Professional Documents
Culture Documents
1. Atomic Emission spectroscopy Chemical analysis that uses intensity of light Particularly used in: the form of inductively coupled
emiited from a flame, plasam or spark in plasma atomic emission spectroscopy (ICP-AES) for
wavelength to determine the quantity of an atomization and excitation
element in a sample
The three most important components of atomic The most commonly used sources: Emission spectrum
emission spectrophotometer are as follows: 1. Flame - are capable of producing is composed of a series of very narrow peaks
1. Source temperatures up to 3,000 K (sometimes known as “lines”) with each line at
the sample is atomized at a sufficient 2. Inductively coupled plasma (ICP) a different wavelength and each line matched
temperature to produce an excited-state to a specific transition
species Gases Each element has its own characteristic
Those species will emit radiation are combined in a specially designed mixing emission spectrum
upon relaxation back to the ground chamber Example: sodium can be detected by tuning the
state A sample is also introduced into the mixing monochromator to a wavelength of 589 nm
2. Wavelength selecting device (monochromator) chamber using a nebulizer that converts
for the spectral dispersion of the radiation liquid into a fine spray Interference-free wavelength
and separation of the analytical line from (atomic or ionic line)
other radiation Typical fuel gases: wavelengths producing suitable analytical
3. Detector permitting measurement of radiation include hydrogen and acetylene performance
intensity Example: limit of quantitation, freedom from
A liquid sample is converted into an aerosol Oxidant gases: interferences, and robustness are selected
and delivered into the source where it include air, oxygen and nitrous oxide
receives energy sufficient to emit radiation The first detectors in AES uses photographic film
Both atomic and ionic excited states can be produced
The intensity of the emitted radiation is correlated to depending on the element and the source Contemporary AES instruments feature
the concentration of an analyte and is the basis for leads to the production of complicated photomultiplier tubes or array-based detector
quantitation emission spectra systems
2. Atomic absorption spectrometry (AAS) an analytical procedure for the quantitative most commonly used instrumentation for trace
determination of elements through the absorption and toxic metal analysis
of optical radiation by free atoms in the gas phase
Spectra of the atoms Typical radiation sources for AAS: Most common sources for AAS:
are line spectra that are specific for the
absorbing elements a) Hollow cathode lamps (HCLs) 1) FAAS
contains a quantity of the target element in Flame atomic absorption spectroscopy
Absorption is governed by the Beer-Lambert law: the form of a hollow cylinder Measures: Copper, Iron, and Zinc
are an ideal source for determining most Graphite tubes are the most commonly used
elements by AA atomizers
Tubes are made of high-purity
b) Electrodeless discharge lamps (EDLs) polycrystalline electrographite and
for volatile elements coated with pyrolytic graphite and
The four most important components of AA can be heated to a high
spectrophotometer are as follows: temperature by an electrical
1. Radiation (light) source current
o which emits the spectrum of A small aliquot (usually 20 µL) of
the analyte element liquid sample is placed in the tube
2. Atomizer at the ambient temperature
o in which the atoms of the Heating program: dry the sample -> pyrolyze ->
element of interest in the vaporize -> atomize the sample -> clean
sample are formed
3. Monochromator 2) GFAAS
o for the spectral dispersion of Graphite furnace atomic absorption
the radiation and separation of spectroscopy
the analytical line from other also called flameless or electrothermal AAS)
radiation Measures: Selenium, Cadmium, and Lead
4. Detector permitting measurement of Allows for measurements of both liquid and
radiation intensity solid samples
Common problem: analyte volatility depends
on the molecular form of the analyte and the
Mixing chamber burner: sample matrix
produces laminar flames of high optical Treatment: Chemical modifiers (palladium
transparency nitrate, magnesium nitrate, or a mixture of
both)
Coupled Plasma Mass Spectrometry ICP-MS : a state-of-the-art analytical technique for elemental analysis
Quadrupole Mass Spectrometers the typical mass spectrometer used for ICP-MS
Analyzer consists of four parallel conducting rods arranged in a square Able to well resolve peaks separated by one m/z unit but not able to
array resolve peaks separated by a small fraction of an m/z unit
Applying RF and constant (DC) voltages to the rods, the instrument can Tends to be relatively simple to use and maintain, but the resolution (the
be tuned so that only ions of a specific m/z ratio can pass through the ability to discriminate between closely spaced m/z values) is limited
device to reach the detector
are usually “double focusing sector field” instruments Commercially available high resolution ICP-MS instruments
separate ions of different m/z values via deflection in a magnetic field, are capable of a resolution of 10,000 (10% valley)
with ions of greater m/z being deflected to a lesser degree than those of This is enough to resolve
lower m/z Example: 75As+ from 75ArCl+, both nominally 75 m/z units,
The magnetic field is adjusted to allow only ions of a selected m/z to but which differ by 1 × 103 units when viewed at high
reach the detection system at any given point in time resolution
Electrostatic analyzer Magnetic sector instruments
corrects for certain nonideal effects, allowing the instrument to achieve are not able to resolve elemental isobaric interferences
high resolution Example: 115Sn/115In and 40Ca/40Ar would require resolution much
higher than 10,000
Interferences
are classified as spectroscopic or nonspectroscopic
1. Spectroscopic
Spectral interferences: Various strategies are used to deal with spectral interferences:
generally result from a spectral overlap with the spectrum of the target A continuum source background corrector may be included in the
analyte instrument design
Example: in AA certain molecular species may have broad absorption Alternative is Zeeman background correction, which relies on shifting
spectra that may overlap the line spectra of the elements of interest, the atomic spectral lines by the application of a magnetic field
leading to false elevations of the target element concentrations
A much less common occurrence would be for the absorption spectrum of one
element to overlap with that of another
2. ICP-MS
3. Nonspectroscopic
4. Nonspectral
Voltammetric methods Ion chromatography Gas chromatography–mass Laser ablation ICP-MS (LA-ICPMS)
spectrometry (GC-MS)
Example: anodic stripping can be used for the determination is capable of determination of a method accommodating direct
voltammetry (ASV) and adsorptive of copper, iron, and zinc in blood, cadmium, chromium, cobalt, analysis of solid samples
stripping voltammetry serum, and plasma copper, lead, and selenium in
can be used in determination of for the determination of zinc in urine, copper in serum, and lead
selected metals and are the basis urine in blood
for some point-of-care devices
PandaMT13