You are on page 1of 6

Issues in the Appraisal

of Laser Diffraction
Particle Sizing Techniques
Articles have been published that purport to demonstrate the difficulties and problems of
laser diffraction without any mention that these apparent issues are inherent in all particle
sizing techniques. Further, no reference has been made to the advantages that laser diffraction
offers in terms of repeatability, ease of verification and speed of measurement. This article sets
out to redress the balance and illustrate some of the many positive reasons why the technique
has become the preferred particle sizing technique in a wide range of industries.

Paul Kippax
is product manager for
laser diffraction products at
L aser diffraction is probably the most widely
used technique for particle size analysis in the
pharmaceutical industry, with applications ranging
helping to reduce the overall time to market of
new products and in improving product quality.
Whereas the use of laser diffraction-based particle
Malvern Instruments Ltd. from drug development through to production and characterization is widely accepted as a standard tech-
quality control. In drug development, it is used to nique both within and outside the pharma industry, it
understand the functionality of new products, for- has nevertheless been the subject of considerable crit-
mulations and delivery systems. In line with the icism. The requirements for method development and
development of FDA’s process analytical tech- data analysis have been highlighted as being difficult
nology (PAT) initiative, the technique is also being to realise. As a result, both the reproducibility and
used in pharmaceutical production as a highly robustness of the technique have been called into
effective tool for process optimization and control, question, raising doubts about its legitimacy for
as well as for routine batch acceptance testing. By assessing product quality. In addition, a lack of com-
providing a robust technique for particle character- parison between laser diffraction and newer methods
ization, laser diffraction plays an important role in is then cited as evidence of the technique’s failings,
LASER DIFFRACTION PARTICLE SIZING

with claims that the newer technologies surements to be averaged when surement, it is assumed that the mea-
provide a “more real” assessment of reporting a single result, delivering sured sample is representative of the
particle size.1–3 excellent repeatability when com- bulk material being processed. Users
It can be noted, however, that the pared with techniques that deliver must therefore consider
nature of particle size analysis in results based on one-off measure- • the way the primary sample is
terms of how “particle size” itself is ments. collected
defined and how measurements • Ease of verification — as a first • the actual number of particles in
should be controlled, is seldom dis- principles technique, laser diffrac- the primary sample that are
cussed. This article aims to provide a tion requires no calibration and analysed to obtain the size distrib-
review of the laser diffraction tech- can be verified easily using a ution. This second point is of great
nique, addressing the concerns that variety of readily available importance, but is rarely consid-
have been expressed. The intention is NIST-traceable standards. ered when techniques are being
to encourage a more informed A good, impartial introduction to the compared.
assessment of the capabilities of laser technique is given in ISO13320-1, the Accurate primary sampling requires
diffraction for particle size analysis. international standard for laser dif- users to understand and control the
In 1997, Allen stated that “novices in fraction.5 This provides an outline of way a sample is obtained for analysis.
the size measurement area must how it works, the key parameters If slurry or emulsion-based products
understand that most errors in size that should be controlled during any are sampled then there is a need to
measurement arise through poor measurements, and performance overcome sedimentation and/or
sampling and dispersion and not expectations in terms of resolution creaming effects. For powder sam-
through instrument inadequacies.”4 and robustness. United States ples, the natural process of particle
Sampling and dispersion issues are Pharmacopeia (USP) General Test segregation during transit must be
often overlooked when assessing the <429> also describes the technique reversed using a device such as a
potential of different techniques. and the requirements for method spinning riffler. If the primary sam-
These aspects of particle size development and validation within pling process is not controlled then
analysis, as well as variations associ- the pharmaceutical industry.6 The size measurement variations of up to
ated with the sensitivity of different key points made are outlined here. 20% or more can result — much
techniques, are discussed below. greater than the variations associ-
Sample preparation ated with any given sizing technique.
Laser diffraction Sampling Having obtained a representative
Before addressing the criticisms The requirement for good sampling sample, it is also necessary to con-
that have been made, it is useful to is common to all particle sizing tech- sider the way in which particles are
consider the particular benefits of niques. When carrying out a mea- sampled for analysis by the sizing
laser diffraction that have led to its
widespread acceptance within the
pharmaceutical industry. These
include Figure 1 Emulsion measurements obtained using tap water and DI water buffered with
• Range of applicability — sprays, surfactant as dispersants.
dry powders, suspensions and
sprays can all be characterized
using the same technique, allowing
different formulation types to be
0.5
compared in a realistic way.
• Dynamic range — size measure-
ments can be made across a range
from 0.02 ␮m to a few millimetres 0.4
Dv50 (␮m)

