You are on page 1of 8

N T SC S y s t e m s

Te l e v i s i o n M e a s u r e m e n t s
Contents

Preface .............................. 3 III. NONLINEAR DISTORTIONS . . . . . .41


Differential Phase . . . . . . . . . . . . . . . .42
Good Measurement Practices ............... 4
Differential Gain . . . . . . . . . . . . . . . .46
EQUIPMENT REQUIREMENTS . . . . . . . . 4
Luminance Nonlinearity . . . . . . . . . .50
CALIBRATION . . . . . . . . . . . . . . . . . . . . . 6
Chrominance Nonlinear Phase . . . . . . . . 52
INSTRUMENT CALIBRATION . . . . . . . . 6
Chrominance Nonlinear Gain . . . . . . . . 53
DEMODULATED RF SIGNALS . . . . . . . . 7
Chrominance-to-Luminance
TERMINATION . . . . . . . . . . . . . . . . . . . . 7
Intermodulation . . . . . . . . . . . . . . . .54
STANDARDS AND
Transient Gain Distortion . . . . . . . . . .56
PERFORMANCE GOALS . . . . . . . . . . . . . 7
Dynamic Gain Change . . . . . . . . . . . .57

Waveform Distortions and


Measurement Methods ............... 8 IV. NOISE MEASUREMENT . . . . . . . . .58
I. VIDEO AMPLITUDE AND Signal-to Noise Ratio . . . . . . . . . . . . .59
TIME MEASUREMENTS ........ 8
Amplitude Measurements . . . . . . . . . 9 V. TRANSMITTER MEASUREMENTS . . 61
Time Measurements . . . . . . . . . . . . . .12 ICPM . . . . . . . . . . . . . . . . . . . . . . . . .62
SCH Phase . . . . . . . . . . . . . . . . . . . . .15 Depth of Modulation . . . . . . . . . . . . .64
GLOSSARY OF TELEVISION TERMS . . . . . . . . . . . . 65
II. LINEAR DISTORTIONS . . . . . . . . . .18 APPENDICES
Chrominance-to-Luminance APPENDIX A - NTSC COLOR BARS . . . .68
Gain and Delay . . . . . . . . . . . . . . . .19
APPENDIX B - SINE-SQUARED PULSES 70
Short Time Distortion . . . . . . . . . . . .24
APPENDIX C - RS-170A . . . . . . . . . . . . . .71
Line Time Distortion . . . . . . . . . . . . .26
APPENDIX D - FCC 73.699, FIGURE 6 . .73
Field Time Distortion . . . . . . . . . . . . .28
Long Time Distortion . . . . . . . . . . . . .30
Frequency Response . . . . . . . . . . . . .31
Group Delay . . . . . . . . . . . . . . . . . . . .36
K Factor Ratings . . . . . . . . . . . . . . . . .38
Preface

To characterize television system New instruments, test signals, A basic knowledge of video is
performance, an understanding and measurement procedures are assumed and a glossary of com-
of signal distortions and mea - introduced as television test monly used terms is included as
surement methods as well as and measurement technology a refresher and to introduce new
proper instrumentation is needed. evolves. This booklet encom- concepts. This booklet does not
This booklet provides informa- passes both traditional measur e- provide detailed instructions on
tion on television test and mea- ment techniques and newer how to use particular instruments.
surement practices and serves as methods. After a discussion of The basics of waveform monitor
a comprehensive reference on good measurement practices, and vectorscope operation are
methods of quantifying signal five general categories of assumed. Consult the instrument
distortions. This publication television measurements manuals for specific operating
deals with NTSC composite ana - are addressed: instructions.
log signals. Analog component, I. Video Amplitude and Time
digital composite and compo- Measurements
nent, and HDTV measurements
are outside its scope. II. Linear Distortions
III.Nonlinear Distortions
IV. Noise
V. Transmitter Measurements

3
Good Measurement Practices

EQUIPMENT REQUIREMENTS good as the measurement toler - supporting measurements in


