Professional Documents
Culture Documents
Operator’s Manual
Version 4.1
227 498 02 GA(e) Revision A
Contents
-0366
indicating its compliance with the provisions of the Council
Directive 93/42/EEC about medical devices and fulfills the
essential requirements of Annex I of this directive.
Revision History
This document is subject to the Marquette Hellige change order system. The revision code, a letter that follows the document
part number, changes with every update of the manual.
1. Introduction
9. Technical Specifications
Appendix
Index
General Information
* This manual is an integral part of the instrument and * The warranty does not cover damages resulting from the
describes its normal use. It should always be kept close to use of accessories and consumables from other manufac-
the equipment. Observance of the manual is a prerequi- turers.
site for proper instrument performance and correct
operation and ensures patient and operator safety. * Marquette Hellige is responsible for the effects on safety,
reliability, and performance of the equipment, only
* The symbol m denotes: Refer to Operator’s Manual!
– if assembly operations, extensions, readjustments,
It serves as an indicator for important facts to be noted
modifications, or repairs are carried out by
when operating the instrument.
Marquette Hellige or by persons authorized by
Marquette Hellige.
* Information which refers only to certain versions of the
instrument is accompanied by the catalog number(s) of – if the instrument is used in accordance with the
the instrument(s) concerned. The catalog number is instructions for use.
given on the instrument nameplate.
* All publications are in conformity with the instrument
* For detailed safety information, please refer to section specifications and IEC publications on safety of electro-
1.3 “For your safety”. medical equipment valid at printing date. All rights are
reserved for instruments, circuits, techniques, and names
* Patient safety, specified measuring accuracy, and interfer- appearing in the manual.
ence-free operation can be guaranteed only if original
Marquette Hellige devices are interconnected (e.g. basic * On request Marquette Hellige will provide a service
units and plug-in modules). manual.
* Only use accessories which are listed in this manual and * The Marquette Hellige quality management system
which have been tested in combination with the device complies with the DIN EN ISO 9001 and EN 46001
(e.g. patient cables, electrodes, transducers, sensors, Standards.
consumables, etc.). If you use accessories or consumables
from other manufacturers, Marquette Hellige does not
guarantee safe operation or functioning of the device.
1. Introduction to CardioSys
CardioSys is a PC-based system for ECG recording, spiro- 3-Lead Exercise ECG (ERG1)
metry tests and ambulatory blood-pressure examinations. The – 3-lead recording and display
following options are available to enhance the program func- – control of all Marquette Hellige ergometers
tions: – creation of custom protocols and phases (depending on
ergometer model)
Basic Software Package (RESB)
– 12-lead ECG acquisition (resting ECG, STAT ECG) Upgrade for 12-Lead Exercise ECG (ERG2)
– recording of spirometry tests – optional recording of 3, 6 or 12 ECG leads
– downloading of data from TONOPORT IV/IVa (am- – 12-lead ST measurement
bulatory blood pressure) – automatic or manual determination of the J+X point
– downloading of data from electrocardiographs EK 53, – presentation of the ST medians
EK 56, EK 512, CardioSmart – presentation of the resting ECG complex and of the
– data management for all operating modes current ST beat for direct comparison
– BDT interface – arrhythmia analysis and presentation during the test
Monitoring of Remote Test Stations (ERGM) Cardio Mailbox (CBOX) requires hardlock
– during the exercise test data are continuously sent via the – transmission of multiple procedures via a modem to
network CardioSys/CardioSoft
– the following data can be viewed at any of the work- – storage of multiple procedures on storage medium
stations in the network: – reception of multiple procedures via a modem (from
• station name CardioSys/CardioSoft or CardioSmart)
• 6-lead ECG, leads selectable
• information on the exercise test and on the ergometer
• HR and blood pressure
• ST measurement results (resting beat, current beat)
F1 F2 F3 F4 F5 F6 F7 F8 F9 F10 2
3 11
4
5
12
6
7 13
14
8
15
9 16
10
Figure 1-1. CardioSys system cart
1 Monitor 9 Shelf with power connector for defibrillator
2 Supports for electrode application system or 10 Conductive castors (front castors locking)
respiration flow sensor 11 Monitor retaining strip
3 Monitor support 12 Connection panel (under lid)
4 Tray for PC keyboard 13 Lid lock
5 Tray for function keyboard and mouse 14 Power switch (removes power from the system)
6 Tray (e.g. for printer) 15 Power input with instrument fuses
7 Accessory drawer 16 Potential equalization pin
8 Power button, floppy disk drive and drive bays
(streamer, MO drive)
The safety information given below is divided into the catego- ☞ CardioSys is a class I system.
ries “Danger”, “Warning” and “Caution”.
Before connecting the device to the power line,
indicates an imminently hazardous situation Caution
check that the voltage and frequency ratings of
Danger
which, if not avoided, will result in death or the power line are the same as those indicated
serious injury. on the system’s nameplate. If this is not the
case, do not connect the system to the power
indicates a potentially hazardous situation line.
Warning
which, if not avoided, could result in death or
serious injury. Do not use extension cables with multiple
Warning
power outlets to connect the system to the
indicates a potentially hazardous situation power line.
Caution
which, if not avoided, may result in minor or
moderate injury and/or damage to the equip- Before putting the system into operation visu-
ment. Danger
ally inspect all connecting cables for signs of
damage. Damaged cables and connectors must
☞ provides application tips or other useful informa- be replaced immediately.
tion to assure that you get the most from your
equipment. When disconnecting the system from the power
Danger
line, remove the plug from the wall outlet first.
Then you may disconnect the power cord from
the device. If you do not observe this sequence,
there is a risk of coming into contact with line
voltage by inserting metal objects, such as the
pins of electrode leads, into the sockets of the
power cord by mistake.
CardioSys can be operated on an uninter- Set up the device in a location which affords
Caution
ruptible power supply (UPS). The UPS should Caution
sufficient ventilation. The ventilation openings
not be disconnected from the power line for of the device must not be obstructed. The
prolonged periods of time to prevent deep ambient conditions specified in the “Technical
discharging of the battery. For this reason, Specifications” section must be ensured at all
switch the CardioSys system off and on with the times.
system’s own power switch and do not use the
switch of the UPS or the main power switch of ☞ The system should be set up to provide the opera-
the office. tor an unobstructed view of the control panel.
As a general rule, the UPS must first be con-
nected to the power line and switched on; then Devices intended for emergency application
you may switch on CardioSys. Danger
must not be stored or transported at tempera-
Do not switch on the CardioSys system during tures which cause moisture condensation at the
a power outage. application site. Wait until all moisture con-
Only the CardioSys system may be connected to densation has evaporated before using the
the UPS. device.
The UPS must be installed within the medi-
cally used room, but outside the patient envi- If the system is used as an emergency device, a
ronment. Warning
second ECG recorder must be available.
Observe the information given in the UPS user
manual. The device is not intended for intracardiac
Warning
application.
The power socket of the CardioSys system cart
Warning
is reserved for a defibrillator for recharging of The device is not suitable for operation in areas
the defibrillator battery (max. power 30 VA). Warning
where an explosion hazard may occur. Explo-
sion hazards may be caused by the use of flam-
All devices of a system must be connected to the mable anesthetics, skin-cleansing agents and
Warning
same power supply circuit. Devices which are disinfectants. Great care must be exercised
not connected to the same circuit must be when the monitor is used in an oxygen-enriched
electrically isolated when operated (electrically atmosphere. The atmosphere is considered to be
isolated RS232 interface). oxygen-enriched when the room air contains
more than 25% of oxygen or nitrous oxide.
The patient signal input is labelled with the If the system cart is installed in the patient envi-
Warning
symbol, indicating that it is protected Warning
ronment, only the special printer modified by
against damage resulting from defibrillation Marquette Hellige and equipped with an addi-
and electrocautery. Nevertheless extreme care tional protective earth connection (potential
must be taken when devices which are directly equalization) may be operated. Observe the as-
connected to the patient remain applied during sembly instructions supplied with the printers.
defibrillation. The distance between ECG elec-
trodes and defibrillation paddles should be at For stress-test examinations, a defibrillator
least 6 in./15 cm. If in doubt, disconnect the Danger
and a pacemaker, regularly checked for proper
patient cable from the device while applying the functioning, must be readily accessible.
defibrillation pulse.
Extreme care must be exercised when using
Magnetic and electrical fields are capable of Warning
medical electrical equipment. Many parts of
Warning
interfering with the proper performance of the the man/machine circuit are conductive, such
device. For this reason do not operate the as the patient, connectors, electrodes, trans-
system in the vicinity of equipment emitting ducers. It is very important that these conduc-
strong electromagnetic radiation (X-ray equip- tive parts do not come into contact with other
ment, diathermy systems or MRI devices). grounded, conductive parts when connected to
the isolated patient input of the device. Such
Only use the original Marquette Hellige pa- contact would bridge the patient’s isolation
Warning
tient cable. Do not connect other signal sources and cancel the protection provided by the
to the cable. isolated input. In particular, there must be no
contact of the neutral electrode and ground.
With all peripheral devices connected,
Warning
CardioSys represents a medical system. It
fulfills the requirements of the system standard
IEC 60601-1-1. The entire system may be
operated in the patient environment.
Devices may only be interconnected with each ☞ Medical technical equipment such as the
Warning
other or to parts of systems when it has been CardioSys system must only be used by persons
made certain that there is no danger to the who have sufficient knowledge, who have received
patient, the operators, or the environment as a adequate training in the use of such equipment
result. In those instances where there is any and who are capable of applying it properly.
element of doubt concerning the safety of
connected devices, the user must contact the Before using the system, the operator is obliged
manufacturers concerned or other informed Warning
to verify that it is in correct working order and
experts as to whether there is any possible operating condition.
danger to the patient, the operator, or the
environment as a result of the proposed combi- The operator must be familiar with the opera-
nation of devices. Standards IEC 60601-1-1/ Caution
tion of the system.
EN 60601-1-1 must be complied with in all
cases. Verify the performance of the system once a
Caution
month.
Liquids must not be allowed to enter the de-
Danger
vices. Devices into which liquids have pen- Dispose of the packaging material, observing
etrated must be checked by a service technician Warning
the applicable waste control regulations and
before being used again. keeping it out of children’s reach.
Chemicals required for the application or At the end of its service life, CardioSys and its
Danger
maintenance of the devices must under all Caution
accessories must be disposed of in compliance
circumstances be stored, prepared and kept at with the applicable waste control regulations.
hand in the containers provided by the manu- If you have any questions in this matter, please
facturer. Failure to observe this instruction contact Marquette Hellige GmbH.
may result in severe consequences for the pa-
tient. Only use bicycle ergometers and treadmills that
Warning
fulfill the applicable IEC 60601-1 require-
CardioSys is not intended for use as a vital ments. CardioSys and the associated bicycle
Warning
signs physiological monitor. ergometer or treadmill must be connected to
the same power supply circuit via separate wall
outlets.
* Before using the system for the first time, please read the – Do not run more than one other Windows application
information given in section 1.3 “For your Safety”. at the same time as CardioSys.
The ECG program too runs a self test when a signal acquisi-
tion screen is called up. If you do not see any error messages,
the program and the computer are intact. Error messages may
also appear during program operation.
2 4 Top View
1
3
5
6 7 8 9 10
printer problem
interface
Chart-Paper Loading
Before loading chart paper, make sure that the * Pull the first sheet of the pad straight out over the paper
Caution
printer is switched off (button 7, Figure 1-2). guide, at the same time closing the paper tray door.
The indicators will go off after a few seconds.
* Turn on the printer.
* Pulling upward on the paper tray door latch (3, Figure
1-2), open the paper tray door. Do not attempt to operate the printer without
Caution
thermal paper.
* Remove the cardboard backing of the old pad. Do not pull on the sheets to remove them from
the printer.
* When inserting the new pad, be aware of the following After loading a new pad, press keys Ctrl +
points: + )9 ] to ensure that the next printout
starts at the top of the page.
– the last sheets of the paper pad have a red stripe, so
this side of the pad should be at the bottom of the
The thermal printer AR 200 does not require
tray
Caution
an additional protective earth connection
– the black squares should be facing down and lo- (device in compliance with IEC 60601).
cated on the side of the printer nearest the paper
tray door latch
V1
To select a function you place the mouse pointer on the
respective icon and press the left mouse button. You will see
V2
that the key on the screen is “pressed” at the same time.
1 2
3
4 5 6
V3 R L
aVR V4
N F
Pressing a mouse button is also called “clicking”. When, later
on in this manual, you are asked to “double-click” this means
aVL V5
Lead Sequence that you should press the left mouse button twice in rapid
STANDARD
aVF V6 Spline
succession.
ON
Select Procedure To select an option, move the mouse pointer to the item and
Patient Baker, Dave, 41yrs
press the left mouse button. This selects the item, and the
Procedure all
window disappears.
Status all
Location
When a list of selections is too long to fit in the window, the
Date (from - to) -
window has a scroll bar at its right edge. Clicking the arrow at
Show new procedure list Display
the upper end of the scroll bar moves the list down one by
12/04/1996 03:38:17pm 0 Exercise Test Confirm
12/04/1996 03:29:24pm x0 Exercise Test one (1, Figure 2-3), clicking the lower arrow moves the list up
12/04/1996
12/04/1996
03:10:13pm
03:00:11pm
0
*0
Exercise
Exercise
Test
Test 1 Send
(2, Figure 2-3).
11/20/1996 10:53:39am 0 Exercise Test Combined Report…
4
3 Delete
Cancel A so-called scroll box between the two arrows allows you to
2 Help
move quickly through the list by clicking the scroll box,
Figure 2-3. Excerpt from the procedure list holding the mouse button depressed, and dragging the box up
1 moves the list down (line by line) or down (3, Figure 2-3).
2 moves the list up (line by line)
3 allows you to scroll to a general location in the list
4 clears the window without saving entries
Select Procedure The scroll box also indicates which section of the list is cur-
Patient Baker, Dave, 41yrs
rently displayed. If the box is just below the upper arrow you
Procedure all
see the top portion of the list, if the box is in the middle of
Status all
the scroll bar, the middle portion of the list is on display, etc.
Location
When you see an horizontal scroll bar below a picture, the
Date (from - to) -
two arrows and the scroll box allow you to shift the picture
Show new procedure list Display
left or right.
12/04/1996 03:38:17pm 0 Exercise Test Confirm
12/04/1996 03:29:24pm x0 Exercise Test
12/04/1996
12/04/1996
03:10:13pm
03:00:11pm
0
*0
Exercise
Exercise
Test
Test 1 Send
Some screens which offer various selections can be cleared by
11/20/1996 10:53:39am 0 Exercise Test Combined Report…
4
3 Delete two methods: either with OK or with Cancel
Figure 2-3. Excerpt from the procedure list When you select OK the program accepts the selec-
1 moves the list down (line by line) tions made, with Cancel , the selections will be ignored.
2 moves the list up (line by line)
3 allows you to scroll to a general location in the list If you wish to highlight text click the first letter of the text
4 clears the window without saving entries block, hold the left mouse button down and drag the pointer
to the last letter. Then release the mouse button. The marked
text is displayed in reverse video.
☞ You clear the HELP window by clicking “File” To obtain information on program operation, click the
and “Exit”. Help button or .
Use the keyboard to enter text and numbers. The following is * or to delete text (Del key) (3, Figure 2-4).
a description of functions which you do not know from a
conventional typewriter. If you are already familiar with a PC * The Home and End keys (2, Figure 2-4) move the
keyboard, you may skip this section. cursor to the first and last position in the text.
* The Backspace key (1, Figure 2-4) clears the characters * With Return (4) and on Windows screens with TAB (6)
from left to right, one by one. you move the cursor to the beginning of a new, blank
line.
* The cursor keys (5, Figure 2-4) are used to move the
cursor up, down, right and left to any desired position in
the text, for instance, to insert text (Ins key)
1 2
Q W E R T Y U I O P { } Page 7 8 9 +
Delete End
[ ] Down Home PgUp
Caps Lock A S D F G H J K L : @ ~ 4 5 6
; ' #
| Z X C V B N M < > ? 1 2 3
\ , . / PgDn
Enter
O ,
Ctrl Alt Alt Gr Ctrl
Ins Del
6 4 3 5 3
Figure 2-4. The keyboard
227 498 02-A CardioSys V 4.1 23
General Information on System Operation
Start program
marquette HELLIGE CardioSoft (Supervisor)
120/80 2.3 Starting the Program,
Explanation of the Function
Keys and Selecting the
Language
Opens a window for patient selection or for Calls up further examination procedures
admission of a new patient (section 3.1, 3.2)
Recording of a resting ECG (section 4.1) Opens a window for selection of stored procedures
Recording of an exercise ECG (section 4.2) Opens the printer menu and allows you to print
the displayed procedure
Recording of a STAT ECG (section 4.3) Allows you to compare the current procedure with
another one
Initiates a pulmonary function test (section 4.4) Allows you to delete the displayed procedure
120/80 Initiates the ambulatory BP program for program- Calls up system services (chapter 6)
ming of the BP monitor and data evaluation
(section 4.5)
Calls up the online help function
Initiates the late potential analysis program (sec-
tion 4.6)
Allows you to quit the program
Initiates a bronchodilation test (section 4.7)
…
Hellige
Patient Procedure System Services
SystemPPG
Services
HelligeMenu
Help Selecting the Language
(Unbenannt)
Procedure Archiving Setup
The program offers different languages for the user interfaces
and for the printed reports.
System Copy Patient Compression Load
Configuration Records
Cancel Help
The window showing the system services opens (Figure 2-6).
Select .
Figure 2-6. System services selection window
You will see Figure 2-7. (This paragraph describes language
selection only. Other settings to be made in this window are
System Configuration discussed in section 6.1)
General Modem MUSE Option Code Country Settings
Institution Name and Address
Name
Street
Database… Select the “Country Settings” tab (arrow, Figure 2-7).
City
Physician ID
User List…
Technician ID
Modify password…
Save
Load
Print
For Service only Help
OK Cancel
System Configuration
A screen similar to Figure 2-8 appears.
General Modem MUSE Option Code Country Settings
Language English Select your language from the “Language” list box and
Time Format 24-Hour Format
Date Format DD.MM.YYYY click OK to quit the menu.
Unit of Measure cm, kg
Speed Unit km/h
ST Level mV
Line Frequency 50 Hz
Lead Label IEC
Help
OK Cancel
Close
Use the spacebar to apply or remove the checkmark in
New Patient…
Last Name Master First Name Jim
Patient ID 8738999 Date of Birth 10/05/1949 MM/DD/YYYY
the Pacemaker checkbox.
V1
V2
1 2
3
4 5 6
V3 R L
N F
aVR V4
aVL V5
Lead Sequence
STANDARD
aVF V6 Spline
ON
Figure 3-4. Window for selection of stored tests Before you are able to edit test data, you will have to select a
1 Selected patient (only this patient’s records are test from the database. You can either display a list of all
displayed) stored tests or of the tests stored for a specific patient.
2 Choose “all” if you want to see all stored tests
3 Selection of the search filter (date range) Select the patient.
4 Selection of the search filter (tests recorded at a
special location) Click on .
5 Displays list when all search filters were set
6 Selected test A list appears showing all the examination procedures stored
7 Click arrow to scroll through list for the selected patient (Figure 3-4). If you wish to see all
8 Control buttons
stored procedures,
Displays the selected test
Displays a combined report
Click on (2, Figure 3-4) and select “all”.
Allows the selected test to be confirmed
Saves the test for MUSE (to the database or
storage medium, as configured)
Allows transmission of procedures
Clears a test
9 Selection of the search filter (confirmed, uncon-
firmed or all tests)
10 Selection of the search filter (only resting ECGs,
only STAT ECGs, only exercise tests)
1 2 4 6 8 10 You can set various filters to display only those tests that meet
3 5 7 9 specific criteria:
Select Test
Displaying Tests
Transmission via modem You can transmit one or several procedures via a modem or
Telephone Number
save them to a storage medium.
External User Name
1
minor left axis deviation
because QRS angle between -30° and 0° 5
down when clicking on a test.
Print
negative T-wave [lateral]
Cancel
Click on Combined Report .
Hellige For examinations made with an external program, you will see
Patient Untersuchung Hilfsfunktionen
External Procedure Hilfe
Summary
Holter ECG: 19.10.1995 • 13:00:30
a procedure summary (Figure 3-8), when you press the
(Unbenannt)
Interpretation (summary) OK button.
no pathological findings
Detailed Report Cancel Use only standard ASCII letters (A through Z) and numbers
(0 through 9) when entering information for the MUSE CV
Figure 3-8. Summary system.
