You are on page 1of 35

UV-Vis Absorption Spectroscopy

Introduction
A.) Absorption: electromagnetic (light) energy is transferred to atoms, ions, or molecules
in the sample. Results in a transition to a higher energy state.

E1

(excited state)

Eo

(ground state)

Energy required of photon


to give this transition:
hE= E1 - Eo

Transition can be change in electronic levels, vibrations, rotations, translation, etc.

Concentrate on Molecular Spectrum in UV/Vis (electronic transition)

Power (P): energy of a beam that reaches a given area per second

Intensity (I): power per unit solid angle

P and I related to amplitude2

B.) Terms:
1.) Beers Law:

A = bc

The amount of light absorbed (A) by a sample is dependent on the path length (b),
concentration of the sample (c) and a proportionality constant ( molar absorptivity)

Amount of light absorbed is dependent on frequency () or wavelength ()


Increasing Fe+2 concentration

Absorbance is directly proportional to concentration Fe+2

B.) Terms:
1.) Beers Law:

A = bc

Transmittance (T) = P/Po

%Transmittance = %T = 100T

Absorbance (A) = log10 Po/P

No light absorbed% transmittance is 100% absorbance is 0


All light absorbed% transmittance is 0% absorbance is infinite

Relationship Described in Terms of Beers Law


A = Absorbance = bc = -log(%T/100)
= molar absorptivity: constant for a compound at a given frequency () or wavelength ()
units of L mol-1 cm-1
b = path length: cell distance in cm
c = concentration: sample concentration in moles per liter.
Therefore, by measuring absorbance or percent transmittance at a given frequency can get
information related to the amount of sample (c) present with an identified and .

Note: law does not hold at high


concentrations, when A > 1

C.) Components of an Instrument for UV/Vis Absorbance Measurements:


1.) Basic Design:

Hitachi Instruments U-3010

Light Source, selector, Sample cell holder, Detector (amplifier, recorder)

a) Desired Properties of Components of UV/Vis:


Light Source

Selector

Creates Proper
Narrow Band pass:
Stable:
Selects Desired
Constant P
Large Light Throughput:
Good Precision
Increase P
Intense:
Increase P
Easier to See Absorbance

Sample Cell Holder


Fixed Geometry:
Constant b
Transmits of Interest:
Increase P

Detector
Stable
Sensitive to of Interest

b) Light Sources UV/Vis (~ 200 800 nm):


1. Deuterium & Hydrogen Lamps (UV range)
- continuous source, broad range of frequencies
- based on electric excitation of H2 or D2 at Low pressure

D2 or H2 Gas

Filament
40V

Electric Arc

Sealed Quartz Tube


Electrode

In presence of arc, some of the electrical energy is absorbed by D2 (or H2)


which results in the disassociation of the gas and release of light
D2 + Eelect D*2 D + D + h(light produced)
Excited state

2. Tungsten Filament Lamp (Vis Near IR)


- continuous source, broad range of frequencies
- based on black body radiation:
heat solid filament to glowing, light emitted will be
characteristic of temperature more than nature of
solid filament

Low pressure (vacuum)

Tungsten Filament

Temperature Dependence of

2. Tungsten Filament Lamp (Vis Near IR)


- max 1/T
Typical Tungsten lamp T ~ 2870K
range: 350 2500 nm
need high temperatures to get high light intensity and low max

- cost ~ $10-15

3. Xenon Arc Lamps (UV Vis Range)


- Continuous source, broad range of frequencies
range: 250 600 nm

- works by passage of current through Xe, causes thermal excitation


Blackbody emission

- Gives Very intense radiation over frequency range.


developed for search lights during WWII

- problems: higher heat, more stray light, higher cost, shorter lifetimes

b) Wavelength Selectors:
1. Monochromator
- separates frequencies () from polychromatic light.
- allows only certain s to be selected and used.
i.) Dispersing Monochromator:
a) Prism: based on refraction of light and fact that different s
have different values of refraction index (i) in a medium.

Dispersion Curve: change in i as a function of


Normally want to work in areas of normal dispersions for prisms.
Anomalous dispersion occurs near where substance itself absorbs light.