Television system performance ance. For example, differential other composite and analog com -
is evaluated by sending test sig- gain measurement to 1% accura - ponent formats. This eliminates
nals with known characteristics cy should be made with a gener- the need for additional signal
through the signal path. The sig- ator having 0.5% or less differ - generation equipment where
nals are then observed at the ential gain distortion. there is the requirement for mea-
output (or at intermediate Television equipment and sys- surements in multiple formats.
points) to determine whether or tem performance is generally For synchronization of the
not they are being accurately assessed on either an out-of-ser - equipment under test, a black
transferred through the system. vice or in-service basis. In broad- burst reference signal is provided
Two basic types of television test cast television applications, by the TG2000 mainframe. For
and measurement equipment are measurements must often be applications requiring the test
required to perform these tasks. made during regular broadcast signal source be synchronous
Test signal generators provide hours or on an in-service basis. with existing equipment, the
the stimulus and specialized This requires a generator capable AGL1 Analog Genlock module
oscilloscopes, known as wave- of placing test signals within the provides the interface needed to
form monitors and vectorscopes, vertical blanking interval (VBI) lock the TG2000 to an external
provide the tools for evaluating of the television program signal. black burst reference signal.
the response. Out-of-service measurements, For in-service measurements, the
those performed on other than Tektronix VITS200 Generator
Test Signal Generators. Television an in-service basis, may be made and Inserter is the recommended
signal generators provide a wide with any suitable full field test product. The VITS200 provides
variety of test and synchroniza - signal generator. a full complement of NTSC test
tion signals. Two key criteria in signals and high degree of flexi-
For out-of-service measurements,
selection of a test signal genera- bility in placement of these
the Tektronix TG2000 Signal
tor for precision measurements signals within the VBI. Signal
Generation Platform with the
are signal complement and accu - accuracy is adequate for most
AVG1 and AGL1 modules is the
racy. The generator should transmission and transmitter
recommended product. The
provide all of the test signals to measurement requirements.
AVG1 Analog Video Generator
support the required measure-
provides comprehensive signal Both the TG2000 and VITS200
ments and the signal accuracy
sets and sufficient accuracy for fully support the measurement
must be better than the toler-
virtually all measurement capabilities of the 1780R and
ances of the measurements to be
requirements. The AVG1 is also VM700 Series Video
made. If possible, the generator
a multiformat unit capable of Measurement Sets.
accuracy should be twice as

4
Waveform Monitors and booklet. Many varieties of wave -
Vectorscopes. The instruments form monitors and vectorscopes
used to evaluate a system's are available today but the
response to test signals make up majority of them are not intend-
the second major category of ed for precision measurement
television test and measurement applications. Most vectorscopes,
equipment. Although some mea- for example, do not have preci-
surements can be performed sion differential phase and gain
with a general purpose oscillo - measurement capabilities.
scope, a waveform monitor is The recommended products for
generally preferred in television precision measurement applica-
facilities. Waveform monitors tions are the Tektronix 1780R
provide TV triggering capabili- and the VM700T. Most of the
ties and video filters that allow measurement procedures in this
evaluation the chrominance and booklet are based on these
luminance portions of the signal instruments.
independently. Most models also A waveform monitor display of color bars.
have a line selector for looking The 1780R provides waveform
at signals on individual lines. monitor and vectorscope func-
tions as well as many special-
A vectorscope, which demodu-
ized measurement features and
lates the signal and displays R-Y
modes that simplify complex
versus B-Y, is another important
measurements.
test and measurement tool. With
a vectorscope, the chrominance The VM700T is an automated
portion of the signal can be measurement set with results
accurately evaluated. available in numeric and graphic
form. Waveform and vector dis-
When selecting a waveform
plays, similar to those of tradi-
monitor and vectorscope, care-
tional waveform monitors and
fully evaluate the feature sets
vectorscopes operating in line
and specifications to make sur e
select mode, are also provided.
they will meet present and
The VM700T Measure mode pr o-
future needs. This is particularly
vides unique displays of mea -
true if making accurate measur e-
surement results, many of which
ments of all signal parameters
are presented in this book. A vectorscope display of color bars.
and distortions described in this

5
CALIBRATION INSTRUMENT CONFIGURATION
Most instruments are quite Most of the functions on analog
stable over time, however, it is waveform monitor and vec-
good practice to verify equip - torscope front panels are fairly
ment calibration prior to ever y straightforward and have obvious
measurement session. Many applications in measurement
instruments have internally gen- procedures. A few controls,
erated calibration signals that however, might need a bit
facilitate this process. In the more explanation.
absence of a calibrator, or as an
additional check, a test signal DC Restorer. The basic function of
directly out of a high quality the DC restorer in a waveform
generator makes a good substi- monitor is to clamp one point of
tute. Calibration procedures vary the video waveform to a fixed
from instrument to instrument DC level. This ensures that the
and the manuals contain display will not move vertically
The 1780R waveform calibrator. detailed instructions. with changes in signal amplitude
or APL (Average Picture Level).
Analog CRT-based instruments
such as the 1780R have a speci- Some instruments offer a choice
fied warm up time, typically of slow and fast DC restorer
20 or 30 minutes. Turn the speeds. The slow setting is used
instrument on and allow it to to measure hum or other low fre-
operate for at least that long quency distortions. The fast
before checking the calibration setting removes hum from the
and performing measurements. display so it will not interfere
This ensures that the measur e- with other measurements. Back
ment instrumentation will porch is the most commonly
have little or no effect on the used clamp point, but sync tip
measurement results. clamping has some applications
at the transmitter.
Computer-based instruments
such as the VM700T also specify
a warm up time but the operator
does not need to verify or adjust
The 1780R vectorscope calibration oscillator.
the calibration settings. The
VM700T will automatically cali-
brate itself when it is turned on
and will continue to do so peri-
odically during operation. For
best results, wait 20 or 30 min-
utes after initial turn-on befor e
making any measurements.