Help
Cancel
When you click on one of the procedure selection buttons,
Close the patient selection screen opens, allowing you to retrieve a
New Patient…
Last Name Master First Name Jim
Patient ID 8738999 Date of Birth 10/05/1949 MM/DD/YYYY
patient from the database or to enter a new patient (sections
3.1 and 3.2). The advantage of this approach is that you can
Gender Male Height 176 in
Race Weight 66/0 lb Pacemaker Accept immediately select the examination procedure without having
to choose a patient first.
Figure 4.0-1. Patient selection screen
On the patient selection screen the bar cursor is always on the
patient selected last. So, if you wish to perform multiple tests
on the same patient, all you have to do is click on the
Select button or press .
If you prefer to run an unassigned procedure, click on
Unnamed .
Choose Close to display the title screen.
Test Information
A
4.1.1 The Basics
Library: Reason for Test ABE F Reason for Test A B C
Therapy outcome control
Routine test
Clarify chest discomfort
Confirm/Rule out CAD
» Choose the icon to activate the resting ECG mode.
Angina Pectoris
Post-MI
Preop. Diagnosis
After selection of a patient, the window for entry of the test
information will open (Figure 4.1-1), if you have chosen this
feature in the system configuration (refer to section 6.1 “Sys-
tem Configuration”). We recommend that you apply the
Blood Press.(mmHg) sys dia
electrodes to the patient as explained in section 4.1.2 prior to
Medication …
Technician: Help
correct incorrect
Figure 4.1-5. Arranging the patient cable
C1 C2
☞ Please observe the information given in sections
C3 white
C3 C5 7.1 “Artifact Due to Poor Electrode Application”,
C6
C4
Nst (Nap) 7.2 “ECGs of Pacemaker Patients” and 7.3 “ECG
C4 white (Nap)
C8 Nax
Recording During Defibrillation” in the appen-
Nax
C5 white C8
C7 dix.
C6 white
C6
L yellow
left arm
F green C5
left leg C1 C2 C3 C4
Nst Nap
Test Information
A
4.1.3 Entering the Test Information
Library: Reason for Test ABE F Reason for Test A B C
Therapy outcome control
Routine test
Clarify chest discomfort » 2 The test information screen offers menus and text boxes for
Confirm/Rule out CAD
Angina Pectoris
Post-MI
entry of all data relevant for the resting ECG procedure.
Preop. Diagnosis
1
On this screen you can
3
Blood Press.(mmHg)
4 – enter the reason for the test
sys dia
Medication …
Technician: Help
– select the medication
Figure 4.1-8. Test information screen
1 Reason for Test Library – enter comments
2 The patient’s reason for test
3 Blood pressure readings – register the physician’s and technician’s names.
4 Medications
5 Boxes for entry of comments and of the physi-
cian’s and technician’s names
☞ A Reason for Test can easily be copied by dou- If you wish to delete a copied reason or part of it, high-
ble-clicking.
light the words in question and click A .
Click ABE F .
Therapy outcome control Move the cursor to the desired position in the list and
Routine test
Clarify chest discomfort
Confirm/Rule out CAD
click the mouse button.
Angina Pectoris
Post-MI
Preop. Diagnosis
OK
Enter a standard statement and press on the key-
Cancel
board. The cursor moves to the beginning of the next
Help line.
Move the cursor to the sys box and click the left mouse
button.
Comment OK
Physician: Cancel
Click on to open the medication list box (Figure
Technician: Help 4.1-10).
☞ The physician is responsible for assigning drugs to If you wish to select a second medication or to assign a
a medication group. drug name to a medication group, click on … and
select the medication as described above.
unknown -->
steroids -->
respective text box, if you wish to enter a name. Then type the
beta-blockers II --> drug name.
beta-blockers III -->
antiarrhythmics Ia -->
antiarrhythmics Ib -->
Furthermore, you are free to arrange the medication groups
antiarrhythmics Ic --> any way you want. To do so
Help
antiarrhythmics III -->
OK
Ca antagonist,verapamil type -->
Cancel
Click on the at the end of the group name box.
Ca antagonist,nifedipine type -->
Medication This brings up a new window where you can choose one of
Medication Group Drug Name
none
the group names (Figure 4.1-12).
-->
none
unknown -->
digitalis
diuretics
-->
Click on the drug name box and enter the name
psychotherap. agents
-->
steroids
beta-blockers II -->
beta-blockers III
antiarrhythmics Ia --> Click on OK to clear the window.
antiarrhythmics Ib
antiarrhythmics Ic -->
antiarrhythmics III
Ca antagonist,verapamil type -->
Ca antagonist,nifedipine type Entering Comments
nitrates -->
ACE
alpha-blockers -->
cytostatics
--> If you wish to enter a comment on the test procedure, click in
Help
--> the “Comment” text box and key in the text.
OK
-->
--> Cancel
Physician’s/Technician’s Name
Figure 4.1-12. Editing the medication group
The names of the physician and ECG technician can either be
entered here (click in the text box and type the name), or they
can be selected from the list of the registered system users
(section 6.1 “System Configuration”).
Once you have entered all test information, you can either
☞ Pressing the key on the PC keyboard also
clears the window and saves the entries.
– clear the window with OK and start recording the
resting ECG or
– clear the window with Cancel (in this case the test
information will be ignored and are not assigned to the
test)
OK
Patient Procedure
Hellige
System Services Help 4.1.4 Recording A Resting ECG
Baker, Dave, 41yrs REST ECG/ Signal Acquisition
HR (BPM)
» Once you have entered the test information and cleared the
Test Information Freeze Traces Options
85
Setup X 40 HZ X 50 HZ 10 mm/mV 25 mm/s
SYS/DIA (mmHg) window with OK , the signal acquisition screen for the
2 V1
130/80 resting ECG (Figure 4.1-13) will appear. Take a few seconds
aVR 6 4 V4
N F
The resting ECG signal acquisition screen
aVL V5
Lead Sequence
Figure 4.1-13. Signal acquisition screen for the resting ECG – the current heart rate (2) and the blood pressure readings
1 Control buttons (3) (if entered manually, section 4.1.3)
2 Current heart rate
3 Blood pressure readings – the applied electrodes and the signal quality (4). Further-
4 Applied electrodes more an arrow (6) points to the leads which are impaired
green: properly applied by improper electrode application or which cannot be
yellow: not properly applied recorded due to disconnected electrodes (or due to a
white: site not used break in the patient cable).
red: electrode off or lead break
5 For selection of the lead sequence – the selected lead sequence (5)
6 Leads impaired by poor signal quality
7 Calibration pulse – the selected sensitivity (the 1-mV calibration pulse (7) is
valid for all channels).
☞ The applied electrodes are shown on the torso only This program allows you to select specific system settings
similar to a conventional ECG recorder. You can enable or
for the “Standard” and “CABRERA” lead se-
quences. In all other cases there is only a sche- disable muscle and AC line filters, select different sweep
matic representation of the electrodes. The refer- speeds and sensitivities, etc. Furthermore, you can set up the
ence electrode R (red) must be applied for the system to suit your personal needs and preferences. These
lead-check function to work properly. settings will be retained and are activated each time the system
is turned on (section 4.1.7 “Setup”).
1 2
3
4 5 6
The arrows at mm/mV select the gain (2.5, 5,
V3 R L
aVL V5 The arrows at ... mm/s select the sweep speed (25 or
aVF V6 Analyzing… 50 mm/s).
Figure 4.1-14. Resting ECG acquisition screen with control The entire ECG recording is controlled with the four buttons:
buttons and “Analyzing...” message
.
☞ Please remember that a filter may also suppress
Test Information Freeze Traces Options >>
Patient Procedure
Hellige
System Services Help
Recording and Evaluating the ECG
V1
most recent 10 seconds of the ECG signal will be analyzed.
Thus, before initiating the analysis, you can make sure that
V2
1 2
the quality of the ECG signal to be evaluated is adequate.
3
4 5 6
V3 R L
aVR V4
N F If you wish to modify the sweep speed or the filter settings
(muscle, AC line filter), click on Options >> to display the
aVL V5
second row of controls.
aVF Analyzing…
V6
Click Freeze Traces to freeze and release the signal traces. The
Figure 4.1-14. Resting ECG acquisition screen with control
freeze function does not affect channel 1.
buttons and “Analyzing...” message
At the end of the analysis you will automatically see the first
page of the procedure report showing the 10-second ECG
strips (Figure 4.1-15).
Patient Procedure
Hellige
System Services Help
ECG Trace Display
V2
– 2 x 6 leads (Figure 4.1-15)
1 2 3 4 5 – 1 x 6 leads (Figure 4.1-16)
V3
6 – 3 x 4 leads
aVR V4
aVR
aVL
aVF 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0
Patient Procedure
Hellige
System Services Help
Zooming and Measuring the ECG
aVL Amplitude 1.5 mV 40 mm/mV 200mm/s 40 Hz 50 Hz The zoom function allows you to display a portion of the
1 ECG at a larger scale.
2 3 4 5 6
Click .
7
701 ms
Move the cursor to the region of interest that you want
to zoom, and click. Your display looks similar to the one
0.1 0.2 0.3 0.4 0.5 0.6 0.7
shown on Figure 4.1-17.
The key functions described above are available for all docu-
ments. An explanation of the report pages follows on the next
few pages.
Vector angles
Baker, Dave, 41 yrs REST ECG / Test Summary The icon allows you to export the most important data
ABE F pertaining to the resting ECG procedure to the clipboard or
Reason for Test to any file. Data exported to the clipboard, for instance, can
Age 48 years
Export Data Sex
Race
Male
Caucasian
be inserted in a text file with the “Edit” —> “Paste” com-
Height 175 cm
Export data to clipboard
Measurements(Analysis Program)
Weight 80.0 kg mand.
Blood Pressure
supraventr premature (SVPB) HROK 84 BPM
because premature beat with same morphology
Export data to file QRS
Cancel 106 ms
left axis deviation QT 392 ms
because QRS angle between -90° and -30° QTC 330 ms Click .
PQ 180 ms
P 136 ms
RR 708 ms
PP 705 ms
Sokolov
QTD
3,50
20
mV
ms
Select either “Export data to clipboard” or “Export data
Medication -90
< QRS -25°
Comment
Physician Dr. Williams <T 50° 0 to file”.
Technician Dorothy <P 25° 90
For data exported to a file you will have to specify the target
directory and confirm it with OK .
Medians (dom.)
Hellige Medians (dominant/subdominant)
Patient Procedure System Services Help
Return… Reanalysis
Click Reanalysis to display a screen which allows you to
QRS 102 ms P 122 ms NB 12
QT 444 ms RR 700 ms QTD 46 ms
< QRS -25° aVR -90 aVL
0I
adapt the analysis criteria to your needs (Figure 4.1-23).
<T 50°
QTC(H) 522 ms PP 700 ms III 90 II
<P 25°
PQ 178 ms Sokolov 1.83 mV aVF
First select the leads from which you want to derive the
median complex for determination of the reference marks.
Only those beats will be superimposed.
Arrhythmia Review
Hellige Arrhythmia Review
Patient Procedure System Services Help
Figure 4.1-25. Arrhythmia review Furthermore you can determine the heart rate.
1 number of identified QRS complexes
2 HR Measuring the Heart Rate in a Normal ECG
3 displayed lead
4 mean RR interval Click on the QRS complex at the beginning of the
interval to be measured; the system automatically posi-
tions the cursor in the correct place.
Arrhythmia Review
Hellige Measuring the Heart Rate in the Presence of
Patient Procedure System Services Help
Atrial Fibrillation or Flutter
Baker, Dave, 41 yrs REST ECG / Arrhythmia Review
Cycles >> 1
Determine HR Number In this situation you will have to define a cardiac cycle your-
QRS Complexes 13 HR 85 BPM
III 10 mm/s 40 Hz 50 Hz self.
PVC
Click Cycles .
HR 85 BPM
Interpretation
Hellige Interpretation
Patient Procedure System Services Help
2 1 10 mm/mV 25 mm/s
Select the name from the list (open the list box with )
or enter a name via the keyboard.
3
Editing the Statement Library
Figure 4.1-28. Interpretation screen with confirmation window
and ECG The icon becomes active as soon as you display the
ABE F
1 Displays the list of registered physicians statement library (1, Figure 4.1-27).
2 Displays the other ECG leads
3 Removes the ECG leads Click on ABE F to open the window for entry of interpre-
tative statements.
Patient Procedure
Hellige
Hilfsfunktionen
Compare TestHilfe 4.1.5 Comparing Resting ECGs
Procedure 1
Baker, Dave, 4810.08.1996
J 11:32:23 Ruhe-EKG / Interpretation
Procedure 2 A B
A
C
You can compare the following details of two resting ECGs
10.08.1996 11:32:51 Resting ECG
Measurements(Analyseprogramm)
10.08.1996 11:12:45 RestingInterpretation
ECG (Analyseprogramm)
supraventr premature (SVPB)
10.08.1996 11:07:42
because premature beat with same morphology
10.08.1996 11:40:18
Resting ECG
Resting ECG
– the 10-second ECG
10.08.1996
left axis deviation 11:35:29 Resting ECG
because QRS-angel >= 90 Grad – the median complexes (dominant and sub-dominant)
because QRS-angel <= 90 Grad
1 2 3 4 Click .
10 mm/mV 25 mm/s 40 Hz 50 Hz A window will open listing all of the patient’s resting ECGs
(Figure 4.1-29).
Hellige For each ECG the screen shows 3 leads at a time (25 mm/s,
Patient Procedure System Services Help
10 mm/mV). ECG 1 is displayed in the top half, ECG 2
Baker, Dave, 41 yrs REST ECG / Compare / Zoom
below. Click buttons and to display the other ECG
Procedure 1 10.08.1996 · 11:32:51
Procedure 2 10.08.1996 · 11:32:51 leads.
40 Hz 50 Hz
1
12.10.1996
2 3 4 5
10 mm/mV 25 mm/s
Hellige On this screen you can check 3D graph (4, Figure 4.1-33) to
Patient Procedure System Services Help
view a three-dimensional presentation of the beat (Figure
Baker, Dave, 41yrs REST ECG / Comparison of Medians
4.1-34).
12.10.1996 - 11:08:14 3D graph
II 0,05 mV
0,10 mV/s
1 2 3 4
Hellige
Patient Procedure System Services Help
II
Sütterlin, Franz, 48 J
Print
Ruhe-EKG / Vergleichen / 10 Sekunden 01.08.94 · 11:40:23
4.1.6 Printing the Resting ECG
Pages to print
Aufnahme 1
1
10.08.1994 · 11:32:51
1 Procedure Date:
Aufnahme
X Standard Page2 10.08.1994
Dom.· 11:32:51
Median 10-sec. ECG segment Interpretation
Subdom. Median The scope of the printed report is determined on the print
ECG Format Display Format Leads/Page10 mm/mV 25 mm/s 35Hz 50 Hz
You can print a page with the dominant average medians and
with the subdominant medians, if these were found as well.
For more information, you may add either the measurements
or the ST profile.
Here you set up the system for printout of the full 10-second
ECG. You can choose to print either 6 or 12 leads/page and
you select the gain and paper speed. The resulting number of
pages to be printed is indicated.
In this window you set up the system for printout of the full
interpretation (interpretation as derived by the system and/or
physician’s interpretation).
The standard page and the interpretation page include the 10-
second ECG strip of the rhythm lead. However, you can
select another channel to be printed.
Printer
Barcode/Patient ID
Setup
Hellige
Setup
4.1.7 Setup
Speed (mm/s) 25
X 1-mV Cal Pulse
Gain (mm/mV) 10 Audible LEAD OFF Alarm From the setup menu you can select a number of system
Manual Override
Lead check noise level low settings to suit your personal needs and preferences. These
QTC Calculation Hegglin-Holzmann
settings will be retained. They are automatically activated each
40-Hz Filter
Lead Sequence STANDARD Modify 50-Hz Filter time the system is turned on.
Spline
Rhythm Lead II
Cancel
Use this screen to select your personal system settings.
Help
Figure 4.1-40. Setup menu Adjustments are made in three different ways:
Speed
selects the trace sweep speed
Gain
selects the signal size (amplitude)
Manual Override
When the function is enabled, the program will start the
signal acquisition, even if not all electrodes are applied.
Manual Override
Lead check noise level low
Lead Sequence
QTC Calculation Hegglin-Holzmann
40-Hz Filter In the lead sequence window you can assign the leads to the
Lead Sequence STANDARD Modify 50-Hz Filter
STANDARD Anti-Drift System
recording channels. Six lead sequences are predefined. Four of
Rhythm Lead II
CABRERA
Modify…
Hellige
Lead Sequence Modifying the Lead Sequence
HIGH Defaults
11 V5 V5' Cancel Cancel channels indicating the lead currently assigned to that
12 V6 V6' Help Help channel. To select another lead click to display all
available leads (Figure 4.1-43).
Figure 4.1-42. Modifying the lead sequence
Hellige
Lead Sequence Enter up to 4 characters in the “Designation” column to
HIGH Defaults
Lead25 Sequence
label the lead.
Speed (mm/s)
X 1-mV Signal
I
Gain (mm/mV) 10 Audible LEAD OFF Alarm
I
Lead check noise level
II
low
Once you have assigned a new lead to each of the channels,
III
aVR
aVL
you can enter a name for the lead sequence in the headline.
QTC Calculation Hegglin-Holzmann
aVF 40-Hz Filter
V1
Lead Sequence STANDARD Modify 50-Hz Filter
V2
Rhythm Lead
V3
II
Anti-Drift System Move the cursor to the text box and click.
V4
V5
Print after Analysis V6 Modify
Report D
A
Enter a name and select OK to clear the window.
J
OK OK
Rhythm Lead
rhythm or reference lead
Report Modify
Here you determine the scope of the report printed with the
“Print after Analysis” command. The special configuration
screen is described in detail in section 4.1.6 “Printing the
Resting ECG”.
40-Hz Filter
enables or disables muscle filter ( X muscle filter on)
50-Hz/60-Hz Filter
enables or disables AC filter ( X AC filter on)
Spline
Enables or disables the Cubic Spline algorithm (automatic
baseline correction, X Spline on); the algorithm causes a
signal delay of 4 seconds. On the signal acquisition and
control screen the algorithm can be disabled temporarily to
view the undelayed ECG signal.
Quit program
– print reports
Test Information
A
4.2.1 Some Basic Facts
Library: Reason for Test ABE F Reason for Test A B C
History of chest discomfort
Evaluate effect of medication »
Evaluate effect of CABG
Evaluate effect of PTCA ☞ Before recording your first exercise ECG, go to the
Evaluate functional capacity
Stress echo study
defaults screen and choose the basic system settings
Stress thallium study Protocol
Rule out/confirm CAD
Bruce (section 4.2.7 “Exercise Test Setup”). These set-
Post-MI study Target HR(%) 100 180/min
... tings will be retained and are automatically
activated each time the system is turned on.
Blood Press [mmHg] sys dia
Medication …
OK
Click on to start the program
Comment
Physician: Cancel
Technician: Help After you have selected the patient, the system displays the
Referring Physician: dialog window for entry of the test information (Figure 4.2-1)
Figure 4.2-1. Dialog window for entry of the test information (unless this option was deselected in the System configura-
tion). We recommend applying the electrodes to the patient,
following the instructions given in section 4.2.2, before
entering the test demographics.
Skin Preparation
Figure 4.2-2. Electrode application points * Carefully degrease the application points with a cotton
C1 in the 4th intercostal space at the right swab soaked in benzine.
sternal edge
C2 in the 4th intercostal space at the left sternal * Prep the application points with skin prepping paste
edge
C3 at the level of the 5th rib midway between (EPICONT). To do so, apply a small amount of paste to
C2 and C4 the tip of a cotton swab. Place the tip against the skin at
C4 in the 5th intercostal space on the left the electrode site and twirl the swab 4 to 5 times between
midclavicular line your thumb and index finger, lightly pressing on the swab.
C5 on the left anterior axillary line between C4
and C6
C6 on the mid-axillary line at the level of C4 * Connect the electrode leads to the electrodes.
C7 in the 5th intercostal space on the left
posterior axillary line * Remove the backing paper from the electrodes and press
C8 in the 5th intercostal space on the left
scapulary line the electrodes firmly into place.
R, L on the right and left shoulders
N, F above the right and left kidneys * Secure each electrode lead with surgical tape.
