Technical Data

BSM-3532/3552/3562
BSM-3733/3753/3763
BSM-3532-Q01, BSM-3552-Q01, BSM-3562-Q01
BSM-3733-Q01, BSM-3753-Q01, BSM-3763-Q01

Bedside Monitor

TD. BSM-3000_F

This technical data may be revised or replaced by Nihon Kohden at any time without notice.
Some products may not be available in your country.
Upgraded arrhythmia detection
FEATURES Upgraded arrhythmia detection allows you to
The BSM-3000 bedside monitor is intended for analyze multi-channel ECG. And the special
use for monitoring, displaying, reviewing, software ec1 with the multi-template matching
storing and alarming of multiple physiological method reduces false alarms and assures high
parameters for single adult, pediatr ic and quality detection. Up to 23 itemes of arrhythmia can
neonatal patients (some of the parameters are be detected. Up to16,384 arrhythmia episodes can
specific for a category of patients e.g. OCRG be stored.
for neonatal use, esCCO for adult/pediatric use,
and so on.). The monitors are to be used in Multigas measurement
healthcare facilities by clinical physicians or You can measure CO 2 , N 2 O, O 2 , and anesthetic
appropriate medical staff under the direction of agents (Halothane, Isoflurane, Enflurane,
physicians. Sevoflurane, Desflurane). Gases can be
measured quickly with small sampling flow rate.
Smart cable concept (MULTI connectors) (Requires optional GF-110/120PA or
Smart cable technology give you high flexibility GF-210/220R) The MAC value is displayed.
in the selection of parameters without
traditional modules. When you plug a sensor 12 lead ECG simultaneous display using 10
cable into a MULTI connector, the monitor electrodes
automatically detects the type of parameter 10 electrodes provide the same quality
and starts measuring. (2 MULTI connectors for simultaneous 12 lead ECG as a multi-channel
BSM-3532/3552/3562. 3 MULTI connectors for electrocardiograph on a screen. This makes it
BSM-3733/3753/3763.) Either IBP, CO, CO 2 , easier to recognize multi-focus VPC, ST
APCO or BIS can be measured per MULTI change and verticular hypertrophy.
connector. For the basic parameters, ECG,
12 lead ECG interpretation
Respiration, SpO 2 , NIBP and Temperature can
12 lead ECG interpretation is available. The 12
be measured with dedicated connectors .
lead ECG with 10 electrodes provides 12 lead
High resolution, TFT color display ECG interpretation results with identical quality
The monitor has a high resolution, to Nihon Kohden electrocardiographs with
wide-viewing angle 15-inch ECAPS 12C program.
(BSM-3733/3753/3763) and 12.1-inch
ST recall
(BSM-3532/3552/3562) TFT color LCD. Max.
Up to 72 hours ST events can be saved. All
15 waveforms (BSM-3733/3753/3763) or Max.
Lead ST Segment analysis is available. You
14 waveforms (BSM-3532/3552/3562) can be
can edit an ST measurement point after
displayed.
reviewing it.
Flexible monitor control
Alarm history
A touch panel screen provides easy and quick
You can review up to 16,384 (72 hours) files for
operation. You can directly write the patient
the numeric data of alarm occurrences from the
name on screen with your finger. Any
alarm history table. You can also review the
characters and figures can be displayed as
wave by jumping to the full disclosure screen
patient name.
from the alarm history table.
(When a monitor is connected by LAN, free
writing patient name is not available.) 72 hours, 5 trace full disclosure
Active parameters appear automatically on the Up to 72 hours of ECG and 4 other parameters
screen with a variety of screen configurations. full disclosure waveform is displayed. Full
disclosure makes it easy to search for
Enlarged numeric screen
waveforms.
Numeric information on the enlarged numeric
screen is useful for viewing from a distance. Trendgraph/trend list
Up to 72 hours of trendgraphs of all parameters,
Alarm indicator
and 72 hours of trend list data and 1,024 (72
An alarm indicator at the top of the monitor can
hours) sets of NIBP list data can be stored and
be seen clearly from a distance from all angles.
reviewed.
It lights in red or yellow depending on the
severity of the alarm. A green lamp blinks in Short trend
synchronization with the QRS when configured. The latest 30 minute parameter data can be
displayed as a trendgraph on the home screen.
.
This short trandgraph can be dragged left or right
by touch panel operation. The OCRG
(oxygen-cardio-respirogram) can also be displayed
in the home screen as short trend.

2
Mainstream EtCO2 for both intubated & Hemodynamics Review Program, Advanced
non-intubated patient (cap-ONE, TG-920P) Intensive Therapy Management Calculation/
With a unique and very small (about 10 g) Trend
mainstream cap-ONE CO 2 sensor (TG-920P), With optional QP-033P, Hemodynamics Review
smooth, fast (within 5 seconds) and reliable Program is available. The Hemodynamics
readings of EtCO 2 & respiration-rate is Review Program with hemodynamics graph
available. provides a graphic view of overall
hemodynamic information. It allows a more
PWTT trigger NIBP measurement intuitive approach to diagnostic and
PWTT is non-invasively measured by therapeutic decision making in hemodynamic
continuous monitoring of ECG and SpO 2. If a management.
sudden critical circulation change occurs
between periodic NIBP measurements, PWTT CVP-ET
detects it and trigger NIBP measurement to End tidal mean CVP measurement is available.
confirm it. The CVP-ET uses the CO 2 signal to calculate
the end tidal mean CVP. In this mode, the mean
8ch EEG CVP value is stable and less affected by
EEG monitoring can be performed by using the respiration. To measure CVP-ET, mainstream
AE-918P neuro unit. Up to 8 channels can be CO 2 measurement (TG-920P, TG-950P or
monitored. TG-970P) is necessary.
esCCO LS-NET (Life Scope Networking)
With optional QP-033P/034P, esCCO You can connect with Nihon Kohden central
measurement is available. esCCO is a new monitors and bedside monitors by LS-NET (Life
technology to calculate cardiac output Scope Networking with Ethernet LAN) or with
continuously. It uses PWTT (Pulse Wave optional WLAN.
Transit Time) which is obtained from the pulse
oximetry and ECG-signals from each cycle of Wireless networking
the ECG and peripheral pulse wave. It can By attaching optional ZS-900PK transmitter to
continuously display CCO, CCI, SV, SVI, SVR the monitor, telemetry networking is available
and SVRI. with an ORG-9100, multiple patient receiver.

Interbed
When Life Scope series bedside monitors are
connected to each other by LS-NET, you can
exchange and view data of other bedside
monitors without a need of the central monitor.
Both individual display and 20 beds display are
available.

SYSTEM COMPOSITION Telemetry


system
Transmitter WEP-5200
*2
Recorder Bedside Monitor ZS-900PK
Multi-patient LS-NET Central
WS-371P Wireless LAN Receiver monitor
station ORG-9100
BSM-3532/3552/3562 *3 CNS-6200/9600 series
Battery QI-320PA
BSM-3733/3753/3763
X075
(SB-671P) Wireless LAN LS-NET Central
Network monitor
isolation unit access point
YS-095P5
HIT-100 CNS-6200/9600 series
Interface QI-372P
LS-NET Central
monitor
Multi-link x2
CNS-6200/9600 series

ECG/IABP Optional External *1: NEC LCD191VXM or equivalent (local purchase)


*1
Output module unit, display *2: DI-590P Holder required
*3: YS-095P3 WLAN adapter kit required
external device

Note: In order to connect an external device via interface QF-9xxP, IF-9xxP, GF-110PA/120PA , GF-210R/220R or
AE-918P Multi-link port is used. QI-372P is necessary for BSM-3500/3700

3
Numerical data display:
SPECIFICATIONS HR, VPC, ST, RR, NIBP (systolic, diastolic,
MAP), IBP (systolic, diastolic, mean), SpO 2 ,
◆ Measuring Parameters PR, PI, TEMP, CO, CI, Ti, Tb, O 2 , FiCO 2 ,
ECG(3/6/10 lead), respiration (impedance ETCO 2 , BIS, SEF90, SEF95, SR, EMG, SQI,
1
method), SpO 2 , NIBP, IBP* , temperature (2ch), N 2 O (I), N 2 O (E), O 2 (I), O 2 (E), Agent (I),
2 1
esCCO* , cardiac output* , CO 2 in mainstream Agent (E), TVi, TVe, MV, Ppeak, Pmean, PEEP,
1 3
method* and sidestream method* , Ri, Re, C, I:E, SEF, MDF, PPF, TP, Absδ, Absθ,
3 1 3
Flow/Paw* , BIS* , anesthetic gas* (CO 2 , O 2 , Absα, Absβ, Absγ, %δ, %θ, %α, %β, %γ, TOF
4 4 4
N 2 O, agent), TOF* , ventilation* , CCO* , rat, CCO, CCI, SVR, SVRI, SV, SVI, SVV,SvO 2 ,
3
EEG* ScvO 2 , EF, EDV, EDVI, tcPO 2 , tcPCO 2 , PPV,
1
* : MULTI connector parameters. 2 MULTI SPV, esCCO, esCCI, esSV, esSVI, esSVR,
connectors are available for esSVRI
BSM-3532/3552/3562. 3 MULTI connectors Synchronization mark: Heart rate sync mark,
are available for BSM-3733/3753/3763. pulse rate sync mark, respiratory sync mark
2
* : Optional software QP-033/034P is required. Alarm configurable start-up values: Master
3
* : Optional module/unit is required. To connect Alarm Neonatal, Master Alarm Pediatric,
module/unit, the main unit needs to be Master Alarm Adult
installed interface board, QI-372P**. Site specialized settings included: OR, ICU,
4
* : External device and interface are required. To NICU
connect interface, the main unit needs to be
installed interface board, QI-372P**. ◆ Language
**: QI-372P interfaces 2x Multi-link ports, RGB Available language: Czech, Danish, English,
and ECG/BP OUT port.(Factory option) Finnish, French, German, Greek, Hungarian,
Italian, Netherlandish, Norwegian, Portuguese,
◆ Display Polish, Romanian, Russian, Slovak, Spanish
Display size: 15-inch (BSM-3733/3753/3763), and Turkish
12.1-inch (BSM-3532/3552/3562)
Touch panel operation: available ◆ Sound
Waveform display mode: Non-fade moving or Type of sound: Alarm, synchronization (HR,
non-fade fixed IBP, SpO 2 ), click, NIBP completion sound
Viewing area: 304.1 x 228.1 mm Alarm sound: 4 types (pip, bing bong, bong,
(BSM-3733/3753/3763), 246.0 x 184.5 mm bing 3 times)
(BSM-3532/3552/3562) Synchronization sound: Variable pitch for
Resolution: 1024 x 768 dots IBP and SpO 2
(BSM-3733/3753/3763), 800 x 600 dots
(BSM-3532/3552/3562) ◆ Alarm
Maximum number of traces: Alarm items: Upper/lower limits, apnea,
15 (BSM-3733/3753/3763), arrhythmia, connector disconnection, noise,
14 (BSM-3532/3552/3562) electrode off, waveform detecting, probe off,
Sweep speed: 6.25, 12.5, 25 or 50 mm/s cuff/hose check, sensor check, low battery,
Respiration/CO 2 sweep speed: 1.56, 6.25, operating environment
12.5 or 25 mm/s Vital sign alarms: HR, PR, ST, RR, APNEA,
Sweep time (at 25 mm/s sweep speed): TEMP, delta TEMP, SpO 2 , NIBP, IBP, ETCO 2 ,
6.5s (BSM-3532/3552/3562), CO 2 (I), O 2 (I), O 2 (E), Tb, MV, Ppeak, PEEP,
9.5s (BSM-3733/3753/3763) N 2 O (I), N 2 O (E), Agent (I), Agent (E), SEF,
Waveform display colors: 12, selectable BIS, CCO, CCI, TP, VPC
Numeric display colors: 12, selectable Arrhythmia alarms: ASYSTOLE, VF, VT, V
Displayed waveforms: ECG (maximum 12 BRADY, EXT TACHY, EXT BRADY, SV TACHY,
traces), respiration, IBP (maximum 2 or 3 VPC RUN, TACHYCARDIA, BRADYCARDIA,
traces, depending on number of MULTI COUPLET, EARLY VPC, MULTIFORM, V
connectors), SpO 2 pulse wave, CO 2 , CO RHYTHM, PAUSE, BIGEMINY, TRIGEMINY,
thermodilution curve, EEG, N 2 O concentration, VPC, IRREGULAR RR, PACER
O 2 concentration, anesthetic agent (Halothane, NON-CAPTURE, PROLONGED RR, NO
Isoflurane, Enflurane, Sevoflurane, PACER PULSER
Desflurane), flow, Paw, volume, P-V loop, F-V Alarm types: Crisis (red, blinking), Warning
loop (yellow, blinking), Advisory (yellow or blue, lit),
Message
Alarm indication: Alarm indicator, highlighted
message, alarm sound
Alarm suspend: Provided (for 1, 2, 3 min, Off)
When connected to a central, this alarms can
be adjusted from the central.

