New Non-Contrast MR Techniques

syngo NATIVE and syngo ASL


One step ahead with Inline Technology

www.usa.siemens.com/healthcare
New Non-Contrast MR Techniques
syngo NATIVE and syngo ASL

MR angiography and Siemens Administration issued new-boxed obtained in patients who are at risk for
warnings for all manufactures of renal insufficiency. NYU’s departmental
Contrast-Enhanced Magnetic Resonance gadolinium, which stated that patients policy for administration of MR contrast
Angiography (ce-MRA) remains the gold with severe kidney insufficiency who agent based on estimated GFR (calculated
standard for evaluating vasculature in receive these agents are at risk for using the MDRD equation) is as follows:
MR as it enables high spatial and developing NSF. Their recommendations
temporal resolution with extremely fast further state that facilities should avoid eGFR >60 (mL/min/1.73 m2)
imaging times. Siemens technological use of gadolinium in patients with known No restriction for administration
innovations in angiography along with risks for developing NSF unless the of contrast.
the implementation of stronger gradients diagnostic information is essential and eGFR >30-60
has pushed imaging times to seconds cannot be obtained with a non-contrast Discuss with the patient the risks
instead of minutes. High resolution enhanced MR or other diagnostic of administration of contrast and
ultrafast ce-MRA techniques from procedures. The FDA also stated that consider using low dose.
Siemens are in routine clinical usage facilities should assess renal dysfunction
eGFR 15-30
today along with improved workflow prior to administration of gadolinium as
Discuss risks and benefits with
with Siemens unique inline processing, well as not exceed recommended doses.
referring physician, nephrologist
subtraction and image composing Go to www.FDA.gov for more information
and patient. Consider using low
speeding up exam time. Innovative on gadolinium-containing contrast agents.
dose but not recommended.
applications like syngo TWIST and syngo
NATIVE now provide the user flexibility Due to recent events involving NSF and eGFR <15 or on dialysis
of choice of technique to do MR gadolinium, facilities such as New York Avoid administration of contrast
angiography. syngo NATIVE enables University Hospital have implemented a unless benefits outweigh risks based
excellent non-contrast angiography screening process for those patients at on assessment of the referring
imaging and reproducible results. risk and modified their routine written physicians, nephrologists and
consent form. NYU assesses for renal radiologist. Peritoneal dialysis is
dysfunction prior to administration of an absolute contraindication.
NSF (Nephrogenic Systemic Fibrosis) MR contrast agents to identify patients
and changes in hospital policies at risk for renal insufficiency and thereby NYU utilizes a non-contrast angiography
potentially at risk for developing NSF. technique called syngo NATIVE when
The use of injected contrast media has The risk factors include increased these patients cannot have contrast.
come under scrutiny as the common serum creatinine, history of renal When low dose contrast is indicated,
contrast agents containing gadolinium disease, acute renal failure or recent renal NYU utilizes syngo TWIST (Time-resolved
have been linked to a condition called surgery, proteinuria, diabetes mellitus, Angiography With Stochastic Trajectories),
Nephrogenic Systemic Fibrosis (NSF) that hypertension, gout or recent intake of a new versatile high-resolution
can occur in patients with renal nephrotoxic drugs. Documentation of angiography technique that further
insufficiency. The Food and Drug recent BUN, creatinine and estimated improves time-resolved ce-MRA.
Glomerular Filtration Rate (GFR) are

2
2A

2B

Figure 1: syngo TWIST, MAGNETOM Avanto. Figure 2A: Time of Flight (ToF).
Figure 2B: Phase Contrast (PC).

syngo TWIST Demand for non-contrast New non-contrast MR techniques


angiography techniques
The new k-space coverage in syngo Siemens is the leader in Innovative
TWIST provides the user greater Utilization for Contrast-Enhanced Applications and Innovative Workflow
flexibility in tailoring temporal and Magnetic Resonance Angiography (CE with the most comprehensive contrast
spatial resolution for dynamic MR MRA) has been affected by the FDA’s new and non-contrast angiography offerings
imaging (Figure 1). syngo TWIST is up to warning. Currently, facilities performing with the flexibility to fit every clinical
10 times faster* compared to a standard, MRA examinations utilizing gadolinium scenario. Due to extensive research
full k-space version of dynamic imaging. desire non-contrast angiography collaborations, Siemens has developed
Excellent background suppression as applications as an alternative backup multiple non-contrast techniques
compared to other available techniques for patients at risk for developing NSF. like syngo NATIVE and syngo ASL for
results in accurate and reproducible clinical need.
diagnostic information even with smaller
amounts of contrast agent. Current MRA techniques
In Magnetic Resonance Imaging, several syngo NATIVE
The key clinical benefits of physical effects of flowing blood are
syngo TWIST are: exploited to image the vessels without For non-contrast MR Angiography
injection of contrast. Traditional non- Siemens currently offers two techniques
• Scalable temporal resolution contrast techniques encompass: under the category of NATIVE (Non-
• Workflow efficient ultrafast contrast MRA of ArTerIes and Veins)—
technique • Time of Flight MRA (ToF MRA) — syngo NATIVE TrueFISP and syngo
• Lower cost and improvements in (Figure 2A) Contrast mechanism is NATIVE SPACE.
diagnostic confidence as additional based on exchanging the signal from
high-res scan can be acquired in the inflowing blood during the syngo NATIVE TrueFISP
same setting measurement. syngo NATIVE TrueFISP is based on the
• Increased speed • Phase Contrast MRA (PC MRA)— sequence TrueFISP (True Fast Imaging
• Head-to-toe applications (Figure 2B) Contrast mechanism is with Steady state Precession) which is a
• Improved image quality based on the change in phase due to balanced steady state gradient echo
• Multiple very small dose contrast motion or flow. technique. The contrast mechanism for
injections are possible to capture syngo NATIVE TrueFISP comes from the
dynamic processes These techniques are adequate for preparation of the imaging volume with
intra- and extra-cranial circulation. a spatially selective inversion pulse,
resulting in the suppression of stationary
tissue within the imaging volume and
suppression of signal from blood in the