227 498 02-A CardioSys V 4.1 83
The Exercise ECG / Applying the Electrodes
C4 white
C5 white (I, II, III, aVR, aVL, aVF, V1 to V6)
C6 white C8
C7
L yellow
F green
on left shoulder
C6
Four limb and six chest electrodes are usually required for the
above left kidney
C5 acquisition of the standard leads. In stress testing, however,
C1 C2 C3 C4
we recommend attaching the limb electrodes on the shoulders
Figure 4.2-3. Connecting the patient cable (10-wire, standard
(where there are no muscles) or above the kidneys (Figure
ECG leads)
4.2-2).
☞ Please observe the information given in sections
7.1 “Artifact Due to Poor Electrode Application”, * Connect the patient cable as shown in Figure 4.2-3.
7.2 “ECGs of Pacemaker Patients” and 7.3 “ECG
Recording During Defibrillation” in the appen- * Secure the electrode distributor of the patient cable
dix. (chest strap or holder).
☞ The red R electrode is the reference electrode. * Connect the patient cable to the device (type CF signal
Placement and signal quality of all other elec- input: highly isolated, suitable for intracardiac applica-
trodes can be assessed on the signal acquisition tion, defibrillation-proof).
screen only when this electrode is applied.
Test Information
4.2.3 Entering the Test Information
Library: Reason for Test A
ABE F Reason for Test A B C
History of chest discomfort
Evaluate effect of medication » 2 The test information window allows you to make all entries
Evaluate effect of CABG
Evaluate effect of PTCA 1 3
Evaluate functional capacity and settings relevant to the test.
Stress echo study
Stress thallium study Protocol Bruce
Rule out/confirm CAD
Target HR(%) 100 180/min
Post-MI study
... 5 These are:
6 4
Blood Press [mmHg] sys dia – the reason(s) for the test
Medication …
OK
Comment
– the test protocol
Physician: 7 Cancel
Technician: Help
ABE F Modify Library: Reason for Test Copying Reasons for Test to the Patient Indication Window
History of chest discomfort
Evaluate effect of medication
Evaluate effect of CABG
Evaluate effect of PCTA
Highlight the statement by clicking.
Evaluate functional capacity
Stress echo study
Stress thallium study
Rule out/confirm CAD OK Click the button.
Post-Ml study
…
Cancel
…
…
Help If you wish to delete a copied statement or part of it,
highlight the words in question and click A .
Figure 4.2-5. Window for entry of standard reasons for test
The deleted words can be inserted in another position by
placing the cursor there and clicking B A C .
☞ A reason for test is easily copied from the library
to the right window by double-clicking!
You can also enter a reason for test directly by moving
the cursor to the desired position and typing the text.
Click ABE F .
Referring Physician:
Adjusting the Target HR
Figure 4.2-6. Protocol selection window
The target heart rate that the patient is expected to attain
during the test can either be adjusted manually, or the system
will calculate this value automatically (section 4.2.7 “Exercise-
Test Setup ”). Furthermore, you can change the calculated or
entered value (e.g. 90% of 180 = 162). This setting will be
retained.
Move the cursor to the “sys” box and click the left mouse
☞ Click on to display and increase the value, or
button.
click on to decrease the value.
Enter the value via the keyboard.
Medication …
☞ The physician is responsible for assigning drugs to a To select additional medications, to enter a medication
medication group. dose or to assign drug names to medication groups, click
… .
none –>
unknown –> A window opens, showing several medication group names
medication stopped –> (Figure 4.2-8). There is a drug name window for each of these
nitrates –>
beta-blockers –>
groups. Click on the respective text box, if you wish to enter a
diuretics –> name. Then type the drug name.
anti-hypertensives –>
digitalis –>
Ca antagonists –> Furthermore, you are free to arrange the medication groups
antiarrhythmics –> OK
any way you want. To do so
glycoside –> Cancel
–>
Medication This brings up a new window where you can choose one of
Medication Group Drug Name the group names (Figure 4.2-9).
none –>
none –>
unknown
unknown
medication stopped –> Click on the drug name box and enter the name
nitrates –>
nitrates
beta-blockers
diuretics
beta-blockers –>
anti-hypertensives
–>
Click on OK to clear the window.
diuretics
digitalis
Ca antagonists
anti-hypertensives –>
antiarrhythmics
digitalis –>
glycoside
Entering Comments
Ca antagonists –>
antiarrhythmics –> OK
–>
glycoside Cancel If you wish to enter a comment on the test procedure, click in
–>
the “Comment” text box and key in the text.
Figure 4.2-9. Editing the medication group
Physician’s/User’s Name
Once you have entered all test information, you can either
– clear the window with Cancel (in this case the test
demographics will be ignored and are not assigned to the
test) to display the preacquisition screen
I
Baker, Dave, 41 yrs. Exercise Test
4.2.4 Running an Exercise Test
W + HR[BPM]
00:421 WHO 4 Phase>>6 0 -
8 85
3 Hold5 Stage>> RPM 7
% of 179
+
00:42
2 #1 - RPP
The signal acquisition screen appears when you have entered
12
10 BP/Comment 15
14 Freeze ECG
Lead Check Interpretation SYS/DIA (mm/Hg)
13 ST/Medians 16 Trends 9 the test information and clear the dialog window with
11Print Event Mark ECG Strips
QRSL VPB/min 1
V1 17 OK . For bicycle ergometers the screen is similar to
V2 1 2 3 4 5s
Figure 4.2-10, for treadmills it is similar to Figure 4.2-11.
V6/Scan 19 18 Please take a few seconds to become familiar with all the
V3
details shown on this screen, before proceeding to start the
V4 1,88 mV/s 1,88
Rest
0,07 mV 0,07 test.
E J J+60 ms
20
V5
10 mm/mV 25 mm/s
V6
X ADS 21
I
Baker,Dave, 41 yrs. Exercise Test The Signal Acquisition Screen of an
+
00:421 P-Bruce 4 Phase>> 6 mph HR[BPM]
9 85
- Stop Treadmill Exercise Test
% 7 + 8
00:422 Supine 3 Hold5 Stage >> - Start Treadmill RPP
12
10 BP/Comment 14 Freeze ECG
Lead Check 15 Interpretation SYS/DIA (mm/Hg)
Print Rhythm
11 13 ST/Medians 16 Trends
Event Marker ECG Strips
The signal acquisition screen (Figure 4-11) of the exercise test
VPB VPB/min 1
V1 17 presents the following information:
1 2 3 4 5s
V2
– the current load and RPM or speed and grade (7)
19
V6/Scan 18
V3
– the current stage (3) and the time elapsed in that stage (2)
V4 1,88 mV/s 1,88
Rest
0,07 mV 0,07
E J J+60ms
J+60 ms
– the current phase (4) and the time elapsed in that phase (1)
V5 20
10 mm/mV 25 mm/s
V6
X ADS 21 – the heart rate, RPP and blood pressure (9)
Figure 4.2-11. Signal acquisition screen for a treadmill exercise – the results of the arrhythmia analysis (complex ectopy in
test red) and number of VPB/min (17)
1 Time elapsed in current phase – the ST analysis results compared with the resting beat
2 Time elapsed in current stage analysis results (18) (“Scan” indicates that the lead with
3 Stage number/name the most prominent ST depression is displayed (leads I,
4 Current phase and button to modify protocol aVR, aVL, V1 and V2 are not considered))
5 Buttons to hold the stage and to proceed to the next stage
6 Button to proceed to the next phase – various procedure parameters (21)
7 Current speed and grade, buttons to increase or decrease the
speed and the grade 16 Buttons to display the ST profiles and medians, the trends
8 Buttons to start and stop the treadmill (protocol continues) and the stored 12-lead ECG
9 Windows indicating the heart rate, rate-pressure product 17 Results of the arrhythmia analysis (complex ectopy in red)
(RPP) and the blood pressure and number of VPB/min
10 Button to abort the test 18 ST median beat window and ST analysis results (resting
11 Button to initiate a printout (laser printer or thermal-array beat analysis results are shown with a different color)
printer) 19 Button for selection of another lead
12 Button for entry of comments and blood pressure 20 Buttons for adjustment of the E, J, and J+X point (only
13 Button to mark an event episode (the episode is saved at the when “Manual” has been selected in the Defaults menu)
same time) 21 Procedure parameters and buttons to modify the signal
14 Button to verify the ECG signal quality (all leads) sensitivity and to temporarily disable the ADS (anti-drift
15 Button to freeze and release the signal traces system)
With these ergometers, the test follows the stored test phases.
This is to say that the initial load and load increase displayed
in the protocol have to be adjusted manually at the ergometer.
Manual modifications of the protocol (load increase or selec-
tion of the next test phase) must also be repeated at the ergo-
meter.
Then press the Phase >> button and ask the test subject to
start pedalling.
Window (2) in Figure 4.2-10/4.2-11 indicates the remaining During the test you can
time in the current stage, while window (1) shows the time
elapsed in that phase. advance to the next load stage with Stage >>
At the end of the load stage, the load increase displayed by the continue at the same load with Hold
Test Information Prot. 1 Phase >> Pedal! Stage >> Hold Phase >>
Ergometers without Digital Communication Interface but with Remote Start Capability (e.g., MEDITRONIC M 40)
As above, but there is no need to select the next phase manu- As trend values are saved at 30-second intervals, we recommend
ally at the ergometer, as this is done by the remote start signal. that you wait approx. 30 seconds before starting to pedal.
Then press the Phase >> button and ask the test subject to
On the setup screen (section 4.2.7) select the “intern.
start pedalling. If your ergometer sends a remote start
remote start” ergometer.
impulse when the patient starts pedalling, there is no
need to click the Phase >> button.
If you wish to obtain a recording at the end of each load
stage only, set the “ECG Rec. Inter. EI” to 00:00 min. Window (2) in Figure 4.2-10 indicates the remaining time in
the current stage, while window (1) shows the time elapsed in
For “Time in stage” (TS) enter a time which is consider- that phase.
ably longer than the corresponding value at the ergometer.
During the test you can
Select a protocol and adjust the corresponding initial modify the load in steps of 5 Watts with + 5 and
load at the ergometer. -5 ; do not forget to make the same adjustments at the
ergometer
Click the button. Wait for the first ST meas-
advance to the recovery phase with Phase >> (upon comple-
urement to be displayed (option) (18, Figure 4.2-10/
tion of the last stage, the program automatically activates
4.2-11).
the recovery phase)
Ergometers with Digital Communication Interface (ERGOCOMP EC 560/ECB 561 / ERGOLINE 900/900L, EC 1200 V3.52)
The EC 1200 will display the version number when turned With ERG 900, select the recovery phase “recovery ERG
on. The following explanations refer to version 3.52 or higher 900” (25 W, 1 minute, then 0 W) only if you expect to
(upgrade, if possible). For older versions, see section “M 700 do a low-impact test. The BP meter will then respond
and EC 1200”. more promptly to the load decrease.
With these ergometers the selected protocol fully controls the Select a protocol.
test procedure. The ergometer operating controls are inopera-
tive. Only blood pressure measurements can be taken manu- Click the button.
ally. Refer to Appendix 6 “Connecting External Devices”.
A blood pressure measurement is now taken automatically, if
EC 1200 configured. Wait until the readings are available or until the
Blood pressure intervals are valid only for external BP meters first ST measurement is displayed (option) (18, Figure 4.2-10).
(appendix, section 6). BP meters must be connected directly
when used with EC 1200 version 3.52 and higher. They are As trend values are saved at 30-second intervals, we recom-
controlled by the stage time. The BP intervals are ignored. mend that you wait approx. 30 seconds before starting to
pedal.
On the setup screen (section 4.2.7) select
“ERGA” for EC 560/ECB 561, * Ask the test subject to start pedalling.
“ERG 900” for ERGOLINE 900/900L and
“EC 1200 (V3.52)” for EC 1200 version 3.52 or higher.
In conjunction with M 700 ergometers too, the selected * modify the load at the ergometer, but only for your own
protocol fully controls the test procedure. During a test, load custom-programmed phases
changes are only possible at the ergometer and only for the
user-programmed phases. If you wish to create custom * advance to the recovery phase with Phase >>
protocols you will have to use the phases stored in the ergo-
meter. * terminate the test with .
Select a protocol. In the program, 10 protocols have been created from these
phases. You are free to modify these protocols as needed.
Click the button.
Phase No. Name on Ergometer Display
A blood pressure measurement is now taken automatically, if 8 OWN PROGRAM 1
configured. Wait until the readings are available or until the 9 OWN PROGRAM 2
first ST measurement is displayed (option) (18, Figure 4.2-10). 10 OWN PROGRAM 3
11 OWN PROGRAM 4
* Ask the test subject to start pedalling. 12 OWN PROGRAM 5
1 WHO
Window (2) in Figure 4.2-10 indicates the remaining time in 2 ICSPE
the current stage, while window (1) shows the time elapsed in 3 ST. FRANCE
that phase. 4 WHO NEW
5 HOLLMANN
6 REHA
7 not available for EC 1200
32 RECOVERY
98 CardioSys V 4.1 227 498 02-A
The Exercise ECG / The Elements of an Exercise Test
If you call up an unconfigured phase (see above), the ergome- Window (2) in Figure 4.2-10 indicates the remaining time in
ter runs idle. the current stage, while window (1) shows the time elapsed in
that phase.
On the setup screen (section 4.2.7) select the “ERGS”
ergometer. During the test you can
If you wish to obtain a recording at the end of each load modify the load at the ergometer
stage only, set the “ECG Rec. Inter. EI” to 00:00 min.
advance to the recovery phase with Phase >>
Select a protocol.
terminate the test with .
Click the button.
Treadmills
☞ The treadmill can be stopped at any time with the
emergency stop button.
From the setup menu (section 4.2.7), item “Ergom./
Treadmill” select:
– “Internal Treadmill” for treadmills without digital
☞ When using the treadmill model TM400E, it is interface
important to adjust the same unit of speed meas-
– “Series 2000” for treadmills 1800, 1900, T2000
urement at the ergometer and at the CardioSys
– “TM 400E” for Trackmaster treadmills TM 310,
system (km/h or mph).
TM 400E
Test Information Stage >> Start Treadmill Phase >> Stage >> Hold Phase >>
Then you can proceed to the next stage ( Stage >> button, 5,
☞ Pressing the + or – button cancels the automatic
Figure 4.2-11) to record the ECG of the patient standing
program sequence and activates the manual mode
upright.
(phase name “Manual”). The program maintains
the current stage for reasons of safety and the
The next stage is reserved for recording the ECG of the
Hold and Stage >> buttons are locked. The test can
hyperventilating patient.
only be controlled manually with the + and –
keys (speed, grade). When you activate + and –
Then the warm-up stage follows, in which the patient gets
again after a pause of approx. 15 seconds, the
used to the treadmill exercise.
program displays the next higher stage number.
With the Phase >> button you advance to the next
Proceed to the next stage and start the treadmill with the
test phase.
Start Treadmill button (8, Figure 4.2-11).
Name:
Date of Birth:
P-Bruce
Dave Baker
05/05/1955 (41 yrs.)
Male, Caucasian, 145.0 lb, 79 in
00:31 Total Time 00:31 HR [BPM] 60
Rhythm Report
Patient ID 12345
During the test you can:
Supine 00:31 0.0 mph 0.0% BP [mmHg]
I
(the reference numbers refer to Figures 4.2-10 and 4.2-11)
II
III
– increase or decrease the speed and grade (7) (if you
AVR
AVF
buttons are disabled and you have to control the rest of
V1
the test manually) or prevent an automatic increase or
V2
decrease with Hold (5)
V3
V6
– proceed to the next phase with Phase >> (6)
09/11/1996 09:51:09 10 mm/mV 25 mm/s
– freeze the ECG (15) (the screen display only, one real-
time lead continues to be displayed for monitoring)
BP/Comment
– display ST measurement results of the other acquired
Baker, Dave, 41yrs. Exercise Test
mph + HR [BPM] leads (19), “Scan” indicates that the lead with the most
01:02 Pretest Phase >> 0.0 - Stop Treadmill
85
01:02 Supine Hold Stage >>
%
0.0 +- Start Treadmill RPP significant ST depression is displayed automatically
BP/Comment Lead Check Freeze ECG Interpretation SYS/DIA (mm/Hg)
Figure 4.2-18. Dialog box for entry of comments and blood Click on BP/Comment to open the dialog box (Figure
pressure 4.2-18).
Name:
Date of Birth:
Dave Baker
05/05/1955 (41 yrs.)
Male, Caucasian, 145.0 lb, 79 in
Recall Report
Patient ID 12345 Recall prints the most recent 10 seconds of the ECG (3,
P-Bruce 01:17 Total Time 01:17 HR [BPM] 60
Supine
I
01:17 0.0 mph 0.0% BP [mmHg]
6, or 12 leads) (Figure 4.2-20).
II
III
12-Lead prints the 10-second ECG acquired after clicking
AVR
on the button in the preconfigured format (4x2.5s, 2x5s,
AVL
V1
V2
Click on to clear the lead check screen.
V3
V4
☞ Recall and 12-Lead reports are printed only when
V5 option “print” or “print & save” is selected for
V6
I aVR V1 V4
II aVL V2 V5
III aVF V3 V6
Name: Dave Baker 12-Lead Report Name: Dave Baker Comparative Medians Report
Date of Birth: 05/05/1955 (41 yrs.) Patient ID 12345 Date of Birth: 05/05/1955 (41 yrs.) Patient ID 12345
Male, Caucasian, 145.0 lb, 79 in Male, Caucasian, 145.0 lb, 79 in
P-Bruce 00:16 Total Time 00:16 HR [BPM] 60 P-Bruce 00:09 Total Time 02:33 HR [BPM] 60
Supine 00:16 0.0 mph 0.0% BP [mmHg] Stagel 00:09 1.7 mph 10.0% BP [mmHg]
Rest Current Beat Rest Current Beat
I V1 I I I V1 V1 V1
II V2 II II II V2 V2 V2
III V3
III III III V3 V3 V3
aVR V4
AVR AVR AVR V4 V4 V4
aVL V5
AVL AVL AVL V5 V5 V5
II
III
AVR
AVL
AVF
V1
V2
V3
V4
V5
V6
ST/Medians
– review the ST analysis results (medians and profile) of all
Baker, Dave, 41yrs Exercise Test
Print Rhythm Event Mark ST/Medians Trends ECG Strips button you can toggle to the ST histogram screen (clear
SPB VPB/min 3
display with ).
V1
V2 1 2 3 4 5s
Comp.Median aVR ST Profile V4* The Comp. Median or Linked Median button is displayed only when
I
V3
a thermal printer is connected to the system. The button
0.06 mV 0.06 mV 0.01 mV -0.06 mV
0.25 mV/s
V4 II -0.13 mV/s
aVL -0.12 mV/s
V2 V5* 1,88mV/s
-0.05
0,07
actually shown depends on the system setup (section 4.2.7
0.12 mV
“Exercise-Test Setup”).
V5III -0.02 mV 0.03 mV
0.32 mV/s
0.26 mV
-0.07 mV/s 0.03 mV/s
V3 V6* 2.28 mV/s
aVF
V6
X The Comp. Median button initiates a printout showing:
0.15 mV 0.07 mV -0.01 mV 0.00 mV
1.13 mV/s 1.30 mV/s 1.38 mV/s -0.12 mV/s – the resting beat for reference
Figure 4.2-25. ST analysis results – the current median complex
– a 2.5-second ECG segment
for all 12 leads.
Trends
HR/min
240–
BO/mmHg mph
–5 20–
VPB PACE
–40 –
0,4–
Level. mV I* Slope. mV/s
–
–4
The Linked Median button initiates a printout showing the
180– –4 15– –30 – –
–3
0,2–
–
–2
– current median complex and the corresponding RR interval
7 120–
60–
1 –2
–1
10–
5–
2 –20
–10
0–
–
-0,2–
–
3 –0
–
–-2
– linked by the computer to form a computer-synthesized
-0,4– –-4
0– –0 0– –0 – –
HR/min
240–
ART%
–100 5–
METS RPP
–200 –
0,4–
Ampl. mV aVR* Slope. mV/s
–
–4
rhythm strip for all 12 leads. The raw data of an additional
180– –75 4– –160 – –
0,2– –2
120– –50
3– –120 –
0–
–
–0
rhythm lead are also printed for reference (this will be the first
6
60– 4 –25
2–
1– 5 –80
–40
–
-0,2–
–
-0,4–
3 –
–-2
–
–-4 manually selected rhythm lead (item “Select Arrhythmia
0– –0 0– –0 – –
0 5 10 15 20 0 5 10 15 20 0 5 10 15 20
Lead” in the “Setup” menu)).