in a single measurement, ensuring


that both well-dispersed and
agglomerated particles are
0.3
detected equally well.
• Speed of measurement — single Deionized water
measurements can be made in Tap water
400 ␮sec, allowing the dynamics of 0.2
drug delivery from aerosol devices
to be followed. The effect of
changing dispersion conditions can
0.1
also be assessed, aiding the devel- 0 2 4 6 8 10
opment of robust measurement Measurement number
methods.
• Measurement repeatability — the
ability to acquire data rapidly The instability of the results obtained in tap water is clearly seen when making repeat mea-
allows many thousands of mea- surements. The result variation is less then 1% for the measurements in deionized water.
LASER DIFFRACTION PARTICLE SIZING

technique itself. Here there are


distinct differences between particle
counting methods, such as
microscopy or time of flight (TOF)
measurements, and ensemble par-
ticle size techniques, such as laser
diffraction.
Typical counting-based methods
consider only a few thousand parti-
cles during the course of a single
analysis, whereas laser diffraction
Figure 2 Particle size measured as a function of pressure for a pharmaceutical powder. measures millions of particles. There
can, therefore, be considerable dif-
ferences between the size distribu-
tions reported by these techniques,
0.5 particularly when measuring polydis-
Dv10 perse samples. Commonly, laser dif-
Dv50 fraction will report material that is
Dv90 not observed by the counting tech-
0.4
nique, particularly at the coarse end
Size (␮m)

of the particle size distribution. This


is often put forward as a reason to
0.3
disregard laser diffraction as a reli-
able technique. However, because it
is sampling more material, laser dif-
0.2 fraction often provides a better
assessment of the size distribution
width.
0
Dispersion
0 1 2 8 10 Good agreement between tech-
Dispersion pressure (bar) niques is possible only if a consis-
tent state of sample dispersion is
Each measurement is the average of 5 repeats. The measurement reproducibility was less than 5% at each point. achieved. While the need for disper-
sion control is often highlighted, it
is seldom explored, with authors
presenting single results as indica-
tive of a technique’s performance.
Figure 3 Agreement between wet dispersion and the dry measurements carried out at 0.2 bar It should also be borne in mind that
dispersion pressure. the dispersion methods used for
different measurement techniques
can vary considerably. For example,
microscopy measurements, with
0.2 bar dry dispersion
particles placed on a glass slide, are
Wet dispersion
12 by their nature made in a low
energy environment where the
break up of agglomerates is often
not achieved.
Volume (%)

In stark contrast to this, TOF


8 techniques accelerate particles at
sonic velocities in an air stream
before measurement, subjecting
them to high shear. In the case of
4 laser diffraction, measurements are
made on particles either held in a
liquid dispersant or entrained at
high shear in a compressed air
stream. It is evident that careful
0 optimization of the dispersion con-
0.1 1 10 100 1000 ditions is an absolute requirement if
Particle size (µm) these different techniques are to be
compared meaningfully.
LASER DIFFRACTION PARTICLE SIZING

When assessing the dispersion


stability of a system repeat
measurements must be performed on
the sample under test to ensure the
correct selection of materials and
parameters.5,7 A good example of this
can be seen in Figure 1 where two
sample dispersants are compared.
Here the median particle size (Dv50)
reported by laser diffraction is shown
for a pharmaceutical emulsion. The
rapid measurement capabilities of the
laser diffraction system allow the
dispersion stability to be monitored in
real time for each of the chosen
dispersants. Since the initial results Figure 4 Volume and number distributions reported for the same product.
obtained for each dispersant are
similar, reliance on a single result
would lead to the erroneous
conclusion that either dispersant 10 Volume distribution
would be suitable for performing size Number distribution
analysis. Only by carrying out repeat
measurements is the instability of the 8
dispersion in tap water observed.
Such instability would adversely
Volume (%)