*Data on file.
3
Figure 3: syngo NATIVE TrueFISP.

Figure 4: syngo NATIVE SPACE. Figure 5: Pulsed ASL perfusion with 3D PACE. Single Shot EPI
pulsed arterial spin labeling sequence (Q2TIPS).

imaging volume (e.g. venous blood). enables dynamic angiography—for Tim is the solution that addresses the
Blood which flows into the imaging example, in the lower legs (Figure 4). real needs of today’s clinical MRI
volume during the inversion time has the environment. Tim provides the flexibility
same high signal characteristics exhibited to scan any part of the body without
in conventional TrueFISP. The contrast is Non-contrast MR perfusion studies— interrupting the exam for coil or patient
further enhanced by the suppression of syngo ASL (Arterial Spin Labeling) repositioning. Tim is also the first RF
the background by the inversion pulse. technology optimized for use with
The sequence can be made selective for Siemens has been at the forefront in parallel imaging.
arteries or veins by appropriate other non-contrast techniques beyond
positioning of the inversion pulse, which angiography. Arterial Spin Labeling
can be positioned independently from (ASL) is an MR technique using the Conclusions
the imaging volume. The sequence water in arterial blood as an endogenous
accommodates 3D, 2D, breath-hold, contrast agent to evaluate perfusion Siemens technological innovations
PACE (Prospective Acquisition and non-invasively. ASL provides unique in angiography along with the
CorEction) navigated and respiratory insight into human brain perfusion and implementation of stronger gradients
triggered approaches depending on function physiology by evaluating has pushed imaging times to seconds
clinical environment (Figure 3). cerebral blood flow (CBF). ASL is capable instead of minutes. High resolution
of high spatial resolution perfusion ultrafast ce-MRA techniques from Siemens
syngo NATIVE SPACE— imaging, making the technique very are in routine clinical usage. Improved
Siemens exclusive appealing in the evaluation of stroke, workflow with Siemens unique inline
syngo NATIVE SPACE is based on the tumors, degenerative diseases and processing, subtraction and image
SPACE technique which is a modified epilepsy, but also in basic neuroscience, composing speeds up exam time.
variable flip angle 3D Turbo Spin Echo e.g. for studies of functional CBF changes. Innovative applications like syngo TWIST
sequence in which the contrast Siemens is proud to introduce syngo ASL, and syngo NATIVE provide the user
mechanism for visualizing vessels is based powered by Tim (Total imaging matrix), flexibility of choice of technique to do
on the difference in intravascular signal thereby allowing the use of noncontrast MR angiography in any clinical scenario.
between maximal and minimal flow enhanced perfusion imaging in both the syngo NATIVE enables excellent non-
during the cardiac cycle. The subtracted clinical and research settings. (Figure 5) contrast angiography imaging and is an
image is calculated in-line and with inline excellent alternative for patients in whom
MIP generation, instantaneous clinical contrast material needs to be avoided.
results are produced for completely Tim technology With Tim and inline processing, non-
non-invasive MR angiography, further contrast MRA, head-to-toe is easy and
improving workflow at the scanner. The syngo NATIVE and syngo ASL are demonstrates excellent image quality.
syngo NATIVE SPACE technique can also powered by Tim (Total imaging matrix).
accommodate multi phase imaging that

4
Head to Toe Coverage with
Non-Contrast Techniques

syngo ASL NEW 3D Time of Flight 2D ToF Carotids

syngo NATIVE SPACE syngo NATIVE TrueFISP

syngo NATIVE TrueFISP

Whole body ce-MRA with Tim syngo NATIVE SPACE syngo NATIVE SPACE

5
On account of certain regional limitations of Contact Address Germany
sales rights and service availability, we cannot Siemens AG, Medical Solutions
guarantee that all products included in this Henkestr. 127, D-91052 Erlangen
brochure are available through the Siemens Germany
sales organization worldwide. Availability and Telephone: +49 9131 84-0
packaging may vary by country and is subject www.siemens.com/medical
to change without prior notice. Some/all of
the features and products described herein Contact Address USA
may not be available in the United States. Siemens Medical Solutions USA, Inc.
Magnetic Resonance Division
The information in this document contains 51 Valley Stream Parkway
general technical descriptions of specifications Malvern, PA 19355-1406 USA
and options as well as standard and optional Telephone: 1-888-826-9702
features which do not always have to be present
in individual cases.

Siemens reserves the right to modify the design,


packaging, specifications and options described
herein without prior notice. Please contact your
local Siemens sales representative for the most
current information.

Note: Any technical data contained in this


document may vary within defined tolerances.
Original images always lose a certain amount