Recall Snells Law of Refraction

1sin1sin2

Also, remember that no refraction occurs if light at normal or = 0


So, light must hit prism at an angle.
Most common is a 60o prism
(glass or quartz).

ii.) Grating Monochromator: based on diffraction of light (constructive


and deconstructive interference)
a) Transmission Grating: groves or slits placed on a
transparent material. Same as earlier example shown in
diffraction
discussion
X
n =2
Order of Interference (n): n = d sin
n =1
n =0
n =1

Different s will have constructive


interference at different points.
Can select desired by letting light at
different points into instrument.

b) Reflection Grating: most commonly used


- grooved surface with reflective coating (Al, Au, Pt)

n = d(sin i + sin r)

Now, spacing of slits (d) is distance from one groove to next.


Typically have 300-2000 grooves/mm.
Constructive and deconstructive interference occurs because
light travels different distances when reflected from each grating
Angle at which constructive interference occurs is now given by:

n = d(sin i + sin r)

Example 5: A monochromator was equipped with an echellete grating of 2500 blazes per
millimeter, where 2.0 cm of the grating was illuminated. At approximately 430 nm, what
minimum wavelength difference could in theory be completely resolved by the instrument?

iii.) Two Types of Monochromators:


a) Czerney-Turner Grating:

Important Components:
i) Entrance slit
ii) Collimating lens or mirror makes radiation parallel before hitting
dispersing element
iii) Grating or Prism
iv) Focusing Lens or mirror to focus light of desired on exit slit.

iii.) Two Types of Monochromators:


b) Bunsen Prism:

Important Components:
i) Entrance slit
ii) Collimating lens or mirror makes radiation parallel before hitting
dispersing element
iii) Grating or Prism
iv) Focusing Lens or mirror to focus light of desired on exit slit.

iv.) Comparison of Gratings vs. Prisms:

Grating
Prism

Type of Dispersion

Size

Stray Light

uniform dispersion vs.

smaller

Higher stray light

shorter better separated

larger

Less of problem

range of
use
unlimited
Limited
( 350 nm)

increase size of either prism or grating will give better dispersion.


stray light can be removed with filters.
glass absorbs light 350 nm

v.) Slits in the Monochromator:


- need to be carefully made, since they control the range of
emerging from the monochromator.

- typical slit widths are 0.01 2 mm are often adjustable.

a) Band pass of the monochromator: range of s transmitted


the half-height of transmitted light band.

at

Typical band pass : 20nm -0.5 nm for UV-vis instruments.


- decrease slit size decrease band pass (good).
- less undesired s, but less intensity (bad).

Two s can be resolved by the


monochromator if they differ by
2 or more times the band pass.
- resolution is directly related to
slit size.

a) Filter: Monochromators
1) Absorption Filters
- material remove undesired s by absorbing them.
- typically made from colored glass or dye
suspended in gelatin between glass plates.
- fixed , much energy lost due to absorption.
- cheap

wide range of allowed through.


can combine filters with different range.
typical band pass (30-250 nm).

Effective bandwidth for two types of


filters and the result of combining
filters.

2) Interference Filters
- made up of thin layers of metal and dielectric (eq. CaF2)
material sandwiched between glass plates
- dielectric material is of uniform, known thickness.
- metal acts as partial mirror

2) Interference Filters
As light enters, some goes through but some is
reflected. The distance the light travels before it
exits generates constructive and destructive
interference on the other side of the filter.
s transmitted through filter:
N = 2d
refractive index
d thickness of dielectric
N integer
- wavelength

Band pass can be 1-20nm. (narrow), but filter is fixed at given value as much intensity
is lost due to reflection

vi.) Sample Cell:


- Must be transparent at used.
quartz or fused silica for UV (< 350 nm)
glass or plastic for visible
- best if flat cells, with matched sample and reference cells.
- many automated instruments have flow-through cells with
temperature control.

vii.) Detectors:
a) earliest detectors were the eye or film.
now use devices that convert light to electrical signal
b) for good detector want:
high sensitivity
good signal to noise ratio
constant response over range of interest
signal light intensity
fast response
little or no signal in absence of light (dark current)

c) Many Types of Available Detectors for UV/VIS


1) photovoltaic cell (Barrier-Layer Cell)

Process:
light of sufficiently high energy passes through the thin transparent silver layer and
hits selenium causing electrons to be released which move across barrier toward
silver layer (electropositive) and collected at iron layer to neutralize selenium layer.
-

Current produced is proportional to photons hitting surface


Maximum response at 550 nm (10% at 350-750 nm ~ same as human eye).