4
◆ Trendgraph
Trend parameters: Heart rate, Pulse rate, ◆ Alarm History
VPC rate, Respiration rate, ST level, apnea Item: Vital sign alarms ,Arrhythmia alarms,
time, apnea frequency, NIBP, CO 2 (E), CO 2 (I), Technical alarms and Operational alarms
O 2 (E), O 2 (I), SpO 2 , PI, Tb, IBP, PPV, SPV, Number of files: 16,384 files
Temperature, N 2 O(E), N 2 O(I), anesthetic
agent(E), anesthetic agent(I), BIS, SQI(BIS), ◆ Full Disclosure
SR, EMG, PWTT, CCO, CCI, SV, SVI, SvO 2 , Storage capacity: 72 hours
ScvO 2 , EDV, SVR, SVRI, SVV, PCCO, PCCI, Number of storing parameters: 5
DO 2 , DO 2 I, VO 2 , VO 2 I, TOF ratio, TOF count, Wave duration: 60 s, up to 6 traces (1
PTC, MV, TVe, Compliance, Airway resistance, parameter)
Expiratory airway resistance, Inspiratory Actual size ECG waveform display:
airway resistance, Ppeak, Pmean, PEEP, Provided
tcPO 2 , tcPCO 2 , OCRG
Storage capacity: 72 hours ◆ ST Recall
Up to 6 parameters can be user selected for 3 Item: ST level waveforms of all monitoring
separate graphical windows. ECG
Storage capacity: 72 hours
◆ Vital Signs List
Parameters: Heart rate, Pulse rate, VPC rate, ◆ 12 LEAD
Respiration rate, ST level, NIBP, CO 2 (E), Item: 12 lead analysis result data
CO 2 (I), O 2 (E), O 2 (I), SpO 2 , PI, Tb, IBP, PPV, measurement, interpretation, diagnostic
SPV, Temperature, anesthetic agent(E), Number of files: 18 files
anesthetic agent(I), BIS, SQI(BIS), SR, EMG, ◆ OCRG
CCO, CCI, SV, SVI, SvO 2 , ScvO 2 , EDV, EDVI, Item: Heart rate and SpO 2 acquired every 1
ESV, ESVI, EF, TOF ratio, TOF count, PTC, minutes for the trendgraph and compressed
PCCO, PCCI, SVR, SVRI,SVV, DO 2 , DO 2 I, respiration waveform
VO 2 , VO 2 I, MV, TVe, Compliance, Airway Scale: 1cm/min, 3cm/min
resistance, Expiratory airway resistance, Storage capacity: 72 hours
Inspiratory airway resistance, Ppeak, Pmean,
PEEP, tcPO 2 , tcPCO 2 ◆ Interbed
Storage capacity: Waveforms and data of other Life Scope series
Vital sign data: 72 hours bedside monitor connected by Ethernet can be
NIBP list: 1,024 files reviewed without the need of a central station.
List interval for periodic list: 1, 5, 10, 15, 30 or Alarm silencing from interbed window is
60 min. available.
List interval for NIBP list: at NIBP Display items for personal display:
measurement Waveform: 2 traces
Numeric data: ECG, SpO 2 , Resp, CO 2 , IBP1,
◆ HEMO Trend IBP2, IBP3, NIBP, TEMP1, TEMP2, O 2 , Gas
Number of files: 1,024 files Display items for 20 other bed display:
Parameters: HR/PR, CO, CI, CCO, CCI, Items: Bed ID, Patient name, Alarm (if any),
ART-MEAN, PAP-MEAN, PCWP, PCWP time, Heart rate and SpO 2
CVP, SV, SVI, LVSW, LVSWI, SVR, SVRI, PVR, Waveform: none
PVRI, RVSW, RVSWI Numeric data: HR, SpO 2
◆ LUNG Trend ◆ ECG
Number of files: 128 files Complies with IEC 60601-2-27 2nd edition:
Parameters: BSA, CI, AaDO 2 , CaO 2 , DO 2 , 2005, ANSI/AAMI EC13: 2002, ANSI/AAMI
DO 2 I, CvO 2 , avDO 2 , VO 2 , VO 2 I, O 2 ER, EC57: 1998.
PaO 2 /O 2 , Qs/Qt Leads:
◆ Recall 3-electrode cable: I, II, III
Item: Arrhythmia 6-electrode cable: I, II, III, aVR, aVL, aVF, 2
Storage capacity: 16,384 files from V1 to V6
Arrhythmia recalled: ASYSTOLE, VF, VT, V 10-electrode cable: I, II, III, aVR, aVL, aVF, V1
BRADY, EXT TACHY, EXT BRADY, SV TACHY, to V6
VPC RUN, TA-CHYCARDIA, BRADYCARDIA, Defibrillation-proof: ECG input protected
COUPLET, EARLY VPC, MULTIFORM, V against 400 Ws/DC 5 kV IEC 60601-2-27
RHYTHM, PAUSE, BIGEMINY, TRI-GEMINY, 17.101 compatible
VPC, IRREGULAR RR, PACER Electrode offset potential tolerance: ≥ ±500
NON-CAPTURE, PROLONGED RR, NO mV
PACER PULSE Input dynamic range: ≥ ±5 mV
Internal noise: ≤ 30 μVp-p (Referred to input)

