Professional Documents
Culture Documents
Pronto-VE/SE/XE
TECHNICAL GUIDE
- Principle of Operation -
Tokyo, Japan
L1E-BW1346-3
Page
Section 1 Introduction.......................................................................................................1 - 1
(1) 1 L1E-BW1346
Section 7 Operation principle of scanner unit...................................................................7 - 1
7.1 Structure and summary of scanner unit ............................................................................. 7 - 1
7.2 Scan drive unit................................................................................................................... 7 - 8
7.3 Gantry tilt unit ................................................................................................................. 7 - 11
7.4 Collimator device ............................................................................................................ 7 - 16
7.5 Slip ring ........................................................................................................................... 7 - 22
7.6 Summary of electric control device of scanner................................................................ 7 - 24
7.7 Power distribution board ................................................................................................. 7 - 32
7.8 Stationary side control unit.............................................................................................. 7 - 34
7.9 PCB unit II (Rotation plate PCB unit) ............................................................................. 7 - 53
7.10 Tube current control circuit ............................................................................................. 7 - 56
7.11 Slip ring ........................................................................................................................... 7 - 62
7.12 Preamplifier signal........................................................................................................... 7 - 63
7.13 Tube shift Mechanism ..................................................................................................... 7 - 64
7.14 Shutter mechanism .......................................................................................................... 7 - 72
(2) 2 L1E-BW1346
Section 1 Introduction
This technical guide has been edited for the persons who are engaged in installation, adjustment and maintenance of the
whole body X-ray CT system to enable them to fully understand the mechanisms and functions of this system.
The X-ray CT system is a diagnostic method indispensable for clinical diagnosis of the head, and the clinical application
fields have been extended by the development of the system for diagnosing the whole body by shortening the scan time.
The scan system in the initial development period of the CT system was the pencil beam type to scan the surrounding of
the head by means of fine X-ray beams while repeating the linear move and rotary move. (See Fig.1.1 (a))
On the other hand, a system has been developed to reduce the traverse count and reduce the scan time to 20 ~ 30
seconds by increasing the number of detectors and transmitting corresponding 3o ~ 10o sector X-ray beams a shown in
Fig. 1.1 (b) for the purpose of shortening the scan time. This system is called the second generation system. The
pencil beam type is called the first generation system. Since this second generation system can obtain good images at
comparatively low costs when it is applied to such inspection objects as their internal movement is negligible like the
head, it serves as a leading scanning system of the CT system for the head and neck as the translate/rotate system.
In order to accelerate the scan to be higher, a system has been developed to terminate the scan by rotating the X-ray tube
and detector group only without any linear scan as a result of widening the X-ray beam open angle and preparing the
number of detectors covering the sections of a patient (object). (See Fig.1.1 (c)) This system uses 300 ~ 1000
o
detectors and rotates the X-ray tube and detector group by 360 in 0.8 ~ 4 seconds. This system is called the third
generation CT system as compared with the preceding two systems. In addition, a system has also been developed to
execute scanning by rotating the X-ray tube only while arranging about 800 detectors around the patient as shown in Fig.
1.1 (d).
This system is called stationary/rotate system or the fourth generation system.
The first stage of developing this system was advantageous over the third generation system from the viewpoints of the
image quality.
However, as a result of improving the image quality by increasing the number of the detector in the third generation
system, the rotate/rotate system (the third generation system) plays a leading role as the scan system of the whole body
CT system at present. Our whole-body CT system also adopts the rotate/rotate system.
This technical guide was originally drafted, approved and supplied by Hitachi Medical Corporation in the English
language.
1−1A L1E-BW1346
1°rotation by every translation scan
Rotate every
3°∼10°scan
X-ray source
X-ray source
走査
1
an
120
Sc
Patient
Patient Head
tor
te c
De Detector
(8 ~ 30 elements)
Linear move
Patient
Patient
Rotation
X-ray source
Detector
(300 ~ 1000 devices)
1−2 L1E-BW1346
Features of CT
Difference of CT from conventional system X-ray radiography and tomography
In the conventional X-ray radiography, a patient (object) is put in the X-ray beam flux irradiated from the X-ray
source as a spot irradiation source to record an image on the two-dimensional recording medium provided at the
back of the object as shown in Fig.1.2. Accordingly, the density at each point on the film is determined by the
total absorption quantity when the X-ray beams are passing through the object in the system consisting of the
intensifying screen as a recording medium and films. In other words, an image obtained by the above
radiographic method is a two-dimensional projection image of a three-dimensional body, so that all pieces of
information in the thickness direction of the object are recorded by overlapping them with each other.
In the tomographic method, on the contrary, only an image of a section perpendicular to the X-ray beams is
obtained by synchronously moving the X-ray source and recording system. In the rotary traverse tomographic
method, a tomographic image can be obtained in the direction perpendicular to the long axis of the human body.
In these conventional tomographic methods, however, (1) out-of-focusing depending upon the distance from the
focal plane occurs, (2) the image on the film is overlapped with out-of-focus images before and after the focal
plane to cause the reproduction of the low contrast information to be difficult even in case of focus image, and (3)
the images other than the image on the focal plane are produced as intrinsic artificial out-of-focus images
according to the radiation source and moving direction and moving quantity of films to cause the image quality to
be limited.
In the CT system, the X-ray fine beams thinly collimated to
X-ray source
the direction perpendicular to the long axis of the human
body are projected in multi-directions and the transmitted
X-ray intensity is measured by high sensitivity X-ray
detectors every projection to produce and display a
distribution image of the X-ray absorption coefficient of the
Object (patient) layer based on individual measured values. Accordingly,
the overlap of images other than the layer does not occur
basically in CT images, so that a precise layer image is
obtained. This is the greatest feature of the CT system.
Image contrast is improved due to the absence of
overlapping of images from those other than the layer. Since
the X-rays are finely (or thinly) collimated, the sharpness of
Recording medium
images is improved without being noticeably affected by
scattered rays.
1−3 5 L1E-BW1346
Section 2 Basic principle of system
606
type detectors are arranged to be a
ng are
suri
opposite to the X-rays to measure the Mea 20
998
φ4
X-ray intensity distribution.
Out of the total number of channels of
these detectors, 4 channels each on Xe gas ionization
chamber detector
both ends are used as reference Ref
4ch Ref
channels. The expansion ratio of the 4ch
568ch
X-ray source is 1.65, and the opening
576ch
size per channel at the measuring area
center is 0.75mm.
Fig.2.1.1 Measuring geographic system
2−1 L1E-BW1346
2.2 Solid detector data collection (Pronto-VE/SE)
2.2.1 Measuring geographic system
The measuring geographic system of this CT X-ray tube
is shown in Fig.2.2.1.
The X-ray source can irradiate sectorial
X-rays and solid type detectors are arranged to 606
Pronto-VE
2−2 L1E-BW1346
2.3 Offset detector
The X-ray source is opposed to the detectors as shown in Fig.2.1.1. This system adopts the offset detector
system to displace the detectors by 1/4 sample in the 1/4 offset
Opposed
addition data
x=Xcosθ−Ysinθ
y=Xsinθ+Ycosθ (2.1)
When the X-ray beams having intensity Io are irradiated in parallel to the X-axis, the X-ray intensity I
after penetrating an object f (x, y) is given by
I / I0
∞
=exp − f (Xcosθ−Ysinθ), (Xsinθ+Ycosθ) dY (2.2)
∞
The logarithmic conversion g (X, θ) of the damping factor of this X-ray intensity (called projection
data) is given by
g (X, θ)
∞
= f (Xcosθ−Ysinθ), (Xsinθ+Ycosθ) dY (2.3)
∞
Image reconstruction is executed to obtain the absorption coefficient distribution f (x,y) by giving g (X,
θ) shown in the above relation within a range of 0≦θ≦2π.
2−3 L1E-BW1346
X
1
g(X, θ) g(X, θ 1) X θ
k(
in
1+
ys
x1,
θ
X=
θ
0 os
co
0 xc y
1
y =
)
sθ
Y X
f(x, y)
1
X 0
+y
sin
P(x, y)
θ
θ 0
1
0 x (x, y) x
x
Back projection
Back projection is available as a simple and intuitive method to estimate the distribution (x,y) of the
absorption coefficient from projection data g (X, θ).
We can obtain a distribution close to (x, y) when reversely projecting the values on projection data by
means of X-ray beams which pass the target point (x, y) ,and then, overlapping them with each other in
all directions as shown in Fig.2.4.2.
The X coordinate of coordinate system XOY tilted by angle θfrom coordinate system x0y to point (x,
y) is represented by;
Thus, the image reconstructed by the back- projection method is represented by;
2−4 L1E-BW1346
2.4.2 Image reconstruction flow
As described above, this system executes high-speed arithmetic operation with improved image quality
by adopting the [angle arrange method] and [filter correction back projection method] as the methods of
obtaining the CT value at each part in the image processor from projection data obtained by the detectors
This image reconstruction flow is explained below, referring to Fig.2.4.3.
In the pre-processing of raw projection data obtained by scan measurement, the sensitivity correction of
detectors and preamplifier and offset correction are executed, and the arrangement processing is executed
to rearrange the fan beams into parallel beams. These results are processed by Fourier transformation to
obtain frequency spatial data. Filtering (equivalent to the convolution operation on the actual space) is
executed by multiplying these data and filter function on the frequency space to correct the out-of-focus
condition, and then, projection data are returned to the actual space by Fourier inverse transformation.
Data returned on the actual space are back-projected on the parallel beam to obtain a CT image.
Frequency
spatial Parallel
Scan Pre- Fourier
Fourier data Filtering beam CT
measu- processing Arrangement inverse
transformation (Multiplication) back image
rement (correction) transformation
Filter projection
function
RAW Data
Parallel beam
X-ray source conversion
G
Sensitivity correction
and offset correction W G
Logarithmic X
Detector
transformation W
G
Filter function
W
2−5 L1E-BW1346
K(μ−μw) μ
CT−No= =K −1 ................... (2.6)
μw μw
where
K : Scale factor (1000)
μ : X-ray absorption coefficient value of the tissue as an object
μw : X-ray absorption coefficient value of water
Assume that K=1000.
CT-No (air) ≒−1000 (μ≒0)
CT-No (water) =0 (μ=μw)
CT-No (bone) ≒+1000 (μbone≒2μw)
(Air) (Water)
− 1000 − 200 − 100 0 40 80 4000
− 400 − 180 6 38
Lung Pancreas
32 58 Kidney
26 84 Blood
40 76 White matter
68 104
※ By X-ray tube voltage of 120kVp Gray matter
16 56 Heart
16 80 Liver
2 32 Small intestine
20 66 Primary tumor or
metastasized tumor
2−6 L1E-BW1346
4000
(White)
60 White
50
Window width
35
CT value range
10 Black
(Black)
− 30
− 1000
2−7 L1E-BW1346
2.5
Scanner
table
control Starter
System configuration
HV HV
generator generator
Fig.2.5.lshows the system configuration.
Detector
Preamplifier
2−8
AC200V
L1E-BW1346
2.6 Communication control
2.6.1 Summary
A microcomputer (hereinafter abbreviated to as μ-CPU) is used in the CT system described in this
chapter for controlling each unit as shown in Fig.2.6.1 (a), and each unit is uniquely controlled by μ
-CPU. As a means of connecting the entire CT system systematically, the image processing unit and
scanner μ-CPU are connected by memory shared controller MKY28, while the scannerμ-CPU and
X-ray control unit as well as the scannerμ-CPU and the rotation plate control unit are connected by
using a serial communication cable conforming to RS-232C standard.
Each unit can be divided into (1) image processing system composed of operation unit and image
processing unit and (2) scanner system composed of the scanner, table, rotation plate, and the X-ray
control unit. The operation unit plays an interface role between the CT system and an operator to
transfer the operator’s desired operation to the image processing unit. Accordingly, the image
processing unit and operation unit communicate with each other frequently. (See Fig.2.6.1 (b)) On the
other hand, the scanner system is connected by the μ-CPU called scanner μ-CPU as a pivotal unit
The scanner μ-CPU shares the information of individual units collected by the scanner, table, rotation
plate, and μ-CPU of X-ray control unit by means of communication, and it offers the information as
occasion demands according to the demand of the image processing unit. (Fig.2.6.1 (b))
2−9 L1E-BW1346
Image processing unit Communication cable Scanner μ-CPU Rotation plate μ-CPU
(shared memory)
SH7032 8085
Pentium
Communication cable(RS-232C)
Internal
communication
X-ray control μ-CPU
8085
Pentium
Operation unit
μ-CPU type
Entire information
Scanner system information
Operation unit information Scanner information
Image processing information X-ray control information
Scanner system information Table information
Table information
2−10 L1E-BW1346
2.6.2 Communication method
An RS232C cable and memory shared controller MKY28 are used for communication. Serial
communication will be described below.
(1) RS-232C
The circuit between scanner μ-CPU and the rotation plate control unit as well as the circuit
between the scanner μ-CPU and X-ray control unit are connected through serial communication.
Fig.2.6.2 shows the hardware configuration.
Since the input/output signal cables of μ-CPU are parallel signal cables, while the communication
cables are serial signal cables, the basic hardware configuration is composed of a conversion circuit
for matching the difference of these signal cables. When μ-CPU signals are sent to the
communication lines, μ-CPU inputs parallel data into a communication IC. This communication
IC sends these parallel data sequentially one by one at specified time intervals according to the
transmission instruction to convert them into serial data and then, sends these data to the
communication cables. On the other hand, the serial signals being input through communication
cables are stored into the communication IC sequentially, and read into μ-CPU after they have
been converted into specified parallel signals.
(2) Memory shared controller MKY28
The image processing unit and scanner μ-CPU are connected to each other by memory shared
controller MKY28. Its configuration is shown in Fig.2.6.3.
This controller executes high-speed high-quality serial transmission by its unique 2-signalss level
synchronous system. If communication is in error, this MKY28 re-sends signals automatically to
ensure enhanced reliability of communication. This MKY28 is provided with a memory to be able
to store the information equivalent to 256 characters to reduce the burden of each μ-CPU.
2−11 L1E-BW1346
Parallel data
Serial data
[Communication cable]
Parallel data
Communication IC
D1
Start marker
Serial data
Stop marker
D1 D8
D8
0 t1 2t1 t
0 t0
2−12 L1E-BW1346
Feed side Bit No 7 8 8 12 15 Receive side
PASS STP ADDRESS DATA CRC - 12 clock SEE
MKY28 MKY28
(For send) (For receive)
MKY28 MKY28
(For receive) (For send)
2−13 L1E-BW1346
Rotation Rotation
Image unit Scanner plate WB Image unit Scanner plate WB
89
C4 07 C3 : X-ray high voltage ON
C4 :Error reset
BE
Offset
06 measurement
87 CC :Status read
07
97 04
08 Regular
measurement
81 05
E1 :Projector goes out
C1 :Preparation for measurement Table feed start
C6 :Measurement end
Scanner rotation CC :Status read
start or X-ray 07
C3 : X-ray high voltage ON
tube positioning
Offset
Collimator setting measurement
06 Waiting for table feed end
CC : ステータスリード
87 CC : ステータスリード
97 8A
87 CC :Status read
97
Shared memory RS - 232C
MKY28 B0
80 C0 : X-ray tube high voltage OFF
CC :Status read
C4 :Error reset
2−14 19 L1E-BW1346
Section 3 Operation principle of X-ray generation system
The X-ray generator system is composed of the X-ray control unit, plus (+) and minus (-) side high-voltage
generators, filament control unit and, X-ray tube assembly mounted on the scanner rotation plate.
This X-ray generation system is characterized with the following features.
① It adopts the micro computer (hereinafter referred to as μ-CPU) as a major control element, and all circuits
are controlled by this μ-CPU directly or indirectly.
② The high voltage generator is designed to be compact by adopting a high frequency resonance type inverter
system on the primary side of main circuit.
3−1 L1E-BW1346
X-ray control unit Scanner rotation plate unit
High voltage
generator
Anode
rotation
Filament drive
3−2
Filament
control unit
kV,mA signals
Anode
rotation
Control unit drive unit
kV signals
Communication line
Main power line
Control power line
Scanner unit
Signal line
(Emergency, X-ray trigger signal, X-ray start/abort, kV signals, and kV,
mA abnormal signals are sent as individual signals)
L1E-BW1346
As described above, X-ray generation sequential control is mainly executed by μ-CPU. If the
tube voltage, tube current, or main circuit primary is abnormal, the system is stopped by hardware
logic for a quick and secure response.
(2) Main circuit DC voltage, tube current, and tube voltage control
Control parameters are output from μ-CPU according to the selected kV, mA conditions. The
main circuit DC voltage, tube current, and tube voltage are feedback-controlled based on these
control parameters. And DA converter is indispensable for controlling them based on the control
parameters from μ-CPU, and a DA converter trouble causes the generated X-ray trouble. For
DA-converted parameters, DA converted analog value parameters are read into μ-CPU through an
AD converter again, and compared with output parameters to monitor the DA converter for its
troubles.
● Maintenance circuit
This maintenance circuit is provided with a maintenance program start switch for interrupt
processing, a digital switch for selecting the maintenance program, and a status indicator.
● Data input interface circuit
This circuit is provided to interface between μ-CPU circuit and peripheral control circuit to
enable μ-CPU to read data obtained by DA-converting analog values such as X-ray generation
system status, main circuit DC voltage, etc.
3−3 L1E-BW1346
● Data output interface circuit
This circuit is provided to interface between μ-CPU circuit and peripheral control circuit to
give the control information from μ-CPU to each control circuit. The control information
comprises parameters of tube voltage and main circuit DC voltage. These parameters are
DA-converted in this circuit and sent to the control circuit. This circuit is also provided with
counter type timers that can be mode-set and preset from CPU. These timers are used for the
following purposes.
Uses Setting
Anode rotation hold timer This timer is set to 10 minutes from CPU in
standby, X-ray ON, and X-ray off (continuous
scan) modes.
In case of Varian 3.5MHu Timer is set to
250sec.
Backup timer This timer is set to 110% of the scan time from
CPU when X-ray is turned on
Scan counter clock A reference clock corresponding to the scan time
is set from CPU when the scan without stop time
is selected
(2) XCONT3 circuit will be described together with the drive circuit by function.
3−4 L1E-BW1346
Scanner unit Communication μ -CPU
circuit circuit
Data input
interface Anode rotation drive circuit
Control circuit
circuit Power supply circuit
Maintenance
circuit
DATA
output
interface Rectifier circuit Inverter
circuit
3−5
XCONT1 circuit XCONT3 circuit
L1E-BW1346
3.2 Anode rotation drive circuit
3.2.1 System with Hitachi-made 1.5M / 2 / 3MHU X-ray tube assembly
System with Varian-made 2MHU X-ray tube assembly
The anode rotation drive circuit (hereinafter referred to as starter) is exclusively used for low speed. It
drives the X-ray tube anode at 50 or 60Hz.
Fig.3.3 shows the operation sequence, and Fig.3.4 shows its main circuit.
When ROTOR START signal turns on, the rotor starts running, and when it turns off, the rotor is braked
to terminate the sequence. When 2DR turns on, a gate current flows through the route shown by arrow
② in Fig.3.4, so that 201FLS ignites at almost zero phase to cause a start voltage close to the power
voltage to be applied between stator coils 7-9 of the X-ray tube assembly. This condition remains
unchanged until 5DR is turned on by BOOST TIMER (1VR start time adjustment). When 5DR turns on,
the status is transferred from the start condition to the steady-state condition. In this case, the gate
current flows through the route shown by arrow ③ in Fig.3.4.
Since the gate current does not flow until the 15C voltage exceeds the break-over voltage of 1TD (trigger
diode), 201FLS ignites with the phase delayed by the time constant by CR. The steady-state voltage can
be changed by the phase control angle of 201FLS. (3VR Steady-stage voltage adjustment)
When ROTOR START signal turns off and 222DR turns on, the gate current flows through the route
shown by arrow ① in Fig.3.4, so that 201FLS turns on in its half wave only and a half-wave rectified
DC voltage is applied to the stator main coil to cause braking. This condition continues until 222DR is
turned off by the brake timer. (2VR braking time adjustment)
1SCR and 2SCR are provided to detect the stator coil (phase coil, main coil) current, while 3SCR is
provided to detect a steady-state voltage failure. (over-voltage) If the ignition angle of the current
flowing to the stator coil is detected to be excessive by 1, 2 SCR under the steady-state condition or if
3SCR turns on, error detection relay 7DR is turned on and held. As a result, main relay 203DR turns off
to stop feeding power to the stator coil. The hold condition of 7DR is reset when ROTOR START
signal turns off to transfer to the brake condition.
3−6A L1E-BW1346
1DR
2DR
3DR
203DR
5DR
6DR
4DR
222DR
7DR
STATOR
COIL
STATOR
COIL
3−7 L1E-BW1346
203DR 1
7∼8ZD
2DR 5DR (at boost time)
201FLS
5DR 3
(Steady-state) 1TD
3VR
AC200V 9 36R
38R
PHASE MAIN
14C 39R
8 7
201C1, 2
222DR
212R
Stator current detection
3−8 L1E-BW1346
3.2.2 System with Varian-made 3.5MHU X-ray tube assembly
3.2.2.1 Control circuit
Function of microcomputer (μ-CPU)
(1) Sequence control from starting to stopping rotation of the X-ray tube anode
Theμ-CPU of the driving device is connected with the control circuit of the X-ray control unit
through the signal line. Thisμ-CPU works according to control signal from the X-ray control unit.
The table below shows operation of theμ-CPU of the driving device responding to control signal
from the X-ray control unit.
As shown in the above table, the anode drive sequence is controlled mainly by theμ-CPU. If an
abnormal drive voltage or current occurs, putting emphasis on responsibility and reliability, stop
operation system by hard logic is employed.
3−9 L1E-BW1346
Inverter output
transformer
Phase delay capacitor
Rectification Current
Power supply unit Inverter
and detection
1φ circuit X-ray tube unit
smoothing circuit High / Low circuit
AC200V / 208V starter coil
Power speed relay
supply selection
circuit circuit
3−10
Control unit
(Signal) Control
circuit
Setting block
Display block
L1E-BW1346
3.2.2.2 Composition of control circuit
The composition of the control circuit is shown in Fig. 3.4.2.
Inverter
Inverter circuit
control
circuit
Abnormality
detection circuit
μ− CPU
circuit A/D
converter
Voltage and current
detection circuit
A / D circuit
Control signal Input / Output
To control circuit of circuit High / Low speed
X-ray control unit selection relay
● μ-CPU circuit
H8 micro computer and its peripheral circuits
This circuit also incorporates the counter-type timer that can set and preset mode by theμ-CPU for
the following purposes.
Inverter controlled interval time μ-CPU sets 1/4 cycle time for High/Low speed
Inverter controlled reference clock μ-CPU sets the reference clock for High/Low speed.
Sequence control timer μ-CPU sets time necessary for sequence control
3−11 L1E-BW1346
● Input / Output circuit
<Functions>
(1) Transmission and reception of signals to and from X-ray control unit (Rotor start, High speed, Rotor
up, High speed verified)
(2) Output of selection signal to relay circuit (High speed selection relay, Brake selection relay)
(3) Input of hard lock signal from A/D circuit (Overvoltage, Overcurrent)
(4) Output of prohibition signal to inverter control circuit
(5) Input of DIP switch signal
(6) Input of rotary switch signal
(7) Output of LED display signal
As shown above, this is the interface circuit between theμ-CPU and its peripheral circuits, the
circuit outputs control signals from theμ-CPU to each control circuit and inputs data from those
peripheral circuits.
● A/D circuit
<Analog signals to be input>
Note: The signals of (1) ~ (4) are also input to abnormal detection circuit. (Overvoltage and
overcurrent)
The abnormal detection circuit is provided, as shown in Fig. 3.4.3, with the variable
resistor for setting abnormality detection. If voltage exceeds the set value, the
circuit prohibits the operation of the inverter, reads the abnormality into the μ-CPU
and displays an error through the input/output circuit.
Analog signals such as set voltage, current and variable resistance values is also
A/D-converted and read into the μ-CPU as data.
3−12 L1E-BW1346
High speed
Low speed
VR setting
Inverter operation
prohibit Inverter control circuit
Voltage / Current Full wave Comparator F/F
detection rectification
circuit
3−13
Input / Output
Display circuit
circuit
A/D
μ− CPU
converter
L1E-BW1346
As shown in Fig. 3.4.4, the output voltage is stabilized by software feedback control (PWM control).
Output voltage
detection
μ -CPU
Full wave
A/D
3−14
rectification circuit
L1E-BW1346
● Inverter control circuit
By setting each data by theμ-CPU, inverter waveforms shown in Fig. 3.4.5 and 3.4.6 are output.
① Phase 1 data (ON time data)
② Phase 2 data (ON time data)
③ Dead time data (to prevent short-circuit of arm)
④ Segment data (By dividing the basic clock, count frequency is produced.)
3−15 L1E-BW1346
Dead time Dead time Dead time Dead time
GTR1 ON Phase 1
(Upper arm) data
OFF
GTR1 ON
(Lower arm)
OFF
GTR2 ON
3−16
Phase 2
(Upper arm) data
OFF
GTR2 ON
(Lower arm)
OFF
Segment data Segment data
1 cycle 1 cycle
Fig. 3.4.5 During Boosting Period Fig. 3.4.6 During Running Period
L1E-BW1346
3.2.2.3 Anode rotation drive sequence
Inputting Rotor start signal starts rotation according to the High/Low speed mode.
When High speed is selected, High speed verified signal is output. If the output voltage, input current
and output current exceed the reference values after a starting period passes, Rotor up signal is output and
the running operation mode is selected.
Turning off Rotor start signal selects the braking mode after the preset internal hold-on time.
After the braking time passes, the braking operation ends and the sequence is completed.
The operation time charts are shown in Fig. 3.4.7 and 3.4.8.
3−17 L1E-BW1346
Power ON Power ON
Hold ON Hold ON
time time
Rotor up 14LD
3−18
Inverter drive 15-18LD
Output
L1E-BW1346
Power ON
Hold ON
Hold ON time
Rotor up 14LD
3−19
Inverter drive 15-18LD
Output
L1E-BW1346
3.2.2.4 Error codes list
Error Detection
Error code Cause Remedy
release method
ROM SUM check error Turn off the power once and then turn it on Power Software
again. down.
Or replace EPROM.
RAM write error Turn off the power once and then turn it on Power Software
again. down.
When powered up, Rotor Turn off Rotor start signal. Same Software
up signal is not turned on. as left.
Over-output voltage Check input voltage. Power Hardware
down. (T2)
Over-input current Check inverter circuit. Hardware
(CT1)
Over-output current in Check coil resistance of main coil of X-ray Power Hardware
main coil tube unit. down. (CT2)
Over-output current in Check resistance of sub-coil of X-ray tube Power Hardware
sub-coil unit. down. (CT3)
Zero output voltage Check F1 fuse for blown down. Power Software
down. (T2)
Zero input current at Check F1 fuse for blown down. Power Software
starting down. (CT1)
Zero input current at Check F1 fuse for blown down. Power Software
braking down. (CT1)
Zero current in main coil Check connection of output cable and Power Software
during operation resistance of main coil. down. (CT2)
Zero current in main coil Check connection of output cable and Power Software
at braking resistance of main coil. down. (CT2)
Zero current in sub-coil Check connection of output cable and Power Software
during operation resistance of sub- coil. down. (CT3)
Insufficient output current Check connection of output cable and Power Software
at starting resistance of both main coil and sub-coil. down. (CT2)
Insufficient output Check connection of output cable and Power Software
voltage and current at resistance of both main coil and sub-coil. down. (CT1, T2)
starting Check input and output voltage
3−20 L1E-BW1346
3.3 X-ray main circuits (primary)
Fig.3.5 shows the X-ray main circuits. The main circuits consist of the power filter circuit, rectifier circuit, DC
smoothing circuit, inverter circuit, serial resonance capacitors, and resonance circuit correction coil.
(1) Rectifier circuit
The rectifier circuit adopts the three-phase full-wave thyristor rectification system. It controls the DC
voltage to the inverter by controlling the thyristor ignition angle.
(2) DC smoothing circuit
The rectified DC voltage is smoothed by an LC filter and supplied to the inverter circuit. The DC voltage
output is feedback-controlled to eliminate an error to its set value.
(3) Inverter circuit
The IGBT (Insulated Gate Bipolar Transistor) full bridge inverter is composed of INV1 ~ 4. This
resonance type inverter uses the leak inductance of a high voltage transformer, the inductance connected in
series, and the capacitor of serial resonance capacitors as resonance elements. The tube voltage is
feedback-controlled by using the output voltage of the high voltage divider of positive and negative high
voltage generators mounted on the scanner rotation plate by connecting the output voltage to the X-ray
control unit via a control circuit.
The feedback control of the tube voltage is performed by controlling the INV4, INV2 turn-on phase
difference when paired INV1 and INV4 turn on or paired INV3 and INV2 turn on. In other words, the
feedback control is performed by controlling the overlapped phase of INV1 and INV4 or the overlapped
phase of INV.3 and INV.2. Fig.3.6 shows the inverter control method.
An inverter circuit over-current is detected by utilizing such a symptom as the IGBT collector ~ emitter
voltage VCE and the gate ~ emitter voltage VGE become high when an over-current flows. The inverter
output current is detected by a current transformer to detect an over-current when it flows.
3−21 L1E-BW1346
3.3.1 DC voltage control circuit
Fig.3.7 shows the DC voltage control. Since the X-ray tube voltage is feedback-controlled by the inverter,
DC voltage does not directly affect the X-ray tube voltage, but its voltage value is controlled according to
the X-ray conditions for the purpose of reducing the switching loss of the inverter. The DC voltage is
controlled by controlling the ignition angle of the thyristor by means of the thyristor gate drive signal
being synchronized with the three-phase power supply. Also, the DC voltage is input to the control
circuit from the smoothing circuit via an isolated amplifier and feedback controlled based on DAC output
REC reference. The thyristor gate drive signal is output when REC START signal is input, and the
output is stopped if kV, mA, INV operation, or DC voltage is abnormal. Also, the DC voltage is read
into μ-CPU by ADC, and the output is stopped, if abnormal.
3.3.2 Inverter control circuit (X-ray tube voltage feedback control circuit)
Fig.3.8 shows the inverter control circuit. The X-ray tube voltage is controlled by controlling the
overlap phase of the inverter. The X-ray tube voltage is input to the control circuit of X-ray control unit
via the control circuit mounted on the scanner rotation plate from the output of the high voltage divider of
positive and negative high voltage generators, and feedback-controlled based on the DAC output kV
reference. INV.1,3 ON signal is obtained by using the reference clock of the oscillator. The error
amplification output of kV reference and high voltage divider output is pulse width-modulated. This
pulse width is used as a phase difference to obtain a phase-shift clock. An INV2,4 ON signal is obtained
by using this phase shift clock. The inverter IGBT drive circuit comprises a delay circuit to delay the
IGBT ON signal, and also, it provides an interlock function to prevent a short-circuit of upper and lower
inverter arms. Accordingly, a pause time to prevent a short-circuit failure of the arms is not provided in
the control circuit.
If the error amplification output of kV reference and high voltage divider output is large, the pulse width
becomes narrow. In other words, the phase difference decreases, while the overlapped phase of the
inverter increases, causing the inverter output to increase and also, causing the X-ray tube voltage to
increase.
3−22 L1E-BW1346
3.3.3 Inverter IGBT drive circuit
Fig. 3.9 shows the inverter IGBT drive circuit. The control circuit and drive circuit are connected to
each other by optical fiber. A circuit to delay the turn-on signal of inverter IGBT is connected to the
arms of inverters INV 1 ~ 4.
When INV ON signal is not input (INV OFF), hybrid IC applies an OFF gate voltage (reverse bias
voltage) to the inverter IGBT by the gate drive circuit. When INV ON signal is input from the control
circuit, the hybrid IC applies the ON gate voltage (forward bias voltage) to the inverter IGBT by the gate
drive circuit after a slight delay to turn on IGBT. The IGBT over-current detection circuit of hybrid IC
becomes effective to detect the inverter IGBT collector ~ emitter voltage VCE and the gate ~ emitter
voltage VGE.
If an over-current flows to IGBT to cause VCE and VGE to be higher than the over-current detection setting
voltage, the over-current detection circuit functions to apply an OFF gate voltage to IGBT to turn off
IGBT.
When the over-current detection circuit operates, an INV OC error detection signal is fed to the control
circuit through optical fiber to stop the sequence. This inverter over-current detection is done for the
purpose of detecting a short-circuit failure of the inverter arms. If the drive circuit power supply drops,
an error signal is sent to the control circuit to stop the sequence.
3−23 L1E-BW1346
DC smoothing circuit
Power filter circuit Rectifier circuit INV.1
2FFB 1KP
INV.1 INV.2 INV.2
Drive Drive
Discharge
3φ resistor
AC200V
INV.3 INV.4
INV.3 Drive INV.4
Drive
Series resonance
capacitor
CT
REC
3−24
Drive
T2 T2
DC voltage
feedback
T1 T1
CONT Main INV output current
UNIT IF
T2
Optical fiber
FBL
INV ON / INV OC
T1
L1E-BW1346
Phase difference Phase difference
Transistor
INV. 1 current current
Diode
current
INV. 2 current
INV. 3 current
INV. 4 current
3−25 L1E-BW1346
Rectifier circuit Smoothing circuit
DC
3 φ power supply Inverter circuit
DC voltage
REC
DRIVE (primary voltage)
Insulated amplifier
31E,
33E, 29E 27E 18C Input buffer
33F
17C
Gate 25E
DC voltage < 50V
Error detection
DC voltage > 250V
INV. ABNORMAL
3−26 L1E-BW1346
Rise-time Level shift Oscillator pulse Inverter operation
constant circuit inversion amplifier width modulation Buffer M. M phase F.F Gate
kV reference 23C
18D 16D 20D 21E 21F 23F INV. 1 ON
23E
Error amplifier
Reference clock
High voltage 6E
Synchronous circuit
divider output 201VR INV. 3 ON
22C
kV
Via scanner turn plate
control circuit Buffer Limiter
6E
Phase-shift clock
INV. 2 ON
Inverter operation
M.M phase F.F
3−27
21F 23F
Gate
X-RAY TRIG INV. 4 ON
REC START
INV. ON
kV, mA ABNORMAL
INV. ABNORMAL
REC ABNORMAL INV. 1, 4 ON phase difference
INV. 1 ON
INV. 2 ON
INV. 3 ON
INV. 3, 2 ON phase difference
INV. 4 ON
L1E-BW1346
Fig.3.8 Inverter control circuit XCONT3 circuit (2/3)
+ Vcc1
+Vcc2
Optical fiber
receiver ON/OFF control
1RXD Delay
INV. ON
(101RXD) circuit
(FBL PCB)
VCE
G
0V Gate drive
OFF
3−28
Hybrid IC VGE detection
1IC
VCE detection
− Vcc
Error
detection 1TXD
(101TXD) INV. OVER CURRENT
circuit (TO FBL PCB)
Optical fiber
Drive power error transmitter
+VCC1, +VCC2, -VCC
L1E-BW1346
Fig.3.9 Inverter IGBT drive circuit
3.4 kV, mA error detection circuit
An error of the tube voltage or tube current is detected by +kV, -kV (tube voltage divider output) and NE+, NE-,
N+, N- (neutral point current) shown in Fig.7.31 in Chapter 7. All kV, mA errors are detected by the error
detection circuit of the scanner rotation plate. A part of kV errors is also detected on the X-ray control unit side
Fig.3.10 shows the kV error detection circuit.
The tube voltage is detected as follows
● An excessive tube voltage (higher than 150kV across positive and negative terminals) is detected.
● A tube voltage lowering waveform error (The tube voltage does not become lower than 75kV when 47ms at
least has passed from INV stop) is detected.
● Inverted polarity of the kV signal sent from the scanner is detected.
Tube voltage
lowering waveform VT WIDE
error
Tube voltage detection 75kV
47ms after INV stop
Tube voltage OV VT
is excessive
Tube voltage higher than 150kV
3−29 48 L1E-BW1346
Section 4 Operation principle of X-ray detection system
The X-ray detection system accurately measures the X-ray absorption quantity at the object corresponding to individual
X-ray beams, converts them into electric signals, and inputs these signals to the image processor. This system is
divided roughly into the X-ray detector and the preamplifier as shown in Fig.4.1.
Preamplifier
Vs
Flat electrode plates are arranged in parallel and used as high-voltage electrode plates (H) and signal
electrode plates (S) alternately as shown in Fig.4.2.
A high voltage (Vs) of several hundreds volts is applied to the high voltage electrode plates to form an
electric fields between the high-voltage electrode plates and the signal electrode plates.
When X-rays are incident to these spaces, the ion pairs generated by the ionization of Xe gas move to the
signal electrode plate and high-voltage electrode plates according to the signs of electric charge to
become an output current (Is). As is understood from Fig.4.2, the output current is the sum of the
ionization currents at respective spaces surrounded by two high voltage electrode plates on both sides of
one signal electrode. The ratio of X-ray absorption along the incident direction in the detector after
passing the object (patient) is given by equation (1).
4−1A L1E-BW1346
I (L)=Io(1−e-μPL) (1)
where, Io ; X-ray dose at the front end of detector electrode
I (L) ; Absorption quantity of X-rays along detector electrode length (L)
μ ; X-ray absorption coefficient of Xe
P ; Pressure
L ; Detector electrode length
Xe gas is characterized with a large X-ray absorption coefficient as compared with other gases. In order
to improve the detection sensitivity, Xe gas is sealed at 2.45MPa (absolute pressure) to absorb the greater
part of the incident X-rays and convert them into an ionization current.
The fluorescent X-ray escape and output characteristics are important factors when the Xe ionization
chamber is used as a detector of the X-ray CT system.
4−2 L1E-BW1346
4.1.3 Output characteristics
When the incident X-ray intensity is kept constant, the output current depends on the applied voltage (V)
between electrodes and gradually increases starting from 0 to saturated current (is) at a voltage higher
than a certain value as shown in Fig.4.3. The generated ion pairs reduce due to ionic rebinding in a
weak electric field, but they reach the electrode plates without ionic rebinding as the electric field
becomes intense sufficently and all the ionization 10
currents generated in the electrode spaces can be 9
8
X-ray tube current:200mA
taken out by the signal electrodes. This area is 7
called saturated area. It is necessary for assuring the 6 X-ray tube current:150mA
5
linearity of the output current to the incident X-ray 4 X-ray tube voltage :120kV
intensity to operate the detector in the saturated area, 3 Xe gas pressure :2.45MPa
2
so that 500V is applied. (all ion pairs are collected by
1
the electrodes at higher than 300V.) 0 1 2 3 4 5 6 7 8(× 100)
Fig.4.4 shows the actual structure. The high voltage electrode plates and signal electrode plates are
alternately arranged on the insulator supports and the entire assembly is housed in an aluminum vessel
being internally filled with Xe gas. Output signals are taken out of the vessel every signal electrode
plate by using a specially air-tight processed printed circuit board without any necessity of internal check
of the detector.
Vessel
Insulator support
High voltage
electrode plate
Signal electrode
X-rays incident window plate
The high voltage electrodes and signal electrodes are arranged alternately, and the high voltage electrodes
on both sides of each signal electrode correspond to one channel of measuring data.
4−3 L1E-BW1346
4.2 Solid detector (Pronto-SE/VE)
4.2.1 Principle
The X-ray detector is a solid state detector consisting of independent 752-channel detection elements to
the fan X-ray beams, and its basic principle is as described below.
Rf
Is
X-rays
This unit consists of a scintillator to convert X-rays into visible light, a separator plate to guide the visible
light to a photo diode, a photo diode to convert visible light into a current, and a grid to eliminate
scattered X-rays.
Assume that X-rays are incident to the scintillator and absorbed, now.
The visible light produced by the fluorescent action is converged to the light receptor of the photo diode
to become an output current (Is) after photoelectric conversion. The ratio of X-ray absorption along the
incident direction (痕) in the detector after passing the object (patient) is given by equation (1).
4−4 L1E-BW1346
4.2.2 Fluorescent X-ray escape
The scintillator in this detector has a K absorption edge at 47keV, and incident X-ray photons having
higher energy than the K absorption ease excite K shell to return the excited atoms to the ground state,
causing the fluorescent X-rays to be irradiated.
The average energy of the fluorescent X-rays of K shell is about 45keV to be slightly lower than the
energy at the K absorption edge. Accordingly, it is not easily absorbed in the scintillator to cause
crosstalk to the adjacent channels to result in a drop of the spatial resolution of the detector.
Accordingly, it is necessary to separate each detector channel by a plate having a high X-ray absorption
coefficient. This detector adopts a molybdenum (Mo) plate having a high X-ray absorption coefficient
to suppress the transmittance of the fluorescent X-rays to be less than 2% in the direction perpendicular
to the electrode plate.
Luminescence wavelength of
the scintillator for this
Luminescence wavelength detector Spectral sensitivity of
of conventional scintillator photo diode for this
detector
Spectral sensitivity of
conventional photo diode
Relative sensitivity
Wavelength (nm)
Fig.4.6 Luminescence wavelength of scintillator and the spectral sensitivity of photo diode
4−5A L1E-BW1346
4.2.4 Structure
X-ray Scintillator
Mo separator plate
Photo diode
Fig.4.7 shows the basic structure of the solid state detector. The scintillators and molybdenum separator
plates which separate the elements from each other are mounted on the photo diode consisting of plural
elements according to the element patterns of the photo diode. Incident X-rays are converted into
visible light by the scintillators, and the visible light reach the photo diodes through the scintillator and
converted into a current. The molybdenum separator plates introduce the visible light to the photo
diodes efficiently and also prevent the crosstalk of X-rays to adjacent channels. The current generated
by the photo diodes is sent to the preamplifier through the PCB pattern, connector and cable.
4−6 L1E-BW1346
4.3 Preamplifier (Pronto-XE, Pronto-SE)
4.3.1 General
The preamplifier is mounted between the X-ray detector and the image processor as shown in Fig.4.8. It
receives the control signal from the preamplifier control unit in the scanner to convert current signals
obtained by the X-ray detector into digital codes, and then, transfers these signals to the image processor.
Fig.4.9 shows the entire block diagram.
The preamplifier is composed of the +5V power supply for 3 analog, digital, and interface systems, the
DSC1 PCB, and IFCON2 PCB, and arranged as shown on Fig.4.10 and Fig.4.11. The preamplifier is
standing by to be ready for receiving VIEW TRIG signal from the control unit of scanner after its power
supply has been turned on. When the CT system is set to the measuring condition (in scan mode,
scanogram mode, etc.), the control unit of scanner generates the VIEW TRIG signal with which the
preamplifier starts measuring operation when starting the measurement, signal current from the detector
is AD-converted every channel in the DSC1 PCB and sent to the image processor via IFCON2 PCB.
The preamplifier executes this series operation every VIEW TRIG and returns to be ready for receiving
the VIEW TRIG signal again when this signal is interrupted.
X-ray tube
Control unit of scanner
Patient
Image processor
Pre-Amp Unit
CRT(Image)
Digital signal
Control signal
4−7 L1E-BW1346
Detector
1 2 3 4 5 6 7 8 9 10 111213141516171819 20 212223242526272829303132333435 36 Power supply unit
+5VA
1 3 5 7 9 11 13 15 17 19 21 23 25 27 29 31 33 35 +5VD
DSC1 No.1 DSC1 No.3 DSC1 No.5
ISO+5V
2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 36
DSC1 No.2 DSC1 No.4 DSC1 No.6
IFCON2
2CN
To control unit of
scanner and image 1CN
processor
Preamplifier
Detector
1 2 3 4 13 14 15 21 22 23 24 31 32
ABCABCABCABC ABCABCABC ABCABCABCABC ABCABC
3 6 33 36 48 ISO+5V
DSC1 No.3 DSC1 No.8
IFCON2
To control unit of
scanner and image 1CN 2CN
processor
Preamplifier
4−8 L1E-BW1346
4.3.2 DSC1 PCB
DSC1 PCB is composed of the △-ΣADC to convert the current from the detector into 20-bit digital
signals, the reference unit to feed a reference voltage to △-ΣADC, and the serial-parallel converter to
convert serial data from △-ΣADC into parallel data.
Reference unit
The reference unit feeds a 4.1V reference voltage to △-ΣADC. The △-ΣADC can collect data of 2
channels per IC. Since one DSC1 PCB is provided with 48△-ΣADC, DSC1 PCB can collect data of
96 channels in total.
The △-ΣADC is provided with two QV converter and an ADC every channel, and data are collected
into two phases A,B. In phase A, detector data integrated by the A side QV converter are converted into
digital data by ADC, and the data from the detector are integrated by the B side QV converter. In phase
B, detector data integrated by the B side QV converter are converted into digital data by ADC, and the
data from the detector are integrated by the A side QV converter.
Phase A and phase B are executed once per VIEW, and 20-bit data are sent from DSC1 PCB to IFCON2
PCB each time, and then, stored into IFCON2 PCB memory.
Average values of phase A data and phase B data are sent to the image processor.
Data output from △-ΣADC are serial data. In order to improve the data transfer efficiency, these data
are converted into parallel data by the serial-parallel converter before they are output from the DSC1
PCB. The △-ΣADC output is daisy-chain data, so that 2IC data of △-ΣADC are input into the shift
register in 1IC while the 1IC output of △-ΣADC is being sent to the serial-parallel converter. In the
next sequence, 2IC data are sent to the serial-parallel converter through the 1IC output.
By repeating the above procedure, data of all 96 channels are sent to the serial-parallel converter and sent
to IFCON2 PCB.
4−9 L1E-BW1346
A phases
-
+
QV converter A
From detector △-Σ To serial-parallel
ADC converter
-
+
QV converter B
B phases
-
+
QV converter A
From detector △-Σ To serial-parallel
ADC converter
-
+
QV converter B
4−10 L1E-BW1346
4.3.3 IFCON2 PCB
IFCON2 PCB is composed of the interface unit to receive control signals from the control unit of scanner
and feed AD-converted data to the image processor, the control unit to control each block every view of
preamplifier, the memory unit to temporarily store data from DSC1 PCB, and the data converter unit to
process the stored data to be fed to the image processor.
Memory unit
Memory 1 Interface unit
Data converter
A
B 16 bit To image
Averaging 16-bit data
20-bit conversion processor
parallel data Memory 2 circuit unit
circuit
D0∼D15
ADC To image
processor
Check data
Selector
Photo coupler
Photo coupler
4−11 L1E-BW1346
When the measuring operation is started, the control unit of scanner issues DATA MODE signal to designate
whether the signal being required to the preamplifier is ADC output data or check pattern data. Then, it issues
VIEW TRIG signals of necessary number of views to the preamplifier. When the preamplifier receives the
VIEW TRIG signal, it issues RTN TRIG to output data by channels together with DATA CLK of the synchronous
signal data. The image processor reads data at the positive edge of this DATA CLK as shown in the time chart
given below
DATA MODE
VIEW TRIG
DATA CLK
Table 4.1 and Table 4.2 show the kinds and functions of preamplifier input/output signals.
When DATA MODE signal is H, the test data shown in Table 4.3 and 4.4 are output.
4−12 L1E-BW1346
Table 4.1 Input signals of preamplifier
4−13 L1E-BW1346
Table 4.3 Test data (Pronto XE)
4−14 L1E-BW1346
Table 4.4 test data (Pronto SE)
4−15 L1E-BW1346
(2) Control unit
The control unit is composed of the start control circuit and control signal generator circuit. The start
control circuit controls the generation timing of control signals.
It generates an internal trigger automatically to operate DAS automatically to prevent the detector and
preamplifier from being charged up when the measuring operation is not executed. When the measuring
operation is started and VIEW TRIG signal is input from the control unit of scanner, the start control circuit
resets the auto run of DAS and starts normal data input operation. This circuit judges as the measuring end
if VIEW TRIG signal is not input for longer than 5ms, and resets the mode to the auto run mode again.
The control signal generator circuit is composed of an address counter and an ROM whose addresses are
counted up by the address counter output, and it generates various control signals from the data of ROM. It
issues a CONV signal to designate either phase A or B to the DSC1 PCB and also issues DMIT signal to
demand the data output to △-ΣADC. Then, it issues DCEN signal to transfer the output to the shift
register every bit and also issues LACK signal to latch data to the register of each DSC1 PCB register when
the data of one channel have been stored into the shift register and converted into parallel data. Pronto-XE
selects signals from BEN1 signal to BEN6 signals (from BEN1 signal to BEN8 signal in case of Pronto-SE)
sequentially to output 20-bit data from each DSC1 PCB to store data into the memory of IFCON2 PCB.