Advantage: cheap, rugged, no external power source, good for portable instruments.
Disadvantage: not very sensitive, shows fatigue (decrease in response with continued
illumination), difficult to amplify signal-small resistance (Ohms law: I=(V/R)).

2) Vacuum Phototube

(Photoemission material)

Process:
- photoemission material (Cs2O) ejects an e- when hit with a photon
(photoelectric effect).
- potential of 90V across cathode(-) and anode(+). As light hits cathode, eare emitted from cathode and attracted to anode. Produces current that can be
measured.
- current number of photons.
- smaller current then photovoltaic cell, but can be amplified because of larger
resistance.
- 90V difference sufficient to collect all e- produced (working at saturation).
- various photoemission material (sensitive to certain photon s)

2) Vacuum Phototube
Process (cont.):
- various photoemission material (sensitive to certain photon s)

- common to use phototubes in UV-Vis instruments


one tube for UV & most visible region, second tube for red end of spectrum.
some phototubes have flat regions throughout entire range (ex.128 Ga/As composite).
sometimes gas is present in tube. As e - collide with gas, more e- and ions produce

results in an increase in current.

Advantages: sensitive, signal easily amplified.


Disadvantages: some dark current (from thermal e- emission & natural decay of 40K in glass
housing

3) Photomultiplier tube (PMT)


Dynodes all covered with photoemissive material

Process:
a) light hits cathode and e- emitted.
b) an emitted e- is attracted to electrode #1
(dynode 1), which is 90V more positive.
Causes several more e- to be emitted.
c) these e- are attracted to dynode 2, which is
90V more positive then dynode 1, emitting
more e-.
d) process continues until e- are collected at
anode after amplification at 9 dynodes.
e) overall voltage between anode and cathode
is 900V.
6
f) one
photon produces
107 electrons.
Advantages: very sensitive to low
intensity,
very fast10
response.
g) current
is supply,
amplified
and measured
Disadvantages: need high voltage
power
intense
light damages

4) Silicon Diode or Photodiode detectors


Semiconductor material conducts current only under certain conditions

Reverse-bias: no current flows

free electrons inside the N-type material need some


extra energy to overcome the repulsion of the P-type's
acceptor atoms.

- Light shining on the silicon diode provides the energy needed for the
electrons to travel into the P region.
- Flow of current is related to intensity of light.

4) Silicon Diode or Photodiode detectors

- Not very sensitive, but useful in Vidicon Tubes


- Vidicon tubes can be used for 1 & 2 dimensional detection
Detect all the s along the focal plane of an instrument

A vidicon tube contains many silicon diodes each insulated from each
other (~ 8m apart). The e- gun or light source sweeps the P-region
and charges the diodes. (Television)

vii.) Single and Double Beam Instruments:


- To determine absorbance both Po and P must be measured.

Absorbance (A) = log10 Po/P


- depends on:
intensity of source
slit width & of monochromator
reflectance of the cell
sensitivity of the detector

Process for a Single Beam Spectrometer:


a) Po is measured with solvent in the cell
b) P, blank everything except compound to be analyzed
- spectrophotometer is adjusted to read
100%T or 0% A.
- 0%T or A is set by blocking the light beam (P=0)
- settings may change over a period of minutes (drift)
change in light source intensity

- settings are wavelength dependent

Schematic of Spectronic 20

Spectronic 20

- Single beam spectrophotometer (340-625nm) other phototube goes to 950


nm (20 nm band pass)
- Reference phototube electronically adjusts for changes in source intensity
- The control turns the grating.
- The 0%T adjust sets the meter to 0%T when the occluder blocks the light
beam.
- The 100% T adjusts moves a V-shaped slot into or out of the light beam
so the meter reads 100%T

Advantages: cheap, rugged.


Disadvantages: must readjust 100%T at every and periodically check for drift, cell is
round so path length can vary if dont have cell aligned the same each time

Double Beam Spectrometer Decrease/Eliminates Single Beam Problems

Schematic of Hitachi 100-60

Selector mirror or beam chopper: mirrors which rotate at 60 cps, reflecting the light
alternatively to the reference and sample cells
All light is combined and goes to a single PMT. Output of PMT is 60 cps square wave. Po and
P are measured alternatively at a rate of 60 times per second.
Advantages: scanning, little drift only one PMT.
Disadvantages: more complex and expensive

You might also like