5
Noise suppression: Pacemaker pulse rejection capability, with
RL driving gain: maximum 40 dB overshoot:
Maximum voltage: 1.23 Vrms Overshoot amplitudes and time constants of
Common mode rejection ratio: ≥ 95 dB ±0.12 mV/100 ms to ±2 mV/4 ms
Input bias current: ≤ 100 nA (As defined by method B of ANSI/AAMI EC13
Frequency response: Sect. 4.1.4.2, this corresponds to the
DIAG mode: 0.05 to 150 Hz (–3 dB) pacemaker pulses amplitudes and widths of ±4
MONITOR mode: 0.3 to 40 Hz (–3 dB) mV/2 ms to amplitudes ±80 mV/0.1 ms.)
MAXIMUM mode: 1 to 18 Hz (–3 dB) Heart rate alarm:
AC hum filter: ≤ –40 dB (at 50 or 60 Hz) Upper limit range: 16 to 300 beats/min, OFF in
Input impedance: ≥ 5 MΩ (at 10 Hz), ≥ 2.5 1 beat/min steps
MΩ (at 0.67 to 40 Hz) Lower limit range: OFF, 15 to 299 beats/min in
ESU protection: Provided (IEC 60601-2-27: 1 beat/min steps
2005 compatible) Alarm items: TACHYCARDIA, BRADYCARDIA,
Recovery time after defibrillation: 10 s Escalating Alarm leads off time
Leads-off sensing: Each leads has own Arrhythmia analysis:
sensing with user configurable automatic lead Analysis method: Multi-template matching
change method
Active electrode: < 100 nA Number of channels: 2
Reference electrode: < 900 nA QRS detection type: Adult, child, neonate
Waveform display: VPC counting rate: 0 to 99 VPCs/min
Display sensitivity: 10 mm/mV ±5% (DIAG Arrhythmia message: ASYSTOLE, VF, VT, V
mode at ×1 sensitivity) BRADY, EXT TACHY, EXT BRADY, SV TACHY,
Number of channels: 3 (maximum, with 6 or 10 VPC RUN, TACHYCARDIA, BRADYCARDIA,
electrodes on home screen), 12 (maximum, COUPLET, EARLY VPC, MULTIFORM, V
with 10 electrodes at 12 LEAD window) RHYTHM, PAUSE, BIGEMINY, TRIGEMINY,
Sensitivity control: ×1/4, ×1/2, ×1, ×2, ×4, or FREQ VPC, VPC, IRREGULAR RR, PACER
AUTO NON-CAPTURE, PROLONGED RR, NO
Pacing mark display: Available PACER PULSE
Recording sensitivity: 10 mm/mV ±5% (same Other messages: NOISE, CHECK
as the display sensitivity) ELECTRODES, LEARNING
Heart rate count: Arrhythmia alarm: Upper limit range: OFF, 1 to
Calculation method: Moving 99 VPC/min
average/Instantaneous beat to beat Number of arrhythmia recall files: 16,384 files
QRS detection (at × 1 sensitivity): (72 hours)
Adult: Width: 70 to 120 ms, Amplitude: 0.5 to 5 Storage time per file: 8 s
mV, rate: 30 to 200 beats/min ST level measurement:
Child and neonate: Width: 40 to 120 ms, Number of measurement channels:
Amplitude: 0.5 to 5 mV, rate: 30 to 250 3-electrode: 1 ch
beats/min 6-electrode: 8 ch
Counting range: 0, 15 to 300 beats/min (±2 10-electrode: 12 ch
beats/min) ST level measuring range: ±2.5 mV
Counting accuracy: ±2 beats/min (0, 15 to 300 Measurement point: automatic with possibility
beats/min) (Essential performance in EMC to adjust manually
standard) ST level alarm:
Heart rate display update cycle: Every 3 s or Upper limit range: –1.99 to 2.00 mV in 0.01 mV
when alarm is generated steps, OFF
Heart rate sync mark delay time: within 100 to Lower limit range: OFF, –2.00 to 1.99 mV in
200 ms (when QRS is detected) 0.01 mV steps
Tall T-wave rejection capability: Complies with Storage capacity: 72 hours
the heights of T-waves from 0 mV to 1.2 mV Cascade ECG Waveform: available
specified in ANSI/ AAMI EC13 Sect. 4.1.2.1(c)
Heart rate averaging: Calculated by using the ◆ 12-lead Interpretation
most recent 4 or 12 beats. 12 lead ECG interpretation: ECAPS 12C
Pacemaker pulse detector rejection of fast (BSM), Measurements and diagnostic,
ECG signals: Available when monitoring 12 leads
Slew rate at which the pacemaker pulse Analysis patient age: 3 years to adult
detector responds: 6 to 8 V/s Analysis time: about 30 sec
Tested as specified in ANSI/AAMI EC13 Sect. Finding items: 200
4.1.4.3 Judging items: 5
Pacemaker pulse rejection capability, Interpretation items: Normal sinus rhythm,
without overshoot: TACHYCARDIA, BRADYCARDIA, VPC
Complies with the amplitudes of pacemaker Display and output: Screen, recorder module,
pulses ±2 to ±700 mV and widths 0.1 to network printer, printer connected to the
2 ms specified in ANSI/AAMI EC13 Sect. 4.1.4.1 central monitor
File storage: 18 files
6
◆ Respiration (Transthoracic Impedance Display range:
BSM-3532/BSM-3552/BSM-3733/BSM-3753:1
Pneumography) to 100%SpO 2
Measuring method: Transthoracic impedance
BSM-3562/BSM-3572/BSM-3763/BSM-3773:
pneumography
0 to 100%SpO 2
Number of channels: Selectable from R-F
Declared range:
and R-L
BSM-3562/BSM-3572/BSM-3763/BSM-3773:
Measuring impedance available range: 220 Ω
70 to 100%SpO 2
to 4 kΩ
Measuring accuracy*:
Excitor current: 45 ±10 μArms at 40 kHz
BSM-3532/BSM-3733:
(sine wave)
Adult: ±2%SpO 2 (No motion), ±3%SpO 2
Internal noise: ≤ 0.1 Ω (Referred to input)
(Motion)
Respiration rate counting range: 0 to 150
Neonate: ±3%SpO2
counts/min
BSM-3552/BSM-3753:
Respiration rate counting accuracy*: ±2
Adult: ±2%SpO 2
counts/min (0 to 150 counts/min)
Neonate: ±3%SpO 2
* Essential performance in EMC standard
BSM-3562/BSM-3572/BSM-3763/BSM-3773:
Frequency response (high frequency
70%SpO 2 ≤ %SpO 2 < 80%SpO 2 ±3%SpO 2
cut-off): 3 Hz ±1 Hz (–3 dB)
80%SpO 2 ≤ %SpO 2 ≤ 100%SpO 2 ±2%SpO 2
Defibrillation proof: Respiration input
SpO 2 accuracy is guaranteed at surrounding
protected against 400 Ws/DC 5 kV
temperature of 18 to 40°C (64.4 to 104°F).
Recovery time after defibrillation: 10 s
* Essential performance in EMC standard
Impedance respiration: Measurement On/Off
SpO 2 alarm:
available
Upper limit range: 51 to 100%SpO 2 in 1%SpO 2
Heart beat rejection: Available
steps , OFF
Waveform display:
Lower limit range: OFF, 50 to 99%SpO 2 in
Display sensitivity: 10 mm/1 Ω ±25% (at ×1
1%SpO 2 steps
sensitivity)
Pulse rate:
Sensitivity control: ×1/4, ×1/2, ×1, ×2, ×4
Display range:
Respiration rate display update cycle: Every 3
BSM-3532/BSM-3733: 25 to 240 beats/min
s or when alarm is generated
BSM-3552/BSM-3753: 20 to 300 beats/min
Alarm:
BSM-3562/BSM-3763/BSM-3572/BSM-3773:
Upper limit range: 2 to 150 counts/min in 2
30 to 300 beats/min
counts/min steps, OFF
Declared range:
Lower limit range: OFF, 0 to 148 counts/min in
BSM-3532/BSM-3733: 25 to 240 beats/min
2 counts/min steps
BSM-3552/BSM-3753: 20 to 250 beats/min
Apnea alarm: OFF, 5 to 40 s in 5 s steps
BSM-3562/BSM-3572/BSM-3763/BSM-3773:
Displayed message: APNEA
30 to 300 beats/min
Escalating Alarm: apnea
Counting accuracy (rms)*:
◆ SpO2 BSM-3532/BSM-3733: ±3 beats/min (No
Complies with ISO 9919: 2005. motion), ±5 beats/min (Motion)
Display: BSM-3552/BSM-3753: ±3 beats/min
Display update cycle: Every 3 s or when alarm BSM-3562/BSM-3572/BSM-3763/BSM-3773:
is generated ±3% ±1 beat/min
Sync tone modulation: Changes tone * Essential performance in EMC standard
depending on SpO 2 value Pulse rate alarm:
Sweep speed: 6.25, 12.5, 25, 50 mm/s Upper limit range:
Waveform sensitivity: ×1/8, ×1/4, ×1/2, ×1, ×2, 16 to 300 beats/min in 1 beat/min steps, OFF
×4, ×8 or AUTO (When SYNC SOURCE is set to ECG)
SpO 2 : 31 to 300 beats/min in 1 beat/min steps, OFF
Measuring method: Two wavelength light (When SYNC SOURCE is set to PRESS or
absorption method SpO 2 )
Wavelength range: Lower limit range:
BSM-3532/BSM-3733: 660/905 nm (LNOP tip OFF, 15 to 299 beats/min in 1 beat/min steps
clip and LNCS tip clip), 663/880 nm (Other (When SYNC SOURCE is set to ECG)
clips) BSM-3552/BSM-3753: 660/900 nm OFF, 30 to 299 beats/min in 1 beat/min steps
BSM-3562/BSM-3763: 660/940 nm (When SYNC SOURCE is set to PRESS or
Averaging time: SpO 2 )
BSM-3552/BSM-3753: 6 to 7 s (approximately Escalating Alarm: SpO 2 drop, SpO 2 sub value
3 seconds in FAST mode), If the dynamic time
averaging time exceeds 20 seconds for SpO 2 , Response time (BSM-3562/BSM-3763 only):
the SpO 2 and pulse rate will continue to be Selectable from “SLOW”, “NORMAL” and
updated every second. “FAST”

7
The following graphs show the response time Maximum measurement time:
example when SpO2 changes 0.6%/s. Adult/Child: ≤ 160 s
Neonate: ≤ 80 s
Operation mode: Manual, STAT (≤ 15 min),
Periodic, PWTT and SIM (depends on the
SITE setting)
Auto remeasurement: 1 time
Air leakage: ≤ 3 mmHg/min
Measurement accuracy with a simulator*: ±10
mmHg
* Essential performance in EMC standard
Systolic range:
Adult: 0 - 300 mmHg
Child: 0 - 300 mmHg
Neonate: 0 - 150 mmHg
Diastolic range:
Adult: 0 - 300 mmHg
Child: 0 - 300 mmHg
Neonate: 0 - 150 mmHg
Mean range:
Adult: 0 - 300 mmHg
Child: 0 - 300 mmHg
Neonate: 0 - 150 mmHg
Initial pressurization value:
Adult: 180 mmHg
Child: 140 mmHg
Neonate: 100 mmHg
Maximum pressurization value:
The following graph shows the response time Adult/Child: 300 mmHg
example when pulse rate changes 10 Neonate: 150 mmHg
bpm/s. Display items: Systolic (SYS), diastolic (DIA),
mean (MAP), cuff pressure during NIBP,
measurement, delta PWTT
NIBP data display update cycle: Updated
every measurement
Measurement completion sound: Generated
at measurement completion (depends on the
setting)
Alarm:
Upper limit range: 15 to 260 mmHg in 5 mmHg
steps, OFF
Lower limit range: OFF, 10 to 255 mmHg in 5
mmHg steps
Safety:
Maximum pressurization value cuff inflation
Recovering time after defibrillation: 10 s limiter:
PI: available for BSM-3532/3733, Adult/Child: 300 to 330 mmHg
BSM-3562/3763 Neonate: 150 to 165 mmHg
Cuff inflation time limiter:
◆ Non Invasive Blood Pressure, NIBP Adult/Child: 161 to 165 s
Complies with IEC 60601-2-30: 1999. Neonate: 81 to 84 s
Interval time limiter: 25 to 29 s
Measuring method: Oscillometric Power discontinuity: Deflate immediately after
Measuring range: 0 to 300 mmHg power down
Cuff pressure display range: 0 to 300 mmHg Recovering time after defibrillation: 10 s
Accuracy: ±3 mmHg (0 mmHg ≤ NIBP ≤ 300
mmHg) ◆ Invasive Blood Pressure, IBP
Cuff inflation time: Complies with IEC 60601-2-34: 2000 except
≤ 11 s (700 cc), 0 to 200 mmHg for clauses 44.6, 45.101 a) and 45.101 b).
≤ 5 s (70 cc), 0 to 200 mmHg Labels: P1, P2, P3, ART, ART2, RAD, DORS,
Measurement mode: Adult, child or neonate AO, FEM, UA, UV, PAP, CVP, RAP, RVP, LAP,
is recognized by connected air hose LVP, ICP, ICP2, ICP3
Typical Measurement time: neonatal 20s, Calculation: CPP, PPV, SPV, CVP-ET
pediatric 25s, adult 30 sec. (normal
conditions)