This operation is repeated 96 times every phase (A, B).
This circuit reads data being input by the previous VIEW from the memory and sends them to the image
processor via the data converter together with DATA CLK signal.
4−16 L1E-BW1346
4.4 Preamplifier (Pronto-VE)
4.4.1 General
The preamplifier is mounted between the X-ray detector and the image processor as shown in Fig.4.17.
It receives control signals from the preamplifier control unit in the scanner to convert current signals
obtained by the X-ray detector into digital codes, and then, transfers these signals to the image processor.
Fig.4.18 shows the entire block diagram.
X-ray tube
Control unit of scanner
Patient
Image processor
Pre-Amp Unit
CRT(Image)
Digital signal
Converter PCB
Interface PCB
Digital
Detector Δ-Σ data Image processor
Digital filter Interface
ADC Control unit of scanner
Control signal
Terminator PCB
4−17 L1E-BW1346
A current signal from the detector is converted into a digital signal through converter PCB △-ΣADC
and sent to the interface after eliminating its unwanted high-frequency components in the next digital
filter. The interface PCB sends digital signals to the image processor and transfers signals to and from
the control unit of scanner to control the other PCB of the preamplifier.
The terminator PCB aids the signal transfer between interface PCB and converter PCB. The
preamplifier unit is composed of a DC power supply having three outputs of analog ±5V and digital
+5V, 33 converter PCB (24 channels/PCB), interface PCB, and terminator PCB. These components are
arranged as shown in Fig.4.19.
DC power
supply
X52
4CN
①, ②∼⑰ ○
18 ∼○
34 , ○
35
3CN
2CN
X49
X50
A B C A B C A B C A B C A B C
1 16 17 32 33
Detector
① :Interface PCB
②∼○
34 :Converter PCB
○
35 :Terminator PCB
4−18 L1E-BW1346
Analog ±5V and digital +5V are supplied to the right and left preamplifier units. Also, the ISO+5V
for interface is supplied to the left unit through a power cable, and control signals and data bus line are
connected between the right and left units through two flat cables.
The interface PCB is mounted in one slot at the left end of the left unit.
An external I/O signal cable is connected to this PCB to control the measuring mode corresponding to
plural units by setting the plug pins inside PCB.
Converter PCBs are arranged in the right and left units separately and 16 PCB are mounted on the left
unit, while 17 PCB are mounted on the right unit. (33 PCB in total) The terminator PCB is mounted
at the right end of the right unit, and this PCB provides a terminator function to securely transmit the
signals from the interface PCB to the converter PCB. Each of respective converter PCBs is provided
with three detector signal input connectors corresponding to 8 channels, and one PCB can input signals of
24 channels.
A detector cable is connected between this connector and the detector connector.
Since each of total 33 right and left converter PCB can input 24-channel measuring data, these PCBs can
input signals corresponding to 24 x 33 = 792 channels. Of these data, data of 768 channels are sent to
the interface PCB and then, sent to the image processor. (Of these data, first 752-channel data are used
in the image processor)
4−19 L1E-BW1346
Table 4.4 Input signals of △-Σpreamplifier
These input/output signals are electrically separated from the other units by a photo coupler and ISO +5V
independent power supply for the purpose of preventing ingress of noises to the preamplifier.
From the converter PCB, 21-bit digital data are sent to the interface PCB periodically. The interface
PCB converts these 21-bit digital data into 16-bit data of floating point (data having a 14-bit data part and
2 gain bits).
4−20 L1E-BW1346
Table 4.6 Correspondence of gain codes
0 0 1 time
0 1 8 time
1 0 64 time ※ Errors comprise an overflow error
1 1 ERROR and an underflow error.
The interface PCB is provided with a memory by two views to save 16-bit data and data are saved by one
view alternately. If the data transmission to the image processor is demanded by a view trigger signal,
saved data are sent. If not, the next new view data are written into the memory. On the other hand, if
the data mode signal is H, and the image processor demands a test pattern output, the interface PCB
outputs data shown in Table 4.7 repeatedly.
4−21 L1E-BW1346
Table 4.7 Test patterns
This preamplifier prepares a calibration signal and a reset signal. When the calibration is input, the preamplifier
executes correction operation (about 2 seconds) to be set to an optimum condition. When the reset signal is
input, the preamplifier initializes its all portions, and then, executes calibration.
Accordingly, it takes about 12 seconds until it is ready for operation.
4−22 L1E-BW1346
The interface PCB is provided with jumper plugs (The number of measuring channels and the measuring cycles
are switched by combined setting) and LED. These function as follows.
No. Conditions
W1 SHORT
W2 OPEN
W3 SHORT
W4 OPEN
W5 OPEN
W6 SHORT
4−23 L1E-BW1346
Table 4.10 Functions of LED of terminator PCB
4−24 72 L1E-BW1346
Section 5 Central Operation Console
5.1 Hardware
5.1.1 Block diagram
Program MD Program MD
CD-
Raw Data Image ROM
PCI-bus PCI-bus
Measuring data IM
DSP DSP BP DSP
I/F
Imager
AC200V
POWER DC POWER
Mouse
5−1 L1E-BW1346
5.1.2 Comparison table
14 Imager interface Exclusive display card Exclusive display card EXP signal
is provided
15 Auto film Provided ( DICOM print) Not provided (EXP signal
only)
5−2 L1E-BW1346
5.1.3 Scanner control system (CPU#0)
The scanner control system (CPU#0) is provided with a scanner I/F circuit, a touch panel, an LCD display,
and a sound source circuit.
Neither keyboard nor mouse is mounted However, START key and other hardware keys are processed as
the keyboard.
Panel unit
To CPU#1
LINK
TO SCANNER LCD
SCIF1 FR-MK2 FR-MK2 display
PCI BUS
LINK
MD MEM
Bold frame shows internal products
N.C.
5−3 L1E-BW1346
○ Component parts of scanner control system (CPU#0)
<CPU>
Mother board RICOH FB2LST INTEL440BX 6PCI
CPU INTEL PentiumIII 850MHz
Memory 128MB DIMM
<PCB>
Display PCB RICOH FB2VT AGP 4MB
Ether PCB 100BASE-T 82559 (PCI connection in mother board)
Sound PCB ESS Mwareo-3 (PCI connection in mother board)
Scanner I/F board SCIF1 internal product
DSP PCB FR-MK2 internal product
M-BUFF PCB FR-MB2 internal product
<Panel>
Touch panel LCD RICOH FL1
Molded panel unit Maker: Wako Rubber Co.
Intercom PCB Internal product (IAMP)
<Others>
Hard disk 20GB or over IDE
Power supply 300W ATX V2 specifications
Table 5.2
Bus
Slot No. PCB names Remarks
specifications
1 Not specified FR-MB2 M-BUFFER
2 ISA
3 PCI FR-MK2 DSP for pre-processing
4 PCI
5 PCI FR-MK2 Arrangement DSP
6 PCI SCIF1 Scanner I/F
7 AGP FB2VT Display
8 Not specified LCD I/F LCD monitor connector
5−4 L1E-BW1346
5.1.4 Image analysis system (CPU#1)
The arithmetic analytical processor (CPU1) is provided with an FR arithmetic unit, a display, a keyboard,
a mouse, an imager I/F, an SCSI I/F, and Ether I/F.
17"
MONITOR
Video signal
To external network
OPTION
From CPU#0 side PC-BP LINK FR-MK2 DISPLAY SCSI ETHER PC-IMG
-RV (DSP) I/F I/F to IMAGER
AGP BUS
PCI BUS
OPTION
KEYBORD
MOUSE
5−5 L1E-BW1346
○ Component parts of image processing system (CPU#1)
<CPU>
Mother board RICOH FB2LST INTEL440BX 6PCI
CPU INTEL PentiumIII 850MHz
Memory 256MB DIMM
<PCB>
Display PCB G450 AGP
DSP PCB Internal product (FR-MK2)
BP PCB Internal product (PC-BP-RV)
Imager I/F PCB Internal product (PC-IMG)
Extended Ether PCB For external Ethernet
Extended SCSI PCB For external SCSI
<Others>
Hard disk 20GB or over IDE
CD-ROM 20X speed or over
FD 3.5FD 1.44MB
Monitor Monitor 17” TRINITRON system
Keyboard 104 keys arrangement PS/2
Mouse PS/2 mouse
Power supply 300W ATX V2 specifications
Table 5.3
Bus
Slot No. PCB names Remarks
specifications
9 ISA PC-BP-RV (No bus is used)
10 PCI PC-IMG Imager I/F
11 PCI FR-MK2 DSP for after-processing
12 PCI SCSI External SCSI (option)
13 PCI ETHER External Ether
14 AGP DISPLAY Display
15 − − −
5−6 L1E-BW1346
5.1.5 Arithmetic unit configuration
The arithmetic unit is composed of FR-MB2 PCB, three FR-MK2 PCB, and PC-BP-RV PCB. Data
connections among these PCB are executed by DSP-LINK signal and PCI buses.
RAW IMAGE
DATA DATA
Measuring Analytical
MD MD
CPU CPU
Table 5.4
5−7 L1E-BW1346
5.1.6 Scanner interface
(1) Communication through common memories
Two independent common memories are arranged between the scanner and image processor without
any conventional RS232C cable. (An interrupt function is provided)
Common memory
A
Commu- Commu-
Scanner unit nication nication
I/F I/F
Common memory
B
PCI-BUS
Measuring
CPU
Hard wire
5−8 L1E-BW1346
○ Details of connection specifications
232C port
Output only
To scanner
SCIF1
AUTO/MANUAL PCB
PCI-BUS
SCANNER RESET
START
START SW LED
SET SW LED
SET
START
SW.
( Note1)
SET SW.
From scanner
X-RAY TRIG
EMERGENCY
X-RAY ON LED
X-RAY ON
X-RAY ON BUZZER
BZ
Note 1) Start signal serves as an OR signal of hardware switch and software I/O
5−9 L1E-BW1346
5.1.7 Imager interface
(1) The imager is connected from the exclusive PCB (PC-IMG PCB). Two kinds of signals (analog
video signal and digital video signal) are provided as signal modes. The PCB is connected by the
image analyzing system CPU bus and image data only are transferred to PC-IMG board.
Video signal
DISPLAY PC-IMG
PCB PCB Imager
multi-camera
PCI BUS
EXP signal
Image processor
ETHER-NET
Image processing
DICOM
system CPU
100BASE-T Imager
External
ETHER I/F
5−10 L1E-BW1346
5.1.8 Interphone auto voice configuration
A sound output of CPU#0 is used as auto voice. It can be recorded by using a microphone inside the
panel unit. An exclusive power supply is used for amplifier. (analog power supply)
Panel unit
Image processor cubicle
Speaker Microphone
Exclusive amplifier
Scanner power supply
microphone
Scanner
Speaker
5−11 L1E-BW1346
5.1.9 Power supplies and reset specifications
(1) Power supplies
The image processor power supply is turned on by pressing POWER switch on the front panel of
image processor, and turned off by pressing POWER switch after the end of software. (Power
supply can be turned off without terminating software However, the memory contents are not held)
The gantry power supply is turned on or off by the momentary switch. CPU ON signal is
connected to the scanner.
(2) Reset
CPU#0, CPU#1, and scanner are reset by RESET switch on the rear panel of the panel unit. Also,
I/O reset to the scanner can be done from CPU#0.
5−12 L1E-BW1346
5.1.10 Connection signal specifications
(1) Connection diagram
①Measuring data
Scanner Image processor
<S> <C>
②Scanner control
5−13 L1E-BW1346
(3) Connector block diagram
CPU#0
50PIN DDK
Scanner control
To scanner
50PIN DDK
Measuring data
To panel unit
Switches
intercommunication
CPU#1
15PIN D-SUB
Monitor signal To monitor
PS/2
Keyboard To keyboard
PS/2
Mouse To mouse
34PIN MIL
Digital video
BNC
Analog video
To imager
BNC
Pixel clock
Exclusive cable EXP signal 9PIN D-SUB
Printer port
100BASE-T
ETHER-NET DICOM I/F
Fig.5.13 Connectors
5−14 L1E-BW1346
Imager interface specifications
0.660V
Black Level
0.054V
Blank Level
0.286V
Sync Level
b c d e
3H 4H 43H 512H
562H
5−15 L1E-BW1346
5.2 Measurement processing
5.2.1 Summary of measurement and image reconstruction arithmetic operation
Fig.5.15 shows the scan measuring sequence. After setting a patient to the table
(initial slice position), set the scan conditions by the touch panel.
By pressing the standby switch, the OC side scan control task starts the scanner control task to set
parameters to FR-MK2 and PC-BP-RV. The scanner CPU instructs the rotation start to the starter.
The X-ray generator starts the X-ray tube voltage control and the scanner unit starts clockwise rotation.
When the image processing, X-ray generator system, and scanner move are set to be ready, the green
lamp of the start switch flickers to show that the start switch is effective. By pressing this switch, the
image processor starts offset measurement and waits for inputting X-ray data after the end of
measurement.
Next, an X-ray data input start signal is input to the scanner CPU. By scanner CPU control, the rotary
encoder signal of the scanner unit sends a data input control signal to the preamplifier unit. The
preamplifier unit sends data to the image processor after executing the AD conversion through a low-pass
filter. When data are input by one scan angle, the scanner control task sends an X-ray off command to
the X-ray control unit to turn off X-rays.
The image processor starts the reconstruction operation and displays an image on the monitor of the
central operation console at the end of reconstruction operation.
Fig.5.16 shows the details of the measuring and arithmetic operation sequences.
Offset data measurement and X-ray data measurement are executed by FR-MK2. Variously corrected
and log-converted data are saved into an HD after total-angle sampling. X-ray measuring data are
parallel-beam converted by FR-MK2, and transferred to PC-BP-RV. PC-BP-RV executes filtering
processing and back projection operation by using this data , and transfers data to the FR-MK2 in the
subsequent stage. FR-MK2 executes after-processing to obtain a CT image. These arithmetic
operations are executed after inputting X-ray data.
5−16 L1E-BW1346
Table slice Start switch
position Scan Condition Standby Start lamp Start switch lamp End
ON setting ON flickering ON
setting extinction
Starter rotation
Steady-state rotation
OK
X-ray X-ray generation
preparation
X-rays
Scanner rotation
5−17
Scanner Clockwise rotation
control
task
start
Data input signal
Scanner OK
preparation
Scanner
Offset
Image measurement
processing OK Measurement Arithmetic Image
preparation processing operation display
HD
Waiting for X-ray data transfer
Image processor
L1E-BW1346
Fig.5.15 Scan measuring sequence (in single scan mode)
Measurement
Standby start
Clockwise rotation
Scanner rotation
X-ray exposure
Scattered rays correction and inter-channel correction Arrangement After-processing CRT display
of image
Filtering Back
projection
5−18
Offset data measurement X-ray data measurement Arrangement processing
437 samples are measured and averaged. The following items are executed in Filtering processing
parallel with data input ● FFT calculation
● Data angle addition ● Filter function
● Offset correction ● FFT-1 calculation
● Log conversion ● Data interpolation
● Reference correction The above processing results are
● Air calibration correction transferred to BP sequentially.
● Non-linearity correction
● Scattered rays correction
● Inter-channel correction
L1E-BW1346
5.2.2 Measuring data flow
(1) CAL measurement
For obtaining a good image by scan measurement and scanogram measurement, measuring data
must be corrected by using calibration data. For this purpose, calibration data are obtained by CAL
measurement. Fig.5.17 shows the data flow.
This CAL measurement is divided into AIR CAL measurement and phantom CAL measurement
Observe the following procedures.
5−19 L1E-BW1346
X-ray tube CAL measurement
Scanner rotation (0 data are set during air measurement) ・Offset measurement
・X-ray measurement
・Offset correction
・Log conversion
Air measuring data
・Reference correction
・AIR CAL correction
(No correction is done in
AIR measuring mode.)
Log table
Detector
HD
5−20
FR-MK2(1) Measuring data Measuring data
Preamplifier AD
CAL phantom
(No phantom is inserted during air measurement)
Fig.5.17 CAL measuring data flow (AIR CAL and phantom CAL measurement)
L1E-BW1346
X-ray tube ● Offset measurement
Scan measurement ● X-ray measurement
Scanner rotation ● Data angle calculation
● Offset correction
● Log conversion
CAL data ● Reference correction
● AIR CAL correction
● Phantom calibration correction
● Non-linearity correction
● Scattered X-ray correction
Detector Log table ● Inter-channel correction HD
Measuring
FR-MK2(1) Measuring
Preamplifier AD data
data
Filtering processing
Angle
parameter Back projection results
5−21
Arrangement
parameters Filter weight FR-MK2(3)
CRT monitor
Image data
PC-BP-RV Display
Filtering
PC-BP-RV memory
Arrangement processing results
FR-MK2(2)
results
Image after-processing
Expansion HD image loading area
weight
Image
data
FR-MK2(3)
L1E-BW1346
● Offset measurement HD
● X-ray measurement
X-ray tube ● Data angle addition
Measuring
● Offset correction data
● Log conversion
Scanogram measurement ● Reference correction
AIR CAL ● AIR CAL correction
Table move data Interpolation
weight
Measuring
FR-MK2(1) data
Preamplifier AD
5−22
Image after-processing
Expansion
weight FR-MK2(1)
CRT monitor
Image data
Display
FR-MK2(1) memory
Filtering
results
Image
data
L1E-BW1346
5.2.3 Scan measuring mode
The scan measuring mode comprises the kinds shown in Table 5.6 and 5.7.
As the scan time becomes longer, data samples increase to be able to obtain a good image, but the
exposure dose and HU wait time increase, on the contrary.
5−23 L1E-BW1346
[{log (XD-OFD)−log (XR-OFR)−CAL}* (1+n) ]−CHD
(1) (1)
(2) (2)
(3)
(4)
(5)
(6)
5−24 L1E-BW1346
(3) Reference correction
Four channels at both ends of the detector are out of the measuring visual field, and no substance
other than air exists on the X-ray path received by 4 channels at both ends. These four channels at
both channel are called reference channels.
The intensity of the X-rays irradiated from the X-ray tube is not kept constant at all times, but it
fluctuates every irradiation (every view).
This fluctuation is corrected every view, and the X-ray fluctuation between views must be corrected.
This is called reference correction.
This correction is executed by dividing the reference channel output.
5−25 L1E-BW1346
5.2.5 Image reconstruction calculation
The processing from offset correction to ring correction is generally called pre-processing. The
following processing is done to the data after pre-processing.
(1) Parallel beam conversion
(2) Filtering processing
(3) Back projection calculation
These three processing are called image reconstruction processing. After-processing is done to the
image reconstruction results.
For obtaining the parallel beams of one view, data are necessary by the opening angle of the
detector.
In the parallel beam conversion, fan beam data are interpolated to convert them into parallel beam
data.
5−26 L1E-BW1346
(2) Filtering processing
Filtering is done to the data converted into parallel beams by parallel beam conversion. The system
contains loaded filter parameters which have been adjusted every image reconstruction part, and the
filters being set on the operation panel are used for filtering processing.
The filter values change according to filter numbers and the filter function has the following profile,
in general.
Filtering is to obtain the composite product of the filter parameters and parallel beam conversion
results. Composite product is calculated by weighting each element of one-dimensional
arrangement data to peripheral data and then converting these data. It is calculated by the
following equation
N-1
R[k] : Σ=ΣFilter[i]*Para[k-i]
i=0
where
R[k] : Filtering results
Filter[i] : Filter parameter
Para[k-i] : Parallel beam conversion results
Since it takes time to calculate the above composite product, filtering is practically done by the
Fourier transformation as shown in Fig.5.23. Fourier transformation results of filter parameter are
loaded into the system first. Then, parallel beam results of one view are Fourier-transformed to
copy into a frequency space.
In the frequency space, the above composite product is a simple multiplication method. Composite
product of filter parameters and parallel beam conversion results is obtained by a simple
multiplication method. Then, these multiplication results are inverse Fourier-transformed finally.
5−27 L1E-BW1346
Parallel beam conversion Filter parameter
results
Fourier transformation
MD
Filtering results
Channel direction
View direction
5−28 L1E-BW1346
Back projection calculation is done by adding the filtering results of each view on the image.
Fig.5.25 shows an example of back projection operation of only 3 views at 0o, 45o, and 90o.
In actual system, this back projection calculation is done over all number of views.
Since back projection results are obtained as data where air part is 0, the CT value is adjusted after
back projection calculation to correspond to a CT value where the air part is -1000.
Processing to be executed to an image after back projection calculation is called after-processing,
and after-processing results are loaded into an HD.
BIOS(ROM) OS Application
WinNT program
The CT system is internally provided with two computers for scanner control and image analysis.
The scanner control system and image analysis system computers execute various control and processing by
incorporating with the scan, image reconstruction, and image display through TCP/IP communication.
TCP/IP communication
The scanner control system computer executes (controls) scan according to the touch panel operation by a user.
Scanner control system monitor display is done on the touch panel. This touch panel also serves as a pointing
device (simulated mouse). No keyboard is provided.
6−1 L1E-BW1346
Scanner,
Table
Monitor display
Scanner control
Touch panel To image analysis system
system computer
HD
The image analysis system computer receives an image from the image reconstruction operator
(DSP/reconstructor), and executes writing into HD, image display and analysis, patient reservation, etc. The
image analysis system is provided with monitor display and keyboard, and operated in almost the same way as in
general computers.
DSP/reconstructor
CRT monitor
Scanner control system
Analysis system
computer
Keyboard/
mouse
HD
6−2 L1E-BW1346
6.2 Disk map
A magnetic disk just after entering a program into the system is mapped as follows.
6.3 OS
Windows NT Workstation 4.0 (manufactured by Microsoft Co.) is used as an OS for CT. For details, refer to the
marketed instruction manual.
The basic instructions of OS are described below.
The following figure shows an internal structure of OS. Various utilities of OS are sorted as applications.
Windows NT can be sorted into the following modes
User mode ・・・ Subsystem (Offering API to the application) operates.
Kernel mode (privilege mode) ・・・ Basic parts of operating system operates.
Kernel mode is called NT Executive System Service, and it provides basic functions of the operating system.
NT Executive System Service can be subdivided into various managers, Kernel, and HAL. Seven kinds of
managers offer basic functions of OS. Kernel is called nuclei of the OS which assists various managers.
HAL offers the codes depending upon the hardware platform.
6−3 L1E-BW1346
Application
USR32.DLL
Stub client DLL
GDI32.DLL
Kernel mode
Security
Win32K Object Process LPC function VM manager I/O manager
reference
(USER, GDI) manager manager
monitor
Kernel
Graphic
device driver
Hardware abstraction layer (HAL)
Hardware
This system maintains the independency from hardware by means of HAL, enhances OS rigidity by dividing NT
executive and subsystem operation mode from each other, and divides the subsystem and application from each
other to prevent the OS from being affected by applications.
The scanner unit is composed of the mechanism to rotate the X-ray tube assembly and detector around the patient
to be able to execute vertical and tilted diagnostic measurement. Its components and summary of functions are
described below.
7−1 L1E-BW1346
⑤X-ray tube assembly ⑱ Projector (for positioning) ⑲Slip ring
Front upper frame
Projector ⑳
(for median line)
③Bearing
⑩Projector
Frame ① (for slice position)
④Large pulley
②
Rotation plate
7−2
⑨1mm collimator High voltage generator(+) ⑭Gantry tilt drive unit
(option)
L1E-BW1346
7.1.2 X-ray geometric system
The X-ray geometric system is shown in Fig. 7.2.1 and Fig. 7.2.2.
Fig. 7.2.1 shows the X-ray geometric system when gas (Xe) detector is mounted, while Fig.7.2.2 shows
the X-ray geometric system when solid detector is mounted.
7−3 L1E-BW1346
Collimate plate
(X-ray tube side)
241.3
998(Detection plane)
606.3
Slice thickness S
d2
Unit mm
7−4
Collimate plate
(Detector side)
40.5°(※ 2)
Slice thickness Collimate plate space mm
41.1°(※ 3)
S mm d1 d2
43°(※ 1) (1) 0.8 1.7 (※4)
2 0.8 −
※1 : X-ray beam angle 43°
3 1.2 −
※2 : Effective number of channels for image 568ch
5 2 −
※3 : No. of all channels 576ch
7 2.8 −
(568ch + right and left channels for reference 4ch each)
※4 : Tube side collimator uses a 2mm slice when 1mm slice is used 10 4 −
L1E-BW1346
241.3
1037(Detection plane)
606.3
Slice thickness S
d2
Unit mm
7−5
Collimate plate
(Detector side)
(※ 2)
41.1° Slice thickness Collimate plate space mm
S mm d1 d2
(※ 3)
41.5°
(1) 0.7 1.5 (※4)
(※ 1)
43°
2 0.7 −
3 1.0 −
5 1.8 −
※1 : X-ray beam angle 43°
7 2.6 −
※2 : Effective number of channels for image 744ch
10 3.8 −
※3 : No. of all channels 752ch
(744ch + right and left channels for reference 4ch each)
L1E-BW1346
※4 : Tube side collimator uses a 2mm slice when 1mm slice is used
E
D A 15°
45°
CW
CW Note : Rotation plate rotates in clockwise
direction only.
C B F
Rectifier
voltage
Measurement
ti
Note : ti≒1s
Measurement starts
Press standby Press start switch
switch
Pronto VE Unit s
Scan speed ta td
0.5, 0.8 8.0 Less than 16
1.0 6.0 Less than 14
1.5 4.0 Less than 12
2.0 3.0 Less than 11
3.0 2.0 Less than 10
7−6 L1E-BW1346
The X-ray tube assembly stops at just above (12 o’clock) normally. The rotation plate starts clockwise
rotation when pressing standby button. Measurement is started by pressing start button. Seven
A ~ ○
measurement start points (○ G ) are provided as shown below, and they are set automatically
Pronto XE, SE
0.6s, 1s Point ○
A, ○
B, ○
C, ○
D Point ○
A, ○
C
1.5s Point ○
E , ○
F , ○
G Point ○
E , ○
F , ○
G
2s Point ○
A, ○
B, ○
C, ○
D −
3s Point ○
E , ○
F , ○
G −
Pronto VE
1.0s Point ○
A, ○
B, ○
C, ○
D Point ○
A, ○
C
1.5s Point ○
E , ○
F , ○
G Point ○
E , ○
F , ○
G
2s Point ○
A, ○
B, ○
C, ○
D −
3s Point ○
E , ○
F , ○
G −
After the rotation plate has started rotating, it stops rotating automatically in 120 seconds if the start key
is not pressed.
7−7 L1E-BW1346
7.2 Scan drive unit
The scan drive system and the detection unit for drive control will be described.
Large pulley
6
65
Small pulley
Drive timing belt
Electromagnetic brake
I/O CPU
Inverter
control unit Speed command rotating
direction command
Induction motor Drive pulley
V-belt
7−8 L1E-BW1346
The large pulley connected to the rotation plate rotates by means of the combination of an induction motor
(1.5kW), single-stage V-belt reduction gears, and drive timing belt.
For drive transmission, the induction motor and small pulley are coupled with each other by a V-belt, while the
small pulley of reduction gears output shaft is coupled with the large pulley by a drive timing belt. The rotating
direction and speed to the induction motor are controlled by a general-purpose inverter.
The number of revolutions N of the rotary plate during 1-sec scan is related to the number of revolutions of the
induction motor as shown below.
An electromagnetic brake is mounted on the reduction gears shaft. This electromagnetic brake normally
operates when the panel power supply is turned off and the X-ray tube assembly is set to the just above position
under the stop condition. It is also operated by the emergency button to stop rotating the rotation plate.
<Pronto-XE, SE>
30 (No. of timing pulley teeth Z1) 891.3 (Large pulley diameter)
P = 1024 pulses× × = 9600 pulses
20 (No. of timing pulley teeth Z2) 142.6 (Small pulley diameter)
<Pronto-VE>
24 (No. of timing pulley teeth Z1) 891.3 (Large pulley diameter)
P = 1800 pulses× × = 10800 pulses
25 (No. of timing pulley teeth Z2) 142.6 (Small pulley diameter)
The photo sensor (tube upside sensor) ② detects the light shielding plate ③ mounted on the large
pulley. The X-ray tube assembly is positioned just above when the sensor ② detects the light
shielding plate ③.
7−9 L1E-BW1346
X-ray tube assembly large pulley
③
Detail of part A
Photo sensor② ON
⑤ (just above)
① ②
7−10A
Small pulley
L1E-BW1346
7.3 Gantry tilt unit
The scanner tilt unit consists of the gantry tilt link mechanism to support and tilt the gantry, the tilt drive device to
tilt the gantry, and the angle detector to detect the tilt quantity. Fig.7.5 shows the gantry tilt unit structure.
Frame
0 (Tilt center)
Angle detector
Angle detector
Link mechanism
Stand
The tilt link mechanism is provided to support and tilt the gantry with the intersecting point of the
rotation center of the rotation plate and the X-ray center employed as the tilt center. The tilt link
mechanism structure is shown in Fig.7.6.
Links A and B are coupled with each other by shafts A, B on the right and left side faces of the frame
supporting the rotation plate, etc. Links C and D are supported to the stand by shafts C and D. Links A,
B and links C, D are coupled with each other by shafts E, F, G respectively, so that parallel links of shafts
A-E and B-F and shafts C-F and D-G are constructed
7−11 L1E-BW1346
X-ray center
Frame
Tilt center
I O
0
J
I J
Stand D
C D
C
H A
A B
Link A
B
Link
Link D
E
Hydraulic cylinder Link C
E F G G
F Link B
Part H of link D is supported by a hydraulic cylinder. As the cylinder expands and contracts link D
rotates around the shaft D being supported by the stand. Link C also rotates around fixing shaft C. As
a result, the point of shaft B of link B coupled by shafts F, G moves with the same curvature radius as the
curvature of link D (inter-shaft distance of shafts D and G). Also, the point of shaft A of link A moves
with the curvature radius (inter-shaft distance of shafts C and E) by the parallel link structure of shafts
B-F. Accordingly, the gantry supported by shafts A and B is tilted around the rotation plate tilt center at
the intersecting point O of the circle (I-I) drawn with the D-G distance as the radius around shaft A as the
center and the circle (J-J) drawn with the C-E distance as the radius around shaft B as the center. Point
0 is set to meet the intersecting point of the rotation center of the rotation plate and the X-ray center.
7−12 L1E-BW1346
Motor pump
Shaft(H)
Electro-magnetic valve
Link(D)
Check valve
Throttle valve
Pump
M
Relief valve
Filter
Tank
Stand
Cam
Drive pulley
backward tilt limit switch
Belt
Forward tilt limit switch
Driven pulley
Potentiometer
7−13 L1E-BW1346
Gantry tilt opetation of rotation plate and the vertical operation of patient table secure their positionaly
relation by mean of vertical move detection potentiometer assembled in the patient table.
Fig.7.9 and Fig.7.10 show the interlock between the gantry tilt and the patient table height.
250 1000
140 890
130 880
Height indication (mm)
120 870
Tilt angle
− 25° + 25°
1000 250
750 0
380
Unit :mm
7−14 L1E-BW1346
250 1000
130 880
Height indication (mm)
120 870
Tilt angle
− 25° + 25°
1000
250
750 0
380
Unit :mm
Fig.7.10 Interlock of gantry and patient table (when solid detector is mounted)
7−15 L1E-BW1346
7.3.4 Positioning projector
A positioning side projector is mounted on the front upper frame of the scanner (See Fig.7.1).
The projection position is 290mm from the slice position (X-ray irradiation center).
7.4.1A X-ray tube side collimator (Applies to Pronto-XE and Pronto-SE ( ~ S/N 2024) )
(1) Compensation filter
Compensation filter ① made of aluminum is mounted below the X-ray irradiation port.
(2) Slice thickness selector
The slice thickness is selected by the crank mechanism shown below.
Reversible motor ② rotates gears (1) ③. Gears (1) ③ rotate gears (2) ④.
Rod ⑥ being set to the rotary contraposition relation with gears (2) ④ pushes collimator plate ⑧
to cause rod ⑤ to pull the collimator plate ⑦.
Five-stage grooves are provided on collimator plates ⑦, ⑧, respectively.
These collimator plates move through the guide of fixed bearing ⑨, and the distance between
collimator plates ⑦, ⑧ forms five kinds of slit width by the aid of this groove guide.
The collimator plate ⑦, ⑧ positions corresponding to the above conditions are detected by
sensors ⑩ ~ ⑭ and shield plate ⑯. When sensor ⑩ detects the hole of gears (2)④, 10mm
slice is selected. Sensor ⑪ applies to 7mm slice, sensor ⑫ applies to 5mm slice, sensor ⑬
applies to 3 mm slice, and sensor ⑭ applies 1, 2mm slices.
7−16A L1E-BW1346
A. In case of Pronto-XE
(for 1, 2 mm slices)
⑦ ⑬
③
A ⑫
CB ⑩ ⑪ Slice thickness 10mm 7mm 5mm 3mm 1, 2mm
ED ⑥
Collimate plates distance
(when Xe detector is mounted) 4mm 2.8mm 2mm 1.2mm 0.8mm
Stop position E D C B A
⑨ Photo sensor ⑩ ON − − − −
Table side
⑧
Photo sensor ⑪ − ON − − −
Photo sensor ⑫ − − ON − −
Photo sensor ⑬ − − − ON −
Photo sensor ⑭ − − − − ON
L1E-BW1346
③
A ⑫ Slice thickness 10mm 7mm 5mm 3mm 1, 2mm Shutter
CB ⑩ ⑪
ED ⑥ Collimate plate distance
F 3.8mm 2.6mm 1.8mm 1.0mm 0.7mm close
(when Solid detector is mounted)
Stop position E D C B A F
⑨ Photo sensor ⑩ ON − − − − −
Table side
Photo sensor ⑪ − ON − − − −
⑧
Photo sensor ⑫ − − ON − − −
Photo sensor ⑬ − − − ON − −
Photo sensor ⑭ − − − − ON −
Photo sensor ⑮ − − − − − ON
L1E-BW1346
7−19A L1E-BW1346
C. In case of Pronto-SE (having S/N 2025 and higher)
Pronto-VE
⑧ Photo sensor ⑪ − ON − − − −
Photo sensor ⑫ − − ON − − −
Photo sensor ⑬ − − − ON − −
Photo sensor ⑭ − − − − ON −
L1E-BW1346
Accordingly, the collimator frame position is held without any play even when the rotary plate rotates.
Z-axis
④
① Collimator frame
② Mechanical section
③
③ Linear guide
④ Stopper A
⑤ Stopper B
⑥ Torque motor
⑦ Pinion
⑧ Rack
①
⑥
⑦
⑧
7−21 L1E-BW1346
7.5 Slip ring
This mechanism transfers signals and electricity between the frame or other stationary side and the rotation plate
or other rotation side. It is composed of slip ring ①, power brush block ②, signal brush block ③, and HF
signal brush block ④. The slip ring outer circumference is covered with a shield cover to prevent noise leak
from the brushes. The shield cover is composed of left upper ring cover ⑤, right upper ring cover ⑥, left
lower ring cover ⑨, right lower ring cover ⑩, left ring cover ⑦, left side ring cover ⑪, and right ring cover
⑧.
②
③ A
⑪
A
⑱
⑭
⑮ ⑥
④
⑧
①
⑩
⑨
⑰ ⑯
7−22 L1E-BW1346
7.5.2 Brush block
The brush is a bundle of φ0.22 wires.
Fig.7.14 shows an approximate profile of each brush block.
3ch
2ch
1ch
23ch
4ch
23ch
20ch
Fig.7.14
7−23 L1E-BW1346
7.6 Summary of electric control device of scanner
The electrical control device of this third generation CT scanner uses a low-voltage slip ring. The electric
control device of the scanner consists of the stationary control device, rotation plate control, and power
distribution board control device. Signals and power supply of these devices are connected via the slip ring.
The main power supply of the high voltage unit on the scanner rotation plate is also supplied via the slip ring.
The power distribution board control device is composed of chassis 7 (mounted at the bottom of the scanner) and
chassis 3 (mounted on the left base of the scanner), and turned on and off by the gantry power switch of the
operator console.
The stationary control device is composed of the PCB unit 1 (mounted at the upper left of the scanner) and chassis
2 (mounted at the upper right of the scanner), and the electric parts mounted on the lift cover, rear cover, frame,
etc. are controlled by the PCB unit 1. The rotation plate control device is composed of PCB unit II,
chassis 1, etc. and the X-ray tube unit, radiator, high-voltage unit, solid detector (or Xe detector), preamplifier,
and collimators are arranged around.
Electric parts mounted on the rotation plate frame are controlled by PCB unit .II.
Internal control of scanner and communication control to the exterior is executed by Hitachi RISC (Reduced
Instruction Set Computer) (SH micro-computer).
This high-speed computer reduces complicated instructions to simplify the hardware. This CPU is internally
composed of 32 bits, and the basic instructions are operated by one instruction/state (1 system clock cycle), so
that the execution speed inside CPU is high (2 state operation in case of the exterior).
Since CPU is designed as a multiple system to disperse the processing, the processing speed of each function is
improved. This CPU is also provided with serial communication ports, and it can execute RS232C
communication with the exterior via driver/receiver only.
Three CPU are used in the stationary control device Inter-CPU communication of the stationary control device is
executed via dual port RAM. The CPU is composed mainly of a flash memory for loading programs and an
SRAM for loading data.
Major specifications of CPU circuit are described below.
・ Multi (3) CPU (stationary side) : Main CPU (M. CPU/MCPUC PCB)
: Scanner CPU (S. CPU/MCPUC PCB)
: Table CPU (T. CPU/TCPUB PCB)
7−24 L1E-BW1346
・ Work memory area (SRAM) : HM628512FLP-7A (QFP) x 2
: Memory capacity (512kW)/M.CPU
M.CPU∼S.CPU(DPRAM1)
Write use only from the first half 1kB M.CPU
(Read use only from S.CPU)
Write use only from the second half 1kB M.CPU
(Read use only from M.CPU)
M.CPU∼T.CPU(DPRAM2)
Write use only from the first half 1kB M.CPU
(Read use only from TCPU)
Write use only from the second half 1kB M.CPU
(Read use only from T.CPU)
The priority sequence of interruptions are represented by priority level values (0 ~ 16). Level 0 is the minimum,
while level 16 (NMI) is the maximum.
7−25 L1E-BW1346
Priority levels of interruptions excluding NMI can be set within a range from 0 to 15.
The interruption sources every CPU are shown below.
(A part of interruption lines (*A~ *D) are used as separate functions)
*1: Interruption occurs when data are written from SCPU/TCPU to SM area/TM area of dual port RAM.
*2: Interruption occurs when data are written from MCPU to the MS area of dual port RAM.
*3: Interruption occurs when the scanner internal switch conditions have changed.
7−26 L1E-BW1346
*4: Interruption occurs when data are written from MCPU to the MT area of dual port RAM
*5: Interruption occurs when limit switch LSW conditions have changed
・ Table longitudinal operation ・ Table vertical move operation
(1); Forward LSW (1); Table travel up LSW
(2); Backward LSW (2); Table travel down LSW
(3); Forward speed LSW
(4); Backward speed LSW
Fig.7.15 shows the entire configuration diagram, Fig.7.16and 7.17 show the mounting layout of units and sensors,
and. Fig.7.20 shows the entire block diagram.
7−27 L1E-BW1346
Rotation plate
is
X-ray tube
ss
Chassis
ha
tc
1 sis
PCB unit 2
en
as
m
Ch
I
la
Fi X-ray collimator
II unit
Co
ol
B
e
PC
PC S2
r
FS
TX
D
PC
B
High voltage High voltage
transformer transformer
(−) (+)
Off
brake
Chassis Detector
3 Scan
motor
Preamplifier
Operator X-ray
Foot switch Table
console control unit
Signal line
7−28 L1E-BW1346
4PC:7mm Collimator sensor
6PC:Shutter sensor
7−29A
push-button switch
SPB2
High PCB High
4SW: voltage voltage
EMG. switch generator generator
(-) (+)
Detector
Hydraulic unit
(L)
Preamplifier Preamplifier
3LS:
Chassis 3 lift cover limit switch
Chassis 7
1M:Scan motor
Cable outlet
Left side view of scanner Front view of scanner
L1E-BW1346
Fig.7.16 Mounting layout of units and sensors-1
X-IND PCB unit I
TILT+ TILT- Chassis 2 PCB
Front Rear
Chassis 2
SPB2
PCB
1LS:TILT+
limit switch
2LS:TILT-
limit switch
Hydraulic unit
(R)
Chassis 3
Chassis 7
1RE:
1PT:for TILT 1B:OFF brake Rotary encoder for zebra pulse SENS2A
potentiometer PCB
1M:scan motor
7−30 L1E-BW1346
Table
Power Stationary side control unit Slip ring Rotation plate side control unit
(WT)
distribution board
7−31
X-ray control unit (・Collimator)
(WB)
Foot switch
Scanner unit
Signal
L1E-BW1346
7.7 Power distribution board
The power distribution board is composed of chassis 7 and chassis 3. Fig.7.19 shows the block diagram of the
power distribution board.
The functions of each unit are described below.
7.7.1 Chassis 7
Chassis 7 connects cables to the external units. Also, this chassis distributes AC power supply, GND,
and signal cables. Each cable is connected into the scanner via this unit. The composition of external
cables is shown below.
・ AC power supply : 4TN・ 1/2/3............Input (main power supply) ; AC200V
: 4TN・ 6/7...............Output (Table) ; AC200V
: 4TN・ 10/11/12......X-ray control unit output (starter)
; AC70∼200V
: 1TN・ T1/T2........... X-ray control unit inverter output
(High voltage generator input)
; AC100∼250V, 10kHZ
・ Earth : E, E2 (Hospital side power distribution board)
: E3 (Table)
・ Signals : 7CN (X-ray control unit) ; Serial communication / kV control / X-ray abnormal
/ power on
: 9CN (Image processor) ; Serial communication / MIC / SP
: 10CN(Image processor) ; Measuring data transfer
: 4TN (External speaker output)
7.7.2 Chassis 3
Chassis 3 generates AC 100V for scanner power supply on-off control and scanner control by means of a
power transformer This isolation transformer having a high attenuation effect of eliminating power line
noises provides a step-down (AC200V → AC 100V) function.
This unit is composed of the following parts.
・ Circuit protector (31FFB) : PCB unit/shutting off the power supply of rotation plate interior
・ Pilot lamp (31PL) : Lights when main power supply is being fed
7−32 L1E-BW1346
Chassis 7 PCB unit Ⅰ
WT (control) WT control
WT(GND)
E3
WT(AC200V)
4TN
OC (Communication) OC communication
OC (Measuring data) Measuring data
AC100V
AC100V
Signal
Power
GND
7−33 L1E-BW1346
・ Pilot lamp (32PL) : Indicates that power is being fed to chassis 2/table power supply
・ Power transformer (31T) : AC200V→AC100V (1.8kVA)
Relay 31DR is controlled by the on-off operation of “GANTRY POWER” switch on the operator console.
Electromagnetic contactors 31,32KP are controlled by relay 31DR
・ 31KP: Chassis 2 power supply/table power supply
・ 32KP: PCB unit 1 power supply
・ 31DR: X-ray control unit output (DC 24V)
7−34 L1E-BW1346
Frame Cover
PB-L17 PB-R17
EMG.SW EMG.SW
Tube position sensor SENS2A
PB-M17
Zebra rotary encoder
SPB2 SPB2
Laser projector cross line (R)
(Traffic light)
SPB2 SPB2
Laser projector cross line (L)
X-IND
TILT potentiometer Touch switch
TILT LSW ( + )
Frame
TILT LSW ( − ) PCB unit I
Scan motor
Inverter communication
7−35
MANU.SW Brake
Chassis 2
Selector switch
Front cover LSW
Push button switch
Stationary Hydraulic unit (R)
Push button switch switch
Brake
release switch Hydraulic unit (L)
AC AC
WT control
Fan Fan
Speaker output
OC communication
L1E-BW1346
Fig.7.20 Stationary side control unit block diagram
OC (SP / MIC)
MIC MIC SP
TILT LSW ( − )
24V
Traffic light (B)
Chassis 2
Traffic light (F)
TCPUB PCB
Operation panel
・Serial port/X-TRG
X - IND ・RESET/EMERGENCY
WT (Control)
Error detection SERI DATA
7−36
TXR2A VIEW TRG (*)
OC (Communication)
PCB
OC (measuring data)
5V・12V・± 15V 5V
24V・5V(B)
AC AC 12V
N.F. DCPS
AC Signal
Fan
AC
L1E-BW1346
Fig.7.21 PCB unit block diagram
Dual port RAM
(DPRAM1)
MS.AREA
(first-half of memory)
M.CPU →S.CPU
SM.AREA
(second-half of memory)
Scanner CPU
S.CPU →M.CPU (S.CPU)
block
MS F / F OUT
MS F / F SET MS MS DATA WR
F/F
MS F / F RESET
SM F / F OUT
SM DATA
WR SM SM F / F SET
F/F
SM F / F RESET
Main CPU MT F / F OUT RESET
(M.CPU) MT DATA
block MT F / F SET MT WR
F/F
MT F / F RESET
TM F / F OUT
TM DATA
WR TM TM F / F SET
F/F
TM F / F RESET
RESET
M.CPU → T.CPU
TM.AREA
(second-half of memory)
T.CPU →M.CPU
7−37 L1E-BW1346
SRAM I/ O
Shared
Flash F I F O Serial Shared memory
memory
memory memory MKY28(OC)
MKY28
CLK;
20MHz
I/ O
RESET
I/ O To S.CPU / T.CPU
DIP SW
MS F / F OUT
MS F / F SET
MS F / F
STATUS
LEDS
(SH7032) SM DATA WR
SM F / F
SM F / F RESET
S.CPU
MT F / F OUT
MT F / F SET
MT F / F
FIFO
FIFO TM DATA WR
memory TM F / F
memory
(OC) TM F / F RESET
T.CPU
PCI
I/F Dual port RAM
M.CPU → S.CPU
(OC)
M.CPU ← S.CPU
FIFO (DPRAM1)
memory FIFO
memory
(OC)
M.CPU ← T.CPU
(DPRAM2)
7−38 L1E-BW1346
Stationary switch/inverter error
brake release switch/EMG
switch/release switch/cover open
touch switch/inverter control Chassis 2
I / F PORT
(SRY4)
Projector
I/ O Resend ERR
TXR2A
SRAM
MANUAL
switch
TUBE POSI
sensor
Flash (SENS2A)
memory
Measuring Measuring
system
signal data port
(OC)
I/ O
CLK;
20MHz
DIP SW RS232C
Scanner CPU driver/ RS232C Serial port
I/ O
(S.CPU) receiver (PC)
STATUS
LEDS
M.CPU ZEBRA
rotary encoder
(SH7032)
MS DATA WR
MS F / F
MS F / F RESET
(DPRAM1)
RESET
From M.CPU
7−39 L1E-BW1346
Each CPU block functions as described below
1) Main CPU (M. CPU) block
・ Serial communication (2 ports)
: Rotation plate communication (RS232C) ; Data (1 line)
: PC communication for terminals (RS232C)
・ Rotation plate control signal
: EXP. ; X-RAY ON signal (high-voltage unit output)
: RESET ; CPU reset
: X-TRG ; X-ray trigger (high voltage unit input)
: EMERGENCY ; (high voltage unit input)
: +12V ; Communication power supply
・ Shared memory MKY28 communication (1 port)
: OC (Image processor unit) communication
; Through 8-bit serial. OC → Scanner. FIFO memory
; Through 8-bit serial. Scanner → OC. FIFO memory
・ Inter-CPU communication (2 ports)
: Scanner CPU (via dual port RAM)
: Table CPU (via dual port RAM)
・ Status LED display
: X-ray trigger
: View trigger
: ZEBRA SENSOR
: ABORT
: STATUS(8BIT)
・ PCB operation switch read : RESET switch
・ Function setting DIP switch read
・ Measuring sequence control : Measuring sequence setting
・ Status send-receive between OC (image processor) and WS (scanner)
: EMERGENCY ; OC→WS
: X-RAY ON ; WS→OC
: DEADMAN ; OC→WS
(Deadman relay is released when the table is fed automatically)
: RESET ; OC→WS (CPU reset)
: START ; OC→WS
7−40 L1E-BW1346
2) Scanner CPU (S CPU) block
・ Serial communication (1 port) : For PC communication and terminals (RS232C)
・ Inter-CPU communication (1 port) : Main CPU (via dual port RAM)
・ Scanner internal switch (SW) input (via SRY4 PCB)
: MANU switch (Manual rotation switch)
: ST SCAN SW (Stationary scan switch)
: BRAKE REL SW (Brake release switch)
: TAPE SW (Touch switch)
: Lift cover SW (Lift cover open switch)
・ Scanner internal contact (trip) monitoring (via SRY4 PCB)
: General-purpose inverter input FFB ; 21FFB
: Hydraulic unit input FFB ; 22FFB
・ Rotation plate rotation angle/X-ray tube position detection
: Zebra rotary encoder (1RE) input
: Tube position (SENS2A PCB)
・ Measuring sync. signal control
: View trigger signal (VIEW-TRG) ; For data input
: X-ray trigger signal (X-TRG) ; For X-ray synchronization
: Start angle setting (SA0/1) ; Data input start position
: Generation of table sync signal (S PULSE) by volume scan
; 1 pulse/rotation (TUBE UP SIDE detection)
・ Rotation plate rotation (general-purpose inverter) control
: Scan motor on-off ; CW
: Speed selection ; I5 ~ I7
: Brake release ; OFBRK RELESE
: General-purpose inverter power supply release
; INV. POW. REL
・ Resend PCB (TXR2A) error detection
: PCB error display contents read ; LD1∼LD10
: Error read pulse generation ; PSW1∼PSW3
7−41 L1E-BW1346
・ Function setting DIP switch read
・ X-ray ON display : PB-M17 PCB (lift cover)
: X-IND PCB (rear cover)
・ Traffic light display : SPB2 PCB
・ Status LED display : Zebra sensor
: X-RAY TRIG
: VIEW TRIG
: ABORT
: 8-bit bus status (D0 ~ D7)
MCPUC PCB
LED
LD1 DO(LSB)
LD2 D1
LD3 D2
LD4 D3
Bus status
LD5 D4
LD6 D5
LD7 D6
LD8 D7(MSB)
LD9
ABORT
TP1
LD10
ZEBRA SENSOR
TP2
LD11
VIEW TRIG
TP3
LD12
X-RAY TRIG
TP4
TP5 VIEW TRIG2
TP6 DEADMAN
DGND DGND
CHECK PIN
RESET SW
SW
(Pushbutton switch)
Note: The above figure shows the MCPUC PCB as viewed from the front of the scanner
Status LED (bus status) of MCPUC PCB indicate the operating conditions of software and mean as described in
the next paragraph.
