm CD
DN 88.9 DN 80 – 106.0 mm CE Further designs Order code
• Varivent connection (Tuchenhagen) Add "-Z" to Order No. and
add Order code.
- D = 50 mm (1.97 inch), KU
for Varivent housing DN 25 and DN 1" • Process connection completely electropol- P01
- D = 68 mm (2.68 inch), KV ished
for Varivent housing DN 40 ... DN 125 • Hygiene version H01
and 1½" ... 6" (Ra < 0.8 µm (3.1 x 10-5 inch))
• Welding piece LA • Certificates
(sphere diameter 30 x 40 mm - Roughness depth measurement Ra C18
(1.2 x 1.6 inch) long) certified by factory certificate to
• Special version: ZA J1Y EN 10204-3.1B
Type of screwed gland and nominal diame- - Material certificate to EN 10204-3.1 C19
ter (add Order code and plain text)
• Specify special version in plain text Y01
Protective tube Measuring insert
• TAG plate made of stainless steel Y15
• Ø F1=6 mm Ø 3/3.2 mm, 1 specify TAG No. in plain text
(0.24 inch) (0.12/0.13 inch)
miner. insul. • Test report (at 0, 50 and 100%) Y33
specify measuring range in plain text
• Ø F1=9 mm Ø 6 mm (0.24 inch) 2
(0.35 inch) To order a temperature transmitter installed in the connection head,
• Ø F1=9 mm Ø 6 mm (0.24 inch) 3 see "Temperature transmitters for installation in the connection
(0.35 inch) miner. insul. head" (page 3/64).
• Ø F1=9 mm Ø 3/3.2 mm, 4
(0.35 inch) (0.12/0.12 inch)
tapered tip miner. insul.
F3=5 Ø x 20 mm
(0.2 x 0.79 inch)
• Special version: 9 L1Y
(add Order code and plain text)

3/96 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Resistance thermometers for food, pharmaceuticals and biotechnology
Resistance thermometers
for installation in pipelines and tanks

■ Dimensional drawings
Connection heads
)RUP%KLJKKLQJHGFRYHUZLWK
)RUP%VFUHZFRYHU
 VORWWHGVFUHZDOXPLQLXP
)RUP%FRYHUZLWK ,3 %8=+ 
VORWWHGVFUHZV SODVWLF,3 %. 
DOXPLQLXP,3 0[IRUFDEOHGLDP  

0[ 0[ 3

DSSUR[  
IRUFDEOHGLDP   IRUFDEOHGLDP  
DSSUR[  

DSSUR[  

DSSUR[
DSSUR[

DSSUR[
DSSUR[

 
 

 
 

0
0
0

)RUP%VWDQGDUGVFUHZFRYHU
VWDLQOHVVVWHHO,3

&DEOHJODQG
 SODVWLF0[
º    IRUFDEOHš

DSSUR[  
DSSUR[
 
0

1HFNWXEHOHQJWK0XSWRVHDOLQJIDFH

Process connections

7
DSHUHGQLSSOHZLWKVORWWHGQXW &ODPSFRQQHFWLRQ 6SKHULFDOZHOGLQJW\SHVOHHYH
',1RUDVHSWLF WR',1 6SKHUH[PP [LQ 
WR',1 RU,6275,FODPS
'1WR ರWRರ '1WR
ರWRರ
0

0

0

GLDP'
8

8

8

GLDP) GLDP) GLDP)

3URWHFWLYHWXEHEDVHGRQ',1
9DULYHQWFRQQHFWLRQ 1HXPD%LR&RQWURO
'IRU9DULYHQWKRXVLQJ'1DQGರ 6L]H'GLDP
'IRU9DULYHQWKRXVLQJ'1WR'1 6L]H'GLDP )RUP 6LPLOLDUWRIRUP
DQGರWRರ 6L]H'GLDP WDSHUHGWLS

GLDP)
0

0

8

GLDP' GLDP'
8
8
8

8





GLDP) GLDP) º ) º )

Connection heads and process connections, dimensions in mm (inches)

Siemens FI 01 · 2008 3/97


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Resistance thermometers for food, pharmaceuticals and biotechnology
Resistance thermometers
for installation in pipelines and tanks

■ Schematics

PT100 two-wire PT100 three-wire PT100 four-wire 2xPT100 two-wire 2xPT100 three-wire

1 2 6 2 1 1 2 4 3 1 2 3 4 1 5 2 4 6 3

yellow
yellow

yellow

yellow

yellow
black
black
3 red
red

red
red

red
red
Connection diagram

3/98 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Resistance thermometers for food, pharmaceuticals and biotechnology
Resistance thermometers with clamp-on system

