You are on page 1of 11

The Journal of EVIDENCE-BASED DENTAL PRACTICE

ORIGINAL ARTICLE

ERBIUM LASER TECHNOLOGY VS


TRADITIONAL DRILLING FOR CARIES REMOVAL:
A SYSTEMATIC REVIEW WITH META-ANALYSIS

SIYING TAO, DDSa,b, LAN LI, MDSa,b, HE YUAN, DDS, PHDa,b, SIBEI TAO, MDc,
YIMING CHENG, MDc, LIBANG HE, DDS, PHDa,b, AND JIYAO LI, DDS, PHDa,b
a
State Key Laboratory of Oral Diseases, National Clinical Research Center for Oral Diseases, West China Hospital of Stomatology, Sichuan University, Chengdu,
China
b
Department of Operative Dentistry and Endodontics, West China Hospital of Stomatology, Sichuan University, Chengdu, China
c
Renal Division, Department of Internal Medicine, West China Hospital of Sichuan University, Chengdu, China

CORRESPONDING AUTHOR: ABSTRACT


Jiyao Li, Department of Operative
Objective
Dentistry and Endodontics, West
The study aimed to assess the efficacy of erbium laser technology compared with
China Hospital of Stomatology, No. 14,
traditional drilling for caries removal.
3rd section, Renmin Road South,
Chengdu 610041, Sichuan Province, Methods
China. A systematic search was conducted through Medline via PubMed, Embase,
E-mail: jiyaoliscu@163.com Cochrane databases, CNKI till December 2016. Randomised controlled trials,
quasi-randomized controlled trials, or controlled clinical trials with data
KEYWORDS comparing the efficacy of erbium laser technology versus traditional drilling for
Laser, Drilling, Caries caries removal were included.
Results
Fourteen studies were selected in our meta-analysis. Erbium laser technology
showed an increased time when removing caries compared with drilling (mean
Conflict of Interest: The authors have no difference: 3.48, 95% confidence interval: 1.90-5.06, P , .0001). However, erbium
actual or potential conflicts of interest. laser technology reduced the requirement for local anesthesia (risk ratio: 0.28,
95% confidence interval: 0.13-0.62, P 5 .002). Erbium laser technology was also
Source of Funding: This work was
not significantly different to traditional drilling with regard to restoration loss,
supported by the National Science &
pulpal vitality, and postoperative sensitivity.
Technology Pillar Program during
the 12th Five-year Plan Period of Conclusions
China (grant number Erbium laser technology showed an increased time for cavity preparation
20911BAZ03171); Science & compared with traditional drilling. However, erbium laser technology reduced the
Technology People-benefit requirement for local anesthesia. There was no significant difference between
Application Demonstration Project erbium laser technology and traditional drilling regarding restoration loss, pulpal
of Chengdu (grant number 2014- vitality, and postoperative sensitivity.
HM02-00041-SF).

Received 19 March 2017; revised 9 INTRODUCTION


May 2017; accepted 9 May 2017

J Evid Base Dent Pract 2017: [324-334] C aries is an infectious bacterial disease that results in the localized dissolution
and destruction of dental hard tissues. The most common methods for
removing infected hard tissues involve the use of excavators and burs, which may
1532-3382/$36.00
remove the largest amount of sound tissue (overpreparation) or leave residue of a
ª 2017 Elsevier Inc. All
certain amount of carious tissue (underpreparation). Moreover, the process is un-
rights reserved.
doi: http://dx.doi.org/10.1016/ comfortable because the pressure and heat on pulp, the vibration, noise, and pain
j.jebdp.2017.05.004 make patients anxious; therefore, it is usually necessary to use local anesthesia,