8
Complied transducer:
P23XL-1 and P10EZ-1 Becton Dickinson ◆ Temperature
reusable transducers Complies with EN 12470-4: 2000 only for clauses
Becton Dickinson disposable transducers DX 6.2, 6.3 a), 6.5, 6.6, 6.7, 6.8, 6.9, 6.10 and 8.
series Labels: T1, T2, Tskin, Tskin2, Trect, Tcore,
5 μV/V/mmHg, bridge resistor: 200 Ω to 20 kΩ, Tnaso, Teso, Ttymp, Tblad, Taxil
defibrillation-proof or the equivalents Thermistor probe: 400 series (YSI)
For the warm-up time, refer to the transducer Number of channels: 2 channels with Delta
manual. Temp
Volume displacement: 0.04 mm3/100 mmHg Measuring range: 0 to 45°C, 32 to 113°F
Auto zero balancing range: ±200 mmHg Measuring accuracy*:
Auto ±0.1°C (25°C ≤ TEMP ≤ 45°C)
zero balancing accuracy: ±1 mmHg ±0.2°C ( 0°C ≤ TEMP < 25°C)
Calculation Method: Peak, Standard * Essential performance in EMC standard
(Average) Internal noise: ≤ 0.014°C (at 37°C)
Measuring range: –50 to 300 mmHg Temperature drift: within ±0.005°C /°C
Measuring accuracy: Display range: 0 to 45°C (32 to 113°F)
±1 mmHg ±1 digit (–50 mmHg ≤ IBP < 100 Display update cycle: Every 3 s or when
mmHg) alarm is generated
±1% ±1 digit (100 mmHg ≤ IBP ≤ 300 mmHg) Time response delay from probe to monitor
Total measuring accuracy*: ±4% or ±4 display: ≤ 6 seconds (sensor time constant is
mmHg (whichever is greater)** not included)
* Essential performance in EMC standard Recovering time after defibrillation: 10 s
** When used with ANSI/AAMI BP-22-1994 Alarm:
complied equipments Upper limit range: 0.1 to 45.0°C (33 to 113°F)
Internal noise: within ±1 mmHg in 0.1°C (1°F) steps, OFF
Temperature zero drift: ±0.1 mmHg/1°C Lower limit range: OFF, 0.0 to 44.9°C (32 to
Frequency response: DC to 12 Hz or 20 Hz 112°F) in 0.1°C (1°F) steps
(selectable)
Recovering time after defibrillation: 10 s ◆ Cardiac Output, CO
Display items: Systolic (SYS), diastolic (DIA), Measuring method: Thermodilution method
mean (MEAN) Measuring parameters: Cardiac output (CO),
Display update cycle: Every 3 s or when injectate temperature (Ti), blood temperature
alarm is generated (Tb), delta Tb
BP sync sound: Systolic value 20 to 120 Number of channel: 1
mmHg, changes in 20 steps every 5 mmHg Measuring range:
Alarm: Injectate temperature (Ti): 0°C to 27°C (32 to
Upper limit range: −48 to 300 mmHg in 2 81°F)
mmHg steps, OFF Blood temperature (Tb): 15°C to 45°C (59 to
Lower limit range: OFF, −50 to 298 mmHg 113°F)
steps in 2 mmHg steps Thermodilution curve (delta Tb): 0°C to 2.5°C
Alarm inactivation: Alarm is inactivated in (32 to 37°F)
certain period when zero balancing is Cardiac output (CO): 0.5 to 20 L/min
performed. Measuring accuracy:
Pulse rate: Ti: ±0.2°C (0°C to 27°C)
Counting range: 0, 30 to 300 beats/min Tb:
Display range: 0 to 300 beats/min ±0.1°C (25°C ≤ TEMP ≤ 45°C)
Counting accuracy (rms): ±2 beats/min (30 ±0.2°C (15°C ≤ TEMP < 25°C)
beats/min ≤ PR ≤ 300 beats/min) CO: ±5%
Alarm: Internal noise:
Upper limit range: Ti: ≤ 0.025°C RMS
16 to 300 beats/min in 1 beat/min steps, OFF Tb: ≤ 0.016°C RMS (correspond to 37°C)
(When SYNC SOURCE is set to ECG) Delta Tb: ≤ 0.005°C RMS
31 to 300 beats/min in 1 beat/min steps, OFF Temperature drift:
(When SYNC SOURCE is set to PRESS or Ti: ±0.005°C /°C
SpO 2 ) Tb: ±0.005°C /°C
Lower limit range: Frequency response (delta Tb): DC to 12 Hz
OFF, 15 to 299 beats/min in 1 beat/min steps (–3 dB)
(When SYNC SOURCE is set to ECG) Injectate volume range: 3, 5, 10 mL
OFF, 30 to 299 beats/min in 1 beat/min steps Display update cycle: Updated every
(When SYNC SOURCE is set to PRESS or measurement
SpO 2 ) Tb alarm:
Upper limit range: 15.1 to 45.0°C (60 to 113°F)
in 0.1°C (1°F) steps, OFF
Lower limit range: OFF, 15.0 to 44.9°C (59 to
112°F) in 0.1°C (1°F) steps
9
◆ Carbon Dioxide, CO2 (Mainstream Respiration rate alarm:
Upper limit range: 2 to 150 counts/min in 2
Method) counts/min steps, OFF
For the TG-900P/TG-920P/TG-950P*/TG-970P
Lower limit range: OFF, 0 to 148 counts/min in
CO2 sensor kit specifications, refer to the kit
2 counts/min steps
manual.
Apnea time:
* TG-950P is not available for Op No. 32A,
OFF, 5 to 40 s in 5 s steps
52A, 72A, 33A, 53A and 73A.
Displayed message: APNEA
Displayed CO 2 *: Values detected by the
Total system response time: ≤ 1.0 second
TG-900P/TG-920P/TG-950P/TG-970P CO 2
Recovering time after defibrillation: 10 s
sensor kit
* Essential performance in EMC standard ◆ Calculation
Calculation method: Drug calculation: Available
TG-900P/TG-920P: semi-quantitative Lung function calculation: Available
TG-950P/TG-970P: quantitative (oxygenation, ventilation)
CO 2 measuring range: Hemodynamics Review Program, Advanced
TG-900P/TG-920P/TG-950P: 0 to 100 mmHg Intensive Therapy Management
TG-970P: 0 to 150 mmHg Calculation/Trend: Available (with optional
CO 2 measuring accuracy: software QP-033P)
TG-900P/TG-920P:
±0.4 kPa (0 ≤ CO 2 ≤ 1.33 kPa) (±3 mmHg (0 ≤ ◆ Bispectral Index, BIS
CO 2 ≤ 10 mmHg)) For the BISx/BIS processor specifications,
±0.53 kPa (1.33 < CO 2 ≤ 5.33 kPa) (±4 mmHg refer to the BISx/BIS processor manual.
(10 < CO 2 ≤ 40 mmHg)) BIS can be also monitored with Covidien’s BIS
±10% reading (5.33 < CO 2 ≤ 13.3 kPa (40 < monitor.
CO 2 ≤ 100 mmHg)) Recovering time after defibrillation: 10 s
(At 1 atmospheric pressure, air inspiration, no BIS alarm:
condensation) Upper limit range: 2 to 100 in 1 steps, OFF
TG-950P/TG-970P: Lower limit range: OFF, 0 to 99 in 1 steps
±0.27 kPa (0 ≤ CO 2 ≤ 5.33 kPa) (±2 mmHg (0 ≤
CO 2 ≤ 40 mmHg)) ◆ ECG/BP Output (When QI-372P is
±5% reading (5.33 < CO 2 ≤ 9.33 kPa (40 < CO 2 Connected)
≤ 70 mmHg)) Outputs 100 mmHg/V IBP waveform and the
±7% reading (9.33 < CO 2 ≤ 13.3 kPa (70 < CO 2 first trace of 1 mV/V ECG waveform. When
≤ 100 mmHg)) more than one IBP waveforms are acquired,
(When no condensation) the IBP waveform of the top MULTI socket on
Warm-up time: the monitor is output (when “FIXED
TG-900P/TG-920P: 5 s POSITION” is set for IBP Analog Output) or
TG-950P: 15 s the IBP waveform is output following the
TG-970P: 10 s highest priority label (when “HIGHEST
Response time: PRIORITY LABEL” is set for IBP Analog
TG-900P: 160 ms (typical) for steps from 10 to Output).
90% Complied medical electrical equipments:
TG-920P/TG-950P/TG-970P: 120 ms (typical) Connecting medical electrical equipment must
for steps from 10 to 90% comply to the following standards:
Respiration rate counting range: IEC 60601-1: 1988
TG-900P/TG-920P: 3 to 150 counts/min IEC 60601-1 Amendment 1: 1991
TG-950P/TG-970P: 0 to 150 counts/min IEC 60601-1 Amendment 2: 1995
Respiration rate counting accuracy: Medical electrical equipment must be
TG-900P/TG-920P: ±10% (3 to 150 connected by specified method in following
counts/min) standards:
TG-950P/TG-970P: ±1 count/min IEC 60601-1-1: 2000
CO 2 value display update cycle: Every 3 s or Output impedance:
when alarm is generated ECG: ≤ 100 Ω
CO 2 alarm: BP: ≤ 100 Ω
Upper limit: Output-waveform:
CO 2 (I): ECG: ±5.0 V (at 1 mV/V ± 5% sensitivity)
1 to 99 mmHg in 1 mmHg steps, OFF BP: –0.5 to +3.0 V (at 100 mmHg/V ± 1%
0.1 to 13.0 kPa in 0.1 kPa steps, OFF sensitivity)
ETCO 2 : HT: 5.0 to 15.0 V (Open collector output: 0.5 to
2 to 99 mmHg in 1 mmHg steps, OFF 50 mA)
0.2 to 13.0 kPa in 0.1 kPa steps, OFF Frequency response:
Lower limit: ECG: ≥ 0.5 to 100 Hz (≥ –3 dB) (No
ETCO 2 : reproducibility of pace maker pulse)
OFF, 1 to 98 mmHg in 1 mmHg steps BP: DC to 20 Hz ±3 Hz (–3 dB)
OFF, 0.1 to 12.9 kPa in 0.1 kPa steps HT pulse width: 15 ms
10
Gain: Spurious emission strength: ≤ 2.5 μW (5
ECG: 1000 MHz to 1.5 GHz)
Offset: Occupied bandwidth: 5.0 to 8.5 kHz
ECG: ≤ ±50 mV Adjacent channel leaking power: ≥ 40 dBR
BP: ≤ ±10 mV Transmission frequency range: 420.0500 to
Sensitivity accuracy: 449.6625 MHz
ECG: ±5% Modulation method: Frequency shift keying
BP: ±1% Parameters sendable to central monitor:
Delay: ECG (1ch), respiration, SpO 2 , NIBP, IBP (2 ch),
ECG, HT: 20 ms max temp (2ch), CO 2
BP: 40 ms max
ART (CCO): 60 ms max ◆ Gas
Gas can be monitored with the AG-920R,
◆ RGB Socket (When QI-372P is GF-110PA or GF-210R multigas unit or
Connected) GF-120PA or GF-220R multigas/flow unit. For
Resolution: the AG-920R, GF-110PA or GF-210R multigas
BSM-3532/3552/3562/3572: 800 × 600 dots unit or GF-120PA or GF-220R multigas/flow
BSM-3733/3753/3763/3773: 1024 × 768 dots unit specifications, refer to the manual.
Measurement method: Sidestream gas
◆ RS-232C Socket (When QI-373P is sampling
Connected) Measured parameters: Inspired/expired CO 2
Serial communication: RS-232C complies partial pressure, inspired/expired N 2 O
Baud rate: 9600, 19200, 38400 bps concentration, inspired expired O 2
concentration, inspired/expired anesthetic
◆ Nurse Call Socket (When QI-373P is agent concentration (Halothane, Isoflurane,
Connected) Enflurane, Sevoflurane, Desflurane),
Nurse call output: Open collector output (Law respiration rate, minimum alveolar
active) concentration
Warm-up time:
◆ When WS-371P Recorder Module is AG-920R, GF-110PA/120PA:
Connected within 45 seconds to first measurement
Recording method: Thermal array recording within 10 minutes to measurement with
Number of channels: 3 traces (maximum) guaranteed accuracy
Recording width: ≥ 46 mm GF-210R/220R:
Paper speed: 12.5, 25, 50 mm/s about 1 minute to CO 2 measurement
Recording mode: Manual, periodic, alarm about 6 minutes to measurement with
Recording density: guaranteed accuracy
Amplitude direction: 8 dots/mm Sampling rate:
Feeding direction: AG-920R, GF-110PA/120PA:
40 dots/mm (≤ 25 mm/s) 70 to 100 mL/min ±10 mL/min
20 dots/mm (50 mm/s) 100 to 200 mL/min ±10%rel
Recording paper: FQW-50-2-100 GF-210R/220R:
Recording: three waveforms selected on 200 mL/min ±20 mL/min
RECORD window, Trendgraphs and tables, Total system response time:
NIBP trend, hemo trend, lung trend, AG-920R, GF-110PA/120PA:
arrhythmia recall, alarm history, full disclosure, ≤ 5.0 seconds (when sampling volume is 200
ST level recall, OCRG, 12 lead ECG mL/min, using adult sampling tube and adult
interpretation, 12 lead analysis waveform, water trap)
DRUG, LUNG function, CO data on CO window, GF-210R/220R:
TOF measurement result on TOF window, AP ≤ 5.0 seconds (when YG-610P sampling line is
waveform on CCO window, FLOW data and connected)
loops on FLOW window, EEG waveforms with CO 2 measurement:
numeric values on EEG window, analog Measurement method: Non-dispersive infrared
waveforms on ANALOG window ray absorption
Measuring range:
◆ When ZS-900P Transmitter is AG-920R, GF-110PA/120PA: 0 to 76 mmHg, 0
Connected to 10.13 kPa
GF-210R/220R: 0 to 10 vol%
ZS-900P transmitter is not available for Op No.
Measuring accuracy:
32A, 52A, 72A, 33A, 53A and 73A.
AG-920R, GF-110PA/120PA:
±2 mmHg (0 ≤ CO 2 ≤ 40 mmHg), ±0.27 kPa (0
Frequency capacity deviation: ≤ ±3 ppm (15
≤ CO 2 ≤ 5.33 kPa)
to 35°C)
±3 mmHg (40 ≤ CO 2 ≤ 55 mmHg), ±0.40 kPa
Transmission power: 1.0 mW +5%, –40% (15
(5.33 ≤ CO 2 ≤ 7.33 kPa)
to 35°C)
±4 mmHg (55 < CO 2 ≤ 76 mmHg), ±0.53 kPa
(7.33 < CO 2 ≤ 10.13 kPa)
11
Response time (10 to 90%): SEV 0 to 8%
AG-920R, GF-110PA/120PA: ≤ 250 ms (under DES 0 to 18%
the condition of sampling flow is 200 mL/min GF-210R/220R:
and sampling line for adult and water trap for HAL, ISO 0 to 8.5 vol%
adult is connected) ENF, SEV 0 to 10 vol%
Alarm: DES 0 to 20% vol%
Upper limit: Measuring accuracy:
CO 2 (I): AG-920R, GF-110PA/120PA:
1 to 99 mmHg in 1 mmHg steps, OFF ±0.2% (0 ≤ GAS ≤ 5%)
0.1 to 13.0 kPa in 0.1 kPa steps, OFF ±0.4% (5 < GAS ≤ 10%)
ETCO 2 : ±0.6% (10 < GAS ≤ 15%)
2 to 99 mmHg in 1 mmHg steps, OFF ±1.0% (15 < GAS ≤ 18%)
0.2 to 13.0 kPa in 0.1 kPa steps, Off Response time (10 to 90%):
Lower limit: AG-920R, GF-110PA/120PA:
ETCO 2 : ≤ 300 ms (HAL, ISO, SEV, DES)
OFF, 1 to 98 mmHg in 1 mmHg steps ≤ 500 ms (ENF) (under the condition of
OFF, 0.1 to 12.9 kPa in 0.1 kPa step sampling flow is 200 mL/min and sampling line
N 2 O measurement: for adult and water trap for adult is connected)
Measurement method: Non-dispersive infrared Alarm:
ray absorption Upper limit:
Measuring range: Agent(I), Agent(E) (HAL, ISO, SEV, ENF): 0.1
AG-920R, GF-110PA/120PA: 0 to 100% to 7.0% in 0.1% steps, OFF
GF-210R/220R: 0 to 100 vol% DES(I), DES(E): 0.1 to 20.0% in 0.1% steps,
Measuring accuracy: OFF
AG-920R, GF-110PA/120PA: ±3% Lower limit:
Response time (10 to 90%): Agent(I), Agent(E) (HAL, ISO, SEV, ENF): OFF,
AG-920R, GF-110PA/120PA: ≤ 250 ms (under 0.0 to 6.9% in 0.1% steps, OFF
the condition of sampling flow is 200 mL/min DES(I), DES(E): OFF, 0.0 to 19.9% in 0.1%
and sampling line for adult and water trap for steps
adult is connected) MAC:
Alarm (N 2 O(I), N 2 O(E)): Uncorrected MAC: When AG-920R multigas
Upper limit: 1 to 100% in 1% steps, OFF unit is connected
Lower limit: OFF, 0 to 99% in 1% steps Uncorrected MAC = %Et(AA1)/x(AA1) +
O 2 measurement: %Et(AA2)/x(AA2) + %Et(N 2 O)/x(N 2 O)
Measurement method: Paramagnetic %Et(AA1): End tidal concentration of primary
Measuring range: anesthetic agent
AG-920R, GF-110PA/120PA: 0 to 100% %Et(AA2): End tidal concentration of
GF-210R/220R: 5 to 100 vol% secondary anesthetic agent
Measuring accuracy: %Et(N 2 O): End tidal concentration of N 2 O
AG-920R, GF-110PA/120PA: x(AA1): Uses the following values with the
±2% (0 ≤ O 2 ≤ 55%) MAC of primary anesthetic agent
±3% (55 < O 2 ≤ 100%) HAL = 0.77%, ENF = 1.7%, ISO = 1.15%, SEV
GF-210R/220R: = 2.1%, DES = 7.3%
±(2.5 vol% + 2.5%rel) x(AA2): Uses the following values with the
Response time (10 to 90%): MAC of secondary anesthetic agent
AG-920R, GF-110PA/120PA: ≤ 500 ms (under HAL = 0.77%, ENF = 1.7%, ISO = 1.15%, SEV
the condition of sampling flow is 200 mL/min = 2.1%, DES = 7.3%
and sampling line for adult and water trap for x(N 2 O): Uses 105% with the MAC of N 2 O
adult is connected) For the GF-110PA or GF-210R multigas unit
Alarm: and GF-120PA or GF-220R multigas/flow unit,
Upper limit: refer to the manual.
O 2 (I): 19 to 100% in 1% steps, OFF Ambient pressure corrected MAC: For
O 2 (E): 11 to 100% in 1% steps, OFF GF-110PA or GF-210R multigas unit and
Lower limit: GF-120PA or GF-220R multigas/flow unit, refer
O 2 (I): 18 to 99% in 1% steps to the manual.
O 2 (E): OFF, 10 to 99% in 1% steps Enhanced MAC correction: For GF-110PA or
Anesthetic agent measurement: GF-210R multigas unit and GF-120PA or
Measurement method: Non-dispersive infrared GF-220R multigas/flow unit, refer to the
ray absorption manual.
Measured items: HAL (Halothane), ISO
(Isoflurane), ENF (Enflurane), SEV
(Sevoflurane), DES (Desflurane)
Measuring range:
AG-920R, GF-110PA/120PA:
HAL, ISO, ENF 0 to 5%