Figure 4.2-26. Graphic trends
1 Heart rate, blood pressure, speed If the system is not able to analyze the ST segment, it will
2 VPB, pacer pulses output the 12-lead report instead of the linked medians.
3 ST level/slope (slope and level of the 2 leads Click on to clear the ST screen.
with the most significant ST changes, all other
leads are accessible with button 6) – display trends (Figure 4.2-26) with the Trends
ECG Strips
– display segments of the stored 12 leads (Figure 4.2-27) with
Baker, Dave, 41yrs Exercise Test
mph + HR[BPM] ECG Strips
01:02 Pretest Phase >> - Stop Treadmill.
+
85
01:02 Stage1 Hold Stage >> %
- Start Treadmill RPP
BP/Comment Lead Check Freeze ECG Interpretation SYS/DIA (mm/Hg)
The recordings are annotated with time and load. The first
Print Rhythm Event Mark ST/Medians Trends ECG Strips 120/80
00:32 P-Bruce : Supine VPB.VBIG VPB/min 3 recording represents the 0 Watt level. With you can
V1
display the other recordings. The button allows you to
V2 1 2 3 4 5s
access the current segment directly. Click on to clear the
V6/scan
V3 window.
V4 1,88 mV/s 1,88
Rest
0,07 mV 0,07
I J J+60ms
J+60 ms
Click on Interpretation
V5
10 mm/mV 25 mm/s
V6
X ADS
– to view the “Statement Library” page (refer to Figure
Figure 4.2-27. Stored 12 leads 4.2-38).
METS values are interpolated between two stages. The full – Test Summary (Figure 4.2-28)
METS value of a test stage is attained after the test subject has – Tabular Summary (Figure 4.2-31)
exercised for a period of 2 minutes in that stage. – Trend Reports (Figure 4.2-32)
– Sample Cardiac Cycles (Figure 4.2-34)
– QRS Morphologies (Figure 4.2-35)
– ECG strips (Figure 4.2-36)
– Arrhythmia Review (Figure 4.2-37)
– Interpretation (Figure 4.2-38)
– Full Disclosure ECG (Figure 4.2-39)
Cancel
Modify Help
Figure 4.2-30. Documents of the exercise test report When you export the data to the clipboard, you can insert
them in a text file for further editing (commands “Edit” —>
“Paste”).
If you choose the second option you will have to specify the
directory and the file name. Confirm the export command
with OK .
You can either scroll through the entire final report, or access
specific pages directly.
Tabular Summary
The Tabular Summary
Patient Procedure System Services Help
Phase: Stage Time Speed Grade METS HR Sys/Dia RPP V4* VPB
shows:
P-Bruce Supine 1.30 0.0 0.0 0.0 85 0.07
Standing 1.30 0.0 0.0 1.0 85 0.07
Hypervent 1.30 0.0 0.0 1.0 85 0.07
Manual
Warm-Up
Stage1
1.30
1.30
0.0
1.9
0.0
10.0
1.0
1.7
85
85
0.07
0.07 Bicycle Ergometer Treadmill
– phase – phase/stage
– time – time
– load – speed
– RPM – grade
– heart rate – METS
– blood pressure – heart rate
– RPP – blood pressure
Figure 4.2-31. Summary Table for treadmill test – ST level – RPP
– number of VPBs – ST level
☞ Only the stage format of the Tabular Summary is – comments, if entered – number of VPBs
transmitted to MUSE. – comments, if entered
Baker,Dave, 41 yrs Exercise Test/Trend Reports Figure 4.2-32 shows page 1 of the trend reports. Here you see
heart rate, blood pressure and speed (1)
HR[BPM] BP/mmHg Speedmph VPS PACE Level mV V3 Slope mV/s
240– –5 20– –40 –
0,4–
–
–4
VPBs and pacer pulses (2)
180– –4 15– –30 – –
0,2– –2
120–
1 –3
–2
10–
2 –20
–
0–
–
–
–0
–
ST amplitude and slope (of the two leads with the most
60– 5– –10 -0,2– –-2
0–
–1
–0 0– –0
–
-0,4–
–
–
–-4
–
significant ST changes) (3)
HR[BPM]
240–
ART%
–100 5–
Mets RPP
–200 –
0,4–
Level[mV] I
3
Slope[mV/s]
–
–4 heart rate and artifacts (4)
180– –75 4– –160 – –
0,2– –2
120– –50
3– –120 –
0–
–
–0 METS and RPP (5)
60– 4 –25
2–
1–
5 –80
–40
–
-0,2–
–
-0,4–
–
–-2
–
–-4
0– –0 0– –0 – –
0 5 10 15 20 0 5 10 15 20 0 5 10 15 20 The program will show you the exact values for any point on
Phase Grade % Total Exercise Time 6:49 min
Time mIn Level mV Duration P-Bruce 1:25 min the trend graph. When, in the selected trend period, the
HR /min Slope mV/s Duration Manual 5:24 min
Speed km/h
system could measure neither ST level nor slope, both values
are set to zero.
Figure 4.2-32. Trend Reports, page 1
– the slope
0
0 60 120 180 HR[EFM] 240
– the number of points (N)
Figure 4.2-33. ST/HR slope according to R.J. LINDEN
– the correlation quotient r
I
METS 1.00 METS 4.90
6 leads are displayed per screen. The beats on the far left are
0.00 mV 0.06
0.31 mV/s 0.00 the QRS complexes at rest, those following to the right are the
II
0.00 0.05
0.03 0.00 median beats for each load stage (Figure 4.2-34). The pro-
III
0.00 0.01
gram can display up to 20 median beats. When this number is
0.23 0.00
aVR reached, the program proceeds as follows: The beat on the far
0.00 -0.06
-0.21 -0.00
left is retained, while the latest beat overwrites the oldest beat
aVL
0.00 0.02
0.00 0.00 in the section on the right. The last beat before a phase
aVF
0.00
0.13
0.03
0.00
change is not overwritten.
With the check box next to “ST values” you can select
X or deselect display of the ST values.
QRS Morphologies
Hellige QRS Morphologies
Patient Procedure System Services Help
Figure 4.2-35. QRS Morphologies Use and to display the other leads of the same
templates.
Arrhythmia Review
Hellige Arrhythmia Review
Patient Procedure System Services Help
Figure 4.2-37. Arrhythmia Review Use and to display the next two event epi-
sodes.
Arrhythmic Events (acc. to priority)
Interpretation
HELLIGE Interpretation
Patient Procedure System Services Help
Measurements Summary
procedure you can choose both from the system-generated
Statement Library Interpretation
HR/RPP/METS:
measurements and from the statement library which you
2
HR at rest: 85 BPM
Max. HR: 86 PBM
3 4 entered yourself.
1 (48 % of Target HR: 179 BPM)
Max. METS: 1.00
BP:
BP at rest: 120/90 mmHg You can copy statements from both libraries by clicking on
Max. BP: 120/80 mmHg
Events: »
QRS 689, VBIG, VPB: 38 the respective lines and on the copy icon .
CPLT: 8
Significant ST Change:
V6: 0.00 mV
Select “System Evaluation“ or “Statement Library”.
☞ If your PC is equipped with a sound card and a A test is considered confirmed, when a physician has reviewed
microphone, the program supports voice recording.
the test interpretation and “signed” it electronically with his/
her name. Tests that have been reviewed and confirmed by a
physician are marked with x.
V2
1 2 3 4 5 Click (Figure 4.2-39).
V3
Pages to Print: 1
Print
Procedure Date:
4.2.5 Printing an Exercise-Test
1 Report
X Standard Page Trend Reports Sample Card.Cycles
6 lines
Phase: Stage: Time: Speed Grade METS HR Sys/Dia RPP V5* VPB
Test Information Medication
min mph % BPM mmHg 1/100 mV
Date/Time 08/05/1996 - 10:14:22 beta-blockers
Total Exercise Time
Duration P-Bruce
6:18 min
2:04 min
P-Bruce Rest ECG 2:04
>>> P-Bruce 00:20
>>> P-BRUCE Rest ECG
0.0 0.0 1.0 85 120/80 0.00 3
Standard Page
Comment
Interpretation
HR/RPP/METS:
HR at rest: 85 BPM
Max. HR: 86 BPM
(66% of Target HR: 130 BPM)
The standard page (Figure 4.2-42) shows, at left, the patient
Max. RPP: 114
BP:
Max. METS: 1.00
Name:
Date of Birth:
Dave Baker
05/05/1955 (41 yrs.)
Male, Caucasian, 145.0 lb, 79 in
EXERCISE TEST / Trend Report
Patient ID 1726712676
Page 2
The ST beats page shows the resting ST beats in the first
240
180
HR(BPM);BP(mmHg) Speed mph
5
4
20
7,5
VPB PACE
40
30
5
4
METS RPP 1/100
200
160
240
180
HR(BPM) ART(%)
100
75
column, the ST beats with the max. ST depression in column
3 3 120
120
60
2
10
5
20
10
2 80
120
60
50
25
no. 2 and in the following columns the ST beats at max. load
1 1 40
0
0
Level (mV)
5 10
I
15
Slope (mV/s)
20
0 0
0
Level (mV)
5 10
AVR
15 20
Slope (mV/s)
0 0
0
Level (mV)
5 10
V1
15
Slope (mV/s)
20
0 0
0
Level (mV)
5 10
V4*
15 20
Slope (mV/s)
0
and for the recovery phase.
0,4 4 0,4 4 0,4 4 0,4 4
0 0 0 0 0 0 0 0
-0,2
-0,4
-2
-4
-0,2
-0,4
-2
-4
-0,2
-0,4
-2
-4
-0,2
-0,4
-2
-4
Trend Reports
0 5 10 15 20 0 5 10 15 20 0 5 10 15 20 0 5 10 15 20
Level (mV) II Slope (mV/s) Level (mV) AVL Slope (mV/s) Level (mV) V2 Slope (mV/s) Level (mV) V5* Slope (mV/s)
0 0 0 0 0 0 0 0
-0,4
0 5 10 15 20
-4 -0,4
0 5 10 15 20
-4 -0,4
0 5 10 15 20
-4 -0,4
0 5 10 15 20
-4
overview (Figure 4.2-43), all trend graphs are printed on a
Level (mV) III Slope (mV/s) Level (mV) AVF Slope (mV/s) Level (mV) V3 Slope (mV/s) Level (mV) V6 Slope (mV/s)
0,4
0,2
4
2
0,4
0,2
4
2
0,4
0,2
4
2
0,4
0,2
4
2
single page. When you select “Detail”, you obtain several
0
-0,2
0
-2
0
-0,2
0
-2
0
-0,2
0
-2
0
-0,2
0
0 5
08/05/1996 10:14:22
10 15 20 0 5 10 15 20 0
Technician: Dorothy, RN
10 15 20
to be printed changes accordingly. (Click and select for-
mat). Check the ST/HR box to obtain a printout of the ST/
Figure 4.2-43. Trend report – overview
HR slope page.
180
4
7,5 30 You can choose the documents (Figures 4.2-45 trough
3
60 5 10
1
0
0 5 10 15 20
0 0
0 5 10 15 20
0
you wish to have a 12 or 6-channel recording (window be-
low).
METS RPP 1/100 HR(BPM) ART (%)
5 200 240 100
4 160
3 120
120 50
1
80
40
60 25 printer (printer port), the number of copies, whether the
0
0 5 10 15 20
0 0
0 5 10 15 20
0
patient ID no. will be annotated in the form of a barcode on
the report and whether an ECG grid is to be printed along
08/05/1996 10:14:22 Physician: Dr. Williams Technician: Dorothy, RN
with the ECG traces. If you deselect the grid, the printout will
Figure 4.2-44. Trend report – detailed format take less time.
Cover Page
Name: Dave Baker EXERCISE TEST / Arrhythmia Review Name: Dave Baker EXERCISE TEST / QRS Morphologies
Date of Birth: 05/05/1955 (41 yrs.) Patient ID 1726712676 Date of Birth: 05/05/1955 (41 yrs.) Patient ID 1726712676
Male, Caucasian, 145.0 lb, 79 in Page 1 Male, Caucasian, 145.0 lb, 79 in Page 3
QRS: 494 28
I I VPB: 0 28
I
II II
III III
AVR AVR II
AVL AVL
III
AVF AVF
V2 V2 AVR
V3 V3
V4 V4 AVL
V5 V5
V6 V6 AVF
Figure 4.2-45. Arrhythmia review report page Figure 4.2-47. QRS morphologies report page
Name: Dave Baker EXERCISE TEST / ECG Strips Name: Dave Baker EXERCISE TEST / Sample Cardiac Cycles
Date of Birth: 05/05/1955 (41 yrs.) Patient ID 1726712676 Date of Birth: 05/05/1955 (41 yrs.) Patient ID 1726712676
Male, Caucasian, 145.0 lb, 79 in Page 2 Male, Caucasian, 145.0 lb, 79 in Page 3
2:00 P-Bruce Rest-ECG 0.0 km/h 85 BPM P-Bruce Bruce Bruce R-Bruce
I AVR V1 V4 Rest ECG Stage1 Stage2 Recovery
2:00 3:20 4:49 5:44
85 BPM 85 BPM 89 BPM 85 BPM
135/80 mmHg
METS 1.00 METS 1.00 METS 1.00 METS 1.00
I
0,05 mV 0,05 mV 0,05 mV 0,05 mV
0,25 mV/s 0,10 mV/s 0,10 mV/s 0,10 mV/s
II
III
0,01 mV 0,01 mV 0,01 mV 0,01 mV
0,26 mV/s 0,31 mV/s 0,31 mV/s 0,31 mV/s
AVR
-0,05 mV -0,05 mV -0,05 mV -0,05 mV
-0,16 mV/s -0,05 mV/s -0,05 mV/s -0,05 mV/s
III AVF V3 V6
AVL
0,02 mV 0,02 mV 0,02 mV 0,02 mV
0,03 mV/s -0,10 mV/s -0,10 mV/s -0,10 mV/s
AVF
0,03 mV 0,03 mV 0,03 mV 0,03 mV
0,11 mV/s 0,06 mV/s 0,06 mV/s 0,06 mV/s
08/05/1996 10:14:22 10 mm/mV 25 mm/s 20 Hz Spline Physician: Dr. Williams Technician: Dorothy, RN 08/05/1996 10:14:22 10 mm/mV 25 mm/s 20 Hz Spline Physician: Dr. Williams Technician: Dorothy, RN
Figure 4.2-46. ECG Strips report page Figure 4.2-48. Sample cardiac cycles report page
Cancel
Patient Procedure System Services
Hellige
Help 4.2.6 Verifying the Exercise-Test
1
Baker, Dave, 41 yrs. Exercise Test
Settings
P-Bruce 50 + 25 03:00 #5 03:00 Modify Bruce
2 1 2
3
3 R
4 5 6
L The control screen (Figure 4.2-49) shows:
4 N F
– the selected protocol (1)
aVR
poor signal quality
L
– the applied electrodes and the signal quality (3) (for the
aVL
5 10 mm/mV 25 mm/s standard and CABRERA leads on the body as shown on the
aVF Spline 50 Hz 6 picture at left, for all other leads, there is a schematic
Figure 4.2-49. Exercise-test control screen representation of the electrodes). Furthermore an arrow
1 Selected protocol (4) points to the leads which are impaired by improper
2 Control buttons electrode application or which cannot be recorded due to
3 Applied electrodes disconnected electrodes (or due to a break in the patient
green: properly applied cable)
yellow: poor signal – the selected gain (the 1-mV calibration pulse (5) is valid
red: electrode off or lead break for all channels)
white: electrode not used
4 Leads impaired by poor signal quality – the most important system settings (6).
5 Calibration pulse
6 System settings Other operations possible on the control screen are
Gain (mm/mV) 10
Before running the first stress test, select several system set-
Writer no
Lead Check noise level low Auto. Report Format 12-Lead Modify tings on the setup screen. These settings are saved and loaded
none Filter (Hz) Manual Recording Print&Save automatically on power up. The setup menu is displayed from
Printmanuell
V2 6
50 Hz Arrhythmia Lead Select Save
Scan V6 6
the exercise-test control screen (section 4.2.6).
X Spline
Manual 60
J+X-Point (ms) F
Save full disclosure ECG
Auto
Select Setup to display page 1 of the menu
Blood Pressure Limit [sys] 300 Manual 180 6
OK Cancel Help
Adjustments can be made in four different ways:
Figure 4.2-50. Exercise-Test Setup, page 1
1 There is an arrow at the end of a box. Clicking this
arrow will open a window which offers various options.
Setup/ Exercise Test
Select the desired setting by clicking the mouse button.
Page 1 Page 2
BP Monitor BP Meter 1
C R-Bruce 3 There is a check box in front of the function. Click on
Ergometer Model Internal Treadmill
Port COM2 the box to enable or disable the function ( X function
Port COM1
OK Cancel Help
4 Click on the Modify… box and proceed as described for
Figure 4.2-51. Exercise-Test Setup, page 2 each individual case.
J+X Point
The measurement point J+X for ST analysis may be deter-
mined by the program or manually by the user.With
“manual”, you can make a preliminary adjustment of the J+X
point on this screen and then shift the E, J and J+X points
during the exercise test. In the automatic mode, the J+X point
is calculated in dependence of the heart rate (for HR of
60 BPM, X=60 ms, X becomes smaller with increasing heart
rates). For calcuation of the ST/HR slope, the “manual”
method must be selected and the reference point should be set
at J + 60 ms.
OK Cancel Help Select the sequence that you wish to modify (any se-
Figure 4.2-52. Lead sequence selection window quence, except 1 and 2).
Select Modify… .
Modify… Setup/ Exercise Test
Modify Lead Sequence
STANDARD
Page 1
Report Label
Page 2
Screen Sequence A window opens showing all channels (Figure 4.2-53). There
Sequence
X 1-mV Cal Pulse I I Monitored LeadsI 12
is one window for each of the channels indicating the lead
Audible LEAD OFF AlarmII II Displayed LeadsII 6
Modify…
Audible HR Alarm III III Lead Sequence III STANDARD currently assigned to that channel.
Diagram: HR in % of max.
aVRHR aVR aVR
Protocol Bruce Modify…
X Remote View aVL aVL aVL
Phases A P-Bruce
X RPP Readout aVF aVF aVF
B Bruce
Start clock at pretest V1 V1 V1
C R-Bruce
V2 V2 V2
BP Monitor BP Meter 1
V3 V3 Ergometer Model V3Internal Treadmill
Port COM2
V4 V4 Port V4 COM1
Thermal Writer V5 V5 V5
none Library: Comments Modify…
(AR200P) local V6 V6 V6
Setup/
Modify Lead Exercise Test
Sequence Click the box or the arrow to display all available
NEHB
Page 1 Report
Sequence
Label Page 2 Screen Sequence leads (Figure 4.2-54).
X 1-mV Cal Pulse I I
Monitored Leads I 12
II
Audible LEAD OFF Alarm III II
Displayed Leads II 6
Modify…
Audible HR Alarm
aVR
aVL
III
Lead Sequence III STANDARD Select the new lead to be recorded in the respective
Diagram: HR in % of max. HRaVF aVR aVR
V1 Protocol Bruce Modify… channel.
X Remote View V2 aVL aVL
V3 Phases A P-Bruce
X RPP Readout aVF aVF
V4
V5 BD Bruce
Start clock at pretest
V6
D
CA R-Bruce
Enter up to 4 characters in the “Label” column to label
D A
BP Monitor BP Meter 1 A
J
J
Ergometer J
Model Internal Treadmill the lead.
Port COM2
I I
Port I COM1
II II II
Thermal Writer
(AR200P) local none
III
Library:
III Comments III Modify… Once you have assigned a lead to each of the channels, you
OK Cancel
OK
Help
Cancel Help can enter a name for the lead sequence in the headline.
Protocol
A protocol comprises up to three phases. These are the pre-
test phase, the exercise phase and the recovery phase. The
program comes with several preconfigured protocols for the
various patient categories (for instance: BRUCE protocol). In
the “Protocol” window you select the default protocol, specify
a new protocol or modify existing protocols.
Modify…
Modify Protocol Click Modify… if you wish to modify an existing proto-
Protocol Bruce col or to define a new one. This will open the entry
Phases A Pretest Ramping window (Figure 4.2-55).
B Bruce Ramping
C Recovery Ramping Click in the box for phase A, then select one of the
Existing Phases P-Bruce New phases in the window by clicking.
Bruce
R-Bruce
Modify…
P-ModBruce
ModBruce
Delete
Select phases B and C in the same way.
OK Cancel Help
Modify… Modify Phase Position the cursor on the parameter to be changed and
Phase P-Bruce Type: TM Square click.