324 Volume 17, Number 4


The Journal of EVIDENCE-BASED DENTAL PRACTICE

especially for children. Presently, the concepts of minimally Literature Search


invasive treatment and preventive dentistry are gaining more We identified relevant articles by searching MEDLINE via
and more recognition by dentists.1 The minimally invasive PubMed, the Cochrane Library, EMBASE, and CNKI from
caries removal techniques include the use of air abrasion,2 the earliest available data indexing to December 2016. Both
new polymer prototype bur,3 chemomechanical agents,4 MeSH heading words and free-text words were included
and lasers.5 during the literature search. The search equations used in
each database are elaborated in Supplementary Material 3.
Since the early days of laser use in dentistry by Goldman Reference lists of relevant studies were examined manually
et al.,6 there are different types of lasers used to remove to identify additional eligible articles. The search
caries, such as ruby6; CO27; neodymium-doped yttrium strategies were confirmed by experts, and all relevant
aluminum-garnet (Nd:YAG)8; argon fluoride (ArF) excimer9; articles were scrutinized.
erbium, chromium, yttrium-scandium-gallium-garnet (Er,
Cr:YSGG)10; and erbium yttrium aluminum-garnet (Er:YAG) Selection Criteria
lasers.11 The most popular one is the Er:YAG laser, which Types of studies
was authorized for use on human teeth by the Food and Data used in this meta-analysis were all from randomized
Drug Administration in 1997. The Er:YAG laser wavelength controlled trials (RCTs), quasi-RCTs, or controlled clinical
(2940 nm) coincides with the peak of water absorption trials of parallel design that compared the effects of erbium
and hydroxyl radicals of hydroxyapatite. Effective ablation laser technology with traditional drilling/rotary instruments
of the carious tissue occurs via microexplosions from the for caries removal.
evaporation of the water contained in the mineralized
tissue.12-14 The efficacy, efficiency, and patient comfort Types of Participants
level are the most important factors considered when This meta-analysis selected studies that involved individuals
patients and dentists choose a method to remove dental with caries lesions. Individuals with symptoms/evidence of
caries. pulpal or periapical lesions were excluded. Individuals with
systemic diseases were also excluded.
Currently, there are many published clinical trials comparing
the use of the erbium laser with the drill for caries removal.
Types of Intervention
However, the efficacy of the erbium laser is still controver-
Erbium laser equipment vs traditional drilling for caries
sial. Some studies have shown a better outcome after laser
removal was selected in our meta-analysis. Both Er:YAG and
treatment,15 and some have found no significant difference
Er, Cr:YSGG laser systems were included because they have
with regard to the clinical success between laser and drill.16
similar working principles. Both of their wavelengths coin-
For the pain experienced during caries removal, some
cide with the maximum absorption by water molecules and
dentists have reported that the laser was more
hydroxyl group in enamel and dentin.
comfortable and reduced the use of local
anesthesia,15,17,18 whereas others have demonstrated that
Outcome Measures
the laser had contributed little in reducing patients’ pain
Outcomes were evaluated during treatment or at follow-up
compared with conventional drilling.19
visits as follows: (1) time for cavity preparation; (2) local
Therefore, we aim to undertake a systematic review and anesthesia requirement; (3) subjective acceptance by pa-
meta-analysis of the published literature to determine tients; (4) restoration loss; (5) pulpal vitality; and (6) post-
whether erbium laser, compared with traditional drilling, can operative sensitivity.
improve clinical success and reduce the use of local anes-
thesia. The treatment time of these 2 methods was esti- Data Extraction and Quality Assessment
mated concurrent to our other investigation. The study Two authors of this article (S.T. and L.L.) conducted a liter-
protocols are shown in Supplementary Material 1. ature search, selected studies, and conducted data extrac-
tion independently. On comparison of the 2 reviewers’
results, consensus was reached through discussion or by
MATERIALS AND METHODS consulting a third reviewer. Titles and abstracts of all rele-
This meta-analysis was performed according to the vant studies were examined for the inclusion in our meta-
Preferred Reporting Items of Systematic Reviews and Meta- analysis, and full-text articles were also inspected, if neces-
Analyses (PRISMA) items.20 The PRISMA is a concise sary. To extract data from each of the published studies
checklist consisting of 27 items deemed essential for included in the review, a data extraction form was used.
reporting a clear and complete systematic review. The Study characteristics, patient characteristics, and treatment
PRISMA checklist of this meta-analysis can be seen in outcomes were extracted from each study included in our
Supplementary Material 2. review. The study characteristics included the first author,

December 2017 325


The Journal of EVIDENCE-BASED DENTAL PRACTICE

publication date, countries where the studies were carried outcome assessment, incomplete outcome data, selective
out, sample size, follow-up period, type of intervention, and reporting, and other possible sources of bias (Table 1).
type of control. The patient characteristics included clinical
features (location of the lesions) and demographic features Statistical Analysis
of the individuals at baseline. We used inverse variance weighted random-effects analysis
The Cochrane Collaboration methodology21 was used to with 95% confidence intervals (CIs) to estimate association
assess the risk of bias of each study included in this meta- for all the studies included. The random-effects models
analysis. The criteria were random sequence generation, were used to measure the effect sizes because there was
allocation concealment, blinding of participants, blinding of heterogeneity in the treatment efficacy.22 Random-effects

Table 1. Ascertainment of the risk of bias in the included studies.

