IOLMaster Advanced Technology software version 5 2 ASCRS 2007, San Diego, Training IOLMaster Advanced Technology Axial Length Measurement Mode Minimum of 5 measurements are required. 1. measurement (red) 3 ASCRS 2007, San Diego, Training IOLMaster Advanced Technology Axial Length Measurement Mode Single AL measurement signal is shown in RED. 2. measurement (red) 4 ASCRS 2007, San Diego, Training IOLMaster Advanced Technology Axial Length Measurement Mode SNR is shown on signal as before. 3. measurement (red) 5 ASCRS 2007, San Diego, Training IOLMaster Advanced Technology Axial Length Measurement Mode Toggle between single / composite signals by scrolling through measurement value list. 4. measurement (red) 6 ASCRS 2007, San Diego, Training IOLMaster Advanced Technology Axial Length Measurement Mode After 5th measurement RED single measurement signal is shown just for 1 second. 5. measurement (red) 7 ASCRS 2007, San Diego, Training IOLMaster Advanced Technology Axial Length Measurement Mode Then BLUE composite measurement signal is shown.
Composite AL measurement value is shown underneath horizontal line.
Enhanced SNR is shown as number and as traffic signal.
composite signal (blue) 8 ASCRS 2007, San Diego, Training IOLMaster Advanced Technology Axial Length Measurement Mode SNR categories
RED
YELLOW
GREEN
Measuring error Perform more single measurements.
Uncertain value Perform more single measurements, especially in case of multiple peaks.
Value with good SNR Look also for quality of signal: Avoid double peaks from RPE / ILM)!!! Axial length measurement succeeded.
While the system is internally calculating the axial length from the interference signal, it automatically analyzes the SNR. 9 ASCRS 2007, San Diego, Training IOLMaster Advanced Technology Keratometry Adjustment Aid The keratometry adjustment aid can be activated in the Program Settings menue: 10 ASCRS 2007, San Diego, Training Alignment status
RED
YELLOW
GREEN
IOLMaster Advanced Technology Keratometry Adjustment Aid A traffic light indicates the status of the system adjustment to the patients eye: System out of focus Adjust instrument head distance to patients eye. All 6 spots should be clearly visible.
System near the best focus position. Fine tune distance to patients eye. All 6 spots should be clearly visible and in focus.
System in best focus position. Ask patient to blink and perform keratometry measurements. Three measurements are required. 11 ASCRS 2007, San Diego, Training IOLMaster Advanced Technology Keratometry Adjustment Aid 12 ASCRS 2007, San Diego, Training Plausibility Check AL measurements: OD and OS difference > 0.3 mm K measurements: OD and OS difference > 1.0D 13 ASCRS 2007, San Diego, Training The IOL Master Version 5 software is
Additional COST is involved 14 ASCRS 2007, San Diego, Training IOL Master version 5 roll out process New system available after 15 th June Hardware upgrade with new software available after 29 th June Orders for new IOL Master version 5 software before 15 th June 1) Same order part number as per V4. Place remark ,IOL Master V5 Orders for new IOL Master version 5 software after 15 th June 1) New part number Up grade from V4 to V5 1) Software upgrade done by engineer Up grade from V1,2,3 to V5 1) Software & hardware upgrade done by engineer