affect the measurement 8


reproducibility.
A similar situation exists for dry
powder analysis. Here, laser
diffraction results are often 8
compared with those from
microscopy. However, the state of
sample dispersion achieved in each 4
of these techniques is quite
different. It is, therefore,
unreasonable to expect the 0
generation of comparable results 0.01 0.1 1 10 100 1000
without first assessing the effect of Particle size (µm)
the applied air pressure on the laser
diffraction measurements. This The number distribution is shifted to smaller sizes compared with the volume distribution.
assessment can be made by The agglomerates reported in the volume distribution are also insignificant in number terms.
following the guidance in
ISO13320-1. Typically, a reduction
in particle size is observed as the air
pressure is increased (Figure 2). It Figure 5 Different equivalent sphere interpretations for an irregularly
is necessary to determine whether shaped particle.
this size reduction is because of
particle dispersion or is the result of
Sphere of same Sphere of same
milling of the sample. This is an minimum length Sphere of
maximum length
often neglected determination that same weight
requires the dry results to be
compared with a stable wet d min dw
d max
dispersion.7 Only when this is done Sphere of
is it possible to establish the air same volume
Sphere have same
pressure needed to achieve sedimentation rate dv
optimum dispersion (Figure 3).
Critically though, this knowledge d sed
allows true assessment of
measurement reproducibility and ds Sphere of same
enables comparison of the Sphere passing same d sieve surface area
performance of laser diffraction sieve aperture
systems with other techniques.
LASER DIFFRACTION PARTICLE SIZING

Data interpretation agreement between different Particle shape


Size distribution representations counting-based methods (micro- The assumption of sphericity, upon
What constitutes an appropriate scope and TOF systems, for which laser diffraction particle sizing
representation of the measured example), these will not always is based, has often been criticised,
particle size distribution for a given agree with either volume or mass- with other techniques being reported
technique depends on the tech- based techniques. as providing a better assessment of
nique’s sensitivity. Laser diffraction Volume-based distributions will the particle size of non-spherical par-
reports the volume of material of a always be shifted to larger particle ticles. However, all techniques use a
given size, since the light energy sizes compared with number similar approach when reporting the
reported by the detector system is distributions (Figure 4). In fact, it is particle size. Particles are three-
proportional to the volume of the mathematically impossible for the dimensional objects and, therefore,
measured particle. This contrasts volume distribution reported by cannot be described by a single
with counting-based techniques, laser diffraction to be shifted to number that equates to the “particle
which report the number of particles smaller sizes compared with the size.” For this reason, all techniques
of a given size. number distribution reported by measure some property of a particle
The differences between number- counting methods, unless and provide the diameter of the
and volume-based size distributions measurements are made at the equivalent sphere as the particle size.
have been discussed at great limits of resolution of the counting It is this approximation that is the
length.8 Clearly, the distributions method. This has not prevented source of differences between sizing
reported by these techniques will results contrary to this basic techniques when non-spherical
not be the same, particularly when principle being reported as objects are measured (Figure 5).
measuring polydisperse materials. evidence of the inaccuracy of laser Each technique is thus subject to
Thus, although there may be good diffraction measurements. some form of error, with no one tech-
nique providing a more real
representation than any other of the
true particle size. USP 776 expresses
Table 1 Correlation between microscope observations and particle absorption. this by stating “For irregularly shaped
Absorption Microscope observation Example particles, characterization of particle
size must include information on par-
0 Transparent spheres Glass beads, latex ticle shape.”9 Selection of a technique
0.001 Spherical, off-white Emulsion droplets should, therefore, be based on its
speed, reproducibility and robustness
or yellow particles
as well as a consideration of the perti-
0.01–0.1 Irregular, translucent milled Most pharmaceuticals nence of the reported size parameter
particles or crystallites and milled materials to the application.
⬎1 Opaque particles Metal particles, pigments
LASER DIFFRACTION PARTICLE SIZING