Stage Speed Grade ST Time in Stage ECG Rec.Int BP Interval
Supine 0.0 0.0 99:00 00:00 00:00
Enter the new value. It is possible to leave a box blank or
1 2 3 to remove a parameter.
Stage Supine
Modify the other parameters in the same way.
Speed SP 0.0 mph
Grade GR 0.0 %
Time in Stage TS 99:00 min OK If you wish to create a new phase, enter a (new) name
ECG Rec.Inter. EI 00:00 min Cancel under which the phase will be saved.
BP Interval PI 00:00 min Help
Stage Speed Grade ST Time in Stage ECG Rec.Int BP Interval Click the stage to be modified (reverse video, 2, Figure
Supine 0.0 0.0 99:00 00:00 00:00
4.2-56). The corresponding values will be displayed at 3
(Figure 4.2-56).
1 2 3
Stage Supine Click the value to be modified and type the new value
Speed SP 0.0 mph (you are free to leave entry boxes open and to delete
Grade GR 0.0 % values).
Time in Stage TS 99:00 min OK
ECG Rec.Inter. EI 00:00 min Cancel Having made all necessary changes, click OK to
BP Interval PI 00:00 min Help clear the window or click the next stage to be modified.
Ergometer/Treadmill
Allows you to select the ergometer/treadmill model used for
the test.
Port
Allows you to select the communications port to which the
ergometer/treadmill is connected (default: COM 2).
Comments Library
Here you may enter or modify comments to be added to the
procedure “during the test” (section 4.2.4 “Running an
Exerise Test”).
Modify…
Setup/ Exercise Test Select Modify… : A window opens showing the library of
Page 1 Page 2 existing comments (Figure 4.2-57).
X 1-mV Cal Pulse Monitored
Library:Leads
Comments 12
Audible LEAD OFF Alarm Displayed Leads 6
Audible HR Alarm
During Test
Lead Sequence STANDARD
Modify…
Move the cursor to the item you wish to change and click.
Chest discomfort
Diagram: HR in % of max. HR Dyspnea
Leg discomfort
Protocol Bruce Modify…
X Remote View Fatigue
X RPP Readout
Nausea Phases A P-Bruce Enter the modification.
Syncope
Contrast agent administered B Bruce
Start clock at pretest
C R-Bruce
BP Monitor BP Meter 1 Select OK to clear the window.
Ergometer Model Internal Treadmill
Port COM2
Port COM1
OK Cancel Help
01:25 Stage 2
1 2 3 •••
RPP 110
SYS/DIA (mm/Hg)
4 5 6 Return The program allows you to monitor exercise tests at remote
I Station: stations from your PC. Click to call up the monitoring
Stress Lab 1
II screen (Figure 4.2-58). The function is available only when
you have entered a station name in the system setup menu at
III
0.15 mV/s
0.01 mV V4 both stations (section 6).
aVR
0.15 mV/s
7
-0.01 mV.63 mV Values
at rest Click on to display the procedure selection menu.
aVL
10 mm/mV
aVR
25 mm/s
Select the remote exercise-test monitoring screen (button
Figure 4.2-58. Monitoring of a remote exercise-test station ).
1 Durations, phase name, stage number
2 Speed and grade In the selection window, click the station you wish to
3 Heart rate, blood pressure, RPP monitor. Then click OK to clear the window.
4 Key to call up further ECG leads
5 Buttons to call up further remote stress-test The remote station screen is made up in the same way as the
stations exercise-test acquisition screen: durations, phase name and
6 Button to terminate the monitoring function stage number (1), speed and grade (2), heart rate, blood
7 ST measurement results (resting beat, current pressure (3), buttons to display more ECG leads (4), buttons
beat) for selection of other remote exercise-test stations (5) and the
Return ... button which is used to quit the monitoring
function (6).
☞ Conditions for monitoring of remote exercise-test ☞ When data are lost while transferred via the net-
workstations are: work (slow PC), you will be alarmed by a red bar.
– a station name was entered
– the monitoring function is enabled
– the correct communication protocol (netbios.exe
or TCP/IP) is installed and selected (System
Services –> System Configuration –> Remote
Test Screen).
load/ next
start exercise
new patient speed + grade + stage * Press new patient or retrieve patient (1) and enter the
load/ hold
retrieve patient start recovery speed -
grade -
stage patient data.
prepare final report start STOP treadmill
system setup
BP
comments
lead
check
freeze
ECG
comparative
medians
* Start the exercise phase with (2) start exercise and ask the
close window
ST ST
trends
writer
recall
patient to start pedalling or start the treadmill with
medians profilr STOP
start (5).
Power
* Freeze and release the ECG signal traces with freeze ECG
load/ next
start exercise
new patient speed + grade + stage (7).
load/ hold
retrieve patient start recovery speed -
grade -
stage
prepare final report start STOP treadmill * Display the ST measurement results with ST medians,
ST profile (8). Press close window (11) to clear the
system defaults
stored
ECG <<
stored
ECG
stored
ECG >> 12 lead window from the screen.
rhythm
BP lead freeze comparative
system setup comments check ECG medians
close window
ST ST
trends
writer
recall
* Display graphic trends with trends (8). Press close win-
medians profilr STOP
Admit new patient or select patient from list Admit new patient or select patient from list
Apply electrodes and check signal quality Apply electrodes and check signal quality
Test Information Enter the test information Test Information Enter the test information
Adjust initial load at the ergometer as required by Adjust initial load and load increase/stage
the selected protocol as required by the protocol selected at the ergo-
meter
Click button and wait for ST values to
be displayed (option) Click button and wait for ST values to
be displayed (option)
Phase >> Click button and start pedalling
Phase >> Click button and start pedalling
Repeat all load changes of the selected program at
the ergometer Interactive load changes (+5W) and adjustment of
load for recovery phase
Stage >> Selects the next stage
+5 -5 Modifies the load in 5-Watt increments
Hold Maintains the stage
Phase >> Selects the recovery phase
+5 -5 Modifies the load in 5-Watt increments
Quits program
Phase >> Selects the recovery phase
Quits program
Admit new patient or select patient from list Admit new patient or select patient from list
Apply electrodes and check signal quality Apply electrodes and check signal quality
Test Information Enter the test information Test Information Enter the test information
Click button and wait for ST values to Click button and wait for ST values to
be displayed (option) be displayed (option)
Hold Maintains the load stage Load changes can be made at the ergometer only
(with M700 only for the custom-programmed
Stage >> Switches to the next load stage (only with protocols)
ERGOLINE 900/900L)
Phase >> Selects the recovery phase
+5 -5 Modifies the load in 5-Watt increments
Quits program
Phase >> Selects the recovery phase
Quits program
Treadmill
Start program
Phase >> Starts exercise test
Enter new patient data or select patient from list Hold Maintains the stage
Apply electrodes and check signal quality + – Modifies the speed and grade (note: interrupts
automatic program control)
Test Information Enter the test information
Stop Treadmill Stops treadmill (without interrupting the test)
Select test protocol
Start Treadmill Restarts treadmill
OK Record resting ECG (supine patient)
Phase >> Selects the recovery phase
Stage >> Record resting ECG (standing patient)
Quits program
Stage >> Record ECG of hyperventilating patient
correct incorrect
Figure 4.3-4. Arranging the patient cable
C1 C2
☞ Please observe the information given in sections
C3 white
C3 C5
C6
7.1 “”Artifact Due to Poor Electrode Applica-
C4
Nst (Nap) tion”, 7.2 “ECGs of Pacemaker Patients” and 7.3
C4 white (Nap)
C8 Nax “ECG Recording During Defibrillation” in the
Nax
C5 white C8
C7 appendix.
C6 white
C6
L yellow
left arm
F green C5
left leg C1 C2 C3 C4
Nst Nap
V1
00:00 tion screen.
1 3 V2
1 2
3
4 5 6
The signal acquisition screen (Figure 4.3-7) shows:
V3 R L
aVR
6 4 V4
N F
– the current heart rate (2) and the remaining storage time
Poor Signal Quality
V5
L for the full-disclosure ECG (3)
aVL
Lead Sequence
aVF
7 5 V6
STANDARD
Spline
ON – the applied electrodes and the signal quality (4). Further-
Figure 4.3-7. Signal acquisition screen for the STAT ECG more an arrow (6) points to the leads which are impaired
1 Control buttons by improper electrode application or which cannot be
2 Current heart rate recorded due to disconnected electrodes (or due to a
3 Remaining full-disclosure ECG storage time break in the patient cable).
4 Applied electrodes
green: properly applied – the selected lead sequence (5) (click on to select
yellow: not properly applied another lead sequence)
white: site not used
red: electrode off or lead break – the selected gain (the 1-mV calibration pulse (7) is valid
5 For selection of the lead sequence for all channels).
6 Leads impaired by poor signal quality
7 Calibration pulse This program allows you to select specific system settings
similar to a conventional ECG recorder. You can enable or
disable muscle and AC line filters, select different sweep
☞ The applied electrodes are shown on the torso
speeds and sensitivities, etc. Furthermore, you can set up the
only for the ”Standard” and ”CABRERA” lead
system to suit your personal needs and preferences. These
sequences. In all other cases there is only a
settings will be retained and are activated each time the system
schematic representation of the electrodes. The
is turned on (section 4.3.4 “Setup”). From the setup menu
reference electrode R (red) must be applied for
you also choose the format for the in-test reports.
the lead-check function to work properly.
The arrows at ... mm/mV select the gain (2.5, 5, 10, 20,
40 mm/mV).
☞
Test Information Freeze Traces Options >>
The test information screen appears automatically
when you check the “Patient Selection Prior to
The following pages explain the operating steps required to
STAT ECG” box in the system setup menu. The
record a STAT ECG.
same screen can also be displayed by clicking on
the Test Information button.
At the end of the session the program asks you whether or not
you wish to save the results (select a patient).
Setup
Setup
Hellige
4.3.4 Setup
Speed (mm/s) 25 X 1-mV Cal Pulse
Number of Leads 12 OK
1 There is an arrow at the end of a box .
Lead Sequence STANDARD Modify …
Cancel Clicking this arrow will open a window which offers
Rhythm Lead II
Top Window
Gain
Audible LEAD OFF Alarm
selects the signal size (amplitude)
audible signal when an electrode is disconnected: yes or no
Lead Check noise level ( X yes)
This is a measure of the quality of the acquired signal (signal
to noise ratio). When there is too much signal noise at the Manual Override
electrode, the electrode is displayed yellow on the signal When the function is enabled, the program will start the
acquisition screen. The setting that you select here (low, signal acquisition, even if not all electrodes are applied.
middle, high) changes the threshold as of which the electrode
is displayed yellow (high sensitivity to noise = low threshold).
Rhythm Lead
STANDARD
II X Spline ments made in this field only refer to the 10-s ECG analysis
CABRERA
X Print after Analysis HIGH
NEHBModify
(e. g. lead sequence).
Report
Number of Leads 12 OK
QTC Calculation
Lead Sequence STANDARD Modify …
Cancel
Equation for calculation of the QT interval: HEGGLIN-
Rhythm Lead II
In the lead sequence window you can assign the leads to the
recording channels. Six lead sequences are predefined. Four of
these (sequences 3 to 6) can be modified.
Modify…
ModifyHellige
Lead Sequence Modifying the Lead Sequence
Setup
Rhythm Lead 6 II
aVF aVF
X Spline quences 3 to 6 only).
X Print after Analysis 7 V1'
Report Modify V1'
8
X Save full disclosure ECG
Terminate after (min)
V2'
20
V2'
Click Modify… to open a window showing all channels
9 V3' V3'
Number of Leads 12
OK OK (Figure 4.3-11). There is a window for each of the
10 V4' V4'
Lead Sequence STANDARD Modify …
11 V5' V5' Cancel Cancel channels indicating the lead currently assigned to that
Rhythm Lead II
12 V6'
Thermal Writer (AR200P) LPT1
V6' Help Help channel. To select another lead
Figure 4.3-11. Modifying the lead sequence Click the arrow to display all available leads
(Figure 4.3-12).
Hellige
Modify Lead Sequence Select the new lead to be recorded in the respective
Setup
channel.
Speed (mm/s) Lead 25 Label X 1-mV Cal Pulse
1 I I Audible LEAD OFF Alarm
Gain (mm/mV) 10
Rhythm Lead
V2
6 V3 II aVF
X Spline Once you have assigned a lead to each of the channels, you
V4
X Print after Analysis 7
V5 V1'
Modify
can enter a name for the lead sequence in the headline.
Report
V6
8 V2'
X Save full disclosure ECG D
Terminate after (min) A 20
9 V3'
Number of Leads
J
12 OK OK
Move the cursor to the text box and click.
10 V4' V4'
Lead Sequence STANDARD Modify …
11 V5' V5' Cancel Cancel
Rhythm Lead II
12 V6' V6' Help Help
Enter a name and select OK to clear the window.
Thermal Writer (AR200P) LPT1
Quit program
– conduct a test
Patient Untersuchung
ProcedureHellige
Selection Menu
Hilfsfunktionen Hilfe 4.4.2 The Basics
ECG External Programs
(Unbenannt)
Rest ECG STAT ECG Right-Heart Click to display the spirometry procedure screen. You
Catheterization Holter ECG
will find this icon on the procedure selection screen, unless
Exercise Test Monitoring
you have configured the icon bar to include that symbol
External Programs
(section 6. “System Services”).
120/80 HELLIGE CARDIOSOFT
T1.0a Setup
Ambulatory Blood Late Potentials
Pressure Click to display the procedure selection screen
Spirometry
(Figure 4.4-2).
…
Spirometry Bronchodilation
Click to display the spirometry procedure screen.
Cancel Help
Figure 4.4-2. Procedure Selection screen Once you select a patient, the system will display the test
information entry screen (Figure 4.4-3).
Test Information
Test Information
A
4.4.3 Entering the Test Information
Library: Reason for Test ABE F Reason for Test A B C
Respiratory problems
Shortness of breath
Pain » 2 The test information screen provides space for entry of impor-
Follow-up
tant information pertaining to the spirometry test.
1
These are:
– the reasons for test
3 – the medication
– comments
Medication …
Technician: Hilfe
Click on Medications… .
Medications… Medication
Medication Group Drug Name A window opens, showing medication group names (Figure
->
none
beta-blockers ->
4.4-6). There is a drug name window for each of these groups.
betamimetics ->
->
Click on the respective text box, if you wish to enter a name.
xanthine derivates
corticoids ->
Then type the drug name.
atropine ->
cromoglycinic acid ->
fenoterole ->
amiodarone -> Furthermore, you are free to arrange the medication groups
unknown ->
see Comment ->
Help
any way you want. To do so
->
OK
->
-> Cancel
Medication This brings up a new window where you can choose one of
Medication Group Drug Name
none ->
the group names (Figure 4.4-7).
none ->
beta-blockers
beta-blockers
betamimetics ->
xanthine derivates
xanthine derivates
corticoids
atropine
->
->
Click on the drug name box and enter the name
corticoids
cromoglycinic acid
atropine
fenoterole ->
amiodarone
cromoglycinic acid ->
unknown
fenoterole
see Comment -> Click on OK to clear the window.
amiodarone ->
unknown ->
see Comment ->
Help
->
OK
Entering Comments
->
-> Cancel
Physician’s/Technician’s Name
☞ Simply press on the keyboard to save the data The names of the physician and technician can either be
and clear the window. entered here (click in the text box and type the name), or they
can be selected from the list of the registered system users
(section 6.1 “System Configuration”).
Once you have entered all test information, you can either
– clear the window with Cancel (in this case the infor-
mation will be ignored and is not assigned to the test) to
display the test control screen.
Scale
0 The Spirometry Test Screen
– + Normal Values (ECCS) 0+R R
1
0
t(s)
MEF50(l/s)
MEF25 (l/s) 5 3.76
1.61
-1
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19
MMF (l/s)
IVC (l/s) 2.46
3.34
2.68 92 On this screen you see
FIVC (l/s) 2.63
-2
FIV1 (l/s) 2.23
-3 FIV1/IVC (%)
-4 PIF (l/s)
-5 MIV75 (l/s) – the control buttons (1)
MIV50 (l/s)
-6 MIV25 (l/s)
-7 MVV (l/min) 92.0
Patient Data
Select patient (local database) Modifying Patient Data
Search for …
Last Name Patient ID
Last Name First Name Date of Birth Patient ID For calculation of the spirometric normal values, the program
Master Jim 10/05/1949 8738999
Miller
Richards
Jane
Chantal
06/04/1943
03/12/1977
654987
7766699
Select
needs the following patient data:
Roberts Jennifer 07/27/1977 543876
Shields Stuart 11/11/1912 765321 New Patient
Stevenson Robert 11/22/1966 333999
Thompson Anthony 09/19/1932 665544 Delete
next page…
Unnamed age
Help height
Cancel
gender.
Close
New Patient…
Last Name Master First Name Jim
Patient ID 8738999 Date of Birth 10/05/1949 MM/DD/YYYY
Without these data which are adopted from the patient file
Gender Male Height 176 in (section 3. “Selecting and Admitting a Patient”) the normal
Race Weight 66/0 lb Pacemaker Accept
values cannot be derived. Height, age and gender are dis-
played on the spirometry procedure screen (Figure 4.4-8).
Figure 4.4-9. Patient data screen
You can measure the vital capacity VC and the forced vital
capacity FVC and combine both to form a bronchodilation
test (section 4.4.5). Click …
to select the bronchodilation
test, or to select a standard spirometry test.
☞ Before each test, attach a new mouthpiece with Before the test, we recommend that you give the patient an
bacterial filter to the sensor. idea of the test by demonstrating the animation.
When the patient is familiar with the test procedure, you can
start the test.
Attach the nose clip, so that the entire tidal volume flows
through the sensor.
-6
MIV50 (l/s)
MIV25 (l/s)
only) and the table lists the parameter readings and percentage
-7 MVV (l/min) 92.0
deviations from the normal values.
Figure 4.4-10. VC measurement
Click to initiate further measurements. The next
curve is depicted with another color and displayed on top of
the first one. The best results for EVC and IVC, which are
not necessarily taken from the same test, give VCmax.
Hellige
Patient Procedure System Services Help
Click Scale – + if you wish to change the curve size.
Baker, Dave, 41 yrs Spirometry / Recording FVC
Test Information End Patient Data 188 cm Male
Interpretation Miller Sq.
VC Measure FVC Measure Animation Setup
0 Select FVC Measure to measure the forced vital capacity.
Scale
– + Flow/Volume Volume/Time Normal Values (ECCS) 0+R R You will see a screen similar to Figure 4.4.-11.
7 F (I/s) Value Act. Norm. %
6 EVC (l) 3.14 2.68 118
FEVC (l) 2.63
5
FEV1 (l) 2.25
4 FEV1/VC (%) 81.0 Demonstrate the FVC procedure as well ( Animation ).
3 PEF (l/s) 5.86
2 MEF75 (l/s) 5.35
MEF50 (l/s) 3.76
1 V(l) MEF25 (l/s) 1.61
0
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19
MMF (l/s) 3.34 Start the procedure, when the patient knows what to do.
-1 IVC (l/s) 3.14 2.68 117
FIVC (l/s) 2.63
-2
FIV1 (l/s) 2.23
-3 FIV1/IVC (%)
-4 PIF (l/s)
MIV75 (l/s)
Check the nose clip is in place.
-5
MIV50 (l/s)
-6 MIV25 (l/s)
-7 MVV (l/min) 92.0
The recording area shows the flow-volume curve and the table
lists the parameter readings and percentage deviations from
the normal values. The program also presents an interpreta-
tion (obstruction/restriction) and marks the results in the
Miller square (Figure 4.4-11).
End
Hellige
Patient Procedure System Services Help Click End to terminate this part of the test and
Baker, Dave, 41 yrs Spirometry / Display / FVC Measurement
display the report (Figure 4.4-12).
188 cm Male
1 Interpretation
Obstruction
Miller Sq.
0
Scale
0+R R
Click to display the test information screen
– + Flow/Volume Volume/Time Normal Values (ECCS)
14 F (I/s) Value Norm. Act. % (Figure 4.4-13).
12 EVC (l) 2.68 3.14 117
FEVC (l) 2.63 2.93 111
10
8 2 FEV1 (l)
FEV1/VC (%)
2.25
81.0
2.57
76.0
114
94
6 PEF (l/s)
MEF75 (l/s)
5.86
5.35
5.62
5.56
96
104
The test information screen shows:
4
MEF50 (l/s) 3.76 5.25 140
2 V(l) MEF25 (l/s) 1.61 1.94 120
0 MMF (l/s) 3.34 4.23 127
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19
-2 IVC (l/s)
FIVC (l/s)
2.68
2.63
3.34
3.28
125
125
– the reason for test
-4
FIV1 (l/s) 2.23 2.61 117
-6 FIV1/IVC (%) 78.0 – the interpretation made by the program
-8 PIF (l/s) 3.48
-10 MIV75 (l/s)
MIV50 (l/s)
2.93
3.28 – the patient data
-12 MIV25 (l/s) 2.63
-14 MVV (l/min) 92.0 74.5 81 – medication, comments, physician, technician.