12
Respiration rate: Measuring accuracy: ±3% rel or ±0.005 L/s
Measuring range: 0, 4 to 60 counts/min whichever is greater (Applicable when 10
Measuring accuracy: ±1 count/min minutes or more has elapsed)
Alarm: Paw measurement:
Upper limit: 2 to 150 counts/min in 2 Ppeak, Pmean, PEEP:
counts/min steps, OFF Measuring range: –20 to +100 cmH 2 O, hPa
Lower limit: OFF, 0 to 148 counts/min in 2 Measuring accuracy: ±1 cmH 2 O, hPa
counts/min steps (Applicable when 10 minutes or more has
Apnea alarm: OFF, 5 to 40 s in 5 s steps elapsed)
Displayed message: APNEA Ppeak alarm:
Upper limit range: 1 to 50 cmH 2 O, hPa in 1
◆ Carbon Dioxide, CO2 (Sidestream cmH 2 O, hPa steps, OFF
Method) Lower limit range: OFF, 0 to 49 cmH 2 O, hPa in
CO 2 in sidestream method can be monitored 1 cmH 2 O, hPa steps
with the AG-400R CO 2 unit. PEEP alarm:
For the AG-400R CO 2 unit specifications, refer Upper limit range: 1 to 100 cmH 2 O, hPa in 1
to the AG-400R CO 2 unit manual. cmH 2 O, hPa steps, OFF
Sampling flow: 50 mL/min +15/−7.5 mL/min Lower limit range: OFF, 0 to 99 cmH 2 O, hPa in
Warm-up time: 30 s average (from power on 1 cmH 2 O, hPa steps
to the measurable state) Volume measurement:
Measuring range: 0 to 99 mmHg Measuring range: 0 to 3000 mL
Total measuring accuracy: Whichever Measuring accuracy: ±5% rel or ±10 mL
greater in following measuring accuracy whichever is greater (Applicable when 10
Measuring accuracy: minutes or more has elapsed)
0 to 38 mmHg ±2 mmHg TVe, TVi measurement:
39 to 99 mmHg ± [5 + 0.08 × (χ− 39)] % of Measuring range: 0 to 3000 mL
reading Display range: 0 to 9999 mL
χ: CO 2 partial pressure of a standard gas with Measuring accuracy: ±5% rel or ±10 mL
a known CO 2 partial pressure (mmHg) whichever is greater (Applicable when 10
ETCO 2 and CO 2 (I) alarm: minutes or more has elapsed. Not applicable
Upper limit: when TVi and TVe is less than 100 mL)
CO 2 (I): MV measurement:
1 to 99 mmHg in 1 mmHg steps, OFF Display range: 0 to 99.9 L/min
0.1 to 13.0 kPa in 0.1 kPa steps, OFF Alarm:
ETCO 2 : Upper limit range: 0.1 to 30.0 L/min in 0.1
2 to 99 mmHg in 1 mmHg steps, OFF L/min steps, OFF
0.2 to 13.0 kPa in 0.1 kPa steps, OFF Lower limit range: OFF, 0.0 to 29.9 L/min in 0.1
Lower limit: L/min steps
ETCO 2 : C measurement:
OFF, 1 to 98 mmHg in 1 mmHg steps Display range: 0.0 to 999.9 mL/cmH 2 O
OFF, 0.1 to 12.9 kPa in 0.1 kPa steps R, Ri, Re measurement:
Respiration rate measuring range: Display range: 0.0 to 999.9 cmH 2 O/L/s
101 to 150 counts/min: ±5% Respiration rate measurement:
71 to 100 counts/min: ±3% Counting range: 0, 4 to 60 counts/min
41 to 70 counts/min: ±2 counts/min Counting accuracy: ±1 counts/min
0 to 40 counts/min: ±1 count/min Alarm:
Respiration rate alarm: Upper limit range: 2 to 150 counts/min in 2
Upper limit range: 2 to 150 counts/min in 2 counts/min steps, OFF
counts/min steps, OFF Lower limit range: OFF, 0 to 148 counts/min in
Lower limit range: OFF, 0 to 148 counts/min in 2 counts/min steps
2 counts/min steps Apnea time: OFF, 5 to 40 s in 5 s steps
Apnea alarm: OFF, 5 to 40 s in 5 s steps Displayed message: APNEA
Displayed message: APNEA
Total system response time: ≤ 4 seconds ◆ EEG
EEG can be monitored with the AE-918P
◆ FLOW/Paw neuro unit. For the AE-918P neuro unit
Flow/Paw can be monitored with the specifications, refer to the AE-918P neuro unit
GF-120PA or GF-220R multigas/flow unit. For manual.
the GF-120PA or GF-220R multigas/flow unit Number of channels: 8
specifications, refer to the GF-120PA or Measuring range:
GF-220R multigas/flow unit manual. SEF, MDF, PPF: 0.0 to 62.5 Hz
FLOW measurement: TP: 0.01 to 9.99 nW
Measurement method: Differential pressure ABS δ, ABS θ, ABS α, ABS β, ABS γ: 1 to 9999
method (fixed orifice) pW% δ, % θ, % α, % β, % γ: 0 to 100%
Measuring range: –3 to +3 L/s Data display update cycle: Every 3 s or when
alarm is generated
13
Electrode impedance check: > 10 kΩ within esCCI upper limit:0.60 to 20.00 L/min in 0.10
±20% L/min steps, OFF
Sensitivity: 10 µV/1 mm within ±5% esCCI lower limit: OFF, 0.50 to 19.90 L/min in
Non distorted maximum input: > ±2 mV 0.10 L/min steps
Polarization voltage: > ±700 mV Signal Quality Index: available (4 levels)
Input impedance: > 15 MΩ at 10 Hz
CMRR: > 110 dB (in isolation mode) ◆ Battery (SB-671P Battery Pack)
Frequency characteristics: Type of battery: Nickel-metal hydride
High range: 70 Hz at 70% amplitude (−3 dB) Number of battery: 1
within ±20% Battery lifetime: 1 year or 200 cycles of full
Low range: 2 s ±20% or 0.08 Hz at 70% discharging/charging
amplitude (−3 dB) within ±20% Battery operation time:
AC filter: attenuation ratio > 26 dB BSM-3532/3552/3562/3572: 90 minutes
Noise: within 3 µVp-p BSM-3733/3753/3763/3773: 60 minutes
SEF alarm: (new battery, fully charged and no options are
Upper limit range: 1.0 to 60.0 Hz in 0.5 Hz used in normal temperature)
steps, OFF DC voltage: 9.6 V
Lower limit range: OFF, 0.5 to 59.5 Hz in 0.5 Charging current: 360 mA ±50 mA (normal
Hz steps use)
TP alarm: Charging time:
Upper limit range: 0.02 to 9.99 nW in 0.01 nW During monitoring: 10 hours
steps, OFF During non-monitoring: 2 hours
Lower limit range: OFF, 0.01 to 9.98 nW in Battery status indication: Battery lamps on
0.01 nW steps the front panel, screen message and alarm
sound, alarm indicator
◆ CCO Operating environment:
For the APCO/IBP processor specifications, Charging temperature: 10 to 55°C (50 to
refer to the APCO/IBP processor manual. 131°F)
CCO alarm: Discharging temperature: 5 to 50°C (41 to
Upper limit range: 1.1 to 20.0 L/min in 0.1 122°F)
L/min steps, OFF Humidity: 30 to 85% RH (noncondensing)
Lower limit range: OFF, 1.0 to 19.9 L/min in 0.1 Atmospheric pressure: 700 to 1060 hPa
L/min steps Transport and storage environment: When
CCI alarm: the battery pack is stored more than 6 months,
2
Upper limit range: 1.1 to 20.0 L/min/m in 0.1 charge and discharge or charge the battery
2
L/min/m steps, OFF once every 6 months.
2
Lower limit range: OFF, 1.0 to 19.9 L/min/m in Temperature:
2
0.1 L/min/m steps –20 to +60°C (–4 to +140°F) (within 30 days)
–20 to +45°C (–4 to +113°F) (within 90 days)
◆ esCCO (Optional QP-033P/QP-034P is –20 to +35°C (–4 to +95°F) (more than 90
required. Available for BSM-3562/3763 days)
only.) Humidity: 20 to 85% RH (noncondensing)
Method: estimated continuous noninvasive Atmospheric pressure: 700 to 1060 hPa
cardiac output, pulse wave transit time method ◆ External Output
Measured parameter: esCCO, esCCI, esSV,
ZS-900PK: Available
esSVI, esSVR, esSVRI (To calculate esCCI,
LS-NET: Available, direct connection to
esSVI and esSVRI, patient information (BSA)
Laserprinter
is required. To calculate esSVR and esSVRI,
External display: Available (Option: Interface
ART and CVP measurement is reqired.)
board, QI-372P )
Calibration CO source: CO from patient
Multi-Link Port: 2 (Option: Interface board,
information, CO from Hemo Trend, CCO,
QI-372P)
Manually enter
Calibration pulse pressure source: ART, ART2, ◆ Power Requirement
RAD, FEM, NIBP Line voltage:
Measuring range: AC: AC 100 to 240 V
esCCO: 0.5 to 20.00 L/min, 0.01 L/min steps DC (SB-671P): 8.5 to 12.6 V
2
esCCI: 0.5 to 20.00 L/min/m2, 0.01 L/min/m Line frequency: 50 or 60 Hz
steps Power input: AC 100 VA
esSV: 0 to 300 mL, 1 mL steps
2 2
esSVI: 0 to 200 mL/m , 1 mL/m steps ◆ Clock Accuracy
Alarm range: At operating temperature 25°C: approx. ±2
esCCO upper limit:0.60 to 20.00 L/min in 0.10 min 40 s/month maximum
L/min steps, OFF At storage temperature –20 to +60°C:
esCCO lower limit: OFF, 0.50 to 19.90 L/min in approx. ±6 min/month maximum
0.10 L/min steps

