DEMCON is an acronym of
Thermostatic liquid
Conductivity electrode
Crude Oil Emulsion
Water phase
Magnetic stirrer
Test procedure (1): ⇒
• The conductivity of the water used for mixing also needs to be known
• The test cell is filled with a measured quantity of water for mixing
and thermostated to the test temperature
Test procedure (2):
• The measuring starts from time zero and will continue as long as the
monitored and logged signal indicates that the separation is still in
progress
Conductivity Calibration Curve for mixtures
of separated water and low conductivity water
12 12
10 10
conductivity x10^4 [µS]
6 6
4 4
y = 0.0998x + 0.3184
2 2
0 0
0 20 40 60 80 100
α = 0.106, the slope of the trend line of the calibration curve [∗104µS / m%]
100% 100%
80% 80%
60% 60%
DEMCON 1
DEMCON 2
40% Bottle test 1 40%
Bottle test 2
Bottle test 3 20% DEMCON 1
20% DEMCON 2
Bottle test 1
0% 0% Bottle test 2
Bottle test 3
-20% -20%
0 20 40 60 80 100 0 20 40 60 80 100
time [min]
time [min]
Comparison results of the Bottletest and the DEMCOM
method 30ppm products E1, E2 and E3
Comparison results of
the Bottle test and the DEMCON method
30 ppm products E1, E2 and E3
100%
80%
60%
separated water [m%]
40%
20%
0%
-20%
0 10 20 30 40 50 60 70 80 90 100
tim e [m in]
Bottle tes t 1, E1 Bottle te s t 2, E1 Bottle tes t 3, E1 DEMCON E1
Bottle tes t 1, E2 Bottle te s t 2, E2 Bottle tes t 3, E2 DEMCON E2
Bottle tes t 1, E3 Bottle te s t 2, E3 Bottle tes t 3, E3 DEMCON E3
Comparison of “Bottle test” and DEMCON
Classic “Bottle test”
• Very traditional
• Well excepted by industry
• Relative simple to perform
• Great number of tests are possible
• Registering results is labour intensive
DEMCON method
• No human interferences in reading results
• Registering results is automatized
• Measuring conductivity has a high accuracy
• More details about the demulsification process
• Demulsifier characteristics fed into data files
• Comparing results easier from data files
Conclusions and Recommendations
• The calibration curves are easy to run and good reproducible
• The DEMCON method has only practical value if more cells are
available
Acknowledgement
The authors want to thank Mr. Marcel Huis in ‘t Veld† posthumously for his support
in performing the tests, his critique on interpretation and in discussing the results.
Literature
Babchin, A.J.; Chow, R.S.; Sawatzky, R.P. Adv. Colloid Interface Sci. 1989, 30, 111
Isaacs, E.E.; Huang, H.; Chow, R.S.;Babchin, A.J. Colloids Surf. 1990, 46, 177-192