You are on page 1of 14

1

ASCRS 2007, San Diego, Training


IOLMaster
Advanced Technology
software version 5
2
ASCRS 2007, San Diego, Training
IOLMaster Advanced Technology
Axial Length Measurement Mode
Minimum of
5 measurements
are required.
1. measurement (red)
3
ASCRS 2007, San Diego, Training
IOLMaster Advanced Technology
Axial Length Measurement Mode
Single AL
measurement signal
is shown in RED.
2. measurement (red)
4
ASCRS 2007, San Diego, Training
IOLMaster Advanced Technology
Axial Length Measurement Mode
SNR is shown
on signal as before.
3. measurement (red)
5
ASCRS 2007, San Diego, Training
IOLMaster Advanced Technology
Axial Length Measurement Mode
Toggle between
single / composite
signals by
scrolling through
measurement
value list.
4. measurement (red)
6
ASCRS 2007, San Diego, Training
IOLMaster Advanced Technology
Axial Length Measurement Mode
After
5th measurement
RED single
measurement
signal is shown
just for 1 second.
5. measurement (red)
7
ASCRS 2007, San Diego, Training
IOLMaster Advanced Technology
Axial Length Measurement Mode
Then BLUE composite
measurement signal
is shown.

Composite AL
measurement value
is shown underneath
horizontal line.

Enhanced SNR is
shown as number and
as traffic signal.



composite signal (blue)
8
ASCRS 2007, San Diego, Training
IOLMaster Advanced Technology
Axial Length Measurement Mode
SNR categories

RED


YELLOW


GREEN


Measuring error
Perform more single measurements.


Uncertain value
Perform more single measurements,
especially in case of multiple peaks.

Value with good SNR
Look also for quality of signal:
Avoid double peaks from RPE / ILM)!!!
Axial length measurement succeeded.

While the system is internally calculating the axial length from
the interference signal, it automatically analyzes the SNR.
9
ASCRS 2007, San Diego, Training
IOLMaster Advanced Technology
Keratometry Adjustment Aid
The keratometry adjustment aid can
be activated in the Program Settings menue:
10
ASCRS 2007, San Diego, Training
Alignment status

RED


YELLOW


GREEN


IOLMaster Advanced Technology
Keratometry Adjustment Aid
A traffic light indicates the status of the
system adjustment to the patients eye:
System out of focus
Adjust instrument head distance to patients eye.
All 6 spots should be clearly visible.


System near the best focus position.
Fine tune distance to patients eye.
All 6 spots should be clearly visible and in focus.

System in best focus position.
Ask patient to blink and perform keratometry
measurements. Three measurements are required.
11
ASCRS 2007, San Diego, Training
IOLMaster Advanced Technology
Keratometry Adjustment Aid
12
ASCRS 2007, San Diego, Training
Plausibility Check
AL measurements: OD and OS difference > 0.3 mm
K measurements: OD and OS difference > 1.0D
13
ASCRS 2007, San Diego, Training
The IOL Master Version 5 software is

Additional COST is involved
14
ASCRS 2007, San Diego, Training
IOL Master version 5 roll out process
New system available after 15
th
June
Hardware upgrade with new software available after 29
th
June
Orders for new IOL Master version 5 software before 15
th
June
1) Same order part number as per V4. Place remark ,IOL Master V5
Orders for new IOL Master version 5 software after 15
th
June
1) New part number
Up grade from V4 to V5
1) Software upgrade done by engineer
Up grade from V1,2,3 to V5
1) Software & hardware upgrade done by engineer