Random Blinding of Blinding of Other


sequence Allocation participants and outcome Incomplete Selective sources of Risk of
Study generation concealment personnel assessment outcome data reporting bias bias

Zhegova, ? ? 2 2 2 2 2 Unclear
201516

Belcheva, 2 ? ? 2 ? 2 2 Unclear
201432

Zhang, 2 ? 2 2 ? 2 2 Unclear
201317

Eren, ? ? 2 2 ? 2 2 Unclear
201326

Bohari, 2 1 ? ? 2 2 2 High
201224

Yazici, ? ? 2 2 2 2 2 Unclear
201027

Dommisch, ? ? ? 2 2 2 2 Unclear
200828

Liu, 200629 ? 2 2 2 ? 2 2 Unclear

DenBesten, ? ? ? 2 2 2 2 Unclear
200119

Hadley, ? ? 2 2 2 2 2 Unclear
200030

Evans, ? ? 2 ? 2 2 2 Unclear
200031

Keller, ? ? ? 2 2 2 2 Unclear
199833

Pelagalli, 2 2 2 2 2 2 2 Low
199725

He, 201315 2 2 2 2 2 2 2 Low

1, this item would increase the risk of bias; 2 means this item would decrease the risk of bias; ?, this item was not clearly reported in the study so that it
could not be assessed accurately.

326 Volume 17, Number 4


The Journal of EVIDENCE-BASED DENTAL PRACTICE

meta-analysis was performed to acquire estimates of out- for inclusion, and 14 studies were finally included in our
comes, and we presented the outcomes as mean differ- meta-analysis (Figure 1). Main characteristics of these 14
ences (continuous outcomes, including time for cavity studies can be seen in Table 2. Table 1 shows the risk of
preparation) or risk ratios (RRs, dichotomous outcomes, bias of each study included. One24 of these 14 studies
including local anesthesia requirement, restoration loss, were at a high risk of bias, 215,25 studies were at a low risk
pulpal vitality, and postoperative sensitivity) with 95% of bias, and the other 11 studies16-19,26-32 were at an
CIs. We assessed heterogeneity using I2 statistical index, unclear risk of bias according to the Cochrane
and I2 . 50% represented a high heterogeneity.23 P , .05 Collaboration’s assessment.
was regarded statistically significant, apart from the
heterogeneity test where P , .1 was considered Effectiveness of Treatment
statistically significant. All the analyses were conducted Time for Cavity Preparation (minutes)
using RevMan statistical software, version 5.3, provided by Six15,17,18,24,28,29 of the 14 studies reported time for cavity
the Cochrane Collaboration. preparation. All the 6 studies reported a shorter preparation
time using conventional rotary instruments than erbium
Sensitivity Analysis laser equipment. Heterogeneity of the 6 studies was indi-
To assess the robustness of our meta-analysis results, cated to be significant (P , .01, I2 5 98%). Meta-analysis
sensitivity analysis was performed: (1) high-quality studies manifested a significantly shorter time for cavity pre-
versus vs low-quality studies and (2) studies with small paration using conventional rotary instruments than erbium
sample size vs studies with large sample size. laser equipment (mean difference: 3.48, 95% CI: 1.90-5.06,
P , .0001; Figure 2).
RESULTS
Local Anesthesia Requirement
Search Results and Study Characteristics Four studies15,17-19 compared patients’ requirement of local
We identified 536 studies in total, among which 83 were anesthesia using 2 kinds of treatment. No evidence of
duplicates. Four hundred fifty-three studies were assessed significant heterogeneity was found within these 4 studies

Figure 1. Flow diagram of the literature search process. CCTs, controlled clinical trials; RCTs, randomized controlled
trials.

December 2017 327


328
Volume 17, Number 4

The Journal of EVIDENCE-BASED DENTAL PRACTICE


Table 2. Main characteristics of included trials.

No. of Preparation Anesthesia Pulpal Postoperative


First author teeth Age time (min) requirement Restoration vitality sensitivity
(year, country/ (laser/ range Male Follow- Location of Intervention Control (laser/ (laser/ loss (laser/ (laser/ (laser/
area) control) (mean) (%) up lesions group group control) control) control) control) control)

Zhegova (2015, 58/58 6-16 y 42 2y cer, occ, pro Er: YAG laser bur n.r n.r 0/0 n.r 0/0
Bulgaria)16 (n.r)

Belcheva (2014, 45/45 6-12 y n.r – occ, pro Er:YAG laser bur n.r n.r n.r n.r n.r
Bulgaria)32 (7.42 y)

Zhang (2013, 60/60 3-15 y 39.60 3 mo n.r Er:YAG laser bur 6.4/3.9 1/4 2/1 n.r n.r
China)17 (8.6 y)

Eren (2013, 10/10 7-12 y 60 – occ Er,Cr:YSGG bur n.r n.r n.r n.r n.r
Turkey)26 (n.r) laser

Bohari (2012, 30/30 5-9 y n.r – occ Er:YAG laser bur 5.998/3.445 n.r n.r n.r n.r
India)24 (n.r)

