Professional Documents
Culture Documents
BIO-CHEMICAL INSTRUMENT
Medical Laboratory
Instrument
SUJITH.M, LECTURER/EEE, VIDYAA VIKAS ENGG COLLEGE Ph.No.
9600286256
Equipment
Spectrophotometer
Is optical device that measure light absorption at various
wavelengths for a given liquid sample.
Blood cell analyzer
Is a device to measures the number of red and white
blood cells per scaled volume.
The aperture impedance and flow cytometery
sujith.m,lecturer/eee,9600286256
Equipment
Glassware, centrifuges, suction devices
Colorimeter
Is an optical devise that measures the color concentration
of a substance in solution
Flame photometer
Is an optical electronic devise that measures the color
intensity of substance that have been aspirated into a
flame (sodium and potassium)
sujith.m,lecturer/eee,9600286256
Equipment
Ph/ blood gas analyzer
Is a device which measure blood Ph, Po2, Pco2
Chromatograph and Autoanalyzer
Is a electromechanical device used to separate, identify,
and measure the concentration of substances in a liquid
medium.
Computer based record and operation system
sujith.m,lecturer/eee,9600286256
Electrophoresis
Electrophoresis is the motion of dispersed particle
relative to a fluid under the influence of a spatially
uniform electric field
sujith.m,lecturer/eee,9600286256
There are several types of electrophoresis, but
the concepts are similar.
The machine has an anode (positive charge)
and a cathode (negative charge).
Negative ions move toward the anode, and
positive-charged ions move towards the
cathode.
The rate and distance traveled by these
molecules help scientists classify and study
different bimolecular.
sujith.m,lecturer/eee,9600286256
What is Gel Electrophoresis?
etcsujith.m,lecturer/eee,9600286256
Principles of Gel Electrophoresis
Electrophoresis is a technique used
to separate and sometimes purify
macromolecules - especially
proteins and nucleic acids - that
differ in size, charge or
conformation
When charged molecules are
placed in an electric field, they
migrate toward either the positive or
negative pole according to their
charge. In contrast to proteins,
which can have either a net positive
or net negative charge, nucleic acids
have a consistent negative charge
imparted by their phosphate
backbone, and migrate toward the
anode.
sujith.m,lecturer/eee,9600286256
Definition
Separation of DNA fragments according
to size, based on movement through a gel
medium when an electric field is applied.
sujith.m,lecturer/eee,9600286256
DNA negatively charged
sujith.m,lecturer/eee,9600286256
DNA cut up using restriction endonucleases
sujith.m,lecturer/eee,9600286256
sujith.m,lecturer/eee,9600286256
Make up the gel which the DNA will be put
into
sujith.m,lecturer/eee,9600286256
sujith.m,lecturer/eee,9600286256
sujith.m,lecturer/eee,9600286256
sujith.m,lecturer/eee,9600286256
Dye added to the DNA
sujith.m,lecturer/eee,9600286256
Buffer solution added to the tank
sujith.m,lecturer/eee,9600286256
DNA samples loaded into wells
sujith.m,lecturer/eee,9600286256
Electrical current applied to the chamber
sujith.m,lecturer/eee,9600286256
DNA is stained using ethidium bromide
sujith.m,lecturer/eee,9600286256
Colorimeter
The colorimeter is a filter photometer that measures
the color concentration of a substance in solution, this
is accomplished electronically by detecting the color
light intensity passing through a sample containing the
reaction products of the original substance and reagent
sujith.m,lecturer/eee,9600286256
sujith.m,lecturer/eee,9600286256
Basic colorimeter analysis
Involves the precise measurement of light intensity.
Transmittance is defined as:
T= (I1/I0) x 100%
I2= T I1
I2=T2 I0
sujith.m,lecturer/eee,9600286256
Basic colorimeter analysis
Absorbance is defined as:
A= log(I0/I1)=log 1/T
Beer’s law
A= aCL
L is cuvette path length
C is concentration of absorbing substance
a is absorbtivity (~ substance)
sujith.m,lecturer/eee,9600286256
Basic colorimeter analysis
Concentration of unknown solution
Cμ= Cs Aμ/As
Cμ is unknown concentration
Cs standard concentration
Aμ is unknown absorbance
As is standard absorbance
sujith.m,lecturer/eee,9600286256
Calibration
1. Ground the amplifier input V1 and adjust
potentiometer R4 and adjust potentiometer R4 for
0V ± 5mV at the amplifier output.
