Setting up Mini Encoder- The

Basics for a One Line Scan


with 2.0 SW
By Nick Bublitz
Getting Started
 Almost all the parameters to set
up the encoder can be found
under two menus-Scan Menu,
and Wizard- Calibration-encoder
Menu.
 The encoder should be attached
to the wedge using the brackets.
Align the wheel in the direction
you want to encode- the scan
axis-direction probe will move.
 Make sure there is some
tension on the wheel when you
press down to ensure adequate
pressure to avoid slipping.
 Insert the encoder cable into the
encoder slot on top of the
OmniScan being conscious of
removed pin and its location.

2
Change Scan- Inspection Menu parameters
Type- one line scan
Scan-Encoder 1

3
Select Encoder and Polarity-Scan/Encoder Menu

Select encoder 1.
Check polarity by looking at axis
movement in right corner while
moving encoder in intended
direction. If the numbers are
moving in a positive direction
polarity is correct, if not
change to inverse. Input encoder
resolution if known.

4
Set Scan Length-Where to Start and End- Scan/Area Menu

Also set resolution in scan


axis.

5
Set Overwrite Parameters-Scan/Data Menu
What would you like the OmniScan
to record when the encoder goes over an
area it has already covered?

Can be last seen data, highest


amplitude in the A gate, or the
minimum or maximum thickness
when doing thickness
inspections

6
Calibrating the Encoder

After
the scan parameters are set the
encoder will need to be calibrated if the
encoder resolution is unknown.

Needed items- small rule or distance


reference, good surface to calibrate on.

7
Calibration-Wizard-Calibration Menu-choose Type-
Encoder-Start

8
Setting Origin
x axis 0.00
After positioning encoder at a 0 position hit next

Start your 0 point at a reference point on the wedge.

9
Set Distance to Travel
Input the distance you plan on moving the encoder.

10
Move this Distance

Move the distance you input and hold

11
Push Calibrate and Verify
When you push calibrate the axis value
will change to your set parameter

Carefully move the probe back to the


0 point and verify x axis reading is at 0.

Accept or Restart

12