Figure 4.4-12. Report page
1 Button to call up the test information You can enter your own interpretation in the “Interpre-
2 Button to display the volume/time or the flow/ tation” window by placing the cursor in the window
volume curve and typing the text.
The icon allows you to export the most important data A test is considered confirmed, when a physician has reviewed
pertaining to the spirometry test to the clipboard or to any the test evaluation and “signed” it electronically with his/her
file. Data exported to the clipboard, for instance, can be name. Tests that have been reviewed and confirmed by a
inserted in a text file with the “Edit” —> “Paste” command. physician are marked with x.
For data exported to a file you will have to specify the target
directory and confirm it with OK .
The bronchodilation test always consists of two measure- The screen for VC measurement after medication will be
ments, one before and one after medication. displayed.
Click …
to select the bronchodilation procedure. Perform measurement of VC and FVC again.
The screen for VC measurement before medication will be Click End to terminate the measurement.
displayed.
You will now see both examinations on one screen so that you
Perform the measurement as described in section 4.4.4. can compare the effects of the medication (section 4.4.6
“Comparing Two Spirograms”).
Measure FVC in the same manner.
If you decide to run a bronchodilation test on a patient whose
Click End to terminate the measurement. spirogram was recorded earlier that day you will be asked
whether the first spirogram is to be taken as the reference test.
These tests are saved and annotated as measurement “before
bronchodilation”.
Patient Procedure
Hellige
Hilfsfunktionen re
Compare Procedure 4.4.6 Comparing Two Spirograms
Procedure 1
Baker, Dave 48 10.08.1996
J 11:32:23 Ruhe-EKG / Interpretation
Procedure 2
You can compare the present spirogram with another one
10.08.1996 11:32:51 Spirometry
Interpretation (Analyseprogramm)
recorded earlier. For the example below we assume that you
09.08.1996 11:12:45 Spirometry
08.08.1996 11:07:42 Spirometry
wish to compare the recent spirogram with an earlier record-
ing of the patient. Of course, you are free to call up any
spirogram (section 3.3 “Editing Stored Tests”) and compare it
» with any other spirogram as described below:
OK
Cancel Click .
Help
Select a spirogram.
OK
Hellige
Patient Procedure System Services Help
Click OK . You will then see a report screen
Baker, Dave, 41 yrs Spirometry / Compare / FVC Measurement
188 cm Male
showing the respiration curves and parameter readings
Test 1: 08.07.1994 · 10:32:54
Miller Sq.
Test 2: 09.07.1994 · 11:26:22
0
from both tests (Figure 4.4-15).
Scale
– + Flow/Volume Volume/Time Normal Values (ECCS) 0+R R
-4
FIVC (l/s)
FIV1 (l/s)
2.63
2.23
3.28
2.61
125
117
2.22 84
1.83 82
pages.
-6 FIV1/IVC (%) 78.0 85.0
-8 PIF (l/s) 3.48 2.35
-10 MIV75 (l/s) 2.93 1.92
MIV50 (l/s) 3.28 2.26
-12 MIV25 (l/s) 2.63 1.88
-14 MVV (l/min) 92.0 74.5 81 53.7 58
Setup
Setup
4.4.7 Spirometry System Setup
Equation for Normal Value Calculation:
ECCS Austrian
Local Settings
Print Spirogram
4.4.8 Printing a Spirogram
Procedure Date: 08.07.1994 – 10:32:38
Printer: Modify …
Number of Copies: 1
Check that the correct procedure is selected.
Print Cancel Help
Figure 4.4-17. Print setup screen Check the “Interpretation” box, if it is to be included in
the printed report.
Admit new patient or select patient from list Perform VC and FVC measurements as described
above
Enter test information
End End measurement
Attach nose clip
…
Administer medication and display
Initiate VC measurement and ask patient to bronchodilation test screen again
breathe in, out and in again as deeply as possible
through the mouthpiece Repeat VC and FVC measurements
FVC Measure Repeat measurement or end this part of the test End End test
with FVC Measure
Display report
1 1
4.5.1 General Information
2 3 2 3
E
ST IN
4 9 E
ST IN
4
The recording unit employed for ambulatory blood-pressure
STO ART STO ART
P P
AU
S ma AU
S ma
TES rqu TES rqu
T ett T ett
TON H
S
D/m /mm
m Hg
eH
EL
TON H
S
D/m /mm
H eH
EL
OPOF min Hg OPOF minmHg g
RT
-1
IV
LIG
E
15
RT
IV
-1 LIG
E
15
measurement is the TONOPORT IV/IVa unit. While the
90.0 .40 90.0 .40
9 9
ProcedureHellige
Selection Menu 4.5.2 Setting Up the Recording Unit
Patient Untersuchung Hilfsfunktionen Hilfe
ECG External Programs
(Unbenannt)
Spirometry Bronchodilation
Cancel Help
The signal acquisition screen is displayed with the key.
120/80
Figure 4.5-3. Ambulatory blood-pressure measurement, signal – information on whether or not the recording unit con-
acquisition screen tains data (2)
1 Control buttons
2 Message indicating the memory status – a separate field for details of the memory status (3)
3 Date and time of downloading from the record- – a summary of the memory status (4)
ing unit
4 Memory status details The memory status field indicates the configuration of the
5 Summary of the stored data recording unit.
227 498 02-A CardioSys V 4.1 177
Ambulatory Blood-Pressure Measurement
Interface COM 2
Patient Monitor Status
Protocol Program0
Data stored in patient monitor
Modify
Figure 4.5-4 will appear.
Date: 07.12.1995 Time: 14:11:59
Duration
Number of BP Measurements
24:00 h
48
Click on OK or Cancel to clear the window.
Number of ECG Strips 24
ECG Strip Length 3s
Ratio of ECG strips to BP measurements 1:2
Setting Up the Recording Unit
Figure 4.5-4. Window for entry of the interface and for protocol
selection Before setting up the recording unit, click on Clear to
delete all data stored in the memory (Figure 4.5-3).
Hellige In order to view the setup of the recording unit for a specific
Patient Procedure System Services Help
protocol, click on Modify… . You will then see the setup win-
Setup
Baker, David, 41 yrs Ambulatory
Langzeit-Blutdruck-Aufnahme
Blood Pressure Recording
dow (Figure 4.5-5) showing all the relevant data (see
No. of BP Periods 0 1 2 3 4 Protocol Program0
Defaults
Period from to
Setup
1
Download data
BP Interval (min)
Date/Time
2 Clear
Defl. Rate Infl. Press Qty
“Configuring Protocols” below). Click on OK to clear
Patient Monitor Status
1 07:00:00
Data stored in patient monitor the window.
Date: 07.12.1995 21:59:00 exact
Time: 14:11:59 30 4 200 30
Configuring Protocols
The two protocols can be configured to suit the clinical The BP interval (exact, approx.) and the deflation rate
picture of the patient to be monitored. The recording unit can be changed by clicking on the list box . Then you
retains these modifications even when switched off. can select the desired setting.
Display the setup screen as described above (Figure For TONOPORT IVa you can also specify the length of the
4.5-5). ECG strips (0 to 10 s) and whether an ECG strip is to be
recorded for every, every second or every third blood-pressure
First of all, select the number of blood-pressure periods measurement.
for the protocol.
☞ The storage capacity for ECG strips is limited: the
The day can be subdivided into up to 4 different periods, and longer the strips, the less strips can be stored (232
different measuring intervals can be selected for each of these 1-second strips or 23 10-second strips).
periods. Furthermore, you can choose whether the measure-
ments are to be taken at “exactly” or at “approximately” the Select OK to clear the setup menu.
selected intervals. Selecting the approximate intervals is advis-
able for apprehensive patients who may be stressed knowing
when a measurement is being taken. The deflation rate and
the cuff inflation pressure are also adjusted in this menu.
120/80
Figure 4.5-6. Signal acquisition screen, Download data button Connect the recording unit to the PC and turn it on.
Test Summary Hellige Display the signal acquisition screen for ambulatory
Patient Procedure System Services Help
Test Information
Test Information On the test summary page you find the test information,
A
Library: Reason for Test ABE F Reason for Test A B C reasons for test, medications and patient data as well as the
Respiratory problems
Shortness of breath statistics intervals (night time, wake-up time, Statistics Intervals
Pain »
Follow-up
button).
OK
Comment
Enter the procedure data as described in section 4.1.3.
Physician: Dr. Williams Cancel
Click .
Click on OK .
For data exported to a file you will have to specify the target
directory and confirm it with OK .
Graphics
Wake-up Time
Interpretation
Cancel
Hellige BP Graph
Patient Procedure System Services Help
Tabular
Hellige
Patient Procedure System Services Help
Return
Hourly Averages
Hellige Graphic Representation of the Hourly Averages
Patient Procedure System Services Help
300
Click on Tabular to display the same values in
250
HR
BP
tabular form (Figure 4.5-13).
200
150
100
50
0
06:00 08:00 10:00 12:00 14:00 16:00 18:00 20:00 22:00 00:00 02:00 04:00 06:00 08:00 10:00
Tabular
Hellige
Patient Procedure System Services Help
Return
Statistics Summary
Hellige Statistics Summary
Patient Procedure System Services Help
20%
values.
21
8 10
0%
< 110 120 130 140 150 160 170 180 190 200 > mmHG < 60 70 80 90 100 110 120 130 140 150 >
Minimum Maximum Average SD The percentage of readings which exceeded the set limits is
sys. BP 76 8:29 2. day 149 9:18 2. day 101.5 14.8
dia. BP
mn BP
57 15:16 1. day
69 15:16 1. day
85 14:56 1. day
105 9:18 2. day
75.1
84.8
5.4
6.7
shown at the bottom.
Heart Rate 62 112 74.7 6.1
21% of systolic blood-pressure readings exceeded 135 mmHg
0.0% of diastolic blood-pressure readings exceeded 85 mmHg
Similar report pages exist for the day time, the night time and
Figure 4.5-14. Statistics summary the wake-up time.
Interpretation Interpretation
Hellige
Patient Procedure System Services Help
Confirming Tests
Hellige
Patient Untersuchung Hilfsfunktionen
Print Hilfe 4.5.4 Documenting the Ambulatory
Pages to Print
Sütterlin, Franz, 48 J
1
Procedure Date: 07.12.1995 · 14:22:10
Baker, Dave
Langzeit-Blutdruck / Gesamtstatistik 00.00.0000 · 00:00:00
Blood-Pressure Measurement
Standard Page Tabular Hourly Averages
Procedure
1 2
Gesamtstatistik )7:51 9:46O
0 0 80% 3 1 You can determine the scope of the printed report by selecting
Statistics Summary Phases
60%
ECG Strips
or deselecting the individual documents. This is done on a
Day Time
Night Time
40%
special screen (Figure 4.5-16).
Wake-up Time 20%
0 0 0% 0 4 Print
110 120 130 140 150 160 170 180 190 200 mmHG 60 70 80 90 100 110 120 130 140 150
Click on to display the setup screen.
ECGBDGrid Barcode / Patient ID [22chrs] Cancel
sys. 76 8:29 2.none 149 9:18 2. 101.5 14.8
dia. BD 57 15:16 1. 85 14:56EPSON
1. EPL-5600 [LPT1:]
75.1 5.4
mit. BD 69 15:16 1. 105 9:18 2. 84.8 6.7 Save
Number of Copies
62
1
112
Modify…
74.7 6.1
Help
If you want to include a document in the report, click on the
2.1% 135 mmHg
0.0% 85 mmHg check box next to the respective title, so that a check mark is
Figure 4.5-16. Printer setup screen for documentation of the displayed. Click again, to deselect a document. At the same
ambulatory blood-pressure measurement proce- time the little text box at the top indicates the number of
dure pages to be printed (1, Figure 4.5-16). The printout is initi-
1 Number of pages ated with the Print button.
2 Recording date Save will save the printer setup.
3 Document included in the printout
4 Selected printer
Hourly Averages To select another printer, click on Modify… . The printer setup
Printout of the hourly averages in the form of tables and graphs window will open and you can click on another printer.
Statistics Summary
Printout of the procedure statistics in the form of histograms,
minimum, maximum and average values
188 CardioSys V 4.1 227 498 02-A
Ambulatory Blood-Pressure Measurement
Start program
120/80
Activate ambulatory blood-pressure measurement
mode
Having an amplitude in the microvolt range, late potentials Due to the small size of late-potential signals, a very careful
occur in the terminal QRS complex and in the early ST electrode technique is required. In most cases a resting ECG
segment. Proof of the existence of late potentials is obtained will be recorded before the ECG is analyzed for late
by signal averaging. For this method, the correlating QRS potentials. Therefore the electrodes for the 12 standard leads
complexes are summed up. Since signal noise is the non- are applied. Late potential analysis is performed on the uncor-
correlating portion, this method assures the required high rected leads X, Y and Z. To obtain these leads, chest elec-
sensitivity. A correlation coefficient can be set to any value trodes C1, C3 and C5 must be moved to other sites (see
between 1 and 99 (default: 95), thus determining the accuracy below) and all 6 electrodes must be carefully applied:
of the measurement: The higher the coefficient, the more
accurate is the measurement. Furthermore, an upper noise – clean and degrease the skin at the application sites, using
limit for the averaged signal is preset. When the noise drops benzine
below this limit, averaging stops and the measurement begins.
The analysis method used is the SIMSON method. A – massage the application sites with EPICONT cream
SIMSON algorithm is also used for determination of the until the perfusion increases (reddening of the skin)
QRS offset.
– use only silver-silver chloride electrodes
The ECG leads are the uncorrected, orthogonal leads X, Y
and Z, they are acquired via chest electrodes attached at the – after electrode application, wait for about 5 minutes
corresponding pick-up points. before acquiring the ECG signal.
After the analysis, the vector diagram is displayed, showing Apply the electrodes as follows:
the measured values and the average beats. Also included is
the procedure summary comprising the patient data, meas- C1 on the back, opposite C2 (signal across C1 and C2 = Z)
ured values, interpretation and medication.
C3 parasternal site, below the clavicula (signal across C3 and
The overall gain of the vector representation is 20 µV/cm, C4 = Y)
which is equivalent to a 50-fold gain of the standard sensitiv-
ity of 10 mm/mV. Alternative settings selectable on the vector C5 at the right, opposite C6 (signal across C5 and C6 = X)
screen are 5, 10, 20 and 40 µV.
Patient Untersuchung
ProcedureHellige
Selection Menu
Hilfsfunktionen Hilfe 4.6.3 The Late Potential Signal
ECG External Programs
(Unbenannt)
Acquisition Screen
Rest ECG STAT ECG Right-Heart
Catheterization Holter ECG
The signal acquisition screen for late potential analysis is
Exercise Test Monitoring
displayed with the button which is located on the proce-
External Programs
dure selection screen, unless you have reconfigured the stand-
120/80 HELLIGE CARDIOSOFT
T1.0a Setup
ard icon bar (section 6 “System Services Menu”).
Ambulatory Blood Late Potentials
Pressure
Spirometry
Entering the Test Information
…
Spirometry Bronchodilation
Click on to display the procedure selection menu
Cancel Help
(Figure 4.6-1).
Figure 4.6-1. Procedure selection menu
Click on to display the signal acquisition screen for
late potential analysis.
Corr. Beats 1
100 % QRS 43 3
Noise (<1.0 µV) 1.7 µV %Corr. Beats (200) The signal acquisition screen shows:
3
1
2
5
4
6
X
R
4
L
– information on the ongoing analysis (2):
N F
number of correlating beats in %
Y the noise limit as configured
6 Lead Sequence
the number of correlating QRS complexes
XYZ - uncorrected
Z
5 HFLA Treshold [µV]
a bar graph illustrating the acquired correlating beats as a
40
percentage of the required beats as configured
Figure 4.6-3. Signal acquisition screen for late potential analysis
1 Control buttons – the current heart rate and blood pressure readings (3) (if
2 Details on late potential analysis entered manually, section 4.1.3)
3 Current heart rate and blood-pressure readings
4 Applied electrodes – the applied electrodes and the signal quality (4) (pro-
green: properly applied vided the red R electrode is applied). Furthermore an
yellow: not properly applied arrow (6) points to the leads which are impaired by
white: site not used improper electrode application or which cannot be
red: disconnected (lead break) recorded due to disconnected electrodes (or due to a
5 Lead sequence break in the patient cable)
6 Lead impaired by poor signal quality
– the selected lead sequence (5)
Baker, Dave, 41 yrs Late Potentials / Signal Acquisition Click on I to initiate the analysis.
HR
Test Information Freeze Traces Options »
Setup Reset QRS 10mm/mV 25mm/s 60
SYS/DIA (mm/Hg)
The signal acquisition screen shows
Corr. Beats 100 % QRS 43
Noise (<1.0µV) 1.7 µV %Corr. Beats (200)
3 – the percentage of correlating beats
1
2
5 6
4
X
R L – the selected noise limit in µV
N F
Z
Lead Sequence
XYZ - uncorrected – a bar graph representing the acquired, correlating beats
HFLA Treshold [µV]
40 as a percentage of the necessary beats
☞ After every fifth correlating QRS complex the You can stop the 2nd and 3rd ECG trace with Freeze Traces .
program checks whether the signal noise lies below
the limit. If so, signal averaging is terminated, You can enable and disable the 2nd row of control buttons
even when the set number of QRS complexes has with Options >> .
not yet been reached.
You can display the setup menu (section 4.6.5) with
Setup .
Noise [µV]
RMS40 [µV]
3 1.6 1.6
7 7
4.6-5).
128 128
QRS Duration [ms]
HFLA Duration [ms]52 52
Y
5 Other documents of the procedure report are the interpreta-
4 tion (Figure 4.6-7) and the test summary (Figure 4.6-8).
40 µV
Z
Vector Diagram
40 ms
6 7
Figure 4.6-5 shows the vector diagram with the marks identi-
Figure 4.6-5. Vector diagram fying onset (6) and offset (7) of the QRS complex set by the
1 Buttons to adjust the gain program as well as the resulting measured values (4). The
2 Buttons to adjust the sweep speed HFLA threshold is a dashed line and the HFLA duration is
3 Buttons to display other report documents displayed with a different color.
4 Parameter table
5 Medians (non-amplified) The signal size can be changed with the buttons.
6 QRS onset
7 QRS offset The sweep speed can be changed with the but-
tons.
☞ This noise value (4) may differ from the one Furthermore, you can correct the position of QRS onset
shown on the signal acquisition screen, as it is and offset as follows:
calculated over all acquired beats.
Position the cursor on the upper portion of the caliper
and press the left mouse button, when the cursor turns
into a double-headed arrow pointer.
Hellige RMS
Patient Procedure System Services Help
Cancel Y You can display the other report pages either by paging
through the report or by accessing them directly.
40 µV
Z
Click on or to page forward or backward
40 ms
through the report.
Figure 4.6-6. Table of contents of the late potential analysis Click on to display the table of contents and select
report a document (Figure 4.6-6).
Test Summary
Hellige Test Summary
Patient Procedure System Services Help
Medication
Comment
Physician Dr. Williams If you wish to add or edit test demographics, click on the
Technician Dorothy
ABE
icon to display the corresponding dialog window.
F
Figure 4.6-7. Test summary
Data Export
Click .
Click on OK .
For data exported to a file you will have to specify the target
directory and confirm it with OK .
Interpretation Interpretation
Hellige
Patient Procedure System Services Help
Confirming Tests
When you display the statement library (1, Figure 4.6-8), the
ABE F
icon becomes active.
Patient Procedure
Hellige
System Services
Print Help 4.6.4 Documenting the Analysis
Pages to Print
Bauer, Erna, 30 yrs
1
Late Potentials / Procedure SummaryProcedure Date: 07.12.1996
Dave Baker
07.12.1995· ·16:02:24
11:40:23
Results
Speed Gain
ABE F
Vector
1400 mm/s 2 Click on the button to obtain a document of the
Indication
Speed Gain
3 Age
Sex
48 yrs
female
analysis results.
