Anti-HBs

Test Setup Test Procedure

Test Components 1 Turn the power


switch ‘On’ 1 Draw the diluent from
diluent vial 100 µL 3 Draw 50 µL
(Serum / Plasma / Control) 5 Mix well by pipetting
10-20 times. 8 Insert the test cartridge

ID Chip
10-20
times

Sample
diluent vial 50µL
100µL
25 Tests
Anti-HBs

9
Diluent

Press ‘Select’
for in vitro diagnostic use only
store at 2~8℃
LOT : EXP :

Detection 2 Insert ID Chip


Buffer Tube (Diluent Vial)

Test Cartridge
100µL

10
Just right Too little or Bubble Just right Too little or Bubble
In the middle or near In the middle or near Read the test result
the top of the tip the top of the tip
6 Load the
sample Mixture
Reader Anti-HBs Result

2 Add it into
detection buffer tube 4 Add it into
detection buffer tube

3 Press ‘Select’ 150.00 mIU/mL (Pos)

100µL Result

100µL
50µL
Important Anti-HBs

Rev.00.170223_Anti-HBs
Allow ‘Detection Buffer Tube’ to attain
room temperature at least for
30 minutes before performing the test

Make sure that lot number of the

7
‘ID Chip’ is exactly same as the lot
number of ‘Test Cartridge’ and Waiting 15 minutes
Detection Buffer Tube’