7−42 L1E-BW1346
Meanings of MCPUC PCB status LED (bus status) display
(Displayed sequentially starting with the numeric value corresponding to the initialization start
・80→81→82→・・・・・・)
・ Scanner initialization in progress (for several tens seconds after power supply has been turned on)
Initialization start Status LED = 80H
RAM check (shift data) Status LED = 81H (built-in RAM)
RAM check (55AAH data) Status LED = 82H (built-in RAM)
RAM check (marching) Status LED = 83H (built-in RAM)
RAM check (shift data) Status LED = 84H (external RAM E000000H)
RAM check (55AAH data) Status LED = 85H (external RAM E000000H)
RAM check (marching) Status LED = 86H (external RAM E000000H)
RAM check (shift data) Status LED = 87H (external RAM ...........SCPU DRAM send area) RAM
RAM check (55AAH data) Status LED = 88H (external RAM ...........SCPU DRAM send area) RAM
RAM check (marching) Status LED = 89H (external RAM ...........SCPU DRAM send area) RAM
RAM check (shift data) Status LED = 8AH (external RAM ..........SCPU DRAM send area) RAM
RAM check (55AAH data) Status LED = 8BH (external RAM...........SCPU DRAM send area) RAM
RAM check (marching) Status LED = 8CH (external RAM...........SCPU DRAM send area) RAM
RAM check Status LED = 8DH
If an RAM is abnormal, the LED corresponding to the RAM remains displayed, and the scanner does not
start.
Initial communication check Status LED = 1FH
(Each LED goes out when initial communication is succeeded)
D0 = LED・OFF ; Rotation side CPU initial communication ends
D1 = LED・OFF ; SCPU initial communication ends
D2 = LED・OFF ; TCPU initial communication ends
D3 = LED・OFF ; WB initial communication ends
(If an LED does not goes out, it shows that a communication error occurs)
If a communication error occurs, the following LED light and the scanner does not start
Rotary plate communication error D4LED・ON
SCPU communication error D5LED・ON
TCPU communication error D6LED・ON
WB communication error D7LED・ON
・ After the scanner has been initialized
* Waiting for a command from OC Status LED = 01H
* A command was received form OC Status LED = (Received command display) H
* An error occurs Status LED = (Internal status display) H
* An emergency occurs Status LED = 1FH
* An abnormal interrupt occurs Status LED = FEH
(+ high-order bit of interrupt No. + low-order but of
interrupt No. flicker)
* Watch dog timer error Status LED= FFH
7−43 L1E-BW1346
(2) TCPUB PCB
TCPUB PCB is composed of table CPU (T CPU) to mainly execute table control and operation panel control.
Since this PCB is provided with LED for operation panel and switches, it can be operated during
maintenance. Fig.7.25 shows the block diagram of table CPU (T CPU).
7−44 L1E-BW1346
(Tilt) Inverter unit
I / F PORT
(Via MCPU)
Table
(SW / LED) operation
panel
I / F PORT
(LSW /interlock)
(Longitudinal move/
I/ O Table ※
vertical move control)
I / F PORT
SRAM
ETC SIG
(X-RAY/EMG/deadman) PORT
(OC)
TILT LSW
Flash
memory
Tilt sensor
CLK;
I/ O 20MHz
RS232C
Table CPU driver/ RS232C Serial port
receiver (PC)
DIP SW (T.CPU)
I/ O
STATUS
LEDS ※
Rotary encoder
(table)
(SH7032)
Tilt potentiometer
MT DATA WR
MT F / F AD Offset/gain
MT F / F RESET
※
Table height
potentiometer
TM F / F OUT
TM F / F SET
TM F / F To S.CPU
Table zebra (after frequency division)
M.CPU
(DPRAM2)
RESET
From M.CPU
7−45 L1E-BW1346
Table CPU block executes the following operation.
1) Table (T CPU) block
・ Serial communication (1 port) : PC communication for terminals (RS232C)
・ Inter-CPU communication (1 port) : Main CPU (via dual port RAM)
・ Scanner operation panel control
: Control of operation panel (LED/SW) mounted on TCPUB PCB
: Gantry side operation panel (LED/SW) control
; PB-L17/PB-R17 PCB (SW)
; PB-M17 PCB (LED)
: Operation panel interface signal
; D0~ D7 (LED display data)
; A0 ~ A3 (I/O address)
; STRB (I/O address strobe signal)
: Operation panel LED signal
; Table position (±1999mm/0.5mm pitch)
; Tilt angle (±25°/0.5°pitch)
; Table height (0∼250mm/1mm pitch)
: Operation panel switch input
; PAT SW (PAT operation switch)
; TILT (+) SW (Forward tilt switch)
; TILT (-) SW (Backward tilt switch)
; TABLE DOWN (↓) SW (Table down switch)
; TABLE UP (↑) SW (Table up switch)
; TABLE HSP (《 》) SW (Table high-speed switch)
; TABLE LF (←) SW (Table low speed forward move switch)
; TABLE LB (→) SW (Table low speed backward move switch)
; HOME POSI SW (Table home position switch)
; PRESET SW (Table preset position switch)
Switch on TCPUB PCB side is called SCAN POSI
; RESET SW (Table position “0” clear/Emergency reset switch)
; LIGHT SW (Projector ON/OFF switch)
; EMERGENCY SW (Emergency switch)
7−46 L1E-BW1346
・ Dual port RAM communication (Main CPU ~ table CPU)
・ Table LSW detection
: FORWARD MOVE LSW (Table forward move limit position detection)
: F MOVE SPEED LOW (Speed switching position in table forward move direction)
: BACKWARD MOVE LSW (Table backward move limit position detection)
: B MOVE SPEED LSW (Table backward move direction speed switching position)
: DOWN LSW (Table down limit position detection)
: UP LSW (Table up limit position detection)
・ Tilt (hydraulic unit) operation control ; TILT+/-
・ Tilt LSW detection
: TILT + (1LS) (Tilt forward limit position detection)
: TILT - (2LS) (Tilt backward limit position detection)
・ AD input
: Tilt angle (potentiometer 1PT) detection
: Table height (potentiometer) detection
・ Table position (longitudinal move position) counter input
: RE-A (rotary encoder 1REA phase)
: RE-B (rotary encoder 1REB phase)
・ Table vertical move control
: Table up control (hydraulic pump)
: Table down control (electromagnetic valve)
・ Table longitudinal move control
: Operating direction signal generation
: Speed command signal generation
: Operation enable signal generation
・ Function setting DIP SW read
・ Emergency processing
・ X-RAY ON LED display
: PB-L17/PB-R17 PCB (lift cover)
: X-IND PCB (Rear cover)
・ EMERGENCY ON LED display
: PB-L17/PB-R17 PCB (lift cover)
7−47 L1E-BW1346
(3) TXR2A PCB
For TXR2APCB, refer to “Preamplifier signals “ (7.12).
1DPS 5V20A
2DPS 5V7A/+15V1.2A/−15V0.5A
3DPS 24V2.2A
4DPS 12V4.2A
7.8.2 Chassis 2
Chassis 2 is composed of the general-purpose inverter to drive a scan motor, power supply consisting of
circuit breaker, magnet switch, noise filter, etc., and SRY4 PCB for executing the drive system control for
motor, brake system, etc., drive system error monitoring, tilt control, etc. The brake and tilt control
power supply are turned on or off via an SSR.
7−48 L1E-BW1346
Slip ring
Scan motor
For brake
21SSR
Control signal
Brake
Stationary switch
General-purpose inverter
MCPUC PCB 21SW (INV.)
AC.10KHz
Brake release switch
Touch switch 22SW
SRY4 PCB
Lift cover switch
Tilt thermal (R)
7−49
INV.
Tilt thermal (L) (3 φ )
For hydraulic unit 21NF
Hydraulic unit (R)
SSR INV. FFB To high voltage unit
Hydraulic unit (L) auxiliary contact
22SSR ~ 23SSR INV. 2TN
21KP (T1,T2,E)
TILT. FFB
auxiliary contact
3 φ AC200V
1 φ AC200V
L1E-BW1346
Fig.7.26 Inverter unit block diagram
Each component unit functions as follows
7−50 L1E-BW1346
7.8.3 Frame
(1) SENS2A PCB
This PCB is provided with a transmittance type photo sensor and mounted on the side face of the
large pulley of the mechanism. It outputs a signal when the X-ray tube is positioned just above.
(2) Potentiometer
The potentiometer is set in such a way as the resistance value becomes a half at a tilt angle of 0°,
and the resistance changed due to tilt operation is calculated as an angle by CPU using the above
resistance value as a reference value.
・ Potentiometer : 1PT (5kΩ, 10 turns)
(3) Rotary encoder
The rotary encoder outputs a zebra pulse for detecting the rotating angle of the rotation plate.
・ Rotary encoder : 1RE
In case of Pronto-XE/SE
(RE resolution ; 1024 pulses/rotation ; Rotation plate 9600 pulses/scan)
In case of Pronto-VE
(RE resolution ; 1800 pulses/rotation ; Rotation plate 10800 pulses/scan)
(4) Limit switches
These limit switches detect an overrun during gantry tilt.
7−51A L1E-BW1346
7.8.4 Cover assembly
(1) SPB2 PCB
This PCB instructs a breathing spell to a patient. It is mounted before and behind the cover and on
both sides of the cover (4 positions in total). It lights in synchronization with scan.
・ Breathing spell is allowable (Green LED lights)
・ Breathing stop is instructed (Orange LED lights)
(2) X-IND PCB
This PCB is mounted on the rear cover and its LED lights during X-ray ON operation
・ “X-RAY ON”
: indicates that X-ray exposure is in progress (Yellow LED lights)
(3) Operation panel PCB
・ PB-L17/PB-R17 PCB
: Table up/down/home switch
・ PB-M17 PCB
: X-ray on/emergency LED
: Emergency switch relaying
(4) Microphone amplifier PCB
This PCB amplifies the input from the condenser microphone for reducing noises between this
amplifier and intercom amplifier, and then, feeds the input to the intercom amplifier. This PCB is
mounted at two upper parts before and behind the gantry open port.
7−52 L1E-BW1346
7.9 PCB unit II (Rotation plate PCB unit)
7.9.1 CPU85A PCB
This PCB is a CPU board composed mainly of microprocessor 8085AC-2, interval timer, ROM and
RAM.
ROM (8kbytes) and RAM (8kbytes) can be mounted up to 6 pieces in total.
This maintenance circuit comprises a maintenance program start switch, a maintenance program selection
digital switch, and a status indicator.
This is one of the interface circuits between μ-CPU circuit and peripheral circuit to enable μ-CPU to
read the data obtained by AD conversion of analog values such as X-ray generation system status, tube
current, filament primary voltage, etc.
This is one of the interface circuits between µ-CPU circuit and peripheral circuit to feed the control
information from μ-CPU to each control circuit. This control information comprises the tube current
parameter which is DA-converted in this circuit and sent to the control circuit. This circuit is also
provided with counter type timers whose modes can be set or preset from CPU and used for the following
purposes.
Uses Setting
Backup timer Sets to 110% of the scan time from CPU when X-ray is turned on.
7−53 L1E-BW1346
Stationary side
COMMU2 CPU85A
PCB unit
CONT
IFB-1 Filament heating circuit
B-1
MAINTE
7−54
IFB-2
CONT
SIFB-2 High voltage generator
D3
7.9.7 CONT B-1, D3 circuit is described by function in each paragraph together with the drive circuit.
7−55 L1E-BW1346
7.10 Tube current control circuit
7.10.1 Filament heating circuit
The filament heating circuit is composed as shown in Fig.7.28. The 48V power output is converted into
a necessary DC voltage by the chopper circuit. This chopper circuit is operated at 5kHz to be able to
obtain a DC voltage proportional to the duty of chopper control signal A in TP1 in the figure. This DC
voltage is converted into AC by the inverter consisting of two transistors, and supplied to the X-ray tube
filament through a heating transformer in the high voltage generator. The inverter is operated at 625Hz
with a fixed duty. Accordingly, the heating condition of the X-ray tube filament is variable by the
operating duty of the chopper circuit and all heating condition operations are done by the operating duty
of the chopper circuit. The filament heating circuit is provided with a built-in resistor to divide the DC
output voltage of the chopper circuit and a resistor to detect a current flowing into the filament heating
circuit. An over-current, an over-voltage, and breakage of filament are detected by these resistors.
7−56 L1E-BW1346
Heating transformer
(High voltage generator)
Chopper
TP1
+
1φ
48V
AC 100V X-ray tube assembly
−
○
A
+ +
○
B ○
C
5kHz
7−57
Inverter
0.1Ω
625Hz
−
− Filament over-current
+
+
301VR
+
Filament disconnect
−
302VR
−
Filament over voltage
+
before exposure
8B
reference
8C 6A
8B ○
A
5kHz
Chopper
Differential amplifier clock
Current-voltag Integration
e conversion circuit 6B
4VR MM (pre-heating)
Frequency
divider MM (0.67ms)
1.25kHz
5kHz ON
5B
5C
clock 625Hz ○
B
ON
○
C
7−58 L1E-BW1346
7.10.3 kV, mA error detection circuit
An abnormal tube voltage or an abnormal tube current is detected by +kV, -kV (tube voltage divider
output) and NE+, NE-, N+, N- (high voltage transformer neutral point current) in Fig.7.31.
Fig.7.30 shows the kV, mA error detection circuit.
For the tube voltage
● Unbalance of the tube voltage between anode side and cathode side (Unbalance: Higher than 15kV
and higher than 35kV).
● An excessive tube voltage (The ground voltage is higher than 75kV) are also detected.
7−59 L1E-BW1346
Comparator
Excessive + tube
OV VT+
voltage
Excessive - tube
OV VT−
voltage
Anode current
NE○
− over-current + side OC IT+
(IG1)
+
N○
Addition
(IG3)
over-current - side OC IT−
Over-current
higher than 1A
+
NE○
(IG2)
−
N○
Addition
(IG4)
Cathode current
Over-current
75/185mA
Excessive center
ICS OC ICS
section current
Current
switching signal
7−60 L1E-BW1346
7.10.4 High voltage generator
This is a single-phase 2-peak rectifying circuit where the anode side and cathode side are independent of
each other. It feeds a composite output to the X-ray tube. It is provided with voltage-dividing resistors
for tube voltage waveform monitors of respective outputs The cathode side is provided with a built-in
filament heating transformer to feed a filament current to the X-ray tube.
Fig.7.31 shows the X-ray high voltage circuit.
T1
100MΩ
T2
NE +kV
NE NE
○
+ ○
−
5KΩ kV+
5Ω 5Ω
5Ω CS
Anode side
ICS
Cathode side 5Ω 5Ω
N N
−
○ +
○ 5KΩ kV−
N
−kV
T1
100MΩ
T2
CL1
C
CL2
X-RAY
H.V.GENERATOR H.V. CABLE TUBE
7−61 L1E-BW1346
7.11 Slip ring
All power supply and signals to the scanner rotation plate are supplied through the slip ring. No. of rings amounts
to 23 channels as shown below.
7−62 L1E-BW1346
7.12 Preamplifier signal
Conventional 16-bit parallel signal is converted into a serial signal through slip ring. This serial-parallel
conversion is done through slip ring by 2 signal cables consisting of a data cable and a response cable
Serial-parallel conversion PCB (TXR2A, TXS2) provide a function of resending data by VIEW TRIG signal, if
data are abnormal in the course of serial-parallel conversion or data send-receive operation. The block diagram
is shown below.
Fig.7.32 shows the preamplifier output block diagram.
SCAN MODE
View TRIG
Data CLK
IDATA 0 ~ 15
Data CLK serial
IDATA0 ~ 15
IDATA0 ~ 15 parallel
parallel
View TRIG
RTN TRIG
7−63 L1E-BW1346
7.12.1 Resend function
The preamplifier signal transfers data every view by means of View TRIG signal via serial-parallel
conversion PCB (TXS2, TXR2A).
TT VN DATA LRC
7−64 L1E-BW1346
7.12.3 TXR2A PCB
TXR2A PCB provides the following functions by PSW 1 ~ 3.
(1) By pressing PSW1 once, display errors can be selected in 4 kinds.
(2) By pressing PSW2 together with DIP SW, data transfer between TXR2A and TXS2 can be checked.
(3) By pressing PSW3, the error count being stored into TXR2A PCB memory can be cleared.
TXR2A PCB indicates the following contents by LED display.
(1) Display address (LED9, 10)
Indicates four kinds of errors by binary 2 bits.
(2) VN non-coincidence error count (LED 1 ~ 8) (Display address 0)
Indicates an error count if the view number of data sent from TXS2 is not a specified number.
(3) LRC error count (LED1 ~ 8) (Display address 1)
Indicates an error count when LRC is not a specified numeric value.
(4) TAXI VOL error count (LED1 ~ 8) (Display address 2)
Indicates an error count detected by IC in TXR2A.
(5) TT receive count during data receive (LED1 ~ 8) (Display address 3)
Indicates an error count when the start of data of one view does not meet the specified signal.
These pieces of error information are read by MCPUC and sent to the image processor as error log
information.
7−65 L1E-BW1346
7.13 Tube shift Mechanism (In case of Pronto-SE 3.5MHU, Pronto-VE 3.5MHU)
7.13.1 General
When the tube gets warm and cool due to continuous scanning or patient change, expansion and
contraction of the anode or housing itself causes the focal spot position of the tube bulb to change. As a
result, the beam moves in the slicing direction with the collimator being as a fulcrum, and a position of
incidence into a solid detector shifts. Sensitivity characteristic of the solid detector is not uniform, and
when the beam position of incidence shifts the sensitivity characteristic comes to differ from that during
calibration, thereby generating ring artifact.
Focal spot moves.
100%
Collimator
Sensitivity
Solid detector
95%
Fig. 7.13.1 Focal spot move and beam shift Fig. 7.13.2 An example of sensitivity characteristic
in slicing direction of the solid detector
To prevent the artifact from generating, it should be necessary to measure the focal spot position and to
adjust the position by correction depending on the amount of displacement so that the focal spot will
never move. The way to make this correction by letting the tube bulb move in the slicing direction is
the tube bulb shift mechanism.
7−66 L1E-BW1346
7.13.2 Composition
Fig. 7.13.4 illustrates the composition of the tube bulb shift mechanism. The tube bulb shift mechanism
is mainly comprised of three sections:
Focal spot position detector which detects the focal spot position,
Control section which measures the displacement of the focal spot based on the output from the detector
and controls the movement of the X-ray tube, and Drive section which actually forces to move the tube
bulb by the signals from the control section.
Control section Drive section
Opening for beam incidence Beam for detecting focal spot position
CCW detector
Reference detector
CW detector
7−67 L1E-BW1346
When the focal spot of the tube is placed at regular position, output from the CW detector equals that
from the CCW detector. But if the focal spot moves in the direction toward the table, the beam shifts
toward the gantry side with the collimator being as the center.
Consequently, the output from the CW detector increases and the output from the CCW detector
decreases. On the contrary, if the gantry moves toward the gantry, the beam shifts toward the table with
the collimator being as the center. As a result, the output from the CCW detector increases and the
output from the CW detector decreases. This change of the outputs is measured by the control section
which the outputs is measured by the control section which the outputs is measured by the control section
which follows, and it sends a command to the tube bulb drive section. Since it is assumed that multiple
solid detectors are used, the last 24th channel has a separator on one side only, which is placed at a
boundary between channels, and the light generated by the scintillator breaks through at the side of no
separator, thereby causing the output sensitivity to be lowered compared with the other channels.
Therefore, in cases that this solid detector is used as a focal spot move detector, the 24th channel with
less output level cannot be used. To balance the CCW and CW detectors, the first channel, the opposite
channel of the 24th channel, is not used.
Separator
1ch 2ch 3ch 4ch 5ch 6ch 7ch 21ch 22ch 23ch 24ch
Light Light
Scintillator
Photodiode
7−68 L1E-BW1346
This hysteresis characteristic is set by the comparator so that it will become valid when the beam
movement on the focal spot detector exceeds the allowable value and will continue to be valid until the
beam position is corrected to the reference position by shifting the tube bulb. The output from
differential amplifier at the moment when the beam movement exceeds the allowable value differs
depending on the tube current and voltage current. So the starting point of operation has been
determined based on the output from the reference detector.
Furthermore, since the output from the focal spot detector differs depending on the tube voltage, the
output difference must be corrected. The gain of the I/V converter is switched by the kV signal from the
scanner to correct the output difference.
In addition, to avoid operation while the X-ray is not generated, a tube bulb shift command is not given
unless the reference output exceeds the specified value.
CW detector I/V
Tube bulb shift Shift enabler
command Operation
starting point
Detection of X-ray output
Reference
I/V
detector
Shift enable
GAIN switching
signal
kV signal I/V
kV checker
The control signal generator section serves to generate pulses to the stepping motor which shifts the tube
bulb.
In controlling the stepping motor, it is necessary to give low frequency pulses to rotate the motor at its
startup, and to give gradually higher frequency pulses. Then the pulses are generated by the V/F
converter. Frequency control is made by using ramp output from the integrator so that it changes the
frequency from lower to higher.
If the stepping motor cannot be started for some reason, the pulse frequency goes high, the stepping
motor is not given by the sufficient torque to rotate it, and it will remain stopped.
7−69 L1E-BW1346
Since the control section is always giving shift command pulse because the stepping motor remain
stopped, the control system will also be stopped. To avoid this, a timer is provided in the pulse generator
circuit so that the integration circuit is reset at a certain period and the pulses are started to be given again
from lower frequency while the shift command is being given.
Reset signal
Timer
Lamp signal
v − V/F
Pulse signal
+ converter
Integrator
7−70 L1E-BW1346
7.13.6 Shift mechanism of X-ray tube unit
The shift mechanism is the mechanism to shift the X-ray tube unit in the back-and-forth (Z) direction,
Fig.7.13.10 shows its mechanism.
The shift mechanism of the X-ray tube unit is comprised of the holder ① and drive unit ②.
The holder ① is supported by the linear guide ③ and drive screw ④, allowing the mechanism to
move in Z-direction.
The both ends of the drive screw ④ are supported by the bearing ⑥ and the one end of the drive screw
④ is connected to the stepping motor ⑧ via the coupling ⑦.
Rotation of the stepping motor is transmitted to linear motion in Z-direction of the holder ① through the
drive screw.
As pitch of the drive screw ④ is 2mm, the holders moves by 2mm per one rotation of the drive screw.
③
⑦
⑧
⑤
Turn plate
Z-direction
④
⑥
Fig. 7.13.10
7−71 L1E-BW1346
7.14 Shutter mechanism
7.14.1 General
The correction of X-ray tube focus movement is made by the X-ray tube shift mechanism. This
correction can be done only while X-ray are being exposed. For this reason, if the focus position is
largely changed in such case as the X-ray tube is cooled due to long scan interval, or immediately after
the gantry is tilted, correction of focus position requires much time. Moreover, a ring artifact is apt to
occur from the first to second image due to incomplete correction.
In this case, prevention of ring artifact generation can be performed by preliminary exposing X-ray to
correct the focus movement prior to exposure.
However, these exposures are not mean to take image.
As a countermeasure, the system is provided with a shutter mechanism not to expose the patient to the
harmful X-rays by closing the shutter when the preliminary exposure is effected. With the aid of this
mechanism, the actual scan is possible from the first scan as the correction of focus position is completed
by ignoring those vain exposure.
7.14.2 Composition
7−72 L1E-BW1346
Collimator (function as collimator)
The shutter composition is described in Fig. 7.14.1 and can shield the main fan shaped beam by sliding
the collimation plate to close position.
7−73 L1E-BW1346
7.14.4 Preliminary exposure sequence
Scan selection
STANDBY ON
START ON
Scan
The sequence from the selection of scan to the beginning of the actual scan is described in Fig. 7.14.3.
After the scan condition is set and the STANDBY button is pressed, the measurement control software
compares it with the former scan and checks whether preliminary exposure is necessary or not. In case
the preliminary exposure is necessary, the shutter turns on to effect preliminary exposure. After
completion of preliminary exposure, the shutter turns off to prepare the actual scan. When preparation
to scan is finished, the START button starts to flickers.
In case the preliminary exposure is not necessary, press the STANDBY button to prepare for actual scan.
When preparation is finished, the START button starts to flickers. Pressing the START button effects a
scan.
The patient table is a fixed canti-lever support type. Its table top can move vertically and longitudinally. This move
is controlled according to the scanner operation.
Table top
8−1 L1E-BW1346
8.2 Longitudinal move device
8.2.1 Longitudinal move drive device
Fig.8.2 shows the longitudinal move device structure.
Table top ○
34 is fixed to table top fixing base ○
35 . Chain ⑲ is wound to the sprockets ⑰ ⑱ on the
motor mounting plate ⑫ and front mounting plate ⑬. The drive motor ⑭ output is transferred to
sprocket ⑰ by spur gears ⑮ ⑯ and then, transferred by chain ⑲. Table top 俗 moves
longitudinally when the chain clamper ⑳ fixed on the rear face of the table top fixing base ○
35 is
engaged with chain ⑲. The moving speed is 60mm/sec in high speed feed and 5mm/sec in low speed
feed.
26 25 22 27 23 31 30 24 21 33 28 29
1 13 18 32 19 20 17 12 14
34 35
20
16 10 15
⑫ Motor mounting plate ⑲ Chain 即 Forward move power supply limit switch
⑬ Longitudinal mounting plate ⑳ Chain clamper 息 Backward move deceleration limit switch
⑩ Longitudinal move rotary 蔵 Pulley 捉 Backward move limit switch
encoder
⑭ Drive motor 贈 Pulley 束 Backward move power supply limit switch
⑮ Motor side spur gear 造 Wire rope 測 Cam (mounted on table top fixing base)
⑯ Sprocket side gear 促 Wire rope fixture 足 C-frame
⑰ Sprocket 側 Forward move 速 Reduction gears
deceleration limit switch
⑱ Sprocket 則 Forward move limit 俗 Table top
switch
属 Table top fixing base
8−2 L1E-BW1346
The number of revolutions of drive motor is as specified below.
× Sprocket ⑰ pitch.
2460 1 64
= × × ×18× 9.525 =60mm/s
60 75 100
209 1 64
Low-speed feed = × × ×18× 9.525=5mm/s
60 75 100
○
23 is wound. Wire rope ○
23 is connected by the wire rope fixture ○
24 on the side face of table top
fixing base ○
35 . As the table top fixing base ○
35 moves longitudinally, pulley ○
21 rotates so that the
table top move quantity is detected by the number of turns of the longitudinal move rotary encoder ⑩
being coupled to pulley ○
21 . Longitudinal move rotary encoder ⑩ generates 1000 pulses per rotation.
Accordingly, 100 pulses are generate per mm of the table move quantity by the following relation.
1000
= ×8 = 100 pulses
π×25.465
8−3A L1E-BW1346
By detecting this number of pulses, the longitudinal move speed and move quantity are controlled. The
move stroke of table top is 1470mm. Three limit switches ○
28 ,○ 30 and ○
29 ,○ 25 ,○
26 ,○
27 are mounted at the
Upper frame
A D Guide rail
Hydraulic cylinder
C B Guide rail
Link
Feedback hose
8−4A L1E-BW1346
8.3.1 Vertical move mechanism
The upper frame is supported by A-B and C-D links. These links are coupled to the base by shaft C and
the upper frame by shaft A. By means of a hydraulic cylinder, the slide rollers of links A-B and C-D
slide along their guide rails to move the C-frame vertically. (See Fig.8.3)
8−5 L1E-BW1346
8.3.3 Mounting layout of parts
2 1 4 10
3
10
9 7 8 ,11
① 11LS forward move power supply limit switch ⑦ 8LS higher-limit limit switch
② 3LS forward move deceleration limit switch ⑧ 9LS lower-limit overrun limit switch
③ 4LS forward move limit switch ⑨ 1VR vertical move potentiometer
④ 10LS backward move power supply limit switch ⑩ 1RX longitudinal move rotary encoder
⑤ 2LS backward move deceleration limit switch ⑪ 5LS lower-limit limit switch
⑥ 1LS backward move limit switch
Fig.8.4 shows the mounting layout of each detector and control device.
8−6 L1E-BW1346
8.4 Electric control of patient table
8.4.1 General
All patient table operation is controlled externally, so that various sensors and motor control PCB only
are mounted inside the patient table. This paragraph describes the motor control PCB and the height
display only. For the other control circuits, refer to each PCB in the scanner.
Pronto – SE / XE / VE
TECHNICAL GUIDE
--- Installation & Adjustment ----
Tokyo, Japan
L2E-BW1348-3
NOTE: The tightening torque of bolts, screws, nuts, and double nuts in the work specified in this manual
shall conform to the values shown in Table 1 and 2.
M3 98 (10) 69 (7)
Flat head screw M4 216 (22) 147 (15)
Pan head screw M5 392 (40) 275 (28)
M6 540 (55) 373 (38)
M3 108 (11) 79 (8)
M4 265 (27) 187 (19)
M5 539 (55) 392 (40)
Hex. head bolt M6 833 (85) 588 (60)
M8 1990 (203) 1392 (142)
M10 3528 (360) 2450 (250)
M12 5586 (570) 3920 (400)
( 1A ) 1 L2E-BW1348
Table 2 Tightening torque of nuts*
Low nut
NOTE: Do not install and store the equipment in the following place.
・ Ambient temperature is below –10℃ or above +40℃.
・ Barometric pressure is below 700hPa or above 1060hPa.
・ The equipment is exposed to harmful gas.
・ Humidity is excessively high.
・ The equipment is exposed to steam.
・ The equipment is exposed to water drops.
・ There is much dust or sand dust.
・ There exists excessive oil mist.
・ The equipment is exposed to saline air.
・ There is explosive gas or dust.
・ The equipment receives excessive vibration or mechanical shock.
・ The floor is inclined more than 10 degree from the horizontal level.
・ The mains voltage fluctuates abnormally (10% or more).
・ The mains voltage excessively drops while the equipment being operated under load.
・ The equipment is exposed to direct sunlight.
(2) 2 L2E-BW1348
Contents
Page
Section 1 CARRYING - IN, MOUNTING, AND ASSEMBLING ................................................................ 1- 1
1.1 Carrying in and mounting of scanner unit ............................................................................. 1- 1
1.2 Leveling of scanner unit ........................................................................................................ 1- 6
1.3 Carrying in the patient table .................................................................................................. 1- 7
1.4 Leveling and fixing of patient table....................................................................................... 1- 8
1.5 Mounting of other units ......................................................................................................... 1- 12
1.5.1 Image processing unit................................................................................................ 1- 12
1.5.2 X-ray control unit ...................................................................................................... 1- 13
1.6 Mounting of X-ray detector................................................................................................... 1- 14
1.6.1 In case of Pronto-XE ................................................................................................. 1- 14
1.6.2 In case of Pronto-SE,VE ........................................................................................... 1- 16
1.7 Mounting of X-ray Tube and cooler...................................................................................... 1- 19
1.7.1 Mounting procedure of X-ray tube assembly............................................................ 1- 19
1.7.2A Mounting procedure of cooler................................................................................... 1- 20
1.7.2B Mounting procedure of cooler................................................................................... 1- 21
1.7.2C Cooler mounting procedure....................................................................................... 1- 22
1.7.2D Mounting procedure of cooler................................................................................... 1- 23
1.7.3 X-ray tube assembly mounting procedure (3.5MHU for Pronto SE/VE)
Tube shift mechanism type........................................................................................ 1- 23-1
1.7.4 Cooler mounting procedure....................................................................................... 1- 23-2
1.8 Removal of protective materials for transport ....................................................................... 1- 24
1.9 Internal inspection of patient table ........................................................................................ 1- 25
1.10 Mounting of covers ............................................................................................................... 1- 27
Section 2 WIRING .......................................................................................................................................... 2- 1A
2.1 Wiring of cables between units ............................................................................................. 2- 1A
2.2 Wiring of cables in each unit................................................................................................. 2- 6
2.2.1 Scanner...................................................................................................................... 2- 6
2.2.2 Table wiring .............................................................................................................. 2- 9
A. In case of Pronto XE/SE ........................................................................................................ 2- 10
2.2.3 Mounting of foot switch and electric parts cover of the table. .................................. 2- 10
B In case of Pronto VE.............................................................................................................. 2- 11-1
2.2.3 Mounting of foot switch and electric parts cover of the table. .................................. 2- 11-1
2.2.4 Image processing unit................................................................................................ 2- 12
2.2.5 X-ray control unit ...................................................................................................... 2- 20A
2.2.6 Step down transformer .............................................................................................. 2- 23
2.3 Internal wiring of system....................................................................................................... 2- 25
2.3.1 Wiring on scanner rotary plate .................................................................................. 2- 25
2.3.2 Detector cabling work ............................................................................................... 2- 27
Section 3 ADJUSTMENT ............................................................................................................................... 3- 1
3.1 Turning on the power supply................................................................................................. 3- 1
3.2 Operation check of image processing unit in central operator console ................................. 3- 1
3.3 Operation check and adjustment of scanner unit................................................................... 3- 2
3.4 Operation check of patient table and adjustment of scanner gantry tilt................................. 3- 10
3.5 Adjustment of X-ray system
(In case of Hitachi 1.5M/2M/3.5MHU or Varian 2MHU X-ray tube is mounted) ............... 3- 17A
3.5.1 X-ray tube assembly check........................................................................................ 3- 17A
( 3A ) 3 L2E-BW1348
3.5.2 Anode rotation drive unit (starter) adjustment .......................................................... 3- 17A
3.6 Adjustment of X-ray system (In case of Varian 3.5MHU X-ray tube is mounted)............... 3- 18
3.6.1 X-ray tube assembly check........................................................................................ 3- 18
3.6.2 Anode rotation drive unit (starter) adjustment .......................................................... 3- 18
3.7 kV, mA waveform check....................................................................................................... 3- 24
3.8 Filament heating conditions adjustment before X-ray exposure ........................................... 3- 25
3.9 X-ray tube aging .................................................................................................................... 3- 26
3.10 kV, mA adjustment and check............................................................................................... 3- 27
3.11 Adjustment of X-ray field ..................................................................................................... 3- 28
3.11.1A Adjustment in X-axis direction (in case of 2 or 3.5MHU X-ray tube assembly) ... 3- 28
3.11.2A Adjustment in Z-axis direction (in case of 2 or 3.5MHU X-ray tube assembly).... 3- 28
3.11.1B Adjustment in X-axis direction (in case of 1.5MHU X-ray tube assembly) .......... 3- 29
3.11.2B Adjustment in Z-axis direction (in case of 1.5 MHU X-ray tube assembly).......... 3- 29
3.11.3 R-direction adjustment (Common to each X-ray tube assembly)........................... 3- 30
3.11.3A Additional item to the adjusting method in X, Z, and R directions in case
of Pronto-VE .......................................................................................................... 3- 31
3.11.4 Normal condition of X-ray tube ............................................................................. 3- 31-1
3.11.5 X-ray beam radiation adjustment (slice thickness direction).................................. 3- 32-1
3.11.5A In case of Pronto SE/VE (See Figure .3.11.7) ..................................................... 3- 32-1
3.11.5B In case of Pronto-XE (See Fig.3.11.9) ................................................................... 3- 34
3.12 Preamplifier check................................................................................................................. 3- 37
3.12.1 Bus line check ........................................................................................................... 3- 37
3.12.2 Output sequence check.............................................................................................. 3- 38
3.13 Checking the influences of vibration noises on measuring data............................................ 3- 41
3.14 Center adjustment.................................................................................................................. 3- 45
3.15 Compensator position adjustment ......................................................................................... 3- 48
3.16 Focus shift detector (FSD) adjustment .................................................................................. 3- 50A-2
3.17 Phantom calibration............................................................................................................... 3- 51
3.17.1 Procedure................................................................................................................... 3- 51
3.17.2 Mounting method of phantom mounting bracket...................................................... 3- 52
3.17.3 X-ray detection system warming up.......................................................................... 3- 55
3.17.4 X-ray tube warm up and air calibration..................................................................... 3- 55
3.17.5 Phantom position adjustment and phantom calibration............................................. 3- 58
3.18 CT number adjustment .......................................................................................................... 3- 67
3.18.1 Procedure................................................................................................................... 3- 67
3.18.2 CT number adjustment scan ...................................................................................... 3- 68
3.18.3 CT number adjustment calculation............................................................................ 3- 70
3.19 Laser marker adjustment ....................................................................................................... 3- 73
3.20 Inter-phone (intercommunication) check .............................................................................. 3- 76
3.21 Hospital name entry............................................................................................................... 3- 77
Section 4 Reference data ................................................................................................................................. 4- 1
4.1 Service scan ([Phantom Calibration][Maintenance Scan] [CT Number Adjustment]) ......... 4- 1
4.1.1 Service scan display/non-display selection ............................................................... 4- 1
4.1.2 Stationary scan .......................................................................................................... 4- 2
4.2 Detector - preamplifier channels correspondence table......................................................... 4- 3
4.3 Connection of laser imager (L.I.) to the image processing unit ............................................ 4- 5
4.3.1 Connection of imager control cable .......................................................................... 4- 5
4.3.2 Video signal connection ............................................................................................ 4- 6
( 4A ) 4 L2E-BW1348
Pronto XE SE VE
Heat Storage 1.5 2.0 2.0 3.5 3.5 2.0 3.5 3.5 3.5
Manufacture Hitachi Varian Hitachi Varian Hitachi Varian Varian
Tube Shift - - ○
Section 1 1.1 Carrying in and mounting of scanner unit ○ ○
CARRYING-IN, 1.2 Leveling of scanner unit ○ ○
MOUNTING, AND 1.3 Carrying in the patient table ○ ○
ASSEMBLING 1.4 Leveling and fixing of patient table ○ ○
1.5 Mounting of other unit ○ ○
1.6 Mounting of X-ray 1.6.1 In case of Pronto-XE(CT-WD-5C) ○ -
detector 1.6.2 In case of Pronto -SE(CT-SD-4A), VE(CT-SD-4) - ○
1.7 Mounting of X-ray Tube 1.7.1 Mounting procedure of X-ray tube assembly Pronto-XE(1.5/2/3.5MHU), Pronto-SE(2MHU) ○ ○ -
and cooler 1.7.2 A Mounting procedure of cooler (XE:U-5SA-01VS) ○ - -
1.7.2 B Mounting procedure of cooler (XE:UKM-5TA-01V/UK-6TA-01VE(A)) - ○ - -
1.7.2 C Mounting procedure of cooler (SE:UKM-5TA-01TF) - ○ -
1.7.2 D Mounting procedure of cooler 3.5MHU X-ray tube(XE) - ○ -
1.7.3 X-ray tube assembly mounting procedure (3.5MHU for Pronto-SE/VE)Tube shift mechanism type - - ○
1.7.4 Cooler mounting procedure (3.5MHU for Pronto-SE/VE)Tube shift mechanism type - - ○
1.8 Removal of protective materials for transport ○ ○
1.9 Internal inspection of patient table ○ ○
1.10 Mounting of covers ○ ○
Section 2 2.1 Wiring of cables between units ○ ○
WIRING 2.2 Wiring of cables in each 2.2.1 Scanner ○ ○
units 2.2.2 Table wiring ○ ○
A 2.2.3 In case of Pronto XE/SE ○ ○
B 2.2.3 In case of Pronto VE - - ○
2.2.4 Image processing unit CASE 1
CASE 2 (Refer to KGC-0385:S/N A11717201∼)
2.2.5 X-ray control unit ○ ○ ○ Note7)
(5)
※3950
※2545 (3750) 1405
830 690 (1205)
(In case of (In case of
backward tilting) forward tilting)
800 300
(In case of (100)
backward tilting)
740
(100)
(1090)
250
1240
1980
500
1000
(2380)
2750
500
1035
1010
(300)
520
※1175 365
Note 1) Don’t put any obstacle within the two-point dot-dash line when mounting the scanner
unit. Numeric values show the required sizes.
2) Parenthesized dimensions in ( ) show the mounting limit sizes.
3) The dotted line portion shows the required area for opening the front panel and rear
cover of the scanner unit during servicing.
4) The shadowed portion shows the movable range of the patient table.
5) For asterisked sizes (*), their required sizes depend upon the option attachments of
the patient table.
1−1 L2E-BW1348
7250N(725Kgf) 370 633 1414
147
35
105
90
420 64
1980
366
426
640
843
2800N(280Kg)
1980
7250N(725Kgf) φ 22 hole (an extra hole)
Scale 1:30
4-M10 anchor bolt Unit mm
3585
1980
1800min.
990
1−2 L2E-BW1348
Dimensions and weight when carrying in the scanner and patient table
1) Standard carrying in
Scanner : 1980mm wide×900mm deep×1870mm high, 1480kg (with cart)
Patient table : 650mm wide×2120mm long×740mm high, 280kg
(with cart having a width of 650mm)
(3) The patient table can be positioned by using the attached templates.
After lowering the scanner unit to the specified position, mount the templates according to the scanner
position and scribe the patient table position. (See Fig.1.1.4)
The patient table position shows an approximate reference position when using the templates.
For precise positioning and anchor bolt fixing, observer [1.4 Leveling and fixing of patient
table].
4 positions (Notch)
Scribe L-holes
Align the notch of the scanner
4 positions base plate with the notch of the
template ①
500
① ②
1−3 L2E-BW1348
(4) Lower the scanner unit onto the floor.
1) After opening the lift cover, lower the scanner unit horizontally by alternately loosening four front, rear,
right, and left caster adjusting screws until its bottom cover touches the floor face.
2) Remove four cart brackets and connecting members 1, 2, 3 connecting the right and left stands.
Connecting member 1
Connecting member 3
Fixing screw
Cart bracket
Connecting member 2
1−4 L2E-BW1348
(5) Remove two suspension jigs and two binding jigs from the scanner unit.
Binding jig
(right and left)
2-M6x20L
hexagon socket head bolt
(plain washer)
Mount these bolts after
removing the jigs
6-M8x30L
hexagon socket head bolt
Mount these bolts after removing the jigs
(Caution)Tightening torque shall be 1960N・cm
( 200kg・cm)
Base retaining jig
Link retaining jig
1−5 L2E-BW1348
1.2 Leveling of scanner unit
Level the scanner unit according to the following procedure before fixing it to the floor with anchor bolts.
(1) Check the level of the scanner gantry by using a level gauge on both front and rear connecting members
provided between the right and left stands and also on both right and left stand bases as shown in Fig.1.2.1.
(2) Level the scanner unit by adjusting the leveling screws of the scanner stand until the bubble of the level
comes to the center. Place an adjusting plate below the stand bottom cover.
Bottom cover
Drive in Tighten
Dome washer Anchor bolt Continue tightening until
dome washer becomes flat
Stand base
1−6 L2E-BW1348
1.3 Carrying in the patient table
The patient table is packaged as mounted on a transport cart. After unpacking, carry in and mount it according
to the following procedure.
(1) Bring the patient table to the mounting position.
(2) Mount the patient table on the floor face.
1) Lift the rear portion of patient table by a hydraulic truck.
2) Remove the truck mounted at the rear portion of the patient table from the table base.
(a) Loosen the set screws of the truck.
(b) Remove the fixing screws from L-bracket.
(c) Pull the pin of L-bracket out of the table base and remove the L-bracket.
(d) Draw out the pin on the opposite side of the truck from the table base.
3) Mount the installing jig to the table base.
4) Tighten the supporting bolts in installing jigs.
5) Put the receiving plate on the floor.
6) Lower the patient table in such a way as the supporting bolts meet the groove in the receiving plate to
support the rear portion of the patient table.
7) Lift the front portion of the patient table by a hydraulic truck.
8) Support the front portion of the table according to steps 2) ~ 6).
9) Remove the hydraulic truck from the patient table.
10) After loosening four supporting bolts alternately to lower the patient table horizontally until the table
base touches the floor.
11) After the table base has landed on the floor, loosen the supporting bolts furthermore and remove the
receiving plates.
12) Remove the installing jig from the table base.
Supporting bolt
Installing jig
Hydraulic cart
Pin
Cart cart
Installing jig
Switchboard cover
Supporting bolt
Receiving plate
L-bracket
Supporting bolt
Receiving plate
1−7 L2E-BW1348
1.4 Leveling and fixing of patient table
Fix the foot switch shown in 2.2.2 temporarily before starting this work, and make sure that the
foot switch can be mounted without any trouble. If the table unit is too close to the scanner side,
it is possible that the foot switch cover cannot be mounted later.
(1) Lift the table to a height of 750mm (display height 0mm) above the floor and turn off the power supply.
(2) Move the tabletop forward up to the extreme front position by turning the manual retreating handle for the
tabletop mounted at the rear of the tabletop. (See Fig.1.4.1)
(5) Put the leveling jig on the rail, and check if the tabletop is leveled properly by putting the level on the rail.
(See Fig. 1.4.2)
X-axis turning
Level Z-axis turning Level
Side cover
leveling jig
Y
X
Leveling jig
Z
Fig. 1.4.2
(6) If the tabletop is not leveled properly, put adjusting plates below the base after lifting the base by turning
the leveling screw, and then, level the patient table. (See Fig.1.4.6)
(7) After the end of horizontal leveling, turn on the patient table power supply.
1−8 L2E-BW1348
(8) Adjust the verticality of the table to the scanner unit by using the scanogram.
1) Take a picture of the tabletop in the scanogram mode.
Set the scano mode to “PA”. (See Fig.1.4.3)
Fig. 1.4.3
2) Display a scanogram image of the table top as grid display. (See Fig.1.4.4)
3) Make sure that the scanogram image of the tabletop satisfies the following requirements (a), (b) with
reference to the grid.
