Non-Invasive Blood Pressure (NIBP) Analyzer


PRODUCT HIGHLIGHTS
High performance, compact design, simple to use Auto-calibration against all NIBP units Real blood pressure waveforms settings Real-life artifact and arrhythmia conditions such as: Bradycardia, Tachycardia, Geriatric Patient, Obese Patient, Strenuous and Mild Exercise, Weak Pulse, Premature Atrial Contraction, Premature Ventricular Contraction, Atrial Fibrillation, Motion/Tremor Artifact, 2-BLK II Actual BP waveforms based on clinical data Performs a complete test including cuff and tubing Digital manometer with automatic or manual inflation of the cuff pressure for Leak Test and Over Pressure Cut-off Test

OVERVIEW
By using innovative new technology the QA-1290 NIBP Simulator accurately tests all types of adult and neonate oscillometric NIBP monitors. Before simulation, the QA-1290 can calibrate both the cuff and tubing in order to verify the NIBP monitor. Once the QA-1290 knows the working environment, it is able to perform calibrated simulations under all conditions. The QA-1290 provides dynamic performance testing using both real and simulated blood pressure waveforms. Blood pressure waveforms include individual settings for systolic BP, diastolic BP, heart rate, and the pulse volume/amplitude. The QA-1290 also simulates calibrated artifact and arrhythmias in a large variety of real-life conditions. It provides high/low pressure release verification and automated leak testing. It also generates user selectable pressure levels to calibrate a variety of pressure monitors.

In the interest of product improvement, Metron claims the right to alter specifications without notice.

QA-1290 Specifications
MEASUREMENTS
Simulation technique: Calibration Tables: Oscillometric DINAMAP, HP-Merlin, Propaq, SpaceLabs, Datascope

LEAK TEST & OVER-PRESSURE CUTOFF TEST


Pressure Range: 0-500 mmHg Automatic or Manual Inflation

Cuff Adapters:
Male/female Luer-Lock Hose barbed fittings, 2 ea Male/female Clippard (Criticon/Siemens) Male/female Serto (Agilent) Male/female Colder (Marquette, Protocol System) Test accessories Inflation bulb Note: Adapters connect to the QA-1290 Pressure Port via a quick-disconnect, airtight, O-ring pressures fitting.

Simulation repeatability: Heart Rate:

0.5 mmHg 10-200 bpm, 1 bpm steps Heart Rate Accuracy: 1 bpm @pp < 2mmHg Preset Pressures: (Systolic/Diastolic (MAP), in mmHg) Adult Neonate 60 / 30 80 / 48 100 / 65 120 / 80 150 / 95 200 / 140 255 / 195 (40) (58) (77) (95) (114) (167) (215) 60 / 30 80 / 48 100 / 65 120 / 80 150 / 95 200 / 140 255 / 195 (40) (58) (77) (95) (114) (167) (215)

Pressure Accuracy: 1 mmHg Pressure Units: mmHg, cmH20, inH20, PSI, kPa, mBar

Digital Readouts: Manometer, Start Pressure, Peak Pressure, Deflation Time and Rate Leak Rate Range: 1.0-300 mmHg/minute Leak Test Time: selectable 1/2, 1, 2, 3, 4, 5 minute

Recommended Printers:
HP Desk Jet, Cannon Bubble Jet or compatible

QA-1290 ORDERING INFORMATION


Order no:

Independent Systolic/Diastolic: Systolic: Diastolic: Pulse Pressure: 20 - 255 mmHg 5 - 220 mmHg 0.05 - 5.0 mmHg

GENERAL INFORMATION
Temperature Requirements:
+15/59 to +35/95 C/F 0 to +50/122 C/F while operating for storage

17500: QA-1290 NIBP Analyzer

Accessories:
17510: 17515: 10500: 17520: 17521: 17530: 17525: QA-1290 Carrying Case Tubing Kit Carrying Case, ext. printer ansur QA-1290 Plug-In ansur QA-1290 Plug-In, Demo User Manual ansur QA-1290 Plug-In User/Service Manual QA-1290

CUFF MEASUREMENTS DURING BP SIMULATION


Real Time and Peak Cuff Pressure: 0 to 500 mmHg Total Measurement Time: 0 to 60 seconds Variable Pulse Pressure: Almost any patient condition can be simulated with the pre-set or user selectable pulse pressures. Motion/Tremor Artifacts: Test exactly how the BP monitor responds to calibrated levels of motion and tremor. Arrhythmia Simulations: Produce the appropriate pulse oscillations for patient arrhythmias such as Premature Arterial Contraction, AFiB, 2 Block II and PVC. Custom Patient Pattern Mode: An arbitrary patient pattern can be downloaded via the Serial Remote Control.

Display:
Type: LCD Alphanumeric format: 4 lines, 40 characters Display control: 7 F-keys and keypad

Data Input/Outputs (2):


Parallel printer-port (1): Bi-directional 25 pin, Type Centronics, RS-232C: (1) for Remote Control

Power: Housing:

From 110 VAC to 240 VAC, 47/63 Hz Aluminium Case

Dimensions: D X W X H
270 x 300 x 95 (mm) 11 x 11.7 x 4 (in)

Weight:

3.5 kg/7.7 lb

Standard Accessories:
QA-1290 User Manual

Cuff Mandrels:
Medium Adult (110 mm 0D, 160 mm width) and Neonate (50 mm 0D, 80 mm width)

MTD 0302.182.06