You are on page 1of 3

 Trace elements

Element is considered essential if a deficiency impairs a biochemical or functional Trace elements


process  Example: Iron, Copper, and Zinc
 replacement of the element corrects this impairment  Found in milligram per liter or parts per million concentration

Deficiencies Ultratrace elements


 typically impair one or more biochemical functions and excess  Example: Selenium, Chromium and Manganese
concentrations are associated with at least some degree of toxicity  Found in microgram per liter or parts per billion concentrations

Conditions that could result in deficiency: Nonessential trace elements


 Decreased intake, Impaired absorption, Increased excretion and Genetic  are of medical interest primarily because many of them are toxic
abnormalities
Essential trace elements
Dietary requirement  are often associated with an enzyme (metalloenzyme) or another
 the smallest amount of the nutrient needed to maintain optimal function protein (metalloprotein) as a cofactor
and health

 Methods and Instrumentation

1. Atomic Emission spectroscopy  Chemical analysis that uses intensity of light Particularly used in: the form of inductively coupled
emiited from a flame, plasam or spark in plasma atomic emission spectroscopy (ICP-AES) for
wavelength to determine the quantity of an atomization and excitation
element in a sample

The three most important components of atomic The most commonly used sources: Emission spectrum
emission spectrophotometer are as follows: 1. Flame - are capable of producing  is composed of a series of very narrow peaks
1. Source temperatures up to 3,000 K (sometimes known as “lines”) with each line at
 the sample is atomized at a sufficient 2. Inductively coupled plasma (ICP) a different wavelength and each line matched
temperature to produce an excited-state to a specific transition
species Gases  Each element has its own characteristic
 Those species will emit radiation  are combined in a specially designed mixing emission spectrum
upon relaxation back to the ground chamber  Example: sodium can be detected by tuning the
state  A sample is also introduced into the mixing monochromator to a wavelength of 589 nm
2. Wavelength selecting device (monochromator) chamber using a nebulizer that converts
 for the spectral dispersion of the radiation liquid into a fine spray Interference-free wavelength
and separation of the analytical line from (atomic or ionic line)
other radiation Typical fuel gases:  wavelengths producing suitable analytical
3. Detector permitting measurement of radiation  include hydrogen and acetylene performance
intensity  Example: limit of quantitation, freedom from
 A liquid sample is converted into an aerosol Oxidant gases: interferences, and robustness are selected
and delivered into the source where it  include air, oxygen and nitrous oxide
receives energy sufficient to emit radiation The first detectors in AES uses photographic film
Both atomic and ionic excited states can be produced
The intensity of the emitted radiation is correlated to  depending on the element and the source Contemporary AES instruments feature
the concentration of an analyte and is the basis for  leads to the production of complicated photomultiplier tubes or array-based detector
quantitation emission spectra systems

2. Atomic absorption spectrometry (AAS)  an analytical procedure for the quantitative  most commonly used instrumentation for trace
determination of elements through the absorption and toxic metal analysis
of optical radiation by free atoms in the gas phase

Spectra of the atoms Typical radiation sources for AAS: Most common sources for AAS:
 are line spectra that are specific for the
absorbing elements a) Hollow cathode lamps (HCLs) 1) FAAS
 contains a quantity of the target element in  Flame atomic absorption spectroscopy
Absorption is governed by the Beer-Lambert law: the form of a hollow cylinder  Measures: Copper, Iron, and Zinc
 are an ideal source for determining most  Graphite tubes are the most commonly used
elements by AA atomizers
 Tubes are made of high-purity
b) Electrodeless discharge lamps (EDLs) polycrystalline electrographite and
 for volatile elements coated with pyrolytic graphite and
The four most important components of AA can be heated to a high
spectrophotometer are as follows: temperature by an electrical
1. Radiation (light) source current
o which emits the spectrum of  A small aliquot (usually 20 µL) of
the analyte element liquid sample is placed in the tube
2. Atomizer at the ambient temperature
o in which the atoms of the  Heating program: dry the sample -> pyrolyze ->
element of interest in the vaporize -> atomize the sample -> clean
sample are formed
3. Monochromator 2) GFAAS
o for the spectral dispersion of  Graphite furnace atomic absorption
the radiation and separation of spectroscopy
the analytical line from other  also called flameless or electrothermal AAS)
radiation  Measures: Selenium, Cadmium, and Lead
4. Detector permitting measurement of  Allows for measurements of both liquid and
radiation intensity solid samples
 Common problem: analyte volatility depends
on the molecular form of the analyte and the
Mixing chamber burner: sample matrix
 produces laminar flames of high optical  Treatment: Chemical modifiers (palladium
transparency nitrate, magnesium nitrate, or a mixture of
both)

 Sample Collection and Processing

Specimens Specimen consideration


 Must be collected with details:  Care arefully evaluate the following for the use in trace and ultratrace
 anticoagulant, collection apparatus, and specimen type (urine, serum, analyses:
plasma, or blood) o selection of needles, evacuated blood collection tubes, anticoagulants
 Extraordinary measures are required to prevent contamination of the and other additives, water and other reagents, pipettes, and sample
specimen: cups
 low concentration in biologic specimens and the ubiquitous presence in  Recommended measures include:
the environment o placing the trace elements laboratory in a separate room incorporating
 includes using special sampling and collection devices, specially cleaned rigorous contamination control features
glassware, and water and reagents of high purity o Example: sticky mats at doors, non-shedding ceiling tiles, carefully
controlled air flow to minimize particulate contamination, disposable
Sample collection booties worn over shoes, and particle monitoring equipment
 Venipuncture can be used to obtain serum, plasma and whole blood
 24 – hour urine can also be used
 Rarely, hair, nails or biopsy samples may be obtained