(a) Measure the distances (A1-A2, B1-B2) between grid center line and the right and left edges of the
tabletop and make sure that these distances are equal to each other.
(b) Make sure that the right and left edges of the tabletop are parallel to the grid.
4) Adjust the lateral position of the table until the requirements (a) (b) are satisfied, and confirm it by the
scanogram image again.
Fig.1.4.4
1−9 L2E-BW1348
(9) Turn off the table power supply.
(10) Fix the table body by driving in anchor bolts on the floor.
Anchor bolt
Dome washer
Base
Fig. 1.4.5
(11) Mount the removed switchboard cover and rear base cover to the table.
(12) Cover the installing jig screw holes on the side face of the base cover with an installing seal.
1−10 L2E-BW1348
A
Switchboard cover
Screw B
Bar
Leveling screw
Anchor
Installing jig
M10 washer
Ratchet handle
base
Anchor bolt
Washer
Adjusting plate
Leveling screw
M10 x 35L hexagon socket bolt
Anchor tongue
Screw A
a
Seal
b (attachment)
Arrow view A
Base cover (rear)
1−11 L2E-BW1348
Fig. 1.4.6
1.5 Mounting of other units
Monitor
Mouse
PC console unit
Keyboard
1−12 L2E-BW1348
1.5.2 X-ray control unit
(1) Since cooling fans are provided at the back, place the X-ray control unit more than 15cm away
from the wall.
(2) Secure a space for wiring and maintenance work at the front and right side face.
(3) After placing the X-ray control unit in position, insert the attached holders to fix the casters.
15cm or over
Work space
X-ray control unit
Work space
Caster
Caster holder
1−13 L2E-BW1348
1.6 Mounting of X-ray detector
1.6.1 In case of Pronto-XE
If the system is delivered under the standard condition, the detector is assembled into the scanner unit,
so that this work is not necessary. When the detector is delivered under its disassembled condition,
mount it referring to Fig.1.6.1.
(1) Insert the center pin into the center pin mounting plate with its notched part facing downward, and
fix it by M4 hexagon socket head screw.
(2) Push the detector inward after aligning the center hole of the detector with the center pin.
Rotary plate
Center pin
Detector
Washer
1−14 L2E-BW1348
(3) Turn the detector clockwise and counterclockwise lightly around the center pin and fix the bolts
securely at the center of the play. (Fix the screws using a spanner so as not to allow the washers
to rotate when fastening the screws.)
(4) Draw the center pin out of the preamplifier fixing bracket, insert the center pin again to the fixing
bracket and fix it with M4 hexagon socket head screw.
(5) Make sure that the rotary switch of the high voltage power supply for detector is turned off.
(6) Connect the HV coaxial cable to 1DCN on the detector PCB.
(7) Turn the rotary switch to “READY” and wait for longer than 10 seconds.
(8) Turn on the rotary switch.
(9) Press the snap switch and make sure that the voltmeter indicator indicates within the green range.
(10) If the indicator does not move in step (9), the battery pack switch is turned off inside the high
voltage power supply. Turn off the rotary switch once, and turn on the battery pack switch.
Then, perform the above steps (7) ~ (9).
For the connection between the detector and the preamplifier, refer to 2.3.2.
Snap switch
Voltmeter
Rotary switch
H.V.cable
1−15 L2E-BW1348
1.6.2 In case of Pronto-SE,VE
Perform the following work by two persons for safety.
An optimum directive position depends upon the detector. Mount the detector at an optimum
directive position by using the designated direction adjusting pin out of five directive positions. The
optimum directive position is indicated on the direction nameplate ⑫. Mount the detector, referring
to Fig.1.6.3 shown below.
⑤
⑩
②
⑧
⑦
①
⑪ ⑫ ④
⑥
Fig.1.6.3
1−16 L2E-BW1348
(1) Fix grip bar ① to mounting brackets CR ⑦ and CL ⑧, respectively.
(2) Set two guide pins ② to rotary plate ③.
(3) Push the detector ④ against rotary plate ③ by inserting the center hole of detector ④ onto the
shaft ⑤, and fix mounting brackets CR ⑦ and CL ⑧ temporarily to the rotary plate ③ by
using M8 x 35L hexagon socket bolts with plain washer ⑥ each to such an extent as the detector
④ can rotate slightly.
(4) Remove guide pins ② and grip bars ①.
(5) Fix the mounting bracket OL ⑩ temporarily to rotary plate ③ by using M8 x 35L hexagon
socket bolts with plain washer ⑥ each to such an extent as the detector ④ can rotate slightly.
(6) Screw in a direction adjusting pin ⑪ to rotary plate ③ through the hole of mounting bracket
OR ⑨ (Select the direction adjusting pin ⑪ to be used, referring to the direction nameplate
⑫).
(7) Apply the inside of the hole of mounting bracket OR ⑨ to the direction adjusting pin ⑪ by
lightly turning the detector ④ clockwise or counterclockwise around the shaft ⑤ according to
Table 1.6.1.
(8) Fasten three M8 x 35L hexagon socket head bolts being fixed temporarily in step (3) and (5) with
tightening torque of 27N・m , while keeping the angle so as not to allow the detector ④ to get out
of shaft ⑤.
(9) Remove the direction adjusting pin ⑪ from rotary plate ③, and fix detector mounting bracket
OR ⑨ with tightening torque of 27N・m by using M8 x 35L hexagon socket head bolts with
plain washer ⑥ each.
Table 1.6.1
1−17 L2E-BW1348
After mounting the detector, connect 552CN of the detector to the temperature controller, referring to
Fig.1.6.4.
Temperature
controller
Detector
552CN
Fig. 1.6.4
It is necessary for stabilizing the performance of the detector (temperature controller) and
preamplifier to apply power to these units for about 4 hours before using them. Continue
feeding power to these units normally.
For the connection between the detector and the preamplifier, refer to 2.3.2.
1−18 L2E-BW1348
1.7 Mounting of X-ray Tube and cooler
Four types of X-ray tube assemblies (1.5MHU, 2MHU, and low focus moving type 2MHU and 3.5MHU) can be
mounted to the scanner. In Pronto SE, either 2MHU or 3.5MHU X-ray tube assembly being characterized with
its small focus move is mounted. Note : Another X-ray tube can not be used in Pronto SE. X-ray tube
assemblies are classified into the type where the high-voltage bushing is mounted on the X-ray radiation port
side and the type where the high-voltage bushing is mounted on the opposite side of the radiation port. Each
type can be mounted to the scanner by just the same method.
Since the X-ray tube assembly and cooler are connected to each other with hoses, etc., be careful not to scratch
cooler hoses or cables.
Be careful not to forget to use the insulation Mylar sheets (insulation ring) when mounting
the X-ray tube assembly. If they are not applied, secondary electronic parts may be
damaged due to tube arcing. Since holders are used to fix the X-ray tube assembly
during delivery, check if the X-ray tube is correctly insulated by using a circuit tester.
1−19 L2E-BW1348
A : In case of 1.5 MHU X-ray tube assembly (for XE)
6B
6A
15B
10B
10A
⑤
15A
−
+ ⑫
③
⑰
⑨
②
① ⑬
⑦⑧
Fig. 1.7.1
1−20 L2E-BW1348
B : In case of 2 MHU X-ray tube assembly (for XE)
HITACHI ⑥
⑤
⑩ Varian
−
CS
H7
H8
+ H9
Power cord
③ ( L0, L100 (HITACHI)
⑰ or L110,H,HO,E(Varian)
②
⑫ ⑨
⑯
⑭
H,H0,L0,L100,E,ED ⑧
⑦ 7,8,9,CS,EX ⑪
④
Fig. 1.7.2
1−21 L2E-BW1348
C: Low focus shift quantity in case of 2MHU X-ray tube assembly (for SE)
⑤
⑩
+
CS
H7
− H8
H9 Power cord
(L0,L100,H,H0,E)
③
⑰
② ⑨
⑫
H,H0,L0,L100,E,ED ⑧
⑦ 7,8,9,CS,EX ⑪
④
Fig. 1.7.3
1−22 L2E-BW1348
D : In case of 3.5 MHU X-ray tube assembly (for XE)
⑩ +
⑤
CS
H7
− H8
H9
Power cord
③ (L0,L110,H,H0,E)
② ⑫ ⑨
H,H0,L0,L100,E,ED ⑧
⑬
⑪
⑦ 7,8,9,CS,EX
④
Fig. 1.7.4
1−23 L2E-BW1348
1.7.3 X-ray tube assembly mounting procedure ( 3.5MHU for Pronto SE/VE ) Tube shift mechanism type
The following mounting procedure applies to the X-ray tube assembly having a tube shift mechanism.
For details, refer to Fig.1.7.5.
(1) Rotate the rotary plate until collimator case ① reaches the highest position
(2) Fix the holder ③ to the rotary plate by screws (Bring the key way of holder ③ into contact with
the key of rotary plate)
(3) Fit the insulation ring ⑩ to X-ray tube assembly ⑤ and fix it with a Mylar tape.
(4) Mount the X-ray tube assembly ⑤ to holder ③
(5) Mount the trunnions ⑥ to holder ③ by four M8 × 50痕 hexagon socket head bolts with
tightening torque of 1960N.cm (200kgf-cm)
(6) Connect the anode and cathode high-voltage cables to the X-ray tube assembly
(7) Connect the starter cable from the X-ray tube assembly to terminal ⑦ (7, 8, 9, CS and EX) and
the cable to terminal ⑧ (ED, E, L, L100, H, H0).
(8) For the wiring and piping, refer to [2.3 Internal wiring of system]
(9) Mount pedestal ⑯ and gauge mount ⑰
(10) Mount the retaining band ⑱ to holder ③ by M8 × 20痕 hexagon socket head bolts
(11) Mount the guard fitting 1 ⑲ to the rotary plate by M16 × 35痕 hexagon socket head bolts.
(This item applies to Pronto-VE)
(1) Be careful not to forget to use the insulation Mylar sheets ( insulation ring ) when
mounting the X-ray tube assembly, otherwise secondary electronic parts may be
damaged due to discharge Since holders are mounted to fix the X-ray tube assembly
during delivery, check if the X-ray tube is correctly insulated by using a circuit tester.
(2) Parts 19 and 21 are removed at the delivery time from the Works.
Mount these parts securely
(3) Parts 19 and 21 may be mounted after center adjustment in Chapter 3.
1−23−1 L2E-BW1348
E: In case of 3.5MHU X-ray tube assembly (for SE/VE) ...... Tube shift mechanism type
⑱ 4-M8X20L
Hexagon socket head bolt
(spring, plain washer)
⑥ 2-M16X35L
Hexagon socket head bolt
a (spring, plain washer)
3-M4X10L
Hexagon socket head bolt
⑤ ⑲ (plain washer)
21
⑩ +
3-M4X10L
Hexagon socket head bolt
CS
H7 (plain washer)
− H9
H8 ⑳
Power cord
(L0,L110,H,H0,E)
③ a
⑫
⑰
⑯
⑭
H,H0,L0,L100,E,ED ⑧
⑬
⑪
⑦ 7,8,9,CS,EX
④
① Collimator case ⑨ Cooler ⑰ Gauge mount
③ Holder (base plate) ⑩ Insulation ring ⑱ Retaining band
④ Mounting plate ⑪ Oil container ⑲ Guard fitting 1 (Pronto VE)
⑤ X-ray tube assembly ⑫ Plate (A) ⑳ Guard fitting 2 (Pronto VE)
⑥ Trunnion ⑬ Plate (B) 蔵 Guard plate (Pronto VE)
⑦ Terminal ⑭ Hexagon bolt M6×14 痕, spring,
plain washer
⑧ Terminal ⑯ Pedestal
Fig. 1.7.5
1.7.4 Cooler mounting procedure
(1) Mount plate A⑫ and plate B⑬ to the bottom face of the cooler by hexagon bolts ⑭
(2) Put the cooler ⑨ on mounting plate ④, and fix plate A⑫ and plate B⑬ by M6 nuts, spring
washers, and plain washers
(3) Connect the power cord from the cooler to terminal ⑧
(4) Mount guard fitting 2 ⑳ to plate A⑫ by M4×10 痕 hexagon socket head bolts.
(Pronto VE only)
(5) Mount guard plate 蔵 to guard fittings 蔵 ⑳ by M4×10 痕 hexagon socket head bolts. (Pronto
VE only)
After fixing the oil hose and high-voltage cables with tie-lap, mount the guard plate 蔵. Make sure
that guard plate 蔵 does not touch hose or cables. If they touch each other, re-cleat the hose and
cables so that they don’t touch guard plate 蔵.
1−23−2 L2E-BW1348
1.8 Removal of protective materials for transport
Remove the vinyl bag covering the electromagnetic brake of the scan drive unit of scanner, orange rotary plate
transport bracket, and drive unit protective plate.
Fig. 1.8.1
1−24 L2E-BW1348
1.9A Internal inspection of patient table
Make sure that no oil leaks from the hydraulic system and check the oil level in the oil tank according to the
procedure at the end of wiring of patient table.
Observe the following procedure before moving the table.
(1) Remove the link supporting bracket.
1) Remove screws.
2) After lifting the table, remove fixtures A and B.
(2) Set the patient table to 900mm above the floor.
(3) Make sure that no oil leaks from the hydraulic system.
(4) Make sure that the oil level of the oil tank gauge is 60mm ±10mm.
(5) Set the patient table to 700mm above the floor and turn off the power supply.
(6) Lift the upper edge pipe of the bellows cover and fit it to the bellows holders.
(Three bellows holders are mounted below the right and left side covers, respectively.)
1−25−1 L2E-BW1348
Bellows fixture
Bellows cover
upper edge pipe
Side cover (left)
Bellows holder
Bellows fixture
Bellows holder
60 ± 10mm
Air vent plug
Link supporting bracket
Lubricant : Daphine Superhydro 22WR
Oil volume: 1.9 liters
Fixture A
Screw
Fixture B
Fig. 1.9.1
(7) Fix the front and rear parts of the upper edge pipe of the bellows cover with the bellows fixtures
(Attachments of patient table) at 2 positions respectively.
1−25−2 L2E-BW1348
1.9B Internal inspection of patient table (in case of the system with CE mark)
Make sure that no oil leaks from the hydraulic system and check the oil level in the oil tank according to the
procedure at the end of wiring of patient table.
Observe the following procedure before moving the table.
(1) Remove the link supporting bracket.
1) Remove screws.
2) After lifting the table, remove fixtures A and B.
(2) Set the patient table to 900mm above the floor.
(3) Make sure that no oil leaks from the hydraulic system.
(4) After removing the tank cover, make sure that the oil level is 60mm±10mm.
(5) Set the patient table to 700mm above the floor and turn off the power supply.
(6) Lift the upper edge pipe of the bellows cover and fit it to the bellows holders.
(Three bellows holders are mounted below the right and left side covers, respectively.)
(7) Mount the L-plate to the right and left side covers while covering the pipe at the upper fringe of the bellows
cover with the L-plate.
1−26−1 L2E-BW1348
Bellows fixture
Bellows cover
upper edge pipe
Side cover (left)
Bellows holder
Bellows fixture
Bellows holder
L-plate
U-plate
60 ± 10mm
Air vent plug
Link supporting bracket
Lubricant : Daphine Superhydro 22WR
Oil volume: 1.9 liters
Fixture A
Tank cover
Screw
Fixture B
(8) Fix the front and rear parts of the upper edge pipe of the bellows cover with the bellows fixtures
(Attachments of patient table) at 2 positions respectively.
1−26−2 L2E-BW1348
1.10 Mounting of covers
The scanner covers consist of those illustrated in Fig.1.10.1. When the system is shipped in the standard style
of packaging, the scanner unit is delivered with the top cover, both side covers, stand covers (right, left, front,
rear), rear cover, cone cover, opening covers (right, left, front, and rear), and front lower cover detached.
Top cover
Side cover
Rear cover
Stand cover
Side cover Stand cover (rear)
Opening cover (front) (Front stand cover on the opposite side)
After completion of wiring and adjustment of the scanner, mount the covers to the scanner referring to Fig.
1.10.2 and Fig. 1.10.3.
(1) Remove four thumb screws and open the lift cover.
(2) Mount the top cover with four M4 truss head screws and toothed washers. Connect the connector 250CN
of the top cooling fan. Connect the grounding wire.
(3) Insert two hooks mounted at the upper part of the side cover into the U-receiving recesses on either end of
the top cover. Fix the side covers to the support base of the scanner gantry with four M4 x 8 痕 truss
screws and toothed washers. Cover the screws with attached screw caps.
(4) Mount the front opening cover to the lift cover.
1−27 L2E-BW1348
Side cover (left) Top cover
Lift cover
1−28 L2E-BW1348
(6) Mount the cone cover to meet the scribed line of the gantry side mounting bracket in such a way as its
center is aligned with the lift cover center.
Connect the traffic light, microphone and mat switch connectors 32CN, 312CN, and 88CN.
Connect mat switch connector 88CN without fail before starting the scan, tilt, and table
operation. They are not operated unless 88CN connector is connected.
(7) Mount the front rod and cable holder assembly to the stand. The length of front rods and rear rods shall be
as shown in Fig. 1.10.4.
After mounting the covers, tilt the system (±25°) and make sure that no excessive load is
applied to the hinge of the opening cover. If an excessive load is applied, adjust the length
of the front and rear rods so as not to apply any excessive load.
(8) Connect cable 163CN of the rear cover X-ray exposure indicator lamp.
(9) Close the rear cover and fix it with truss head screws (M4 x 8 痕) at 5 places.
Cover these screws with attached screw caps. (Mount the attached rear cover truck to the rear cover in
advance. After mounting the rear cover, remove the rear cover truck and store it.)
(10) Fix the side face of the rear cover to the side cover by using four sems screws M14 x 12 痕, and cover these
screws with the attached screw caps.
(11) Close the lift cover and fix it with thumb screws.
(12) Mount the front lower cover to the lift cover.
(13) Fix the rear opening cover with truss screws (M4 x 8 痕)
Cone cover
Rear cover
Screw cap
Front rod
Rear rod
Screw cap
1−29 L2E-BW1348
(14)-A Mount the X-ray transmission window to the lift cover and cone cover. .....(Pronto XE, SE)
1) Label the convex side of the magic tape to the concave side of the magic tape of the cone cover and
lift cover. (Be careful with the laser projector window position when labeling the tape.)
(1) Mount the X-ray transmission window closely so as not to produce any clearance from
the cover.
(2) Overlap both ends of the X-ray transmission window with each other as illustrated below
so as not to allow any liquid like a vomited substance to enter the scanner interior even if
it flows down from the upper part.
2) Place the adhesive agent side of the belt to the X-ray transmission window side, and fix the belt
center to the lower part of the cone cover (lift cover) with M3 sems screws.
3) Label the belt to the X-ray transmission window by the adhesive agent on the rear face of the belt.
4) Overlap both ends of the belt with each other and fix them to the upper part of the cone cover (lift
cover) by using M3 sems screws.
∼
Belt
Belt
Table side
Table side
Arrow view A
1−30 L2E-BW1348
(14)-B Mount the X-ray transmission window to the lift cover and cone cover. .....(Pronto VE)
1) Label the convex side of the magic tape to the concave side of the magic tape of the cone cover and
lift cover. (Be careful with the laser projector window position when labeling the tape.)
(1) Mount the X-ray transmission window closely so as not to produce any clearance from
the cover.
(2) Overlap both ends of the X-ray transmission window with each other as illustrated below
so as not to allow any liquid like a vomited substance to enter the scanner interior even if
it flows down from the upper part.
2) Place the adhesive agent side of the belt to the X-ray transmission window side, and fix the belt
center to the lower part of the cone cover (lift cover) with M3 screws.
3) Label the belt to the X-ray transmission window by the adhesive agent on the rear face of the belt.
4) Adjust both ends of the belt with each other and use two press plates to fix them to the upper part of
the cone cover (lift cover) by using M3 sems screws.
Belt
A
Belt
1−30−1 38 L2E-BW1348
Section 2 WIRING
SIG 150
HV POW GND
50 100 100
SIG 100
GND
POW 100
HV 50
Unit : mm
150
2−1A L2E-BW1348
X-RAY CONTROL UNIT
3TN H7
(M4) H8
H9
Note 1) Since the starter is changed when
LU 4TN Note 1
Run #100 LV (M10) Varian 3.5MHU X-ray tube is mounted,
HP-PB LW 1TN T1 T2 WB-3CN
U G1 (M8) the external cable is connected to
V Run #105 Run #302
36P
DDK
W H7, H8, H9 of terminal TB1 (M5).
(M10)
Run #003
HP-PB SCANNER GANTRY
36P
DDK
E
(M8) Run #106 E T1 T2 WS-7CN
1TN
10 (M10)
11
12 4TN
(M4)
2−2A
U1 Run #131 1 4TN WS-17
V1 2 (M4)
3
5TN W1
(M4) V8 1SP
W8 *4 4 4TN OPERATOR CONSOLE
5 (M4) Shield clamp
DDK Run #331 DDK
6 4TN WS-9CN 50P 50P 56CN
7 (M4) DDK DDK
WS-10CN 50P Run #362 50P 57CN
E3
(M8)
Run #031 E2 OC-12
HP-PB (M8) TCPU-66CN
E GND (M6)
(M8) *1 *2 *3 1 1TN (M5)
2
PATIENT TABLE
50P
DDK
WT-2CN
R
S 1TN
E (M4) WT-12F
HP-PB
E
(M8) E (M5) Run #061
Run #161
L2E-BW1348
X-RAY CONTROL UNIT
36P
V
DDK
W cable is connected to H7,
E E
(M10) H8, H9 of terminal TB1
Run #003 (M5).
AUTO VOLTAGE REGURATOR SCANNER GANTRY
36P
DDK
Run #106 E T1 T2 WS-7CN
A. V. R 1TN
10 (M10)
11
12 4TN
(M4)
Run #050 R Run #111 US1 4TN
E S VS1 (M6) U1
HP-PB (M6) (M10) T WS1 Run #131 1 4TN WS-17
V1 2 (M4)
E
2−3A
5TN W1 3
(M8) U Run #112 US2 4TN
V VS2 (M6) (M4) 1SP
(M10) W WS2 V8
W8 *4 4 4TN OPERATOR CONSOLE
5 (M4) Shield clamp
DDK Run #331 DDK
6 4TN WS-9CN 50P 50P 56CN
7 (M4) DDK DDK
WS-10CN 50P Run #362 50P 57CN
E3
(M8)
Run #031 E2 OC-12
(M8) TCPU-66CN
GND (M6)
*1 *2 *3 1 1TN (M5)
2
PATIENT TABLE
50P
DDK
WT-2CN
R
S 1TN
E (M4) WT-12F
(ⅰ) In case of the addition of step down transformer and AVR ( Export specifications)
L2E-BW1348
X-RAY CONTROL UNIT
36P
V
DDK
W cable is connected to H7,
E E
(M10) H8, H9 of terminal TB1
Run #003 (M5).
SCANNER GANTRY
36P
DDK
Run #106 E T1 T2 WS-7CN
1TN
10 (M10)
11
12 4TN
(M4)
U1 Run #131 1 4TN WS-17
V1 2
2−4A
3 (M4)
5TN W1
(M4) V8 1SP
W8 *4 4 4TN OPERATOR CONSOLE
5 (M4) Shield clamp
DDK Run #331 DDK
6 4TN WS-9CN 50P 50P 56CN
7 (M4) DDK DDK
WS-10CN 50P Run #362 50P 57CN
E3
(M8)
Run #031 E2 OC-12
(M8) TCPU-66CN
GND (M6)
*1 *2 *3 1 1TN (M5)
2
PATIENT TABLE
50P
DDK
WT-2CN
R
S 1TN
E (M4) WT-12F
L2E-BW1348
External cables
Max. Standard Outer Connector dimensions
CABLE Destinations Internal length used Maximum Kinds of cables
permissible length diameter No. of conductors and H×W×D [mm]
RUN external
Wiring Wiring length Wiring Wiring of cable wire thickness Wiring Wiring
No. length POW. SIG. GND
sources destinations [m] [m] sources destinations [mm] sources destinations
003 HP-PB WS 15 12 0 1.0 14.0 4×22 UL1683 6AWG *
031 HP-PB WS 15 12 0 0.6 14.4 4×22 UL1683 6AWG *
061 HP-PB OC 15 10 0.3 0 14.7 4×22 UL1683 6AWG *
NOTE
100 HP-PB WB 15 10 1.5 0 13.5 41.2 ST-SB 4×2AWG
2*
NOTE
105 WB WS 15 15 0.3 1.0 13.7 35.0 ST-SB 2×2AWG
1*
106 WB WS 15 12 1.3 0.6 13.1 18.6 UL-ST-SB 3×10AWG *
107 WB WL 15 10 1.5 0.0 13.5 6.6 UL2464SB 2×20AWG *
131 WB WS 15 12 1.8 0.6 12.6 18.6 UL-ST-SB 3×10AWG *
161 WB OC 15 10 1.8 0.3 12.9 13.5 ST-SB 2×14AWG *
20 × 20×61.7
302 WB WS 15 12 2.0 0.7 12.3 11.5 UL2464SB 17P×28AWG * 61.7 ×51.5
×51.5
303 WB DS 15 10 1.5 0 13.5 6.6 UL2464SB 2×20AWG *
20 × 20×76.8
331 WS OC 15 12 0.7 0.5 13.8 11.5 UL2464SB 25P×28AWG * 76.8 ×51.5
×51.5
20 × 20×76.8
362 WS OC 15 12 0.7 0.5 13.8 11.5 UL2464SB 25P×28AWG * 76.8 ×51.5
×51.5
In case of the addition of step down transformer and AVR ( Export specifications)
050 HP-PB A.V.R 5 3 0 1.5 3.5 4×22 UL1683 6AWG *
NOTE
100 TR WB 15 10 0.5 1.5 13 41.2 ST-SB 4×2AWG
2*
110 HP-PB TR 5 3 0 0.8 4.2 41.2 ST-SB 4×2AWG *
111 WB A.V.R 5 3 1.5 0 3.5 25.7 ST 2×6AWG *
112 A.V.R WB 5 3 0 1.5 3.5 25.7 ST 2×6AWG *
In case of Varian 3.5MHU X-ray tube is mounted
106 WB WS 15 12 1.5 0.6 12.9 18.6 UL-ST-SB 3×10AWG *
NOTE 1. Cable Run No. 105 should be regarded as included in the HV system and laid in a pit
isolated from pits where other power and signal cables are laid.
NOTE 2. Change RUN No. 100 cable when adding the step-down transformer.
NOTE 3. Change RUN No.003, 031, 061 cable’s wiring from HP-PB to TR when adding the step down
transformer.
NOTE 4. Since the starter is changed when Varian 3.5MHU X-ray tube is mounted, RUN No. 106
cable is changed.
Internal cables (scanner)
Max. Standard Outer Connector dimensions
CABLE Destinations Internal length used Standard Kinds of cables
permissible length diameter No. of conductors and H×W×D [mm]
RUN external
Wiring Wiring length Wiring Wiring of cable wire thickness Wiring Wiring
No. length POW. SIG. GND
sources destinations [m] [m] sources destinations [mm] sources destinations
*1 WS WT 1.5 1.5 − − − 4×22 UL1683 6AWG *
*2 WS WT 1.5 1.5 − − − 13.5 ST-SB 2×14AWG *
7.5×68 20×76.8
*3 WS WT 5.5 5.5 − − − 11.5 UL2464SB 25P×28AWG *
×15.6 ×51.5
2−5A L2E-BW1348
14.7 × 20×76.8
*4 WS SP 6.0 6.0 − − − 6.5 UL2464SB 2×18AWG * 21 51.5
×27.6
2−6A L2E-BW1348
2.2 Wiring of cables in each unit
2.2.1 Scanner
All the cables for connection to external units should be laid via the power distribution board (chassis
7) at the lower rear part of the scanner. For the block diagram of wiring routes, refer to Page 2-2 For
the wiring routes in detail, refer to Page 2-3. Connect each cable after removing chassis 7 cover.
View A
Chassis 7
Fig. 2.2.1
2−7 L2E-BW1348
7CN 10CN 9CN
Part B
Part C
Fig. 2.2.2 View A
1 2 3 4 5 6 7 8 9 10 11 12
T2
70G 76G
T1
105
Fig. 2.2.3 Detail of part B
106 131 *4
1 2 3 4 5 6 7 8 9 10 11 12
76G
4TN
2−8 L2E-BW1348
RUN NO. 105, 106, 131 RUN NO. 302, 331, 362
Fig. 2.2.5
2−9 L2E-BW1348
2.2.2 Table wiring
All table cables are internally connected from the scanner.
(These cables are enclosed in the scanner and table without any external protrusion.)
Base Table
*3(SIG.)
*1(GND.)
*2(POW.)
WT-2CN
WT-1TN
Cable cleat
1TN
R S
Be careful not to apply any excessive force to the connectors or terminals when holding
the cables, since the cable cleat is mounted to be close to the connectors and terminals.
2−10 L2E-BW1348
A. In case of Pronto XE/SE
2.2.3 Mounting of foot switch and electric parts cover of the table.
(1) Mount the foot switch base to the power distribution board at the front of the table.
(2) Connect cables to the power distribution board.
(3) Connect the foot switch cable connector
(4) Fix cables to the foot switch base properly.
Arrange cables regularly on the stand base at the lower part of gantry as shown in the
following figure so as not to overlap or coil them.
(All cables are internal cables from the scanner)
WS-5CN
121G
Signal cable (WT-2CN)
(5) Mount the L-bracket mounted on the scanner side of the foot switch base to the stand base side.
The L bracket can move longitudinally. Position it according to the table position.
Scanner side
L bracket
2−11 L2E-BW1348
B In case of Pronto VE
2.2.3 Mounting of foot switch and electric parts cover of the table.
(1) Mount the foot switch base to the power distribution board at the front of the table.
(2) Fix the foot switch base by driving the grip anchor to the floor.
① Prepare a φ10.5×30 deep hole corresponding to the anchor hole on the floor.
② Drive the grip anchor (M6) into the φ10.5 hole
③ Fix the foot switch base to the floor by M6 x 30L hexagon socket head bolt with washer
Scanner side
8mm
Anchor
Stand base
Don’t allow the foot switch base to touch the stand base or stand cover. The clearance
between the foot switch base and the stand base is 8mm, while the clearance between
the foot switch base and the stand cover is 2mm.
Arrange cables regularly on the stand base at the lower part of gantry as shown in the
following figure so as not to overlap or coil them. (All cables are internal cables from the
scanner)
2−11−1 L2E-BW1348
M4 x 10L screw with spring and plain washer
WS-5CN
121G
Signal cable (WT-2CN)
(5) Mount the power distribution board cover to the base cover.
(6) over the screw hole on the side face of the table base cover with the seal (attachment to the table)
Seal
2−11−2 L2E-BW1348
2.2.4 Image processing unit
(1) Method of removing the image processing unit covers
1) Remove the rear cover by shifting it downward after removing two screws.
Rear cover
PC cover
2−12 L2E-BW1348
(2) Cabling work procedure
NOTE: This work differs according to the system versions. Perform the work referring
to CASE 1 and CASE 2.
CASE 1
1) Connect the external cables from the gantry unit to the image processing unit.
Connect the cables (connector #56CN and 57CN) from the rear panel of the gantry unit to
connectors 56CN and 57CN at the lower part on the side panel of the main unit as illustrated
below. Cleat these cables on the rear panel of the main unit as shown by arrow view A.
57CN
56CN
Arrow view A
1FFB
57CN
56CN
RUN# RUN#
Fix with tie lap 362 331
To gantry unit
2−13 L2E-BW1348
CASE 2
(1) Connect the external cables from the gantry unit to the image processing unit Connect the
external cables to the connectors ( 56CN, 57CN) from the rear panel of the main body
Cleat the cables according to arrow view A
Arrow view A
57CN 56CN
2TN
56CN
57CN
56CN
1FFB
To gantry unit
2−13−1 L2E-BW1348
2) Connect cables to the connectors on the upper panel of the image processing unit.
Insert these cables through the rubber clearances on the rear panel of the main unit as illustrated
below. Observe the following procedure. (The following figure shows the upper panel of the
main unit.)
① Connect cable A (Touch panel unit ~ 9CN) to 9CN of the main unit.
② Connect cable C (touch panel unit ~ 22CN) to 22CN of the main unit through core 5FC.
Front
82CN 83CN
7FC
82CN
83CN
22CN
5FC
22CN
21CN
21CN
9CN
9CN
FG4
FG4 Rear
A B D C G F
2−14 L2E-BW1348
③ Connect cable D (monitor ~ 21CN) to 21CN of the main unit.
④ Connect cable F (keyboard ~ 83CN) to 83CN of the main unit through core 7FC.
⑤ Connect cable G (mouse ~ 82CN) to 82CN of the main unit through core 7FC.
⑥ Fix cable B (shielded cable from the touch panel unit) to FG4 by using a screw.
22CN
21CN
9CN
FG4
A B D C G F
2−15 L2E-BW1348
4) Magnetic disk mounting procedure
(This work is not necessary if magnetic disk units are mounted at the delivery time.)
① Remove the cover of the image processing unit.
② Connect the cables to the magnetic disks.
③ Mount the magnetic disks in the PCB unit.
Note 1. Mount the magnetic disks on both sides of the PCB unit. Mount the 2MD
magnetic disk on the opposite side in the following figure.
Note 2. Connect the signal cable with its connector label plane facing upward in the
same way as in the power cable.
PCB unit
Magnetic disk
(1MD)
2−16 L2E-BW1348
5) Connect cables to the monitor
Connect the monitor power cable and monitor signal cable to the specified connectors on the rear
panel of the monitor as illustrated below. (The following figure shows the rear panel of the
monitor.)
Note: Connect the power cable only to the monitor whose signal cable directly
comes out of the rear panel of the monitor.
2−17 L2E-BW1348
6) Connect the power cable to the image processing unit.
NOTE: This work differs according to the system versions. Perform the work referring
to CASE 1 and CASE 2.
CASE 1
① Fix the power cable with clamp A. Fix the cable shield by the shield clamp.
② Connect the GND cable as shown in the following figure, and fix it with clamp B.
1FFB
Shield clamp
Clamp B
1TN
2TN
57CN
Clamp A
56CN
RUN#
Power cable
061 GND cable
RUN#
161
2−18 L2E-BW1348
CASE 2
① Fix the power cable by clamp fitting A. Fix the cable shield securely by the shield fitting.
② Connect the GND cable as shown in the following figure, and fix it by clamp fitting B.
1TN
Clamp fitting A
2TN
57CN
56CN
1FFB
2−18−1 L2E-BW1348
Fixing of image processing unit
After turning the front casters sideways, fix them by inserting the attached caster holders.
2−19 L2E-BW1348
2.2.5 X-ray control unit
RUN#
106 (Note 7)
RUN#
100 131
RUN#
161
107
303 RUN#
105
RUN#
131 RUN#
RUN# 100 106
RUN#
161
RUN#
105
2−20A L2E-BW1348
Starter (Note 7)
RUN#
302
100
131
161
2−21A L2E-BW1348
For cabling to the X-ray control unit, remove the front, right, and left covers, and observe the
following procedure.
Note 1) For the cables being connected from the front lower part of the X-ray control unit,
separate these cables from the front of the inverter unit and the front of the rectifier
circuit as much as possible for the purpose of preventing ingress of noises.
2) Cleat Run #100 after cleating Run #131, 161, and 106.
3) Short 2TN DI terminals if the door switch is not used, otherwise a “DOOR OPEN”
error occurs.
4) AC 100V is output to 2TN - WL terminals in the following cases.
・ AC 100V is output when No.1 of DIP switch 201SW of XCONT3 PCB is turned on
and X-rays are turned on.
・ AC 100V is output when No. 1 of DIP switch 201SW is turned off and GANTRY
POWER is turned on.
For setting the mode where the AC 100V is output when turning on X-rays,
don’t connect any fluorescent lamp between WL terminals, otherwise the
system does not function normally. Connect a lamp whose power
consumption is lower than 60W.
2−22A L2E-BW1348
7) The power supply unit and starter are changed when Varian 3.5MHU X-ray tube is
mounted. The starter is mounted at the lower part on the right side panel of the
X-ray control unit, and external cable RUN#106 is connected to starter terminal TB1
(H7,H8,H9).
Starter
Terminal TB1
(H7,H8,H9)
RUN# 106
2−22A−1 L2E-BW1348
2.2.6 Step down transformer
Run#
110
Run# Run#
Install the step down transformer and remove the front cover for cabling to the step down transformer.
2−23 L2E-BW1348
Note 1) When installing the step down transformer, open the top cover and make sure without fail that the
power supply voltage at the installation site meets the setting of the step down transformer’s primary
voltage tap. If setting of the step down transformer tap is to be changed according to the voltage at
the site, change the taps of all three phases as shown in the following example.
Tap Tap
Primary side
Secondary side
2−24 L2E-BW1348
2.3 Internal wiring of system
2.3.1 Wiring on scanner rotary plate
Connect the following cables after assembling the X-ray tube assembly and detector unit if these units
are delivered separately without being assembled to the system (This work is not necessary for
standard delivery type where these units are assembled to the system at the delivery time.)
(1) Connect the anode and cathode high voltage cables to the X-ray tube assembly. (Remove dirt
from the high voltage cable plugs by using gauze or the like and coat them with silicon compound
before connecting these cables.)
Refer to Maintenance & Inspection (Appendix S).
(2) Connect X-ray tube assembly and cooler cables to terminal boards 91TN and 92TN.
(See Fig.2.3.1)
(3) Connect the oil container ⑧ and relief valve ⑨ of the X-ray tube assembly to each other by
using vent hose ⑦, and fix the hose by hose clip ⑩. (This does not apply to 1.5MHU X-ray
tube assembly.)
(4) Fix high voltage cables ②, ③, oil hose ⑥, and vent hose ⑦ securely by using tie laps so as
not to allow them to come outside.
(5) Turn the rotary plate clockwise (as viewed from the gantry unit front) slowly by hand, and make
sure that neither high voltage cables nor vent hose touches any other part.
2−25 L2E-BW1348
Front of scanner 1 9 10 2 6
4 5 9 2
Rear of scanner
8 7 6
2−26 L2E-BW1348
1 2 11 13
Front side of scanner
12
92TN 91TN
Rear side of scanner
8 10 7 6
2−26−1 L2E-BW1348
Fig. 2.3.1B Wiring of X-ray tube assembly
2−26−2 L2E-BW1348
2.3.2 Detector cabling work
・ In case of Pronto XE (CT-WA-14A)
Connect detector cables as shown in Fig. 2.3.1.
DETECTOR
1
2
11
12
13
14
23
24
25
26
35
36
DETECTOR
CONNECTOR No.
DETECTOR CABLE
1CN
2CN
3CN
11CN
12CN
13CN
14CN
15CN
22CN
23CN
24CN
25CN
26CN
27CN
35CN
36CN
CONNECTOR No.
Connect the earth wire receptacle of each detector cable to each tab, referring to Fig.2.3.2 and
2.3.3. Connect the receptacles of continuous connectors (like in 1CN and 2CN, and 3CN and
4CN) to the same tab on the detector side (See Fig.2.3.3).
Connect the receptacles of adjacent connectors (like in1CN and 3CN, and 5CN and 7CN)to the
same tab between connectors on the preamplifier side (See Fig.2.3.3).
When replacing the earth plate during detector exchange, fasten the hexagon socket head bolts
with the torque of 1274N・cm ~ 1568N・cm (130kgf・cm ~ 160kgf・cm).
Each receptacle is provided with a handy lock device. Don’t pull the earth wire but draw out the
receptacle part. (The receptacle is unlocked automatically when drawing it out by holding it.)
1CN
Earth wire receptacle
2CN
Earth plate
4CN
3CN
2−27 L2E-BW1348
No.1 P.C.B.
No.2 P.C.B.
Tab
DETECTOR
DETECTOR
ELEMENT No. 1 12 13 20 21 32
C
C
A
B
A
B
A
B
A
B
A
B
A
B
CONNECTOR No.
DETECTOR CABLE
18CN
19CN
20CN
21CN
28CN
29CN
30CN
31CN
32CN
33CN
34CN
47CN
48CN
1CN
2CN
3CN
4CN
5CN
CONNECTOR No.
Connect the preamplifier side earth wire receptacle of each detector cable to each tab, referring to
Fig.2.3.5. Connect the receptacles of adjacent connectors (like in 1CN and 4CN, and 7CN and
10CN) to the same tab between connectors.
Each receptacle is provided with a handy lock device. Don’t pull the earth wire but draw out the
receptacle part. (The receptacle is unlocked automatically when drawing it out by holding it.)
2−28 L2E-BW1348
Tab No.1 P.C.B.
No.4 P.C.B.
No.2 P.C.B.
No.3 P.C.B.
No.5 P.C.B.
Fig. 2.3.6
(2) Connect the preamplifier cable referring to 2.3.7 Detector cable connection work
(3) Mount the preamplifier cover with due care so as not to allow the cable to be caught
2−29 L2E-BW1348
DETECTOR
DETECTOR
ELEMENT No. 1 16 17 32
C
A
B
A
B
A
B
A
B
CONNECTOR No.
DETECTOR CABLE
106CN
107CN
108CN
13CN
14CN
15CN
58CN
59CN
60CN
61CN
62CN
63CN
CONNECTOR No.
2−30 72 L2E-BW1348
Section 3 ADJUSTMENT
POWER ON
NO
Is the unit initialized?
(Screen display)
YES
Check if PCB are inserted securely and
also cables are connected securely
Check DC power supply
Image display
YES
NO Is the unit
Normal? initialized?
YES NO
Check by the
check program
OK
Fig. 3.2.1
3−1 L2E-BW1348
3.3 Operation check and adjustment of scanner unit
(1) Turning on the power supply
Open the front cover and side covers of the scanner unit to make sure that circuit breakers 21FFB, 22FFB,
31FFB are turned on (set to the upper position) and switches 21SW and 22SW are set to NORMAL (lower
position). When the front cover is opened, front cover limit switch 3LS is turned off. Turn on the limit
switch by using a tie lap or the like.
Inverter
PCB unit Ⅰ 1SW:Manual scan push-button switch
Chassis 2
Check window
MCPUC
PCB 21FFB
22FFB
TCPUB
PCB
21SW:Stationary/normal selector switch
2SW:
Manual scan push- 22SW:Brake free selector switch
button switch
3SW:Manual-normal selector switch
4SW:EMG switch
31FFB
Chassis 3
31PL 32PL
Front view of scanner Right side view of scanner
3−2 L2E-BW1348
(3) Check parameter values according to the following procedure after power supply has been applied to the
motor controller (inverter).
Change the parameter values according to the same procedure when the motor controller has been replaced
or when the values have been changed. Data n025, 026, 027, 028, and 031 (Pronto-VE includes data n024,
too.) have been finely adjusted by scan speed adjustment and scanner stationary position adjustment.
Don’t change them, except for readjustment. (Adjusted values remain held even after the power supply
has been turned off)
3−3A L2E-BW1348
* In case of inverter type CIMR-V7AA22P2
The following figure shows the names of keys and their functions on the inverter operator panel. By
turning on the power supply, FREF lamp of handy run lamps lights. By pressing DSPL key, the lamp
indications change in the order of FREF →FOUT → IOUT → MNTR → F/R→ LO/RE →
PRGM →FREF. For running the inverter, reset the lamp to the FREF lighting condition.
Indicator
Data indicator
DIGITAL
Handy run lamps OPERATOR
Lighting of the lamps of these items moves
JVOP-140 Frequency setting
one by one each time DSPL key is FREF FOUT IOUT MNTR
control knob
pressed once. Also, the upper data F/R LO/RE PRGM Changes the frequency
DATA STOP
Increases constant No/ data value ENTER RESET
Names of handy run lamps Parenthesized colors in ( ) indicate the lighting colors of lamps
3−4 L2E-BW1348
For referring to setting parameters or changing the setting, light PRGM lamp, select a parameter number
according to the following procedure, and refer to or change data. After the end, reset to FREF lighting condition
by pressing DSPL key.
Parameter No.
/data
PRGM PRGM ENTER PRGM PRGM ENTER
PRGM PRGM
NOTE: Don’t change to [LO], but keep [RE] unchanged as preset at the delivery time from the
works in remote/local selection. Change it to [RE] if the initial value is set to [LO]. If
[LO] remains set, the scanner does not rotate even if the mode is returned to FREF
mode.
3−5 L2E-BW1348
List of inverter setting parameters (For inverter type CIMR-V7AA22P2)…In case of Pronto-SE/XE
Parameter Maker’s set Adjusting
No. Parameter names Remarks
No. values values
1 Constant write 001 1 4 Settable up to 4: n001 ~ n179
2 Run command selection 003 0 1 1: Input terminals are effective
Frequency command selection 1: Frequency command 1 is
3 004 0 1
effective
4 Reverse run disable selection 006 0 1 1: Reverse run is disabled
5 Maximum output frequency 011 60.0 50.0
6 Maximum voltage 012 200.0 180.0
7 Maximum voltage output frequency 013 60.0 50.0
8 Intermediate output frequency 014 1.5 15.0
9 Intermediate output frequency voltage 015 12.0 54.0
10 Minimum output frequency 016 1.5 0.1
11 Minimum output frequency voltage 017 12.0 18.0
Acceleration/
12 018 0 1 1: 0.01s unit
deceleration time unit selection
13 Acceleration time 1 019 10.0 5.00
14 Deceleration time 1 020 10.0 5.00
15 Frequency command 1 024 6.00 0.00
16 Frequency command 2 025 0.00 45.60 *Note 2
17 Frequency command 3 026 0.00 30.40 *Note 2
18 Frequency command 4 027 0.00 22.80 *Note 2
19 Frequency command 5 028 0.00 15.20 *Note 2
20 Frequency command 6 029 0.00 11.60
21 Frequency command 7 030 0.00 0.60
22 Frequency command 8 031 0.00 0.80 *Note 3
23 Motor rated current 036 8.5 7.0
24 Input terminal 7 function selection 056 10 8 8: Multi-stage speed command 3
25 Carrier frequency 080 3 13 13:15kHz
26 DC braking current 089 50 100
Stall prevention function selection
27 092 0 1 1: Not provided
during deceleration
Stall prevention operation level during
28 093 170 200
acceleration
Stall prevention operation level during
29 094 160 200
run
30 Motor rating slip 106 2.9 2.4
31 Motor phase component resistance 107 0.800 1.175
32 Motor no-load current 110 35 52
NOTE 1) This table shows initial set values when replacing the inverter. Don’t change any parameter
values other than indicated here. Don’t change any parameter during motor rotation.
2) Parameters n025, n026, n027 and n028 are finally adjusted in case of speed check described
later.
3) Parameter n031 is finally adjusted during stop adjustment described later.
4) When the content of parameter number other than the number in the table is erroneously altered,
set the parameter No. n001 to [8] once. All parameter values return to the set values at the time
3−6A L2E-BW1348
of shipment from the factory of inverter maker.
3−7A L2E-BW1348
List of inverter setting parameters … In case of Pronto-VE
Parameter Maker’s set Adjusting
No. Parameter names Remarks
No. values values
Constant write Settable over a range from 4.n001 to
1 001 1 4
n179
2 Run command selection 003 0 1 1: Input terminals are effective
3 Frequency command selection 004 0 1 1: Frequency command 1 is effective
4 Reverse run disable selection 006 0 1 1: Reverse run is disabled
5 Maximum output frequency 011 60.0 60.0
6 Maximum voltage 012 200.0 200.0
7 Maximum voltage output frequency 013 60.0 60.0
8 Intermediate output frequency 014 1.5 15.0
9 Intermediate output frequency voltage 015 12.0 54.0
10 Minimum output frequency 016 1.5 0.1
11 Minimum output frequency voltage 017 12.0 18.0
Acceleration/
12 018 0 1 1: 0.01s unit
deceleration time unit selection
13 Acceleration time 1 019 10.0 6.00
14 Deceleration time 1 020 10.0 6.00
15 Frequency command 1 024 6.00 56.50 *Note 2
16 Frequency command 2 025 0.00 45.60 *Note 2
17 Frequency command 3 026 0.00 30.40 *Note 2
18 Frequency command 4 027 0.00 22.80 *Note 2
19 Frequency command 5 028 0.00 15.20 *Note 2
20 Frequency command 6 029 0.00 11.60
21 Frequency command 7 030 0.00 0.60
22 Frequency command 8 031 0.00 0.80 *Note 3
23 Motor rated current 036 8.5 7.0
24 Input terminal 7 function selection 056 10 8 8: Multi-stage speed command 3
25 Carrier frequency 080 3 13 13:15kHz
26 DC braking current 089 50 100
Stall prevention function selection
27 092 0 1 1: Not provided
during deceleration
Stall prevention operation level during
28 093 170 200
acceleration
Stall prevention operation level during
29 094 160 200
run
30 Motor rating slip 106 2.9 2.4
31 Motor phase component resistance 107 0.800 1.175
32 Motor no-load current 110 35 52
NOTE 1) This table shows initial set values when replacing the inverter Don’t change any parameter
values other than indicated here. Don’t change any parameter during motor rotation
2) Parameters n024, n025, n026, n027, and n028 are finally adjusted in case of speed check
described later.
