2
Change Scan- Inspection Menu parameters
Type- one line scan
Scan-Encoder 1
3
Select Encoder and Polarity-Scan/Encoder Menu
Select encoder 1.
Check polarity by looking at axis
movement in right corner while
moving encoder in intended
direction. If the numbers are
moving in a positive direction
polarity is correct, if not
change to inverse. Input encoder
resolution if known.
4
Set Scan Length-Where to Start and End- Scan/Area Menu
5
Set Overwrite Parameters-Scan/Data Menu
What would you like the OmniScan
to record when the encoder goes over an
area it has already covered?
6
Calibrating the Encoder
After
the scan parameters are set the
encoder will need to be calibrated if the
encoder resolution is unknown.
7
Calibration-Wizard-Calibration Menu-choose Type-
Encoder-Start
8
Setting Origin
x axis 0.00
After positioning encoder at a 0 position hit next
9
Set Distance to Travel
Input the distance you plan on moving the encoder.
10
Move this Distance
11
Push Calibrate and Verify
When you push calibrate the axis value
will change to your set parameter
Accept or Restart
12