Yazici (2010, 54/54 19-21 y 22.20 2y occ Er,Cr:YSGG bur n.r n.r 0/1 n.r 0/0
Turkey)27 (n.r) laser

Dommisch 102/102 22-56 y 53.80 – n.r Er:YAG laser bur 3/1.02 n.r n.r n.r n.r
(2008, (n.r)
Germany)28

Liu (2006, 40/40 3-12 y 50 – pro Er:YAG laser bur 2/1 n.r n.r n.r n.r
Taiwan)29 (6 y)

DenBesten 82/42 4-18 y 53 3 mo n.r Er:YAG laser bur n.r 4/11 1/0 1/1 1/1
(2001, USA)19 (10.4 y)

Hadley 75/75 20-84 y n.r 6 mo occ, buc, Er,Cr:YSGG bur n.r n.r 0/0 0/0 n.r
(2000, USA)30 (n.r) lin, pro laser

Evans 77/77 3.5-68 y n.r – n.r Er:YAG laser bur n.r n.r n.r n.r n.r
(2000, UK)31 (n.r)

Keller (1998, 97/97 18-72 y 47 – n.r Er:YAG laser bur 7.5/4.3 6/11 n.r 0/1 n.r
Germany)33 (32.8 y)
The Journal of EVIDENCE-BASED DENTAL PRACTICE

(P 5 .30, I2 5 18%). A significant difference between the 2

buc, buccal; cer, cervical; Er: YAG laser, erbium: yttrium-aluminum-garnet laser, wavelength: 2.94 mm; Er, Cr: YSGG laser, erbium, chrome: yttrium-scandium-gallium-garnet laser, wavelength: 2.79 mm; lin,
kinds of treatment was noted. Fewer persons in the laser
2/2 group experienced pain during cavity preparation and
n.r

asked for the use of local anesthesia (RR: 0.28, 95% CI:
0.13-0.62, P 5 .002; Figure 3).

Subjective Acceptance by Patients


Two trials31,32 investigated subjective acceptance by pa-
1/1

0/0

tients of these 2 kinds of treatments. Belcheva et al.


demonstrated that erbium YAG laser had a better level of
subjective acceptance among patients compared with
conventional rotary instruments for caries removal. Vibra-
0/1
n.r

tion, sight, and sound, which were the most common


annoyance factors for patients during conventional cavity
preparation, were eliminated using laser technology. How-
ever, the study by Belcheva et al. reported that smell and
taste during laser preparation were increased compared
n.r

0/8

with drilling (prevalence of vibration in bur/laser group:


86.7%, 2.2%; prevalence of sight in bur/laser group: 40%,
20%; prevalence of sound in bur/laser group: 62.2%, 15.6%;
prevalence of smell in bur/laser group: 17.8%, 66.7%; and
14.7/4.7

prevalence of taste in bur/laser group: 22.2%, 42.2%). Evans


n.r

et al. also reported that laser technology showed a better


level of subjective acceptance compared with traditional
drilling among patients older than 10 years. Significantly
bur

bur

less vibration was felt during laser preparation than with


drilling, but treatment time was significantly longer in the
laser group. Smell, taste, and sound showed no significant
Er:YAG laser

Er:YAG laser

difference between the 2 groups in this study. Meta-analysis


could not be performed comparing subjective acceptance
of patients in 2 kinds of treatment because data available in
both studies were limited.
occ,buc,lin,
pro,cer

Restoration Loss
n.r

Six15-17,19,27,30 of the included 14 studies reported restoration


loss. Follow-up time in the studies of DenBesten et al.,19
Hadley et al.,30 He et al.,15 Yazici et al.,27 Zhang et al.,17
1y

1y

and Zhegova et al.16 was 3 months, 6 months, 1 year,


2 years, 3 months, and 2 years, respectively (Table 2).
34.20

lingual; n.r, not reported; occ, occlusal; pro, proximal.

Heterogeneity could not be observed among the 6 studies


37

(P 5 .73, I2 5 0%). There was no significant difference


between the 2 treatment groups considering restoration
12-60 y

18-60 y
(33.6 y)
(29 y)

loss (RR: 0.90, 95% CI: 0.21–3.84, P 5 .89; Figure 4).

Pulpal Vitality
Five trials15,19,25,30,33 assessed pulpal vitality. Follow-up time in
45/61

60/60

the studies of DenBesten et al.,19 Hadley et al.,30 He et al.,15


and Pelagall et al.25 was 3 months, 6 months, 1 year, and
1 year, respectively. Keller et al.33 tested pupal vitality
Pelagalli (1997,

immediately after the treatment (Table 2). Heterogeneity of


He (2013,

these 5 studies was not significant (P 5 .97, I2 5 0%). Meta-


China)15
USA)25

analysis did not show a statistically significant difference be-


tween the 2 kinds of treatment comparing pulpal vitality (RR:
0.46, 95% CI: 0.09-2.40, P 5 .35; Figure 5).