3) Parameter n031 is finally adjusted during stop adjustment described later
4) When the content of parameter number other than the number in the table is erroneously altered,
set the parameter No. n001 to [8] once. All parameter values return to the set values at the time
3−6−1 L2E-BW1348
of shipment from the factory of inverter maker.
3−6−2 L2E-BW1348
(4) Select 3SW switch to the manual side
Perform manual scan operation by pressing 1SW and 2SW switches at the same time to make sure that the
rotating part does not contact any fixed part.
(5) Perform a series of scan operation at a scan speed of 0.8sec (in case of VE only) 1sec, 1.5sec, 2sec, and 3sec
in NORMAL SCAN mode, to make sure that the operation ends normally without being interrupted
halfway. If the X-ray system adjustment and check is not completed yet, perform scanning under the
turn-off condition of X-rays (by turning off 2FFB of the X-ray control unit). Perform scan once each.
(6) Check the scanner speed from the operator console after checking the manual scan operation in (4).
1) Display the basic screen for scanner check in the operation panel screen.
2) Select SCANNER CHECK from MAIN MENU. Select SCANNING TIME CHECK out of actual
check items in SCANNER CHECK being displayed in SUB MENU.
3−7A L2E-BW1348
3) By selecting SCANNING TIME CHECK in SUB MENU, SELECT appears below SUB MENU.
Check the desired check time out of 1.0sec, 1.5sec, 2.0 sec., and 3.0sec. (0.8sec, 1.0sec, 1.5sec, 2.0sec,
and 3.0sec in case of VE) After clicking a selection item. Click EXECUTE button, and the check
operation is started.
4) When check ends, the required CW scan time is displayed in STATUS part.
Make sure that respective speed values conform to the specified values (A message “WITHIN A
SPECIFICATION” is displayed).
5) If the speed does not conform to the specified values, set it to the value, whichever smaller in the
adjusting value range by adjusting the motor controller parameters according to the following table.
3−8A L2E-BW1348
In case of Pronto-SE/XE In case of Pronto-VE
(6) Adjust the parameters so that the stop position detection sensor center is closest to the striker center
when the scanner has stopped after rotating it.
If the distance between these centers is deviated from each other more than 2mm, adjust the inverter
parameter n031 every 0.05 so that the deviation between the sensor center and the striker center does
not exceed 2mm.
As n031 increases as 0.80 → 0.85 →0.90, the striker position shifts in the plus direction.
As n031 decreases as 0.80 → 0.75 →0.70, the striker position shifts in the minus direction.
If the striker does not stop at the center completely, set it to stop after being deviated toward minus
direction. If the scan rotation is started and stopped repeatedly, the motor and regenerative resistor
are heated to decrease the braking force, causing the striker position to be gradually deviated toward
the plus direction. Don’t repeat the stop operation at the sensor position more than 5 times.
If it is desired to repeat the stop operation more than 5 times, pause it for longer than 3 minutes before
restarting the operation.
Striker sensor
3−9 L2E-BW1348
3.4 Operation check of patient table and adjustment of scanner gantry tilt
3−10 L2E-BW1348
TDRV1D PCB DC power supply (5V) Table control unit
Table interior
Protective cover
Scanner side
Bellows cover
DSW1 VR7
1 2 3 4 5 6 7 8 (height display gain adjustment)
TDRV1D PCB
VR1,VR2 VR3
(DA output adjustment) (Speed offset adjustment)
3−11 L2E-BW1348
(2) Gantry tilting
(a) Set an angle gauge on the scanner gantry and set the tilt angle to 0°.
Adjust VR3 of TCPUB PCB so that the angle indicator lamp goes out (0°).
(b) Tilt the scanner gantry toward the patient table and check if the tilt angle indication becomes +25 when
the angle gauge indicates +25° and the READY lamp for tilting toward the patient table turns off.
If not, adjust VR4 of TCPUB PCB to set the angle indication to +25. After adjusting VR4, check
adjustment item (a).
For tilting the scanner gantry before mounting the lift cover, fix the gas spring to the frame
(square pipe) with Tyraps or the like, and be careful not to allow the gas spring to collide
against the scanner gantry to be broken.
(c) Check the gantry tilting in the same way down to –25° when the scanner gantry tilts in the direction
opposite to the patient table.
3−12 L2E-BW1348
(3) Longitudinal move of tabletop
Check the tabletop feed speed by SCANNER CHECK software (MOVING SPEED CHECK).
Operation procedure
1) Display the basic screen of SCANNER CHECK on the operation panel screen.
2) Select TABLE CHECK from MAIN MENU. Select MOVING SPEED CHECK out of the actual
check items in TABLE CHECK being displayed in SUB MENU.
3−13 L2E-BW1348
3) EXECUTE button becomes effective when selecting MOVING SPEED CHECK in SUB MENU.
Check operation is started by clicking EXECUTE button.
Start checking the speed from the position more than 400mm forward from the extreme
retreat position of the tabletop (in the direction opposite to the gantry).
Speed check does not function if the tabletop is set to the forward or backward limit
position.
3−14 L2E-BW1348
(4) Operate the patient table by the gantry and adjust each item by TDRV1D PCB in the patient table.
1) Make sure that table height display indicates “1” and tilt display indicates “1.5”.
(See the following table A).
2) Retreat the tabletop and adjust VR3 of TDRV1D PCB so that the table speed becomes 1.5±0.1 (mm/s).
VR3
3) Set the table height display to “13” by pressing RESET key several times.
4) Retreat the tabletop and adjust DSW1 of TDRV1D PCB so that the table speed becomes 60±3 (mm/s).
(Adjust DIP SW value so that the table speed becomes closest to 60mm/s, referring to the following
table B.)
5) Move the tabletop forward and adjust DSW1 of TDRV1D PCB so that the table speed becomes 60±3
(mm/s). (Adjust DIP SW value so that the table speed becomes closest to 60mm/s, referring to the
following table C.)
6) Move the patient table longitudinally and make sure that each speed is within the error range.
3−15 L2E-BW1348
Table C DSW1 adjustment (2)
Bit No.
Speed correction values
3 4 5
OFF OFF OFF Speed change quantity:0%
OFF OFF ON Speed change quantity:+1% (Approx.+0.6mm/s)
OFF ON OFF Speed change quantity:+2% (Approx.+1.2mm/s)
OFF ON ON Speed change quantity:+3% (Approx.+1.8mm/s)
ON OFF OFF Speed change quantity:0%
ON OFF ON Speed change quantity:−1% (Approx.−0.6mm/s)
ON ON OFF Speed change quantity:−2% (Approx.−1.2mm/s)
ON ON ON Speed change quantity:−3% (Approx.−1.8mm/s)
Speed 0 ~ 60 (mm/s) is set to 0 ~ 23 being displayed in table height indicator. However, the
values shown in Table D only are used for adjustment work.
3−16 L2E-BW1348
(5) Check and adjustment of stable stop position
1) Move the tabletop from the console side by index 10mm feeding in several times. Check by the
display each tabletop feeding is correct as the previously set value.
2) If the tabletop does not move correctly as the previously set value, adjust DIPSW (17F) on the TCPUB
PCB referring to the index correction value in the table A.
Bit No.
Index correction value
8 7
OFF OFF Predetermined value -0.1mm
OFF ON Predetermined value -0.2mm
ON OFF Predetermined value -0.3mm
ON ON Predetermined value -0.4mm
3) After the adjustment, repeat the same work in 1) to check by the display that the tabletop moves
correctly as the previously set value.
3−16−1 L2E-BW1348
3.5 Adjustment of X-ray system
(In case of Hitachi 1.5M/2M/3.5MHU or Varian 2MHU X-ray tube is mounted)
3.5.1 X-ray tube assembly check
Check the following operation.
(1) The cooler fan is rotating.
(2) The cooler pump operates and cooling oil is circulating.
Approx. 10 minutes
Boost
Run
Braking
Operate the starter once more after braking, and check the anode rotation and braking by
applying your hand to the X-ray tube housing.
Note 1. Change the adjusting value to 70+5/-0 when mounting the Varian X-ray tube.
3−17A L2E-BW1348
3.6 Adjustment of X-ray system (In case of Varian 3.5MHU X-ray tube is mounted)
3.6.1 X-ray tube assembly check
Check the following operation
(1) The cooler fan is rotating
(2) The cooler pump operates and the cooling oil is circulating
Approx. 290sec.
Boost
Run
Braking
Fig.3.6.1
3−18 L2E-BW1348
(1) Output voltage adjustment (CPU CONT PCB)
Note 1) Check the symbols of adjustment VR and RSW sufficiently with due care so as
not to mistake them. Refer to the figure (VR and RSW positions on CPU
CONT PCB).
2) Connect a TRUE RMS meter across H7 and H9 securely, and measure the
output voltage.
3) Adjust VR and RSW while turning them gradually. VR can be adjusted
continuously, but RSW changes stepwise from 0 to F.
4) The boost voltage cannot be adjusted in real time during boosting. After
adjusting VR first, check the output voltage by boosting. Boosting is
necessary about 3 times until the feedback control is stabilized. Check the
output voltage after boosting three times, accordingly.
5) The running voltage and braking voltage can be adjusted in real time,
respectively. However, since the output voltage changes stepwise, set RSW
to a position being closest to the specifications.
6) Select the correct drive frequency.
For details, refer to 3.6.2.1 and subsequent paragraphs.
3−19 L2E-BW1348
3.6.2.1 Starter CPU-CONT PCB adjusting method
CPU-CONT PCB (inside the anode rotation drive unit)
Numerical
display LED
1RSW for
display
setting
change
Low-speed
Low-speed braking boosting voltage Low-speed hold
time adjustment adjustment ON time
adjustment Low-speed boosting
time adjustment
3−20 L2E-BW1348
3.6.2.2 Control volume on PCB and dial time adjustment (Adjust VR while checking the LED display)
Set 1RSW (16 position code switch) on CPU CONT PCB to the position corresponding to the time to
be adjusted, and display the set time. Adjust respective time, referring to the following table. Set
the starter to be idle.
1RSW Contents of numerical VR to be Display of numerical
Setting methods
position display LED adjusted display LED
0 Standby condition − − −
4 Low-speed boosting time 10VR 4L☆☆☆ Set an actual adjusting value
Set it assuming that the running
5 Low-speed hold ON time 12VR 5L☆☆☆
rotation time is 40 sec.
6 Low-speed braking time 14VR 6L☆☆☆ Set an actual adjusting value
* Since the ROM of PCB is set to 40 seconds as a fixed value, the hold ON time becomes 290 seconds
(total) - 40 seconds (ROM) = 250 seconds (set value), assuming that it is desired to set the running
rotation time to 290 seconds, for example.
* ☆☆☆ becomes the adjusting value (the time to be adjusted is displayed).
* 7 ~ F positions of 1RSW are used for adjusting in the Works.
* After the adjustment has been completed, reset 1RSW to 0 position.
3−21 L2E-BW1348
3.6.2.4 CPU-CONT PCB adjustment
3−22 L2E-BW1348
F 0 1 2
3 4 5 6
C D
B
7 8 9 A
1DSW
1RSW 3RSW 5RSW
1 2 3 4 2RSW 4RSW
Fig. 3.6.4
3−23 L2E-BW1348
3.7 kV, mA waveform check
Perform the following check in the stationary scan mode without fail.
Connect probe CH1 of an oscilloscope to TP6 of CONT-D3 board in PCB unit of the scanner turn plate.
Connect the probe CH2 to TP8 of CONT-B1 board. Connect GND of the probe to GND terminal on the board
(Use an oscilloscope having the storage mode for the purpose of avoiding the X-rays exposure).
Set the oscilloscope to CHOP mode, the time range to 0.1s/div, and the voltage range to CH1: 2V/div, CH2:
1V/div. Perform the X-ray exposure by the following operation on the central operator console and check the
following waveforms.
(1) Select 120kV, 100mA, 10mm slice, 2-second stationary scan in the MAINTENANCE SCAN mode.
CH1 kV(Anode)
20kV/div
CH2 mA(Cathode)
50mA/div
3−24 L2E-BW1348
3.8 Filament heating conditions adjustment before X-ray exposure
Carry out the following adjustment in the stationary mode without fail in the same way as in kV,
mA waveform check.
(1) Connect a digital voltmeter across TP1 and GND of the filament board in the scanner turn plate. (DC volt
range)
(2) Select the conditions of 120kV, 150mA, 10mm slice thickness, and 2-sec stationary scan on the central
operator console, and make a scan. Measure the voltage across TP1 and GND during X-ray exposure.
(3) Turn off 2FFB of the X-ray control unit, make a scan under the same condition as specified in (2), and
adjust 3VR of CONT B-1 board of the scanner turn plate so that the voltage during the exposure sequence
becomes equal to the value measured in (2).
(4) Turn on 2FFB.
Variation value
Stable value
Variation value
×100≦4%
Stable value
Table 3.8.1
Pronto-SE 376
Pronto-XE 288
3−25 L2E-BW1348
3.9 X-ray tube aging
Perform the X-ray tube aging according to the following procedure.
1. In case of 1.5MHU
(1) Repeat warming up twice.
(2) Wait for 10 minutes while keeping the power supply turned on.
(3) Perform serial scan 20 times under the conditions of 130kV, 175mA, 10mm slice, and 3.0 sec. scan at
41 seconds interval without scanning.
2. In case of 2MHU
(1) Perform serial scan 20 times under the conditions of 100kV, 120mA, 10mm slice, and 3.0 sec. scan at
11 seconds interval without scanning.
(2) Wait for 50 seconds without scanning.
(3) Perform scanning 6 times under the X-ray conditions specified in (1) at 50 seconds interval without
scanning.
(4) Wait for 30 minutes without scanning while keeping the power supply turned on.
(5) Perform serial scan 4 times under the conditions of 120kV, 200mA, 10mm slice, and 2.0 sec. scan.
(6) Wait for 30 seconds without scanning.
(7) Perform serial scan 4 times under the conditions of 130kV, 200mA, 10mm slice, and 2.0 sec. scan at
11 seconds interval without scanning.
3. In case of 3.5 MHU
(1) Perform serial scan 26 times under the conditions of 100kV, 200mA, 10mm slice, and 3.0 sec. scan at
11 seconds interval without scanning.
(2) Wait for 50 seconds without scanning.
(3) Perform scanning 6 times under the X-ray conditions specified in (1) at 30 seconds interval without
scanning.
(4) Wait for 30 minutes without scanning while keeping the power supply turned on.
(5) Perform serial scan 4 times under the conditions of 120kV, 200mA, 10mm slice, and 2.0 sec. scan at
10 seconds interval without scanning.
(6) Wait for 30 seconds without scanning.
(7) Perform serial scan 4 times under the conditions of 130kV, 200mA, 10mm slice, and 2.0 sec. scan at
10 seconds interval without scanning.
3−26 L2E-BW1348
3.10 kV, mA adjustment and check
(1) Measure the anode tube voltage and cathode side neutral point current by an oscilloscope in the same way
as in 3.7.
(CONTD3 TP6・・・+kV, CONTB1 TP8・・・cathode side mA)
(2) Select 120kV, 25mA, 10mm slice, 3-sec scan to make a scan. Observe kV and mA waveforms.
(3) Since the tube current can be finely adjusted by mA (coordinates 11A) of the DIP switch on IF B-l board,
set it to be as close to 25mA as possible.
NOTE: DIP SW No. 1 ~ 4 are effective to be able to set from 0 (all OFF) to F (all ON).
The tube current increases every 2 ~ 3mA as the switch is set with an increment of 1
bit.
NOTE: kV is the peak value and mA is a mean value except for a position where a rush
current flows. The tolerance value of 25mA is ±15%.
3−27 L2E-BW1348
3.11 Adjustment of X-ray field
3.11.1A Adjustment in X-axis direction (in case of 2 or 3.5MHU X-ray tube assembly)
(i) Loosen four M8 hexagon socket head bolts ①.
(ii) Loosen nuts ②.
(iii) Adjust the X-axis direction position of X-ray tube by using bolts ○
3A ○3B .
(For this adjustment, mount a dial gauge as shown in the following figure, and check the moving
quantity of the X-ray tube. Remove the dial gauge after adjustment.)
(iv) After adjustment, tighten nuts ② to lock bolts ○ 3A ○
3B securely.
R direction
Z direction
X direction
A
Bolt 5R
(Bolt 5F on the front side)
Nut ⑥
Nut ②
Dial gauge
X direction
3.11.2 A Adjustment in Z-axis direction (in case of 2 or 3.5MHU X-ray tube assembly)
(i) Loosen four M8 hexagon socket head bolts ④, and loosen nuts ⑥.
(ii) Move the X-ray tube toward the Z-axis direction by using bolt ○
5F (on the table side of the X-ray
(a) For moving the X-ray tube toward the table, loosen bolt ○
5F and fasten bolt ○
5R .
(b) For moving the X-ray tube toward the scanner backward, loosen bolt ○
5R and then, fasten
bolt ○
5F .
(iv) Fasten four M8 hexagon socket head bolts ④ (with the tightening torque of 1960N・cm (200kgf・
cm))
3−28 L2E-BW1348
○ ○
Bolt 5F (Bolt 5R on the rear side)
and nut ⑥
Fig. 3.11.3 Positioning of 2, 3.5MHU X-ray tube assembly
3.11.1 B Adjustment in X-axis direction (in case of 1.5MHU X-ray tube assembly)
(i) Loosen four M8 hexagon socket head bolts ①.
(ii) Loosen nuts ② and adjust the X-axis direction position of X-ray tube by using M6 hexagon bolts
○
3A ○
3B . (Check the moving quantity of the X-ray tube by using a dial gauge.)
(iii) After adjustment, tighten nuts ② to lock bolts ○
3A ○
3B securely.
R direction
3−29 L2E-BW1348
3.11.3 R-direction adjustment (Common to each X-ray tube assembly).
(i) Loosen M8 hexagon socket head bolts ①.
(ii) Align the center mark line of the X-ray tube housing with the end face of the X-ray tube holder.
(iii) Fasten M8 hexagon socket head bolts ①.
Fig.3.11.5
3−30 L2E-BW1348
3.11.3A Additional item to the adjusting method in X, Z, and R directions in case of Pronto-SE and VE
For Pronto-SE and VE, perform the following work in addition to the adjusting method described later.
(See Figure 3.11.5A, 5B)
(i) For X-direction adjustment, loosen bolt ○
A once , and retighten it after adjusting the X-direction
position of the X-ray tube (Guard plate ○
B may be removed once for this work)
(ii) For R-direction adjustment, loosen ○
C bolt once, and retighten it after adjusting the R-direction
position of the X-ray tube.
Retaining band
Guard plate B A 3-M4 x 10L hexagon socket head bolt (plain washer)
Fig. 3.11.5B
3−31 L2E-BW1348
3.11.4 Normal condition of X-ray tube
When Pronto-SE (including the system having sub-second option) is provided with the focus shift
detector (FSD), certain system adjusting work must be done under the stable extended condition of the
anode shaft of the X-ray tube.
This stable extended condition of the anode shaft of X-ray tube is called normal condition in this
manual. For setting the X-ray tube to the normal condition, the X-ray tube must be cooled down
suitably after fully warming it up. For this purpose, warm up the X-ray tube according to (1A) of
3.17.4, and set the accumulated heat quantity of anode to 43%~58% by interrupting the warm up just
before starting the air calibration. This condition is called normal condition.
The normal condition must be maintained in 3.11.5 and 3.16 in SE ( with FSD).
If the normal condition is being already held, it is not necessary any more to repeat the warm-up work
from the beginning even if the “setting to normal condition” is demanded for the adjustment work.
3−31−1 L2E-BW1348
3.11.5 X-ray beam radiation adjustment (slice thickness direction)
3.11.5A In case of Pronto SE/VE (See Figure .3.11.7)
(1) Perform (4) ~ (9) when the focus shift detector (FSD) is not provided
Perform (2) ~ (9) when the focus shift detector (FSD) is provided.
(2) Stop the FSD control by turning on No. 1 of 3SW of FSD PCB
2CN
(FSD PCB)
1CN
3SW
2SW 1SW
(3) Set the X-ray tube to the NORMAL condition. (See 3.11.4)
Keep this NORMAL condition hereafter by using [stationary scan] suitably during the adjustment
of the X-ray beam radiation in the slice thickness direction.
The accumulated heat of anode increases as shown below when 100kV, 200mA,3s scan is done.
3−32−1 L2E-BW1348
(4) Set the switch 21SW at the front right of the scanner unit to the STATIONARY SCAN side so that
the scanner does not rotate (See Fig.3.11.6).
(5) Apply the beam adjusting jig ① to beam adjusting jig positioning plate ②.
The 1mm groove of beam adjusting jig ① serves as the X-ray beam center.
(6) Set the film ③ between the detector and beam adjusting jig ①.
(7) Select the “maintenance scan” from the main menu on the operation panel.
(If the maintenance scan is not displayed, display it according to 4.1.1.)
(7.1) Set the following conditions and press STANDBY button.
(7.2) START button flickers when the system has been set to be ready for measurement.
Press START button.
(7.3) After the end of stationary scan under the conditions in (7.1), reverse the film ③ position,
and make stationary scan with 2mm slice thickness (The other conditions remain
unchanged from those in (7.1)).
(7.4) After the end of stationary scan in (7.3), develop the film.
3−32−2 L2E-BW1348
(8) Measure the deviation between the narrow X-ray part and the broad X-ray part. The move
quantity of X-ray tube corresponds to about 40% of the above deviation. Move the X-ray tube
toward the rear of the gantry if the narrow X-ray part is displaced toward the front of gantry unit.
Move the X-ray tube toward the front of the gantry if the X-ray narrow part is displaced toward
the rear of gantry unit. For the method of moving the X-ray tube, observe 3.11.2A or 3.11.2B.
The X-ray tube moves about 1mm per rotation of X-ray tube moving bolt.
Repeat steps (5) ~ (8) until the deviation between the narrow X-ray part and the broad X-ray part
becomes within 0.5mm.
(9) After the end of adjustment, perform (9.1) ~ (9.3) if not equipped with Focus Shift Detector. If
equipped with Foucus Shift Detector, perform (9.3).
(9.1) Fasten the X-ray tube fixing bolts.
(9.2) Loosen the X-ray tube moving bolts and fix them with lock nuts at the position where is
kept free of contacting the X-ray tube.
(9.3) Reset switch 21SW to NORMAL (See Fig.3.11.6).
3−32−3 L2E-BW1348
②
③ ①
①
②
③
Detector Detector
Photosensitive film
Less than 0.5mm
3−33 L2E-BW1348
3.11.5B In case of Pronto-XE (See Fig.3.11.9)
(1) Set the switch 21SW at the front right of the scanner unit to the STATIONARY SCAN side so that
the scanner does not rotate (See Fig.3.11.8).
(2) Insert the film holder ② into the X-ray incident window of the detector. The wire on the film
holder serves as the X-ray beam center.
(3) Insert film ③ into film holder ②.
(4) Select the “maintenance scan” from the main menu on the operation panel.
(If the maintenance scan is not displayed, display it according to 4.1.1.)
(4.1) Set the following conditions and press STANDBY button.
(4.2) START button flickers when the system has been set to be ready for measurement.
Press START button.
(4.3) After the end of stationary scan under the conditions in (4.1), reverse the film ③ position,
and make stationary scan with 2mm slice thickness (The other conditions remain
unchanged from those in 4.1).
(4.4) After the end of stationary scan in (4.3), develop the film.
3−34 L2E-BW1348
(5) Measure the deviation between the wire and the X-ray exposed part. The move quantity of
X-ray tube corresponds to about 40% of the above deviation. Move the X-ray tube toward the
rear of the gantry if the wire shadow is displaced toward the front of gantry unit. Move the
X-ray tube toward the front of the gantry if the wire shadow is displaced toward the rear of gantry
unit. For the method of moving the X-ray tube, observe 3.11.2A or 3.11.2B. The X-ray tube
moves about 1mm per rotation of X-ray tube moving bolt.
Repeat steps (2) ~ (5) until the deviation between the wire shadow and the X-ray exposured part
center of the film becomes within 0.5mm.
3−35 L2E-BW1348
Z
1 X-ray tube assembly
0.5mm max
4
Figure as viewed from A direction
A
Insertion
Wire
3
1 Bolt
2 Film holder
2 3 Film
4 Bolt Detector
3−36 L2E-BW1348
3.12 Preamplifier check
3.12.1 Bus line check
Start the PREAMP CHECK of check program.
Select BUS-LINE CHECK of MENU, and click EXECUTE button to make sure
that no error occurs.
This check is looped until an error is detected or STOP button is clicked and the
loop count is displayed in the window. An error list is displayed in the status
part if an error is detected.
3−37 L2E-BW1348
3.12.2 Output sequence check
Check the data sequence from the detector channel to the image processing unit via the preamplifier
according to the following procedure.
(1) Execute air calibration according to the following procedure.
(1.1) Select the items in the order of “warm up (calibration)”→“calibration select mode” from
the main menu on the operation panel.
(1.2) Select 120kV, 10mm slice thickness on the operation panel and press STANDBY
button.
(1.3) After the system has been set to be ready for measurement, START button flickers.
Press it.
(2) After the end of air calibration, fix either Cu wire or a solder wire having a size of about 1mm
diameter to the right end or the left end of the tabletop using an adhesive tape, and adjust the
patient table height to obtain the maximum field of view.
(3) Select the “maintenance scan” from the main menu of the operation panel.
(3.1) Set the scan conditions on the operation panel as shown below and press STANDBY
button.
(Note) For the items indicated as [Anything], they need not to be changed.
For the item indicated as [Maximum], input a settable maximum value.
(3.2) After the system has been set to be ready for measurement, START button flickers.
Press it.
3−38 L2E-BW1348
(4) At the end of scan, start “Raw Data Analysis” of the service program to indicate raw data.
(4.1) Select items in the order of [File] - [Open]- [Raw Data] from the menu bar of “Raw Data
Analysis” (click [File] by the left button of mouse, point [Open], and click [Raw Data] by
the left button of mouse).
(4.2) Select the raw data scanned in (3) (Click the corresponding patient number and slice
number data by the left button of mouse), and click OK by the left button of mouse to
display raw data.
(4.3) Set a window width (W/W) and window level (W/L) suitable for observing the wire
trajectory (Be careful since the input is not completed unless ENTER key is pressed
when inputting them from the keyboard).
(4.4) Make sure that the wire trajectory changes smooth without any step.
3−39 L2E-BW1348
The projection range to display raw data can be changed by moving up or down the slider
on the right side of the window by means of mouse.
Slider
(4.5) For terminating the “Raw Data Analysis”, click items in the order of [File] - [Exit] (Click
[File] by the left button of mouse, and then, click [Exit] by the left button of mouse).
3−40 L2E-BW1348
3.13 Checking the influences of vibration noises on measuring data
Check the vibration noises of detector, detector cable, and preamplifier according to the following procedure.
(1) Turn off 2FFB of the X-ray control unit to turn off X-rays.
(2) Select the “Maintenance Scan” from the main menu of the operation panel.
(If the “Maintenance Scan” is not displayed, display it according to 4.1.1.)
(2.1) Set the scan conditions on the operation panel as shown below and press STANDBY button.
Set values
Items
XE/SE VE
Scans 1 1
Table index (mm) 0 0
Angle (°) 0 0
Current (mA) 175 175
Voltage (kV) 120 120
Scan time (s) 1 0.8
Thickness (mm) 2 2
Interval (s) Anything Anything
Stationary scan OFF OFF
Delay (s) Anything Anything
Start mode Anything Anything
F.O.V. (mm) Maximum Maximum
Filter 4 4
B.H.C. OFF OFF
B.G.C. OFF OFF
Variation code 3 3
Phantom correction OFF OFF
Bias 320 320
ADPP (Measure) OFF OFF
ADPP (Recon) OFF OFF
Scatter corr. OFF OFF
CT No. Adj. OFF OFF
ADNR OFF OFF
(Note) For the items indicated as [Anything], they need not to be changed. For the
item indicated as [Maximum], input a settable maximum value.
(2.2) After the system has been set to be ready for measurement, START button flickers. Press it.
(3) At the end of scan, start “Raw Data Analysis” of the service program to indicate raw data.
(3.1) Select items in the order of [File] - [Open]- [Raw Data] from the menu bar of “Raw Data Analysis”
(click [File] by the left button of mouse, point [Open], and click [Raw Data] by the left button of
mouse).
3−41 L2E-BW1348
(3.2) Select the raw data scanned in (2) (Click the corresponding patient number and slice number data by
the left button of mouse), and click OK by the left button of mouse to display raw data.
(3.3) Set a window width (W/W) to 280 and window level (W/L) to 21430 (Be careful since the input is
not completed unless ENTER key is pressed when inputting them from the keyboard).
(3.4) Make sure that the displayed raw data are free of extreme longitudinal streaks, lateral streaks, black
lines, and white lines, and execute VCK processing as described in and after (3.5). The projection
range to display raw data can be changed by moving up or down the slider on the right side of the
window by means of mouse.
Slider
3−42 L2E-BW1348
(3.5) Select items in the order of [Analysis] - [VCK] from the menu bar of Raw Data Analysis (Click
[Analysis] by the left button of mouse and then, click [VCK] by the left button of mouse).
(3.6) Set the threshold values of MAX-MIN and SD as shown below ( ENTER input is not necessary),
and left-click the CHECK key with the mouse.
MAX-MIN threshold
Model SD threshold value
value
Pronto-XE 800 160
Pronto-SE 800 160
Pronto-VE 700 140
(3.7) If no channel exceeds the threshold values being set in (3.6), a message
All channels are OK.
is displayed. If some channels exceed these threshold values, data of the channels are displayed.
If none of channels exceeds MAX-MIN threshold values, check results are acceptable (OK).
(3.8) The ordinate range of MAX-MIN graph or SD graph can be changed by clicking DFAULT ,
MAX/MIN , or KEY IN as desired using the left button of mouse.
3−43 L2E-BW1348
(3.9) For returning to the RAW DATA display after terminating VCK processing, select items in the order
of [Analysis] - [VCK] from the menu bar of Raw Data Analysis (Click [Analysis] by the left button
of mouse and then, click [VCK] by the left button of mouse).
(3.10) For terminating “Raw Data Analysis”, select items in the order of [File]-[Exit] (Click [File] by the
left button of mouse and then, click [Exit] the left button of mouse).
3−44 L2E-BW1348
3.14 Center adjustment
Adjust the center for correcting the alignment of X-ray focus, scanner rotation center, and detector center
according to the following procedure.
*)If equipped Focus Shift Detector (FSD), switch on No.1 of 3SW on the FSD PCB.
(2) After the end of air calibration, set the center adjustment rod to 20 ~ 25mm deviated from the scanner
rotation center.
(3) Select the “Maintenance Scan” from the main menu of the operation panel.
(3.1) Set the scan conditions on the operation panel as shown below and press STANDBY button.
(Note) For the items indicated as [Anything], they need not to be changed.
(3.2) After the system has been set to be ready for measurement, START button flickers. Press it.
3−45 L2E-BW1348
(4) At the end of scan, start “Raw Data Analysis” of the service program to indicate raw data.
(4.1) Select items in the order of [File] - [Open]- [Raw Data] from the menu bar of “Raw Data Analysis”
(click [File] by the left button of mouse, point [Open] and click [Raw Data] by the left button of
mouse).
(4.2) Select the raw data scanned in (3) (Click the corresponding patient number and slice number data by
the left button of mouse), and click OK by the left button of mouse to display raw data.
(4.3) Set a window width (W/W) and window level (W/L) suitable for observing the rod trajectory (Be
careful since the input is not completed unless ENTER key is pressed when inputting them from
the keyboard).
(4.4) Select items in the order of [Analysis] - [CACK] from the menu bar of “Raw Data Analysis” (Click
[Analysis] by the left button of mouse first, and then, click [CACK] by the left button of mouse).
Now, the calculation of X-ray tube move quantity is started.
3−46 L2E-BW1348
(4.5) At the end of calculation, the X-ray move quantity is displayed when the scanner is viewed from the
table side. Move the X-ray tube according to the display value (See 3.11). Perform the work
while monitoring the move quantity on the dial gauge.
For closing the X-ray tube move quantity display window, click Close by the left button of
mouse.
Fasten fixing bolts at 4 positions without fail after moving the X-ray tube.
(4.6) For terminating the “Raw Data Analysis”, click items in the order of [File] - [Exit] (Click [File] by
the left button of mouse, and then, click [Exit] by the left button of mouse).
(5) Repeat (1)~ (4) until the indicating value of the X-ray tube move quantity becomes within 0.05mm. Be
careful since the air calibration must be repeated without fail after moving the X-ray tube.
(7) If equipped Focus Shift Detector (FSD), switch off No.1 of 3SW on the FSD PCB.
3−47 L2E-BW1348
3.15 Compensator position adjustment
Adjust the compensator position according to the following procedure.
(1) After making sure that the phantom, tabletop of patient table, etc. are not put in the gantry opening to be
ready for air scan, select the “Maintenance Scan” from the main menu of the operation panel. (Display the
“Maintenance Scan” according to 4.1.1, if not displayed.)
(1.1) Set the scan conditions on the operation panel as shown below and then, press STANDBY
button.
(Note) For the items indicated as [Anything], they need not to be changed.
(1.2) After the system has been set to be ready for measurement, START button flickers. Press it.
(2) At the end of scan, start “Raw Data Analysis” of the service program to indicate raw data.
(2.1) Select items in the order of [File] - [Open]- [Raw Data] from the menu bar of “Raw Data Analysis”
(click [File] by the left button of mouse, point [Open] and click [Raw Data] by the left button of
mouse).
3−48 L2E-BW1348
(2.2) Select the raw data scanned in (1) (Click the corresponding patient number and slice number data by
the left button of mouse), and click OK by the left button of mouse to display raw data.
(2.3) Select items in the order of [Analysis] - [CLB] from the menu bar of Raw Data Analysis (Click
[Analysis] by the left button of mouse, and then, click [CLB] by the left button of mouse).
(2.4) Check the values being displayed at the lower end of the CLB window.
These values are Left, Right, Center, and Diff. in 4 kinds.
A “Diff” of less than 5% is acceptable. If “Diff” exceeds 5%, compare Left and Right values with
each other, and move the X-ray tube assembly as shown below.
If Left >Right, move the compensator rightward as viewed from the table side.
If Left < Right, move the compensator leftward as viewed from the table side.
Move the compensator according to the following procedure (See Fig.3.15).
1) Remove the collimator box cover.
2) Loosen the right and left sems screws.
3) Loosen the M4 hexagon socket head bolts in the moving direction to produce a suitable
clearance at the bolt tips.
4) Move the compensator by fastening M4 hexagon socket head bolts on the opposite side.
5) Fix the compensator by fastening sems screws.
6) Loosen the right and left M4 hexagon socket head bolts to produce a suitable clearance at the
bolt tips Then, fix the bolts with nuts.
3−49 L2E-BW1348
7) After confirming that Diff in CLB processing is less than 5%, mount the collimator box cover.
Sems screw
Nut
Fig. 3.15
(2.5) For returning to the raw data display at the end of CLB processing, select items in the order of
[Analysis] - [CLB] from the menu bar of raw data analysis in the same way as in (2.3) (Click
[Analysis] by the left button of mouse and then, click [CLB] by the left button of mouse).
(2.6) For returning to the raw data display , select items in the order of [File] - [Exit] (Click [File] by the
left button of mouse and then, click [Exit] by the left button of mouse).
(2.7) Repeat steps (1) and (2) until the criterion in (2.4) is satisfied. The Center value should be within
the range shown in Table 3.15 when the criterion in (2.4) has been satisfied.
3−50A−1 L2E-BW1348
3.16 Focus shift detector (FSD) adjustment
Perform the following adjusting method (1) ~ (19) when the focus shift detector (FSD detector and FSD PCB) is
not adjusted at all.
(a) If the system was installed under the mounted condition of the X-ray tube, this adjustment is not necessary
because FSD was adjusted at the delivery time from the Works. (Adjustment is not necessary when the
X-ray beam is within the reference value in 3.11 X-ray beam adjustment)
(b) If the X-ray tube was removed when installing the system, perform (1) ~ (3), (10) ~ (12), and (14) ~ (19)
only in 3.16 Focus shift detector (FSD) adjustment.
(Other items are not normally affected by mounting or dismounting the X-ray tube.
If adjustment cannot be done by the control knob in (10), position the FSD in (7) ~ (9)
(c) When FSD was replaced, perform (1) ~ (4) in 3.16 FSD adjustment, and finely adjust the FSD position
without turning the FSD PCB VR in (7) to fix the FSD at the position where the FSD PCB 3TP output
becomes 0±2V when1.5 seconds have passed after turning on the X-ray during 120kV, 150mA, 10mm
slice, 3-sec scan. Then, finely adjust VR and perform (10) ~ (12) and (19)
(d) When FSD PCB was replaced, the FSD is positioned correctly.
Perform (1) ~ (6) and (10) ~ (19) in 3.16 FSD adjustment
Perform all FSD adjustment to the KV condition, otherwise ring artifact occurs
(1) Make sure that No. 1 of 2SW of FSD PCB is turned on and the collimator case cover is mounted.
(2) Set the X-ray tube to the NORMAL condition (See 3.11.4) Connect the storage oscilloscope probe to (3) ~
(5) during stop.
(3) Set the X-ray tube by turning it clockwise by 270°as viewed from the top as illustrated below, and set
21SW to “STATIONARY SCAN”
270°
3−50A−2 L2E-BW1348
(4) Offset adjustment
Adjust 1VR so that the 3TP output becomes 0±10mV when X-ray is turned off (The output increases when
turning 1VR clockwise) (Use a digital voltmeter)
(5) KV selector adjustment
Adjust 5VR so that 6TP becomes 6.25±0.05V (The output increases when turning 5VR clockwise) Make
sure that 7TP is 5.5±0.2V (Use a digital voltmeter)
* Keep the tube rating to 45 ~ 65% during adjustment in (7) ~ (12)
Perform stationary scan in (6) ~ (13) and (18)
(6) Adjust 6VR so that the 4TP output becomes 1 ~ 3V in case of 10mm slice, 100kV, 150mA, 3-sec scan.
(The output decreases when turning 6VR clockwise)
(7) Turn 3VR (for 120kV) fully counterclockwise (until a click is heard to show idle turning after about 30
turns) Turn 3VR clockwise by 12 turns from the above condition, and fix the detector for FSD to the
position where 3TP output (point A) becomes 0 ~ ±2V 1.5 seconds after X-ray has been turned on in 10mm
slice, 120kV, 150mA, and 3-sec scan. (For checking the 3TP waveform, trigger the 4TP output as shown in
< An example of FSD adjustment> in 50A-8, 9 on P3) (Oscilloscope screen) X-ray ON X-ray OFF 1.5sec.
3sec. Point A. Detector for FSD Adjusting screw M4 screw Loosen.
(Oscilloscope screen)
Adjusting screw
2V
FSD
Point A
M4 screw
0V
Loosen
3−50A−3 L2E-BW1348
(8) If the 3TP output is 100KV 1.5 sec after X-ray has been turned on in case of 10mm slice, 150mA, 3-sec
scan, make sure that the 3TP output becomes higher than 1V when turning 4VR counterclockwise and
becomes lower than -1V when turning 4VR clockwise. Make sure in the same way that the 3TP output
changes from being higher than 1V to be lower than -1V when turning 2VR counterclockwise and
clockwise at 120KV.
(9) If the output does not become higher than 1V under a KV condition in (8), readjust the position of the
detector for FSD to increase the 3TP output, and perform the confirmation in (8). If the output does not
become lower than -1V under a KV condition in (8), readjust the position of the detector for FSD to
decrease the 3TP output, and perform the confirmation in (8).
3−50A−4 L2E-BW1348
(10) Adjust the following control knobs (VR) so that the 3TP 10mm
output becomes 0 ±0.1V at 100KV or 0 ±0.2V at 120 slice
2VR : 130KV
3VR : 120KV
4VR : 100KV
Reference
FSD PCB Rev
A/8
2VR Approx. 2.0V/turn
3VR Approx. 1.0V/turn
4VR Approx.0.6V/turn
0°
3−50A−5 L2E-BW1348
(13) Adjust 13VR so that the 9TP output rises with a delay of 200ms±20ms after 4TP rise in 10mm slice,
120KV, 150mA, 1-sec scan. (The output rise is delayed when turning 13VR clockwise)
(If only one probe appears, make sure that the 9TP output is 800ms±20ms during 1-sec scan.)
Output
4TP output
200ms ± 20ms
9TP output
(TTL level)
Time
(14) Turn off No. 1 of 3SW of FSD PCB (Tube shift turns on), and set 21SW to “NORMAL”.
(15) Perform 120KV, 10mm slice air calibration.
(16) Fix the beam adjusting jig to the center of the solid detector by using the fixing screw of the solid detector
cover.
PPC
3−50A−6 L2E-BW1348
(17) (In case of Pronto SE)
After turning off No.1 of 3SW ,repeat air scan 3 times by 10mm slice, 120kV, 150mA, 3-sec scan, 5 times
by 1.5 -sec scan, and 5 times by 1-sec scan.
Select a channel nearest to 1850 as “c” in the 3rd raw data of 3-sec scan, assume “d” be the 5th value of
1.5-sec scan of the channel, and also assume “e” be the 5th value of 1-sec scan of the channel.
(In case of Pronto SE sub-second)
After turning off No.1 of 3SW ,repeat air scan 3 times by 10mm slice, 120kV, 150mA, 3-sec scan, 5 times
by 1-sec scan, and 5 times by 0.8-sec scan.
Select a channel nearest to 1850 as “c” in the 3rd raw data of 3-sec scan, assume “d” be the 5th value of
1-sec scan of the channel, and also assume “e” be the 5th value of 0.8-sec scan of the channel.
(18) Calculate d-c and e-c, and perform the following processing.
① 12VR adjustment is not necessary when 20≧d-c≧−20
② In case of d-c >20
i) Adjust 12VR by turning it clockwise so that 20≧d-c≧-20 when No.7 of 3SW is turned off
ii) Adjust 12VR by turning it counterclockwise so that 20≧d-c≧−20 when No.7 of 3SW is turned
on. If 20≧d-c≧−20 is not obtained even if 12VR is turned fully counterclockwise, turn off
No.7 of 3SW , and turn 12VR counterclockwise until 20≧d-c≧−20 is obtained
③ In case of −20>d-c
i) Adjust 12VR by turning it counterclockwise so that 20≧d-c≧−20 when No.7 of 3SW is turned
off. If 20≧d-c≧−20 is not obtained even if 12VR is turned fully counterclockwise, turn on
No.7 of 3SW , and turn 12VR clockwise until 20≧d-c≧−20 is obtained
ii) Adjust 12VR by turning it clockwise so that 20≧d-c≧−20 when No.7 of 3SW is turned on
④ 10VR adjustment is not necessary when 40≧e - c≧−40
⑤ In case of e - c≧40
i) Adjust 10VR by turning it clockwise so that 40≧e-c≧−40
when No.5 of 3SW is turned off
ii) Adjust 10VR by turning it counterclockwise so that 40≧e-c≧−40 when No.5 of 3SW is turned
on. If 40≧e-c≧−40 is not obtained even if 10VR is turned fully counterclockwise, turn on No.
5 of 3SW, and turn 10VR clockwise until
40≧e-c≧−40 is obtained
⑥ In case of −40>e-c
i) Adjust 10VR by turning it counterclockwise so that 40≧e-c≧−40 when No.5 of 3SW is turned
off. If 40≧e-c≧−40 is not obtained even if 10VR is turned fully counterclockwise, turn on
No.5 of 3SW , and turn 10VR clockwise until 40≧e-c≧−40 is obtained
ii) Adjust 10VR by turning it clockwise so that 40≧e-c≧−40 when No.5 of 3SW is turned on
* When 10 and 12VR were adjusted, repeat scanning in (17) and make sure that the c~ e values
are set to be within the range where no adjustment is necessary.
(19) Make sure that No. 1 of 3SW of FSD PCB is turned off, and mount the FSD PCB cover.
3−50A−7 L2E-BW1348
<An example of FSD adjustment>
● 100kV (Adjustment is necessary)
TP3 output
Fig. 3.16.7
● 100kV OK
TP3 output
Fig. 3.16.8
3−50A−8 L2E-BW1348
● 120kV OK
TP3 output
Fig. 3.16.9
● 130kV OK
TP3 output
Fig. 3.16.10
3−50A−9 L2E-BW1348
3.17 Phantom calibration
3.17.1 Procedure
Start
X-ray tube NO
warm up
OK?
Warm up
YES
(calibration)
YES
Within 0.1mm in NO
the vertical and horizontal
directions?
YES
YES
End
Perform X-ray warm up (or air calibration) and phantom calibration measurement under the
mounted conditions of the scanner corn cover, lift cover, and X-ray transmission window for
executing calibration under the X-ray conditions during user’s use.
3−51 L2E-BW1348
3.17.2 Mounting method of phantom mounting bracket
(1) Mounting
(a) Turn the knob counterclockwise and loosen fixing plate L under the condition shown below.
Knob
Fixing plate
Fixing plate L
(b) Lift up the phantom mounting bracket and insert the phantom mounting in the
direction until the end bracket touches the tabletop end as illustrated below.
Tabletop
End bracket
3−52 L2E-BW1348
(c) Adjust the fixing plate R to the right edge of tabletop and clip the tabletop by sliding fixing
plate L in the direction. Turn the knob to fix the phantom mounting bracket.
If the bracket is not fixed properly, the bracket may possibly fall down during the
calibration resulting in the damage of scanner cover.
Knob
Fixing plate
Fixing plate L
Tabletop
Fixing plate R
3−53 L2E-BW1348
(2) Removal
Loosen the knob to make the fixing plate L free to move, and slide the fixing plate L outward.
Remove the phantom mounting bracket from the tabletop.
Knob
Fixing plate L
Tabletop
Fixing plate L
3−54 L2E-BW1348
3.17.3 X-ray detection system warming up
Table 3.17.1 shows the required time from turning on the X-ray detector power supply and
preamplifier power supply to be ready for calibration. With a longer warm-up time, better results may
be expected. When it is necessary for any work to turn off the power supply, the off time should be as
short as possible.
Table 3.17.1 Required time from turning on the power supplies to be ready for calibration
Table 3.17.2 When it has passed shorter than 2 hours after the latest X-ray exposure
The X-ray tube warm up and air calibration can be operated systematically by Warm up (calibration)
in the measurement main menu on the operation panel.
Since the air calibration is executed protocol mode during continuous operation of warm
up and air calibration, the air calibration is executed with conforming to the [combinations
of X-ray tube voltage and slice thickness] being entered in the measurement protocol. If
air calibration is necessary under the other conditions, execute it in the calibration select
mode ((2) described later).
3−55 L2E-BW1348
(1A) Continuous operation of X-ray tube warm up and air calibration (In case of Pronto SE/VE)
(1A.1) Make sure that none of tabletop, phantoms, etc. is present in the gantry aperture.
(1A.2) If the “maintenance scan” is not displayed on the measurement main menu on the operation
panel, display it according to 4.1.1. Be careful since the “maintenance warm up” cannot
be selected in (1A.4) without this display.
(1A.3) Select Warm up (calibration) in the measurement main menu on the operation panel,
and display the mode selection menu.
(1A.4) Select the “maintenance warm up” and press STANDBY button.
Air calibration before phantom calibration must be executed under the stable
condition (normal condition) of the elongation of the anode axis of X-ray tube.
