You are on page 1of 44

UNIT-3

BIO-CHEMICAL INSTRUMENT

Medical Laboratory
Instrument
SUJITH.M, LECTURER/EEE, VIDYAA VIKAS ENGG COLLEGE Ph.No.
9600286256
Equipment
Spectrophotometer
Is optical device that measure light absorption at various
wavelengths for a given liquid sample.
Blood cell analyzer
Is a device to measures the number of red and white
blood cells per scaled volume.
The aperture impedance and flow cytometery

sujith.m,lecturer/eee,9600286256
Equipment
Glassware, centrifuges, suction devices
Colorimeter
Is an optical devise that measures the color concentration
of a substance in solution
Flame photometer
Is an optical electronic devise that measures the color
intensity of substance that have been aspirated into a
flame (sodium and potassium)

sujith.m,lecturer/eee,9600286256
Equipment
Ph/ blood gas analyzer
Is a device which measure blood Ph, Po2, Pco2
Chromatograph and Autoanalyzer
Is a electromechanical device used to separate, identify,
and measure the concentration of substances in a liquid
medium.
Computer based record and operation system

sujith.m,lecturer/eee,9600286256
Electrophoresis
Electrophoresis is the motion of dispersed particle
relative to a fluid under the influence of a spatially
uniform electric field

This electro-kinetic phenomenon was observed for


the first time in 1807 by Reuss

sujith.m,lecturer/eee,9600286256
There are several types of electrophoresis, but
the concepts are similar.
The machine has an anode (positive charge)
and a cathode (negative charge).
Negative ions move toward the anode, and
positive-charged ions move towards the
cathode.
The rate and distance traveled by these
molecules help scientists classify and study
different bimolecular.

sujith.m,lecturer/eee,9600286256
What is Gel Electrophoresis?

Electro = flow of electricity,

phoresis, from the Greek = to carry across

A gel is a colloid, a suspension of tiny particles in a

medium, occurring in a solid form, like gelatin


Gel electrophoresis refers to the separation of charged

particles located in a gel when an electric current is


applied
Charged particles can include DNA, amino acids, peptides,

etcsujith.m,lecturer/eee,9600286256
Principles of Gel Electrophoresis
Electrophoresis is a technique used
to separate and sometimes purify
macromolecules - especially
proteins and nucleic acids - that
differ in size, charge or
conformation
 When charged molecules are
placed in an electric field, they
migrate toward either the positive or
negative pole according to their
charge. In contrast to proteins,
which can have either a net positive
or net negative charge, nucleic acids
have a consistent negative charge
imparted by their phosphate
backbone, and migrate toward the
anode.
sujith.m,lecturer/eee,9600286256
Definition
Separation of DNA fragments according
to size, based on movement through a gel
medium when an electric field is applied.

sujith.m,lecturer/eee,9600286256
DNA negatively charged

sujith.m,lecturer/eee,9600286256
DNA cut up using restriction endonucleases

sujith.m,lecturer/eee,9600286256
sujith.m,lecturer/eee,9600286256
Make up the gel which the DNA will be put
into

sujith.m,lecturer/eee,9600286256
sujith.m,lecturer/eee,9600286256
sujith.m,lecturer/eee,9600286256
sujith.m,lecturer/eee,9600286256
Dye added to the DNA

sujith.m,lecturer/eee,9600286256
Buffer solution added to the tank

sujith.m,lecturer/eee,9600286256
DNA samples loaded into wells

sujith.m,lecturer/eee,9600286256
Electrical current applied to the chamber

sujith.m,lecturer/eee,9600286256
DNA is stained using ethidium bromide

sujith.m,lecturer/eee,9600286256
Colorimeter
The colorimeter is a filter photometer that measures
the color concentration of a substance in solution, this
is accomplished electronically by detecting the color
light intensity passing through a sample containing the
reaction products of the original substance and reagent

Eg: urine test

sujith.m,lecturer/eee,9600286256
sujith.m,lecturer/eee,9600286256
Basic colorimeter analysis
 Involves the precise measurement of light intensity.
Transmittance is defined as:
T= (I1/I0) x 100%
I2= T I1
I2=T2 I0

I0 is initial light intensity


I1 is first attenuated light intensity
I2 is second attenuated light intensity
T is the transmittance in percent

sujith.m,lecturer/eee,9600286256
Basic colorimeter analysis
Absorbance is defined as:
A= log(I0/I1)=log 1/T
Beer’s law
A= aCL
L is cuvette path length
C is concentration of absorbing substance
a is absorbtivity (~ substance)

sujith.m,lecturer/eee,9600286256
Basic colorimeter analysis
Concentration of unknown solution
Cμ= Cs Aμ/As