Analysis parameters materials these techniques provide


For all optically-based particle sizing an orientation-averaged RI that is Laser diffraction provides a robust
techniques, the optical properties of valid for use in laser diffraction means of particle size analysis that has
the materials under test must be con- analysis. The particle absorption
sidered. In microscopy the refractive need only be specified to the nearest
many advantages compared with other
index (RI) difference between the order of magnitude and can be esti- techniques in terms of the degree of
particle and dispersant phases defines mated easily from microscope obser- information obtained.
how well the particle boundary can vations (Table 1). Thus, an informed
be resolved and, therefore, the preci- choice of optical properties can be
sion of any measurements. In TOF or made based on evidence that is 3. D.A. Sesholtz, “Sizing Up,”
light obscuration techniques the RI simple to collect. Pharmaceutical Formulation & Quality,
difference defines the intensity of the If these properties were less easy to May, 79–80 (2004).
light scattering signal observed as determine, it is unlikely that laser dif- 4. T. Allen, Particle Size Measurement 5th
particles pass through the measure- fraction would ever have achieved Edition, Volume 1 (Chapman and Hall,
ment zone, an effect that must be such widespread application in the London, UK, 1997) p. 38.
allowed for by calibrating with latex pharmaceutical industry, and the incor- 5. ISO13320-1 Particle Size Analysis —
particles. In laser diffraction the RI rect selection of optical properties is Laser Diffraction Methods. Part 1:
difference and particle absorption largely inexcusable. It should be noted General Principles, ISO Standards
must be known to calculate the par- that the use of incorrect optical prop- Authority (1999). www.iso.ch
ticle size from the measured scat- erties could never account for some of 6. USP General Chapter <429> “Light
tering pattern. the large differences reported between Diffraction Measurement of Particle
The need to select optical proper- laser diffraction and other techniques Size,” Pharmacopoeial Forum 28(4),
ties for laser diffraction is often mis- (up to 2–3 orders of magnitude is some 1293–1298 (2002).
understood. ISO13320-1 clearly cases). The errors observed are much 7. A. F. Rawle, “Attrition, Dispersion and
states that, for particles greater than more likely to be because of poor Sampling Effects in Dry and Wet Particle
approximately 50 ␮m in size, the sample preparation or the incorrect Size Analysis Using Laser Diffraction,”
Fraunhofer Approximation can be selection of the range lens within the paper presented at the 14th International
used to calculate particle size distrib- laser diffraction system. Congress of Chemical and Process
utions from light scattering data Engineering (Praha, Czech Republic),
without knowledge of the optical Conclusions 27–31 (2000).
properties. At less than 50 ␮m, ana- Laser diffraction provides a robust 8. A. F. Rawle, “The Basic Principles of
lysts need to use the Mie Theory, means of particle size analysis that Particle Size Analysis,” Malvern
requiring the specification of the par- has many advantages compared with Instruments Application Note: MRK038,
ticle RI and absorption (imaginary other techniques in terms of the Malvern Instruments, UK (1993).
refractive index) together with the degree of information obtained. www.malvern.co.uk
dispersant RI, to obtain accurate While it is not being argued that 9. General Chapter <776>, “Optical
results. Papers comparing laser dif- laser diffraction is the method of Microscopy,” USP 24 (US
fraction measurements often neglect choice for every particle sizing appli- Pharmacopoeial Convention, 12601
this requirement, comparing distrib- cation, misuse of the technique is to Twinbrook Parkway, Rockville, Maryland
utions that have been calculated blame for some of the poor compar- 20852, USA, 2000) pp. 1965–1967.
using different models (for example, isons that have been made against 10. www.tde.lth.se/teorel/Publications/
using both the Fraunhofer other methods of analysis. Laser dif- TEAT-7000-series/TEAT-7120.pdf)
Approximation and Mie Theory) and fraction can provide precise, rapid 11. H. Saveyn et al., “Determination of the
presenting these as evidence of the results as long as sampling and dis- Refractive Index of Water Dispersible
poor reproducibility and robustness persion are controlled, and the Granules for Use in Laser Diffraction
of the technique. This is clearly not requirements of the analysis are cor- Experiments,” Part. Part. Syst. Charact. 19,
valid as selection of reasonable rectly understood. As such, it repre- 426–432 (2002).
optical properties is a prerequisite sents an enabling technology, which 12. “Rapid Refractive Index Determination
for accurate measurements. has and will continue to bring huge for Pharmaceutical Actives,” Malvern
Selection of the correct optical benefits to the pharmaceutical Instruments application note: MRK529,
properties need not be an onerous industry within both product devel- Malvern Instruments, UK (2004).
task. The refractive index of pharma- opment and manufacturing control. www.malvern.co.uk ■
ceuticals is generally in the range of
1.38–1.65 and is required only to an References
accuracy of ⫾0.2 to achieve reliable 1. F.M. Etzler and R. Deanne, “Particle Size
Article Reprinted from the
Article Reprint Number: 0643

results. Simple tests such as optical Analysis: A Comparison of Various


index matching or the measurement Methods II,” Part. Part. Syst. Charact. 14, ©January 2005 issue of:
of the RI of solutions containing 278–282 (1997).
known concentrations of the phar- 2. F.M. Etzler, “Particle Size Analysis: A
maceutical under test,9–12 can be Comparison of Methods,” American
used to estimate the RI to the Pharmaceutical Review 7(1), 104–108
required accuracy. For non-isotropic (2004).

You might also like