“Maintenance warm up” is executed to realize the normal condition for
executing air calibration. Keep the X-ray tube rate to be within a range of 43 ~
75% after executing the maintenance warm up until all necessary phantom
calibration measurement ends.
(1A.5) When the system is ready for measurement, START button flickers. Press it. Now,
the X-ray tube warm up and air calibration in protocol mode are executed automatically.
If air calibration under the conditions other than protocol mode is necessary, perform
necessary air calibration in the “calibration select mode”.
If the lapsed time from the latest X-ray radiation is shorter than 2 hours, perform
the following procedure instead of maintenance, warm-up.
(1) Execute 120kV, 200mA, 2s, 2mm slice thickness continuous scan until the
accumulated heat of anode ≧90%
(2) Perform (3) when focus shift detector (FSD) is not provided,
If FSD is provided, continue scanning after (1) above, and execute (3)
after continuing the condition of accumulated heat of anode ≧85% for
longer than 12 minutes.
(3) Stop operating the system for about 10 minutes, and execute air
calibration after the accumulated heat of anode has become within a
range of 43 ~ 58%
After (3) above, keep the accumulated heat of anode within a range of 43 ~
75% until all necessary phantom calibration measurement ends.
3−56 L2E-BW1348
(1B) Continuous operation of X-ray tube warm up and air calibration (In case of Pronto-XE)
(1B.1) Make sure that none of tabletop, phantoms, etc. is present in the gantry aperture.
(1B.2) Select Warm up (calibration) in the measurement main menu on the operation panel,
and display the mode selection menu.
(1B.3) Select the “warm up” and press STANDBY button.
(1B.4) When the system is ready for measurement, START button flickers.
Press it. Now, the X-ray tube warm up and air calibration in protocol mode are executed
automatically. If air calibration under the conditions other than protocol mode is
necessary, perform necessary air calibration in the “calibration select mode”.
(2) Air calibration (select mode)
(2.1) Make sure that none of tabletop, phantoms, etc. is present in the gantry aperture
(2.2) Select Warm up (calibration) in the measurement main menu on the operation panel, and display
the mode selection menu. When it has passed shorter than 2 hours after the latest X-ray exposure,
execute air calibration without warm up. In this case, the conditions setting screen in the calibration
select mode is displayed.
(2.3) Select a desired tube voltage and slice thickness, and press STANDBY button.
(2.4) After the system has become ready for measurement, START button flickers. Press it. Now,
the air calibration is done under the selected conditions.
3−57 L2E-BW1348
3.17.5 Phantom position adjustment and phantom calibration
Be sure to carry out the air calibration before starting the phantom calibration.
Observe the following caution, in particular, in case of Pronto-SE/VE.
Perform the air calibration after setting the X-ray tube to the normal condition according to
3.17.4 (1A). Keep the tube rating to be within a range of 43 ~ 75% until all necessary
phantom calibration measurement ends.
(1) If Phantom calibration is not displayed in the measurement main menu on the operation panel,
display it according to 4.1.1.
(2) Set the calibration phantom shown in Table 3.17.4 at the center of the gantry aperture.
Use the beam of the light localizer (for slice planes) to adjust the phantom so that it is kept in
parallel with the X-ray beam (Make the scan starting from smaller phantom).
(3) Use maintenance scan for positioning the phantom. Set the scan conditions as shown in Table
3.16.3 and execute maintenance scan.
The indication value can be inaccuracy if the maintenance scan does not be carried out.
3−58 L2E-BW1348
Table 3.17.4 Phantom position adjustment vs FOV
Phantom F.O.V.(mm)
φ230 polyethylene 250
φ260 polyethylene 300
φ350 polyethylene 420
φ410 polyethylene (*1) 430
(*1) Applies to Pronto-SE/VE
It is possible that the latest scan image is not displayed if the phantom position program is
started just after scan.
In such a case, select Yes on the above warning window, and perform the operation in
and after (4.3).
(4.3) A patient directory window is displayed. Designate the patient number of the calibration phantom
measured in (3) by clicking it with the left button of mouse, and then, click OK button by the left
button of mouse.
(4.4) A slice directory window corresponding to the designated patient number is displayed. Designate
the slice number of the calibration phantom measured in (3) by clicking it with the left button of
mouse, and then, click OK button by the left button of mouse.
(4.5) A phantom move indicating value is displayed as shown below when clicking
Calculation by the left button of mouse after the designated image has been displayed.
3−59 L2E-BW1348
(4.6) Move the phantom by using the micro gauge according to the displayed phantom move indicating
value.
For closing the information window displaying the phantom move indicating value, click
OK button of the information window by the left button of mouse.
(4.7) Repeat steps (3) ~ (4.6) until the phantom move indicating values become less than 0.1mm in the
vertical and horizontal directions.
In case of Proton-SE/VE, calculate the phantom calibration data after measuring the
phantom of all diameters (See Table 3.17.4).
3−60 L2E-BW1348
(6) Calculate the phantom calibration data.
(6.1) Click [Start] of the task bar by the left button of mouse, point the items in the order of [Programs]→
[2. Service], and then, click [Phantom Calibration] by the left button of mouse.
(6.2) A phantom calibration calculation menu window is displayed. By clicking Calculation
with the left button of mouse, the following data list window is displayed.
The data list remarks column indicates the following contents according to the calculation date and
time, and measurement date and time.
“Impossible to Calc.” : Non-measured data
“Calculation required” : Measured, but not calculated data
“Measured” : Data whose measuring date and time are later than the calculation date
and time
“Calculated” : Data whose calculation date and time are later than the measurement
date and time
(6.3) By clicking the check box at the left of window with the left button of mouse, the phantom
calibration data calculation conditions (X-ray tube voltage, slice thickness, and also phantom
diameter in case of Pronto-XE) are selected.
A check mark is displayed on the [Sel.] column of the data conforming to the selected calculation
conditions.
(6.4) By clicking Calc. start button with the left button of mouse after the end of conditions selection,
calculation of phantom calibration data is started.
3−61 L2E-BW1348
If an error occurs halfway in the course of calculation, the following window is displayed.
For erasing the error display window, click Abort button with the left button of mouse.
(6.5) After calculation ends, make sure that “End” is displayed in [Remarks] column of data selected in
(6.3), and terminate the phantom calibration calculation by clicking Close button with the left
button of mouse.
(7.2) Click [File] in the menu bar of Raw Data Analysis by the left button of mouse, point [Open], and
then, click [Cal. Data] by the left button of mouse.
Now, the calibration data display mode is selected.
(7.3) A window for selecting the kinds of data to be displayed appears. Select “Phantom Cal. Data” by
selecting the corresponding radio button with the left button of mouse, select a tube voltage (120kV,
for example) by clicking the corresponding radio button with the left button of mouse, and then,
click
OK by the left button of mouse.
3−62 L2E-BW1348
(7.4) After the calibration data group conforming to the conditions selected in (7.3) has been displayed,
click [Analysis] in the menu bar of the calibration data display window with the left button of mouse,
and then, click [PLP] with the left button of mouse for observing data shape.
(7.5) When the PLP window has been displayed, click the data having a desired profile to be observed by
the left button of mouse. A desired data graph is displayed in PLP window. For selecting the data
to be observed, the data contents being pointed by the mouse cursor are displayed in the format of
the X-ray tube voltage, slice thickness – alpha or beta : and calibration phantom diameter in the
“Phantom Cal. Data” window as a hint. The contents of the selected data are displayed at the lower
end of the PLP window in the same format.
3−63 L2E-BW1348
(7.6) After a desired data graph has been displayed on the PLP window, click MAX/MIN key of PLP
window by the left button of mouse to set the graph to the MAX/MIN display condition.
(7.7) Make sure that the calibration data graph is free of abnormal convex or concave. If abnormal,
repeat the above procedure, starting with air calibration (3.16.4). If the abnormal condition cannot
be solved yet, check the hardware.
3−64 L2E-BW1348
(7.8) For terminating raw data analysis, click [File] of the menu bar by the left button of mouse, and then,
click [Exit] by the left button of mouse or click the × button at the right upper end of the raw data
analysis window by the left button of mouse.
3−65 L2E-BW1348
Caution Supplement of Pronto-SE/VE phantom calibration measurement
(X-ray tube rate control)
Until application soft ware V1.00, an X-ray tube rate change of [Phantom calibration measurement in progress] is not
reflected on the system monitor in real time. Accordingly, use the system monitor without executing the maintenance
scan program or phantom calibration measurement program in the following work.
(1) Perform step (2) and subsequent steps after the X-ray detection system has been warmed up (See 3.17.3).
(2) Execute X-ray tube warm up and air calibration (See 3.17.4).
(3) Adjust the calibration phantom position (See 3.17.5). Make sure that the X-ray tube rate is within a range of 50%
~ 72% by the system monitor before executing the phantom position measuring scan (maintenance scan).
Execute X-ray exposure under 120kV, 175mA, 2s conditions (same conditions as in phantom position measuring
scan) as occasion demands for the purpose of keeping the X-ray tube rate within a range of 43% ~ 75%.
Reference: The energy to be input to the X-ray tube during 120kV, 175mA, 2s scan corresponds to 3% X-ray tube
rate in 2MHU X-ray tube or 2% X-ray tube rate in 3.5MHU X-ray tube.
It takes about 3 minutes to decrease the anode accumulated heating quantity from 50% to 43% in
2MHU X-ray tube and about 4 minutes in 3.5MHU X-ray tube.
(4) Execute phantom calibration measurement (See 3.17.5). Table 1 shows the input energy for one phantom
calibration scan. By executing the phantom calibration measurement in the following sequence, it can be
measured within a range of about 50% ~ 70% of the X-ray tube rate without deviating from 43% ~ 75% range of
the X-ray tube rate.
Wait for 5 minutes after 3 scans from 50% X-ray tube rate in case of 2MHU X-ray tube
Wait for 7 minutes after 6 scans from 50% X-ray tube rate in case of 3.5 MHU X-ray tube
When the phantom calibration measurement (18 scans) is executed under all kV, all slice thickness (including 1mm
thick option) conditions about one phantom in the above sequence, it takes about 25 ~ 30 minutes in 2MHU X-ray
tube or about 15 ~ 20 minutes in 3.5MHU X-ray tube.
(5) Repeat steps (3) ~ (4) for all necessary phantoms. (In Pronto-SE/VE, the phantom calibration calculation cannot
be executed unless the measurement of all phantoms ends.)
3−66 L2E-BW1348
3.18 CT number adjustment
3.18.1 Procedure
(1) Scan all water phantoms (See table 3.18.1) by using the CT number adjustment scan (a kind of
service scans) under the combined conditions of X-ray tube voltage and the slice thickness to be
adjusted.
(2) Calculate CT number adjusting parameter by inputting the combined conditions of X-ray tube
voltage and slice thickness measured in (1) in [Calculation] of CT number adjustment calculation
program of service programs.
(3) Execute off-line image reconstruction using raw data of the water phantoms measured in CT
number adjustment scan, and check the CT number at the water phantom center (Radius of
measuring R.O.I is 75 pixels in X and Y). If the CT number is deviated from 0±4, repeat the
procedure from the CT number adjustment scan of the corresponding water phantom.
Start
N.G.
Check CT number by off-line
reconstructed images
O.K.
End
Fig. 3.18.1
Execute CT number adjustment scan under the mounted condition of scanner cone
cover, lift cover, and X-ray transmission window for the purpose of adjusting the CT
number under the X-ray conditions during user’s use.
3−67 L2E-BW1348
3.18.2 CT number adjustment scan
(1) If CT Number Adjustment is not displayed in the measurement main menu on the operation
panel, display CT Number Adjustment according to 4.1.1.
(2) Mount the water phantom shown in table 3.18.1 to the phantom fixture, and set it to the gantry
aperture center. Set the phantom to be parallel to the X-ray beam by using the projection line
(for slice plane) of the light localizer.
(3) Adjust the water phantom position.
(3.1) For adjusting the phantom position, scan the water phantom by using maintenance scan.
Set F.O.V to the value shown in Table 3.18.1. Other conditions are as specified below.
(Note) For the items indicated as [anything], they need not to be changed.
F.O.V.(mm) F.O.V.(mm)
Water phantoms
(Pronto-XE) (Pronto-SE/VE)
φ210 250 250
φ230 (*1) 300 300
φ305 420 420
φ380 (*2) 420 430
(*1) Used for 1mm slice (option) only.
(*2) Not used for 1mm slice (option).
3−68 L2E-BW1348
(3.2) Click [Start] of the task bar by the left button of mouse, point the items in the order of [Programs]→
[2. Service], and then, click [Phantom Position] by the left button of mouse.
(3.3) When the latest scan CT image, if any, is displayed, proceed to (3.6).
If the latest scan CT image is not displayed, the following warning window is displayed.
Click Yes when designating the image by patient slice directory or Abort when terminating
processing by the left button of mouse.
It is possible that the latest scan image is not displayed if the Phantom Position
program is started just after scan. In such a case, select Yes on the above
warning window, and perform the operation in and after (3.4).
(3.4) A patient directory window is displayed. Designate the patient number of the calibration phantom
measured in (3.1) by clicking it with the left button of mouse, and then, click OK switch by the
left button of mouse.
(3.5) A slice directory window corresponding to the designated patient number is displayed. Designate
the slice number of the calibration phantom measured in (3.1) by clicking it with the left key of
mouse, and then, click OK button by the left button of mouse.
(3.6) A phantom move indicating value is displayed as shown below when clicking.
Calculation by the left button of mouse after the designated image has been displayed.
(3.7) Move the phantom by using the micro gauge according to the displayed phantom move indicating
value. For closing the information window displaying the phantom move indicating value, click
OK button of the information window by the left button of mouse.
(3.8) Repeat steps (3.1) ~ (3.7) until the phantom move indicating values become less than 5 mm in the
vertical and horizontal directions.
3−69 L2E-BW1348
(4) Execute CT number adjustment scan
(4.1) Select CT Number Adjustment in the measurement main menu on the operation panel to display
the water phantom selection menu.
(4.2) Display the tube voltage and slice thickness selection menu by pressing the button of the phantom
diameter ( 210 for example) to be adjusted, select the tube voltage and slice thickness ( 120 kV,
10 mm for example) to be adjusted, and then, press STANDBY button.
(4.3) After the system has been ready for measurement, START button flickers. Press START
button. The CT number adjustment scan is executed under the selected conditions.
Measure all necessary water phantoms by CT number adjustment scan under the tube
voltage and slice thickness where the CT number adjustment is desired, and then,
execute CT number adjustment calculation.
(1) Click [Start] of the task bar by the left button of mouse, point the items in the order of [Programs]→[2.
Service], and then, click [CT No. Adjustment] by the left button of mouse.
(2) A CT number adjustment processing menu window is displayed. By clicking Calculation with the left
button of mouse, the following data list window is displayed.
The data list remarks column indicates the following contents according to the calculation date and time,
and measurement date and time.
“Calculation disable” : Non-measured data
“Calculation required” : Measured, but not calculated data
“Measured” : Data whose measuring date and time are later than the calculation date and
time
“Calculated” : Data whose calculation date and time are later than the measurement date
and time
3−70 L2E-BW1348
(3) By clicking the check box at the left of window with the left button of mouse, the CT number adjustment
parameter calculation conditions (X-ray tube voltage, slice thickness) are selected. A check mark is
displayed on the [Selection] column of the data conforming to the selected calculation conditions (See the
display example in (2)).
(4) By clicking Calc. Start button with the left button of mouse after the end of conditions selection,
calculation of CT number adjustment parameters is started.
If an error occurs halfway in the course of calculation, the following window is displayed.
For erasing the error display window, click Abort button with the left button of mouse.
(5) After calculation ends, make sure that “End” is displayed in [Remarks] column of data selected in (3), and
terminate the CT number adjustment calculation program by clicking Close button with the left button
of mouse.
(6) Check the water CT number after CT number adjustment calculation processing by using raw data of the
water phantoms measured in CT number adjustment scan.
(6.1) Execute off-line image reconstruction by using the raw data of water phantoms requiring water CT
number check (raw data measured by CT number adjustment scan in 3.18.2). Then, A CT value
adjusted water phantom image is obtained. Set off-line image reconstruction conditions as shown
below.
3−71 L2E-BW1348
(6.2) Set circular R.O.I of radius 75 pixel at the center of the off-line reconstructed water phantom image,
measure the mean value of CT number (MEAN) in R.O.I, and make sure that the mean value is
0.0±4.0. If not, repeat the procedure starting with 3.18.2 [CT number adjustment scan].
If the mean value does not become within a range of 0.0±4.0 even after repeating the procedure
starting with the CT number adjustment scan, execute air calibration under the corresponding
conditions (combinations of x-ray tube voltage and slice thickness) and execute the CT number
adjustment scan, CT number adjustment calculation, and CT number check.
(7) Execute the CT number adjustment scan, CT number adjustment calculation, and CT number check under
all necessary conditions (combinations of X-ray tube voltage and slice thickness).
3−72 L2E-BW1348
3.19 Laser marker adjustment
Fig.3.19.1 shows the mounting layout of laser markers.
Fig.3.19.4 ~ Fig.3.19.7 show the mounting method of each laser marker.
After mounting the laser marker adjustment jig, adjust the laser marker to align the laser marker beams with the
black lines of the jig.
Laser marker ③
(patient positioning)
Laser marker ①
(slice plane)
Laser marker ④
(median line)
Laser marker ②
(scan center height)
Adjustment jig
Fig.3.19.1
Black line A
Black line C
Black line B
Black line D
Fig.3.19.2
Basic application method of adjustment jig ・・・ Common to each laser marker
(i) Align the laser marker beam with the black line
on the front plate (on the plate close to the laser Black line of rear plate
3−73 L2E-BW1348
(1) Adjustment of laser marker ①.
(This laser marker projects the slice plane.)
Move the laser marker after loosening the screws shown in Fig.3.19.4, and align the laser marker beam with
black line A in Fig.3.19.2. Align the beam passing the hole with the black line of the rear plate at the same
time.
Set to
36mm Fig. 3.19.4
Bracket
(This laser marker projects the horizontal line showing the scan center height.)
Move the laser marker after loosening the screws shown in Fig.3.19.5, and align the laser marker beam with
black line B in Fig.3.19.2 Align the beam passing the hole with the black line of the rear plate at the same
time.
3−74 L2E-BW1348
Fig. 3.19.5
Laser marker
Set to
36mm Bracket
3−75 L2E-BW1348
(3) Adjustment of laser marker ③.
(This laser marker projects the beam perpendicular to the patient body axis.)
Move the laser marker after loosening the screws shown in Fig.3.19.6, and align the laser marker beam with
black line D in Fig.3.19.2. Align the beam passing the hole with the black line of the lower plate at the
same time.
Laser marker
Move the laser marker after loosening the screws shown in Fig.3.19.7, and align the laser marker beam with
black line C in Fig.3.19.2. Align the beam passing the hole with the black line of the rear plate at the same
time.
Set to
36mm Bracket
Fig.3.19.7
3−76 L2E-BW1348
3.20 Inter-phone (intercommunication) check
(1) Operation room → Scanner room
Operation room sound comes out to the scanner room when pressing TALK switch. Adjust the
sound volume by “OC-GANTRY” control knob on the rear face of touch panel.
When TALK is pressed, the sound from the scanner room is automatically attenuated.
3−77 L2E-BW1348
3.21 Hospital name entry
Enter a hospital name according to the following procedure
(1) After clicking [Start] switch, point [Programs] and [2 Service] Click [Hospital Name Entry].
The preset hospital name is displayed For changing it, input a hospital name within 26 half-sized characters
and click [OK].
(3) The entered hospital name becomes effective after starting the system again. Check the hospital name by
an image scanned after starting the system again.
(3) By setting the service scan to [ON] in (2), the service scans ([Phantom Calibration][Maintenance
Scan][CT Number Adjustment]) are
displayed in the maintenance
measurement menu on the operation
panel.
4−1 L2E-BW1348
4.1.2 Stationary scan
(1) If [Maintenance scan] is not displayed in the measurement main menu on the operation panel,
display it according to 4.1.1.
(2) Set 21SW mounted at the front right of the scanner unit to STATIONARY SCAN to set the
scanner to be stationary (See Fig.4.1.1).
21SW:stationary/normal selector switch
22SW:
X-ray Brake free selector switch
tube
PCB unit Ⅰ
Chassis 2
Detector
(3) Select maintenance scan from the main menu on the operation panel.
(4) After selecting a desired protocol on the operation panel, set the scan conditions to desired values,
and press STANDBY button. Turn [Stationary scan] item “ON”.
Table
Scans Angle Current Voltage Scan time Thickness Interval
index
Stationary Start
Delay FOV Filter B.H.C. B.G.C. Recon moment
scan mode
Variation Phantom ADPP ADPP CT No.
Bias Scatter corr. ADNR
code correction (measure) (recon) Adj.
(5) After the system has been set to be ready for measurement, START button flickers. Press it.
(6) When the stationary scan is not necessary, reset 21SW changed in (2) to NORMAL
(See Fig.4.1.1).
(7) For deleting [Maintenance scan] from the measurement main menu on the operation panel,
observe 4.1.1.
[Maintenance scan] cannot be deleted individually, but [phantom calibration], [maintenance scan],
and [CT number adjustment scan] belonging to the service scan are deleted collectively.
4−2 L2E-BW1348
4.2 Detector - preamplifier channels correspondence table
Pronto-XE
4−3 L2E-BW1348
Pronto-SE (not sub-second)
4−4 L2E-BW1348
Pronto-VE
4−4−1 L2E-BW1348
4.3 Connection of laser imager (L.I.) to the image processing unit
4.3.1 Connection of imager control cable
Connect the imager control cable being attached to 85CN on the upper panel of the image processing
unit.
Front
85CN
83CN
82CN
7
22CN
13
21CN
61CN
9CN
FG4
Rear
4−5 L2E-BW1348
4.3.2 Video signal connection
(a) Analog video signal connection.
(1) Connect analog video signals to 18CN on the rear panel of the main unit.
(2) Connect the synchronous signal (pixel clock) to 19CN, if necessary.
18CN
19CN
1FFB
Coaxial cable
(Prepared by the maker)
1TN
To L.I.
4−6 L2E-BW1348
(b) Digital video signal connection.
Connect digital video signals to 20CN on PC-IMG PCB of slot No. 10 on the side panel of main
unit.
20CN
Front
Slot No.10
PC-IMG PCB
Signal cable
(Prepared by the maker)
A
Rear
1FFB
1TN
To L.I.
Model Pronto
TECHNICAL GUIDE
- - - Maintenance & Inspection - - -
Tokyo, Japan
L3E-BW1349-3
( 1A ) 1 L3E-BW1349
Person in
Manager
charge
Pre-work Checklist
Contents
Workers (Section, name) (Person in charge)
(Remarks)
NOTE: This checklist is for checking the purpose of work, instructions, preliminary work, etc. by holding a
meeting before starting inspection work. Fully take care of safety in performing inspection work.
(2) 2 L3E-BW1349
Tables of Maintenance Inspection Items and Methods
CONTENTS
Page
Appendix
A Checking Hydraulic Drive Unit for Tilting (Check item No.101) ..............................................13
B Belt Tension Adjustment of Scan Drive Unit (Check item No.102)...........................................14
C Cleaning Sensor PCB (Check item No.104) ...............................................................................17
D Checking and Replacing Slip Ring Brush (Check item No.107) ................................................18
E-1 Interlock Specification of XE (Check item No.115)...................................................................26
E-2 Interlock Specification of SE and VE (Check item No.115).......................................................27
F Checking Chain Tension and Wire Rope Tension (Check item No.211) ....................................28
G Lubrication to Table Hydraulic Drive Unit (Check item No. 214) .............................................31
H (1) Pronto (1.5MHU) Permissible Load Range ..........................................................................32
(2) Pronto (2MHU, 3.5MHU) Permissible Load Range .............................................................33
I Checking Primary Voltage of X-ray Controller (Check item No.408)........................................34
J Checking Detector Temperature Controller (SE/VE) (Check item No. 121)..............................35
K Fastening Torque of Bolts and Nuts............................................................................................36
L Correspondence between Detector and Preamplifier Channels ..................................................38
M Checking Voltage of Scanner DC Power Supply (Check item No. 120) ....................................41
N Checking Voltage of Table DC Power Supply (Check item No. 210) ........................................43
O Checking Electromagnetic Valve of Table Hydraulic System ....................................................44
P Replacing Battery Pack of Detector High Voltage Power Supply (XE) (Check item No.122)...46
Q Replacing the Capacitor Microphone (Check item No.125).......................................................47
R Air Scan Image for Reference.....................................................................................................49
S Connecting High Voltage Cable..................................................................................................50
(3) 3 L3E-BW1349
Removing Covers in Maintenance Work
In maintenance inspection, remove or open the covers listed below. Fig. 1 shows the gantry covers.
− 1 − L3E-BW1349
⑧
⑥
④
③
− 2 − L3E-BW1349
Periodical Maintenance Inspection Work
1. Scanner
No. Check item Purpose Method Cautions Cycle Man-hours
101 Hydraulic (1) Oil leak Check if there is no oil leak at While turning on 6 months 10 min
cylinder (for the tube joints of cylinder, the power, keep
tilting drive) pump, etc. and joints of you off the
electromagnetic valves, etc. scanner to check
(2) Electro- Follow Appendix A at the end it. 12 months 10 min
magnetic of this manual to check if the
valve and gantry does not tilt.
check
valve
(3) Applying Apply spray type lubricant 12 months 10 min
lubricant (Accessory to the table) to the
hydraulic cylinder rod.
102 Belt for scan (1) Scratch Check if there is no scratch or Turn OFF the 6 months 10 min
drive crack on the belt. power switch on
(2) Belt Follow Appendix B at the end the control panel 6 months 10 min
tension of this manual to check if belt before work.
tension is 40~50N {4~5kgf}.
103 Electro-mag Operation Use 3s scan to check if rotation Do not enter into 6 months 10 min
netic brake stops with the EMERGENCY the scanner room
(Scan drive button on the control panel during the check.
unit) (Try twice.)
104 Scan (1) Dust If the position detecting photo Turn OFF the 6 months 30 min
controller sensor is covered with dust, use power switch on
a piece of cloth to clean it. the control panel
When removing the sensor before check.
PCB, see Appendix C at the
end of this manual.
(2) Operation Check signal level of the 6 months 10 min
position detecting photo
sensor.
(3) Looseness Further tighten the fixing screw
of each signal detector.
(4) Contact Check if the position detecting
photo sensor does not touch the Keep you off the 6 months 10 min
striker. turn plate to
check this.
105 Interlock Operation Check that scan operations are 6 months 5 min
switch (Lift disabled if opening the lift
cover) cover.
106 Touch Operation Check if operation stops by 6 months 5 min
switch (Lift pressing the touch switch
cover and during scanner tilt and table
cone cover) movement.
− 3 − L3E-BW1349
No. Check item Purpose Method Cautions Cycle Man-hours
107 Slip ring (1) Checking Check wearing of the brush Turn OFF all 6 months 8 hours
brush wear of and clean the slip ring (See breakers of the
brush and Appendix D at the end of this power
cleaning manual.). distribution
(2) Replacing Periodically replace the left board before 160k scans 1 hour
power and right power brushes. check.
brush
(3) Replacing Periodically replace the upper 320k scans 2 hours
signal and lower signal brushes.
brush
108 Collimator (1) Dust If there is dust on the Turn OFF the 6 months 15 min
collimator bar, use a piece of power switch on
cloth to wipe it off. the control panel
(2) Scratch Check if there is no scratch in before work.
the slit of collimator bar.
(3) Loose Further tighten the fixing
screw screws.
109 Light Projected line Check the position of projected Keep you off the 6 months 5 min
localizer line is not shifted. If any shift turn plate to
is found, adjust it. check it.
Do not turn your
eyes directly to
the laser light.
110 X-ray Dust and stain Use a piece of soft cloth to 6 months 5 min
window clean the transparent sheet.
111 Cabling (in Loose cleat Check if there is loose cable 6 months 15 min
the turn cleat or the cable does not
plate and touch the scanner cover and so
stand) on.
112 X-ray tube Loose screw Further tighten the fixing Turn OFF the 6 months 15 min
assembly screws of X-ray tube assembly power switch on
support support. the control panel
before work.
113 Chassis and (1) Dust Remove dust on the chassis 6 months 15 min
chassis (gantry and stand).
terminal (2) Loose Further tighten the fixing
screw screws of connection cable at
the terminal.
114 Tilt angle Broken wire, Check if the indicator 6 months 2 min
indicator functions increments or decrements by
0.5 according to tilt of the
scanner and if no segment
fails.
− 4 − L3E-BW1349
No. Check item Purpose Method Cautions Cycle Man-hours
115 Range of Operation Check if angle agrees with 6 months 2 min
tilting indication (See Appendix E at
the end of this manual.).
116 EMERGENCY Operation When pressing the ● Wear the 6 months 2 min
EMERGENCY button, X-ray protective
exposure and scan operation clothing to
must stop and the indicator protect you
must light. from
exposure.
● Keep the
X-ray
condition at
its lowest
position.
117 Scan time Operation Run the check program ● Be careful 6 months 15 min
Measurement “Scanning Time Check” from since the
the central operation console to scanner can
check four types (five : VE) of rotate (Do
scan time. not emit
If measured value is out of X-rays.)
specification, change Reference:
parameter values of the motor TECHNICAL
inverter. GUIDE (Check
& Service
Programs)
(Installation &
Adjustment)
118 Table control Operation Run the check program “Table Reference: 6 months 5 min
panel Check” – “LED Check” to TECHNICAL
check if LED’s light in turn. GUIDE (Check
& Service)
119 PCB Cleaning Remove dust on PCB. Turn OFF the 12 months 10 min
200V breaker of
the power
distribution board
for check.
− 5 − L3E-BW1349
No. Check item Purpose Method Cautions Cycle Man-hours
120 DC power Voltage check Use the digital voltmeter to See Appendix M. 12 months 20 min
supply check DC voltages of PCB
unit.
(Gantry)
1DPS: 5V ± 0.1V
2DPS: 5V ± 0.1V
+15V ± 0.5V
-15V ± 0.5V
3DPS: 24V ± 1.0V
4DPS: 12V ± 0.5V
51DPS: +15V ± 0.5V
52DPS: 5V ± 0.1V
53DPS: -15V ± 0.5V
54DPS: 5V ± 0.1V
101DPS: 5V ± 0.1V
301DPS: 48V ± 2.0V
103DPS: +15V ± 0.1V
-15V ± 0.1V
104DPS: 5V ± 0.1V
* Use 103DPS and 104DPS in
case of VE only.
(Preamplifier)
XE, SE
1APS: +5V±0.1V
2APS: +5V±0.1V
3APS: +5V±0.1V
VE
DC OUT #1pin: +5V±0.1V
DC OUT #10pin: -5V±0.1V
DC OUT #5pin: +5V±0.2V
DC OUT #2pin to GND
Measure with pins connected
as above.
121 Indication of Check Check if temperature controller See Appendix J. 6 months 2 min
detector reads 35ºC±1.
temperature
controller
(SE/VE)
122 High voltage Check Press the snap switch of high 6 months 2 min
power supply Recording voltage power supply to check
for detector the indicator of voltmeter is
(XE) within the green range.
Also record the voltage.
See Appendix P for procedure
to replace the battery pack.
123 Gas pressure Check Check if the indicator of 12 months 2 min
of detector pressure gauge is within the
(XE) green range.
124 Intercom Check Check if communication in the 12 months 3 min
direction of the scanner Æ
operation console is available.
125 Capacitor Periodical Periodically replace the Lift cover and 24 months 20 min
microphone replacement capacitor microphones at two cone cover
locations. See Appendix Q.
− 6A − L3E-BW1349
2. Table
No. Check item Purpose Method Cautions Cycle Man-hours
201 Tabletop Broken wire, Check if indication increments or 6 months 2 min
slide functions decrements by 0.5 according to
indicator longitudinal move of the tabletop
and if no segment fails.
202 Table Broken wire, Check if indication increments or 6 months 2 min
up/down functions decrements by 1 according to
indicator up/down move of the table and if
no segment fails.
203 Table Stroke Check if actual stroke agrees with 6 months 2 min
longitudinal indication indication.
move stroke
204 Up/down Stroke Check if actual stroke agrees with 6 months 2 min
move stroke indication indication.
205 Segment Broken wire Check if indication is correctly 6 months 5 min
indicator performed according to each
operation.
206 PAT Operation Check if the PAT button moves Check no one in 6 months 5 min
operation the table by 290mm of preset the scanner room
value. during X-ray
Check if PAT operation is exposure.
disabled during X-ray exposure or
manual indexing, scanogram and
SET indexing.
207 SET Operation Check if the table moves 6 months 5 min
indexing correctly following the stroke
operation setting using the SET switch at
the touch panel side.
208 Abnormal Abnormal Perform each operation to check 6 months 5 min
sound sound if no abnormal sound is heard.
during
operation
209 PCB Cleaning Remove dust on PCB. Turn OFF the 12 months 10 min
200V breaker in
the power
distribution board
before check.
210 DC power Voltage check Use the digital voltmeter to check See Appendix N.
supply DC voltage of the control unit
1DPS: 5V±0.1V
− 7 − L3E-BW1349
No. Check item Purpose Method Cautions Cycle Man-hours
211 Tabletop (1) Tension Check if tension of the drive ● See 6 months 10 min
longitudinal chain and wire rope is Appendix F
move drive appropriate. at the end of
chain and (2) Applying Apply the machine oil to the this manual.
wire rope machine chain.
oil
(3) Scratch Check if there is no scratch and
broken element wires on the wire
rope.
212 CAL Damage Check if there is no damage 6 months 5 min
phantom affecting functions.
mount
213 CAL Attached Check no foreign article is 6 months 5 min
phantom foreign article, attached or no damage is found at
damage the periphery of phantom.
214 Checking oil Oil level (1) When setting the table height 6 months 30 min
level of the to 900mm (indicator reads
table 150mm), oil level in the tank
hydraulic must be within 60mm±
drive and oil 10mm.
leak from (2) Check if there is no oil leak (See Appendix G
cylinder, at the cylinder and joints. at the end of this
joint, etc. manual.)
215 Hydraulic Lubrication Apply spray type lubricant 12 months 5 min
cylinder (accessory to the table) to the rod.
216 Fixing of Loose screw Check the bolts fixing tabletop For the back of 12 months 5 min
tabletop (on the surface and back of tabletop, shift the
tabletop) and further tighten them. tabletop to its
foremost
position.
− 8 − L3E-BW1349
3. Central Operation Console (Control unit and image processing unit)
No. Check item Purpose Method Cautions Cycle Man-hours
301 Operation of Checking Perform scanogram and scan to See 6 months 15 min
scan and functions and check the following operations. INSTRUCTION
image display (1) Check date and time. If MANUAL.
display any of them is incorrect,
correct it.
(2) Check both-way
communication of the
intercom and auto voice.
(3) Check if the SET switch
carries out table indexing
correctly.
(4) Check indication and tone
of X-RAY ON, and check
indication of WAIT TIME
when waiting HU.
(5) Check the EMERGENCY
button stops scan and
lights the indicator lamp.
(6) Check the keyboard and
mouse. Try patient
registration and window
change to check them.
302 Internal Cleaning (1) Clean the inside if there is Turn OFF the 6 months 20 min
check Operation any dust. power before
check (2) Clean PCB in the PCB work.
unit if there is any dust on
it. Also check if the
connector is correctly
inserted.
(3) Check if the cable is not
poorly connected. Also
check if screws on the
terminal block are not
loose.
(4) If dust is attached to each
cooling fan, clean it and
check if it rotates
smoothly.
(5) Check there is no damage
on the other electric
appliances by eyes.
303 Checking Output of error Run the service program to Take appropriate 6 months 5 min
error log log check the error in detail. action according
to contents.
304 Check Operation Use the check program to Implement in an Reference ─
program check operation as necessary. abnormal
situation.
− 9 − L3E-BW1349
4. X-ray Control Unit
No. Check item Purpose Method Cautions Cycle Man-hours
401 Inside in Cleaning Remove dust. Turn OFF all 12months 30 min
general FFB’s of the
402 Electric Damage Check any damaged parts. power 6 months 10 min
parts distribution unit
403 Relay Rough contact Replace any contact the extent in the X-ray 6 months 5 min
contact of roughness is big. control unit
404 Terminal, Looseness, Check there is no looseness before work. 6 months 15 min
connector damage and damage. Further tighten Check if the
terminal screws. power
indicator
“MAINS” in
the X-ray
control unit is
not lit.
405 Checking Operation, See [3.5] [3.6] of Turn ON the 6 months 20 min
anode Measurement TECHNICAL GUIDE power to the CPU
rotation (Installation & Adjustment) to and gantry before
drive unit check boost, running, braking check.
(Starter) voltages and boost and braking
time (Use the RMS meter to
measure voltage.)
406 DC power Voltage check Use the digital voltmeter to Turn ON the 6 months 10 min
supply check DC voltages. power to the CPU
PCB unit: and gantry before
501DPS- +5±0.1V check.
502DPS- +15±0.2V
503DPS- -15±0.2V
407 Cooling fan Cleaning, check Clean two cooling fans at the 6 months 10 min
lower part of inverter unit and
six cooling fans at the back of
cabinet and check their
rotation.
408 Checking Measurement See Appendix I to check 6 months 15 min
primary primary voltage.
voltage
409 Checking Measurement See TECHNICAL GUIDE Care should be 6 months 30 min
kV, mA (Installation & Adjustment) to taken because
check kV and mA. X-rays are
emitted.
410 Record of Record of scan Record of scan counter 6 months 5min
scan counter numbers indication
− 10 − L3E-BW1349
5. High Voltage Generator and Filament Heating Circuit (in the scanner)
No. Check item Purpose Method Cautions Cycle Man-hours
501 Oil leak Checking oil leak Check it by eyes. 6 months 15 min
502 Terminal, Looseness, Check if there is no looseness or Turn OFF all 12 months 15 min
connector damage damage. Further tighten breakers of the
terminals. power distribution
board.
503 Filament Measurement With digital voltmeter connected Turn on the power 4 months 20 min
heating between TP1 and GND of supplies of CPU and
Filament PCB, measure the gantry in advance.
voltage there. Perform this
(1) With only power on, adjust measurement in the
4VR on CONT B1 PCB for stationary scan
the voltage between TP1 and
GND to be 7±0.1V DC. mode.
504 HV plug (1) Cleaning After disconnecting the plug, Ditto 6 months 20min
and socket (2) Replacement check the appearance. Reset the
of silicon plug as before after replacing
grease silicon grease.
− 11 − L3E-BW1349
7. Overall Adjustment
No. Check item Purpose Method Cautions Cycle Man-hours
701 X-ray beam Check Use the instant film to check Use the instant 6 months 5 min
and adjust X-ray beam. film. Do not
(1) Use stationary scan. expose yourself to
(2) X-ray conditions: 100kV, X-rays.
50mA and 1.0s See TECHNICAL
GUIDE
(Installation &
Adjustment).
Shift between the center of
X-ray beam and the center of
window must be within
0.5mm.
X-ray beam X-ray beam
narrow area wide area
− 12A − L3E-BW1349
Appendix A Checking Hydraulic Drive Unit for Tilting (Check item No.101)
Follow the steps below to check the hydraulic drive unit for tilting located at the left and right stands.
(1) While setting the scanner with the TILT CENTER lit, turn OFF the scanner power.
(2) Remove the side cover of the left stand and remove the connector of power cable for hydraulic circuit (214CN).
Short the #3 and #4 terminals of connector at the cable side.
(3) Turn ON the power and stick the gummed tape to the back tilt button of the control panel to keep the ON state for
10 minutes. If the TILT CENTER keeps lit, there is no abnormality in the left side hydraulic drive unit.
(4) Turn the power OFF and return the connector to its original position.
(5) Remove the connector of power cable (215CN) for hydraulic circuit in the right stand.
(6) As well as Item (3) above, check if there is no abnormality in the right side hydraulic drive unit.
NOTE: If back tilting of the gantry is extraordinary, press the EMERGENCY button to turn the panel
power OFF. Return the connector to its original position and inform the factory of this
trouble.
Control panel
− 13 − L3E-BW1349
Appendix B Belt Tension Adjustment of Scan Drive Unit (Check item No.102)
1. Follow (1) below to check belt tension. If tension is not normal, follow 2 to work for it.
(1) Measuring tension of the scan drive timing belt
Press the spring balance against the belt halfway between the large pulley and small pulley as shown in Fig.
B.1.
Tension is appropriate if deflection of the belt is 7mm when pressing the spring balance with the force of
40~50N (4~5kgf).
(2) Measuring tension of the V-belt and timing belt (Rotary encoder)
Belt tension is appropriate if deflection of the belt δ1 and δ2 halfway between pulleys as shown in Fig. B.2
are those shown in Table B.1.
Timing belt
δ =7mm
Timing belt
Spring balance
Spring balance
40∼50N
Electromagnetic brake
Tension is normal if deflection δ is 7mm when
pressing the spring balance by 40~50N.
V-belt Press the center part so as not to twist the
timing belt.
Drive pulley
Induction motor
− 14 − L3E-BW1349
δ2
F2 Rotary encoder
F1
M4 screw
M8 nut
M12 hexagonal socket bolt
M8 hexagonal bolt
δ1 M8 hexagonal socket bolts (4 pieces)
Motor base
F (N)
Spring balance
Scale
δ (mm)
Belt
Marking line
Fig. B.2
Table B.1
− 15A − L3E-BW1349
2. Adjusting belt tension
(1) Timing belt (for scan drive)
1) Loose two M12 hexagonal socket bolts fixing the scan drive base to the frame.
2) Use the adjusting M8 hexagonal bolt to adjust tension.
3) After adjustment, fasten the M12 hexagonal socket bolt to fix the base.
4) Manually turn the pulley clockwise two or three turns to check belt tension is appropriate.
(2) V-belt
1) Loose four motor base fixing bolts (M8 hexagonal socket bolt).
2) Loose the M8 nut fixing the M8 hexagonal bolt for adjustment, and use the M8 hexagonal bolt to adjust
tension of the V-belt.
3) After adjustment, fasten the M8 nut to fix the M8 hexagonal bolt so as not to be loosened.
Fig. B.3
− 16 − L3E-BW1349
Appendix C Cleaning Sensor PCB (Check item No.104)
Fig.C.1 shows layout of SENS2A PCB viewing from the rear of scanner.
① SENS2A PCB (TUBE UPSIDE)
Rotary encoder ①
When removing SENS2A PCB to clean the photo-sensor, move the shading plate off the sensor first.
Shading plate
Mounting plate
Photo-sensor
SENS2A PCB
− 17 − L3E-BW1349
Appendix D Checking and Replacing Slip Ring Brush (Check item No.107)
1. Preliminary work
(1) Turn off all breakers of the power distribution board.
(2) Remove the following gantry covers.
Side cover (Right) Side cover (Left) Top cover
(3) Prepare a stepladder 80cm high.
(4) Remove ⑤ Ring cover (Upper left), ⑥ Ring cover (Upper right), ⑨ Ring cover (Lower left),
⑩ Ring cover (Lower right), ⑪ Mounting bracket (Upper) and ⑯ Mounting bracket (Lower).
②
③ A
⑱
A
⑰
⑬
⑭ ⑥
④
⑧
①
⑩
⑨
⑯ ⑮
Fig. D.1
− 18 − L3E-BW1349
2. Brushes to be checked and replaced
(1) Brushes to be checked and replaced are the following 8 pieces of 3 types
② Power brush block 2 pieces........... (Parts code: 1227086B)
③ Signal brush block 5 pieces........... (Parts code: 1513808D×5) Parts code for 6-piece set of
④ HF signal brush block 1 piece ............ (Parts code: 1513808D) 1513808D : 1513808C
3ch
2ch
1ch
23ch
5ch
4ch
④ HF signal brush block
(1513808D)
Brush fiber
11 pieces/bundle × 1 bundle/ch = 11 pieces/ch
23ch
20ch
Fig. D.2
− 19 − L3E-BW1349
3. Checking and replacing brushes
3.1 Power brush block
3.1.1 Removing power brush block
1) Check items
Check items of the brush fiber include ① Wear of brush fiber, ② Shape of the brush fiber and
③ Worn dust attached. If there is any abnormality, follow Table D.1 to take appropriate action.
Table D.1
Base
− 20 − L3E-BW1349
NOTE: Number of cut brush fibers not affecting brush function is 2/bundle for the
power brush and 6/channel in total. For the signal brush and HF brush,
4/bundle and 4/channel in total.
2) Action to be taken
To cut the brush fiber
a. Use the tweezers to pick up the brush fiber and raise it in the arrow direction.
b. Use the nippers to cut the base of brush fiber.
Cut
Tweezers
Fig. D.3
Cotton swab
Fig. D.4
− 21 − L3E-BW1349
3.1.3 Mounting power brush block
When mounting the power brush block again after checking and cleaning, or when mounting a new
power brush block for replacement, follow the steps below.
1) Let the power brush block directly face with the mount and put the power brush fibers to the
mount so that the brush fit with each ring.
NOTE: Take care for the brush fibers not to be bent by catching any part.
2) Use the M4×12L hexagonal socket bolt to fix the power brush block to the mount.
3) Use the M10 hexagonal bolt to fix the lead wire of power brush to the terminal block.
4) Check if the brush fibers correctly fit in the channel of slip ring.
(Check mirror and penlight are accessories to the system.)
Check and correct any of brush fibers so that it;
① Does not touch the partition between slip ring channels.
② Does not come into the adjacent channel.
③ Does not touch any of brushes of the adjacent channel.
In many cases, fiber trouble can be corrected by raise the power brush block a little from the
brush mount and fit it again.
Brush fiber
②Crossing
①Touching
③Touching
Partition
Fig. D.5
5) After temporarily fix the lead wire to the terminal block, fix the terminal block and secure the lead
wire again.
− 22 − L3E-BW1349
3.2 Signal and HF signal brush block
Remove the brush block fixing screw and slowly raise the signal brush block ③ right upward and turn
it upside down.
3.2.2 Checking and cleaning brush fibers
Upon mounting all shield covers after cleaning brushes, slowly turn the turn plate
clockwise by hands to check if the shield covers never touch any rotating parts.
− 23 − L3E-BW1349
4. Cleaning slip rings
If water, oil or any other fine particles are considerably attached to the slip ring, clean the slip ring at the same
time as checking and replacing the brush.
(1) Turn OFF all breakers of the power distribution board.
(2) Fasten three M5 screws of the scan drive brake to release braking (See Fig. D.6).
Brake
Screw
Fig. D.6
(3) Insert the cleaning pad 1 and 2 into the cleaning jig and fix it with the holding plate.
NOTE: Give attention to a slant at the end of "cleaning pad 2" (See A-View) before inserting it
into the cleaning jig.
Cleaning pad 2
Cleaning pad 1
Cleaning jig
Cleaning jig
Holding plate
A-view
Fig. D.7
− 24 − L3E-BW1349
(4) Remove the ring cover (Left B) and attach the cleaning jig to the mounting bracket.
NOTE: Press the cleaning jig so that the cleaning pad touches all slots and fix it with M3×20L
screws.
Cleaning jig
Mounting bracket
Fig. D.8
(5) Use the syringe to fully pour I.P.A. (Isopropyl alcohol) into each cleaning pad.
Turn the turn plate clockwise three turns by hands.
I.P.A
Rotating direction
of slip ring
Cleaning pad
Center frame
Fig. D.9
− 25 − L3E-BW1349
Appendix E-1 Interlock Specification of XE (Check item No.115)
250 1000
140 890
130 880
120 870
Displayed height (mm)
Tilt angle
− 25° + 25°
1000
250
750
0
380
Unit mm
− 26 − L3E-BW1349
Appendix E-2 Interlock Specification of SE and VE (Check item No.115)
250 1000
130 880
Displayed height (mm)
120 870
Tilt angle
− 25° + 25°
1000
250
750
0
380
Unit mm
− 27 − L3E-BW1349
Appendix F Checking Chain Tension and Wire Rope Tension (Check item No.211)
(3) Withdraw the upper left and right of bellows, and detach the bellows frame from the fixing bracket to lower
the bellows cover.