December 2017 329


The Journal of EVIDENCE-BASED DENTAL PRACTICE

Figure 2. Time for cavity preparation (minutes) in erbium laser treatment vs drilling treatment. CI, confidence interval;
SD, standard deviation.

Postoperative Sensitivity cavity preparation. Erbium laser was first introduced in


Four studies15,16,19,27 evaluated postoperative sensitivity. restorative dentistry in 1989.33,34 After more than 2 decades
Follow-up time in the studies of DenBesten et al.,19 He of research, it is now widely applied clinically. In vitro and
et al.,15 Yazici et al.,27 and Zhegova et al.16 was 3 months, in vivo studies have shown mixed results and conclusions
1 year, 2 years, and 2 years, respectively (Table 2). No about the effects of erbium laser on caries removal.35-40
significant heterogeneity was found (P 5 .7, I2 5 0%). Although there are still limited data to make a conclusion
Meta-analysis demonstrated no significant difference be- at this moment in regard to subjective acceptance by pa-
tween the 2 treatment groups with regard to postoperative tients, our meta-analysis concluded that erbium laser
sensitivity (RR: 0.80, 95% CI: 0.17-3.88, P 5 .78; Figure 6). equipment can reduce the use of local anesthesia during
caries removal. Therefore, it is more comfortable for patients
Sensitivity Analysis compared with dental drilling objectively speaking. How-
We conducted sensitivity analysis for studies with large ever, it requires a longer cavity preparation time. The effi-
sample size and low risk of bias. When eliminating trials with cacy from follow-up evaluations, including postoperative
small sample size or high risk of bias, all results were sensitivity, pulpal vitality, and restoration loss, showed no
consistent to the meta-analysis results using all 14 studies. significant difference between these 2 methods. Several
other systematic reviews/meta-analyses41,42 have been
performed about laser technique in caries removal. They
DISCUSSION supported the use of laser as an alternative option for caries
Various methods for caries removal have been used in removal in pediatric dentistry, due to minimal invasion,
clinical practice in the hope of compensating for the nega- lower risk of pulp infection, and more cooperation from
tive aspects of traditional bur, such as pain, vibration, noise, children. Other methods to remove caries were also re-
and risk of accidental damage to soft tissues or pulp during ported. A meta-analysis43 suggested that Carisolv, a chemo-

Figure 3. Local anesthesia requirement in erbium laser treatment vs drilling treatment. CI, confidence interval; M-H,
MantelHaensze.

330 Volume 17, Number 4


The Journal of EVIDENCE-BASED DENTAL PRACTICE

Figure 4. Restoration loss in erbium laser treatment vs drilling treatment. CI, confidence interval; M-H,
MantelHaensze.

mechanical caries removal system, was an effective alter- limitations. This review was limited by both the data quantity
native. Ultrasonic instrumentation was found to be favorable and data type available till December 2016, like all other
with regard to the reduced vibration and noise compared meta-analyses. We converted time units in 2 studies24,29 to
with dental drilling.44 Other kinds of laser technology, such make meta-analysis possible, which was a major risk source
as Nd:YAG laser, were also demonstrated to have that might lower the quantity of evidence. The effects of
advantages, including minimal cellular destruction.45 other methods of caries removal (eg, Carisolv and air
abrasion) or other kinds of laser technology (eg, Nd:YAG
Our review selected 14 studies comparing 2 methods of laser) were not discussed.
caries removal applied in 1646 teeth with dental caries. This
sample size was sufficient to obtain robust conclusions. Cavity preparation time was longer using erbium laser
Table 2 (main characteristics of included studies) and technology than using rotary instruments according to our
Table 1 (risk of bias for each study) show the quality of review. The high-speed bur ablation speed is approximately
evidence in our meta-analysis. The sensitivity analysis out- 10 times of that of the laser, and in dentin, the laser ablation
comes indicated that our meta-analysis results were of high speed is approximately the same as a slow rotating bur,18
quality. We tried to avoid any possible source of bias by which can explain our results. Nevertheless, the outcome
following a rigid protocol. Two review authors screened of time for preparation indicated high heterogeneity.
studies and extracted data independently, and any Many factors can affect the time for caries removal, such
disagreement was resolved either through discussion or by as patients’ age, carious site, carious area, carious stage,
consulting a third assessor. However, there are still some and different energy settings for the laser equipment. We

Figure 5. Pulpal vitality in erbium laser treatment vs drilling treatment. CI, confidence interval; M-H, MantelHaensze.

