Professional Documents
Culture Documents
Block Diagram
System Description
The collimators, scintillation crystal and PM Tubes
are housed in a cylindrical shaped housing
commonly called the camera head.
Collimator normally consists of very large piece
of lead with many small parallel holes of equal
cross section
Gamma rays perpendicular to collimator hole
surface reach to crystal and remaining rays
absorbed in collimator
Only small percentage typically 0.01 % rays
detected by crystal and used for image formation
System Description
The crystal can be between about 25 cm and
40 cm in diameter and about 1 cm thick.
The crystal diameter is dependent on the
application of the device. For example a 25
cm diameter crystal might be used for a
camera designed for cardiac applications
while a larger 40 cm crystal would be used
for producing images of the lungs.
The thickness of the crystal is chosen so that
it provides good detection for the 140 keV rays emitted from 99mTc - which is the most
common radioisotope used today.
System Description
Scintillations produced in the crystal are detected
by a large number of PM tubes which are arranged
in a two-dimensional array.
There is typically between 37 and 91 PM tubes in
modern gamma cameras.
The output voltages generated by these PM tubes
are fed to a position circuit which produces four
output signals called X and Y.
These position signals contain information about
where the scintillations were produced within the
crystal.
In the most basic gamma camera design they are
fed to a cathode ray oscilloscope (CRO).
System Description
We should note that the position signals also
contain information about the intensity of
each scintillation.
This intensity information can be derived from
the position signals by feeding them to a
summation circuit (marked in the figure)
which adds up the four position signals to
generate a voltage pulse which represents
the intensity of a scintillation.
This voltage pulse is commonly called the Zpulse which following pulse height analysis
(PHA) is fed to the CRO.( control intensity )
Position Determination by a
Gamma Camera
To determine where the gamma was absorbed
on the crystal, consider a system of X and Y
axes superimposed on the crystal surface.
To obtain the Y position of the gamma,
compare the outputs of PM tubes above the X
axis to those below the X axis.
To obtain the X position of the gamma,
compare the outputs of the PM tubes to the
right of the Y axis to those to the left of the Y
axis.
Position Signals + X, - X,
+ Y, - Y
If a gamma strikes the very centre of the
crystal, the PM tubes above and below
the X axis will receive the same number
of photons and will generate signals that
have the same signal strength.
In this case, the output of + Y and Y
pulses will be the same, no matter
where the gamma strikes on the X
axis.
Energy Normalization
The absolute value of the pulses depends
upon the energy of the gamma.
To make positioning independent of energy, the
ratio between the individual pulses and the sum of
all the pulses is determined.
For example, the 280 keV gamma absorbed in the
crystal at the same location as the 140 keV
produces positioning pulses that are twice the
height of the 140 keV gamma pulse. However, the
normalized pulses are dependent only on gamma
position and are independent of incident gamma
energy.
Collimator
If the gamma camera is used without a collimator,
the radiation coming from the tumor will strike the
surface of the entire camera crystal. In this
situation, the crystal would be flooded with
radiation, and the camera image would not show
the tumor.
Adding a collimator makes a big difference. The
tumor is now clearly visible on the camera image.
The collimator absorbs all the radiation that
reaches the crystal except that which is
located directly opposite the position of the
tumor.
Collimator
Collimators
The collimator can be made from
lead foil or from cast lead.
The walls of each channel in the
collimator are called septa, and if a
photon manages to penetrate the
wall, it is called septal penetration
Performance Parameters
Sensitivity
Resolving time
Uniformity
Spatial resolution
Spatial distortion
Sensitivity
Governed by counting efficiency of
camera components
Defined as the number of counts
per second that system obtains
for each unit of activity being
viewed
Dependent on geometrical efficiency
of the collimator, efficiency of crystal
and width of the PHA
Resolving time
Counts observed per unit time will be directly
proportional to the amount of activity
As the intensity of gamma ray increases the
probability that two rays will arrive the same time
increases
This produces two overlapping flashes in the
crystal which are interpreted in the system
coming from a single higher energy photon
These are normally rejected by PHA
Finite dead time of electronics leads to counting
loss
Uniformity
Ideally, camera should have a
uniform response across its field of
view
System uniformity varies by as much
as 15 % over the entire crystal
Counts per unit area in response to
uniform gamma flux depends on
crystal response, spatial photo peak
alignment and linearity
Spatial resolution
Systems ability to distinguish
between two closely spaced points
It is limited by collimator
characteristics, scatter and systems
ability to accurately determine point
in the crystal
As gamma energy increases,
collimation become more difficult,
septal penetration increases
Spatial distortion
Co-ordinates calculated by decoder
for any event are subjected to both
random and systematic errors
Hence events are plotted at improper
locations in the image
Energy resolution
Flood field uniformity
Spatial resolution
Spatial distortion
Counting rate loss
Sensitivity
Energy resolution
A measure of its ability to distinguish between
interactions depositing different energies in its
crystal
Refers to ability of system to reject events associated
with scattered photons
Depends upon width of the photo peak in the energy
spectrum
Energy spreads expressed in terms of FWHM
Causes of such spread are intrinsic fluctuation
in the light emitted from event to event, light
collection efficiency and electron multiplication
spread from PMT
Spatial resolution
It refers to the systems ability to distinguish
between two closely spaced points or line
sources
Spatial resolution is a measure of the cameras
ability to resolve small objects in the field of view.
It is inversely proportional to gamma energy
It is determined primarily by measuring the FWHM
of the recorded count profile along X or Y coordinate
Spread is measured right angles to the line
sources
Rs2 = Rc2 + Ri2
Rs: system resolution
Rc: collimator resolution
Ri: Intrinsic resolution
Deterioration in the intrinsic resolution
is due to the noise contributed by the
photomultipliers, poor optical coupling,
incorrect setting of window, high count
rate
Spatial distortion
It represents a non-linearity or inaccuracy in
the positional output
To find non-linearity in the system, use
transmission type bar phantom with lead bars,
so as to produce about 5 mm width exposure of
the crystal
When the co-ordinate signal is fed to an MCA,
the spatial spectrum is obtained
Then determine spacing between two bars and
hence the error associated with respect to the
actual distance of the bars
Once a week
Countrate loss
When the output countrate is lower
than the input countrate, some
counts are lost due to the constraint
of minimum time requirement for
signal processing
This means that two events which
take place rapidly cannot be resolved
Dead time of 2 microsec will result
counting loss of 10%
SPECT
Rotational uniformity
Photomultiplier tubes (PMT) are known to
undergo gain shifts when their spatial
orientation changes with respect to an
external magnetic field.
These effects can be produced by the
earths magnetic field (0.05 mT) as well
as fields around magnetic resonance
imaging systems and particle accelerators
such as cyclotrons.
Spatial resolution
In planar imaging it depends on the intrinsic capabilities
of the scintillation camera, the geometrical properties of
the collimator, and the presence of scatter.
In SPECT, additional factors affect resolution because
the information is gathered over multiple angles. Precise
positioning of the gantry, detector, and table and
calibration of the center of rotation (COR) is required.
In addition, the reconstruction matrix size, filter, and
use of pre and post-processing procedures can affect
the measured spa
Tomographic uniformity
The uniformity of reconstructed
images is very sensitive to
scintillation camera field uniformity.