2. Remove the ground and place reference
concentration in cuvettes 1 and 2
3. Adjust potentiometer (R1) for 0V ± 10mv at the
amplifier output.
sujith.m,lecturer/eee,9600286256
Calibration
4. Leave the reference concentration in cuvette 1 and
replace cuvette 2 with a cuvette containing the
sample.
5. Read the unbalanced voltage on the meter in
percentage of transmittance or absorbance units.
sujith.m,lecturer/eee,9600286256
Calibration
Example: if V1 = +1 mV with reference cuvettes
and V1 = +25 mV with reference and sample
cuvette, R2 = 2 kΩ, R3 = 1
kΩ
Calculate the voltage read on the meter display for
both conditions of V1.
Solution: Av = 1+ R2/R3 = 1+ 2/1 = 3
Av = Vm/V1
Condition 1: Vm = Av * V1 = 3 * 1mV = 3mV
Condition 2: Vm = Av * V1 = 3 * 25 mV = 75 mV
sujith.m,lecturer/eee,9600286256
Maintenance
Calibration
Adjustment
Replacement of burned-out lamp and photo detector
Electronic problems are rare
sujith.m,lecturer/eee,9600286256
Medical Laboratory Instrumentation
Spectrophotometer
sujith.m,lecturer/eee,9600286256
Definition, Purpose, Use
Spectrophotometer measures light absorption by a
liquid substance at various wavelengths.
The Components of unknown material can be
determined, or the concentration of a number of
known substances can be measured.
Eg: blood cell resistivity
sujith.m,lecturer/eee,9600286256
Operation, Calibration
A monochromator uses a diffraction grating or prism to
disperse the light from the lamp (slit S1).
The light is broken into its spectral components as it
arises from slit S2 and falls on the sample in the cuvette.
Narrower slits give rise to shorter wavelenghts.
The angle of the diffraction grating determines light
wavelength if all other parameter are fixed.
The mirror reduces equipment size
Light output, Photodetector sensitivity, and sample
substance absorption change with wavelengths
measurement
sujith.m,lecturer/eee,9600286256
Schematic
sujith.m,lecturer/eee,9600286256
Maintenance
Calibration
Adjustment
Replacement of light source bulbs
Replacement of the Photodetector
Mechanically rotating assemblies (mirrors,…)
Electronics faulty
sujith.m,lecturer/eee,9600286256
Flame photometer
It is measure the color intensity of a flame
that is supported by oxygen and a specific
substances
The basic schematic shows that a reference
gas containing a lithium salt causes a red light
to shine on the reference photodetector
through the optical filter
A yellow or violet light from sample sodium
or potassium falls on the sample photodetector
It is similar to colorimeter
sujith.m,lecturer/eee,9600286256
sujith.m,lecturer/eee,9600286256
Continuous calibration can be accomplished by
inspiration of air( oxygen to support
combustion) and lithium.
The output read in units of sodium or
potassium concentration
Maintenance
Adjustment of bulbs & photo detectors
Aspiration device and flame chambers
occasionally cleaning
Electronics failure usually frequent
sujith.m,lecturer/eee,9600286256
Auto analyzer
Purpose of Autoanalyzer
The autoanalyzer is sequentially measures
blood chemistry and displays this on a graphic readout
This is accomplished by
– Mixing
– reagent
– Reaction
– Colorimetric measurement in continuous stream
sujith.m,lecturer/eee,9600286256
Elements of Autoanalyzer
Sampler
Proportioning pump and manifold
Dialyzer
Heating bath
Colorimeter
Record
sujith.m,lecturer/eee,9600286256
sujith.m,lecturer/eee,9600286256
Sampler
Aspirates samples, standards, and wash solutions to the
autoanalyzer system
Proportioning pump
Introduces samples with reagents to effect the proper chemical
color reaction to be read by the colorimeter.
Pumps fluids at precise flow rates to other modules, as proper
color development depends on reaction time and temperature
Dialyzer
Separates interfacing substances from the sample material by
permitting selective passage of sample components through a
semipermeable membrane
sujith.m,lecturer/eee,9600286256
Heating bath
Heats fluids continuously to exact temperature (37 degree).
Colorimeter
Monitors the changes in optical density of the fluid stream
flowing through a tubular flow cell.
Color intensities proportional to substance concentrations
Colorimeter convert the color intensity to equivalent
electrical voltages
Recorder
Converts the electrical signal from the colorimeter into a
graphic display on moving chart
sujith.m,lecturer/eee,9600286256
Maintenance
Calibration and adjustment
Mechanical
– Tubing
– Moving pump parts
Electrical
– Switches
– Motors
Electronic failures are few Problems
Identification of samples
Sterilization for sample and glassware and
equipment parts
sujith.m,lecturer/eee,9600286256