December 2017 331


The Journal of EVIDENCE-BASED DENTAL PRACTICE

Figure 6. Postoperative sensitivity in erbium laser treatment vs drilling treatment. CI, confidence interval; M-H,
MantelHaensze.

tried to conduct subgroup analyses according to different bonding performance of adhesives can be improved in the
factors, but high heterogeneity still existed, which long run using erbium laser. No significant difference could
confirmed that there was more than 1 factor affecting be detected between the 2 methods regarding post-
treatment time in each study included. The included operative sensitivity and pulpal vitality in our meta-analysis,
studies recruited both children and adults, but treatment indicating that erbium laser may not do more harm to pulpal
time for children is usually longer than for adults because tissues than conventional drilling during caries removal. This
of their limited self-control abilities and lower tolerance of finding corresponded with conclusions of 2 previous animal
pain. The carious site, area, and stage are also crucial pa- experiments.10,51 The power of the meta-analyses is limited
rameters for caries removal time. Different energy settings by the quantity of qualified studies; therefore, more future
for laser equipment were suggested to cause different studies are still needed.
treatment time.18,46 These are all possible sources of high
heterogeneity in our outcome of time for cavity preparation. Erbium laser technology is a safe and effective carious
removal method in clinical practice. According to our find-
Pain when removing caries is a major complaint and a major ings, it can significantly reduce pain but requires a longer
factor leading to fear of dental procedures, especially for treatment time. Dentists are encouraged to consider using
children and patients with deep caries lesions. Fewer pa- laser technology to remove caries, especially when treating
tients required local anesthesia in the laser group compared children, patients with deep caries, and patients with anxi-
with the traditional drilling group according to our out- ety. However, we still look forward to more well-designed
comes, which suggested that laser technology can signifi- studies to confirm our results.
cantly reduce pain during caries removal, thus increasing
patients’ subjective acceptance of treatment.31,32,47 Tradi-
tional drilling treatment removes caries through bur rota- CONCLUSIONS
tion, which generates a lot of heat and vibration. These can Erbium laser technology makes fewer patients suffer from
act on the nerve fibers of dentin to cause pain. But most of pain during caries removal compared with traditional dril-
the energy of the erbium laser is absorbed by water and ling, making it preferred by patients. However, a longer
converted to kinetic energy, reducing pain. Moreover, it treatment time is required for this method. There was no
possesses shallow force of penetration due to its noncontact significant difference between these 2 caries removal
caries removal mode and high biocompatibility because methods considering postoperative sensitivity, pulpal vital-
water is used as the energy mediator.48 ity, and restoration loss during follow-up. Further well-
designed, large-sample, long-term RCTs are needed to
There were no significant differences in our results between strengthen our conclusions.
the 2 methods with regard to retention rates of restoration,
suggesting that erbium laser may not negatively affect the
bonding performance of adhesives to dentin or enamel, SUPPLEMENTARY DATA
which was confirmed in 2 previous in vitro studies.49,50 Supplementary data related to this article can be found at
Future long-term trials are needed to explore whether the http://dx.doi.org/10.1016/j.jebdp.2017.05.004.