Detach the left and right bellows supports. (only CE marking, CCIB certificated system)
Bellows keeper
Bellows fixture
Bellows keeper
− 28 − L3E-BW1349
(4) Checking chain tension and wire rope tension
Fig. F.3 shows check position of the chain and wire rope.
Position to check wire rope tension
Spring fixing pin Movable bed
Spring
Wire rope
Chain
Fore
Tabletop
Side cover
Position to check chain tension M8 hexagonal socket bolt
Sprocket
700mm
15mm
− 29 − L3E-BW1349
(4-2) Checking wire rope tension (See Fig. F.5)
Check that the space between the inner edge of front mounting plate and edge of movable bed is
24~26mm.
L=24∼26mm
Tabletop
− 30 − L3E-BW1349
Appendix G Lubrication to Table Hydraulic Drive Unit (Check item No. 214)
Follow the steps below to check no oil leak at the hydraulic unit and the oil level of oil tank.
(1) Lower the bellows. (See Appendix F.1 (3) above)
(2) After setting the table height to 900mm above the floor surface, turn the power OFF.
(3) Check if there is no oil leak at the hydraulic unit.
(4) Check if the oil level is 60mm±10mm.
60 ± 10mm
− 31 − L3E-BW1349
Appendix H (1) Pronto (1.5MHU) Permissible Load Range
0.6sec
25mA 50mA 75mA 100mA 125mA 150mA 175mA 200mA 225mA
100kV ○ ○ ○ ○ ○ ○ ○ ○ ○
Overall slice
120kV ○ ○ ○ ○ ○ ○ ○ ○ ×
thickness
130kV ○ ○ ○ ○ ○ ○ ○ × ×
1.0sec
25mA 50mA 75mA 100mA 125mA 150mA 175mA 200mA 225mA
100kV ○ ○ ○ ○ ○ ○ ○ ○ ○
Overall slice
120kV ○ ○ ○ ○ ○ ○ ○ ○ ×
thickness
130kV ○ ○ ○ ○ ○ ○ ○ × ×
1.5sec
25mA 50mA 75mA 100mA 125mA 150mA 175mA 200mA 225mA
100kV ○ ○ ○ ○ ○ ○ ○ ○ ○
Overall slice
120kV ○ ○ ○ ○ ○ ○ ○ ○ ×
thickness
130kV ○ ○ ○ ○ ○ ○ ○ × ×
2.0sec
25mA 50mA 75mA 100mA 125mA 150mA 175mA 200mA 225mA
100kV ○ ○ ○ ○ ○ ○ ○ ○ ○
Overall slice
120kV ○ ○ ○ ○ ○ ○ ○ ○ ×
thickness
130kV ○ ○ ○ ○ ○ ○ ○ × ×
3.0sec
25mA 50mA 75mA 100mA 125mA 150mA 175mA 200mA 225mA
100kV ○ ○ ○ ○ ○ ○ ○ ○ ○
Overall slice
120kV ○ ○ ○ ○ ○ ○ ○ ○ ×
thickness
130kV ○ ○ ○ ○ ○ ○ ○ × ×
− 32 − L3E-BW1349
Appendix H (2) Pronto (2MHU, 3.5MHU) Permissible Load Range
0.5、0.6、0.8sec
25mA 50mA 75mA 100mA 125mA 150mA 175mA 200mA 225mA
100kV ○ ○ ○ ○ ○ ○ ○ ○ ○
Overall slice
120kV ○ ○ ○ ○ ○ ○ ○ ○ ○
thickness
130kV ○ ○ ○ ○ ○ ○ ○ ○ ×
1.0sec
25mA 50mA 75mA 100mA 125mA 150mA 175mA 200mA 225mA
100kV ○ ○ ○ ○ ○ ○ ○ ○ ○
Overall slice
120kV ○ ○ ○ ○ ○ ○ ○ ○ ○
thickness
130kV ○ ○ ○ ○ ○ ○ ○ ○ ×
1.5sec
25mA 50mA 75mA 100mA 125mA 150mA 175mA 200mA 225mA
100kV ○ ○ ○ ○ ○ ○ ○ ○ ○
Overall slice
120kV ○ ○ ○ ○ ○ ○ ○ ○ ○
thickness
130kV ○ ○ ○ ○ ○ ○ ○ ○ ×
2.0sec
25mA 50mA 75mA 100mA 125mA 150mA 175mA 200mA 225mA
100kV ○ ○ ○ ○ ○ ○ ○ ○ ○
Overall slice
120kV ○ ○ ○ ○ ○ ○ ○ ○ ○
thickness
130kV ○ ○ ○ ○ ○ ○ ○ ○ ×
3.0sec
25mA 50mA 75mA 100mA 125mA 150mA 175mA 200mA 225mA
100kV ○ ○ ○ ○ ○ ○ ○ ○ ○
Overall slice
120kV ○ ○ ○ ○ ○ ○ ○ ○ ○
thickness
130kV ○ ○ ○ ○ ○ ○ ○ ○ ×
− 33 − L3E-BW1349
Appendix I Checking Primary Voltage of X-ray Controller (Check item No.408)
(1) Remove the front cover of X-ray controller and connect the digital voltmeter between the plus (indicated by the
label + ) and minus (indicated by the label − ) terminal of 601TN on the lower CAP chassis to measure DC
voltage (See Fig. I.1)
(2) Turn the CPU and GANTRY POWER switches ON.
(3) Run the check program RECTIFIER CHECK for X-ray controller from the operation console (See TECHNICAL
GUIDE (Check & Service Program).
(4) Check if the DC voltage measured is 200V±1Vdc. If the DC voltage measured differs from the reference
voltage, use 1VR of MAIN IF PCB on the REC chassis to adjust DC voltage to the reference value.
Capacitor chassis
Digital voltmeter
Fig. I.1
− 34 − L3E-BW1349
Appendix J Checking Detector Temperature Controller (SE/VE) (Check item No. 121)
Follow the flow chart shown below to check the detector temperature controller.
NO
Set temperature of
NO (A) Follow temperature setting procedure
detector temperature controllers are 35ºC.
(*1) to set temperature to 35ºC
YES
Temperature display
Set key (UP)
Display mode
(B) Check broken cable in the detector
Temperature display
℃
changeover key
35
(*1) Checking set temperature SP AL
− 35 − L3E-BW1349
Appendix K Fastening Torque of Bolts and Nuts
Follow Table K.1 and K.2 for fastening torque of bolts, screws, nuts and double nuts in working for this manual.
− 36 − L3E-BW1349
Table K.2 Fastening Torque of Nuts*
Normal nut
Thin nut
− 37 − L3E-BW1349
Appendix L Correspondence between Detector and Preamplifier Channels
ProntoXE
− 38 − L3E-BW1349
ProntoSE
Detector – Preamplifier corresponding table
− 39 − L3E-BW1349
ProntoVE
PCB. No. CHANNEL PCB CN DET CN BLOCK PCB. No. CHANNEL PCB CN DET CN BLOCK
1 1 ∼ 8 15 C 1 393 ∼ 400 64 A
9 ∼ 16 16 A 18 401 ∼ 408 65 B 18
2 17 ∼ 24 17 B 2 409 ∼ 416 66 C
25 ∼ 32 18 C 417 ∼ 424 67 A
33 ∼ 40 19 A 19 425 ∼ 432 68 B 19
3 41 ∼ 48 20 B 3 433 ∼ 440 69 C
49 ∼ 56 21 C 441 ∼ 448 70 A
57 ∼ 64 22 A 20 449 ∼ 456 71 B 20
4 65 ∼ 72 23 B 4 457 ∼ 464 72 C
73 ∼ 80 24 C 465 ∼ 472 73 A
81 ∼ 88 25 A 21 473 ∼ 480 74 B 21
5 89 ∼ 96 26 B 5 481 ∼ 488 75 C
97 ∼ 104 27 C 489 ∼ 496 76 A
105 ∼ 112 28 A 22 497 ∼ 504 77 B 22
6 113 ∼ 120 29 B 6 505 ∼ 512 78 C
121 ∼ 128 30 C 513 ∼ 520 79 A
129 ∼ 136 31 A 23 521 ∼ 528 80 B 23
7 137 ∼ 144 32 B 7 529 ∼ 536 81 C
145 ∼ 152 33 C 537 ∼ 544 82 A
153 ∼ 160 34 A 24 545 ∼ 552 83 B 24
8 161 ∼ 168 35 B 8 553 ∼ 560 84 C
169 ∼ 176 36 C 561 ∼ 568 85 A
177 ∼ 184 37 A 25 569 ∼ 576 86 B 25
9 185 ∼ 192 38 B 9 577 ∼ 584 87 C
193 ∼ 200 39 C 585 ∼ 592 88 A
201 ∼ 208 40 A 26 593 ∼ 600 89 B 26
10 209 ∼ 216 41 B 10 601 ∼ 608 90 C
217 ∼ 224 42 C 609 ∼ 616 91 A
225 ∼ 232 43 A 27 617 ∼ 624 92 B 27
11 233 ∼ 240 44 B 11 625 ∼ 632 93 C
241 ∼ 248 45 C 633 ∼ 640 94 A
249 ∼ 256 46 A 28 641 ∼ 648 95 B 28
12 257 ∼ 264 47 B 12 649 ∼ 656 96 C
265 ∼ 272 48 C 657 ∼ 664 97 A
273 ∼ 280 49 A 29 665 ∼ 672 98 B 29
13 281 ∼ 288 50 B 13 673 ∼ 680 99 C
289 ∼ 296 51 C 681 ∼ 688 100 A
297 ∼ 304 52 A 30 689 ∼ 696 101 B 30
14 305 ∼ 312 53 B 14 697 ∼ 704 102 C
313 ∼ 320 54 C 705 ∼ 712 103 A
321 ∼ 328 55 A 31 713 ∼ 720 104 B 31
15 329 ∼ 336 56 B 15 721 ∼ 728 105 C
337 ∼ 344 57 C 729 ∼ 736 106 A
345 ∼ 352 58 A 32 737 ∼ 744 107 B 32
16 353 ∼ 360 59 B 16 745 ∼ 752 108 C
361 ∼ 368 60 C
369 ∼ 376 61 A
17 377 ∼ 384 62 B 17
385 ∼ 392 63 C
− 40 − L3E-BW1349
Appendix M Checking Voltage of Scanner DC Power Supply (Check item No. 120)
Intercom amplifier
PCB-DC unit
1DPS(5V)
2DPS(5V/15V/-15V)
PCB unit 1
(Left side viewing
from the front)
Chassis 1
(Turn plate) 3DPS(24V)
4DPS(12V)
− 41 − L3E-BW1349
PCB-unit II (Turn plate)
CONTD3 PCB
CPU85A PCB
54DPS(5V)
53DPS(-15V)
PCB unit II
Filament PCB
PCB unit II
Chassis 1
301DPS(48V)
− 42 − L3E-BW1349
Chassis 1 (Turn plate)…In case of Revision “B” and following
PCB unit Ⅱ
CLCB1 PCB
101DPS(5V)
(Layout viewed from rear side of the gantry by removing the side cover.)
*…Position of the volume controller for output voltage.
102DPS
(Stepping motor driver) * *
− 42 – 1A − L3E-BW1349
Appendix N Checking Voltage of Table DC Power Supply (Check item No. 210)
Protection cover
Scanner side
Bellows cover
− 43 − L3E-BW1349
Appendix O Checking Electromagnetic Valve of Table Hydraulic System
Two electromagnetic valves are provided in the hydraulic circuit of table up/down move. The following describes
procedure to check the electromagnetic valve A and B respectively.
The hydraulic circuit is shown in Fig. O.1 Both of the electromagnetic valve A and B open only when the tabletop
goes down. Therefore, with the electromagnetic valve A and B closed alternately, press the tabletop DOWN button on
the scanner front control panel. At this time, if the tabletop does not go down, the electromagnetic valve can be
checked to be normal.
Directly connected
Electromagnetic valve A
Electromagnetic valve B
P R
Fig. O.1
− 44 − L3E-BW1349
1. Checking the electromagnetic valve A:
① Disconnect the connector 122CN.
② Keep the table DOWN button pressed for one minute.
③ If the table does not go down, the electromagnetic valve is normal.
Fig. O.2
Fig. O.3
− 45 − L3E-BW1349
Appendix P Replacing Battery Pack of Detector High Voltage Power Supply (XE)
(Check item No.122)
Greatest care should be taken not to get an electric shock since handling high voltage.
NOTE: The connector of spiral DC cable is shaped to avoid reverse connection of plus/minus for
safety.
NOTE: The battery pack is a MnO2 dry cell battery not including Hg or Cd. The battery pack no
longer required should be disposed of according to the local ordinance or the domestic law.
− 46A − L3E-BW1349
Appendix Q Replacing the Capacitor Microphone (Check item No.125)
6. Remove the mounting screws at four locations to remove the PCB mounting plate. Then remove the wiring
screw to disconnect the grounding wire.
Relay connector
Sponge Microphone case M4 wiring screw
− 47 − L3E-BW1349
7. Take out the microphone from the microphone case. Also remove the rubber pad around the microphone and
replace the microphone and relay connector assembly.
Parts code for set of relay connector assembly: (Lift cover assembly) 1387375A
(Cone cover assembly) 1387375C
Microphone
Relay connector
Microphone case
Rubber pad
8. After replacement, install the microphone in the reverse order of removing it. (Insert the concave of microphone
case into the cover hole and put sponge and PCB mounting plate on it.)
9. Restore all covers. Use Tyrap to fix the cable so as not to project.
10. Turn on the power to the gantry and table.
Turn the scanner to check if no cable or any part is caught.
11. Hear voice to check if the capacitor microphone works normally.
− 48A − L3E-BW1349
Appendix R Air Scan Image for Reference
The X-ray window is installed overlaid about 3 cm at the upper right of scan aperture. As shown in the picture below,
radiated black pattern is generated due to this structure, but it is not an artifact due to contrast medium or stain. Also, it
does not affect phantom scan and clinical imaging.
− 49 − L3E-BW1349
Appendix S Connecting High Voltage Cable
Follow the steps below to connect the high voltage cable with the high voltage bushing of X-ray generator for CT.
− 50 − L3E-BW1349
(4) Drying the surface
Use the dryer to dry the entire surface for about 5 minutes. If it is a 300W dryer, keep it about 10cm from the
target so as not to get a blast of hot air over 100ºC close to the target.
Fig. S.2
Fig. S.3
− 51 − L3E-BW1349
(6) Mounting the packing
Put the packing onto the base and tip (high voltage cable side) of cable head. Use the packing designated by us.
Do not use the packing used once because it swells.
(Use the packing after cleaning it with alcohol or the like and drying it.)
Apply a small amount of silicone grease to the cable head so that air does not enter between the tip packing and
cable head and insert the tip packing as shown in Fig. 5, and confirm no air remained. Also apply silicone
grease thin on the tip packing. Do not apply it to the pin.
(To prevent poor contact and corona discharge due to air remained)
If applying too much grease to the tip, the packing is hard to be inserted due to air remained. Apply a small
amount of grease.
Apply a small
amount of
grease so that
no air remained.
Fig. S.4
For the base For the tip
Fig. S.5
− 52 − L3E-BW1349
(7) Applying the compound
Apply the compound to the entire circumference of cable head evenly and use the spatula to spread it. Refer to
Fig. 4 on the previous page to apply 10cc (1/5 of SC104 tube) as a standard. Do not additionally apply it to the
tip.
Fig. S.6
Fig. S.7
− 53 − L3E-BW1349
(10) Fastening the fixture
Screw the fixture (about 2 and half turns as a standard).
(11) Baffle
Tighten two M4 screws to lock turning to fix the fixture.
Fig. S.8
− 54 − 57 L3E-BW1349
Whole-body X-ray CT System
Model Pronto
TECHNICAL GUIDE
--- Check & Service Program ---
To k y o , J a p a n
L4E-BW1360-2
(1) 1 L4E-BW1360
At the beginning
Caution to the readers of this manual
<CAUTION>
(2) 2 L4E-BW1360
Section 1 Registration of System Program
Therefore, the System installation must be made for the hard disks of both the CPU0
and CPU1. Location of the hard disks is shown below.
(The picture is shown with the cover so that the direction of the unit can be recognized.)
In installation of application software, after the system is booted up, the software for
CPU0 system and CPU1 system is installed. Development application and various
short-cut software are installed.
*1 Before installing the system software, be sure to keep a record of the study No.
It is used in 1.1.2.4 Registration of study No. Study No. can be checked by Study
No. in ID of the latest image or Study No. displayed by starting the Study No.
Setting program in Service Program. For usage of Study No. Setting program,
refer to “3.6 Study No. registration”
*2 Also save the user data for 3D Display program. (Refer to 1.1.2.7.)
*3 System installation and application installation may initialize the user data,
service data, Study No., etc. Be sure to take a backup copy of the service data and
user data, and also keep a record of Study No.
FD drive
1-1 L4E-BW1360
The flow chart of the installation is shown below.
START
*1, *2, *3
Save the service backup, user data and 3D
display program user data, and check < CAUTION >
Study No. Be sure to carry out the system
installation in both CPU0 and CPU1.
System installation only in CPU1 will
not result in normal operation.
Yes
Do you install system
software in CPU0 and CPU1?
No
Start up application installation
on Service menu.
< 2.1 >
CPU0 installation
Automatic startup of
application software
END
1-2 L4E-BW1360
1.1.2 Installation procedure
Install the CPU0 system software by using the hardware environment in the CPU1.
Use the boot FD (DOS) and system installation CD-ROM.
(1) Connect the CPU0 disk drive unit to the CPU1. Do not connect anything to the
CPU0.
1. CPU0 (Measurement)
3. END
After the above message is displayed, type “1”. Then, when the following message
is displayed, type “1”.
2. SYSTEM INSTALL
3. END
• Select “1. DISK INITIALIZE AND SYSTEM INSTALL” when an unused disk is
used or partition is destructed.
When the above message is displayed, set “CPU0 System Install CD-ROM” in the
drive, and then press Enter key. CPU0 installation starts.
1-3 L4E-BW1360
CPU0 System has been installed.
• Connect 83CN cable connector (CPU1 mother board) to 63CN connector (CPU0
mother board.)
• Press and hold down “+” mark at the upper left with your finger about 5 sec.
Releasing your finger moves “+” mark onto the upper center. (If it does not
move, you have pressed and held down too short. Press and hold down it a
little longer.)
• Pressing and holding “+” mark about 5 sec. and releasing it similarly moves
the mark to the upper right.
• Repeat the same operation to move it to the center left, center, center right,
lower left, lower center and lower right, 9 points in total.
If Task bar or other window appears on the touch panel, hide them. (Following
the above operation procedures can hide Task bar and windows.
1-4 L4E-BW1360
③ Select “Shutdown (S)” radio button.
(11) After the message box displaying "Please turn off the CT system." appears, shut
down the power of the OC.
1. CPU0 (Measurement)
3. END
After the above message is displayed, type “2”. Then, when the following message
is displayed, type “1”.
2. SYSTEM INSTALL
3. END
• Select “1. DISK INITIALIZE AND SYSTEM INSTALL” when an unused disk is
used or partition is destructed.
1-5 L4E-BW1360
Insert System Install CD-ROM.
Strike a key when ready …
When the above message is displayed, set “CPU1 System Install CD-ROM” in the
drive, and then press Enter key. CPU1 installation starts.
• Input Serial No. (4 digit number) printed on the label placed on the back panel
of the OC unit.
• Make sure that the input No. is correct. (Start up the Serial No. setting
program and check the number displayed as default.)
• Entry from Service menu after normal start-up (for installation of only
application software)
(Entry from Service menu can be done by selecting Start → Programs → 2. Service
→ Install the application program (Default password is “medico”.).
Judging the system status, the installation program executes either STEP1
program or STEP2 program suitable for each entry.
1-6 L4E-BW1360
Use the application installation CD-ROM.
After the above message is displayed, select Yes. The system automatically starts
and moves to STEP 2.
• Upon checking that the CPU0 system has started, select Yes. Application
installation starts and the copies are processed on DOS window.
The installation operation of application software has completed and the system
automatically restarts.
Before starting the system installation operation, register No. that is Study No.
recorded in Precaution *1 in 1.1.1 “+ 1”. In case the recorded Study No. is missing,
register Study No. that is Study No. appearing in ID of the latest image “+ 1”.
To register Study No., click Start → Programs → 2. Service → Study No. Setting,
and then register No. that is the recorded Study No. “+ 1”. For detailed registration
procedures, refer to “3.6 Study No. Registration”.
1.1.2.5 Also, load the user data of 3D display process. (Refer to 1.1.2.7.)
1.1.2.6 Miscellaneous
In the case “2. SYSTEM INSTALL” is selected in System Installation of the CPU0 or
CPU1, restart the system and then carry out Image Table Reset.
1-7 L4E-BW1360
1.1.2.7 Saving and writing user data of 3D Display process
This processing is used to prevent user data from being lost due to system software
or 3D processing software upgrading and to set individual user data in the system to
share a single unit among two or more operators.
Click on the [File] and [Save user data...], the [Save user's data] window as shown
below will be displayed.
The data names saved on the floppy disk will be displayed in [FD contents list].
By clicking on the tab at the top of the dialog box, the MARP data and customized
data names saved on the floppy disk will be displayed.
(b) To save data set in the current system onto a floppy disk
(1) Click on the tab corresponding to the data you want to save (MARP data or
customized data).
(2) Click on the [Save to FD...] button. The dialog box as shown below will be be
displayed.
(3) Enter a data name in the [Data name:] input box and click on the [OK] button.
If the data of the same name as the data name entered have been saved onto the
floppy disk, the overwrite confirmation dialog box will be displayed.
1-8 L4E-BW1360
Note : Up to 32 characters can be entered in the [Data name:] input box.
The eleven characters shown below cannot be used for data names:
(c) To Load data saved on a floppy disk into the current system
(1) Click on the tab corresponding to the data you want to load MARP data or
customized data).
(2) Highlight the data you want to load in [FD contents list].
(3) Click on the [Load from FD...]button. The dialog box as shown below will be
displayed:
(4) Confirm that the name of the data to load is displyed in the [Data name:] box, and
click on the [OK] button.
Note : By closing [Save user’s data] window after loading the data from FD, 3D
Display program is closed.
1-9 L4E-BW1360
1.2 System Setting
For system setup, click [Start] and point [Program] [Service] [System config.] then point
each item and click its detailed setting.
When registering the system program, the system settings are those by default. Be sure to
perform system setup after registering the system program. If using the system without
performing system setup, heat unit (HU) calculation of X-ray tube may not be correctly
carried out, scan may not be effected due to inconsistency of correspondence between the
scanner and software. This can cause system failures due to system malfunction or broken
X-ray tube. Be sure to perform system setup matching to the system status.
Also, perform system setup matching to the system status including “With(Add on)” and
“Without” of each function.
Pointing the arrow at each item displays detailed selection items or a list of candidate
setting values of each item. Click any value matching to the system.
Settings in this process are to match software to each item. Perform each setting after
confirming system parameters to be set since the system may not work correctly if no
hardware installed at the scanner side such as sub-second option, FOV500 and arcing
recovery.
When loading User Data into the system by using “Save User Data” function, make
sure to carry out it after having terminated System Setting.
If System Setting is carried out after User Data is loaded into the system, it is
likely to become an error.
1 - 10 L4E-BW1360
<Settings for X-ray tube assembly>
Depending on the type (HU) of X-ray tube to be mounted to the system, click an appropriate item
value referring to the table shown below.
1 - 11 L4E-BW1360
<Setting for detector>
The following three types are available for the detector. Click any value matching to the system.
1 - 12 L4E-BW1360
<Setting of standard scan protocol value>
Click to set values while paying attention to the following points.
(1) The protocol to be set for the combination of each detector and X-ray tube is applied with
values when shipped from the factory. Therefore, the standard protocol has been set.
(2) Normally restore the old protocol data using user data having been saved.
(3) Only when the user requests because any change occurs especially relating to the detector
and X-ray tube or availability of sub-second feature and so on, use this system setting to
register the standard protocol.
1 - 13 L4E-BW1360
<Setting for 1mm slice>
Click either of “Available” if the 1mm slice option is provided or “Unavailable” if it is not
provided.
1 - 14 L4E-BW1360
<Setting for patient table>
For presence of the sub-second scan, click any item value with Pronto or Robusto as a standard.
1 - 15 L4E-BW1360
<Setting of arcing recovery>
The arcing recover function temporally halts operation of the X-ray high voltage unit when high
voltage discharge occurs in the X-ray tube assembly and restarts it after predetermined time.
This function allows scan to continue without being interrupted if the number of occurrence is
within the permissible range even when high voltage discharge occurs in the X-ray tube assembly
in the volume scan, dynamic scan and so on.
(However, if arcing recovery is activated during scan, any abnormal data may enter into the
image data. Therefore, a warning message is displayed.)
Set “Yes” if the scanner is applicable to the arcing recovery, while “No” if it is not applicable.
As a hardware revision of applicable scanner, the following scanner hardware after the following
revision is necessary.
1 - 16 18 L4E-BW1360
Section 2 Check Program
The check program can be started by selecting "Start”, “Programs” and "1. Check" in this order.
Upon completion of the check operation, be sure to enable the password by Password setting for
Check program.
This program carries out check programs for data and route of the preamplifier, and
includes the following processes.
By loading the test pattern data from the ROM in the preamplifier, the program checks
if the data transfer route works normally.
This program checks if leakage current from the detector and offset voltage of the
integrator in the preamplifier are within the specified values.
2-1 L4E-BW1360
2.1.2 Basic screen
(1) MENU
(2) THRESHOLD
Click this button to re-input the threshold value defined in THRESHOLD entry.
2-2 L4E-BW1360
(7) HELP button
(8) MESSAGE
(9) STATUS
When either an error is detected or the check program is stopped by the STOP button,
the EXECUTE button is enabled, and operations related to the menu are also enabled.
2-3 L4E-BW1360
(2) OFFSET DATA check
Selecting this menu displays the dialog box to input the threshold
value. Input the desired value and press ENTER key.
The average value obtained by the last OFFSET DATA check is
also displayed in this dialog box. Note that this average is not
displayed in the first OFFSET DATA check operation after the
Preamp Check program is started.
Pronto-SE : 700
Pronto-XE : 700
Pronto-VE : 50
Upon completion of the check process, the result is displayed in the status area, and
then the threshold entry dialog box is displayed. The operations related to the menu
are now enabled.
To carry out the check again, follow the operations from entry of the threshold value.
When time out error is detected, the EXECUTE button is enabled, and the operations
related to the menu are now enabled.
2-4 L4E-BW1360
2.2 Scanner Check
2.2.1 Introduction
The scanner check performs checking of the major functions of the scanner unit and
patient table and includes the following processes.
Communication between the scanner unit and patient table and between the scanner
unit and X-ray control unit are checked.
The scanner is rotated in CW direction for the specified scan time, and the required
time is measured.
The switch on the scanner control panel is read and the switch No. is displayed on the
control panel.
2-5 L4E-BW1360
2.2.2 Basic screen
Table check includes three check modes, that are LED lighting
check, gantry switch read check and moving speed check.
2-6 L4E-BW1360
(3) General box
According to the item selected on Sub Menu, the parameter entry box and small menu
are displayed.
Selecting the scanning time check displays the scan time selection
menu.
(4) PROGRESS
Click this button to reenter the parameters defined in the general box.
Click this button to suspend or stop the check operation being executed.
(10) MESSAGE
(11) STATUS
2-7 L4E-BW1360
2.2.3 Operation procedure
Selecting this menu displays the entry box for number of checking
under the SUB MENU. Input a desired number of checking and
press ENTER key.
Unless STOP button is clicked during check operation, the check operation for the
specified number of checking is completed, and a message showing normal completion
of the operation is displayed in the status display area.
If any error is detected during check operation, the error message is displayed in this
area and the check operation is suspended.
2-8 L4E-BW1360
(2) Scanning time check
Selecting this menu displays the scanning time selection box under
the SUB MENU. Click the time to be checked.
Clicking any of the options enables EXECUTE button. Clicking
EXECUTE button starts the check operation.
During executing the program, operations related to the menu are
disabled.
Care must be exercised as the scanner rotates.
NOTE: It takes more than the selected scan time (CW) at least until the result is
displayed.
(3) Timing pulses generation
2-9 L4E-BW1360
As generation of timing pulses starts, EXECUTE button is disabled,
while STOP button is enabled. Clicking STOP button during check
operation suspends generation of the control signal.
NOTE: Unless STOP button is clicked while control signals are being generated,
the process is not completed.
NOTE: Unless STOP button is clicked during the check operation, the process is not
completed.
2 - 10 L4E-BW1360
The lighting sequence of the LEDs is repeated from (a) to (d) as follows.
(a)
+001.1→+002.2→…+009.9→+110.0→+220.0→…+990.0 →+1000.0→-1000
(d) HEIGHT
011→022→…099→100→200
2 - 11 L4E-BW1360
No. of each switch is shown below.
1 5 9 13 SET *
2 6 10 14 START *
3 7 11
4 8 12
NOTE 1: If more than one switch are pressed, priority is given to the switch with
smaller No.
2 - 12 L4E-BW1360
Operation procedures
The set table speed and actual moving speed are displayed in TILT and POSITION
indicators respectively.
Press the table backward move button or forward move button together with high speed
move button to move the table backward or forward.
In this case, actual moving speed is displayed in POSITION indicator. If this value is
correct within the range shown in the following table, it is normal, while it is out of the
range, judge it as error and adjust the table move speed.
Next, press RESET button several times and display “12” in HEIGHT indicator and
“20.0” in TILT indicator. Then, check if the moving speed by the above operation is
within the value shown in the table or not.
Repeating the above operations, check the actual speed for each of the 4 moving speeds.
NOTE: Other values are displayed while pressing RESET button to display the
specified values in HEIGHT and TILT indicators, but those values are not
subjected to the check.
2 - 13 L4E-BW1360
As the check operation starts, EXECUTE button is disabled, while
STOP button is enabled.
Clicking this STOP button during check operation, the check
operation is suspended.
2.3.1 Introduction
Board check is to check the high speed computing unit, memories and data transfer, and
includes the following check modes.
NOTE: The table showing correspondence between each item and actual hardware is
provided in the 2.3.3 Reference.
Two modes are available for selection of check item and operation of start/stop .
One is operation on Menu Bar displayed on Windows and the other is operation on Sub
window displayed from Menu. Selection of either mode is made on Display menu to be
described later.
2 - 14 L4E-BW1360
NOTE: This version does not include the filing operation function.
Therefore, only Exit is available in the pull-down menu.
① Exit
① Start
The programs with check mark are sequentially processed. Check operation is
looped until the operation is stopped or an error is detected. In case more than
one program are selected and an error is detected during operation, the check
operation is stopped after the last one of the selected processes is completed.
The name of the check program in operation, loop count and status of in-operation
or stop are displayed in the status bar.
This menu is disabled only during execution of the check program.
② The check process is terminated after the check operation being executed is
completed.
The result of the check operation being executed when Stop is clicked is displayed.
This menu is disabled while the check process being operated.
NOTE: Only this menu window is available in this version for such selection.
① Menu
2 - 15 L4E-BW1360
NOTE: The menu operation is disabled while the check operation being executed.
① Tile
Tile display of the displayed menu window and check result window are made
The windows are adjusted to the same size so that the whole window can be seen.
② Cascade
Cascade display of the displayed menu window and check result window are made.
Windows are adjusted to the same size and arranged so that they are overlapped.
③ Arrange Icons
④ Window list
The list of the windows being opened is displayed. The active window among them
is shown with a check mark.
Clicking the window without a check mark can make it active.
This is also used to display the hidden window on the top.
If the window is made active by clicking the displayed window, the check mark in
the list also moves to the active window.
① Help Topic
② About
2 - 16 L4E-BW1360
(8) Operation on window
Every time each of items is clicked, the check mark alternately appears or disappears.
More than one items can be selected, and the check processes with the check mark are
executed. The check process being executed changes its color.
The check operation is executed in loop until the check operation is stopped or an error
is detected.
In case more than one item are selected and an error is detected during check operation,
the process is terminated after the last check operation is completed.
① Start button
② Stop button
The process operation is stopped after the check operation being executed is
completed.
The result of the check operation being executed when Stop button is pressed is
displayed.
This button is disabled while the check operation is stopped.
① FR PCB 1
② FR PCB 2
③ FR PCB 3
④ Back Projector
CPU0 CPU1
① ② ④ ③
2 - 17 L4E-BW1360
2.4 Ideal reconstruction check
2.4.1 Introduction
Ideal reconstruction check is the function to check reconstructed image by using the
dedicated raw data.
① Exit
Ideal Reconstruction Check program is closed.
① Start
The check program is executed. The check program is looped until it is stopped or an
error is detected. While the program being executed, statuses of loop count and
Executing/Stop are displayed in the status bar. This menu is disabled only during
execution of the check program.
② Stop
The program is stopped after the check program being operated finishes.
The result of the check being executed when Stop is clicked is displayed.
This menu is disabled when the check program is stopped
2 - 18 L4E-BW1360
① Help Topic
Help for Ideal Reconstruction Check is started.
② About
Version information of Ideal Reconstruction Check is displayed.
① Start button
This button has the same function as that of Start menu.
This button is disabled only while the check program being executed.
② Stop button
This button has the same function as that of Stop menu.
This button is disabled while Check program is stopped.
(6) Result
As check result, the reconstructed image and data check result
are displayed.
Each display is altered by the tab.
① Direct input
Move the focus to each of the boxes and directly input a value from the keyboard.
By pressing Enter key on the keyboard, the image reconstructed with the input
values is displayed again.
Inputting any character other than numeral causes an error.
2 - 19 L4E-BW1360
② Changing by spin button
Move the mouse cursor to each of the spin buttons and press the left mouse button.
Then, value in the box can be changed up and down. When the left mouse button is
released, the image is displayed again.
Numeral can be changed in two steps. Pressing and holding down the spin button
change the value by 1 each up to the first 30 counts and then in 10 steps up to the
next 30 counts. Pressing and holding down the button change the value in 100 steps
from the next 30 counts.
2 - 20 L4E-BW1360
2.4.3 Contents to be checked and description of error displays
Ideal Reconstruction Check program executes image reconstruction process based on the
data prepared in advance. The image reconstruction process for check is executed as
shown in the following flow chart.
Setting raw data and filter required for Check program and preparation
of software within the calculation PCB.
2 - 21 L4E-BW1360
A display example of check result is shown below.
(As it is an example with an error found, correct value, contents of error data, and type of
error vary depending on occurrence condition of the actual error.)
Note: (1) One hundred errors at maximum can be registered in an error list.
(2) Total number of errors on the screen is the total number of data that were
found error in the check process.
Error data
Total number of errors CPU data Board data
CPU data
CPU data Description
CPU0 Error possibly related to the environment in CPU0 side
CPU1 Error possibly related to the environment in CPU1 side
2 - 22 L4E-BW1360
Board data
Board data Description
FR-MK(2) PCB mainly executing Image reconstruction process
step ① in the above-mentioned flow chart
PC-BP-RV(3) PCB mainly executing Image reconstruction process
step ② in the above-mentioned flow chart
FR-MK(4) PCB mainly executing Image reconstruction process
step ③ in the above-mentioned flow chart
Error data
Type of error Description
“NO ERROR DETECTED” When the check process has finished normally.
“TIME OUT ERROR” When the system hangs up because a response from
the other cannot be obtained within the specified time
during processing. Possibility of the error due to the
hanged up process related to the CPU No. and board
data displayed at that time is high.
“DSP BOARD NOT FOUND” When any image process board to execute
reconstruction process cannot be found.
When the board displayed in the board data is not
found mainly by software.
“PARAMETER INSUFFICIENT” When any of parameters to execute Ideal
reconstruction check is not found in the system disk.
“DSP BOOT ERROR” When program loading has failed even if the program
is tried to be loaded in the hardware for reconstruction.
“ILLEGAL RESULT” Error displayed when illegal reconstruction result is
obtained in comparison of the reconstruction result
with the correct value.
Board data for error display:
FR-MK2(2): Error occurred in Reconstruction result
check ① in the flow chart.
PC-BP-RV(3): Error occurred in Reconstruction result
check ② in the flow chart.
FR-MK2(4): Error occurred in Reconstruction result
check ③ in the flow chart.
“SYSTEM ERROR” Error occurred due to other causes than the above
occurred in execution of Check program.
Cautions:
(1) If the image of Ideal reconstruction check is displayed with a smaller window width,
faint ring pattern can be seen, but such pattern appears even in normal image
reconstruction process. In case of abnormal image reconstruction result, it is judged as
2 - 23 L4E-BW1360
check error in the reconstruction result check, therefore, make a judgment of normal
image or abnormal image in the check result.
(Although it is not caused usually, if a displayed image is mixed with very faint streak
artifact, it may not be judged as check error, therefore be sure to check the displayed
image.)
(2) If an “ILLEGAL RESULT” error is found, the data is displayed in both integer type and
floating decimal type in the error list display. In such case, for each board data,
refer to:
FR-MK2(2) : Floating decimal point display part (part with Real display)
PC-BP RV(3) : Floating decimal point display part (part with Real display)
FR-MK2(4) : Integer type (part with HEX display)
(The displayed contents in the other part mean nothing.)
(3) When this Check program has been executed (including only startup), the system must
be rebooted. To start scan operation or other check service programs thereafter, reboot
the system.
2 - 24 L4E-BW1360
Image display example:
2 - 25 43 L4E-BW1360
Section 3 Service Program
The Service program can be started by selecting "Start"," Programs" and "2. Service" in this order.
After Service work is completed, be sure to enable Password by “Setting Password” program.
Raw Data analysis is the function to display raw data acquired by the CT scanner
system and to analyze the image. The following processes are available.
(1) PTD
(2) PLD
Value for any specified single projection on the displayed raw data is displayed.
(3) CHD
Value for any specified single channel of the displayed raw data is displayed.
(4) PLP
Profile of the data for any specified single projection of the displayed raw data is
displayed.
(5) CHP
Profile of the data for any specified single channel of the displayed raw data is
displayed.
3-1 L4E-BW1360
(6) PLC
The maximum value, minimum value, difference (maximum value - minimum value),
mean value, variance and standard deviation for each projection in the range of all
projections of the displayed raw data are calculated and displayed.
(7) CHC
The maximum value, minimum value, difference (maximum value - minimum value),
mean value, variance and standard deviation for each channel (data for all displayed
projections) in the range of any channel of the displayed data are calculated and
displayed.
(8) CLB
This is used for center alignment of the compensator. Deviations of the compensator
toward left and right are shown with the difference in outputs between the first half
channels and second half channels.
(9) VCK
After threshold values are set for the difference (max. value - min. value) and standard
deviation, the difference (max. value - min. value) and standard deviation for each of all
channels (all scanned projection data) are calculated, and if the result exceeds the set
threshold, the channel No. and calculated result are displayed. These data are usually
used for vibration check.
(10) FCE
(11) CACK
This is used for position alignment of the X-ray tube unit. After raw data obtained by
scanning the adjustment rod supplied with the system is displayed, this process is
started up.
3-2 L4E-BW1360
NOTE: Number of channels and number of views of raw data described in this
instruction manual differ depending on model of the scanner system.
Placing the focus on each of the entry boxes and input a value directly from the
keyboard.
Pressing ENTER key displays the image reconstructed with the newly input W/W
and W/L.
Inputting any character other than numeral causes an error.
Adjustment by spin button
Placing the mouse cursor on each of the spin buttons, click the left mouse button to
change the value up and down. When releasing the left mouse button, the
reconstructed image is displayed.
The value can be adjusted in 2 modes. By pressing and holding down the spin
button, the value changes up to 30 counts in 1 steps and it changes up to the next
30 counts in 10 steps. Further holding down the button, it changes in 100 steps
from the next 30 counts (that is, it advances by 300 counts).
NOTE: With PTD selected on Main window, adjustment by from the keyboard is
disabled.
② Data display
As moving the mouse cursor on the displayed image, the coordinates (channel and
view) at the pointer of the mouse cursor and data value are displayed under the
menu bar.
3-3 L4E-BW1360
(2) Basic menu
This menu functions to open and close raw data and Cal. data to be
analyzed. As shown in the left Fig., it contains 3 menus of Open,
Close and Exit.
Close menu is enabled while raw data being displayed.
While raw data being displayed, Open menu is disabled.
① Open
Raw data
Clicking a line on the patient data table displays the slice data table of the selected
patient in the line.
Clicking the desired line on the slice data table and then clicking OK button
displays the raw data of the selected slice.
Pressing Cancel button closes the dialog box without displaying any data.
3-4 L4E-BW1360
NOTE: The above Fig. is one of examples and it may be different from the one
actually displayed.
Cal. data
NOTE: 1) "All the data related to the selected conditions" means, for example,
all the files of the slices corresponding to the tube voltage in case of
Phantom Cal. data, and data of all the phantoms with different
diameter in one file.
2) The above Fig. is one of examples and it may be different from the
one actually displayed.
② Close
The currently displayed Raw data or Cal. data is closed. In case other analysis
windows are displayed, they are also closed at the same time.
On this menu, the Raw data analysis process cannot be closed.
This menu is not enabled unless Raw data or Cal. data is displayed.
③ Exit
This menu is used to closes the Raw data analysis process. The Raw data, Cal.
data or Analysis window being analyzed is automatically closed.
This menu is used to open and close the windows for various types
of analysis.
The menu contains analysis processes of PTD, PLD, CHD, PLP,
CHP, PLC, CHC, CLB, VCK, FCE and CACK.
Every time each item is clicked, the check mark alternately
appears or disappears. Simultaneously the corresponding window
is opened or closed.
This menu is not enabled unless either Raw data or Cal. data is
displayed.
① PTD
3-5 L4E-BW1360
The status bar is displayed under the image. As the mouse cursor moves onto the
image the cursor changes to a cross mark. Clicking the left mouse button at any
position displays values of that coordinates (channel and view) in the status bar.
And, a red cross caret flashes at the clicked position.
To close PTD, click the PTD menu again.
NOTE: 1) Note that if other processes such as PLD, PLP, CHD and CHP which
are described later are simultaneously displayed the caret may be
hidden by the line on the image drawn in each process and not
displayed.
2) Key entry operation for W/W and W/L is disabled during PTD
operation.
3-6 L4E-BW1360
② PLD
Checking PLD displays the blank sub-window of PLD in the lower half of the
window. Clicking any position on the image displays values in the view direction
in the PLD sub-window, and a red line appears on the view displayed on the image.
To close the PLD window, click either the PLD menu again or the × button at the
upper right of the PLD sub-window.
③ CHD
3-7 L4E-BW1360
Checking CHD displays the blank sub-window of the CHD in the right half of the
window. Clicking any position on the image displays values in the channel
direction on the CHD sub-window. Also, a red line appears on the displayed
channel on the image.
To close the CHD window, click the CHD menu again or click the × button at the
upper right of the CHD sub-window.
④ PLP
Checking PLP displays the blank sub-window of the PLP in the lower half of the
window. Clicking any position on the image displays the profile of the data in the
view direction in the PLP sub-window. A red line is displayed on the displayed
view on the image.
By moving the mouse cursor onto the profile image in the sub-window and pressing
the left mouse button, a red line appears in the vertical direction on the profile, and
its channel No. and data are displayed in the status bar.
By moving the mouse in the horizontal direction with the left mouse button pressed
and held down, the red line moves along the mouse movement and the values in the
status bar changes accordingly in real time. As the left mouse button is released,
the red line disappears.
Four buttons are provided in the sub-window, of which functions are as follows.
DEFAULT
A width between the minimum value and maximum value on the ordinate is set
from -32768 to 32767.
3-8 L4E-BW1360
MAX/MIN
A width between the minimum value and maximum value on the ordinate is set
between the minimum value and maximum value in the displayed profile.
KEY IN
A width between the minimum value and maximum value on the ordinate is freely
set.
Redisplay
If the size of the sub-window is changed, the scale of the ordinate is adjusted
according to the length of the window and the profile is redisplayed.
To close the PLP window, click either the PLP menu again or the × button at
the upper right of the PLP sub-window.
⑤ CHP
Checking CHP display the blank sub-window of the CHP in the right half of the
window. Clicking any position on the image displays values in the channel
direction in the CHP sub-window. A red line is displayed on the displayed channel
on the image.
3-9 L4E-BW1360
By moving the mouse cursor onto the profile image in the sub-window and pressing
the left mouse button, a red line appears in the view direction on the profile, and its
view No. and data are displayed in the status bar.
By moving the mouse in the vertical direction with the left mouse button pressed
and held down, the red line moves along the mouse movement and the values in the
status bar changes accordingly in real time. As the left mouse button is released,
the red line disappears.
Four button are provided in the sub-window, of which functions are as follows.
DEFAULT
A width between the minimum value and maximum value on the abscissa is set
from -32768 to 32767.
MAX/MIN
A width between the minimum value and maximum value on the abscissa is set
between the minimum value and maximum value in the displayed profile.
KEY IN
A width between the minimum value and maximum value on the abscissa is freely
set.
Redisplay
If the size of the sub-window is changed, the scale of the abscissa is adjusted
according to the width of the window and the profile is redisplayed.
To close the CHP window, click either the CHP menu again or the × button at
the upper right of the CHP sub-window.
3 - 10 L4E-BW1360
⑥ PLC
Checking PLC displays the sub-window, and max. value,. min. value, difference
(max. value - min. value), mean value, variance and standard deviation are
calculated for each projection in the range of all projections of the displayed raw
data (data for all channels) and displayed in the table.
To close the PLC window, click either the PLC menu or the × button at the upper
right of the PLC sub-window.
⑦ CHC
3 - 11 L4E-BW1360
Checking CHC displays the sub-window, and max. value,. min. value, difference
(max. value - min. value), mean value, variance and standard deviation are
calculated for each channel in the range of all channels of the displayed raw data
(data for all displayed projections) and displayed in the table.
To close the CHC window, click either the CHC menu or the × button at the upper
right of the CHC sub-window.
⑧ CLB
Checking CLB displays the CLB sub-window in the lower half of the window and
the profile of averaging all views of the displayed raw data is displayed.
In the status bar are displayed the average value of 10 channels each of the left side,
center part and right side and the difference between the data of the left side and
right side.
Four buttons are provided in the sub-window and their functions are as follows.
DEFAULT
A width between the minimum value and maximum value on the ordinate is set
from -32768 to 32767.
MAX/MIN
A width between the minimum value and maximum value on the ordinate is set
between the minimum value and maximum value in the displayed profile.
KEY IN
A width between the minimum value and maximum value on the ordinate is freely
set.
3 - 12 L4E-BW1360
Clicking this button opens the dialog box for entry of
the min. value and max. value. Enter values and click
OK button. Clicking CANCEL button closes this
dialog box without setting any value.
Redisplay
If the size of the sub-window is changed, the scale of the ordinate is adjusted
according to the length of the window and the profile is redisplayed.
To close the CLB window, click either the CLB menu again or the × button at the
upper right of the CLB sub-window.
⑨ VCK
Checking VCK displays the sub-window, and the difference between the max. value
and min. value, profile of standard deviation, channel No. exceeding the default-set
threshold and table showing the calculated results of all channels for the displayed
raw data are displayed.
Also, the set threshold is shown with dotted line on each of the profiles.
The following buttons and input boxes are available at the top of the sub-window.
Diff Profile
This button is of toggle operation type, and every time it is clicked, On and Off are
alternately selected.
With this switch pushed in, the profile of the difference (max. value - min. value) is
displayed in the VCK sub-window. With the switch not pushed in, the profile is
not displayed.
3 - 13 L4E-BW1360
SD Profile
This button is of toggle operation type, and every time it is clicked, On and Off are
alternately selected.
With this switch pushed in, the profile of standard deviation is displayed in the
VCK sub-window. With the switch not pushed in, the profile is not displayed.
Error List
This button is of toggle operation type, and every time it is clicked, On and Off are
alternately selected.
With this switch pushed in, channel No. of which data exceed the set threshold
and calculation results are displayed in the VCK sub-window. With the switch
not pushed in, the list is not displayed.
MAX-MIN
The threshold value of the difference (max. value - min. value) is designated.
SD
Based on the difference (max. value - min. value) and the value input in the
threshold entry area for standard deviation, Error List is displayed. The dotted
line showing the threshold displayed in each profile is also updated.
A width from the minimum value to the maximum value in the ordinate range is
set to 0 ~ 65535.