Race
Medians 100 mm/s 20 mm/mV Height 168 cm
Interpretation Weight 56.0 kg
Blood Pressure
HR 60 BPM
The printer setup screen displays (Figure 4.6-9) where you
Interpretation
Number of QRS 43 can change the signal size and the sweep speed of the vector
4
Noise [µV]
RMS40 [µV]
Print
1.6
7 diagram and the median complexes before the printout.
Cancel
ECG Grid 5 mm Barcode / Patient ID [22chrs] QRS Duration [ms] 128
EPSON EPL-5600 [LPT1:] HFLA Duration [ms] 52
Save
Number of Copies
Medication
Comment
1 Modify…
Help On the same screen you include or exclude the interpretation,
Physician Dr. Braun
User Moessner you select or deselect the ECG grid and you choose the
Figure 4.6-9. Printer setup screen number of copies as well as the printer.
Setup
Patient Procedure System Services
Hellige
Help 4.6.5 Setup
Baker, David, 48 yrs SetupAcquisition
Late Potential / Signal
Test Demographics25
Speed (mm/s) Freeze ECG Options » HR Click on Options >> to display the second line of control
Gain (mm/mV) Reset QRS
Defaults 10 10mm/mV 25mm/s
SYS/DIA (mm/Hg)
60 buttons.
Lead Check noise level low
Corr. Beats 100 % QRS 43
Noise (<1.0µV) 1.7 µV %Corr.
Cut-off Frequency (Hz) 40 Beats (200)
Noise Limit (µV) 1.0 (0.1…5.0)
Click on Setup to call up the setup screen (Fig-
3
4
6
ure 4.6-10).
X
HFLA Treshold (µV) 40 (5…99) R L
N F
Number of QRS 200 (20…500)
OK Use this screen to select your personal system settings. These
Y
X 1-mV Cal Pulse Cancel
settings will be saved.
Audible LEAD OFF Alarm
Help Lead Sequence
Print after Recording XYZ - uncorrected
Z
HFLA Treshold [µV]
40 Adjustments are made in three different ways:
Figure 4.6-10. The defaults menu
1 There is an arrow at the end of a box. Clicking this
arrow will open a window which offers various options.
Select the desired setting by clicking the mouse button.
The Marquette Hellige program has a special BDT interface Dialog between Marquette Hellige program and external
for communication with external programs. program:
ProcedureHellige
Patient
ECG
Untersuchung
Selection Menu
Hilfsfunktionen Hilfe
External Programs
4.7.2 Setting Up External Programs
(Unbenannt)
Rest ECG STAT ECG Right-Heart Click on to display the procedure selection menu
Catheterization Holter ECG
(Figure 4.7-1).
Exercise Test Monitoring
External Programs
120/80 HELLIGE CARDIOSOFT All icons of the external programs are dimmed, which means
T1.0a Setup
that they are inactive. To activate an icon, the corresponding
Ambulatory Blood Late Potentials
Pressure program must be set up first.
Spirometry
Cancel Help
4.7-2).
Figure 4.7-1. Procedure selection menu In the list box, click on the program to be configured,
then type the program path and name in the text box
below (e.g., c:\ecg\holter.exe). For “External Program”
Setup Hellige you will have to add a name for the examination under
Procedure
PPGSelection
Hellige Menu
Patient Untersuchung Hilfsfunktionen Hilfe
ECG External Programs which it will be saved.
Baker External Program Setup
RestPrograms
ECG STAT ECG Holder ECG Ultrasound
External
Select OK or Cancel to clear the window.
Holter ECG
Ultrasound
Stress Echocardiography
X-Ray
Stress Test Monitoring
Cardiac Catheterization
Cath Lab Data
Stress
Echocardiography
X-Ray
☞ When changing the external program setup, please
External
120/80 HELLIGE CARDIOSOFT
Program pay attention to the compatibility with archived
Ambulatory Blood Late Potentials
T1.0a
Cardiac Cath. Lab. Data procedures.
Program path and name
Pressure c:\|ecg\holter.exe
Catheterization
Spirometry CARDIO-
SYS
OK Cancel
external Program Help
… Configuration
Spirometry Bronchodilation
Cancel Help
Deleting Examinations
Select .
Password
Patient Procedure
Hellige
System Services
SystemPPG
Services
HelligeMenu
Help 6.1 System Configuration
(Unbenannt)
Procedure Archiving Setup
Click : the window showing the system services
opens (Figure 6-1).
System Copy Patient Compression Load
Configuration Records
General Tab
Cancel Help
Name
Database…
Default Procedure
Street
City Here you select the procedure which the system selects auto-
Remote Cart Name
matically when turned on. Click the arrow , then click the
Physician ID
User List…
Technician ID
Modify password…
desired procedure.
Default Procedure Windows control menu
(Restart program for changes to become
Trace Background black-yellow (antial.) effective)
Trace Background
Screen Diagonal 15 in Menu labels
Remote Test Screen NETBIOS-Protocol You can select various colors for the background on which the
Test Patient (temporary) Enable Password Function signal traces are displayed. Click the arrow , then click the
Anti-Aliasing of ECG Waveforms Patient Selection prior to STAT ECG
Save
Load Screen Diagonal
Print
For Service only Help Select the screen size of your PC system’s monitor to assure
OK Cancel that the traces are displayed true to scale. Click the arrow ,
Figure 6-2. System configuration screen then click the dimension.
Setup
Technician
2
Click on the User List… button to display the list of
External User
Privileges
3 registered users (technician, physician, privileges).
Edit Data
Configure System
To register a new user,
Password …
Figure 6-3. Dialog window A user registration form (Figure 6-3) will appear.
Password (5)
To change the user’s password.
Menu labels
Above the icons you can display the menu labels, which
allows you to click the words and select options from the
drop-down menus.
Icon Bar
Here you choose the icons that will appear in the icon bar and
the order of icons.
Port
Baud Rate
COM2
19200
If you wish to send data via a modem, select the port, the
Modem Type
Initialization
MultiTech
baud rate and the modem model here (set sender and receiver
AT&FMO&E7X3VO&Q1S
Dial Mode ATDT to the same baud rate).
Hang-up Prefix +++
Hang-up Suffix ATHO For Multitech and ELSA modems, the system automatically
enters all other necessary data. If you choose another modem,
you will have to enter them yourself.
Help
Init string for ELSA modem:
AT&FM0\X1X3V0S0=1
OK Cancel
1. factory defaults
Only modems complying with the requirements
Warning
of IEC 60950 or UL1950 may be connected to
2. response code in digital form
the system. Furthermore, the regulations appli-
cable in the respective countries must be ob-
3. for telephone system with extensions: display additional
served.
BUSY message
The modem must be set up inside the medically
used room, but outside the patient environ-
4. modem answers after 1st ring
ment.
System Configuration
General Modem MUSE Option Code Country Settings
MUSE Tab
Setup for MUSE
On this tab you make all selections for the transfer of data to a
MUSE User Name
2 MUSE Password
MUSE CV database. First of all you should obtain the follow-
3 ing information: MUSE server network address, login name,
Store procedure for MUSE
4 No data transfer to MUSE and password. Verify that the network configuration is OK
Save MUSE data to medium
5 Drive A and that the Internet Explorer and Acrobat Reader are prop-
Data transfer to MUSE via the network
10 Delete local test data after transfer to MUSE 2 Entry fields for MUSE database queries.
Start modem connection before transfer
Location Number 0 Cart Number 0
11 OK
Help
Cancel
3 Entry field for MUSE site number.
12
4 Click radio button if you do not want to transfer data to
Figure 6-5. MUSE tab MUSE.
OK Cancel
System Configuration
General Modem MUSE Option Code Country Settings
On this screen you select all country-specific system settings.
Language English
Time Format 24-Hour Format 1 Enable or disable the Swedish patient ID.
Date Format DD.MM.YYYY
Unit of Measure cm, kg
Speed Unit km/h
ST Level mV 2 Enable or disable the Card Reader function (patient data
Line Frequency 50 Hz
Lead Label IEC acquired from patient ID card).
Enable Swedish Patient ID
Help
4 Select the AC line filter frequency (50/60 Hz).
OK Cancel
5 Select the language.
Figure 6-7. Country Settings tab
6 Select the time format.
8 Select the units for height (cm, inches) and weight (kg,
lb).
9 Select the speed unit (km/h, mph). Ensure that the same
unit is selected at the treadmill.
Patient Procedure
Hellige
System Services Help 6.2 The Voice Recorder
SystemPPG
Services
HelligeMenu
(Unbenannt)
Procedure Archiving Setup
The voice recorder feature allows you to record comments,
save them with the date and under a specific heading, and
System
Configuration
Copy Patient
Records
Compression Load play them back. If the program runs in a network environ-
ment, you can call up the comments from all connected
HELLIGE CARDIOSOFT
Voice from ECG T1.0a
Archive Patient Save
stations.
Recorder System/ECG Records
Recorder
Click the icon: you will see a list of all stored voice
Figure 6-8. System Services Menu recordings (Figure 6-9).
Hellige
Patient Procedure System Services Help
Voice
PPGRecorder
Hellige
Date / Time of Recording, Category
(Unbenannt)
07.08.1994 14:01:00 Info to personnel
HELLIGE CARDIOSOFT
Modify… Cancel Help
T1.0a
Referral
Info to telephone switchboard The voice recording window appears (Figure 6-10). The
Info to staff
Standard Categories box contains up to 6 categories or head-
Private
HELLIGE CARDIOSOFT ings for standard texts that you may want to record. To copy
T1.0a
Modify… OK one of these standard headings into the Category box, simply
Cancel double-click it in the left-hand column. It is possible to copy
Help more than one standard heading into the Category box.
Click Modify… .
You will now see the table of contents, including the new
recording.
Click the icon: you will see a list of all saved speech
recordings.
Click or , as required.
Patient Procedure
System
PPG
Hellige
System Services
Services
HelligeMenu
Help 6.3 Copying Patient Records
(Unbenannt)
Procedure Archiving Setup
to/from Storage Medium
Baker
Patient Medium Directory Copying Records from the PC to the Storage Medium
Find… Modify…
4
Jane Miller, 42 yrs. 1 Check that the correct patient is selected.
Patient Records
10.08.1998,
HELLIGE CARDIOSOFT
13:22:01, Spirometry
Records on Storage Medium
Click Find ... , if you wish to select another patient.
10.08.1998, 12:24:13, Stress Test
T1.0a
10.08.1998,
10.08.1998, 2
12:05:05,
12:03:48, » Stress Test
Stress Test
10.08.1998, 11:45:19, 3 5
Stress Test This opens the patient selection window. Select a patient as
10.08.1998,
10.08.1998,
11:41:42,
11:32:36, » Rest ECG
Rest ECG
described in section 3.1. All records of this patient appear in
10.08.1998, 11:21:07, Stress Test
10.08.1998, 11:12:45, Rest ECG
the list below.
Cancel Help
– the storage medium is write-protected Click >> to initiate transfer of the selected file(s).
The copied record(s) will appear in the “Records on
– the storage medium is unformatted
Storage Medium” window.
– the storage capacity is insufficient
Restoring Files from the Storage Medium to the PC
– the storage medium is defective
Patient Procedure
Hellige
System Services Help 6.4 Transferring Resting ECGs from
SystemPPG
Services
HelligeMenu
(Unbenannt)
the ECG Recorder
Procedure Archiving Setup
Cancel Help
In general, the system will automatically assign the incoming
records to the appropriate patients, if they exist in the data-
base.
Figure 6-13. System Services Menu
If the incoming record is without patient ID (and without
blanks), the patient ID 000000 will be entered.
☞ Some of the patient data are not transferred (e.g.
the medication). Always verify the records sent
If the patient ID of the incoming record exists in the database,
from an ECG system/recorder.
the record will be assigned to that patient. The patient data of
the record will be updated with those in the database.
☞ If a user ID was not entered, the ECG recorders
will ocasionally send the record with a number of
“Automatically assign procedure to patient” disabled:
blanks in place of the ID.
If the patient ID of the incoming record does not exist in the
database, the record will be registered to the list of received
records and the user has to assign it manually to the correct
patient.
Modify…
First of all, select the transfer system
(ECG system = CardioSys/Soft,
OK Cancel Help ECG recorder = Marquette Hellige electrocardiograph).
Figure 6-14. Transfer setup window (from ECG recorder)
Select the appropriate baud rate (19200, if possible; refer
to section “Specifications” in the ECG recorder manual).
☞ A device number (e.g. number 12) can be entered
for ECG recorders, by writing “Hospital %12%” Select the communications port .
in the Hospital/Practice box.
Select the directory, if you wish to transfer data from a
storage medium.
Patient Procedure
Hellige
System Services Help 6.5 Data Compression,
SystemPPG
Services
HelligeMenu
(Unbenannt)
Archiving Patient Records
Procedure Archiving Setup
Data Compression
System Copy Patient Compression Load
Configuration Records
Help
Click on Cancel to terminate data compression.
Figure 6-17. Data compression setup screen
Hellige
Patient Procedure System Services
System
PPG
Services
HelligeMenu
Help
Archiving Patient Records (e. g. to a magneto-optical disk)
(Unbenannt)
Procedure Archiving Setup
If you select one of these files for display, the program tells
Cancel Help
you where they are stored. Before archived records can be
transferred to MUSE, they must be restored to the system.
Figure 6-18. System Services Menu Click : the window showing the System Services
opens (Figure 6-18).
☞ In a network environment, the program must be
terminated at all other workstations. Click the icon. The archive setup screen appears
(Figure 6-19).
Import… 01.01.1999 Export… Insert the storage medium and, after the archive proce-
dure, label it with the number displayed.
Cancel
Help
Patient Procedure
Hellige
System Services
SystemPPG
Services
HelligeMenu
Help 6.6 Saving and Loading Custom
(Unbenannt)
Setups
Procedure Archiving Setup
The program allows you to save setups and text libraries under
System
Configuration
Copy Patient
Records
Compression Load a user or ward name and to quickly load these settings when-
ever needed (exceptions to this rule are explained at the end of
HELLIGE CARDIOSOFT
Voice from ECG T1.0a
Archive Patient Save
this chapter). With this feature you can adjust the system to
Recorder System/ECG Records
Recorder suit a specific user’s requirements in a matter of seconds.
When running in a network environment, the program
Cancel Help
settings, including the factory settings, can be called up at any
connected station.
Figure 6-20. System Services Menu Click : the window showing the System Services
opens (Figure 6-20).
2. In the “Save As” field, enter the name for this setup.
☞ When reactivating the factory settings, the exercise
test protocols and phases that you created yourself
3. Click the Save button.
will be deleted with some ergometer models.
Load
* Before cleaning and disinfecting the device, disconnect it * Discard disposable electrodes immediately after use to
from the power line. prevent their being reused.
* Clean the instrument surface with a cloth moistened in a * Clean reusable electrodes immediately upon removal
cleaning solution. Take care that the solution does not from the patient.
enter the device. All hospital-grade cleaning solutions
and disinfectants containing up to 70% alcohol are * Peel off the adhesive foil before cleaning them (residual
suitable. adhesive can be removed with benzine).
* Disinfectants with a phenol base or peroxide compounds * Use warm water and a small brush to remove electrode
are not suitable for disinfection of the instrument sur- cream. Do not use pointed metal objects to remove solid
face. particles of dirt from the electrodes.
7.2 Patient Cable and Electrodes * Disinfect the electrodes with alcohol-free disinfectants.
Make sure that plugs and sockets do not become wet.
* Disconnect the cable from the device before cleaning or
disinfecting it. To do this, pull on the plug, not on the * Use only gas to sterilize the electrodes (frequently steriliz-
cable itself. ing the electrodes with ethylene oxide shortens the life of
the plastic material).
* Rub the cable clean with a cloth soaked in soapy water.
To disinfect the cable, rub it down with disinfectant. Do * For information on cleaning, disinfection and steriliza-
not immerse the cable in liquids! tion of other sensors and transducers, please refer to the
instructions for use included with these devices.
This applies on the condition that the patient does not suffer
from infections of the respiratory tract and that not more than
15 respirations are measured on the patient.
7.4 Maintenance
Visually inspect the system, all cables, and electrodes for signs For safety, the devices require regular maintenance. To ensure
of mechanical damage before each use. functional and operational safety of the system, technical
inspections should be carried out annually.
If you detect defects or malfunctions which may impair the
patient’s or the user’s safety, the system must be repaired These inspections should be referred to independent persons
before being used again. with adequate training and experience.
The system does not require other regular maintenance. Within the framework of a service contract Marquette Hellige
service representatives would be pleased to carry out these
preventive maintenance inspections for you.
8. Troubleshooting Tips
System Messages
Malfunctions
9. Technical Specifications
ECG Processing
Lead selection
Automatic functions
* formatted reports
simultaneous transmission of all electrode signals to digital * noise of signal transmission path less than specified in
signal processing system after lead derivation and digitization; IEC and AHA recommendations: < 2.5 µV rms
a muscle filter and an AC filter can be activated simultane-
ously for all leads, pacing pulse detection, automatic or * common-mode rejection for 50 or 60-Hz signals
manual sensitivity adjustment, automatic baseline adjustment, (depending on model) with AC filter enabled >140 dB
drift compensation by cubic spline
ECG calibration
* lower cut-off frequency (-3-dB limits) 0.04 Hz, corre-
sponding to a time constant of 4 s automatic recording of a defined voltage jump, valid for all
channels
* upper cut-off frequency (-3-dB limits)
resting ECG/STAT ECG: 150 Hz (IEC/AHA) * calibration voltage, referred to the ECG signal input:
stress-test: 75 Hz (IEC) 1 mV
* resolution, referred to the input: 5 µV automatic adjustment to the optimal writing range, in
dependance of signal amplitude
* sensitivity simultaneously adjustable for all leads:
40 – 20 – 10 – 5 mm/mV Cubic Spline
* 3-dB drop of the amplitude frequency response with automatic compensation of baseline fluctuations caused by
active muscle filter (low-pass) at about 40 Hz variations of the polarization voltage at the pick-up electrodes
ECG signal input * pulse voltage resistance of all pick-up electrode connec-
tions and N-connection referred to chassis of either
isolated patient connection type CF according to IEC, all polarity (e.g., defibrillation) 5000 V
pick-up electrode connections and N-connection high-
voltage-proof, electrodes are monitored for signal quality and ECG signal output
lead breaks
ECG signal output for synchronization of ultrasonic units
* electrode connections for R, L, F, N, C1 to C6, Nax, (not for diagnostic purposes)
Nst, Nap (=C4)
The initialization that takes place each time a test mode is
* input impedance for differential signals between any two started briefly blanks the ECG signal at the analog output.
electrodes at 10 Hz > 10 MOhms
The medical system created by connection of external devices
* input impedance for common-mode signals referred to must fulfill the requirements of standard IEC 60601-1-1/
N up to 60 Hz > 50 MOhms EN 60601-1-1
* dynamic range for differential signals between any – ECG lead selectable
electrode connections ±10 mV for AC voltage, ±600 mV – electrically isolated (±1500 V referred to chassis)
for superimposed DC voltage (polarization voltage) – input signal of 1 mV results in 1-Volt output signal
– Uoutmax ±10 V
* dynamic range for common-mode signals referred to – offset accuracy < 300 mV (typical, full scale)
N ±1V, referred to chassis 264 V rms – accuracy over entire width < 3% (typical)
– R l min. 2 kOhms
* quiescent input current via any electrode connection for – delay <10 ms
1-kOhm termination referred to N < 50 nA – pacing pulses removed by filtering
* patient leakage current (rms values) according to IEC (The low-pass characteristic of the input stage of the con-
class CF: in normal condition <10 µA, in single-fault nected trigger detection circuit should be fg <500 Hz.)
condition (e.g., patient in contact with line voltage)
ECG
<20 µA 3 1
2
Visual and audible indication of disconnected electrodes and 15-inch or 17-inch monitors can be connected
broken leads. Individual monitoring of each electrode. (only the monitor models supplied by Marquette Hellige are
allowed to be installed on the system cart)
Interfaces
Power supply
PS/2 mouse
from the power line, system design in compliance with
Serial interface IEC 60601 protection class I
* power consumption
LPT1 printer
w/o. monitor 60 W
LPT2 CORINA
with 15-inch monitor 130 W
with 17-inch monitor 140 W
Printer
Environment
Operation
height: 1150 mm
width: 680 mm
depth: 825 mm
weight: 60 kg (w/o. monitor and printer)
10. Equipment
Accessories
General accessories
System carts
701 249 16 Controller for MO or streamer drive 455 024 01 Resting ECG interpretation
226 169 01 Paper for thermal printer AR 200 System cart accessories
384 018 92 Network option PCI (upgrade) 915 418 12 Adapter 9/25 pole, D-SUB interface
384 015 80 Network option BNC (requires system PC 915 418 11 Adapter (AT keyboard to PS2)
with network board 701 180 66)
919 202 47 Mouse pad
384 017 33 Network option TP (requires system PC with
701 244 01 Printer cart
network board 701 180 72)
919 200 37 Potential equalization cable 384 018 17 Set of 10 leadwires, 4-mm connector, defib-
protected, AHA
223 330 03 Connection cable for M 700
223 362 03 Connection cable for EC 560/ECB 561/ 217 225 03 Adhesive electrode for babies, 13 mm diameter,
CardioSmart with 4-mm connector, 60-cm lead, to be
attached with adhesive rings 927 224 00
223 368 01 Connection cable for TM 400 E
217 110 03 Adhesive electrode for children, 22 mm
223 380 01 Connection cable for ERGOLINE ERG 900
diameter, 4-mm connector, to be attached with
223 381 01 Connection cable for SUNTECH 4240 adhesive rings 217 123 01
223 390 01 Connection cable for T2000 (RS 422) 217 320 01 Adhesive electrode for children, 22 mm
diameter, with press-stud connector, to be
700609-001 Connection cable for T2000 (RS 232)
attached with adhesive rings 217 123 01
223 372 01 Connection cable for EK 51, EK 56, EK 512
217 321 01 Adhesive electrode for adults, 35 mm diameter,
223 298 04 Network connection cable with press-stud connector, to be attached with
adhesive rings 217 123 01
223 363 03 Connection cable for bosotron BP monitor
303 442 96 Adapter for electrode with press-stud connector
Patient Cables to patient cable with 4-mm connector
223 418 01 Patient trunk cable, 10-lead, IEC Multi-Link, 504 648 56 Limb-lead electrode for adults (stainless-steel
2.2 m plate electrode), 31 x 40 mm, with 4-mm
socket
223 418 02 Patient trunk cable, 10-lead, AHA Multi-Link,
2.2 m
9490-210 Clip adapter for SILVER MACTRODE, 10 ECG Accessories, Cava Catheter
adapters
223 408 04 Patient trunk cable, 5-lead, for cava catheter
NEHB Accessories 384 017 78 Set of 5 leadwires, IEC, for electrodes with
press stud
223 403 03 Patient trunk cable, 12-lead, NEHB, IEC
384 017 66 Set of 2 leadwires, NEHB, IEC, with 4-mm 217 123 01 Adhesive rings, pkg. of 500, for electrodes
connector (in addition to 384 017 65) 217 110 .., 217 320 .. and 217 321 ..