14
Degree of protection against electrical
◆ Environment shock:
Operating environment: Defibrillator-proof type CF applied part: ECG,
Temperature: 10 to 40°C (50 to 104°F) IBP, respiration (impedance method),
SpO 2 accuracy is guaranteed at surrounding temperature, SpO2, CO2, BIS, NIBP, CCO
temperature of 18 to 40°C (64 to 104°F). (APCO)
Humidity: 30 to 85% RH (10 to 40°C, CF applied part: CO
noncondensing) Degree of protection against harmful
Atmospheric pressure: 700 to 1060 hPa ingress of water: IPX1 (protected against
Transport and storage environment: vertically falling water drops)
Temperature: Degree of safety of application in the
–20 to +65°C (–4 to +149°F) presence of FLAMMABLE ANAESTHETIC
–15 to +55°C (Recording paper) MIXTURE WITH AIR, OR WITH OXYGEN OR
Humidity: 10 to 95% RH NITROUS OXIDE: Equipment not suitable for
Atmospheric pressure: 700 to 1060 hPa use in the presence of FLAMMABLE
Cooling: Convection, no fan/filter ANAESTHETIC MIXTURE WITH AIR, OR
WITH OXYGEN OR NITROUS OXIDE
◆ Mechanical Strength Mode of operation: CONTINUOUS
Mechanical strength: Indoor mobile type OPERATION
◆ Electromagnetic Compatibility ◆ Dimensions and Weight (Approximate)
IEC 60601-1-2: 2001 BSM-3532/3552/3562/3572 bedside monitor:
IEC 60601-1-2 Amendment 1: 2004 Dimensions: 370 W × 310 H × 172 D mm
◆ Safety Standard (excluding protruding parts)
Safety standard: Weight: 6.2 kg
EN 12470-4: 2000*
1 BSM-3733/3753/3763/3773 bedside monitor:
IEC 60601-1: 1988 Dimensions: 430 W × 350 H × 172 D mm
IEC 60601-1 Amendment 1: 1991 (excluding protruding parts)
IEC 60601-1 Amendment 2: 1995 Weight: 7.4 kg
IEC 60601-1-1: 2000 WS-371P recorder module:
IEC 60601-1-2 Amendment 1: 2004 Dimensions: 77 W × 73 H × 120 D mm
IEC 60601-1-4: 1996 (excluding protruding parts)
IEC 60601-1-4 Amendment 1: 1999 Weight: 0.35 kg
IEC 60601-1-6: 2006 X075 Battery:
IEC 60601-1-8: 2006* *
2 3 Dimensions: 75 W × 36 H × 96 D mm
IEC 60601-2-27: 2005 - Particular Weight: 0.56 kg
requirements for the safety, including essential QI-372P/QI-373P interface:
performance, of electrocardiographic Dimensions: 32 W × 103 H × 140 D mm
monitoring equipment (excluding protruding parts)
IEC 60601-2-30: 1999 - Particular Weight: 0.17 kg
requirements for the safety of automatic RY-910PA remote controller:
cycling in in-direct blood pressure monitoring Dimensions: 45 W × 35 H × 135 D mm
equipment Weight: 0.08 kg
IEC 60601-2-34: 2000 - Particular Interface QF series:
requirements for the safety of direct blood Dimensions: 65 W × 23 H × 44 D mm
pressure monitoring equipment*
4 (excluding cables)
IEC 60601-2-49: 2001 - Particular Weight: 0.13 kg
requirements for the safety of multifunction Communication cable IF series:
patient monitoring equipment Dimensions: 65 W × 23 H × 44 D mm
ISO 9919: 2005 (excluding cables)
ISO 21647: 2004 ◆ Electromagnetic Emissions
1
* This monitor complies with EN 12470-4: For Electromagnetic Emissions, refer to the
2000 only for clauses 6.2, 6.3 a), 6.5, 6.6, 6.7, Operator's manual.
6.8, 6.9, 6.10 and 8.
2
* Only the “IEC standard” alarm sound ◆ Electromagnetic Immunity
complies with clause 6.3.3.2. For Electromagnetic Immunity, refer to the
3
* This monitor complies with IEC 60601-1-8: Operator's manual.
2006 except for interbed alarm.
4
* This monitor complies with IEC 60601-2-34:
2000 except for clauses 44.6,45.101 a) and
45.101 b).
STANDARD ACCESSORIES
Bedside monitor BSM-3532/3562/3733/3763
Type of protection against electrical shock:
Operator’s Manual 1
CLASS I EQUIPMENT (AC Powered)
BSM-3000 series Manuals CD-ROM 1
Internally Powered EQUIPMENT (BATTERY
Recorder module, WS-371P
Powered)
Recording paper 1
15
ACCESSORIES OPTIONS
Accessory set for BSM-3562/3763 (NK Interface, QI-372P interface (2 x Multi-link
SpO 2 ) ports, RGB, ECG/BP out port), factory option.
(An accessory set must be ordered with Interface, QI-373P interface
main unit.) Hemodynamics Review Program including
esCCO program, QP-033P
Order code Y212A Y212B Y213A Y213B esCCO Program, QP-034P
Model YO-60IY1 YO-60IY2 YO-60AY1 YO-60AY2 Ground lead, 544582
Electrode lead,
Battery, X075 (SB-671P)
3 electrode, clip, IEC,  
K911, B R-903P
Transmitter, ZS-900PK
Electrode lead, 3 Holder for transmitter, DI-590P
electrode, clip, AHA,   Wireless LAN access point, YS-095P5
K911A, B R-903PA
Wireless LAN station, QI-320PA
ECG co nnection cord
  WLAN adapter kit, YS-095P3
K922, J C-906P
ECG co nnection cord Recorder module, WS-371P
 