332 Volume 17, Number 4


The Journal of EVIDENCE-BASED DENTAL PRACTICE

REFERENCES 17. Zhang S, Chen T, Ge LH. [Evaluation of clinical outcomes for Er:
YAG laser application in caries therapy of children] (Chinese).
1. Murdoch-Kinch CA, Mclean ME. Minimally invasive dentistry.
J Peking Univ Health Sci 2013;45:87-91.
J Am Dent Assoc 2003;134(1):87-95.
18. Keller U, Hibst R, Geurtsen W, et al. Erbium:YAG laser appli-
2. Rainey JT. Air abrasion: an emerging standard of care in con-
cation in caries therapy. Evaluation of patient perception and
servative operative dentistry. Dent Clin North Am 2002;46(2):
acceptance. J Dent 1998;26(8):649-56.
185-209.
19. DenBesten PK, White JM, Pelino JE, Furnish G, Silveira A,
3. Meller C, Welk A, Zeligowski T, Splieth C. Comparison of dentin
Parkins FM. The safety and effectiveness of an Er: YAG laser for
caries excavation with polymer and conventional tungsten car-
caries removal and cavity preparation in children. Med Laser
bide burs. Quintessence Int 2007;38(7):565-9.
Appl 2001;16(3):215-22.
4. Maragakis GM, Hahn P, Hellwig E. Chemomechanical caries
20. Liberati A, Altman DG, Tetzlaff J, et al. The PRISMA statement
removal: a comprehensive review of the literature. Int Dental J
for reporting systematic reviews and meta-analyses of studies
2001;51(4):291-9.
that evaluate health care interventions: explanation and elab-
5. Kornblit R, Trapani D, Bossù M, Muller-Bolla M, Rocca J, oration. Ann Intern Med 2009;151:W-65-94.
Polimeni A. The use of Erbium: YAG laser for caries removal in
21. Green S. Cochrane handbook for Systematic Reviews of In-
paediatric patients following Minimally Invasive Dentistry con-
terventions Version 5.1.0 [updated March 2011]. New Jersey,
cepts. Eur J Paediatr Dent 2008;9(2):81-7.
USA: Wiley-Blackwel: The Cochrane Collaboration; 2011.
6. Goldman L, Hornby P, Meyer R, Goldman B. Impact of the laser
22. Higgins JP, Thompson SG, Deeks JJ, Altman DG. Measuring
on dental caries. Nature 1964;203(4943):417.
inconsistency in meta-analyses. BMJ 2003;327:557-60.
7. Pogrel M, Muff D, Marshall G. Structural changes in dental
23. Huedo-Medina TB, Sánchez-Meca J, Marín-Martínez F,
enamel induced by high energy continuous wave carbon di-
Botella J. Assessing heterogeneity in meta-analysis: Q statistic
oxide laser. Lasers Surg Med 1993;13(1):89-96.
or I2 index? Psychol Methods 2006;11(2):193.
8. Hess JA. Subsurface morphologic changes of ND: YAG laser-
24. Bohari MR, Chunawalla YK, Ahmed BMN. Clinical evaluation of
etched enamel. Lasers Surg Med 1997;21(2):193-7.
caries removal in primary teeth using conventional, chemo-
9. Palamara J, Phakey P, Orams H, Rachinger W. The effect on the mechanical and laser technique: an in vivo study. J Contemp
ultrastructure of dental enamel of excimer-dye, argon-ion and Dent Pract 2012;13(1):40-7.
CO2 lasers. Scanning Microsc 1992;6(4):1061-70.
25. Pelagalli J, Gimbel CB, Hansen RT, Swett A. Investigational
10. Rizoiu I, Kohanghadosh F, Kimmel AI, Eversole LR. Pulpal study of the use of Er:YAG laser versus dental drill for caries
thermal responses to an erbium, chromium: YSGG pulsed laser removal and cavity preparation–phase I. J Clin Laser Med Sur
hydrokinetic system. Oral Surg Oral Med Oral Pathol Oral 1997;15(3):109-15.
Radiol Endod 1998;86(2):220-3.
26. Eren F, Altinok B, Ertugal F, Tanboga I. The effect of erbium,
11. Cozean C, Arcoria CJ, Pelagalli J, Powell GL. Dentistry for the chromium:yttrium-scandium-gallium-garnet (Er,Cr:YSGG) laser
21st century? Erbium: YAG laser for teeth. J Am Dent Assoc therapy on pain during cavity preparation in paediatric dental
1997;128(8):1080-7. patients: a pilot study. Oral Health Dent Manag 2013;12:80-4.

12. Bolortuya G, Ebihara A, Ichinose S, et al. Effects of dentin sur- 27. Yazici AR, Baseren M, Gorucu J. Clinical comparison of bur- and
face modifications treated with Er: YAG and Nd: YAG laser laser-prepared minimally invasive occlusal resin composite
irradiation on fibroblast cell adhesion. Photomed Laser Surg restorations: two-year follow-up. Oper Dent 2010;35(5):500-7.
2012;30(2):63-70.
28. Dommisch H, Peus K, Kneist S, et al. Fluorescence-controlled
13. Rodríguez-Vilchis LE, Contreras-Bulnes R, Sánchez-Flores I, Er:YAG laser for caries removal in permanent teeth: a ran-
Samano EC. Acid resistance and structural changes of human domized clinical trial. Eur J Oral Sci 2008;116(2):170-6.
dental enamel treated with Er: YAG laser. Photomed Laser Surg
29. Liu JF, Lai YL, Shu WY, Lee SY. Acceptance and efficiency of Er:
2010;28(2):207-11.
YAG laser for cavity preparation in children. Photomed Laser
14. Omae M, Shinnou Y, Tanaka K, et al. XPS analysis of the dentin Surg 2006;24(4):489-93.
irradiated by Er: YAG laser. Dent Mater J 2009;28(4):471-6.
30. Hadley J, Young DA, Eversole LR, Gornbein JA. A laser-pow-
15. He HY. The Clinical Comparative Study of Er: YAG Laser and ered hydrokinetic system for caries removal and cavity prepa-
Traditional Drilling to Treat Dental Caries. Chinese: [D]: Medical ration. J Am Dent Assoc 2000;131(6):777-85.
School of Nanchang University; 2013.
31. Evans D, Matthews S, Pitts N, Longbottom C, Nugent Z.
16. Zhegova GG, Rashkova MR. Er-YAG laser and dental caries Restorative dentistry: a clinical evaluation of an Erbium: YAG
treatment of permanent teeth in childhood. J IMAB 2015;21(1): laser for dental cavity preparation. Br Dent J 2000;188(12):
699-704. 677-9.

