You are on page 1of 2

Product Spe ci fi ca ti ons

Thermo Scientific
Indiko

Thermo Scientific Indiko is a fully


automated, sample oriented random
access analyzer for clinical and special
chemistries, including enzymes,
substrates, electrolytes, specific
proteins, therapeutic drug monitoring,
drugs of abuse and immunosuppressive
drug tests(*. Many different tests can
be run simultaneously from a single
sample. Indiko employs colorimetric
end-point and kinetic as well as
turbidimetric and bichromatic reactions
with or without sample blanking.
Optional ISE unit uses ion selective
electrodes for sodium, potassium and
chloride measurements.
Measurement ON-BOARD CAPACITY: max 6 racks -
Single channel interference filter photometer 9-position sample rack or 6-position reagent
with beam splitting reference, rack in the disk cooled about 8 °C below
12 filter positions. ambient temperature.
FILTER RANGE: 340 - 700 nm SAMPLE VOLUMES: 2 - 120 µL
MEASUREMENT TEMPERATURE: 37 °C SAMPLE CONTAINERS: 0.5 mL, 2.0 mL
LIGHT SOURCE: Flash lamp cups and sample tubes (diameter 12 - 16
ABSORBANCE RANGE: 0 - 3.5 A, resolution mm, length 75 - 100 mm).
of 0.001 A and reproducibility of SD <0.005 SAMPLE BARCODES: Code 128 and
A at 2 A barcodes USS Codabar, Interleaved 2 of 5
and Code 39 with a check digit
Reaction vessels
REAGENT VOLUMES: 2 - 240 µL
Discrete disposable cuvette. Continuous
REAGENT CONTAINERS: 10 mL and 20 mL
access to cuvettes without interrupting test
vials
processing.
SAMPLE AND REAGENT DISPENSING: CV
ON-BOARD CAPACITY: 360 measurement
≤2% for volumes ≥2 µl
cells - 36 cuvettes with 10 reaction cells,
up to 2 hours walk-away time.
REACTION END VOLUME: 120 - 300 µL
Application availability may vary from
(*
Samples and Reagents
country to country. Please contact your local
Continuous access to samples and reagents
sales representative for more information.
without interrupting test processing.
Automatic identification via internal barcode
reader. Clearly displayed real-time reagent
volume and remaining test capacity. Up to
four reagent additions per test.
Calibration Dilutions LIS INTERFACE: CLSI LIS02-A2 (former
Factor, bias, linear, logit-log, spline, Automatic sample pre-dilution. Automatic ASTM 1394-91)

Product Spe ci fi ca ti ons


polynomial and point-to-point calibration. dilution of over-range tests with automatic HARDWARE INTERFACE: RS-232 or TCP/IP
Method-dependent use of individual rerun. Addition of manual pre-dilution value RESULT REPORTS: Collated by sample,
calibrators or automatically diluted series for result calculation. manual entry of off-line results allowing
from a stock calibrator. Previous curve for fully collated result reports, results
Data Management calculated from both measured and off-line
comparison available.
Windows® workstation with graphical user
results. Abnormal values and repeats flagged
Quality Control interface. Data input online, via mouse,
automatically.
Real-time QC program with multiple, user- touchscreen, keyboard and barcode reader.
TRACEABILITY: Full traceability with long
definable Westgard rules. Control frequency Different user groups can have different
term storage of results including associated
user-definable. Out-of-specification control access rights. Different user interface
calibrations and reagent lot data.
results flagged. QC chart printouts, daily and language versions available.
cumulative reports.

Capacity: Up to 200 tests/hour depending on the workload.


Dimensions and Weight: 75 cm (width) x 70 cm (depth) x 62/130 cm (height/with open cover), 85 kg (weight). Separate workstation.
Power Requirements: 100 - 240 V ±10%, 50-60 Hz ±5%, 250 W
Deionized Water Consumption: 1.5 litres/ hour
Average Noise Level at 1 meter: <60 dB(A)
Environmental Conditions: Operating temperature range of 18 - 30 °C, humidity 40-80% (non-condensing)
Regulatory: Conforms to CAN/CSA-C22.2 No.61010-1-04
UL Std. No. 61010-1:2004
FCC 47 CFR Part 15, subpart B
FDA cleared, 510(k) number: k110035 (Indiko 9863 0000)
98/79/EC IVDMD Directive
Ordering Codes: 9863 0000 Indiko
9863 1000 Indiko with the ISE unit - not available in the USA

thermoscientific.com/indiko

© 2012 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. The information
and technical specifications are subject to change without notice.

Europe: Tel: +358 9 329 100 | email: sales.cdx.fi@thermofisher.com


North America: Tel: 1-510-979-5000 | email: sales.diagnostics.fmt@thermofisher.com
Australia: Tel: 800-333-110 | info@microgenics.com.au

Part of Thermo Fisher Scientific

05 / 2012 CODE: N11232_03