Key Features


Highly conductive wet gel

Superior adhesion

Occlusive backing material

Unique offset concept

High-quality Ag/AgCl sensor

Ambu® Blue Sensor M


The small Ambu Blue Sensor M features highly Dimensions
conductive wet gel and superior adhesion to ensure Electrode size (max. L x W or diameter in mm) 40.8 x 34
optimal signal quality during short to medium- Skin contact size (max. L x W or diameter in mm) 34
term applications. Thanks to the special occlusive Adhesive area (in mm2) 754
backing material, the electrode will not absorb Height excluding connector/wire (in mm) 1
liquids.
sensor
Sensor material Silver/silver chloride
Gel system Wet gel
Sensor area (in mm2) 13.2
Gel area/measuring area (in mm2) 154

Lifetime
In opened pouches 1 month ELECTRICAL DATA (Ansi/aami)
Unopened pouches* 15 months AC impedance - typical 400 Ω
*from date of production DC offset voltage - typical 0.2 mV
Defibrillation overload recovery - typical 7.8 mV
Rate of change of polarization
X-ray & mri potential - typical 0.2 mV/s
X-ray-translucent No Combined offset instability and internal noise <15μV
MRI-safe No Bias current tolerance (over 8 hours) <5 mV

Airway Management |  Patient Monitoring & Diagnostics | Emergency Care


environment materials - electrode Connectors
PVC-free electrode No Bio-compatible Yes
Latex-free electrode Yes Sensor Silver/silver chloride (Ag/AgCl)
PVC-free lead wire - Sponge Polyurethane foam (PUR)
Latex-free lead wire - Outer carrier Polyvinyl chloride foam (PVC)
PVC-free packaging Yes Outer adhesive Acrylate
Inner carrier (if any) -
standard packaging Inner adhesive (if any) -
Quantity/liner 1 Upper part* Polyvinyl chloride film (PVC)
Quantity/pouch 10, 50 Fitting Polypropylene (PP) S = Stud
Quantity/box 1000/2000 Connector (stud) Nickel-plated brass
Minimum sales 2000 Lead wire core -
Lead wire insulation -
options Lead wire connector -
Connectors S, A, F Liner Polyester/polyethylene film (PE/PET)
Pre-wired -
Pre-wired connectors - materials - packaging
Pouches, inner layer Polyethylene (PE)
Available configurations* Pouches, centre layer Aluminium (Al)
Quantity/pouch 10 50 Pouches, outer layer Polyester (PET)
A = 4 mm fitting
Qty per inner/outer box Boxes Cardboard
M-00-S 1000/2000 1000/2000
M-00-A,F 1000/2000
* Some configurations may not be available worldwide

 PRODUcT
M-00-S  Wire length
 Connector

precaution F = 3 mm fitting
Single patient use only

493 1011 01 - V01 - 09/2010 - Ambu A/S - Technical data may be modified without further notice.