Professional Documents
Culture Documents
SPECIAL STAINS
GiemsaStaining
Cat. No. SS016
Doc. No. 932-SS016 Rev. No: B
Release Date: 06-Nov-2015
Intended Use
This product is intended for in vitro diagnostic use on the i6000 BioGenex Automated Staining System & also for Manual
staining.
Reagents Provided
Store all reagents at 2-8°C. Do not use after expiration dates as indicated on the reagent labels
Specimen Preparation
Fixation plays an important role in preserving the tissue structures to be visualized using this stain. 10% neutral buffered
formalin is the preferred fixative, however, other fixatives like Carnoy’s may also be used. Ensure that the fixed sections are
adequately embedded in paraffin. Cut tissue sections to 4-5 microns.
1
49026 Milmont Drive, Fremont, CA 94538 Emergo Europe, Molenstraat 15, NL-2513 BH The Hague, The Netherlands
Tel : +1 (800) 421-4149, Fax: +1 (510) 824-1490,
support@biogenex.com
Precautions
Not regarded as a health or environmental hazard under current legislation. This product contains no hazardous material at a
reportable concentration according to U.S.29 CFR 1910.1200 and EC Directive 91/155/EC.
1. Bring reagents to room temperature. Deparaffinize the tissues and hydrate to distilled water.
2. Load the barcode-labeled slides for the appropriate stain into the BioGenex Automated Staining System slide racks. Load
the barcode-labeled reagent vials into the BioGenex Automated Staining System reagent racks.
3. Start Special Stains software on the instrument and check the protocol parameters against the factory default settings listed
below. Change any of the parameters as required. Please note that any parameter change, once saved, becomes the default
setting until changed again. Use of parameter settings other than factory default requires validation by the user.
4. Select “Start Scan” and the instrument will perform the steps listed in the table (if the factory default settings have been
selected).
Mounting Instructions
After staining, dehydrate and clear in xylene as usual and apply permanent mounting medium.
Quality Control
A positive control slide—one that will display positive staining with this stain—should be included in every run.
Trouble shooting
1. Follow data sheet instructions correctly.
2. Gently invert all reagents prior to use.
3. Use the correct parameters as given in the table above
Expected Results
Helicobacter pylori and nuclei stain dark blue, and cytoplasm stains pink
2
49026 Milmont Drive, Fremont, CA 94538 Emergo Europe, Molenstraat 15, NL-2513 BH The Hague, The Netherlands
Tel : +1 (800) 421-4149, Fax: +1 (510) 824-1490,
support@biogenex.com
Bibliography
1. Carson FL: Histotechnology: A self Instructional Text, ASCP Press, Chicago (IL), 1990:191