K922A, JC-906PA Cart, KC-600P
Disposable electrode
    Adaptor for cart, DI-370P
G203, L -150
Counter top mount, KG-600P
SpO 2 connection cord
K931, JL-900P
    Wall mount kit, KG-951P
Air hose for NIBP Wall mount kit, product of GCX (local
   
S902, Y N -901P purchase)
NIBP cuff, adult, 13 cm Mount adapter for GF-210/220R, DH-220P
cuff width,  
S944B, YP-963T
Unit mount for GF-210R/220R with cart,
NIBP cuff, adult large,
DH-223P
15 cm cuff width,   Wall mount adapter for GF-210R/220R,
S944C, Y P-964T DH-221P
Order power cord separately or purchase Touch pen, Y075
locally Multigas unit, GF-110PA
Power cord Multigas/Flow unit, GF-120PA
L936 European version 200-240V Multigas Unit, GF-210R
L935 US version 100-110V Multigas/Flow unit, GF-220R
Connection cable for GF-210R/220R (0.3 m),
Accessory set for BSM-3552/3753 (Nellcor K977 (YJ-600P)
SpO 2 )/BSM-3532/3733 (Masimo SpO 2 ) Connection cable for GF-210R/220R (2.5 m),
(An accessory set must be ordered with K978 (YJ-601P)
main unit.) Sidestream CO 2 unit, AG-400RA/RK
Neuro Unit, AE-918P
Order code Y212C Y212D Unit stand, DM-912P
Model Y O-60IY3 Y O-60IY4 Isolation transformer, SM-800RJ/RK
Electrode lead, 3 Network isolation unit, HIT-100
electrode, clip, IEC,  
BIS processor, QE-910P
K911, B R-903P
Electrode lead, 3
Connection cable to BIS processor
electrode, clip, AHA, (QE-910P), YJ-671P
K911A, B R-903PA APCO module, JP-600P/PK
ECG connection cord
  Interface for Evita series/Savina, QF-901P
K922, J C-906P
ECG connection cord
Interface for A-2000, QF-902P
K922A, JC-906PA Interface for Edwards CCO monitor,
Disposable electrode
 
Vigilance, Vigilance II, Vigileo, QF-903P
G203, L -150 Interface for AG-920RK, QF-904P
Air hose for NIBP
  Interface for AG-400RK, QF-905P
S902, Y N -901P
NIBP cuff, adult, 13 cm
Interface for PB740/760/840 ventilator,
cuff width,  QF-907P
S944B, YP-963T Interface for Servo I/S ventilator, QF-908P
NIBP cuff, adult large, 15 Interface for TOF-Watch SX, QF-909P
cm cuff width, S 944C, 
YP-964T Interface for Pulsion PiCCO monitor,
Order power cord separately or purchase QF-911P
locally Communication cable for connecting an
Power cord external device which outputs analog
L936 European version 200-240V voltage signal, IF-912P
Communication cable for connecting
L935 US version 100-110V Radiometer MicroGas 7650 rapid
transcutaneous monitor, IF-913P

16
Communication cable for connecting K925 ECG connection cord, JC-916P, 3/6
Radiometer TCM4 or TCM40 electrodes, IEC, 1.5 m
transcutaneous monitor, IF-914P K921 ECG connection cord, JC-900P, 10
Interface for Hamilton ventilator, IF-917P electrodes, IEC, 3 m, use with disposable
Communication cable for connecting electrode with lead
Heinen + Lowenstein anesthesia K921A ECG connection cord, JC-900PA,
workstation, IF-918P 10 electrodes, AHA, 3 m, use with disposable
Interface for Drager anesthesia workstation, electrode with lead
IF-920P K901 ECG patient cable, BJ-900P, 10
Interface for Oridion Nicrocap monitor, electrodes, IEC, 3.8 m
IF-921P K901A ECG patient cable, BJ-900PA, 10
Communication cable for Hospira electrodes, AHA, 3.8 m
Q2/Q2Plus/Q-VUE, IF-922P K974 ECG/BP output cable, YJ-910P, 5 m
Communication cable for PB7200 ventilator,
K975 ECG/BP output cable, YJ-920P, 0.3
IF-923P
Communication cable for connecting a m
Disposable electrodes (Vitrode series)
MAQUET FLOW-i anesthesia system,
G203 L-150, for general use, 30 electrodes
IF-932P
Handle for BSM-3500, DH-350P x 5/set, 35 mm dia, Vitrode L
G207 L-150X, for general use, X-ray
QI-RGB cable, YS-080P3, 10 m
(radiolucent), 30 electrodes x 5/set, 35 mm dia,
QI-RGB cable, YS-080P4, 2 m
External monitor (local purchase), NEC Vitrode L
G210D F-150M, for infant/neonate to adult,
LCD191 VXM or equivalent
3 electrodes x 50/set, 25 x 45 mm, Vitrode F
Minimum specifications of external monitor:
• Resolution: 800 x 600 G210C F-150S, for NICU, 3 electrodes x
50/set, 18 x 36 mm, Vitrode F
(BSM-3532/3552/3662), 1024 x768
G221 G-600, for ICU/OR, 30 electrodes x
(BSM-3733/3753/3763)
• Input: Analog RGB 0.7Vp-p 20/set, 47 x 47 mm, Vitrode G
G272A V-090M3, for general use, 3
• V-sync: 60.3 Hz (BSM-3532/3552/3662),
electrodes x 30/set, DIN, 25 x 45 mm, Vitrode V
60.0 Hz (BSM-3733/3753/3763)
• H-sync: 37.9 kHz (BSM-3532/3552/3662), G272B V-040M4, for general use, 4
electrodes x 10/set, DIN, Vitrode V
48.4 Hz (BSM-3733/3753/3763)
G278A V-09IO3, for general use, 3
• Sync signal: negative
–Some commercially available monitors do not electrodes x 30/set, DIN, 25 x 45 mm, Vitrode V
G273A V-04IO4, for general use, 4
work well with BSM-3000 series. Before
electrodes x 10/set, DIN, 25 x 45 mm, Vitrode
purchasing an external monitor locally, you must
confirm whether the monitor works with the V
G272C V-060M6, for general use, 6
BSM-3000 series.
electrodes x 10/set, DIN, Vitrode V
Safety standards: IEC60601-1-1 or must use with
a medical grade isolation transformer. G274A V-06IO6, for general use, 6
electrodes x 10/set, DIN, 25 x 45 mm, Vitrode
Network printer, HP LaserJet 4200N or
V
equivalent
Remote controller, RY-910PA G271A V-120S3, for NICU, 3 electrodes x
40/set, DIN, 20 x 20 mm, Vitrode V
Upgrade software, QS-058P
G300A N-03IS3, for NICU, 3 electrodes x
10/set, DIN, Vitrode N
CONSUMABLES SpO 2
Cord
ECG and Respiration (Impedance) K931 SpO 2 connection cord, JL-900P, 2.5
m
Cable/cord K935 Nellcor SpO 2 connection cord,
K911 Electrode lead, BR-903P, for 3 JL-650P, 3 m
electrodes, clip, IEC, 0.8 m K936 Masimo SpO 2 connection cord, for
K911A Electrode lead, BR-903PA, for 3 LNOP series, JL-630P, 3.6 m
electrodes, clip, AHA, 0.8 m K937 Masimo SpO 2 connection cord, for
K916 Electrode lead, BR-963P, for 3 LNCS series, JL-631P, 3 m
electrodes on one lead, clip, IEC, 0.8 m Probes (BluPRO series)
K912 Electrode lead, BR-906P, for 6 P225F Finger probe, TL-201T, attached to
electrodes, clip, IEC, 0.8 m finger or toe by clip
K912A Electrode lead, BR-906PA, for 6 P311C Finger probe, TL-631T3, regular
electrodes, clip, AHA, 0.8 m size, attached to finger or toe by tape
K922 ECG connection cord, JC-906P, 3/6 P310C Finger probe, TL-630T3, large size,
electrodes, IEC, 3 m attached to finger or toe by tape
K922A ECG connection cord, JC-906PA, P225G Multi-site probe, TL-220T, attached
3/6 electrodes, AHA, 3 m to finger or toe by tape
17
P259 Cottony tape, 340703, secures
probe and cord for stable monitoring, NIBP
stretchable, 20 pcs/set Reusable Cuffs (Latex free)
P267 Probe fastener, 30 pcs/set S943A YP-960T: for infant, 5 cm cuff width
Reusable Probe
TL-651T3 Finger-tip type, 40 kg or more S943B YP-961T: for child, 7 cm cuff width
Disposable probes (BluPRO series) S943C YP-962T: for child, 10 cm cuff width
P203A TL-271T, for adult finger or toe, 24
pcs/set, 0.8 m S944B YP-963T: for adult, 13 cm cuff width
P204A TL-271T, for adult finger or toe, 5 S944C YP-964T: for adult, 15 cm cuff width
pcs/set, 0.8 m
P203E TL-271T3, for adult finger or toe, 24 S944D YP-965T: for thigh, 19 cm cuff
pcs/set, 1.6 m width
P204E TL-271T3, for adult finger or toe, 5 Reusable Cuffs (YAWARA-CUFF series,
pcs/set, 1.6 m Latex free)
P203B TL-272T, for child finger or toe, 24 S949A YP-702T: for child, 10 cm cuff width
pcs/set, 0.8 m S949B YP-703T: for adult, 13 cm cuff width
P204B TL-272T, for child finger or toe, 5 Disposable Cuffs
pcs/set, 0.8 m S945C YP-810P: for infant, 6 cm cuff width,
P203F TL-272T3, for child finger or toe, 24 20 pcs
pcs/set, 1.6 m S945D YP-811P: for child, 8 cm cuff width,
P204F TL-272T3, for child finger or toe, 5 20 pcs
pcs/set, 1.6 m S946E YP-812P: for adult, 10 cm cuff width,
P203C TL-273T, for neonate instep, 24 20 pcs
pcs/set, 0.8 m S946F YP-813P: for adult, 14 cm cuff width,
P204C TL-273T, for neonate instep, 5 20 pcs
pcs/set, 0.8 m S946G YP-814P: for adult, 15 cm cuff width,
P203G TL-273T3, for neonate instep, 24 20 pcs
pcs/set, 1.6 m S946H YP-815P: for adult, 17 cm cuff width,
P204G TL-273T3, for neonate instep, 5 20 pcs
pcs/set, 1.6 m S946I YP-816P: for adult, 18 cm cuff width,
P203D TL-274T, for infant finger or toe, 24 20 pcs
pcs/set, 0.8 m S946J YP-817P: for thigh, 20 cm cuff width,
P204D TL-274T, for infant finger or toe, 5 20 pcs
pcs/set, 0.8 m S948A YP-820P: for neonate, 2 cm cuff
P203H TL-274T3, for infant finger or toe, width, 10 pcs
24 pcs/set, 1.6 m S948B YP-821P: for neonate, 3 cm cuff
P204H TL-274T3, for infant finger or toe, 5 width, 10 pcs
pcs/set, 1.6 m S948C YP-822P: for neonate, 4 cm cuff
Multi-site Y probe width, 10 pcs
P205A Multi-site Y probe, TL-260T (for S948D YP-823P: for neonate, 4.5 cm cuff
adult, child, neonate), 5 pcs/set width, 10 pcs
P260A Sponge attachment tape S for S948E YP-824P: for neonate, 5 cm cuff
TL-260T multi-site Y probe, 24 pcs/set width, 10 pcs
P260B Sponge attachment tape L for Air hoses:
TL-260T multi-site Y probe, 24 pcs/set S903 YN-990P: extension hose, 1.5 m
P256 Ear clip adapter for TL-260T,
reusable S901 YN-900P: 1.5 m
Disposable probes
S902 YN-901P: 3.5 m
P228A TL-051S, for adult/neonate, 5
pcs/set, 0.8 m cable S904 YN-920P: for neonate, 1.5 m
P228B TL-052S, for adult/neonate, 5
pcs/set, 1.6 m cable S905 YN-921P: for neonate, 3.5 m
P229A TL-061S, for child/infant, 5 pcs/set, IBP
0.8 m cable
P229B TL-062S, for child/infant, 5 pcs/set, Becton Dickinson:
1.6 m cable K951 IBP connection cord, JP-900P,
P260C Foam tape, for Becton Dickinson transducer, 3.5 m
TL-051S/052S/061S/062S, 100 pcs Becton Dickinson blood pressure transducers
Nellcor SpO 2 probes are available from are available from Becton Dickinson Company
NELLCOR (www.nellcor.com) or their (http://www.bd.com) or their supplier.
suppliers
Masimo SpO 2 probes are available from
MASIMO (www.masimo.com) or their suppliers