December 2017 333


The Journal of EVIDENCE-BASED DENTAL PRACTICE

32. Belcheva A, Shindova M. Subjective acceptance of pediatric 41. Nazemisalman B, Farsadeghi M, Sokhansanj M. Types of lasers
patients during cavity preparation with er:YAG laser and con- and their applications in pediatric dentistry[j]. J Lasers Med Sci
ventional rotary instruments. J IMAB 2014;1(20):631-3. 2015;6(3):96-101.

33. Keller U, Hibst R. Experimental studies of the application of the 42. Montedori A, Abraha I, Orso M, et al. Lasers for caries removal
Er: YAG laser on dental hard substances: II. Light microscopic in deciduous and permanent teeth[J]. Cochrane Database Syst
and SEM investigations. Laser Surg Med 1989;9(4):345-51. Rev 2016;9(9):Cd010229.

34. Hibst R, Keller U. Experimental studies of the application of the 43. Li R, Zhao Y, Ye L. How to make choice of the carious removal
Er: YAG laser on dental hard substances: I. Measurement of the methods, Carisolv or traditional drilling? A meta-analysis. J Oral
ablation rate. Laser Surg Med 1989;9(4):338-44. Rehabil 2014;41(6):432-42.

44. Oman CR, Applebaum E. Ultrasonic cavity preparation II.


35. Olivi G, Genovese M. Laser restorative dentistry in children and
Progress report. J Am Dent Assoc 1955;50(4):414-7.
adolescents. Eur Arch Paediatr Dent 2011;12(2):68-78.
45. Myers T, McDaniel J. The pulsed Nd: YAG dental laser: review
36. Lepri TP, Souza-Gabriel AE, Atoui JA, Palma-Dibb RG,
of clinical applications. J Calif Dent Assoc 1991;19(11):25-30.
Pecora JD, Milori Corona SA. Shear bond strength of a sealant
to contaminated-enamel surface: influence of erbium : yttrium- 46. Raucci-Neto W, Chinelatti MA, Ito IY, Pécora JD, Palma-
aluminum-garnet laser pretreatment. J Esthet Restor Dent Dibb RG. Influence of Er: YAG laser frequency on dentin caries
2008;20(6):386-92. removal capacity. Microsc Res Tech 2011;74(3):281-6.

37. Borsatto MC, Corona SA, de Araujo FP, de Souza-Gabriel AE, 47. Mosskull Hjertton P, Bågesund M. Er: YAG laser or high-speed
Pecora JD, Palma-Dibb RG. Effect of Er:YAG laser on tensile bur for cavity preparation in adolescents. Acta Odontol Scand
bond strength of sealants in primary teeth. J Dent Child (chic) 2013;71(34):610-5.
2007;74:104-8.
48. Matsumoto K, Hossain M, Hossain MI, Kawano H, Kimura Y.
38. Borsatto MC, Corona SA, Chinelatti MA, et al. Comparison of Clinical assessment of Er, Cr: YSGG laser application for cavity
marginal microleakage of flowable composite restorations in preparation. J Clin Laser Med Sur 2002;20(1):17-21.
primary molars prepared by high-speed carbide bur, Er:YAG 49. Ergücü Z, Celik E, Ünlü N, Türkün M, Özer F. Effect of Er, Cr:
laser, and air abrasion. J Dent Child (Chic) 2006;73(2):122-6. YSGG laser on the microtensile bond strength of two different
adhesives to the sound and caries-affected dentin. Oper Dent
39. Kohara EK, Hossain M, Kimura Y, Matsumoto K, Inoue M,
2009;34(4):460-6.
Sasa R. Morphological and microleakage studies of the cavities
prepared by Er:YAG laser irradiation in primary teeth. J Clin 50. Chou JC, Chen CC, Ding SJ. Effect of Er, Cr: YSGG laser pa-
Laser Med Surg 2002;20(3):141-7. rameters on shear bond strength and microstructure of dentine.
Photomed Laser Surg 2009;27(3):481-6.
40. Dostalova T, Jelinkova H, Krejsa O, et al. Dentin and pulp
response to Erbium:YAG laser ablation: a preliminary evalu- 51. Eversole L, Rizoiu I. Pulpal response to cavity preparation by an
ation of human teeth. J Clin Laser Med Surg 1997;15(3): erbium, chromium: YSGG laser-powered hydrokinetic system.
117-21. J Am Dent Assoc 1997;128(8):1099-106.

334 Volume 17, Number 4

You might also like