■ Overview Response time / accuracy


Application example
Measured medium: Steam
T1: Conventional welding-type protective tube form 2 acc. to
DIN 43772 (pipe 9 x 1mm (0.35 to 0.04 inch))
T2: Pipe collar for tube OD 25.4 mm (1.0 inch) acc. to BS 4825
with measuring insert (fast response)
3
130 °C (266 °F)
130 (266)
Temp. T1 129,5 °C (265.1 °F)
°C (°F)
100 (212)

T2

Result
The pipe collar provides almost
40 (104) identical results compared
to an inline temperature

0 (32)
0 30 60 90 120 Time
[sec]

The temperature sensor with pipe collar is used for temperature Results with steam as measured medium
monitoring, particularly for sterile applications in the food and
pharmaceutical industries.
Temperature measurement is possible following simple, fast as-
sembly on any existing pipeline system. It is not necessary to in-
T2
terrupt the pipeline or to carry out welding. Assembly on the
pipeline is simply carried out by adapting the two halves of the
collar.
Medium T3
The device can be optionally supplied with an integral transmit-
ter.

■ Function T1

Features
• Temperature measurement free of dead space
• Can be retrofitted onto existing pipelines Schematic representation

• Matches all common nominal pipe diameters The result for the actual temperature of the medium T3 depends
on several parameters:
• Optimum temperature measurement using metallic pressing
system • Heat transfer of temperature sensor/pipe
• Insulating collar made of temperature-resistant plastic, easy to • Conductivity of medium
clean • Flow velocity
• Measuring range -20 to +200 °C (-4 to +392 °F) • Pipe wall thickness
• Replaceable Pt100 measuring insert; thus no changes in mea- • Pipe material
suring setup
• Ambient temperature
• Calibration possible
Installation and operating instructions
• Transmitter can be integrated (4 to 20 mA or PROFIBUS PA)
Assembly of the two halves of the pipe collar to the pipe is car-
Mode of operation ried out simply and rapidly by connection of the two M6 hexagon
The temperature measurement is carried out using a Pt100 mea- screws. Any positioning is possible, but the Pt100 measurement
suring element. An integral metallic pressing system guarantees should preferably be made "underneath" the pipeline to achieve
optimum heat transfer between the pipeline and Pt100. The an optimum result.
measuring insert is protected by an insulating sleeve made of Following mounting on the pipeline, the complete Pt100 housing
temperature-resistant plastic. unit can be dismounted on site by loosening a clamping screw.
The complete Pt100 unit can be easily dismounted at any time It is not necessary to disconnect the cable during the usual cal-
for calibration; it is not necessary to disconnect the cables from ibration procedures.
the sensor.
The temperature sensor with pipe collar is available with all stan-
dard DIN connection heads.

Siemens FI 01 · 2008 3/99


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Resistance thermometers for food, pharmaceuticals and biotechnology
Resistance thermometers with clamp-on system

■ Technical specifications Selection and ordering data Order No. Order code

Design with connection head Measuring insert in connection Pipe collar Pt100 thermometer 7 MC 8 0 1 5 -
head with neck tube as one unit Stainless steel pressing system 777 0 0 - 7A7 0 777
for insertion into pipe collar.
All standard DIN connection Connection head
heads are available, see Selec- • Form B, cast light alloy, screw cover, 1
tion and Ordering data. IP54, cable gland