18
Edwards Lifesciences: P924 CO 2 sensor, TG-221T, 3 m
L901 IBP connection cord, JP-920P, for
K987 CO 2 adapter, JG-970P, 0.5 m
Edwards Lifesciences transducer, 3.5 m
Edwards Lifesciences blood pressure R805 Airway adapter, adult, YG-211T, 30
transducers are available from Edwards pcs/set
Lifesciences Corporation (www.edwards.com) R806 Airway adapter, pediatric, YG-213T,
or their suppliers. 30 pcs/set
Biosensors International: V910 Connector, YG-270P, 30 pcs/set, for
K957 IBP connection cord, JP-960P, for YG-213T
Biosensor transducer, 3.5 m
Biosensors blood pressure transducers are
available from BIOSENSORS CO 2 (Sidestream measurement)
INTERNATIONAL (www.biosensorsintl.com) or TM
FilterLine H set, 25 pcs/set
their suppliers. V908F Infant/neonate
Others:
K952 IBP connection cord, JP-910P, 0.3 m V908E Adult/pediatric
TM
FilterLine set, 25 pcs/set
JP-911P IBP interface isolation cable
V908D Adult/pediatric
TM
FilterLine , 12 pcs/set
Temperature
V905A OR/EMS, YG-910P
Thermistor Probe TM
CapnoLine H set, 25 pcs/set
P240B For adult rectum/esophagus, 401J V909D Adult
P241B For child rectum/esophagus, 402J V909E Child
P242D Disc type, 409J V909F Infant/neonate
TM
P249A Probe cover, for 401J, 10 pcs/set Microstream airway adapter, 25 pcs/set
V903F Infant/neonate
P252 Temperature insulation pad, for 409J,
60 pcs/set CO
CO 2 (Mainstream measurement) K962 CO connection cord, JT-950P, 2.0 m

Semi-quantitative method Becton Dickinson:


P903 CO 2 sensor kit, TG-900P (TG-101T + Becton Dickinson injectate supply s ystems
JG-900P), 3 m and thermodilution cathetes are available from
P922A CO 2 sensor, TG-101T, 1 m Becton Dickinson Company (www.bd.com).
K981 CO 2 adapter, JG-900P, 2 m
Edwards Lifesciences:
R801 Airway adapter, YG-101T, 50 pcs/set Edwards Lifesciences injectate supply
systems and thermodilution cathetes are
P907 CO 2 sensor kit, TG-920P (TG-121T + available from Edwards Lifesciences
JG-920P), 3.5 m
(www.edwards.com)
P923 CO 2 sensor, TG-121T, 2 m
Gas monitoring
K984 CO 2 adapter, JG-920P, 1.5 m
GF-110PA/120PA
R804 Airway adapter, YG-111T, 30 pcs/set
V915A Dryline A-Adult, water trap, 613712,
V922 Nasal adapter, YG-121T, for 10 pcs/set
nasal/oral, 30 pcs/set V915B Dryline N-Neonate, water trap, 6137,
V923 Nasal adapter, YG-122T, for 10 pcs/set
nasal/oral, 30 pcs/set, with O 2 supply tube V916A T-piece S-Straight, 613739, 10
adapter pcs/set
V927 Oxygen cannula, 25 pcs/set V916B T-piece E-Elbow, 613748, 10
Quantitative method pcs/set
P905 CO 2 sensor kit, TG-950P (TG-201T + V903F T-piece, for neonate, 618877, 25
JG-950P), 4 m pcs/set
V917C Sampling tube, Adult, 25 pcs/set
P921 CO 2 sensor, TG-201T, 2 m
V917D Sampling tube, Neonate, 25 pcs/set
K982 CO 2 adapter, JG-950P, 2 m Exhaust gas adapter, 6144-008197A
R802 Airway adapter, adult, YG-201T, 30 GF-210R/220R
pcs/set K977 Connection cable, YJ-600P, 0.3 m
R803 Airway adapter, pediatric, YG-202T, K978 Connection cable, YJ-601P, 2.5 m
30 pcs/set
P909 CO 2 sensor kit, TG-970P (TG-221T + V918 Water trap, YG-600P, 12 pcs/set
JG-970P), 3.5 m

19
V920A T-piece S-Straight, YG-620P, 25
pcs/set
V920B T-piece E-Elbow, YG-621P, 25
pcs/set
V919 Sampling line, YG-610P, 3.0 m, 25
pcs/set
Flow/Paw
GF-120PA/220R
P913 Flow adapter, TF-120P
P914 Flow tube, YF-120P, 5 pcs/set
P915 Luer cap, YS-094PO, 10 pcs/set
Recording
A721 Recording paper, FQW50-2-100, 10
pcs/set, 50 mm x 20 m
Y011 Themal head clening pen, 5 pcs/set
BIS
BIS sensors are available from Covidien
(www.covidien.com) or their suppliers.
EEG
L461 EEG connection cord, JE-905P, 2 ch,
3m
L462 EEG connection cord, JE-906P, 8 ch,
3m
H503A EEG disk electrode, NE-113A, 26
pcs/set, 1.5 m, DIN
H503B EEG disk electrode, NE-114A, 26
pcs/set, 0.8 m, DIN
H527 EEG collodion electrode, NE-136A,
12 pcs/set, 0.7 m, DIN
H537A EEG needle electrode, NE-224S, 20
pcs/set, 1.5m, DIN
H471 Disk electrode, BE-910P, 0.8 m
H472 Collodion electrode, BE-920P, 0.8 m
K461 EEG electrode lead, BE-930P, 0.8
m
F507 EEG paste, Z-181BE, 180 g tube x
2/box, ELEFIX

ORDERING INFORMATION
When ordering, please specify the following
information:
1) Power line voltage and its frequency
2) Requirements for consumables
3) Requirements for optional accessories
4) Transmission frequency of any existing
telemetry system—Nihon Kohden’s or any
other manufacturers’
5) Specify part order code in xxxx

20