MAX/MIN
A width from the minimum value to the maximum value in the ordinate range is
set between the min. value and max. value in the displayed profile.
KEY IN
A width between the minimum value and maximum value in the ordinate range is
freely set.
Redisplay
If the size of the sub-window is changed, the scale of the ordinate is adjusted
according to the length of the window and the profile is redisplayed.
3 - 14 L4E-BW1360
To close the VCK window, click either the VCK menu again or the × button at
the upper right of the VCK sub-window.
⑩ FCE
Checking FCE displays the sub-window. This sub-window contains a combo box,
and by clicking the downward arrow button at the right side the currently
registered defective channel is displayed.
By inputting the channel to be registered in the combo box and clicking the
ENTRY button at the top, the input data is set.
DELETE
Click the downward arrow button at the right side of the combo box to display the
registered data, click and highlight the data to be deleted and click the DELETE
button at the top. Then, the highlighted data is deleted.
3 - 15 L4E-BW1360
⑪ CACK
Before executing this process, display the raw data acquired by scanning the rod for
adjustment supplied with the system.
Clicking CACK displays the trace part of the path of the rod for adjustment in red,
and the check result is displayed in the adjustment data dialog box. According to
the displayed result, make adjustment of the X-ray tube unit position.
This dialog box can be closed by clicking CLOSE button.
NOTE: 1) Without display of the raw data acquired by scanning the rod for
adjustment, correct result cannot be obtained.
2) Set the rod for adjustment at the position 20 ~ 25mm off the scanner
rotation center.
① Tile
3 - 16 L4E-BW1360
Size of each of the sub-windows is adjusted to the same size so that all windows can
be seen. Tile display operation hides raw data or Cal. data window. To display Raw
data or Cal. data window, move its sub-window.
② Cascade
The currently displayed sub-windows for analysis are cascade displayed. Size of
each window is adjusted same and they are rearranged for cascade display.
3 - 17 L4E-BW1360
③ Arrange Icons
Icons of the minimized sub-windows for analysis are arranged in the same intervals
so that they are not overlapped.
3 - 18 L4E-BW1360
The list is also used to display any hidden sub-window on the top.
If an opened sub-window is clicked to be made active, the check mark also moves to
the activated sub-window in the list.
① Help Topic
② About
3 - 19 L4E-BW1360
3.2 CT Number Adjustment
(1) Click "Start" button and point “Programs” and "2. Service".
Then, click "CT Number Adjustment".
(a) Clicking "Calculation" on the CT Number Adjustment Process Menu window displays
the following window.
Following contents are displayed in the Remark column of the list view depending on
calculation date and time and measurement date and time.
3 - 20 L4E-BW1360
"Calculation impossible" : No scan data.
"Calculation required" : Data of which scan date and time is later than
calculation date and time.
(b) By setting the parameters for calculation in the left part of the window, the items
corresponding to the selected calculation condition are displayed in the list view with a
check mark.
In the case more than one check mark are included in the list view, "Calc. Start"
button is enabled. However, the item marked with "Calculation impossible" in
Remarks column is displayed with a check mark, the "Calc. Start" button is
displayed in light color.
Clicking "Calc. Start" button starts the calculation.
(c) While the calculation being executed, the following window is displayed.
• In case an error occurred, retry the CT number adjustment scan with this
condition.
(d) Upon completion of the calculation or by clicking "Abort" button on the error display
window, the following Calculation Process window is displayed.
3 - 21 L4E-BW1360
• The following contents are displayed in the Remarks for the items with a check
mark in Selection column in the list view.
To make calculation again, click [Redo]. In this case the screen returns to the
window in the step (a). To close the calculation program, click [Close].
To set default parameters for calculation, click "Default Parameter Setting" on the
"Settings" menu in the step (a).
3 - 22 L4E-BW1360
The calculation parameters to be displayed as default on the window displayed in the
step (a) are set.
Checking "Use previous calculating parameters." displays the parameters used in the
previous calculation are displayed as default.
If "Set individually." is checked, check each parameter of tube voltage and slice
thickness. These parameters are displayed as default.
The data set here are saved even after the CT number adjustment process is closed.
(a) Clicking "Date Check" on the CT Number Adjustment Process Menu window displays
the following window.
The following contents are displayed in Remarks of the List View depending on
calculation date and time and scan date and time.
"Calculated" : Calculation date and time are later than the latest scan date
and time with the same tube voltage and slice thickness
including other phantoms.
"Calculation required"
: Scan date and time are earlier than calculation date and
time in the case other than the above.
3 - 23 L4E-BW1360
(b) Selection of list view display parameter
Scan and calculation date and time corresponding to each parameter of the tube voltage,
slice thickness and phantom diameter highlighted in the left side of the window are
displayed in the List View. By clicking any parameter in the list, the item to be
highlighted can be changed.
Pointing the "Priority Order Selection" in "View(V)" menu displays the following menu.
Clicking the desired display order reflects on the display order in the List View.
3 - 24 L4E-BW1360
3.3 Error List Display
• If none of error data has been registered, a message “Error data has not been
registered.” is displayed.
(3) Selection of display order (descending and ascending) of date and time and display of
second in ms steps
By either clicking [Date] in List View or pointing [Set display order] in [View] menu, the
following dialog is displayed.
To change display order of date and time (descending/ascending), select the desired
display order in setting display items. By checking [display down to msec], time can be
displayed down to in ms unit. After selection of a display order, click [OK] button.
3 - 25 L4E-BW1360
(4) Display of only the specified process name
By either clicking [Process Name] in List View or pointing [Set display order] in [View (V)]
menu and clicking [Process], the following dialog is displayed.
To display errors by process name where errors have been registered, select the desired
process name in setting display items. By selecting [All], the contents of all the process
names where errors have been registered are displayed. After selection, click [OK]
button.
By either clicking [Message] in List View or pointing [Set display order] in [View (V)]
menu and clicking [Message (M) …], the following dialog is displayed.
To display errors by message where errors have been registered, select the desired
message in setting display items. By selecting [All], the contents of all the messages
where errors have been registered are displayed. After selection, click [OK] button.
3 - 26 L4E-BW1360
(6) Setting information in Error Log file
To change the set contents, check the desired information box and change the contents of
the desired item.
• Analysis side
• Measurement side
3 - 27 L4E-BW1360
※ Caution about the maximum size (byte) of error log file
• Error log file is managed by process name, and if file size under a process name
exceeds the maximum size, the current file is stored as the file before one
generation and the current file is initialized. Because of this reason the file before
two generations is not stored, if errors occur frequently or old error information is
needed to store, make the file size larger.
With the default file size of 16384 bytes set, the current file + file for one previous
generation can store error data up to about 300 errors in total
The button to carry out phantom calibration scan and CT number adjustment scan is
usually not on the touch panel. These scans are called Service scan. To enable these
scans, Service scan function must be set to ON.
However, after completion of the Service scan, be sure to set the Service scan function to
OFF to prevent the user from use of the function by mistake.
(1) Click [Start] button and point [Programs] and [2. Service].
Then, click "Service Scan Selection".
The window is displayed with the [OFF] checked if the service scan buttons (Phantom
Cal, Maintenance scan and CT Number Adjustment scan) are not currently displayed on
the touch panel, while it is displayed with the [ON] checked if they are displayed.
To display each of the buttons for Service Scan on the touch panel, check [ON] and click
[OK] button.
To erase each of the buttons for Service Scan on the touch panel, check [OFF] and click
[OK] button.
3 - 28 L4E-BW1360
3.5 Service Backup
3.5.1 Outline
This function is used to store CT number adjustment values, phantom calibration data
and phantom data for calibration stored in the magnetic disk into the FD cartridge, to
read those data from the FD cartridge and register them in the magnetic disk.
INITIALIZE FD CARTRIDGE
This function initialize FD cartridge. This operation is carried out to use a new FD
cartridge or using the used cartridge after initializing it.
VIEW FD CARTRIDGE
This function allows the data in the FD cartridge to be checked.
WRITE FD CARTRIDGE
This function writes various data in the FD cartridge.
READ FD CARTRIDGE
This function reads various data from the FD cartridge and writes them into the
magnetic disk.
3 - 29 L4E-BW1360
(3) INITIALIZE FD CARTRIDGE
Insert the service backup FD into the FD drive, and click [VIEW FD CARTRIDGE] on
Service Backup Process window. Then, the data stored in the FD cartridge are
displayed as follows.
The data with blank date column and message "No data." in the comment column
means that no data is stored in the FD cartridge.
3 - 30 L4E-BW1360
(5) WRITE FD CARTRIDGE
Insert the FD for service backup into the FD drive, and click [WRITE FD CARTRIDGE]
on the Service Backup Process window. Then, Write Process window is displayed as
follows.
The check box for the data to be written into the FD is checked.
Checking the check box displays the current date in the date column and enables key
entry in the comment column. If necessary, enter comments.
Clicking [Start] button starts to write the checked data into the FD. While the process
being executed, the following window is displayed and Pause/Continue/Stop operation
are enabled.
After FD write operation is completed or clicking [Stop] button, the process returns to
display of the first window in the step (5).
3 - 31 L4E-BW1360
(6) READ FD CARTRIDGE
Insert the FD for service backup into the FD drive, and click [READ FD CARTRIDGE]
on the Service Backup Process Menu window. Then, the Read Process window is
displayed as follows.
• The data displayed in light color are not stored in the FD cartridge, and it shows that
they cannot be read.
The check box for data to read from the FD is checked.
The data displayed in light color cannot be checked.
Clicking [Start] button starts to read the checked data from the FD. While the process
being executed, the following window is displayed and Pause/Continue/Stop operation
are enabled.
After FD read operation is completed or clicking [Stop] button, the process returns to
display of the first window in the step (6).
3 - 32 L4E-BW1360
3.5.3 Error message dialog
FD is not ready.
FD is write-protected.
3 - 33 L4E-BW1360
FD has not been initialized for service backup.
Verify error
3 - 34 L4E-BW1360
3.6 Study No. Registration
Study No. that is one of image data to be affixed to DICOM-format image is set.
(1) Click [Start], and point [Programs], [2. Service], and then click [Study No. Setting].
(2) On the following displayed window, input Study No. and click [OK] button.
(Input a value made by adding “+1” to the No. recorded before system software
installation.)
Because the above window is the one for Study No. 2, input “3” made by 2 “+1” and click
[OK] button. (The Study No. set here will be effective after rebooting the system next
time.)
3 - 35 L4E-BW1360
3.7 Setting Password
3.7.1 Introduction
The password setting is used to change or make the password effective that is requested
for entry in starting the Check and Service Process program.
By setting the password, a limitation is given to starting the Check and Service Process
program, and data tampering by the user or erroneous deletion of data are prevented.
To change the password, input all of three items: the old password, new password and
confirm, and click [OK] button.
For old password, input the password used before change.
For new password, input the password to be used newly.
For confirm, input the new password again for reconfirmation (to prevent erroneous
entry).
If all the entered data are correct, the new password is set.
If any error is found in the entered data, the error message is displayed and reentry of
the passwords is requested.
※ The password set at the factory before shipment is "medico".
(2) Taking effect
The setting is made to request entry of the password in starting the Check and Service
Process program.
Click the button to make the password effective.
Once the password is input in starting the Check and Service Process program, the
effective setting of the password is automatically cancelled, and entry of the password
is thereafter not requested until the operation "(2) Setting password effective" is
performed.
※ After the above work is done, be sure to carry out the operation "(2) Setting
password".
3 - 36 L4E-BW1360
3.8 Phantom Calibration
(a) Clicking [Calculation] on the Phantom Calibration Process Menu window displays the
following window.
3 - 37 L4E-BW1360
• The above window is an example for Pronto-XE system.
• The following contents are displayed in Remarks column in List View according to
calculation date and time and scan date and time.
"Scan has been made." : Data of which scan date and time are later than
calculation date and time.
(b) By setting parameters for calculation in the left screen of the window, the items related
to the selected calculation parameters are displayed with a check mark in the Selection
column on the List View.
• When more than one check mark is in List View, [Calc. Start] button is enabled.
However, in case the condition "Impossible to Calc." in Remarks column is displayed
with a check mark, [Calc. Start] button is displayed in light color.
(c) Clicking [Calc. Start] starts calculation. While calculation being executed, the
following window is displayed.
When any error occurs during calculation, the following window is displayed.
• If any error occurred, retry the phantom calibration calculation with these
parameters.
3 - 38 L4E-BW1360
(d) Clicking [Abort] button after completion of the calculation or on Error Display window,
the calculation process window is displayed as follows.
• The following contents are displayed in Remarks column for the item with a check
mark in Selection column in List View.
"Calculation aborted : Data for which calculation has not been completed due to
error.
To retry the calculation, click [Redo] button. Then, the screen returns to the step (a).
To close the calculation process, click [Close] button.
To set default parameters for calculation, click [Calculating condition…] button on the
[Settings (S)] menu in the step (a).
3 - 39 L4E-BW1360
• The above window is an example for Pronto-XE system.
On the window of the step (a), set the calculation parameters to be displayed as default.
• If [Use previous calculating parameters.] is checked, the parameters used for the
last calculation are displayed as default.
• If [Set individually.] is checked, check each parameter for phantom diameter, tube
voltage and slice thickness. These parameters are displayed as default.
The parameters set here are kept saved even if the phantom calibration process is
closed.
(a) Clicking [Date Check] button on Phantom Calibration Process Menu window displays
the following window.
3 - 40 L4E-BW1360
• The above window is an example for Pronto-XE system.
• The following contents are displayed in Remarks column in List View depending on
calculation date and time and scan date and time.
"Scan has been made." : Data of which scan date and time are later than
calculation date and time.
Scan and calculation date and time corresponding to each parameter of tube voltage,
slice thickness and phantom diameter highlighted in the left screen of the window are
displayed in List View. By clicking each parameter list, the highlighted item can be
changed.
(c) Change of display order
Pointing the [Priority of display] in [View(V)] menu displays the following menu.
By clicking the desired display order, the display order is reflected in List View.
3 - 41 L4E-BW1360
Example: If [slice → kV → phantom] is clicked:
3 - 42 L4E-BW1360
3.9 Phantom Position
This function is used to position the phantom for Phantom Cal and CT Number Adjustment
scan.
After setting the phantom, make a scan and start this program.
(1) Click [Start] button and point “Programs” and "2. Service".
Then, click [Phantom Position] button.
(2) If any latest scan CT image is available, the image is displayed. In this case, jump to the
step (5). If any newest scan CT image is not available, the following window is displayed.
To designate an image on the patient directory and slice directory, click [Yes(Y)] button.
To end the process, click [Abort(A)] button.
CAUTION
When the program is started within 1 or 2 seconds immediately after scan, the latest
scan image may not be displayed. In this case, restart the phantom position program .
3 - 43 L4E-BW1360
(3) The patient directory is displayed. Select a patient on the directory.
3 - 44 L4E-BW1360
(4) The slice directory window for the selected patient is displayed. Select a desired slice.
3 - 45 L4E-BW1360
(5) The selected image is displayed.
3 - 46 L4E-BW1360
(6) Clicking [Calculation] button displays the following values for phantom displacement.
(7) To display another image, click [Open(O)] button in the [File(F)] menu to return to the
display of the step (3) window. Retry the operation from patient selection.
(8) To display the latest scan CT image, click [Open the latest scan image…] button in the
[File(F)] menu.
If any latest CT image is available, the image is displayed and the process moves to the
step (5).
If any latest CT image is not available, the following window is displayed. To select an
image on the patient and slice directories, click [Yes(Y)] button to return to display of the
step (3) window. Then, retry the operation from patient selection.
NOTE: If this operation is made within 1 ~ 2 seconds immediately after scan, the latest
scan image may not be displayed.
3 - 47 L4E-BW1360
3.10 Screen Copy
3.10.1 Introduction
The screen copy function is used to store the whole screen on the CRT or the selected
window as image file in the FD or MO disk.
As image file mode, bit map or JPEG image can be selected.
Press Alt + Print Screen keys together : Copy of only the selected window
3 - 48 L4E-BW1360
To paste an image onto the screen area, select Edit(E)
and then Paste(P). (Pressing Ctrl + V keys
simultaneously makes the same operation.)
Either FD or MO can be
selected as the drive for saving
image.
3 - 49 L4E-BW1360
Set in this item below the
destination drive to designate
directories. Insert the "/"
mark between directories in
setting.
(Example: Directory:
Date/image
(Example: Image001)
The JEPG image (*.jpg) converts the bit map in the screen to JEPG image and save
it. In the latter mode, image can be saved with a relatively smaller data volume as
compare with the former case.
(Example of Bit map (*.bmp))
Upon inputting all the data necessary for saving, click Save(S) button to start the
saving operation.
(2) To return to the basic screen
To delete the image in the screen area, select Edit (E) and
then Clear (C). (Pressing Delete key makes the same
operation.)
3 - 50 L4E-BW1360
(4) To cut the image on the screen
To cut the image in the screen area, select Edit(E) and then
Cut(C). (Pressing Ctrl+X keys makes the same operation.)
To copy the image in the screen area, select Edit(E) and then
Copy(C). (Pressing Ctrl + C keys together makes the same
operation.)
When closing the Create New or Screen Copy process and if unsaved bit map is in the
screen area, the following message is displayed.
If the directory designated in "Save as" does not exist, the following message is
displayed. To create the directory and save it, select [Yes(Y)], while not to save it,
select [No(N)].
If the file designated in "Save as" already exists, the following message is displayed.
To overwrite save the file, select Yes(Y), while not to save it select No(N).
(Display example)
3 - 51 L4E-BW1360
In the case Bit map is selected as saving format of the image file and the data volume of
the bit map image to be saved exceeds the file capacity of the drive, the following
message is displayed.
In this case, change the saving format to JPEG image (*.jpg) and retry the saving
operation.
To save the image file as bit map, change the disk of the saving drive to the one with a
larger memory capacity.
In the case JPEG image format is selected as the saving format of the image and the
data volume of the JPEG image to be saved exceeds the file capacity of the drive, the
following message is displayed. In such case, change the disk of the saving drive to the
one with a larger memory capacity.
After completion of the saving operation, be sure to carry out the operation "3.7.3 (2)
Setting Password Effective".
3 - 52 L4E-BW1360
3.11 Selection of Cu-Filter
3.11.1 Outline
The default software setting in Pronto system has been made without Cu-filter.
Unless any special instruction is given, do not change the setting.
Unless any special instruction is given, only check the set contents and do not change
the setting.
Click [Start], and select [Programs], [2. Service] and [Cu-Filter Selection].
According to the instruction, select either [Apply] or [Not apply] and click [OK]
button. If the setting is changed, reboot the system and then make scans.
3 - 53 L4E-BW1360
3.12 Entry of Hospital Name
(1) Click [Start] button, and point “Programs” and "2. Service".
Then, click [Hospital Name Entry] button.
The currently registered hospital name is displayed. To change it, input a new hospital
name within 26 half-size characters from the keyboard and click [OK] button.
(3) As the hospital name registered here becomes effective after rebooting the system, make
sure of the hospital name displayed on the image acquired by scan after reboot
3 - 54 97 L4E-BW1360
Section 4 Error Codes List (Appendix)
4-1 L4E-BW1360
Error code Message Description Recovery Procedure
S 09 “Stationary SW is ON” Error occurred in the scanner unit. User: Contact Service
representative.
Service Office: Turn off
Stationary switch.
S 10 “Scanner is not Ready” Error occurred in the scanner unit. User: Retry the operation.
Service office: Retry the
operation. Check Stationary
switch.
S 11 “Scanner is not Ready” Error occurred in the scanner unit. User: Retry the operation.
Service office: Retry the
operation. Check Stationary
switch.
S 12 “Scanner is not Ready” Error occurred in the scanner unit. Retry the operation.
S 15 “Relay Off of Invertor” Error occurred in the scanner unit. Retry the operation.
S 16 “Relay OFF of Invertor” Error occurred in the scanner unit. Retry the operation.
S 17 “22FFB OFF for Tilt” Error occurred in the scanner unit. Retry the operation.
S 18 “21FFB OFF for Invertor” Error occurred in the scanner unit. Retry the operation.
S 19 “Stationary SW is ON” Error occurred in the scanner unit. User: Retry the operation.
Service office: Turn off
Stationary switch.
S 1A “Over Heat Tube” Error occurred in the scanner unit. Retry the operation.
S 1B “Servopack Power is OFF” Error occurred in the scanner unit. Retry the operation.
S 1C “Over Heat L-Tilt” Error occurred in the scanner unit. Keep the system for a while
without operation with power
turned on.
S 1D “Over Heat R-Tilt” Error occurred in the scanner unit. Keep the system for a while
without operation with power
turned on
S 1E “Manual SW is ON” Error occurred in the scanner unit. User: Contact Service office.
Service office: Set Manual
switch to Normal.
S 1F “Emergency” Error occurred in the scanner unit. After making sure of safety,
release the switch.
S 20 “Brake Release is ON” Error occurred in the scanner unit. User: Contact Service office.
Service Office: Turn off Brake
Release switch.
S 27 “Tape SW is ON” Tape switch is actuated. After making sure of safety,
release the switch.
S 28(08) “Miss Sence code:08” Error occurred in the scanner unit. Retry the operation.
S 28(09) “unresolved code:09” Error occurred in the scanner unit. Retry the operation.
S 28(0A) “unresolved code:0A” Error occurred in the scanner unit. Retry the operation.
S 28(0B) “unresolved code:0B” Error occurred in the scanner unit. Retry the operation.
S 28(0C) “unresolved code:0C” Error occurred in the scanner unit. Retry the operation.
S 28(0D) “unresolved code:0D” Error occurred in the scanner unit. Retry the operation.
S 28(0E) “unresolved code:0E” Error occurred in the scanner unit. Retry the operation.
S 28(0F) “unresolved code:0F” Error occurred in the scanner unit. Retry the operation.
S 28(10) “unresolved code:10” Error occurred in the scanner unit. Retry the operation.
S 28(11) “unresolved code:11” Error occurred in the scanner unit. Retry the operation.
S 28(12) “unresolved code:12” Error occurred in the scanner unit. Retry the operation.
S 28(13) “unresolved code:13” Error occurred in the scanner unit. Retry the operation.
S 28(14) “unresolved code:14” Error occurred in the scanner unit. Retry the operation.
S 28(15) “unresolved code:15” Error occurred in the scanner unit. Retry the operation.
S 28(16) “unresolved code:16” Error occurred in the scanner unit. Retry the operation.
S 28(17) “unresolved code:17” Error occurred in the scanner unit. Retry the operation.
S 28(18) “unresolved code:18” Error occurred in the scanner unit. Retry the operation.
S 28(19) “unresolved code:19” Error occurred in the scanner unit. Retry the operation.
4-2 L4E-BW1360
Error code Message Description Recovery Procedure
S 28(1A) “unresolved code:1A” Error occurred in the scanner unit. Retry the operation.
S 28(1B) “unresolved code:1B” Error occurred in the scanner unit. Retry the operation.
S 28(1C) “unresolved code:1C” Error occurred in the scanner unit. Retry the operation.
S 28(1D) “Over Speed Error:1D” Error occurred in the scanner unit. Retry the operation.
S 28(1E) “Under Speed Error:1E” Error occurred in the scanner unit. Retry the operation.
S 28(1F) “unresolved code:1F” Error occurred in the scanner unit. Retry the operation.
S 28(20) “unresolved code:20” Error occurred in the scanner unit. Retry the operation.
S 28(21) “unresolved code:21” Error occurred in the scanner unit. Retry the operation.
S 28(22) “unresolved code:22” Error occurred in the scanner unit. Retry the operation.
S 28(23) “unresolved code:23” Error occurred in the scanner unit. Retry the operation.
S 28(25) “unresolved code:25” Error occurred in the scanner unit. Retry the operation.
S 28(26) “unresolved code:26” Error occurred in the scanner unit. Retry the operation.
S 28(27) “unresolved code:27” Error occurred in the scanner unit. Retry the operation.
S 28(28) “unresolved code:28” Error occurred in the scanner unit. Retry the operation.
S 28(29) “unresolved code:29” Error occurred in the scanner unit. Retry the operation.
S 28(2A) “unresolved code:2A” Error occurred in the scanner unit. Retry the operation.
S 28(2B) “unresolved code:2B” Error occurred in the scanner unit. Retry the operation.
S 28(2C) “unresolved code:2C” Error occurred in the scanner unit. Retry the operation.
S 28(2D) “unresolved code:2D” Error occurred in the scanner unit. Retry the operation.
S 28(30) “unresolved code:30” Error occurred in the scanner unit. Retry the operation.
S 28(31) “unresolved code:31” Error occurred in the scanner unit. Retry the operation.
S 28(32) “unresolved code:32” Error occurred in the scanner unit. Retry the operation.
S 28(33) “unresolved code:33” Error occurred in the scanner unit. Retry the operation.
S 28(34) “unresolved code:34” Error occurred in the scanner unit. Retry the operation.
S 28(35) “unresolved code:35” Error occurred in the scanner unit. Retry the operation.
S 28(36) “unresolved code:36” Error occurred in the scanner unit. Retry the operation.
S 28(37) “unresolved code:37” Error occurred in the scanner unit. Retry the operation.
S 28(38) “unresolved code:38” Error occurred in the scanner unit. Retry the operation.
S 28(39) “unresolved code:39” Error occurred in the scanner unit. Retry the operation.
S 28(3A) “unresolved code:3A” Error occurred in the scanner unit. Retry the operation.
S 28(3B) “unresolved code:3B” Error occurred in the scanner unit. Retry the operation.
S 28(3C) “unresolved code:3C” Error occurred in the scanner unit. Retry the operation.
S 28(3D) “unresolved code:3D” Error occurred in the scanner unit. Retry the operation.
S 28(3E) “unresolved code:3E” Error occurred in the scanner unit. Retry the operation.
“unresolved code:41” Error occurred in the scanner unit. Retry the operation.
“unresolved code:42” Error occurred in the scanner unit. Retry the operation.
S 28(43) “Invertor Thermal Relay ON:43” Error occurred in the scanner unit. Keep the system for a while
without operation with power
turned on.
S 28(44) “Invertor Trip ON:44” Error occurred in the scanner unit. Power up the system again.
S 28(45) “Front Cover Open:45” Error occurred in the scanner unit. User: Contact Service office.
Service Office: Close the
scanner cover.
S 28(46) “Side Cover Open:46” Error occurred in the scanner unit. User: Contact Service office.
Service Office: Close the
scanner cover.
S 28(54) “DPRAM Commu. Init Error:54” Error occurred in the scanner unit. Retry the operation.
S 28(57) “Scan Seq. Timeout:57” Error occurred in the scanner unit. Retry the operation.
S 28(58) “Timer Counter not Synchronize:58” Error occurred in the scanner unit. Retry the operation.
S 28(59) “Bord Cover Open:59” Error occurred in the scanner unit. Retry the operation.
S 28(60) “unresolved code:60” Error occurred in the scanner unit. Retry the operation.
S 28(71) “Communication Error to IPU:71” Error occurred in the scanner unit. Retry the operation.
S 28(72) “Illegal Command:72” Error occurred in the scanner unit. Retry the operation.
S 28(73) “Illegal Seq. Number:73” Error occurred in the scanner unit. Retry the operation.
S 28(74) “Parameter Error:74” Error occurred in the scanner unit. Retry the operation.
4-3 L4E-BW1360
Error code Message Description Recovery Procedure
S 28(76) “The X-ray Abort:76” Error occurred in the scanner unit. Retry the operation.
S 28(77) “1 Scan Timeout:77” Error occurred in the scanner unit. Retry the operation.
S 28(78) “Move Table Timeout:78” Error occurred in the scanner unit. Retry the operation.
S 28(79) “Parameter Error from IPU:79” Error occurred in the scanner unit. Retry the operation.
S 30 “Command Error to Scanner CPU” Error occurred in the scanner unit. Retry the operation.
S 50 “Check Seq. Now” Error occurred in the scanner unit. Check Sequence is in
operation. Do not execute
other operation.
S 60 “Communication Error to Scanner Error occurred in the scanner unit. Retry the operation.
CPU”
X 10 “Timing Error” Error occurred in the scanner unit. Retry the operation.
X 11 “Rotor not Standby” Error occurred in the scanner unit. Retry the operation.
X 14 “Commu.Error” Error occurred in the scanner unit. Retry the operation.
X 17 “Door Open” Door of The X-ray room is open. Close the door.
X 1A “Over Heat Tube” Error occurred in the X-ray generation Keep the system for a while
system. without operation with power
turned on
X 1F “Emergency” Emergency button was pressed. After making sure of safety,
release the button.
X 21 “Rotor Error Exposure” Error occurred in the X-ray generation Retry the operation.
system.
X 22 “Primary Error (REC or INV)” Error occurred in the X-ray generation Retry the operation.
system.
X 23 “Focus Error Miss Selection” Error occurred in the X-ray generation Retry the operation.
system.
X 24 “Breaker OFF Control Circuit” Error occurred in the X-ray generation Power up the scanner unit
system. again.
X 25 “Memory Error” Error occurred in the X-ray generation Retry the operation.
system.
X 26 “Arcing” Error occurred in the X-ray generation Retry the operation.
system.
X 27 “Filament Power OFF” Error occurred in the X-ray generation Power up the scanner unit
system. again.
X 28 “High Voltage Trans.r Bellows SW ON” Error occurred in the X-ray generation Retry the operation.
system.
X 29 “mA Error Neutral” Error occurred in the X-ray generation Retry the operation.
system.
X 2A “Tube Volt. Error” Error occurred in the X-ray generation Retry the operation.
system.
X 2B “Filament Error” Error occurred in the X-ray generation Retry the operation.
system.
X 2C “Backup Timer Error” Error occurred in the X-ray generation Retry the operation.
system.
X 2D “ADC Error” Error occurred in the X-ray generation Retry the operation.
system.
X 2E “Miss Radiation” Error occurred in the X-ray generation Retry the operation.
system.
X 30 “Command Error unresolved” Error occurred in the X-ray generation Retry the operation.
system.
X 32 “Command Error” Error occurred in the X-ray generation Retry the operation.
system.
X 3C “Collimator Error” Error occurred in the X-ray generation Retry the operation.
system.
X 3D “ADC Error (Turn plate)” Error occurred in the X-ray generation Retry the operation.
system.
4-4 L4E-BW1360
Error code Message Description Recovery Procedure
X 3E “Ref. Error (Turn plate)” Error occurred in the X-ray generation Retry the operation.
system.
X 3F “Command Error Emergency” Error occurred in the X-ray generation Retry the operation.
system.
X 60 “The X-ray Error not Commu.(Turn Error occurred in the X-ray generation Retry the operation.
plate)” system.
X 61 “The X-ray Error not Commu.(WB)” Error occurred in the X-ray generation
system.
T 00 “Table Error” Clear button was pressed during scan Retry the operation.
operation. Positional information may be
inaccurate.
T 02 “Index Move Now” Error occurred in the table control system. Retry the operation.
T 03 “Gantry Tilt Now” Error occurred in the table control system Retry the operation.
T 04 “Scan Seq. Now” Error occurred in the table control system Retry the operation.
T 0A “Table Limit Inward” Tabletop reached the limit in inward direction. Make sure of table move
direction and retry the
operation.
T 0B “Table Limit Outward” Tabletop reached the limit in outward Make sure of table move
direction. direction and retry the
operation.
T 0C “Tilt Limit AF/PH” Gantry tilt angle reached the limit of forward Make sure of the set gantry
tilt angle tilt angle and retry the
operation
T 0D “Tilt Limit AH/PH” Gantry tilt angle reached the limit of backward Make sure of the set gantry
tilt angle tilt angle and retry the
operation
T 0E “Table Error not run” Error occurred in the table. Retry the operation.
T 10 “Irregular OP.” Table is in operation from the gantry. After making sure that the
operation from the gantry has
completed, retry the
operation.
T 14 “Manual OP. done in Scanning” Error occurred in the table. Retry the operation.
T 1A “Latch & Table Limit Inward” Table is latched. Unlatch the table.
T 1B “Latch & Table Limit Outward” Table is latched. Unlatch the table.
T 1E “Table Latch Error” Table is latched. Unlatch the table.
T 1F “Emergency” Emergency switch was pressed. After making sure of safety,
release the switch.
T 20 “Table Speed Error” Error occurred in the table. Retry the operation.
T 27 “Tape SW ON” Error occurred in the table. Retry the operation.
T 30 “Command Error” Error occurred in the table. Retry the operation.
T 50 “Check Seq. Now” Error occurred in the table. Retry the operation.
T 60 “Table Error not Cummunicate(Main1)” Error occurred in the table. Retry the operation.
T 61 “Table Error not Cummunicate(Main2)” Error occurred in the table. Retry the operation.
T 62 “Scanner-Table Error Cummunicate” Error occurred in the table. Retry the operation.
T 63 “Table CPU Error Cummunicate” Error occurred in the table. Retry the operation.
T 64 “Scanner-Table Error Retry Over” Error occurred in the table. Retry the operation.
T 65 “Table NACK Recv.” Error occurred in the table. Retry the operation.
T 66 “Move Command Retry Over” Error occurred in the table. Retry the operation.
T 67 “Table Reset Error Communicate” Error occurred in the table. Retry the operation.
T 68 “Scanner-Table Error Cummunicate2” Error occurred in the table. Retry the operation.
T 69 “Scanner-Table Error Retry Over2” Error occurred in the table. Retry the operation.
SD00 “Scanner Error” Error occurred in the scanner control system.
SD01 “Scanner Driver Open Error” Error occurred in the scanner control system. Retry the operation.
SD02 “RS-232C Open Condition Read Error” Error occurred in the scanner control system. Retry the operation.
SD03 “RS-232C Open Condition Set Error” Error occurred in the scanner control system. Retry the operation.
SD04 “RS-232C Open Timeout Set Error” Error occurred in the scanner control system. Retry the operation.
4-5 L4E-BW1360
Error code Message Description Recovery Procedure
SD05 “Status Read Timeout” Error occurred in the scanner control system. Retry the operation.
SD06 “Status(Table) Read Timeout” Error occurred in the scanner control system. Retry the operation.
SD07 “Scanner Driver Init Error” Error occurred in the scanner control system. Retry the operation.
SD08 “Scanner Driver Close Error” Error occurred in the scanner control system. Retry the operation.
SD09 “RS-232C Close Timeout Set Error” Error occurred in the scanner control system. Retry the operation.
SD0A “Scanner Abnomal” Error occurred in the scanner control system. Retry the operation.
SD11 “ENQ Recv Timeout” Error occurred in the scanner control system. Retry the operation.
SD12 “ENQ Recv Error” Error occurred in the scanner control system. Retry the operation.
SD13 “ENQ ClearCommError” Error occurred in the scanner control system. Retry the operation.
SD14 “1st ACK Send Timeout” Error occurred in the scanner control system. Retry the operation.
SD15 “1st ACK Send Error” Error occurred in the scanner control system. Retry the operation.
SD16 “Body Recv Error” Error occurred in the scanner control system. Retry the operation.
SD17 “Body ClearCommError” Error occurred in the scanner control system. Retry the operation.
SD18 “Body ClearCommError Timeout (3 Error occurred in the scanner control system. Retry the operation.
char)”
SD19 “Body Recv Timeout (3 char)” Error occurred in the scanner control system. Retry the operation.
SD1A “Body ClearCommError Timeout (N Error occurred in the scanner control system. Retry the operation.
char)”
SD1B “Body Recv Timeout (N char)” Error occurred in the scanner control system. Retry the operation.
SD1C “2nd ACK Send Timeout” Error occurred in the scanner control system. Retry the operation.
SD1D “2nd ACK Send Error” Error occurred in the scanner control system. Retry the operation.
SD21 “ENQ Send Timeout” Error occurred in the scanner control system. Retry the operation.
SD22 “ENQ Send Error” Error occurred in the scanner control system. Retry the operation.
SD23 “1st ACK Recv Timeout” Error occurred in the scanner control system. Retry the operation. Make
sure if the gantry is not
powered down.
SD24 “1st ACK Recv Error” Error occurred in the scanner control system. Retry the operation.
SD25 “1st ACK ClearCommError” Error occurred in the scanner control system. Retry the operation.
SD26 “Body Send Timeout” Error occurred in the scanner control system. Retry the operation.
SD27 “Body Send Error” Error occurred in the scanner control system. Retry the operation.
SD28 “2nd ACK Recv Timeout” Error occurred in the scanner control system. Retry the operation.
SD29 “2nd ACK Recv Error” Error occurred in the scanner control system. Retry the operation.
SD2A “2nd ACK ClearCommError” Error occurred in the scanner control system. Retry the operation.
SD2B “NACK Recv 3 times” Error occurred in the scanner control system. Retry the operation.
SD30 “Check Sum Error” Error occurred in the scanner control system. Retry the operation.
SD31 “Scanner Invalid Code” Error occurred in the scanner control system. Retry the operation.
SD32 “Gantry power is OFF” Error occurred in the scanner control system. Retry the operation. Make
sure if the gantry is not
powered down.
S 1F “Emergency” Emergency switch was pressed. After making sure of safety,
release the switch.
SD34 “Abnormal Table Position” Error occurred in the scanner control system. Retry the operation.
SD35 “Table Pos. is not expected” Error occurred in the scanner control system. Retry the operation.
SD36 “Gantry Ang. is not expected” Error occurred in the scanner control system. Retry the operation.
SD37 “Table not Move” Error occurred in the scanner control system. Retry the operation.
SD38 “TapeSW ON Error” Error occurred in the scanner control system. Retry the operation.
SD39 “In Scanner Wakeup” Scanner unit is being started up. Wait for a while.
SD40 “Parallel Port Open Error” Error occurred in the scanner control system. Retry the operation.
SD41 “Parallel Port Close Error” Error occurred in the scanner control system. Retry the operation.
SD42 “Set LED Error” Error occurred in the scanner control system. Retry the operation.
SD43 “Set LED Para. Error” Error occurred in the scanner control system. Retry the operation.
SD44 “Start LED Error” Error occurred in the scanner control system. Retry the operation.
SD45 “Start LED Para. Error” Error occurred in the scanner control system. Retry the operation.
SD46 “DeadMan Set Error” Error occurred in the scanner control system. Retry the operation.
SD47 “DeadMan Unset Error” Error occurred in the scanner control system. Retry the operation.
4-6 L4E-BW1360
Error code Message Description Recovery Procedure
SD48 “The X-ray Soft Switch On Error” Error occurred in the scanner control system. Retry the operation.
SD49 “The X-ray Soft Switch Off Error” Error occurred in the scanner control system. Retry the operation.
SD4A “LED Flick Error” Error occurred in the scanner control system. Retry the operation.
SD4B “LED Flick Para. Error” Error occurred in the scanner control system. Retry the operation.
SD4C “CPU Ready On Error” Error occurred in the scanner control system. Retry the operation.
SD4D “CPU Ready Off Error” Error occurred in the scanner control system. Retry the operation.
SD4E “Scanner ResetOn Error” Error occurred in the scanner control system. Retry the operation.
SD4F “Scanner ResetOff Error” Error occurred in the scanner control system. Retry the operation.
SD61 “Send Command Parameter Error” Error occurred in the scanner control system. Retry the operation.
SD62 “Send Command Device Error” Error occurred in the scanner control system. Retry the operation.
SD63 “Send Command Scanner Ready Error occurred in the scanner control system. Retry the operation.
Error”
SD64 “Send Command Other Error” Error occurred in the scanner control system. Retry the operation.
SD65 “Recv Command Device Error” Error occurred in the scanner control system. Retry the operation.
SD66 “Recv Command NoData” Error occurred in the scanner control system. Retry the operation.
SD67 “Recv Command Other Error” Error occurred in the scanner control system. Retry the operation.
SD68 “Recv Command Open Event Error” Error occurred in the scanner control system. Retry the operation.
SD69 “Recv Command Clear Event Error” Error occurred in the scanner control system. Retry the operation.
SD6A “Recv Command WaitObject Timeout” Error occurred in the scanner control system. Retry the operation.
SD6B “Send Command Timeout Error” Error occurred in the scanner control system. Retry the operation.
SD6C “Invalid TableData Code” Error occurred in the scanner control system. Retry the operation.
SD6D “Invalid TableData Data” Error occurred in the scanner control system. Retry the operation.
OT01 “Process Timeout” Error occurred in the scanner control system. Retry the operation.
OT02 “Offset Scan Timeout” Error occurred in the data acquisition system. Retry the operation.
OT03 “Scan Start Timeout” Error occurred in the data acquisition system. Retry the operation.
OT04 “Scan End Timeout” Error occurred in the data acquisition system. Retry the operation.
OT05 “HU Exceeded” Error occurred in the scanner control system. Retry the operation.
OT06 “Generator Overload” .
OT07 “HU calc Error” Error occurred in the scanner control system. Retry the operation.
OT08 “Scanner Commu. Check Timeout” Error occurred in the scanner control system. Retry the operation.
OT09 “Scanner Rot. Time Check (CW) Error occurred in the scanner control system. Retry the operation.
Timeout”
OT0A “Scanner Rot. Time Check (CCW) Error occurred in the scanner control system. Retry the operation.
Timeout”
OT0B “Table Moving Speed Check Timeout” Error occurred in the scanner control system. Retry the operation.
OT0C “Status Read Timeout” Error occurred in the scanner control system. Retry the operation.
OT0D “Move Table Timeout” Error occurred in the scanner control system. Retry the operation.
OT0E “Tilt Gantry Timeout” Error occurred in the scanner control system. Retry the operation.
OT0F “Emergency Release Timeout” Error occurred in the scanner control system. Retry the operation.
OT10 “Tube Return Timeout” Error occurred in the scanner control system. Retry the operation.
OT11 “Standby Timeout” Error occurred in the scanner control system. Retry the operation.
OT12 “Raw Data Timeout” Error occurred in the scanner control system. Retry the operation.
OT13 “Image Recon Timeout” Error occurred in the scanner control system. Retry the operation.
OT14 “RotStop Timeout” Error occurred in the scanner control system. Retry the operation.
OT15 “SeqStart Timeout” Error occurred in the scanner control system. Retry the operation.
OP01 File space is insufficient. Reset appropriate parameter.
OP02 Generator is overloaded. Change scan Reset appropriate parameter
parameter.
OP03 Gantry tilt angle has exceeded the limit value. Reset appropriate parameter
OP03 Set gantry angle at 0º. Reset appropriate parameter
OP04 Heat storage has exceeded the max. HU. Reset appropriate parameter
Change scan parameter.
OP05 Number of scans is too many. Reset appropriate parameter
OT20 “AvScanPn.DB Open Error 1” Auto-voice data file cannot be opened. Retry the operation.
OT21 “AvScanPn.DB Open Error 2” Auto-voice data file cannot be opened. Retry the operation.
4-7 L4E-BW1360
Error code Message Description Recovery Procedure
OT22 “AvScanPn.DB Open Error 3” Auto-voice data file cannot be opened. Retry the operation.
OT23 “AvScngmPn.DB Open Error 1” Auto-voice data file cannot be opened. Retry the operation.
OT24 “AvScngmPn.DB Open Error 2” Auto-voice data file cannot be opened. Retry the operation.
OT25 “AvScngmPn.DB Open Error 3” Auto-voice data file cannot be opened. Retry the operation.
OT26 “AvVd.DB Open Error 1” Auto-voice data file cannot be opened. Retry the operation.
OT27 “AvVd.DB Open Error 2” Auto-voice data file cannot be opened. Retry the operation.
OT28 “AvVd.DB Open Error 3” Auto-voice data file cannot be opened. Retry the operation.
OT29 “Scan PnNo. Error” Error occurred in auto-voice data. Retry the operation.
OT2A “Scngm PnNo. Error” Error occurred in auto-voice data. Retry the operation.
OT30 “Scan StandPh No. Error” Error occurred in access to voice data file Retry the operation.
OT31 “Scan StandPh No. Error” Error occurred in access to voice data file Retry the operation.
OT32 “StandPh not Exist” Error occurred in access to voice data file Retry the operation.
OT33 “StandPh Open Error” Error occurred in access to voice data file Retry the operation.
OT34 “StandPh Play Error” Error occurred in access to voice data file Retry the operation.
OT35 “StandPh Close Error” Error occurred in access to voice data file Retry the operation.
OT36 “StandPh Wait Timeout” Error occurred in access to voice data file Retry the operation.
OT37 Error occurred in access to voice data file Retry the operation.
OT38 “Scan StartPh No. Error” Error occurred in access to voice data file Retry the operation.
OT39 “Scan StartPh No. Error” Error occurred in access to voice data file Retry the operation.
OT3A “StartPh not Exist” Error occurred in access to voice data file Retry the operation.
OT3B “StartPh Open Error” Error occurred in access to voice data file Retry the operation.
OT3C “StartPh Play Error” Error occurred in access to voice data file. Retry the operation.
OT3D “StartPh Close Error” Error occurred in access to voice data file. Retry the operation.
OT3E “StartPh Wait Timeout” Error occurred in access to voice data file. Retry the operation.
OT3F Error occurred in access to voice data file. Retry the operation.
OT40 “Scan XoffPh No. Error” Error occurred in access to voice data file. Retry the operation.
OT41 Error occurred in access to voice data file. Retry the operation.
OT42 “XoffPh not Exist” Error occurred in access to voice data file. Retry the operation.
OT43 “XoffPh Open Error” Error occurred in access to voice data file. Retry the operation.
OT44 “XoffPh Play Error” Error occurred in access to voice data file. Retry the operation.
OT45 “XoffPh Close Error” Error occurred in access to voice data file. Retry the operation.
OT46 “XoffPh Wait Timeout” Error occurred in access to voice data file. Retry the operation.
OT47 Error occurred in access to voice data file. Retry the operation.
OT48 “Scan IntSeqPh No. Error” Error occurred in access to voice data file. Retry the operation.
OT49 Error occurred in access to voice data file. Retry the operation.
OT4A “IntSeqPh not Exist” Error occurred in access to voice data file. Retry the operation.
OT4B “IntSeqPh Open Error” Error occurred in access to voice data file. Retry the operation.
OT4C “IntSeqPh Play Error” Error occurred in access to voice data file. Retry the operation.
OT4D “IntSeqPh Close Error” Error occurred in access to voice data file. Retry the operation.
OT4E “IntSeqPh Wait Timeout” Error occurred in access to voice data file. Retry the operation.
OT4F Error occurred in access to voice data file. Retry the operation.
OT50 “Scan EndPh No. Error” Error occurred in access to voice data file. Retry the operation.
OT51 “Scan EndPh No. Error” Error occurred in access to voice data file. Retry the operation.
OT52 “EndPh not Exist” Error occurred in access to voice data file. Retry the operation.
OT53 “EndPh Open Error” Error occurred in access to voice data file. Retry the operation.
OT54 “EndPh Play Error” Error occurred in access to voice data file. Retry the operation.
OT55 “EndPh Close Error” Error occurred in access to voice data file. Retry the operation.
OT56 “EndPh Wait Timeout” Error occurred in access to voice data file. Retry the operation.
OT60 “FileRead Error” Error occurred in reading file. Retry the operation.
In accordance with the agreement between Hitachi Medical Corporation and Microsoft Corporation,
the password for administrator log-on must be disclosed when it is required by the Pronto system
administrator.
However, for the purpose of preventing both surreptitious use and malfunction of the system, the
disclosure must not be for all the users but should be limited to only the administrator.
If any system administrator has not been appointed, the user is requested to appoint the system
administrator and to disclose the password only to the administrator.
Then, thoroughly explain the following responsibilities to the administrator:
2. Hitachi shall in no event be liable for malfunction of the system which may be caused by
operation of the system software (including installation of commercially available software).
3. Any work to correct such malfunction (including reinstallation of the system software) shall
be born by the user.
Password: ct3g1g
6.1 Introduction
Event Viewer is the program for monitoring system events which are incorporated in
Windows NT as standard.
Using the Event Viewer, detailed data of system startup and errors occurred during booting
the Pronto system can be analyzed.
6.2 Startup
After booting the system, select [Start] of the task bar → [Programs] → [Administrative
Tools] → [Event Viewer] to start up the Event Viewer program. The list of the events that
have occurred is displayed.
6-1 L4E-BW1360
6.3 Detailed View of Event
Select the event of which details is to be displayed, and double-click the mouse or [View] →
[Detail].
Then, the following window displaying detailed data is displayed.
6.5 Notes
For details of Event Viewer, refer to the commercially available Instruction Manual of
Windows NT.