3 Pipe collar Temperature-resistant plastic for • Form B, plastic, screw cover, 2


range: IP54, cable gland
• Form BUZ, cast light alloy, hinged cover, 3
• -20 ... +160 °C (-4 ... +320 °F), IP65, cable gland
material PVDF
• Form BUZH, cast light alloy, high hinged 4
• -20 ... +200 °C (-4 ... +392 °F), cover, IP 65, cable gland
material PEEK • Form B, stainless steel, standard, IP67, 5
Neck tube Material: Stainless steel 1.4571 cable gland
• without, with M12 plug (without mating 6
Measuring insert The replaceable measuring insert connector)
is designed as a fast-response • Special version 9 H1Y
jacket element of 3 mm diameter. (add Order code and plain text)
It guarantees optimum heat trans-
fer from the jacket pipe to the Collar material
Pt100 measuring element (see • PVDF, -20 ... +160 °C (-4 ... +320 °F) B
"Response time / accuracy" under • PEEK, -20 ... +200 °C (-4 ... +392 °F) C
"Function" for reference measure-
ment with measuring insert ). Nominal diameter for pipes acc. to
Commercially available measur- DIN 11850 Series 2
ing inserts can also be used as Nom. diam. Outer pipe diameter
alternatives. • DN 10 13 mm (0.51 inch) A
Measuring resistor Pt100 acc. to DIN EN 60751 • DN 15 19 mm (0.75 inch) B
Class A in three-wire design, • DN 20 23 mm (0.91 inch) C
ceramic
• DN 25 29 mm (1.14 inch) D
Integration of transmitter Suitable Pt100 transmitters for • DN 32 35 mm (1.38 inch) E
head mounting can be fitted in • DN 40 41 mm (1.61 inch) F
the connection head, see Selec-
• DN 50 53 mm (2.09 inch) G
tion and Ordering data
• DN 65 70 mm (2.76 inch) H
Nominal pipe diameters Matches all standard nominal • DN 80 85 mm (3.35 inch) J
pipe diameters acc. to DIN 11850
series 2. • DN 100 104 mm (4.09 inch) K
Adaptation of the pipe collar is • Special version: Nominal diameter Z K1Y
possible for pipes acc. to (add Order code and plain text)
DIN 2463, ISO 1127 series 1, 2 Neck tube length
and 3 and pipes to BS 4825 Part
1 O.D.Tubing. • 60 mm (with plug version 0 mm) 1
The design of the pipe collar • Special version 9 N1Y
takes into account the usual toler- (add Order code and plain text)
ances for pipe outer diameters Sensor
acc. to EN ISO 1127 D3 and D4,
and acc. to ASTM-A213/A450, Ceramic type, fast response
ASTM-A249/A450 and ASTM- • 1 x Pt100, class A, three-wire F
A269. • 2 x Pt100, class A, three-wire G
Deviating tolerances must be • Special version Z Q1Y
specified. (add Order code and plain text)
See Selection and Ordering data
for dimensions. Further designs Order code
Add "-Z" to Order No. and
Pressing system Guarantees optimum heat trans- Order code.
fer between outer wall of pipe and
PT100, material: Stainless steel • Specify special version in plain text Y01
Weights Design with light alloy connection • TAG plate made of stainless steel Y15
head: Form B specify TAG No. in plain text
• Test report (at 0, 50 and 100%) Y33
• DN 10 Approx. 0.4 kg (0.88 lb) specify measuring range in plain text
• DN 15 Approx. 0.5 kg (1.10 lb) To order a temperature transmitter installed in the connection head,
• DN 20 Approx. 0.6 kg (1.32 lb) see "Temperature transmitters for installation in the connection
head" (Page 3/64).
• DN 25 Approx. 0.7 kg (1.54 lb)
• DN 32 Approx. 0.8 kg (1.76 lb)
• DN 40 Approx. 0.8 kg (1.76 lb)
• DN 50 Approx. 0.9 kg (1.98 lb)
• DN 65 Approx. 0.9 kg (1.98 lb)
• DN 80 Approx. 1.0 kg (2.20 lb)
• DN 100 Approx. 1.5 kg (3.31 lb)
Response time / accuracy see "Response time / accuracy"
under "Function" for reference
measurement with mesauring
insert

3/100 Siemens FI 01 · 2008


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Resistance thermometers for food, pharmaceuticals and biotechnology
Resistance thermometers with clamp-on system

■ Dimensional drawings

Connection heads

&RQQHFWLRQKHDGIRUP% &RQQHFWLRQKHDGIRUP%8=+
DOXPLQLXP DOXPLQLXP

DSSUR[  
0[ 0[
DSSUR[  
  
DSSUR[

  
DSSUR[
5RXQGFRQQHFWRUZLWK
VFUHZW\SHFDS0[
  

DSSUR[  
  
DSSUR[

DSSUR[


)RUP%VWDQGDUGVFUHZFRYHU
VWDLQOHVVVWHHO,3

&DEOHJODQG
 SODVWLF0[
º    IRUFDEOHš

DSSUR[  
  
DSSUR[
0

  
DSSUR[

Pipe collar Pressure system

0RXQWLQJVFUHZV
&ODPSLQJVFUHZ RIFROODU 3WPHDVXUHPHQW

3WWKUHHZLUH
ഋ; 3LSH2' ZLWKRSWLPL]HGKHDW
      WUDQVIHUš 
3LSH2'

     


ഋ;

     


      3LSH
     
  
   
      3UHVVXUH
         V\VWHP
3WPHDVXUHPHQW      
9LEUDWRQSURWHFWLRQ
0HGLXP

Connection heads/pipe collar/pressing system, dimensions in mm (inches)

Siemens FI 01 · 2008 3/101


© Siemens AG 2007

SITRANS T measuring instruments for temperature


Resistance thermometers for food, pharmaceuticals and biotechnology
Resistance thermometers with clamp-on system

■ Schematics

Pt100 three-wire 2x Pt100 three-wire

yellow
black
black
3
red
red

red
red

Connection diagram connection head

Pt100 three-wire 2 x Pt100 three-wire

4 3 5
6 4
7 3
J

1 2
1 2

J J

Connection diagram round connector M12, 4-/8-pin

3/102 Siemens FI 01 · 2008

You might also like