217 083 14 Electrode cream, 5-liter container 208 019 01 Respiration Flow Sensor LF 501
930 115 82 Dispenser, 30 ml 928 056 62 Support for Respiration Flow Sensor
217 083 06 Electrode gel, pkg. of 10 tubes, 100 ml each 303 440 83 Disposable respiration filters (100 pcs.)
217 007 01 Electrode paper, 200 sheets, for electrode 701 217 37 Connection cable TONOPORT IV to PC
504 648 56 (9 pin)
217 148 01 Electrode paper, 200 sheets, for electrode 710 217 31 TONOPORT IV recording unit
217 144 01/02
710 217 32 TONOPORT IVa recording unit
217 043 02 ECG Filing Cards (50 pcs.)
Do not install CardioSoft on your PC if you are not familiar * Select “Start” —> “Settings” —> “Control Panel”.
with the Windows NT operating system!
* Double-click “Printer”.
Prerequisite for operation in conjunction with MUSE:
Install the Internet Explorer 4.0 and the Acrobat Reader, if * Select the printer you will use.
you want to view test data stored in the MUSE CV database.
* Select “File” —> “Properties” —> “Scheduling”.
* Switch on the PC and the monitor. Exit ALL applica-
tions. * Select “Print directly to the printer” and confirm with
“OK”.
* Insert the CardioSoft CD in the CR ROM drive.
We recommend a resolution of 800 x 600 (15" monitor) or
* Select “Start” —> “Run”. 1024 x 768 (17" monitor). For resolutions of 1024 x 768 or
better we recommend to select “large font” in Windows –
* At the command line, enter Setup – Display.
X:\setup
(X = CD ROM drive letter) ☞ The file V24.DLL is protected under the provi-
sions of the copyright law. Copyright © 1992-
* Click “OK”. 1996 Langner GmbH. All rights reserved. Permis-
sion is granted to use the file V24.DLL only in
* Select “Next” to confirm the suggested directory (when combination with the CardioSoft program, and
reinstalling the program, be sure to select the same not in combination with any other program.
drive). Copying and making the file V24.DLL available
to third parties is prohibited.
CardioSoft will now be installed on your computer.
Be sure to properly terminate the program before switching
* Restart Windows. off the PC.
CardioSys/Soft is supplied with a special BDT interface Communication between reimbursement program and
which allows it to be operated in conjunction with a medical examination program:
reimbursement program or clinical information system. Please
contact the supplier of these programs to find out about the – start of the reimbursement program
conditions for full compatibility.
– selection of a patient
When running the program in conjunction with a medical
reimbursement program, the patient data file is handled by – display of file card
the reimbursement program, whereas the procedure data files
are handled by the Marquette Hellige program (CardioSys/ – selection of a procedure (start of the examination
CardioSoft). In this case patients are selected via the reim- program)
bursement program. The patient’s height and weight should
be entered in the reimbursement program, but they can also – new procedures (ECG, spirogram, etc.)
be added in CardioSys/CardioSoft.
– termination of examination program
Having selected the patient, you choose the procedure to be
performed. This takes you automatically to the Marquette – data transfer to reimbursement program.
Hellige program, which is handled as described in this
manual. When the Marquette Hellige program runs in conjunction
with a reimbursement program, the following functions are
You can call up previous examinations of the selected patient disabled:
and compare them or conduct new procedures, but you
cannot select another patient. To do this, you have to quit the – selecting another patient, admitting a new patient
examination program first. Then you can choose a new
patient in the reimbursement program. The following patient data are adopted by the reimbursement
program: last name, first name, patient ID, date of birth,
After the procedure and quitting the examination program the height, weight.
most important procedure results (not the signal traces!) are
transferred to the reimbursement program. Some of the
reimbursement programs automatically suggest a bill.
3. System Maintenance
We recommend making one backup of the operating system, * Select the backup method (for backups of the operating
and making daily backups of the CardioSys/Soft program and system “normal“, for the first backup of CardioSys/Soft
patient files. data “normal“, for subsequent backups “differential”.
When making a “differential” backup, select “append“,
Use the Windows backup application and refer to the online for normal backups select “replace”.
help for instructions.
* Click the OK button.
* Insert the tape.
The backup session begins. After the session, the backup will
* On the Windows screen, select “Start” —> “Programs” be verified.
—> “Administrative Tools (Common)” —> “Backup”.
* Click the OK button to clear the window.
* Select drive C or drive D.
* Insert the tape. Before saving data to an optical disk, the disk must be
formatted.
* On the Windows screen, select “Start” —> “Programs”
—> “Administrative Tools (Common)” —> “Backup”. * Insert the disk.
* Open the Tapes window (icon in the bottom left-hand * On the Windows screen, select “Start” —> “Programs”
corner of the screen). —> “Administrative Tools (Common)” —> “Disk
Administrator”.
* Select the appropriate tape.
The disk administrator displays symbols for the different
* Click the Restore button. drives:
drive 0 = hard drive
* Select the target drive. drive 1 = optical disk (with free storage capacity)
* When restoring the operating system, select “Restore * Click the “Drive 1” box.
registry” and “Restore file authorization”. When restor-
ing CardioSys/Soft and patient files, select “Restore file * In the “Partition” menu select “Create” and confirm the
authorization”. max. storage capacity shown with OK.
* Click the OK button. The field for drive 1 now says “unformatted xxx MB”.
* After the restore, click the OK button to clear the * In the “Partition” menu, select “Make changes now” and
window. click “OK”.
The field for drive 1 now shows the drive letter, e.g. E.
The documents usually sent to the printer can be directly Explanations: AT command start
routed to another PC or to a fax machine. Required accesso- &F default setting
ries are a modem (recommended model: ELSA Microlink X3 for telephone systems with extensions
14.4TQ) and the WINFAX Pro 4.0 fax software. &C1 identify connection
&DO ignore DTR status change
When installing the fax software from the program disk, you S=3 fax answers after 3rd ring
can configure the fax port as the standard printer. With this \Q3 RTS-CTS handshake, bidirectional
setting, every print command is sent to the fax program.
Enter frequently used fax numbers in the “telephone direc- Sending the FAX
tory”.
Select the “high” printer resolution, the A4 format and the * Display the image to be printed.
landscape orientation. If the same image is transferred
* Initiate the printout.
between 2 PC’s more than twice, you must select the portrait
orientation after the second transfer. The program converts the ECG into an image file. Depend-
ing on the computer model, this procedure may take several
If the modem is reserved for communication with the fax, minutes. The printer status box indicates the page currently
there is no need for any special programming. being converted. The fax software is activated automatically
after the conversion.
However, if it is also used for transmission of data from the
* Select the addressee (entries in telephone directory).
ECG recorder to the PC, the init string should be modified
accordingly (menu, fax program setup, fax modem): * Select “Send”. The status window and the audible signals
from the modem inform you of the transfer procedure.
example for the ELSA modem
After the transmission, the program hangs up automatically.
Connecting the SunTech 4240 Sphygmomanometer Connecting the ERGOLINE 900/900L Ergometer
The sphygmomanometer needs to be configured once. Adjust the following settings at the ergometer:
* When the program screen is displayed, select “Change – EKG model P10
Test Parameters”.
– baud rate 4800
LAST NEXT
You move the cursor with the and keys. You
confirm entries with .
YES
The ergometer software must be version V436 or higher, or
version V236 (call in an ERGOLINE specialist to install the
* Confirm your selection with YES
. latest version).
OK
* Display the program screen again (press twice) and
select “Utilities”.
YES NO
* Select SUNTECH (with or ).
OK
* Display the program screen again (press twice).
7. Application Tips
When electrodes are not properly applied these measures may * If it becomes necessary to verify the unprocessed “raw”
not fully compensate for artifact. High polarization voltages ECG signal, switch off the Cubic Spline algorithm and
induced by electrodes applied without conductive gel may all filters (40 Hz, AC filter) (refer to “Setup” in section
cause the amplifier to overrange, so that a straight line will be 4.1.7).
recorded instead of the ECG. Cubic spline will return this
line to its normal position, and a baseline ensues for approx.
1 second (message Excess DC Offset).
It is not possible to display pacing pulses directly. At a sweep or The patient signal input is defibrillation-proof. Therefore it is
paper speed of 50 mm/s and a pulse duration of just 0.5 ms, the not necessary to remove the ECG electrodes during defibril-
width of the recorded pacing pulse would be a mere 0.025 mm. lation.
For this reason the device reduces the pulse amplitude and
expands the pulse width, so that the pacing pulse is easier to When using stainless steel or silver electrodes, however, the
identify. The pulse is displayed with the correct polarity and a defibrillator discharge current may cause complete polariza-
width of 5 ms (depending on the polarity of the pacing pulse tion at the electrode/skin interface. This will block acquisition
in leads I and II, the pacing pulse in lead III may be sup- of the ECG signal for several minutes. This can be avoided by
pressed). Figure 7-1 shows an ECG recording with pacing using silver/silver-chloride electrodes.
pulses.
If electrodes made of other materials are used, we recommend
disconnecting the patient cable from the recorder while the
shock is applied.
The ST/HR slope was originally developed by Dr. R.J. The large increments in heart rate between stages of the Bruce
Linden at the University of Leeds, United Kingdom in the protocol were found to yield an inadequate number of points
late 1970s. It is reported to yield more accurate electrocardio- for proper slope evaluation. A modification of the Bruce
graphic determination of the presence and severity of coro- protocol (half work loads in 2-minute exercise stages) more
nary heart disease. While originally applied only to bicycle closely approximate the roughly 10 beats per minute incre-
ergometry testing, it has recently been adapted for treadmill ments found in ergometry testing. This protocol, developed at
exercise by Dr. Paul Kligfield and colleagues at Cornell Cornell University by Okin, et al, is reproduced here and is
University. strongly recommended for the highest predictive accuracy.
EC Declaration of Conformity
Index
Symbole
3D presentation ................................................................. 71 CE marking ......................................................................... 2
A Chart-paper loading ........................................................... 17
AC line filter frequency .................................................... 220 Cleaning and disinfection
Accessories ....................................................................... 245 Instrument surface ..................................................... 233
Acitvate program options ................................................. 219 Patient cable and electrodes ........................................ 233
Admitting new patients ...................................................... 32 Respiration flow sensor LF 501 .................................. 234
Ambulatory blood pressure measurement ......................... 175 Clinical information system ............................................. 252
Arrhythmia episodes ........................................................ 117 Combined report ............................................................... 37
Arrhythmia results ............................................................. 64 Comments entry
Assigning records to patients ............................................ 226 during Exercise tests ..................................................... 89
B during Resting ECG .................................................... 51
Backup ............................................................................ 254 Comp. Median (button) .................................................. 106
Bar code wand ................................................................... 31 Comparing examinations ................................................... 68
Baud rate selection ................................................... 216, 227 Comparing tests
Blood pressure Spirometry ................................................................. 170
Graphic presentation (Amb. blood pressure) .............. 184 Compression of data ........................................................ 229
Hourly averages (diagram) .......................................... 185 Configuration of test protocols
Blood pressure readings Ambulatory blood pressure......................................... 180
Entry ............................................................................ 49 Confirming test results ................................. 35, 67, 118, 168
Entry during exercise test ..................................... 87, 103 Connecting peripheral devices ......................................... 258
Brief operating instructions Consumables ................................................................... 248
Ambulatory blood pressure......................................... 189 Copying procedures to/from storage media ...................... 224
Bronchodilation test ................................................... 173 Correlation coefficient ..................................................... 203
Exercise test ................................................................ 137 Country settings .............................................................. 220
Resting ECG ................................................................ 79 Cubic spline ............................................... 78, 126, 153, 240
Spirometry ................................................................. 173 Custom setup (save/load) ................................................. 231
STAT ECG ................................................................ 154 Cut-off frequency ............................................................ 203
Bronchodilation test ........................................................ 169 D
C Data backup .................................................................... 254
Calibration pulse ........................................ 76, 128, 150, 203 Data compression ............................................................ 229
Card reader ................................................................ 32, 220
Keyboard .................................. 23. See alos Function keypad Transmission .............................................................. 257
L Mouse ................................................................................ 20
Language selection ............................................................. 26 MUSE
Late potential analysis ...................................................... 195 Communication ......................................................... 218
Late potential mode ......................................................... 191 Selecting patients from MUSE database ....................... 31
Lead check (ECG signal qualiy screen) ............................. 103 Sending test data to ...................................................... 38
Lead check noise level ................................ 76, 126, 150, 203 N
Lead Off alarm .......................................... 76, 128, 150, 203 Name (physician, technician, user) .............. 51, 89, 161, 214
Lead sequence (selection, modification) NEHB ....................................................................... 46, 145
Exercise test ................................................................ 129 New patient (admission) .................................................... 32
Resting ECG ................................................................ 76 Normal value calculation (spirometry) ............................. 171
Lead sequence (selection/modification) O
STAT ECG ................................................................ 151 Option code (entry) ......................................................... 219
Linden, R.J. ............................................................. 113, 261 Optional software modules .................................................. 6
Linked medians ....................................................... 106, 126 Options, activate .............................................................. 219
M Order information ........................................................... 245
Maintenance .................................................................... 235 Order list ......................................................................... 253
Malfunctions ................................................................... 238 P
Measurement and analysis programs ................................ 256 Pacemaker patients .......................................................... 260
Measurement of the ECG .................................................. 57 Password (modification) ................................................... 215
Measurements result (tabular form) ................................... 61 Patient cable connection
Median complexes (dominant/subdominant) ..................... 61 Exercise test .................................................................. 83
Medians ........................................................................... 106 Resting ECG .......................................................... 43–46
Medical reimbursement program ..................................... 252 STAT ECG ................................................................ 143
Medication Patient data
Ambulatory blood pressure......................................... 182 Edit ........................................................................ 30, 31
Exercise test .................................................................. 88 Entering ....................................................................... 32
Late potentials ............................................................ 193 Entering (spirometry) ................................................. 163
Resting ECG ................................................................ 50 Patient deletion .................................................................. 30
Spirometry ................................................................. 160 Patient retrieval from database ........................................... 30
METS calculation ............................................................ 108 Patient selection window .................................................... 30
Modem Performance test ................................................................ 15
Setup .......................................................................... 216 Peripheral devices (connection) ........................................ 258
Physician. See Name Respiration flow sensor LF 501 ................................ 156, 234
Physician and user/technician name (entry) ....................... 51 Resting ECG mode ............................................................ 41
Printer driver installation ................................................... 18 Resting ECG, signal acquisition ......................................... 52
Printer problems ................................................................ 18 Retrieving patients ............................................................. 30
Printing the test report Retrieving the factory settings .......................................... 231
Ambulatory blood pressure ......................................... 188 Revision history ................................................................... 2
Exercise test ................................................................ 121 S
Late potentials ............................................................ 201 Safety information ............................................................. 10
Resting ECG ................................................................ 72 Sample cardiac cycles (exercise test) .................................. 114
Spirometry ................................................................. 172 Screen diagonal (adjustment) ........................................... 212
STAT ECG ................................................................ 149 Selecting stored tests .......................................................... 34
Procedure list, view ............................................................ 33 Self-test .............................................................................. 15
Program end .................................................................... 209 Sending procedures ............................................................ 36
Program start ..................................................................... 24 Setup
Q Ambulatory blood pressure ......................................... 178
QRS morphologies (exercise test) ..................................... 115 Exercise test ................................................................ 125
QTC calculation ........................................................ 76, 151 Late potentials ............................................................ 202
Quitting the program ...................................................... 209 Resting ECG ................................................................ 75
R Spirometry ................................................................. 171
Reanalysis .......................................................................... 62 STAT ECG ................................................................ 150
Reason for test Software options .................................................................. 6
Ambulatory blood pressure ......................................... 182 Sphygmomanometers ....................................................... 262
Late potentials ............................................................ 193 Spirometry
Reason for test library General ...................................................................... 155
Exercise test .................................................................. 86 On-screen demonstration ................................... 164–165
Late potentials ............................................................ 193 Running a test ............................................................ 164
Resting ECG ................................................................ 48 Spline. See Cubic spline
Spirometry ................................................................. 158 Square-wave profiles ......................................................... 132
Recording from ECG recorder ......................................... 226 ST profile, medians (exercise test) .................................... 106
Recording problems ......................................................... 259 ST/HR slope ............................................................ 113, 261
Registered system users .................................................... 214 Standard leads .................................................................... 44
Remote exercise test workstations, monitoring of ............. 134 Starting the program .......................................................... 24
Reports ............................................................ 104, 112, 122 STAT ECG mode ............................................................ 141
STAT ECG recording .............................................. 146, 148 TONOPORT IV/IVa ...................................................... 176
Statement library Trace background (color selection) ................................... 212
Exercise test ................................................................ 118 Transfer of resting ECGs .................................................. 226
Late potentials ............................................................ 200 Trends ...................................................................... 106, 112
Resting ECG ................................................................ 67 Troubleshooting ............................................................... 237
Spirometry ................................................................. 159 U
Statistics summary (Amb. blood pressure) ........................ 186 Units of measurement ...................................................... 220
Stored tests, view .......................................................... 33, 35 User list ........................................................................... 214
System configuration ....................................................... 212 V
System maintenance ........................................................ 254 VC measurement ............................................................. 165
System messages ............................................................... 237 Vector ECG ..................................................................... 196
System operating routines .................................................. 20 Voice recorder .................................................................. 221
T Z
Tabular summary (exercise test) ....................................... 111 Zoom display
Target HR .......................................................................... 87 Exercise test ................................................................ 120
Technical inspections ....................................................... 235 Resting ECG ................................................................ 56
Technical specifications .................................................... 239
Technician. See Name
Test information
Ambulatory blood pressure......................................... 182
Exercise test .................................................................. 85
Late potentials ............................................................ 193
Resting ECG ................................................................ 47
Spirometry ................................................................. 158
STAT ECG ................................................................ 147
Test patient ...................................................................... 213
Test protocol (selection, modification) ............................... 87
Test results (confirming) .................................................... 35
Test summary
Exercise test ................................................................ 109
Resting ECG ................................................................ 59
Thermal printer ................................................................. 16
Time format .................................................................... 220