6
AFC/Direct. This control provides 100% bars may overload the the most frequently used, ar e
selection of the method of trig- transmitter. However, some test reproduced in Appendices C
gering the waveform monitor signal generators produce both. and D of this booklet. Both docu-
horizontal sweep. The ramp that It is important to know what ments define blanking and
produces the horizontal sweep is amplitude color bar is being synchronizing signal parameters.
always synchronous with the H used and to select the corre- RS-170A includes references to
or V pulses of the reference sponding gain setting on the SCH phase and is generally used
video and is started either by the vectorscope. Otherwise chromi- in studio environments. The
pulses themselves (Direct) or by nance gain can easily be FCC diagram is used to verify
their average (AFC). misadjusted. the quality of transmitted sig -
In the direct mode, the video nals. Use them as a reference but
sync pulses directly trigger the DEMODULATED RF SIGNALS exercise caution in assuming
waveform monitor's horizontal All of the baseband measurements that compliance with these stan -
sweep. The direct setting should discussed in this booklet may dards is mandatory or that com-
be used to remove the effects of also be made on demodulated pliance is sufficient to ensur e
time base jitter from the display RF signals. It is important, how- signal quality.
in order to evaluate other para- ever, to eliminate the demodula- Acceptable levels of distortion
meters. Since a new trigger point tor itself as a possible source of are usually determined subjec-
is established for each sweep, distortion. Measurement-quality tively but a number of organiza-
line-to-line jitter is not visible in instruments such as the Tektronix tions publish documents that
this mode. TV1350 and 1450 Television provide recommended limits.
Demodulators will eliminate the These standards, which include
In the AFC (Automatic Frequency
likelihood that the demodulator EIA-250-C, are frequently edited
Control) mode, a phase-locked
is introducing distortion. and revised. Each facility ulti -
loop generates pulses that repr e-
sent the average timing of the mately needs to determine its
TERMINATION own performance goals, however,
sync pulses. These averaged
pulses are used to trigger the Improper termination is a ver y these documents can provide
sweep. The AFC mode is useful common source of operator er ror some good guidelines.
for making measurements in the and frustration. Always make While there is usually agreement
presence of noise as the effects sure the signal under measur e- about the nature of each distor-
of noise-induced horizontal jitter ment is terminated with a 75 Ohm tion, definitions for expressing
are removed from the display. terminator in one location. It is the magnitude of the distortion
generally best to terminate at the vary considerably from standard
The AFC mode is also useful for final piece of equipment in the
evaluating the amount of time to standard. A number of ques-
signal path. tions should be kept in mind. Is
base jitter in a signal. The leading
edge of sync will appear wide The quality of the terminator is the measurement absolute or rel -
(blurred) if much time base jitter also important, particularly ative? If it is relative, what is the
is present. This method is very when measuring very small dis - reference? Under what conditions
useful for comparing signals or tortions. Select a terminator with is the reference established? Is
for indicating the presence of jit- the tightest practical tolerance as the peak-to-peak variation or the
ter but be cautious about actually incorrect termination impedance largest single deviation to be
trying to measure it. The band- can cause amplitude errors, fre- quoted as the distortion?
width of the AFC phase-locked quency response problems, and A misunderstanding of any one
loop also affects the display. pulse distortions. Terminators of these issues can seriously
in the 1/2% to 1/4% tolerance affect measurement results so it
75%/100% Bars. Some vectorscopes range are widely available and is important to become familiar
have a 75%/100% selection on are generally adequate for with the definitions in whatever
the front panel. This setting routine testing. standards are used. Make sur e
changes the calibration of the those involved in measuring sys -
vectorscope chrominance gain to STANDARDS AND
tem performance agree on how
accommodate two different PERFORMANCE GOALS
to express the amount of distor-
types of color bars. The 75%/ No one standard defines all tion. It is good practice to record
100% distinction refers to ampli - amplitude and timing relation- this information along with the
tude, not to saturation or white ships for the NTSC signal. There measurement results.
bar level. These issues are dis- are a number of reference docu-
cussed in detail in Appendix A. ments produced by different
75% bars are most frequently organizations, several of which
used in NTSC systems as the are in common use today. RS-170A
large chrominance peaks in and FCC 73.699 Figure 6, two of

You might also like