Cμ is unknown concentration
Cs standard concentration
Aμ is unknown absorbance
As is standard absorbance

sujith.m,lecturer/eee,9600286256
Calibration
1. Ground the amplifier input V1 and adjust
potentiometer R4 and adjust potentiometer R4 for
0V ± 5mV at the amplifier output.
2. Remove the ground and place reference
concentration in cuvettes 1 and 2
3. Adjust potentiometer (R1) for 0V ± 10mv at the
amplifier output.

sujith.m,lecturer/eee,9600286256
Calibration
4. Leave the reference concentration in cuvette 1 and
replace cuvette 2 with a cuvette containing the
sample.
5. Read the unbalanced voltage on the meter in
percentage of transmittance or absorbance units.

sujith.m,lecturer/eee,9600286256
Calibration
Example: if V1 = +1 mV with reference cuvettes
and V1 = +25 mV with reference and sample
cuvette, R2 = 2 kΩ, R3 = 1

Calculate the voltage read on the meter display for
both conditions of V1.
Solution: Av = 1+ R2/R3 = 1+ 2/1 = 3
Av = Vm/V1
Condition 1: Vm = Av * V1 = 3 * 1mV = 3mV
Condition 2: Vm = Av * V1 = 3 * 25 mV = 75 mV

sujith.m,lecturer/eee,9600286256
Maintenance
Calibration
Adjustment
Replacement of burned-out lamp and photo detector
Electronic problems are rare

sujith.m,lecturer/eee,9600286256
Medical Laboratory Instrumentation

Spectrophotometer

sujith.m,lecturer/eee,9600286256
Definition, Purpose, Use
Spectrophotometer measures light absorption by a
liquid substance at various wavelengths.
The Components of unknown material can be
determined, or the concentration of a number of
known substances can be measured.
Eg: blood cell resistivity

sujith.m,lecturer/eee,9600286256
Operation, Calibration
A monochromator uses a diffraction grating or prism to
disperse the light from the lamp (slit S1).
The light is broken into its spectral components as it
arises from slit S2 and falls on the sample in the cuvette.
Narrower slits give rise to shorter wavelenghts.
The angle of the diffraction grating determines light
wavelength if all other parameter are fixed.
The mirror reduces equipment size
Light output, Photodetector sensitivity, and sample
substance absorption change with wavelengths
measurement
sujith.m,lecturer/eee,9600286256
Schematic

sujith.m,lecturer/eee,9600286256
Maintenance
Calibration
Adjustment
Replacement of light source bulbs
Replacement of the Photodetector
Mechanically rotating assemblies (mirrors,…)
Electronics faulty

sujith.m,lecturer/eee,9600286256
Flame photometer
It is measure the color intensity of a flame
that is supported by oxygen and a specific
substances
The basic schematic shows that a reference
gas containing a lithium salt causes a red light
to shine on the reference photodetector
through the optical filter
A yellow or violet light from sample sodium
or potassium falls on the sample photodetector
It is similar to colorimeter
sujith.m,lecturer/eee,9600286256
sujith.m,lecturer/eee,9600286256
Continuous calibration can be accomplished by
inspiration of air( oxygen to support
combustion) and lithium.
The output read in units of sodium or
potassium concentration
Maintenance
Adjustment of bulbs & photo detectors
Aspiration device and flame chambers
occasionally cleaning
Electronics failure usually frequent

sujith.m,lecturer/eee,9600286256
Auto analyzer
Purpose of Autoanalyzer
The autoanalyzer is sequentially measures
blood chemistry and displays this on a graphic readout

This is accomplished by
– Mixing
– reagent
– Reaction
– Colorimetric measurement in continuous stream

sujith.m,lecturer/eee,9600286256
Elements of Autoanalyzer
Sampler
Proportioning pump and manifold
 Dialyzer
 Heating bath
 Colorimeter
 Record

sujith.m,lecturer/eee,9600286256
sujith.m,lecturer/eee,9600286256
Sampler
Aspirates samples, standards, and wash solutions to the
autoanalyzer system

Proportioning pump
Introduces samples with reagents to effect the proper chemical
color reaction to be read by the colorimeter.
 Pumps fluids at precise flow rates to other modules, as proper
color development depends on reaction time and temperature

Dialyzer
 Separates interfacing substances from the sample material by
permitting selective passage of sample components through a
semipermeable membrane
sujith.m,lecturer/eee,9600286256
Heating bath
Heats fluids continuously to exact temperature (37 degree).
Colorimeter
 Monitors the changes in optical density of the fluid stream
flowing through a tubular flow cell.
 Color intensities proportional to substance concentrations
 Colorimeter convert the color intensity to equivalent
electrical voltages
Recorder
Converts the electrical signal from the colorimeter into a
graphic display on moving chart

sujith.m,lecturer/eee,9600286256
Maintenance
 Calibration and adjustment
Mechanical
– Tubing
– Moving pump parts
 Electrical
– Switches
– Motors
Electronic failures are few Problems
 Identification of samples
 Sterilization for sample and glassware and
equipment parts

sujith.m,lecturer/eee,9600286256

You might also like