Other Methods and Instrumentation:

Coupled Plasma Mass Spectrometry ICP-MS : a state-of-the-art analytical technique for elemental analysis

Plasma Argon plasma


 refers to an ionized gas (typically argon) in which a certain proportion of  induced by commercial ICP instruments (both ICP-AES and ICP-MS)
electrons are free  generates temperatures ranging from 6,000 to 10,000 K
 measures the mass-to-charge ratio (molecular mass divided by ionic  Steps:
charge [m/z]) of selected analyte ions) and includes the following 1. dries and then vaporizes the droplets produced by the nebulizer
components: 2. atomization of any molecular species
(1) an ion source 3. atoms are thermally ionized at which point they are ready for
(2) a mass analyzer introduction into the mass spectrometer
(3) an ion detector
Nearly, all ICP torches consist of three concentric quartz tubes surrounded by a coil
Quantitative analysis for clinical samples carrying radiofrequency (RF) power The middle tube of the torch carries the argon
 is best performed with the use of an internal standard (Ar) that forms the plasma

All patient samples, calibrators, and controls


 are diluted with an internal standard, usually a solution of an uncommon
element such as yttrium

Quadrupole Mass Spectrometers  the typical mass spectrometer used for ICP-MS

 Analyzer consists of four parallel conducting rods arranged in a square  Able to well resolve peaks separated by one m/z unit but not able to
array resolve peaks separated by a small fraction of an m/z unit
 Applying RF and constant (DC) voltages to the rods, the instrument can  Tends to be relatively simple to use and maintain, but the resolution (the
be tuned so that only ions of a specific m/z ratio can pass through the ability to discriminate between closely spaced m/z values) is limited
device to reach the detector

High-Resolution Mass Spectrometers  an ICP-MS instruments

 are usually “double focusing sector field” instruments Commercially available high resolution ICP-MS instruments
 separate ions of different m/z values via deflection in a magnetic field,  are capable of a resolution of 10,000 (10% valley)
with ions of greater m/z being deflected to a lesser degree than those of  This is enough to resolve
lower m/z  Example: 75As+ from 75ArCl+, both nominally 75 m/z units,
 The magnetic field is adjusted to allow only ions of a selected m/z to but which differ by 1 × 103 units when viewed at high
reach the detection system at any given point in time resolution
Electrostatic analyzer Magnetic sector instruments
 corrects for certain nonideal effects, allowing the instrument to achieve  are not able to resolve elemental isobaric interferences
high resolution  Example: 115Sn/115In and 40Ca/40Ar would require resolution much
higher than 10,000

 Interferences
 are classified as spectroscopic or nonspectroscopic

1. Spectroscopic

Spectral interferences: Various strategies are used to deal with spectral interferences:
 generally result from a spectral overlap with the spectrum of the target  A continuum source background corrector may be included in the
analyte instrument design
 Example: in AA certain molecular species may have broad absorption  Alternative is Zeeman background correction, which relies on shifting
spectra that may overlap the line spectra of the elements of interest, the atomic spectral lines by the application of a magnetic field
leading to false elevations of the target element concentrations

A much less common occurrence would be for the absorption spectrum of one
element to overlap with that of another

2. ICP-MS

 spectral interferences include:


a) polyatomic species whose m/z may overlap m/z of the target
analyte
b) arises from nearby elements in the periodic table
c) comes from doubly charged ions

3. Nonspectroscopic

Matrix interferences Properties of significance


 involve the bulk physical properties of the sample to be analyzed  viscosity, presence of easily ionized elements, and presence of carbon
 aqueous samples may behave differently than organic and biological
specimens depending upon the technology used and the analyte of
interest
 Overcome matrix interference by: Matrix matching of the calibrators,
controls, and specimens
 Minimize matrix effect by: Dilution of the specimens but it is only
applicable to certain analytical techniques and to the determination of
analytes with higher concentrations

4. Nonspectral

 anything that could interfere with atomization of the sample  Examples:


 In AA: a flame may not be hot enough for efficient atomization
 Difference in sample viscosity between standards and unknown samples,
resulting in differing rates of sample introduction
 In AES: anything that would prevent the efficient excitation or emission
of spectral lines used for the analysis

 Alternative Analytical Techniques

Voltammetric methods Ion chromatography Gas chromatography–mass Laser ablation ICP-MS (LA-ICPMS)
spectrometry (GC-MS)

 Example: anodic stripping  can be used for the determination  is capable of determination of  a method accommodating direct
voltammetry (ASV) and adsorptive of copper, iron, and zinc in blood, cadmium, chromium, cobalt, analysis of solid samples
stripping voltammetry serum, and plasma copper, lead, and selenium in
 can be used in determination of  for the determination of zinc in urine, copper in serum, and lead
selected metals and are the basis urine in blood
for some point-of-care devices

PandaMT13

You might also like